Skip to main content
Materials logoLink to Materials
. 2021 Sep 29;14(19):5683. doi: 10.3390/ma14195683

Particle Size Measurement Using Dynamic Light Scattering at Ultra-Low Concentration Accounting for Particle Number Fluctuations

Mengjie Wang 1, Jin Shen 1,*, John C Thomas 1,2, Tongtong Mu 1, Wei Liu 1, Yajing Wang 1, Jinfeng Pan 1, Qin Wang 1, Kaishi Liu 1
Editors: E Andrew Payzant, Paweł Głuchowski, Robert Tomala
PMCID: PMC8510349  PMID: 34640079

Abstract

Dynamic light scattering (DLS) is a popular method of particle size measurement, but at ultra-low particle concentrations, the occurrence of number concentration fluctuations limits the use of the technique. Number fluctuations add a non-Gaussian term to the scattered light intensity autocorrelation function (ACF). This leads to an inaccurate particle size distribution (PSD) being recovered if the normal DLS analysis model is used. We propose two methods for inverting the DLS data and recovering the PSDs when number fluctuations are apparent. One is to directly establish the relationship between the non-Gaussian ACF and the PSD by the kernel function reconstruction (KFR) method while including the non-Gaussian term to recover the PSD. The other is to remove the effect of the non-Gaussian term in the ACF by the baseline reset (BR) method. By including the number fluctuation term, the ideal recovered PSD can be obtained from the simulated data, but this will not happen in the experimental measurement data. This is because the measured intensity ACF contains more noise than the simulated ACF at ultra-low concentration. In particular, the baseline noise at the tail of long delay time of ACF overwhelms the number fluctuation term, making it difficult to recover reliable PSD data. Resetting the baseline can effectively remove the digital fluctuation term in ACF, which is also a feasible method to improve PSD recovery under ultra-low concentration. However, increasing noise at ultra-low concentrations can lead to errors in determining an effective baseline. This greatly reduces the accuracy of inversion results. Results from simulated and measured ACF data show that, for both methods, noise on the ACF limits reliable PSD recovery.

Keywords: dynamic light scattering, ultra-low particle concentration, kernel function reconstruction, baseline reset

1. Introduction

Dynamic light scattering (DLS) was pioneered in the 1960s [1,2,3] and is now also commonly called photon correlation spectroscopy (PCS) because the experiment is done using digital correlation of photon count signals [4,5]. This technique has been widely applied to measure particle size in many fields such as materials, chemical industry, food, biology and medicine [6,7,8,9,10]. It is an effective technology [11] and a standard method for measuring the particle size and distribution of submicron particles in suspension [12,13,14,15]. Particle size and particle size distribution (PSD) determination essentially occurs by measuring the diffusion coefficients of the particles undergoing Brownian motion in a small scattering volume. This is done by analyzing the timescale of the fluctuations in the intensity of the light scattered from the particles. The particle concentration is high enough that the number of particles in the scattering volume is essentially constant. For an ultra-low concentration particle system, there are less than ~50 diffusively moving particles in the scattering volume, and particle diffusion in and out of the volume may result in substantial changes in the relative local particle concentration, which will also give rise to changes in the scattered light intensity. Thus, intensity fluctuations will arise now from both the Brownian motion of the particles and the fluctuations in the number concentration of particles [16,17,18,19]. These fluctuation components will occur on different timescales and with different amplitudes. Willemse et al. proposed a method where the output signals of the photon detectors are divided by high and low frequencies by using the frequency difference between signals caused by Brownian motion and fluctuation of particle number. They removed the number fluctuation noise located in the low frequency band from signals and carried out a series of studies [20,21,22,23,24]. Although this method is feasible in theory, additional devices are needed between detector and correlator in the DLS measurement system and additional noise will be introduced. Furthermore, it changes the real-time signal acquisition and measurement nature of DLS as the signal represented by the pulse density is low-pass filtered before entering the photon correlator, which makes DLS measurement at low concentration different to what is normally done. This method is difficult to implement and operate in practice and has not been realized in a commercial particle size analyzer.

The ideal DLS measurement at ultra-low concentration is to obtain the PSD information by analyzing the light intensity autocorrelation function (ACF) data calculated in real-time by the correlator without changing the real-time operation mechanism of DLS. For ultra-low concentration samples, the amplitude change of the scattered field is no longer a Gaussian random process due to the small number of particles in the scattering volume [25]. In this case, the light intensity ACF and electric field ACF are no longer related by the Siegert relation. However, an additional relaxation term can be added to the relation to account for number fluctuations. If provision is not made for this term, the PSD results will be meaningless. For this reason, it is clearly stipulated in the corresponding international standards [14] that the number of particles in the measured volume must be kept constant. If there are large particles or the sample concentration is too low in the scattering volume, this situation may contribute an artifact to the PSD.

Recently, interest in the measurement of ultra-low concentration samples has increased [26,27,28]. Earlier studies suggest that there are two approaches to obtain PSD information of ultra-low concentration samples. One is to construct an equivalent electric field ACF model using the modified Siegert relation, and reconstruct the kernel function of the inversion equation, and then obtain the PSD. The second is to effectively separate the contribution of particle number fluctuation in the measured intensity ACF, and then recover the PSD through the usual data processing method. In this paper we investigate the two ways to carry out DLS measurements at ultra-low concentrations through theoretical analysis, numerical simulation and experimental verification. In addition, for DLS measurement at ultra-low concentration, the key inhibiting factors were analyzed and feasible ways to improve the accuracy were discussed.

2. Theory and Methods

2.1. Classical DLS Theory and Number Fluctuations in the Scattering Volume

Light scattered from a suspension of Brownian particles exhibits intensity fluctuations [15]. The time scale of the fluctuations is related to the diffusion coefficient of the Brownian particles, which is related to the size of the particles. These fluctuations can be characterized by the light intensity ACF [15]

G(2)(τ)=i(t)i(t+τ)=limT00T0i(t)i(t+τ)dt/T0. (1)

Here, i(t) is the scattered light intensity, T0 and τ are measurement time and delay time, respectively. The intensity ACF can be obtained by photon correlation. For a Gaussian light field, the normalized light intensity ACF, g(2)(τ), and the normalized electric field ACF, g(1)(τ), satisfy the Siegert relation

g(2)(τ)=1+β|g(1)(τ)|2, (2)

where β is an instrumental coherence factor. For Brownian particles and, in the absence of number fluctuations, the electric field ACF can be expressed as

g(1)(τ)=exp(Γτ). (3)

Here, Γ=DTq2 is the decay constant. DT is the diffusion coefficient and is related to particle diameter d and medium viscosity η by the Stokes–Einstein relation DT=kBT3πηd. Here, kB is the Boltzmann constant, and T is the absolute temperature of the scattering medium. q is the scattering vector, whose amplitude is |q|=4πnλ0sinθ2. λ0, n, and θ are wavelength of incident light in vacuum, refractive index, and scattering angle of solution, respectively.

For polydisperse particle systems, the relation between particle size and the normalized electric field ACF is [15,29]

g(1)(τ)=0G(Γ)exp(Γτ)dΓ, (4)

where G(Γ) is normalized intensity distribution function of Γ. Using the Stokes–Einstein relation and substituting in the various parameters allows Equation (4) to be written in discrete form as,

g(1)(τj)=i=1Mexp(16πn2kBT3ηλ02disin2(θ2)τj)f(di). (5)

Here, g(1)(τj) is the normalized electric field ACF, j(1jCh) is the correlator channel number. f(di) is the discrete PSD and satisfies i=1Mf(di)=1. di(i=1,2,,M) is the particle diameter. Equation (5) can be simplified as

g=Af. (6)

Here, g is the vector composed of the normalized electric field ACF data, whose elements are g(1)(τj) and dimension is Ch×1. The elements of f, a vector made up of discrete PSDs, are f(di). The dimension of f is M×1 and A is the kernel matrix corresponding to the electric field ACF data, whose dimension is Ch×M. The elements of A are A(j,i)=exp(16πn2kBT3ηλ02disin2(θ2)τj). Equation (6) is a Fredholm integral equation of the first kind, and is a typical ill-posed problem that can be solved by the Tikhonov regularization method. The ill-posed problem is transformed into a well-posed problem, and the approximate solution closest to the real solution of Equation (6) is obtained by optimizing the objective function,

Mα(f,g)=Af - g22+αLf22,s.t.f0. (7)

Here, α, L, 22, and Lf22 are regularization parameter, regularization matrix, Euclidean norm, and penalty factor, respectively.

At ultra-low concentrations, when number fluctuations are significant, the scattered field amplitude is no longer Gaussian, and the intensity ACF and field ACF no longer satisfy the usual Siegert relation. Under this condition, the relation between the intensity ACF and the field ACF can be expressed as [30]

gd(2)(τ)=1+β|g(1)(τ)|2+γN(1+4DTτω02)1(1+4DTτa2)12, (8)

where γ=23/2 is a constant related to the effective scattering volume [30]. N is the average number of particles in the scattering volume, ω0 is the waist radius of the laser beam, and a is the radius of the detector aperture. In Equation (8), the second term on the right represents the contribution of Brownian motion to the ACF, and the third term is the number fluctuation term. The latter is usually referred to as the non-Gaussian term. For N50 particles in the scattering volume, the term is negligible and can be ignored for routine DLS measurements. Comparing Equations (8) and (2), it can be seen that the number fluctuation term also contains PSD information and that this term becomes larger as the number concentration decreases making it impossible to obtain the field ACF using the usual Siegert relation at very low particle concentrations. Also, the wrong PSD will be recovered by using the kernel function matrix in Equation (6). Thus, the analysis of DLS is not straight forward at ultra-low concentration.

2.2. PSD Recovery from DLS Measurements at Ultra-Low Concentrations

2.2.1. Kernel Function Reconstruction of Inversion Equation

Equation (8) shows that both Brownian motion and particle number fluctuations potentially provide particle size information. If we can construct the field ACF to include Brownian motion and number fluctuations and reconstruct the corresponding kernel function matrix on the basis of Equation (8), we establish the inversion equation relationship between the equivalent field ACF and the PSD. The PSD information would be obtained from both Brownian motion and number fluctuations.

In discrete form, Equation (8) can be written as

gd(2)(τj)=1+βexp(32πn2kBT3ηλ02dsin2(θ2)τj)+γN(1+4kBTτj3πηdω02)1(1+4kBTτj3πηda2)12. (9)

For a polydisperse particle system, the intensity ACF can be expressed as

gd(2)(τj,di)=1+βi=1M(exp(32πn2kBT3ηλ02disin2(θ2)τj)+γβN(1+4kBTτj3πηdiω02)1(1+4kBTτj3πηdia2)12)f(di). (10)

If we define

A1(τj,di)=exp(32πn2kBT3ηλ02disin2(θ2)τj), (11)
e(τj,di)=γβN(1+4kBTτj3πηdiω02)1(1+4kBTτj3πηdia2)12, (12)

and by analogy with the Siegert equation we can write an equivalent field ACF as

g1(τj)=i=1MA1(τj,di)+e(τj,di)f(di), (13)

The inversion equation in the form of Equation (6) is obtained, and the reconstructed kernel function matrix is expressed as

Ar=A1+e, (14)

and its elements are

Ar(j,i)=exp(32πn2kBT3ηλ02disin2(θ2)τj)+γβN(1+4kBTτj3πηdiω02)1(1+4kBTτj3πηdia2)12. (15)

The PSD can be obtained from Equation (7).

