Skip to main content
Scientific Reports logoLink to Scientific Reports
. 2021 Sep 24;11:19035. doi: 10.1038/s41598-021-98530-7

Optimization of collimator angles in dual-arc volumetric modulated arc therapy planning for whole-brain radiotherapy with hippocampus and inner ear sparing

Wuji Sun 1,#, Kunzhi Chen 1,#, Yu Li 1, Wenming Xia 1, Lihua Dong 1,2,3, Yinghua Shi 1, Chao Ge 1, Xu Yang 1, Libo Wang 1, Huidong Wang 1,
PMCID: PMC8463591  PMID: 34561504

Abstract

To optimize the collimator angles in dual-arc volumetric modulated arc therapy (VMAT) plans for whole-brain radiotherapy with hippocampus and inner ear sparing (HIS-WBRT). Two sets of dual-arc VMAT plans were generated for 13 small-cell lung cancer patients: (1) The collimator angles of arcs 1 and 2 (θ1/θ2) were 350°/10°, 350°/30°, 350°/45°, 350°/60°, and 350°/80°, i.e., the intersection angle of θ1 and θ2θ) increased. (2) θ1/θ2 were 280°/10°, 300°/30°, 315°/45°, 330°/60°, and 350°/80°, i.e., Δθ = 90°. The conformity index (CI), homogeneity index (HI), monitor units (MUs), and dosimetric parameters of organs-at-risk were analyzed. Quality assurance for Δθ = 90° plans was performed. With Δθ increasing towards 90°, a significant improvement was observed for most parameters. In 350°/80° plans compared with 350°/10° ones, CI and HI were improved by 1.1% and 25.2%, respectively; MUs were reduced by 16.2%; minimum, maximum, and mean doses (D100%, Dmax, and Dmean, respectively) to the hippocampus were reduced by 5.5%, 6.3%, and 5.4%, respectively; Dmean to the inner ear and eye were reduced by 0.7% and 5.1%, respectively. With Δθ kept at 90°, the plan quality was not significantly affected by θ1/θ2 combinations. The gamma-index passing rates in 280°/10° and 350°/80° plans were relatively lower compared with the other Δθ = 90° plans. Δθ showed a significant effect on dual-arc VMAT plans for HIS-WBRT. With Δθ approaching 90°, the plan quality exhibited a nearly continuous improvement, whereas with Δθ = 90°, the effect of θ1/θ2 combination was insignificant.

Subject terms: Non-small-cell lung cancer, Radiotherapy, Techniques and instrumentation

Introduction

Whole-brain radiotherapy (WBRT) has been used for the treatment of brain metastases or the prophylactic cranial irradiation of patients with small-cell lung cancer in the past decades13. However, the neurocognitive deficits caused by the radiation, including memory loss and cognitive impairment, have always been a concern4,5. The damage to the neural stem cell compartment in the hippocampus is generally believed to be the main cause6,7. The Radiation Therapy Oncology Group (RTOG) 0933 trial has demonstrated promising results in reducing the adverse neurocognitive effect of WBRT with hippocampus and inner ear sparing (HIS-WBRT) compared with conventional WBRT79. In the RTOG 0933 protocol, the target coverage and the dose to the organs-at-risk (OARs) were strictly constrained to achieve a reasonable dose distribution, and intensity-modulated radiation therapy (IMRT) modalities are required9. Volumetric modulated arc therapy (VMAT) is usually preferred because of its higher delivery efficiency than fixed-gantry IMRT10,11.

Current VMAT implementation does not allow collimator rotation during delivery, and the collimator angle in each arc is required to be determined before plan optimization12. Numerous studies have investigated the effect of collimator angle in many aspects, including target coverage, normal tissue sparing, monitor units (MUs), and delivery accuracy, and the optimal collimator angle varied for different target sites and modalities1223. For the complex-shaped target volume in HIS-WBRT, an optimal setting for the collimator angles must be applied, which, to the knowledge of the authors, has yet to be investigated.

Apart from the specific angle of each arc, the intersection angle between arcs should be investigated to optimize the collimator angle. In particular, for dual-arc VMAT, the intersection angle between two collimator settings may have a significant effect on the plan optimization. By rotating the collimator between arcs, a better dose distribution could be achieved with the multi-leaf collimator (MLC) moving directions orthogonal to each other24.

In this study, the collimator angle of dual-arc VMAT for HIS-WBRT was optimized in two steps. First, the plan quality with different intersection angles between arcs was evaluated to find the most adequate choice. Second, with the optimal intersection angle, the specific collimator setting of each arc was investigated on the basis of plan quality evaluation and quality assurance (QA).

