. 2021 Jun 14;12(7):4046–4055. doi: 10.1364/BOE.427907

Composite polarization systems for independent controlling polarization of two beams with different wavelengths

Evelina A Bibikova 1,2, Nataliya D Kundikova 1,2,*, Yurii V Mukhin 1,2, Vladimir V Chirkov 3
PMCID: PMC8367228  PMID: 34457398

Abstract

The usage of independent and simultaneous control of the state of light polarization at different wavelengths can expand the capabilities of polarization methods for biomedical application. Unfortunately, all known methods of polarization conversion cannot convert the state of light polarization at different wavelengths independently. We propose a method and device for independent and simultaneous control of the polarization state at two wavelengths. We have theoretically proved the possibility of maintaining the phase shift at the first wavelength unchanged while simultaneously and independently changing the phase shift at the second wavelength from 0 to 180 degrees. The capabilities of the method were for the first time demonstrated for radiation with wavelengths λ = 632.8 nm and λ = 488 nm. At the wavelength λ = 632.8 nm, the phase shift remained equal to 180° whereas at the wavelength λ = 488 nm, it varied in the range from 121° to 136°.

1. Introduction

Polarized light methods, various types of polarimetry, and polarized light imaging in biomedical applications have sharply increased in recent years. They include developed laboratory techniques for biochemical and biomedical studies [1,2], histological analysis of tissue samples [3,4], ex vivo survey, and detection of diseases [59]. Polarization methods can be used for for determination of the optical parameters of collagen solutions [10], quantitative detection of biomarkers [11], for analysis of prostate adenoma and carcinoma diffuse tissues [12]. Mueller matrix polarimetry has been used for studies of spatial heterogeneities in hydrogel remodeling by smooth muscle cells [13]. Biosensors based on imaging ellipsometry have been used for antibody screening, hepatitis B markers detection, cancer markers spectrum determination, virus recognition, to image and quantify pathogenic viruses and disease biomarkers [14,15].

Independent and simultaneous control of the state of polarization at two wavelengths can expand the functionality of polarization methods for biomedical application, reduce the time for conducting research, and lead to the miniaturization of polarization devices operating at several wavelengths.

Unfortunately, all known methods of polarization conversion cannot convert the state of light polarization at different wavelengths independently. Optical systems that linearly transform polarization of light may be described in the formalism of 2×2 complex Jones matrices [16,17]. According to the generalized polarization matrix equivalence theorem [18], any polarization system comprising any number of polarization elements can be presented by only four optical elements. All those elements generally depend on the wavelength of light λ. The most prominent factor is the dependence of phase retardation Γ on the wavelength, namely, Γ=Δnd/λ, here Δn is the material birefringence, and d is the thickness of the retardation plate. That is why if an optical system transforms the polarization of light at one wavelength in a particular way, the change of polarization for another wavelength would be set and would depend on the values of the system parameters for that second wavelength. Thus, polarization states for two different wavelengths would become entangled.

The use of broadband (achromatic) polarization devices has been proposed to overcome this principal limitation. It turns out that combinations of half-wave and quarter-wave plates - if their main axes are oriented at particular angles - can equally transform the states of polarization in a wide range of wavelengths [19,20]. Increasing the number of quarter-wave and half-wave plates allows one to create broadband achromatic polarization systems with tunable phase retardation [2124]. It is also possible to achieve broadband achromatic polarization devices using modern polymer and liquid crystal cells [25,26]. Combining phase plates with almost arbitrary phase retardations allows one to create polarization systems tunable in a wide wavelength range [2731]. The system of LC spatial light modulators and a deformable mirror have been proposed to eliminate parasitic polarization inhomogeneities in the beam cross-section [32].

Thus, while much attention has been devoted to the development of various broadband polarization devices, the problem of independent and simultaneous conversion of the polarization of two-wave radiation has not been raised or properly studied. Yet, few separate issues on the subject have been experimentally addressed [3335].

We propose a new method and device based on the use of several birefringent plates for simultaneous and independent polarization conversion of two beams with different wavelengths. The main principle of the method is based on the idea that several free parameters determine and govern the properties of composite polarization systems. It turns out that a set of four phase plates can make up a polarization system that simultaneously and independently transforms states of polarization for two wavelengths.

2. Theory of polarization systems

A polarization device based on a system of four phase plates is considered in our study (Fig. 1). Let the effective phase retardation of the system at the first wavelength λ1 to be Γ1,2,3,4(1)=π. Let us find out if this is possible with the predetermined Γ1,2,3,4(2) at the second wavelength λ2. We will use the upper index (1) to denote all values related to the first wavelength and the upper index (2) to denote all values related to the second wavelength.

Fig. 1.

Fig. 1.

Polarisation system consisted of four phase plates with the same retardation.

We chose the coordinate system so that the direction of light propagation coincides with the z-axis. The x-axis and the y-axis coincide with the direction of the slow axis and the fast axis of the first plate, respectively. The values φ2, φ3, and φ4 denote the angles between the x-axis and the slow axes of the second, third, and fourth plates, respectively. The Jones matrices formalism [17] is used to analyze the polarization characteristics of the system. We will define the range of changes of system parameters, under which the condition Γ1,2,3,4(1)=π is satisfied, and investigate the dependency of the value Γ1,2,3,4(2) on the system parameters, when they are changed in this range.

We should use the modified Jones theorem [30] of the polarization systems equivalence to describe the properties of a four-component system. This theorem states that any polarization system consisting of two phase-retardation plates can be presented as a system of one effective phase-retardation plate, and a rotator of the plane of polarization (an element with optical activity). First, we consider the system [28] that consists of two phase-retardation plates whose slow axes are oriented at an angle φ. Such a two-component polarization system has the following Jones matrix:

W(Γ1,Γ2,φ)=R(φ)T(Γ2)R(φ)T(Γ1), (1)

where

T(Γi)=(100exp(iΓi))

is the polarization matrix of the phase retardation plate in the coordinate system associated with its eigen axes; Γ1 and Γ2 are phase retardations of the first and second plates, respectively. The rotation matrix R(φ) are used for transit from one coordinate system to another:

R(φ)=(cosφsinφsinφcosφ). (2)

The theorem of the equivalence [30] allows the matrix W(Γ1,Γ2,φ) to be equated to the matrix of the system consisting of one effective phase-retardation plate, and a rotator of the plane of polarization V(ε,θ,Γ1,2):

W(Γ1,Γ2,φ)=V(ε,θ,Γ1,2). (3)

Here V(θ,ε,Γ1,2)=R(θ)R(ε)T(Γ1,2)R(ε) where R(±ε), R(θ) are the rotation matrices defines like Eq. (2) , θ is an angle of the optical activity, Γ1,2 is the effective phase retardation and ε is the angle between the eigen axis of the effective phase-retardation plate and the x-axis of the coordinate system. By solving Eq. (3), we can find the values of these variables

Γ1,2=acos(cosΓ1cosΓ2cos2φsinΓ1sinΓ2),ε=12atan[sin2φcotΓ2sinΓ1+cos2φcosΓ1],θ=atan[sin2φcot(Γ1/2)cot(Γ2/2)cos2φ]. (4)

Further, we will consider the polarization properties of the four-component system for the first wavelength (Fig. 1). Let us assume for simplicity that the values of the phase retardations of all four plates are equal to each other and are equal to Γ(1) for the first wavelength and are equal to Γ(2) for the second wavelength. Thus, we can write the Jones matrix of the system as:

W(1)(Γ(1),φ2,φ3,φ4)==R(φ4)T(Γ(1))R(φ4)R(φ3)T(Γ(1))R(φ3)R(φ2)T(Γ(1))R(φ2)T(Γ(1)), (5)

where R(φi) is the rotation matrix (i=2,3,4), and T(Γ(1)) is the matrix of the single phase-retardation plate.

Consistently applying the above equivalence theorem to the Eq. (5), it is possible to bring the polarization matrix of the system to a more convenient form for analysis. Taking into account the property of the rotation matrices R(ψ1)R(ψ2)=R(ψ1+ψ2), we can rewrite Eq. (5) as

W(1)(Γ(1),φ2,φ3,φ4)==R(φ4)T(Γ(1))R(φ4φ3)T(Γ(1))R(φ3)R(θ1,2(1)ε1,2(1))T(Γ1,2(1))R(ε1,2(1))==R(φ4)T(Γ(1))R(φ4+θ1,2(1)ε1,2(1))R(γ1,2(1))T(Γ(1))R(γ1,2(1))T(Γ1,2(1))R(ε1,2(1))==R(φ4)T(Γ(1))R(φ4+θ1,2(1)ε1,2(1))R(θ1,2,3(1)ε1,2,3(1))T(Γ1,2,3(1))R(ε1,2,3(1))R(ε1,2(1))==R(φ4)T(Γ(1))R(β(1))T(Γ1,2,3(1))R(ε1,2,3(1)+ε1,2(1)). (6)

Here, the ε with indices denote the angles between the slow axes of the effective retardation plates and the x-axis. The variables θ with indices indicate the azimuths of the plane of polarization rotation by the effective rotator. Γ(1) with indices denote the effective phase retardations of the corresponding polarization systems. The subscripts 1,2 and 1,2,3 indicate the parameters related to the polarization systems consisting of the first and second or the first, second and third plates. The variable γ(1)=φ3+θ1,2(1)ε1,2(1) denotes the angle between the effective slow axis of the composite polarization system 1,2 and the slow axis of the third retardation plate, thus taking into account the presence of the effective optical activity θ1,2 for the system 1,2; while β(1)=φ4+θ1,2(1)+θ1,2,3(1)ε1,2(1)ε1,2,3(1) has the same definition only for the system 1,2,3 and the fourth plate.

The variables in Eq. (6) are related to each other in the same way as in Eq. (4):

Γ1,2(1)=acos(cos2Γ(1)cos2φ2sin2Γ(1)), (7)
ε1,2(1)=12atan[sin2φ2(1+cos2φ2)cosΓ(1)], (8)
θ1,2(1)=atan[sin2φ2cot2(Γ(1)/2)cos2φ2], (9)
Γ1,2,3(1)=acos(cosΓ(1)cosΓ1,2(1)cos2γ(1)sinΓ(1)sinΓ1,2(1)), (10)
ε1,2,3(1)=12atan(sin2γ(1)cotΓ(1)sinΓ1,2(1)+cos2γ(1)cosΓ1,2(1)), (11)
θ1,2,3(1)=atan[sin2γ(1)cot(Γ1,2(1)/2)cot(Γ(1)/2)cos2γ(1)]. (12)

Further, it is necessary to find the values of the system parameters when the system has phase retardation Γ1,2,3,4(1)=π at the first wavelength. For this purpose, we note that according to the last line of Eq. (6), the polarization system under investigation is equivalent to the consecutively placed effective phase-retardation plate with Γ1,2,3(1) and the fourth phase-retardation plate with Γ(1). Such polarization system can have the effective phase retardation π only if the angle β(1) between the slow axes of these plates is zero, and the sum of their phase retardations is equal to π. It means that the condition Γ1,2,3,4(1)=π is satisfied for those values of the system parameters for which the following equalities are valid:

φ4=ε1,2(1)+ε1,2,3(1)θ1,2(1)θ1,2,3(1), (13)
Γ1,2,3(1)+Γ(1)=π. (14)

From Eq. (10) and Eq. (14) we obtain the following equation:

cosΓ(1)=cosΓ(1)cosΓ1,2(1)cos2γ(1)sinΓ(1)sinΓ1,2(1). (15)

From Eq. (15), it is possible to obtain an expression for the value of the angle γ(1):

γ(1)=12acos(cotΓ(1)1+cosΓ1,2(1)sinΓ1,2(1)). (16)

Since according to Eqs. (7) the values Γ1,2(1) depend only on the angle φ2 for the fixed Γ(1) then γ(1) is a function of φ2. The angle φ3 can also be expressed in terms of γ(1), ε1,2(1), θ1,2(1) and it is also a function of φ2:

φ3=γ(1)+ε1,2(1)θ1,2(1). (17)

The angle φ4 is expressed in Eq. (13), and it is also a function of φ2. Thus, if the four-component system has phase shift Γ1,2,3,4(1)=π, then all the other system parameters (φ3 and φ4) are uniquely determined for the given parameter φ2. It means that there is only one independent free parameter (the angle φ2) for the system.

Now we investigate how the effective phase retardation of the four-component system at the second wavelength changes when the parameters of the system are varied. Since the polarization properties of the system under investigation are uniquely determined by the angles φ2, φ3, φ4 and phase retardation of its components for the chosen wavelength, so the effective phase retardation Γ1,2,3,4(2) of the system for the second wavelength is a function of the angle φ2 and quantities Γ(1) and Γ(2), when the system is adjusted according to the condition Γ1,2,3,4(1)=π.

