Skip to main content
Journal of Research of the National Institute of Standards and Technology logoLink to Journal of Research of the National Institute of Standards and Technology
. 1994 Jan-Feb;99(1):19–25. doi: 10.6028/jres.099.003

The NIST 30 MHz Linear Measurement System

Jeffrey A Jargon 1, Ronald A Ginley 1, Douglas D Sutton 1
PMCID: PMC8345255  PMID: 37404355

Abstract

An automated linear measurement system (LMS) has been developed to determine the nonlinearity of a tuned 30 MHz power detector over a 6.021 dB range. This detector uses a single thermistor bead design with thermal isolation to obtain nearly linear tracking over a 4:1 change in input power. The nonlinear correction for this change, determined by the LMS, is on the order of 1.00030 (+130 μB) for the detector presently in use. Initial experiments indicate an expanded uncertainty of ±0.138% (±598 μB), which is based upon Type A and Type B components.

Keywords: attenuation, automated, calibration, linear, measurement, power, thermistor, uncertainty

1. Introduction

There has been a recent interest in and demand for a calibration service at NIST to support rf attenuators and voltage doublers that operate specifically at 30 MHz, The first step required to offer such a service is to develop a reference standard. For the best possible accuracy, a tuned single-element thermistor mount was chosen. A linear measurement system was designed and constructed at NIST to calibrate the nonlinearity of this mount. This paper contains a description of the LMS, an explanation of the measurement scheme, calibration results, and an uncertainty analysis.

2. System Description

A diagram of the system is shown in Fig. 1. The signal generator provides a stable 30 MHz rf signal that is amplified and Filtered before the signal is split into two channels. One channel consists of a variable phase shifter and a fixed attenuator. A coaxial switch either terminates the signal with a 50 Ω termination or feeds the signal into a power divider, which splits the signal again. Half of the signal is detected by a single-element thermistor mount, P1, and the other half is fed into the hybrid. The second channel consists of a variable attenuator. Like the first channel, a coaxial switch either terminates the signal with a 50 Ω load or feeds the signal into a power divider, which splits the signal. Half of the signal is detected by a single-element thermistor mount, P2, and the other half is fed into the hybrid.

Fig. 1.

Fig. 1

Block diagram of 30 MHz linear measurement system.

The hybrid takes the sum and difference of the two input signals. The difference is fed into a diode detector to rectify the signal, and then into a null meter. The sum is fed into a coaxial switch, and when the switch is in position 1, the signal is detected by a third thermistor mount, P3, which is the thermistor to be calibrated.

Each thermistor mount is connected to a NIST Type IV bridge and a digital voltmeter to measure rf power.

The computer controls the signal generator, the digital voltmeters, and the switch controller, and handles the data acquisition and processing through an IEEE-488 bus.

3. Design of 30 MHz Single-Element Thermistor Mount

The dc response of detector P3 must be nearly linear with changes in input rf power. A single 50 Ω thermistor bead design was selected as a linear detection scheme [1]. This detector is used in conjunction with a NIST Type IV self-balancing dc-substitution rf power meter modified to bias a 50 Ω detector [2]. The Type IV power meter is designed to change the thermistor bead bias current so that the thermistor always maintains the same resistance. The detectors are placed in the LMS housing where the temperature is held to ±0.2 °C.

It is difficult to filter the rf signal from the power meter leads due to the nature of the single bead. The total rf signal should appear across the bead in the ideal case. Other considerations in the detector design include thermal stability and any forms of rf leakage into or out of the detector.

Several ideas are incorporated in the detector to eliminate these problems. Figure 2 shows the circuit diagram of a single-element thermistor mount. LC filter sections are inserted to filter the rf signal in the power meter leads. No ferrite material (such as a ferrite core inductor) is used near the thermistor, since ferrite components experience changes in impedance with changes in the rf power. These impedance changes lead to nonlinearities in the detector response. Therefore, air-cored inductors are inserted near the thermistor. Ferrite-cored conductors are allowable in sections following the first filter section because the rf power is sufficiently reduced and renders any impedance variation negligible. The parallel LC filter is tuned to resonate at 30 MHz. A special thermistor-mounting structure has been developed, and consists of an electrically insulated doughnut and two copper blocks, one on either side of the doughnut. The thermistor is placed inside the doughnut hole and is encapsulated in an air pocket by the addition of the copper blocks. The copper blocks also provide a large thermal mass so that the entire structure cannot experience rapid changes in temperature. The thermal time constant is much longer than the time required to perform a single measurement cycle. Double-sided copper-clad fiberglass boards are used in the exterior detector housing and in the internal compartment walls which separate various filter sections. This type of construction and the use of capacitive feedthroughs in the compartment walls reduce the amount of rf leakage though the detector. These physical and electrical construction considerations provide a stable linear detector with low rf leakage.

Fig. 2.

Fig. 2

Circuit diagram of single-clement thermistor mount. All inductors are hand-wound and all capacitors are 1 μF unless otherwise labeled.

4. Measurement Methods

4.1. Calibration of P3

Three power meters are used in the LMS–P1, P2, and P3. When a given power meter is read, the notation used is PXYT, where X denotes the power meter (1, 2, or 3), Y is 1 if channel 1 is switched on and 0 if channel 1 is switched off, Z is 1 if channel 2 is switched on and 0 if channel 2 is switched off, and “r” denotes that this power is a reading and not a true value. Powers without a subscript “r” are true values.

The calibration of P3 is achieved using a three-stage method. First, with switches 1, 2, and 3 each in position 1, enough power is applied so that a nominal 12 mW is incident on the three thermistor mounts. The phase shifter and variable attenuator are adjusted to balance the two channels, thus obtaining a null on the null meter. Readings are taken on P1, P2, and P3 and are designated P111r, P211r, and P311r, respectively. Next, switch 2 is moved to position 2, so that power is only applied to the first channel. Readings are taken on P1 and F3 and are designated P110r and P310r, respectively, A nominal 12 mW will be incident on P1 and approximately 3 mW will be incident on P3. Finally, switch 2 is moved back to position 1 and switch 1 is moved to position 2, so that power is only applied to the second channel. Readings are taken on P2 and P3 and are designated P201r and P310r, respectively. A nominal 12 mW will be incident on P2 and approximately 3 mW will be incident on P3.

The calibration constant of P3, denoted CHL2, is calculated using

CHL2=[(P310rP110rP311rP111r)1/2+(P301rP201rP311rP211r)1/2]2. (1)

The derivation of this formula can be found in Appendix A. The calibration constant is a measure of the detector’s nonlinearity over the 6.021 dB power change, and is used as a multiplication factor to correct the ratio measured by P3, where

P311P310=CHL2P311rP310r. (2)

4.2. Power Measurements

The NIST Type IV power meter must be connected to an external dc voltmeter. The substituted dc power, Pdc, is calculated from measured voltages using

Pdc=Voff2Von2R0, (3)

where Voff is the output voltage with no rf power applied, Von is the output voltage with rf applied, and R0 (50 Ω) is the resistance of the thermistor mount. Figure 3 shows the measurement sequence for a power calculation [3]. An initial Voff is taken; rf power is applied and Von is measured; rf power is removed and a final Voff is taken. The initial and final dc measurements are used with the Von measurement to calculate the power and correct for any mount drift, which is assumed to be linear. The calculated value of Voff in Eq. (3) is given by

Voff=Voff,i+t2t1t3t1(Voff,fVoff,i), (4)

where Voff,i is the voltage reading taken before rf is applied at time t1 Voff,f is the voltage taken after rf is removed at time t3, and t2 is the time at which Von is taken.

Fig. 3.

Fig. 3

Sequence for measuring power meter dc voltages.

5. Results

The calibration constant, CHL2, which is a measure of the nonlinearity of mount P3, was obtained by repeated measurements. The average value of the three hundred trials taken so far is CHL2=1.00030 or +130 μB. Long-term data are being accumulated to validate the calibration of P3.

Table 1 shows a sample calibration. Powers read by the three thermistor mounts are displayed for each of the three stages. The calculated calibration constant is shown, along with the actual step in power for each leg of the system.

Table 1.

