Relationship of Catheter Contact Angle and Contact Force with Contact Area on the Surface of Heart Muscle Tissue in Cardiac Catheter Ablation

Kriengsak Masnok 1, Nobuo Watanabe 1,2,
PMCID: PMC8354874  PMID: 33723738

Abstract

Purpose

The aims of this study were to develop an experimental procedure for setting the catheter angle with respect to the surface of the heart muscle and the catheter contact force and to investigate the catheter contact area on the heart muscle as a function of catheter contact angle and force.

Methods

Visualization tests were performed for 5 contact angles (0°, 30°, 45°, 60°, and 90°) and 8 contact forces (2, 4, 6, 10, 15, 20, 30, and 40 gf). Each experiment was repeated 6 times with 2 different commercially available catheter tips.

Results

The morphology of the contact area was classified into rectangular, circular, ellipsoidal, and semi-ellipsoidal. The correlation between contact force and contact area was a logarithmic function; increasing contact force was associated with increased contact area. At the same contact force, the correlation between contact angle and contact area was inverse; decreasing contact angle was associated with a corresponding increase in contact area.

Conclusion

Both the catheter contact angle and contact force substantially impact the contact area and morphology in catheter ablation procedures.

Keywords: Catheter ablation, Contact force, Contact angle, Contact area, Tachyarrhythmia

Introduction

Over the past three decades, cardiac catheter ablation therapy has become a widely used and effective treatment for tachyarrhythmia.7,9,10,14,24,31 In this treatment, radiofrequency current is applied to the heart, heating the target area to a temperature exceeding 50 °C through resistive heating,19 thereby creating a lesion that isolates the abnormal electric pathway.8,17 Earlier studies have revealed several factors that correlate with lesion size, evaluated in terms of ablated area, volume, and depth.2,18,28,30 These factors include ablation circuit impedance,8,30 electrical power,16,21,25 energy delivery,11,13 catheter diameter,15,26 exposure time,8,16 contact force,1,5,16,21,22,30 ablation electrode temperature,12,29 irrigation saline flow amount,5,25,28 and blood flow near the myocardial surface.6,11,20,23 Among these factors, catheter contact force is reported to show a strongly positive correlation with lesion size.1,35,22,27

In addition to these factors, we hypothesized that the catheter contact angle with respect to the surface of the heart muscle would also have a substantial effect. During radiofrequency current catheter ablation, the catheter tip should contact the heart tissue surface at a variety of angles. However, no studies to date have investigated the relationship between the catheter tip and contact area with the heart muscle.

Against this backdrop, the purpose of this study was to develop an experimental procedure for setting the catheter angle with respect to the surface of the heart muscle and the catheter contact force, as well as to investigate the catheter contact area on the heart muscle as a function of catheter contact angle and contact force.

Methods

Heart Muscle Surface Flattener and Different Open-Loop Irrigated Catheter Tips

Most of the surface of the heart is round, and the state of catheter contact would vary according to clinical conditions. Therefore, to provide better reproducibility of our in vitro experiments, we developed a special instrument that precisely adjusts the catheter angle between the catheter tip and the heart muscle.

The instrument consists of a heart muscle surface flattener and catheter tip angle setter. As part of the heart muscle surface flattener, a circular crystalline acrylic plate with a thickness of 12 mm and a diameter of 130 mm was used to flatten the surface of porcine heart tissue and fix its position at a specific location and orientation, ensuring that all experiments using this plate will maintain uniformity.

The porcine heart was obtained from a slaughterhouse, cut into pieces, and stored in a refrigerator. Before the experiment, the pieces were removed from the refrigerator and kept at room temperature under moist conditions to prevent drying. The porcine heart tissue was sandwiched between the acrylic plate and a soft sponge placed in a stainless bowl. The surface of the heart muscle surface was flattened by adjusting the amount of the sponge. The catheter contact experiments were performed through a hole (20 mm × 50 mm) in the acrylic plate, as shown in Fig. 1a. All experiments were performed within 3 days after the pig was sacrificed at the slaughterhouse.

Figure 1.

Figure 1

Heart muscle surface flattener and shape of the catheter tip; (a) photograph of the heart muscle surface flattener. The heart muscle tissue is sandwiched between a flat acrylic plate and a soft sponge, which are placed in a stainless bowl, and the surface of the heart muscle was flattened by adjusting the amount of the sponge in the bottom of the bowl. (b) Ablation catheter tips with two shapes were used: flat (top) and round (bottom).

Two different open-loop irrigated catheter tips were used in this study (Fig. 1b). One was the IntellaNav Mifi™ OI (7 Fr/4.5 mm 7.5 Fr; PMR9620, Boston Scientific, Inc.; top of Fig. 1b), which is representative of flat-tip catheters. It was 110 cm long, with a tip length of 2.33 mm, and had a standard curve style. The other catheter was an Abbott TactiCath™ (7 Fr/3.5 mm 7.5 Fr; Quadripolar, PN-004075, St. Jude Medical, Inc.; bottom of Fig. 1b), which is representative of round-tip catheters. It was 115 cm long, with a tip length of 2.33 mm, and had a steerable curve style. Both catheters were open-loop irrigated catheters, with six small irrigation holes circumferentially located on the lateral surface of the tip. Irrigation of the catheter tip was designed to reduce excessive heating of the tissue and blood at the catheter tip. The main difference between the two catheters is the shape of the end tip.

Experimental Procedure and Evaluation

To elucidate the effects of the catheter contact angle and contact force on the contact area of the heart tissue surface, we developed a special experimental procedure that enables the setting of various catheter contact angles (0°, 30°, 45°, 60°, and 90°) using a special acrylic tube guide, as well as the measurement of the contact force. In the experimental setup, a digital force sensor (FGP-0.5, Nidec-Shimpo Corporation) was mounted on a motion stage (FGS-5000TV, Nidec-Shimpo Corporation), the position of which can be controlled vertically. Using this setup, the catheter contact force and contact angle could be precisely controlled. The system was operated using commercial software (FGT-TV) running on a computer, as shown in Fig. 2.

Figure 2.

Figure 2

Experimental setup. A compact desktop test stand equipped with a digital force gauge was controlled using an FGT-TV software link from a computer. In this picture, the acrylic tube guide for 90° was used.

White soluble ink (Pen Cure, Japan Pen Company) was overlaid on the metal electrode of the catheter tip to visualize the contacted area on the heart tissue surface. Then, the 8 levels of contact force within the clinically used range (2, 4, 6, 10, 15, 20, 30, and 40 gf) were applied to the cardiac muscle in line with the typical clinical contact force ranges.3,4,27 Using this process, the catheter contact area-visualization test was repeated 6 times each for the 5 contact angles and 8 contact forces to ensure equal distribution of contact force. In the final step, all of the catheter contact areas for each condition were photographed for the evaluation of contact area through image analysis. In total, 480 experiments were performed.

Morphological Evaluation of Catheter Contact Area

Image analysis of photographs of the catheter contact area was performed to evaluate the morphology of the contact area. The image analysis program MATLAB (Version 2019a) was used to perform the following actions. First, the raw color image of the contact area was manually segmented into individual lesion images and converted into a grayscale image, and finally the grayscale image was binarized. Next, the catheter contact area on the heart tissue surface was calculated. To understand the morphology of the contact area, the centroid of each contact area image was aligned to create a reference point for comparison. Then, the image was rotated about the centroid to align each area's longest axis parallel to the vertical direction. The average morphology of the contact area was derived from six experimentally acquired images. The morphological characteristics corresponding to physical parameters were also evaluated. All statistical analyses were performed using GraphPad Prism software (Version 8.4.3).

Results

Contact Area Morphology

Figure 3 shows the four distinct morphologies of the contact area with a contact angle of 0° using the flat-tip catheter and contact angles of 0°, 30°, and 90° using the round-tip catheter. These images show that the contact angle and shape of the catheter tip can affect the contact area morphology. For example, the morphology differs according to the shape of the catheter tip, even when both are applied at a contact force of 2 gf and a contact angle at 0°. In contrast, the morphology was similar when both shapes were applied at a contact force of 2 gf and a contact angle at 90. Further details about the differences in contact area morphology will be discussed later in the “Discussion” section.

Figure 3.

