Abstract
OBJECTIVES
Driveline infections continue to be a significant complication following left ventricular assist device (LVAD) implantation. Driveline exit-site care is crucial for the prevention of infections; however, there are no uniform guidelines. The goal of this study was to provide an overview of the currently published driveline exit-site care protocols in patients with LVAD.
METHODS
A systematic literature review was performed. Studies before 15 December 2020 were included if the number of driveline infections was a primary outcome and the driveline exit-site care protocol was explained.
RESULTS
Eleven articles were included in the systematic review, including 1602 patients with LVADs. The median of the frequency of driveline infections in the articles was 13.8% with a range of 0–52.6%. There was a marked variability in the methods of care of driveline exit sites, without a standardized driveline dressing technique in patients with LVADs. The frequency of driveline infections was 6–7.5% in studies using a dressing kit that included chlorhexidine, a silver-based dressing and an anchoring device. Furthermore, there was variability in the anchoring devices and the frequency of dressing changes, which varied from daily to weekly. No specific anchoring device or change frequency was found to be superior.
CONCLUSIONS
Based on this systematic review, driveline exit care protocols that included chlorhexidine, a silver-based dressing, the use of an anchoring device and dressing kits might be best in reducing driveline infection rates. However, prospective studies with larger cohorts are needed to establish the optimal protocol for driveline exit-site care.
Keywords: Left ventricular assist device, Dressing, Driveline, Driveline infection, Protocols, Care
INTRODUCTION
Driveline infections continue to be a significant complication following left ventricular assist device (LVAD) implants and are a limiting factor to successful long-term LVAD support [1, 2]. The International Society for Heart and Lung Transplantation Registry for Mechanically Assisted Circulatory Support reported driveline infection rates as high as 29% after LVAD implantation for 3 months [2]. The Interagency Registry for Mechanically Assisted Circulatory Support reported that driveline infections occur in ∼19% of recipients of LVAD by 12 months after implant [3].
The driveline exit site is frequently the entry site of pathogens that may cause local infection, and these infections could track to the pocket and the pump. Therefore, driveline infections increase the risk for pump/cannula, pocket and bloodstream infections [2]. We know that care of the driveline exit site is paramount for the prevention and treatment of driveline infections [4, 5]. Despite the importance of its care, there are few specific recommendations for the management of the LVAD driveline exit site [4, 6–8]. Research on driveline exit-site care has shown that driveline exit-site management is not standardized, resulting in a wide variety of management protocols among LVAD centres [9]. The goal of this systematic review was to provide an overview of the currently published driveline exit-site care protocols in patients with an LVAD.
MATERIALS AND METHODS
Search strategy
This systematic literature review was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines [10]. The search strategy was developed with a librarian for inclusion sensitivity. A literature search was performed using Embase, Medline Ovid, Cinahl EBSCOhost, Web of Science Core Collection, Cochrane Central register of trials and Google scholar databases using the following search terms: ‘left ventricular assist device’, ‘ventricular assist device’, ‘heart assist device’, ‘driveline’, ‘wound care’, ‘infection prevention’, ‘driveline infection’, ‘device infection’, ‘exit site’, ‘wound infection’, ‘wound care’, ‘care’, ‘wound management’, ‘dressing’, ‘protocol’.
Study selection
Two reviewers (Z.O.K. and Y.C.Y.) independently reviewed potentially eligible studies for evaluation. Titles and abstracts were examined for possible inclusion before the full-text versions of the remaining articles were obtained. All authors were involved in the final selection of and data extraction from included articles. Any discrepancies among the authors regarding inclusion were resolved by consensus among all authors. Full-text clinical research articles written in English and published before 15 December 2020 were included in the systematic review. Studies were included if a driveline exit-site care protocol was explained, and if driveline-related and specific infections were a primary outcome. Case reports, review articles, animal studies and conference abstracts were excluded. In addition, studies were excluded if they only discussed surgical interventions for care management, if they examined pulsatile flow devices or if fewer than 30 patients were included.
Data extraction
Data that were extracted included study and LVAD characteristics, sample size, follow-up time, device type and strategy, definition criteria of driveline infections and perioperative antibiotic prophylaxis. The primary outcomes reported in the studies were driveline care protocols and driveline infections. We also evaluated time to first infection, infection relapse rates and microorganisms causing driveline infections.
Quality assessment
The quality of each article was appraised using the Strengthening the Reporting of Observational Studies in Epidemiology Statement checklist (Table 1). A higher score from the checklist indicates higher quality [19].
