Measuring Acetabular Version after Total Hip Arthroplasty: A Comparison of Two Radiographic Methods *

Afonso Cardoso 1,, Mafalda Duarte 2, Rui Viegas 1, Filipe Oliveira 1, Pedro Pinto 1, Paulo Rego 1
PMCID: PMC8405260  PMID: 34483397

Abstract

Objective  The objective of the present work was to compare the measurement of acetabular component version on anteroposterior (AP) and on cross-table radiographs after total hip arthroplasty (THA).

Methods  Radiographs of 60 hips with a primary THA were selected. Version was calculated on the AP radiograph using the Lewinnek method and, on the cross-table, using the Woo and Morrey direct method.

Results  Mean and standard deviation (SD) were different on both radiographs, being 9.7° ± 5.5° on the AP, whereas in the cross-table the measurements were 20.6° ± 8.4° ( p  < 0.001). Minding our aim of 10°, the cross-table measurements were statistically different from it ( p  < 0.001), while the AP measurement did not differ ( p  = 0.716).

Conclusion  The present study showed that the best way to correctly evaluate the acetabular component positioning following a THA is by measuring anteversion and abduction on an AP radiograph after confirming, in a cross-table radiograph, that the component is not retroverted.

Keywords: arthroplasty, replacement, hip; acetabulum; radiography

Introduction

Stability is one of the most important factors and objectives in a total hip arthroplasty (THA) and, to achieve it, correct component placement is of paramount importance. 1 An important cause of dislocation has been shown to be poor acetabular component positioning, which also leads to a limited range of motion and increased polyethylene wear. 2 3 4 Acetabular position is defined by its abduction and its version. 5 Abduction is defined as the angle between the face of the cup and the transverse axis, whereas version is defined as the angle between the axis of the component and the coronal plane. 5 The importance of these parameters lie in the recognition that subtle orientation differences can lead to a higher rate of dislocations. 6 Lewinnek et al. have suggest that an ideal cup should have an abduction of 40° and an anteversion of 15°. 7 While acetabular abduction is relatively easy to measure on anteroposterior (AP) radiographs, the same is not true for version. Numerous studies have focused on different methods to measure version on radiographs, although not all have been shown to be consistent and accurate 5 7 8 9 10 11 12 An ideal method should be accurate, reproducible, and feasible within a day-to-day practice. To measure abduction, a pelvis AP radiograph is used, whereas to measure version, the methods consist of different mathematical equations that use AP radiographs or a direct measure on a cross-table radiograph. 5 7 8 9 10 11 12 Although it is true that the measurement of version is made simple with cross-table radiographs, these radiographs are often dependent on good quality images, which can be compromised by contralateral hip joint stiffness. 13

The objective of the present work was to compare the measure of acetabular component version on an AP radiograph using the Lewinnek method and the direct measure on a cross-table radiograph. Our hypothesis is that the cross-table radiograph is more influenced by patient positioning, what could lead to an incorrect measure of acetabular version.

Material and Methods

Between June 2018 and July 2018, patients who attended our hospital were screened for inclusion in the present study. The inclusion criteria were previously primary THA for primary coxarthrosis done at our institution. Patients with a history of instrumented spine surgery were excluded. Selected patients underwent a pelvic AP and a cross-table radiograph. The present study was approved by the institution's ethics committee and informed consent was obtained from all patients. In all patients, the same acetabular component was used (Epifit, Smith & Nephew, London, United Kingdom). We use the alignment guide, according to the surgical technique, to place the acetabular component in 10° of anteversion.

All radiographs were taken in the same radiology department according to a standard protocol. Pelvic AP radiographs were done in the supine position, with both hips extended, with a source-image distance of 100 cm from the film, perpendicular to the patient, centered on the superior aspect of the pubic symphysis. Cross-table lateral radiographs were taken with the contralateral hip flexed as much as possible up to 90°, with the central ray angled perpendicularly to the long axis of the neck of the femur (45° cephalad), parallel to the examination table. All images were acquired, and all measurements were made using the Picture Archiving and Communication System (PACS) software (syngo.plaza, Siemens Healthineers, Erlangen, Germany).

Version was calculated on the AP radiograph using the Lewinnek method. 7 In this method, the ellipse formed by the acetabular component is used so that version = sin −1 (short axis/long axis) ( Fig. 1 ) On the cross-table radiograph, version was the measured angle between a line perpendicular to the examination table and a tangential line to the opening face of the acetabular component as described by Woo et al. 4 ( Fig. 2 ). Cross-table radiographs were also analyzed with the objective of excluding retroversion, which would be undetectable in the AP radiographs.

Fig. 1.

Fig. 1

Representation of the Lewinnek method: version = sin −1 (AB / CD).

Fig. 2.

Fig. 2

Representation of the Woo and Morrey method: version = angle between a perpendicular line and a tangential line to the opening face of the acetabular component.

The normality of the data was tested with the Kolmogorov-Smirnov test. The data were compared and analyzed using the Student t -test for one and dependent samples, and the Mann-Whitney or the Wilcoxon test, according to the normality of the data. The correlation of data was analyzed with the Pearson or Spearman correlation coefficient according to its normality. Nominal variables were compared using the chi-squared test. P-values < 0.05 were considered significant. Dedicated statistical software (IBM SPSS Statistics for Windows, Version 21, IBM Corp., Armonk, NY, USA) was used.

Results

We identified and included 54 patients (60 hips). There were 36 men and 18 women and the average age of the patients at surgery was 68,5 years old. No differences were found between the measurements and age or gender. The mean follow-up was of 45.3 months (24–71 months).

For the anteversion measure by the Lewinnek method on AP radiographs, the mean and standard deviation (SD) were 9.7° ± 5.5° (2°–21°). For the anteversion measure on cross-table radiographs, it was 20.6° ± 8.4° (0°–40°). As such, no cases of retroversion were present. There was a positive correlation between both methods ( n  = 60, r = 0.642, p  < 0.0001). However, both measures gave statistically different results ( p  < 0.001). The mean and SD for the difference of both methods was 10.8° ± 6.6° (-13°–29°). Minding our aim of 10° of anteversion, we tested both measurements against that value and found the cross-table measurement to be statistically different from it ( p  < 0.0001), while the AP measurement did not differ ( p  = 0.716).

Discussion

Following THA, there are several imagological exams that can be performed to assess the status of the arthroplasty and evaluate the position of the components. Although those exams, such as computed tomography (CT) scans, can be more accurate, they represent increased costs and radiation for the patient. Therefore, traditional radiography continues to be the most important exam in the follow-up of arthroplasties.

Component positioning is essential for its stability, function and longevity. 2 3 7 However, how to achieve it, namely, how much acetabular anteversion to give, remains controversial. Various authors recommend between 0° and 30° of anteversion. 1 14 15 At our center, we aim for an anteversion of 10°. While the Lewinnek concept of safe zone has guided acetabular component positioning for many years, there is evidence that an important proportion of THAs dislocate within that safe zone 16 17 Furthermore, acetabular anteversion should not be analyzed on its own, as combined femoral and acetabular anteversion might better predict dislocation risk. 18 On the other hand, other authors try to place the acetabular component in an anatomic position according to specific patient anatomic references (such as the transverse acetabular ligament). 19 20 Which of the two methods produce better results, however, is still a matter of debate. Surgeons should keep in mind that the stability of a THA is an interplay between specific patient anatomy and component positioning, and is also dependent on proper restoration of soft tissue tension and balance.

While it is easy to measure acetabular abduction, the same is not true for anteversion. Nho et al. studied various methods and concluded that the Lewinnek, Hassan and Liaw methods, based on AP radiographs, were similar to measurements made on CT scans. 10 In their study, they also included the Woo et al. method for cross-table radiographs, which also gave similar results. 10 However, another study by Arai et al. found a difference of 5° between AP and cross-table measurements. 13 Furthermore, and perhaps more clinically relevant, they also found a correlation between a contralateral stiff hip and the measurement of acetabular version on the cross-table radiograph due to pelvic tilting. 13

In our study, we found a mean acetabular anteversion of 9.7° ± 5.5° while using the Lewinnek method on an AP radiograph, which is not significantly different from our aim of 10°. On the other hand, using the cross-table radiograph, we measured an anteversion of 20.6° ± 8.4° which significantly differs from our surgical target. Both measurements were statistically different from one another. We also note a wider range of measurements in the cross-table radiographs (40° versus 19°). We found a moderate correlation between both methods, which is below expectations when measuring the same parameter.

Cross-table radiographs, when proper positioning is stressed, have been shown to accurately determine acetabular anteversion. 21 However, given the results of the present study, we believe that this method is prone to inaccuracy, due to poor positioning, probably related to contralateral hip or lumbar stiffness. Additionally, it is our opinion that pelvic tilt is harder to detect on cross-table radiographs than on AP incidences, which precludes negligent errors in the later, but not in the former radiographs. Notwithstanding, AP radiographs have some limitations because retroversion cannot be detected, and the apex of the ellipse is somewhat difficult to identify when a metal liner is used.

Regarding the limitations of our study, a measurement standard, such as a CT scan, could further strengthen our conclusions.

Conclusion

The present study showed that the best way to correctly evaluate the acetabular component positioning following a THA is by measuring anteversion and abduction on an AP radiograph after confirming, in a cross-table radiograph, that the component is not retroverted.

Conflito de Interesses Os autores não têm conflito de interesses a declarar.

Suporte Financeiro

Não houve suporte financeiro de fontes públicas, comerciais, ou sem fins lucrativos.

Financial Support

There was no financial support from public, commercial, or non-profit sources.

*

Trabalho desenvolvido no Departamento de Ortopedia e Traumatologia do Hospital Beatriz Ângelo, Loures, Portugal.

*

Work developed at the Department of Orthopedics and Traumatology, Hospital Beatriz Ângelo, Loures, Portugal.

Referências

  • 1.Charnley J. Total hip replacement by low-friction arthroplasty. Clin Orthop Relat Res. 1970;72(72):7–21. [PubMed] [Google Scholar]
  • 2.Del Schutte H, Jr, Lipman A J, Bannar S M, Livermore J T, Ilstrup D, Morrey B F. Effects of acetabular abduction on cup wear rates in total hip arthroplasty. J Arthroplasty. 1998;13(06):621–626. doi: 10.1016/s0883-5403(98)80003-x. [DOI] [PubMed] [Google Scholar]
  • 3.Widmer K H, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22(04):815–821. doi: 10.1016/j.orthres.2003.11.001. [DOI] [PubMed] [Google Scholar]
  • 4.Woo R Y, Morrey B F. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64(09):1295–1306. [PubMed] [Google Scholar]
  • 5.Murray D W. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75(02):228–232. doi: 10.1302/0301-620X.75B2.8444942. [DOI] [PubMed] [Google Scholar]
  • 6.Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stöckl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87(06):762–769. doi: 10.1302/0301-620X.87B6.14745. [DOI] [PubMed] [Google Scholar]
  • 7.Lewinnek G E, Lewis J L, Tarr R, Compere C L, Zimmerman J R. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60(02):217–220. [PubMed] [Google Scholar]
  • 8.Ackland M K, Bourne W B, Uhthoff H K. Anteversion of the acetabular cup. Measurement of angle after total hip replacement. J Bone Joint Surg Br. 1986;68(03):409–413. doi: 10.1302/0301-620X.68B3.3733807. [DOI] [PubMed] [Google Scholar]
  • 9.Müller O, Reize P, Trappmann D, Wülker N. Measuring anatomical acetabular cup orientation with a new X-ray technique. Comput Aided Surg. 2006;11(02):69–75. doi: 10.3109/10929080600640618. [DOI] [PubMed] [Google Scholar]
  • 10.Nho J H, Lee Y K, Kim H J, Ha Y C, Suh Y S, Koo K H. Reliability and validity of measuring version of the acetabular component. J Bone Joint Surg Br. 2012;94(01):32–36. doi: 10.1302/0301-620X.94B1.27621. [DOI] [PubMed] [Google Scholar]
  • 11.Visser J D, Konings J G. A new method for measuring angles after total hip arthroplasty. A study of the acetabular cup and femoral component. J Bone Joint Surg Br. 1981;63B(04):556–559. doi: 10.1302/0301-620X.63B4.7298684. [DOI] [PubMed] [Google Scholar]
  • 12.Widmer K H. A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplasty. 2004;19(03):387–390. doi: 10.1016/j.arth.2003.10.016. [DOI] [PubMed] [Google Scholar]
  • 13.Arai N, Nakamura S, Matsushita T. Difference between 2 measurement methods of version angles of the acetabular component. J Arthroplasty. 2007;22(05):715–720. doi: 10.1016/j.arth.2006.07.011. [DOI] [PubMed] [Google Scholar]
  • 14.Coventry M B. Late dislocations in patients with Charnley total hip arthroplasty. J Bone Joint Surg Am. 1985;67(06):832–841. [PubMed] [Google Scholar]
  • 15.Seki M, Yuasa N, Ohkuni K. Analysis of optimal range of socket orientations in total hip arthroplasty with use of computer-aided design simulation. J Orthop Res. 1998;16(04):513–517. doi: 10.1002/jor.1100160418. [DOI] [PubMed] [Google Scholar]
  • 16.Abdel M P, von Roth P, Jennings M T, Hanssen A D, Pagnano M W. What Safe Zone? The Vast Majority of Dislocated THAs Are Within the Lewinnek Safe Zone for Acetabular Component Position. Clin Orthop Relat Res. 2016;474(02):386–391. doi: 10.1007/s11999-015-4432-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Esposito C I, Gladnick B P, Lee Y Y. Cup position alone does not predict risk of dislocation after hip arthroplasty. J Arthroplasty. 2015;30(01):109–113. doi: 10.1016/j.arth.2014.07.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Jolles B M, Zangger P, Leyvraz P F. Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty. 2002;17(03):282–288. doi: 10.1054/arth.2002.30286. [DOI] [PubMed] [Google Scholar]
  • 19.Ha Y C, Yoo J J, Lee Y K, Kim J Y, Koo K H. Acetabular component positioning using anatomic landmarks of the acetabulum. Clin Orthop Relat Res. 2012;470(12):3515–3523. doi: 10.1007/s11999-012-2460-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Bhaskar D, Rajpura A, Board T. Current Concepts in Acetabular Positioning in Total Hip Arthroplasty. Indian J Orthop. 2017;51(04):386–396. doi: 10.4103/ortho.IJOrtho_144_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.McArthur B, Cross M, Geatrakas C, Mayman D, Ghelman B. Measuring acetabular component version after THA: CT or plain radiograph? Clin Orthop Relat Res. 2012;470(10):2810–2818. doi: 10.1007/s11999-012-2292-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
Rev Bras Ortop (Sao Paulo). 2021 Mar 22;56(4):513–516. [Article in Portuguese]

