Skip to main content
Microorganisms logoLink to Microorganisms
. 2021 Aug 10;9(8):1700. doi: 10.3390/microorganisms9081700

Reduced Relative Sensitivity of the Elecsys SARS-CoV-2 Antigen Assay in Saliva Compared to Nasopharyngeal Swabs

Annette Audigé 1, Jürg Böni 1, Peter W Schreiber 2, Thomas Scheier 2, Roberto Buonomano 3, Alain Rudiger 4, Dominique L Braun 2, Gerhard Eich 5, Dagmar I Keller 6, Barbara Hasse 2, Christoph Berger 7, Huldrych F Günthard 1,2, Amapola Manrique 1, Alexandra Trkola 1,*,, Michael Huber 1,*,
Editor: Didier Hober
PMCID: PMC8401978  PMID: 34442779

Abstract

Early identification and isolation of SARS-CoV-2-infected individuals is central to contain the COVID-19 pandemic. Nasopharyngeal swabs (NPS) serve as a specimen for detection by RT-PCR and rapid antigen screening tests. Saliva has been confirmed as a reliable alternative specimen for RT-PCR and has been shown to be valuable for diagnosing children and in repetitive mass testing due to its non-invasive collection. Combining the advantages of saliva with those of antigen tests would be highly attractive to further increase test capacities. Here, we evaluated the performance of the Elecsys SARS-CoV-2 Antigen assay (Roche) in RT-PCR-positive paired NPS and saliva samples (N = 87) and unpaired NPS (N = 100) with confirmed SARS-CoV-2 infection (Roche cobas SARS-CoV-2 IVD test). We observed a high positive percent agreement (PPA) of the antigen assay with RT-PCR in NPS, reaching 87.2% across the entire cohort, whereas the overall PPA for saliva was insufficient (40.2%). At Ct values ≤ 28, PPA were 100% and 91.2% for NPS and saliva, respectively. At lower viral loads, the sensitivity loss of the antigen assay in saliva was striking. At Ct values ≤ 35, the PPA for NPS remained satisfactory (91.5%), whereas the PPA for saliva dropped to 46.6%. In conclusion, saliva cannot be recommended as a reliable alternative to NPS for testing with the Elecsys Anti-SARS-CoV-2 Antigen assay. As saliva is successfully used broadly in combination with RT-PCR testing, it is critical to create awareness that suitability for RT-PCR cannot be translated to implementation in antigen assays without thorough evaluation of each individual test system.

Keywords: SARS-CoV-2, saliva, RT-PCR, antigen assay

1. Introduction

Severe Acute Respiratory Syndrome Coronavirus-2 (SARS-CoV-2) infection is routinely diagnosed by reverse-transcription polymerase chain reaction (RT-PCR) in nasopharyngeal swabs (NPS). In early infection, NPS are considered the optimal material for detection [1]. However, NPS collection requires trained personnel, exposes the personnel to a potential risk of nosocomial transmission, and is invasive. Saliva circumvents these problems and has proven advantageous for children, for whom conducting an NPS is challenging, as well as for adults, when frequent testing is required. A considerable advantage of saliva is that it does not require trained personnel for sampling and is amenable for home-collection. The overall sensitivity and thus suitability of saliva as an alternative to NPS has been differentially debated, with most studies reporting an acceptable performance of saliva despite a generally lower sensitivity and viral load compared to NPS [2,3,4,5,6,7,8,9,10,11,12,13,14,15,16], some studies showing a substantially lower sensitivity [17,18], and others reporting higher viral loads in saliva relative to NPS [19,20]. Differential saliva collection strategies, RT-PCR test systems, and small and heterogeneous study populations may cause discrepancies in the sensitivity of detecting SARS-CoV-2 infection in saliva. Therefore, controlled evaluations in large, well-defined cohorts are critical. To this end, we recently conducted a large-scale head-to-head comparison of SARS-CoV-2 detection by RT-PCR in NPS and saliva of adults and children in a test center setting [21]. We demonstrated that saliva is a reliable alternative specimen for the detection of SARS-CoV-2 by RT-PCR with particular advantages for testing children.

While saliva tested with RT-PCR has become a widely used strategy, additional testing options are in demand. Mass testing strategies would benefit from test formats that are rapid, simple, accurate, and non-invasive, which could be offered, for instance, by testing saliva with a SARS-CoV-2 antigen assay. Provided that saliva allows SARS-CoV-2 detection by antigen tests with sufficient sensitivity, this would combine the advantages of antigen testing, namely the ability to provide cost-effective results within minutes, with the advantages of non-invasive, self-collected saliva.

Here, we investigated the suitability of saliva as an alternative specimen for SARS-CoV-2 antigen detection by the Elecsys SARS-CoV-2 Antigen assay (Roche, Basel, Switzerland). We evaluated the performance of the assay for saliva samples in comparison to NPS by determining the agreement and correlation between the Elecsys Anti-SARS-CoV-2 Antigen assay and the cobas SARS-CoV-2 IVD test (Roche, Basel, Switzerland) stratified by sample type and SARS-CoV-2 E-gene Ct values. We confirmed that NPS samples have a high Positive Percent Agreement (PPA) between the antigen assay and the RT-PCR test. In contrast, the antigen assay had a much lower performance with saliva samples. In the latter sample type, SARS-CoV-2 could only be detected reliably in samples with Ct values ≤ 26. Saliva is thus not a suitable specimen in combination with the Roche Elecsys Anti-SARS-CoV-2 Antigen assay.

2. Materials and Methods

2.1. Study Population

Probed specimen were leftover from routine diagnostics and a previous study [21] approved with an ethics waiver by the Zurich Cantonal Ethics Commission (Req-2020-00398). No additional personal data beyond the usual information on symptoms and duration required by the Swiss Federal Office of Public Health for all SARS-CoV-2 tests in Switzerland were collected.

For nasopharyngeal sampling, conventional swab and virus transport medium (VTM)/universal transport medium (UTM) were used. Transport media included in-house VTM (HEPES, DMEM, FCS, antibiotics, antimycotics) and Virus Preservative Medium (Improviral) for paired PCR-positive NPS; Liquid amies preservation medium (Copan) for unpaired PCR-positive NPS; in-house VTM for paired PCR-negative NPS. Saliva collection was conducted as previously described [21]. Briefly, individuals were asked to clear the throat thoroughly and collect saliva up to three times into an empty saliva collection tube to collect 0.5–1 mL (approx. a teaspoon full). Immediately after saliva collection, VTM was added to the crude saliva and the content mixed through gentle tilting. Saliva was collected directly after NPS for paired analysis and both specimens were immediately sent for SARS-CoV-2 RT-PCR testing. Residual material was stored at −80 °C until use in the present study.