2.2.2. Intensity ACF Baseline Reset

Comparing Equation (2) with Equation (8), it can be seen that the last term on the right of Equation (8), the number fluctuation term, is not present for DLS measurements at higher concentrations. The number fluctuation term has an amplitude that decreases with concentration and a long decay time.

We define the number fluctuation term in the intensity ACF as

g(τ)=γN(1+4DTτω02)1(1+4DTτa2)12. (16)

Figure 1 shows the theoretical intensity ACFs, gd(2)(τ), and number fluctuation terms, g(τ), for 10 nm and 1000 nm particles at average particle number N = 5, 15, and 50 in the scattering volume.

Figure 1.

Figure 1

Theoretical (a) intensity ACFs gd(2)(τ) and (b) number fluctuation terms g(τ) for 10 nm particles, (c) intensity ACFs gd(2)(τ) and (d) number fluctuation terms g(τ) for 1000 nm particles at average particle number N = 5, 15, and 50 in the scattering volume.

Figure 1 shows that the number fluctuation and Brownian motion terms contribute significantly different attenuation processes to the intensity ACF. This can be further seen in Table 1, which shows the delay time, td, for the number fluctuation term to decrease by 10−3 and its decrease in tc, the Brownian motion decay time. It is observed that the largest magnitude variation in g(τ) over the time tc, is only on the order of 10−8 for both the 10 and 1000 nm particles when N=5. In other words, on the timescale tc, number fluctuations essentially contribute a small constant term to the ACF. This suggests that we could obtain a PSD by removing the number fluctuation term by resetting the baseline of gd(2)(τ) from 1.0 and then calculate g(1)(τ) using the usual Siegert relation.

Table 1.

The decrease in magnitude, Δg, of the number fluctuation term, g(τ), for 10 and 1000 nm particles in the Brownian motion decay time, tc and the delay time, td, at which the intensity ACFs, gd(2)(τ), decay to 0.001 for 10 and 1000 nm particles at N = 5, 15, and 50 particles in the scattering volume.

Particle Size (nm) <N> Δg tc (s) td (s)
10 5 1.0 × 10−8 3.0 × 10−5 1.0 × 10−4
15 3.3 × 10−9 3.0 × 10−5 5.0 × 10−3
50 1.0 × 10−9 3.0 × 10−5 2.0 × 10−3
1000 5 1.6 × 10−8 4.0 × 10−3 1.0 × 10−6
15 5.2 × 10−9 4.0 × 10−3 4.0 × 10−5
50 1.6 × 10−9 4.0 × 10−3 2.0 × 10−5

To determine a new, adjusted baseline, we can examine the change in adjacent channels of gd(2)(τ) and find the truncation point where g(2)(τ) at τtc. Figure 2 shows the difference between adjacent channels, gd(2)(τj+1)gd(2)(τj), with the delay time of g(2)(τ). At τj=τs, when gd(2)(τj+1)gd(2)(τj)103, gd(2)(τ) has decayed to g(τ), the effective baseline term due to number fluctuations. The new ACFs, g(2)(τ), with new baseline gd(2)(τs) are obtained as shown in Figure 3. Then, g(1)(τ) is calculated by the usual Siegert relation and the PSD is obtained by Equation (7). In the actual DLS measurement, four consecutive delay channels are required to satisfy gd(2)(τj+1)gd(2)(τj)103 in order to minimize the effect of noise on the ACF in choosing the correct baseline point.

Figure 2.

Figure 2

The truncation point τs determined by the difference between gd(2)(τj+1) and gd(2)(τj).

Figure 3.

Figure 3

The new baseline gd(2)(τs) of gd(2)(τ) determined by τs.

3. Results

3.1. Simulation

The usual method, the kernel function reconstruction (KFR) method, and the baseline reset (BR) method were compared using eight samples of simulated data consisting of two unimodal PSDs at ultra-low concentration. To simulate real conditions, noise at levels of 0, 10−4, 10−3, 10−2, and 10−1 was added to the ACF data. In analyzing the ACF data to reconstruct the PSDs, both the usual and the BR methods used the A1 matrix (Equation (11)) and the KFR method used the Ar matrix (Equation (15)).

Johnson’s SB function [31] was used to simulate PSDs as follows:

f(d)=σ(dmaxdmin)2π[t(1t)]1exp[0.5[μ+σln(t1t)]2]. (17)

Here, t is the normalized particle size satisfying t=ddmindmaxdmin, where dmax and dmin denote maximum and minimum particle size, respectively. σ and μ are distribution parameters.

The intensity ACF was obtained using Equation (10). Gaussian random noise was

Gnoise(2)(τ)=G(2)(τ)+δn(τ). (18)

Here, Gnoise(2)(τ) is the noisy intensity ACF, δ is the noise standard deviation, and n(τ) denotes Gaussian random noise.

The ACF data were simulated with λ0 = 532 nm, ω0 = 54 μm, a = 200 μm, T = 298.15 K, n = 1.334, θ = 90°, kB = 1.3807 × 10−23 J/K, η = 0.89 × 10−3 cP, and β = 0.9. A discrete PSD with M = 150 data points was used. Four concentration values were used, N = 6, 12, 24, and 48.

Table 2 summarizes the properties of the simulated PSDs. P is the peak position of the PSD. To characterize the accuracy of PSD recovery, we introduce two indices: the relative error of peak position (EP) and the PSD recovery error (VE). These two indices are defined as follows:

EP=|PtruePmeas|Ptrue, (19)
VE=(1M[ftrue(di)fmeas(di)]2M)1/2, (20)

where ftrue(di) is the true PSD; fmeas(di) is the recovered PSD.

Table 2.

Parameters and properties of the simulated PSDs.

