Efficiencies of Four Versions of the AMPLICOR HIV-1 MONITOR Test for Quantification of Different Subtypes of Human Immunodeficiency Virus Type 1

K Triques 1,*, J Coste 2, J L Perret 3, C Segarra 2, E Mpoudi 4, J Reynes 5, E Delaporte 1, A Butcher 6, K Dreyer 6,, S Herman 6, J Spadoro 6, M Peeters 1
PMCID: PMC84180  PMID: 9854073

Abstract

Three versions of a commercial human immunodeficiency virus (HIV) type 1 (HIV-1) load test (the AMPLICOR HIV-1 MONITOR Test versions 1.0, 1.0+, and 1.5; Roche Diagnostics, Branchburg, N.J.) were evaluated for their ability to detect and quantify HIV-1 RNA of different genetic subtypes. Plasma samples from 96 patients infected with various subtypes of HIV-1 (55 patients infected with subtype A, 9 with subtype B, 21 with subtype C, 2 with subtype D, 7 with subtype E, and 2 with subtype G) and cultured virus from 29 HIV-1 reference strains (3 of subtype A, 6 of subtype B, 5 of subtype C, 3 of subtype D, 8 of subtype E, 3 of subtype F, and 1 of subtype G) were tested. Detection of subtypes A and E was significantly improved with versions 1.0+ and 1.5 compared to that with version 1.0, whereas detection of subtypes B, C, D, and G was equivalent with the three versions. Versions 1.0, 1.0+, and 1.5 detected 65, 98, and 100% of the subtype A-infected samples from patients, respectively, and 71, 100, and 100% of the subtype E-infected samples from patients, respectively. Version 1.5 yielded a significant increase in viral load for samples infected with subtypes A and E (greater than 1 log10 HIV RNA copies/ml). For samples infected with subtype B, C, and D and tested with version 1.5, only a slight increase in viral load was observed (<0.5 log10). We also evaluated a prototype automated version of the test that uses the same PCR primers as version 1.5. The results with the prototype automated test were highly correlated with those of the version 1.5 test for all subtypes, but were lower overall. The AMPLICOR HIV-1 MONITOR Test, version 1.5, yielded accurate measurement of the HIV load for all HIV-1 subtypes tested, which should allow the test to be used to assess disease prognosis and response to antiretroviral treatment in patients infected with a group M HIV-1 subtype.


Human immunodeficiency virus (HIV) type 1 (HIV-1) is characterized by extensive genetic variability. Phylogenetic analysis of many isolates collected from different geographic regions worldwide shows that HIV-1 can be divided into two groups. Group M (for major) represents the vast majority of characterized isolates, and it comprises at least 10 different phylogenetically related subtypes (subtypes A to J). Group O (for outlier) represents the most divergent strains, which are mainly found in Cameroon and Cameroonians (23, 24, 36, 37). Preliminary results indicate a heterogeneous distribution of genetic subtypes, with different prevalences in each country (6). Subtype B is predominant in Europe and North America, while all known subtypes are present in Central Africa, with a predominance of subtype A in west and central Africa and of subtype C in south and eastern Africa. Subtype C is also widely prevalent in India, which is the country with the largest number of HIV-infected people in the world, whereas subtype E is predominant in southeastern Asia.

Recent studies showed the introduction of subtype A, D, and E viruses and even group O viruses in the United States (8, 38). Similar data have been reported from several European countries such as Belgium (17), France (21, 43), England (3), Sweden (2), and Germany (15), with subtype A being the predominant non-B HIV-1 subtype in those countries. A study of blood donors in France showed a 16% increase in the prevalence of individuals infected with non-subtype B viruses between 1984 and 1995 (5). The impact of the different genetic subtypes on the pathogenesis, the course of HIV-1 infection, the transmissibility, and vaccine efficacy are not yet well known and must be further studied (16). However, several reports have shown that the performance of diagnostic assays can be influenced by genetic variability (18, 25, 26, 42).

To assess the progression of HIV-1 infection in patients, the CD4 cell count is the most commonly used marker. In recent years studies have shown that measurement of the viral load in plasma is also a valuable marker for monitoring the progression of AIDS disease (19, 33, 34). To measure the response to antiretroviral therapy, a decrease in the HIV RNA level is a better predictor than an increase in the CD4 lymphocyte count (34, 40, 41). Therefore, several methods for the detection of virus in plasma have been developed (9, 39). HIV-1 p24 antigen detection in plasma by means of an antigen binding enzyme immunoassay is easy to perform but is not sensitive enough since it does no detect the p24 antigen in most asymptomatic patients (9). Quantitative culture of virus from plasma is very time-consuming and expensive, and its large-scale application is limited due to the short transport time needed to maintain the integrity of the samples. For these reasons, several quantitative nucleic acid amplification assays have been developed to quantify the HIV RNA load. The Quantiplex HIV RNA assay (Chiron)—based on branched DNA (44), the HIV-1 RNA QT (Organon Teknika) NASBA (nucleic acid sequence-based amplification) procedure (20), and the AMPLICOR HIV-1 MONITOR Test (Roche Diagnostics, Branchburg, N.J.)—is based on reverse transcription-PCR technology with coamplification of an internal RNA quantitation standard.

