Performances of automated digital imaging of Gram-stained slides with on-screen reading against manual microscopy

Adrien Fischer 1, Nouria Azam 1, Lara Rasga 1, Valérie Barras 1, Manuela Tangomo 1, Gesuele Renzi 1, Nicolas Vuilleumier 2, Jacques Schrenzel 1,3, Abdessalam Cherkaoui 1,
PMCID: PMC8449764  PMID: 33963927

Abstract

The objective of this study was to evaluate the performances of the automated digital imaging of Gram-stained slides against manual microscopy. Four hundred forty-three identified Gram-stained slides were included in this study. When both methods agreed, we considered the results as correct, and no further examination was carried out. Whenever the methods gave discrepant results, we reviewed the digital images and the glass slides by manual microscopy to avoid incorrectly read smears. The final result was a consensus of multiple independent reader interpretations. Among the 443 slides analyzed in this study, 101 (22.8%) showed discrepant results between the compared methods. The rates of discrepant results according to the specimen types were 5.7% (9/157) for positive blood cultures, 42% (60/142) for respiratory tract specimens, and 22% (32/144) for sterile site specimens. After a subsequent review of the discrepant slides, the final rate of discrepancies dropped to 7.0% (31/443). The overall agreement between the compared methods and the culture results reached 78% (345/443) and 79% (349/443) for manual microscopy and automated digital imaging, respectively. According to culture results, the specificity for automated digital imaging and manual microscopy were 90.8% and 87.7% respectively. In contrast, sensitivity was 84.1% for the two compared methods. The discrepant results were mostly encountered with microorganism morphologies of rare occurrence. The results reported in this study emphasize that on-screen reading is challenging, since the recognition of morphologies on-screen can appear different as compared to routine manual microscopy. Monitoring of Gram stain errors, which is facilitated by automated digital imaging, remains crucial for the quality control of reported Gram stain results.

Keywords: Gram stain, Manual microscopy, Automated digital imaging, On-screen reading, Digitalization

Introduction

By any measure, this decade has been outstanding in the history of automation in clinical microbiology. Automation enabled not only to customize each analytical step but also to force the laboratory managers to concentrate all the pre-analytical steps onto a unique physical interface that has become the entry door to all further analytical activities of conventional bacteriology. Two automated systems are currently available for clinical specimens streaking and slides preparation: Inoculated media are loaded onto conveyors for transfer to automated incubators where cultures are imaged with high-resolution digital pictures at pre-defined times. Direct consequences of total laboratory automation can be measured as improved productivity, traceability, quality, and reduced turn-around times [16]. But despite the implementation of these novel technologies, some traditional techniques (e.g., Gram stain) continue to bear an important role in the diagnostic process. The Gram-stained smears remain important as a pre-analytical indicator of respiratory tract specimen quality (e.g., sputum), for presumptive etiologic diagnosis, to guide empirical therapy and to indicate the presence of mixed aerobic and anaerobic infections. The Gram stain has therefore been a cornerstone for clinical bacteriology laboratories for over a century, despite the subjectivity of the results interpretation (highly operator-dependent) and the manual nature of the staining process. The interpretation of Gram stain results continues to be labor-intensive, time-consuming, and strongly dependent on the quality of the samples. To face the increasing workload in clinical microbiology laboratories, automated slide scanning and imaging might provide several advantages and adequately complement manual testing. Nonetheless, many technical challenges should be overcome before Gram-stain automation can be systematically deployed in bacteriology. For instance, the quality of the staining remains strongly affected by the smear preparation (markedly different for a thick sputum or a biopsy versus for a blood culture or a body fluid). Therefore, Gram-stained smears can display tremendously variable and heterogeneous background staining, which can obviously affect the algorithm that may target areas without bacteria and miss the most relevant microscopic zones.

The overarching objective of this study was to assess the performances of automated digital imaging of Gram-stained slides with on-screen reading against manual microscopy. Automated digital imaging was performed by the Metafer slide scanning platform that permits scanning, digitalization, and archiving of slides automatically, even in a batch mode.

Materials and methods

Slide collection and workup

A total of 443 identified Gram-stained slides from positive blood cultures (n = 157), respiratory tract specimens (n = 142), and sterile site specimens (n = 144) were collected in the clinical bacteriology laboratory of Geneva University Hospitals between February and June 2020. All the slides included in this study were prepared by the Copan WASP® during the routine clinical workup. Importantly, the slides were chosen without any preselection (e.g., staining quality, abundance, or identity of the microorganisms), and they were not pre-screened by automated digital imaging in order to ideally capture the variability of routine specimens. One hundred additional slides, encompassing all the specimen types analyzed in this study, were used during the training period to validate and evaluate the different parameters of the Metafer slide scanning and the imaging platform, according to the manufacturer’s instructions (MetaSystems Hard & Software GmbH, Altlussheim, Germany). These slides were not included in the subsequent study period. Slides from respiratory tract and sterile site specimens were stained using a manual method. In contrast, all slides from positive blood cultures were stained using the PREVI® Color Gram (BioMérieux, Marcy L’Etoile, France). The slides analysis for each workflow was performed by four experienced laboratory technologists and two clinical microbiologists by rotating after a training period, in order to avoid any learning bias. Importantly, all were blinded from the results obtained using the other method.

Culture diagnostic workup

All specimens included in this study were processed on the WASPLab following the protocols previously published [13].

Discrepant results

The results of automated digital imaging with on-screen reading were compared to the manual microscopy. When both methods agreed, we considered the results as correct, and no further examination was carried out. Whenever the methods gave discrepant results (i.e., negative smear or one or more morphologies was/were not reported), we reviewed the digital images and the glass slides by manual microscopy to avoid incorrectly read smears. For the remaining discrepant slides, the Gram strain results were assessed against culture results (Fig. 1). The final result was a consensus of multiple independent reader interpretations.

Fig. 1.

Fig. 1

Algorithm of results assessment between automated digital imaging with on-screen reading and manual microscopy

Metafer slide scanning and imaging platform

In this study, we used a commercial off-the-shelf software. All the 443 slides included in this study plus the 100 slides used during the training period were imaged without coverslips using a Metafer slide scanning and imaging platform with a 160-slide-capacity automated slide loader equipped with × 10 and × 40 magnification objectives (Carl Zeiss AG, Oberkochen, Germany) and automatic random access of slides. The × 10 magnification was used for collecting images spanning the whole slide. Collected images were then stitched together to create a single digital picture of the Gram smear. Right after, 20 images were taken with seven focal planes for each such picture using a × 40 oil immersion lens magnification for defined areas, according to the manufacturer’s instructions. On-screen reading of such digital images was performed using the Metafer 5 software.

Results

Among the 443 slides analyzed in this study, 101 (22.8%) showed discrepant results between manual microscopy and automated digital imaging using on-screen reading. The rates of discrepant results according to the specimen types were 5.7% (9/157) for positive blood cultures, 42% (60/142) for respiratory tract specimens, and 22% (32/144) for sterile site specimens. A large part of incorrectly read slides were those where one or more microorganism morphologies was/were not reported. After a subsequent review of these discrepant slides, the rate of discrepant results dropped drastically to 7.0% (31/443). More specifically, the rates according to the specimen types became 0.6% (1/157) for positive blood cultures, 15% (21/142) for respiratory tract specimens, and 6.3% (9/144) for sterile site specimens (Table 1). As depicted in Table 2, the commonly missed microorganism morphologies by manual microscopy versus automated digital imaging with on-screen reading were Gram-negative rods (39% (12/31) versus 23% (7/31)), Gram-positive rods (6.5% (2/31) versus 19.4% (6/31)), and Gram-positive cocci (6.5% (2/31) versus 16.1% (5/31)) respectively. In contrast, less commonly missed microorganism morphologies were Gram-negative cocci ((0% versus 6.5% (2/31)) and yeasts ((0% versus 3.2% (1/31)). The discrepant results were seen mostly in microorganism morphologies of rare occurrence.

Table 1.

Rates of discrepant results broken by specimen types

Specimen types Total number of specimens Number of discrepant results Number of discrepant results after review of the digital images Number of discrepant results after review of the slides by manual microscopy Final agreement including reviewed slides (%)
Blood cultures 157 9 1 1 99.4
Respiratory tract specimens 142 60 46 21 85.2
Sterile site specimens 144 32 15 9 93.8
Total 443 101 62 31 93

Table 2.

Analysis of discrepant results

Specimen types Manual microscopy result Digital images result Culture result
Bronchoalveolar lavage Gram-positive cocci, and Gram-negative rods Gram-positive cocci Oropharyngeal flora
Bronchoalveolar lavage Gram-positive cocci Gram-positive cocci, and Gram-negative rods Oropharyngeal flora
Bronchoalveolar lavage Gram-positive cocci, Gram-negative rods, and Gram-negative cocci Gram-positive cocci, and Gram-negative rods Oropharyngeal flora
Bronchoaspiration Gram-positive cocci Negative smear Oropharyngeal flora
Bronchoaspiration Gram-positive cocci, and Gram-positive rods Gram-positive cocci Oropharyngeal flora
Bronchoaspiration Gram-positive cocci, and Gram-positive rods Gram-positive cocci, Gram-positive rods, and Gram-negative rods Oropharyngeal flora
Tracheal aspiration Gram-positive cocci, Gram-negative rods, and yeast Yeast Candida glabrata, oropharyngeal flora
Tracheal aspiration Gram-positive cocci Gram-negative rods Enterococcus faecium
Sputum Gram-positive cocci Negative smear Pseudomonas aeruginosa, and oropharyngeal flora
Sputum Gram-positive cocci, and Gram-negative rods Gram-positive cocci Enterobacterales, and oropharyngeal flora
Sputum Gram-positive cocci, and Gram-positive rods Gram-positive cocci, and Gram-negative rods Candida albicans, Enterobacteriales, and oropharyngeal flora
Sputum Gram-positive cocci, and yeast Gram-positive cocci, and Gram-negative rods Staphylococcus aureus, Candida albicans, and oropharyngeal flora
Sputum Gram-positive cocci, and Gram-negative rods Gram-positive cocci, Gram-positive rods, and Gram-negative rods Staphylococcus aureus, Enterobacterales, and oropharyngeal flora
Sputum Gram-positive cocci, and yeast Gram-positive cocci, Gram-negative rods, and yeast Pseudomonas aeruginosa, Candida albicans, and oropharyngeal flora
Sputum Gram-positive cocci, and Gram-positive rods Gram-positive cocci Oropharyngeal flora
Sputum Gram-negative cocci, and yeast Gram-negative rods, and yeast Candida albicans, and Enterobacterales
Sputum Gram-positive cocci, Gram-positive rods, Gram-negative rods, and yeast Gram-positive cocci, Gram-negative rods, and yeast Haemophilus influenzae, and oropharyngeal flora
Sputum Gram-positive cocci, Gram-negative rods, and yeast Gram-positive cocci, and yeast Candida glabrata, and oropharyngeal flora
Sputum Gram-positive cocci, and Gram-positive rods Gram-positive cocci, Gram-positive rods, and Gram-negative rods Oropharyngeal flora
Sputum Gram-positive cocci, and yeast Gram-positive cocci, Gram-negative rods, and yeast Enterobacterales, and oropharyngeal flora
Sputum Gram-positive cocci, and Gram-positive rods Gram-positive cocci, and Gram-negative rods Pseudomonas aeruginosa, Enterobacterales, and oropharyngeal flora
Oropharyngeal smear Gram-positive cocci, and Gram-negative rods Gram-positive cocci, Gram-positive rods, and Gram-negative rods Oropharyngeal flora
Oropharyngeal smear Gram-positive cocci Gram-positive cocci, and Gram-negative rods Oropharyngeal flora
Deep wound smear Gram-positive cocci, Gram-positive rods, and Gram-negative rods Gram-positive cocci Streptococcus anginosus, Prevotella buccae, Candida albicans, and Gram-positive flora
Deep wound smear Gram-positive cocci Gram-positive cocci, Gram-negative rods Staphylococcus aureus, and Streptococcus agalactiae
Deep wound smear Negative smear Gram-positive cocci Gram positive flora
Deep wound smear Negative smear Gram-negative rods Morganella morganii
Deep wound smear Gram-positive rods Gram-positive cocci, and Gram-positive rods Staphylococcus aureus, and Bacillus simplex
Biopsy Gram-negative rods Negative smear Negative culture
Deep wound smear Gram-positive cocci, and yeast Gram-positive cocci, Gram-negative rods, and yeast Mixed flora, and Candida albicans
Positive blood culture Gram-negative rods Negative smear Fusobacterium nucleatum

According to culture results, 6.5% (2/31) and 9.7% (3/31) of the remaining discrepant results were smear negative and culture positive for manual microscopy and automated digital imaging with on-screen reading, respectively. Additionally, for 32.3% (10/31) of the remaining discrepant results, only oropharyngeal flora was observed on culture media. The assessment of the 31 discrepant reviewed slides against culture highlighted that 42% (13/31) and 38.7% (12/31) were discordant with the culture results for manual microscopy and automated digital imaging with on-screen reading, respectively (Table 2). The performances of the two compared methods according to culture results are depicted in Table 3.

Table 3.

The performances of the two compared methods according to culture results

Manual microscopy % Sensitivity % Specificity % PPV % NPV
Blood cultures 100 100 100 100
Respiratory tract specimens 75 100 100 17.1
Sterile site specimens 58.2 88.7 89.8 55.3
Total 81.3 90.8 98 46.8
Automated digital imaging % Sensitivity % Specificity % PPV % NPV
Blood cultures 98.7 100 100 75
Respiratory tract specimens 80.2 83.3 98.9 17.9
Sterile site specimens 51.6 83 83.9 50
Total 80.7 84.6 96.7 44.4
Performances after a subsequent review of the discrepant slides
Manual microscopy % sensitivity % specificity % PPV % NPV
Blood cultures 100 100 100 100
Respiratory tract specimens 81.9 83.3 99 19.2
Sterile site specimens 60.4 86.8 88.7 56.1
Total 84.1 87.7 97.4 50
Automated digital imaging % sensitivity % specificity % PPV % NPV
Blood cultures 99.3 100 100 85.7
Respiratory tract specimens 81 83.3 98.9 18.5
Sterile site specimens 62.6 90.6 91.9 58.5
Total 84.1 90.8 98 50.9

NPV negative predictive value, PPV positive predictive value

Discussion

The Gram stain belongs to those tests routinely performed in clinical microbiology laboratories that are prone to variability and to the subjectivity of their interpretation [7, 8]. In many cases, Gram stain errors can have a significant clinical impact, especially for sterile site specimens and blood cultures, underlining why clinical microbiology laboratories perform frequent quality controls to monitor the correlation between Gram stain results and cultures. Several reports have examined Gram stain errors rates and highlighted the major drivers of such errors [912]. Moreover, the interpretation of Gram stain results remains labor-intensive, time-consuming, highly subjective, and strongly dependent on the specimen types and on the smear quality. Nowadays, the steadily increasing workload for clinical analyses challenges clinical microbiology laboratories, facing the divergent needs to improve quality, productivity, and turn-around times while simultaneously rationalizing the laboratory technologists’ workforce. Some of these challenges can therefore further impact the Gram stain errors rates by precluding a daily review of smears that showed discrepant results with cultures, and decrease the ongoing quality control. To decrease the rates of discrepant results for sterile fluid samples, an additional slide is systematically performed and stained with acridine orange in our laboratory, even if the sensitivity of Gram and acridine orange-staining remains suboptimal compared to culture in the rapid diagnosis of septic arthritis [13].

Using automated digital imaging with on-screen reading, we assessed the overall slide classification accuracy on the 443 Gram-stained smears which were previously classified by manually microscopy. The overall agreement between both methods was 77%. However, after a subsequent review of the discrepant slides, the overall agreement reached 93%. The rate of discrepant results was markedly different between the three specimen types included in this study. Specifically, the agreements between the compared methods according to the specimen types were 99.4%, 85.2%, and 93.8% for positive blood cultures, respiratory tract specimens, and sterile site specimens, respectively. Despite the fact that not all observed bacteria in a specimen may be recovered in culture due to either a lack of viability or overgrowth by a more predominant organism(s), the overall agreement between the compared methods and the culture results reached 78% (345/443) and 79% (349/443) for manual microscopy and automated digital imaging using on-screen reading, respectively.

In the context of a multicenter evaluation of Gram stain error study, Samuel et al. reported that 24% of discrepant Gram strain results were linked to interpretation errors by the technologists, across the different study sites [11, 12]. Based on the observations made during our study, specific factors were highlighted as the cause of Gram stain errors using the automated digital imaging with on-screen reading: 1) the recognition of microorganism morphologies on-screen can appear very different and more challenging to identify as compared to routine manual microscopy, 2) the nature of smear preparation, and 3) the thick smears with high cellular content are also especially challenging. To mitigate such errors, smears with inadequate material should be repeated in order to increase the number and the quality of fields examined in addition to the double review of the smears. While this approach might reduce error rates, the logistics appears arduous. Double review of smears can therefore be routinely performed only for a subset of specimens, focusing for example only on blood cultures and sterile fluids. Finally, the reporting and categorization of Gram stain errors by types of error and technologists can help revealing patterns, for targeted review or for additional training.

Conclusion

While the results reported in this study were not surprising given the subjective nature of Gram stains, they emphasize that on-screen reading is challenging even to experienced professionals; the laboratory technologists should therefore benefit from additional and specific training coupled to performance assessment. Additionally, the monitoring of Gram-stain errors, which is facilitated by automated digital imaging, represents a crucial step in the process of improving the quality of Gram stain results. Automated digital imaging of Gram-stained slides permits improved diagnostic workflow by facilitating the slides review and the exchange of information and by building educational picture libraries containing challenging smears that are the source of the most frequent errors. Finally, on-screen reading of digital images affords huge practical and ergonomic advantages as compared to the tedious manual microscopy and constitutes a useful complement to manual microscopy.

Acknowledgments

Code availability

Not applicable.

Availability of data and material

Not applicable.

Authors’ contributions

AF: performed the analysis and monitored, compiled, and analyzed the data. NA, LR, VB, MT, and GR: performed the analysis. NV and JS: reviewed the manuscript. AC: designed and conceptualized the study, supervision of the procedures, and validation of the data and wrote the manuscript. All authors contributed to the article and approved the submitted version.

Funding

Open Access funding provided by Université de Genève. This study was performed by using internal funding.

Declarations

Ethics approval

In accordance with local ethical committee, routine clinical laboratories of our institution may use biological sample leftovers for method development after irreversible anonymization of data. The official name of the ethics committee is “Commission cantonale d’éthique de la recherche (CCER)” https://www.hug-ge.ch/ethique

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Cherkaoui A, Renzi G, Viollet A, Fleischmann M, Metral-Boffod L, Dominguez-Amado D, Vuilleumier N, Schrenzel J (2020) Implementation of the WASPLab and first year achievements within a university hospital. Eur J Clin Microbiol Infect Dis [DOI] [PubMed]
  • 2.Cherkaoui A, Renzi G, Vuilleumier N, Schrenzel J. Copan WASPLab automation significantly reduces incubation times and allows earlier culture readings. Clin Microbiol Infect. 2019;25(11):1430 e1435-1430 e1412. doi: 10.1016/j.cmi.2019.04.001. [DOI] [PubMed] [Google Scholar]
  • 3.Cherkaoui A, Renzi G, Azam N, Schorderet D, Vuilleumier N, Schrenzel J (2020) Rapid identification by MALDI-TOF/MS and antimicrobial disk diffusion susceptibility testing for positive blood cultures after a short incubation on the WASPLab. Eur J Clin Microbiol Infect Dis [DOI] [PubMed]
  • 4.Cherkaoui A, Renzi G, Charretier Y, Blanc DS, Vuilleumier N, Schrenzel J. Automated incubation and digital image analysis of chromogenic media using Copan WASPLab enables rapid detection of vancomycin-resistant enterococcus. Front Cell Infect Microbiol. 2019;9:379. doi: 10.3389/fcimb.2019.00379. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Bailey AL, Burnham CD. Reducing the time between inoculation and first-read of urine cultures using total lab automation significantly reduces turn-around-time of positive culture results with minimal loss of first-read sensitivity. Eur J Clin Microbiol Infect Dis. 2019;38(6):1135–1141. doi: 10.1007/s10096-019-03512-3. [DOI] [PubMed] [Google Scholar]
  • 6.Burckhardt I, Last K, Zimmermann S. Shorter incubation times for detecting multi-drug resistant bacteria in patient samples: defining early imaging time points using growth kinetics and total laboratory automation. Ann Lab Med. 2019;39(1):43–49. doi: 10.3343/alm.2019.39.1.43. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Nagendra S, Bourbeau P, Brecher S, Dunne M, LaRocco M, Doern G. Sampling variability in the microbiological evaluation of expectorated sputa and endotracheal aspirates. J Clin Microbiol. 2001;39(6):2344–2347. doi: 10.1128/JCM.39.6.2344-2347.2001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Musher DM, Montoya R, Wanahita A. Diagnostic value of microscopic examination of gram-stained sputum and sputum cultures in patients with bacteremic pneumococcal pneumonia. Clin Infect Dis. 2004;39(2):165–169. doi: 10.1086/421497. [DOI] [PubMed] [Google Scholar]
  • 9.Rand KH, Tillan M. Errors in interpretation of gram stains from positive blood cultures. Am J Clin Pathol. 2006;126(5):686–690. doi: 10.1309/V4KE2FPM5T8V4552. [DOI] [PubMed] [Google Scholar]
  • 10.Goodyear N, Kim S, Reeves M, Astion ML. A 2-year study of gram stain competency assessment in 40 clinical laboratories. Am J Clin Pathol. 2006;125(1):28–33. doi: 10.1309/40WD3015CH1RYH58. [DOI] [PubMed] [Google Scholar]
  • 11.Samuel LP, Balada-Llasat JM, Harrington A, Cavagnolo R. Multicenter assessment of gram stain error rates. J Clin Microbiol. 2016;54(6):1442–1447. doi: 10.1128/JCM.03066-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Samuel LP, Balada-Llasat JM, Harrington A, Cavagnolo R. Correction for Samuel et al., multicenter assessment of gram stain error rates. J Clin Microbiol. 2016;54(9):2405. doi: 10.1128/JCM.01465-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Cunningham G, Seghrouchni K, Ruffieux E, Vaudaux P, Gayet-Ageron A, Cherkaoui A, Godinho E, Lew D, Hoffmeyer P, Uckay I. Gram and acridine orange staining for diagnosis of septic arthritis in different patient populations. Int Orthop. 2014;38(6):1283–1290. doi: 10.1007/s00264-014-2284-3. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable.


Articles from European Journal of Clinical Microbiology & Infectious Diseases are provided here courtesy of Springer

RESOURCES