P (nm) μ σ (dmin, dmax) (nm)
156 0.50 9.0 (2, 320)
696 0.17 5.0 (400, 1000)

For the simulated 156 nm PSD, the PSDs recovered by the usual method, the KFR method and the BR method are shown in Figure 4, Figure 5 and Figure 6, respectively. Table 3 shows the performance parameter values of the true and recovered PSDs. PU, EPU, and VEU are the parameters of PSD obtained by the usual method; PKFR, EPKFR, and VEKFR are the parameters of PSD obtained by the KFR method; PBR, EPBR, and VEBR are the parameters of PSD obtained by the BR method. Figure 4 shows that the PSDs obtained by the usual method have false peaks much larger than the real peaks for all values of N investigated. The significant deviation of the main peak from the actual position is caused by the large false peak. Also, with increasing noise level, the PSDs obtained by the usual method deviate more from the true PSDs. In Figure 5, an ideal inversion result is obtained by the KFR method at zero noise level. It can be seen from Table 3 that the relative error (EPKFR) and the recovery error (VEKFR) are only 0.01 and 0.002, even when the concentration is as low as N=6. At concentration N=24, the KFR method gives a close estimation to the true PSD. Figure 5 and Table 3 also show that the inversion results of the KFR method are not significantly affected by noise. At low noise level (10−3), even if the sample concentration is extremely low (N=6), EPKFR increases considerably less than EPU and VEKFR is limited to <0.01. As the noise level increases to 10−2, the EPKFR and VEKFR show a clear increase. In Table 3, the performance parameter values of the recovered PSDs with the BR method are clearly worse than with the KFR method at the same sample concentrations and noise levels. When the noise level is 0, the relative error (EPBR) of the BR method is twice that of the KFR method and the recovery error (VEBR) can be up to four times that of the KFR method at very low concentrations. Moreover, with increasing noise, the differences of EP and VE between the two methods tends to increase. It can also be seen that, with increasing noise, the BR method moves the peak to smaller sizes and broadens the PSD.

Figure 4.

Figure 4

PSD recovered by the usual method for a 156 nm unimodal sample at number concentrations of <N> = 6, 12, 24, 48, and different levels of noise. (a) 0, (b) 10−4, (c) 10−3, (d) 10−2, and (e) 10−1 noise levels.

Figure 5.

Figure 5

Figure 5

PSD recovered by the KFR method for a 156 nm unimodal sample at number concentrations of <N> = 6, 12, 24, 48, and different levels of noise. (a) 0, (b) 10−4, (c) 10−3, (d) 10−2, and (e) 10−1 noise levels.

Figure 6.

Figure 6

Figure 6

PSD recovered by the BR method for a 156 nm unimodal sample at number concentrations of <N> = 6, 12, 24, 48, and different levels of noise. (a) 0, (b) 10−4, (c) 10−3, (d) 10−2, and (e) 10−1 noise levels.

Table 3.

Performance parameter values for the recovery of the 156 nm PSD by using three methods.

δ <N> PU (nm) PKFR (nm) PBR (nm) E PU E PKFR E PBR V EU V EKFR V EBR
True 156 156 156 0 0 0 0 0 0
0 6 68/- 154 152 0.6 0.01 0.03 0.03 0.002 0.009
12 88/- 154 152 0.4 0.01 0.03 0.03 0.003 0.009
24 101/- 156 152 0.4 0 0.03 0.03 0.003 0.009
48 115/- 156 152 0.3 0 0.03 0.02 0.005 0.009
1 × 10−4 6 55/- 154 152 0.6 0.01 0.03 0.03 0.006 0.009
12 74/- 154 152 0.5 0.01 0.03 0.03 0.004 0.009
24 91/- 154 152 0.4 0.01 0.03 0.03 0.004 0.009
48 105/- 154 152 0.3 0.01 0.03 0.02 0.005 0.009
1 × 10−3 6 55/- 154 150 0.6 0.01 0.04 0.07 0.01 0.01
12 76/- 154 150 0.5 0.01 0.04 0.06 0.006 0.01
24 95/- 156 154 0.4 0 0.01 0.05 0.003 0.01
48 103/- 156 154 0.3 0 0.01 0.05 0.005 0.01
1 × 10−2 6 61/- 173 141 0.6 0.1 0.1 0.03 0.02 0.03
12 84/- 158 78 0.5 0.01 0.5 0.03 0.006 0.04
24 120/- 180 122 0.2 0.2 0.2 0.03 0.03 0.04
48 114/- 163 71 0.3 0.05 0.5 0.03 0.01 0.04
1 × 10−1 6 -/- 14/190 -/150/- - - 0.04 0.04 0.03 0.08
12 -/99 4/190 -/178 0.4 - 0.1 0.04 0.04 0.07
24 103/- 174/- 173 0.3 0.1 0.1 0.03 0.03 0.03
48 108/- 155/- 161 0.3 0.006 0.03 0.03 0.007 0.01

For the simulated 696 nm PSD, the PSDs recovered by the usual method, the KFR method and the BR method are shown in Figure 7, Figure 8 and Figure 9, respectively. Corresponding performance indices are shown in Table 4. At different noise levels, it is clear that false peaks obtained by the usual method exceed the upper limit of the PSD range at ultra-low concentrations. The appearance of false peaks causes the main peak to decrease and shift towards smaller particle sizes, and gradually move out of the lower limit of the PSD range as the concentration decreases. However, the KFR method gives better results. At 10−3 noise level, the relative error (EPKFR) and the recovery error (VEKFR) of the KFR method are as low as 0.006 and 0.01, respectively. When the noise level increased to 10−1, the performance indices deteriorated significantly. The BR method results are similar to those of the KFR method, but are slightly less affected by noise increase.

Figure 7.

Figure 7

PSD recovered by the usual method for a 696 nm unimodal sample at number concentrations of <N> = 6, 12, 24, 48, and different levels of noise. (a) 0, (b) 10−4, (c) 10−3, (d) 10−2, and (e) 10−1 noise levels.