However, several studies have reported that some commercial viral load assays (NASBA and AMPLICOR version 1.0) were not able to detect and/or correctly quantify viral RNA in plasma from patients infected with non-subtype B HIV-1 strains (1, 4, 12, 25, 45). Due to the increase in international travel and the mixing of different populations, it is evident that the different subtypes will spread into Europe and the United States. Therefore, it is important that these commercial viral load assays be modified so that they can detect and accurately quantify all known HIV-1 variants. In the present study, we evaluated the performance of three versions of the AMPLICOR HIV-1 MONITOR Test, which contain different PCR primer pairs for detection of the HIV-1 gag gene, with plasma samples from patients infected with different genetic subtypes of HIV-1. In addition, we evaluated a prototype automated version of the test, in which amplification and detection take place on the COBAS AMPLICOR instrument.

MATERIALS AND METHODS

Origin and genetic subtypes of clinical samples.

Ninety-six plasma samples obtained from 96 different HIV-1-infected patients were studied. Sixty of these patients live in France, where they receive medical care; the remaining 36 patients live in Cameroon and are followed in a hospital in that country. Fifty of the 60 patients living in France were receiving antiretroviral therapy, whereas none of the patients in Cameroon were receiving any antiretroviral drugs. Genetic subtypes were determined in the gp120 region of the envelope gene by the Heteroduplex Mobility Assay described by Delwart et al. (13). Some samples were also characterized by sequencing the C2-V3 region of the envelope followed by phylogenetic analysis as described previously (21). The distribution of the genetic subtypes among the samples was as follows: 55 subtype A, 9 subtype B, 21 subtype C, 2 subtype D, 7 subtype E, and 2 subtype G. For each patient 20 ml of blood was drawn into tubes containing EDTA as an anticoagulant, and the plasma samples were stored at −80°C until use.

Quantified HIV-1 reference strains.

A reference panel of 29 HIV-1 isolates of subtypes A through G was prepared as part of a study to prepare a set of quantified, well-characterized HIV-1 subtype standards and determine the quantitative performance of the HIV-1 MONITOR Test with HIV-1 subtypes. Details on the preparation and characterization of the HIV-1 subtype reference panel will be described elsewhere (35). Briefly, each isolate was cultured in vitro by standard methods, and the culture supernatants were clarified and frozen. The virion concentration in the culture supernatants was determined by electron microscopic particle counting (22). Each virus stock was diluted to 20,000 virus particles per ml in defibrinated, normal human plasma. The 29 reference strains used in this study represented the following subtypes: subtype A (UG273, DJ258, and DJ263), subtype B (US1, US3, US4, CM237, BK132, and BZ167), subtype C (ZAM18, UG278, ETH2220, SE364, and SM145), subtype D (SE365, UG270, and UG274), subtype E (CM235, CM238, CM240, CM243, POC30506, RA12, RA17, and NP1465), subtype F (BZ126, BZ162, and BZ163), and subtype G (HH8793).

Quantification of HIV-1 RNA.

Four versions of the AMPLICOR HIV-1 MONITOR Test (Roche Diagnostics) were evaluated. The lower detection limit for all the versions is 200 RNA copies/ml, and the upper limit is 750,000 HIV RNA copies/ml. Version 1.0 is the currently available commercial test (product codes 83088 and 83102; Roche Diagnostics). Versions 1.0+ and 1.5 are modifications to version 1.0 to improve the performance with non-B subtypes of HIV-1. The prototype automated test is a modification to version 1.5 in which amplification, detection, and calculation of results are automated on the COBAS AMPLICOR instrument. The COBAS AMPLICOR technique, which allows fully automated RNA and DNA amplification and detection, has been previously described in detail (14). The four versions differ in the internal quantitation standard RNA, the PCR primers, and the thermal cycling conditions that they use. The sample processing procedure is identical for all four versions. Versions 1.0 and 1.0+ use the version 1.0 quantitation standard RNA. Version 1.5 and the prototype automated test use a modified quantitation standard RNA that contains an additional 20 nucleotides of the HIV-1 sequence to include the primer binding site of the new downstream PCR primer. All four versions amplify the same region in the gag gene of HIV-1. Version 1.0 uses primers SK462 and SK431 and yields a 142-bp amplification product. Version 1.0+ is identical to version 1.0, except that the amplification reaction contains two additional primers, SK145 and SK151, which are the same lengths as SK462 and SK431, respectively, and which bind to the same sites. The nucleotide sequences of SK145 and SK151 have been modified to be more homologous to the non-B subtypes of HIV-1. Version 1.5 uses primers SK145 and SKCC1B and yields a 155-bp amplification product. The downstream primer (SKCC1B) was shifted downstream by 13 nucleotides to a region more highly conserved among the group M subtypes of HIV-1. The prototype automated test uses the same amplification reagents as version 1.5. Versions 1.0 and 1.0+ use the version 1.0 thermal cycling profile. The thermal cycling profiles for version 1.5 and the prototype automated test have been reoptimized with lower annealing temperatures to improve the performance with non-B subtypes of HIV-1. All four versions use the same oligonucleotide hybridization probes (SK102 and CP35) for capture of the HIV and quantitation standard amplification products, respectively. The detection procedure and reagents are identical for versions 1.0, 1.0+, and 1.5. The prototype automated test uses similar detection chemistries, except that oligonucleotide probe-coated magnetic microparticles instead of probe-coated microwell plates are used for capture of the amplification products.

The study protocol is illustrated in Fig. 1. Each sample was extracted in duplicate with the version 1.0 and version 1.5 quantitation standard RNA, according to the manufacturer’s instructions. The samples processed with the version 1.0 quantitation standard were used for version 1.0 and 1.0+ amplification; the samples processed with the version 1.5 quantitation standard were used for version 1.5 and amplification on the COBAS instrument. Negative, low-positive, and high-positive controls were included in each run of 10 tests.

