Skip to main content
Frontiers in Neurology logoLink to Frontiers in Neurology
. 2021 Aug 12;12:686253. doi: 10.3389/fneur.2021.686253

Efficacy and Safety of a Novel Thrombectomy Device in Patients With Acute Ischemic Stroke: A Randomized Controlled Trial

Yongxin Zhang 1,, Weilong Hua 1,, Zifu Li 1, Ya Peng 2, Zhian Han 3, Tong Li 4, Congguo Yin 5, Shoucun Wang 6, Guangxian Nan 7, Zhenwei Zhao 8, Hua Yang 9, Bin Zhou 10, Tianxiao Li 11, Yiling Cai 12, Jianmin Zhang 13, Guifu Li 14, Xiaoxiang Peng 15, Sheng Guan 16, Junshan Zhou 17, Ming Ye 10, Liqin Wang 10, Lei Zhang 1, Bo Hong 1, Yongwei Zhang 1, Jieqing Wan 18, Yang Wang 19, Qing Zhu 20, Jianmin Liu 1,*, Pengfei Yang 1,*
PMCID: PMC8397519  PMID: 34456847

Abstract

Purpose: The Tonbridge stent is a novel retriever with several design improvements which aim to achieve promising flow reperfusion in the treatment of acute ischemic stroke (AIS). We conducted a randomized controlled, multicenter, non-inferiority trial to compare the safety and efficacy of the Tonbridge stent with the Solitaire FR.

Methods: AIS patients aged 18–85 years with large vessel occlusion in anterior circulation who could undergo puncture within 6 h of symptom onset were included. Randomization was performed on a 1:1 ratio to thrombectomy with either the Tonbridge stent or the Solitaire FR. The primary efficacy endpoint was successful reperfusion using a modified thrombolysis in cerebral infarction score (mTICI) of 2b/3. Safety outcomes were symptomatic intracranial hemorrhage (sICH) within 24 ± 6 h and all-cause mortality within 90 days. A clinically relevant non-inferiority margin of 12% was chosen as the acceptable difference between groups. Secondary endpoints included time from groin puncture to reperfusion, National Institutes of Health Stroke Scale (NIHSS) score at 24 h and at 7 days, and a modified Rankin Scale (mRS) score of 0–2 at 90 days.

Results: A total of 220 patients were enrolled; 104 patients underwent thrombectomy with the Tonbridge stent and 104 were treated with the Solitaire FR. In all test group patients, the Tonbridge was used as a single retriever without rescuing by other thrombectomy devices. Angioplasty with balloon and/or stent was performed in 26 patients in the Tonbridge group and 16 patients in the Solitaire group (p = 0.084). Before angioplasty, 86.5% of those in the Tonbridge group and 81.7% of those in the Solitaire group reached successful reperfusion (p = 0.343). Finally, more patients in the Tonbridge group achieved successful reperfusion (92.3 vs. 84.6%, 95% CI of difference value 0.9–16.7%, p < 0.0001). There were no significant differences on sICH within 24 ± 6 h between the two groups. All-cause mortality within 90 days was 13.5% in the Tonbridge group and 16.3% in the Solitaire group (p = 0.559). We noted no significant differences between groups on the NIHSS at either 24 h or 7 days and the mRS of 0–2 at 90 days.

Conclusion: The trial indicated that the Tonbridge stent was non-inferior to the Solitaire FR within 6 h of symptom onset in cases of large vessel occlusion stroke.

Clinical Trial Registration:ClinicalTrials.gov, number: NCT03210623.

Keywords: acute ischemic stroke, large vessel occlusion, thrombectomy, reperfusion, stent

Introduction

Five landmark trials have established endovascular thrombectomy as one of the most powerful treatments for acute ischemic stroke (AIS) due to large vessel occlusion in anterior circulation (15). The benefits shown in these trials were driven by improved stent-retriever thrombectomy devices combined with patient selection. The Solitaire FR (Medtronic Inc., Irvine, CA, USA) has been one of the most frequently used stent retrievers. Recently, various novel thrombectomy devices have been developed and put into use.

The Tonbridge stent (Tonbridge, Tonbridge Medical Technology, Zhuhai, Guangzhou, China) has been modified into a new design with a longitudinal spiral opening along its tubular surface (Figure 1). It has also been modified so that the finished temperature of nitinol increases the radial force, and has a broad size ranging from 3/4/5/6 mm. The maximum length of the series 4/5/6 mm is 30 mm. In in vitro tests, the Tonbridge stent had similar maximum friction within the 0.021-in. microcatheter and a slight increase in radial force when compared with the Solitaire FR. An in vivo comparative study in beagle models showed that the Tonbridge stent was safe and had a similar number of retriever attempts and similar recanalization rates when compared with the Solitaire FR (6). To evaluate the true efficacy and safety of this new device compared with that of the Solitaire FR in a clinical setting, a multicenter randomized controlled trial was designed and carried out.

