Vitamin D Standardization Program (VDSP) Intralaboratory Study for the Assessment of 25-Hydroxyvitamin D Assay Variability and Bias

Stephen A Wise a,b,*, Johanna E Camara c, Christopher T Sempos a,d, Pierre Lukas e, Caroline Le Goff e, Stephanie Peeters e, Carolyn Q Burdette c, Federica Nalin c, Grace Hahm c, Ramón A Durazo-Arvizu f, Adam J Kuszak a, Joyce Merkel a, Étienne Cavalier e
PMCID: PMC8403635  NIHMSID: NIHMS1710726  PMID: 34010687

Abstract

An intralaboratory study assessing assay variability and bias for determination of serum total 25-hydroxyvitamin D [25(OH)D] was conducted by the Vitamin D Standardization Program (VDSP). Thirteen assays for serum total 25(OH)D were evaluated in a single laboratory including 11 unique immunoassays and one liquid chromatography – tandem mass spectrometry (LC-MS/MS) assay. Fifty single-donor serum samples, including eight samples with high concentrations of 25(OH)D2 (> 30 nmol/L), were assigned target values for 25(OH)D2 and 25(OH)D3 using reference measurement procedures (RMP). Using four replicate measurements for each sample, the mean total percent coefficient of variation (%CV) and mean % bias from the target values were determined for each assay using the 50 single-donor samples and a 42-sample subset, which excluded 8 high 25(OH)D2 concentration samples, and compared with VDSP performance criteria of ≤ 10% CV and ≤ ±5% mean bias. All 12 assays achieved the performance criterion for %CV, and 9 of the 12 assays were within ≤ ±5 % mean bias. The Fujirebio Inc. assay exhibited the lowest %CV and highest percentage of individual measurements within ≤ ±5% mean bias. Ten immunoassays exhibited changes in response due to the high 25(OH)D2 samples with Abbott, Biomérieux, DiaSorin, DIAsource, and IDS-iSYS assays having the largest deviations. The Fujirebio Inc. and Beckman Coulter assays were only minimally affected by the presence of the high 25(OH)D2 samples. Samples with high concentrations of 25(OH)D2 provided a critical performance test for immunoassays indicating that some assays may not have equal response or recovery for 25(OH)D2 and 25(OH)D3.

Keywords: 25-hydroxyvitamin D3, 25-hydroxyvitamin D2, 3-epi-25-hydroxyvitamin D3, serum total 25-hydroxyvitamin D, liquid chromatography – tandem mass spectrometry (LC-MS/MS), immunoassay, reference measurement procedure (RMP), Vitamin D Standardization Program (VDSP)

Graphical Abstract

graphic file with name nihms-1710726-f0001.jpg

1. Introduction

Since 2010, the Vitamin D Standardization Program (VDSP), established by the U.S. National Institutes of Health, Office of Dietary Supplements (NIH-ODS), has coordinated activities to assist in the standardization of measurements of serum total 25-hydroxyvitamin D [25(OH)D], the primary marker of vitamin D status. Studies have demonstrated that assay results for the determination of serum total 25(OH)D, which is defined as the sum of 25-hydroxyvitamin D2 [25(OH)D2] and 25-hydroxyvitamin D3 [25(OH)D3], may vary depending on the assay used [15]. The analytical challenges and difficulties in assessing vitamin D status have been the subject of several recent reviews[4, 614]. The VDSP is a collaborative effort among the U.S. National Institutes of Health, Office of Dietary Supplements (NIH-ODS), the U.S. National Institute of Standards and Technology (NIST) [15], the U.S. Centers for Disease Control and Prevention (CDC), national survey laboratories in several countries, and vitamin D researchers worldwide [16]. Through the VDSP, a reference measurement system has been established consisting of reference measurement procedures (RMPs) at NIST [17], Ghent University [18], and CDC [19]; NIST Standard Reference Materials (SRMs) [2022]; the CDC Vitamin D Standardization – Certification Program (VDSCP) [23]; and collaborations with two accuracy-based performance testing/external quality assessment (PT/EQA) programs, i.e., the U.S. College of American Pathologists (CAP) accuracy-based vitamin D (ABVD) program [24] and the U.K.-based Vitamin D External Quality Assessment Scheme (DEQAS) [2527]. Importantly, the VDSP also established assay performance criteria for measurement variability and bias, i.e., Coefficient of Variation (CV) ≤ 10% and mean bias ≤ ±5% [28, 29].

One of the initial activities of the VDSP was the coordination of an interlaboratory comparison study in 2011 to benchmark measurement variability and bias for immunoassays and LC-MS/MS assays used to determine 25(OH)D [3] and to evaluate the commutability of SRMs and PT/EQA program study materials [30]. As a follow up to the first VDSP intercomparison/commutability studies, a second set of intercomparison/commutability studies was undertaken, denoted as Intercomparison Study 2 and Commutability Study 2. As part of Intercomparison 2, an intralaboratory study evaluated 12 immunoassays and one LC-MS/MS assay to assess variability and bias compared with VDSP criteria. Results of this single laboratory, multi-assay comparison study are reported in this paper. Results of the VDSP multi-laboratory Intercomparison Study 2 [31, 32] and Commutability Study 2 [33] are reported elsewhere.

2. Materials and Methods

2.1. Measurands

The measurand for the intralaboratory comparison study was human serum total 25(OH)D in concentration units of nanomoles per liter (nmol/L). Serum total 25(OH)D is defined as the sum of the concentrations of 25(OH)D2 and 25(OH)D3, without the inclusion of the concentration of 3-epi-25-hydroxyvitamin D3 [3-epi-25(OH)D3].

2.2. Intralaboratory Comparison Study - Coordination and Responsibilities

The intralaboratory study was coordinated by NIST and NIH-ODS, including acquisition of the single-donor serum samples and compilation and evaluation of the results. NIST was responsible for analyzing the 50 single-donor samples to assign values for the concentrations of 25(OH)D2 and 25(OH)D3. The University of Liège (Liège, BE) analyzed the 50 single-donor samples using the 12 different assays. Samples were distributed to University of Liège in November 2016 and the results were reported to NIST in January 2017. NIH-ODS and VDSP LLC were responsible for conducting the data analyses.

2.3. Single-donor Serum Samples

The single-donor serum samples used in the single-laboratory comparison study were the same sample set used in Intercomparison Study 2 [31] and Commutability Study 2 [33]. Single-donor serum samples from 50 human donors containing only endogenous vitamin D metabolites, which were prepared according to Clinical and Laboratory Standards Institute (CLSI) C37-A guidelines [34, 35], were obtained with a distribution of total 25(OH)D concentrations across the clinically-relevant range of 15 nmol/L to 150 nmol/L. However, eight samples were included with concentrations of 25(OH)D2 > 30 nmol/L. A detailed description of the acquisition of the 50 single-donor samples is reported elsewhere [31]. The single-donor samples were stored at NIST at −80 °C until shipped to the University of Liege in November 2016 frozen on dry ice. The samples arrived frozen and were stored at −80 °C until the time of analysis.