Table 1:
Summary of baseline characteristics of the articles in the systematic literature review
| Article/country | Study design | Study period | Sample size | LVAD characteristics | Patient characteristics | Definition of LVAD infection | Perioperative antibiotic prophylaxis | STROBE score |
|---|---|---|---|---|---|---|---|---|
|
Schlöglhofer et al. (2020) Austria [11] |
Retrospective analysis, single centre | January 2013–July 2017 | Cohort n = 183 |
Device: HM II, HM 3, HW Device strategy: BTC: 35.5%, BTT: 32.8%, DT: 27.9%, BTR: 0.5% |
Mean age: 58.0 years BMI: 26.6 DM: 32.7% Length of support: N/R |
ISHLT | N/R | 19 |
|
Lander et al. (2018) USA [8] |
Retrospective cohort study, single centre | January 2010–October 2015 |
Intervention n = 92 Control n = 61 |
Device: HM II, HW Device strategy: BTC: 59%, DT: 41% |
Mean age: 58 years 1 BMI: 29.7 DM: 43.8% Length of support: N/R |
INTERMACS ISHLT |
N/R | 20 |
|
Aburjania et al. (2017) USA [12] |
Retrospective cohort study, single centre | November 2006–September 2015 |
Intervention n = 120 Control n = 163 |
Device: HM II Device strategy: N/R |
Mean age: 59 years BMI: 29.1 DM: N/R Length of support: 713 days |
Institution’s standard definition | Yes | 20 |
|
Durand et al. (2017) USA [13] |
Retrospective cohort study, single centre | August 2005–December 2014 |
Intervention n = 65 Control n = 19 |
Device: N/R Device strategy: BTT: 82%, DT: 18% |
Mean age: 55 years BMI: 25.8 DM: 22% Length of support: 370 days |
ISHLT | Yes | 20 |
|
Cagliostro et al. (2016) USA [14] |
Prospective cohort study, single centre | January 2009–December 2013 |
Intervention n = 159 Control n = 107 |
Device: HM II, HW Device strategy: BTT: 67%, DT: 33% |
Mean age: 58.2 years BMI: N/R DM: 34.2% Length of support: N/R |
INTERMACS | N/R | 18 |
|
Son et al. (2017) USA [15] |
Retrospective analysis, single centre | January 2009–December 2013 |
CHG group n = 37 PVP-I group n = 7 |
Device: HM II Device strategy: N/R |
Mean age: 60 years BMI: 25.0 DM: 36.4% Length of support: 521 days |
INTERMACS | Yes | 19 |
|
Stahovich et al. (2016) USA [7] |
Prospective study, multicentre | N/R | Cohort n = 50 |
Device: HM II Device strategy: N/R |
Mean age: 62 years BMI: N/R DM: 25% Length of support: 496 days |
N/R | N/R | 18 |
|
Menon et al. (2015) Germany [16] |
Retrospective cohort study, single centre | January 2008–April 2011 |
Intervention n = 31 Control n = 17 |
Device: HM II Device strategy: BTT/BTC/BTR = 77.5%, DT = 22.5% |
Mean age: 58.1 years BMI: 27.2 DM: 29% Length of support: 198 days |
The Cleveland Clinic Classification | Yes | 20 |
|
Stulak et al. (2013) USA [17] |
Retrospective cohort study, single centre | February 2007–September 2011 |
Intervention n = 144 Control n = 141 |
Device: HM II Device strategy: BTT 59%, DT 41% |
Mean age: 54 years BMI: N/R DM: N/R Length of support: N/R |
N/R | Yes | 18 |
|
Sharma et al. (2012) USA [4] |
Retrospective analysis, single centre | January 2007–January 2011 | Cohort n = 143 |
Device: HM II Device strategy: BTT 39%, DT 61% |
Mean age: 61.3 years BMI: 30.8 DM: 33.6% Length of support: N/R |
ISHLT | Yes | 18 |
|
Hozayen et al. (2012) USA [18] |
Retrospective cohort study, single centre | N/R |
Utah protocol n = 16 Minnesota protocol n = 47 |
Device: HM II, HW, VentrAssist Device strategy: N/R |
Mean age: 57.1 years BMI: 29.7 DM: 39.6% Length of support: 483 days |
ISHLT | N/R | 15 |
Categorical values are reported as n (%).
BMI: body mass index (kg/m2); BTC: bridge to candidacy; BTR: bridge to recovery; BTT: bridge to transplant; CHG: chlorhexidine gluconate; DM: diabetes mellitus; DT: destination therapy; HM: HeartMate; HW: HeartWare; INTERMACS: Interagency Registry for Mechanically Assisted Circulatory Support; ISHLT: International Society for Heart and Lung Transplantation; LVAD: left ventricular assist device infection; N/R: not reported; PVP-I: povidone-iodine; STROBE: Strengthening the Reporting of Observational Studies in Epidemiology Statement.
Statistical analyses
Data from the articles were analysed with SPSS 22.0 (SPSS Inc., Chicago, IL, USA). Categorical variables were presented in numbers and percentages. Continuous variables were presented as median and mean. A meta-analysis could not be performed due to the substantial heterogeneity of the exit-site care methods reported in the studies.
RESULTS
A total of 846 articles were identified through the literature search after duplicates had been removed and were assessed by title and abstract for the review. The full texts of 47 articles were reviewed based on the selection criteria, and 36 were excluded based on full manuscript assessment. Eleven articles fulfilled the inclusion criteria, underwent quality assessment and were included in the final review (Fig. 1).
Figure 1:
Flow diagram summarizing the review process.
Nine articles included studies conducted in USA: one study was conducted in Germany and one in Austria. The definitions of LVAD driveline infections used in the article are shown in the Supplementary Table 2. Nine articles were retrospective cohort studies and 2 were prospective studies (Table 1). There was a substantial range in types of dressing methods and of management care methods discussed in the articles (Table 2). Three of the studies evaluated the entire care protocol, whereas others compared cleaning agents, covering materials, showering properties and the use of a dressing kit (Table 3). There was no obvious change over time in the driveline exit-care protocols that made a substantial difference. Interagency Registry for Mechanically Assisted Circulatory Support and International Society for Heart and Lung Transplantation criteria were used for definitions of driveline infections in 7 of the articles. Table 1 summarizes the final articles and the Strengthening the Reporting of Observational Studies in Epidemiology Statement scores. The mean number of patients described in the 11 articles was 145 patients with LVAD (range 44–285 patients). The LVAD strategies mentioned in the articles were bridge to transplant/candidacy/recovery (63%) and destination therapy (37%). The median of the driveline infection frequencies reported in the articles was 13.8%, with a range of 0–52.6% in a follow-up of 6–44 months in this cohort. The causative microorganisms of driveline infections were reported in 8 out of 11 articles. The organisms reported as the most common causes of driveline infections were Staphylococcus aureus and Pseudomonas aeruginosa (Table 3). The types of the devices were HeartMate (HM) II and 3, HeartWare (HW) and VentrAssist in the included articles; HM II in 6, HM II and HW in 2, HM II, 3 and HW HVAD in 1 and, HM II, HW and VentrAssist in one study. In one study, the device type was not reported (Table 1). Readmission for a driveline infection was evaluated according to the device type in one study. Those authors reported that patients with an HM 3 had a higher risk for driveline infection readmissions compared to those with an HW HVAD or an HM II [11].
Table 2:
Summary of driveline exit-site dressing change methods described in the articles and at the Erasmus University Medical Center
| Article | Cleaning agent | Dressing closure | Anchoring device | Dressing change frequency | Showering | Sterile technique | Use of a dressing kit |
|---|---|---|---|---|---|---|---|
| Schlöglhofer et al. (2020) [11] | Octenidine solution | Absorptive non-adherent slit compress (Askina Pad 5 × 5 cm) | Secutape Nanoplast fixation | Twice or 3 times weekly | N/R | Yesa | N/R |
| Lander et al. (2018) [8] | CHG and saline solution |
Control: sterile gauze pad dressing (the Telfa island occlusive sandwich dressing) Intervention: the Kendal fenestrated foam dressing |
Control: Centurion Foley holder Intervention: Centurion secure view port window |
Weekly | N/R | Yesb | N/R |
| Aburjania et al. (2017) [12] |
CHG PVP-I (if there is a skin irritation) |
N/R | N/R | N/R | Not allowed, exit site must be kept dry (using a hand-held shower attachment to wash their hair and lower extremities) | Yesc | N/R |
| Durand et al. (2017) [13] | Polymyxin-trimethoprim solution (polymyxin B 10 000 units/ml + trimethoprim 1 mg/ml) | N/R | N/R | Once or twice weekly | Allowed once per week when the exit site has healed | N/R | N/R |
| Cagliostro et al. (2016) [14] |
Control: CHG 2% Intervention: CHG 2% and 70% isopropyl alcohol |
Control: sterile gauze pad dressing Intervention: silver-based dressing |
Control: bio-occlusive dressing, binder or StatLock device Intervention: centurion Foley holder | N/R | N/R | N/R | Yes |
| Son et al. (2017) [15] |
CHG group: CHG 2% PVP-I group: povidone-iodine 10% |
Sterile gauze sponge dressing | Centurion Foley holder | Three times weekly | Allowed with occlusive covering of the driveline exit site | Yesc | Yes |
| Stahovich et al. (2016) [7] | CHG 3.15%, saline solution and isopropyl alcohol 70% |
Silver-based dressing SorbaView ultimate dressing |
Centurion Foley holder | Weekly | Allowed with occlusive covering of the driveline exit site | Yesd | Yes |
| Menon et al. (2015) [16] |
Control: octenidine dihydrochloride 0.1% Intervention: merbromin 2% |
Silver-based dressing and sterile gauze pad dressing | Hollister plate stabilizer | Every 5–6 days | N/R | Yesb | N/R |
| Stulak et al. (2013) [17] | CHG 4% | N/R | N/R | N/R | N/R | Yesc | N/R |
| Sharma et al. (2012) [4] | CHG and saline swabs | Sterile gauze pad dressing | Abdominal binder | Daily | N/R | Yesb | N/R |
| Hozayen et al. (2012) [18] |
Utah protocol: N/R Minnesota protocol: soap and antimicrobial spray |
Utah protocol: foam-based dressing Minnesota protocol: sterile gauze pad dressing |
N/R |
Utah protocol: every third day Minnesota protocol: daily |
N/R | Yesc | N/R |
| Erasmus University Medical Centre | CHG 4% | Silver-based and sterile gauze pad dressing | Hollister plate stabilizer | Two times a week | Allowed once per week when the exit site healed | Yese | No |
Using sterile dressing closure, face mask, hair cover and sterile gloves.
Using sterile dressing closure, face mask and sterile gloves.
Using sterile dressing closure.
Using sterile dressing closure, face mask, hair cover and non-latex gloves.
Using sterile dressing closure and sterile gloves.
CHG: chlorhexidine gluconate; N/R: not reported; PVP-I: povidone-iodine.
Table 3:
Summary of results of the articles included in the systematic literature review
| Article | Intervention/implementation | Median follow-up time (months) | Driveline infection rate | Time to first infection (days) | Infection relapse rate | Microorganisms |
|---|---|---|---|---|---|---|
| Schlöglhofer et al. (2020) [11] | To characterize risk factors for DLI readmission 2 years postimplant | 24 | 27.3% (50/183) | N/R | N/R |
S. aureus 56% (32/57)a P. aeruginosa 12% (7/57)a Others 32% (18/57)a |
| Lander et al. (2018) [8] |
Control: historical LVAD patients (using Telfa Island dressing) Intervention: fenestrated hydrophilic foam dressing |
Control: 23.8 Intervention: 39.1 |
Control: 21.3% (13/61) Intervention: 7.6% (7/92) (P = 0.032) |
N/R | 45% (9/20) |
P. aeruginosa 14%,aProteus 11%a CoNS 7%,a Others 25%a |
| Aburjania et al. (2017) [12] |
Control: historical LVAD patients Intervention: no conventional showers and keep the exit site dry while bathing |
Control: 44.4 Intervention: 13.2 |
Driveline infection (P = 0.06) Control: 42% (69/163) Intervention: 14 % (17/120) Pseudomonas infection (P = 0.077) Control: 9% (15/163) Intervention: 1% (1/120) |
347 (entire sample) | N/R |
P. aeruginosa Control: 17% (15/86) Intervention: 1.1% (1/86) |
| Durand et al. (2017) [13] |
Control: no topical antibiotics Intervention: topical polymyxin trimethoprim solution |
12.3 |
Control: 52.6 (10/19) Intervention: 13.8% (9/65) (P = 0.001) |
Control: 164 Intervention: 320 |
37% (7/19) |
P. aeruginosa: 37% (7/19) MSSA 16% (3/19) Skin flora: 26% (5/19) Others: 21% (4/19) |
| Cagliostro et al. (2016) [14] |
Control: historical LVAD patients (using gauze pad dressing and not using a standard kit) Intervention: using a standard kit for dressing (includes silver-based gauze dressing and a standard anchoring device) |
Control: 8.7 Intervention: 11.6 |
Control: 15.8% (17/107) Intervention: 7.5% (12/159) |
Control: 154 Intervention: 181 |
Control: 65% (11/17) Intervention: 50% (6/12) |
N/R |
| Son et al. (2017) [15] |
CHG group: using CHG for cleaning PVP-I group: using povidone-iodine in patients with CHG intolerance for cleaning |
17.3 |
CHG group= 5.4% (2/37) PVP-I group= 42.9% (3/7) (P = 0.02) |
336 (entire sample) | N/R |
CHG group: Acinetobacter 50% (1/2) Stenotrophomonas 50% (1/2) PVP-I group: S. aureus 100% (3/3) |
| Stahovich et al. (2016) [7] | Using the percutaneous lead management kit for dressing | 6 | 6% (3/50) | 180 | N/R | N/R |
| Menon et al. (2015) [16] |
Control: using octenidine solution for cleaning Intervention: using Merbromin solution for cleaning |
Control: 6.7 Intervention: 6.5 |
Control: 11.8% (2/17) Intervention: 0% (P = 0.043) |
Control: 130.5 Intervention: infection free |
N/R | S. aureus 100% (2/2) |
| Stulak et al. (2013) [17] |
Control: sterile dressing changes without continued (long-term) prophylactic antibiotics Intervention: sterile dressing changes with continued (long-term) prophylactic antibiotics |
Control: 11 Intervention: 12.3 |
Control: 13% (19/141) Intervention: 18% (26/144) (P = 0.15) |
N/R | N/R | Staphylococci (no rate) |
| Sharma et al. (2012) [4] | Experience with the management of driveline infections (sterile dressing changes with CHG and saline application, without prophylactic oral antibiotics) | 11.3 | 12% (18/143) | 182 | 22% (4/18) |
CoNs: 44.5% (4/9b) S. aureus: 33.3% (3/9b) P. aeruginosa: 22.2% (2/9b) |
| Hozayen et al. (2012) [18] |
Utah protocol: foam-based dressing Minnesota protocol: Gauze-based dressing |
18 |
Utah protocol: 19% (3/16) Minnesota protocol: 13% (6/47) (P = 0.68) |
N/R | N/R | N/R |
The study reported cultures with multiple organisms.
The study reported microorganisms for 9 cases.
CHG: chlorhexidine gluconate; CoNS: coagulase-negative staphylococci; DLI: driveline infection; LVAD: left ventricular assist device; MRSA: methicillin-resistant Staphylococcus aureus; MSSA: methicillin-sensitive Staphylococcus aureus; N/R: not reported; PVP-I: povidone-iodine; P. aeruginosa: Pseudomonas aeruginosa; S. aureus: Staphylococcus aureus.
Driveline exit-site cleaning agents
The most frequently used cleaning agent was chlorhexidine gluconate (CHG) (n = 7) (Table 2) [4, 7, 8, 12, 14,15, 17]. CHG was used with the saline solution in 3 studies [4, 7,8]. CHG characteristics (saline or alcohol-based) were not reported in the articles. In 2 studies, if the patient had a skin irritation or CHG intolerance, povidone-iodine was used as an alternative cleaning agent [12, 15]. Driveline infection frequency differed among the studies, with a range of 5.4–21.3% in the studies using CHG as a cleaning agent (Table 3). Studies using CHG and a silver-based dressing for the care of driveline exit sites reported a driveline infection frequency of 6–7.5% [7, 14]. In the studies using CHG and a sterile gauze dressing for exit-site care, driveline infection frequencies were between 5.4% and 21.3% [4, 8, 14, 15]. Son et al. [15] reported a higher driveline infection frequency when povidone-iodine was used as an alternative cleaning agent in patients intolerant to CHG (42.9%). Durand et al. [13] evaluated the effect of topical polymyxin-trimethoprim (poly) solution on driveline infections. The driveline infection rate was reported in 9 patients (13.8%) in the group using the poly solution and in 10 patients (52.6%) in the non-poly group (P = 0.001). Menon et al. [16] compared merbromin with octenidine solutions for cleaning the driveline exit site. The frequency of driveline infections in patients using octenidine was 11.8%; no infections were found in patients using a merbromin solution. In the study by Schlöglhofer et al. [11], octenidine was used as a cleaning agent and the driveline infection frequency was reported to be 27.3%. Hozayen et al. [18] reported a driveline infection frequency of 13% when soap and antimicrobial spray were used to clean the driveline exit site.
Dressing materials for the driveline exit site
Sterile gauze pads (n = 5) and silver-based dressings (n = 3) were the most commonly used covering materials for dressing the driveline exit site (Table 2). Cagliostro et al. [14] compared driveline dressing protocols using silver-based dressings with those using a sterile gauze dressing. The frequency of driveline infections (7.5%) in the silver-based group was lower than that in the group using sterile gauze (15.8%). In the study by Stahovich et al. [7], silver-based dressings and CHG were used for the care of the driveline exit site. With this protocol, driveline infection frequency and the time to first infection were 6% and 180 days, respectively. Two studies evaluated driveline exit-site dressing protocols that included foam-based dressings [8, 18]. Hozayen et al. [18] compared foam and sterile gauze dressings for covering the driveline exit site. Driveline infection frequency was reported to be 19% for foam-based dressings and 13% for sterile gauze dressings (P = 0.68). Lander et al. [8] used foam-based dressings and CHG for dressings; the frequency of driveline infections was 7.6%. Schlöglhofer et al. [11] reported a driveline infection rate of 27.3% when absorptive non-adherent compresses and octenidine were used to care for the driveline exit site.
Anchoring devices used for stabilization of the driveline
For immobilization of the driveline, the Centurion Foley holder, Hollister plate stabilizer, abdominal binder, Centurion secure view port and Secutape Nanoplast fixation were used (Table 2). The most frequently used anchoring device for stabilization of the driveline was the Centurion Foley holder in 4 studies [7, 8, 14, 15]. In 2 studies, the driveline exit-site dressing protocol included the Centurion Foley holder, a silver-based dressing and CHG. These studies reported a driveline infection frequency and a time to first infection of 6–7.5% and 180 days, respectively [7, 14]. The driveline care protocol in 2 studies included the Centurion Foley holder, sterile gauze dressing and CHG. These studies reported driveline infection frequencies 5.4–21.3% [8, 15]. The Hollister plate stabilizer was used to immobilize the driveline in one study; the reported frequency of driveline infections was 0–11.8% [16]. The study that used an abdominal binder for immobilizing the driveline reported a driveline infection rate of 12% [4]. In the study of Schlöglhofer et al. [11], Secutape Nanoplast fixation was used; the reported driveline infection frequency was 27.3%.
Frequency of driveline exit-site dressing change
The frequencies of dressing changes differed among the studies and varied from daily to weekly (Table 2) [4, 7, 8, 11, 13, 15, 18]. Two studies reported weekly dressing changes in driveline exit-site care protocol. Stahovich et al. [7] evaluated the effect of using a dressing kit and a weekly dressing change in their study; they reported 6% driveline infection frequency and 180 days to first driveline infection. Lander et al. [8] compared weekly fenestrated foam dressings and weekly occlusive sterile gauze dressings. Whereas the driveline infection frequency in the weekly fenestrated foam dressing group was 7.6%, an infection frequency of 21.3% was reported in the weekly occlusive sterile gauze dressing group. In 2 studies with daily dressing changes, driveline infection frequencies of 12–13% were reported [4, 18]. Hozayen et al. [18] reported a 3-times-weekly foam-based dressing change and a 19% driveline infection frequency in their study. In the study of Son et al. [15], a 3-times-weekly dressing change protocol that included CHG and povidone-iodine as the cleaning agent was evaluated. The study reported a driveline infection frequency of 5.4% in the CHG group and 42.9% in the povidone-iodine group. In the study of Schlöglhofer et al. [11], a 2- to 3-times-weekly dressing change with octenidine solution and 27.3% driveline infection frequency were reported.
Showering strategies for patients with left ventricular assist devices
Showering strategies in the driveline exit-site care protocols were reported in 4 studies (Table 2) [7, 12, 13, 15]. Aburjania et al. [12] investigated the effect of abstaining from conventional showers and keeping the driveline exit site dry. Driveline infection frequency and Pseudomonas infection frequency were 14% and 1% in the intervention group and 42% and 9% in the control group, respectively. An occlusive covering over the driveline exit-site dressing during showering was used in 2 studies; the driveline infection frequencies reported were between 5.4% and 42.9% [7, 15].
Using a kit for driveline exit-site dressing
The utilization of a kit for driveline exit-site dressing was reported in 2 studies (Table 2) [7, 14]. Cagliostro et al. [14] compared a group that used a standard kit that included a silver-based dressing and an anchoring device with an historical control group. Driveline infection frequency and time to infection were 7.5% and 181 days in the standard dressing kit group and 15.8% and 154 days in the historical control group, respectively. In the study of Stahovich et al. [7], the use of a percutaneous lead management kit for dressing was evaluated. They reported a driveline infection frequency and time to infection of 6% and 180 days, respectively.
DISCUSSION
In this systematic review, we found a marked variability in the care protocols for LVAD driveline exit sites and no standardized driveline dressing technique. But, driveline exit care protocols including chlorhexidine, silver-based dressings, the use of an anchoring device and dressing kits might be the best for reducing driveline infection rates.
CHG appeared to be the most commonly used cleaning agent for driveline care. CHG has a broad spectrum of activity against gram-positive, gram-negative non-spore-forming bacteria, yeast and selective lipid envelope viruses [20, 21]. Furthermore, CHG is considered to be advantageous in the care because of poor absorption from the skin and no evidence of systemic accumulation and adverse events [20]. Additionally, CHG has already been proposed as an effective agent for the prevention of surgical site infections [20, 21]. Unfortunately, data on the concentration of CHG, time of evaporation and whether it was saline or alcohol-based or not, were not available in the included studies. In case of CHG intolerance, povidone-iodine was used for driveline care [12, 15]. Povidone-iodine is an effective bactericidal solution against gram-positive and gram-negative organisms and does not delay healing. However, the absorption of iodine from the skin is a disadvantage [22]. Studies comparing the efficacy of CHG and povidone-iodine in surgical site cleaning and prevention of infections show the superiority of CHG [23, 24]. The use of merbromin in driveline exit-site care is controversial because it is a toxic agent due to the brome and mercury content and is therefore prohibited in many countries [16, 25]. Octenidine, which has been used frequently as an antiseptic in recent years, is another cleaning agent used in driveline care. It is recommended for use in prophylactic antisepsis because it is not absorbed by the skin and mucosa, and it is well tolerated and suitable for topical use [26, 27]. In addition, a polymyxin-trimethoprim solution was effective in the prevention of driveline infections. However, no other studies suggested the use of either of these solutions in LVAD care. Based on the foregoing discussion, the standardized driveline care protocol should include CHG as the main cleaning and antiseptic agent, because it is advantageous over other solutions in terms of both cost and effectiveness [23]. Alternatively, the use of octenidine or povidone-iodine solutions in CHG-intolerant patients may be suggested.
In this systematic review, sterile gauze and silver-based dressings were the most frequently used materials to cover the driveline exit site. The use of sterile gauze dressings in a non-infected dry exit site that completed the healing process can be cost-effective in the driveline exit-site care protocol [28]. In our analysis, the silver-based dressing was more effective. The use of a silver-based dressing was recommended for preventing colonization and improving healing [29, 30]. However, some researchers assumed the evidence for silver-based dressings in the prevention of infections to be insufficient and the costs too high [31, 32]. Foam-based dressing is generally recommended for exudate wounds due to its absorbent property and is not recommended for use in dry wounds [28, 33]. According to the results of our study and those reported in the literature, wound characteristics should be taken into consideration when choosing the covering material in driveline care protocol. Therefore, the use of a silver-based dressing only in the first 6 months after LVAD implantation may be (cost) effective in preventing driveline infections [34]. Silver or foam-based dressings may be preferred depending on whether the exit site has infection or exudate.
An anchoring device is one method for preventing driveline exit-site trauma and is thereby effective in reducing the driveline exit-site infections [34–37]. In our study, there was great variability in the anchoring devices used for the driveline stabilization. However, no clear data for the superiority of any one of the anchoring devices were found.
The frequency of changing a driveline exit-site dressing varies considerably according to the institution [9, 36, 38]. None of the particular dressing change frequencies was more effective than another in the prevention of driveline infections. The study by Wus et al. [6], not included in the systematic review because of the short follow-up time and inclusion of hospitalized patients, reported that dressing change frequency had no effect on driveline infection frequencies. In determining the optimal dressing change frequency, driveline exit-site features and whether there is infection should be taken into consideration [34, 36]. Daily dressing change is recommended until the driveline exit site heals completely for effective exit-site cleaning and preventing wet dressings. Once the driveline exit site has healed and there is no drainage, a lower dressing change frequency may be feasible and safe and also increases caregiver and patient satisfaction [18, 36].
In the management of driveline exit-site care, keeping the driveline exit site dry should be considered in the prevention of driveline exit-site infections. Therefore, showers are recommended only after the driveline exit site has healed completely in patients with LVAD [34, 36]. Keeping the driveline exit site as dry as possible during the shower and changing the dressing immediately after the shower may be effective in preventing driveline infections, in particular Pseudomonas infections.
The driveline care protocol requires the use of many different materials. This systematic review suggests that using a dressing kit might be effective in reducing the driveline infections. The use of a dressing kit in driveline care can be effective in increasing patient compliance and reducing infection frequency [7].
Clinical implications
Taking the findings of this systemic review into account, driveline exit-site care is best performed within a standardized protocol, using sterile dressing materials and sterile gloves and a dressing kit (Fig. 2, Supplementary Table 1). CHG should be used for driveline exit-site cleaning; octenidine or povidone-iodine should be used in case of CHG intolerance. The properties of the exit site should be taken into consideration when choosing the covering material. A silver-based dressing can be used, particularly in the first months after implantation. An anchoring device should be used to prevent driveline exit-site trauma. The driveline exit site should be kept as dry as reasonably possible. The dressing change frequency can be based on the properties of the exit site. For a dry exit site that has completed the healing process, the dressing change frequency can be once or twice a week. The proposed changes, despite the increased cost for materials and agents (e.g. silver-based dressings), may significantly reduce frequent readmissions and long hospitalizations.
Figure 2:
Recommendations based on the systematic review.
Limitations
This systematic review has limitations that should be considered when interpreting the results. The studies included in the systematic review were mostly retrospective cohort studies. In addition, the studies had small sample sizes and did not compare the same exit-site care protocols. Due to the substantial heterogeneity in the exit-site care methods of the studies, a meta-analysis could not be performed, and the centre-specific findings of the systematic review are perhaps not generalizable. Furthermore, the studies had fewer destination therapy patients and hence may under-represent these patients.
CONCLUSION
Based on this systematic review, driveline exit-site care protocols including chlorhexidine, silver-based dressings, use of an anchoring device and dressing kits might be best in reducing the frequency of driveline infections. However, no strong evidence for a standardized driveline exit-site protocol exists. Prospective studies with larger cohorts are needed to establish the optimal protocol for LVAD driveline exit-site care.
SUPPLEMENTARY MATERIAL
Supplementary material is available at EJCTS online.
Supplementary Material
ACKNOWLEDGEMENTS
The authors wish to thank Wichor Bramer from the Erasmus MC Medical Library for developing and updating the search strategies.
Conflict of interest: none declared.
ABBREVIATIONS
- CHG
Chlorhexidine gluconate
- HM
HeartMate
- HW
HeartWare
- LVAD
Left ventricular assist device
Author contributions
Zeliha Ozdemir Koken: Conceptualization; Data curation; Formal analysis; Investigation; Methodology; Project administration; Resources; Software; Validation; Visualization; Writing—original draft; Writing—review & editing. Yunus C. Yalcin: Investigation; Methodology; Resources; Software; Writing—original draft. Diana van Netten: Investigation; Methodology; Resources; Supervision; Writing—review & editing. Chantal C. de Bakker: Investigation; Resources; Writing—review & editing. Maaike van der Graaf: Investigation; Resources; Writing—review & editing. Umit Kervan: Methodology; Resources; Supervision; Writing—review & editing. Nelianne J. Verkaik: Conceptualization; Methodology; Resources; Supervision; Writing—review & editing. Kadir Caliskan: Conceptualization; Data curation; Investigation; Methodology; Project administration; Resources; Software; Supervision; Writing—original draft; Writing—review & editing.
Reviewer information
European Journal of Cardio-Thoracic Surgery thanks Massimiliano Meineri and the other, anonymous reviewer(s) for their contribution to the peer review process of this article.
REFERENCES
- 1.Kirklin JK, Pagani FD, Kormos RL, Stevenson LW, Blume ED, Myers SL. et al. Eighth annual INTERMACS report: special focus on framing the impact of adverse events. J Heart Lung Transplant 2017;36:1080–6. [DOI] [PubMed] [Google Scholar]
- 2.Hannan MM, Xie R, Cowger J, Schueler S, de By T, Dipchand AI. et al. Epidemiology of infection in mechanical circulatory support: a global analysis from the ISHLT Mechanically Assisted Circulatory Support Registry. J Heart Lung Transplant 2019;38:364–73. [DOI] [PubMed] [Google Scholar]
- 3.Goldstein DJ, Naftel D, Holman W, Bellumkonda L, Pamboukian SV, Pagani FD. et al. Continuous-flow devices and percutaneous site infections: clinical outcomes. J Heart Lung Transplant 2012;31:1151–7. [DOI] [PubMed] [Google Scholar]
- 4.Sharma V, Deo SV, Stulak JM, Durham LA 3rd, Daly RC, Park SJ. et al. Driveline infections in left ventricular assist devices: implications for destination therapy. Ann Thorac Surg 2012;94:1381–6. [DOI] [PubMed] [Google Scholar]
- 5.Hieda M, Sata M, Seguchi O, Yanase M, Murata Y, Sato T. et al. Importance of early appropriate intervention including antibiotics and wound care for device-related infection in patients with left ventricular assist device. Transplant Proc 2014;46:907–10. [DOI] [PubMed] [Google Scholar]
- 6.Wus L, Manning M, Entwistle JW 3rd. Left ventricular assist device driveline infection and the frequency of dressing change in hospitalized patients. Heart Lung 2015;44:225–9. [DOI] [PubMed] [Google Scholar]
- 7.Stahovich M, Sundareswaran KS, Fox S, Hallinan W, Blood P, Chen L. et al. Reduce driveline trauma through stabilization and exit site management: 30 days feasibility results from the multicenter RESIST study. ASAIO J 2016;62:240–5. [DOI] [PubMed] [Google Scholar]
- 8.Lander MM, Kunz N, Dunn E, Althouse AD, Lockard K, Shullo MA. et al. Substantial reduction in driveline infection rates with the modification of driveline dressing protocol. J Card Fail 2018;24:746–52. [DOI] [PubMed] [Google Scholar]
- 9.Wilcox JE, Cameron KA, Harap RS, Shanklin KL, Grady KL, Cohen ER. et al. Ventricular assist device driveline dressing-change protocols: a need for standardization. A report from the SimVAD Investigators. J Card Fail 2019;25:695–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Moher D, Liberati A, Tetzlaff J, Altman DG; PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. Open Med 2009;3:e123–30. [PMC free article] [PubMed] [Google Scholar]
- 11.Schlöglhofer T, Michalovics P, Riebandt J, Angleitner P, Stoiber M, Laufer G. et al. Left ventricular assist device driveline infections in three contemporary devices. Artif Organs 2020:1–9. doi:10.1111/aor.13843. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Aburjania N, Sherazi S, Tchantchaleishvili V, Alexis JD, Hay CM.. Stopping conventional showering decreases Pseudomonas infections in left ventricular assist device patients. Int J Artif Organs 2017;40:282–5. [DOI] [PubMed] [Google Scholar]
- 13.Durand ML, Ennis SC, Baker JN, Camuso JM, McEachern KM, Kotton CN. et al. Topical polymyxin-trimethoprim prophylaxis may decrease the incidence of driveline infections in patients with continuous-flow left ventricular assist devices. Artif Organs 2017;41:169–75. [DOI] [PubMed] [Google Scholar]
- 14.Cagliostro B, Levin AP, Fried J, Stewart S, Parkis G, Mody KP. et al. Continuous-flow left ventricular assist devices and usefulness of a standardized strategy to reduce drive-line infections. J Heart Lung Transplant 2016;35:108–14. [DOI] [PubMed] [Google Scholar]
- 15.Son AY, Stein LH, DeAnda A, Katz SD, Smith DE, Reyentovich A. et al. Impact of chlorhexidine gluconate intolerance on driveline infection during chronic HeartMate II left ventricular assist device support. Int J Artif Organs 2017;39:570–4. [DOI] [PubMed] [Google Scholar]
- 16.Menon AK, Baranski SK, Unterkofler J, Autschbach R, Moza AK, Goetzenich A. et al. Special treatment and wound care of the driveline exit site after left ventricular assist device implantation. Thorac Cardiovasc Surg 2015;63:670–4. [DOI] [PubMed] [Google Scholar]
- 17.Stulak JM, Maltais S, Cowger J, Joyce LD, Daly RC, Park SJ. et al. Prevention of percutaneous driveline infection after left ventricular assist device implantation: prophylactic antibiotics are not necessary. ASAIO J 2013;59:570–4. [DOI] [PubMed] [Google Scholar]
- 18.Hozayen SM, Soliman AM, Eckman PM.. Comparison of two ventricular assist device dressing change protocols. J Heart Lung Transplant 2012;31:108–9. [DOI] [PubMed] [Google Scholar]
- 19.von Elm E, Altman DG, Egger M, Pocock SJ, Gotzsche PC, Vandenbroucke JP; STROBE Initiative. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Lancet 2007;370:1453–7. [DOI] [PubMed] [Google Scholar]
- 20.Edmiston CE Jr, Bruden B, Rucinski MC, Henen C, Graham MB, Lewis BL.. Reducing the risk of surgical site infections: does chlorhexidine gluconate provide a risk reduction benefit? Am J Infect Control 2013;41:S49–55. [DOI] [PubMed] [Google Scholar]
- 21.Milstone AM, Passaretti CL, Perl TM.. Chlorhexidine: expanding the armamentarium for infection control and prevention. Clin Infect Dis 2008;46:274–81. [DOI] [PubMed] [Google Scholar]
- 22.Bigliardi PL, Alsagoff SAL, El-Kafrawi HY, Pyon JK, Wa CTC, Villa MA.. Povidone iodine in wound healing: a review of current concepts and practices. Int J Surg 2017;44:260–8. [DOI] [PubMed] [Google Scholar]
- 23.Lee I, Agarwal RK, Lee BY, Fishman NO, Umscheid CA.. Systematic review and cost analysis comparing use of chlorhexidine with use of iodine for preoperative skin antisepsis to prevent surgical site infection. Infect Control Hosp Epidemiol 2010;31:1219–29. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Darouiche RO, Wall MJ Jr, Itani KM, Otterson MF, Webb AL, Carrick MM. et al. Chlorhexidine-alcohol versus povidone-iodine for surgical-site antisepsis. N Engl J Med 2010;362:18–26. [DOI] [PubMed] [Google Scholar]
- 25.European Environmental Bureau Zero Mercury Working Group. Mercury in Products. 2019. https://www.zeromercury.org/resources/ (15 December 2020, date last accessed).
- 26.Hubner NO, Siebert J, Kramer A.. Octenidine dihydrochloride, a modern antiseptic for skin, mucous membranes and wounds. Skin Pharmacol Physiol 2010;23:244–58. [DOI] [PubMed] [Google Scholar]
- 27.Koburger T, Hubner NO, Braun M, Siebert J, Kramer A.. Standardized comparison of antiseptic efficacy of triclosan, PVP-iodine, octenidine dihydrochloride, polyhexanide and chlorhexidine digluconate. J Antimicrob Chemother 2010;65:1712–19. [DOI] [PubMed] [Google Scholar]
- 28.Dhivya S, Padma VV, Santhini E.. Wound dressings—a review. Biomedicine (Taipei) 2015;5:22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Sibbald RG, Contreras-Ruiz J, Coutts P, Fierheller M, Rothman A, Woo K.. Bacteriology, inflammation, and healing: a study of nanocrystalline silver dressings in chronic venous leg ulcers. Adv Skin Wound Care 2007;20:549–58. [DOI] [PubMed] [Google Scholar]
- 30.Leaper D.An overview of the evidence on the efficacy of silver dressings. J Wound Care 2011;20:8–14. [DOI] [PubMed] [Google Scholar]
- 31.Storm-Versloot MN, Vos CG, Ubbink DT, Vermeulen H.. Topical silver for preventing wound infection. Cochrane Database Syst Rev 2010;CD006478. doi:10.1002/14651858.CD006478.pub2. [DOI] [PubMed] [Google Scholar]
- 32.Michaels JA, Campbell B, King B, Palfreyman SJ, Shackley P, Stevenson M.. Randomized controlled trial and cost-effectiveness analysis of silver-donating antimicrobial dressings for venous leg ulcers (VULCAN trial). Br J Surg 2009;96:1147–56. [DOI] [PubMed] [Google Scholar]
- 33.Lee SM, Park IK, Kim YS, Kim HJ, Moon H, Mueller S. et al. Superior absorption and retention properties of foam-film silver dressing versus other commercially available silver dressing. Biomater Res 2016;20:22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Kusne S, Mooney M, Danziger-Isakov L, Kaan A, Lund LH, Lyster H. et al. An ISHLT consensus document for prevention and management strategies for mechanical circulatory support infection. J Heart Lung Transplant 2017;36:1137–53. [DOI] [PubMed] [Google Scholar]
- 35.Potapov EV, Antonides C, Crespo-Leiro MG, Combes A, Farber G, Hannan MM. et al. 2019 EACTS Expert Consensus on long-term mechanical circulatory support. Eur J Cardiothorac Surg 2019;56:230–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Bernhardt AM, Schloglhofer T, Lauenroth V, Mueller F, Mueller M, Schoede A. et al. ; Driveline Expert STagINg and carE DESTINE Study Group, a Ventricular Assist Device Driveline Infection Study Group. Prevention and early treatment of driveline infections in ventricular assist device patients—the DESTINE staging proposal and the first standard of care protocol. J Crit Care 2020;56:106–12. [DOI] [PubMed] [Google Scholar]
- 37.Baronetto A, Centofanti P, Attisani M, Ricci D, Mussa B, Devotini R. et al. A simple device to secure ventricular assist device driveline and prevent exit-site infection. Interact CardioVasc Thorac Surg 2014;18:415–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Schloglhofer T, Robson D, Bancroft J, Sorensen G, Kaufmann F, Sweet L. et al. International coordinator survey results on the outpatient management of patients with the HeartWare(R) ventricular assist system. Int J Artif Organs 2017;39:553–7. [DOI] [PubMed] [Google Scholar]