Technique for measuring hippocampal T2 relaxation time.

J S Duncan 1, P Bartlett 1, G J Barker 1
PMCID: PMC8337552  PMID: 8933861

Abstract

PURPOSE

To implement and validate methods for producing calculated coronal T2 maps with complete coverage of the hippocampus and the rest of the brain.

METHODS

T2 relaxation times were estimated on three occasions for 12 quality assessment test objects with the use of fast spin-echo (FSE) and conventional spin-echo (CSE) sequences. Hippocampal T2 relaxation times were measured in 15 healthy subjects with FSE and CSE sequences. Ten of these subjects were scanned twice with CSE imaging. Twenty-two patients with temporal lobe epilepsy, in whom hippocampal T2 relaxation times were previously estimated by means of a multiecho sequence, were studied with CSE.

RESULTS

CSE data were more reliable than FSE data. Reliability of hippocampal T2 measures on repeated acquisitions was superior for the CSE sequence with echo time 30,120 than for the 30,80 sequence (test-retest reliability, 4.0%; interrater reliability, 4.6%). CSE images were suitable for qualitative reporting, so additional T2-weighted sequences were not required. There was good correlation between these hippocampal T2 data and values obtained on a previously validated multiecho technique, with reliable identification of hippocampal sclerosis.

CONCLUSION

T2 mapping with CSE 30/120 sequences may be readily applied, produces reliable T2 values in contiguous 5-mm sections, and may be useful in the assessment of the hippocampus and other cerebral structures.