2.4. Intralaboratory Comparison Study Design and Assays Evaluated

Analyses for the study were performed at the University of Liège during December 2016 for ten immunoassays and an LC-MS/MS assay; two additional assays were evaluated in June 2019 using the same 50 single-donor samples and protocol. The LC-MS/MS assay was based on the method reported by Fabregat-Cabello et al. [36], which has been certified by the Centers for Disease Control and Prevention (CDC) Vitamin D Standardization – Certification Program for the measurement of 25(OH)D [23]. The Diazyme assay was added as a new assay to the study. The Biomérieux assay was re-evaluated out of concern that there may have been a problem with the VIDAS instrument during the first evaluation; however, the results of the second evaluation using a new VIDAS instrument were not significantly different from the first evaluation, and therefore, both sets of results were included designated as Biomérieux I and Biomérieux II, respectively. The assay manufacturers, assay kits, and instrument models used in this study are summarized in Table 1, and additional details on the assay calibrators and reagents used are provided in Supplemental Table S1. The immunoassays included seven CLIA-based assays, and one each for ELFA-, ELISA-, EIA-, and ITA-based assays (see Table 1 for assay types).

Table 1.

Commercial Assays for Serum Total 25(OH)D used in VDSP Intralaboratory Study

Assay No. Assay Manufacturer Assay Kit/Instrument M odel Assay Typea
1 Abbott Architect 25-OH Vitamin D; Architect i1000 CLIA
2 Beckman Coulter Access 25(OH) Vitamin D Total; Access-2 CLIA
3 Biomérieux I VIDAS 25 OH Vitamin D Total; Vides ELFA
4 Biomérieux IIb VIDAS 25 OH Vitamin D Total; Vides ELFA
5 DiaSorin Liaison 25 OH Vitamin D Total; Liaison XL CLIA
6 DIAsource 25OH Vitamin D Total ELISA; Thermo Fischer Multiskan FC w incubator ELISA
7 Diazymeb EZ Vitamin D Total ITA
8 Fujirebio Inc. 25-OH Vitamin D; Lumipulse G1200 CLIA
9 IDS-EIA 25-Hydroxy Vitamin D EIA (IDS); Thermo Fischer Multiskan FC EIA
10 IDS-iSYS IDS 25 VitDS CLIA
11 Roche Vitamin D Total II; Cobas e411 CLIA
12 Siemens Vitamin D Total (VitD); ADVIA Centaur XPT CLIA
13 LC/MS/MS ABSciex Q-Trap 6500/UPLC LC-MS/MS
a

CLIA = Chemiluminescence Immunoassay

ELFA = Enzyme-Linked Fluorescence Assay

ELISA = Enzyme-Linked Immunosorbent Assay

EIA = Electrochemical immunoassay

ITA = Immunoturbidimetric assay

b

Biomérieux II and Diazyme assays were evaluated in June 2019; all other assays evaluated in December 2016

For each assay, duplicate measurements were performed on two separate days (n = 4) for each of the 50 single-donor serum samples (≈200 measurements per assay). Prior to analysis, the serum samples were removed from the −80 °C storage, thawed at room temperature, vortexed 30 s, and centrifuged at 3500 rpm for 10 min. The assays were performed using the routine laboratory operation procedures with normal internal QC criteria. Results were reported in nmol/L with three significant figures. The 11 unique immunoassays evaluated in this study included the most frequently represented immunoassays in recent DEQAS exercises [26], i.e., DiaSorin, Roche, Siemens, IDS-iSYS, and Abbott. To avoid repetition of the assay names, only the manufacturer’s name will be used since only one assay from a specific manufacturer was used.

2.5. Value Assignment of the Single-Donor Samples

The 50 single-donor serum samples were analyzed using NIST RMPs for determination of 25(OH)D2 and 25(OH)D3 [17]. The assignment of target values to the 50 single-donor samples is reported elsewhere [31] and the results are summarized in the Supplemental Table S2. The distribution of 25(OH)D concentrations in the 50 single-donor serum samples, arranged from low to high concentration, is shown in Supplemental Figure 1, with the contributions of 25(OH)D2 and 25(OH)D3 indicated.

2.6. Data Analysis

Mean bias (%) was determined using the following equation:

Mean Bias (%)=j=1j=50i=1i=4[( TestLaboratory NIST)/NIST]×100NTotal

where j = donor samples, i = sample replicates, NIST represents the assigned value, and NTotal = the total number of assays performing the measurements. The percent Coefficient of Variation (% CV) for each laboratory was calculated as the mean of the CVs for each of the 50 single-donor samples as calculated in the VDSCP. Ordinary least squares linear regression analysis was performed on both the 50- and 42-sample sets. The standardized residual plots from the linear regression analysis indicated heteroscedasticity in the regression model for the majority of the assays, particularly for the 50-sample set. We then performed a weighted least squares linear regression analysis using the 42-sample set, excluding the samples with high 25(OH)D2 content, and observed that the heteroscedasticity was reduced, but still present in 10 of the 14 assays. For the 10 assays with significant heteroscedasticity, the maximum change in slope of the regression line when using the weighted regression analysis compared with the linear regression analysis was 13% with an average of 3.9% change across the 10 assays. As a result, we concluded that although weighting was recommended, it did not result in significant changes in the regression line slopes. The samples with high concentrations of 25(OH)D2 introduce additional heteroscedasticity and when these samples are plotted with the linear regression line along with the 95% prediction intervals, they generally fall outside the prediction intervals. Therefore, we chose to use the linear (unweighted) linear regression analyses which allows the calculation of 95% prediction intervals for visual qualitative comparison of the differences in the performances of the various assays. All calculations, including CV, % bias, and linear regression with 95% prediction intervals, were performed using Stata software (College Station, TX) and some analyses were confirmed using Analyze-It, a statistical analysis add-in for Microsoft Excel (Analyze-It Software, Leeds, UK).

3. Results and Discussion

The focus of this comparison study was to assess the variability and bias in a single-laboratory setting of commonly used immunoassays and a commercial LC-MS/MS assay for the determination of serum total 25(OH)D. The study protocol of duplicate analyses on two days provided four replicate measurements for each single-donor sample for evaluation of each assay. Serum total 25(OH)D concentrations in the 50 single-donor samples ranged from 16 nmol/L to 148 nmol/L, with 25(OH)D3 ranging from 10 nmol/L to 141 nmol/L and with most 25(OH)D2 concentrations between 0.30 nmol/L to 3.0 nmol/L; however, 8 of the 50 samples had high concentrations of 25(OH)D2 concentrations of > 30 nmol/L ranging from 32 nmol/L to 137 nmol/L, which provided a test of the assays’ capabilities and response to both 25(OH)D2 and 25(OH)D3. [37]. For the assay performance evaluations described below, the assay results are reported and compared for both the 50 single-donor sample set and for the 42-sample subset after removal of the samples with 25(OH)D2 content > 30 nmol/L. The 42-sample subset may be considered as more representative set of clinical patient samples regarding the distribution of concentrations of 25(OH)D2 [38, 39].