Medição da versão acetabular após artroplastia total do quadril: Comparação de dois métodos radiográficos *

Resumo

Objetivo  O objetivo do presente trabalho foi comparar a medição da versão do componente acetabular em radiografias em incidência anteroposterior (AP) e cross-table após artroplastia total do quadril (ATQ).

Métodos  Foram selecionadas radiografias de 60 quadris com ATQ primário. A versão foi calculada na radiografia AP usando o método de Lewinnek e, na cross-table , usando o método direto do Woo e Morrey.

Resultados  A média e o desvio padrão (DP) foram diferentes em ambas as radiografias, sendo 9,7° ± 5,5° no AP, enquanto na cross-table foram 20,6° ± 8,4° ( p  < 0,001). Considerando nosso objetivo de 10°, as medidas da cross-table foram estatisticamente diferentes dele ( p  < 0,001), enquanto a medição AP não diferiu ( p  = 0,716).

Conclusão  O presente estudo mostrou que a melhor maneira de avaliar corretamente o posicionamento do componente acetabular após uma ATQ é medindo a anteversão e a abdução em uma radiografia AP após confirmar, em uma radiografia cross-table , que o componente não é retrovertido.

Palavras-chave: artroplastia de quadril, acetábulo, radiografia

Introdução

A estabilidade é um dos fatores e objetivos mais importantes em uma artroplastia total do quadril (ATQ) e, para alcançá-la, a colocação correta do componente é de suma importância. 1 O posicionamento ruim do componente acetabular tem se mostrado uma causa importante de luxação, e também leva a uma limitação do movimento e ao aumento do desgaste do polietileno. 2 3 4 A posição acetabular é definida por sua abdução e sua versão. 5 A abdução é definida como o ângulo entre a face do copo e o eixo transversal, enquanto a versão é definida como o ângulo entre o eixo do componente e o plano coronal. 5 A importância desses parâmetros reside no reconhecimento de que diferenças sutis de orientação podem levar a uma taxa de deslocamentos maior. 6 Lewinnek et al. sugerem que um componente acetabular ideal deve ter um abdução de 40° e uma anteversão de 15°. 7 A abdução acetabular é relativamente fácil de medir em radiografias anteroposteriores (APs), mas isso não é verdade para a versão. Inúmeros estudos têm se concentrado em diferentes métodos para medir a versão em radiografias, embora nem todos tenham se mostrado consistentes e precisos. 5 7 8 9 10 11 12 Um método ideal deve ser preciso, reprodutível e viável dentro de uma prática cotidiana. Para medir a abdução, usa-se uma radiografia AP pélvica, enquanto que para medir a versão, os métodos consistem em diferentes equações matemáticas que usam radiografias AP ou uma medida direta em uma radiografia cross-table . 5 7 8 9 10 11 12 Embora seja verdade que a medição da versão é simples com radiografias cross-table , essas radiografias muitas vezes dependem de imagens de boa qualidade que podem ser comprometidas pela rigidez da articulação do quadril contralateral. 13

O objetivo do presente trabalho foi comparar a medida da versão do componente acetabular em uma radiografia AP usando o método Lewinnek e a medida direta em uma radiografia cross-table . Nossa hipótese é a de que a radiografia cross-table é mais influenciada pelo posicionamento do paciente, o que poderia levar a uma medida incorreta da versão acetabular.

Material e Métodos

Entre junho de 2018 e julho de 2018, os pacientes que compareceram ao hospital foram examinados para inclusão no presente estudo. Os critérios de inclusão foram ATQ primários para coxartrose primária feitos em nossa instituição. Foram excluídos pacientes com histórico de cirurgia instrumentada na coluna vertebral. Os pacientes selecionados foram submetidos a um AP pélvico e a uma radiografia cross-table . O estudo foi aprovado pelo comitê de ética da instituição e o consentimento informado foi obtido de todos os pacientes. Em todos os pacientes, foi utilizado o mesmo componente acetabular (Epifit, Smith & Nephew, Londres, Reino Unido). Utilizamos o guia de alinhamento, de acordo com a técnica cirúrgica, para colocar o componente acetabular em 10° de anteversão.

Todas as radiografias foram tiradas no mesmo departamento de radiologia de acordo com um protocolo padrão. As radiografias AP pélvicas foram feitas na posição supina, com ambos os quadris estendidos, com uma distância de imagem de origem de 100 cm do filme, perpendicular ao paciente, centrada no aspecto superior da sínfise púbica. As radiografias laterais cross-table foram tomadas com o quadril contralateral flexionado o máximo possível até 90°, com o raio central inclinado perpendicularmente ao eixo longo do pescoço do fêmur (45° cefálico), paralelo à mesa de exame. Todas as imagens foram adquiridas, e todas as medições foram feitas utilizando o software Picture Archiving and Communication System (PACS) (syngo.plaza, Siemens Healthineers, Erlangen, Alemanha).

A versão foi calculada na radiografia AP usando o método de Lewinnek. 7 Neste método, utiliza-se a elipse formada pelo componente acetabular para que a versão seja = sen −1 (eixo curto/eixo longo) ( Fig. 1 ). Na radiografia cross-table , a versão foi o ângulo medido entre uma linha perpendicular à mesa de exame e uma linha tangencial à face de abertura do componente acetabular, conforme descrito por Woo et al. 4 ( Fig 2) . A radiografia cross-table também foi analisada com o objetivo de excluir a retroversão, que seria indetectável nas radiografias AP.

Fig. 1.

Fig. 1

Representação do método de Lewinnek: versão = sen −1 (AB / CD).

Fig. 2.

Fig. 2

Representação do método de Woo e Morrey: versão = ângulo entre uma linha perpendicular e uma linha tangencial à face de abertura do componente acetabular.

A normalidade dos dados foi testada com o teste de Kolmogorov-Smirnov. Os dados foram comparados e analisados utilizando-se o teste- t de Student para uma amostra e amostra dependente, e o teste de Mann-Whitney ou Wilcoxon, de acordo com a normalidade dos dados. A correlação dos dados foi analisada com o coeficiente de correlação de Pearson ou Spearman de acordo com sua normalidade. As variáveis nominais foram comparadas utilizando-se o teste qui-quadrado. Valores-p < 0,05 foram considerados significativos. Um software estatístico dedicado (IBM SPSS Statistics for Windows, versão 21, IBM Corp., Armonk, NY, EUA) foi utilizado.

Resultados

Identificamos e incluímos 54 pacientes (60 quadris). Havia 36 homens e 18 mulheres e a média de idade dos pacientes em cirurgia era de 68,5 anos. Não foram encontradas diferenças entre as medidas e idade ou gênero. O seguimento médio foi de 45,3 meses (24 a 71 meses).

Para a medida de anteversão pelo método Lewinnek em radiografias AP, a média e o desvio-padrão (DP) foram 9,7° ± 5,5° (2°–21°). Para a medida de anteversão em radiografias cross-table , a média e o DP foram de 20,6° ± 8,4° (0°–40°). Como tal, não houve casos de retroversão. Houve correlação positiva entre ambos os métodos ( n  = 60, r = 0,642, p  < 0,0001). No entanto, ambas as medidas deram resultados estatisticamente diferentes ( p  < 0,001). A média e o DP para a diferença de ambos os métodos foram de 10,8° ± 6,6° (- 13°–29°). Considerando nosso objetivo de 10° de anteversão, testamos ambas as medidas contra esse valor e descobrimos que a medição cross-table era estatisticamente diferente dele ( p  < 0,0001), enquanto a medição AP não difere ( p  = 0,716).