FIG. 1.

FIG. 1

Outline of study protocol.

Statistical analysis.

In order to compare to what extent the data obtained with the AMPLICOR version 1.5 test agree between the microwell plate version and the prototype automated test, the data were analyzed as described by Bland and Altman (7). The analysis was based on the differences between measurements for the same samples by plotting the differences against the average.

RESULTS

Sensitivity for detection of HIV-1 RNA from different genetic subtypes.

Overall, the version 1.0, 1.0+, and 1.5 tests detected viral RNA in 73 (76%), 94 (98%), and 95 (99%) of the 96 patient samples, respectively (Table 1). The only sample in which HIV-1 RNA was not detected with version 1.5 was also negative by the two other tests. It was obtained from a patient who was infected with HIV-1 subtype B and who was receiving antiretroviral treatment. For this same patient we were also unable to isolate the virus from the corresponding peripheral blood mononuclear cells (PBMCs). The version 1.0+ and 1.5 tests detected significantly more subtype A- and E-infected samples compared to the numbers detected by version 1.0 (Table 1). Thirty-six (65%), 54 (98%), and 55 (100%) of the 55 subtype A-infected samples were detected with the version 1.0, 1.0+, and 1.5 tests, respectively; and 5 (71.4%), 7 (100%), and 7 (100%) of the 7 subtype E-infected samples were detected by the three tests, respectively. The only subtype C-infected sample not detected with version 1.0 had very low HIV RNA levels with versions 1.0+ and 1.5: 288 and 294 copies per ml, respectively.

TABLE 1.

Number and percentage of patient samples in which HIV-1 RNA was detected with versions 1.0, 1.0+, and 1.5 of the AMPLICOR HIV-1 MONITOR Test

Genetic subtype No. of samples tested No. (%) of samples in which HIV-1 RNA was detected by the following version of the AMPLICOR HIV-1 MONITOR Test:
Version 1.0 Version 1.0+ Version 1.5
A 55 36 (65) 54 (98) 55 (100)
B 9 8 (89) 8 (89) 8 (89)
C 21 20 (95) 21 (100) 21 (100)
D 2 2 (100) 2 (100) 2 (100)
E 7 5 (71) 7 (100) 7 (100)
G 2 2 (100) 2 (100) 2 (100)
Total 96 73 (76) 94 (98) 95 (99)

Table 2 presents the HIV-1 loads, expressed as log10 HIV RNA copies per milliliter, obtained with the panel of quantified reference strains of HIV-1. Each test sample contained 20,000 virus particles per ml. Two of the three subtype A strains were not detected with the version 1.0 test; all three subtype A strains were detected with versions 1.0+ and 1.5. All of the subtype B, C, D, E, F, and G strains were detected by versions 1.0, 1.0+, and 1.5.

TABLE 2.

HIV-1 RNA levels determined by the three AMPLICOR HIV-1 MONITOR microwell plate tests and the prototype automated AMPLICOR HIV-1 MONITOR Test

Virus strain Refer-ence(s) Sub-type RNA level (log10 copies/ml)
Version 1.0 Version 1.0+ Microwell plate version of 1.5 Prototype automated version of 1.5
UG273 29 A 2.975 4.862 5.604 5.149
DJ25 8 28, 29 A <dla 4.584 4.665 4.678
DJ263 29 A <dl 4.372 4.781 4.640
US1 30 B 4.761 5.109 5.211 4.894
US3 30 B 4.643 4.767 5.514 4.841
US4 30 B 4.436 4.433 4.731 4.581
CM237 30 B 4.742 4.649 5.132 4.600
BK132 32 B 4.799 4.865 5.487 5.068
BZ167 27, 28 B 5.041 4.778 5.123 5.143
ZAM18 29 C 4.598 4.540 5.183 4.728
UG278 28, 29 C 4.767 4.827 4.677 4.665
ETH2220 41a C 4.707 4.902 5.374 4.941
SE364 29 C 4.347 4.684 4.767 4.461
SM145 28, 29 C 5.065 5.109 5.142 5.017
SE365 28, 29 D 4.613 4.708 4.621 4.618
UG270 28 D 4.512 4.692 4.831 4.650
UG274 28, 29 D 4.617 4.729 4.656 5.111
CM235 31 E 5.274 5.294 5.853 5.274
CM238 27, 30 E 3.484 4.915 4.635 4.533
CM240 27, 30 E 3.309 4.486 4.655 4.609
CM243 32 E 3.217 4.797 4.891 4.723
POC30506 37a E 3.450 4.821 5.238 4.886
RA12 37a E 2.937 4.882 4.790 4.473
RA17 37a E 3.356 4.776 4.838 4.583
NP1465 37a E 3.293 4.642 4.724 3.930
BZ126 28 F 3.147 4.071 5.612 4.572
BZ162 27, 28 F 3.872 5.133 5.166 5.140
BZ163 27 F 3.173 3.759 5.076 4.996
HH8793  9a G 4.216 4.830 4.819 5.196
Mean 4.124 4.725 5.027 4.783
SD 0.75 0.31 0.35 0.29
a

<dl, less than the detection limit. 

Efficiency of quantitation of HIV-1 RNA from different genetic subtypes.

To determine to what extent AMPLICOR HIV-1 MONITOR Test versions 1.0, 1.0+, and 1.5 were able to quantify HIV-1 RNA of various subtypes, we compared the viral loads obtained with the patient samples with each version (Fig. 2 and 3). Figure 2 presents the individual results obtained for all 96 samples; Fig. 3 presents the average results obtained with each test version by subtype. A significant increase in viral load of at least 1 order of magnitude was observed for the subtype A- and E-infected samples with the version 1.0+ and 1.5 tests. For the subtype B-, C-, and D-infected samples, versions 1.0 and 1.0+ yielded equivalent results. The version 1.5 results for subtypes B, C, and D were highly correlated to those for versions 1.0 and 1.0+, but the loads were slightly higher. The results obtained with versions 1.5 and 1.0+ were very highly correlated for all subtypes, but the loads obtained with version 1.5 were generally higher, by approximately 0.4 log10 HIV RNA copies/ml. For some subtype A-infected samples the loads obtained with version 1.5 were more than 1 log10 HIV RNA copies/ml higher than those obtained with version 1.0 (Fig. 2C).

FIG. 2.

FIG. 2

HIV loads in clinical samples with AMPLICOR HIV-1 MONITOR Test versions 1.0, 1.0+, and 1.5 and the prototype automated test. The dotted lines indicate the upper limit of the linear range of the tests, at 750,000 HIV RNA copies/ml. (A) Version 1.0+ compared to version 1.0. (B) Version 1.5 compared to version 1.0. (C) Version 1.5 compared to version 1.0+. (D) Prototype automated test compared to version 1.5. Data for the four samples in which HIV-1 RNA was not detected by the prototype automated test were excluded from the linear regression analysis in panel D. The different samples are represented as follows: ○, subtype A (n = 55); •, subtype B (n = 9); ▴, subtype C (n = 21); ▾, subtype D (n = 2); □, subtype E (n = 7); and ▵, subtype G (n = 2).

FIG. 3.

FIG. 3

Average log10 HIV RNA copies per milliliter for subtypes A, B, C, D, E, and G from clinical specimens. For subtypes D, E, and G the results for individual samples are shown. For subtypes A, B, and C, the brackets represent the 95% confidence interval of the average. Only data for samples that were positive and below the upper limit of all four versions (750,000 HIV RNA copies/ml) were included in the figure. V, version.

As shown in Table 2, a similar trend was observed for the quantification of HIV-1 RNA from the reference strains. Each test sample contained 20,000 virus particles per ml. Therefore, the expected result was approximately 40,000 RNA copies per ml (or 4.6 log10 HIV RNA copies/ml). The overall mean viral loads measured in the viral stocks were 4.124 ± 0.750, 4.725 ± 0.379, and 5.027 ± 0.350 log10 copies per ml for the version 1.0, 1.0+, and 1.5 tests, respectively. The viral load increased mainly for subtype A-, E-, and F-infected samples, with increases of 1.89, 1.29, and 0.92 log10 copies/ml, respectively, between version 1.0 and version 1.0+ of the test and with 2.63, 1.41, and 1.89 log10 increases observed between version 1.0 and version 1.5 of the test. Only a small increase in viral load was observed for subtype B-, C-, and D-infected samples.

Correlation between AMPLICOR HIV-1 MONITOR Test version 1.5 and the prototype automated test.

HIV RNA was not detected by the prototype automated test in four of the patient samples: one sample each infected with subtypes A, B, C, and E. None of the four test methods detected HIV RNA in the subtype B-infected sample, which was from a patient receiving antiretroviral therapy (described above). The subtype A-, C-, and E-infected samples had low but detectable viral loads with version 1.5 of the test of 2,630, 294, and 499 HIV RNA copies/ml, respectively.

The prototype automated test and version 1.5 yielded results that were highly correlated for all subtypes, demonstrating that the correlation does not depend on subtype (Fig. 2D and 3). The correlation between the two tests was calculated by the method of Bland and Altman (7) by plotting differences against the average (Fig. 4). The correlation coefficient is 0.05 (95% confidence interval, −0.18 to 0.28), and the high P value (P = 0.655) indicates that the differences observed between the two tests are not significant, independent from the average. However, the loads obtained with the prototype automated test were generally lower than those obtained with version 1.5 by approximately 0.3 log10 HIV RNA copies/ml (from 0.24 to 0.38 log10 HIV RNA copies/ml) (Fig. 2D and 3).

FIG. 4.

FIG. 4

Comparison of AMPLICOR HIV-1 MONITOR Test version 1.5 (microwell plate) and the prototype automated version as described by Bland and Altman (7). Only data for samples that were positive and that had loads below the upper limit for all four versions (750,000 HIV RNA copies/ml) were included in the figure. The broken lines indicate the 95% confidence interval of the mean difference between the two tests for each sample.

As observed for the plasma samples, a good correlation between the two techniques was observed, with slightly higher viral loads for the reference strains also being detected by version 1.5 compared to the loads detected by the prototype automated test (Table 2).

Reproducibility.

A total of 10 test runs have been done by one technician, and in each run, one high-positive and one low-positive control and one negative control were included. The reproducibilities of the different test versions were measured by calculating the mean and standard deviation from the values obtained for each of the controls. In each run of the different tests all the positive controls were positive and all the negative controls were negative. For the low-positive control the values observed were as follows: 3.594 ± 0.18, 3.554 ± 0.18, 3.930 ± 0.16, and 3.735 ± 0.11 log10 copies/ml with versions 1.0, 1.0+, and 1.5 and the prototype automated test, respectively; and for the high-positive control the averages were 5.520 ± 0.12, 5.752 ± 0.15, 5.917 ± 0.12, and 5.762 ± 0.17 log10 copies/ml for versions 1.0, 1.0+, and 1.5 and the prototype automated version, respectively.