3.1. Descriptive Statistics

Descriptive statistics (N, mean, SD, and minimum and maximum values) for each assay are summarized in Table S3 (Supplementary Material) for the measurements of serum total 25(OH)D in both the 50-sample set and in the 42-sample subset and in Table S4 for the 8 high concentration 25(OH)D2 samples (> 30 nmol/L). The mean values for serum total 25(OH)D in the 50 single-donor samples ranged from 69.0 nmol/L to 93.3 nmol/L for the different assays compared to the NIST mean value of 77.1 nmol/L as determined by using LC-MS/MS-based RMPs.

3.2. Measurement Variability

The results for the mean Coefficient of Variation (CV) as percent for each assay are summarized in Table 2 including the percent of individual sample with mean CVs less than 10% for both the 50- and 42-sample sets. A similar table with %CV and mean % bias for the 8-sample subset is provided in Table S5 (Supplemental Material). Box and Whisker plots of the %CV for the analysis of all 50 single-donor samples for each assay are shown in Figure 1A. The four immunoassays with the lowest %CVs for the 50 single-donor samples were Fujirebio Inc. (1.9 %), DiaSorin (2.6%), Abbott (2.7%), and Diazyme (2.9%) with SDs ranging from 1.0% to 1.6%. For comparison, the LC-MS/MS assay achieved a %CV of 4.5% for both the 50- and 42-sample sets with SDs of less than 1.8%. Similar Box and Whisker plots for %CV by assay for the analysis of the 42-sample subset and ordered by increasing NIST serum 25(OH)D3 are provided in Supplemental Figure 2. When comparing the Box and Whisker plots for %CV for the 50-sample set (Figure 1A) versus the 42-sample set (Supplemental Figure 2A), there are no significant differences (compare also the %CV values in Table 2 for both sample sets). There was no apparent trend observed for %CV as a function of concentration of serum total 25(OH)D in the 50 single-donor samples (Supplemental Figure 2B). All 13 assays achieved the VDSP criterion of %CV less than 10%; however, for the IDS-EIA assay, only 56% of the individual measurements (i.e., 200 total measurements) were ≤ 10% (see Table 2).

Table 2.

Mean Coefficient of Variation (%) and Mean Percent Bias for Serum Total 25(OH)D based on Four Individual Replicate Measurements and Assessment of VDSP Performance Criteria by Assay

Assay Manufacturer All 50 Donor Samples 42 Donor Samples (excluding 25(OH)D2 > 30 nmol/L)
Total CV (%) % Value ≤ 10%a VDSP Criteria Γ Total CV (%) % Values ≤ 10%a VDSP Criteria
N Mean SD Min Max N Mean SD Min Max
Abbott 50 2.71 1.4 0.5 5.9 100 Pass 42 2.61 1.3 0.5 5.9 100 Pass
Beckman Coulter 50 5.65 3.4 1.0 15.2 86 Pass 42 6.03 3.3 1.2 15.2 86 Pass
Biomérieux I 49 6.02 8.4 0.7 47.3 90 Pass 41 6.26 9.1 0.7 47.3 91 Pass
Biomérieux II 49 4.30 3.4 0.2 16.4 90 Pass 41 4.36 3.4 0.2 16.4 90 Pass
DiaSorin 50 2.65 1.2 0.9 5.6 100 Pass 42 2.67 1.2 0.9 5.6 100 Pass
DIAsource 50 6.48 3.1 1.0 18.6 90 Pass 42 6.61 3.3 1.2 18.6 88 Pass
Diazyme 50 2.89 1.6 0.0 11.4 96 Pass 42 3.25 2.6 0.3 11.4 95 Pass
Fujirebio Inc. 50 1.91 1.0 0.3 4.5 100 Pass 42 1.99 1.1 0.4 4.5 100 Pass
IDS-EIA 50 8.82 4.6 0.9 18.4 56 Pass 42 8.88 4.6 1.6 18.4 57 Pass
IDS-iSYS 50 5.50 2.6 0.6 12.5 94 Pass 42 5.71 2.7 0.6 12.5 93 Pass
Roche 50 3.43 1.9 0.7 8.2 100 Pass 42 3.54 2.0 0.7 8.2 100 Pass
Siemens 50 7.72 5.8 1.8 37.7 70 Pass 42 8.00 6.2 1.8 37.7 67 Pass
LC-MS/MS 50 4.50 1.7 0.7 10.6 98 Pass 42 4.48 1.8 0.7 10.6 98 Pass
Mean 4.81 3.1 0.7 16.3 90 4.95 3.3 0.8 16.3 90
Mean Bias (%) % Values ≤ 5% Mean Bias (%) % Values ≤ 5%
Abbott 200 − 2.89 17.8 −48.8 28.4 33 Pass 168 3.44 10.0 −17.9 28.4 39 Pass
Beckman Coulter 200 −1.98 18.7 −50.2 73.2 21 Pass 168 −3.27 19.7 −50.2 73.2 18 Pass
Biomérieux I 196 9.45 28.8 −58.8 121 12 Fail 164 18.6 24.3 −58.8 121 15 Fail
Biomérieux II 193 1.46 24.1 −44.3 59.4 16 Pass 162 8.03 19.3 −32.9 59.4 19 Fail
DiaSorin 200 −2.91 15.8 −34.0 38.2 24 Pass 168 2.04 13.5 −27.8 38.2 26 Pass
DIAsource 200 16.7 32.0 −36.9 151 15 Fail 168 22.3 30.6 −26.0 151 17 Fail
Diazyme 196 −2.78 20.7 −56.4 54.7 18 Pass 166 −2.04 22.1 −56.4 54.7 15 Pass
Fujirebio Inc. 200 −3.72 6.8 −27.8 12.2 49 Pass 168 −3.62 7.2 −27.8 12.2 46 Pass
IDS-EIA 200 −8.52 18.1 −42.5 66.9 17 Fail 168 −5.08 −17.3 −37.5 66.9 20 Fail
IDS-iSYS 200 19.5 16.8 −23.1 77.6 9 Fail 168 14.8 12.5 −23.1 42.0 11 Fail
Roche 200 −1.79 11.8 −28.2 53.8 31 Pass 168 −0.13 11.9 −28.2 53.8 34 Pass
Siemens 200 −0.47 20.1 −59.4 57.3 30 Pass 168 −3.44 19.5 −59.4 57.3 30 Pass
LC-MS/MS 199 1.78 6.4 −20.4 27.7 58 Pass 167 1.20 6.4 −20.4 27.7 60 Pass
Mean 1.83 18.3 40.8 63.2 4.06 16.5 35.9 60.4
a

Percentage of individual measurements ≤ 10% CV

Figure 1.

Figure 1.

Box and whisker plots for the percent Coefficient of Variation (%CV) (A) and mean % bias (B) by test assay for the analysis of 50 single-donor samples. The box encompasses the 25th to 75th percentile of all results, the whiskers represent 1.5 × interquartile range above the 75th percentile and 1.5 × interquartile range below the 25th percentile. The bar in the box represents the median value (CV% or mean % bias) for each assay. The dots outside the box and whiskers plots represent individual measurements outside the whisker range. For plot A, CVs greater than 35% are excluded and the red line denotes the 10% VDSP criterion for %CV. Four points are excluded with %CVs of 38% for Siemens, 41% and 47% for Biomérieux I, and 67% for Biomérieux II. For plot B, biases greater than 100% are excluded and the red solid line denotes zero bias and the red dashed lines represent ± ≤5% bias. Six points are off the scale and excluded with % biases of 141%, 148%, 149%, and 151% for DIAsource and 120% and 121% for Biomérieux I. Plot C is box and whisker plots for the mean % bias by test assay for the analysis of the eight single-donor samples with concentrations 25(OH)D2 > 30 nmol/L.

3.3. Measurement Bias

The mean % bias (compared to the NIST target values) for the results of the analysis of the 50- and 42-sample sets using all individual replicate measurement for each assay (n = ≈200) are summarized in Table 2, including the number of measurements for each assay, standard deviation of the mean % bias, minimum and maximum % bias values, and percentage of individual samples within ± 5% bias. Four of the 13 assays evaluated failed to meet the VDSP criterion of mean % bias ≤ ±5%, i.e., Biomérieux I, DIAsource, IDS-EIA, and IDS-iSYS with mean % biases of 9.4%, 16.7%, −8.5%, and 19.5%, respectively, for the 50-sample set. (Biomérieux II passed for the 50-sample set but failed for the 42-sample set.) These four assays failed the criterion even when the high 25(OH)D2 concentration samples were excluded from the evaluation; however, the mean % biases for the IDS-EIA and IDS-iSYS assays decreased (with IDS-EIA almost achieving the criterion) whereas the mean % biases for the Biomérieux I and DIAsource assays showed a higher mean % bias for the 42-sample subset compared to the 50-sample set.

The mean % bias for each assay for the 50-sample set is shown in Figure 1B as Box and Whisker plots. Similar Box and Whisker plots are provided in Supplemental Figure S3 for the 42-sample subset; comparison of Figure 1B and Supplemental Figure S3A indicates that there are only minor differences. However, when only the eight samples with high concentrations of 25(OH)D2 are included in the Box and Whisker plots (Figure 1C), significant differences are observed for the Abbott, Biomérieux (I and II), DiaSorin, DIAsource, IDS-EIA, and IDS-iSYS assays. The Fujirebio Inc. and the LC-MS/MS assays appear to have comparable behavior for the samples with high 25(OH)D2 concentrations. The mean % bias results as Box and Whisker plots for all assays for the 50 single-donor samples as a function of increasing concentration of 25(OH)D are shown in Supplemental Figure 3B, with no apparent trend observed.

The % bias results with all four replicate measurements for each of the 50 single-donor samples are shown in Figure 2 for three assays (Abbott, LC-MS/MS, and Fujirebio Inc.). The results in Figure 2A indicate that the Abbott assay has a significant negative bias (> 30%) for 7 of the 8 samples with concentrations of 25(OH)D2 > 30 nmol/L. Similar negative biases were observed for the Biomérieux, DiaSorin, and DIAsource assays (see Table 2) but not as pronounced as with the Abbott assay. The immunoassay with the largest percentage of sample measurements within ± 5% bias, Fujirebio Inc. at 49%, is shown in Figure 2B, and the assay performance does not appear to be affected by the high 25(OH)D2 concentrations. However, the Fujirebio Inc. assay appears to have significant negative biases for low concentrations of 25(OH)D (< 40 nmol/L). A similar mean % bias plot for the LC-MS/MS assay is shown in Figure 2C. Similar plots of % bias for the remaining nine assays are provided in Supplemental Figures S4 to S6. The IDS-iSYS assay has a positive bias for the high concentration 25(OH)D2 samples (Supplemental Figure S2B).

Figure 2.

Figure 2.

Comparison of % bias from NIST target values for individual replicate measurements for 50 single-donor samples using the (A) Abbott (B) IDS-iSYS, (C) Fujirebio Inc., and (D) LC-MS/MS (C) assays. Red circles represent the samples with 25(OH)D2 concentrations greater than 30 nmol/L. Red solid line represents the mean % bias for all 50 samples; black solid line represents the mean % bias for 42 samples excluding the samples with 25(OH)D2 concentrations > 30 nmol/L. Dashed blue lines represent ±5% bias from NIST target values.

The four assays that do not meet the VDSP mean % bias criterion, i.e., Biomérieux (I and II), DIAsource, IDS-iSYS, and IDS-EIA, have significantly lower percentages of samples within ≤ ±5% bias (12% and 16%, 15%, 9%, and 17%, respectively) as would be expected (see Table 2). However, six of the assays that do meet the criterion of ≤ ±5% bias still have less than 33% of the individual sample measurements within ≤ ±5% bias. Only Fujirebio Inc. (49%) and the LC-MS/MS assay (58%) approach having approximately 50% of the samples within the ≤ ±5% bias criterion. The Centers for Disease Control and Prevention (CDC) Vitamin D Standardization – Certification Program (VDSCP) uses a criterion of ≤ ±5% bias for the mean of the replicate measurements for each single-donor sample rather than using the individual measurements. The percentage of individual sample measurements of serum total 25(OH)D within various limits of bias (5% to 50%) by assay are summarized in Supplemental Table S6 including the percentage of samples within ≤ ±5% mean bias using the mean of the four replicate measurements. Using the CDC approach does not significantly alter the percentage of values with bias ≤ ± 5% with only 6 of the 13 assays slightly increasing the percentage with ≤ ±5%. Only for the LC-MS/MS assay did the percentage ≤ ±5% bias increase significantly (from 58% to 78%) by using the mean rather than the individual measurements.

3.4. Regression Analysis of Test Assays vs. NIST Target Values

Using the individual replicate measurements for 25(OH)D (n = 200) and the NIST target value for each of the 50 single-donor samples, the ordinary least squares linear regression line and 95% prediction interval were calculated for each test assay [40]. The results of the linear regression analysis (slope and R2) are summarized in Table 3 for both the 50 single-donor samples and for the 42-sample subset. In comparing the linear regression analysis for the 50 samples versus 42 samples in Table 3, only minor differences (< 5%) are observed for the slopes for the LC-MS/MS, Beckman Coulter, and Fujirebio Inc. assays indicating that the samples with high levels of 25(OH)D2 have minimal effect on the performance of these assays. The Diazyme and Roche assays have minor slope changes (<12%) while the remaining assays show a significant change in regression line slope (16% to 38%) with removal of the high 25(OH)D2 samples.

Table 3.