Discussão

Após a ATQ, existem vários exames de imagem que podem ser realizados para avaliar o estado da artroplastia e a posição dos componentes. Embora exames como tomografia computadorizada (TC) possam ser mais precisos, eles representam aumento de custos e radiação para o paciente. Portanto, a radiografia tradicional continua sendo o exame mais importante no acompanhamento das artroplastias.

O posicionamento dos componentes é essencial para sua estabilidade, função e longevidade. 2 3 7 No entanto, como alcançá-lo, ou seja, quanto dar de anteversão acetabular, permanece controverso. Vários autores recomendam entre 0° e 30° de anteversão. 1 14 15 Em nosso centro, buscamos uma anteversão de 10°. Embora o conceito de zona segura de Lewinnek tenha guiado o posicionamento de componentes acetabulares por muitos anos, há evidências de que uma importante proporção de ATQ se desloca dentro dessa zona segura. 16 17 Além disso, a anteversão acetabular não deve ser analisada por conta própria, pois a combinação das anteversões femoral e acetabular pode prever melhor o risco de luxação. 18 Por outro lado, outros autores tentam colocar o componente acetabular em posição anatômica de acordo com referências anatômicas específicas do paciente (como o ligamento acetabular transversal). 19 20 Qual dos dois métodos produz melhores resultados, no entanto, ainda é uma questão de debate. Os cirurgiões devem ter em mente que a estabilidade de uma ATQ é uma interação entre a anatomia específica do paciente e o posicionamento de componentes, e também depende da restauração adequada da tensão e do equilíbrio do tecido mole.

Considerando que é fácil medir a abdução acetabular, não se pode dizer o mesmo para a anteversão. Nho et al. estudaram vários métodos e concluíram que os métodos de Lewinnek, Hassan e Liaw, baseados em radiografias AP, eram semelhantes às medições feitas em TCs. 10 Em seu estudo, eles também incluíram o método de Woo e Morrey para radiografias cross-table que também deram resultados semelhantes. 10 No entanto, outro estudo de Arai et al. encontrou uma diferença de 5° entre as medidas de AP e de cross-table . 13 Além disso, e talvez mais clinicamente relevante, eles também encontraram uma correlação entre um quadril duro contralateral e a medição da versão acetabular na radiografia de cross-table devido à inclinação pélvica. 13

Em nosso estudo, encontramos uma anteversão acetabular média de 9,7° ± 5,5° ao usar o método de Lewinnek em uma radiografia AP, o que não é significativamente diferente da nossa meta de 10°. Por outro lado, utilizando a radiografia cross-table , medimos uma anteversão de 20,6° ± 8,4°, o que difere significativamente do nosso alvo cirúrgico. Ambas as medidas foram estatisticamente diferentes umas das outras. Observamos também uma gama mais ampla de medições nas radiografias transversais (40° versus 19°). Encontramos uma correlação moderada entre ambos os métodos que está abaixo das expectativas ao medir o mesmo parâmetro.

A radiografia cross-table , quando o posicionamento adequado é enfatizado, tem sido mostrada para determinar com precisão a anteversão acetabular. 21 No entanto, dados os resultados do presente estudo, acreditamos que este método é propenso à imprecisão, devido ao mau posicionamento, provavelmente relacionado ao quadril contralateral ou rigidez lombar. Além disso, é nossa opinião que a inclinação pélvica é mais difícil de detectar em radiografias cross-table do que em incidências de AP, o que impede erros negligentes nessa última, mas não na primeira. Não obstante, as radiografias AP têm algumas limitações porque a retroversão não pode ser detectada e o ápice da elipse é um pouco difícil de identificar quando um forro de metal é usado.

Quanto às limitações do nosso estudo, um padrão de medição, como uma TC, poderia fortalecer ainda mais nossas conclusões.

Conclusão

O presente estudo mostrou que a melhor maneira de avaliar corretamente o posicionamento do componente acetabular após um ATQ é medindo a anteversão e a abdução em uma radiografia AP após confirmar, em uma radiografia cross-table , que o componente não é retrovertido.