DISCUSSION

Viral load in plasma is an essential marker for the prediction of disease progression and assessment of the potential of therapeutic drugs and vaccines (41), and the quantification of HIV RNA in plasma has become a main tool in predicting disease outcome and monitoring the efficacy of antiretroviral therapy in HIV-1-infected patients. However, the current AMPLICOR HIV-1 MONITOR Test is unable to quantitate accurately the RNA of some non-B subtypes of HIV-1.

In this study we evaluated three versions of the AMPLICOR HIV-1 MONITOR Test, version 1.0 (the current test), version 1.0+ (the current test with an additional pair of PCR primers), and version 1.5 (an upgraded test with new PCR primers), for their performance with group M subtypes of HIV-1. We also evaluated a prototype automated test, in which amplification and detection are automated on the COBAS AMPLICOR instrument.

Subtype A- and E-infected patient samples were poorly detected by the version 1.0 test (66% [41 of 62] were positive). Detection of subtype A- and E-infected samples was significantly improved with the version 1.0+ and 1.5 tests (98% [61 of 62] and 100% [62 of 62], respectively). Detection of subtype C-, D-, and G-infected samples was equivalent with versions 1.0, 1.0+, and 1.5 of the test (96% [24 of 25], 100% [25 of 25], and 100% [25 of 25], respectively), as was detection of subtype B-infected samples (89% [8 of 9] with all three versions). One subtype B-infected patient sample was negative by these three tests, but it came from a patient undergoing antiviral treatment and for whom the viral culture from the corresponding PBMCs remained negative, suggesting a very low viral load in this patient.

The version 1.0 test uses the PCR primers SK462 (upstream, sense primer) and SK431 (downstream, antisense primer). Alignment of these primers with HIV-1 gag sequences in GenBank shows that the upstream primer (SK462) contains three to six mismatches with subtype A sequences and that the downstream primer (SK431) contains a mismatch two bases from the 3′ end in many subtype A, F, and G isolates (18). The presence of more than four mismatches at critical positions in a primer decreases the efficiency of the amplification (10). This decrease in efficiency can result in decreases in viral load measurements of more than 10-fold. The lower sensitivity for subtype A- and E-infected samples and the lower viral load results for the detectable subtype A- and E-infected samples are most likely explained by the mismatches with the primers. All the subtype A-infected samples from Cameroon were from AIDS patients receiving no antiretroviral treatment, but the viral loads obtained by the version 1.0 test were lower than expected for such a population. However, the viral loads in these patients observed by AMPLICOR HIV-1 MONITOR Test version 1.5 were similar to the viral loads expected in subtype B-infected patients at terminal stages of the disease.

Plasma RNA levels obtained with version 1.0+ with the subtype A- and E-infected samples were significantly greater than those obtained with version 1.0 but were slightly less than those obtained with version 1.5. Version 1.0+ includes the downstream primer SK151. SK151 differs from SK431 by one nucleotide, which improves the homology to all HIV-1 isolates. However, the mismatch with subtype A samples near the 3′ end is still present in SK151, which may explain the slightly lower results for subtype A- and E-infected samples obtained with version 1.0+ compared to those obtained with version 1.5.

AMPLICOR HIV-1 MONITOR Test version 1.5 is able to detect and quantify HIV-1 RNA from all the HIV-1 group M genetic subtypes. However, the numbers of HIV-1 RNA copies per milliliter are slightly larger than those determined by the viral particle count assay, suggesting that the quantification of the viral load with this version could be slightly overestimated.

In this study we also evaluated a prototype automated test on the COBAS AMPLICOR instrument. The prototype automated test uses the same amplification reaction as version 1.5, except that the amplification takes place in the COBAS instrument. The results with version 1.5 and the prototype automated test were highly correlated; however, the counts obtained with the automated test were generally 0.3 log10 lower than those obtained with version 1.5, and the HIV-1 RNA in four samples with low viral loads with version 1.5 was not detected with the automated test. The lower counts obtained with the prototype automated test can be due to the difference in how the amplicons are captured (microwells versus microbeads) or to the detection reagents used in the prototype automated test. Evaluation of the modified test is in progress at Roche Molecular Systems. However, the counts obtained by the prototype automated test are very close to the viral particle counts, and this result therefore makes us believe that the automated test on the COBAS instrument may become the future HIV-1 MONITOR test for the quantification of the viral load in plasma.

In conclusion, the new gag primers and other changes in version 1.5 significantly improved the detection and quantification of non-B subtypes of HIV-1 by the AMPLICOR HIV-1 MONITOR Test. This improvement was especially observed for subtype A, E, G, and also F strains, whereas previously, several studies reported difficulties in the detection and quantification HIV-1 RNA from subtype A-, E-, and G-infected samples (1, 12, 18, 25). All subtype E strains are recombinant viruses with a gag region from subtype A viruses; also, several subtype G viruses have been described as recombinants with a subtype A gag region (11, 32). This new version permits the accurate measurement of viral load for all group M subtypes tested and should therefore allow the accurate assessment of disease prognosis and response to antiretroviral treatment in patients infected with a group M subtype of HIV-1.

ACKNOWLEDGMENTS

We express our gratitude to Christian Laurent for help with the statistical analysis.

This work was partially supported by a grant from the ANRS (Agence National de Recherche sur le SIDA).