Figure 1.

Figure 1

The Tonbridge device and the unique design of the longitudinal spiral opening along its tubular surface.

Methods

Study Design and Patients

We conducted a randomized, prospective, controlled, multicenter, single-blind, non-inferiority trial with blinded outcome assessment that enrolled patients with AIS. This study was designed with input from an academic steering committee and overseen by an independent clinical events committee as well as an independent core laboratory. The sponsor of the study, Tonbridge Medical Technology, was responsible for the logistical operations and monitoring of the trial. An independent contract research organization (OSMUNDA Medical Device Service Group, en.osmundacn.com/) and a site management organization (Excellence Future International Consulting Co. Ltd., http://www.chncro.com) were also involved in monitoring the study in order to ensure the quality of the trial. All statistical analyses were conducted by four independent external statisticians (Medical Research & Biometrics Center, National Center for Cardiovascular Disease, China).

We enrolled patients from 17 tertiary care centers, which were each required to have performed at least 30 endovascular thrombectomy procedures during the previous year. The protocol was approved by the respective ethics committee of each participating site. Written informed consent was obtained from all patient participants or their legal representatives prior to enrollment. The trial was designed to enroll 220 patients with the following eligibility criteria: (1) adults aged 18–85 years of age; (2) a baseline National Institutes of Health Stroke Scale (NIHSS) score <30; (3) an angiographically proven occlusion in the internal carotid artery (ICA), middle cerebral artery (MCA) (M1 or M2), or the anterior cerebral artery (MCA) (A1 or A2); (4) a prestroke modified Rankin Scale (mRS) (7) score of <2; and (5) patients able to undergo puncture within 6 h of symptom onset. Key exclusion criteria were the following: a massive cerebral infarction (8, 9); an Alberta Stroke Program Early CT Score (ASPECTS) (10) of <6, or infarct volume ≥70 ml, or >1/3 of blood supplying areas on CT/diffusion-weighted imaging; simultaneous acute bilateral carotid occlusion; uncontrolled hypertension (defined as SBP >185 mmHg or DBP >110 mmHg after medication); concomitant use of oral anticoagulation medications; an INR >3.0; and a platelet count of <40 ×109.

Randomization and Blinding

Randomization was performed utilizing a 1:1 ratio to mechanical thrombectomy with either the Tonbridge stent or the Solitaire FR. This was accomplished by employing a web-based system with stratification according to the participating site. Treatment-group assignment was known to the operating physicians but blinded to the patients. Three postprocedure clinical follow-up exams were performed by independent physicians who were unaware of the treatment-group assignment of the patient.

Procedures

According to guidelines, intravenous thrombolysis with rtPA was administered as a bridging therapy in patients who had no contraindications for it. The use of general anesthesia was performed according to standard practices based on local practice. A baseline angiogram was obtained before device deployment in order to assess angiographic inclusion and exclusion criteria. The choice of thrombectomy device was made according to random allocation. The instructions of the manufacturers regarding the Tonbridge stent were very similar to those of the Solitaire FR. Other retrievers, such as the Trevo stent (Stryker, Kalamazoo, MI, USA), or other techniques were allowed after three unsuccessful attempts with the Tonbridge stent. Aspiration with an intermediate catheter was allowed in both stent retriever arms. If there was an underlying stenosis or insufficient reperfusion, salvage measures, including additional balloon (Gateway, Boston Scientific, Natick, MA, USA) angioplasty and/or placement of a permanent stent (Enterprise, Johnson and Johnson, Raynham, MA, USA; Wingspan, Stryker, Kalamazoo, MI, USA) or an Apollo balloon-mounted stent (MicroPort, Shanghai, China), were allowed. If permanent stent deployment was performed, tirofiban (glycoprotein IIb/IIIa inhibitors) was provisionally administered, followed by a loading dose of clopidogrel, and aspirin was immediately administered orally. Daily oral dual antiplatelet therapy was started postprocedure and continued for 3 months. Subsequently, 100 mg of aspirin was prescribed for the rest of the lifetime of the patient (11).

