Characterizing spatial differences between SPECT-ventilation and SPECT-perfusion in patients with lung cancer undergoing radiotherapy

Farnoush Forghani 1, Taylor Patton 2, Jennifer Kwak 3, David Thomas 1, Quentin Diot 1, Chad Rusthoven 1, Richard Castillo 4, Edward Castillo 5, Inga Grills 5, Thomas Guerrero 5, Moyed Miften 1, Yevgeniy Vinogradskiy 1
PMCID: PMC8489737  NIHMSID: NIHMS1700997  PMID: 33964328

Summary:

This study investigates agreement between ventilation and perfusion for lung cancer patients undergoing radiotherapy. Ventilation-perfusion scans of nineteen patients with stage III lung cancer from a prospective protocol were compared using voxel-wise Spearman correlation- coefficients. The presented results show in about 25% of patients, ventilation and perfusion exhibit lower agreement.

Keywords: SPECT, Ventilation, Perfusion, functional avoidance, lung cancer, Spearman correlation, Dice Similarity, 4D-CT

Introduction:

Functional avoidance thoracic radiotherapy is a paradigm that proposes to use functional imaging to generate radiation treatment plans that preferentially avoid the functional portions of the lung. The hypothesis of functional avoidance radiotherapy is that reduced dose to functional portions of the lung can lead to reduced probability that patient develop pulmonary side-effects1,2. One form of functional imaging that has been proposed for functional avoidance is 4-dimensional computed tomography (4DCT)-derived ventilation. 4DCT images contain the local density changes in the pulmonary parenchyma due to respiration. These changes can be extracted mathematically by correspondence of image intensity values between spatially registered 4DCT phase images to obtain a surrogate for pulmonary ventilation 3,4. This technique is well suited for integration into the radiotherapy workflow because 4DCT scans are routinely acquired for treatment planning for most lung cancer patients. Therefore, obtaining 4DCT-ventilation images can provide functional information while eliminating the extra cost and reducing imaging-based radiation exposure.

Early research efforts focused on validation of 4DCT-ventilation 59 followed by modeling studies that demonstrated the potential clinical utility of 4DCT-ventilation-based functional avoidance for reducing treatment-related toxicity outcomes2,1012. Based on the retrospective data, several clinical trials (NCT02528942, NCT02308709, NCT02843568) investigating the use of 4DCT-ventilation in the context of functional avoidance radiotherapy 13,14 are underway. One potential limitation of current formulations of 4DCT-ventilation imaging is that only ventilation information is generated. An implicit assumption in most 4DCT-ventilation studies has been that for most lung cancer patients, the spatial profiles of ventilation and perfusion are similar. The similarity between ventilation and perfusion for lung cancer patients has been noted in other clinical scenarios1519. While ventilation and perfusion have been shown to be spatially similar for most patients, there are clinical scenarios where spatial differences exist. The underlying cause of dissimilarity between ventilation and perfusion can be due to perfusion defects for patients with a pulmonary embolism (blood clot in one of the pulmonary vessels)20 or due to ventilation defects for patients with pulmonary edema (alveoli filling with fluid), atelectasis (deflation of the alveoli), pleural effusion (a build-up of fluid around the lung), or obstruction of the airways21. While there have been studies comparing 4DCT-ventilation to other lung function imaging modalities (both ventilation and perfusion)57, there has been no work to quantitatively test the assumption that ventilation profiles are spatially equivalent to perfusion profiles for lung cancer patients undergoing radiotherapy. In order to accurately characterize the physiological spatial differences between ventilation and perfusion, a study is needed that can evaluate spatial differences between ventilation and perfusion using the same imaging modality. The purpose of this study was therefore to use single-photon emission computerized tomography (SPECT) imaging to quantitatively characterize the agreement between ventilation and perfusion for lung cancer patients undergoing radiotherapy.

Methods and Materials:

Patient cohort

Patients for the current study were selected from those enrolled on a 2-institution, prospective clinical trial (NCT02528942) for functional avoidance thoracic radiation therapy. 67 patients with biopsy proven lung cancer included in the trail. Informed consent was obtained for all subjects for this study. The trial aim was to use 4DCT-based ventilation imaging to reduce dose to functional portions of the lung with the idea that reduced doses to functional lung may lead to reduced rates of radiation-related thoracic toxicity. The primary objective of the trial was to determine whether 4DCT-ventilation-based functional avoidance results in reduced radiation pneumonitis when compared to current standard of care techniques. The primary end-point of the trial was to evaluate the rate of radiation pneumonitis. The trial was approved to enroll patients by the institutional review boards (IRB) at the University of Colorado (Aurora, CO) (IRB #14–1586) and Beaumont Health System (Royal Oak, MI) (IRB #2016–037). Of the patients enrolled on the functional avoidance clinical trial, nineteen patients with stage III lung cancer who underwent pre-treatment SPECT-CT imaging were eligible for the current analysis.

Pulmonary SPECT imaging

SPECT imaging started with ventilation imaging using Technetium-labeled (Tc-99m) diethylenetriaminepentaacetic (DTPA) acid aerosol inhalation with the approximate activity of 30 MBq. Immediately after ventilation, perfusion images were acquired using Technetium-labeled particles of macroaggregates of albumin (Tc-99m, MAA) with the approximate activity of 100 MBq that were given intravenously. The number of expected ventilation and perfusion counts is on the order of few 10^6 per 10 min. Perfusion counting rate is about 4 times of ventilation counting rate. A Siemens Simbia T6 SPECT—CT scanner (Siemens, Hoffman Estates, IL) was used to acquire the SPECT images. Patients were in the supine position with normal resting breathing. Two detectors with low-energy and high-resolution collimation were used to obtain the raw emission data. Image reconstruction was performed using the 3D ordered-subset expectation—maximization (3D OSEM) iterative reconstruction method. Co-registered CT images were acquired to perform attenuation correction of the SPECT images. The voxel size in the SPECT data was 7.75 mm × 7.75 mm × 3 mm.

Quantitative functional comparison

The ventilation and perfusion components of the SPECT-CT were registered using MIM Software (Cleveland, OH). Each ventilation-perfusion registration was manually reviewed and adjusted if necessary. The next step was to appropriately segment the lung volume of interest on both the ventilation and perfusion scans. Lung segmentation for the current analysis included the lung parenchyma and excluded the gross tumor volume as well as any aerosol deposition artifacts. Both SPECT-ventilation and SPECT-perfusion are susceptible to aerosol clumping artifacts5,22,23. The clumping of radioactive aerosol results in locally enhanced counts that manifest as bright image hotspots. We carefully excluded radioaerosol deposition artifacts from the SPECT images based on the location of these hotspots in the central airways (trachea, bronchi). Central airways are the sites for radioaerosol deposition due to turbulent air flow in the airways caused by obstructive airways processes such as COPD (chronic obstructive pulmonary disease) and chronic bronchitis. In addition, hotspots located in the posterior aspects of the lower lobes of the lung caused by dependent atelectasis from supine positioning during imaging were excluded. A board-certified nuclear medicine radiologist oversaw the identification of artifacts and the review of the SPECT-CT images.

In order to improve the quantitative accuracy of the analysis, each ventilation and perfusion artifact was manually removed from the lung segmentation. The final step in the segmentation was to crop the contour 5 mm away from the lung boundary in order to reduce the partial volume artifact24. Once the lungs were appropriately segmented, normalization of ventilation and perfusion raw data was done by converting both images to percentile images to facilitate direct comparison between the modalities25,26. Two metrics previously used for image comparison studies59,25 were employed to compare the ventilation and perfusion distributions: Spearmen correlation coefficients and Dice Similarity Coefficients (DSC). DSCs were calculated for different functional zones defined as 0–25th percentile, 25–50th percentile, 50–75th percentile, and >75th percentile25. Previous studies show the most reproducible functional profiles are in the highest and lowest functional zones 5,6,27. Therefore, this study focuses on 0–25th percentile and >75th percentile functional zones. In addition to analyzing the whole lung, ipsilateral and contralateral lungs were analyzed separately. The correlation data are presented in the form of histogram distributions with the results summarized as mean ± standard deviation (range). All computational analysis was done using MIM Extensions.

Spearman correlation and Dice similarity metrics may not elucidate spatial differences between SPECT ventilation and perfusion images and could be influenced by non-relevant outlier such as the residual of hotspot artifacts. Therefore, voxel-wise scatter plots of percentile perfusion versus ventilation values were also generated for each patient and included in the supplementary materials.

Results and Discussion

The median age of the cohort was 70 (range 54–84) years, all patients had stage III lung cancer, and 63% of patients had chronic obstructive pulmonary disease.

The SPECT-CT scans in Figure 1 represent an example of high spatial correlation (Figures 1A and 1B) and an example of low spatial correlation (Figures 1C and 1D) between ventilation and perfusion. In the top row of Figure 1 the Spearman correlation coefficients for the presented patients are 0.84 and 0.82 for the right and left lungs, respectively. The Spearman correlation coefficient for the presented patient in the bottom row is 0.19 for both the right and left lung. The high (>75%) and low functional zones (<0.25%) (the 25%-50% and 50%-75% functional zones contours are omitted for clarity) for the patients are overlaid on the scans. Similar to the visual percentile correlations, the high and low functional zones show good agreement for the patient presented in the top row and poor agreement between the patient presented in the bottom row.

Figure 1: Percentile SPECT-CT image sets: A) and B) show the images for a patient with high spatial agreement between ventilation and perfusion (Spearman correlation coefficient of 0.84 for the right lung and 0.82 for the left lung). The tumor was located in the right lung. The Dice similarity coefficients for the low and high functional regions of the whole lung are 0.83 and 0.57, respectively. C) and D) show the images for a patient with low spatial agreement between ventilation and perfusion (Spearman correlation coefficient of 0.19 for the right and left lungs). The tumor was located in the right lung. The Dice similarity coefficients for the low and high functional regions of the whole lung are 0.40 and 0.24, respectively. The segmented low and high functional lung regions are shown in blue and light green, respectively.

Figure 1:

The mean ± standard deviation (range) of Spearman coefficients was 0.64±0.19 (0.19, 0.91) for the whole lung, 0.66±0.23 (0.19, 0.91) for the ipsilateral, and 0.62±0.16 (0.19, 0.82) for the contralateral lungs. For the low functional zones (0–25 percentile), the mean ± standard deviation (range) of DSC was 0.65±0.13 (0.38, 0.92) for the whole lung, 0.67± 0.16 (0.38, 0.92) for the ipsilateral, and 0.63±0.09 (0.47,0.83) for the contralateral lungs. The mean and range of DSC for the high functional zones (75–100 percentile) was 0.54±0.14 (0.19, 0.80) for the whole lung, 0.55±0.15 (0.24, 0.80) for the ipsilateral, and 0.53± 0.13(0.19,0.75) for the contralateral lungs.

Histogram distributions of Spearman correlation coefficients between ventilation and perfusion for the ipsilateral, contralateral, and the whole lung for the entire patient cohort are shown in Figure 2. For the ipsilateral lungs, 73.7% of the scans exhibit a higher correlation (≥0.5) between ventilation and perfusion and 26.3% of the scans exhibit lower correlation (<0.5). For the contralateral lungs, 79.0% of the scans exhibit a higher correlation and 21.0% of the scans exhibit lower correlation. For the whole lung, 76.3% of the scans exhibit a higher correlation and 23.7% of the scans exhibit lower correlation.

Figure 2: Histogram of Spearman correlation coefficients, illustrating the similarity distribution of ventilation and perfusion scans among all patients for the A) ipsilateral, B) contralateral, and C) whole lung.

Figure 2:

Examples of the scatter plots for two patients with lower and two patients with higher similarity coefficients between ventilation and perfusion are provided in Figure 2 of the supplementary materials. These examples showed agreement between the scatter plot and quantitative analysis by Spearman and Dice coefficients in determining the similarity between the perfusion and ventilation scans.

The presented work quantitatively evaluated the spatial agreement between SPECT ventilation and perfusion using 19 SPECT scans for stage III lung cancer patients. The data can be interpreted from the perspective of 4DCT-based functional avoidance, that uses functional imaging to generate radiation treatment plans that preferentially avoid the functional portions of the lung. The motivation for this work stems from 4DCT-ventilation because with current formulations of 4DCT-based functional imaging, only the ventilation component can be computed. One argument that 4DCT-ventilation studies have made is acceptable to use ventilation as a surrogate for both ventilation and perfusion because ventilation and perfusion are spatially similar for most lung cancer patients. However, this assumption (of ventilation/perfusion being spatially similar) has not been validated/quantified. Our results were consistent between the ipsilateral, contralateral, and the whole lungs among the patient cohort (histograms in Figure 2) and quantitatively revealed that about 74% of patients had ≥ 0.5 correlations for the ipsilateral lung, 79% of patients had≥ 0.5 correlations for the contralateral lung, and about 76% had ≥ 0.5 correlations for the whole lung. The 74%-79% rate of patients with correlations ≥ 0.5 between ventilation and perfusion can be used as a baseline figure to evaluate in what percentage of cases does perfusion provide unique information when compared to ventilation alone. In other words, in about 21%-26% of patients both ventilation and perfusion would be needed to provide complete (ventilation and perfusion) lung function information. The subsequent extension of the current work is to propagate the ventilation-perfusion spatial differences into differences in the treatment planning domain. Previous studies 28,29 have evaluated differences between 4DCT-ventilation-based and SPECT-perfusion-based dose-function metrics. Nakajima et al 28 found strong correlations (range, r=0.94–0.97) between ventilation-based and perfusion-based dose-function metrics but noted wide variability possible among individual patients with differences as large as 6.6 Gy possible between ventilation and perfusion in functional dose-metrics. Using the data from the dose-function comparison studies 28,29, it is difficult to isolate whether the potential individual differences in dose-function metrics are due to differences between the two modalities (4DCT versus SPECT), true clinical differences between the patient’s ventilation and perfusion profiles, differences caused by the treatment plan itself, or whether the studies potentially did not have a large enough sample size (as may be the case for 8-patient study by Kida et al 29) to capture potential ventilation-perfusion differences. The importance of our study is that it used the same modality to isolate physiological differences between ventilation and perfusion for lung cancer patients. The data from the current work will allow for future studies to de-couple and isolate dose-function differences due to physiological differences, differences due to using various imaging modalities, or differences due to treatment plan generation techniques.

The correlation of SPECT ventilation and perfusion was a research question specific to this study and was not defined as part of the protocol for the prospective clinical trial. Therefore, this study was limited only to nineteen patients who underwent pre-treatment SPECT-CT imaging. This sample size may not be optimal for comparison of a wide range of perfusion-ventilation SPECT scenarios. The current analysis is prone to uncertainties due to artifacts generated by deposition of the Technetium isotope in airways (ventilation) and blood vessels (perfusion). The strategy for the current study was to mitigate the impact of the artifacts by manually removing them during segmentation and employing normalization strategies. Future studies employing either numerical or imaging-based methods to reduce the deposition artifacts could further improve the physiological comparison between ventilation and perfusion differences5. As was done in previous functional imaging comparison studies25, the comparison metrics in our work included Spearmen correlation coefficients, DSC.

Conclusions:

The current work performed a spatial comparison of SPECT-based ventilation and perfusion for 19 scans (38 individual lungs) acquired for nineteen patients with stage III lung cancer. The quantitative results demonstrated that about 74%-79% of scans had ventilation-perfusion correlations ≥0.5. In other words, in about 21%-26% of patients both ventilation and perfusion would be needed to provide complete (ventilation and perfusion) lung function information. Within the context of 4DCT ventilation functional avoidance, the results of the current investigation suggest that both ventilation and perfusion may be necessary to achieve complete lung function information. However, future studies may be beneficial to investigate the methodologies that identify patients for whom ventilation provides similar functional information to perfusion.

Supplementary Material

1

Acknowledgements:

This work was funded by NIH Grant R01CA236857.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References:

  • 1.Marks LB, Spencer DP, Bentel GC, et al. The utility of SPECT lung perfusion scans in minimizing and assessing the physiologic consequences of thoracic irradiation. Int J Radiat Oncol Biol Phys. 1993;26(4):659–668. [DOI] [PubMed] [Google Scholar]
  • 2.Vinogradskiy Y, Castillo R, Castillo E, et al. Use of 4-Dimensional Computed Tomography-Based Ventilation Imaging to Correlate Lung Dose and Function With Clinical Outcomes. International Journal of Radiation Oncology Biology Physics. 2013;86(2):366–371. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Guerrero T, Sanders K, Noyola-Martinez J, et al. Quantification of regional ventilation from treatment planning CT. International Journal of Radiation Oncology* Biology* Physics. 2005;62(3):630–634. [DOI] [PubMed] [Google Scholar]
  • 4.Reinhardt JM, Ding K, Cao K, Christensen GE, Hoffman EA, Bodas SV. Registration-based estimates of local lung tissue expansion compared to xenon CT measures of specific ventilation. Medical Image Analysis. 2008;12(6):752–763. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Castillo R, Castillo E, McCurdy M, et al. Spatial correspondence of 4D CT ventilation and SPECT pulmonary perfusion defects in patients with malignant airway stenosis. Physics in Medicine and Biology. 2012;57(7):1855–1871. [DOI] [PubMed] [Google Scholar]
  • 6.Vinogradskiy Y, Koo PJ, Castillo R, et al. Comparison of 4-Dimensional Computed Tomography Ventilation With Nuclear Medicine Ventilation-Perfusion Imaging: A Clinical Validation Study. International Journal of Radiation Oncology Biology Physics. 2014;89(1):199–205. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Yamamoto T, Kabus S, Lorenz C, et al. Pulmonary Ventilation Imaging Based on 4-Dimensional Computed Tomography: Comparison With Pulmonary Function Tests and SPECT Ventilation Images. International Journal of Radiation Oncology* Biology* Physics. 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Mathew L, Wheatley A, Castillo R, et al. Hyperpolarized He-3 Magnetic Resonance Imaging: Comparison with Four-dimensional X-ray Computed Tomography Imaging in Lung Cancer. Academic Radiology. 2012;19(12):1546–1553. [DOI] [PubMed] [Google Scholar]
  • 9.Kipritidis J, Siva S, Hofman MS, Callahan J, Hicks RJ, Keall PJ. Validating and improving CT ventilation imaging by correlating with ventilation 4D-PET/CT using 68Ga-labeled nanoparticles. Medical physics. 2014;41(1):011910. [DOI] [PubMed] [Google Scholar]
  • 10.Faught AM, Miyasaka Y, Kadoya N, et al. Evaluating the toxicity reduction with CT-ventilation functional avoidance radiotherapy. International Journal of Radiation Oncology* Biology* Physics. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Faught AM, Yamamoto T, Castillo R, et al. Evaluating which dose-function metrics are most critical for functional-guided radiotherapy with CT ventilation imaging. International Journal of Radiation Oncology* Biology* Physics. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Patton TJ, Gerard SE, Shao W, Christensen GE, Reinhardt JM, Bayouth JE. Quantifying ventilation change due to radiation therapy using 4 DCT Jacobian calculations. Medical physics. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Yamamoto T, Kabus S, Bal M, Keall P, Benedict S, Daly M. The first patient treatment of computed tomography ventilation functional image-guided radiotherapy for lung cancer. Radiotherapy and Oncology. 2015. [DOI] [PubMed] [Google Scholar]
  • 14.Vinogradskiy Y, Rusthoven CG, Schubert L, et al. Interim Analysis of a Two-Institution, Prospective Clinical Trial of 4DCT-Ventilation-based Functional Avoidance Radiation Therapy. Int J Radiat Oncol Biol Phys. 2018;102(4):1357–1365. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Wernly JA, DeMeester TR, Kirchner PT, Myerowitz PD, Oxford DE, Golomb HM. Clinical value of quantitative ventilation-perfusion lung scans in the surgical management of bronchogenic carcinoma. J Thorac Cardiovasc Surg. 1980;80(4):535–543. [PubMed] [Google Scholar]
  • 16.Ohno Y, Koyama H, Takenaka D, et al. Coregistered Ventilation and Perfusion SPECT Using Krypton-81m and Tc-99m— Labeled Macroaggregated Albumin With Multislice CT: Utility for Prediction of Postoperative Lung Function in Non— Small Cell Lung Cancer Patients. Academic radiology. 2007;14(7):830–838. [DOI] [PubMed] [Google Scholar]
  • 17.Win T, Tasker AD, Groves AM, et al. Ventilation-perfusion scintigraphy to predict postoperative pulmonary function in lung cancer patients undergoing pneumonectomy. American Journal of Roentgenology. 2006;187(5):1260–1265. [DOI] [PubMed] [Google Scholar]
  • 18.Jackson MW, Schubert LK, Jones B, et al. A Novel Lung Function Imaging Modality for Surgical Lung Cancer Evaluation. International journal of radiation oncology, biology, physics. 2016;96(2S):S46. [Google Scholar]
  • 19.Brunelli A, Kim AW, Berger KI, Addrizzo-Harris DJ. Physiologic evaluation of the patient with lung cancer being considered for resectional surgery: diagnosis and management of lung cancer: American College of Chest Physicians evidence-based clinical practice guidelines. CHEST Journal. 2013;143(5_suppl):e166S–e190S. [DOI] [PubMed] [Google Scholar]
  • 20.King GG, Harris B, Mahadev S. V/Q SPECT: utility for investigation of pulmonary physiology. Semin Nucl Med. 2010;40(6):467–473. [DOI] [PubMed] [Google Scholar]
  • 21.Carvalho P, Lavender JP. The incidence and etiology of the ventilation/perfusion reverse mismatch defect. Clin Nucl Med. 1989;14(8):571–576. [DOI] [PubMed] [Google Scholar]
  • 22.Petersson J, Sánchez-Crespo A, Larsson SA, Mure M. Physiological imaging of the lung: single-photon-emission computed tomography (SPECT). J Appl Physiol (1985). 2007;102(1):468–476. [DOI] [PubMed] [Google Scholar]
  • 23.Coghe J, Votion D, Lekeux P. Comparison between radioactive aerosol, technegas and krypton for ventilation imaging in healthy calves. Vet J. 2000;160(1):25–32. [DOI] [PubMed] [Google Scholar]
  • 24.Pretorius PH, King MA. Diminishing the impact of the partial volume effect in cardiac SPECT perfusion imaging. Med Phys. 2009;36(1):105–115. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Kipritidis J, Tahir BA, Cazoulat G, et al. The VAMPIRE challenge: A multi-institutional validation study of CT ventilation imaging. Medical physics. 2019;46(3):1198–1217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Vinogradskiy YY, Castillo R, Castillo E, Chandler A, Martel MK, Guerrero T. Use of weekly 4DCT-based ventilation maps to quantify changes in lung function for patients undergoing radiation therapy. Medical Physics. 2012;39(1):289–298. [DOI] [PubMed] [Google Scholar]
  • 27.Lee E, Zeng J, Miyaoka RS, et al. Functional lung avoidance and response-adaptive escalation (FLARE) RT: Multimodality plan dosimetry of a precision radiation oncology strategy. Medical physics 2017;44(7):3418–3429. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Nakajima Y, Kadoya N, Kimura T, Hioki K, Jingu K, Yamamoto T. Variations between dose-ventilation and dose-perfusion metrics in radiotherapy planning for lung cancer. Advances in Radiation Oncology. 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Kida S, Bal M, Kabus S, et al. CT ventilation functional image-based IMRT treatment plans are comparable to SPECT ventilation functional image-based plans. Radiotherapy and Oncology. 2016;118(3):521–527. [DOI] [PubMed] [Google Scholar]
  • 30.Du K, Reinhardt JM, Christensen GE, Ding K, Bayouth JE. Respiratory effort correction strategies to improve the reproducibility of lung expansion measurements. Med Phys. 2013;40(12):123504. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1

RESOURCES