Qualification of ELISA and neutralization methodologies to measure SARS-CoV-2 humoral immunity using human clinical samples

Sasha E Larsen a,1, Bryan J Berube a,b,1, Tiffany Pecor a, Evan Cross a, Bryan P Brown a, Brittany D Williams a,c, Emma Johnson a, Pingping Qu d, Lauren Carter e, Samuel Wrenn e, Elizabeth Kepl e, Claire Sydeman e, Neil P King e, Susan L Baldwin a, Rhea N Coler a,c,f,
PMCID: PMC8481082  PMID: 34599915

Abstract

In response to the SARS-CoV-2 pandemic many vaccines have been developed and evaluated in human clinical trials. The humoral immune response magnitude, composition and efficacy of neutralizing SARS-CoV-2 are essential endpoints for these trials. Robust assays that are reproducibly precise, linear, and specific for SARS-CoV-2 antigens would be beneficial for the vaccine pipeline. In this work we describe the methodologies and clinical qualification of three SARS-CoV-2 endpoint assays. We developed and qualified Endpoint titer ELISAs for total IgG, IgG1, IgG3, IgG4, IgM and IgA to evaluate the magnitude of specific responses to the trimeric spike (S) antigen and total IgG specific to the spike receptor binding domain (RBD) of SARS-CoV-2. We also qualified a pseudovirus neutralization assay which evaluates functional antibody titers capable of inhibiting the entry and replication of a lentivirus containing the Spike antigen of SARS-CoV-2. To complete the suite of assays we qualified a plaque reduction neutralization test (PRNT) methodology using the 2019-nCoV/USA-WA1/2020 isolate of SARS-CoV-2 to assess neutralizing titers of antibodies in plasma from normal healthy donors and convalescent COVID-19 individuals.

Keywords: Qualification, SARS-CoV-2, COVID, ELISA, Neutralization, Pseudovirus

Graphical abstract

Image 1

Highlights

  • Precision, Linearity, and Specificity are essential for Clinical Assay Qualification.

  • Vaccine or Infection-induced humoral response magnitude can be evaluated by high-throughput ELISAs.

  • Neutralization of SARS-CoV-2 is the gold-standard for in vitro vaccine efficacy evaluations.

  • ELISA, pseudovirus neutralization and PRNT assays are Clinically Qualified for SARS-CoV-2 vaccine trials.

  • Positive WHO control sample of 250 ABU equals 4.7 EPT for total IgG against SARS-CoV-2 trimeric spike antigen in ELISAs.

1. Introduction

In December 2019, a series of severe viral pneumonia cases were detected in Wuhan, Hubei Province, China, and later discovered to be induced by a novel coronavirus (Zhou et al., 2020). Since then, the severe acute respiratory syndrome corona virus 2 (SARS-CoV-2) has spread into a pandemic (Cucinotta and Vanelli, 2020). As SARS-CoV-2 continues to devastate the world and its economies, the pressure for effective and equitable vaccine distribution remains a global priority (Wang et al., 2020; Schwartz, 2020). Many vaccine and therapeutic candidates are designed to induce a robust humoral immune response against the SARS-CoV-2 spike glycoprotein, which binds to human angiotensin-converting enzyme 2 (ACE2) and mediates viral entry into pulmonary alveolar epithelial cells (Letko et al., 2020; Ou et al., 2020; Shang et al., 2020; Yan et al., 2020). These vaccines attempt to neutralize the receptor-binding domain (RBD) (Roy et al., 2020) of spike and reduce viral entry into host cells upon exposure (Poland et al., 2020), a strategy similar to that employed for SARS-CoV (Du et al., 2009). When measuring the efficacy of therapeutics and vaccines, high-throughput and clinically qualified reproducible immunological assays provide crucial standardized information regarding the quantity and quality of antibody and cell-mediated responses (Andreasson et al., 2015; Burd, 2010). Specifically, the quantification of antibodies in a spike antigen-based endpoint titer (EPT) enzyme-linked immunosorbent assay (ELISA) in tandem with efficacy measurements via plaque reduction neutralization tests (PRNT) or pseudovirus neutralization can be leveraged to evaluate responses to SARS-CoV-2 infections and interventions (Liu et al., 2020; Okba et al., 2020; Henss et al., 2020; Perera et al., 2020; Sholukh et al., 2020; Earle et al., 2021).

Indeed, many ELISAs have been developed against SARS-CoV-2 (Amanat et al., 2020; Mohit et al., 2021), including those that characterize seroconversion in response to infection and quantify the magnitude of the humoral response (Roy et al., 2020). However, few are clinically qualified using trimeric SARS-CoV-2 spike antigen as our studies report. In order to meet this need, we have developed and qualified a comprehensive suite of high-throughput 384-well format ELISAs for detecting abundance of total human immunoglobulin G (IgG), IgG subclasses (IgG1, IgG3, IgG4), IgM, and IgA recognizing SARS-CoV-2 spike antigen and total IgG specific for RBD. Like other research groups (Mazzini et al., 2021; Chen et al., 2020) we were unable to identify a sufficient number of samples with robust IgG2 responses to qualify the assay for that subclass. Importantly, we have directly compared our calculated endpoint titer values with WHO international reference standards to make future comparisons between labs more accessible (WHO/BS/2020.2403). In tandem we have qualified clinical SARS-CoV-2 PRNT and pseudovirus neutralization assays evaluating the ability of antibodies in a given sample to reduce viral cell entry as a quality gold-standard surrogate for vaccine-mediated immunogenicity and efficacy (Sholukh et al., 2020; Gach et al., 2014). Each assay was qualified across precision (Chesher, 2008), linearity (Jhang et al., 2004) and specificity (Burd, 2010) endpoints to establish they 1) were accurate and repeatable across users and days, 2) showed a linear response with respect to sample concentrations, and 3) were explicitly detecting responses to SARS-CoV-2. We believe this combination of clinically qualified serology assays will be valuable in comparing responses from various vaccines (Polack et al., 2020; Keech et al., 2020; Sadoff et al., 2021; Erasmus et al., 2020; Folegatti et al., 2020), cutting-edge boost regimens (Logunov et al., 2021; Tanne, 2021), as well as evaluating preclinical vaccine candidates (Jain et al., 2020; Rice et al., 2020) poised to enter clinical trials. Furthermore, these assays will serve to aid the scientific community by allowing the assessment long term vaccine durability (Widge et al., 2020; Dan et al., 2021) and, with adaptation, the intermediate vaccine efficacy against circulating and emerging viral variants (Williams and Burgers, 2021).

2. Methods and materials

2.1. Specimens and controls

Positive controls were obtained through the Seattle Children's Research Institute's (SCRI) Center for Global Infectious Disease Research Biorepository. These samples were drawn from Seattle Children's workforce and adult community members diagnosed with COVID-19 by detection of SARS-CoV-2 nucleic acids in nasopharyngeal specimens, who subsequently enrolled into the Seattle Children's SARS2 Recovered Cohort with informed consent and Institutional Review Board approval through SCRI. Commercially sourced (Bloodworks Northwest) positive controls were also used from confirmed and recovered COVID-19 patients, termed convalescent plasma. Commercial convalescent plasma was also pooled (CCPP) as a standardized positive control across assays. A sample of pooled plasma from 11 SARS-CoV-2 positive patients from the National Institute for Biological Standards and Controls (NIBSC 20/136), given an arbitrary WHO international standard value of 250 international units (IU) of binding antibody activity per ampule, was also obtained for use in selected assays. Negative plasma originated from historical in-house biorepository of samples isolated from individuals before November 2019, predating the SARS-CoV-2 pandemic and thereby reducing the opportunity for existing cross-reactive antibody responses. Commercially available Normal Human Plasma (Boston Biomedical, Cambridge, MA) was pooled from several donors (NHPP) and used as a standard negative control.

2.2. ELISA methodology

2.2.1. Coating

Sterile High Binding 384 well plates (Corning, Corning, NY) were filled with 50 μL per well of ELISA coating buffer containing full-length trimeric SARS-CoV-2 spike protein 11.8 μM or RBD, expressed and purified as described previously (Walls et al., 2020), with a final concentration of 1 μg/mL. Coating buffer was made in house using one standard packet of ELISA coating buffer powder (ebioscience, San Diego, CA) mixed with 1 L of distilled water and filtered sterile with a 22 μm filter to generate 1 L of 0.01 M PBS, pH 7.4. Plates were incubated for at least 2 h at room temperature, or up to 3 days at 4 °C.

2.2.2. Blocking

Wash Buffer A was made by diluting 1 L of 20× Wash Buffer A solution (Teknova, Hollister, CA) in 19 L of H2O from a Barnstead Nanopure Water system (Thermo Scientific, Waltham, MA). After incubation with SARS-CoV-2 Spike protein, plates were washed three times with 100 μL per well of 1× Wash Buffer A. Plates were washed using a BioTek EL406 plate washer (BioTek, Winooski, VT). Blocking buffer was formulated on site by adding 10.0 g bovine serum albumin (Sigma, St. Louis, MO) in 1000 mL of phosphate buffered saline (PBS, Lonza, Basel, Switzerland) with 0.05% Tween. Each well received 100 μL of blocking buffer for all plates. Plates were incubated for at least 2 h, or wrapped in plastic wrap and incubated at 4 °C overnight.

2.2.3. Sample addition

After blocking, plates were washed three times with 1× Wash Buffer A and subsequently 50 μL of diluent (mixture of 1.0 g bovine serum albumin, 500 mL 1× Wash Buffer A and 500 mL PBS) was added to every well of the washed plates. Next, plasma samples were diluted 1:20 in diluent (5 μL of plasma to 95 μL of diluent for total IgG, other Ig sample dilutions listed in Table 1) in a 96-well master block before their addition to the 384-well ELISA plate. Using a multichannel pipette, 12.5 μL of the master block sample dilution was manually pipetted into the first column of each sample and subsequently diluted 1:5 from left to right across the plate, discarding 12.5 μL from the final dilution column. Plate layouts varied depending on which aspect of the qualification was being examined. Plates were incubated overnight at 4 °C.

Table 1.

Parameters for each immunoglobulin qualified in ELISA methodology.

Antibody of interest Dilution of secondary antibody Development time (minutes) Initial plasma dilution per antibody
Total IgG 1:4000 2.5–3.5 1:20
IgG1 1:1000 10 1:2
IgG3 1:1000 10 1:2
IgG4 1:2000 10 1:2
IgM 1:2000 3.5–4.5 1:20
IgA 1:8000 3.5–4.5 1:20

RBD qualification was only completed for total IgG.

2.2.4. Development

After incubation, plates were removed from the refrigerator and left to warm to room temperature. Once warmed, plates were washed five times with 1× Wash Buffer A. HRP conjugated rec-Protein G antibodies (Invitrogen, 101,223) were diluted (diluent mixture described above) by a factor of 4000, HRP conjugated IgM antibodies (ThermoFisher, A18835) were diluted by a factor of 2000, HRP conjugated IgA antibodies (ThermoFisher, PAI-74395) were diluted by a factor of 8000, HRP conjugated IgG1 antibodies (Invitrogen, MH1715) were diluted by a factor of 1000, HRP conjugated IgG3 antibodies (Invitrogen, 05–3620) were diluted by a factor of 1000, and HRP conjugated IgG4 antibodies (ThermoFisher, MH1742) were diluted by a factor of 2000 (Table 1). Plates were incubated for one hour in the dark at room temperature, then washed five times with 1× Wash Buffer A followed by one wash with PBS. Plates then received 100 μL of Tetramethylbenzidine (TMB) (SeraCare, Milford, MA) per well. After 4–10 min; (Table 1) depending on secondary antibody, 25 μL of 1 N H2SO4 (Sigma, St. Louis, MO) was added to each well to halt the TMB reaction. Plates were read at a wavelength of 450 nm with a reference filter set at 570 nm using a SpectraMax i3x plate reader (Molecular Devices, San Jose, CA) and SoftMax Pro 6.4.2 analysis software.

2.2.5. EPT calculations

Each diluted sample O.D. value was calculated using 450 nm minus 570 nm values and the average O.D. value for all NHPP dilutions was used to set a minimum cutoff value for each plate. These plate cutoff values were then used to calculate each sample EPT using a 4-parameter logistic model in XL-fit software (model 208) as a Microsoft Excel add-in, as used by our team in previous clinical trial evaluations (Coler et al., 2018; Day et al., 2020). The EPT value calculated for each duplicate plate is then averaged for a final EPT run value.

2.2.6. Variations for competition ELISA

The competition ELISA followed procedures outlined as above except for the following changes. Before adding plasma samples to coated and blocked plates, they were pre-incubated with SARS-CoV-2 Spike protein or RBD at decreasing concentrations. In a 96-well setup plate, soluble SARS-CoV-2 spike or RBD antigens were serially 5-fold diluted. Plasma samples were diluted 1:250 in diluent, and 100 μL was added to the setup plate containing spike protein or RBD. The setup/co-incubation plate was incubated at room temperature for 4 h with gentle rotation. The plasma/soluble spike or soluble RBD mixtures were added to spike or RBD-coated and blocked plates and the remainder of the ELISA methodology was followed as described above.

2.3. Neutralization methodologies

2.3.1. Cell lines and growth conditions

Vero E6 and HEK293T cells were obtained from ATCC and HEK293T cells stably transfected with human Angiotensin-Converting Enzyme 2 (HEK293T-hACE2) were obtained from BEI Resources (NR-52511). All cells were maintained in Dulbecco's Modified Eagle Media (DMEM) containing 10% fetal bovine serum (FBS), 1% penicillin/streptomycin, and 2 mM l-glutamine (complete DMEM; cDMEM) in a 37 °C + 5% CO2 incubator. Cells were maintained 25 to 90% confluent and used for the assay up to but not past 14 passages.

2.3.2. Pseudovirus production

HIV-1-based SARS-CoV-2 Spike pseudoviruses were prepared as previously described (Crawford et al., 2020). In brief, HEK293T cells were co-transfected with plasmids in the SARS-CoV-2 Spike-pseudotyped lentiviral particle kit (BEI Resources; NR-52948) using BioT transfection reagent (Bioland Scientific). Plasmids encode for that SARS-CoV-2 Spike protein (Wuhan-Hu-1; Genbank: NC 045512), a lentiviral backbone containing Luc2, and HIV-1 Gag, Pol, Tat1b, and Rev1b. After 48–56 h, the supernatant was collected and passed through a 0.45 μm filter and aliquots were made and stored at −80 °C until use.

2.3.3. Pseudovirus titration and neutralization assay

One day prior to infection, black-walled 96-well plates were coated with 0.01% poly-l-lysine for 5 min. Wells were washed twice with sterile water, and HEK293T-hACE2 cells were plated at 2.5 × 104 cells/well in 50 μL cDMEM. Plates were incubated at 37 °C + 5% CO2 overnight. To titer the pseudovirus, aliquots were thawed and 2-fold dilutions of pseudovirus in cDMEM were prepared in a 96-well plate. Virus dilutions (100 μL) were added to the HEK293T-hACE2 cells followed immediately by addition of polybrene at a concentration of 5 μg/mL. Cells were incubated at 37 °C + 5% CO2. After 48 h, 100 μL of supernatants were removed from the wells, and cells were lysed with 30 μL Bright Glo Luciferase Reagent. Relative luminescence units (RLU) were measured immediately on a Spectramax i3x plate reader, as a surrogate for pseudovirus entry and replication.

For pseudovirus neutralization assays, plasma samples were diluted in cDMEM in a 10-point 3-fold dilution series in a 96-well setup plate. SARS-CoV-2 Spike pseudovirus was diluted to 1 × 105 RLU/mL and added to wells of the setup plate in a 1:1 ratio. The setup plate was incubated at 37 °C + 5% CO2 for 1 h prior to addition to HEK293T-hACEs cell as above. Percent inhibition was calculated by dividing RLU values in each well to the average RLU value of virus only wells. Pseudovirus inhibition curves and the concentration of plasma required to inhibit pseudovirus entry by 50% (IC50) was determined by plotting the data and fitting a curve using the neutcurve Python package (https://jbloomlab.github.io/neutcurve/).

2.3.4. SARS-CoV-2 virus production

SARS-CoV-2 isolate 2019-nCoV/USA-WA1/2020 was obtained from BEI Resources and housed under standard BSL-3 laboratory conditions. SARS-CoV-2 virus was propagated and titered by plaque assay in Vero E6 cells. The original stock virus was added to cells at an MOI of 0.1 and incubated at 37 °C + 5% CO2 for 72 h. Supernatants were harvested, spun at 1,200 rpm for 20 min, and the supernatant was aliquoted and frozen at −80 °C to create a Passage 1 (P1) virus stock. P1 virus was titered and propagated a second time as above to create Passage 2 (P2) virus. Aliquots of the P2 virus were diluted to 4.5 × 103 plaque-forming units (PFU)/mL to create a PRNT virus stock.

2.3.5. PRNT assay

One day prior to infection, Vero E6 cells were plated at 4 × 105 cells/well in 2 mL cDMEM in a 6-well plate. The following day, heat-inactivated serum/plasma samples (incubated at 56 °C for 1 h) were diluted in dilution media (DMEM +1% FBS) in a 6-point 2-fold dilution series in a 96-well setup plate to a total volume of 115 μL per well. SARS-CoV-2 PRNT stocks were thawed and diluted 5-fold in dilution media and 115 μL added to the setup plate (9.0 × 102 PFU/mL). Virus and plasma were co-incubated for 1 h at 37 °C + 5% CO2. Media was carefully removed from the 6-well plates and 200 μL of the virus/plasma mixture was added to the cells. The plates were incubated at 37 °C + 5% CO2 with rocking every 15 min. After 1 h, 2.0 mL overlay media (dilution media +0.2% agarose, kept at a minimum of 45 °C) was added to the wells and plates were subsequently incubated at 37 °C + 5% CO2 for 72 h. After 72 h, cells were fixed by the addition of 2.0 mL of 10% formaldehyde and incubated for 30 min. Fixative and overlay were removed, wells were stained for 20 min with 1.0 mL of 0.3% Crystal Violet (BD) and washed once with 1.0 mL PBS per well. Plates were inverted and plaques indicating cell lysis from viral infection were counted for each well and recorded. The lowest dilution of plasma to reduce PFU by 80% compared to virus only wells was deemed the PRNT80. The calculation for the threshold is as follows

PRNT80threshold=Avg.PFUVirus OnlyX0.2

2.4. Qualification

Precision is the reproducibility of an assay across different variables, including between users and days of execution. Given sample endpoint and neutralization titers are unknown, they are not compared to a known standard EPT value but rather evaluated for a consistent performance of the assay to return a similar value for specific control samples across conditions. Intra-assay variability measures the variation of the assay under the same conditions, while inter-assay variability measures variation between different operators and days. Intra-assay variability was measured across three runs performed by a single technician on the same day with like samples. Inter-day variability was measured across three runs performed by a single technician across three separate days with like samples. Finally, for inter-operator variability each of three operators generated their own independent plates and reagents and performed the assay on the same day with like samples. Plate layouts and samples used remained consistent across each measurement of variability. This evaluates the reproducibility and repeatability of the assay and demonstrates that the procedure generates consistent results. When evaluating precision, the inverse or imprecision is actually measured and reported as described below (Andreasson et al., 2015).

2.4.1. ELISA precision endpoint

Each sample must return a coefficient of variation (CV) no greater than 20% for any variable (operator, day, run) evaluated. Coefficient of variation is calculated as

%CV=100×Standard DeviationEPT1EPT2EPT3AverageEPT1EPT2EPT3

where EPT 1–3 are the calculated endpoint values for an individual sample across the three runs (operator, day) being evaluated. For example, CCPP sample EPT values across each of the three operator runs must not exceed 20% CV. This value is partially based on European Medicines Agency (EMA) guidelines that state precision and repeatability must be below this 20% CV threshold, as well as the U.S. Food and Drug Administration's Bioanalytical Method Validation guidance for Industry (2018) (Meesters and Voswinkel, 2018) where precision within and between runs should be below 20% CV.

2.4.2. Pseudovirus neutralization precision endpoint

The average from duplicate samples on any given plate must be within 2-fold (+/−) of the geometric mean within the respective category (operator, run, reproducibility). For example, the average IC50 value from duplicate samples for CCPP on each plate within an intra-operator variability test must be within 2-fold of the geometric mean IC50 value of all CCPP samples tested within the intra-operator variability test.

2.4.3. PRNT precision endpoint

Each sample on any given plate must be within 2-fold (+/−) of the geometric mean within the respective category. For example, each PRNT80 value for CCPP within an intra-operator variability test must be within 2-fold of the geometric mean PRNT80 value of all CCPP samples tested within the intra-operatory variability test.

Linearity helps evaluate how well an assay is able to discern variable concentrations of an analyte and if the detected quantity well represents a value across a dilution series. These samples do not have a reference ‘true value’ so instead we leveraged known positive samples across a standardized dilution curve to determine if the detected response was proportional to the dilution. For ELISA assays, curves were created using O.D. values for positive samples from precision plates and analyzed by linear regression (Jhang et al., 2004) with calculated R2 values for individual samples as well as an aggregate analysis of an adjusted multiple R2 following established calculation methods (Wherry, 1931). Only samples with three or more data points were included. Similarly, for PRNT and pseudovirus neutralization assays precision plate dilutions were used from positive samples analyzed by linear regression (Jhang et al., 2004) with calculated R2 values.

2.4.4. Linearity endpoint

A graphical analysis of the line of best fit (by least squares criterion) coupled with an R2 value of 0.90 or greater by linear regression analysis, demonstrates the acceptable linearity of the assay.

Specificity is determined in two distinct ways to ensure SARS-CoV-2 antigen-specific responses are being captured, including: a robust screening of presumed negatives, and competition assays with soluble antigen pre-competing for antibody binding. This helps promote that responses are not due to cross-reactive humoral responses that are present in unexposed blood donors. A set of 92 (used for each SARS-CoV-2 spike antibody ELISA) or 82 (used for total IgG RBD) negative samples from a pre-COVID-19 biorepository and four positive SARS-CoV-2 convalescent samples (2 positives for IgG4) were used in determining specificity of our ELISA methodology. For total IgG, IgA and IgM ELISAs, samples were diluted 5-fold starting at 1:100 for a 4-point dilution. For IgG1, IgG3, and IgG4 ELISAs, samples were diluted 5-fold starting at 1:10 for a 4-point dilution. A single plate was run by one technician. A competitive ELISA was also performed to further demonstrate specificity of the immune response to SARS-CoV-2. In this specificity assay, we are expecting serum or plasma antibodies to bind and be sequestered from the coated plate by higher concentrations of soluble spike or RBD in solution. A positive sample was run in duplicate on the same plate by a single operator for this analysis.

We evaluated the specificity of serum/plasma samples to neutralize SARS-CoV-2 via a robust negative sample size. We used 20 negative plasma samples from a historical (pre-COVID) biorepository and 4 positive control samples from patients who became infected and recovered from COVID-19 as a characterization of the pseudovirus neutralization and PRNT assay specificity. For pseudovirus neutralization, each sample was assessed at a 1:20 dilution and was run in duplicate by a single operator on the same day. In addition, we tested plasma from a convalescent patient, as well as a pool of plasma from a set of convalescent patients, to evaluate specificity against a non-specific pseudovirus harboring the VSV G cell entry protein instead of the SARS-CoV-2 spike protein. For PRNT specificity, 20 negative samples from a pre-COVID-19 biorepository and four positive samples were used. Each sample was assessed at a 1:40 dilution for and run in duplicate by a single operator.

2.4.5. ELISA specificity endpoints

For ELISAs ≥90% of negative sample O.D.s must be lower than the collective average of all negative samples +3 SD (plate limit of detection [LOD]), while all 4 positives must exceed this value. For the competition ELISA the results must demonstrate that as the amount of exogenous SARS-CoV-2 Spike antigen or RBD decreases, there will be an increase in positive signal measured by O.D. 450 nm–570 nm. Graphical analysis of O.D. values versus dilution values and a linear relationship with an R2 value >0.90 demonstrate an acceptable competition specificity qualification of the ELISA. Samples are recorded as pass-fail for the negative-screening specificity criteria.

2.4.6. Neutralization specificity endpoints

For pseudovirus neutralization specificity, all 20 negative samples must not achieve an IC50 value at the 1:20 dilution of plasma, whereas all 4 positive samples must meet this threshold. Samples are recorded as pass-fail for these criteria. For the VSV G pseudovirus specificity test, all COVID-positive samples must not achieve an IC50 value at the 1:20 dilution of plasma. For PRNT, the observed PFU from all 20 negative samples must not achieve a PRNT80 at the 1:40 dilution of plasma, whereas all 4 positive samples must meet this threshold.

3. Results and discussion

3.1. ELISA

3.1.1. Precision

Assay repeatability, or the degree to which an assay will produce the same result when identical test samples are evaluated under the same operating conditions, was assessed in terms of run to run variation. Samples from 14 subjects and 2 pooled controls (CCPP and NHPP) were run by a single operator in triplicate across three separate days (interday), by a single operator in triplicate on the same day (intraday), and by three individual operators on the same day (interoperator). Representative data for Total IgG SARS-CoV-2 spike antigen-specific responses for each precision test is shown in Fig. 1 including two individual convalescent samples (M and N) and pooled controls (CCPP, NHPP). Fig. 2 shows representative samples assess for SARS-CoV-2 RBD responses across the different variables. The CV for each sample (16 total) across the precision variables evaluated are reported in Table 2, Table 3, Table 4, and cumulative EPT graphs containing all data points are in Supplemental Fig. 1. Importantly, across the robust set of samples tested, all were below the 20% CV endpoint established for this assay, demonstrating the assay is reproducible and precise.

Fig. 1.

Fig. 1

SARS-CoV-2 spike total IgG endpoint titers evaluated for two individual samples (M, N) as well as pooled positive (CCPP) and negative (NHPP) control samples. Dashed line respresents lower limit of detection.

Fig. 2.

Fig. 2

SARS-CoV-2 RBD total IgG endpoint titers evaluated for three individual samples (L, M, N) as well as pooled positive (CCPP) and negative (NHPP) control samples. Dashed line respresents lower limit of detection.

Table 2.

ELISA intraday precision analysis using coefficient of variation (%).

Antigen
RBD
Spike
Sample ID Total IgG Total IgG IgM IgA IgG1 IgG3 IgG4
A 7.21 2.40 5.04 1.95 8.86 7.47 ND
B 2.93 2.36 6.30 4.47 8.51 ND ND
C 1.86 1.71 1.48 1.26 7.24 18.46 ND
D 1.84 2.28 2.02 3.38 6.80 16.75 ND
E 3.59 1.21 2.28 0.54 5.46 12.21 ND
F 0.86 0.82 5.19 1.67 4.67 13.3 2.68
G 2.14 1.41 1.68 2.27 7.97 9.92 ND
H 2.53 0.62 1.22 2.11 4.14 ND ND
I 5.92 2.07 1.12 0.77 4.44 9.10 ND
J 2.44 1.04 1.00 1.30 4.92 13.24 ND
K 4.12 2.12 1.10 1.08 3.51 2.61 ND
L 0.004 1.65 2.20 1.57 6.86 9.72 ND
M 0.85 0.73 3.28 1.51 1.88 1.91 7.39
N 0.69 4.13 0.63 1.37 4.82 4.81 ND
CCPP 0.69 4.83 1.51 3.78 5.80 6.38 8.09
NHPP 0.28 0.05 7.09 0.12 3.07 0.00 10.14
Average 2.37 1.84 2.7 1.82 5.56 8.39 5.66
Standard Error 0.51 0.31 0.51 0.29 0.50 1.48 1.58

ND: not done, sample values below the limit of detection or not used for this assessment.

Table 3.

ELISA interday precision analysis using coefficient of variation (%).

Antigen
RBD
Spike
Sample ID Total IgG Total IgG IgM IgA IgG1 IgG3 IgG4
A 5.94 2.24 8.71 3.24 10.69 12.94 ND
B 6.70 2.64 3.24 2.28 13.26 4.56 ND
C 7.50 3.71 12.92 3.80 3.29 9.90 ND
D 3.39 0.82 7.73 2.23 6.89 17.39 ND
E 5.42 0.99 10.36 1.95 4.06 8.76 ND
F 9.50 3.03 13.40 1.56 5.11 17.68 2.12
G 3.45 2.82 9.84 0.10 3.73 2.43 ND
H 3.67 1.79 3.70 0.97 ND ND ND
I 1.63 0.91 7.98 2.19 10.65 13.21 ND
J 5.28 0.29 10.29 2.28 6.21 17.39 ND
K 6.14 4.31 9.47 3.06 2.80 4.88 ND
L 4.72 1.41 5.06 1.53 8.76 6.69 ND
M 4.73 1.96 12.93 4.30 7.12 10.73 5.10
N 8.40 0.50 6.28 2.43 8.28 13.12 ND
CCPP 6.00 5.86 8.81 6.02 8.54 9.80 ND
NHPP 0.69 0.065 6.07 1.45 2.71 3.57 7.44
Average 5.20 2.08 8.55 2.46 6.81 10.20 4.88
Standard Error 0.58 0.40 0.78 0.35 0.83 1.32 1.54

ND: not done, sample values below the limit of detection or not used for this assessment.

Table 4.

ELISA interoperator precision analysis using coefficient of variation (%).

Antigen
RBD
Spike
Sample ID Total IgG Total IgG IgM IgA IgG1 IgG3 IgG4
A 1.48 9.12 4.82 10.15 ND 3.66 ND
B 12.1 3.50 9.53 3.40 ND ND ND
C 5.03 7.15 10.82 12.49 10.53 6.05 ND
D 3.98 5.68 12.58 10.03 19.67 ND ND
E 2.55 7.43 10.76 12.94 6.97 1.68 ND
F 2.08 0.42 9.28 8.39 8.54 4.65 7.02
G 4.05 2.38 12.59 12.34 8.72 ND ND
H 1.33 15.18 7.00 5.76 ND 1.24 ND
I 1.46 2.94 9.63 13.41 14.67 7.85 ND
J 5.05 8.73 7.91 9.23 ND ND ND
K 3.06 7.91 11.85 14.50 ND 7.80 ND
L 3.77 5.61 7.42 7.33 19.33 3.35 ND
M 4.68 8.39 10.36 6.27 12.76 4.42 13.18
N 0.43 4.85 7.93 8.28 4.53 4.75 ND
CCPP 5.55 10.21 7.26 14.58 10.62 2.19 1.49
NHPP 0.92 0.19 1.26 0.84 1.29 0 13.2
Average 3.60 6.23 8.81 9.37 10.69 3.66 8.73
Standard Error 0.70 0.97 0.74 1.00 1.72 0.71 2.82

ND: not done, sample values below the limit of detection or not used for this assessment.

3.2. Intraday variability

One operator completed three runs, totaling three sets of duplicate plates on the same day using like samples (Sample ID A-N, CCPP and NHPP). Endpoint titers were calculated as described above to evaluate the magnitude of total IgG, IgM, IgA, IgG1, IgG3 and IgG4 present in each sample (Supplemental Fig. 2) and the CV was evaluated across the three runs executed (Table 2). Importantly, the % CV for all samples tested and the average across samples was well below our endpoint of <20% variability (Table 2).

3.3. Interday variability

One operator performed three runs with like samples over three different days (Sample ID A-N, CCPP and NHPP). Endpoint titers were calculated as described (Supplemental Fig. 3) and the CV was evaluated across the three runs executed (Table 3). Importantly, all samples and the average across samples %CV was well below our endpoint of <20% variability (Table 3).

3.4. Interoperator variability

Three different technicians ran two duplicate plates using like samples on the same day. Similarly, endpoint titers were calculated (Supplemental Fig. 4) and CV evaluated for each antibody ELISA across the three runs executed (Table 4). These data demonstrate that the EPT ELISA methodology yields results within an acceptable range of variance, all below 20%, meeting the EMA guidelines.

3.4.1. Linearity

We used a linear regression analysis of O.D. compared to dilution to calculate the R2 for a positive samples across each antibody ELISA evaluated for spike and RBD (Total IgG only) responses. Importantly, only sample containing three data points in the linear range were used to calculate R (Cucinotta and Vanelli, 2020), and those values in plateau ranges were excluded (Fig. 3). The adjusted multiple R2 values calculated for spike antigen-specific responses were as follows: total IgG 0.930, IgM 0.862, IgA 0.927, IgG1 0.863, and IgG3 0.882 (Fig. 3). Given the low responder rate for IgG4 linearity requirements for sample inclusion were not met. The adjusted multiple R2 value calculated for total IgG RBD-specific responses was 0.903. Adjusted multiple R2 values for spike specific IgA, total IgG and RBD specific total IgG responses exceeded the 0.90 linearity qualification endpoint, demonstrating acceptable linearity for this methodology. While spike specific IgM, IgG3 and IgG1 adjusted multiple R2 values did not meet 0.90, they did approach this threshold. Importantly, the positive plate control CCPP exceeded an R2 value of 0.90 for each antibody class, except IgG4, suggesting these assays are linear.

Fig. 3.

Fig. 3

ELISA linearity for SARS-CoV-2 spike responses. Linear regression comparing O.D. 450-570 nm values to dilution of plasma for each antibody evaluated is shown. Linear model lines for each sample included for IgA, IgG1, IgG3, IgM and total IgG for both spike and RBD. Each line of best fit represents a unique sample. The adjusted multiple R2 value for each antibody class is shown.

3.4.2. Specificity

We evaluated the effectiveness of our assay to specifically distinguish antibodies in samples that recognize SARS-CoV-2 spike or RBD. Each antibody was screened against 92 total presumed negative (pre-pandemic) samples and 4 convalescent samples (or 2 in the case of IgG4) for spike antigen (Table 5), while 82 negative samples were leveraged for RBD ELISA analysis (Table 6). Each antibody evaluated met our endpoint of having ≥90% of negative samples screened return true negative values (below 3 SD of the plate LOD, while all SARS-CoV-2 convalescent samples evaluated returned 100% positive values (Table 5 and Table 6), establishing the specificity of our methodology.

Table 5.

Robust negative sample screen for SARS-CoV-2 spike ELISA EPT specificity.

Antibody Pre-pandemic samples (n = 92)
Convalescent samples (n = 4)
Negative Positive % negative Negative Positive % positive
Total IgG 84 8 91.0 0 4 100
IgG1 88 4 95.6 0 4 100
IgG3 89 3 96.7 0 4 100
IgG4 88 4 95.6 0 2 100
IgM 90 2 97.8 0 4 100
IgA 87 5 94.6 0 4 100

All ELISAs included 4 convalescent samples except IgG4 due to limited positive sample availability.

Table 6.

Robust negative sample screen for SARS-CoV-2 RBD ELISA EPT specificity.

Antibody Pre-pandemic samples (n = 82)
Convalescent samples (n = 4)
Negative Positive % negative Negative Positive % positive
Total IgG 74 8 91 0 4 100

3.4.3. Correlation to WHO reference standard

We ran a WHO reference positive standard (NIBSC 20/136) with an assigned antibody binding activity of 250 IU per ampule alongside a set of samples used for qualification to equate IU antibody binding activity to EPT. We observed a Total IgG endpoint titer of 4.73 and 4.47 against SARS-CoV-2 trimeric spike and RBD, respectively for the WHO sample (Fig. 4). Not surprisingly we saw a consistently lower EPT against RBD than Spike for the same plasma samples, including pooled controls (CCPP and NHPP) as well as the WHO sample (Fig. 4).

Fig. 4.

Fig. 4

SARS-CoV-2 spike and RBD total IgG endpoint titers evaluated for three individual samples (A, B, C) as well as positive WHO reference sample, pooled positive (CCPP) and negative (NHPP) control samples. Dashed line represents lower limit of detection.

3.5. Pseudovirus neutralization

3.5.1. Precision

Assay repeatability was assessed in terms of run to run variation. Samples from 2 subjects and 2 pooled plasma controls (CCPP and NHPP) were run by a single operator in duplicate across three separate days (interday), by a single operator in 3 different runs on the same day (intraday) and by three individual operators on the same day (interoperator). Data from individual samples (Positive, Negative) and pooled plasma controls (CCPP, NHPP) for all intraday, interday, and interoperator precision tests are shown in Fig. 5A, B, and C, respectively. Data from all 3 precision tests are compiled in Fig. 5D. In each individual test and in the compiled data shown in Fig. 5D, all samples fall within 2-fold of the geometric mean, demonstrating the assay is reproducible and precise.

Fig. 5.

Fig. 5

Pseudovirus neutralization reciprocal titer IC50 values evaluated for two individual samples (Positive, Negative) as well as pooled plasma positive (CCPP) and pooled plasma negative (NHPP) control samples. Data shows intraday (A), interday (B), and interoperator (C) variability. Panel D represents IC50 values from all qualification runs combined. Dashed line respresents lower limit of detection.

3.5.2. Linearity

RLU values from the Positive plasma samples of the interoperator precision run were used to evaluate assay linearity using linear regression analysis between Log-transformed RLU values and reciprocal plasma dilution concentrations. Only reciprocal plasma dilution concentrations within the linear range were used to generate representative graphs and calculate R2 values (Fig. 6). As the plasma became more dilute, the PFU values plateaued, and the assay was no longer in the linear range (data not shown). Duplicate samples for each operator were averaged for an R2 of 0.939, 0.929, and 0.975 for Operators 1, 2, and 3, respectively. The adjusted multiple R2 value across all positive samples in all qualification runs came to 0.896, demonstrating the pseudovirus neutralization assay methodology is linear.

Fig. 6.

Fig. 6

Graphical representations of linear regression models for the positive control sample for each of the 3 operators. The calculated R2 value for each operator is shown in the legend.

3.5.3. Specificity

Across our robust negative sample screen (pre-COVID-19 samples) for specificity, all 20 samples (100%) were negative for SARS-CoV-2 pseudovirus neutralization. All four positive samples (100%) inhibited SARS-CoV-2 pseudovirus by ≥50%. Neither of the two tested COVID-convalescent samples inhibited the VSV G pseudovirus by 50% at any plasma concentration (Fig. 7), demonstrating the specificity of the assay for SARS-CoV-2 Spike antigen and not overlapping with an unrelated viral antigen.

Fig. 7.

Fig. 7

Inhibition of VSV G pseudovirus entry and replication in HEK293T-ACE2 cells as a function of plasma concentration from healthy human donors (NHPP and Negative) or COVID-19 convalescent plasma samples (CCPP and Positive).

3.6. PRNT

3.6.1. Precision

To examine the sources of variation present in the assay, we determined the geometric mean of the PRNT80 values for each sample within each category of precision evaluated (run, operator, day). Data from individual samples (Positive, Negative) and pooled plasma controls (CCPP, NHPP) for all intraday, interday, and interoperator precision tests are shown in Fig. 8A, B, and C, respectively. Data from all 3 precision tests are compiled in Fig. 8D. In each individual test and in the compiled data shown in Fig. 8D, all samples fall within 2-fold of the geometric mean, demonstrating the assay is reproducible and precise.

Fig. 8.

Fig. 8

Reciprocal titer PRNT80 values evaluated for two individual samples (Positive, Negative) as well as pooled plasma positive (CCPP) and negative pooled plasma (NHPP) control samples. Data show intraday (A), interday (B), and interoperator (C) variability. Panel D represents PRNT80 values from all qualification runs combined. Dashed line respresents lower limit of detection.

3.6.2. Linearity

We evaluated the ability of the assay to return PFU values that were proportional to the dilution being examined using linear regression. Only the first four dilutions from each operator (1:80, 1:160, 1:320, and 1:640) were used to generate representative graphs and calculate R2 values (Fig. 9). As the plasma became more dilute, the PFU values plateaued, and the assay was no longer in the linear range (data not shown). Positive control samples from each of the three operators were assessed using a simple linear regression. Each operator's positive sample demonstrated a linear relationship (R2 = 0.922, 0.965, 0.950) between Log PFU and Log reciprocal dilution (Fig. 9). These data demonstrate that the PRNT assay generates results that are linear with respect to sample dilutions and that this feature is consistent across assay operators.

Fig. 9.

Fig. 9

Graphical representation of linear regression models for the Positive control sample run by each of the 3 operators. The adjusted multiple R2 value across this sample set is 0.9157.

3.6.3. Specificity

We next evaluated the specificity of serum/plasma samples to SARS-CoV-2 using a robust number of negative samples. Across our negative sample screen (pre-pandemic samples) for specificity, all 20 samples (100%) were negative for SARS-CoV-2 neutralization. All four positive samples (100%) inhibited SARS-CoV-2-PFUs by ≥80%. Both of these metrics meet our required endpoints for qualification of specificity of the methodology.

3.7. Assay alignment

Many researchers have demonstrated that higher magnitude antibody responses often correlate with enhanced protection in endpoint assays (Earle et al., 2021; Lee et al., 2020; Wajnberg et al., 2020; Salazar et al., 2020; Padoan et al., 2021). Consistent with recent reports (Sholukh et al., 2021), we aimed to assess this correlation in our own suite of qualified assays. Pooled plasma controls (CCPP and NHPP) were used across all assays and a single convalescent plasma sample was use in the pseudovirus neutralization and total IgG ELISA for RBD and Spike responses. We used these overlapping samples to demonstrate a correlation between EPT values and neutralization for High, Medium and Low responder samples (Fig. 10, Supplementary Fig. 5). We observed that CCPP consistently returned High responder values, the single convalescent samples consistently returned Medium responder values and NHPP was consistently Low across all endpoint assays evaluated (Supplemental Fig. 5). Furthermore, we are able to demonstrate a significant correlation between Reciprocal Titer IC50 from pseudovirus neutralization and both Total IgG Spike EPT (adjusted R2 = 0.9623) and Total IgG RBD EPT (adjusted R2 = 0.8921), as well as a correlation between Total IgG for Spike and RBD EPT (adjusted R2 = 0.9814) (Fig. 10).

Fig. 10.

Fig. 10

Total IgG EPT correlates with pseudovirus neutralization IC50 values. Correlations between Reciprocal Titer IC50 and (A) Total IgG Spike EPT, (B) Total IgG RBD EPT and (C) Total IgG RBD EPT and Total IgG Spike EPT. In each panel the blue line represents the line of best fit and shaded grey region is the 95% confidence interval. Adjusted R2 values shown for each correlation. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)

4. Conclusions

Our collective understanding of the required magnitude and composition of a protective immune response against SARS-CoV-2 continues to develop. Early investigations of infection-induced responses for diagnostic purposes included evaluation of IgA, IgM and IgG humoral immunity (Ma et al., 2020; Bichara et al., 2021; Maeda et al., 2021) and many commercial kits have been developed for these endpoints. SARS-CoV-2 specific IgA seems to be a high performing early diagnostic endpoint (Ma et al., 2020; Sterlin et al., 2021), while the kinetics of post-infection IgM and IgG better reflect peak and waning immunity in convalescent cohorts (Bichara et al., 2021; Duysburgh et al., 2021; De Greef et al., 2021). Interestingly, some studies have observed pro-inflammatory IgG subsets (IgG1 and IgG3) trend higher with severity of clinical COVID-19 disease while others (IgG2 and IgG4) are rarely detected (Yates et al., 2021; Luo et al., 2021). Many clinical vaccine immunogenicity and efficacy studies have focused on total IgG as a primary endpoint, and recent evaluations suggest this is a quality correlate with vaccine efficacy rates (Earle et al., 2021). Given the heterogenous health and infection history of the global population, we decided to include total IgG, IgG subsets, IgA and IgM in our qualifications to provide a comprehensive breadth of primary and secondary humoral immunity endpoints.

Here we have described robust methodologies qualified to evaluate the essential protective features of the humoral response against SARS-CoV-2. The EPT ELISA, pseudovirus neutralization and PRNT assays were all determined to be precise, linear, and specific for antibody-specific SARS-CoV-2 responses. These assays can be reliably used across days and by different trained operators to measure the effectiveness of clinical vaccine candidates both in the pipeline and approved for emergency use. In addition, the use of a WHO reference standard makes these qualified assays more accessible to other labs and researchers for use in comparing SARS-CoV-2 vaccine candidates. We have also established robust high, medium and low responder samples or sample pools that can be leveraged for future work.

Author contribution statement

SEL, BJB, TP, EC and RNC designed experiments; SEL, BJB, TP, EC, BDW, and EJ carried out experiments and SEL, PQ, BJB and BPB executed statistical analysis; LC, SW, EK, CS and NPK provided reagents; SEL, BJB and EC wrote the manuscript and prepared figures and tables with assistance from PQ, BPB, SLB, and RNC. All authors reviewed the manuscript.

Additional information

All authors declare no financial or alternative competing interests with the data described in this manuscript.

Declaration of Competing Interest

None.

Acknowledgements

The authors would like to express their gratitude to Seattle Children's Research Institute Leadership for their guidance, compassion, and mentorship. The authors would like to thank the Harrington Lab at Seattle Children's Research Institute as well as our colleague participants in the Center for Global Infectious Disease Research Biorepository for their commitment to science and resilience during a pandemic. The authors would like to thank their long-time colleague Dr. Jesse Erasmus for invaluable technical help when adopting and optimizing neutralization assays.

The datasets generated and/or analyzed during the study presented here are available from the corresponding authors on reasonable request.

Research reported here was supported by: National Institute of Allergy and Infectious Diseases (NIAID) of the National Institutes of Health under award numbers 1UM1AI148373-01 and 3UM1AI148373-01S1, a Bill & Melinda Gates Foundation (BMGF) grant to N.P.K. (OPP1156262), and B.P·B is supported through award number F32HD102290 from the National Institute of Child Health and Human Development (NICHD). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health or BMGF.

Footnotes

Appendix A

Supplementary data to this article can be found online at https://doi.org/10.1016/j.jim.2021.113160.

Appendix A. Supplementary data

Supplementary material

mmc1.docx (612.4KB, docx)

References

  1. Amanat F., et al. A serological assay to detect SARS-CoV-2 seroconversion in humans. Nat. Med. 2020;26:1033–1036. doi: 10.1038/s41591-020-0913-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Andreasson U., et al. A practical guide to immunoassay method validation. Front. Neurol. 2015;6:179. doi: 10.3389/fneur.2015.00179. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Bichara C.D.A., et al. Dynamics of anti-SARS-CoV-2 IgG antibodies post-COVID-19 in a Brazilian Amazon population. BMC Infect. Dis. 2021;21:443. doi: 10.1186/s12879-021-06156-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Burd E.M. Validation of laboratory-developed molecular assays for infectious diseases. Clin. Microbiol. Rev. 2010;23:550–576. doi: 10.1128/CMR.00074-09. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Chen Y., et al. A comprehensive, longitudinal analysis of humoral responses specific to four recombinant antigens of SARS-CoV-2 in severe and non-severe COVID-19 patients. PLoS Pathog. 2020;16 doi: 10.1371/journal.ppat.1008796. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Chesher D. Evaluating assay precision. Clin. Biochem. Rev. 2008;29(Suppl. 1):S23–S26. [PMC free article] [PubMed] [Google Scholar]
  7. Coler R.N., et al. The TLR-4 agonist adjuvant, GLA-SE, improves magnitude and quality of immune responses elicited by the ID93 tuberculosis vaccine: first-in-human trial. NPJ Vaccines. 2018;3:34. doi: 10.1038/s41541-018-0057-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Crawford K.H.D., et al. Protocol and reagents for Pseudotyping Lentiviral particles with SARS-CoV-2 spike protein for neutralization assays. Viruses. 2020;12:513. doi: 10.3390/v12050513. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Cucinotta D., Vanelli M. WHO declares COVID-19 a pandemic. Acta Biomed. 2020;91:157–160. doi: 10.23750/abm.v91i1.9397. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Dan J.M., et al. Immunological memory to SARS-CoV-2 assessed for up to 8 months after infection. Science. 2021;371 doi: 10.1126/science.abf4063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Day T.A., et al. Safety and immunogenicity of the adjunct therapeutic vaccine ID93 + GLA-SE in adults who have completed treatment for tuberculosis: a randomised, double-blind, placebo-controlled, phase 2a trial. Lancet Respir. Med. 2020 doi: 10.1016/s2213-2600(20)30319-2. [DOI] [PubMed] [Google Scholar]
  12. De Greef J., et al. Determinants of IgG antibodies kinetics after severe and critical COVID-19. J. Med. Virol. 2021;93:5416–5424. doi: 10.1002/jmv.27059. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Du L., et al. The spike protein of SARS-CoV — a target for vaccine and therapeutic development. Nat. Rev. Microbiol. 2009;7:226–236. doi: 10.1038/nrmicro2090. [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. Duysburgh E., et al. Persistence of IgG response to SARS-CoV-2. Lancet Infect. Dis. 2021;21:163–164. doi: 10.1016/S1473-3099(20)30943-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Earle K.A., et al. Evidence for antibody as a protective correlate for COVID-19 vaccines. Vaccine. 2021;39:4423–4428. doi: 10.1016/j.vaccine.2021.05.063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Erasmus J.H., et al. An Alphavirus-derived replicon RNA vaccine induces SARS-CoV-2 neutralizing antibody and T cell responses in mice and nonhuman primates. Sci. Transl. Med. 2020;12 doi: 10.1126/scitranslmed.abc9396. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Folegatti P.M., et al. Safety and immunogenicity of the ChAdOx1 nCoV-19 vaccine against SARS-CoV-2: a preliminary report of a phase 1/2, single-blind, randomised controlled trial. Lancet. 2020;396:467–478. doi: 10.1016/S0140-6736(20)31604-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Gach J.S., et al. HIV-1 specific antibody titers and neutralization among chronically infected patients on long-term suppressive antiretroviral therapy (ART): a cross-sectional study. PLoS One. 2014;9 doi: 10.1371/journal.pone.0085371. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Henss L., et al. Analysis of Humoral immune responses in patients with severe acute respiratory syndrome coronavirus 2 infection. J. Infect. Dis. 2020;223:56–61. doi: 10.1093/infdis/jiaa680. [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Jain S., Batra H., Yadav P., Chand S. COVID-19 vaccines currently under preclinical and clinical studies, and associated antiviral immune response. Vaccines. 2020;8:649. doi: 10.3390/vaccines8040649. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Jhang J.S., Chang C.-C., Fink D.J., Kroll M.H. Evaluation of linearity in the clinical laboratory. Arch. Pathol. Lab. Med. 2004;128:44–48. doi: 10.1043/1543-2165(2004)128&#x0003c;44:Eolitc&#x0003e;2.0.Co;2. [DOI] [PubMed] [Google Scholar]
  22. Keech C., et al. Phase 1–2 trial of a SARS-CoV-2 recombinant spike protein nanoparticle vaccine. N. Engl. J. Med. 2020;383:2320–2332. doi: 10.1056/NEJMoa2026920. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Lee W.T., et al. Neutralizing antibody responses in COVID-19 convalescent sera. J. Infect. Dis. 2020;223:47–55. doi: 10.1093/infdis/jiaa673. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Letko M., Marzi A., Munster V. Functional assessment of cell entry and receptor usage for SARS-CoV-2 and other lineage B betacoronaviruses. Nat. Microbiol. 2020;5:562–569. doi: 10.1038/s41564-020-0688-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. Liu W., et al. Evaluation of Nucleocapsid and spike protein-based enzyme-linked Immunosorbent assays for detecting antibodies against SARS-CoV-2. J. Clin. Microbiol. 2020;58 doi: 10.1128/jcm.00461-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Logunov D.Y., et al. Safety and efficacy of an rAd26 and rAd5 vector-based heterologous prime-boost COVID-19 vaccine: an interim analysis of a randomised controlled phase 3 trial in Russia. Lancet. 2021 doi: 10.1016/S0140-6736(21)00234-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Luo H., et al. The characterization of disease severity associated IgG subclasses response in COVID-19 patients. Front. Immunol. 2021;12:460. doi: 10.3389/fimmu.2021.632814. [DOI] [PMC free article] [PubMed] [Google Scholar]
  28. Ma H., et al. Serum IgA, IgM, and IgG responses in COVID-19. Cell. Mol. Immunol. 2020;17:773–775. doi: 10.1038/s41423-020-0474-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Maeda K., et al. Neutralization of SARS-CoV-2 with IgG from COVID-19-convalescent plasma. Sci. Rep. 2021;11:5563. doi: 10.1038/s41598-021-84733-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Mazzini L., et al. Comparative analyses of SARS-CoV-2 binding (IgG, IgM, IgA) and neutralizing antibodies from human serum samples. J. Immunol. Methods. 2021;489:112937. doi: 10.1016/j.jim.2020.112937. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Meesters R., Voswinkel S. Bioanalytical method development and validation: from the USFDA 2001 to the USFDA 2018 guidance for industry. J. Appl. Bioanal. 2018;4:67–73. [Google Scholar]
  32. Mohit E., Rostami Z., Vahidi H. A comparative review of immunoassays for COVID-19 detection. Expert. Rev. Clin. Immunol. 2021 doi: 10.1080/1744666x.2021.1908886. [DOI] [PubMed] [Google Scholar]
  33. Okba N.M.A., et al. Severe acute respiratory syndrome coronavirus 2-specific antibody responses in coronavirus disease patients. Emerg. Infect. Dis. 2020;26:1478–1488. doi: 10.3201/eid2607.200841. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Ou X., et al. Characterization of spike glycoprotein of SARS-CoV-2 on virus entry and its immune cross-reactivity with SARS-CoV. Nat. Commun. 2020;11:1620. doi: 10.1038/s41467-020-15562-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Padoan A., et al. Analytical and clinical performances of a SARS-CoV-2 S-RBD IgG assay: comparison with neutralization titers. Clin. Chem. Lab. Med. (CCLM) 2021 doi: 10.1515/cclm-2021-0313. [DOI] [PubMed] [Google Scholar]
  36. Perera R.A., et al. Serological assays for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), march 2020. Euro. Surveill. 2020;25:2000421. doi: 10.2807/1560-7917.ES.2020.25.16.2000421. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. Polack F.P., et al. Safety and efficacy of the BNT162b2 mRNA COVID-19 vaccine. N. Engl. J. Med. 2020;383:2603–2615. doi: 10.1056/NEJMoa2034577. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. Poland G.A., Ovsyannikova I.G., Kennedy R.B. SARS-CoV-2 immunity: review and applications to phase 3 vaccine candidates. Lancet. 2020;396:1595–1606. doi: 10.1016/S0140-6736(20)32137-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  39. Rice A., et al. A next generation bivalent human Ad5 COVID-19 vaccine delivering both spike and nucleocapsid antigens elicits Th1 dominant CD4+, CD8+ T-cell and neutralizing antibody responses. bioRxiv. 2020 doi: 10.1101/2020.07.29.227595. 2020.2007.2029.227595. [DOI] [Google Scholar]
  40. Roy V., et al. SARS-CoV-2-specific ELISA development. J. Immunol. Methods. 2020;484-485:112832. doi: 10.1016/j.jim.2020.112832. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Sadoff J., et al. Interim results of a Phase 1–2a trial of Ad26.COV2.S COVID-19 vaccine. N. Engl. J. Med. 2021 doi: 10.1056/NEJMoa2034201. [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Salazar E., et al. Relationship between anti-spike protein antibody titers and SARS-CoV-2 in vitro virus neutralization in convalescent plasma. bioRxiv Preprint Serv. Biol. 2020 doi: 10.1101/2020.06.08.138990. 2020.2006.2008.138990. [DOI] [Google Scholar]
  43. Schwartz J.L. Equitable global access to coronavirus disease 2019 vaccines. BMJ. 2020;371:m4735. doi: 10.1136/bmj.m4735. [DOI] [PubMed] [Google Scholar]
  44. Shang J., et al. Structural basis of receptor recognition by SARS-CoV-2. Nature. 2020;581:221–224. doi: 10.1038/s41586-020-2179-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  45. Sholukh A.M., et al. Evaluation of SARS-CoV-2 neutralization assays for antibody monitoring in natural infection and vaccine trials. medRxiv Preprint Serv. Health Sci. 2020 doi: 10.1101/2020.12.07.20245431. 2020.2012.2007.20245431. [DOI] [Google Scholar]
  46. Sholukh A.M., et al. Evaluation of cell-based and surrogate SARS-CoV-2 neutralization assays. J. Clin. Microbiol. 2021 doi: 10.1128/jcm.00527-21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  47. Sterlin D., et al. IgA dominates the early neutralizing antibody response to SARS-CoV-2. Sci. Transl. Med. 2021;13 doi: 10.1126/scitranslmed.abd2223. [DOI] [PMC free article] [PubMed] [Google Scholar]
  48. Tanne J.H. COVID-19: moderna plans booster doses to counter variants. BMJ. 2021;372:n232. doi: 10.1136/bmj.n232. [DOI] [PubMed] [Google Scholar]
  49. Wajnberg A., et al. Robust neutralizing antibodies to SARS-CoV-2 infection persist for months. Science. 2020;370:1227–1230. doi: 10.1126/science.abd7728. [DOI] [PMC free article] [PubMed] [Google Scholar]
  50. Walls A.C., et al. Elicitation of potent neutralizing antibody responses by designed protein nanoparticle vaccines for SARS-CoV-2. Cell. 2020;183:1367–1382. doi: 10.1016/j.cell.2020.10.043. e1317. [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. Wang W., et al. Global, regional, and national estimates of target population sizes for COVID-19 vaccination: descriptive study. BMJ. 2020;371:m4704. doi: 10.1136/bmj.m4704. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. Wherry R.J. A new formula for predicting the shrinkage of the coefficient of multiple correlation. Ann. Math. Stat. 1931;2(440–457):418. [Google Scholar]
  53. Widge A.T., et al. Durability of responses after SARS-CoV-2 mRNA-1273 vaccination. N. Engl. J. Med. 2020;384:80–82. doi: 10.1056/NEJMc2032195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  54. Williams T.C., Burgers W.A. SARS-CoV-2 evolution and vaccines: cause for concern? Lancet Respir. Med. 2021 doi: 10.1016/S2213-2600(21)00075-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  55. Yan R., et al. Structural basis for the recognition of SARS-CoV-2 by full-length human ACE2. Science. 2020;367:1444–1448. doi: 10.1126/science.abb2762. [DOI] [PMC free article] [PubMed] [Google Scholar]
  56. Yates J.L., et al. Serological analysis reveals an imbalanced IgG subclass composition associated with COVID-19 disease severity. Cell Rep. Med. 2021;2:100329. doi: 10.1016/j.xcrm.2021.100329. [DOI] [PMC free article] [PubMed] [Google Scholar]