Abstract
Compared with photon stereotactic body radiotherapy (SBRT) plans that may have to use many more penetrating x-ray beams for each isocenter, proton SBRT with ultrahypofractionated doses use fewer beam angles and offer significantly reduced low-dose radiation bath to normal liver tissue. We demonstrate techniques to deliver safe and effective proton SBRT, where planning and organ motion complexity further increased with multiple liver lesions. For treatment planning, we recommend robust and logical beam angles, avoiding devices and encouraging entry perpendicular to the dominant motion, as well as volumetric repainting to mitigate the interplay effect to clinically acceptable levels. This report highlights the significant technical challenges with ultrahypofractionated proton pencil beam scanning liver therapy, how they are managed, and the effectiveness of this treatment.
Keywords: proton therapy, liver metastasis, SBRT, stereotactic body radiotherapy, PBS, pencil beam scanning
Introduction
Ultrahypofractionated proton stereotactic body radiotherapy (SBRT) for liver metastasis is attractive, but technically challenging given organ motion, localization, and worry about the interplay effect. We highlight a case and technical challenges and requisite solutions to treat patients with this technique. For this patient and for our liver proton SBRT program, at simulation we recommend custom whole-body immobilization and a thin abdominal compression belt to reduce intra-abdominal organ motion and avoid bulky stereotactic frames. Challenges to overcome at treatment planning include choosing logical and short beam paths, motion management, and overcoming tumor and normal organ motion during scanning therapy. On-board localization with cone beam computed tomography (CT) is critical for target matching.
Using these methods, we report the case of a patient with oligometastatic cancer with 2 liver lesions. The patient was prescribed pencil beam scanning (PBS) proton liver SBRT with 50 GyRBE in 5 fractions to each lesion, with a generic proton relative biological effectiveness (RBE) of 1.1. The patient has been disease free for 2 years after therapy completion followed with positron emission tomography–computed tomography (PET/CT) and clinical surveillance; no radiation-related sequelae were noted. This therapy can be very attractive for the appropriate patient, and further prospective evaluation is warranted.
Case Report
The patient was a 77-year-old woman diagnosed with right estrogen receptor positive (ER+) breast cancer in 1998, which was treated with mastectomy. In 2010, she experienced right nodal recurrence, which was treated with axillary dissection, postmastectomy radiation (52 Gy to chest and nodes and 46 Gy to supraclavicular nodes), and systemic therapy. In 2018, the patient had recurrence with liver magnetic resonance imaging (MRI) showing a 2.2-cm lesion in segment 8 (biopsy confirmed) and a 0.8-cm lesion in segment 7 (Figure 1). PET/CT showed no other disease. The patient started on palbociclib and letrozole. Given oligometastatic disease and prior radiotherapy, patient was offered proton PBS liver SBRT. She had no history of liver disease.
Fiducial markers were placed adjacent to the anterior lesion. The patient was simulated supine, arms up, in full-body custom immobilization using a compression belt for intra-abdominal organ and respiratory motion management with contrast-enhanced 4-dimensional (4D) computed tomography (CT) and MRI simulation. Presimulation fluoroscopic assessment ensured significant reduction in target motion with goal interaction range variations <3 mm for 95% of the target volume [1].
With generic proton RBE of 1.1, the patient was treated with PBS 50 GyRBE in 5 fractions to each lesion, completing therapy September 2018, using a 360° rotational gantry (Ion Beam Applications, Louvain-La-Neuve, Belgium) (Figure 2). Key treatment-planning goals are shown in the Table. Key metrics were planning target volume 50 with 95% dose > 50 Gy and normal liver gross target volume (GTV) of the primary tumor was V15 Gy < 700 cm3. Three right anterolateral oblique beams were used to treat the anterior lesion, and 2 posterior oblique beams were used for the posterior. 4D robust single-field optimization was used with field weighting divided equally among all fields in each plan (each field delivering approximately 1/3 of the dose in the anterior lesion plan and 1/2 of the dose in the posterior lesion plan). Robustness settings used to optimize for internal GTV coverage included 0.5 cm isocenter shifts and 3.5% range uncertainty perturbations. Internal GTV coverage was optimized to ensure 95% of the target volume received at least 47.5 Gy in the worst-case uncertainty dose volume histogram scenario. Given the extent of tumor motion due to respiration, each plan was also optimized for nominal 95% coverage of the PTVs to 47.5 Gy as an additional measure to ensure plan robustness (Figure 2). We used methods of 4D dynamic dose calculation to reduce interplay effects via volumetric repainting as previously described [2]. Liver GTV volume was 985 cm3. Mean liver GTV dose was 11.6 Gy. Volume of liver GTV receiving ≥15Gy was 284 cm3, and the volume receiving <15 Gy was 701 cm3. The lesions were treated simultaneously, every other day, using orthogonal imaging and kilovoltage CBCT for alignment. The anterior lesion was aligned to fiducials, and the posterior lesion was aligned to stable vascular and bony anatomy. Mean time on the treatment table was 36.9 minutes and 31.6 minutes for the anterior and posterior lesions, respectively.
Table.
Target and critical normal tissue constraints
| ||||
Structure
|
Dose volume histogram objective, cGy |
Evaluator
|
Variation acceptable
|
Priority
|
PTV_5000 | D95% | >5000 | 3000 | 2 |
PTV_5000_Eval | D95% | >5000 | 3000 | 2 |
Spinal cord | D0.03 cm3 | <3000 | 3150 | 1 |
Spinal cord | D0.1 cm3 | <2500 | 2625 | 1 |
Spinal cord | D0.35 cm3 | <2300 | 2415 | 1 |
Spinal cord | D1.2 cm3 | <1450 | 1522.5 | 1 |
Esophagus | D0.03 cm3 | <3500 | 3675 | 3 |
Esophagus | D5 cm3 | <1950 | 2047.5 | 3 |
Heart | D0.03 cm3 | <3800 | 3990 | 3 |
Heart | D15 cm3 | <3200 | 3360 | 3 |
Chest wall | D0.03 cm3 | <4300 | 4515 | 3 |
Chest wall | D30 cm3 | <3250 | 3675 | 3 |
Skin | D0.03 cm3 | <3950 | 4147.5 | 3 |
Skin | D10 cm3 | <3650 | 3832.5 | 3 |
Stomach | D0.03 cm3 | <3200 | 3360 | 1 |
Stomach | D0.1 cm3 | <2750 | 2887.5 | 1 |
Stomach | D10 cm3 | <1800 | 1890 | 1 |
Duodenum | D0.03 cm3 | <3200 | 3360 | 2 |
Duodenum | D0.1 cm3 | <3000 | 3150 | 2 |
Duodenum | D5 cm3 | <1800 | 1890 | 2 |
Duodenum | D10 cm3 | <1250 | 1312.5 | 2 |
Bowel, small | D0.03 cm3 | <3500 | 3675 | 1 |
Bowel, small | D5 cm3 | <1950 | 2047.5 | 1 |
Bowel, large | D0.03 cm3 | <3800 | 3990 | 2 |
Bowel, large | D20 cm3 | <2500 | 2625 | 2 |
Gallbladder | D0.1 cm3 | <4000 | 4200 | 3 |
Lungs | D1500 cm3 | <1250 | 1312.5 | 3 |
Lungs | D1000 cm3 | <1350 | 1417.5 | 3 |
Kidneys | D200 cm3 | <1750 | 1837.5 | 3 |
Liver gross tumor volume | CV1500 cGy (cm3) | >700 | 1 |
Abbreviations: PTV_5000, planning target volume XXXX; Eval, XXXX; CV, XXXX.
Follow-up PET/CT and liver MRI at 3 months showed complete response to proton SBRT. Subsequent MRIs (Figure 3) showed stable proton tracks in liver, most pronounced for the posterior lesion. The patient experienced acute grade 1 fatigue, but no long-term sequelae. Two weeks after SBRT, she began palbociclib and letrozole. She experienced transient elevation of liver enzymes, which resolved after palbociclib discontinuation. The patient resumed letrozole and has been disease free for 2.5 years after treatment followed by PET/CT.
Discussion
To our knowledge, this is the first published report of simultaneously treated lesions using ultrahypofractionated PBS-proton SBRT. We highlight the value of abdominal compression to mitigate organ motion, 4D CT, repainting, and image-guided therapy to optimize ultrahypofractionated dose delivery in a patient with multiple lesions. Proton SBRT for liver metastasis is attractive. Compared with photon SBRT plans that may use more penetrating x-rays for each isocenter, proton SBRT plans have little exit dose beyond target for each beam, significantly reducing low dose to normal liver.
Proton liver therapy has been evaluated in several prior studies, including PBS proton therapy, although proton ultrahypofractionated SBRT experience is limited, and much long-term clinical data is limited to passively scattered beams [3, 4]. Early proton liver therapy reports demonstrated reduced side effects concordant with reduced liver dose to 700 cm3 normal liver, and other organs at risk, particularly dome and central tumors >3 cm [5, 6]. In a phase II study of proton liver SBRT for patients with limited extrahepatic disease and median dose 40 GyRBE in 5 fractions (median biologically effective dose = 72 Gy) using passively scattered beams, local control was 70.1% at 1 year. Treatment was well tolerated, and patients with large liver tumors had excellent control, although the authors indicated that further dose escalation could result in further improved local control. Most PBS liver studies have involved fractionated proton therapy. Dionisi et al [7] reported on 18 patients with hepatobiliary tumors treated with 15 fraction PBS liver therapy with median dose 58 GyRBE. One-year local control was 90%, with excellent toxicity. Kanai et al [8] reported on 20 liver cancer patients comparing 4D robust optimization (4DRO) compared to range-adapted internal target volume , previously treated with gated PBS carbon-therapy, finding 4D robust optimization had higher mean clinical target volume 95% dose using 48 GyRBE in 2 fractions.
However, there are technical challenges to effectively delivering ultrahypofractionated proton liver therapy, such as on-board imaging, motion management, and technology (scattered beams vs PBS) [5]. While respiratory management with deep breath-hold, gating, and compression is well-characterized in photon therapy, it is less well characterized in protons, even though geometric distortion is the predominant driver for poor proton conformality [9–11].
We used rational beam selection and abdominal compression to mitigate organ motion, which we demonstrated to reduce interplay effects levels to that acceptable in clinical practice [1]. Our group has described optimal beam angle selection in PBS liver therapy and found that coplanar right anterolateral and posterior oblique angles are optimal. We used abdominal compression, which we found critical to reduce perpendicular amplitude in moderate (7-10 mm) and significant (>10 mm) motion, although it is likely not necessary in patients with small amplitude motion. Thin, soft, homogeneous compression belts have reduced influence on proton range compared with bulky heterogeneous stereotactic frames. For proton liver SBRT, we recommend robust and logical right anterolateral and posterior oblique angles, avoiding devices. Predominant motion direction (ie, perpendicular to dominant motion) must be reduced, and a compression belt is an attractive solution.
Additional planning techniques are critical to further optimize motion mitigation, including volumetric repainting (using the statistical average of positional errors), spot sequence delivery optimization, and robust optimization. For large lesions PBS produces more conformality than scattered beams, as the scattered spread-out Bragg peak cannot be adjusted for different portions of the target. Motion issues are more pronounced in PBS than scattered beams, given the interplay effect, which is when target motion during active beam scanning and treatment delivery results in differences in the projected doses [12]. The 4D dose must be calculated to evaluate the interplay effect of treatment and repainting, which is beam scanning multiple times during a single fraction, and can be used to reduce uncertainty to clinically acceptable levels. PBS may be more advantageous than scattered beams for additional reasons (eg, scattered beams would require heavy mostly blocked fields producing more neutrons).
Finally, on-board imaging is important for localization and understanding anatomic changes around the target, leading to significant dose perturbation (eg, lung tissue or effusions) [13, 14]. The Imaging and Radiation Oncology Core proton liver phantom demonstrates only 38% pass rate in quality assurance, given that the phantom incorporates multiple lesions and a moving target [15]. The pass rate for all moving phantoms was 63%. CBCT is increasing in proton therapy centers; however, quality of CBCT, pretreatment verification, and robotic repositioning vary. Quality assurance CT provides high-fidelity dose recalculation, but not real-time localization. For proton SBRT, we recommend orthogonal imaging for alignment, if available, and then on-board CBCT matched to liver fiducials, immobile adjacent structures, or liver contour for final localization.
Multiple-lesion SBRT takes the aforementioned technical challenges and then adds liver-sparing challenges and increased time requirements. Even with photons, multiple lesion treatment with photons requires careful treatment planning to ensure adequate liver and organ-at-risk sparing due to intersecting or overlapping beams [16, 17]. We chose optimal beams, as described earlier, to reduce end-of-range uncertainties from coplanar opposed beams, but note that time of delivery was significant given that multiple fields were used to treat each lesion in a single session in a multi-room center.
In summary, to our knowledge, we present the first case of simultaneous treatment of multiple liver metastases treated with PBS-proton SBRT. The patient has remained disease free after proton SBRT for over 2 years on endocrine therapy alone with no chronic toxicity related to radiation. This case represents a technical challenge due to the need for respiratory motion management, mitigating interplay effect, multiple lesion treatment, and use of PBS over scattered beams, but it also highlights an excellent treatment option in proton centers. From the Miami Liver Proton Therapy Conference, there was consensus that radiation therapy is effective for liver tumors and that proton therapy should be considered for select patients with metastasis [18]. There continues to be a need to improve technical barriers to treatment delivery, and additional clinical data are needed considering that outcomes data are largely based on passive scattered beams.
ADDITIONAL INFORMATION AND DECLARATIONS
CRediT: Neil K. Taunk: conceptualization, writing – draft, review, editing; Brendan Burgdorf: data curation, methodology; Lei Dong: methodology, writing – draft, review, editing; Edgar Ben-Josef: conceptualization, writing – draft, review, editing.
Conflicts of Interest: Lei Dong, PhD, is an associate editor of the International Journal of Particle Therapy. The authors have no other relevant conflicts of interest to disclose.
Funding: The authors have no funding to disclose.
Ethical Approval: Institutional review approval was previously obtained by our institution to report this single case.
References
- 1.Lin L, Souris K, Kang M, Glick A, Lin H, Huang S, Stutzer K, Janssens G, Sterpin E, Lee JA, Solberg TD, McDonough JE, Simone CB, 2nd, Ben-Josef E. Evaluation of motion mitigation using abdominal compression in the clinical implementation of pencil beam scanning proton therapy of liver tumors. Med Phys. 2017;44:703–12. doi: 10.1002/mp.12040. [DOI] [PubMed] [Google Scholar]
- 2.Zeng C, Plastaras JP, Tochner ZA, White BM, Hill-Kayser CE, Hahn SM, Both S. Proton pencil beam scanning for mediastinal lymphoma: the impact of interplay between target motion and beam scanning. Phys Med Biol. 2015;60:3013–29. doi: 10.1088/0031-9155/60/7/3013. [DOI] [PubMed] [Google Scholar]
- 3.Fukumitsu N, Okumura T, Takizawa D, Makishima H, Numajiri H, Murofushi K, Ohnishi K, Mizumoto M, Aihara T, Ishikawa H, Tsuboi K, Sakurai H. Proton beam therapy for metastatic liver tumors. Radiother Oncology. 2015;117:322–7. doi: 10.1016/j.radonc.2015.09.011. [DOI] [PubMed] [Google Scholar]
- 4.Chadha AS, Gunther JR, Hsieh CE, Aliru M, Mahadevan LS, Venkatesulu BP, Crane CH, Das P, Herman JM, Koay EJ, Taniguchi C, Holliday EB, Minsky BD, Suh Y, Park P, Sawakuchi G, Beddar S, Odisio BC, Gupta S, Loyer E, Kaur H, Raghav K, Javle MM, Kaseb AO, Krishnan S. Proton beam therapy outcomes for localized unresectable hepatocellular carcinoma. Radiother Oncology. 2019;133:54–61. doi: 10.1016/j.radonc.2018.10.041. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Arscott WT, Thompson RF, Yin L, Burgdorf B, Kirk M, Ben-Josef E. Stereotactic body proton therapy for liver tumors: dosimetric advantages and their radiobiological and clinical implications. Phys Imaging Radiat Oncol. 2018;8:17–22. doi: 10.1016/j.phro.2018.11.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Gandhi SJ, Liang X, Ding X, Zhu TC, Ben-Josef E, Plastaras JP, Metz JM, Both S, Apisarnthanarax S. Clinical decision tool for optimal delivery of liver stereotactic body radiation therapy: photons versus protons. Pract Radiat Oncol. 2015;5:209–18. doi: 10.1016/j.prro.2015.01.004. [DOI] [PubMed] [Google Scholar]
- 7.Dionisi F, Brolese A, Siniscalchi B, Giacomelli I, Fracchiolla F, Righetto R, Morganti AG, Pravadelli C, Avancini I, Rozzanigo U, Mattiuzzi A, Frisinghelli M, Pertile R, Ciarleglio FA, Vitale A, Schwarz M, Amichetti M. Clinical results of active scanning proton therapy for primary liver tumors. Tumori. 2021;107:71–9. doi: 10.1177/0300891620937809. [DOI] [PubMed] [Google Scholar]
- 8.Kanai T, Paz A, Furuichi W, Liu CS, He P, Mori S. Four-dimensional carbon-ion pencil beam treatment planning comparison between robust optimization and range-adapted internal target volume for respiratory-gated liver and lung treatment. Phys Med. 2020;80:277–87. doi: 10.1016/j.ejmp.2020.11.009. [DOI] [PubMed] [Google Scholar]
- 9.Parkes MJ, Green S, Cashmore J, Stevens AM, Clutton-Brock TH, Bel A, Lens E, Lohr F, Boda-Heggemann J. In Regard to Boda-Heggemann et al. Int J Radiat Oncol Biol Phys. 2016;96:709–10. doi: 10.1016/j.ijrobp.2016.07.001. [DOI] [PubMed] [Google Scholar]
- 10.Engelsman M, Schwarz M, Dong L. Physics controversies in proton therapy. Semin Radiat Oncol. 2013;23:88–96. doi: 10.1016/j.semradonc.2012.11.003. [DOI] [PubMed] [Google Scholar]
- 11.De Ruysscher D, Sterpin E, Haustermans K, Depuydt T. Tumour movement in proton therapy: solutions and remaining questions: a review. Cancers (Basel) 2015;7:1143–53. doi: 10.3390/cancers7030829. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Kang M, Huang S, Solberg TD, Mayer R, Thomas A, Teo BK, McDonough JE, Simone CB, II, Lin L. A study of the beam-specific interplay effect in proton pencil beam scanning delivery in lung cancer. Acta Oncol. 2017;56:531–40. doi: 10.1080/0284186X.2017.1293287. [DOI] [PubMed] [Google Scholar]
- 13.Veiga C, Janssens G, Teng CL, Baudier T, Hotoiu L, McClelland JR, Royle G, Lin L, Yin L, Metz J, Solberg TD, Tochner Z, Simone CB, II, McDonough J, Kevin Teo BK. First clinical investigation of cone beam computed tomography and deformable registration for adaptive proton therapy for lung cancer. Int J Radiat Oncol Biol Phys. 2016;95:549–59. doi: 10.1016/j.ijrobp.2016.01.055. [DOI] [PubMed] [Google Scholar]
- 14.Molitoris JK, Diwanji T, Snider JW, III, Mossahebi S, Samanta S, Badiyan SN, Simone CB, II, Mohindra P. Advances in the use of motion management and image guidance in radiation therapy treatment for lung cancer. J Thorac Dis. 2018;10:S2437–S50. doi: 10.21037/jtd.2018.01.155. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Taylor PA, Kry SF, Alvarez P, Keith T, Lujano C, Hernandez N, Followill DS. Results from the Imaging and Radiation Oncology Core Houston's anthropomorphic phantoms used for proton therapy clinical trial credentialing. Int J Radiat Oncol Biol Phys. 2016;95:242–8. doi: 10.1016/j.ijrobp.2016.01.061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Rusthoven KE, Kavanagh BD, Cardenes H, Stieber VW, Burri SH, Feigenberg SJ, Chidel MA, Pugh TJ, Franklin W, Kane M, Gaspar LE, Schefter TE. Multi-institutional phase I/II trial of stereotactic body radiation therapy for liver metastases. J Clin Oncol. 2009;27:1572–8. doi: 10.1200/JCO.2008.19.6329. [DOI] [PubMed] [Google Scholar]
- 17.Crane CH, Koay EJ. Solutions that enable ablative radiotherapy for large liver tumors: fractionated dose painting, simultaneous integrated protection, motion management, and computed tomography image guidance. Cancer. 2016;122:1974–86. doi: 10.1002/cncr.29878. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Chuong MD, Kaiser A, Khan F, Parikh P, Ben-Josef E, Crane C, Brunner T, Okumura T, Schreuder N, Bentzen SM, Gutierrez A, Mendez Romero A, Yoon SM, Sharma N, Kim TH, Kishi K, Moeslein F, Hoffe S, Schefter T, Hanish S, Scorsetti M, Apisarnthanarax S. Consensus report from the Miami Liver Proton Therapy Conference. Front Oncol. 2019;9:457. doi: 10.3389/fonc.2019.00457. [DOI] [PMC free article] [PubMed] [Google Scholar]