Sample calibration results of the 30 MHz Linear Measurement System’s standard mount

Switches 1&2 Switch 1 Switch 2
P(1) 11.5194 11.5089 0.0000
P(2) 11.4170 0.0001 11.3991
P(3) 11.0907 2.7644 2.7756
Leg 1: 6.0335 dB
Leg 2: 6.0160 dB
CHL2:1.00029

6. Uncertainty Analysis

6.1. Evaluation of Type A Standard Uncertainty

Evaluation of a Type A standard uncertainty may be based on any valid statistical method for treating data. Examples are calculating the standard deviation of the mean of a series of independent observations, using the method of least squares to fit a curve to data in order to estimate the parameters of the curve and their standard deviations, and carrying out an analysis of variance in order to identify and quantify random effects in certain kinds of measurements [4].

The calibration of the standard mount, P3, has been repeated three hundred times to determine the repeatability of the system. Tests were performed at various times of the day over several days to cover as many random factors as possible, including variations of environmental conditions and the operator’s ability to renull the system. The sample standard deviation of the mean is ±0.000335% or ±2 μB. Long-term data are being accumulated to validate this calibration, and a control chart is being developed to track any possible outliers or drift.

6.2. Evaluation of Type B Standard Uncertainty

Evaluation of a Type B standard uncertainty is based upon scientific judgment using all of the available relevant information. This includes previous measurement data, manufacturers’ specifications, data provided in calibration reports, knowledge of the behavior of relevant instruments and materials, and uncertainties assigned to reference data taken from handbooks [4].

The Type B evaluation of standard uncertainty accounts for the following factors:

  1. Uncertainty in the dc voltmeter measurements.

  2. Uncertainty in the Type IV power meters.

  3. Imperfect isolation between the two legs.

  4. Uncertainty because P301≠P310

  5. Effects of impedance changes in P3.

  6. RF leakage.

  7. Spurious signals and harmonics.

6.2.1 Voltmeter Uncertainty

The uncertainty in the individual voltmeter readings may be determined by taking the total differential of the power expression, Eq. (3), which gives

dP=2R0(VoffdVoffVondVon). (5)

The total differential of power, Eq. (5), may be determined by taking the differential of Voff, Eq. (4), which gives

dVoff=(1T)dVoff,i+TdVoff,f, (6)

where

T=t2t1t3t1. (7)

The uncertainties, dVoff,i. dVoff,f, and dVon, in the measured values of Voff,i, Voff,r, and Von, are based on the voltmeter manufacturer’s specifications.

Figure 4 shows the uncertainty in the power measurement as a function of power level, assuming the powers are ratioed as they are in the CHL2 calculation, Eq. (1). The power measurements, P301 and P310, which are approximately 3 mW, result in uncertainties of 0.036%. The other power measurements, which are approximately 12 mW, result in uncertainties of 0.008%. The uncertainty of CHL2 due to the voltmeter readings may be found by inserting the individual power uncertainties into Eq. (1), which gives

CHL2+Δv=[(P310+Δ310P110Δ110P311+Δ311P111Δ111)1/2+(P301+Δ301P201Δ201P311+Δ311P211Δ211)1/2]2. (8)

This results in an uncertainly of Δr = ±0.028% or ± 122 μB.

Fig. 4.

Fig. 4

Power measurement uncertainty duc to DVM when ratios are taken.

6.2.2 Type IV Powcr-Mctcr Uncertainty

The four possible sources of uncertainties internal to the Type IV power meter are the reference resistors, the operational amplifier open-loop gain, input offset voltage, and input bias current. Larsen has shown that the uncertainties due to the Type IV power meters are negligible compared to those of the voltmeters [2].

6.2.3 Imperfect Isolation

An uncertainty in the calculation of CHL2 may result if P1 and P2 are not zero when they are assumed to be. The thermistor mounts are square-law detectors and are not sufficiently sensitive to determine whether P1 and P2 are low enough in the “off” condition to avoid significant errors. This uncertainty is derived in Appendix B.