Outcomes

The primary efficacy endpoint was successful reperfusion, defined as achieving modified thrombolysis in cerebral infarction (mTICI) (12) 2b or 3, in AIS patients assessed by an independent angiography core laboratory. All of the images were read by two experienced neuroradiologists, with consensus required in cases of discrepancy. First-pass effect and modified first-pass effect were also compared. First-pass effect was defined as achieving complete recanalization (mTICI 3) with a single thrombectomy device pass. Modified first-pass effect was defined as meeting mTICI 3/2b after the first pass (13, 14). Major safety outcomes included symptomatic intracranial hemorrhage (sICH) within 24 ± 6 h and all-cause mortality within 90 days, which were assessed by an independent clinical events committee. An sICH was defined as any ICH identified by CT scan combined with a four-point increase in NIHSS or death. The secondary endpoints were time from groin puncture to reperfusion, NIHSS at 24 h and at 7 days, favorable clinical outcomes (defined as an mRS of 0–2), and median mRS score at 90 days. The mRS score at 90 days was done by outpatient follow-up or telephone interview.

Statistical Analysis

The primary study hypothesis was that the rate of successful reperfusion in the Tonbridge group would be non-inferior to the rate in the Solitaire group. According to results previously reported in the literature, both devices were assumed to have a successful reperfusion rate of 90%. A clinically relevant non-inferiority margin of 12% was chosen as the acceptable difference between groups simultaneously, under a one-sided significance level alpha of 2.5% and an estimated 10% withdrawal or loss to follow-up rate. Under these conditions, randomizing a total of 220 patients would provide 80% power to demonstrate non-inferiority of the Tonbridge stent to that of the Solitaire FR.

Baseline data are presented as descriptive statistics according to treatment assignment, as appropriate. All statistical analyses followed the intention-to-treat principle. Continuous variables are presented as mean ± SD and categorical variables are presented as either counts or percentages. Normally distributed continuous variables were compared using Student's t-test. Categorical variables were compared using a chi-squared test or Fisher's exact test. For the primary endpoint, two-sided 95% confidence intervals (CIs) of differences in the rate of successful reperfusion between the groups were estimated by a Cochran–Mantel–Haenszel chi-squared test with adjusting center. The non-inferiority test was based on an asymptotic Z test. All the analyses were performed assuming a significance level of two-sided 0.05, using SAS software, version 9.4 (SAS Institute, Cary, NC, USA).

Results

Between August 3, 2017, and August 27, 2018, a total of 220 patients were enrolled in the trial (Figure 2). Eight patients had thrombus dissolution before the devices reached the target vessels, and four had protocol violations. A total of 104 patients were treated with the Tonbridge stent, while 104 underwent thrombectomy with the Solitaire FR. No patients crossed over or were lost to follow-up. The baseline characteristics of the patients are detailed in Table 1. The patient median age was 62 years in the Tonbridge group and 61 years in the Solitaire group. There were 52 target vessels in the ICA cases, 133 in the MCA M1 cases, 21 in the MCA M2 cases, and 2 in cases of ACA A2. There were no differences in baseline characteristics observed between the two groups.

Figure 2.

Figure 2

Randomization and treatment of the patients.

Table 1.

Baseline characteristics of the 208 patients.

Characteristic Tonbridge group
(n = 104)
Solitaire group
(n = 104)
P
Age—years, median (IQR) 63.98 (53.96–71.39) 66.50 (55.82–72.76) 0.910
Male sex, no. (%) 62 (59.6%) 61 (58.7%) 0.888
BMI, mean 24.1 ± 3.3 23.8 ± 3.8 0.631
Prestroke mRS 1, no. (%) 10 (9.6%) 14 (13.5%) 0.385
NIHSS—median (IQR) 15 (12–19) 16 (13–19) 0.431
ASPECTS—median (IQR) 9 (8–10) 9 (8–10) 0.580
Systolic blood pressure at hospital arrival—mmHg, median (IQR) 143 (129–160) 143 (127–153) 0.416
Glucose level at hospital arrival—mmol/L, median (IQR) 6.9 (6.0–8.5) 6.8 (6.1–7.8) 0.130
Time from stroke onset to groin puncture—min, median (IQR) 256 (193–315) 248 (187–306) 0.462
Intracranial arterial occlusion, no. (%)
ICA
MCA M1 segment
MCA M2 segment
ACA A1 segment
ACA A2 segment
21 (20.2%)
71 (68.3%)
11 (10.6%)
0 (0%)
1 (1.0%)
31 (29.8%)
62 (59.6%)
10 (9.6%)
0 (0.0%)
1 (1.0%)
0.461
Preprocedure mTICI, no. (%)
0
1
2a
2b
3
90 (86.5%)
10 (9.6%)
4 (3.8%)
0 (0%)
0 (0%)
97 (93.3%)
6 (5.8%)
1 (1.0%)
0 (0%)
0 (0%)
0.196

BMI, body mass index; IQR, interquartile range; mRS, modified Rankin Scale; NIHSS, National Institutes of Health Stroke Scale; ASPECTS, Alberta Stroke Program Early CT score; ICA, internal carotid artery; MCA, middle cerebral artery; ACA, anterior cerebral artery; mTICI, modified thrombolysis in cerebral infarction score.

Procedural results and outcomes are shown in Table 2. In 37 patients (18 in the Tonbridge group and 19 in the Solitaire group), bridging intravenous fibrinolysis was started before thrombectomy. There were no differences in bridging therapy between the two groups. No balloon guide catheters were used in either arm. In all 104 patients in the Tonbridge group, the Tonbridge was used as the single retriever with no rescuing needed using other thrombectomy devices. The use of a retriever in conjunction with aspiration was similar between the two groups. There were no significant differences observed in the median number of passes needed with the assigned study device between the groups. First-pass effect was achieved in 39/208 patients, with no significant differences noted between the Tonbridge group and the Solitaire group (18.3 vs. 19.2%, respectively). There was a similar result in the modified first-pass effect. Angioplasty with balloon and/or stent was performed as a remedial measure to maintain a stable flow in 26 patients in the Tonbridge group and 16 patients in the Solitaire group (p = 0.084). Before angioplasty, 90 patients (86.5%) in the Tonbridge group and 85 (81.7%) in the Solitaire group reached successful reperfusion (p = 0.343). Median time from groin puncture to successful reperfusion was similar between treatment groups. Finally, more patients in the Tonbridge group than those in the Solitaire group achieved successful reperfusion (92.3 vs. 84.6%, 95% CI of difference value 0.9%−16.7%, p < 0.0001).

Table 2.

Procedural and outcomes data.

Tonbridge group
(n = 104)
Solitaire group
(n = 104)
P
Bridging intravenous fibrinolysis, no. (%) 18 (17.3%) 19 (18.3%) 0.856
Retriever in conjunction with aspiration, no. (%) 4 (3.8%) 7 (6.7%) 0.353
Number of passes by retriever, median 1.5 1 0.641
Balloon and/or stent angioplasty, no. (%) 26 (25.0%) 16 (15.4%) 0.084
Primary outcome:Final successful reperfusion, no. (%)
96 (92.3%)

88 (84.6%)

<0.0001a
mTICI 2b/3 with a first pass, no. (%) 50 (48.1%) 46 (44.2%) 0.578b
mTICI 3 with a first pass, no. (%) 19 (18.3%) 20 (19.2%) 0.859b
mTICI 2b/3 before angioplasty, no. (%) 90 (86.5%) 85 (81.7%) 0.343b
Secondary outcomes:
Time from groin puncture to successful reperfusion—min, median (IQR)
NIHSS at 24 h, median (IQR)c
NIHSS at 7 days, median (IQR)d
mRS 0–2 at 90 days, no. (%)
mRS at 90 days, median (IQR)

67.5 (46.5–95.5)
10 (4–15)
6 (2–13)
61 (58.7%)
2 (1–4)

67 (44.0–99.0)
8 (4–16)
4 (1–13)
64 (61.5%)
2 (1–4)

0.970
0.993
0.777
0.671
1.000
Postprocedure mTICI, no. (%)
0
1
2a
2b
3
1 (1.0%)
3 (2.9%)
4 (3.8%)
64 (61.5%)
32 (30.8%)
4 (3.8%)
3 (2.9%)
9 (8.7%)
58 (55.8%)
30 (28.8%)
0.3138
Major safety outcomes:
Symptomatic intracranial hemorrhage within 24 h, no. (%)e
Death within 90 days, no. (%)

2 (1.9%)
14 (13.5%)

5 (5.0%)
17 (16.3%)

0.238
0.559

mTICI, the modified thrombolysis in cerebral infarction score; NIHSS, National Institutes of Health Stroke Scale; mRS, modified Rankin Scale.

a

The p-value was for noninferiority and calculated by the CMH chi-square test.

b

The p-value was for difference and calculated by the chi-square test.

c

Data were missing for one patient in the Tonbridge group and four patients in the Solitaire group.

d

Data were missing for seven patients in the Tonbridge group and 11 patients in the Solitaire group.

e

Data were missing for one patient in the Tonbridge group and three patients in the Solitaire group.

Regarding other group comparisons, differences in sICH within 24 h (1.9% in the Tonbridge group and 5.0% in the Solitaire group) and all-cause mortality within 90 days (13.5% in the Tonbridge group and 16.3% in the Solitaire group) were not significant. We also noted no significant differences between groups in terms of NIHSS at 24 h and NIHSS at 7 days. Rates of favorable outcome (mRS 0–2) at 90 days were 58.7% in the Tonbridge group and 61.5% in the Solitaire group (p = 0.671).