2.2. Specimen

RT-PCR tested (Roche cobas SARS-CoV-2 IVD test), cryo-preserved (−80 °C) NPS and saliva specimens were included. Paired SARS-CoV-2 NPS and saliva included positive (N = 87) and negative tested samples (N = 220) collected in the framework of a prior study that conducted a prospective head-to-head comparison of the specimen by RT-PCR [21]. In addition, SARS-CoV-2 positive NPS samples from routine diagnostic testing were included (N = 100).

2.3. Quantitative SARS-CoV-2 PCR

NPS and saliva were processed identically using the procedures established for NPS in the diagnostics laboratory of the Institute of Medical Virology, University of Zurich. A total of 500 µL of NPS or saliva in VTM were diluted in 500 µL of Nuclisens easyMAG Lysis Buffer (bioMérieux, Marcy-l’Étoile, France), centrifuged (2000 rpm, 5 min), and analyzed with the cobas SARS-CoV-2 IVD test (Roche, Basel, Switzerland) on a Cobas 6800. RT-PCR testing for paired NPS and saliva was done in parallel on the same day as part of a previously published study [21]. SARS-CoV-2 detection was further quantified using SARS-CoV-2 Frankfurt 1 RNA as a calibrator (European Virus Archive, 004N-02005) allowing to report both Ct and genome equivalents. E-gene Ct values were used for comparison. In one NPS sample, the E-gene Ct value was considerably higher than the ORF1-gene Ct value (37.31 vs. 26.8); in this case, the ORF1-gene Ct value was used for the analysis.

2.4. SARS-CoV-2 Antigen Assay

Samples were thawed at room temperature and analyzed with the Elecsys SARS-CoV-2 Antigen assay (Roche, Basel, Switzerland) on a cobas e 411 analyzer (Roche, Basel, Switzerland) according to the manufacturer’s instructions. Briefly, Elecsys SARS-CoV-2 Antigen is an electrochemiluminescence immunoassay, which detects the nucleocapsid (N) antigen of SARS-CoV-2. The assay uses an antibody sandwich principle. Results are recorded as the Cutoff Index (COI).

Saliva samples were treated identically to NPS. Single measurements were conducted.

2.5. Data Analysis

Positive Percent Agreement (PPA), Negative Percent Agreement (NPA), and Overall Percent Agreement (OPA) between the Elecsys Anti-SARS-CoV-2 Antigen assay and the cobas SARS-CoV-2 IVD test were calculated. PPA was determined and stratified by Ct values (PPAoverall and PPA < Ct). PPA, NPA, and OPA were calculated using R (version 4.0.2); 95% confidence intervals (CI, two-sided) were calculated with the epiR package (version 1.0.15). Analysis of correlation (Spearman correlation) between the E-gene Ct value and antigen assay Cutoff Index (COI) and linear regression analysis were performed with GraphPad Prism (version 8.4.3). p values < 0.05 were considered statistically significant. All raw data are available in Supplemental Table S1.

3. Results

3.1. Lower Qualitative Agreement of the Elecsys Anti-SARS-CoV-2 Antigen Assay with RT-PCR in Saliva Than in NPS

The Elecsys SARS-CoV-2 Antigen assay has been established for the detection of SARS-CoV-2 in nasopharyngeal and oropharyngeal swab samples from patients with signs and symptoms suggestive of COVID-19 or known or suspected exposure to SARS-CoV-2 (Elecsys SARS-CoV-2 Antigen Assay Method sheet v1 (December 2020)). Here, we probed its capacity in detecting SARS-CoV-2 in saliva by comparing the performance in specimens previously tested by RT-PCR (cobas SARS-CoV-2 IVD). PCR-positive samples included paired NPS-saliva samples (N = 87) and unpaired NPS samples (N = 100). PCR-negative samples comprised paired NPS-saliva samples (N = 220). Symptomatic and asymptomatic donors were included (Table 1).

Table 1.

Donor demographics.

SARS-CoV-2 Positive Positive Negative
Unpaired NPS
(N = 100)
Paired NPS/Saliva
(N = 87)
Paired NPS/Saliva
(N = 220)
Male/Female (%) 45 (45%)/
55 (55%)
56 (64%)/
31 (36%)
117 (53%)/
103 (47%)
Age median (range) 52 (13–98) 35 (16–76) 29 (16–77)
Symptomatic (%) 94 (94%) 77 (88%) 167 (76%)
Asymptomatic (%) 5 (5%) 6 (7%) 50 (23%)
No information on symptoms (%) 1 (1%) 4 (5%) 3 (1%)
Median days of symptoms (range) 2 (1–28) 2 (1–15) 3 (1–14)
No information on duration of symptoms (%) 2 (2%) 16 (21%) 33 (20%)

We first established the positive percent agreement (PPA) of the Elecsys Anti-SARS-CoV-2 Antigen assay using NPS or saliva with the cobas SARS-CoV-2 IVD test. The overall PPA for NPS (including paired and unpaired NPS; N = 187) was 87.2%, with 100% PPA for samples with Ct values ≤ 28 and respectable 91.5% PPA for samples with Ct values ≤ 35 (Table 2). In the paired NPS-saliva cohort (N = 87), these values were similar for NPS, reaching PPAoverall = 86.2%, PPA<Ct28 = 100%, and PPA<Ct35 = 93.7%, whereas saliva (N = 87) performed clearly less well, yielding PPAoverall = 40.2%, PPA<Ct26 = 100%, PPA<Ct28 = 91.2%, and PPA<Ct35 = 46.6%, only. Thus, a reliable detection of SARS-CoV-2 in saliva with the Roche Elecsys Anti-SARS-CoV-2 Antigen assay was only possible for samples with Ct values ≤ 26, showing the lower performance of the antigen assay with saliva compared to NPS.

Table 2.

Overview of positive percent agreement (PPA) of the Roche Elecsys Anti-SARS-CoV-2 Antigen assay with the Roche cobas SARS-CoV-2 IVD test in NPS and saliva.

Cobas E-Gene Ct Value a NPS (All) NPS (Paired) Saliva
N
(cum. b)
PPA, %
(95% CI)
N
(cum.)
PPA, %
(95% CI)
N
(cum.)
PPA, %
(95% CI)
≤26 134 100
97.3–100
57 100
93.7–100.0
22 100
84.6–100
≤27 138 100
97.4–100
60 100
94.0–100.0
26 96.2
80.4–99.9
≤28 150 100
97.6–100
65 100
94.5–100.0
34 91.2
76.3–98.1
≤29 156 99.4
96.5–100
70 98.6
92.3–100
37 86.5
71.2–95.5
≤30 160 98.8
95.6–99.8
73 97.3
90.4–99.7
42 78.6
63.2–89.7
≤31 162 98.8
95.6–99.9
75 97.3
90.7–99.7
50 66.0
51.2–78.8
≤32 165 97.0
93.1–99
75 97.3
90.7–99.7
60 55.0
41.6–67.9
≤33 170 94.7
90.2–97.6
75 97.3
90.7–99.7
65 50.8
38.1–63.4
≤34 172 93.6
88.8–96.8
76 96.1
88.9–99.2
72 47.2
35.3–59.3
≤35 177 91.5
86.4–95.2
79 93.7
85.8–97.9
73 46.6
34.8–58.6
Overall
(≤40)
187 87.2
81.5–91.6
87 86.2
77.1–92.7
87 40.2
29.9–51.3

a sample type-specific Ct values; b cumulative.

To assess the specificity of the Roche Elecsys Anti-SARS-CoV-2 Antigen assay, we next determined the Negative Percent Agreement (NPA) and Overall Percent Agreement (OPA) of the Roche Elecsys Anti-SARS-CoV-2 Antigen assay in reference to the cobas SARS-CoV-2 IVD test. NPA was 100% and 99.5% for NPS and saliva, respectively (Table 3). The lower NPA in saliva was due to a single saliva sample that yielded a positive result (COI = 1.41). A repeat analysis of this sample retrieved a negative result (COI = 0.612). As all other samples were measured only once, the first positive result was considered for analysis. In sum, the Elecsys Anti-SARS-CoV-2 Antigen assay showed a high specificity both in NPS and saliva. The Overall Percent Agreement (OPA) was 94.1% and 82.7% for NPS and saliva, respectively (Table 3).

Table 3.

NPA and OPA of the Roche Elecsys Anti-SARS-CoV-2 Antigen assay with the Roche cobas SARS-CoV-2 IVD test.

N NPA, %
(95% CI)
OPA, %
(95% CI)
NPS 220 (NPA)
407 (OPA)
100
98.3–100
94.1
91.4–96.2
Saliva 220 (NPA)
307 (OPA)
99.5
97.4–100
82.7
78.0–86.8

3.2. Low Qualitative Agreement of Antigen Measurements between Paired NPS and Saliva

To assess the capacity of the antigen assay to detect SARS-CoV-2 more directly, we next determined the agreement between the antigen assay results obtained for NPS and saliva in the paired sample cohort. This analysis yielded very low percentages for PPA (41.3%), NPA (66.7%), and OPA (44.8%), underlining the limited performance of saliva testing with the antigen assay (Table 4).

Table 4.

PPA, NPA, and OPA between Roche Elecsys Anti-SARS-CoV-2 Antigen results in paired NPS and saliva.

N
NPS/Saliva
PPA, %
(95% CI)
NPA, %
(95% CI)
OPA, %
(95% CI)
87/87 41.3
30.1–53.3
66.7
34.9–90.1
44.8
34.1–55.9

3.3. Lower Quantitative Agreement of the Elecsys Anti-SARS-CoV-2 Antigen Assay with RT-PCR Diagnosis in Saliva Than in NPS

We next evaluated the quantitative agreement of the antigen assay and RT-PCR for both NPS and saliva specimens. Antigen measurements (COI values) and RT-PCR (E-gene Ct values) correlated generally well in both specimen types (Figure 1A); however, saliva yielded a lower correlation coefficient compared to NPS (Figure 1B).

Figure 1.

Figure 1

Analysis of correlation between Roche Elecsys Anti-SARS-CoV-2 Antigen assay and cobas SARS-CoV-2 IVD test for NPS and saliva samples and linear regression. Spearman correlation between Elecsys Anti-SARS-CoV-2 Antigen assay COI values (X axis, log scale, cutoff indicated by dotted line) and cobas SARS-CoV-2 IVD test Ct values (Y axis) was analyzed for all NPS (N = 187 *, p < 0.0001) (A) and paired NPS (orange circles, N = 87 *, p < 0.0001) and saliva samples (blue triangles, N = 87, p < 0.0001) (B). * Including one sample for which the ORF1-gene Ct value was used. Linear regression analysis of only positive antigen tests (COI values ≥ 1) was performed for NPS (left plot) and saliva (right plot); black solid line shows the regression line; dotted black lines show the 95% confidence bands (C).

Linear regression of only positive antigen tests (COI values ≥ 1) yielded different E-gene Ct cutoffs in NPS vs. saliva (Figure 1C). This shift of 1.45 Ct could potentially indicate a matrix effect of saliva that affects detection in the antigen assay at equal sensitivity as in NPS.

We previously reported that Ct values were on average 4.87 higher in saliva than in the corresponding NPS, which corresponds to, on average, 30-fold lower viral loads in saliva [21]. This was also evident in the paired NPS-saliva cohort analyzed in the present study, where we observed, on average, 4.6 higher Ct values in saliva (data not shown).

3.4. Low Capacity of Saliva Antigen Testing in Clinical Diagnosis

The current gold standard for SARS-CoV-2 diagnosis remains RT-PCR detection in NPS, and the sensitivity of other test formats, including antigen tests, are evaluated in reference to it. For SARS-CoV-2 antigen detecting rapid diagnostic tests (Ag-RDTs) of respiratory specimens, the WHO defines as the minimum performance requirements a sensitivity of ≥80% and a specificity of ≥97% compared to a nucleic acid amplification test (NAAT) reference assay (https://www.who.int/publications/i/item/antigen-detection-in-the-diagnosis-of-sars-cov-2infection-using-rapid-immunoassays, accessed on 12 May 2021). Considering that saliva contains less virus than NPS, the performance of saliva antigen testing needs to be put into relation to corresponding levels of the virus in NPS to yield relevant information.

We therefore assessed the diagnostic agreement between NPS RT-PCR as a reference and the paired saliva antigen assay results. The analysis highlighted the markedly reduced clinical performance of saliva antigen testing. PPA overall was insufficient (40.2%) (Table 5). Even for samples with Ct values (of the corresponding NPS) ≤ 26, PPA was only 47.4%. In line with these data, analysis of correlation between COI values of saliva and Ct values of the corresponding NPS yielded a very low correlation coefficient (Figure 2). Thus, saliva testing with the Elecsys Anti-SARS-CoV-2 Antigen assay has overall a low capacity in clinical diagnosis.

Table 5.

PPA of the Roche Elecsys Anti-SARS-CoV-2 Antigen assay with the Roche cobas SARS-CoV-2 IVD test in saliva based on Ct values of corresponding NPS.

Cobas E-Gene
Ct Value a
Saliva
N
(cum.)
PPA, %
(95% CI)
≤26 57 47.4
34.0–61.0
≤27 60 45.0
32.1–58.4
≤28 65 44.6
32.2–57.5
≤29 70 44.3
32.4–56. 7
≤30 73 42.5
31.0–54.6
≤31 75 42.7
31.3–54.6
≤32 75 42.7
31.3–54.6
≤33 75 42.7
31.3–54.6
≤34 76 42.1
30.9–54.0
≤35 79 40.5
29.6–52.1
Overall
(≤40)
87 40.2
29.9–51.3

a Ct values of corresponding NPS.

Figure 2.

Figure 2

Analysis of correlation between Roche Elecsys Anti-SARS-CoV-2 Antigen assay and cobas SARS-CoV-2 IVD test for saliva samples based on Ct values of corresponding NPS. Spearman correlation between Elecsys Anti-SARS-CoV-2 Antigen assay COI values (X axis, log scale, cutoff indicated by dotted line) and cobas SARS-CoV-2 IVD test Ct values of the corresponding NPS (Y axis) was analyzed for saliva samples (N = 87, p = 0.0002).

The study was designed to assess the analytical performance of saliva antigen testing. While an assessment of the performance of the antigen test in saliva at different disease stages would require a dedicated study protocol and enrollment, we had basic data on symptom status available for our cohort (Table 1) that allowed a first analysis (Figure 3). Comparing diagnostic performance of RT-PCR and the antigen test in paired NPS and saliva samples that tested positive in symptomatics (N = 77) and asymptomatics (N = 6), we observed the expected lower sensitivity of the antigen test in samples with lower viral load (higher Ct value) irrespective of disease pattern and duration (Figure 3).

Figure 3.

Figure 3

SARS-CoV-2 detection in asymptomatic and symptomatic disease stages. (A,B) Analysis of paired NPS (A) and saliva (B) samples from asymptomatic (N = 6) and symptomatic (N = 77) individuals that tested positive for SARS-CoV-2 by RT-PCR. Roche Elecsys Anti-SARS-CoV-2 Antigen assay results (COI value) and corresponding cobas SARS-CoV-2 IVD test results (Ct value) are depicted. Cutoff of the antigen test is indicated by a dotted line. Green triangles: asymptomatic, violet squares: symptomatic. Comparison of duration of symptoms (days) and outcome of cobas SARS-CoV-2 IVD test (C) or Roche Elecsys Anti-SARS-CoV-2 Antigen assay (D) in paired symptomatic SARS-CoV-2 positive NPS (orange circles, N = 65) and saliva (blue triangles, N = 65) samples. For 12 of the 77 symptomatic patients, no information on duration of symptoms was available.

4. Discussion

In the present study, we evaluated the suitability of saliva for detection of SARS-CoV-2 with the Elecsys Anti-SARS-CoV-2 Antigen assay. The combination of a specimen that can be easily and safely obtained with an accurate and rapid detection method would expand the range of strategies for diagnosing early SARS-CoV-2 infection. NPS is considered the optimal material for detection of SARS-CoV-2 in early infection [1]. However, its collection requires trained personnel, which are exposed to a potential risk of nosocomial transmission and may create discomfort for the patient. Saliva circumvents these problems, but its sensitivity and thus suitability has been differentially reported [2,3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41]. We recently conducted a large-scale head-to-head comparison of SARS-CoV-2 detection by RT-PCR in NPS and saliva and could show that saliva is a valid alternate specimen for SARS-CoV-2 detection by RT-PCR with particular advantages for testing children [21].

The majority of the currently used antigen tests are Ag-RDTs of respiratory specimens to be used at the Point of Care (POC). A recent Cochrane study reported a sensitivity of only 56.2% (95% confidence interval (CI): 29.5–79.8%) for point-of-care antigen tests [33]. Several studies have evaluated saliva as an alternative to NPS for detection of SARS-CoV-2 with antigen tests, including automated tests [11,12,13,14,15,20,36,37,38,40,42,43,44]. However, none of the current Emergency Use Authorization (EUA) tests for SARS-CoV-2 antigen detection is authorized for saliva.

The Elecsys Anti-SARS-CoV-2 Antigen assay allows for a rapid detection of the SARS-CoV-2 nucleocapsid protein in NPS and oropharyngeal swab samples with a good sensitivity for symptomatic patients. Our study confirms the high agreement between the Elecsys Anti-SARS-CoV-2 Antigen assay and the cobas SARS-CoV-2 IVD test for NPS samples as reported by the manufacturer. Notably, in our study cohort, the PPA (86.2% for all NPS samples, N = 187) was even higher than the relative sensitivity reported by the manufacturer (PPA of 60.5%, N = 390). This was also evident from the observed thresholds of detection, which were for a PPA of 100% at Ct ≤ 28 in our cohort compared to Ct < 26 reported by the manufacturer.

In contrast to the NPS samples, the Elecsys Anti-SARS-CoV-2 Antigen assay performed much less well with saliva. For the saliva samples, the PPA (with RT-PCR in saliva) was overall only 40.2%, and SARS-CoV-2 could be detected reliably only in samples with Ct values ≤ 26. It is important to note that saliva collection was not tailored for the use with this specific assay but was based on a method that we optimized for detection of SARS-CoV-2 by RT-PCR and entails a dilution of saliva upon collection. The ability to use the same material for different test systems is preferable in diagnostic laboratories, because it provides flexibility to choose different test systems for reanalysis.

Our study shows that lower sensitivity can occur in antigen tests using saliva as a medium and highlights the need for developers to establish appropriate protocols for saliva testing. It must be considered that a reduced sensitivity of SARS-CoV-2 antigen testing in saliva is likely not restricted to the assay evaluated here, as several studies noted lower sensitivity in saliva [11,12,13,14,15,20,22,36,37,38,39,40,41,42,43,44,45]. Alternate saliva sampling strategies that allow concentrated saliva processing may thus be of advantage in combination with antigen tests to overcome the sensitivity restrictions. Nevertheless, processing undiluted saliva is challenging due to its viscosity and may not be applicable for automated processing. Optimization of saliva pretreatment will be most effective if tailored by the manufacturers to the respective assay system to avoid interference with detection reagents.

Our results show a clear advantage of NPS as a specimen in combination with the probed antigen test. This is in concordance with other antigen test systems which, in general, have shown lower sensitivity compared to NPS [35,42,43,44,45,46,47,48,49,50,51,52,53,54,55,56]. One of the automated antigen tests previously evaluated for saliva samples is the Lumipulse G SARS-CoV-2 Ag assay, a chemiluminescent enzyme immunoassay which received the CE marking for qualitative and quantitative detection of the SARS-CoV-2 N antigen on both saliva and NPS samples. Sensitivities relative to the corresponding molecular reference test ranged from 41.3% to 88.9%; specificities were reported in four of the five studies and ranged from 96.9% to 98.6% [42,43,44,46,51]. It has to be noted, however, that in the study reporting the highest sensitivity (i.e., 88.9%), only nine RT-PCR-positive saliva samples were evaluated. Another automated antigen test evaluated for saliva is the Simoa SARS-CoV-2 N Protein Antigen Test, which is an EUA for the qualitative detection of the N antigen from SARS-CoV-2 in NPS specimens. PPA and NPA of this assay with molecular testing were 92.3% and 98.1% (N = 26) for days 1–7 [53]. Thus, compared to these two SARS-CoV-2 antigen assays, the Elecsys Anti-SARS-CoV-2 Antigen assay has a lower sensitivity but higher specificity. The sensitivity comparison must, however, be treated with caution, as sensitivity will vary depending on the virus load (Ct) of the included test samples.

The current study was designed as proof of principle analysis to define the capacity of the Elecsys antigen tests in detecting SARS-CoV-2 in saliva. We based our analysis on comparing efficacy of the test in NPS and saliva with a known viral load. We ascertained a representative range of viral loads in the test cohort samples. The study was not designed to assess the performance of the antigen test in saliva of specific patient groups (children versus adults) or disease stages (asymptomatic versus symptomatic). While extended analyses on the use of saliva antigen testing in different disease stages would have been of interest, the low sensitivity we observed in saliva rendered these study extensions unfortunately unnecessary.

The lower sensitivity in saliva with the antigen test probed here highlights the general importance of carefully evaluating saliva as a test material. As we showed, conducting head-to-head comparisons of NPS and saliva for both PCR and antigen test is critical to dissect if matrix effects in saliva exist. As only few studies have assessed saliva in the context of SARS-CoV-2 antigen tests, and those that have, recorded variable results [35,41,42,43,44,45,46,47,48,49,50,51,52,53,54,55,56], it can currently not be excluded that antigen tests generally show a reduced sensitivity across diverse tests systems. Alternatively, these effects may be test-dependent and can be modest enough to allow application in mass testing as a recent study suggests [41].

In sum, there is a general agreement based on published studies and the data presented here that SARS-CoV-2 antigen tests in saliva are less sensitive and, thus, should only be considered when a clear benefit for clinical diagnosis can be established.

Acknowledgments

We thank staff of the participating test centers for coordinating the sample collection, the staff of the Institute of Medical Virology diagnostics unit, sample triage, and administration for their support.

Supplementary Materials

The following are available online at https://www.mdpi.com/article/10.3390/microorganisms9081700/s1, Table S1: Raw data and donor demographics.

Author Contributions

Conceptualization, M.H., J.B. and A.T.; Data curation, A.A. and M.H.; Formal analysis, A.A. and M.H.; Funding acquisition, A.T.; Investigation, A.A.; Resources, P.W.S., T.S., R.B., A.R., D.L.B., G.E., D.I.K., B.H., C.B. and H.F.G.; Supervision, A.T.; Visualization, A.A., M.H. and A.M.; Writing—original draft, A.A., M.H. and A.T.; Writing—review and editing, A.A., J.B., P.W.S., T.S., R.B., A.R., D.L.B., G.E., D.I.K., B.H., C.B., H.F.G., A.M., A.T. and M.H. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the Swiss federal office of public health (FOPH) and the University of Zurich Foundation to A.T. Roche Diagnostics supported a prior study with PCR kits and consumables for the paired NPS and saliva analyses, data of which are shown in this study. Roche Diagnostics supported the present study with Elecsys SARS-CoV-2 Antigen test kits. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, or in the decision to publish the results.

Institutional Review Board Statement

Ethical review and approval were waived for this study by the Zurich Cantonal Ethics Commission as specimen were leftover from routine diagnostics and a previous study [21] approved with an ethics waiver by the Zurich Cantonal Ethics Commission (Req-2020-00398).

Informed Consent Statement

No patient consent was required as ethical review and approval were waived for this study by the Zurich Cantonal Ethics Commission (Req-2020-00398).

Data Availability Statement

All raw data are provided as Supplementary Materials.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.WHO COVID-19 Target Product Profiles for Priority Diagnostics to Support Response to the COVID-19 Pandemic v.1.0. WHO; Geneva, Switzerland: 2020. pp. 1–38. [Google Scholar]
  • 2.Berenger B.M., Conly J.M., Fonseca K., Hu J., Louie T., Schneider A.R., Singh T., Stokes W., Ward L., Zelyas N. Saliva collected in universal transport media is an effective, simple and high-volume amenable method to detect SARS-CoV-2. Clin. Microbiol. Infect. 2021;27:656–657. doi: 10.1016/j.cmi.2020.10.035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Chen J.H., Yip C.C., Poon R.W., Chan K.H., Cheng V.C., Hung I.F., Chan J.F., Yuen K.Y., To K.K. Evaluating the use of posterior oropharyngeal saliva in a point-of-care assay for the detection of SARS-CoV-2. Emerg. Microbes. Infect. 2020;9:1356–1359. doi: 10.1080/22221751.2020.1775133. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Iwasaki S., Fujisawa S., Nakakubo S., Kamada K., Yamashita Y., Fukumoto T., Sato K., Oguri S., Taki K., Senjo H., et al. Comparison of SARS-CoV-2 detection in nasopharyngeal swab and saliva. J. Infect. 2020;81:e145–e147. doi: 10.1016/j.jinf.2020.05.071. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Leung E.C., Chow V.C., Lee M.K., Lai R.W. Deep throat saliva as an alternative diagnostic specimen type for the detection of SARS-CoV-2. J. Med. Virol. 2021;93:533–536. doi: 10.1002/jmv.26258. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.McCormick-Baw C., Morgan K., Gaffney D., Cazares Y., Jaworski K., Byrd A., Molberg K., Cavuoti D. Saliva as an Alternate Specimen Source for Detection of SARS-CoV-2 in Symptomatic Patients Using Cepheid Xpert Xpress SARS-CoV-2. J. Clin. Microbiol. 2020;58:e01109-20. doi: 10.1128/JCM.01109-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Migueres M., Mengelle C., Dimeglio C., Didier A., Alvarez M., Delobel P., Mansuy J.M., Izopet J. Saliva sampling for diagnosing SARS-CoV-2 infections in symptomatic patients and asymptomatic carriers. J. Clin. Virol. 2020;130:104580. doi: 10.1016/j.jcv.2020.104580. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Pasomsub E., Watcharananan S.P., Boonyawat K., Janchompoo P., Wongtabtim G., Suksuwan W., Sungkanuparph S., Phuphuakrat A. Saliva sample as a non-invasive specimen for the diagnosis of coronavirus disease 2019: A cross-sectional study. Clin. Microbiol. Infect. 2021;27:285.e1–285.e4. doi: 10.1016/j.cmi.2020.05.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Procop G.W., Shrestha N.K., Vogel S., Van Sickle K., Harrington S., Rhoads D.D., Rubin B.P., Terpeluk P. A Direct Comparison of Enhanced Saliva to Nasopharyngeal Swab for the Detection of SARS-CoV-2 in Symptomatic Patients. J. Clin. Microbiol. 2020;58:e01946-20. doi: 10.1128/JCM.01946-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Takeuchi Y., Furuchi M., Kamimoto A., Honda K., Matsumura H., Kobayashi R. Saliva-based PCR tests for SARS-CoV-2 detection. J. Oral Sci. 2020;62:350–351. doi: 10.2334/josnusd.20-0267. [DOI] [PubMed] [Google Scholar]
  • 11.Uwamino Y., Nagata M., Aoki W., Fujimori Y., Nakagawa T., Yokota H., Sakai-Tagawa Y., Iwatsuki-Horimoto K., Shiraki T., Uchida S., et al. Accuracy and stability of saliva as a sample for reverse transcription PCR detection of SARS-CoV-2. J. Clin. Pathol. 2021;74:67–68. doi: 10.1136/jclinpath-2020-206972. [DOI] [PubMed] [Google Scholar]
  • 12.Villar L.M., da Costa V.D., Marques B.C.L., da Silva L.L., Santos A.C., Mendonca A., Marques V.A., do Nascimento G.P., Lewis-Ximenez L.L., de Paula V.S. Usefulness of Saliva Samples for Detecting SARS-CoV-2 RNA among liver disease patients. J. Infect. 2021;82:e4–e5. doi: 10.1016/j.jinf.2020.07.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Williams E., Bond K., Zhang B., Putland M., Williamson D.A. Saliva as a Noninvasive Specimen for Detection of SARS-CoV-2. J. Clin. Microbiol. 2020;58:e00776-20. doi: 10.1128/JCM.00776-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Yee R., Truong T.T., Pannaraj P.S., Eubanks N., Gai E., Jumarang J., Turner L., Peralta A., Lee Y., Dien Bard J. Saliva Is a Promising Alternative Specimen for the Detection of SARS-CoV-2 in Children and Adults. J. Clin. Microbiol. 2021;59:e02686-20. doi: 10.1128/JCM.02686-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Yokota I., Shane P.Y., Okada K., Unoki Y., Yang Y., Inao T., Sakamaki K., Iwasaki S., Hayasaka K., Sugita J., et al. Mass screening of asymptomatic persons for SARS-CoV-2 using saliva. Clin. Infect. Dis. 2020;73:e559–e565. doi: 10.1093/cid/ciaa1388. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Zhu J., Guo J., Xu Y., Chen X. Viral dynamics of SARS-CoV-2 in saliva from infected patients. J. Infect. 2020;81:e48–e50. doi: 10.1016/j.jinf.2020.06.059. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Jamal A.J., Mozafarihashjin M., Coomes E., Powis J., Li A.X., Paterson A., Anceva-Sami S., Barati S., Crowl G., Faheem A., et al. Sensitivity of Nasopharyngeal Swabs and Saliva for the Detection of Severe Acute Respiratory Syndrome Coronavirus 2. Clin. Infect. Dis. 2021;72:1064–1066. doi: 10.1093/cid/ciaa848. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Landry M.L., Criscuolo J., Peaper D.R. Challenges in use of saliva for detection of SARS CoV-2 RNA in symptomatic outpatients. J. Clin. Virol. 2020;130:104567. doi: 10.1016/j.jcv.2020.104567. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Rao M., Rashid F.A., Sabri F., Jamil N.N., Zain R., Hashim R., Amran F., Kok H.T., Samad M.A.A., Ahmad N. Comparing Nasopharyngeal Swab and Early Morning Saliva for the Identification of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Clin. Infect. Dis. 2021;72:e352–e356. doi: 10.1093/cid/ciaa1156. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Wyllie A.L., Fournier J., Casanovas-Massana A., Campbell M., Tokuyama M., Vijayakumar P., Warren J.L., Geng B., Muenker M.C., Moore A.J., et al. Saliva or Nasopharyngeal Swab Specimens for Detection of SARS-CoV-2. N. Engl. J. Med. 2020;383:1283–1286. doi: 10.1056/NEJMc2016359. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Huber M., Schreiber P.W., Scheier T., Audigé A., Buonomano R., Rudiger A., Braun D.L., Eich G., Keller D.I., Hasse B., et al. High Efficacy of Saliva in Detecting SARS-CoV-2 by RT-PCR in Adults and Children. Microorganisms. 2021;9:642. doi: 10.3390/microorganisms9030642. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Azzi L., Baj A., Alberio T., Lualdi M., Veronesi G., Carcano G., Ageno W., Gambarini C., Maffioli L., Saverio S.D., et al. Rapid Salivary Test suitable for a mass screening program to detect SARS-CoV-2: A diagnostic accuracy study. J. Infect. 2020;81:e75–e78. doi: 10.1016/j.jinf.2020.06.042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Azzi L., Carcano G., Gianfagna F., Grossi P., Gasperina D.D., Genoni A., Fasano M., Sessa F., Tettamanti L., Carinci F., et al. Saliva is a reliable tool to detect SARS-CoV-2. J. Infect. 2020;81:e45–e50. doi: 10.1016/j.jinf.2020.04.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Fakheran O., Dehghannejad M., Khademi A. Saliva as a diagnostic specimen for detection of SARS-CoV-2 in suspected patients: A scoping review. Infect. Dis. Poverty. 2020;9:100. doi: 10.1186/s40249-020-00728-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Han P., Ivanovski S. Saliva-Friend and Foe in the COVID-19 Outbreak. Diagnostics. 2020;10:290. doi: 10.3390/diagnostics10050290. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Harikrishnan P. Saliva as a Potential Diagnostic Specimen for COVID-19 Testing. J. Craniofac. Surg. 2020;31:e653–e655. doi: 10.1097/SCS.0000000000006724. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Iwata K., Yoshimura K. A concern regarding estimated sensitivities and specificities of nasopharyngeal and saliva specimens for SARS-CoV-2 infection. Clin. Infect. Dis. 2020 doi: 10.1093/cid/ciaa1655. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Lai C.K.C., Chen Z., Lui G., Ling L., Li T., Wong M.C.S., Ng R.W.Y., Tso E.Y.K., Ho T., Fung K.S.C., et al. Prospective Study Comparing Deep Throat Saliva with Other Respiratory Tract Specimens in the Diagnosis of Novel Coronavirus Disease 2019. J. Infect. Dis. 2020;222:1612–1619. doi: 10.1093/infdis/jiaa487. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Lopez-Martinez B., Guzman-Ortiz A.L., Nevarez-Ramirez A.J., Parra-Ortega I., Olivar-Lopez V.B., Angeles-Floriano T., Vilchis-Ordonez A., Quezada H. Saliva as a promising biofluid for SARS-CoV-2 detection during the early stages of infection. Bol. Med. Hosp. Infant. Mex. 2020;77:228–233. doi: 10.24875/BMHIM.20000204. [DOI] [PubMed] [Google Scholar]
  • 30.Mao M.H., Guo J.J., Qin L.Z., Han Z.X., Wang Y.J., Yang D. Serial semiquantitative detection of SARS-CoV-2 in saliva samples. J. Infect. 2021;82:414–451. doi: 10.1016/j.jinf.2020.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Moreno-Contreras J., Espinoza M.A., Sandoval-Jaime C., Cantu-Cuevas M.A., Baron-Olivares H., Ortiz-Orozco O.D., Munoz-Rangel A.V., Hernandez-de la Cruz M., Eroza-Osorio C.M., Arias C.F., et al. Saliva Sampling and Its Direct Lysis, an Excellent Option to Increase the Number of SARS-CoV-2 Diagnostic Tests in Settings with Supply Shortages. J. Clin. Microbiol. 2020;58:e01659-20. doi: 10.1128/JCM.01659-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Ott I.M., Strine M.S., Watkins A.E., Boot M., Kalinich C.C., Harden C.A., Vogels C.B.F., Casanovas-Massana A., Moore A.J., Muenker M.C., et al. Simply saliva: Stability of SARS-CoV-2 detection negates the need for expensive collection devices. medRxiv. 2020 doi: 10.1101/2020.08.03.20165233. [DOI] [Google Scholar]
  • 33.To K.K., Tsang O.T., Leung W.S., Tam A.R., Wu T.C., Lung D.C., Yip C.C., Cai J.P., Chan J.M., Chik T.S., et al. Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: An observational cohort study. Lancet Infect. Dis. 2020;20:565–574. doi: 10.1016/S1473-3099(20)30196-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.To K.K., Tsang O.T., Yip C.C., Chan K.H., Wu T.C., Chan J.M., Leung W.S., Chik T.S., Choi C.Y., Kandamby D.H., et al. Consistent Detection of 2019 Novel Coronavirus in Saliva. Clin. Infect. Dis. 2020;71:841–843. doi: 10.1093/cid/ciaa149. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Uwamino Y., Nagata M., Aoki W., Nakagawa T., Inose R., Yokota H., Furusawa Y., Sakai-Tagawa Y., Iwatsuki-Horimoto K., Kawaoka Y., et al. Accuracy of rapid antigen detection test for nasopharyngeal swab specimens and saliva samples in comparison with RT-PCR and viral culture for SARS-CoV-2 detection. J. Infect. Chemother. 2021;27:1058–1062. doi: 10.1016/j.jiac.2021.04.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Valentine-Graves M., Hall E., Guest J., Adam E., Valencia R., Hardee I., Shinn K., Sanchez T., Siegler A.J., Sullivan P. At-home self-collection of saliva, oropharyngeal swabs and dried blood spots for SARS-CoV-2 diagnosis and serology: Post-collection acceptability of specimen collection process and patient confidence in specimens. medRxiv. :2020. doi: 10.1371/journal.pone.0236775. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Vaz S.N., Santana D.S., Netto E.M., Pedroso C., Wang W.K., Santos F.D.A., Brites C. Saliva is a reliable, non-invasive specimen for SARS-CoV-2 detection. Braz. J. Infect. Dis. 2020;24:422–427. doi: 10.1016/j.bjid.2020.08.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Wong R.C., Wong A.H., Ho Y.I., Leung E.C., Lai R.W. Evaluation on testing of deep throat saliva and lower respiratory tract specimens with Xpert Xpress SARS-CoV-2 assay. J. Clin. Virol. 2020;131:104593. doi: 10.1016/j.jcv.2020.104593. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Wong S.C.Y., Tse H., Siu H.K., Kwong T.S., Chu M.Y., Yau F.Y.S., Cheung I.Y.Y., Tse C.W.S., Poon K.C., Cheung K.C., et al. Posterior Oropharyngeal Saliva for the Detection of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Clin. Infect. Dis. 2020;71:2939–2946. doi: 10.1093/cid/ciaa797. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Yoon J.G., Yoon J., Song J.Y., Yoon S.Y., Lim C.S., Seong H., Noh J.Y., Cheong H.J., Kim W.J. Clinical Significance of a High SARS-CoV-2 Viral Load in the Saliva. J. Korean Med. Sci. 2020;35:e195. doi: 10.3346/jkms.2020.35.e195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Yokota I., Shane P.Y., Okada K., Unoki Y., Yang Y., Iwasaki S., Fujisawa S., Nishida M., Teshima T. A novel strategy for SARS-CoV-2 mass screening with quantitative antigen testing of saliva: A diagnostic accuracy study. Lancet Microbe. 2021;2:e397–e404. doi: 10.1016/S2666-5247(21)00092-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Amendola A., Sberna G., Lalle E., Colavita F., Castilletti C., Menchinelli G., Posteraro B., Sanguinetti M., Ippolito G., Bordi L., et al. Saliva Is a Valid Alternative to Nasopharyngeal Swab in Chemiluminescence-Based Assay for Detection of SARS-CoV-2 Antigen. J. Clin. Med. 2021;10:1471. doi: 10.3390/jcm10071471. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Asai N., Sakanashi D., Ohashi W., Nakamura A., Kawamoto Y., Miyazaki N., Ohno T., Yamada A., Chida S., Shibata Y., et al. Efficacy and validity of automated quantitative chemiluminescent enzyme immunoassay for SARS-CoV-2 antigen test from saliva specimen in the diagnosis of COVID-19. J. Infect. Chemother. 2021;27:1039–1042. doi: 10.1016/j.jiac.2021.03.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Basso D., Aita A., Padoan A., Cosma C., Navaglia F., Moz S., Contran N., Zambon C.F., Maria Cattelan A., Plebani M. Salivary SARS-CoV-2 antigen rapid detection: A prospective cohort study. Clin. Chim. Acta. 2021;517:54–59. doi: 10.1016/j.cca.2021.02.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Agullo V., Fernandez-Gonzalez M., Ortiz de la Tabla V., Gonzalo-Jimenez N., Garcia J.A., Masia M., Gutierrez F. Evaluation of the rapid antigen test Panbio COVID-19 in saliva and nasal swabs in a population-based point-of-care study. J. Infect. 2021;82:186–230. doi: 10.1016/j.jinf.2020.12.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Ishii T., Sasaki M., Yamada K., Kato D., Osuka H., Aoki K., Morita T., Ishii Y., Tateda K. Immunochromatography and chemiluminescent enzyme immunoassay for COVID-19 diagnosis. J. Infect. Chemother. 2021;27:915–918. doi: 10.1016/j.jiac.2021.02.025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Kashiwagi K., Ishii Y., Aoki K., Yagi S., Maeda T., Miyazaki T., Yoshizawa S., Aoyagi K., Tateda K. Immunochromatographic test for the detection of SARS-CoV-2 in saliva. J. Infect. Chemother. 2021;27:384–386. doi: 10.1016/j.jiac.2020.11.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Lee L., Liu F., Chen Y., Roma G. Quantitative and Ultrasensitive In-situ Immunoassay Technology for SARS-CoV-2 Detection in Saliva. Res. Sq. 2021 doi: 10.21203/rs.3.rs-138025/v1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Masia M., Fernandez-Gonzalez M., Sanchez M., Carvajal M., Garcia J.A., Gonzalo-Jimenez N., Ortiz de la Tabla V., Agullo V., Candela I., Guijarro J., et al. Nasopharyngeal Panbio COVID-19 Antigen Performed at Point-of-Care Has a High Sensitivity in Symptomatic and Asymptomatic Patients With Higher Risk for Transmission and Older Age. Open Forum Infect. Dis. 2021;8:ofab059. doi: 10.1093/ofid/ofab059. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Nagura-Ikeda M., Imai K., Tabata S., Miyoshi K., Murahara N., Mizuno T., Horiuchi M., Kato K., Imoto Y., Iwata M., et al. Clinical Evaluation of Self-Collected Saliva by Quantitative Reverse Transcription-PCR (RT-qPCR), Direct RT-qPCR, Reverse Transcription-Loop-Mediated Isothermal Amplification, and a Rapid Antigen Test to Diagnose COVID-19. J. Clin. Microbiol. 2020;58:e01438-20. doi: 10.1128/JCM.01438-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Norizuki M., Hachiya M., Motohashi A., Moriya A., Mezaki K., Kimura M., Sugiura W., Akashi H., Umeda T. Effective screening strategies for detection of asymptomatic COVID-19 travelers at airport quarantine stations: Exploratory findings in Japan. Glob. Health Med. 2021;3:107–111. doi: 10.35772/ghm.2020.01109. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Pinals R.L., Ledesma F., Yang D., Navarro N., Jeong S., Pak J.E., Kuo L., Chuang Y.C., Cheng Y.W., Sun H.Y., et al. Rapid SARS-CoV-2 Spike Protein Detection by Carbon Nanotube-Based Near-Infrared Nanosensors. Nano Lett. 2021;21:2272–2280. doi: 10.1021/acs.nanolett.1c00118. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Shan D., Johnson J.M., Fernandes S.C., Suib H., Hwang S., Wuelfing D., Mendes M., Holdridge M., Burke E.M., Beauregard K., et al. N-protein presents early in blood, dried blood and saliva during asymptomatic and symptomatic SARS-CoV-2 infection. Nat. Commun. 2021;12:1931. doi: 10.1038/s41467-021-22072-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Singh N.K., Ray P., Carlin A.F., Magallanes C., Morgan S.C., Laurent L.C., Aronoff-Spencer E.S., Hall D.A. Hitting the diagnostic sweet spot: Point-of-care SARS-CoV-2 salivary antigen testing with an off-the-shelf glucometer. Biosens. Bioelectron. 2021;180:113111. doi: 10.1016/j.bios.2021.113111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Stokes W., Berenger B.M., Portnoy D., Scott B., Szelewicki J., Singh T., Venner A.A., Turnbull L., Pabbaraju K., Shokoples S., et al. Clinical performance of the Abbott Panbio with nasopharyngeal, throat, and saliva swabs among symptomatic individuals with COVID-19. Eur. J. Clin. Microbiol. Infect. Dis. 2021;40:1721–1726. doi: 10.1007/s10096-021-04202-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Torrente-Rodriguez R.M., Lukas H., Tu J., Min J., Yang Y., Xu C., Rossiter H.B., Gao W. SARS-CoV-2 RapidPlex: A Graphene-Based Multiplexed Telemedicine Platform for Rapid and Low-Cost COVID-19 Diagnosis and Monitoring. Matter. 2020;3:1981–1998. doi: 10.1016/j.matt.2020.09.027. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

All raw data are provided as Supplementary Materials.


Articles from Microorganisms are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES