. 2021 Sep 4;101(4):115536. doi: 10.1016/j.diagmicrobio.2021.115536

Abbott ID now COVID-19 assay performance: a year in review

Phyu M Thwe a, Elphas Maiyo b, Ping Ren a,
PMCID: PMC8418095  PMID: 34634713

As new commercial SARS-CoV-2 assays have received Emergency Use Authorization (EUA) from U.S Food and Drug Administration (FDA) throughout the pandemic, some existing assays have also been updated for improved sensitivity and specificity. While several studies have assessed the initial performance of new testing platforms and/or assays during the early phase of the pandemic, there is very little information about the reassessment of any test systems. In 2020, we reported the performance characteristics of ID NOW COVID-19 assay (Abbott, Lake Forest, IL), in comparison to those of RT-PCR platforms (Thwe and Ren, 2020). Here, we analyzed the data gathered within the past year to reassess the ID NOW COVID-19 assay performance.

While our 2020 report was based on a limited sample size, the data collected for this study represented the testing performed from May 2020 through April 2021, with a total of 3320 paired nasopharyngeal swabs (NPS). The dry NPS were first tested by ID NOW, and second testing was performed on 1 of the RT-PCR/TMA platforms (GeneXpert Infinity, Hologic Panther Fusion or Hologic Panther TMA) within 2-18 hours of recollection of NPS in viral and/or universal transport media (VTM/UTM) after initial testing by ID NOW. A major difference between this study and our 2020 report was that our institution imposed a time limitation (1 hour maximum) for testing on ID NOW, as recommended by the manufacturer (Abbott ID NOW COVID-19 Package Insert. 2020. https://www.fda.gov/media/136525/download). However, in general, we did not reject the samples that could not be tested within the 1 hour time frame. Therefore, there were 23.5% of total samples in this study were tested by ID NOW beyond 1 hour of collection.

The overall percent agreement (OPA) between all RT-PCR/TMA platforms against ID NOW was 92.3%, with 70.3% positive percent agreement (PPA) and 95.3% negative percent agreement (NPA) (kappa value 0.64, 95% CI: 0.60 to 0.68) (Table 1 A). A breakdown of OPA, PPA, and NPA for each platform (GeneXpert Infinity, Panther Fusion, and Hologic Panther TMA) against ID NOW is presented in Table 1B and D.

Table 1.

(A-D): 2 × 2 tables of (A) All RT-PCR/TMA methods versus ID NOW; (B) Xpert Xpress by GeneXpert Infinity versus ID NOW; (C) Panther Fusion versus ID NOW; and (D) Panther TMA versus ID NOW.

All Reference Methods
POS NEG TOTAL
ID NOW POS 282 138 420
NEG 119 2781 2900
TOTAL 401 2919 3320

Xpert Xpress RT-PCR
POS NEG TOTAL
ID NOW POS 56 23 79
NEG 41 700 741
TOTAL 97 723 820

Panther (Fusion) RT-PCR
POS NEG TOTAL
ID NOW POS 75 56 131
NEG 36 879 915
TOTAL 111 935 1046

Panther TMA
POS NEG TOTAL
ID NOW POS 151 59 210
NEG 42 1202 1244
TOTAL 193 1261 1454

POS = positive; NEG = negative; PPA = positive percentage agreement; NPA = negative percent agreement; OPA = overall percent agreement; TMA = Transcription mediated amplification).

(a)PPA: 70.3% (95% CI: 65.6 – 74.8%)

NPA: 95.3% (95% CI: 94.4 – 96.0%)

OPA: 92.3% (95% CI: 91.3 – 93.2%)

(b) PPA: 57.7% (95% CI: 47.3 – 67.7%)

NPA: 96.8% (95% CI: 95.3 – 98.0%)

OPA: 92.2% (95% CI: 90.1 – 93.9%)

(c) PPA: 67.6% (95% CI: 58.0 – 76.2%)

NPA: 94.0% (95% CI: 92.3 – 95.4%)

OPA: 91.2% (95% CI: 89.3 – 92.9%)

(d)PPA: 78.2% (95% CI: 71.7 – 83.8%)

NPA: 95.3% (95% CI: 94.0 – 96.4%)

OPA: 93.1% (95% CI: 91.6 – 94.3%)

Among the total of 257 discrepant results, 119 samples were ID NOW-negative and/or PCR-positive whereas 138 were ID NOW-positive and/or PCR-negative. Upon review of the cycle threshold (Ct) values of thirty-one ID NOW-negative and/or PCR-positive samples performed by GeneXpert Infinity Xpert Xpress SARS-CoV-2 assay, 98% of them had Ct values of high 30s to low 40s. Throughout the pandemic, the application of Ct values in determining COVID-19 disease severity has been vastly controversial among laboratory and clinical professionals (Pujadas et al., 2020; Rhoads et al., 2021; Westblade et al., 2020). While the prediction of viral loads based on Ct values is discouraged for qualitative assays (IDSA and AMP joint statement, 2021), there is still a relative relationship between the 2 factors. It is highly unlikely that any samples with Ct values near the detectable threshold of 1 platform will necessarily reproduce the same results on a different platform. On the contrary, ID NOW-positive and/or PCR-negative samples were not repeated on a different PCR platform. Since the sample collection requires 2 different sample types – dry NPS versus NPS in VTM/UTM, several variables could have contributed to discrepancies.