Figure 8.

Figure 8

Figure 8

PSD recovered by the KFR method for a 696 nm unimodal sample at number concentrations of <N> = 6, 12, 24, 48, and different levels of noise. (a) 0, (b) 10−4, (c) 10−3, (d) 10−2, and (e) 10−1 noise levels.

Figure 9.

Figure 9

Figure 9

PSD recovered by the BR method for a 696 nm unimodal sample at number concentrations of <N> = 6, 12, 24, 48, and different levels of noise. (a) 0, (b) 10−4, (c) 10−3, (d) 10−2, and (e) 10−1 noise levels.

Table 4.

Performance indices for the recovery of the 696 nm PSD by using three methods.

δ <N> PU (nm) PKFR (nm) PBR (nm) E PU E PKFR E PBR V EU V EKFR V EBR
True 696 696 696 0 0 0 0 0 0
0 6 - 696 692 - 0 0.006 0.07 0.005 0.007
12 - 692 692 - 0.006 0.006 0.05 0.003 0.006
24 - 692 692 - 0.006 0.006 0.04 0.002 0.006
48 420/- 692 692 0.4 0.006 0.006 0.03 0.003 0.006
1 × 10−4 6 - 696 688 - 0 0.01 0.07 0.005 0.009
12 - 692 692 - 0.006 0.006 0.05 0.003 0.006
24 - 692 692 - 0.006 0.006 0.04 0.002 0.006
48 424/- 692 692 0.4 0.006 0.006 0.03 0.003 0.006
1 × 10−3 6 - 696 688 - 0 0.01 0.07 0.006 0.01
12 - 692 692 - 0.006 0.006 0.05 0.002 0.006
24 - 692 688 - 0.006 0.01 0.04 0.003 0.009
48 424/- 692 692 0.4 0.006 0.006 0.03 0.003 0.009
1 × 10−2 6 - 704 608/- - 0.01 0.1 0.07 0.004 0.02
12 - 704 -/728 - 0.01 0.05 0.05 0.005 0.01
24 - 696 688 - 0 0.01 0.04 0.003 0.003
48 - 708 -/740 - 0.02 0.06 0.03 0.006 0.009
1 × 10−1 6 - 840 744 - 0.2 0.07 0.08 0.03 0.03
12 408/- 756 644 0.4 0.09 0.08 0.07 0.03 0.04
24 500/- 868 776 0.3 0.2 0.1 0.06 0.03 0.03
48 -/- -/908 772 - 0.3 0.1 0.06 0.02 0.03

3.2. Experiment

Real data were obtained from an experimental setup including a solid-state laser with a wavelength of 532 nm (model MGL-III-532 nm–15 mW), a photon counter (model CH326, Hamamatsu Photonics), and a 512 channels digital correlator. The samples were dilute suspensions of latex spheres in distilled water. They were made from 152 ± 5 nm (Duke 3150A) and 693 ± 10 nm (GBW(E)120087) standard polystyrene latex spheres. The laser power was 15 mW, the scattering angle was 90°, the focal length of the incident lens was 175 mm, the receiving aperture of the detector was 400 μm, and the sample cell temperature was 298.15 K. The PSDs recovered for the 152 and 693 nm diameter spheres are shown in Figure 10 and the corresponding performance indices are shown in Table 5.

Figure 10.

Figure 10

PSDs recovered for standard polystyrene latex spheres at number concentrations of <N> = 6, 12, 24, 48: 152 nm diameter sample; (a) the usual method, (b) the KFR method, and (c) the BR method; and the PSDs recovered for the 693 nm diameter sample; (d) the usual method, (e) the KFR method, and (f) the BR method.

Table 5.

Performance indices for the recovery of 152 and 693 nm PSDs.

P (nm) <N> PU (nm) PKFR (nm) PBR (nm) E PU E PKFR E PBR
152 6 40/- 220 180 0.7 0.4 0.2
12 42/- 239 173 0.7 0.6 0.1
24 114/- 220 169 0.3 0.4 0.1
48 135/- 184 156 0.1 0.2 0.03
693 6 - 820 784 - 0.2 0.1
12 - 876 764 - 0.3 0.1
24 - 908 768 - 0.3 0.1
48 - 864 760 - 0.3 0.1

As is shown in Figure 10 and Table 5, the PSDs recovered by the KFR method and the BR method are better than those by the usual method. Strong false peaks in the PSDs recovered by the usual method did not appear in the PSDs recovered by the KFR method and BR method. At all concentrations, the relative errors of peak positions obtained by BR method are smaller than those obtained by KFR method, which means the BR method is better than the KFR method. For the 152 nm particles, the appearance of strong false peaks causes the main peak reflecting the true particle size to nearly disappear, and the peak position of the main peak shifts significantly to small sizes as the concentration decreases. For the 693 nm particles, the PSDs recovered by the usual method showed the main peak reflecting the true particle size in the PSD results has completely disappeared and only the false peaks remain. The experimental results from the KFR method are different from those with the simulated data. Although there are no more false peaks, the recovered peak positions are obviously shifted from the true positions. These PSDs with large peak position errors meaningless. For the BR method, the relative errors between the experimental PSDs and the recovered PSDs are closer to the simulations at high noise level. Compared with the recovered PSDs of simulations at low-noise conditions, the relative errors of the experimental PSDs using the BR method show a significant increase.

4. Discussion

The results from the simulated data show obvious false peaks in the PSDs obtained by the usual method at ultra-low concentrations and the peak positions exceed the upper limits of the PSD range. Since the normalized PSD satisfies i=1Mf(di)=1, large false peaks that exceed the upper limits of the PSD range are compensated by peaks at small sizes and accurate PSDs are not obtained. The KFR method, using a theoretical model that includes number fluctuations, should account for the number fluctuation term in the intensity ACF and produce reliable PSDs. This did occur with the simulated data, especially with low added noise. However, this was not the case with measured data. The KFR method did not produce the false peaks in the large particle size region, but it did not obtain peak positions that were consistent with or even close to those in the true PSDs. Figure 11 shows the measured intensity ACFs and it is clear that they contain more noise than the simulated ACFs, especially at long delay times. This noise is greater than normally found in ACFs from DLS due to the very low sample concentrations involved. The scattered light signal is weaker and the background noise is relatively larger due to the lower concentrations. Furthermore, at long delay times, the noise is more apparent in the intensity ACF. This region is essentially the ACF baseline at normal particle concentrations and it is usually dealt with using truncation or weighting methods [32,33]. However, this segment is located in the number fluctuation region of the ACFs at ultra-low concentration. If the data is noisy, it will be difficult for any model to recover the number fluctuation term successfully. Figure 12 shows the intensity ACFs for 152 and 693 nm particles with <N> = 6, 12, 24, and 48 calculated from Equation (8). In Figure 11 and Figure 12, the vertical dashed line marks the delay time τ0, when the ACF decays to 10−3 in the long delay time region. The τ0 of the measured and simulated ACFs are 0.75 s, 1.5 s for the 152 nm particles, and 1400 s, 3000 s for the 693 nm particles, respectively.

Figure 11.

Figure 11

Measured ACFs of standard polystyrene latex particles for concentrations of <N> = 6, 12, 24, and 48: (a) 152 nm diameter and (b) 693 nm diameter.

Figure 12.

Figure 12

Simulated ACFs of particles for concentrations of <N> = 6, 12, 24, and 48: (a) 152 nm diameter and (b) 693 nm diameter.

So, the measured ACFs appear to decay down to the level where the number fluctuation term would be significant. However, as can be seen from Figure 11, at that point the ACFs are dominated by noise and oscillations, so it is difficult to observe the number fluctuation term, much less estimate its level. Nijman et al. [34] also believed that noise was an issue and this was because the experiment time was too short. Our measurement time was 120 s, which was very short compared with the ideal measurement time of ~105 s, assuming measuring an accurate ACF requires a measurement time of at least 10 correlation times.

Number fluctuation exists in scattering volume at any concentration. At normal concentrations (N<<N2), number fluctuations may be ignored since there are a vast number of particles in the scattering volume. At ultra-low concentrations, the proportion of scattered light fluctuations caused by number fluctuations compared with those from Brownian motion increases as the concentration decreases and this is manifested in the ACF and cannot be ignored.

As seen in the simulations, the inversion results obtained by the KFR method are better than those of the BR method. Under the same conditions, the KFR method includes the full number fluctuation component that contains particle size information, while the BR method removes it. Since the number fluctuation component contains particle size information, although the information in this part is far less than it in the Brownian motion component, the BR method cannot give more accurate PSD than those of the KFR method. However, with increasing noise, compared with the benefits of including PSD information, the noise in number fluctuation component will gradually damage the accuracy of the PSDs. At the noise level of 10−1, inversion results by the KFR method get worse. Compared with the normal concentration, the noise in the measured data at low concentrations is more serious and appears more at the end section of the Brownian motion component in the intensity ACF, which aggravates the deterioration of the KFR method inversion results. The influence of noise also affects the reset of the ACF baseline. The accuracy of the inversion result of the BR method gets worse with increasing noise, which is also obvious in the inversion of the measured data. Although both methods cannot give ideal inversions for measured data, compared with KFR method, the BR method is only affected by noise on the truncation point of ACF, and is not affected by the negative effect of noise in the number fluctuation term. Therefore, in actual measurement, it appears that the BR method, as a theoretically unsatisfactory method, can give relatively better PSD results than a theoretically ideal method, the KFR method. Compared with the inversion results of simulated data, it can also be seen that the inversion results of measured data are closer to the inversion results of simulated data at 10−1 noise level, which reflects that the noise at ultra-low concentration is much higher than that at normal concentrations, and becomes an important inhibiting factor for accurate DLS measurements at ultra-low concentration.

5. Conclusions

In DLS measurements, particles entering and leaving the scattering volume give rise to number fluctuations and, therefore, scattered light intensity fluctuations which will manifest in the intensity ACF. When the number of particles in the scattering volume is large, the number fluctuation term in the ACF is usually many orders of magnitude less than the Brownian motion term and can be ignored. When N<50, the contribution of number fluctuations to the ACF cannot be ignored. This is the higher limit of concentration used in our experiments. The key to this limitation is that the field ACF obtained from the measured intensity ACF cannot match with the usual kernel function matrix at ultra-low concentrations, which is based only on the diffusion of the Brownian particles. It is feasible to recover particle sizing information from ultra-low concentration DLS measurements if the number fluctuation term in the intensity ACF is accounted for.

This paper introduces two methods for doing this. The first is the KFR method, which specifically incorporates the number fluctuation term in the intensity ACF model used to reconstruct the kernel function matrix. The second is the BR method, which recognizes that the number fluctuation term has very low amplitude and decays very slowly so that it may be considered a baseline component of the ACF and resetting the baseline may account for it. The two methods were used to recover PSDs from simulated and measured ACF data. As is shown in results, ideal recovered PSDs can be obtained from the simulated data by including the number fluctuation term under the normal noise level, but this does not occur with experimental measured data, and the inversion results of experimental data are closer to those of simulated data with high level noise. The reason is that the measured intensity ACF contains more noise than the simulated ACF at ultra-low concentration. In particular, the baseline noise in the long delay time tail of the ACF overwhelms the number fluctuation term and makes it difficult to recover reliable PSD data. The effective removal of the number fluctuation term from the ACF by resetting the baseline, is also a feasible way to improve PSD recovery at ultra-low concentration. However, the increasing noise at ultra-low concentration leads to errors in determining the effective baseline. This significantly reduces the accuracy of the inversion results. To effectively use the PSD information in intensity ACF at ultra-low concentration, the first priority is to minimize the noise on the ACF. Then, we can further explore the methods proposed above.

Author Contributions

Conceptualization, M.W., J.S. and J.C.T.; Methodology, M.W. and J.S.; Software, M.W. and T.M.; Validation, M.W. and W.L.; Formal analysis, M.W., J.S. and J.C.T.; Investigation, M.W. and Q.W.; Resources, J.S. and J.P.; Data curation, M.W. and T.M.; Writing—original draft preparation, M.W. and J.S.; Writing—review and editing, J.S., J.C.T., Q.W. and K.L.; Visualization, M.W.; Supervision, J.S.; Project administration, J.S., W.L. and Y.W.; Funding acquisition, J.S., J.P. and Y.W. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by National Natural Science Foundation of China, grant number 61801272; the Natural Science Foundation of Shandong Province, grant number ZR2020MF124; the Key Technology Research and Development Program of Shandong, grant number 2019GGX104017; and the Shandong University of Technology and Zibo City Integration Development Project, grant number 2019ZBXC011.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Pecora R. Doppler shifts in light scattering from pure liquids and polymer solutions. J. Chem. Phys. 1964;40:1604–1614. doi: 10.1063/1.1725368. [DOI] [Google Scholar]
  • 2.Cummins H.Z., Knable N., Yeh Y. Observation of diffusion broadening of raleigh scattered light. Phys. Rev. Lett. 1964;12:150–153. doi: 10.1103/PhysRevLett.12.150. [DOI] [Google Scholar]
  • 3.Ford N.C., Benedek G.B. Observation of the spectrum of light scattered from a pure fluid near its critical point. Phys. Rev. Lett. 1965;15:649–653. doi: 10.1103/PhysRevLett.15.649. [DOI] [Google Scholar]
  • 4.Pike E.R. The Malvern correlator. Case study in development. Phys. Technol. 1979;10:104–109. doi: 10.1088/0305-4624/10/3/I03. [DOI] [Google Scholar]
  • 5.Pike E.R. Lasers, photon statistics, photon-correlation spectroscopy and subsequent applications. J. Eur. Opt. Soc. Rapid. 2010;5:138. doi: 10.2971/jeos.2010.10047s. [DOI] [Google Scholar]
  • 6.Gómez-Merino A.I., Rubio-Hernández F.J., Velázquez-Navarro J.F., Aguiar J., Jiménez-Agredano C. Study of the aggregation state of anatase water nanofluids using rheological and DLS methods. Ceram. Int. 2014;40:14045–14050. doi: 10.1016/j.ceramint.2014.05.132. [DOI] [Google Scholar]
  • 7.Hou J., Ci H., Wang P., Wang C., Lv B., Miao L., You G. Nanoparticle tracking analysis versus dynamic light scattering: Case study on the effect of Ca2+ and alginate on the aggregation of cerium oxide nanoparticles. J. Hazard. Mater. 2018;360:319–328. doi: 10.1016/j.jhazmat.2018.08.010. [DOI] [PubMed] [Google Scholar]
  • 8.Zhan S., Hu J., Li Y., Huang X., Xiong Y. Direct competitive ELISA enhanced by dynamic light scattering for the ultrasensitive detection of aflatoxin B1 in corn samples. Food Chem. 2021;342:128327. doi: 10.1016/j.foodchem.2020.128327. [DOI] [PubMed] [Google Scholar]
  • 9.Carvalho J.W.P., Santiago P.S., Batista T., Salmon C.E.G., Barbosa L.R.S., Itri R., Tabak M. On the temperature stability of extracellular hemoglobin of Glossoscolex paulistus, at different oxidation states: SAXS and DLS studies. Biophys. Chem. 2012;163–164:44–55. doi: 10.1016/j.bpc.2012.02.004. [DOI] [PubMed] [Google Scholar]
  • 10.Bai M., Bai C., Daisuke A., Hiromu T., Kensuke M., Takashi Y. Role of a long-chain alkyl group in sulfated alkyl oligosaccharides with high anti-HIV activity revealed by SPR and DLS. Carbohyd. Polym. 2020;245:116518. doi: 10.1016/j.carbpol.2020.116518. [DOI] [PubMed] [Google Scholar]
  • 11.Pecora R. Dynamic Light Scattering Measurement of Nanometer Particles in Liquids. J. Nanopart. Res. 2000;2:123–131. doi: 10.1023/A:1010067107182. [DOI] [Google Scholar]
  • 12.International Organization for Standardization . Particle Size Analysis–Photon Correlation Spectroscopy: ISO 13321[S] International Organization for Standardization; Geneva, Switzerland: 1996. [Google Scholar]
  • 13.International Organization for Standardization . Particle Size Analysis–Dynamic Light Scattering (DLS): ISO 22412:2008[S] International Organization for Standardization; Geneva, Switzerland: 2008. [Google Scholar]
  • 14.International Organization for Standardization . Particle Size Analysis–Dynamic Light Scattering (DLS): ISO 22412:2017[S] International Organization for Standardization; Geneva, Switzerland: 2017. [Google Scholar]
  • 15.Thomas J.C. Photon correlation spectroscopy: Technique and instrumentation. Proc. SPIE. 1991;1430:2–18. doi: 10.1117/12.44153. [DOI] [Google Scholar]
  • 16.Bunkin N.F., Shkirin A.V., Chirikov S.N., Sendrovitz A.L. Effect of the spatial distribution of probe beam on the results of measurements of the disperse composition of nanoparticles by dynamic light scattering method. Bull. Lebedev Phys. Inst. 2016;43:252–255. doi: 10.3103/S1068335616080054. [DOI] [Google Scholar]
  • 17.Kirichenko M.N., Sanoeva A.T., Chaikov L.L. Appearance of an Artifact Peak in the Particle Size Distribution Measured by DLS at Low Concentrations. Bull. Lebedev Phys. Inst. 2016;43:256–260. doi: 10.3103/S1068335616080066. [DOI] [Google Scholar]
  • 18.Bunkin N.F., Shkirin A.V., Suyazov N.V., Chaikov L.L., Chirikov S.N., Kirichenko M.N., Nikiforov S.D., Tymper S.I. Influence of low concentrations of scatterers and signal detection time on the results of their measurements using dynamic light scattering. Quantum Electron. 2017;47:949–955. doi: 10.1070/QEL16408. [DOI] [Google Scholar]
  • 19.Yang H., Yang H., Kong P., Zhu Y., Dai S., Zheng G. Concentration measurement of particles by number fluctuation in dynamic light backscattering. Powder Technol. 2013;246:499–503. doi: 10.1016/j.powtec.2013.06.006. [DOI] [Google Scholar]
  • 20.Willemse A.W., Drunen M.A.V., Tuinman I.L., Marijnisse J.C.M., Merkus H.G., Scarlett B. Photon correlation spectroscopy for analysis of low concentration aerosols. J. Aerosol Sci. 1995;26:S31–S32. doi: 10.1016/0021-8502(95)96923-U. [DOI] [Google Scholar]
  • 21.Wuyckhuyse A.L.V., Willemse A.W., Marijnissen J.C.M., Scarlett B. Determination of on-line particle size and concentration for sub-micron particles at low concentrations. J. Aerosol Sci. 1996;27:S577–S578. doi: 10.1016/0021-8502(96)00361-8. [DOI] [Google Scholar]
  • 22.Willemse A.W., Marijnissen J.C.M., Roos R., Wuyckhuyse A.L.V., Scarlett B. Photon correlation spectroscopy for analysis of low concentration submicrometer samples. J. Aerosol Sci. 1996;27:S535–S536. doi: 10.1016/0021-8502(96)00340-0. [DOI] [Google Scholar]
  • 23.Willemse A.W., Marijnissen J.C.M., Wuyckhuyse A.L.V., Roos R., Merkus H.G., Scarlett B. Low-concentration photon correlation spectroscopy. Part. Part. Syst. Charact. 1997;14:157–162. doi: 10.1002/ppsc.199700036. [DOI] [Google Scholar]
  • 24.Willemse A.W., Marijnissen J.C.M., Wuyckhuyse A.L.V., Roos R., Merkus H.G., Scarlett B. Photon correlation spectroscopy for analysis of low concentration submicrometer samples. Progr. Colloid Polym. Sci. 1997;104:113–116. doi: 10.1007/BF01182424. [DOI] [Google Scholar]
  • 25.Schaefer D.W., Berne B.J. Light scattering from Non-Gaussian concentration fluctuations. Phys. Rev. Lett. 1972;28:475–478. doi: 10.1103/PhysRevLett.28.475. [DOI] [Google Scholar]
  • 26.Chaikov L., Kirichenko M.N., Krivokhizha S.V., Zaritskiy A.R. Dynamics of statistically confident particle sizes and concentrations in blood plasma obtained by the dynamic light scattering method. J. Biomed. Opt. 2015;20:57003. doi: 10.1117/1.JBO.20.5.057003. [DOI] [PubMed] [Google Scholar]
  • 27.Kirichenko M.N., Masalov A.V., Chaikov L.L., Zaritskii A.R. Relation between particle sizes and concentration in undiluted and diluted blood plasma according to light scattering data. Bull. Lebedev Phys. Inst. 2015;42:33–36. doi: 10.3103/S1068335615020013. [DOI] [Google Scholar]
  • 28.Surianarayanan R., Shivakumar H.G., Vegesna N.S.K.V., Srivastava A. Effect of sample Concentration on the Characterization of Liposomes using Dynamic light Scattering Technique. Pharm. Methods. 2016;7:70–74. doi: 10.5530/phm.2016.7.11. [DOI] [Google Scholar]
  • 29.Schatzel K. Correlation techniques in dynamic light scattering. Appl. Phys. B. 1987;42:193–213. doi: 10.1007/BF00693937. [DOI] [Google Scholar]
  • 30.Nijman E.J., Merkus H.G., Marijnissen J.C.M., Scarlett B. Simulations and experiments on number fluctuations in photon-correlation spectroscopy at low particle concentrations. Appl. Opt. 2001;40:4058–4063. doi: 10.1364/AO.40.004058. [DOI] [PubMed] [Google Scholar]
  • 31.Yu A.B., Standish N. A study of particle size distribution. Powder Technol. 1990;62:101–118. doi: 10.1016/0032-5910(90)80073-8. [DOI] [Google Scholar]
  • 32.Xu M., Shen J., Thomas J.C., Huang Y., Zhu X., Clementi L.A., Vega J.R. Information-weighted constrained regularization for particle size distribution recovery in multiangle dynamic light scattering. Opt. Express. 2018;26:15–31. doi: 10.1364/OE.26.000015. [DOI] [PubMed] [Google Scholar]
  • 33.Xu M., Shen J., Huang Y., Xu Y., Zhu X., Wang Y., Liu W., Gao M. Weighting inversion of dynamic light scattering based on particle-size information distribution character. Acta Phys. Sin. 2018;67:134201. doi: 10.7498/aps.67.20172377. [DOI] [Google Scholar]
  • 34.Nijman E.J., Merkus H.G., Marijnissen J.C.M., Scarlett B. Photon. Correlation and Scattering. Optical Society of America; Whistler, BC, Canada: 2000. Investigation of the influence of low particle concentration on photon correlation spectroscopy. [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable.


Articles from Materials are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES