Skip to main content
AJNR: American Journal of Neuroradiology logoLink to AJNR: American Journal of Neuroradiology
. 1991 Jan-Feb;12(1):83-5.

Xenon-enhanced CT of the brain: effect of flow activation on derived cerebral blood flow measurements.

W F Good 1, D Gur 1
PMCID: PMC8367570  PMID: 1899526

Abstract

The errors associated with derivation of cerebral blood flow values by the xenon-enhanced CT method have been evaluated through computer simulations as a function of flow-activation patterns and different scanning protocols. The results of this study indicate that actual flow activation during inhalation increases the derived flow values in a systematic way. Compared with the errors introduced by CT noise and/or variations in scanning protocols, flow activation introduces relatively small errors in the derived flow value when the washin technique is used.

Full Text

The Full Text of this article is available as a PDF (1.4 MB).