Although the OPA and NPA in the group of samples tested beyond 1 hour of collection by ID NOW are similar to those tested within 1 hour of collection, the PPA was significantly lower for the samples tested beyond 1 hour than within 1 hour of collection (Table 2 ). Statistical analysis showed a significant correlation between testing beyond an hour of collection and potentially false-negative results by ID NOW (Fisher's test, P < 0.05). These findings indicated that deviating from the recommended testing time likely contributed to the potential ID NOW false-negative results. One potential reason for the poor ID NOW sensitivity and/or PPA when the dry swabs were stored longer than 1 hour could be that, after swabbing the nasopharyngeal, the swabs not only carry viruses but also various cellular enzymes (such as proteinases and RNases) which can degrade the viruses and viral RNA.

Table 2.

Overall, positive, negative percentage agreements between ID NOW and RT-PCR/TMA within and beyond ONE hour of collection.

OPA (%) between ID NOW and RT-PCR/TMA PPA (%) between ID NOW and RT-PCR/TMA NPA (%) between ID NOW and RT-PCR/TMA
ID NOW performed within ONE hour of collection 91.7 74.9 94.4
ID NOW performed beyond ONE hour of collection 94.2 26.3 97.7

One caveat of this study is that we could not determine if ID NOW positive/RT-PCR/TMA negative samples (137 total) were true or false positive by repeat testing from the same samples since dry NPS were utilized on ID NOW. It is also likely that sampling error during the second collection or sample dilution in VTM/UTM for confirmation with RT-PCR/TMA testing could contribute to the negative RT-PCR/TMA results. Meanwhile, we observed a higher percentage of false-positive rate by ID NOW within 1 hour of collection (5.6%) than beyond 1 hour of collection (2.3%). One speculation is that cross-contamination might arise from testing under pressure to meet the goal of 1 hour.

With a robust sample size of 3320, we believe that the performance of ID NOW COVID-19 assay is better characterized, in reference to our previous report. Performance agreement of all RT-PCR/TMA assays versus ID NOW as well as that of Fusion versus ID NOW had increased (Table 1). Overall, we believe that the performance of ID NOW has improved by adhering to the manufacturer's instruction of testing within 1 hour of sample collection. This information would be extremely helpful to clinical laboratories in strategic planning to assure adequate testing in the upcoming respiratory season.

Funding

This project did not receive any funding support from any agencies in the public, commercial, or not-for-profit sectors.

Declaration of competing interest

All authors in this manuscript declared no conflict of interest. All authors have seen and approved this manuscript.

References

  1. Abbott ID NOW COVID-19 Package Insert. 2020. [5th October 2021]. Available https://www.fda.gov/media/136525/download.
  2. IDSA and AMP joint statement. [24th September 2021]. Available at: https://www.idsociety.org/globalassets/idsa/public-health/covid-19/idsa-amp-statement.pdf
  3. Pujadas E, Chaudhry F, McBride R, Richter F, Zhao S, Wajnberg A, et al. SARS-CoV-2 viral load predicts COVID-19 mortality. Lancet Respir Med. 2020;8(9):e70. doi: 10.1016/S2213-2600(20)30354-4. Epub 2020 Aug 6. PMID: 32771081; PMCID: PMC7836878. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Rhoads Daniel, Peaper David R, She Rosemary C, Nolte Frederick S, Wojewoda Christina M, Anderson Neil W, et al. College of American pathologists (CAP) microbiology committee perspective: caution must be used in interpreting the cycle threshold (Ct) value. Clin Infect Dis. 2021;72(Issue 10):e685–e686. doi: 10.1093/cid/ciaa1199. [DOI] [PubMed] [Google Scholar]
  5. Thwe Phyu M, Ren Ping. How many are we missing with ID NOW COVID-19 assay using direct nasopharyngeal swabs? Findings from a mid-sized academic hospital clinical microbiology laboratory. Diagn Microbiol Infect Dis. 2020;98(2) doi: 10.1016/j.diagmicrobio.2020.115123. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Westblade LF, Brar G, Pinheiro LC, Paidoussis D, Rajan M, Martin P, et al. SARS-CoV-2 viral load predicts mortality in patients with and without cancer who are hospitalized with COVID-19. Cancer Cell. 2020;38(5) doi: 10.1016/j.ccell.2020.09.007. 661-671.e2Doi:Epub 2020 Sep 15. PMID: 32997958; PMCID: PMC7492074. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Diagnostic Microbiology and Infectious Disease are provided here courtesy of Elsevier