The corrected formula for CHL2, taking imperfect isolation into account, is

CHL2=[(P310rP110rP311rP111r)1/21|1+β1|+(P301rP201rP311rP211r)1/21|1+γ|]2, (9)

where

β1=Q12Δb2b110 (10)

and

γ1=Δb1b201Q12. (11)

Here, γ1 is the measure of P1’s contribution to CHL2 if it is not zero when it is assumed to be, and β1 is the measure of P2’s contribution to CHL2 if it is not zero when it is assumed to be. The terms, b110 and b201 are the corresponding voltages to the powers P110 and P201, respectively. Assuming P110 and P201 are 12 mW, b110 and b201 are equal to 0.7746 V. The term, Q12, is defined in Appendix B and its value is approximately −1. The Δb’s represent the b’s which were assumed to be zero in Appendix A. If the isolation between the channels is 65 dB, as is stated in the power divider manufacturer’s specifications, then Δb1 = Δb2 = 0.000436. Using nominal values for the P’s and substituting the values into Eq. (9), an uncertainty of ± 0.112% or ± 488 μB is obtained.

6.2.4 Uncertainty Because P301≠P310

An uncertainty in the calculation of CHL2 may result if P301≠P310. The derivation of this uncertainty can be found in Appendix C.

The measure of nonlinearity, α, of thermistor mount three is

α=1CHL2CHL2PLrPHr, (12)

where PLr is the reading of mount three at the low level and PHr, is the power reading of mount three at the high level. Assuming CHL2=1.0004 worst case, PLr =3 mW, and PHr = 12 mW, α is calculated to be 3.334 × 10−5. The ratio K301/K310 is given by

K301K310=1+αP301r1+αP310r. (13)

If P301r is 10% greater than P310r, then

K301K310=1+α(3×1.10)1+α(3)=1.000010. (14)

The effect of this being nonunity results in a corrected formula for CHL2, where

CHL2=[(P310rP110rP311rP111r)1/2+(K301K310P310rP201rP311rP211r)1/2]2=1.000005 (15)

using nominal values for the P’s. The result is an uncertainty of ±0.0005% or ±3 μB.

6.2.5 Uncertainty Due to Impedance Changes in P3

An uncertainty in the calculation of CHL2. may result if the impedance changes in P3 with power level. The derivation of this uncertainty can be found in Appendix D.

The corrected formula for CHL2, taking impedance changes in P3 into account, is

CHL2=[(P310rP110rP311rP111r)1/2+(P310rP201rP311rP211r)1/2]2|1ΓLMSΓ3111ΓLMSΓ310|2(1|Γ310|2)(1|Γ311|2), (16)

where Γ311 is the reflection coefficient of mount three at the high level, Γ310 is the reflection coefficient of mount three at the low level, and ΓLMS is the reflection coefficient of the system looking into port three.

The measured value of Γ310 was actually taken 10 dB below Γ310 instead of the 6 dB step, so the uncertainty should be conservative. Using nominal values for the P’s and the measured reflection coefficients, the calculated uncertainty is ±0.014% or ± 62 μB.

Impedance changes in the diode detector with respect to power level are assumed to have a negligible effect on the overall uncertainty.

6.2.6 Uncertainty Due to RF Leakage

It is difficult to assign a quantitative uncertainty due to rf leakage. In order to reduce leakage, all coaxial cables in the system were replaced with semirigid lines and, wherever possible, SMA connectors were used. For now, the uncertainty due to rf leakage is assumed to be ±10 μB or ±0.0023%.

6.2.7 Uncertainty Due to Spurious Signals and Harmonics

After the signal is amplified and filtered, any harmonics are at least −92 dBc, while spurious signals are no greater than − 84 dBc. This will result in a measurement error that will affect the uncertainty of CHL2.

From Eq, (3), the calculated dc power is

Pdc=20(Voff2Von2), (17)

where Pdc is in mW. Assuming Voff = 1 V, then Von = 0.63246 V for a calculated power of 12 mW and Von = 0.92195 V for a calculated power of 3 mW.

A harmonic at −92 dBc results in an uncertainty of ±0.0025% for a voltage measurement. This uncertainty translates to power uncertainties of (12 ±0.0004) mW and (3 ±0.0008) mW. Using the nominal powers to calculate CHL2, an uncertainty of ±0.023% or ±100 μB is obtained.

A spurious signal at −112 dBc results in an uncertainty of ±0.00025% for a voltage measurement. This uncertainty translates to power uncertainties of (12 ±0.00004) mW and (3 + 0.000085) mW. Using the nominal powers to calculate CHL2, an uncertainty of ±0.0025% or ±11 μB is obtained.

6.2.8. Overall Type B Uncertainty

For each Type B component, an estimated range, ±aj, is given, assuming that the quantity in question has a 100% probability of lying within that interval. The quantity is treated as if it is equally probable for its value to lie anywhere within the interval. Therefore, it is modeled by a rectangular probability distribution. The best estimate of the standard deviation, uj is

uj=aj3. (18)

Table 2 shows all of the Type B components along with their corresponding uncertainties and standard deviations. The overall standard deviation of the Type B components, calculated using the root-sum-of-squares method (RSS), is ±0.069% or ±299 μB.

Table 2.

Type B components of the 30 MHz Linear Measurement System with corresponding uncertainty ranges and standard deviations

Component Range (%) Standard deviation (%)
dc voltage measurements ±0.028     0.016
Imperfect isolation ±0.112     0.065
P301P310 ±0.0005     0.0003
Impedance changes in P3 ±0.014     0.008
rf leakage ±0.0023     0.0013
Spurious signals ±0.0025     0.0014
Harmonics ±0.023     0.013
Combined Type B Standard Uncertainty (RSS)   ± 0.069
(±299 μB)

6.3 Combined Standard Uncertainty

The combined standard uncertainty, ur, is taken to represent the estimated standard deviation of the result. It is obtained by combining the individual standard deviations, ui, whether arising from a Type A or a Type B evaluation [4]. The technique used to combine the standard deviations is the RSS method.

The total uncertainty reported is the expanded uncertainty, U, which is obtained by multiplying uc by a coverage factor, k. To be consistent with current international practice, the value of k used at NIST for calculating U is k = 2. The total expanded uncertainty, U, of the mount’s nonlinearity is calculated to be ±0.138% or ±598 μB.

7. Conclusion

The first step toward offering a calibration service at NIST to support rf attenuators and voltage doublers that operate at 30 MHz has been completed. The reference standard, a tuned single-element thermistor mount, has been calibrated using an LMS, designed and constructed at NIST. The next step is to modify the LMS so that a device under test may be inserted into the system and calibrated against the reference standard.

8. Appendix A. Scattering Coefficient Analysis of the LMS

The 30 MHz Linear Measurement System can be considered a five port network, as shown in Fig. 5. The analysis assumes that the LMS is an ideal system—that is, the system is linear, and when b1 and b2 are assumed to be 0, they truly are.

Fig. 5.

Fig. 5

LMS port nomenclature.

The pertinent scattering coefficients are

b1=S11a1+S12a2+S13a3+S14a4+S15a5b2=S21a1+S22a2+S23a3+S24a4+S25a5b3=S31a1+S32a2+S33a3+S34a4+S35a5b4=S41a1+S42a2+S43a3+S44a4+S45a5b5=S51a1+S52a2+S53a3+S54a4+S55a5. (19)

If a1 = Γ1b1 and a2 = Γ2b2, then Eq. (19) may be written as

b1(1Γ1S11)=S12Γ2b2+S13a3+S14a4+S15a5b2(1Γ2S22)=S21Γ1b1+S23a3+S24a4+S25a5b3=S31Γ1b1+S32Γ2b2+S33a3+S34a4+S35a5b4=S41Γ1b1+S42Γ2b2+S43a3+S44a4+S45a5b5=S51Γ1b1+S52Γ2b2+S53a3+S54a4+S55a5 (20)

This gives five equations and eight unknown variables–a3, a4, a5, b1, b2, b3, b4, and b5, Four variables can be eliminated from Eq. (20), say a4, b4, a5, and b5. Eq. (20) becomes

b1=P13a3+Q12b2+Q13b3, (21)

where the primed quantities are functions of the scattering coefficients and Γ1i and Γ2.

It is assumed that Γ3 is constant with respect to varying power levels, If now a3 = Γ3b3,

b1=(P13Γ3+Q13)b3+Q12b2Q13b3+Q12b2 (22)

or

b3=b1Q12b2Q13, (23)

where two complex constants, Q12 and Q13, describe the relationship among b1, b2, and b3.


Articles from Journal of Research of the National Institute of Standards and Technology are provided here courtesy of National Institute of Standards and Technology

RESOURCES