Abstract
Purpose
We investigate two margin-based schemes for optimization target volumes (OTV), both isotropic expansion (2 mm) and beam-specific OTV, to account for uncertainties due to the setup errors and range uncertainties in pancreatic stereotactic pencil beam scanning (PBS) proton therapy. Also, as 2-mm being one of the extreme sizes of margin, we also study whether the plan quality of 2-mm uniform expansion could be comparable to other plan schemes.
Methods and Materials
We developed 2 schemes for OTV: (1) a uniform expansion of 2 mm (OTV2mm) for setup uncertainty and (2) a water equivalent thickness–based, beam-specific expansion (OTVWET) on beam direction and 2 mm expansion laterally. Six LAPC patients were planned with a prescribed dose of 33 Gy (RBE) in 5 fractions. Robustness optimization (RO) plans on gross tumor volumes, with setup uncertainties of 2 mm and range uncertainties of 3.5%, were implemented as a benchmark.
Results
All 3 optimization schemes achieved decent target coverage with no significant difference. The OTV2mm plans show superior organ at risk (OAR) sparing, especially for proximal duodenum. However, OTV2mm plans demonstrate severe susceptibility to range and setup uncertainties with a passing rate of 19% of the plans meeting the goal of 95% volume covered by the prescribed dose. The proposed dose spread function analysis shows no significant difference.
Conclusions
The use of OTVWET mimics a union volume for all scenarios in robust optimization but saves optimization time noticeably. The beam-specific margin can be attractive to online adaptive stereotactic body proton therapy owing to the efficiency of the plan optimization.
Introduction
Pancreatic adenocarcinoma is one of the most lethal cancers and has a 5-year survival rate of less than 10%.1 To improve the overall survival, there is increasing interest in exploring the hypofractionated regime of radiation therapy in treating local advanced pancreatic cancer. Stereotactic body radiation therapy (SBRT) is one of the techniques allowing delivery of higher dose to target and steeper dose falloff to normal tissues in 3 to 5 fractions, thereby maximizing the therapeutic ratio. The alternative treatment options are to deliver an ablative dose in 15 to 28 fractions2 for better local control.2
Proton therapy has the clinical advantage of depositing the entire prescription dose to the target and yielding no exit dose. Therefore, proton therapy can potentially reduce the dose to normal tissues, resulting in ameliorated local control and decreased toxicities. The study by Thompson et al3 shows that compared with conventional photon therapy, proton plans could significantly decrease the low and intermediate dose to critical organs (ie, duodenum, stomach and liver, etc), while maintaining the dose levels to target. Craneet al2 show that proton therapy can reduce exposure to normal tissue compared with intensity-modulated radiation therapy with 10 -15 mm margin size from GTV to PTV for pancreatic head cancer. Bouchard et al4 confirm this finding and claim it is feasible to boost the therapeutic ratio. Recently, Jethwa et al5 implemented a dosimetric analysis between intensity modulated proton therapy (IMPT) plans and volumetric modulated arc therapy plans for localized intact pancreatic cancer, showing IMPT offers superiority to intensity-modulated radiation therapy in reduction of dose to OARs.
Combining the techniques of SBRT with proton therapy as an emerging novel technique is believed to further boost the gain of local control and minimize toxicity.6 Sio et al7 provided a systematic quantitative comparison between photon stereotactic body radiation therapy (SBRT) and proton stereotactic body proton therapy (SBPT), and the results suggest comparable organs at risk (OAR) sparing in the high-dose region and improved dosimetric sparing for low- and medium-dose regions.
To accurately deliver the prescribed proton dose to the target, one must incorporate setup and range uncertainties. Owing to the inherent uncertainty of conversion of computed tomography (CT) image to stopping power, a range uncertainty of 3.5% is often considered in treatment planning.8, 9, 10 For instance, for margin-based treatment planning, a target volume used for optimization target volumes (OTV) iscreated from expansion from clinical target volume or gross tumor volumes (GTVs) . Sio et al7 compared the plan quality of 3 mm, 5 mm, and 7 mm expansions for SBPT in pancreatic cancer. Thompson et al3 used the expansion of 5 mm as photon therapy to produce OTV for proton pencil beam scanning (PBS). A beam-specific margins tehcnique is proposed to account for proton beam range uncertainties explicitly for by Park et al.11 They tested this technique on prostate cases to demonstrate its superiority to the geometric margin used by Thompson et al3 and Sio et al7 for target coverage. However, the efficacy and robustness of this method on the abdominal case remain to be investigated.
Other than designing margins explicitly to account for uncertainties in proton therapy, a novel way of treatment planning that incorporates the uncertainty into the process of optimization was recently introduced,12 namely robustness optimization (RO). It aims to achieve a robust dose distribution that is insensitive to setup and range uncertainties. The input parameters are determined based on the scenarios of estimated setup and range margins. For example, during the optimization process, to account for range uncertainties, the optimizer automatically expands the margin implicitly without specifying the OTV.5,9 If an OAR is adjacent to the target, instead of placing the margins directly, the optimizer will place dose falloff to shape the dose distribution.9 Although robust optimization outperforms margin-based methods in the robustness of target coverage,13 a large number of error scenarios impose a huge computation burden on clinical implementation of robust optimizations.
The amount of setup-uncertainty margins used in margin-based planning or robustness optimization planning are often derived from imaging guidance (ie, stereoscopic, kilovoltage, 2-dimensional x-ray imaging).5 With the help of 3-dimensional imaging guidance and surrogate fiducials implanted close to the lesion, the setup uncertainties can be further reduced. It is yet known how the reduction of setup margin can benefit the robustness optimization process of proton treatment planning for abdominal targets. To battle with the duodenum toxicity, a novel, absorbable iodinated polyethylene glycol–based hydrogel for tissue marking and spacing was studied to further increase the separation between panceas and duodenum.14,15 The feasibility of incorporating this spacer into proton planning has not systematically investigated. As a first step, if the mere uncertaines other than range uncertainty are studied, it remains unclear whether the spacer can behave a “buffer” to help to reduce the duodenum toxicity, as in photon therapy.
We also investigated 2 margin-based optimization schemes, isotropic expansion and beam-specific OTV, to account for uncertainties associated with setup errors and range uncertainties in pancreatic stereotactic PBS proton therapy. The size of the expansion is inferred from the tolerance of the 3-dimensional cone-beam CT imaging guidance device used in our center. The robustness, target dose coverage, and OAR sparing of the plans generated by the 2 margin-based optimization schemes are quantitatively compared with plans using robustness optimization.
Methods and Materials
Six patients with clinically diagnosed, localized, advanced pancreatic adenocarcinoma who underwent SBRT active breath control technique were chosen in this planning study. All patients were positioned in supine with arms above the head under breath hold during the simulation.
The GTV, OAR, i.e. duodenum, small bowel, etc were contoured by physicians. GTV volumes ranged from 7.6 mL to 82.2 mL. The OARs, including duodenum, small intestine, stomach, kidneys, and spinal cord, were delineated. The CT images and structures were imported into Raystation 9A (RaySearch Laboratories AB, Stockholm, Sweden) for PBS proton treatment planning.
All patients were prescribed with a total dose of 33 Gy (RBE) in a treatment course of 5 fractions. The nominal goal of dose coverage of the GTVs should be 98% of volume receiving at least 100% of the prescribed dose (D98). More details can be found below for each optimization scheme. The OAR constraints include the volume of receiving 100% prescribed dose V33Gy(RBE) < 1 cm3; V20 Gy(RBE) < 20 cm3 and V15 Gy (RBE) < 15 cm3 in compliance with our institutional guidelines.
Beam configuration
The plans were designed to be delivered via PBS using the Hitachi Probeat-CR system equipped with volumetric imaging guidance of the cone-beam computed tomography system. Two posterior oblique fields at gantry of 210 and 150 degrees were used in the treatment planning for all patients except one, who was planned with a posterior (180) and posterior oblique beam (150) to reduce dose to the right kidney. The spot scanning pattern is set hexagonal. The spot size is decided by the optimization algorithm. The use of range shifter is determined on a patient-specific basis.
OTV definitions
Two OTV schemes were investigated. The first scheme is the OTV2mm, which was designed in a fashion of geometric uniform 2-mm expansion around the GTV. With the help of cone-beam computed tomography and implanted fiducial markers, a tolerance of 2 mm was then used in the treatment planning process for the setup uncertainty. The second scheme is a patient-specific scheme (OTVWET), proposed by Park et al11 and based on water-equivalent thicknesses (WETs). Along the beam direction, distal and proximal WETs were determined based on the stoichiometric calibration method and converted to geometric margins by multiplying 3.5% added to GTV as the OTVWET. On the beam's eye-view directions, an expansion of 2 mm was used to account for lateral setup uncertainty.
Plan optimization
The OTV2mm and OTVWET were used for the target dose coverages in the plan optimization. The homogeneous dose objectives (minimum, maximum, and uniform doses) were applied to the whole OTV2mm, and 2 beam-specific OTVWET with equal dose weight. The OTV-optimized plans were compared with the robustness optimization (RO) method, which incorporates the setup and range uncertainties. In this study, a setup uncertainty of 2 mm and a range uncertainty of 3.5% on GTV were used in the RO optimization process. A total of 42 scenarios have to be computed to account for all possible uncertainties, including isocenter with no shifts, and shifted toward patient's anterior, posterior, right, left inferior directions, and diagonal directions (14 scenarios), each of which is with 3 scenarios of −3.5%, 0, +3.5% scaling of WET. The single-field optimization was used for all plans to have the uniform dose distribution in target volumes from each field.
Plan evaluation
The target (GTV) dose coverage of the target volume (GTV) and the dose sparing of OARs were compared among 3 schemes for 6 patients. A Bonferroni-corrected t test was used to evaluate the statistical significance.
There is an agreement that plan robustness as a plan quality metric should be included.16 Although there is no consensus on what exact scenarios should be included in the evaluation for robustness evaluation,16 a separate exam of setup and range uncertainties were used. To evaluate the dose coverage of target under the various conditions of uncertainties, simultaneously 14 setup errors (in both parallel and diagonal directions) and 3 range uncertainties (including a nominal plan with no range uncertainty) with 42 totality of scenarios are examined. All plans aim to achieve at least 95% of the volume of the GTV receiving 100% of the prescribed dose.
The plan robustness could also be evaluated using the dose-volume histogram (DVH) deviation, “bandwidth,” due to the perturbations of isocenter shift and range uncertainties for target and duodenum. A separate independent of 6 setup errors and 3.5% for an under and over range were analyzed. Of all the 8 scenarios, the bandwidth of DVH is defined as the standard deviation at specified dose parameters.
(1) |
Because one of the most main critical structures duodenum, abuts tumor target, a quantity, termed as dose spread function, was proposed to describe dose falloff in the abutment region. In Fig. 1a, the spatial relationship between a typical 100% iso-dose line and duodenum is shown. A line dose profile can be sampled and shown in Fig. 1b. The dose profile can be obtained by fitting the sampled line dose. By differentiating the fitting line dose, the dose spread function (DSF) can be analytically obtained. Mathematically, this can be expressed as follows:
(2) |
Here r is the distance from either end of the dose line. LD refers to the line dose profile. Via DSF, the steepness of the dose falloff gradient, can be described quantitatively. For example, on Figure 1c, the FWHM is referring to the width needed to reduce 100% dose to 50%, whereas FW20M presents the width for the dose to drop from 100% to 20%. To show the practical use of DSF, the line dose, representing dose dropping from 100%, is chosen to cross the abutment region between 100% isodose to duodenum structure. The dose distribution, DVH and line dose profile were exported from Raystation TPS and analyzed by using in-house MATLAB tools.
Results
The derived OTVWET has a more substantial expansion: >2 mm from GTV due to targets being deep-seated in general. It is estimated that the geometric distance from the beam entrance to GTV distal is about 16 to 19 cm. Not like uniform expansion, OTVWET has less conformality to target due to the expansion for range uncertainties. This can be observed from one of the patients shown in Figure 2. The more expansion in the beam direction, the more degradation of dose conformality can be seen in Figure E1 a of the Supplementary Materials.
Figure 3 a exemplifies the DVH similarities for target coverage among 3 plans for one of the patients. Table E1 in the Supplementary Materials summarizes the results of the planning comparison. DVHs of the GTV show that for D98, Dmax and Dmean, these 3 plans have very similar coverage. As expected, no statistically significant differences among the plans were found for GTVs, except D98 for RO and OTVWET plans. A slightly larger Dmax can be found for the plan using robust optimization (RO plan).
Figure 3b shows an example of dose sparing for critical OARs achieved by 3 plans. Due to a smaller margin than OTVWET, the OTV2mm yields the best dose avoidance for duodenum in the high, medium, and low dose regions. In addition to the setup uncertainties of 2 mm, the RO plan also considers the range uncertainties explicitly, leading to the performance of dose sparing similar to the plan of OTVWET. However, in this case, the plan of OTVWET is the worst of sparing small bowel compared with the other 2. According to Table E1 in the Supplementary Materials, OTV2mm plans show statistically significant in dose sparing of the duodenum and small bowel over OTVWET due to the smaller margin in OTV2mm. The plans of OTV2mm also outperform the RO plans in high dose region sparing for small bowel and stomach.
Among all the plans, the response of the nominal plans to perturbations may be plan scheme specific. For example, for target coverage in Figure 4, the compactness or bandwidth of D98 and D50 on DVH shows the least for RO plans. It demonstrates that the RO plans are more forgivable to the setup and range uncertainties as expected. The bandwidths of OTV2mm and OTVWET are similar, mainly, for D50, the widths are 3.80% and 3.84% for OTV2mm and OTVWET, respectively. In other words, the worst resultant target dose could be under- or overdose around 4%. The bandwidths at D98 for OTV2mm and RO are 4.8% and 2.8%, respectively. For this case, the loss of the plan quality due to the perturbation could be up to around 5% for OTV2mm, while for RO plan, such loss could be around 3%. According to our plan objectives, the goal of the target coverage could be maintained.
The average passing rate of target coverage for all patient with the goal of 95% target covered by 100% prescribed dose is only 19% in OTV2mm compared to 100% for OTVWET and RO. This result shows that OTV2mm is not robust enough to offset the perturbation to the dose coverage. One example can be found in Figure 5, where an intentional isocenter shift posteriorly 2 mm and overranged by 3.5% are introduced. In this case, the target dose coverage (D95) can dramatically drop to 70%, leading to huge plan quality degradation.
The impact of perturbation on OAR (i.e. duodenum) could be revealed in the example case of Figure 6. The OTVWET (b) plan has the most significant bandwidth of medium-dose V20Gy(RBE) among the 3 plans. In the high dose region, the use of OTV2mm (a) has more advantages compared with the other plans in response to perturbation scenarios. Again, this is due to a smaller margin of plans of OTV2mm. For the low dose region, 3 plans show similar performance.
Table E2 in the Supplementary Materials summarizes the statistical significance of the resultant impact on the duodenum. For V20Gy(RBE), OTV2mm plan shows the difference from the RO plans. No other significant difference was found among the 3 plans.
The DSF from the RO plan in Figure E2 in the Supplementary Materials shows slightly larger than the other 2, indicating that RO has a more gradual dose falloff region on this particular dose plane. In Table E3 in the Supplementary Materials OTV2mm shows the fastest dose-off among all the patients and plans. For FW20M, the RO plans need about more than 2 mm space to reduce the dose to 20% of the prescribed.
Discussion
Designing an appropriate margin for treatment plan optimization in proton therapy is quite challenging.9,10,16 Although accurate in delivery of high dose to target, the proton beam is also sensitive to any variation along the beam path. Such variation could be due to change of anatomy, setup, and range uncertainties.17 Thus, image guidance in proton therapy plays a crucial role in clinical to achieve the intended treatment plan dose distribution.
The plan quality of OTV2mm shows it not suitable for clinical use. Our motivation to attempt an aggressive 2 mm margin size in this study is rooted in 3 aspects. First, during the SBRT planning, a higher dose to the tumor while a relatively lower dose to the OARS is highly desired. SBRT is ideally suitable for tumors in parallel organs. The pancreas, however, is in close promity to serially functioning OARs (ie, duodenum, small bowel). The ablative BED dose around 106 Gy with current fractionation scheme to these OARs can produce impairment in the organ function. As such, limiting the OAR dose in LAPC SBPT has become increasingly crucial to the improvements in oncological endpoints.18 A more aggressive reduction of margin size is an alternative approach to sparing more OARs dose.
Second, it is worth noting that OTV2mm plan is sensitive to range uncertainties owing to its sufficient margin size to cover the uncertainty along the beam direction. One way to improve the robustness and maintain the low toxicity OTV2mm to duodenum is to place hydrogel not only in the space between duodenum and pancreas, but also along the beam direction by any means. In other words, the function of hydrogel in proton therapy could be viewed as 2-folds: to reduce the toxicity and uncertainties.
Third, the geometric limit of combing fiducial markers and a developed real-time gated floruoscopy at our institution could be within 2 to 3 mm.18, 19, 20 Therefore, as a motion management strategy, OTV2mm as the extreme case of margin design is investigated, whereas this aggressive margin size is not realistic in current proton community.
In line with tolerance allowed by the 2-dimensional imaging device, various sizes of margins up to 5 mm or more expansions from clinical target volume or GTV for pancreas proton therapy have been implemented.3,5,7,21 As for comparison with OTV2mm, we implemented OTV5mm plan study as well. The summary of all OTV5mm plans is updated in Table E1 in Supplementary Materials. One of the patient plans has been shown in Figure 7.
It can be seen that target coverage (D98) is similar to Figure 4, but the overdose of duodenum could be up to 5 times compared with OTV2mm. The evaluation of the robustness shows that using a 5-mm margin size is immune to the perturbations of 2-mm setup errors and 3.5% range uncertainties. In this study, it is our expectation that with larger margin size, we are sacrificing OAR sparing for robustness. The robustness evaluation for OTV5mm plans shows resiliency of the plans to the setup error of 2 mm and range uncertainties with passing rate close to 100%.
Owing to the combined uncertainties of both patient positioning and proton range, the construction of a beam-specific OTV is more appropriate.22 In addition to the margin-based a priori method OTVWET, a robustness optimization that considers the range uncertainties and setup error without specifying the margins for optimization was proposed and implemented.9,12 By minimizing the maximum of objective function for scenarios accounting for setup and range uncertanities, robustness optimization could yield plans that are resilient to the perturbation within the specification. Comparing the plans between OTVWET and RO in DSF, it was found that RO plans have less dose gradient in the beam direction. For the case shown herein, the dose falloff is roughly steeper for RO plan than OTVWET. As a result, when beams overshoot or undershoot, the dose distribution can be only moderately affected by shifting dose distribution as shown in Figure 8. In other words, RO plan can automatically extend the irradiated region distally along the beam direction to compensate the dosimetric perturbations introduced by range uncertainties. Meanwhile, RO plans can reshape the dose distribution accordingly.9 The OTVWET plans predefine the margin with priori knowledge of range uncertainties and shift the dose distribution.22 In the Figure 8c, this is demonstrated by a bare coverage of prescribed dose to GTV.
However, one disadvantage of this method is that it requires computation of all the scenarios to determine the maximum of the objective function. In other words, it could be rather time-consuming to minimize or optimize the objective function for a plan. For instances, using Monte Carlo optimization with the running time about 2 hours, while using beam-specfic OTVwet method, the computation time can be decreased to within a half hour. In the pipeline of online adaptive proton therapy, robustness optimization becomes clinically challenging.23
Robustness optimization and margin-based plans achieve a similar target dose coverage in D98, Dmean. In terms of OAR dose sparing of the duodenum, OTV2mm outperforms the other types of plans owing to its smaller expansion. On the other hand, OTVWET plans show inferior to other plans owing to the relatively large expansion. However, the RO plans are robust to setup and range uncertainties without sacrificing the quality of the plans too much. The OTV2mm plans fail most of the scenarios in the robustness evaluation owing to the plans that are extremely sensitive to the perturbations from range uncertainties. The OTVWET plans are insensitive to setup errors and range uncertainties for the beam arrangements used in the pancreatic patients. The dose spread function proposed in the study reveals that RO plans have less gradient of dose falloff at the specified location, indicating that more room between critical structure and 100% isodose line is needed to spare the duodenum in general. The OTV2mm and OTVWET plans have only one scenario to compute, while the RO scheme has 21 scenarios. If the Monte Carlo dose engine was used, it is anticipated that the RO process would take even longer.24 In summary, the quality of OTVWET plans falls in the category between OTV2mm and RO; it has similar target dose coverage compared with the other 2 but has the inferior normal tissue sparing compared with OTV2mm.
The OTVWET plan explicitly accounts for setup and range uncertainties; it is less susceptible to dosimetric perturbation. As indicated in the results, OTVWET plans have the degradation of quality in robustness evaluation, although the light computation is maintained. In general, the plan quality of OTVWET is closer to RO plans than that of OTV2mm plans.
The proposed quantity DSF can correlate the spatial information of anatomical structure and dose distribution. The edge spread function (ie, dose profile) is used to quantify the dose-response at the edge of the critical structure. At the same time, DSF indicates the rate of dose falloff on the selected image with spatial and directional information. There are other useful tools to quantify the dose falloff behavior,25 which may only focus on the distance between the isodose lines, not on the simultaneous effect of the anatomic specific and dosimetric performance. It is noted that recent studies from Rao et al14, 15, 26, 27, 28, 29 showcased a novel biodegradable hydrogel material injected at head-of-duodenum interface that can limit the dose to the duodenum. However, it is necessary to correlate the amount of injected hydrogel with the dosimetric consequence for each patient. By combining the information from DSF in a volumetric format on a patient-specific basis, the injection of the hydrogel can be optimized more accurately. This will remain in our future studies.
In this study, the WET calculation is based on the stochiometric conversion from single-energy CT images, which is believed to introduce 3.5% uncertainties.8,30 To further improve the dose sparing, a possible method is to incorporate dual-energy CT images into this study. It is accepted that the DECT technique can decrease the range uncertainty from 3.5% to 2%.31, 32, 33, 34, 35, 36, 37
To the authors’ best knowledge, this study is the first one to compare the 2 types of proton treatment planning techniques for pancreas cancer: margin-based versus robustness optimization. However, this study has several limitations. First, the optimization of beam angle selection is not implemented, as it shows an essential role in the quality of the plans.22 Second, the dose falloff region of interest is selected at the abutment of 100% isodose line and duodenum. The choice of this region could be optimized in future investigation.38
Conclusions
We evaluated the plan quality among 3 optimization schemes using uniform expandsion OTV2mm, beam-specific OTVWET, and robustness optimization on GTV. Although the margin-based method (OTV2mm) provides superiority in fast computation and low toxicity, the quality of plans fails to match the ones from robustness optimization in the evaluation of susceptibility to perturbation. A beam-specific margin also remains further investigations for online adaptive stereotactic body proton therapy owing to the efficiency of the plan optimization.
Footnotes
Sources of support: Research reported in this publication was supported by the National Cancer Institute of the National Institutes of Health (award number R37CA229417).
Disclosures: None.
Research data are stored in an institutional repository and will be shared upon request to the corresponding author.
Supplementary material associated with this article can be found in the online version at doi:10.1016/j.adro.2021.100757.
Appendix. Supplementary materials
References
- 1.Siegel RL, Miller KD. Cancer Statistics, 2020. CA Cancer J Clin. 2020:1–24. doi: 10.3322/caac.21590. [DOI] [PubMed] [Google Scholar]
- 2.Crane CH. Hypofractionated ablative radiotherapy for locally advanced pancreatic cancer. J Radiat Res. 2016;57:i53–i57. doi: 10.1093/jrr/rrw016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Thompson RF, Zhai H, Both S. A dosimetric comparison of proton and photon therapy in unresectable cancers of the head of pancreas. Med Phys. 2014;41:1–10. doi: 10.1118/1.4887797. [DOI] [PubMed] [Google Scholar]
- 4.Bouchard M, Amos RA, Briere TM, Beddar S, Crane CH. Dose escalation with proton or photon radiation treatment for pancreatic cancer. Radiother Oncol. 2009;92:238–243. doi: 10.1016/j.radonc.2009.04.015. [DOI] [PubMed] [Google Scholar]
- 5.Jethwa KR, Tryggestad EJ, Whitaker TJ. Initial experience with intensity modulated proton therapy for intact, clinically localized pancreas cancer: Clinical implementation, dosimetric analysis, acute treatment-related adverse events, and patient-reported outcomes. Adv Radiat Oncol. 2018;3:314–321. doi: 10.1016/j.adro.2018.04.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Ng SP, Herman JM. Stereotactic radiotherapy and particle therapy for pancreatic cancer. Cancers (Basel) 2018:10. doi: 10.3390/cancers10030075. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Sio TT, Merrell KW, Beltran CJ. Spot-scanned pancreatic stereotactic body proton therapy: A dosimetric feasibility and robustness study. Phys Medica. 2016;32:331–342. doi: 10.1016/j.ejmp.2015.12.009. [DOI] [PubMed] [Google Scholar]
- 8.Paganetti H. Range uncertainties in proton therapy and the role of Monte Carlo simulations. Phys Med Biol. 2012;57:R99–R117. doi: 10.1088/0031-9155/57/11/R99. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Unkelbach J, Paganetti H. Robust proton treatment planning: Physical and biological optimization. Semin Radiat Oncol. 2018;28:88–96. doi: 10.1016/j.semradonc.2017.11.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Moyers MF, Miller DW, Bush DA, Slater JD. Methodologies and tools for proton beam design for lung tumors. Int J Radiat Oncol Biol Phys. 2001;49:1429–1438. doi: 10.1016/s0360-3016(00)01555-8. [DOI] [PubMed] [Google Scholar]
- 11.Park PC, Zhu XR, Lee AK. A beam-specific planning target volume (PTV) design for proton therapy to account for setup and range uncertainties. Int J Radiat Oncol Biol Phys. 2012;82:329–336. doi: 10.1016/j.ijrobp.2011.05.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Fredriksson A, Forsgren A, Hårdemark B. Minimax optimization for handling range and setup uncertainties in proton therapy. Med Phys. 2011;38:1672–1684. doi: 10.1118/1.3556559. [DOI] [PubMed] [Google Scholar]
- 13.Cao N, Saini J, Bowen SR, Apisarnthanarax S, Rengan R, Wong TP. CTV-based robustness optimization versus PTV-based conventional optimization for intensity modulated proton therapy planning. Int J Radiat Oncol. 2015;93:E617–E618. [Google Scholar]
- 14.Rao AD, Shin EJ, Beck SE. Demonstration of safety and feasibility of hydrogel marking of the pancreas-duodenum interface for image guided radiation therapy (IGRT) in a porcine model: Implications in IGRT for pancreatic cancer patients. Int J Radiat Oncol Biol Phys. 2018;101:640–645. doi: 10.1016/j.ijrobp.2018.02.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Rao AD, Feng Z, Shin EJ. A novel absorbable radiopaque hydrogel spacer to separate the head of the pancreas and duodenum in radiation therapy for pancreatic cancer. Int J Radiat Oncol Biol Phys. 2017;99:1111–1120. doi: 10.1016/j.ijrobp.2017.08.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Langen K, Zhu M. Concepts of PTV and robustness in passively scattered and pencil beam scanning proton therapy. Semin Radiat Oncol. 2018;28:248–255. doi: 10.1016/j.semradonc.2018.02.009. [DOI] [PubMed] [Google Scholar]
- 17.Meyer J, Bluett J, Amos R. Spot scanning proton beam therapy for prostate cancer: Treatment planning technique and analysis of consequences of rotational and translational alignment errors. Int J Radiat Oncol Biol Phys. 2010;78:428–434. doi: 10.1016/j.ijrobp.2009.07.1696. [DOI] [PubMed] [Google Scholar]
- 18.Narang AK. Technical aspects of modern radiation therapy for pancreatic adenocarcinoma: Field design, motion management, dosing, and concurrent therapy. Ann Pancreat Cancer. 2018;1:21. [Google Scholar]
- 19.Moningi S, Marciscano AE, Rosati LM. Stereotactic body radiation therapy in pancreatic cancer: The new frontier. Expert Rev Anticancer Ther. 2014;14:1461–1475. doi: 10.1586/14737140.2014.952286. [DOI] [PubMed] [Google Scholar]
- 20.Yoshimura T, Shimizu S, Hashimoto T. Analysis of treatment process time for real-time-image gated-spot-scanning proton-beam therapy (RGPT) system. J Appl Clin Med Phys. 2019:1–11. doi: 10.1002/acm2.12804. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Houweling AC, Crama K, Visser J. Comparing the dosimetric impact of interfractional anatomical changes in photon, proton and carbon ion radiotherapy for pancreatic cancer patients. Phys Med Biol. 2017;62:3051–3064. doi: 10.1088/1361-6560/aa6419. [DOI] [PubMed] [Google Scholar]
- 22.Tryggestad EJ, Liu W, Pepin MD, Hallemeier CL, Sio TT. Managing treatment-related uncertainties in proton beam radiotherapy for gastrointestinal cancers. J Gastrointest Oncol. 2019:212–224. doi: 10.21037/jgo.2019.11.07. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Liu W, Liao Z, Schild SE. Impact of respiratory motion on worst-case scenario optimized intensity modulated proton therapy for lung cancers. Pract Radiat Oncol. 2015;5:e77–e86. doi: 10.1016/j.prro.2014.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Buti G, Souris K, Montero AMB, Lee JA, Sterpin E. Towards fast and robust 4D optimization for moving tumors with scanned proton therapy. Med Phys. 2019;46:4–6. doi: 10.1002/mp.13850. [DOI] [PubMed] [Google Scholar]
- 25.Sung KH, Choi YE. Dose gradient curve: A new tool for evaluating dose gradient. PLoS One. 2018;13:1–16. doi: 10.1371/journal.pone.0196664. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Feng Z, Rao AD, Cheng Z. Dose prediction model for duodenum sparing with a biodegradable hydrogel spacer for pancreatic cancer radiation therapy. Int J Radiat Oncol Biol Phys. 2018;102:651–659. doi: 10.1016/j.ijrobp.2018.07.184. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Kim SH, Ding K, Rao A. EUS‐guided hydrogel microparticle injection in a cadaveric model. J Appl Clin Med Phys. 2021;22:83–91. doi: 10.1002/acm2.13266. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Rao AD, Shin EJ, Meyer J. Evaluation of a novel absorbable radiopaque hydrogel in patients undergoing image guided radiation therapy for borderline resectable and locally advanced pancreatic adenocarcinoma. Pract Radiat Oncol. 2020;10:e508–e513. doi: 10.1016/j.prro.2020.01.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Kerdsirichairat T, Narang AK, Thompson E. Feasibility of using hydrogel spacers for borderline-resectable and locally advanced pancreatic tumors. Gastroenterology. 2019;157:933–935. doi: 10.1053/j.gastro.2019.07.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Yang M, Virshup G, Clayton J, Zhu XR, Mohan R, Dong L. Theoretical variance analysis of single- and dual-energy computed tomography methods for calculating proton stopping power ratios of biological tissues. Phys Med Biol. 2010;55:1343–1362. doi: 10.1088/0031-9155/55/5/006. [DOI] [PubMed] [Google Scholar]
- 31.Yunhe X, Ainsley C, Yin L. Ex vivo validation of a stoichiometric dual energy CT proton stopping power ratio calibration. Phys Med Biol. 2018;63 doi: 10.1088/1361-6560/aaae91. [DOI] [PubMed] [Google Scholar]
- 32.Hünemohr N, Krauss B, Tremmel C, Ackermann B, Jäkel O, Greilich S. Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogates. Phys Med Biol. 2014;59:83–96. doi: 10.1088/0031-9155/59/1/83. [DOI] [PubMed] [Google Scholar]
- 33.Han D, Siebers JV, Williamson JF. A linear, separable two-parameter model for dual energy CT imaging of proton stopping power computation. Med Phys. 2016;43:600–612. doi: 10.1118/1.4939082. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Zhang S, Han D, Williamson JF. Experimental implementation of a joint statistical image reconstruction method for proton stopping power mapping from dual-energy CT data. Med Phys. 2019;46:273–285. doi: 10.1002/mp.13287. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Bär E, Lalonde A, Royle G, Lu HM, Bouchard H. The potential of dual-energy CT to reduce proton beam range uncertainties. Med Phys. 2017;44:2332–2344. doi: 10.1002/mp.12215. [DOI] [PubMed] [Google Scholar]
- 36.Zhang Shuangyue, Han Dong, Politte David, Williamson Jeffrey, O’Sullivan Joseph. Impact of joint statistical dual-energy CT reconstruction of proton stopping power images: Comparison to image- and sinogram-domain material decomposition approaches. Medical Physics. 2018;45(5):2129–2142. doi: 10.1002/mp.12875. In this issue. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Han Dong, Porras-Chaverri Mariela, O’Sullivan Joseph, Politte David, Williamson Jeffrey. Technical Note: On the accuracy of parametric two-parameter photon cross-section models in dual-energy CT applications. Medical Physics. 2017;44(6):2438–2446. doi: 10.1002/mp.12220. In this issue. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Hooshangnejad Hamed, Youssefian Sina, Guest James, Ding Kai. FEMOSSA: Patient-specific finite element simulation of the prostate–rectum spacer placement, a predictive model for prostate cancer radiotherapy. Medical Physics. 2021;48(7):3438–3452. doi: 10.1002/mp.14990. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.