Abstract
The performance of one-dimensional photonic crystal for plasma cell application is studied theoretically. The geometry of the structure can detect the change in the refractive index of the plasma cells in a sample that infiltrated through the defect layer. We have obtained a variation on the resonant peak positions using the analyte defect layer with different refractive indices. The defect peak of the optimized structure is red-shifted from 2195 to 2322 nm when the refractive index of the defect layer changes from 1.3246 to 1.3634. This indicates a high sensitivity of the device (S = 3300 nm/RIU) as well as a high Q-factor (Q = 103). The proposed sensor has a great potential for biosensing applications and the detection of convalescent plasma.
Keywords: Biosensor, Photonic crystal, Plasma cells, Blood components, Transfer matrix method (TMM)
Introduction
Blood plasma is one of the human blood components that plays an important role in the osmotic effect inside the vessels (Kenner 1989). The concentration of blood plasma is an indication for some diagnoses such as disorder of cardiovascular, leukemia, and autoimmune (Cakmak et al. 2013). Also, exchanging convalescent plasma has attracted more attention as a potential therapy for COVID-19 that created a sense of panic globally (Chen et al. 2020; Shen et al. 2020; Keith et al. 2020; Ahn, et al. 2020). Researchers are attempting to get immunity from recovered COVID-19 patients with "convalescent" plasma (Kupferschmidt and Cohen 2020). Therefore, studying and detecting an abnormal concentration of blood plasma is vitally important for COVID-19 therapy. One of the efficient approaches to estimate the concentration of blood plasma is by measuring its refractive index (RI) as RI of the blood plasma is a function of its concentration ( according to this equation (El-Khozondar et al. 2019):
1 |
In the last few decades, 1D, 2D, and 3D photonic crystals (PCs) and phononic crystals have been received extensive attention in sensing applications as a biochemical sensor, salinity sensor, and filters (El-Khozondar et al. 2019; Arafa et al. 2017; Beheiry et al. 2010; Natesan et al. 2018; Arunkumar et al. 2019; Amiri et al. 2019; Elsayed and Mehaney 2018; Zaky et al. 2020; Abd El-Ghany et al. 2020; Zaky and Aly 2020; Khateib et al. 2020). The PCs are periodic structures of alternating dielectrics (Abd El-Ghany et al. 2020; Aly and Elsayed 2012; Aly et al. 2020a; Aghajamali et al. 2015; Nayak et al 2017, 2018, 2019; Guo et al. 2020a; Srivastava and A. J. J. O. S. Aghajamali, and N. Magnetism 2017; Shiri et al. 2019; Zhou et al. 2021), nanocomposites (Aly et al. 2018a), superconductors (Aly et al. 2019, 2018b), or metal with different refractive indices in one, two or three directions (Yablonovitch 1987; Shaban et al. 2017; Winn et al. 1998). There is a forbidden region for a specific range of wavelengths in a propagated electromagnetic wave. this forbidden region is called the photonic bandgap (PBG) (Aly 2008; Aly and Hanafey 2011). Owing to the multiple Bragg scattering mechanism, This PBG is created (Hassan et al. 2015; Zaky and Aly 2021a, 2021b; Zaky et al. 2021).
Photons are localized through the PBG as a resonant peak by introducing a defect layer inside the PCs (Aly and Elsayed 2012). Any change in the concentration of the defect layer samples increases the refractive index (Aly et al. 2020b). This increase in the refractive index causes strong electric field confinement and shifts the defect peak to the longer wavelengths (red-shift) according to Bragg law (Aly et al. 2020b). Therefore, the defective PCs are used as an optical refractive index sensor (Fan et al. 2015).
In this work, we employed a one-dimensional photonic crystal 1D-PC as a blood plasma sensor. The blood plasma layer is introduced as a defect layer within the 1D-PC and the change in the RI is investigated. Inside the PBG, a resonant defect peak is created (Aly et al. 2020b; Peng et al. 2018). By changing the concentration of the blood plasma sample layer, the RI changes, as well as the resonant defect peak position will be red-shifted because of the strong light confinement through it. Recently, many researchers developed experimentally, highly stable, low cost, and highly efficient 1D-PC for different applications (Peng et al. 2018; Faneca et al. 2018).
In 2019, Khozondar et al. (El-Khozondar et al. 2019) proposed a plasma blood sensor by using a ternary photonic crystal. They achieved a sensitivity of 51.49 nm/RIU. Ahmed et al. (Ahmed et al. 2019) used D-shaped photonic crystal fiber (PCF) to sense the blood plasma. They estimated a sensitivity value equal to 5.8 nm/gL-1. In 2020, Ying, et al. proposed a PCF filled with liquid crystal as a RI sensor with a sensitivity of 2275 nm/RIU (Guo et al. 2020b). Yongbo, et al. designed PCF based on surface plasmon resonance that has a high sensitivity of 21 200 nm/RIU (Fu et al. 2020). Abadla, et al. investigated a binary PC comprising a defect layer for hemoglobin detection with sensitivity 167 nm/RIU (Abadla and Elsayed 2020). As shown in Table 3, The proposed sensor was designed with high sensitivity, low detection limit, and high quality-factor comparing with many published reports (Abadla and Elsayed 2020; Mohamed et al. 2016; Banerjee 2009).
Table 3.
Reference | S (nm/RIU) | Q-factor | DL(RIU) | Sensing analyte |
---|---|---|---|---|
El-Khozondar et al. (2019) | 51 | *** | *** | Blood plasma |
Abadla and Elsayed (2020) | 167 | *** | *** | Hemoglobin concentration |
Aly et al. (2020b, 2019) | 1100 | 106 | 10–8 | Glucose concentration |
Banerjee (2009) | 24 | *** | *** | Blood glucose |
Mohamed et al. (2016) | 1118 | *** | 10–3 | Blood glucose |
This work | 3300 | 103 | 10–5 | Blood plasma |
*** means not be calculated
Design and theoretical model
Figure 1 shows the schematic diagram of the proposed sensor which consists of . The first layer A is , the second layer B is P is a defect layer filled with a sample of blood analytes with different concentrations. The thickness of the layers are , and dp, while the RI are , and np, respectively. N is the number of periods. The analyte of plasma cells is sandwiched as a defect layer between two identical binary periodic structures. This structure was deposited on (layer S) as a substrate with a thickness of 2 mm. The PC was theoretically designed (Sharma et al. 2012). We select and as dielectric materials because they have high contrast in refractive indices and will make a wide photonic band gap in the selected region (infra-red) with low loss (near to zero Extinction coefficient) in this regions, as will mentioned in the next section (Nunley, et al. 2016; Kruschwitz and Pawlewicz 1997).
Homogeneous films of can be prepared by spray pyrolysis deposition, spin-coating, radio frequency (RF) magnetron sputtering, plasma-enhanced chemical vapor deposition (PECVD), and some others as we listed in our last paper (Aly et al. 2020b). During the fabrication of the design, a layer of a certain material can be deposited as a defect film, and this film can be etched with a strong chemical solution (Etching prosses). Then, the blood sample is injected into this etched layer.
Transfer matrix method (TMM) is used to study the interaction between the incident optical wave (IR-region) and the current PC structure as the following equation (Li and Lin 2003):
2 |
where , , and are the transfer matrix elements, are the characteristic matrix for , and the defect layer which are given by:
3 |
where Φi is the phase difference at each layer given by:
q for s-polarization of light is qi = ni cosδi. Also, δ1, δ2, and δp are the incident angle on the Ge, CaF2, and P layers according to Snell’s law:
4 |
For N periods of the structure, we used the Chebyshev polynomials of the second kind to get the matrix ()(Wu et al. 2005). Now, the transmittance coefficient can be given by:
5 |
where the values of q for the s-polarization of air (qs) and substrate (qs) can be calculated as: , and . Finally, the transmittance can be written in terms of transmission coefficient as:
6 |
Results and discussions
The proposed structure consists of two layers and the lattice parameter is , where ,, m, and . Additionally, Fig. 2 shows various RI values of the plasma samples which depend on the concentration of the plasma sample in the blood (cp) that measured in the unit of (g/L) according to Eq. 1.
The Matlab simulation tool is used to study the transmission spectra of the proposed defective PC (Aly and Elsayed 2012; Shaban et al. 2020). To enhance the sensing performance of our sensor, three types of transmission calculations were performed. Firstly, we study the effect of the number of layers (N) at the normal incident light and dp = 1100 nm on sensor performance. Secondly, we repeat the above calculations for different thicknesses of the blood plasma sample by keeping the number of periods at the selected optimized value. Thirdly, we do the same calculations for different incident angles by keeping both the number of periods and thickness at the selected optimized values.
Figure 3 shows the transmittance of the structure as a function of different incident wavelengths at 0, 5, 10, 15, and 20 g/L plasma concentration with m and incident angle δ0 = 0˚. The PBG extended within the infrared regions from 1695 nm(left) to 2380 nm (right) in the absence of the defect layer(black line). The width of the bandgap is equal to 685 nm with cutoff transmittance amplitude through it. Multiple Bragg scattering between the incident, scattered and reflected light prevents the incident light to travel through the structure (Mulder et al. 2012). This large width of the PBG due to the high contrast between the two materials of the unit cell refractive indices.
To evaluate the sensor performance, some parameters such as sensitivity, quality factor, detection limit, and figure of merit should be examined quantitatively. The equation of sensitivity is given by Ayyanar et al. (2018):
7 |
where is the shift in the position of the defect peak (), and is the RI difference between different blood plasma samples.
When the defect layer is immersed with water( , a defect peak appears at λp = 1782 nm with transmittance amplitude (Tp = 98.8%). The resonant peak is due to the light confinement through the blood plasma layer.
The ability of any sensor for measuring a very small change in the resonant wavelength also depends on the figure of merit (FOM). FOM is a very important parameter because it is directly proportional to the sensitivity (S) of the proposed sensor and inversely proportional to the full width at half maximum of the transmission of the resonant peak ()according to the following equation:
8 |
The decrease in FWHM leads to increasing the quality factor (Q-factor) of the sensor as the relation (Beheiry et al. 2010; White and Fan 2008):
9 |
Another parameter related to S and Q is called the detection limit, which indicates the smallest change in RI can be detected, as following equation (Beheiry et al. 2010; White and Fan 2008);
10 |
A very low DL indicates a good sensor for resolving small RI changes. the value of signal-to-noise (SNR) is calculated by Abadla and Elsayed (2020):
11 |
Finally, the resolution(RS) of the proposed sensor that represents the smallest resonant shift that can be detected accurately is given by Abadla and Elsayed (2020):
12 |
Effect of the number of layers (N)
By increasing the number of layers (N) from 2 to 5 layers, the resonance of defect peaks is slightly blue-shifted (decrease in wavelength), decreases and the PBG shrink, as shown in Fig. 4. Besides, the sensitivity and slightly increase but Q-factor significantly increases due to the decrease of , as well as the detection limit decreases. But at values of N higher than N = 6, FOM seems to be constant.
We have considered the number of layers (N = 6) is the optimum condition to maintain a high- FOM (high S and Q but low FWHM and DL), as cleared in Fig. 5. Blue-line is 4th-order polynomial fitted data between N and FWHM (R2 = 99%) that can be considered as:
13 |
Effect of the plasma layer concentration:
According to Eq. 1, the blood plasma RI changes from 1.3246 to 1.3634. When the concentration of blood plasma changed from 0 to 20 (g/L), the resonant peak is shifted to the right region (redshift) to maintain the optical path difference constant as the formula of standing wave states:
14 |
where are the optical path difference, an integer, the wavelength of the incident light, the effective RI of the design, and geometric path difference, respectively (Zeng et al. 2014). The transmission spectrum in Fig. 3 illustrates the defect peak shift from 1782 m to 1810 m with =28 m when the blood plasma concentration varied from 0 to 20 (g/L). The sensitivity at these conditions equals 721 m/RIU.
Effect of plasma layer thickness
When the thickness of the blood plasma layer thickness increases, the incident IR waves passes through a long geometric path inside the plasma layer, a strong interaction occurs between the plasma layer and the incident light, and the difference of the resonant defect peak position sift (Δλp) increases (Aly and Zaky 2019), as shown in Figs. 6a, b. The resonant defect peak position is red-shited as well due to the standing wave formula (Eq. 14).
Figure 7 shows the change of sensitivity with the increase of the blood plasma sample layer thickness from 1100 to 10,000 m. At defect layer thicknesses higher than 5000 nm, the value of sensitivity seems to be constant, and new resonant defect peaks enter the PBG. By increasing the defect layer thickness than 10,000 nm, sensitivity slightly increases, and much sample volume will be needed. For these reasons, this thickness is considered the optimum defect layer thickness. The highest sensitivity for normal incidence (1528 nm/RIU) corresponding to defect mode thickness of 10,000 nm.
Effect of the incident angle on sensitivity
Finally, to take a closer look at the dependence of resonant peak on the incident angle, the transmittance powers as a function of the angle of incidence are studied in the Table. 1 at the defect layer thickness of 10,000 nm. The effect of changing the incident angle on the resonant defect peak as well as on the position of the bandgap can be explained by using Bragg–Snell law (Schroden et al. 2002; Aly et al. 2012):
15 |
where q is the order of diffraction, λ is the wavelength, L is the interplanar spacing, neff is the effective index of refraction, and δ0 is the incident angle. According to Eq. 15, by increasing the incident angle (δ0), λ decreases (blue shifted).
Table 1.
δ0 (degree) | Δλp (nm) | S(nm/RIU) | Transmission power (%) |
---|---|---|---|
0 | 59.3 | 1528 | 98.8 |
30 | 68.5 | 1765 | 94.6 |
60 | 98.1 | 2528 | 82.6 |
85 | 126.6 | 3263 | 27 |
It can be observed that by increasing the angle from 0° to 85°, the sensitivity of the sensor significantly increases from 1528 nm/RIU to 3263 nm/RIU with a decrease in the transmission power (from 96 to 27%). In contrast, a further increase in the angle makes the sensitivity constant and leads to a significant decrease in transmission power until it reaches zero at 90°. Other papers reported this behavior (Armstrong and O'Dwyer 2015; Sharma et al. 2015).
From the above studies, N = 6, thickness dp = 10,000 nm, and incident angle δ0 = 85° are considered as optimum conditions (Fig. 8). In Table 2, Eq. 7 is used to calculate the sensitivity of different blood plasma concentrations. The Q-factor of the proposed sensor is very significant from 1.8 × 103 to 5.1 × 103, which makes it accurate. This is due to the resonant narrow peak (low FWHM). Also, This device recorded a high FOM from 1.9 × 103 to 8.1 × 103, a very low DL ( about RIU). This means that this plasma sensor has a high potential for RI sensing applications.
Table 2.
Cp (g/L) | N (RIU) | λp (nm) | S (nm/RIU) | FWHM (nm) | FOM (/RIU) | Q-Factor | SNR | SR |
---|---|---|---|---|---|---|---|---|
0 | 1.3246 | 2195 | – | 0.432 | – | 5.1 × 103 | – | – |
5 | 1.3343 | 2229 | 3505 | 0.604 | 8.1 × 103 | 3.7 × 103 | 56.3 | 0.147 |
10 | 1.3440 | 2261 | 3299 | 0.868 | 3.8 × 103 | 2.6 × 103 | 36.9 | 0.235 |
15 | 1.3537 | 2292 | 3196 | 1.292 | 2.5 × 103 | 1.8 × 103 | 24.0 | 0.389 |
20 | 1.3634 | 2322 | 3093 | 1.676 | 1.9 × 103 | 3.4 × 103 | 17.9 | 0.543 |
We perform linear fitting to the RI of the plasma blood sample (Fig. 9) and the defect peak position values from (Table 2). It is cleared that the position of the defect peak is red-shifted by increasing the RI. The slope of the fit represents the sensitivity of the sensor (3300 nm/RIU) and we get the following equation;
16 |
where is the defect peak position, is the plasma blood sample RI.
Since the major innovation of this manuscript is to reduce the cost of the proposed sensor, we use the microcavity structure. the 1D-PC is very simple in fabrication than others (2D and 3D) because of its simplest structure and very low cost compared to 2D and 3D structures (Pandey et al. 2017; Li 2014). As cleared in Table 3, most of the previous works that used the 1D-PC and some of the others that used 2D-PC recorded a very low sensitivity and Q-factor, but in this proposed structure, we recorded a super sensitivity (S = 3300 nm/RIU) at the optimum conditions (Abadla and Elsayed 2020; Mohamed et al. 2016; Banerjee 2009).
Conclusion
In the present study, we have investigated a plasma blood sample sensor by using a binary one-dimensional photonic crystal sensor and study the effect of the plasma cell concentration. The layer thickness, number of periods, and incidence angle are significant parameters on the sensitivity of the proposed sensor. The defect peak of the optimized structure is red-shifted from 2195 to 2322 nm with the RI of the defect layer changes from 1.3246 to 1.3634. These results mean that the high sensitivity of the device (S = 3300 nm/RIU) and a high Q-factor is possible (Q = 103). These advantages make this design is very important for biosensing applications, especially in case of a very limited change in blood plasma concentration that may help in COVID-19 therapy.
Acknowledgements
The authors extend their appreciation to the Deanship of Scientific Research at King Khalid University, Saudi Arabia for funding this work through the Research Group Program under grant no. R.G.P 2/127/42.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- Abadla MM, Elsayed HA. Detection and sensing of hemoglobin using one-dimensional binary photonic crystals comprising a defect layer. Appl. Opt. 2020;59(2):418–424. doi: 10.1364/AO.379041. [DOI] [PubMed] [Google Scholar]
- Abd El-Ghany SE, Noum WM, Matar Z, Zaky ZA, Aly AH. Optimized bio-photonic sensor using 1D-photonic crystals as a blood hemoglobin sensor. Physica Scripta. 2020;96(3):35501. [Google Scholar]
- Aghajamali A, Zare A, Wu C-J. Analysis of defect mode in a one-dimensional symmetric double-negative photonic crystal containing magnetized cold plasma defect. Appl. Opt. 2015;54(29):8602–8606. doi: 10.1364/AO.54.008602. [DOI] [PubMed] [Google Scholar]
- Ahmed K, et al. Design of D-shaped elliptical core photonic crystal fiber for blood plasma cell sensing application. Results Phys. 2019;12:2021–2025. [Google Scholar]
- Ahn JY, et al. Use of convalescent plasma therapy in two COVID-19 patients with acute respiratory distress syndrome in Korea. J. Kor. Med. Sci. 2020 doi: 10.3346/jkms.2020.35.e149. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Aly AH. Metallic and superconducting photonic crystal. J. Supercond. Novel Magn. 2008;21(7):421. [Google Scholar]
- Aly AH, Elsayed HA. Defect mode properties in a one-dimensional photonic crystal. Physica B. 2012;407(1):120–125. [Google Scholar]
- Aly AH, Hanafey HS. Polarization modes control on the transmittance characteristics of one-dimensional photonic crystal. J. Comput. Theor. Nanosci. 2011;8(10):1916–1919. [Google Scholar]
- Aly AH, Ismaeel M, Abdel-Rahman E. Comparative study of the one-dimensional dielectric and metallic photonic crystals. Opt. Photon. J. 2012;2(02):105. [Google Scholar]
- Aly AH, Mohamed D, Elsayed HA, Mehaney A. Fano resonance by means of the one-dimensional superconductor photonic crystals. J. Supercond. Novel Magn. 2018;31(12):3827–3833. [Google Scholar]
- Aly AH, Sayed FA, Elsayed HA. Defect mode tunability based on the electro-optical characteristics of the one-dimensional graphene photonic crystals. Appl. Opt. 2020;59(16):4796–4805. doi: 10.1364/AO.393689. [DOI] [PubMed] [Google Scholar]
- Aly AH, Zaky ZA. Ultra-sensitive photonic crystal cancer cells sensor with a high-quality factor. Cryogenics. 2019;104:102991. [Google Scholar]
- Aly AH, Elsayed HA, Malek C. Defect modes properties in one-dimensional photonic crystals employing a superconducting nanocomposite material. Opt. Appl. 2018;48(1):869. [Google Scholar]
- Aly AH, Ameen AA, Elsayed HA, Mohamed S, Singh MR. One-dimensional Metallo-superconductor photonic crystals as a smart window. J. Supercond. Novel Mag. 2019;58:1–16. [Google Scholar]
- Aly AH, Zaky ZA, Shalaby AS, Ahmed AM, Vigneswaran D. Theoretical study of hybrid multifunctional one-dimensional photonic crystal as a flexible blood sugar sensor. Physica Scripta. 2020;95(3):35510. [Google Scholar]
- Amiri IS, et al. Tri-core photonic crystal fiber-based refractive index dual sensor for salinity and temperature detection. Microw. Opt. Technol. Lett. 2019;61(3):847–852. [Google Scholar]
- Arafa S, Bouchemat M, Bouchemat T, Benmerkhi A, Hocini A. Infiltrated photonic crystal cavity as a highly sensitive platform for glucose concentration detection. Opt. Commun. 2017;384:93–100. [Google Scholar]
- Armstrong E, O'Dwyer C. Artificial opal photonic crystals and inverse opal structures–fundamentals and applications from optics to energy storage. J. Mater. Chem. C. 2015;3(24):6109–6143. [Google Scholar]
- Arunkumar R, Suaganya T, Robinson S. Design and analysis of 2D photonic crystal based biosensor to detect different blood components. Photon. Sens. 2019;9(1):69–77. [Google Scholar]
- Ayyanar N, Raja GT, Sharma M, Kumar DS. Photonic crystal fiber-based refractive index sensor for early detection of cancer. IEEE Sens. J. 2018;18(17):7093–7099. [Google Scholar]
- Banerjee A. Enhanced temperature sensing by using one-dimensional ternary photonic band gap structures. Progr. Electromag. Res. 2009;11:129–137. [Google Scholar]
- Cakmak O, et al. Microcantilever based disposable viscosity sensor for serum and blood plasma measurements. Methods. 2013;63(3):225–232. doi: 10.1016/j.ymeth.2013.07.009. [DOI] [PubMed] [Google Scholar]
- Chen L, Xiong J, Bao L, Shi Y. Convalescent plasma as a potential therapy for COVID-19. Lancet. Infect. Dis. 2020;20(4):398–400. doi: 10.1016/S1473-3099(20)30141-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- El Beheiry M, Liu V, Fan S, Levi O. Sensitivity enhancement in photonic crystal slab biosensors. Opt. Expr. 2010;18(22):22702–22714. doi: 10.1364/OE.18.022702. [DOI] [PubMed] [Google Scholar]
- El-Khozondar HJ, Mahalakshmi P, El-Khozondar RJ, Ramanujam N, Amiri I, Yupapin P. Design of one-dimensional refractive index sensor using ternary photonic crystal waveguide for plasma blood samples applications. Physica E. 2019;111:29–36. [Google Scholar]
- Elsayed HA, Mehaney A. A new method for glucose detection using the one-dimensional defective photonic crystals. Mater. Res. Expr. 2018;6(3):36201. [Google Scholar]
- Fan Z, et al. High sensitivity of refractive index sensor based on analyte-filled photonic crystal fiber with surface plasmon resonance. IEEE Photon. J. 2015;7(3):1–9. [Google Scholar]
- Faneca J, Perova TS, Tolmachev V, Baldycheva A. One-Dimensional multi-channel photonic crystal resonators based on silicon-on-insulator with high quality factor. Front. Phys. 2018;6:33. [Google Scholar]
- Fu Y, Liu M, Shum P, Chu L. An ultrahighly sensitive photonic crystal fiber based surface plasmon resonance sensor. Optik. 2012;212:164649. [Google Scholar]
- Guo Y, et al. Amphibious sensor of temperature and refractive index based on D-shaped photonic crystal fibre filled with liquid crystal. Liq. Cryst. 2020;47(6):882–894. [Google Scholar]
- Guo S, Hu C, Zhang HJJB. Analysis of the features of a multifunctional device based on the regulation of the magnetic field in one-dimensional photonic crystals containing only plasma with a novel quasi-periodic structure. JOSA B. 2020;37:1996. [Google Scholar]
- Hassan A-KS, Mohamed AS, Maghrabi MM, Rafat NH. Optimal design of one-dimensional photonic crystal filters using minimax optimization approach. Appl. Opt. 2015;54(6):1399–1409. doi: 10.1364/AO.54.001399. [DOI] [PubMed] [Google Scholar]
- Keith P, Day M, Perkins L, Moyer L, Hewitt K, Wells A. "A novel treatment approach to the novel coronavirus: an argument for the use of therapeutic plasma exchange for fulminant COVID-19. Crit. Care. 2020 doi: 10.1186/s13054-020-2836-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kenner T. The measurement of blood density and its meaning. Basic Res. Cardiol. 1989;84(2):111–124. doi: 10.1007/BF01907921. [DOI] [PubMed] [Google Scholar]
- Khateib F, Mehaney A, Amin RM, Aly AHJPS. Ultra-sensitive acoustic biosensor based on a 1D phononic crystal. Physica Scripta. 2020;95(7):75704. [Google Scholar]
- Kruschwitz JDT, Pawlewicz WT. Optical and durability properties of infrared transmitting thin films. Appl. Opt. 1997;36(10):2157–2159. doi: 10.1364/ao.36.002157. [DOI] [PubMed] [Google Scholar]
- Kupferschmidt K, Cohen J. Race to find COVID-19 treatments accelerates. Science. 2020;367:6485. doi: 10.1126/science.367.6485.1412. [DOI] [PubMed] [Google Scholar]
- Li J-S. Fast-tunable terahertz wave filter based on Kerr medium. Opt. Laser Technol. 2014;56:263–268. [Google Scholar]
- Li Z-Y, Lin L-L. Photonic band structures solved by a plane-wave-based transfer-matrix method. Phys. Rev. E. 2003;67(4):046607. doi: 10.1103/PhysRevE.67.046607. [DOI] [PubMed] [Google Scholar]
- Mohamed MS, Hameed MFO, Areed NF, El-Okr M, Obayya S. Analysis of highly sensitive photonic crystal biosensor for glucose monitoring. Appl. Comput. Electromag. Soc. J. 2016;31(7):584. [Google Scholar]
- Mulder HK, Ymeti A, Subramaniam V, Kanger JS. Size-selective detection in integrated optical interferometric biosensors. Opt. Expr. 2012;20(19):20934–20950. doi: 10.1364/OE.20.020934. [DOI] [PubMed] [Google Scholar]
- Natesan A, Govindasamy KP, Gopal TR, Dhasarathan V, Aly AH. Tricore photonic crystal fiber based refractive index sensor for glucose detection. IET Optoelectron. 2018;13:118. [Google Scholar]
- Nayak C, Aghajamali A, Alamfard T, Saha AJPBCM. Tunable photonic band gaps in an extrinsic Octonacci magnetized cold plasma quasicrystal. Physica B: Condens. Matter. 2017;525:41–45. [Google Scholar]
- Nayak C, Saha A, Aghajamali AJIJP. Periodic multilayer magnetized cold plasma containing a doped semiconductor. Ind. J. Phys. 2018;92(7):911–917. [Google Scholar]
- Nayak C, Costa CH, Aghajamali AJITOPS. Robust photonic bandgaps in quasiperiodic and random extrinsic magnetized plasma. IEEE Trans. Plasma Sci. 2019;47(4):1726–1733. [Google Scholar]
- Nunley TN, et al. (2016) "Optical constants of germanium and thermally grown germanium dioxide from 0.5 to 6.6 eV via a multisample ellipsometry investigation,". J. Vacuum Sci. Technol. B, Nanotechnol. Microelectron.: Mater., Process., Measure., and Phenom. 2016;34(6):061205. [Google Scholar]
- Pandey A, Tripathi A, Srivastava S, Jit S. Analysis of ternary layer photonic band gap tunable filters for wavelength division multiplexing applications. J. Nanoelectron. Optoelectron. 2017;12(4):331–336. [Google Scholar]
- Peng J, Lyu D, Qu Y, Wang W, Sun T, Yang M. Thin films based one-dimensional photonic crystal for refractive index sensing. Optik. 2018;158:1512–1518. [Google Scholar]
- Schroden RC, Al-Daous M, Blanford CF, Stein A. Optical properties of inverse opal photonic crystals. Chem. Mater. 2002;14(8):3305–3315. [Google Scholar]
- Shaban M, Ahmed AM, Abdel-Rahman E, Hamdy H. Tunability and sensing properties of plasmonic/1D photonic crystal. Sci. Rep. 2017;7:41983. doi: 10.1038/srep41983. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shaban SM, Mehaney A, Aly AH. Determination of 1-propanol, ethanol, and methanol concentrations in water based on a one-dimensional phoxonic crystal sensor. Appl. Opt. 2020;59(13):3878–3885. doi: 10.1364/AO.388763. [DOI] [PubMed] [Google Scholar]
- Sharma S, Kumar R, Singh KS, Kumar A, Kumar V. Omnidirectional reflector using linearly graded refractive index profile of 1D binary and ternary photonic crystal. Optik. 2015;126(11–12):1146–1149. [Google Scholar]
- Sharma S, Kumar R, Singh KS, Kumar V, and Jain D, “Design of a Narrow-Band Photonic Crystal Based Omni-Directional Mirror for Optical Fiber Communication.” (2012)
- Shen C, et al. Treatment of 5 critically ill patients with COVID-19 with convalescent plasma. JAMA. 2020;323(16):1582–1589. doi: 10.1001/jama.2020.4783. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shiri, R., Khalkhali, T. F., Shahrokhabadi, H., & Bananej, A.: Designing one-dimensional magnetized plasma photonic crystals for compensating second-and third-order dispersion effects in ultra-short pulse lasers. Optik 188, 137–146 (2019)
- Srivastava SK, Aghajamali AJJOS, Magnetism N. Analysis of reflectance properties in 1D photonic crystal containing metamaterial and high-temperature superconductor. J. Supercond. Noval Mag. 2017;30(2):343–351. [Google Scholar]
- White IM, Fan X. On the performance quantification of resonant refractive index sensors. Opt. Expr. 2008;16(2):1020–1028. doi: 10.1364/oe.16.001020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Winn JN, Fink Y, Fan S, Joannopoulos J. Omnidirectional reflection from a one-dimensional photonic crystal. Opt. Lett. 1998;23(20):1573–1575. doi: 10.1364/ol.23.001573. [DOI] [PubMed] [Google Scholar]
- Wu C-J, Chen M-S, Yang T-J. Photonic band structure for a superconductor-dielectric superlattice. Physica C. 2005;432(3–4):133–139. [Google Scholar]
- Yablonovitch E. Inhibited spontaneous emission in solid-state physics and electronics. Phys. Rev. Lett. 1987;58(20):2059. doi: 10.1103/PhysRevLett.58.2059. [DOI] [PubMed] [Google Scholar]
- Zaky ZA, Aly AH. Theoretical study of a tunable low-temperature photonic crystal sensor using dielectric-superconductor nanocomposite layers. J. Supercond. Novel Magn. 2020;33:2983–2990. [Google Scholar]
- Zaky ZA, Aly AH. Modeling of a biosensor using Tamm resonance excited by graphene. Appl. Opt. 2021;60(5):1411–1419. doi: 10.1364/AO.412896. [DOI] [PubMed] [Google Scholar]
- Zaky ZA, Ahmed AM, Shalaby AS, Aly AH. Refractive index gas sensor based on the Tamm state in a one-dimensional photonic crystal: theoretical optimisation. Sci. Rep. 2020;10(1):9736. doi: 10.1038/s41598-020-66427-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zaky, Z. A., Aly, A. H.: “Highly sensitive salinity and temperature sensor using Tamm resonance”. Plasmonics (2021). 10.1007/s11468-021-01487-6
- Zaky, Z. A., Ahmed, A. M., Aly, A. H.: Remote temperature sensor based on Tamm resonance. Silicon (2021). 10.1007/s12633-021-01064-w.
- Zeng C, Luo C, Hao L, Xie Y. The research on magnetic tunable characteristics of photonic crystal defect localized modes with a defect layer of nanoparticle. Chin. Opt. Lett. 2014;12:11602. [Google Scholar]
- Zhou Z, Wan B, Xu Y, Ma Y, Zhang H, Zhang DJJB. Tunable zero-phase delay of one-dimensional photonic crystals containing InSb material. J. Opt. Soc. Am. B. 2021;38(1):114–122. [Google Scholar]