Skip to main content
AJNR: American Journal of Neuroradiology logoLink to AJNR: American Journal of Neuroradiology
. 1986 Jul-Aug;7(4):665-8.

Radiation exposure from conventional and digital subtraction angiography of cerebral vessels.

M B Plunkett, J E Gray, D B Kispert
PMCID: PMC8334653  PMID: 3088945

Abstract

Patients with atherosclerotic cerebrovascular disease were studied to determine the radiation exposure associated with conventional and digital subtraction angiography of the cerebral vessels. The median exposure-area product was 3198 R cm2 (range, 616-5665 R cm2) in the conventional angiography group and 1831 R cm2 (range, 366-4198 R cm2) in the IV digital subtraction angiography (DSA) group. This difference in exposure resulted from increased use of fluoroscopy in the conventional screen-film angiography group. The actual difference in exposure between the radiographic and digital imaging portions of the examinations was much smaller. The contributions of fluoroscopy to the radiation exposure in conventional angiography and IV-DSA in this study were 37% (range, 8.8-76%) and 6% (range, 1.5-25%), respectively.

Full Text

The Full Text of this article is available as a PDF (2.3 MB).


Articles from AJNR: American Journal of Neuroradiology are provided here courtesy of American Society of Neuroradiology