Equation (5) remains valid for the second wavelength when replacing Γ(1) with Γ(2). Then the values of the angles φ3 and φ4 are determined from Eq. (13), Eq. (17) as a function of φ2 and Γ(1). Then, the modified equivalence theorem Eq. (3) is applied to the resulting polarization matrix, taking into account the indicated substitutions. Having done these calculations, we get the following equation from Eq. (6):

W(2)(φ2,Γ(1),Γ(2))==R(θ1,2(2)+θ1,2,3(2)ε1,2(2)ε1,2,3(2))R(β(2))T(Γ(2))R(β(2))T(Γ1,2,3(2))R(ε1,2(2)+ε1,2,3(2))==R(θ1,2(2)+θ1,2,3(2)+θ1,2,3,4(2)ε1,2(2)ε1,2,3(2)ε1,2,3,4(2))T(Γ1,2,3,4(2))R(ε1,2(2)+ε1,2,3(2)+ε1,2,3,4(2)). (18)

The expressions for variables with the superscript (2) in Eq. (18) are essentially the same as Eqs. (7) –(12) for the first wavelength only substituting superscripts (1) for (2). Here

Γ1,2(2)=acos(cos2Γ(2)cos2φ2sin2Γ(2)),ε1,2(2)=12atan[sin2φ2(1+cos2φ2)cosΓ(2)],θ1,2(2)=atan[sin2φ2cot2(Γ(2)/2)cos2φ2],γ(2)=φ3ε1,2(2)+θ1,2(2),Γ1,2,3(2)=acos(cosΓ(2)cosΓ1,2(2)cos2γ(2)sinΓ(2)sinΓ1,2(2)),ε1,2,3(2)=12atan(sin2γ(2)cotΓ(2)sinΓ1,2(2)+cos2γ(2)cosΓ1,2(2)),θ1,2,3(2)=atan[sin2γ(2)cot(Γ1,2(2)/2)cot(Γ(2)/2)cos2γ(2)].β(2)=φ4ε1,2(2)ε1,2,3(2)+θ1,2(2)+θ1,2,3(2),Γ1,2,3,4(2)=acos(cosΓ1,2,3(2)cosΓ(2)cos2β(2)sinΓ1,2,3(2)sinΓ(2)),ε1,2,3,4(2)=12atan(sin2β(2)cotΓ(2)sinΓ1,2,3(2)+cos2β(2)cosΓ1,2,3(2)),θ1,2,3,4(2)=atan[sin2β(2)cot(Γ1,2,3(2)/2)cot(Γ(2)/2)cos2β(2)]. (19)

Thus, we have obtained all the relationships necessary to determine the conditions under which an adjustable polarization system consisting of four phase-retardation plates has the effective phase retardation π for the first wavelength and any requiring phase retardation Γ1,2,3,4(2) for the second wavelength by varying independent free parameter (the angle φ2).

Figure 2 shows the dependence of the effective phase retardation Γ1,2,3,4(2) on the adjusting angle φ2 for values Γ(1)=55, Γ(2)=125, provided that the system has Γ1,2,3,4(1)=π. It can be seen from Fig. 2 that when the adjusting angle φ2 is changed, the effective phase retardation Γ1,2,3,4(2) continuously varies in the range from almost 0 to 180.

Fig. 2.

Fig. 2.

The dependence of the effective phase retardation Γ1,2,3,4(2) on the adjusting angle φ2 at the values Γ(1)=55, Γ(2)=125.

3. Experimental results

The described approach was used to study experimentally a system consisting of four phase retardation plates with equal phase retardation. The phase retardation of the birefringent plates was chosen to be Γ(1)=52.8 at the first wavelength λ1=632.8 nm and Γ(2)=72.9 at the second wavelength λ2=488 nm. The tuning of the system to have a certain effective phase retardation at the second wavelength was performed by changing the angle φ2 (the free parameter). The phase retardation of the system at the second wavelength was measured and calculated, assuming the system has phase shift π at the first wavelength.

The experimental setup is shown in Fig. 3. A helium-neon laser (the wavelength λ1=632.8 nm) and an argon ion laser (λ1=488 nm) were used as light sources. Both lasers were generated in the fundamental transverse mode. The radiation of the helium-neon laser had a sufficiently high degree of linear polarization. The intensity-related ellipticity of its polarization eI was 104. The argon laser radiation was almost completely depolarized. The polarizer P1 (Glan prism) was used to produce linearly polarized light at the wavelength λ2=488 nm. A quarter wave plate converted the linearly polarized radiation at both wavelengths into circularly polarized. By turning polarizer P2 (Glan prism), it was possible to set the required azimuth of the plane of polarization of linearly polarized light at the input of the four-element polarization system under investigation. An analyzer A was installed after the polarization system. The polarization system consisted of four identical mica waveplates. The retardations of the waveplates at both wavelengths (Γ(1)=52.8 and Γ(2)=72.9) were measured using our method described in [36]. Using the values of the parameters Γ(1) and Γ(2), we calculated the values of the angles φ2, φ3 and φ4 and installed the retardation plates.

Fig. 3.

Fig. 3.

Experimental setup.

At first, we proved that the polarization system has the property of a half-wave plate at the first wavelength λ1=632.8 nm for the different sets of the angles φ2, φ3 and φ4. We determined the azimuth of the effective eigen axes ε1,2,3,4(1) and the angle of the effective optical activity of such system θ1,2,3,4(1). The quality and azimuth of the linearly polarized light at the output of the system were checked for the different orientations of the input linear polarization. We had high-quality linear polarization at the system output; the intensity-related ellipticity of polarization eI did not exceed 104. The azimuth of the linearly polarized light changed in the same way as after an ordinary half-wave plate if we take into account additional optical activity θ1,2,3,4(1).

The dependence of the azimuth of one of the eigen axes ε1,2,3,4(1) on the adjusting angle φ2 is shown in Fig. 4. Fig. 4 shows that the value ε1,2,3,4(1) depends on the adjusting angle φ2 and experimentally measured values of ε1,2,3,4(1) are in good agreement with the calculated values.

Fig. 4.

Fig. 4.

The dependence of the azimuth ε1,2,3,4(1) of one of the polarization system axes on the adjusting angle φ2 at the wavelength λ1.

Three polarization parameters of the polarization system, namely, the direction of one of the effective eigen axes ε1,2,3,4(2), effective optical activity θ1,2,3,4(2) and effective phase retardation Γ1,2,3,4(2) at wavelength λ2=488 nm were measured for the different sets of angles φ2, φ3 and φ4. The effective phase retardation Γ1,2,3,4(2) of the system under investigation was determined by measuring the maximum ellipticity, which the system provides. To determine the ellipticity eI we rotated the analyzer and measured the maximum and minimum light intensity passed through the polarization system. The value of eI was estimated as the ratio of the minimum and maximum measured intensity. To exclude the influence of the lasers output power fluctuations, the output intensity was determined as the ratio of signals from photodiodes D1 and D2.

Figure 5 shows the dependence of the polarization system effective phase retardation Γ1,2,3,4(2) on the adjusting angle φ2. The obtained experimental results are in good agreement with theoretical data. The observed deviation of experimental results from the calculated ones could be due to neglecting the linear amplitude anisotropy in the theoretical calculations.

Fig. 5.

Fig. 5.

The dependence of the effective phase retardation Γ1,2,3,4(2) of the polarization system on the adjusting angle φ2 at wavelength λ2.

Thus, the experimental results demonstrate the validity of a new method for independent and simultaneous polarization control of two light beams with different wavelengths. It should be stressed that achieving the effective phase retardation at the second wavelength up to Γ1,2,3,4(2)=180 is possible using LC-cell as a retardation plate as was theoretically shown in our paper [37].

4. Conclusions

We have proposed a composite polarization system for independent and simultaneous polarization control of two light beams with different wavelengths. It was shown that the polarization system should consist of four phase plates with the same birefringence but with different axes orientations relative to each other. We were successful in demonstrating the independent and simultaneous polarization control at wavelength λ1=632.8 nm and λ2=488 nm keeping the retardation at the first wavelength Γ1,2,3,4(1)=180 unchanged while the retardation at the second wavelength changed in the range from Γ1,2,3,4(2)=121 to Γ1,2,3,4(2)=136. To extend this range, it is necessary to use the elements with variable phase shift (LC-cell) as the components of the system.

In many biomedical applications of optical scanning and tomography, the use of multiple wavelengths is required to increase sensitivity and accuracy of measurements. On the other hand, the great advantage of proposed approach here is clear possibility for miniaturization of desired polarization systems. For example, the use of dual-wavelength optical polarimetry is considered extensively for real-time, non-invasive in-vivo glucose sensing [3840]. Implementation of our approach for designing polarization systems may help to develop these table-top experimental prototypes into real palm-sized devices for clinical use.

Acknowledgments

This work was supported in part by RFBR and Chelyabinsk Region, project number 20-42-740010.

Funding

in part RFBR and Chelyabinsk Region (20-42-740010).

Disclosures

The authors declare no conflicts of interest.

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

References

  • 1.Karsten Hinrichs, Ellipsometry of Functional Organic Surfaces and Films, vol. 52 (Springer, 2018). [Google Scholar]
  • 2.Ogieglo W., Wormeester H., Eichhorn K. J., Wessling M., Benes N. E., “In situ ellipsometry studies on swelling of thin polymer films: A review,” Prog. Polym. Sci. 42, 42–78 (2015). 10.1016/j.progpolymsci.2014.09.004 [DOI] [Google Scholar]
  • 3.Li P., Lee H. R., Chandel S., Lotz C., Groeber-Becker F. K., Dembski S., Ossikovski R., Ma H., Novikova T., “Analysis of tissue microstructure with Mueller microscopy: logarithmic decomposition and Monte Carlo modeling,” J. Biomed. Opt. 25(07), 1 (2020). 10.1117/1.JBO.25.7.070501 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Schucht P., Lee H. R., Mezouar H. M., Hewer E., Raabe A., Murek M., Zubak I., Goldberg J., Kövari E., Pierangelo A., Novikova T., “Visualization of white matter fiber tracts of brain tissue sections with wide-field imaging Mueller polarimetry,” IEEE Trans. Med. Imaging 39(12), 4376–4382 (2020). 10.1109/TMI.2020.3018439 [DOI] [PubMed] [Google Scholar]
  • 5.Ivanov D., Dremin V., Bykov A., Borisova E., Genova T., Popov A., Ossikovski R., Novikova T., Meglinski I., “Colon cancer detection by using Poincaré sphere and 2D polarimetric mapping of ex vivo colon samples,” J. Biophotonics 13(8), 1–10 (2020). 10.1002/jbio.202000082 [DOI] [PubMed] [Google Scholar]
  • 6.Kupinski M., Rehbinder J., Haddad H., Deby S., Vizet J., Teig B., Nazac A., Pierangelo A., Moreau F., Novikova T., “Tasked-based quantification of measurement utility for ex vivo multi-spectral Mueller polarimetry of the uterine cervix,” in Proc. SPIE-OSA, Clin. Preclin. Opt. Diagn. 10411, 104110N–1 (2017). [Google Scholar]
  • 7.Borovkova M., Bykov A., Popov A., Pierangelo A., Novikova T., Pahnke J., Meglinski I., “Evaluating β-amyloidosis progression in Alzheimer’s disease with Mueller polarimetry,” Biomed. Opt. Express 11(8), 4509 (2020). 10.1364/BOE.396294 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Kupinski M., Boffety M., Ossikovski R., Pierangelo A., Rehbinder J., Vizet J., Goudail F., Novikova T., “Diagnostics of high grade cervical intraepithelial neoplasia with Mueller matrix polarimetry,” Proc. SPIE Nov. Biophotonics Tech. Appl. V 1175, 1107509 (2019). [Google Scholar]
  • 9.Kruchinin V. N., Kruchinina M. V., Prudnikova Y. I., Spesivtsev E. V., Rykhlitskii S. V., Volodin V. A., Shekhovtsov S. V., Pel’tek S. E., “The use of spectral ellipsometry and raman spectroscopy in screening diagnostics of colorectal cancer,” Opt. Spectrosc. 127(1), 170–176 (2019). 10.1134/S0030400X19070142 [DOI] [Google Scholar]
  • 10.Derman D., Opar E., Ferhanoglu O., Polat Ö, Kazanci M., “Optical characterization of heat dependent collagen denaturation via Mueller matrix polarimetry in combination with principal component analysis,” Appl. Opt. 60(9), 2543 (2021). 10.1364/AO.418805 [DOI] [Google Scholar]
  • 11.Bai Y., Shu T., Su L., Zhang X., “Functional nucleic acid-based fluorescence polarization/anisotropy biosensors for detection of biomarkers,” Anal. Bioanal. Chem. 412(25), 6655–6665 (2020). 10.1007/s00216-020-02754-x [DOI] [PubMed] [Google Scholar]
  • 12.Ushenko V. A., Hogan B. T., Dubolazov A., Piavchenko G., Kuznetsov S. L., Ushenko A. G., Ushenko Y. O., Gorsky M., Bykov A., Meglinski I., “3D Mueller matrix mapping of layered distributions of depolarisation degree for analysis of prostate adenoma and carcinoma diffuse tissues,” Sci. Rep. 11(1), 5162 (2021). 10.1038/s41598-021-83986-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Chashchina O., Mezouar H., Vizet J., Raoux C., Park J., Ramón-Lozano C., Schanne-Klein M. C., Barakat A. I., Pierangelo A., “Mueller polarimetric imaging for fast macroscopic mapping of microscopic collagen matrix remodeling by smooth muscle cells,” Sci. Rep. 11(1), 5901 (2021). 10.1038/s41598-021-85164-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Jin G., Meng Y. H., Liu L., Niu Y., Chen S., Cai Q., Jiang T. J., “Development of biosensor based on imaging ellipsometry and biomedical applications,” Thin Solid Films 519(9), 2750–2757 (2011). 10.1016/j.tsf.2010.12.175 [DOI] [Google Scholar]
  • 15.Liu H., Shen J., Liu W., Niu Y., Jin G., “Imaging ellipsometry biosensor: Basic theory, principles of operation, and applications,” J. Vac. Sci. Technol., B: Nanotechnol. Microelectron.: Mater., Process., Meas., Phenom. 38(2), 024002 (2020). 10.1116/1.5129596 [DOI] [Google Scholar]
  • 16.Azzam R. M. A., Bashara N. M., Ellipsometry and Polarized Light (North Holland, 1987). [Google Scholar]
  • 17.Gerrard A., Burch J. M., Introduction to Matrix Methods in Optics (John Wiley and Sons, 1975). [Google Scholar]
  • 18.Savenkov S. N., Marienko V. V., Oberemok E. A., Sydoruk O., “Generalized matrix equivalence theorem for polarization theory,” Phys. Rev. E 74(5), 056607 (2006). 10.1103/PhysRevE.74.056607 [DOI] [PubMed] [Google Scholar]
  • 19.Pancharatnam S., “Achromatic combinations of birefringent plates Part II waveplate QWP,” Proc. - Indian Acad. Sci., Sect. A 41(4), 137–144 (1955). 10.1007/BF03047098 [DOI] [Google Scholar]
  • 20.Koester C. J., “Achromatic combinations of half-wave plates,” J. Opt. Soc. Am. 49(4), 405 (1959). 10.1364/JOSA.49.000405 [DOI] [Google Scholar]
  • 21.Messaadi A., Sánchez-López M. M., Vargas A., García-Martínez P., Moreno I., “Achromatic linear retarder with tunable retardance,” Opt. Lett. 43(14), 3277 (2018). 10.1364/OL.43.003277 [DOI] [PubMed] [Google Scholar]
  • 22.Saha A., Chakraborty S., Bhattacharya K., “Achromatic half-wave combination of birefringent plates,” Optik 125(16), 4534–4537 (2014). 10.1016/j.ijleo.2014.02.012 [DOI] [Google Scholar]
  • 23.Dimova E. S., Ivanov S. S., Popkirov G. S., Vitanov N. V., “Highly efficient broadband polarization retarders and tunable polarization filters made of composite stacks of ordinary wave plates,” J. Opt. Soc. Am. A 31(5), 952 (2014). 10.1364/JOSAA.31.000952 [DOI] [PubMed] [Google Scholar]
  • 24.Al-Mahmoud M., Coda V., Rangelov A., Montemezzani G., “Broadband polarization rotator with tunable rotation angle composed of three wave plates,” Phys. Rev. Appl. 13(1), 014048 (2020). 10.1103/PhysRevApplied.13.014048 [DOI] [Google Scholar]
  • 25.Kraemer M., Baur T., “Achromatic retarders in polarization optics,” Opt. Eng. 58(08), 1–15 (2019). 10.1117/1.OE.58.8.082406 [DOI] [Google Scholar]
  • 26.Baur T. G., Kraemer M., “The power of polymers and liquid crystals for polarization control,” Proc SPIE 11132, 111320L (2019). 10.1117/12.2530705 [DOI] [Google Scholar]
  • 27.Gol’tser I. V., Darsht M. Y., Zel’dovich B. Y., Kundikova N. D., Rogacheva L. F., “Quarter-wave plate tunable in a wide wavelength range,” Quantum Electron. 25(2), 187–190 (1995). 10.1070/QE1995v025n02ABEH000321 [DOI] [Google Scholar]
  • 28.Goltser I. V., Darscht M., Kundikova N., Zel’dovich B., “An adjustable quarter-wave plate,” Opt. Commun. 97(5-6), 291–294 (1993). 10.1016/0030-4018(93)90490-V [DOI] [Google Scholar]
  • 29.Kundikova N., Popkov I., Popkova A., “Composite quarter-wave systems with adjustable parameters,” Appl. Opt. 49(33), 6504–6511 (2010). 10.1364/AO.49.006504 [DOI] [PubMed] [Google Scholar]
  • 30.Darsht M. Y., Goltser I. V., Kundikova N., Zel’dovich B. Y., “Adjustable half-wave plate,” Appl. Opt. 34(19), 3658–3661 (1995). 10.1364/AO.34.003658 [DOI] [PubMed] [Google Scholar]
  • 31.Bibikova E., Kundikova N., “Properties of an adjustable quarter-wave system under conditions of multiple beam interference,” Appl. Opt. 52(9), 1852 (2013). 10.1364/AO.52.001852 [DOI] [PubMed] [Google Scholar]
  • 32.Hu Q., He C., Booth M. J., “Arbitrary complex retarders using a sequence of spatial light modulators as the basis for adaptive polarisation compensation,” J. Opt. 23(6), 065602 (2021). 10.1088/2040-8986/abed33 [DOI] [Google Scholar]
  • 33.Valeev A. I., Churikov V. M., “Simultaneous recording and readout of χ(2)-holograms in glasses,” Tech. Phys. Lett. 23(1), 29–31 (1997). 10.1134/1.1261604 [DOI] [Google Scholar]
  • 34.Antonov S., Kotov V., Sotnikov V. N., “Bragg polarizational light splitters based on the TeO2 crystal,” Zhurnal Tekhnicheskoi Fiziki. In Russian. 61, 168–173 (1991). [Google Scholar]
  • 35.Kotov M., Shkerdin G., Kotov E., Tikhomirov A., “Frequency bandwidth of the acousto-optic splitter of the two-colored radiation,” Prikladnaya fizika (Applied Physics) in Russian pp. 20–25 (2011).
  • 36.Bibikova E. A., Kundikova N. D., Rogacheva L. F., “Method for determining polarization parameters of thin phase plates,” Bull. Russ. Acad. Sci.: Phys. 70, 1469–1472 (2006). [Google Scholar]
  • 37.Bibikova E., Kundikova N., Mukhin Y., Betleni P., “Independent controlling of polarization states of two beams with different wavelengths,” Proc. SPIE 11646, 116460F (2021). 10.1117/12.2577152 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Wan Q., Coté G. L., Dixon J. B., “Dual-wavelength polarimetry for monitoring glucose in the presence of varying birefringence,” J. Biomed. Opt. 10(2), 024029 (2005). 10.1117/1.1891175 [DOI] [PubMed] [Google Scholar]
  • 39.Malik B. H., Coté G. L., “Real-time, closed-loop dual-wavelength optical polarimetry for glucose monitoring,” J. Biomed. Opt. 15(1), 017002 (2010). 10.1117/1.3290819 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Pirnstill C. W., Malik B. H., Gresham V. C., Coté G. L., “In vivo glucose monitoring using dual-wavelength polarimetry to overcome corneal birefringence in the presence of motion,” Diabetes Technol. Ther. 14(9), 819–827 (2012). 10.1089/dia.2012.0070 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.


Articles from Biomedical Optics Express are provided here courtesy of Optica Publishing Group