REFERENCES

  • 1.Alaeus A, Lidman K, Sönnerborg A, Albert J. Subtype specific problems with quantification of plasma HIV-1 RNA. AIDS. 1997;11:859–865. doi: 10.1097/00002030-199707000-00004. [DOI] [PubMed] [Google Scholar]
  • 2.Alaeus A, Leitner T, Lidman K, Albert J. Most HIV-1 genetic subtypes have entered Sweden. AIDS. 1997;11:199–202. doi: 10.1097/00002030-199702000-00010. [DOI] [PubMed] [Google Scholar]
  • 3.Arnold C, Barlow K L, Parry J V, Clewley J P. At least five HIV-1 sequence subtypes (A, B, C, D, A/E) occur in England. AIDS Res Hum Retroviruses. 1995;11:427–429. doi: 10.1089/aid.1995.11.427. [DOI] [PubMed] [Google Scholar]
  • 4.Arnold C, Barlow K L, Kaye S, Loveday C, Balfe P, Clewley J P. HIV type 1 sequence subtype G transmission from mother to infant: failure of variant sequence species to amplify in the Roche AMPLICOR test. AIDS Res Hum Retroviruses. 1995;8:999–1001. doi: 10.1089/aid.1995.11.999. [DOI] [PubMed] [Google Scholar]
  • 5.Barin F, Courouce A M, Pillonel J, Buzelay L. The retrovirus study group of the French society of blood transfusion: increasing diversity of HIV-1M serotypes in French blood donors over a 10-year period (1985–1995) AIDS. 1997;11:1503–1508. doi: 10.1097/00002030-199712000-00015. [DOI] [PubMed] [Google Scholar]
  • 6.Birx, D. L. 1996. Summary of track A: basic science. AIDS 10(Suppl. 3):S85–S106. [PubMed]
  • 7.Bland J M, Altman D G. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995;346:1085–1087. doi: 10.1016/s0140-6736(95)91748-9. [DOI] [PubMed] [Google Scholar]
  • 8.Brodine S K, Mascola J R, Weiss P J, Ito S I, Porter K R, Artenstein A W, Garland F C, McCutchan F E, Burke D S. Detection of diverse HIV-1 genetic subtypes in the United States. Lancet. 1995;346:1198–1199. doi: 10.1016/s0140-6736(95)92901-0. [DOI] [PubMed] [Google Scholar]
  • 9.Bush C E, Donovan R M, Manzor O, Baxa D, Moore E, Cohen F, Saravdatz L D. Comparison of HIV type 1 RNA plasma viremia, p24 antigenemia and unintegrated DNA as viral load markers in pediatric patients. AIDS Res Hum Retroviruses. 1996;12:11–15. doi: 10.1089/aid.1996.12.11. [DOI] [PubMed] [Google Scholar]
  • 9a.Carr J K, Salminer M O, Albert J, Sanders-Buell E, Gotte D, Birx D L, McCutchan F E. Full genome sequences of human immunodeficiency virus type 1 subtypes G and A/G intersubtype recombinant. Virology. 1998;247:22–31. doi: 10.1006/viro.1998.9211. [DOI] [PubMed] [Google Scholar]
  • 10.Christopherson C D, Kwok S. Abstracts of the XI International Conference on AIDS. 1996. The effects of mutations on HIV-1 quantitation by RT/PCR, abstr. MoA161. [Google Scholar]
  • 11.Cornelissen M, Kampinga G, Zorgdrager F, Goudsmit J the UNAIDS Network for HIV Isolation and Characterization. Human immunodeficiency virus type 1 subtypes defined by env show high frequency of recombinant gag genes. J Virol. 1996;70:8209–8212. doi: 10.1128/jvi.70.11.8209-8212.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Coste J, Montes B, Reynes J, Peeters M, Segarra C, Vendrell J P, Delaporte E, Segondy M. Comparative evaluation of three assays for the quantitation of human immunodeficiency virus type 1 RNA in plasma. J Med Virol. 1996;50:293–302. doi: 10.1002/(SICI)1096-9071(199612)50:4<293::AID-JMV3>3.0.CO;2-3. [DOI] [PubMed] [Google Scholar]
  • 13.Delwart E L, Shpaer E G, Louwagie F, McCutchan F E, Grez M, Rubsamen-Waigmann H, Mullins J I. Genetic relationships determined by a DNA heteroduplex mobility assay: analysis of HIV-1 env genes. Science. 1993;262:1257–1261. doi: 10.1126/science.8235655. [DOI] [PubMed] [Google Scholar]
  • 14.DiDomenico N, Link H, Knobel R, Caratsch T, Weschler W, Loewy Z G, Rosentraus M. COBAS AMPLICOR: fully automated RNA and DNA amplification and detection system for routine diagnostic PCR. Clin Chem. 1996;42:1915–1923. [PubMed] [Google Scholar]
  • 15.Dietrich U, Ruppach H, Gehring S, Knechten H, Knickmann M, Jager H, Wolf E, Husak R, Orfanos C E, Brede H D, Rubsamen-Waigman H, von Briesen H. Large proportion of non-B HIV-1 subtypes and presence of zidovudine resistance mutations among German seroconvertors. AIDS. 1997;11:1532–1533. [PubMed] [Google Scholar]
  • 16.Expert Group of the Joint United Nations Programme on HIV/AIDS. Implications of HIV variability for transmission: scientific and politic issues. AIDS. 1997;11:UNAIDS1–UNAIDS15. [PubMed] [Google Scholar]
  • 17.Fransen K, Buvé A, Nkengasong J N, Laga M, van der Groen G. Longstanding presence in Belgians of multiple non-B HIV-1 subtypes. Lancet. 1996;347:1403. doi: 10.1016/s0140-6736(96)91042-9. . (Letter.) [DOI] [PubMed] [Google Scholar]
  • 18.Gobbers E, Fransen K, Oosterlaken T, Janssens W, Heyndrickx L, Ivens T, Vereecken K, van de Wiel P, van der Groen G. Reactivity and amplification system with regard to different HIV-1 subtypes. J Virol Methods. 1997;66:293–301. doi: 10.1016/s0166-0934(97)00072-4. [DOI] [PubMed] [Google Scholar]
  • 19.Harrigan R. 1995. Measuring viral load in the clinical setting. J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 10(Suppl):S34–S40. [PubMed]
  • 20.Kievits T, Rothbart P, Quint W, Huisman H, Goudsmit J, Lens P. NASBA isothermal enzymatic in vitro nucleic acid amplification optimized for the diagnosis of HIV-1 infection. J Virol Methods. 1991;35:273–286. doi: 10.1016/0166-0934(91)90069-c. [DOI] [PubMed] [Google Scholar]
  • 21.Lasky M, Perret J L, Peeters M, Bibollet-Ruche F, Liegeois F, Patrel D, Molinier S, Gras C, Delaporte E. Presence of non-B subtypes and divergent subtype B strains of HIV-1 in individuals infected after overseas deployment. AIDS. 1997;11:43–51. doi: 10.1097/00002030-199701000-00007. [DOI] [PubMed] [Google Scholar]
  • 22.Layne S P, Merges M J, Dembo M, et al. Factors underlying spontaneous inactivation and susceptibility to neutralization of human immunodeficiency virus. Virology. 1992;189:695–714. doi: 10.1016/0042-6822(92)90593-e. [DOI] [PubMed] [Google Scholar]
  • 23.Leitner T. Human retroviruses and AIDS. Los Alamos, N.M: Los Alamos National Laboratory; 1996. Genetic subtypes of HIV-1. [Google Scholar]
  • 24.Loussert-Ajaka I, Descamps D, Simon F, Brun-Vezinet F, Ekwalanga M, Saragosti S. Genetic diversity and HIV detection by polymerase chain reaction. Lancet. 1995;346:912–913. doi: 10.1016/s0140-6736(95)92762-x. [DOI] [PubMed] [Google Scholar]
  • 25.Loussert-Ajaka I, Chaix M L, Korber B, Letourneur F, Gomas E, Allen E, Ly T D, Brun-Vezinet F, Simon F, Saragosti S. Variability of human immunodeficiency virus type-1 group O strains isolated from Cameroonian patients living in France. J Virol. 1995;69:5640–5649. doi: 10.1128/jvi.69.9.5640-5649.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Loussert-Ajaka I, Ly T D, Chaix ML, Ingrand D, Saragosti S, Courouce A M, Brun-Vezinet F, Simon F. HIV-1/HIV-2 seronegativity in HIV-1 subtype O infected patients. Lancet. 1994;343:1393–1394. doi: 10.1016/s0140-6736(94)92524-0. [DOI] [PubMed] [Google Scholar]
  • 27.Louwagie J J, Delwart E L, Mullins J I, McCutchan F E, Eddy G, Burke D S. Genetic analysis of HIV-1 isolates from Brazil reveals presence of two distinct genetic subtypes. AIDS Res Hum Retroviruses. 1994;10:561–567. doi: 10.1089/aid.1994.10.561. [DOI] [PubMed] [Google Scholar]
  • 28.Louwagie J J, McCutchan F E, Peeters M, Brennan T, Sanders-Buell E, Eddy G, van der Groen G, Fransen K, Gershy-Damet M, Deleys R, Burke D. Phylogenetic analysis of gag genes from seventy international HIV-1 isolates provides evidence for multiple genotypes. AIDS. 1993;7:769–780. doi: 10.1097/00002030-199306000-00003. [DOI] [PubMed] [Google Scholar]
  • 29.Louwagie J J, Janssens W, Mascola J J, Fischer C L, Hegerich P A, van der Groen G, McCutchan F E, Eddy G, Burke D. Genetic diversity of the HIV-1 envelope glycoprotein. J Virol. 1995;69:263–271. doi: 10.1128/jvi.69.1.263-271.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Mascola J J, Louwagie J J, McCutchan F E, Fischer C L, Hegerich P A, Wagner K F, Fowler A K, McNeil J G, Burke D S. Two antigenically distinct subtypes of human immunodeficiency virus type 1: viral genotype predicts neutralization serotype. J Infect Dis. 1993;169:48–54. doi: 10.1093/infdis/169.1.48. [DOI] [PubMed] [Google Scholar]
  • 31.McCutchan, F. E., M. O. Salminen, J. K. Carr, and D. S. Burke. 1996. HIV genetic diversity. AIDS 10(Suppl. 3):S13–S20. [PubMed]
  • 32.McCutchan F E, Hegerich P A, Brennan T P, Phanuphak P, Singharaj P, Jugsudee A, Berman P W, Gray A M, Fowler A K, Burke D S. Genetic variants of HIV-1 in Thailand. AIDS Res Hum Retroviruses. 1992;8:1887–1895. doi: 10.1089/aid.1992.8.1887. [DOI] [PubMed] [Google Scholar]
  • 33.Mellors J, Kingsley L, Rinaldo C, Todd J, Hoo B, Kokka R, Gupta P. Quantitation of HIV-1 RNA in plasma predicts outcome after seroconversion. Ann Intern Med. 1995;122:573–579. doi: 10.7326/0003-4819-122-8-199504150-00003. [DOI] [PubMed] [Google Scholar]
  • 34.Mellors J W, Rinaldo C R, Gupta P, White R M, Todd J A, Kingsley L A. Prognosis in HIV-1 infection predicted by the quantity of virus in plasma. Science. 1996;272:1167–1170. doi: 10.1126/science.272.5265.1167. [DOI] [PubMed] [Google Scholar]
  • 35.Michael, N. (Walter Reed Army Institute of Research, Rockville, Md.). Personal communication.
  • 36.Myers G, Korber B, Hahn B, Jeang K T, Mellors J W, McCutchan F E, Henderson L E, Pavlakis G N. Human retrovirus and AIDS. Los Alamos, N.M: Los Alamos National Laboratory; 1995. A compilation of analysis of nucleic acid and amino acid sequences. [Google Scholar]
  • 37.Peeters M, Gueye A, Mboup S, Bibollet-Ruche F, Ekaza E, Mulanga C, Ouedrago R, Gandji R, Mpele P, Dibanga G, Koumare B, Saidou M, Esu-Williams E, Lombart J P, Badombena W, Luo N, van den Haesevelde M, Delaporte E. Geographical distribution of HIV-1 group O viruses in Africa. AIDS. 1997;11:493–498. doi: 10.1097/00002030-199704000-00013. [DOI] [PubMed] [Google Scholar]
  • 37a.Porter K R, Mascola J R, Hupudio H, Ewing D, VanColt T C, Anthony R L, Gorwin A L, Widodo S, Ertóno S, McCutchan F E, Burke D S, Hayes C G, Wignall F S, Graham R R. Genetic, antigenic and serologic characterization of human immunodeficiency virus type 1 from Indonesia. J Acquired Immune Defic Syndr Hum Retrovirol. 1997;14:1–6. doi: 10.1097/00042560-199701010-00001. [DOI] [PubMed] [Google Scholar]
  • 38.Rayfield M A, Sullivan P, Bandea C I, Britvan L, Otten R A, Pau C P, Pieniazek D, Subbarao S, Simon P, Schable C A, Wright A C, Ward J, Schochetman G. HIV-1 group O virus identified for the first time in the United States. Emerg Infect Dis. 1996;2:209–212. doi: 10.3201/eid0203.960307. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Revets H, Marissens D, De Wit S, Lacor P, Clumeck N, Lauwers S, Zissis G. Comparative evaluation of NASBA HIV-1 RNA QT, AMPLICOR-HIV Monitor, and QUANTIPLEX HIV RNA assay, three methods for quantification of human immunodeficiency virus type 1 RNA in plasma. J Clin Microbiol. 1996;34:1058–1064. doi: 10.1128/jcm.34.5.1058-1064.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Ruiz L, Romeu J, Clotet B, Balagué M, Cabrera C, Sirera G, Ibanez A, Martinez-Picado J, Raventos A, Tural C, Segura A, Foz M. Quantitative HIV-1 RNA as a marker of clinical stability and survival in a cohort of 302 patients with a mean CD4 cell count of 300 × 106/l. AIDS. 1996;10:F39–F44. doi: 10.1097/00002030-199609000-00001. [DOI] [PubMed] [Google Scholar]
  • 41.Saag M S, Holodniy M, Kuritzkes D R, O’Brien W A, Coombs R, Poscher M E, Jacobsen D M, Shaw G M, Richman D D, Volberding P A. HIV viral load markers in clinical practice. Nat Med. 1996;2:625–629. doi: 10.1038/nm0696-625. [DOI] [PubMed] [Google Scholar]
  • 41a.Salminen M O, Johansson B, Sonnerborg A, Ayehunie S, Gotte D, Leinikki P, Burke D S, McCutchan F E. Full-length sequence of an Ethiopian human immunodeficiency virus type 1 (HIV-1) isolate of genetic subtype C. AIDS Res Hum Retroviruses. 1996;12:1329–1339. doi: 10.1089/aid.1996.12.1329. [DOI] [PubMed] [Google Scholar]
  • 42.Schable C, Zekeng L, Pau C P, Hu D, Kaptue L, Gurtler L, Dondero T, Tsague J M, Schochetman G, Jaffe H, George J R. Sensitivity of United States HIV antibody tests for detection of HIV-1 group O infections. Lancet. 1994;344:1333–1334. doi: 10.1016/s0140-6736(94)90695-5. [DOI] [PubMed] [Google Scholar]
  • 43.Simon F, Loussert-Ajaka I, Damond F, Saragosti S, Barin F, Brun-Venizet F. HIV type 1 diversity in northern Paris. AIDS Res Hum Retroviruses. 1996;12:1427–1433. doi: 10.1089/aid.1996.12.1427. [DOI] [PubMed] [Google Scholar]
  • 44.Urdea, M. S. 1993. Direct and quantitative detection of HIV-1 RNA in human plasma with a branched DNA signal amplification assay. AIDS 7(Suppl. 2):S11–S14. [DOI] [PubMed]
  • 45.Vandamme A M, Schmit J C, Van Dooren S, Van Laethem K, Gobbers E, Kok W, Goubau P, Witvrouw M, Peetersmans W, De Clercq E, Desmyter J. Quantification of HIV-1 RNA in plasma: comparable results with the NASBA HIV-1 RNA QT and AMPLICOR HIV Monitor test. J Acquired Immune Defic Syndr Hum Retrovirol. 1996;13:127–139. doi: 10.1097/00042560-199610010-00003. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES