EHR-integrated monitor data to measure pulse oximetry use in bronchiolitis

Andrew S Kern-Goldberger a, Irit R Rasooly a,b,c,d, Brooke Luo a,b,d, Sansanee Craig a,b,d, Daria F Ferro a,b,d, Halley Ruppel c,e, Padmavathy Parthasarathy a,d, Nathaniel Sergay a,f, Courtney M Solomon g,h, Kate E Lucey i,j, Naveen Muthu a,b,c,d, Christopher P Bonafide a,b,c,d, for the Pediatric Research in Inpatient Settings (PRIS) Network
PMCID: PMC8487905  NIHMSID: NIHMS1694378  PMID: 34583959

Abstract

Background:

Continuous pulse oximetry (SpO2) monitoring in hospitalized children with bronchiolitis not requiring supplemental oxygen is discouraged by national guidelines, but determining monitoring status accurately requires in-person observation.

Objective:

To determine whether electronic health record (EHR) data can accurately estimate the extent of actual pulse oximetry (SpO2) use in bronchiolitis.

Methods:

This repeated-cross sectional study included infants aged 8 weeks through 23 months hospitalized with bronchiolitis. In the validation phase at 3 children’s hospitals, we calculated the test characteristics of the SpO2 monitor data streamed into the EHR each minute when monitoring was active compared to in-person observation of SpO2 use. In the application phase at 1 children’s hospital, we identified periods when supplemental oxygen was administered using EHR flowsheet documentation and calculated the duration of SpO2 monitoring that occurred in the absence of supplemental oxygen.

Results:

Among 668 infants at 3 hospitals (validation phase), EHR-integrated SpO2 data from the same minute as in-person observation had a sensitivity of 90%, specificity of 98%, PPV of 88%, and NPV of 98% for actual SpO2 monitoring use. Using EHR-integrated data in a sample of 317 infants at 1 hospital (application phase), infants were monitored in the absence of oxygen supplementation for a median 4.1 hours (IQR 1.4–9.4 hours). Those who received supplemental oxygen experienced a median 5.6 hours (IQR 3.0–10.6 hours) of monitoring after oxygen was stopped.

Conclusion:

EHR-integrated monitor data is a valid measure of actual SpO2 monitoring use that may help hospitals more efficiently identify opportunities to deimplement guideline-inconsistent use.

Introduction

Overuse of continuous pulse oximetry (SpO2) monitoring in hospitalized children with bronchiolitis is increasingly recognized as an appropriate target for widespread deimplementation efforts.13 Bronchiolitis is highly prevalent, causing over 100,000 infants to be admitted each year, earning it the distinction of being the leading reason infants are hospitalized other than being born.4 Despite the existence of 3 sets of formal recommendations discouraging continuous SpO2 monitoring in bronchiolitis patients not requiring supplemental oxygen57 backed by evidence demonstrating absence of benefit,8,9 potential harm,1012 and nurse alarm fatigue that can result,13,14 the practice remains widespread.1

The first step toward deimplementing any overused practice is measuring the current extent of the practice.15,16 Prior quality improvement initiatives used the presence of physician orders as a measure of SpO2 monitoring use.17,18 Unfortunately, a more recent study revealed that orders do not accurately represent actual SpO2 monitoring use at the bedside.19 A recent prospective study using in-person observation of monitoring status in 56 hospitals revealed that 46% of children hospitalized with bronchiolitis received SpO2 monitoring even though monitoring was unnecessary or inappropriate according to the guidelines.1 While these data provided valuable insights into the prevalence of guideline-inconsistent monitoring, the in-person observational methods were resource-intensive and not sustainable for ongoing measurement.

In contrast, SpO2 data from bedside monitors has the potential to provide comprehensive, longitudinal, automated measurement of actual SpO2 monitoring use to drive audit and feedback interventions. While these data have been difficult to access in the past, increasingly hospitals have chosen to integrate monitor data into their electronic health records (EHRs) to reduce nurses’ work effort spent manually documenting vital signs. When the monitor data are stored and saved in the EHR or data warehouse at high frequency, identifying patients who are actively monitored becomes feasible.

In this study of hospitalized infants with bronchiolitis, we aimed to: (1) Measure the validity of SpO2 monitor data integrated into the electronic health record (EHR) against the gold standard of in-person observation to determine actual SpO2 monitoring use, and (2) apply this measure to estimate the extent of SpO2 monitoring use in the absence of supplemental oxygen using EHR-integrated monitor data.

Methods

This work occurred in the context of two separate larger projects. Validation was performed as a substudy using data from a 6-center pilot single arm clinical trial in the Eliminating Monitor Overuse (EMO) research portfolio (National Clinical Trial Identifier NCT04178941) that measured the feasibility, acceptability, appropriateness, and outcomes of audit and feedback with educational outreach as a strategy to align continuous SpO2 use in stable bronchiolitis patients with evidence and guideline recommendations. All participating centers were members of the Pediatric Research in Inpatient Settings (PRIS) Network, an independent, hospital-based research network that aims to improve the health of and health care delivery to hospitalized children and their families. As part of this study, we validated EHR-integrated monitor data against in-person observation in the 3 hospitals with integrated EHRs actively receiving physiologic monitor data. The study was approved by the institutional review board (IRB) at the coordinating site, with reliance agreements established with the other sites. Waivers of consent, assent, parental permission and HIPAA authorization were granted at all sites.

After validation, we applied the validated measure to a full season of bronchiolitis patients’ monitor data as part of a quality improvement initiative conducted by the Patient Safety Learning Laboratory (PSLL) at the aforementioned coordinating site, a tertiary children’s hospital. The PSLL is funded by the Agency for Healthcare Research and Quality, and one of its primary aims is to locally re-engineer the system of monitoring hospitalized children on acute care units, with a focus on reducing non-informative alarms and accelerating nurse responses to critical events. The hospital’s IRB determined that the PSLL’s work described in this manuscript was consistent with quality improvement and did not meet criteria for human subjects research.

1. Validation of EHR-integrated monitor data in 3 hospitals

Setting and Participants

In the trial that provided the data for the validation, we collected data on infants 8 weeks through 23 months hospitalized with acute bronchiolitis on 8 units that cared for bronchiolitis patients on generalist services at 3 geographically diverse tertiary children’s hospitals between December 1, 2019 and March 14, 2020. Consistent with prior work in the EMO portfolio,1 we only observed infants who were not currently receiving supplemental oxygen therapy (or room air flow). We excluded patients with premature birth (<28 weeks gestation or documented “premature” without gestational age listed), cyanotic congenital heart disease, pulmonary hypertension, home oxygen or positive pressure ventilation requirement, tracheostomy, neuromuscular disease, immunodeficiency, cancer, or diagnosis of heart failure, myocarditis, arrhythmia, or COVID-19 disease.

Data Collection

After identifying eligible patients meeting the criteria above, observers at each hospital walked to each patient’s bedside, visualized the physiologic monitor display to determine whether SpO2 tracings were present, and recorded the date and time of the observation. Observers then returned to the EHR to obtain demographic and clinical information, including determining presence of integrated monitor data displayed in the EHR. Specifically, observers recorded whether monitor data were captured in the EHR during the same minute as in-person observation (“minute 0”); during each of the 15 minutes before in-person observation (“minutes −1 to −15”); and during each of the 15 minutes after in-person observation (“minutes +1 to +15”). Observers were also prompted to complete an optional free-text field with comments pertaining to monitor data.

Analysis

Using in-person observation as the gold standard for actual SpO2 use, we determined the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of SpO2 monitor data visible in the EHR at (a) minute 0; (b) minutes −1 to −15; (c) minutes +1 to +15; and (d) minutes −15 to +15, encompassing time periods (a), (b) and (c). We calculated these test characteristics overall, and stratified by each participating hospital. Two investigators reviewed and classified observer comments to provide context for cases in which EHR-integrated monitor data and in-person observation were discordant. When there were no associated observer comments, discrepancies were considered “unexplained.” We decided a priori that a PPV and NPV of at least 80% would be sufficient for use in future improvement and research efforts.

2. Application of validated measure to bronchiolitis patients’ monitor data in 1 hospital

Setting and Participants

In order to apply the new method for bronchiolitis quality improvement efforts at a single hospital, we accessed nursing flowsheet documentation data and physiologic monitor data from the EHR on all hospitalized patients aged 8 weeks to 23 months admitted to a general pediatrics service with a discharge diagnosis of acute bronchiolitis between September 3, 2019 and January 1, 2020. Discharge diagnoses were defined by the International Classifications of Diseases (ICD) 9th and 10th revision codes (493.x, ICD-9; J45.x, ICD-10). We excluded patients with discharge diagnoses corresponding to comorbid respiratory conditions (Supplemental Table 3), patients with complex chronic conditions (CCCs) as previously defined by the CCC version 2 classification system,20 and patients with an intensive care unit (ICU) stay during the hospitalization so as to study monitoring practices in a sample of patients who were (a) representative of hospitalized infants with bronchiolitis nationally and (b) less likely to have other reasons to have SpO2 monitoring due to comorbidities or recovery from critical illness. For context, physiologic monitors are present in every inpatient ward bed space at this hospital. Monitors are used if ordered by the physician, and/or if the nurse deems its use clinically appropriate.

Data Collection

We defined our cohort of interest using EHR data from our institutional data warehouse. We extracted flowsheet data for nursing respiratory assessments with timestamps, where nurses manually document information including supplemental oxygen FiO2, flow rate, and support devices (e.g. nasal cannula). We then used a unique visit identifier to link visits in our cohort directly to the EHR’s central database containing the synchronized SpO2 data automatically transmitted every minute from the bedside monitor to the EHR (Clarity, Epic Systems, Verona, WI). Flowsheet data from the data warehouse were merged with the synchronized SpO2 data from Clarity and filtered based on inpatient unit admission and discharge timestamps for each patient, allowing us to eliminate data points corresponding to time accrued in the emergency department.

Analysis

1. Determination of when a patient is receiving supplemental oxygen

In order to determine the time intervals when a patient was receiving supplemental oxygen, we first identified initiation times, defined as the time of the first flowsheet assessment documenting its use. There were 3 flowsheet rows that could be used to indicate supplemental oxygen: “support device,” “flow rate,” and “FiO2.” Given the imperfection of clinical data for research, we took a conservative approach to identifying supplemental oxygen that would favor sensitivity over specificity. With that guiding principle in mind, the following entries were defined as indicating supplemental oxygen use: “nasal cannula” or “high flow nasal cannula” support device, or any documented flow rate greater than 0, or FiO2 > 21%. We conservatively assumed that supplemental oxygen continued for the entire interval between subsequent entries consistent with supplemental oxygen administration. The time point of supplemental oxygen discontinuation was defined as flowsheet entries with simultaneous documentation of FiO2 of “NA” or “21,” flow rate of “NA” or “21” (a common documentation error), without simultaneous documentation of support device “nasal cannula” or “high flow nasal cannula.” If a patient on supplemental oxygen had no flowsheet assessment documenting its discontinuation, a patient was assumed to remain on supplemental oxygen until the time of discharge. We did not make assumptions about supplemental oxygen use prior to the first flowsheet assessment documented during the hospitalization.

2. Determination of when a patient is SpO2 monitored

Each minute for which a patient had an SpO2 value streamed from the bedside monitor into the EHR was considered to have been “monitored” during that minute.

3. Determination of when SpO2 monitoring is occurring in the absence of supplemental oxygen

For each minute of the hospitalization, a patient’s supplemental oxygen status was directly compared to the patient’s monitoring status to determine monitoring use during periods without supplemental oxygen. To approximate consensus recommendations from a recent expert panel, we added a transitional hour after the final documented use of supplemental oxygen during each hospitalization before considering monitoring use to be possibly inconsistent with guideline reccomendations.7 For patients who never received supplemental oxygen during the hospitalization, based on all 3 existing guideline recommendations,57 we measured every minute of SpO2 monitoring following the first documented flowsheet assessment.

Using these definitions, for each hospitalization we calculated the total number of monitored minutes in the absence of supplemental oxygen. We calculated monitored time overall, and classified monitored time by whether supplemental oxygen was ever received during the hospitalization (yes vs. no) and patient age group (2–11 months vs. 12–23 months). We performed a sensitivity analysis excluding hospitalizations exceeding the 75th percentile for length of stay, which we presumed may reflect atypical hospitalization courses. We used REDCap 9.10.0 software for data management and Stata (StataCorp LLC, College Station, TX) version 16 for all analyses.

Results

1. Validation of EHR-integrated monitor data

We observed 668 patients across 3 hospitals: 410 patients at Hospital A, 161 patients at Hospital B, and 97 patients at Hospital C. Figure 1 demonstrates the test characteristics of EHR-integrated data for actual SpO2 monitoring use, considering each of the 4 time periods surrounding in-person observation and indicating discrepancies noted by observers between EHR data and in-person observation. The overall PPV among all hospitals over 4 time periods ranged from 88% to 93%; NPV ranged from 94% to 98%. EHR-integrated data present at minute 0 among all hospitals was the most sensitive marker of actual SpO2 monitoring use (90%). The decrements in sensitivity observed for minutes −1 to −15, +1 to +15, and −15 to +15 were explained in part by observer comments noting monitoring use for transient interventions including vital signs and suctioning, and the discontinuation of SpO2 monitoring following in-person observation. Conversely, EHR data was most specific for actual SpO2 monitoring use for minutes −15 to +15 (99%), with minimal decrements to specificity for the other time periods investigated (98% specificity for actual SpO2 monitoring use at minute 0). Test characteristics stratified by each participating hospital are displayed in Table 1; all PPV and NPV were >80% except for “minute 0” at Hospital A (79%).

Figure 1.

Figure 1.

Test Characteristics of EHR-Integrated Data Compared to Actual Pulse Oximetry Monitoring Use.

EHR-integrated monitor data was compared to in-person observation (as the gold standard for actual pulse oximetry monitoring status) in hospitalized infants with bronchiolitis at 3 hospitals. Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were obtained for (A) Minute 0, the minute at which in-person observation was completed; (B) Minutes −1 to −15, each of the 15 minutes preceding in-person observation; (C) Minutes +1 to +15, each of the 15 minutes following in-person observation; and (D) Minutes −15 to +15, the minute at which in-person observation was completed, and each of the 15 minutes preceding and following inperson observation. Boxes with the dashed outline indicate reasons for discrepancy between actual pulse oximetry monitoring status and EHR-integrated data that were obtained from observer comments during chart review. Discrepancies were considered “unexplained” when there were no associated observer comments.

Table 1.

Test Characteristics of EHR-Integrated Monitor Data as a Measure of Actual Pulse Oximetry Monitoring Use

Test Characteristic by Hospital Minutes during which pulse oximetry data are present in EHR
0a −1 to −15b +1 to +15c −15 to +15d
Hospital A (n=410)
Sensitivity 41/44 (93%) 36/44 (82%) 27/44 (61%) 25/44 (57%)
Specificity 355/366 (97%) 358/366 (98%) 360/366 (98%) 361/366 (99%)
PPV 41/52 (79%) 36/44 (82%) 27/33 (82%) 25/30 (83%)
NPV 355/358 (99%) 358/366 (98%) 360/377 (95%) 361/380 (95%)
Hospital B (n=161)
Sensitivity 29/30 (97%) 24/30 (80%) 22/30 (73%) 19/30 (63%)
Specificity 131/131 (100%) 131/131 (100%) 131/131 (100%) 131/131 (100%)
PPV 29/29 (100%) 24/24 (100%) 22/22 (100%) 19/19 (100%)
NPV 131/132 (99%) 131/137 (96%) 131/139 (94%) 131/142 (92%)
Hospital C (n=97)
Sensitivity 22/28 (79%) 22/28 (79%) 19/28 (68%) 19/28 (68%)
Specificity 68/69 (99%) 69/69 (100%) 67/69 (97%) 69/69 (100%)
PPV 22/23 (96%) 22/22 (100%) 19/21 (90%) 19/19 (100%)
NPV 68/74 (92%) 69/75 (92%) 67/76 (88%) 69/78 (88%)
All Hospitals (n=668)
Sensitivity 92/102 (90%) 82/102 (80%) 68/102 (67%) 63/102 (62%)
Specificity 554/566 (98%) 558/566 (99%) 558/556 (99%) 561/566 (99%)
PPV 92/104 (88%) 82/90 (91%) 68/76 (89%) 63/68 (93%)
NPV 554/564 (98%) 558/578 (97%) 558/592 (94%) 561/600 (94%)

Abbreviations: EHR, electronic health record; PPV, positive predictive value; NPV; negative predictive value

a

The minute at which in-person observation was completed.

b

Each of the 15 minutes preceding in-person observation

c

Each of the 15 minutes following in-person observation

d

The minute at which in-person observation, and each of the 15 minutes preceding and following in-person observation

2. Application of validated measure to bronchiolitis patients’ monitor data in 1 hospital

317 hospitalizations were included in the final dataset. We excluded 3 hospitalizations for which there were minimal or no monitor data despite >24 hours of supplemental oxygen administration, suggesting that data may not have been transmitted from the bedside monitor to the EHR. As demonstrated in Table 2, patients were monitored overall for a median 17.2 hours (IQR 2.4–34.4 hours). Excluding a transitional hour after the final documented use of supplemental oxygen, a median 4.1 hours (IQR 1.4–9.4 hours) of monitoring occurred in the absence of supplemental oxygen. Monitoring in the absence of supplemental oxygen accounted for a median 10.6% of patients’ hospital length of stay (IQR 3.7–24.3%). Patients who received supplemental oxygen during the hospitalization experienced a median 5.6 hours (IQR 3.0–10.6 hours) of monitoring after supplemental oxygen was stopped, and patients who did not receive supplemental oxygen experienced a median 1.4 hours (IQR 0.3–6.7 hours) of monitoring. 94% of patients who received supplemental oxygen experienced over 1 hour of monitoring after supplemental oxygen was stopped, and 64% experienced over 4 hours of additional monitoring. The extent of SpO2 monitoring was similar by patient age category (2–11 months and 12–23 months). A sensitivity analysis that excluded patients in the top 25th percent of length of stay demonstrated consistent findings.

Table 2.

Pulse Oximetry Monitoring in Hospitalized Infants with Bronchiolitis Using EHR-Integrated Data

Admission Characteristics Monitoring Measure
Hospital Length of Stay in Hours, median (IQR) Monitored Time in Hours, median (IQR) Monitored Time Without Supplemental O2a in Hours, median (IQR) Proportion of Hospitalization Monitored Without Supplemental O2, median, (IQR) > 1 hour Monitored Without Supplemental O2, n (%) >4 hours Monitored Without Supplemental O2, n (%)
Overall (n=317) 33.1 (21.0–49.5) 17.2 (2.4–34.4) 4.1 (1.4–9.4) 10.6 (3.7–24.3) 246 (78) 162 (51)
Ever on supplemental O2
 Yes (n=186) 42.2 (31.8–60.5) 30.4 (19.7–42.2) 5.6 (3.0–10.6) 13.1 (6.9–22.1) 175 (94) 119 (64)
 No (n=131) 20.4 (16.0–30.7) 1.4 (0.3–6.7) 1.4 (0.3–6.7) 5.4 (1.4–30.0) 71 (54) 43 (33)
Age (months)
 2–11 (n=220) 34.2 (20.5–54.3) 17.1 (2.0–34.8) 4.2 (1.4–9.9) 10.1 (3.8–23.9) 172 (78) 116 (53)
 12–23 (n=97) 31.8 (21.5–43.8) 17.4 (6.3–28.3) 3.8 (1.2–7.8) 11.8 (3.5–24.7) 74 (76) 46 (47)

Abbreviations; EHR, electronic health record; IQR, interquartile range; O2, supplemental oxygen

a

Refers to monitored time >1 hour after supplemental oxygen was stopped, or any monitored time for patients who did not receive supplemental oxygen.

Discussion

We validated minute-level EHR-integrated SpO2 monitor data as a measure that can approximate actual SpO2 monitoring use. Using this method of identifying SpO2 monitoring use, we estimated that infants hospitalized with bronchiolitis at 1 children’s hospital were monitored in the absence of supplemental oxygen for a median 10.6% of their hospital length of stay.

Compared to orders, the EHR method we describe here appears to be favorable. In our prior study evaluating the presence of monitoring orders to in-person observation, sensitivity was 49%, specificity was 89%, PPV was 77% and NPV was 69%; the test characteristics in our study outperform monitoring orders for each of these parameters. Our study also builds on prior work describing the use of SpO2 data streamed from physiologic monitors to estimate aggregate rates of monitored patients at the unit and hospital levels,21,22 and a single-center quality improvement initiative that used this method to estimate time on pulse oximetry monitoring in a population of children with wheezing who were not on supplemental oxygen.23 However, to our knowledge, this is the first study to validate this method against in-person observation, and to specifically estimate overuse based on monitoring practices following the transition off supplemental oxygen, as recently recommended by consensus guidelines.7 Moreover, as recent guideline recommendations also address continuous SpO2 monitoring use in hospitalized children experiencing other diseases including asthma, community-acquired pneumonia, and croup,7 leveraging EHR-integrated data may enhance existing quality measurement tools for these diseases and ameliorate the limitations of existing quality indicators in predicting clinical outcomes.24 Finally, in estimating that transitioning patients weaned off supplemental oxygen to intermittent SpO2 monitoring could save up to a median 5.6 hours of monitored time per patient, this study highlights the opportunity for deimplementation efforts in this population to reduce unnecessary SpO2 monitoring and the associated costs,10 iatrogenic harm,11 and alarm fatigue13,14 described in the literature.

We note several limitations of this study. First, while the overall PPV and NPV of EHR-integrated data for actual SpO2 use were otherwise >80% in the validation phase, the PPV for “minute 0” was 79% at Hospital A, which upon review of observer comments is attributable to at least 4 of 11 discordant instances in which the bedside monitor visual display screen was turned off (leading to a determination that the patient was unmonitored according to our technical definition) but the patient was in fact monitored, representing an area to target for improvement. Second, our full understanding of why EHR-integrated data differed from in-person observation is limited by the large number of discrepancies for which observers did not offer an explanation. Third, while generalizability may be limited by the current availability of EHR-integrated data at some hospitals, this study demonstrates the value of this tool as hospitals increasingly adopt interoperable health information technology.25,26 Fourth, neither SpO2 data stored in the EHR nor data collected in cross-sectional in-person observation distinguishes between continuous SpO2 monitoring versus intermittent SpO2 measurement. However, our methodology overall provides a conservative estimate of SpO2 use in the absence of supplemental oxygen, which outweighs the likely trivial contribution of intermittent SpO2 measurements captured by coincidence, such as those obtained during routine vital sign measurements. Finally, our estimates of SpO2 use at a single hospital, such as a low (5.4%) proportion of hospitalization monitored among patients who never received supplemental O2, may reflect this hospital’s local monitoring culture and prior participation in the EMO:SpO2 research portfolio, and may differ at other sites.

Conclusions

EHR-integrated data can be reliably used to determine actual SpO2 use at the bedside, and in a sample of infants hospitalized with bronchiolitis, demonstrates a high burden of SpO2 monitoring that appears to be outside of guideline recommendations after oxygen supplementation has been discontinued. This measure can be leveraged to inform targeted efforts to deimplement unnecessary use of continuous pulse oximetry monitoring in hospitalized patients with bronchiolitis, as well as in other patient populations.

Supplementary Material

Supplemental Table 3

Funding Source:

Research reported in this publication was supported in part by a Cooperative Agreement from the National Heart, Lung, and Blood Institute (NHLBI) of the National Institutes of Health under award number U01HL143475 (Bonafide, PI). As a Cooperative Agreement, NIH scientists participated in study conference calls and provided ongoing feedback on the conduct and findings of the study. Research reported in this publication was supported in part by a grant from the Agency for Healthcare Research and Quality under award number R18HS026620 (Bonafide, PI).

Funders had no role in the project design; in the collection, analysis, or interpretation of data; in the decision to submit the manuscript for publication; or in the writing of the manuscript.

The authors additionally acknowledge the following individuals for their assistance with data collection:

Ann and Robert H. Lurie Children’s Hospital of Chicago, Chicago, IL

Waheeda Samady, MD, MSCI

Kristin Van Genderen, MD

Children’s Medical Center Dallas, Dallas, TX

Gaylan Dascanio, MD

Yomna Farooqi, MD

Mayra Garcia, DNP, RN, PCNS-BC

Sarah Khan, MD

Caitlin Reaves, BSN, RN, CPN

Hailee Scoggins, BSN, RN, MSN

Children’s Hospital of Philadelphia, Philadelphia, PA

Laura Goldstein, MD

Muida Menon, RN

Financial Disclosures:

Dr. Bonafide discloses additional grant funding from the National Science Foundation for research related to physiologic monitoring.

References

  • 1.Bonafide CP, Xiao R, Brady PW, et al. Prevalence of Continuous Pulse Oximetry Monitoring in Hospitalized Children With Bronchiolitis Not Requiring Supplemental Oxygen. JAMA. 2020;323(15):1467–1477. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Rasooly IR, Beidas RS, Wolk CB, et al. Measuring overuse of continuous pulse oximetry in bronchiolitis and developing strategies for large-scale deimplementation: study protocol for a feasibility trial. Pilot Feasibility Stud. 2019;5:68. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Cheston CC, Vinci RJ. Overuse of Continuous Pulse Oximetry for Bronchiolitis: The Need for Deimplementation Science. JAMA. 2020;323(15):1449–1450. [DOI] [PubMed] [Google Scholar]
  • 4.Hasegawa K, Tsugawa Y, Brown DF, Mansbach JM, Camargo CA, Jr. Trends in bronchiolitis hospitalizations in the United States, 2000–2009. Pediatrics. 2013;132(1):28–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Quinonez RA, Garber MD, Schroeder AR, et al. Choosing wisely in pediatric hospital medicine: five opportunities for improved healthcare value. J Hosp Med. 2013;8(9):479–485. [DOI] [PubMed] [Google Scholar]
  • 6.Ralston SL, Lieberthal AS, Meissner HC, et al. Clinical practice guideline: the diagnosis, management, and prevention of bronchiolitis. Pediatrics. 2014;134(5):e1474–1502. [DOI] [PubMed] [Google Scholar]
  • 7.Schondelmeyer AC, Dewan ML, Brady PW, et al. Cardiorespiratory and Pulse Oximetry Monitoring in Hospitalized Children: A Delphi Process. Pediatrics. 2020;146(2). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.McCulloh R, Koster M, Ralston S, et al. Use of Intermittent vs Continuous Pulse Oximetry for Nonhypoxemic Infants and Young Children Hospitalized for Bronchiolitis: A Randomized Clinical Trial. JAMA Pediatr. 2015;169(10):898–904. [DOI] [PubMed] [Google Scholar]
  • 9.Principi T, Coates AL, Parkin PC, Stephens D, DaSilva Z, Schuh S. Effect of Oxygen Desaturations on Subsequent Medical Visits in Infants Discharged From the Emergency Department With Bronchiolitis. JAMA Pediatr. 2016;170(6):602–608. [DOI] [PubMed] [Google Scholar]
  • 10.Cunningham S, Rodriguez A, Boyd KA, McIntosh E, Lewis SC, Group BC. Bronchiolitis of Infancy Discharge Study (BIDS): a multicentre, parallel-group, double-blind, randomised controlled, equivalence trial with economic evaluation. Health Technol Assess. 2015;19(71):i-xxiii, 1–172. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.McBride SC, Chiang VW, Goldmann DA, Landrigan CP. Preventable adverse events in infants hospitalized with bronchiolitis. Pediatrics. 2005;116(3):603–608. [DOI] [PubMed] [Google Scholar]
  • 12.Schroeder AR, Marmor AK, Pantell RH, Newman TB. Impact of pulse oximetry and oxygen therapy on length of stay in bronchiolitis hospitalizations. Arch Pediatr Adolesc Med. 2004;158(6):527–530. [DOI] [PubMed] [Google Scholar]
  • 13.Bonafide CP, Lin R, Zander M, et al. Association between exposure to nonactionable physiologic monitor alarms and response time in a children’s hospital. J Hosp Med. 2015;10(6):345–351. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Bonafide CP, Localio AR, Holmes JH, et al. Video Analysis of Factors Associated With Response Time to Physiologic Monitor Alarms in a Children’s Hospital. JAMA Pediatr. 2017;171(6):524–531. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Parsons Leigh J, Niven DJ, Boyd JM, Stelfox HT. Developing a framework to guide the de-adoption of low-value clinical practices in acute care medicine: a study protocol. BMC Health Serv Res. 2017;17(1):54. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Prasad V, Ioannidis JP. Evidence-based de-implementation for contradicted, unproven, and aspiring healthcare practices. Implement Sci. 2014;9:1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Mittal S, Marlowe L, Blakeslee S, et al. Successful Use of Quality Improvement Methodology to Reduce Inpatient Length of Stay in Bronchiolitis Through Judicious Use of Intermittent Pulse Oximetry. Hosp Pediatr. 2019;9(2):73–78. [DOI] [PubMed] [Google Scholar]
  • 18.Ralston SL, Garber MD, Rice-Conboy E, et al. A Multicenter Collaborative to Reduce Unnecessary Care in Inpatient Bronchiolitis. Pediatrics. 2016;137(1). [DOI] [PubMed] [Google Scholar]
  • 19.Brady PW, Schondelmeyer AC, Landrigan CP, Xiao R, Brent C, Pediatric Research in Inpatient Settings N. Validity of Continuous Pulse Oximetry Orders for Identification of Actual Monitoring Status in Bronchiolitis. J Hosp Med. 2020;15(11):665–668. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Feudtner C, Feinstein JA, Zhong W, Hall M, Dai D. Pediatric complex chronic conditions classification system version 2: updated for ICD-10 and complex medical technology dependence and transplantation. BMC Pediatr. 2014;14:199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Schondelmeyer AC, Bonafide CP, Goel VV, et al. The frequency of physiologic monitor alarms in a children’s hospital. J Hosp Med. 2016;11(11):796–798. [DOI] [PubMed] [Google Scholar]
  • 22.Schondelmeyer AC, Brady PW, Goel VV, et al. Physiologic Monitor Alarm Rates at 5 Children’s Hospitals. J Hosp Med. 2018;13(6):396–398. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Schondelmeyer AC, Simmons JM, Statile AM, et al. Using quality improvement to reduce continuous pulse oximetry use in children with wheezing. Pediatrics. 2015;135(4):e1044–1051. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Mangione-Smith R, Zhou C, Williams DJ, et al. Pediatric Respiratory Illness Measurement System (PRIMES) Scores and Outcomes. Pediatrics. 2019;144(2). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.DeSalvo KB, Dinkler AN, Stevens L. The US Office of the National Coordinator for Health Information Technology: Progress and Promise for the Future at the 10-Year Mark. Ann Emerg Med. 2015;66(5):507–510. [DOI] [PubMed] [Google Scholar]
  • 26.Holmgren AJ, Patel V, Adler-Milstein J. Progress In Interoperability: Measuring US Hospitals’ Engagement In Sharing Patient Data. Health Aff (Millwood). 2017;36(10):1820–1827. [DOI] [PubMed] [Google Scholar]