Small target repeatability of 68Ga and 18F: effects of target concentration and imaging time on SUV measurements in clinically relevant phantoms

Michael S Silosky 1, Luke W Patten 2, Bennett B Chin 1
PMCID: PMC8414403  PMID: 34513281

Abstract

Quantification of tumor uptake using PET imaging is important for the evaluation of therapy response. For 18F FDG PET scans, a change in uptake of 25% is commonly considered significant. For scans using novel radiopharmaceuticals, the threshold of significance is unclear. Factors including imaging time, tumor size, activity concentration, and radiopharmaceutical may affect the repeatablity of uptake metrics. This work evaluates the effect of these parameters on the repeatablity of maximum SUV (SUVmax) and mean SUV (SUVmean) in phantoms using 18F and 68Ga. An Esser PET phantom (Data Spectrum, Durham NC) was scanned on a Biograph Horizon PET/CT scanner (Siemens Medical Solutions, Malvern PA) using 18F and 68Ga. Data were acquired for 5 minutes with reconstructions between 0.5-5 minutes. The background activity mimicked clinical scans with target-to-background (T/B) ratios from 1.7-19.8. The SUVmax and SUVmean were measured for 5 slices. The mean, standard deviation, and coefficient of variation (COV) were calculated. The effects of radionuclide, imaging time, activity concentration, and target size on COV were evaluated using multivariate gamma regressions. COV for 68Ga was 40% higher and 54% higher on average than for 18F for SUVmax and SUVmean, respectively. Decreased lesion size, imaging time, and activity concentration were significantly associated with increased COV for both metrics (P < 0.001). COV was substantially reduced at high T/B for 68Ga. At the highest T/B the COV for SUVmax and SUVmean was within the typical range seen for 18F. COV is relatively high for small targets (8 mm) but is dramatically reduced with high radiotracer uptake.

Keywords: 68Ga, DOTATATE, SUV, repeatability, phantom, high target-to-background, small lesions

Introduction

Early detection of small lesions by PET is important because therapeutic interventions may be altered based on the early detection of small metastatic lesions [1-4]. For example, 18F DCFPyl PET/CT identified early oligometastatic prostate lesions for targeted external beam radiation and demonstrated improved progression free survival and decreased the risk of new lesions in at 6 months [5]. Quantification of tumor uptake with 18F FDG PET imaging is also important to assess response to therapy and has numerous uses in drug development [6-8]. Changes in tumor uptake for FDG PET scans are not usually considered “significant” unless a greater than approximately 25% increase or decrease in Standardized Uptake Value (SUV) can be demonstrated in relatively large lesions, typically ≥ 2 cm in diameter [8].

The quantitative reproducibility of small lesions, less than 2 cm, has not been well characterized. Physical factors which may influence measurements include lesion size due to partial volume effect, target-to-background ratio (T/B), and imaging counts which are affected by scanner sensitivity, imaging time, and administered doses. Furthermore, quantitative repeatability has not been established for novel pharmaceuticals which use 68Ga as the imaging radionuclide. The physical properties of 68Ga compared to 18F include a shorter half-life, longer mean free path of the positron effecting partial volume, and the presence of high energy gamma emissions. Previous work has demonstrated that these properties may adversely affect signal-to-noise ratio, potentially impacting the ability to detect small lesions [9,10]. These properties, coupled with a typically lower administered activity for 68Ga DOTATATE studies, and more rapid physical decay result in lower true coincidence events compared to 18F FDG, also potentially adversely affecting the quantitative reproducibility of uptake measurements.

Both physical and biological effects (radiopharmaceutical biodistribution properties, tumor biology, and individual tumor heterogeneity, and others) influence the repeatability of SUV measurements. The purpose of this study is to evaluate the physical effects of repeatability of SUV measurements for 68Ga compared to 18F. Specifically for small targets, we evaluate the effects of target activity concentration, and imaging counts on the repeatability of maximum SUV (SUVmax) and mean SUV (SUVmean) in clinically relevant phantom studies.

Methods

Data acquisition

Phantom scans were performed using an Esser PET phantom (Data Spectrum, Durham NC) with data acquired and reconstructed on a Biograph Horizon PET/CT scanner (Siemens Medical Solutions, Malvern PA). The phantom was dosed in accordance with the ACR testing instructions assuming an administered activity of 370 MBq (10 mCi) for the 18F phantom and 185 MBq (5 mCi) for 68Ga phantom, resulting in a T/B of approximately 2.4 to 1 in both cases [11]. These activities where selected as they represent typical administered activities for common scans. A 60-minute decay period between phantom activity calibration and the scan was used to simulate decay during uptake. Data were acquired for 5 minutes in list mode using a single bed position centered over the hot cylinders. Following decay of the phantoms to background, these scans were repeated with the activity concentration in the cylinders adjusted to result in additional T/B ranging from 1.69 to 4.82 in 18F and 1.62 to 19.38 in 68Ga. These T/B ratios were selected to result in similar activity concentrations as seen in 18F FDG and 68Ga DOTATATE scans [12]. Table 1 lists the radionuclide, activity concentrations at scanning start time, nominal SUV in the background and cylinders, and T/B for each scan. The activity concentration in the background was chosen to yield a SUVmean of 1 and is typical of 70 kg patient following an uptake time of one hour.

Table 1.

Phantom dosing information for each scan performed including radionuclide, activity concentration at scanning start time in both background and cylinders, nominal SUV for both background and cylinders, and target-to-background ratio

Scan Radionuclide Cylinder Concentration (kBq/ml) Cylinder Nominal SUV Background Concentration (kBq/ml) Background Nominal SUV Target to Background Ratio
1 18F 5.96 1.65 3.53 0.97 1.69
2 18F 8.41 2.32 3.46 0.96 2.43
3 18F 12.59 3.49 3.48 0.97 3.59
4 18F 16.82 4.64 3.49 0.96 4.82
5 68Ga 2.23 1.55 1.37 0.96 1.62
6 68Ga 3.21 2.24 1.36 0.95 2.37
7 68Ga 4.88 3.41 1.41 0.98 3.48
8 68Ga 6.56 4.57 1.37 0.96 4.78
9 68Ga 12.88 8.98 1.36 0.95 9.51
10 68Ga 25.93 18.08 1.34 0.93 19.38

Scans using 18F use activity concentrations determined assuming a standard administered activity of 370 MBq (10 mCi). Scans using 68Ga use activity concentrations determined assuming a standard administered activity of 185 MBq (5 mCi). The background concentrations are selected to yield a nominal SUV of approximately 1. Consequently, cylinder nominal SUV is similar to the target-to-background ratio.

Image reconstruction

Images for all phantoms were reconstructed with a 180 × 180 image matrix and a voxel size of 4.1 mm × 4.1 mm × 4 mm using manufacturer (Siemens Medical Systems) point spread function correction and time-of-flight (TOF) using OSEM with 4 iterations and 10 subsets and a 5 mm gaussian post reconstruction filter. CT based attenuation correction was used with scatter modelled from the attenuation map. Each 5-minute data set was reconstructed utilizing variable imaging time with the data centered temporally in the 5-minute window as follows: 5 minutes, 4 minutes, 3 minutes, 2 minutes, 1 minute, and 0.5 minutes.

Measurement and analysis

To quantify uptake for each imaging time, target activity concentration, and lesion size, measurements were made on five adjacent slices centered in the axial extent of the hot cylinders. Circular regions of interest (ROI) were drawn on the CT images with a diameter set to match each cylinder (25 mm, 16 mm, 12 mm, and 8 mm). These ROIs were transferred to the PET images and their positioning was verified. The SUVmax and SUVmean were recorded for each ROI. The mean, standard deviation, and coefficient of variation (COV) of the 5 slices were calculated for each SUVmax and SUVmean.

Statistical analyses evaluated whether the following four design variables were associated with the COV: radionuclide type (18F and 68Ga), imaging time, activity concentration, and lesion size. More specifically, COV in SUV was modeled separately for both SUVmax and SUVmean using multivariate gamma regressions, with a random intercept for each cylinder to account for repeated measures under the different imaging times. For each SUVmax and SUVmean, the interaction between radionuclide type and each of the other three parameters were investigated, which answers whether the radionuclide type had a modifying effect on the other parameters. If the interactions were not significant, then the effect of 18F vs. 68Ga is interpreted independent of the other parameters. Moreover, the simplest model would include each of the four design variables (radionuclide type, time, size, and T/B) without any interactions, where each parameter’s effect on the COV is also accounting for the variance explained by the other parameters. For more statistical modeling details, the standard deviation was used as the outcome in a gamma regression (link = log) with the logarithm of the mean used as an offset, which avoids the concern of modeling a ratio as the outcome (COV = standard deviation/mean). This modeling approach does not change any objectives since the interpretation remains the same: in terms of the COV. The effects were estimated with empirical standard errors (or “robust sandwich estimators”). Ultimately, the interpretation of the results provides an estimated rate of change in the COV for each of the four design variables.

Additionally, two higher T/B were measured for 68Ga only: 9.51:1 and 19.38:1 to mirror the T/B measured from clinical 68Ga DOTATATE PET/CT studies [12]. Like the approach described above (using the same multivariate gamma regression format), the changes in COV for 68Ga were modeled versus the parameters of imaging time, size, and T/B. However, two slightly different modeling strategies were used to accommodate the additional T/B. First, to assess the effect of T/B, the COV of SUVmax and SUVmean from all six T/B were compared while accounting for imaging time and cylinder size. More specifically, both the 9.51:1 and 19.38:1 were each compared to the five other T/B using Dunnett’s adjustment for multiple comparisons (i.e., 9.5:1 versus all, and 19.4:1 versus all). Second, to assess the effect of imaging time and cylinder size at high T/B, only the data from the two additional T/B were used to evaluate the changes in COV.

Lastly, as an ad-hoc analysis, the effect of radionuclide type was assessed for the smallest cylinder size only (8 mm). Using the same multivariate regression approach, changes in COV were modeled versus the parameters of radionuclide type, imaging time, and T/B. Statistical modeling was conducted in SAS 9.4 software (SAS Institute Inc., Cary, NC, USA) and R. Using the Bonferroni correction for the two outcomes (SUVmax and SUVmean), the threshold for determining statistical significance was set at 0.025.

Results

Effect of Radionuclide on COV

The COV for 18F and 68Ga were found to be significantly different for both SUVmax and SUVmean as an independent effect (separate from time, T/B, or target size). On average, the COV for 68Ga scans is 40% higher for SUVmax and 54% higher for SUVmean compared to 18F scans, after accounting for imaging time, cylinder size, and T/B (P < 0.01). While these values represent the average difference across all measurements, it should be noted that, when considering the 8 mm targets alone, there is not a significant difference in COV of SUVmax (P = 0.699) or SUVmean (P = 0.858).

Figure 1 shows the five consecutive slices used for COV calculation for the 3-minute frame duration using 18F at a T/B of 2.4:1 (top row), 68Ga at a T/B of 2.4:1 (middle row), 68Ga at a T/B of 19.4:1 (bottom row). Image window width and level is adjusted to maintain consistent background intensity. As shown, 18F scans are observably less noisy than the 68Ga scans. In addition, the 8 mm target is clearly observable in 18F scan at a T/B of 2.4:1 but not in the 68Ga scan at 2.4:1. Increasing the T/B to 19.4:1 increases the visibility of all targets and renders the 8 mm target clearly observable when using 68Ga.

Figure 1.

Figure 1

Five consecutive slices used for COV calculation for the 3-minute frame duration using 18F at a T/B of 2.4:1 (top row), 68Ga at a T/B of 2.4:1 (middle row), 68Ga at a T/B of 19.4:1 (bottom row). While 68Ga scans are generally noisier that the 18F scans, high T/B may still result in clear visibility and repeatability of small targets.

Figure 2 illustrates the difference in COV calculated for both SUVmax and SUVmean between 68Ga and 18F. Here, data from the 30 second images has been omitted as the 68Ga images were very noisy and visualization of both the 8 mm cylinder and the 12 mm cylinder was not possible at the standard concentration. As shown, COV is relatively high for both radionuclides for the smallest 8 mm targets.

Figure 2.

Figure 2

The COV for SUVmax (left column) and SUVmean (right column) calculated for 5,4,3,2, and 1 minute for each target size is shown at the standard target-to-background ratio of 2.4:1 for 68Ga and 18F. As shown, COV was greater for 68Ga than for 18F at the 3 larger cylinder sizes. At the smallest target (8 mm), the difference in COV was not found to be significant for either SUVmax (P = 0.699) or SUVmean (P = 0.858).

Effects of imaging time, T/B, and target size on COV

The COV for SUVmax ranged from 0.006 to 0.152 for 18F and from 0.012 to 0.206 for 68Ga. The COV for SUVmean ranged from 0.002 to 0.147 for 18F and from approximately 0.004 to 0.189 for 68Ga. For both radionuclides, combinations of small lesion size, short imaging time, and low activity concentration in the targets tended to result in the highest COV. Table 2 shows the estimated rate of change in COV for both SUVmax and SUVmean resulting from decreases in time, T/B, and target size for both 18F and 68Ga. The radionuclide type did not significantly alter the effect from the imaging time, T/B, or size parameters on the COV (all P > 0.025). Consequently, radionuclide independent rates of change were calculated for each parameter and listed in Table 3. The changes listed for each parameter are associated with a significant increase in COV (all P < 0.001).

Table 2.

Regression estimated rates of change in COV for specific changes in time, target-to-background ratio, and target size for both SUVmax and SUVmean are shown with their confidence intervals in parentheses

Parameter Radionuclide COV Rate of Change for SUVmax (C.I.) P-Value COV Rate of Change for SUVmean (C.I.) P-Value
Time (1-minute decrease) 18F 32.4% (25.2, 40.1) 0.039 31.9% (24.6, 39.7) 0.108
68Ga 23.1% (16.4, 30.2) 24.5% (17.6, 31.8)
T/B (1-unit decrease) 18F 25.1% (11.2, 40.7) 0.202 29.5% (13.4, 47.9) 0.875
68Ga 13.8% (1.2, 27.9) 27.8% (12.0, 45.9)
Size (4 mm decrease) 18F 16.8% (6.8, 27.6) 0.143 38.2% (25.0, 52.8) 0.112
68Ga 7.6% (-1.6, 17.5) 25.0% (13.1, 38.2)

It should be noted that these are 97.5% confidence intervals, as opposed to the common 95%, since our level of statistical significance was 0.025. These rates where not found to be statistically different between 18F and 68Ga (all P > 0.025).

Table 3.

Radionuclide-independent estimated rates of change in COV for specific changes in time, target-to-background ratio, and target size for both SUVmax and SUVmean are show with their confidence intervals in parentheses

Parameter COV Rate of Change (SUVmax) P-Value COV Rate of Change (SUVmean) P-Value
Time (1-minute decrease) 27.5% (22.5, 32.8) < 0.001 28.2% (23.0, 33.5) < 0.001
T/B (1-unit decrease) 19.1% (9.4, 29.8) < 0.001 28.8% (17.2, 41.6) < 0.001
Size (4 mm decrease) 12.0% (5.0, 19.4) < 0.001 31.5% (22.4, 41.2) < 0.001

It should be noted that these are 97.5% confidence intervals, as opposed to the common 95%, since our level of statistical significance was 0.025. P-values less than 0.025 indicate that the rate of change is significantly different than zero (zero = no change in COV).

Effect of high target-to-background ratios on COV for 68Ga

As above, two additional T/B (9.51:1 and 19.38:1) were tested for 68Ga. The rationale for these additional scans is based on 68Ga DOTATATE lesions which are reported to result in higher uptake values than typically seen FDG 18F studies [12]. Overall, the addition of the two higher T/B did not significantly alter the results of this study. The results for T/B and imaging time are consistent with the primary findings described above. In general, greater T/B resulted in decreased COV across both SUVmax and SUVmean. Figure 3 illustrates this effect across multiple imaging times and target sizes with the higher T/B, common of 68Ga studies consistently resulting in lower COV. Figure 4 provides additional illustration of the advantage of high T/B, plotting COV for both SUVmax and SUVmean for the small, 8 mm targets using a 3-minute imaging time at each T/B. As shown, COV is substantially reduced for small targets at the high T/B.

Figure 3.

Figure 3

For 68Ga, the COV for SUVmax (left column) and SUVmean (right column) calculated for 5, 4, 3, 2, and 1 minute for each target size is shown at both the standard target-to-background ratio of 2.4:1 as well as the highest target to background ratio of 19.4 to 1.

Figure 4.

Figure 4

For small, 8 mm targets and a 3-minute imaging duration, the mean COV predicted by regression for both SUVmax and SUVmean is plotted at each target to background ratio using 68Ga. Error bars represent the 95% confidence intervals. COV is substantially reduced at the highest target to background ratio, demonstrating high repeatability even for small targets. High target-to-background ratios are commonly seen in 68Ga DOTATATE lesions.

Discussion

This study evaluated the COV of the SUVmax and SUVmean while varying imaging time, cylinder size, and T/B for both 68Ga and 18F. Although the phantoms used in this study are not subject to the biological variability of tumors in human patients, the ranges of the three design parameters were chosen to provide supporting evidence for SUV repeatability relevant to clinical practice. Using clinically relevant target sizes including the very small 8 mm targets, imaging times, and target activity concentrations, this study demonstrated that each of these variables have a significant, independent influence on the repeatability of uptake measurements as measured by COV.

While the COV for 68Ga scans was generally higher than that of 18F scans, it is notable that at the high 68Ga T/B, the COV for SUVmax was very low (~5%), even when tested at very short imaging times (30 seconds) and in very small targets (8 mm). This can be contrasted with the results from the standard concentration where the COV for the 8 mm object was ≥ 5% for both 18F and 68Ga. A contributing factor to this relatively high COV in the 8 mm target may have been a combination of the partial volume effect and relatively modest T/B of 2.4:1. Also, as T/B decreases, smaller targets become less distinguishable from background and COV theoretically approaches that of background. This is illustrated in Figure 1, which shows the small 8 mm 68Ga target at 2.4:1 is indistinguishable from background (second row).

These results have several important clinical implications. First, T/B may be the most important independent factor when considering variability of SUV measurements. New radiotracers may have relatively low COV if the tumor uptake is very high. Our data provides supporting evidence that small lesions (8 mm) may be reproducibly measured in clinical studies if counting statistics and tumor uptake are sufficiently high. This observation is particularly important when considered with the results of other studies evaluating the detectability of small objects at high concentrations. Previous work has shown that small objects (~0.6 cm3) that cannot be observed at relatively low activity concentrations may be consistently detected at higher T/B [9,10]. When considering our results in the context of prior studies, high T/B can not only result in small objects being readily detected, but uptake metrics for small objects (8 mm) can be quantified reproducibly.

Next, while images acquired using 68Ga radiopharmaceuticals may have reduced counts due to several factors, the overall measured COV can be sufficiently low even with shorter acquisition times, due to high T/B. The physical disadvantages of 68Ga compared to 18F may be overcome if 68Ga radiotracer ligand properties provide high tumor uptake.

Finally, the magnitude of the effect of imaging time on COV may be mitigated at high concentrations. As expected, decreases in imaging time were associated with in increases in COV for both SUVmax and SUVmean. However, at high concentrations, COV for both SUVmax and SUVmean are still quite low, even at 1 minute of imaging time, as shown in Figure 3.

Repeatability in detection and quantification of small lesions is particularly relevant for novel 68Ga radiotracers. For example, 68Ga PSMA has shown very high sensitivity in detection of prostate cancer early biochemical recurrence [3,13]. Even though 18F radiotracers have superior physical properties compared to 68Ga, 68Ga PSMA has very high affinity, is trapped and retained intracellularly, and is rapidly washed out from normal background tissue. These in vivo biological properties of 68Ga PSMA favor high target to background ratio and have ultimately proven to be superior to the 18F FACBC amino acid analogue in patient with biochemically recurrent prostate cancer [14,15]. Biologically important properties of high target-to-background ratio translate into clinical high lesion detectability which has a significant clinical impact by changing the treatment planning in radiation therapy [2,16]. It is also important for in identifying early bone metastases where very small newly appearing lesions may indicate disease progression, despite an overall decrease in quantified disease burden [17]. In clinical practice, despite the physical limitations of 68Ga compared to 18F, the 68Ga PSMA and 18F DCFPyL PET radiotracers which target PSMA show overall similar lesion detectability [18-20].

Limitations

This exploratory study has several limitations. First, these studies to evaluate the COV of SUV were performed to evaluate parameters relevant to clinical imaging. The ability to determine the effects of the small 8 mm lesions, however, may be limited by the voxel size (4 × 4 × 4 mm) which is typically 64 mm3. Error in measurements could be introduced by the relatively large voxel size compared to the 8 mm target, however, these are voxel sizes used in clinical practice. To better evaluate the effect of voxel size on 8 mm and smaller targets, phantom experiments and clinical PET studies should be performed to specifically address this question. Additionally, one factor not evaluated in this study was the effect of various reconstruction parameters in terms of iterations and subsets as well as different post-reconstruction filters. While all phantoms were reconstructed with the same parameters, it should be noted that increasing the magnitude of the post-reconstruction filter would like lead to reduced COV for both radionuclides due to further smoothing.

Second, the results reported here are specific to the experimental conditions used. The phantom data include clinically relevant target sizes ranging from 8 mm to 25 mm and activities selected to reflect clinical imaging, however clinical lesion activities and sizes may fall outside of these ranges. Additionally, these data were acquired on a single modern analog PET/CT system which uses conventional photomultiplier technology. Modern digital cameras with SiPM detectors have higher sensitivity (counts per unit activity), better TOF temporal resolution, and advanced correction techniques such as depth dependent resolution recovery, and improved image reconstruction algorithms. Overall, these improvements in both hardware and software are likely to have a favorable impact on COV compared to conventional PET/CT scanners.

Third, this study was designed to reflect clinical protocols, activity concentrations, and imaging times of clinical studies, substantial differences exist in clinical practice. Physical factors such as attenuation and scatter, are much different in clinical studies, and may contribute to overall variability. Additionally, biologic variability can also have a major impact, and may be larger and less predictable compared to the reproducibility of the instrumentation. Patient biologic variability includes radiotracer uptake properties, physical lesion uptake, background activity, patient motion, physiologic variability, and other factors which are beyond the scope of this study. This study, however, evaluates the physical aspects of scanner reproducibility, and characterizes the independent effects of lesion size, target activity, and scanning time on the variability of uptake measurements.

In conclusion, the physical properties of 68Ga, when compared to 18F, are suboptimal and may result in higher variability as measured by COV. Target size, imaging time, and T/B are independent factors contributing to COV and decreasing these parameters significantly increases the COV. While the COV is relatively high for small targets (8 mm) for both radionuclides, this can be dramatically reduced if radiotracer target uptake is very high as is commonly found in clinical 68Ga DOTATATE scans. This has clinical implications for repeatability of SUV measurements using other novel 68Ga radiotracers with high tumor uptake.

Disclosure of conflict of interest

None.

References

  • 1.Armstrong W, Thin P, Alano R, Nguyen K, Booker K, Gartmann J, Lok V, Lira S, Sonni I, Czernin J, Calais J. Incidence of local prostate cancer recurrence by 68Ga-PSMA-11 PET/CT is higher after definitive radiation therapy than after radical prostatectomy: a single center post-hoc retrospective analysis of recurrence patterns of 787 patients. J Nucl Med. 2020;61:41. [Google Scholar]
  • 2.Calais J, Czernin J, Fendler WP, Elashoff D, Nickols NG. Randomized prospective phase III trial of (68)Ga-PSMA-11 PET/CT molecular imaging for prostate cancer salvage radiotherapy planning [PSMA-SRT] . BMC Cancer. 2019;19:18. doi: 10.1186/s12885-018-5200-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Fendler WP, Calais J, Eiber M, Flavell RR, Mishoe A, Feng FY, Nguyen HG, Reiter RE, Rettig MB, Okamoto S, Emmett L, Zacho HD, Ilhan H, Wetter A, Rischpler C, Schoder H, Burger IA, Gartmann J, Smith R, Small EJ, Slavik R, Carroll PR, Herrmann K, Czernin J, Hope TA. Assessment of 68Ga-PSMA-11 PET accuracy in localizing recurrent prostate cancer: a prospective single-arm clinical trial. JAMA Oncol. 2019;5:856–863. doi: 10.1001/jamaoncol.2019.0096. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Sonni I, Eiber M, Fendler WP, Alano RM, Vangala SS, Kishan AU, Nickols N, Rettig MB, Reiter RE, Czernin J, Calais J. Impact of 68Ga-PSMA-11 PET/CT on staging and management of prostate cancer patients in various clinical settings: a prospective single-center study. J Nucl Med. 2020;61:1153–1160. doi: 10.2967/jnumed.119.237602. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Phillips R, Shi WY, Deek M, Radwan N, Lim SJ, Antonarakis ES, Rowe SP, Ross AE, Gorin MA, Deville C, Greco SC, Wang H, Denmeade SR, Paller CJ, Dipasquale S, DeWeese TL, Song DY, Wang H, Carducci MA, Pienta KJ, Pomper MG, Dicker AP, Eisenberger MA, Alizadeh AA, Diehn M, Tran PT. Outcomes of observation vs. stereotactic ablative radiation for oligometastatic prostate cancer: the ORIOLE phase 2 randomized clinical trial. JAMA Oncol. 2020;6:650–659. doi: 10.1001/jamaoncol.2020.0147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.O JH, Lodge MA, Wahl RL. Practical PERCIST: a simplified guide to PET response criteria in solid tumors 1.0. Radiology. 2016;280:576–84. doi: 10.1148/radiol.2016142043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Liu E, Paulson S, Gulati A, Freudman J, Grosh W, Kafer S, Wickremesinghe PC, Salem RR, Bodei L. Assessment of NETest clinical utility in a U.S. registry-based study. Oncologist. 2019;24:783–790. doi: 10.1634/theoncologist.2017-0623. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Wahl RL, Jacene H, Kasamon Y, Lodge MA. From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med. 2009;50(Suppl 1):122S–50S. doi: 10.2967/jnumed.108.057307. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Adler S, Seidel J, Choyke P, Knopp MV, Binzel K, Zhang J, Barker C, Conant S, Maass-Moreno R. Minimum lesion detectability as a measure of PET system performance. EJNMMI Phys. 2017;4:13. doi: 10.1186/s40658-017-0179-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Silosky M, Karki R, Morgan R, Anderson J, Chin BB. Physical characteristics of 68Ga DOTATATE PET/CT affecting small lesion detectability. Am J Nucl Med Mol Imaging. 2021;11:27–39. [PMC free article] [PubMed] [Google Scholar]
  • 11.American College of Radiology. American College of Radiology PET accreditation program testing instructions. 2020 [Google Scholar]
  • 12.Sadowski SM, Neychev V, Millo C, Shih J, Nilubol N, Herscovitch P, Pacak K, Marx SJ, Kebebew E. Prospective study of 68Ga-DOTATATE positron emission tomography/computed tomography for detecting gastro-entero-pancreatic neuroendocrine tumors and unknown primary sites. J. Clin. Oncol. 2016;34:588–96. doi: 10.1200/JCO.2015.64.0987. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Eiber M, Maurer T, Souvatzoglou M. Evaluation of hybrid 68Ga-PSMA ligand PET/CT in 248 patients with biochemical recurrence after radical prostatectomy. J Nucl Med. 2015;56:668–74. doi: 10.2967/jnumed.115.154153. [DOI] [PubMed] [Google Scholar]
  • 14.Calais J, Ceci F, Eiber M, Hope TA, Hofman MS, Rischpler C, Bach-Gansmo T, Nanni C, Savir-Baruch B, Elashoff D, Grogan T, Dahlbom M, Slavik R, Gartmann J, Nguyen K, Lok V, Jadvar H, Kishan AU, Rettig MB, Reiter RE, Fendler WP, Czernin J. 18F-fluciclovine PET-CT and 68Ga-PSMA-11 PET-CT in patients with early biochemical recurrence after prostatectomy: a prospective, single-centre, single-arm, comparative imaging trial. Lancet Oncol. 2019;20:1286–1294. doi: 10.1016/S1470-2045(19)30415-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Calais J, Fendler WP, Herrmann K, Eiber M, Ceci F. Comparison of 68Ga-PSMA-11 and 18f-fluciclovine PET/CT in a case series of 10 patients with prostate cancer recurrence. J Nucl Med. 2018;59:789–794. doi: 10.2967/jnumed.117.203257. [DOI] [PubMed] [Google Scholar]
  • 16.Emmett L, van Leeuwen PJ, Nandurkar R, Scheltema MJ, Cusick T, Hruby G, Kneebone A, Eade T, Fogarty G, Jagavkar R, Nguyen Q, Ho B, Joshua AM, Stricker P. Treatment outcomes from 68Ga-PSMA PET/CT-informed salvage radiation treatment in men with rising PSA after radical prostatectomy: prognostic value of a negative PSMA PET. J Nucl Med. 2017;58:1972–1976. doi: 10.2967/jnumed.117.196683. [DOI] [PubMed] [Google Scholar]
  • 17.Bieth M, Krönke M, Tauber R, Dahlbender M, Retz M, Nekolla SG, Menze B, Maurer T, Eiber M, Schwaiger M. Exploring new multimodal quantitative imaging indices for the assessment of osseous tumor burden in prostate cancer using 68Ga-PSMA PET/CT. J Nucl Med. 2017;58:1632–1637. doi: 10.2967/jnumed.116.189050. [DOI] [PubMed] [Google Scholar]
  • 18.Dietlein F, Kobe C, Neubauer S, Schmidt M, Stockter S, Fischer T, Schomäcker K, Heidenreich A, Zlatopolskiy BD, Neumaier B, Drzezga A, Dietlein M. PSA-stratified performance of 18F- and 68Ga-PSMA PET in patients with biochemical recurrence of prostate cancer. J Nucl Med. 2017;58:947–952. doi: 10.2967/jnumed.116.185538. [DOI] [PubMed] [Google Scholar]
  • 19.Dietlein M, Kobe C, Kuhnert G, Stockter S, Fischer T, Schomacker K, Schmidt M, Dietlein F, Zlatopolskiy BD, Krapf P, Richarz R, Neubauer S, Drzezga A, Neumaier B. Comparison of [(18)F]DCFPyL and [(68)Ga]Ga-PSMA-HBED-CC for PSMA-PET imaging in patients with relapsed prostate cancer. Mol Imaging Biol. 2015;17:575–584. doi: 10.1007/s11307-015-0866-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Alberts IL, Seide SE, Mingels C, Bohn KP, Shi K, Zacho HD, Rominger A, Afshar-Oromieh A. Comparing the diagnostic performance of radiotracers in recurrent prostate cancer: a systematic review and network meta-analysis. Eur J Nucl Med Mol Imaging. 2021;48:2978–2989. doi: 10.1007/s00259-021-05210-9. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from American Journal of Nuclear Medicine and Molecular Imaging are provided here courtesy of e-Century Publishing Corporation

RESOURCES