Skip to main content
AJNR: American Journal of Neuroradiology logoLink to AJNR: American Journal of Neuroradiology
. 1988 Sep;9(5):825-8.

Truncation artifact in MR images of the intervertebral disk.

R K Breger 1, L F Czervionke 1, E G Kass 1, S W Yu 1, P S Ho 1, J A Strandt 1, J B Kneeland 1, V M Haughton 1
PMCID: PMC8367676  PMID: 3140625

Abstract

A cadaver's vertebral column, a phantom, and a volunteer were imaged on a 1.5-T MR scanner to study the thin, uniform, dark, transverse lines that characterize some intervertebral disks. An artifactual dark line appears when the field of view (FOV) and matrix steps (n) are chosen so that d = FOV/n, where d equals the intervertebral disk height, or spacing between phantom vertebrae. The artifact is caused by the truncation effect. An artifactual dark line is differentiated from a dark line caused by anatomic variables, and means for reducing such lines by modifying imaging parameters are discussed.

Full Text

The Full Text of this article is available as a PDF (2.2 MB).


Articles from AJNR: American Journal of Neuroradiology are provided here courtesy of American Society of Neuroradiology