Abstract
Introduction
Mammography is one of the most technically demanding radiographic processes, and mammography quality assurance initiatives are priorities in the implementation of public health screening services. In the optimisation of image quality (IQ), radiographers play a major role. Between 1998 and 2009, the steering committee for mammography of a large population‐based screening service in northern Italy undertook several audit checks of the imaging facilities. In 2009, the target age range of the screening service was extended. The mammogram volume was projected to increase steeply but with no substantial increase in the radiographer workforce.
Methods
In view of the potential impact on mammographic IQ, the passive audit approach was abandoned in favour of an active radiographer‐oriented IQ review programme. Its technical basis consists of regularly repeated rounds of review of random samples of digital mammograms performed by each first‐level radiographer and by more experienced local reference radiographers, with IQ classification, followed by a training effort and a monitoring work.
Results
The mammogram volume grew from 140,822 in 2008 to 319,394 in 2014 (+127%) and then stabilised. In 2012, the proportion of mammograms with a poor IQ rose from 0.6% to 19.3%, paralleled by a substantial decrease of mammograms interpreted to have a moderate and perfect IQ. Conversely, a generalised improvement occurred in both rounds of 2016 and in the first round of 2018.
Conclusion
In the new challenging scenario, the programme proved to be effective. A successful IQ review initiative is one that encourages radiographers to participate with a positive and confident attitude.
Keywords: Image quality, mammography, quality assurance, radiographer, screening, training
A regional multicentre image quality review programme for mammography was developed consisting of a process of review of a sample of mammograms with image quality classification, independently performed by all first‐level radiographers and by more experienced local reference radiographers, followed by a training effort and a monitoring work. The programme places emphasis on motivating and training the radiographers and not on testing them. Its results demonstrate that a successful image quality review initiative for radiographers is one that encourages them to participate with a positive and confident attitude.

Introduction
Early detection of breast cancer through mammography screening relies on the radiologist’s ability to recognise subtle changes in the mammographic image. Because these changes are only perceptible with high‐quality imaging,1, 2 mammography is one of the most technically demanding radiographic processes.3 A mammogram is of high quality when it enables the radiologist to discern the presence or absence of the mammographic features of breast cancer with high sensitivity and specificity,3 and when it yields adequate diagnostic information with the least possible radiation exposure to the breast.1, 4
Mammography quality assurance programmes aim at increasing the probability that these two requirements are met.1 In 2006, the fourth edition of the European guidelines for quality assurance in breast cancer screening and diagnosis established the principle that quality assurance initiatives, audit and training are priorities in the implementation of public health screening services.5
In the optimisation of mammographic image quality (IQ), radiographers play a major role, because they position the patients, determine the proper voltage and exposure time for each mammogram and assist the radiologist in special procedures. In this article, we report on an Italian regional radiographer‐oriented mammographic IQ review programme. We illustrate the ideas and concepts that underlie the programme, the objectives, the design, the actors and their roles, the implementation and the results.
Materials and Methods
Setting and background
In the Emilia‐Romagna Region (northern Italy), a two‐yearly, two‐view, double‐read mammography screening service for women aged 50–69 years was established in the second half of the 1990s.6 According to data of 2018, the programme is served by 47 imaging facilities and 79 digital mammography systems, including 68 direct radiography systems and 11 computed radiography systems. The transition to digital mammography was completed in 2010–2011. There are 107 dedicated and 89 non‐dedicated radiographers, for a total 196. In Italy, both male and female radiographers are employed.
Between 1998 and 2009, the regional steering committee for screening mammography, that is composed of radiologists, radiographers, medical physicists and epidemiologists, undertook several onsite audit visits to the imaging facilities in order to review their technical characteristics. IQ, breast dosimetry and mammography sensitivity, with evaluation of incidence and radiologic review of interval breast cancers7, were particularly addressed. In 2009, the target age range of the screening service was extended to 45–74 years.8 The increase in mammography examinations was projected to be over 100%. However, no substantial adjustment of the radiographer workforce was expected. In anticipation of likely adverse effects on mammographic IQ, the audit approach was abandoned in favour of an active IQ review programme.
In its present form and in its entirety, the programme has been introduced in late 2016. Between 2010 and 2015, it has been preceded by schemes that had increasing similarities to the current model.
Rationale
The programme conforms to the 2006 European guidelines for quality assurance in breast cancer screening and diagnosis.9 According to these, three radiographer’s characteristics influence IQ, namely: training, experience and – important to note – motivation.9 In fact, there are neither recognised nor obvious ways to increase a radiographer’s motivation. Whether this key guideline is implemented or not across Europe is virtually unknown. Building radiographers’ motivation is one of the expected effects of this programme.
In brief, the protocol consists of a process of mammography review and IQ classification followed by a training effort and a monitoring work. The programme has a four‐stage design (Fig. 1):
-
1.
Twice a year, each first‐level radiographer, that is, a radiographer who processes the routine screening workload, reviews a set of digital mammograms performed by him/herself. This enables him/her to critically reflect on his/her own performance and to find increased motivation;
-
2.
Individual training needs are identified by a more experienced local reference radiographer with a review of the same mammograms and an evaluation of the discrepancies;
-
3.
Retraining activities are planned and carried out; and
-
4.
The impact of these efforts is monitored.
Figure 1.

Technical scheme depicting the four‐stage design of the Emilia‐Romagna Region mammographic image quality review programme.
The ideas linking these stages are that:
Training is more effective if tailored to the educational needs of a motivated radiographer9 and
This assumption should be confirmed, because monitoring the effects of training allows to identify additional educational needs.
The programme has not administrative functions. It is not designed to set up a monitoring system for central control purposes, to rank the radiographers, to certify their competence and to collect single screening centre’s data for accreditation purposes.
Actors and roles
The actors of the programme are connected in a pyramidal structure, which facilitates a fluid communication and ensures the standardisation of procedures. The actors include the following:
The programme head, who develops the methods and acts as a scientific guarantor;
The programme coordinator, who is responsible for coordinating the agenda;
The programme monitor, who receives the IQ assessment forms from the local reference radiographers and generates standard statistics;
The local reference radiographers (two per screening unit), who occupy an intermediate position between the apical level and first‐level radiographers;
The radiographers’ working group, which is composed by all of the above actors, and meets twice a year to discuss the results and plan the educational interventions; and
The first‐level radiographers, who select the samples of mammograms, perform part of the review work and participate in training activities.
Baseline training
The first implementation of the programme is preceded by a set of initiatives aimed at providing all radiographers with basic training and instructions. Events included are as follows: a preparatory region‐wide meeting, targeted to the directors of the screening units and all radiographers; a centralised theoretical course of four to eight hours, targeted to local reference radiographers; and local practical courses in which groups of three first‐level radiographers at once are trained to the use of the IQ assessment form.
First round
The programme has an annual duration, is repeated each year and is divided in two rounds. In the first round, each first‐level radiographer in each imaging facility selects by him/herself a systematic sample of five bilateral two‐view mammography examinations, for a total of 20 digital mammograms, which must have been performed consecutively in the middle of an arbitrarily selected working day with average‐high workload, and must have been obtained in the standard craniocaudal and mediolateral oblique views. Among these mammograms, those from patients with pacemaker, breast implants, breast scars and large breasts requiring multiple views are excluded and replaced with the next closest mammogram.
The evaluation of mammographic IQ is subjective and prone to variations in the individual perception of observers.10 To overcome this problem, standard assessment tools have been developed.11, 12, 13 With these instruments, the review of images takes place against a set of well‐established and widely accepted criteria. The IQ criteria used in this programme for reviewing the mammograms are listed in Table 1.
Table 1.
Image quality criteria used for reviewing the mammograms in the Emilia‐Romagna Region mammographic image quality review programme, with the view(s) of the breast to which they apply and the weight that is specifically assigned to each technical error (criterion not met) to reflect its importance for the accuracy of the diagnostic process.
| Code* | Relevant view of the breast | Image quality criterion† | Weight |
|---|---|---|---|
| A | CC, MLO | Nipple in profile | 1.00 |
| B | CC | Inframammary fold well demonstrated in internal quadrants and/or retromammary spaces | 0.50 |
| B1 | CC | Inframammary fold well demonstrated in external quadrants and/or retromammary spaces | 0.50 |
| C | CC, MLO | Symmetrical images | 0.10 |
| D | CC, MLO | Absence of overlying artefacts (skin folds altering/obscuring the breast tissue, shoulder, nose, etc.) | 0.50 |
| D1 | CC, MLO | Absence of small artefacts | 0.10 |
| E | CC, MLO | Breast compression ≥5 decaNewtions‡ | 0.25 |
| F | CC | Absence of internal/external glandular rotation | 0.25 |
| F | MLO | Absence of upper/lower glandular rotation | 0.25 |
| G | MLO | Pectoral muscle shadow to the nipple level, full width of the pectoral muscle | 1.00 |
| H | MLO | Complete visibility of the submammary sulcus | 0.75 |
| H1 | MLO | Partial visibility of the submammary sulcus profile | 0.25 |
CC, craniocaudal; MLO, mediolateral oblique.
Criterion identification code being indicated for clarity in all materials of the programme.
For each criterion, the definition of the technical standard is based on the 2006 European guidelines for quality assurance in breast cancer screening and diagnosis.14
The decaNewton is a unit of force in the SI equal to 10 newtons. The amount of breast compression is automatically recorded on the mammogram or can be retrieved from the archiving and communications system (PACS) server at the imaging facility.
A facsimile copy of the IQ assessment form, filled and translated in English, is shown in Figure 2. The middle section should provide the radiographer’s evaluation of the mammographic IQ criteria, whose simplified names are written vertically, for both views of the breast and both breasts. The number 0 indicates that the criterion has been met, 1 indicates that it has not been met. All numbers 1 are multiplied by a weight (Table 1) varying from 0.10 to 1.00, which reflects its importance for the accuracy of the diagnostic process. The column headed ‘Score’ shows the sum of weighted numbers. The column headed ‘Image quality category’ shows the categorisation of the sum of weighted numbers, that is, the final classification of IQ (perfect indicates a sum of 0.00–0.10; excellent indicates 0.20–1.05; good indicates 1.10–2.50, moderate indicates 2.55–3.95, poor indicates 4.00–4.95 and inadequate with need for repeat examination indicates ≥5.00). The validity of this tool relies on its close similarity with validated instruments that are in use in other countries.11, 12, 13
Figure 2.

Facsimile of a filled image quality assessment form, translated in English, being used in the Emilia‐Romagna Region mammographic image quality review programme. CR, computed radiography; DR, direct radiography; CC, craniocaudal; MLO, mediolateral oblique; R, right; L, left.
Each set of 20 mammograms undergoes two independent reviews: one performed by the radiographer him/herself and the other by one of the two local reference radiographers. Two separate IQ assessment forms are used. The initial self‐review serves, in the first place, radiographer‐oriented purposes: to foster his/her professional growth; to improve his/her awareness of the technical determinants of IQ; and to provide him/her with an opportunity to critically reflect on his/her performance at regular intervals.
The forms are compared by the local reference radiographer. In the case of discordant interpretations, the mammograms are reviewed by the programme head. The results of reviews have two distinct uses: determining the level of technical performance of the first‐level radiographer and identifying his/her errors in interpreting the level of IQ.
The two forms are loaded in Excel files and sent to the programme monitor, who applies the appropriate weights, calculates the total IQ score of each mammogram and classifies the overall IQ into five categories (Fig. 2). This approach is derived from the IQ grading system proposed by the UK Mammography Trainers Group.15
The programme monitor generates standard statistics. The data are analysed descriptively. The main outcome measure in data analysis is the overall regional per cent distribution by IQ classification obtained in the round. This is directly compared with a standard distribution derived from the radiographic quality objectives put forth by the 2006 European guidelines for quality assurance in breast cancer screening and diagnosisis.9 The standard distribution is as follows:
Inadequate, repeat needed, <1%;
Poor + inadequate, repeat needed, <3%;
Moderate, <12%;
Perfect + excellent + good, >85%;
Perfect + excellent + good + moderate, >97%.
The results are discussed in the first annual meeting of the radiographers’ working group in order to plan the appropriate educational interventions.
Second round
In each screening centre, the educational interventions include the following: one‐two meetings, targeted to all radiographers (including local reference radiographers) in order to discuss with them of their performance and of the most common technical errors; and an advanced course in which the IQ of the mammograms is reviewed interactively.
While first‐level radiographers undergo these educational interventions, the two local reference radiographers select a new systematic sample of five mammography examinations per first‐level radiographer, according to the same criteria as in the first round, and perform their review. In the case of discordant opinions, the mammograms are evaluated on‐site by the programme head as an arbitration reading.
The programme monitor processes the new set of assessment forms and evaluates the changes occurring between the first and the second round. The results of both rounds of the programme are presented in a meeting of all staff engaged.
The Ethics Committee at the Romagna Cancer Institute of Meldola, Forlì, Italy was asked for ethics approval. The Committee granted a waiver as ethics approval is not required, because this project involves established and regular educational practices that are done according to a law from an Italian Regional Government and are targeted at institutional health care staff as part of the employment duties.
Results
In the 47 imaging facilities involved in the regional screening service, the annual number of two‐view bilateral screening mammography examinations grew from 140,822 in 2008 (target age, 50–69 years) to 319,394 in 2014 (45–74 years) (+127%), unbalanced by an equal adjustment of the radiographer workforce and then roughly stabilised.
Table 2 shows the regional IQ distribution of the mammograms reviewed in 2010, 2012, 2016 and 2018 as compared with the quality objectives set by the 2006 European guidelines for breast cancer screening and diagnosis.9 In 2012, the proportion of mammograms with a poor IQ rose from 0.6% to as much as 19.3%. This was paralleled by a substantial decrease of mammograms interpreted to have a moderate and, more important, a perfect level of IQ. Conversely, a generalised improvement was observed in both rounds of 2016 and, to an even greater extent, in the first round of 2018. In the second round of 2018, two opposite (albeit limited) variations occurred, that is, the ≥good category dropped slightly below the standard level of >85.0%, and the inadequate (repeat needed) category exceeded 2.0% versus a standard of <1.0%. These two observations warrant monitoring. Overall, however, the programme proved to be effective in maintaining the levels of IQ needed to meet current European standards.
Table 2.
Participating screening centres and radiographers, number of mammography examinations and mammograms selected for review and image quality distribution in 2010, 2012, 2016 (after the programme assumed the current form) and 2018 in the Emilia‐Romagna Region mammographic image quality review programme.
| 2010* | 2012† | 2016, 1st round | 2016, 2nd round | 2018, 1st round | 2018, 2nd round | |
|---|---|---|---|---|---|---|
| Participating radiographers, n | 181 | 182 | 187 | 195 | 183 | 171 |
| Participating screening centres, n | 12 | 11 | 11 | 11 | 11 | 9 |
| Mammography examinations reviewed, n | 905 | 910 | 935 | 975 | 915 | 855 |
| Mammograms reviewed, n | 3620 | 3640 | 3740 | 3900 | 3660 | 3420 |
| IQ distribution, n (%) | ||||||
| Inadequate (repeat needed) (standard, <1.0%)† | 0 (0.0) | 0 (0.0) | 8 (0.9) | 12 (1.2) | 10 (1.1) | 23 (2.7) |
| Poor | 5 (0.6) | 176 (19.3) | 23 (2.5) | 19 (1.9) | 29 (3.2) | 32 (3.7) |
| Moderate | 309 (34.1) | 170 (18.7) | 138 (14.8) | 118 (12.1) | 93 (10.2) | 103 (12.0) |
| Good | 369 (40.8) | 319 (35.1) | 366 (39.1) | 376 (38.6) | 35 (38.8) | 327 (38.2) |
| Excellent | 0 (0.0) | 173 (19.0) | 318 (34.0) | 367 (37.6) | 367 (40.1) | 318 (37.2) |
| Perfect | 222 (24.5) | 72 (7.9) | 82 (8.8) | 83 (8.5) | 61 (6.7) | 52 (6.1) |
| Categorised IQ distribution, % | ||||||
| Poor + inadequate (standard, <3.0%)‡ | 0.6 | 19.3 | 3.3 | 3.2 | 4.3 | 6.4 |
| Perfect + excellent + good (standard, >85.0%)‡ | 64.9 | 60.3 | 78.6 | 90.3 | 91.6 | 77.9 |
| Perfect + excellent + good + moderate (standard, >97.0%)‡ | 99.4 | 80.7 | 96.7 | 96.8 | 95.7 | 93.6 |
IQ, image quality.