Figure 3

Morphology of the contact area on a porcine heart under various contact conditions. T1–T4 represent the morphology of the four types of contact area.

Average Contact Area

Figure 4 shows the image analysis flow process for evaluating the morphology of the contact area and the average contact area on a porcine heart with a contact angle of 90° and a contact force of 30 gf using the round-tip catheter.

Figure 4.

Figure 4

Chart of the flow process for evaluating the morphology of the contact area and the average contact area at a contact angle of 90° and a contact force of 30 gf using a round-tip catheter.

Data of average contact areas, standard deviations, and percentage of contact area for flat-tip and round-tip catheters are shown in Tables 1 and 2, respectively. The ratio of the area of the catheter in contact with heart muscle to the catheter tip surface area was calculated by the following equation:

Percentagecontactarea=Catheter/heartcontactareaCathetertipsurfacearea×100,

where the catheter tip surface area of the flat-tip catheter is 37.26 mm2 and that of the round-tip catheter is 25.67 mm2.

Table 1.

Average contact area and percentage contact area when using the flat-tip catheter (mm2).

Contact angle Contact force (gf)
2 4 6 10
AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%)
11.424 1.123 31 10.861 1.188 29 10.547 1.079 28 14.041 1.245 38
30° 6.002 0.908 16 7.406 0.594 20 7.941 1.625 21 9.607 1.284 26
45° 7.058 0.928 19 7.027 0.608 19 10.199 0.873 27 10.463 0.883 28
60° 4.641 0.605 12 7.247 1.228 19 8.163 1.791 22 10.094 1.996 27
90° 3.445 0.831 9 4.392 0.514 12 5.627 0.555 15 6.879 0.874 18
Contact angle Contact force (gf)
15 20 30 40
AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%)
15.458 1.392 41 15.358 1.624 41 16.405 1.050 44 19.097 1.294 51
30° 13.178 2.103 35 14.644 2.609 39 11.769 1.914 32 14.759 1.196 40
45° 11.820 0.680 32 12.842 1.438 34 14.860 1.672 40 19.533 1.361 52
60° 11.131 1.590 30 11.431 1.699 31 12.114 1.443 33 13.515 0.895 36
90° 8.363 0.982 22 7.405 1.126 20 9.508 1.040 26 12.589 0.812 34

AVG CA average contact area, SD standard deviation, PCA percentage contact area

Table 2.

Average contact area and percentage contact area when using the round-tip catheter (mm2).

Contact angle Contact force (gf)
2 4 6 10
AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%)
6.365 0.508 25 8.730 0.840 34 9.039 1.040 35 10.325 1.159 40
30° 6.055 1.075 24 6.999 0.825 27 9.982 0.768 39 12.579 1.107 49
45° 3.592 0.493 14 4.906 0.421 19 5.547 0.629 22 6.815 0.898 27
60° 3.699 0.724 14 5.869 0.719 23 6.782 0.821 26 8.222 0.625 32
90° 3.829 0.294 15 6.309 0.320 25 7.797 1.134 30 7.990 0.553 31
Contact angle Contact force (gf)
15 20 30 40
AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%) AVG CA SD PCA (%)
11.779 1.292 46 12.343 1.042 48 13.661 0.852 53 15.578 1.274 61
30° 14.417 2.141 56 13.697 1.857 53 17.693 1.394 65 16.788 0.581 65
45° 8.123 0.652 32 9.704 0.841 38 12.730 1.583 50 13.725 1.954 53
60° 9.354 0.811 36 10.275 0.868 40 10.464 1.740 41 14.310 2.286 56
90° 8.535 0.751 33 9.340 0.638 36 10.357 1.039 40 11.914 0.909 46

AVG CA average contact area, SD standard deviation, PCA percentage contact area

The standard SI unit for force is the Newton (N), but gram-force (gf) is frequently used to measure contact force in the field of catheter ablation research (1 gf = 0.00981 N, which is the force acting on a mass of 1 g under the Earth’s gravitational acceleration of 9.81 m/s2). The contact forces tested in this study were 2, 4, 6, 10, 15, 20, 30, and 40 gf, which correspond to 0.0196, 0.0392, 0.0588, 0.098, 0.147, 0.196, 0.294, and 0.392 N, respectively.

Figure 5 shows a plot of the catheter contact areas created with the contact forces on the x-axis and the contact area on the y-axis for both catheter shapes. This plot illustrates the correlation between contact force, contact angle, and contact area. The results revealed a positive correlation between contact force and contact area, in which increased contact force was associated with increased contact area. Moreover, the contact angle had as strong an effect on the contact area as the contact force did. At the same contact force, the correlation between contact angle and contact area was inverse; that is, a smaller contact angle was associated with an increased contact area. The logarithmic approximation formulas for expressing the relationship between contact force and contact area for each catheter contact angle are shown in Table 3. The data reveal that the correlation between contact force and the contact area is a logarithmic function with R-squared (R2) being nearly equal to 1.

Figure 5.

Figure 5

Plot of average contact area and contact force for (a) flat-tip catheters and (b) round-tip catheters at each contact angle.

Table 3.

Approximation formulas expressing the relationship between catheter contact force and contact area for each catheter contact angle, where x is catheter contact force (gf), y is catheter contact area (mm2), and R2 is coefficient of determination, respectively.

Angle (°) Flat-tip Round-tip
Approximation formula R2 Approximation formula R2
0 y = 2.685ln(x) + 7.782 0.837 y = 2.837ln(x) + 4.251 0.974
30 y = 3.036ln(x) + 3.465 0.845 y = 3.903ln(x) + 2.890 0.961
45 y = 3.689ln(x) + 2.952 0.867 y = 3.429ln(x) + 0.012 0.926
60 y = 2.807ln(x) + 3.137 0.984 y = 3.062ln(x) + 1.363 0.936
90 y = 2.693ln(x) + 0.892 0.893 y = 2.341ln(x) + 2.709 0.953

Figure 6 shows binarized images of the average contact areas under the various conditions (8 contact forces and 5 contact angles) when using the flat-tip and round-tip catheters. These data clearly show that the contact angle had as much influence as the contact force on the contact area. For example, using the flat-tip catheter at a contact angle of 90°, the contact force increased from 2 to 4, 6, 10, 15, 20, 30, and 40 gf, and the average contact area increased from 3.445 to 4.392, 5.627, 6.879, 8.363, 7.405, 9.508, and 12.589 mm2, respectively. In addition, using the round-tip catheter at a contact angle of 90°, the contact force increased from 2 to 4, 6, 10, 15, 20, 30, and 40 gf, and the average contact area increased from 3.829 to 6.309, 7.797, 7.990, 8.535, 9.340, 10.357, and 11.914 mm2, respectively. Similar trends were seen for both shapes of catheters at contact angles of 60°, 45°, 30°, and 0°.

Figure 6.

Figure 6

Average contact area morphology using flat-tip and round-tip catheters under various conditions.

Discussion

Major Findings

Our major findings are as follows: (i) the morphology of the contact area can be divided into four types: rectangular, circular, ellipsoidal, and semi-ellipsoidal. The morphology of the contact area indicates that (ii) the correlation between contact force and contact area is a logarithmic function; that is, increased contact force was associated with increased contact area, and the contact angle has as strong an effect on the contact area as contact force does. (iii) There is an inverse correlation between contact angle and contact area; smaller contact angle is associated with increased contact area.

Morphological Characterization of the Contact Area

To elucidate the effects of catheter contact angle and contact force on the contact area, we constructed a heart muscle surface flattener to maintain a flat surface to test a range of contact angles. This instrument was designed to achieve improved experimental reproducibility. Although in routine clinical ablation procedures, the surface of the heart tissue is not flat, our data clearly demonstrated that the contact angle and shape of the catheter tip substantially affected the contact area morphology. In summary, we categorized the morphology of the contact area into four types, as shown in Fig. 7. A notable difference occurred when the catheter angle became parallel to the heart surface. Contact area morphology became rectangular when using a flat end tip and semi-ellipsoidal when using a round end tip. This observation clearly shows the effect of the shape of the catheter. When contact is made at a perpendicular angle, the contact area morphology is circular because the projected area of both catheters is a circle, and thus the contact area becomes circular. When the catheter is inclined, the contact area becomes ellipsoidal like an egg. Those morphological character trends changed similarly for both round- and flat-tip catheters except in the parallel (0°) direction.

Figure 7.

Figure 7

Contact area morphologies. The T1 morphology represents the contact area of the flat-tip catheter at a contact angle of 0°. The T2 morphology represents the contact area of the round-tip catheter at a contact angle of 0°. The T3 morphology represents the contact area of both the round- and flat-tip catheters at a contact angle of 90°. The T4 morphology represents the contact area of both the round- and flat-tip catheters at a contact angle of between 30° and 60°.

Correlation Between Contact Force and Contact Area

Catheter contact force shows a strong positive correlation with contact area. When the contact force was increased, the contact area also increased. These results are similar to those in earlier reports1,35,22,27; however, it is essential to consider the small changes in contact area that occurred at higher contact forces. The contact area increased monotonically but logarithmically. The slope of the graph changes slightly when the contact force is between 15 and 40 gf, which is in contrast to the greater change in slope when during initial contact when the contact force ranges from 2 to 15 gf. The equation for estimating contact area might help those performing this procedure to understand the relationships among the parameters and to calculate the contact area as a function of contact force at each contact angle. Our data suggest a limit to the extent by which lesion size can be increased by increasing the contact force. The catheter contact angle relative to the heart muscle tissue surface can also needs to be considered when calculating the desired lesion size.

Correlation Between Contact Angle and Contact Area

The results clearly demonstrate that the contact angle is a key determinant of the contact area morphology. In addition, the contact angle substantially affects the contact area of the catheter tip regardless of the contact force. For the flat-tip catheter, the minimum contact area was produced at a contact angle of 90° and increased with decreasing contact angle from 90° to 60°, 45°, 30°, and 0°. For the round-tip catheter, the minimum contact area was produced at a contact angle of 90° and increased with decreasing contact angle from 90° to 60°, 45°, 0°, and 30°. For both catheter shapes, the contact area progressively increased when the contact angle was decreased. However, our results show a difference between the flat- and round-tip catheters at 0° and 30°. These differences were due to the difference in shape and size between the two shapes of catheter tip. The two catheters used for this study were made by different manufacturers and differ in size according to their shape, especially at a contact angle between 0° and 30°. The round-tip catheter makes less surface contact with the heart tissue surface compared with the flat-tip catheter. Despite this fact, the results of the experiment as a whole show a similar tendency.

Clinical Implications

Our data should be useful for those performing this procedure to understand the relation among the parameters and plan their treatment strategy beforehand. From our experiments, the contact area morphology was derived as a function between the contact angle and contact force. It is reasonable to assume that the contact area is directly related to the area of resulting lesion.

Study Limitations

This study has several limitations. First, we aimed to test our assumptions about contact angle, contact force, and contact area by using two different shapes of commercially available catheter tips. The two catheters were made by different manufacturers, and thus have some differences in design. Accordingly, we did not compare the differences in results between the catheters. Second, to provide better reproducibility of our in vitro experiments, we developed a special instrument that precisely adjusts the catheter angle between the catheter tip and the heart muscle. The instrument consists of a heart muscle surface flattener and a catheter-tip-angle setter. In clinical practice the shape of the heart tissue surface varies according to the part of the heart, and thus the catheter tip orientation can rarely be optimized due to restricting structures such as trabeculated muscle, valves, or the papillary muscle. Nevertheless, at the present stage of research on catheter ablation (pre-clinical experiment studies), it necessary to perform tests on flat surfaces to clearly demonstrate the specific effects of the catheter contact angle and contact force on the contact area of the heart tissue surface. Lastly, to produce effective ablation lesions, the depth of the lesion is at least as important as the ablation size. In this study, we did not investigate whether the catheter contact angle and contact force affected the depth of the ablation lesion; however, we conducted experiments to elucidate the effect of the catheter contact angle and contact force on the contact area. Our findings might be validated in the near future through numerical simulations such as the Finite Elemental Method, which can be used to estimate cardio-muscular deformation in response to catheter tip contact or a practical investigation through an in vitro heart muscle ablation experiment.

Conclusion

This study clearly demonstrated a substantial impact of the contact angle and contact force of a catheter on the size and morphology of the contact area in catheter ablation procedures. The contact area should be directly related to the lesion area. Our data may help doctors understand the relationships among contact angle, contact force, and contact area in ablation therapy procedures. Such information should help doctors plan appropriate treatment strategies in consideration of each patient’s conditions.

Acknowledgments

The authors thank Abbott Medical Japan and Boston Scientific Japan for their kind provision of catheters.

Author Contributions

Both authors contributed to the study conception and design. Material preparation, data collection, and analysis were performed by KM and NW. The first draft of the manuscript was written by KM and both authors commented on subsequent drafts of the manuscript. All authors read and approved the final manuscript.

Funding

This research was supported by the Research and Development Supporting Fund from the Shibaura Institute of Technology, and a Scholarship from the Ministry of Education, Culture, Sports, Science and Technology of Japan.

Data Availability

The datasets generated during and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Code Availability

MATLAB (Version 2019a) and GraphPad Prism software (Version 8.4.3) were used in this study. Code generated and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Conflict of interest

Both authors received consulting/advisory support from Abbott Medical Japan and Boston Scientific Japan. The authors declare that they have no conflicts of interest.

Ethical Approval

Not applicable.

Informed Consent

No human studies were carried out by the authors for this article.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Ariyarathna N, Kumar S, Thomas SP, Stevenson WG, Michaud GF. Role of contact force sensing in catheter ablation of cardiac arrhythmias: evolution or history repeating itself? JACC Clin. Electrophysiol. 2018;4(6):707–723. doi: 10.1016/j.jacep.2018.03.014. [DOI] [PubMed] [Google Scholar]
  • 2.Avitall B, Mughal K, Hare J, Helms R, Krum D. The effects of electrode-tissue contact on radiofrequency lesion generation. Pacing Clin. Electrophysiol. 1997;20(12):2899–2910. doi: 10.1111/j.1540-8159.1997.tb05458.x. [DOI] [PubMed] [Google Scholar]
  • 3.Borregaard R, Jensen HK, Tofig BJ, Thrysøe SA, Gerdes C, Nielsen JC, et al. Is the knowledge of contact force beneficial in pulmonary vein antrum isolation? Scand. Cardiovasc. J. 2017;51(3):129–137. doi: 10.1080/14017431.2017.1285043. [DOI] [PubMed] [Google Scholar]
  • 4.Conti S, Weerasooriya R, Novak P, Champagne J, Lim HE, Macle L, et al. Contact force sensing for ablation of persistent atrial fibrillation: a randomized, multicenter trial. Heart Rhythm. 2018;15(2):201–208. doi: 10.1016/j.hrthm.2017.10.010. [DOI] [PubMed] [Google Scholar]
  • 5.Franco E, Rodríguez Muñoz D, Matía R, Hernández-Madrid A, Sánchez Pérez I, Zamorano JL, et al. Contact force-sensing catheters: performance in an ex vivo porcine heart model. J. Interv. Card. Electrophysiol. 2018;53(2):141–150. doi: 10.1007/s10840-018-0435-y. [DOI] [PubMed] [Google Scholar]
  • 6.Fuller IA, Wood MA. Intramural coronary vasculature prevents transmural radiofrequency lesion formation: implications for linear ablation. Circulation. 2003;107(13):1797–1803. doi: 10.1161/01.CIR.0000058705.97823.F4. [DOI] [PubMed] [Google Scholar]
  • 7.Grubb CS, Lewis M, Whang W, Biviano A, Hickey K, Rosenbaum M, et al. Catheter ablation for atrial tachycardia in adults with congenital heart disease: electrophysiological predictors of acute procedural success and post-procedure atrial tachycardia recurrence. JACC Clin. Electrophysiol. 2019;5(4):438–447. doi: 10.1016/j.jacep.2018.10.011. [DOI] [PubMed] [Google Scholar]
  • 8.Haines D. Biophysics of ablation: application to technology. J. Cardiovasc. Electrophysiol. 2004;15(s10):S2–S11. doi: 10.1046/j.1540-8167.2004.15102.x. [DOI] [PubMed] [Google Scholar]
  • 9.Hosseini SM, Rozen G, Saleh A, Vaid J, Biton Y, Moazzami K, et al. Catheter ablation for cardiac arrhythmias: utilization and in-hospital complications, 2000 to 2013. JACC Clin. Electrophysiol. 2017;3(11):1240–1248. doi: 10.1016/j.jacep.2017.05.005. [DOI] [PubMed] [Google Scholar]
  • 10.Ichijo S, Miyazaki S, Kusa S, Nakamura H, Hachiya H, Kajiyama T, et al. Impact of catheter ablation of atrial fibrillation on long-term clinical outcomes in patients with heart failure. J. Cardiol. 2018;72(3):240–246. doi: 10.1016/j.jjcc.2018.02.012. [DOI] [PubMed] [Google Scholar]
  • 11.Iso K, Okumura Y, Watanabe I, Nagashima K, Sonoda K, Kogawa R, et al. Wall thickness of the pulmonary vein-left atrial junction rather than electrical information as the major determinant of dormant conduction after contact force-guided pulmonary vein isolation. J. Interv. Card. Electrophysiol. 2016;46(3):325–333. doi: 10.1007/s10840-016-0147-0. [DOI] [PubMed] [Google Scholar]
  • 12.Iwasawa J, Koruth JS, Petru J, Dujka L, Kralovec S, Mzourkova K, et al. Temperature-controlled radiofrequency ablation for pulmonary vein isolation in patients with atrial fibrillation. J. Am. Coll. Cardiol. 2017;70(5):542–553. doi: 10.1016/j.jacc.2017.06.008. [DOI] [PubMed] [Google Scholar]
  • 13.Kalman JM, Fitzpatrick AP, Olgin JE, Chin MC, Lee RJ, Scheinman MM, et al. Biophysical characteristics of radiofrequency lesion formation in vivo: dynamics of catheter tip-tissue contact evaluated by intracardiac echocardiography. Am. Heart J. 1997;133(1):8–18. doi: 10.1016/s0002-8703(97)70242-4. [DOI] [PubMed] [Google Scholar]
  • 14.Murakoshi N, Aonuma K. Catheter ablation for ventricular tachyarrhythmia in patients with channelopathies. J. Arrhythm. 2016;32(5):404–410. doi: 10.1016/j.joa.2016.01.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Müssigbrodt A, Grothoff M, Dinov B, Kosiuk J, Richter S, Sommer P, et al. Irrigated tip catheters for radiofrequency ablation in ventricular tachycardia. BioMed. Res. Int. 2015;2015:389294. doi: 10.1155/2015/389294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Nakagawa H, Jackman WM. The role of contact force in atrial fibrillation ablation. J. Atr. Fibrillation. 2014;7(1):1027. doi: 10.4022/jafib.1027. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Nath S, Lynch C, Whayne JG, Haines DE. Cellular electrophysiological effects of hyperthermia on isolated guinea pig papillary muscle. Implications for catheter ablation. Circulation. 1993;88(4):1826–1831. doi: 10.1161/01.CIR.88.4.1826. [DOI] [PubMed] [Google Scholar]
  • 18.Okumura Y, Johnson SB, Bunch TJ, Henz BD, O’Brien CJ, Packer DL. A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. J. Cardiovasc. Electrophysiol. 2008;19(6):632–640. doi: 10.1111/j.1540-8167.2008.01135.x. [DOI] [PubMed] [Google Scholar]
  • 19.Page RL, Joglar JA, Caldwell MA, Calkins H, Conti JB, Deal BJ, et al. 2015 ACC/AHA/HRS Guideline for the Management of Adult Patients with Supraventricular Tachycardia: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2016;133(14):e471–e505. doi: 10.1161/CIR.0000000000000310. [DOI] [PubMed] [Google Scholar]
  • 20.Sacher F, Wright M, Derval N, Denis A, Ramoul K, Pascale P, et al. Endocardial versus epicardial ventricular radiofrequency ablation: utility of in vivo contact force assessment. Circ. Arrhythm. Electrophysiol. 2013;6(1):144–151. doi: 10.1161/CIRCEP.11. [DOI] [PubMed] [Google Scholar]
  • 21.Shah DC, Lambert H, Nakagawa H, Langenkamp A, Aeby N, Leo G. Area under the real-time contact force curve (force–time integral) predicts radiofrequency lesion size in an in vitro contractile model. J. Cardiovasc. Electrophysiol. 2010;21(9):1038–1043. doi: 10.1111/j.1540-8167.2010.01750.x. [DOI] [PubMed] [Google Scholar]
  • 22.Shah DC, Namdar M. Real-time contact force measurement: a key parameter for controlling lesion creation with radiofrequency energy. Circ. Arrhythm. Electrophysiol. 2015;8(3):713–721. doi: 10.1161/CIRCEP.115.002779. [DOI] [PubMed] [Google Scholar]
  • 23.Simmers TA, De Bakker JMT, Coronel R, Wittkampf FHM, Van Capelle FJ, Janse MJ, et al. Effects of intracavitary blood flow and electrode-target distance on radiofrequency power required for transient conduction block in a Langendorff-perfused canine model. J. Am. Coll. Cardiol. 1998;31(1):231–235. doi: 10.1016/S0735-1097(97)00435-X. [DOI] [PubMed] [Google Scholar]
  • 24.Sohns C, Nürnberg JH, Hebe J, Duckeck W, Ventura R, Konietschke F, et al. Catheter ablation for atrial fibrillation in adults with congenital heart disease: lessons learned from more than 10 years following a sequential ablation approach. JACC Clin. Electrophysiol.. 2018;5(4):733–743. doi: 10.1016/j.jacep.2018.01.015. [DOI] [PubMed] [Google Scholar]
  • 25.Takami M, Lehmann HI, Parker KD, Welker KM, Johnson SB, Packer DL. Effect of left atrial ablation process and strategy on microemboli formation during irrigated radiofrequency catheter ablation in an in vivo model. Circ. Arrhythm. Electrophysiol. 2016;9(1):e003226. doi: 10.1161/CIRCEP.115.003226. [DOI] [PubMed] [Google Scholar]
  • 26.Tsai CF, Tai CT, Yu WC, Chen YJ, Hsieh MH, Chiang CE, et al. Is 8-mm more effective than 4-mm tip electrode catheter for ablation of typical atrial flutter? Circulation. 1999;100(7):768–771. doi: 10.1161/01.CIR.100.7.768. [DOI] [PubMed] [Google Scholar]
  • 27.Ullah W, McLean A, Tayebjee MH, Gupta D, Ginks MR, Haywood GA, et al. Randomized trial comparing pulmonary vein isolation using the SmartTouch catheter with or without real-time contact force data. Heart Rhythm. 2016;13(9):1761–1767. doi: 10.1016/j.hrthm.2016.05.011. [DOI] [PubMed] [Google Scholar]
  • 28.Weiss C, Antz M, Eick O, Eshagzaiy K, Meinertz T, Willems S. Radiofrequency catheter ablation using cooled electrodes: impact of irrigation flow rate and catheter contact pressure on lesion dimensions. Pacing Clin. Electrophysiol. 2002;25(4):463–469. doi: 10.1046/j.1460-9592.2002.00463.x. [DOI] [PubMed] [Google Scholar]
  • 29.Wittkampf FH, Nakagawa H. RF catheter ablation: lessons on lesions. Pacing Clin. Electrophysiol. 2006;29(11):1285–1297. doi: 10.1111/j.1540-8159.2006.00533.x. [DOI] [PubMed] [Google Scholar]
  • 30.Yokoyama K, Nakagawa H, Shah DC, Lambert H, Leo G, Aeby N, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ. Arrhythm. Electrophysiol. 2008;1(5):354–362. doi: 10.1161/CIRCEP.108.803650. [DOI] [PubMed] [Google Scholar]
  • 31.Yonas E, Pranata R, Siswanto BB, Abdulgani HB. Comparison between surgical and catheter based ablation in atrial fibrillation, should surgical based ablation be implemented as first line? A meta-analysis of studies. Indian Pacing Electrophysiol. J. 2020;20(1):14–20. doi: 10.1016/j.ipej.2019.12.001. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets generated during and/or analyzed during the current study are available from the corresponding author upon reasonable request.


Articles from Cardiovascular Engineering and Technology are provided here courtesy of Springer