Simple Least Squares Linear Regression Analysis for Assays using All Individual Measurements for Each Sample (n = 200)

Assay 50 Samples (nmol/L) 42 Samples excluding High 25(OH)D2 (nmol/L) Difference 50 – 42 Sample Sets
SLS Regression 95% PI SLS Regression 95% PI
Slope int.a R2 Minb Maxb Widthc Slope Int.a R2 Minb Maxb Widthc Slope Widthc R2
Abbott 0.838 9.01 0.699 −30.1 48.1 78.2 1.157 −7.33 0.958 −22.7 8.0 30.7 −0.321 52.4 −0.260
Beckman Coulter 1.141 −11.5 0.888 −40.3 17.3 57.6 1.089 −8.58 0.834 −38.8 21.6 60.4 0.040 37.5 0.041
Biomérieux I 0.955 10.9 0.621 −41.1 63.0 104.1 1.304 −7.13 0.853 −40.6 26.3 66.9 −0.351 40.4 −0.236
Biomérieux II 0.930 6.26 0.635 −42.7 55.2 97.9 1.284 −12.4 0.876 −42.0 17.1 59.1 −0.353 42.4 −0.245
DiaSorin 0.918 4.24 0.807 −27.7 36.2 63.9 1.114 −5.63 0.899 −29.2 17.9 47.1 −0.198 17.3 −0.094
DIAsource 1.208 −2.07 0.692 −59.4 55.2 114.6 1.514 −17.3 0.826 −61.2 26.5 87.7 −0.310 28.6 −0.137
Diazyme 0.964 0.67 0.817 −31.7 33.0 64.7 1.078 −5.57 0.855 −33.3 22.1 55.4 −0.092 0.2 0.002
Fujirebio Inc. 1.025 −3.57 0.987 −12.4 5.3 17.8 1.065 −5.68 0.986 −13.6 2.2 15.8 −0.050 1.6 0.000
IDS-EIA 0.762 10.2 0.740 −21.9 42.4 64.3 0.902 3.17 0.790 −26.3 −32.6 58.9 −0.142 5.7 −0.052
IDS-iSYS 1.309 −7.66 0.888 −40.7 25.4 66.1 1.095 3.17 0.922 −17.0 23.4 40.4 0.220 28.3 −0.039
Roche 0.999 −1.47 0.929 −21.1 18.2 39.3 1.089 −6.01 0.940 −23.5 11.4 34.9 −0.091 4.1 −0.011
Siemens 1.122 −9.09 0.843 −43.5 25.3 68.8 0.944 0.70 0.826 −26.7 28.1 54.8 0.178 14.5 0.013
LC-MS/MS 1.017 0.03 0.981 −10.0 10.0 20.0 0.998 0.80 0.980 −6.7 8.5 15.2 0.018 0.9 0.002
a

Intercept

b

Minimum and maximum y-intercept for the 95% Prediction Interval

c

Width of the 95% Prediction Interval

The linear regression analysis plots for the 50- and 42-sample sets using the four replicate measurements are shown in Figures 3 and 4 for the Fujirebio Inc., DiaSorin, Abbott, and LC-MS/MS assays. These plots provide a convenient qualitative visual comparison of the differences in the performance of the various assays. Similar regression analysis plots for the remaining assays are provided in Supplemental Figures S7 to Figure S11. In addition, regression analysis plots using all individual replicate measurements are provided in Supplemental Figures S12 through S15 with the linear regression lines for the 50-sample and 42-sample sets and the identity line (y=x) on the same plot for easy comparison of the changes in slope. As shown in Figure 3 for the Fujirebio Inc. (A and B) and in Figure 4 for the LC-MS/MS (C and D) assays, there are only minor differences between the plots for 50- versus 42-sample sets. Similar minor reductions in prediction interval width are observed for the Beckman Coulter (Supplemental Figures S7A and S7B) and Diazyme (Supplemental Figure S9A and S9B) assays. For the DiaSorin (Figures 3C and 3D) and Abbott (Figures 4A and 4B) assays, however, the width of the prediction intervals decreases significantly with the removal of the high 25(OH)D2 concentration samples and the slopes increase by 21% and 38%, respectively (see Table 3).

Figure 3.

Figure 3.

Results for determination of serum total 25(OH)D in single-donor samples versus the NIST assigned target value for the Fujirebio Inc. (A and B) and DiaSorin (C and D) assays. The four replicate measurements for each of the 50 single-donor sample are represented by open black or red circles. Red circles represent the samples with 25(OH)D2 concentrations > 30 nmol/L. The shaded area is the 95% prediction interval for the regression line. Plot on the left is for all 50 single-donor samples; plot on the right is for 42 single-donor sample subset excluding the 8 samples with 25(OH)D2 > 30 nmol/L.

Figure 4.

Figure 4.

Results for determination of serum total 25(OH)D in single-donor samples versus the NIST assigned target value for the Abbott. (A and B) and the LC-MS/MS (C and D) assays. The four replicate measurements for each of the 50 single-donor sample are represented by open black or red circles. Red circles represent the samples with 25(OH)D2 concentrations > 30 nmol/L. The shaded area is the 95% prediction interval for the regression line. Plot on the left is for all 50 single-donor samples; plot on the right is for 42 single-donor sample subset excluding the 8 samples with 25(OH)D2 > 30 nmol/L.

The removal of the high 25(OH)D2 samples produces a similar significant reduction in the prediction interval width for the Biomérieux (I and II) (Supplemental Figures S7C, S7D, S8A, and S8B) and DIAsource (Supplemental Figures S8C and S8D) assays. For the remaining immunoassays, the prediction intervals width decreases slightly with the removal of the high 25(OH)D2 samples i.e., IDS-EIA (Supplemental Figures S10A and S10B), IDS-iSYS (Supplemental Figures S10C and S10D), Roche (Supplemental Figures S11A and S10B), and Siemens (Supplemental Figures 11C and 11D).

3.5. Comparison to Other Studies and VDSP Intercomparison Study 1

A number of recent studies have assessed the performance of various 25(OH)D assays with particular emphasis on recovery for 25(OH)D2 [8, 41, 42]. Freeman et al. [41] evaluated four 25(OH)D immunoassays (Siemens, DiaSorin, Roche, and Abbott) and compared them with an ID LC-MS/MS assay using samples from donors supplemented with vitamin D2 over 6-month period. They found that the Siemens, DiaSorin, Roche, and Abbott assays were negatively biased relative to the ID LC-MS/MS assay by −5.7%, −20.3%, −12.1%, and −17.8%, respectively, for supplemented donor samples having median concentrations of 25(OH)D2 of 57 nmol/L. Our results for the eight samples with 25(OH)D2 concentrations > 30 nmol/L showed similar biases for the DiaSorin, Roche, and Abbott assays (see Figure 1C and Supplemental Table S5). Bjerg et al. [8] analyzed 200 patient serum samples using seven different assays for 25(OH)D (including Siemens, Roche, DiaSorin, and IDS-iSYS) and reported that all achieved the precision requirement of the VDSP (CV ≤ 10%); however, only the IDS-iSYS and DiaSorin assays achieved an accuracy bias of ≤ ±5 % when compared with results for the analysis of SRM 972a. In our study, Abbott, DiaSorin, and Roche met the VDSP bias criterion, whereas the IDS-iSYS assay failed with a significant overestimation of 25(OH)D2 as illustrated in Figure 1C and 2B. Even though the Abbott assay met the bias criterion in our study, it had a significant underestimation (negative bias) for 25(OH)D2 as shown in Figure 1C and Figure 2A. Garnett et al. [42] evaluated the Abbott and Roche assays for their recoveries of 25(OH)D2 and 25(OH)D3 and concluded that caution should be used in interpreting results using the Abbott assay in patients supplemented with vitamin D2.

VDSP Intercomparison Study 1 was a multi-laboratory study with results from 8 immunoassays and 8 LC-MS/MS assays reported [3]. The comparison study reported in this paper was a single-laboratory study using 13 assays (11 unique immunoassays and an LC-MS/MS assay). In the first study, only 50% of the immunoassays met the criterion for CV ≤10% and only three of eight immunoassays achieved the ≤5% bias. The results from this intralaboratory study indicate that there was some improvement in the immunoassay performance with all assays evaluated within the CV ≤10% criterion and 9 of 13 immunoassays achieving the ≤ ±5% bias criterion. To assess the overall assay performance by combining both bias and precision, the mean % bias vs. the %CV for each assay in both Intercomparison Study 1 and this intralaboratory study is plotted in Supplemental Figure S15. Using this plot, it is easy to assess whether a laboratory’s performance meets the VDSP criteria of ≤ 10% for %CV and ≤ ±5% for mean % bias.

4. Conclusions

All 13 assays achieved the VDSP criterion of %CV ≤ 10% while only 9 assays satisfied the criterion of ≤ ±5% mean bias (i.e., Abbott, Beckman Coulter, Biomérieux I, DiaSorin, Diazyme, Fujirebio Inc., Roche, Siemens, and LC-MS/MS). However, 3 of the 9 assays meeting the bias criterion had only 16% to 24% of the individual measurements within ±5% bias indicating a potential need to strengthen the VDSP bias criterion. The presence of a significant number of single-donor samples with high levels of 25(OH)D2 in this study provided a critical performance test for these assays with ten immunoassays exhibiting some change in response due to the high 25(OH)D2 concentration samples with the Abbott, Biomérieux, and DIAsource assays having the largest deviations (25% to 38%). Two immunoassays (Fujirebio Inc. and Beckman Coulter) and the LC-MS/MS assay were only minimally affected by the presence of the high 25(OH)D2 concentration samples. One interpretation of these deviations with high 25(OH)D2 concentration samples would be that some assays do not have equal response or recovery for both 25(OH)D2 and 25(OH)D3. For the 12 immunoassays, the Fujirebio Inc. assay exhibited the lowest %CV and the highest percentage of individual measurements within ±5% mean bias. Expanded assay performance evaluation and comparison with 34 additional assays used in multiple laboratories Intercomparison Study 2 are detailed elsewhere [31, 32].

Supplementary Material

1

Highlights.

  • Single-laboratory study assessed variability and bias for 12 assays for total serum 25(OH)D

  • All 12 assays achieved ≤ 10% CV and 9 achieved ≤ ±5 % mean bias

  • Ten immunoassays exhibited changes in response for samples with high concentrations of 25(OH)D2

Acknowledgements

Funding for this study was provided by the NIH-ODS to the NIST. Assay kits for this study were donated by the assay manufacturers. Véronique Huberty and Bertrand Szabo of the CHR Verviers East Belgium Hospital are acknowledged for allowing use of the Siemens Centaur assay. David L. Duewer (NIST) is acknowledged for his suggestions regarding presentation of some statistical results.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Publisher's Disclaimer: Disclaimer

Publisher's Disclaimer: Certain commercial equipment or materials are identified in this paper to specify adequately the experimental procedure. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology or the National Institutes of Health, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

Declaration of Competing Financial Interests

The authors declare that they have no competing financial interests.

References

  • [1].Binkley N, Dawson-Hughes B, Durazo-Arvizu R, Thamm M, Tian L, Merkel JM, Jones JC, Carter GD, Sempos CT, Vitamin D measurement standardization: The way out of the chaos, J. Steroid Biochem. Mol. Biol, 173 (2017) 117–121. [DOI] [PubMed] [Google Scholar]
  • [2].Binkley N, Krueger D, Cowgill CS, Plum L, Lake E, Hansen KE, DeLuca HF, Drezner MK, Assay variation confounds the diagnosis of hypovitaminosis D: A call for standardization, J. Clin. Endocrinol. Metab, 89 (2004) 3152–3157. [DOI] [PubMed] [Google Scholar]
  • [3].Wise SA, Phinney KW, Tai SSC, Camara JE, Myers GL, Durazo-Arvizu R, Tian L, Hoofnagle AN, Bachmann LM, Young IS, Pettit J, Caldwell G, Liu A, Brooks SPJ, Sarafin K, Thamm M, Mensink GBM, Busch M, Rabenberg M, Cashman KD, Kiely M, Kinsella M, Galvin K, Zhang JY, Oh K, Lee SW, Jung CL, Cox L, Goldberg G, Guberg K, Prentice A, Carter GD, Jones J, Brannon PM, Lucas RM, Crump PM, Cavalier E, Merkel J, Betz JM, Sempos CT, Baseline assessment of 25-hydroxyvitamin D assay performance: A Vitamin D Standardization Program (VDSP) interlaboratory comparison study, J. AOAC Int, 100 (2017) 1244–1252. [DOI] [PubMed] [Google Scholar]
  • [4].Sempos CT, Binkley N, 25-Hydroxvitamin D assay standardization and vitamin D guidelines paralysis, Public Health Nutrition; 23 (2020) 1153–1164. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Sempos CT, Heijboer AC, Bikle DD, Bollerslev J, Bouillon R, Brannon PM, DeLuca HF, Jones G, Munns CF, Bilezikian JP, Giustina A, Binkley N, Vitamin D assays and the definition of hypovitaminosis D: results from the First International Conference on Controversies in Vitamin D, Br. J. Clin. Pharmacol, 84 (2018) 2194–2207. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Altieri B, Cavalier E, Bhattoa HP, Perez-Lopez FR, Lopez-Baena MT, Perez-Roncero GR, Chedraui P, Annweiler C, Della Casa S, Zelzer S, Herrmann M, Faggiano A, Colao A, Holick MF, Vitamin D testing: advantages and limits of the current assays, European Journal of Clinical Nutrition, 74 (2020) 231–247. [DOI] [PubMed] [Google Scholar]
  • [7].Bivona G, Lo Sasso B, Iacolino G, Gambino CM, Scazzone C, Agnello L, Ciaccio M, Standardized measurement of circulating vitamin D 25(OH)D and its putative role as a serum biomarker in Alzheimer’s disease and Parkinson’s disease, Clin. Chim. Acta, 497 (2019) 82–87. [DOI] [PubMed] [Google Scholar]
  • [8].Bjerg LN, Halgreen JR, Hansen SH, Morris HA, Jorgensen NR, An evaluation of total 25-hydroxyvitamin D assay standardization: Where are we today?, J. Steroid Biochem. Mol. Biol, 190 (2019) 224–233. [DOI] [PubMed] [Google Scholar]
  • [9].Dirks NF, Martens F, Vanderschueren D, Billen J, Pauwels S, Ackermans MT, Endert E, den Heijer M, Blankenstein MA, Heijboer AC, Determination of human reference values for serum total 1,25-dihydroxyvitamin D using an extensively validated 2D ID-UPLC-MS/MS method, J. Steroid Biochem. Mol. Biol, 164 (2016) 127–133. [DOI] [PubMed] [Google Scholar]
  • [10].Fraser WD, Tang JCY, Dutton JJ, Schoenmakers I, Vitamin D measurement, the debates continue, new analytes have emerged, developments have variable outcomes, Calcified Tissue International, 106 (2020) 3–13. [DOI] [PubMed] [Google Scholar]
  • [11].Herrmann M, Farrell CJL, Pusceddu I, Fabregat-Cabello N, Cavalier E, Assessment of vitamin D status - a changing landscape, Clin. Chem. Lab. Med, 55 (2017) 3–26. [DOI] [PubMed] [Google Scholar]
  • [12].Makris K, Sempos C, Cavalier E, The measurement of vitamin D metabolites: part I-metabolism of vitamin D and the measurement of 25-hydroxyvitamin D, Horm.-Int. J. Endocrinol. Metab, 19 (2020) 81–96. [DOI] [PubMed] [Google Scholar]
  • [13].Makris K, Sempos C, Cavalier E, The measurement of vitamin D metabolites part II-the measurement of the various vitamin D metabolites, Horm.-Int. J. Endocrinol. Metab, 19 (2020) 97–107. [DOI] [PubMed] [Google Scholar]
  • [14].Stokes CS, Lammert F, Volmer DA, Analytical methods for quantification of vitamin D and implications for research and clinical practice, Anticancer Res, 38 (2018) 1137–1144. [DOI] [PubMed] [Google Scholar]
  • [15].Wise SA, Tai SSC, Burdette CQ, Camara JE, Bedner M, Lippa KA, Nelson MA, Nalin F, Phinney KW, Sander LC, Betz JM, Sempos CT, Coates PM, Role of the National Institute of Standards and Technology (NIST) in support of the vitamin D initiative of the National Institutes of Health, Office of Dietary Supplements, J. AOAC Int, 100 (2017) 1260–1276. [DOI] [PubMed] [Google Scholar]
  • [16].Sempos CT, Vesper HW, Phinney KW, Thienpont LM, Coates PM, Vdsp, Vitamin D status as an international issue: National surveys and the problem of standardization, Scand. J. Clin. Lab. Invest, 72 (2012) 32–40. [DOI] [PubMed] [Google Scholar]
  • [17].Tai SSC, Bedner M, Phinney KW, Development of a candidate reference measurement procedure for the determination of 25-hydroxyvitamin D-3 and 25-hydroxyvitamin D-2 in human serum using isotope-dilution liquid chromatography-tandem mass spectrometry, Anal. Chem, 82 (2010) 1942–1948. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Stepman HCM, Vanderroost A, Van Uytfanghe K, Thienpont LM, Candidate reference measurement procedures for serum 25-hydroxyvitamin D-3 and 25-hydroxyvitamin D-2 by using isotope-dilution liquid chromatography-tandem mass spectrometry, Clin. Chem, 57 (2011) 441–448. [DOI] [PubMed] [Google Scholar]
  • [19].Mineva EM, Schleicher RL, Chaudhary-Webb M, Maw KL, Botelho JC, Vesper HW, Pfeiffer CM, A candidate reference measurement procedure for quantifying serum concentrations of 25-hydroxyvitamin D-3 and 25-hydroxyvitamin D-2 using isotope-dilution liquid chromatography-tandem mass spectrometry, Anal. Bioanal. Chem, 407 (2015) 5615–5624. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Phinney KW, Bedner M, Tai SSC, Vamathevan VV, Sander LC, Sharpless KE, Wise SA, Yen JH, Schleicher RL, Chaudhary-Webb M, Pfeiffer CM, Betz JM, Coates PM, Picciano MF, Development and certification of a Standard Reference Material for vitamin D metabolites in human serum, Anal. Chem, 84 (2012) 956–962. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Phinney KW, Tai SSC, Bedner M, Camara JE, Chia RRC, Sander LC, Sharpless KE, Wise SA, Yen JH, Schleicher RL, Chaudhary-Webb M, Maw KL, Rahmani Y, Betz JM, Merkel J, Sempos CT, Coates PM, Durazo-Arvizu RA, Sarafin K, Brooks SPJ, Development of an improved Standard Reference Material for vitamin D metabolites in human serum, Anal. Chem, 89 (2017) 4907–4913. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Tai SSC, Nelson MA, Bedner M, Lang BE, Phinney KW, Sander LC, Yen JH, Betz JM, Sempos CT, Wise SA, Development of Standard Reference Material (SRM) 2973 Vitamin D Metabolites in Frozen Human Serum (High Level), J. AOAC Int, 100 (2017) 1294–1303. [DOI] [PubMed] [Google Scholar]
  • [23].CDC, Vitamin D Standardization — Certification Program https://www.cdc.gov/labstandards/vdscp.html,Centers for Disease Control and Prevention. [Google Scholar]
  • [24].Erdman P, Palmer-Toy DE, Horowitz G, Hoofnagle A, Accuracy-based vitamin D survey six years of quality improvement guided by proficiency testing, Arch. Pathol. Lab. Med, 143 (2019) 1531–1538. [DOI] [PubMed] [Google Scholar]
  • [25].Burdette CQ, Camara JE, Nalin F, Pritchett J, Sander LC, Carter GD, Jones J, Betz JM, Sempos CT, Wise SA, Establishing an accuracy basis for the Vitamin D External Quality Assessment Scheme (DEQAS), J. AOAC Int, 100 (2017) 1277–1287. [DOI] [PubMed] [Google Scholar]
  • [26].Carter GD, Berry J, Durazo-Arvizu R, Gunter E, Jones G, Jones J, Makin HLJ, Pattni P, Sempos CT, Twomey P, Williams EL, Wise SA, Hydroxyvitamin D assays: An historical perspective from DEQAS, J. Steroid Biochem. Mol. Biol, 177 (2018) 30–35. [DOI] [PubMed] [Google Scholar]
  • [27].DEQAS, Vitamin D External Quality Assessment Scheme (DEQAS); www.deqas.org.
  • [28].Binkley N, Sempos CT, Vitamin DSP, Standardizing vitamin D assays: The way forward, J. Bone Miner. Res, 29 (2014) 1709–1714. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Stockl D, Sluss PM, Thienpont LM, Specifications for trueness and precision of a reference measurement system for serum/plasma 25-hydroxyvitamin D analysis, Clin. Chim. Acta, 408 (2009) 8–13. [DOI] [PubMed] [Google Scholar]
  • [30].Phinney KW, Sempos CT, Tai SSC, Camara JE, Wise SA, Eckfeldt JH, Hoofnagle AN, Carter GD, Jones J, Myers GL, Durazo-Arvizu R, Miller WG, Bachmann LM, Young IS, Pettit J, Caldwell G, Liu A, Brooks SPJ, Sarafin K, Thamm M, Mensink GSM, Busch M, Rabenberg M, Cashman KD, Kiely M, Galvin K, Zhang JY, Kinsella M, Oh K, Lee SW, Jung CL, Cox L, Goldberg G, Guberg K, Meadows S, Prentice A, Baseline assessment of 25-hydroxyvitamin D reference material and proficiency testing/external quality assurance material commutability: A Vitamin D Standardization Program Study, J. AOAC Int, 100 (2017) 1288–1293. [DOI] [PubMed] [Google Scholar]
  • [31].Wise SA, Camara JE, Sempos CT, Burdette CQ, Hahm G, Nalin F, Kuszak AJ, Merkel J, Durazo-Arvizu R, Hoofnagle AN, Williams EL, Ivison F, Fischer R, Van den Ouweland JMW, Ho CS, Law EWK, Simard J-N, Gonthier R, Holmquist B, Meadows S, Cox L, Robyak K, Creer MH, Fitzgerald R, Clarke MW, Breen N, Lukas P, Cavalier E, Interlaboratory Comparison of 25-Hydroxyvitamin D Assays: Vitamin D Standardization Program (VDSP) Intercomparison Study 2 – Part 1 Liquid Chromatography – Tandem Mass Spectrometry (LC-MS/MS) Assays – Impact of 3-epi-25-Hydroxyvitamin D3 on Assay Performance, to be submitted toAnal. Bioanal. Chem, (2021). [DOI] [PubMed] [Google Scholar]
  • [32].Wise SA, Camara JE, Sempos CT, Burdette CQ, Hahm G, Nalin F, Kuszak AJ, Merkel J, Durazo-Arvizu R, Williams EL, Popp C, Beckert C, Schultess C, Van Slooten G, Tourneur C, Pease C, Kaul R, Villarreal A, Batista MC, Pham H, Bennett A, Jansen E, Khan DA, Kilbane M, Freeman J, Parker N, Yuan J, Mushtaq S, Simpson C, Lukas P, Cavalier E, Interlaboratory Comparison of 25-Hydroxyvitamin D Assays: Vitamin D Standardization Program (VDSP) Intercomparison Study 2 - Part 2 Immunoassays – Impact of 25-Hydroxyvitamin D2 and 24R,25-Dihydroxyvitamin D3 on Assay Performance, to be submitted toAnal. Bioanal. Chem, (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Camara JE, Wise SA, Hoofnagle AN, Williams EL, Carter GD, Jones J, Burdette CQ, Hahm G, Nalin F, Kuszak AJ, Cavalier E, Lukas P, Durazo-Arvizu R, Popp C, Beckert C, Schultess C, Van Slooten G, Tourneur C, Pease C, Kaul K, Villarreal A, Emanuelli B, Maggio A, Ivison F, Fischer R, van den Ouweland JMW, Ho CS, Law EWK, Simard J-N, Gonthier R, Holmquist B, Batista MC, Pham H, Bennett A, Meadows S, Cox L, Jansen E, Khan DA, Robyak K, Creer MH, Kilbane M, Freeman J, Parker N, Chen V, Fitzgerald R, Mushtaq S, Clarke MW, Breen N, Simpson C, Sempos CT, Assessment of serum total 25-hydroxyvitamin D assay commutability of Standard Reference Materials, College of American Pathologists Accuracy-Based Vitamin D (ABVD) Scheme and Vitamin D External Quality Assessment Scheme (DEQAS) materials: Vitamin D Standardization Program (VDSP) commutability study 2, Submitted toAnal. Bioanal. Chem, (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].CLSI, Preparation and Validation of Commutable Frozen Human Serum Pools as Secondary Reference Materials for Cholesterol Measurement Procedures, CLSI Document C37-A, Clinical and Laboratory Standards Institute, Wayne, PA, 1999. [Google Scholar]
  • [35].Danilenko U, Vesper HW, Myers GL, Clapshaw PA, Camara JE, Miller WG, An updated protocol based on CLSI document C37 for preparation of off-the-clot serum from individual units for use alone or to prepare commutable pooled serum reference materials, Clin. Chem. Lab. Med, 58 (2020) 368–374. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [36].Fabregat-Cabello N, Farre-Segura J, Huyghebaert L, Peeters S, Le Goff C, Souberbielle JC, Cavalier E, A fast and simple method for simultaneous measurements of 25(OH)D, 24, 25(OH)(2)D and the Vitamin D Metabolite Ratio (VMR) in serum samples by LC-MS/MS, Clin. Chim. Acta, 473 (2017) 116–123. [DOI] [PubMed] [Google Scholar]
  • [37].Cavalier E, Wallace AM, Carlisi A, Chapelle JP, Delanaye P, Souberbielle JC, Cross-reactivity of 25-hydroxy vitamin D2 from different commercial immunoassays for 25-hydroxy vitamin D: an evaluation without spiked samples, Clin. Chem. Lab. Med, 49 (2011) 555–558. [DOI] [PubMed] [Google Scholar]
  • [38].Schleicher RL, Sternberg MR, Lacher DA, Sempos CT, Looker AC, Durazo-Arvizu RA, Yetley EA, Chaudhary-Webb M, Maw KL, Pfeiffer CM, Johnson CL, The vitamin D status of the US population from 1988 to 2010 using standardized serum concentrations of 25-hydroxyvitamin D shows recent modest increases, Am. J. Clin. Nutr, 104 (2016) 454–461. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [39].Schleicher RL, Sternberg MR, Looker AC, Yetley EA, Lacher DA, Sempos CT, Taylor CL, Durazo-Arvizu RA, Maw KL, Chaudhary-Webb M, Johnson CL, Pfeiffer CM, National Estimates of Serum Total 25-Hydroxyvitamin D and Metabolite Concentrations Measured by Liquid Chromatography-Tandem Mass Spectrometry in the US Population during 2007–2010, J. Nutr, 146 (2016) 1051–1061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [40].CLSI, Measurement Procedure Comparison and Bias Estimation using Patient Samples,. CSLI Document EP09c, Clinical and Laboratory Standards Institute, Wayne, PA, 2018. [Google Scholar]
  • [41].Freeman J, Wilson K, Spears R, Shalhoub V, Sibley P, Performance evaluation of four 25-hydroxyvitamin D assays to measure 25-hydroxyvitamin D-2, Clin. Biochem, 48 (2015) 1097–1104. [DOI] [PubMed] [Google Scholar]
  • [42].Garnett E, Li JL, Rajapakshe D, Tam E, Meng QH, Devaraj S, Efficacy of two vitamin D immunoassays to detect 25-OH vitamin D2 and D3, Pract. Lab. Med, 17 (2019) 4. [DOI] [PMC free article] [PubMed] [Google Scholar]