One round only was conducted in order to accurately test the early protocol.
One round only was conducted because part of the Emilia‐Romagna Region was hit by an earthquake.
The standards for IQ distribution are the radiographic quality objectives put forth by the 2006 European guidelines for quality assurance in breast cancer screening and diagnosis.9
Discussion
This programme fulfils some important theoretical requirements for an effective IQ review initiative.16 With its background and long‐standing experience, the steering committee offers a specific professional leadership to the participating radiographers. The committees operate on behalf of the Regional Department of Health and works in a supportive environment.
The experience gained in 2010–2015 was used to refine the programme model. In this way, the initiative has become more effective, more sustainable and more acceptable for participants. In its present form, the programme has been implemented in late 2016. Since then, it has been exported to five more administrative regions and healthcare districts of northern and southern Italy. The protocol has proved to be feasible in screening settings other than that of origin.
On an international scale, the variety of approaches proposed for IQ assessment in the area of mammography screening suggests that a widely accepted method to monitor and optimise the performance of radiographers has not been developed yet.1, 17 It remains necessary to disseminate and evaluate the methods and the results of all ongoing experiences. The programme described here is not designed to set up a central control system. It places the emphasis on motivating and training the radiographers and not on testing them or ranking them or certifying their competence. This strategy made it possible to face a rapid growth of over 125% in mammogram volume – a sort of natural experiment – with only a transient decrease in IQ.
Conclusion
In the new challenging scenario, the programme proved to be effective. The key to this result was that the programme places emphasis on motivating and training the radiographers. A successful IQ review initiative is one that encourages radiographers to participate with a positive and confident attitude.
Conflict of Interest
The authors declare non‐conflict of interest.
Acknowledgements
The authors thank the Emilia‐Romagna Region radiographers’ working group. The membership of the working group is as follows (in parentheses, the name of the health care district of origin identifying the local screening unit): Vania Galli, Monica Pini, Fiorella Vecchiè, Elisabetta Aldrovandi (Modena); Maria Cristina Buzzetti, Barbara Cavedo (Piacenza); Erika Spaggiari, Alessandra Cuttone (Parma); Alessandra Ronzoni, Monica Canovi (Reggio Emilia); Giuseppe Di Bari, Silvia Landini, Stefano Trazzi, Silvia Salimbeni (Bologna); Marco Campomori (Imola); Daniele De Metrio, Annalisa Marigo, Lara Marzocchi, Stefania Polastri (Ferrara); Raffaella Obbi, Paola Baga (Ravenna); Daniela Severi, Denis Mariotti (Forlì); Donatella Castagnoli, Stefania Moretti (Cesena); and Anna Flenghi, Barbara Renzi and Cristiana Piastra (Rimini). The authors also thank Adriana Giannini, Carlo Naldoni, Alba Carola Finarelli (Department of Health, Regional Administration, Regione Emilia‐Romagna, Bologna), Stefano Ferretti (University of Ferrara and Local Health Authority, Ferrara), Paola Golinelli (Medical Physics Service, Local Health Authority, Modena), Valeria Bellelli and Ennio Gallo (Mammography Screening Centre, Local Health Authority, Modena, Italy) and Marco Zappa (Italian National Centre for Screening Monitoring). The authors wish to extend their gratitude and appreciation to the radiographers who participated in the programme.
J Med Radiat Sci. 68(2021) 253–259
References
- 1.Reis C, Pascoal A, Sakellaris T, Koutalonis M. Quality assurance and quality control in mammography: a review of available guidance worldwide. Insights Imaging 2013; 4: 539–53. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Jayadevan R, Armada MJ, Shaheen R, Mulcahy C, Slanetz PJ. Optimizing digital mammographic image quality for full‐field digital detectors: artifacts encountered during the QC process. Radiographics 2015; 35: 2080–9. [DOI] [PubMed] [Google Scholar]
- 3.Caldwell CB, Fishell EK, Jong RA, Weiser WJ, Yaffe MJ. Evaluation of mammographic image quality: pilot study comparing five methods. AJR Am J Roentgenol 1992; 159: 295–301. [DOI] [PubMed] [Google Scholar]
- 4.Ng KH, Jamal N, DeWerd L. Global quality control perspective for the physical and technical aspects of screen‐film mammography: image quality and radiation dose. Radiat Prot Dosimetry 2006; 121: 445–51. [DOI] [PubMed] [Google Scholar]
- 5.Executive summary. In: Perry N, Broeders M, de Wolf C , Törnberg S, Holland R, von Karsa L (eds). European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th ed. Office for Official Publications of the European Communities, Luxembourg, 2006; 5–14. [Google Scholar]
- 6.Bucchi L, Falcini F, Baraldi GP, et al. Integrating self‐referral for mammography into organised screening: results from an Italian experience. J Med Screen 2003; 10: 134–8. [DOI] [PubMed] [Google Scholar]
- 7.Bucchi L, Ravaioli A, Foca F, Colamartini A, Falcini F, Naldoni C. Incidence of interval breast cancers after 650,000 negative mammographies in 13 Italian health districts. J Med Screen 2008; 15: 30–5. [DOI] [PubMed] [Google Scholar]
- 8.Sassoli de Bianchi P, Ravaioli A, Ferretti S, et al. Estensione dell'età‐bersaglio del programma di screening mammografico e governo della pratica mammografica in Emilia‐Romagna. Epidemiol Prev 2017; 41: 38–45. [DOI] [PubMed] [Google Scholar]
- 9.Rijken H, Caseldine J, Laird O. Radiographical guidelines. In: Perry N, Broeders M, de Wolf C, Törnberg S, Holland Rvon Karsa L (eds). European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th edn. Office for Official Publications of the European Communities, Luxembourg, 2006; 167–80. [Google Scholar]
- 10.Grahn A, Hemdal B, Andersson I, et al. Clinical evaluation of a new set of image quality criteria for mammography. Radiat Prot Dosimetry 2005; 114: 389–94. [DOI] [PubMed] [Google Scholar]
- 11.NHS Breast Screening Programme Radiographers Quality Assurance Coordination Committee . Quality Assurance Guidelines for Radiographers, 2nd edn. NHS Breast Screening Programme, Sheffield, 2000. [Google Scholar]
- 12.The National Quality Management Committee of BreastScreen Australia . BreastScreen Australia National Accreditation Standards. BreastScreen Australia, Canberra, 2008. [Google Scholar]
- 13.The National Breast Screening Programme . Guidelines for Quality Assurance in Mammography Screening, 4th edn. National Screening Service, Dublin, 2015. [Google Scholar]
- 14.van Engen R, van Woudenberg S, Bosmans H, Young K, Thijssen M. European protocol for the quality control of the physical and technical aspects of mammography screening: screen‐film mammography. In: Perry N, Broeders M, de Wolf C, Törnberg S, Holland Rvon Karsa L (eds). European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th edn. Office for Official Publications of the European Communities, Luxembourg, 2006; 61–104. [Google Scholar]
- 15.NHS Breast Screening Programme . Quality Assurance Guidelines for Radiographers. NHS Breast Screening Programme, Sheffield, 1994. [Google Scholar]
- 16.Bowling A. Research Methods in Health: Investigating Health and Health Services, 2nd edn. Open University Press, Buckingham, 2002. [Google Scholar]
- 17.Klabunde C, Bouchard F, Taplin S, Scharpantgen A, Ballard‐Barbash R. Quality assurance for screening mammography: an international comparison. J Epidemiol Community Health 2001; 55: 204–12. [DOI] [PMC free article] [PubMed] [Google Scholar]