Discussion

Our randomized controlled trial demonstrated that the Tonbridge stent had a comparable rate of successful reperfusion when compared with the Solitaire FR in a clinical setting for the treatment of AIS due to large vessel occlusion in the anterior circulation. In terms of safety endpoints, sICH within 24 ± 6 h and all-cause mortality within 90 days were comparable for both devices. Clinical outcomes within 7 days and at 90 days follow-up were comparable between the two groups.

Although several approved thrombectomy devices are currently available, improvements designed to increase the effectiveness of reperfusion have continued. The Tonbridge is a novel stent-based clot retriever with a hybrid closed and partially open cell design, which forms a longitudinal spiral opening along its tubular surface. It was designed to allow increases in stent radial force strength and the tensile breaking force of the connection point between the stent and its push wire. In in vitro tests, the radial force and flexibility of the Tonbridge stent were superior to those of the Solitaire FR (6). These improvements may have the following potential advantages: the promise of tighter clot engagement and reduced chance of clot fall-off during retrieval; they make the device better adapted to curved vessel lumen; and they prevent connection site break-off during repeated retriever manipulations. Animal experiments revealed a slightly higher rate of recanalization (100 vs. 88.9%), but without statistical significance (6). To determine the value of this novel device in the treatment of AIS, this clinical comparative study was essential.

Our study is one of the few prospective randomized controlled trials that focuses on the efficacy and safety of a new stent-like device. Trials of thrombectomy devices have usually been single-arm studies whose aim was to show the safety of a device for regulatory approval purposes (1523). In these studies, the data from a single group were usually compared with the data from previous studies of other devices, such as the Solitaire FR or the Trevo. These two devices have been evaluated in numerous trials and prospective registries (16, 18). However, direct comparison of prognoses between single-arm studies and previous registries can produce uncontrolled effects from inhomogeneous baselines and different endpoints. In our study, the two groups had well-balanced baseline characteristics because of the prospective randomization design. To minimize bias, the assessments of key endpoints, including postprocedural mTICI and sICH, were evaluated by an independent core laboratory and clinical events committee.

The Tonbridge stent achieved a similar rate of successful reperfusion as the Solitaire FR before angioplasty. We also compared first-pass effect and modified first-pass effect between the two groups. These indicators were first described by Zaidat et al. (13) and have subsequently been shown to be independent factors of favorable outcomes (14, 24, 25). In this trial, a first-pass effect was achieved in one-fifth of all patients, which was slightly lower than that reported by Zaidat as well as a recently published systematic review (13, 26). Potential reasons may be a lack of the use of balloon guide catheters, which has been proven to contribute to a greater first-pass effect. Our trial indicates a similar rate of modified first-pass effect to that of previous studies (13, 14, 2426).

The fact that acute arterial occlusions due to intracranial atherosclerotic disease (ICAD) are more prevalent in Eastern populations was an issue. Recent studies have shown that ICAD-related occlusion accounts for 12%−30% of all large vessel occlusion etiologies in Asia (2730). However, while the thrombectomy devices could achieve initial reperfusion quickly, reocclusion is often encountered due to subsequent platelet aggregation at the lesion site or elastic recoil. In our study, about one-fifth of patients received angioplasty with a balloon and/or stent in order to maintain successful reperfusion. In line with a study published by Yoon et al., these patients achieved a higher recanalization rate (31). The final successful reperfusion rate we report is higher than those reported in the MR CLEAN, REVASCAT, and ESCAPE trials (13) and comparable with those seen in the SWIFT, PRIME, and EXTEND-IA trials (4).

Another concern about the safety of the device was whether the high radial force of the Tonbridge stent would lead to vessel wall injury or hemorrhage. Gascou et al. estimated that >10% of perioperative complications are associated with thrombectomy devices (32), including arterial perforation (0.9–4.9%), subarachnoid hemorrhage (0.6–4.9%), arterial dissection (0.6–3.9%), and vasospasm (3.9–23%) (27, 33). It is well-known that high radial force is associated with higher rates of clot removal (34). However, high radial force is also associated with endothelial vessel wall injury (35, 36). Therefore, minimizing radial force may reduce the incidence of vessel wall injury complications while still maintaining optimal clot retrieval rates. In an in vitro test, the radial force of the Tonbridge stent was noted to be slightly higher than that of the Solitaire FR. In the present clinical trial, we did not note any significant differences between the two groups in terms of symptomatic intracranial hemorrhage, including subarachnoid hemorrhage. These results indicate that the Tonbridge stent has the potential to improve the rate of reperfusion without increasing vessel wall injury complications.

Our study was limited in many aspects by the restrictive nature of a randomized controlled trial design, such as strict inclusion and exclusion criteria, center selection, and operator experience. The image core lab evaluated the DSA only without CTA or MRA, and the analysis item was very limited. We acknowledged that these were all the limitations. Postmarketing clinical registries allow for more inclusive criteria by including a range of clinical sites, operators, and varying patient populations, which will provide valuable information on the generalizability and reproducibility of our results, and allow additional opportunities to explore clinical hypotheses for patients treated outside of this trial.

Conclusions

In this randomized clinical trial, the Tonbridge stent was non-inferior to the Solitaire FR in the treatment of large vessel occlusion stroke within 6 h of symptom onset.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics Statement

The studies involving human participants were reviewed and approved by Shanghai Changhai Hospital Ethics Committee. The patients/participants provided their written informed consent to participate in this study.

Author Contributions

YongxZ and WH: manuscript writing and revision. JL and PY: conception and design of the work. All authors: data collection. All authors read and approved the final manuscript.

Conflict of Interest

QZ was employed by the company Zhuhai Ton-Bridge Medical Tech. Co., Ltd. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest. The authors declare that this study received funding from Zhuhai Ton-Bridge Medical Tech. Co., Ltd. The funder had the following involvement in the study: study design.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Footnotes

Funding. This work was supported by Zhuhai Ton-Bridge Medical Tech. Co., Ltd. and National Natural Science Foundation (82071287).

References

  • 1.Berkhemer OA, Fransen PS, Beumer D, van den Berg LA, Lingsma HF, Yoo AJ, et al. A randomized trial of intraarterial treatment for acute ischemic stroke. N Engl J Med. (2015) 372:11–20. 10.1056/NEJMoa1411587 [DOI] [PubMed] [Google Scholar]
  • 2.Jovin TG, Chamorro A, Cobo E, de Miquel MA, Molina CA, Rovira A, et al. Thrombectomy within 8 hours after symptom onset in ischemic stroke. N Engl J Med. (2015) 372:2296–306. 10.1056/NEJMoa1503780 [DOI] [PubMed] [Google Scholar]
  • 3.Goyal M, Demchuk AM, Menon BK, Eesa M, Rempel JL, Thornton J, et al. Randomized assessment of rapid endovascular treatment of ischemic stroke. N Engl J Med. (2015) 372:1019–30. 10.1056/NEJMoa1414905 [DOI] [PubMed] [Google Scholar]
  • 4.Campbell BC, Mitchell PJ, Kleinig TJ, Dewey HM, Churilov L, Yassi N, et al. Endovascular therapy for ischemic stroke with perfusion-imaging selection. N Engl J Med. (2015) 372:1009–18. 10.1056/NEJMoa1414792 [DOI] [PubMed] [Google Scholar]
  • 5.Saver JL, Goyal M, Bonafe A, Diener HC, Levy EI, Pereira VM, et al. Stent-retriever thrombectomy after intravenous t-pa vs. T-pa alone in stroke. N Engl J Med. (2015) 372:2285–95. 10.1056/NEJMoa1415061 [DOI] [PubMed] [Google Scholar]
  • 6.Zhou G, Li M, Lu H, Deng J, Zhao Y, Wang YXJ, et al. No inferiority of tonbridge thrombectomy device for acute thrombus retrial compared with solitaire device: an experimental evaluation with a canine distal external carotid-maxillary artery occlusion model. J Neurointerv Surg. (2018) 10:1085–91. 10.1136/neurintsurg-2017-013673 [DOI] [PubMed] [Google Scholar]
  • 7.van Swieten JC, Koudstaal PJ, Visser MC, Schouten HJ, van Gijn J. Interobserver agreement for the assessment of handicap in stroke patients. Stroke. (1988) 19:604–7. 10.1161/01.STR.19.5.604 [DOI] [PubMed] [Google Scholar]
  • 8.Yang P, Zhang Y, Liu J. Last2 ch2ance: a summary of selection criteria for thrombectomy in acute ischemic stroke. Am J Neuroradiol. (2017) 38:E58–9. 10.3174/ajnr.A5249 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Powers WJ, Rabinstein AA, Ackerson T, Adeoye OM, Bambakidis NC, Becker K, et al. 2018 guidelines for the early management of patients with acute ischemic stroke: a guideline for healthcare professionals from the american heart association/american stroke association. Stroke. (2018) 49:e46–99. 10.1161/STR.0000000000000172 [DOI] [PubMed] [Google Scholar]
  • 10.Barber PA, Demchuk AM, Zhang J, Buchan AM. Validity and reliability of a quantitative computed tomography score in predicting outcome of hyperacute stroke before thrombolytic therapy. aspects study group. Alberta stroke programme early ct score. Lancet. (2000) 355:1670–4. 10.1016/S0140-6736(00)02237-6 [DOI] [PubMed] [Google Scholar]
  • 11.Yang W, Zhang Y, Li Z, Zhang L, Li H, Hua W, et al. Differences in safety and efficacy of endovascular treatment for acute ischemic stroke: a propensity score analysis of intracranial atherosclerosis-related occlusion versus embolism. Clin Neuroradiol. (2020) 31:457–64. 10.1007/s00062-020-00899-x [DOI] [PubMed] [Google Scholar]
  • 12.Zaidat OO, Yoo AJ, Khatri P, Tomsick TA, von Kummer R, Saver JL, et al. Recommendations on angiographic revascularization grading standards for acute ischemic stroke: a consensus statement. Stroke. (2013) 44:2650–63. 10.1161/STROKEAHA.113.001972 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Zaidat OO, Castonguay AC, Linfante I, Gupta R, Martin CO, Holloway WE, et al. First pass effect. Stroke. (2018) 49:660–6. 10.1161/STROKEAHA.117.020315 [DOI] [PubMed] [Google Scholar]
  • 14.Zaidat OO, Haussen DC, Hassan AE, Jadhav AP, Mehta BP, Mokin M, et al. Impact of stent retriever size on clinical and angiographic outcomes in the stratis stroke thrombectomy registry. Stroke. (2019) 50:441–7. 10.1161/STROKEAHA.118.022987 [DOI] [PubMed] [Google Scholar]
  • 15.Mueller-Kronast NH, Zaidat OO, Froehler MT, Jahan R, Aziz-Sultan MA, Klucznik RP, et al. Systematic evaluation of patients treated with neurothrombectomy devices for acute ischemic stroke. Stroke. (2017) 48:2760–8. 10.1161/STROKEAHA.117.016456 [DOI] [PubMed] [Google Scholar]
  • 16.Zaidat OO, Castonguay AC, Nogueira RG, Haussen DC, English JD, Satti SR, et al. Trevo stent-retriever mechanical thrombectomy for acute ischemic stroke secondary to large vessel occlusion registry. J Neurointerv Surg. (2018) 10:516–24. 10.1136/neurintsurg-2017-013328 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Zaidat OO, Bozorgchami H, Ribó M, Saver JL, Mattle HP, Chapot R, et al. Primary results of the multicenter arise ii study (analysis of revascularization in ischemic stroke with embotrap). Stroke. (2018) 49:1107–15. 10.1161/STROKEAHA.117.020125 [DOI] [PubMed] [Google Scholar]
  • 18.Zaidat OO, Castonguay AC, Gupta R, Sun CJ, Martin C, Holloway WE, et al. North american solitaire stent retriever acute stroke registry: post-marketing revascularization and clinical outcome results. J Neurointerv Surg. (2018) 10:i45–i9 10.1136/neurintsurg-2018-014098 [DOI] [PubMed] [Google Scholar]
  • 19.Sakai N, Ota S, Matsumoto Y, Kondo R, Satow T, Kubo M, et al. Efficacy and safety of revive se thrombectomy device for acute ischemic stroke: river japan (reperfuse ischemic vessels with endovascular recanalization device in japan). Neurol Med-Chir. (2018) 58:164–172 10.2176/nmc.oa.2017-0145 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Siemonsen S, Forkert ND, Bernhardt M, Thomalla G, Bendszus M, Fiehler J. Eric acute stroke recanalization: a study using predictive analytics to assess a new device for mechanical thrombectomy. Int J Stroke. (2017) 12:659–66. 10.1177/1747493017700661 [DOI] [PubMed] [Google Scholar]
  • 21.Kühn AL, Wakhloo AK, Lozano JD, Massari F, De Macedo Rodrigues K, Marosfoi MG, et al. Two-year single-center experience with the ‘baby trevo' stent retriever for mechanical thrombectomy in acute ischemic stroke. J Neurointerv Surg. (2017) 9:541–6. 10.1136/neurintsurg-2016-012454 [DOI] [PubMed] [Google Scholar]
  • 22.Kabbasch C, Mpotsaris A, Liebig T, Söderman M, Holtmannspötter M, Cronqvist M, et al. First-in-man procedural experience with the novel embotrap® revascularization device for the treatment of ischemic stroke—a european multicenter series. Clin Neuroradiol. (2016) 26:221–8. 10.1007/s00062-014-0352-0 [DOI] [PubMed] [Google Scholar]
  • 23.Gupta R, Saver JL, Levy E, Zaidat OO, Yavagal D, Liebeskind DS, et al. New class of radially adjustable stentrievers for acute ischemic stroke: primary results of the multicenter tiger trial. Stroke. (2021) 52:1534–44. 10.1161/STR.0000000000000376 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Nikoubashman O, Dekeyzer S, Riabikin A, Keulers A, Reich A, Mpotsaris A, et al. True first-pass effect. Stroke. (2019) 50:2140–6. 10.1161/STROKEAHA.119.025148 [DOI] [PubMed] [Google Scholar]
  • 25.Ducroux C, Piotin M, Gory B, Labreuche J, Blanc R, Ben Maacha M, et al. First pass effect with contact aspiration and stent retrievers in the aspiration versus stent retriever (aster) trial. J Neurointerv Surg. (2020) 12:386–91. 10.1136/neurintsurg-2019-015215 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Abbasi M, Liu Y, Fitzgerald S, Mereuta OM, Arturo LJ, Rizvi A, et al. Systematic review and meta-analysis of current rates of first pass effect by thrombectomy technique and associations with clinical outcomes. J Neurointerv Surg. (2021) 13:212–6. 10.1136/neurintsurg-2020-016869 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Goyal M, Menon BK, van Zwam WH, Dippel DW, Mitchell PJ, Demchuk AM, et al. Endovascular thrombectomy after large-vessel ischaemic stroke: a meta-analysis of individual patient data from five randomised trials. Lancet. (2016) 387:1723–31. 10.1016/S0140-6736(16)00163-X [DOI] [PubMed] [Google Scholar]
  • 28.Suri MF, Qiao Y, Ma X, Guallar E, Zhou J, Zhang Y, et al. Prevalence of intracranial atherosclerotic stenosis using high-resolution magnetic resonance angiography in the general population: the atherosclerosis risk in communities study. Stroke. (2016) 47:1187–93. 10.1161/STROKEAHA.115.011292 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Horie N, Tateishi Y, Morikawa M, Morofuji Y, Hayashi K, Izumo T, et al. Acute stroke with major intracranial vessel occlusion: characteristics of cardioembolism and atherosclerosis-related in situ stenosis/occlusion. J Clin Neurosci. (2016) 32:24–9. 10.1016/j.jocn.2015.12.043 [DOI] [PubMed] [Google Scholar]
  • 30.Kim GE, Yoon W, Kim SK, Kim BC, Heo TW, Baek BH, et al. Incidence and clinical significance of acute reocclusion after emergent angioplasty or stenting for underlying intracranial stenosis in patients with acute stroke. AJNR Am J Neuroradiol. (2016) 37:1690–5. 10.3174/ajnr.A4770 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Yoon W, Kim SK, Park MS, Kim BC, Kang HK. Endovascular treatment and the outcomes of atherosclerotic intracranial stenosis in patients with hyperacute stroke. Neurosurgery. (2015) 76:680–6. 10.1227/NEU.0000000000000694 [DOI] [PubMed] [Google Scholar]
  • 32.Zaidi SF, Castonguay AC, Jumaa MA, Malisch TW, Linfante I, Marden FA, et al. Intraarterial thrombolysis as rescue therapy for large vessel occlusions. Stroke. (2019) 50:1003–6. 10.1161/STROKEAHA.119.025820 [DOI] [PubMed] [Google Scholar]
  • 33.Balami JS, White PM, McMeekin PJ, Ford GA, Buchan AM. Complications of endovascular treatment for acute ischemic stroke: prevention and management. Int J Stroke. (2018) 13:348–61. 10.1177/1747493017743051 [DOI] [PubMed] [Google Scholar]
  • 34.Machi P, Jourdan F, Ambard D, Reynaud C, Lobotesis K, Sanchez M, et al. Experimental evaluation of stent retrievers' mechanical properties and effectiveness. J Neurointerv Surg. (2017) 9:257–63. 10.1136/neurintsurg-2015-012213 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Peschillo S, Tomasello A, Diana F, Hernandez D, Toccaceli G, Rosal-Fontana M, et al. Comparison of subacute vascular damage caused by adapt versus stent retriever devices after thrombectomy in acute ischemic stroke: histological and ultrastructural study in an animal model. Intervent Neurol. (2018) 7:501–512. 10.1159/000491028 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Chi Hang Chon, Zhen Qin, Lam AK, Kwok JC, Yuen MM, Lam DC. In-vitro testing of rf-enabled low force mechanical thrombectomy for ischemic stroke. Annu Int Conf IEEE Eng Med Biol Soc. (2015) 2015:1349–52. 10.1109/EMBC.2015.7318618 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.


Articles from Frontiers in Neurology are provided here courtesy of Frontiers Media SA

RESOURCES