Materials and methods

Patient selection and contouring

A total of 13 patients with small-cell lung cancer, who had been previously treated in 2018–2020 in the First Hospital of Jilin University, were retrospectively included. The patients were immobilized in supine position by using a thermoplastic mask (Klarity Medical & Equipment Co. Ltd., Guangzhou, China). Computed tomography (CT) images of the entire head region were acquired using Philips Brilliance Big Bore CT scanner (Philips Healthcare, Cleveland, OH) with 3 mm slice thickness. The T1-weighted contrast-enhanced magnetic resonance imaging sequences, obtained with 1 mm slice thickness, were fused to the plan CT in the Eclipse version 13.5 (Varian Medical Systems, Palo Alto, CA) treatment planning system (TPS).

Anatomical contouring was carried out in accordance with the RTOG 0933 protocol9, including the whole brain, hippocampi, inner ears, lenses, optical nerves, and eyes. The planning target volume (PTV) for HIS-WBRT was defined as the whole brain with a 5 mm margin expansion excluding the hippocampus-sparing region. The hippocampus-sparing region was obtained by adding a 5 mm margin to the hippocampi, for the dose gradient between the hippocampus and PTV. The volumes (mean ± standard deviation (minimum and maximum)) of PTV, hippocampus, and inner ear were 1366.9 ± 114.3 (1099.3 and 1572.5), 3.7 ± 1.1 (1.0 and 5.2), and 1.3 ± 0.6 (0.2 and 2.4) cm3, respectively.

Treatment planning

VMAT plans were optimized on Eclipse version 13.5 for 6 MV X-ray beams of a Varian TrueBeam linear accelerator equipped with a Millennium MLC-120 MLC (Varian Medical Systems, Palo Alto, CA). The maximum dose rate was 600 MUs/min. The dose distributions were calculated using the anisotropic analytic algorithm with a grid of 2.5 mm. The jaw tracking was enabled, and the maximum field size (X × Y) was 15 cm × 40 cm to adequately cover the whole brain.

For all patients, a dose of 30 Gy was prescribed in 10 fractions, and all plans complied with the dose criteria of the RTOG 0933 protocol and additional constraints used in this study, as shown in Table 19,25. Dual-arc VMAT plans for HIS-WBRT were generated for each patient with two coplanar whole arcs. The settings of two collimator angles are denoted by θ1/θ2, i.e., the angles of the first and second arcs. The intersection angle between these two angles, denoted by Δθ, ranged from 0° to 90°. Two sets of plans with different θ1/θ2 were generated as follows:

  1. θ1/θ2 were set as 350°/10°, 350°/30°, 350°/45°, 350°/60°, and 350°/80°, i.e., Δθ increased from 20° to 90°; and

  2. θ1/θ2 were set as 280°/10°, 300°/30°, 315°/45°, 330°/60°, and 350°/80°, i.e., Δθ was fixed at 90°.

Table 1.

Dose criteria of RTOG 0933 protocol and additional constraints used in the current study.

Per protocol Variation acceptable Deviation unacceptable
RTOG 0933 protocol
Whole-brain PTV D2% ≤ 37.5 Gy D2% ≤ 40 Gy D2% > 40 Gy
D98% ≥ 25 Gy D98% < 25 Gy V30 Gy ≤ 90%
Hippocampus D100% ≤ 9 Gy D100% ≤ 10 Gy D100% > 10 Gy
Dmax ≤ 16 Gy Dmax ≤ 17 Gy Dmax > 17 Gy
Optic nerve and chiasm Dmax ≤ 37.5 Gy Dmax ≤ 37.5 Gy Dmax > 37.5 Gy
Additional constraints
Inner ear Dmean ≤ 15 Gy Dmean ≤ 16 Gy Dmean > 16 Gy
Lens Dmax ≤ 9 Gy Dmax ≤ 10 Gy Dmax > 10 Gy
Eye Dmean ≤ 11 Gy Dmean ≤ 12 Gy Dmean > 12 Gy

For each patient, all plans were optimized with identical dosimetric constraints and normalized to 92% of the PTV receiving 30 Gy, while there were certain adjustments in the optimization methods for different patients to assure that all plans complied with the dose criteria.

Treatment plan evaluation

Several parameters, including the conformity index (CI), homogeneity index (HI), MUs, and dose-volume histogram (DVH) parameters of OARs, were selected for the plan evaluation with respect to the collimator setting.

The CI and HI were defined as26,27

CI=TVPIV2TV×PIV 1

and

HI=D2\%-D98\%D50\%, 2

where TV is the target volume, PIV is the prescription isodose volume, TVPIV is the target volume within the prescription isodose volume, and Dn% represents the dose delivered to n% of the target volume. CI closer to 1 indicates better dose conformity. HI closer to 0 indicates better dose homogeneity.

The evaluated dosimetric parameters of OAR included the minimum, maximum, and mean doses (D100%, Dmax, and Dmean, respectively) of the hippocampus; the Dmean of the inner ear and eye; and the Dmax of the optical nerve and lens.

Quality assurance

QA was performed for Δθ = 90° plans by using local gamma-index analysis with 3%/3, 3%/2, and 2%/2 mm criteria28, using an a-Si 1200 Electronic Portal Imaging Device with an active area of 40 cm × 40 cm and a pixel number of 1190 × 1190. Dose images were acquired for each arc with a source-to-detector distance of 100 cm and later retrieved to the Portal Dosimetry module of the Eclipse TPS. The Portal Dose Image Prediction version 13.6.23 algorithm was used to calculate the predicted dose images.

Statistical analysis

Friedman test was first carried out for the above parameters within each set of plans, i.e., with different Δθ and same Δθ (= 90°), to investigate if Δθ and the specific angle of each arc could affect the plan quality of dual-arc VMAT. Wilcoxon signed-rank test was performed for multiple comparisons between the Δθ = 90° and Δθ < 90° plans. The comparisons were conducted in two groups, that is, (A) between plans with same θ1 but different θ2 and (B) between those with same θ2 but different θ1. The obtained p-values were evaluated with Bonferroni correction.

The above two tests were also used for the gamma-index passing rate evaluation in the plan QA for Δθ = 90° plans. All statistical analyses were performed using IBM SPSS Statistics version 26.0 software (IBM Corporation, Armonk, NY). p < 0.05 was defined as statistically significant.

Ethics approval and consent to participate

All experimental protocols were approved by the Ethics Committee of the First Hospital of Jilin University. The Ethics Committee of the First Hospital of Jilin University waived the need for informed consent. All research was performed in accordance with relevant guidelines and regulations.

Results

Table 2 shows the evaluated parameters and their corresponding Friedman test results for the VMAT plans with increasing Δθ. Figure 1 presents the predicted dose distribution of each arc in these plans for a typical patient. The results for plans with the same Δθ (= 90°) are shown in Table 3. Supplementary Fig. S1 presents the calculated dose distribution in one 350°/80° plan for a typical patient and the corresponding DVH compared with that in Δθ < 90° plans.

Table 2.

Dosimetric parameters and MUs of VMAT plans with intersection angle of two collimators (Δθ) changing from 20° to 90°, along with their Friedman test results.

Parameter Result for VMAT with collimator angles θ1/θ2 a p-value
350°/10° 350°/30° 350°/45° 350°/60° 350°/80°
PTV CI 0.848 ± 0.009 0.853 ± 0.007 0.856 ± 0.007 0.858 ± 0.004 0.857 ± 0.008 0.002*
HI 0.259 ± 0.023 0.212 ± 0.016 0.202 ± 0.016 0.199 ± 0.012 0.194 ± 0.019  < 0.001*
Hippocampus D100% (cGy) 929.7 ± 40.1 891.1 ± 32.8 884.2 ± 25.1 880.5 ± 31.7 878.8 ± 29.2  < 0.001*
Dmax (cGy) 1507.2 ± 105.6 1433.0 ± 77.7 1422.0 ± 65.7 1434.2 ± 76.7 1412.7 ± 79.3  < 0.001*
Dmean (cGy) 1103.3 ± 43.3 1056.5 ± 31.0 1053.8 ± 27.6 1049.6 ± 29.7 1043.3 ± 30.3  < 0.001*
Inner ear Dmean (cGy) 1437.6 ± 40.3 1433.0 ± 20.1 1429.5 ± 30.0 1427.4 ± 29.8 1427.3 ± 29.1 0.008*
Lens Dmax (cGy) 705.3 ± 29.3 697.2 ± 33.3 705.5 ± 32.6 703.6 ± 24.7 702.4 ± 38.1 0.699
Optical nerve Dmax (cGy) 3143.5 ± 156.3 3106.8 ± 118.2 3175.1 ± 40.4 3123.6 ± 103.1 3087.3 ± 127.0 0.053
Eye Dmean (cGy) 970.9 ± 37.5 943.2 ± 18.3 934.0 ± 17.6 930.5 ± 11.6 921.0 ± 20.9  < 0.001*
Monitor Unit MU 1073.6 ± 118.8 929.5 ± 44.6 908.7 ± 33.9 900.6 ± 28.5 899.5 ± 28.2  < 0.001*

aθ1 and θ2 indicate the collimator angles of arc 1 and arc 2, respectively.

*Statistically significant (p < 0.05).

Figure 1.

Figure 1

Predicted dose distribution for each arc of dual-arc VMAT plans in coronal view with collimator settings of (a) 350°/10°, (b) 350°/30°, (c) 350°/45°, (d) 350°/60°, and (e) 350°/80° for a typical patient.

Table 3.

Dosimetric parameters and MUs of Δθ = 90° plans, along with their Friedman test results.

Parameter Result for VMAT with collimator angles θ1/θ2 a p-value
280°/10° 300°/30° 315°/45° 330°/60° 350°/80°
PTV CI 0.859 ± 0.006 0.859 ± 0.008 0.858 ± 0.007 0.855 ± 0.007 0.857 ± 0.008 0.501
HI 0.189 ± 0.013 0.193 ± 0.013 0.189 ± 0.013 0.194 ± 0.011 0.194 ± 0.019 0.289
Hippocampus D100% (cGy) 879.4 ± 34.5 873.5 ± 27.1 872.2 ± 28.2 869.8 ± 31.3 878.8 ± 29.2 0.112
Dmax (cGy) 1415.1 ± 71.1 1432.9 ± 69.6 1411.6 ± 81.8 1415.8 ± 65.5 1412.7 ± 79.3 0.074
Dmean (cGy) 1041.7 ± 29.8 1045.4 ± 27.5 1039.6 ± 31.0 1041.6 ± 30.8 1043.3 ± 30.3 0.412
Inner ear Dmean (cGy) 1411.0 ± 56.0 1412.6 ± 31.1 1407.8 ± 38.1 1403.5 ± 49.4 1427.3 ± 29.1 0.106
Lens Dmax (cGy) 669.3 ± 26.1 684.6 ± 20.4 683.9 ± 24.5 698.5 ± 27.8 702.4 ± 38.1 0.001*
Optical nerve Dmax (cGy) 3106.7 ± 87.2 3180.2 ± 55.6 3180.7 ± 96.1 3171.5 ± 63.8 3087.3 ± 127.0 0.001*
Eye Dmean (cGy) 919.4 ± 22.8 922.8 ± 13.9 918.2 ± 21.0 925.0 ± 9.4 921.0 ± 20.9 0.421
Monitor unit MU 905.2 ± 37.5 878.4 ± 44.4 878.5 ± 36.7 881.3 ± 36.7 899.5 ± 28.2 0.003*

aθ1 and θ2 indicate the collimator angles of arc 1 and arc 2, respectively.

*Statistically significant (p < 0.05).

Plan quality with increasing Δθ

As shown in Table 2, Δθ showed a significant effect on the CI, HI, MUs, and dosimetric parameters of the hippocampus (D100%, Dmax, and Dmean), inner ear (Dmean), and eye (Dmean), whereas no significance was found in the Dmax of lens and optical nerve.

Among the five collimator settings, the 350°/80° (Δθ = 90°) plans exhibited the best HI, MUs, and dosimetric parameters of the hippocampus, inner ear, and eye and the second-best CI. In the 350°/80° plans compared with the 350°/10° ones, the CI and HI were improved by 1.1% (p = 0.007) and 25.2% (p < 0.001), respectively; MUs were reduced by 16.2% (p < 0.001); the D100%, Dmax, and Dmean to the hippocampus were reduced by 5.5% (p < 0.001), 6.3% (p < 0.001), and 5.4% (p < 0.001), respectively; the Dmean to the inner ear was reduced by 0.7% (p = 0.039); and the Dmean to the eye was reduced by 5.1% (p < 0.001).

For further investigation of the relation between the plan quality and Δθ, the parameter values in the Δθ = 90° plans minus those in the Δθ < 90° plans are shown in Fig. 2 in two groups: group A, plans with the same θ1 indicated by the solid lines; and group B, plans with the same θ2 indicated by the dashed lines. The difference in the plan Δθ was denoted by δθ = 90° − Δθ (< 90°).

Figure 2.

Figure 2

Variation of parameter differences between Δθ < 90° plans and Δθ = 90° plans with respect to δθ (= 90° − Δθ). Group A shows the differences between plans with the same arc 1 collimator, indicated by the filled symbols and solid lines. Group B shows the differences between plans with the same arc 2 collimator, indicated by the open symbols with dashed lines.

With decreasing δθ, i.e., Δθ increasing towards 90°, improvements were clearly demonstrated for the CI, HI, MUs, and dosimetric parameters of the hippocampus and eye. A high degree of similarity was observed between groups A and B. One-tailed Wilcoxon signed-rank tests were performed for the comparisons within groups A and B, and the results are shown in Supplementary Tables S1 and S2, respectively. Compared with the Δθ < 90° plans, the Δθ = 90° plans generally showed significantly better or similar results depending on δθ.

Plan quality with same Δθ (90°) and different θ1/θ2

As shown in Table 3, θ1/θ2 only showed a significant correlation with the MUs and maximum doses to the lens and optical nerve. The 300°/30°, 315°/45°, and 330°/60° plans required higher MUs but reduced Dmax to the optical nerve compared with the 280°/10° and 350°/80° plans. As to the Dmax of lens, the 280°/10°, 300°/30°, and 315°/45° plans exhibited better results than the 330°/60° and 350°/80° plans. In general, the difference in plan quality was relatively trivial.

Quality assurance for Δθ = 90° plans

The QA results for the Δθ = 90° plans are displayed in Table 4. The passing rate was statistically affected by the collimator setting, and for all gamma criteria, the 280°/10° and 350°/80° plans had significantly lower passing rates than the 300°/30°, 315°/45°, and 330°/60° plans (p < 0.005 for Wilcoxon signed-rank tests). The passing rates of each arc in the 280°/10° and 350°/80° plans are shown in Table 5. The results for 280° and 80° were generally similar and lower than those for 350° and 10°.

Table 4.

Passing rates of Δθ = 90° plans using 3%/3, 3%/2, and 2%/2 mm gamma criteria, along with their Friedman test results.

Gamma criteria Passing rate for VMAT with collimator angles θ1/θ2 (%) p-value
280°/10° 300°/30° 315°/45° 330°/60° 350°/80°
3%/3 mm 97.6 ± 0.6 98.3 ± 0.6 98.4 ± 0.5 98.5 ± 0.5 97.6 ± 0.6  < 0.001*
3%/2 mm 95.4 ± 1.0 96.9 ± 1.0 97.1 ± 0.8 97.2 ± 0.8 95.5 ± 1.1  < 0.001*
2%/2 mm 90.4 ± 2.2 92.7 ± 2.1 92.9 ± 1.8 93.1 ± 1.9 90.4 ± 2.2  < 0.001*

*Statistically significant (p < 0.05).

Table 5.

Passing rates of each arc in 280°/10° and 350°/80° plans using 3%/3, 3%/2, and 2%/2 mm gamma criteria.

Gamma criteria Passing rate for 280°/10° (%) Passing rate for 350°/80° (%)
Arc 1 (280°) Arc 2 (10°) Arc 1 (350°) Arc 2 (80°)
3%/3 mm 98.7 ± 0.6 99.1 ± 0.3 99.3 ± 0.3 97.8 ± 1.4
3%/2 mm 95.6 ± 2.2 97.6 ± 0.7 98.0 ± 0.6 95.0 ± 2.6
2%/2 mm 92.2 ± 2.8 96.1 ± 1.1 96.7 ± 1.0 91.6 ± 3.1

Discussion

Considering the positive results demonstrated in the RTOG 0933 trail, HIS-WBRT has gradually become a common practice for the treatment of brain metastases and prophylactic cranial irradiation79,2931. The plan quality is especially important, and choosing the collimator angle is a vital part of the plan optimization to achieve an ideal dose distribution due to the complexity of HIS-WBRT. In the present study, the collimator angle had a significant effect on the dose distribution quality and MUs, in agreement with the previous studies. Zhang et al.32 developed a collimator trajectory selection method for VMAT, and an enhanced dose distribution could be achieved by aligning the collimator with the target shape. Several studies showed that the VMAT optimization involving sectional optimization of collimator angle could provide delivery efficiency and dosimetric improvements33,34. Recent studies on the dynamic collimator rotation approach have shown positive results on improving the plan quality24,3537.

Table 2 shows that among the five sets of plans with different Δθ, the 350°/80° (Δθ = 90°) plans had the best overall dosimetric performance and MUs. Figure 2 demonstrates the differences between the Δθ = 90° and Δθ < 90° plans with respect to δθ. A visible similarity was present between groups A and B in the improvement of CI, HI, MUs, and doses to the hippocampus and eye with decreasing δθ, indicating that with Δθ increasing towards 90°, the plan quality could be gradually improved.

The five sets of Δθ = 90° plans were compared to investigate the possible effect from the θ1/θ2 combination with the same Δθ, and the results are shown in Table 3. Thus, when the intersection Δθ was the same, most parameters showed no significant correlation with θ1/θ2 combinations, except the MUs and maximum doses to lens and optical nerve. Considering the behavior of all parameters, the effect of Δθ was much greater than that of the specific angle of each arc.

The collimator angle should allow the MLC aperture to encompass the PTV while avoiding the sparing region. For the near-spherical target volume in HIS-WBRT, with the avoidance of two hippocampi, which exist bilaterally in the middle of the brain, the plan complexity could be considerably increased compared with conventional WBRT. Many studies proposed that high plan complexity could affect the accuracy of dose calculation and beam delivery3840. The predicted dose distribution shown in Fig. 1 indicated that regardless of the collimator setting, the plan optimization tended to form mutually orthogonal dose distributions in two arcs, and the integrated dose distribution could then conform with the complex-shape target volume of HIS-WBRT. With the Δθ = 90° plans, i.e., the MLC orientations of two arcs perpendicular to each other, the motion complexity of MLC could be potentially reduced compared with the Δθ < 90° plans, which could be part of the reason for the improved dose distribution quality of Δθ = 90° plans.

Previous studies on the effect of collimator angle on VMAT plan quality proposed various optimal settings depending on the tumor site and shape12,17,41. For vertebrae metastases, Mancosu et al. suggested a collimator angle of around 90° to align the MLC leaf motion with the spinal cord41. For prostate cancer, Li et al. suggested an angle of 45° for the optimal dose distribution plan complexity17. For nasopharynx cancer, Otto found that 45° was preferable in most cases12. In the present study, the optimal collimator setting is affected by the complex-shaped target of HIS-WBRT and thus different from the above findings. For the dual-arc VMAT technique adopted for HIS-WBRT, the optimal plan quality was acquired with Δθ = 90°, while the effect from the specific setting of each arc was relatively insignificant.

For the Δθ = 90° plans with different collimator angle combinations, the gamma-index analysis results shown in Tables 4 and 5 indicated that the passing rates could be affected by the specific angle of each arc. Although the passing rate with the collimator closer to 0° showed preferable results in this study, it should be noted that the QA result could be associated with many factors, such as the dosimetric leaf gap, tongue-and-groove effect, interleaf leakage, and intra- and inter-leaf transmission settings of MLC4245. For the PDIP algorithm in the Eclipse TPS, these MLC configuration parameters are constant for all collimator and gantry angles, but in practice, they may vary for different situations. The leaf position accuracy could be affected by the gantry rotation with different collimator angles due to gravitational effects imposed on the leaf carriage system45. The varying MLC parameters could yield a significant dose error, especially for highly modulated plans. Therefore, with regard to the plan QA for other institutions, the preferable result may vary.

Conclusions

This study demonstrated the role of collimator settings in dual-arc VMAT planning for HIS-WBRT. The intersection angle between two collimator settings, Δθ, showed a significant influence on the plan quality. With Δθ approaching 90°, a nearly continuous improvement was observed in the dose distributions and MUs. However, with the same Δθ = 90°, the effect of specific collimator angle of each arc was relatively trivial. The QA results indicated that the 300°/30°, 315°/45°, and 330°/60° plans had a better delivery accuracy than the 280°/10° and 350°/80° plans.

Supplementary Information

Supplementary Information. (567.8KB, docx)

Acknowledgements

This work was supported by the National Natural Science Foundation of China [No. 81773353]; the Jilin Scientific and Technological Development Program [Nos. 20190201204JC, 20200601005JC]; and the Jilin Province Special Project of Medical and Health Talents [No. JLSCZD2019-032].

Author contributions

H.W. conceived and designed the study. W.S. and K.C. wrote the paper. Y.L, W.X. and C.G. are responsible for data acquisition and statistical analyses. X.Y. and L.W. prepared the figures. L.D. and Y.S. made critical revisions to the paper. All authors reviewed the manuscript.

Data availability

Research data analyzed in this study are included as supplementary materials. Other data that support the findings of this study are available from the corresponding author upon reasonable request.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

These authors contributed equally: Wuji Sun and Kunzhi Chen.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-021-98530-7.

References

  • 1.Slotman B, et al. Prophylactic cranial irradiation in extensive small-cell lung cancer. N. Engl. J. Med. 2007;357:664–672. doi: 10.1056/NEJMoa071780. [DOI] [PubMed] [Google Scholar]
  • 2.Kotalik J, Yu E, Markman BR, Evans WK. Practice guideline on prophylactic cranial irradiation in small-cell lung cancer. Int. J. Radiat. Oncol. Biol. Phys. 2001;50:309–316. doi: 10.1016/S0360-3016(01)01448-1. [DOI] [PubMed] [Google Scholar]
  • 3.Andrews DW, et al. Whole brain radiation therapy with or without stereotactic radiosurgery boost for patients with one to three brain metastases: Phase III results of the RTOG 9508 randomised trial. Lancet. 2004;363:1665–1672. doi: 10.1016/S0140-6736(04)16250-8. [DOI] [PubMed] [Google Scholar]
  • 4.Welzel G, et al. Memory function before and after whole brain radiotherapy in patients with and without brain metastases. Int. J. Radiat. Oncol. Biol. Phys. 2008;72:1311–1318. doi: 10.1016/j.ijrobp.2008.03.009. [DOI] [PubMed] [Google Scholar]
  • 5.Roman DD, Sperduto PW. Neuropsychological effects of cranial radiation: Current knowledge and future directions. Int. J. Radiat. Oncol. Biol. Phys. 1995;31:983–998. doi: 10.1016/0360-3016(94)00550-8. [DOI] [PubMed] [Google Scholar]
  • 6.Kirby N, Chuang C, Pouliot J, Hwang A, Barani IJ. Physics strategies for sparing neural stem cells during whole-brain radiation treatments. Med. Phys. 2011;38:5338–5344. doi: 10.1118/1.3633946. [DOI] [PubMed] [Google Scholar]
  • 7.Gondi V, Hermann BP, Mehta MP, Tome WA. Hippocampal dosimetry predicts neurocognitive function impairment after fractionated stereotactic radiotherapy for benign or low-grade adult brain tumors. Int. J. Radiat. Oncol. Biol. Phys. 2012;83:e487–e493. doi: 10.1016/j.ijrobp.2011.10.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Gondi V, et al. Hippocampal-sparing whole-brain radiotherapy: A "how-to" technique using helical tomotherapy and linear accelerator-based intensity-modulated radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2010;78:1244–1252. doi: 10.1016/j.ijrobp.2010.01.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Gondi V, et al. Preservation of memory with conformal avoidance of the hippocampal neural stem-cell compartment during whole-brain radiotherapy for brain metastases (RTOG 0933): A phase II multi-institutional trial. J. Clin. Oncol. 2014;32:3810–3816. doi: 10.1200/JCO.2014.57.2909. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Rao M, et al. Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: Plan quality, delivery efficiency and accuracy. Med. Phys. 2010;37:1350–1359. doi: 10.1118/1.3326965. [DOI] [PubMed] [Google Scholar]
  • 11.Wolff D, et al. Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. Radiother. Oncol. 2009;93:226–233. doi: 10.1016/j.radonc.2009.08.011. [DOI] [PubMed] [Google Scholar]
  • 12.Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med. Phys. 2008;35:310–317. doi: 10.1118/1.2818738. [DOI] [PubMed] [Google Scholar]
  • 13.Kim H, Li R, Lee R, Xing L. Beam's-eye-view dosimetrics (BEVD) guided rotational station parameter optimized radiation therapy (SPORT) planning based on reweighted total-variation minimization. Phys. Med. Biol. 2015;60:N71–82. doi: 10.1088/0031-9155/60/5/N71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Otto K. Letter to the editor on 'single-arc IMRT?'. Phys. Med. Biol. 2009;54:L37–L41. doi: 10.1088/0031-9155/54/8/L03. [DOI] [PubMed] [Google Scholar]
  • 15.Okumura M, Obata Y, Shimomura K, Tamura M, Nishimura Y. The effect of gantry and collimator angles on leaf limited velocity and position in dynamic multileaf collimator intensity-modulated radiation therapy. Phys. Med. Biol. 2010;55:3101–3113. doi: 10.1088/0031-9155/55/11/008. [DOI] [PubMed] [Google Scholar]
  • 16.Sandrini ES, da Silva AX, da Silva CM. Evaluation of collimator rotation for volumetric modulated arc therapy lung stereotactic body radiation therapy using flattening filter free. Appl. Radiat. Isot. 2018;141:257–260. doi: 10.1016/j.apradiso.2018.05.026. [DOI] [PubMed] [Google Scholar]
  • 17.Li MH, Huang SF, Chang CC, Lin JC, Tsai JT. Variations in dosimetric distribution and plan complexity with collimator angles in hypofractionated volumetric arc radiotherapy for treating prostate cancer. J. Appl. Clin. Med. Phys. 2018;19:93–102. doi: 10.1002/acm2.12249. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Murtaza G, Mehmood S, Favretto MS, Cora S. Optimal VMAT delivery for Elekta MLC beam modulator: A study of collimator rotation for head and neck planning. J. Med. Imaging Radiat. Sci. 2020;51:289–298. doi: 10.1016/j.jmir.2020.02.001. [DOI] [PubMed] [Google Scholar]
  • 19.Li QL, et al. Collimator rotation in volumetric modulated arc therapy for craniospinal irradiation and the dose distribution in the beam junction region. Radiat. Oncol. 2015;10:235. doi: 10.1186/s13014-015-0544-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Tas B, Bilge H, Ozturk ST. An investigation of the dose distribution effect related with collimator angle in volumetric arc therapy of prostate cancer. J. Med. Phys. 2016;41:100–105. doi: 10.4103/0971-6203.181635. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Du WL, Johnson JL, Jiang W, Kudchadker RJ. On the selection of gantry and collimator angles for isocenter localization using Winston–Lutz tests. J. Appl. Clin. Med. Phys. 2016;17:167–178. doi: 10.1120/jacmp.v17i1.5792. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Ohira S, et al. Effect of collimator angle on HyperArc stereotactic radiosurgery planning for single and multiple brain metastases. Med. Dosim. 2020;45:85–91. doi: 10.1016/j.meddos.2019.07.004. [DOI] [PubMed] [Google Scholar]
  • 23.Bai S, et al. Effect of MLC leaf position, collimator rotation angle, and gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma. Med. Dosim. 2013;38:143–147. doi: 10.1016/j.meddos.2012.10.002. [DOI] [PubMed] [Google Scholar]
  • 24.Lyu Q, et al. VMAT optimization with dynamic collimator rotation. Med. Phys. 2018;45:2399–2410. doi: 10.1002/mp.12915. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Lambrecht M, et al. Radiation dose constraints for organs at risk in neuro-oncology; the European particle therapy network consensus. Radiother. Oncol. 2018;128:26–36. doi: 10.1016/j.radonc.2018.05.001. [DOI] [PubMed] [Google Scholar]
  • 26.ICRU ICRU report 83: Prescribing, recording, and reporting photon-beam intensity-modulated radiation therapy (IMRT) J. Int. Comm. Radiat. Units Meas. 2010;10:1–106. [Google Scholar]
  • 27.ICRU ICRU report 91: Prescribing, recording, and reporting of stereotactic treatments with small photon beams. J. Int. Comm. Radiat. Units Meas. 2014;14:1–145. [Google Scholar]
  • 28.Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med. Phys. 1998;25:656–661. doi: 10.1118/1.598248. [DOI] [PubMed] [Google Scholar]
  • 29.Moon SY, Yoon M, Chung M, Chung WK, Kim DW. Comparison of the extent of hippocampal sparing according to the tilt of a patient's head during WBRT using linear accelerator-based IMRT and VMAT. Phys. Med. 2016;32:657–663. doi: 10.1016/j.ejmp.2016.04.009. [DOI] [PubMed] [Google Scholar]
  • 30.Jiang A, et al. Dosimetric evaluation of four whole brain radiation therapy approaches with hippocampus and inner ear avoidance and simultaneous integrated boost for limited brain metastases. Radiat. Oncol. 2019;14:46. doi: 10.1186/s13014-019-1255-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Krayenbuehl J, Di Martino M, Guckenberger M, Andratschke N. Improved plan quality with automated radiotherapy planning for whole brain with hippocampus sparing: A comparison to the RTOG 0933 trial. Radiat. Oncol. 2017;12:161. doi: 10.1186/s13014-017-0896-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Zhang P, et al. Optimization of collimator trajectory in volumetric modulated arc therapy: Development and evaluation for paraspinal SBRT. Int. J. Radiat. Oncol. Biol. Phys. 2010;77:591–599. doi: 10.1016/j.ijrobp.2009.08.056. [DOI] [PubMed] [Google Scholar]
  • 33.Ahn BS, et al. Dosimetric effects of sectional adjustments of collimator angles on volumetric modulated arc therapy for irregularly-shaped targets. PLoS ONE. 2017;12:e0174924. doi: 10.1371/journal.pone.0174924. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Kim JI, Ahn BS, Choi CH, Park JM, Park SY. Optimal collimator rotation based on the outline of multiple brain targets in VMAT. Radiat. Oncol. 2018;13:88. doi: 10.1186/s13014-018-1039-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.de Boer J, Wolf AL, Szeto YZ, van Herk M, Sonke JJ. Dynamic collimator angle adjustments during volumetric modulated arc therapy to account for prostate rotations. Int. J. Radiat. Oncol. Biol. Phys. 2015;91:1009–1016. doi: 10.1016/j.ijrobp.2014.11.020. [DOI] [PubMed] [Google Scholar]
  • 36.Murtaza G, Toftegaard J, Khan EU, Poulsen PR. Volumetric modulated arc therapy with dynamic collimator rotation for improved multileaf collimator tracking of the prostate. Radiother. Oncol. 2017;122:109–115. doi: 10.1016/j.radonc.2016.11.004. [DOI] [PubMed] [Google Scholar]
  • 37.Bokrantz R, van Amerongen JHM, Craft D. Technical note: Improving VMAT delivery efficiency by optimizing the dynamic collimator trajectory. Med. Phys. 2019;46:3877–3882. doi: 10.1002/mp.13671. [DOI] [PubMed] [Google Scholar]
  • 38.Hernandez V, et al. What is plan quality in radiotherapy? The importance of evaluating dose metrics, complexity, and robustness of treatment plans. Radiother. Oncol. 2020;153:26–33. doi: 10.1016/j.radonc.2020.09.038. [DOI] [PubMed] [Google Scholar]
  • 39.Gotstedt J, Hauer AK, Back A. Development and evaluation of aperture-based complexity metrics using film and EPID measurements of static MLC openings. Med. Phys. 2015;42:3911–3921. doi: 10.1118/1.4921733. [DOI] [PubMed] [Google Scholar]
  • 40.Masi L, Doro R, Favuzza V, Cipressi S, Livi L. Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med. Phys. 2013;40:071718. doi: 10.1118/1.4810969. [DOI] [PubMed] [Google Scholar]
  • 41.Mancosu P, et al. Collimator angle influence on dose distribution optimization for vertebral metastases using volumetric modulated arc therapy. Med. Phys. 2010;37:4133–4137. doi: 10.1118/1.3462560. [DOI] [PubMed] [Google Scholar]
  • 42.Bayouth JE. Siemens multileaf collimator characterization and quality assurance approaches for intensity-modulated radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2008;71:S93–97. doi: 10.1016/j.ijrobp.2007.07.2394. [DOI] [PubMed] [Google Scholar]
  • 43.Kumaraswamy LK, Schmitt JD, Bailey DW, Xu ZZ, Podgorsak MB. Spatial variation of dosimetric leaf gap and its impact on dose delivery. Med. Phys. 2014;41:111711. doi: 10.1118/1.4897572. [DOI] [PubMed] [Google Scholar]
  • 44.Gao Z, Szanto J, Gerig L. Using multileaf collimator interleaf leakage to extract absolute spatial information from electronic portal imaging device images. J. Appl. Clin. Med. Phys. 2005;8:1–9. doi: 10.1120/jacmp.v8i1.2226. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Klein EE, et al. Task group 142 report: Quality assurance of medical accelerators. Med. Phys. 2009;36:4197–4212. doi: 10.1118/1.3190392. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Information. (567.8KB, docx)

Data Availability Statement

Research data analyzed in this study are included as supplementary materials. Other data that support the findings of this study are available from the corresponding author upon reasonable request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES