Abstract
The human cerebellum is involved in a wide array of functions, ranging from motor control to cognitive control, and as such is of great neuroscientific interest. However, its function is underexplored in vivo, due to its small size, its dense structure and its placement at the bottom of the brain, where transmit and receive fields are suboptimal. In this study, we combined two dense coil arrays of 16 small surface receive elements each with a transmit array of three antenna elements to improve BOLD sensitivity in the human cerebellum at 7 T. Our results showed improved B 1 + and SNR close to the surface as well as g‐factor gains compared with a commercial coil designed for whole‐head imaging. This resulted in improved signal stability and large gains in the spatial extent of the activation close to the surface (<3.5 cm), while good performance was retained deeper in the cerebellum. Modulating the phase of the transmit elements of the head coil to constructively interfere in the cerebellum improved the B 1 +, resulting in a temporal SNR gain. Overall, our results show that a dedicated transmit array along with the SNR gains of surface coil arrays can improve cerebellar imaging, at the cost of a decreased field of view and increased signal inhomogeneity.
Keywords: 3D EPI; 7 T, BOLD; cerebellum; parallel imaging; receive arrays
Human cerebellar fMRI is hindered by B 1 + cancellation in ultra‐high fields and the need for high resolution. Using a dedicated 3Tx array improved the transmit field close to the cerebellar surface compared with a commercial 8Tx coil. The signal‐to‐noise and parallel imaging performance were further increased by the usage of a high‐density surface receive array. The combination of improved transmit and receive fields resulted in increased BOLD sensitivity close to the surface of the cerebellum.
Abbreviations
- BOLD
blood oxygenation level dependent
- EPI
echo planar imaging
- FOV
field of view
- SENSE
sensitivity encoding
- SNR
signal‐to‐noise ratio
- tSNR
temporal SNR
- UHF
ultra‐high field
1. INTRODUCTION
The cerebellum, or little brain, is a striking structure that, while of relatively small volume, contains approximately 80% of the neurons in the brain of most mammals. 1 The function of the cerebellum is typically associated with the motor system, but it is increasingly accepted that the cortex and the cerebellum tightly coordinate in several cognitive domains, including executive functions. 2 The cerebellum is involved in several neurological diseases, both brain wide (eg multiple sclerosis 3 ) and cerebellum specific (eg cerebellar ataxia 4 ). As such, the cerebellum is of wide neuroscientific interest, but its function is relatively understudied in vivo with MRI due to its high degree of gyrification and thin grey matter (which necessitate high spatial resolution) and suboptimal transmit fields. High‐resolution fMRI relies on the increased signal‐to‐noise ratio (SNR) afforded by ultra‐high‐field (UHF) magnetic fields and the advent of parallel imaging. In this study, we examined the transmit, SNR and parallel imaging advantages offered by a dedicated coil in UHF for cerebellar fMRI.
In recent years, UHF fMRI has been gaining traction due to the roughly linear SNR gain and supralinear contrast‐to‐noise gain in susceptibility‐based contrasts, such as blood oxygenation level dependent (BOLD) imaging. 5 , 6 These gains, however, are usually unequally spread across the brain: typical volume transmit coils produce B 1 + cancellation in the periphery (such as the cerebellum) due to the higher Larmor frequency. This can be mitigated with dielectric pads 7 or multiple transmit elements with a phase or amplitude modulation (B 1 shimming) so that the B 1 + field constructively interferes within the area of interest. 8 Even so, circularly polarized volume transmit coils are still largely the norm. 9 Another approach to further optimize the B 1 + field is to optimize the shape and number count of the transmission array for the region of interest. Such an approach was used for example to improve fMRI imaging of the visual cortex at 7 T, 10 and may be promising for peripheral regions such as the cerebellum. 11
On the receive side, high SNR can be achieved in the periphery of the brain: surface coils (essentially wire loops) show high sensitivity close to the surface, with a penetration depth roughly equivalent to their diameter. Due to the reduced penetration depth, noise from deep in the sample does not contribute to the signal picked up by the receiver. Reducing the coil diameter therefore results in increased SNR close to the surface 12 , 13 at the cost of decreased sensitivity deeper in the tissue. It has been shown that the sensitivity loss deeper in the tissue can be mitigated by increasing the number of receive elements, 14 while still retaining the high surface SNR. In agreement with the above, a dense array of 16 receive channels of 2 cm diameter has been shown to be beneficial for high‐resolution fMRI imaging of the visual cortex at 7 T. 15 For small areas such as the cerebellum that are sited close to the skull, small high‐count surface coils arranged in dense and flexible receive arrays can be highly beneficial at the cost of less uniform signal. 16 , 17 , 18
In neuroimaging MRI studies, high‐count receive arrays are routinely used for parallel imaging through spatial encoding in the coil dimension. This is partially fueled by the numerous benefits of parallel imaging for BOLD fMRI, in its typical T 2*‐weighted echo planar imaging (EPI) implementation: long readouts associated with high‐resolution scans suffer from k‐space blurring, susceptibility artifacts and distortions due to off‐resonance effects. The above artifacts are exacerbated in UHF imaging; it is thus necessary to shorten the readout length with parallel imaging to achieve high spatial resolution. 19 Spatial encoding in the coil dimension (and therefore the parallel imaging acceleration factors) depends on achieving distinct sensitivity profiles between coils. A high count of small receive elements (each with reduced spatial sensitivity) may, therefore, be advantageous for reaching high acceleration factors in a small region such as the cerebellum, if distinct spatial sensitivity profiles and adequate decoupling between the receivers are achieved. 20 , 21 This tradeoff is easier to balance at UHF since the higher frequencies at higher field strengths enhance far‐field conditions for radiofrequency waves.
The optimization of multi‐transmit and multi‐receive arrays is critical to reap the UHF benefits for neuroimaging. In this study, we examine a dedicated back‐of‐the‐head multi‐transmit along with a flexible receive array of high‐density small surface coils for BOLD fMRI imaging of the cerebellum at 7 T. This setup was compared with a commercially available multi‐transmit whole‐head coil, both with and without B 1 shimming in the cerebellum.
2. METHODS
2.1. Coil design and optimization
A 3‐channel transmit coil array and a 32‐channel receive surface coil array were developed by MR Coils (Zaltbommel, The Netherlands). The design of the transmit antenna was optimized to fit in a half‐cylinder (to not obscure the subject's field of view, FOV) and to maximize B 1 + at the back of the head. Three antennas were used with two larger outer fractionated dipole antennas of 27.5 by 4.5 cm, 22 placed at a 45° angle. The middle element was a snake antenna (17.5 cm by 4 cm) 23 positioned near the bottom of the half‐cylinder (Figure 1A and 1C). Electromagnetic simulations were performed using Sim4Life (ZMT, Zürich, Switzerland) to validate the coil for transmit field and SAR limits 24 , 25 (see Supporting Figures S1, S2 and S3 in supplementary material for more details). Each of the two receive arrays consisted of four sets of four coils, based on a previous design (see References 15 , 26 for more details on the receive array's performance). Each receive coil had a dimension of 2 × 3 cm2. The coils in each set partially overlapped (0.5 cm). The receive elements were decoupled with high‐impedance preamplifier decoupling that added between −11 and −17 dB of isolation. The sets were arranged in two flexible arrays of approximately 8 × 10 cm2 each (Figure 1A and 1D).
FIGURE 1.
A, Coronal, sagittal and axial views of the cerebellar coil. Red, transmit elements. Blue, receive elements. B, Distribution of voxels in the cerebellum with respect to distance from the skull. The median distance was 2.58 cm (dotted line). C, Picture of the transmit array of the cerebellar coil (housing was partially removed). D, Picture of the receive array (the covers were opened for the picture). E, Picture of the 8Tx/32Rx commercial head coil
Padding was used to press the surface coil array against the skull for optimal coil loading and signal reception. For MR measurements, a 32‐channel Philips interface box was used. To maximize B 1 +, optimal phase offsets were simulated using Sim4Life on the human model Duke 27 , 28 (Supporting Figure S1) and then finetuned on a person during the MR measurement. The optimum B 1 + was found to be stable across participants and settled on a phase shift of 112° between the two outer elements.
2.2. Experimental design
The performance of the cerebellar coil was examined in terms of the B 1 + field produced, the scattering matrix of the transmit elements, noise correlation between the receive elements, the SNR, the g‐factor maps for different acceleration factors, the temporal SNR (tSNR) of the T 2*‐weighted EPI data and finally the BOLD sensitivity. Six healthy volunteers were scanned with a Philips Achieva 7 T (Philips, Best, The Netherlands). The study protocol and informed consent were approved by the local ethical committee. The participants underwent two consecutive imaging sessions. In each session, either a commercially available 8 Tx/32 Rx whole‐head coil 29 (Nova Medical, Wilmington, MA; Figure 1E) or the cerebellar coil (3 Tx/32 surface Rx) were used, along with the standard 2 kW peak RF power amplifiers per channel. The order of the sessions was alternated between participants. The reference RF power was optimized in each session with a vendor‐provided optimization approach that determines the flip angle by taking the ratio between a spin‐echo and a stimulated echo during the pre‐scans. A B 0 map was acquired (FOV = 224 × 224 × 224 mm3, voxel size 3.5 mm isotropic, T R/T E = 4 ms/1.54 ms, flip angle = 8°) and the B 0 field was homogenized within the brain with a second‐order shim by minimizing the variance of the B 0 distribution in a least‐squares way using MRCodeTool v1.5.7 (Tesla Dynamic Coils, Zaltbommel, The Netherlands). During the head coil session, data were acquired with and without B 1 shimming (quadrature mode) to examine if shimming of the eight transmit elements could also result in a sufficient B 1 + in the cerebellum. For one participant, no anatomical or functional data were acquired without B 1 shimming.
Optimal phase modulations were calculated to minimize the cost function over the cerebellum as implemented in MRCodeTool. To calculate the individual‐channel B 1 + fields, a DREAM B 1 + map was acquired while all coils were transmitting (FOV = 224 × 224 × 168 mm3, voxel size = 3.5 mm3, T R/T E = 6 ms/3 ms, flip angle = 7°) 30 as well as a spoiled gradient echo while transmitting with each channel separately (FOV = 224 × 224 × 168 mm3, voxel size = 3.5 mm isotropic, T R/T E = 8 ms/1.97 ms, flip angle = 1.5°). From these, the relative channel‐specific transmit fields were estimated. 31
The participants performed a motor task while BOLD‐weighted 3D‐EPI data were acquired (FOV = 200 × 200 × 176 mm3, voxel size = 1.8 mm isotropic, T Rvolume/T R/T E = 1300 ms/44 ms/17 ms, flip angle = 13°, sensitivity encoding (SENSE) y/z = 3.2/2.6). The motor task consisted of 13 s of bilateral finger tapping movements interleaved with 13 s of rest, with a total duration of 4 min for each run. This task is known to elicit activation in the anterior (Lobule V) and posterior lobes of the cerebellum (Lobules V and VIII). 32 , 33 The first five 3D‐EPI volumes were discarded to ensure a longitudinal steady state during excitation. Five volumes with reversed phase encoding in the LR direction were also recorded to estimate the off‐resonance field due to susceptibility. In the head‐coil session, an MPRAGE with an elliptical k‐space shutter was acquired (FOV = 200 × 221 × 180 mm3, voxel size = 0.9 mm isotropic, T R/T E = 150 ms/3 ms, flip angle = 7°, SENSE y/z = 2/2.5, T I1 = 1300 ms) as an anatomical reference.
For a B 1 + comparison between the head coil and the cerebellar coil, DREAM B 1 + maps were recorded, as described above. Furthermore, for one participant, a noise 3D EPI image (same parameters as above) was acquired with both coil setups by turning off the gradients and the RF pulses. The vendor‐supplied scattering and noise correlation matrices were extracted (noise correlation calculated from a pre‐scan without RF excitation).
Finally, for another participant, we compared the coils' parallel imaging capability by acquiring 2D EPI volumes with increasing undersampling factors in the LR direction (2 × 1, 3 × 1, 4 × 1 and 6 × 1). The data from each coil channel were used to create g‐factor maps.
2.3. Image processing and quantification
mage preprocessing of the BOLD data was performed with FSL 6.0.1 (https://fsl.fmrib.ox.ac.uk/fsl/fslwiki/). 34 The fMRI images were motion corrected towards the volume that was acquired just before the reversed phase encoding scan with a six degrees of freedom (6‐dof) transform using MCFLIRT. 35 To reduce the EPI distortions, a displacement field was calculated and applied with FSL‐TOPUP by employing a reversed phase encoding scan and their five temporally adjacent volumes for each fMRI scan. 36 EPI‐to‐MPRAGE 6‐dof transforms were calculated for both coils, following initial manual realignment (ITK‐SNAP 3.6.0; http://www.itksnap.org/). MPRAGE‐to‐MNI template registrations were calculated employing a diffeomorphic transform (ANTS 2.1; http://stnava.github.io/ANTs/). 37 , 38 The transforms were concatenated to a single warp. Cerebellar and skull ROIs were created in the MNI space (SUIT atlas 39 and FSL‐FAST respectively) and projected to each native EPI and B 1‐map space. The Euclidean distance from each voxel of the cerebellar ROI towards the inner surface of the skull was calculated to interpret the distance‐dependent characteristics of the coils (median distance of the cerebellar voxels from the skull = 2.58 cm; Figure 1B).
A general linear model was fitted at the individual level for each voxel (finger tapping > rest) with the FMRI Expert Analysis Tool, using generalized least squares with a voxel‐wise, temporally and spatially regularized autocorrelation model. The regressors were convolved with a double gamma hemodynamic response function before they were entered in the model. The framewise displacement was calculated and entered as a confound regressor. 40 A high‐pass filter (cutoff at 100 s) was applied before the fit. Only the voxels within the native‐space cerebellum mask were considered, to ensure a fair comparison between coils, given their coverage difference. The subject‐level z‐maps were cluster‐mass thresholded (cluster‐defining threshold = 3.1) and their volume extracted.
Low B 1 + tends to be more evident in the posterior cerebellum and more in the right than the left hemisphere, due to destructive interferences or asymmetric transmission. 41 As an additional measure of sensitivity, we created ROIs for Lobules V (anterior cerebellum) and VIII (posterior cerebellum), where finger tapping is expected to elicit activation, 32 with the SUIT atlas. 39 The number of activated voxels within each lobule was extracted and the ratio between voxels in the left/right hemisphere was calculated. The SNR within the cerebellum was calculated by dividing the EPI amplitude with the standard deviation of the noise image (normalized and not normalized with the flip angle ratio), while the tSNR was calculated as the mean EPI amplitude over the task fMRI timeseries divided by the standard deviation of the amplitude over time. All of the above measures, along with z‐maps, g‐factor maps and B 1 maps, were projected to the MNI space.
3. RESULTS
3.1. B 1 field
Within the cerebellum, the B 1 + variation of the cerebellar coil (difference between 75% and 25% quartiles = 45.91%—median = 75.53%, quartile1,3 = 52.56, 98.5%—calculated from a between‐subjects average of the percentage to the nominal B 1 +) was similar to that of the 8Tx head coil both following RF shimming (difference between 75% and 25% quartiles = 48.18%—median = 90.16%, quartile1,3 = 66.07, 114.25%) and in quadrature mode (difference between 75% and 25% quartiles = 36.97%—median = 71.33%, quartile1,3 = 52.85, 89.82%). The coils showed different transmit behavior, with the cerebellar coil being particularly efficient close to the surface (up to 3 cm deep), where the head coil's transmit field suffers the most (Figure 2). Sample EPI images produced by each coil are shown in Supporting Figure S4.
FIGURE 2.
Estimated flip angle from B 1 + maps, expressed as the percentage of the nominal flip angle. A, Group flip angle % distribution of coils and their ratio (columns) along axial slices at the height of the cerebellum (rows). B, Group flip angle % within the cerebellum for the different coil setups. C, Group flip angle % plotted against distance from the skull
3.2. Radiofrequency correlation
The scattering matrix of the transmit array and the noise correlation of the receive array were calculated for one individual (Figure 3). The coupling coefficients of the transmit (S ii , S ij ) were S 11 = −11.7 dB, S 12 = −17.9 dB, S 13 = −22.5 dB, S 22 = −5.71 dB, S 23 = −22.85 dB, S 33 = −5.1 dB. The median noise correlation between the receive elements of the cerebellar coil (surface coil array) was 0.05 (max. = 0.65, min. = 0.002). For comparison, the median correlation of the head coil was 0.03 (max. = 0.43, min. < 0.001).
FIGURE 3.
Noise correlation matrices. A, Cerebellar coil. B, Head coil
3.3. SNR
For one individual, noise scans were acquired along with the 3D‐EPI to calculate SNR maps. The SNR achieved by the conformal and smaller loops of the cerebellar coil was higher (median = 3.72, IQR = 0.1‐7.33) compared with the head coil either with RF shimming (median = 2.91, IQR = 1.35‐4.48) or without (unshimmed, median = 2.43, IQR = 1‐3.86), resulting in a 27.51% median increase over the whole cerebellum for the shimmed head coil and 52.8% for the unshimmed coil. Note that SNR was substantially increased closer to the surface (up to 308.42% higher) for the cerebellar coil, while it reached similar values to the shimmed head coil at 3 to 4 cm depth (Figure 4). SNR data that are normalized for B 1 + are presented in Supporting Figure S5.
FIGURE 4.
SNR maps for one participant. A, Coils and their ratios (columns) along axial slices at the height of the cerebellum (rows). B, SNR distribution within the cerebellum for the different coil setups. C, SNR plotted against the distance from the skull
3.4. G‐factor
To compare the coils' performance in parallel imaging, g‐factor maps were calculated (Figure 5). The g‐factor of the cerebellar coil stayed close to 2 within the cerebellum up to an acceleration factor of 6 (median = 2.1, IQR = 1.91‐2.3 for the cerebellum coil, versus median = 2.73, IQR = 2.52‐2.95 for the head coil). Furthermore, a more homogeneous performance was achieved, as can be seen in the tighter histogram distribution of g‐factors.
FIGURE 5.
g‐factor maps for one participant. A, Different acceleration factors in the L‐R direction (columns) for several axial slices at the height of the cerebellum (rows). B‐E, g‐factor distributions
3.5. tSNR
The median tSNR over the cerebellum, as a measure of BOLD signal stability, was slightly higher on average for the shimmed head coil (median = 31.98, IQR = 21.42‐42.45) compared with the cerebellar coil (median = 29.88, IQR = 17.43‐42.34) and the unshimmed head coil (median = 28.03, IQR = 15.29‐40.77), though the cerebellar coil showed a tSNR advantage up to 2 cm in depth (Figure 6).
FIGURE 6.
BOLD fMRI metrics. A, Group tSNR maps for each coil setup and their ratios (columns) along axial slices at the height of the cerebellum (rows). B, Group tSNR distribution. C, Group tSNR standard variability between participants. D, Group tSNR in relation to the distance from the skull
A practical concern when using transmit and receive arrays that cover a reduced FOV is that this may lead to increased signal variability between subjects, since the placement of the coil in relation to the structure of interest may differ. To evaluate this, we estimated the tSNR variability between subjects. The median between‐subjects variability in tSNR along the cerebellum was higher for the cerebellar coil (median = 9.51, IQR = 3.8‐15.22) compared with the head coil (shimmed head coil, median = 7.7, IQR = 3.57‐11.81; unshimmed head coil, median = 7, IQR = 1.97‐11.95), leading to a median 23.64% tSNR variability increase compared with the shimmed head coil and a 36.6% variability increase compared the unshimmed head coil.
One final consideration is that the use of a half‐volume coil transmit setup along with conformal (and looser) receive elements may result in a less‐tight fit and increased head motion. The median framewise displacement along participants was indeed slightly higher for the cerebellar coil (median = 0.34; IQR = 0.28‐0.36) compared with the head coil (shimmed head coil (median = 0.22; IQR = 0.19‐0.26) and unshimmed head coil (median = 0.26; IQR = 0.23‐0.26)).
3.6. BOLD sensitivity
The spatial extent of the activation was estimated for each coil and participant by normalizing the number of active voxels (z > 3.1) with the total number of voxels within the cerebellum (MNI cerebellar ROI projected to individual space; Table 1). Overall, the spatial extent of the significant clusters for the cerebellar coil was increased compared with the shimmed head coil by a median factor of 2.08 (IQR = 1.52‐5.22) and with the unshimmed head coil by a factor of 1.32 (IQR = 0.78‐2.96). Cerebellar activity for this bilateral hand motor task is expected in both left and right cerebellar Lobule V in the anterior lobe and left and right cerebellar Lobule VIII in the posterior lobe.
TABLE 1.
BOLD sensitivity. Each row represents a participant. The percentage of activated voxels normalized with the number of voxels within the mask is given per coil (Columns 1, 2 and 3). The ratios between coil combinations are also given (Columns 4 and 5)
Cerebellar coil | Head coil shimmed | Head coil unshimmed | Cerebellar coil/head coil shimmed | Cerebellar coil/head coil unshimmed | |
---|---|---|---|---|---|
S1 | 5.65 | 3.47 | NA | 1.63 | NA |
S2 | 0.1 | 0.27 | 0.97 | 0.37 | 0.1 |
S3 | 3.11 | 1.23 | 0.49 | 2.53 | 6.29 |
S4 | 0.38 | 0.06 | 0.49 | 6.12 | 0.78 |
S5 | 1.34 | 0.12 | 1.02 | 11.28 | 1.32 |
S6 | 3.27 | 2.2 | 1.1 | 1.48 | 2.96 |
Median (IQR) | 2.23 (0.62‐3.23) | 0.75 (0.16‐1.96) | 0.97 (0.49‐1.02) | 2.08 (1.52‐5.22) | 1.32 (0.78‐2.96) |
The z‐statistic distribution of the increase in the number of activated voxels across groups is shown in Figure 7B: more activated voxels were detected with the cerebellar coil for all z‐values. This spatial‐extent gain was more pronounced for distances up to 3‐4 mm from the skull in accordance with the SNR and B 1 + profiles described earlier (Figure 7C). Individual clusters for each coil are shown in Supporting Figure S6.
FIGURE 7.
A, (Finger tapping > rest) BOLD fMRI significant clusters (z > 3.1) for 2 sample participants along axial slices at the height of the cerebellum, crossing the hand region in cerebellar Lobule V (left) and cerebellar Lobule VIII (right). B, Group z‐statistic distribution for each coil. C, Active voxels at group level in relation to the distance from the skull. The participant for whom no data were acquired before B 1 shimming was excluded
The median right/left hemisphere ratio of activation within Lobule V was 58.69% for the cerebellar coil, 62.1% for the shimmed head coil and 55.11% for the unshimmed head coil (100% denoting matched volume of activation between left and right hemispheres). The median right/left ratio of activation within Lobule VIII (where signal cancellation is a concern) was 75.28% for the cerebellar coil and 35% for the shimmed head coil, while no ratio could be extracted for the unshimmed head coil, due to low sensitivity (no active voxels found in right Lobule VIII). The right/left ratio suggests an improvement in the homogeneity of the sensitivity for the cerebellar coil in Lobule VIII.
4. DISCUSSION
In this study, we have demonstrated that a relatively simple array of small receive elements combined with a dedicated transmit array can markedly enhance BOLD imaging in the cerebellum compared with a whole‐head coil. These benefits were evident even when comparing the cerebellar coil with the head coil driven with an optimized RF shim. fMRI research in critical brain regions such as the cerebellum has so far been hindered by low SNR and spatial resolution. Our results show that the usage of dedicated coil arrays can improve BOLD sensitivity and SNR by a factor of approximately two or three, which can readily be translated to higher resolution.
Surface coil arrays are widely known to show high sensitivity close to the surface. Decreasing the diameter increases SNR as long as the sample noise dominates the coil resistance loss. For 7 T applications, the SNR has been predicted to increase up to a diameter close to 2.5 cm. 42 In line with these predictions, dense arrays of small receive elements have been shown to be advantageous for BOLD imaging of the visual cortex, 15 , 43 , 44 as well as in the cerebellum. 20 To realize these SNR benefits, it is crucial that the arrays are placed adjacent to the tissue to ensure sufficient loading and optimal coupling. Using flexible arrays, we managed to reach up to a threefold SNR increase close to the surface. It has been demonstrated using simulations that high‐count arrays can retain good sensitivity even deeper in the tissue. 14 , 26 In our own dense 32‐channel array, the SNR and BOLD sensitivity were comparable to the head coil's performance at approximately 4 cm into the tissue (Figures 4 and 7), while benefiting from high SNR closer to the surface. To reap the SNR benefits of the surface coil arrays over the cerebellum, accurate placement is important. In practice, the tSNR variability between individuals was only slightly increased for the cerebellar coil (Figure 6C), suggesting that good signal stability was achieved between participants. A head‐shaped cradle for the receivers may further reduce the variability between participants. In this paper we opted instead to use the receivers in a flexible array to set them as close as possible to the head to maximize tissue coupling and allow their usage in other brain regions.
The advent of high‐resolution BOLD was achieved by the popularization of sequences that allow undersampling in two directions (multiband 45 and 3D‐EPI 46 ) and high‐count receive arrays that can disentangle the aliased images through inhomogeneous coil sensitivity profiles. The high‐count arrays of small loops that we used here provide many degrees of freedom, which enables parallel imaging when distinct sensitivity profiles are present in the accelerated direction. Note that we used a gapped (instead of an overlapped) design between receiver columns in the left‐right direction to minimize the g‐factor and increase coverage, in a tradeoff with the need to achieve homogeneous SNR (and inductive decoupling via overlapping).. 26 , 47 In practice, SNR was homogeneous within the brain (Figure 4A), and high acceleration (up to 6) was achieved in the left‐right direction (Figure 5) with little signal degradation despite the higher noise correlation between channels of the surface coil array, demonstrating the advantages of using more receive elements per unit distance. Sampling schemes that increase the coil sensitivity differences between aliased slices, such as CAIPIRINHA, 48 may be particularly beneficial to further increase the acceleration factors allowed in the restricted FOVs of such small arrays. 44 , 49 In future, surface coil arrays may be adapted to cover the whole of the cortex with great accelerations gains. 10 , 26 , 50
One of the biggest challenges in UHF is achieving homogeneous excitation. The average B 1 + map of our head coil showed the well known spatial pattern of destructive interferences in the periphery compared with the center. 51 Our dedicated back‐of‐the‐head transmit array achieved high transmit efficiency up to 3 cm deep in the cerebellum, but the B 1 + uniformity did not improve compared with the head coil (Figure 2B). In pilots, individual static RF shimming was found to not improve the results further (data not shown), likely due to the initial optimization of phase offsets (over the relatively small volume of the cerebellum) and the limited number of elements. 52 Dynamically modulating the RF waveforms in each channel may further homogenize the flip angle distribution. 53 This is particularly the case for large regions of interest (where the finite number of transmit elements cannot be statically modulated to constructively interfere over the whole area), but may also be beneficial for small regions such as the cerebellum (as is suggested by the non‐homogeneous B 1 + group average after static RF shimming 11 ; Figure 2A). Individually optimized and transmit‐element‐specific RF pulses are costly because of the time required to calculate them online during a scanning session. However, the fact that spatial patterns can be clearly observed on average suggests that “universal” pulses can be calculated offline. 54 This may be a promising approach to further optimize the flip angle distribution in cerebellar BOLD imaging. 55
Overall, the SNR and parallel imaging advantages of the cerebellar coil led to increased BOLD sensitivity in most of the cerebellum (up to 3‐3.5 cm deep, with the median distance of the cerebellum to the skull being 2.58 cm), while good performance was retained at greater depth (Figure 7). Optimization of the transmit phases of the head coil, while it improved tSNR, did not achieve the same improvements. Furthermore, the SNR gains of the cerebellar coil were achieved without individual B 1 shimming, which reduces the total scan time. Our two receiver arrays, when placed in parallel, easily covered the required FOV for cerebellar imaging. The cerebellar coil may also be more comfortable for the participants: since only three transmit elements were placed at the back of the head, the transmit array took the form of a half‐cylinder to allow for flexible arrangement of the surface coil arrays. In comparison, typical head coils are tighter (to optimize the transmit fields and signal reception) and are frequently uncomfortable. Space within the bore comes at a premium: the half cylinder coil frees up valuable space to allow the placement of larger mirrors and additional equipment (eg eyetracking devices, goggles). 10 We observed a small increase in fractional displacement during the task with the cerebellar coil that may relate to this looser arrangement, though it remained below worrisome levels. 40 The cerebellar coil can readily be adapted to image cortical regions at the back of the head, such as the visual cortex, since most of the cortex is 3‐4 cm thick and lies close to the surface of the brain. 56
The human cerebellum is a complicated structure that packs four‐fifths of the brain's neurons into an area much smaller than the cortex, a feat that is made possible by intense gyrification and decreased neuronal volume due to the reduced number of long‐range axons. 1 The grey matter of the cerebellum is just 0.5 mm thick; this has led to a scarcity of fMRI studies, even though there is evidence of its involvement in tasks across the cognitive domain. 57 , 58 Furthermore, due to the lack of spatial resolution, many of these analyses are focused on the cerebellar lobules, 20 while more detailed, individual‐level analysis of the cerebellar cortex surface may be beneficial. 59 , 60 , 61 , 62 The SNR benefits shown here can be readily translated to higher resolution and increased sensitivity to the BOLD signal in the cerebellar cortex. Variability in the placement and spatial extent of activations was observed between individuals, particularly in the posterior lobe, the most inferior part of the cerebellum, even though the tSNR standard deviation between individuals was largely consistent. This has been reported before 58 and may partially relate to sub‐optimal spatial registration, the large veins adjacent to the cerebellum, physiological noise, as well as attention differences between individuals. Increasing the BOLD fMRI resolution, optimizing the physiological‐to‐thermal signal ratio 63 , 64 and further improving cerebellar anatomical imaging 62 , 65 , 66 may help in this regard.
There are limitations in the current study: the mean flip angle within the cerebellum for each coil was not forced to match the nominal flip angle. This renders SNR comparisons more challenging. It should be noted, however, that the average B 1 + is similar between the cerebellar (75%) and unshimmed head coils (71%). Given our T R/T 1 and the signal equation of the spoiled GRE, the resulting signal difference due to this flip angle variation is less than 5%. Instead, large depth‐dependent variations (25% of the nominal flip angle) were observed in both coils' flip angle distribution (Figure 2B). SNR comparisons normalized with the flip‐angle ratio largely reproduced these results (Supporting Figure S5). Furthermore, we employed small receive elements that are not likely to benefit the intrinsic SNR for a large part of the cerebellum at 7 T. 42 Larger elements might have been beneficial in this regard, though the smaller loops are still likely to result in SNR benefits for highly accelerated fMRI due to the reduced g‐factor. 67 Even with our current, relatively conservative acceleration, good sensitivity was retained across the whole cerebellum. Ultimately, the optimum loop size is determined by the array size, field strength and intended application of the coil. 42
The human cerebellum is an important structure that is frequently ignored. The cerebellar coil, consisting of 32 small surface coil arrays arranged in two dense, high‐count arrays and three back‐of‐the‐head transmit elements, achieved improved BOLD imaging of the human cerebellum, compared with a commercial head coil with 8 transmit and 32 receive channels. The stronger BOLD signal we achieved will facilitate future research in the human cerebellum.
FUNDING INFORMATION
This work is supported by a Nederlandse Organisatie voor Wetenschappelijk Onderzoek grant (VIDI 198.016) to Wietske van der Zwaag.
Supporting information
Figure S1. Maximum intensity projections of B1 normalized for 2 W of input power. A‐B, Snake antenna array (no shim/shimmed). C‐D, Loop coil array (no shim/shimmed). Note that the snake antenna provides a more homogeneous transmit profile over the cerebellum (white boxes).
Figure S2. Maximum intensity projections of SAR10g for the Duke model (Sims4Life). A, Worst‐case phase shim cerebellar transmit (maximum SAR10g = 1.68 W/kg/W of input power), B, Phase shimmed for maximum B1 cerebellar transmit (maximum SAR10g = 1.58 W/kg/W of input power).
Figure S3. A, Successive slices of the receive array detuning ratio (flip angle map with the surface receive array present divided by the flip angle map with the surface receive array replaced with a pad). The ratio was calculated over the area with sufficient signal in the DREAM images and is overlaid as a colormap over the second magnitude image of the DREAM acquisition. B, example slices of the magnitude 1 and 2 of the DREAM acquisition. C, Examples slices of the resulting flip angle maps with and without the receive array present. D, Histogram of the detuning ratio values within the ROI (colormap).
Figure S4. Example sagittal, coronal and axial slices (columns) of the EPI for each coil. A, Cerebellar coil. B, Head coil unshimmed. C, Head coil shimmed.
Figure S5. B1 + normalized SNR maps for one participant. A, coils and their difference (columns) along axial slices at the height of the cerebellum (rows). B, SNR distribution within the cerebellum for the different coil setups. C, SNR plotted against the distance from the skull.
Figure S6. (Finger tapping > Rest) BOLD fMRI significant clusters (z > 3.1) for all participants (rows) and coils (columns). Axial slices are shown at the level of the anterior (A) and posterior cerebellum (B).
Priovoulos N, Roos T, Ipek Ö, et al. A local multi‐transmit coil combined with a high‐density receive array for cerebellar fMRI at 7 T. NMR in Biomedicine. 2021;34(11):e4586. 10.1002/nbm.4586
Funding information Nederlandse Organisatie voor Wetenschappelijk Onderzoek, Grant/Award Number: VIDI 198.016
DATA AVAILABILITY STATEMENT
The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.
REFERENCES
- 1. Herculano‐Houzel S. The remarkable, yet not extraordinary, human brain as a scaled‐up primate brain and its associated cost. Proc Natl Acad Sci U S A. 2012;109(Suppl 1):10661‐10668. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Stoodley CJ, Schmahmann JD. Functional topography of the human cerebellum. Handb Clin Neurol. 2018;154:59‐70. [DOI] [PubMed] [Google Scholar]
- 3. Fartaria MJ, O'Brien K, Şorega A, et al. An ultra‐high field study of cerebellar pathology in early relapsing‐remitting multiple sclerosis using MP2RAGE. Invest Radiol. 2017;52(5):265‐273. [DOI] [PubMed] [Google Scholar]
- 4. Deistung A, Stefanescu MR, Ernst TM, et al. Structural and functional magnetic resonance imaging of the cerebellum: considerations for assessing cerebellar ataxias. Cerebellum. 2016;15(1):21‐25. [DOI] [PubMed] [Google Scholar]
- 5. van der Zwaag W, Francis S, Head K, et al. fMRI at 1.5, 3 and 7 T: characterising BOLD signal changes. Neuroimage. 2009;47(4):1425‐1434. [DOI] [PubMed] [Google Scholar]
- 6. Budde J, Shajan G, Hoffmann J, Uğurbil K, Pohmann R. Human imaging at 9.4 T using T2*‐, phase‐, and susceptibility‐weighted contrast. Magn Reson Med. 2011;65(2):544‐550. [DOI] [PubMed] [Google Scholar]
- 7. Vaidya MV, Lazar M, Deniz CM, et al. Improved detection of fMRI activation in the cerebellum at 7T with dielectric pads extending the imaging region of a commercial head coil. J Magn Reson Imaging. 2018;48(2):431‐440. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Padormo F, Beqiri A, Hajnal JV, Malik SJ. Parallel transmission for ultrahigh‐field imaging. NMR Biomed. 2016;29(9):1145‐1161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Ipek Ö. Radio‐frequency coils for ultra‐high field magnetic resonance. Anal Biochem. 2017;529:10‐16. [DOI] [PubMed] [Google Scholar]
- 10. Sengupta S, Roebroeck A, Kemper VG, et al. A specialized multi‐transmit head coil for high resolution fMRI of the human visual cortex at 7T. PLoS ONE. 2016;11(12):e0165418. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Clément JD, Gruetter R, Ipek Ö. A human cerebral and cerebellar 8‐channel transceive RF dipole coil array at 7T. Magn Reson Med. 2019;81(2):1447‐1458. [DOI] [PubMed] [Google Scholar]
- 12. Hayes CE, Axel L. Noise performance of surface coils for magnetic resonance imaging at 1.5 T. Med Phys. 1985;12(5):604‐607. [DOI] [PubMed] [Google Scholar]
- 13. Wald LL, Carvajal L, Moyher SE, et al. Phased array detectors and an automated intensity‐correction algorithm for high‐resolution MR imaging of the human brain. Magn Reson Med. 1995;34(3):433‐439. [DOI] [PubMed] [Google Scholar]
- 14. Wright SM, Wald LL. Theory and application of array coils in MR spectroscopy. NMR Biomed. 1997;10(8):394‐410. [DOI] [PubMed] [Google Scholar]
- 15. Petridou N, Italiaander M, van de Bank BL , Siero JCW, Luijten PR, Klomp DWJ. Pushing the limits of high‐resolution functional MRI using a simple high‐density multi‐element coil design. NMR Biomed. 2013;26(1):65‐73. [DOI] [PubMed] [Google Scholar]
- 16. Clément J, Gruetter R, Ipek Ö. A combined 32‐channel receive‐loops/8‐channel transmit‐dipoles coil array for whole‐brain MR imaging at 7T. Magn Reson Med. 2019;82(3):1229‐1241. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Movahedian Attar F, Kirilina E, Haenelt D, et al. Mapping short association fibers in the early cortical visual processing stream using in vivo diffusion tractography. Cereb Cortex. 2020;30(8):4496‐4514. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Frass‐Kriegl R, Navarro de Lara LI, Pichler M, et al. Flexible 23‐channel coil array for high‐resolution magnetic resonance imaging at 3 Tesla. PLoS ONE. 2018;13(11):e0206963. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Golay X, de Zwart JA, Ho YC, Sitoh YY. Parallel imaging techniques in functional MRI. Top Magn Reson Imaging. 2004;15(4):255‐265. [DOI] [PubMed] [Google Scholar]
- 20. Batson MA, Petridou N, Klomp DW, Frens MA, Neggers SF. Single session imaging of cerebellum at 7 Tesla: obtaining structure and function of multiple motor subsystems in individual subjects. PLoS ONE. 2015;10(8):e0134933. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Wiggins GC, Triantafyllou C, Potthast A, Reykowski A, Nittka M, Wald LL. 32‐channel 3 Tesla receive‐only phased‐array head coil with soccer‐ball element geometry. Magn Reson Med. 2006;56(1):216‐223. [DOI] [PubMed] [Google Scholar]
- 22. Raaijmakers AJ, Italiaander M, Voogt IJ, et al. The fractionated dipole antenna: a new antenna for body imaging at 7 Tesla. Magn Reson Med. 2016;75(3):1366‐1374. [DOI] [PubMed] [Google Scholar]
- 23. Steensma B, van de Moortele P‐F, Ertürk A, et al. Introduction of the snake antenna array: geometry optimization of a sinusoidal dipole antenna for 10.5T body imaging with lower peak SAR. Magn Reson Med. 2020;84(5):2885‐2896. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Meliado EF, Sbrizzi A, van den Berg CAT, Luijten PR, Raaijmakers AJE. Real‐time assessment of potential peak local specific absorption rate value without phase monitoring: Trigonometric maximization method for worst‐case local specific absorption rate determination. Magn Reson Med. 2021;85(6):3420–3433. 10.1002/mrm.28635 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Meliado EF, van den Berg CAT, Luijten PR, Raaijmakers AJE. Intersubject specific absorption rate variability analysis through construction of 23 realistic body models for prostate imaging at 7T. Magn Reson Med. 2019;81(3):2106‐2119. [DOI] [PubMed] [Google Scholar]
- 26. Hendriks AD, Luijten PR, Klomp DWJ, Petridou N. Potential acceleration performance of a 256‐channel whole‐brain receive array at 7 T. Magn Reson Med. 2019;81(3):1659‐1670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Christ A, Kainz W, Hahn EG, et al. The Virtual Family—development of surface‐based anatomical models of two adults and two children for dosimetric simulations. Phys Med Biol. 2010;55(2):N23‐N38. [DOI] [PubMed] [Google Scholar]
- 28. Gosselin MC, Neufeld E, Moser H, et al. Development of a new generation of high‐resolution anatomical models for medical device evaluation: the Virtual Population 3.0. Phys Med Biol. 2014;59(18):5287‐5303. [DOI] [PubMed] [Google Scholar]
- 29. Ledden P, Mareyam A, Wang S, Van Gelderen P, Duyn J. 32 channel receive‐only SENSE array for brain imaging at 7T. Paper presented at: Joint Annual Meeting ISMRM‐ESMRMB; May 19‐25, 2007; Berlin, Germany.
- 30. Nehrke K, Bornert P. DREAM—a novel approach for robust, ultrafast, multislice B 1 mapping. Magn Reson Med. 2012;68(5):1517‐1526. [DOI] [PubMed] [Google Scholar]
- 31. Setsompop K, Alagappan V, Gagoski B, et al. Slice‐selective RF pulses for in vivo B 1 + inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16‐element coil. Magn Reson Med. 2008;60(6):1422‐1432. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Boillat Y, van der Zwaag W. Whole brain measurements of the positive BOLD response variability during a finger tapping task at 7 T show regional differences in its profiles. Magn Reson Med. 2019;81(4):2720‐2727. [DOI] [PubMed] [Google Scholar]
- 33. Wiestler T, McGonigle DJ, Diedrichsen J. Integration of sensory and motor representations of single fingers in the human cerebellum. J Neurophysiol. 2011;105(6):3042‐3053. [DOI] [PubMed] [Google Scholar]
- 34. Woolrich MW, Jbabdi S, Patenaude B, et al. Bayesian analysis of neuroimaging data in FSL. Neuroimage. 2009;45(1 Suppl):S173‐S186. [DOI] [PubMed] [Google Scholar]
- 35. Jenkinson M, Bannister P, Brady M, Smith S. Improved optimization for the robust and accurate linear registration and motion correction of brain images. Neuroimage. 2002;17(2):825‐841. [DOI] [PubMed] [Google Scholar]
- 36. Andersson JL, Skare S, Ashburner J. How to correct susceptibility distortions in spin‐echo echo‐planar images: application to diffusion tensor imaging. Neuroimage. 2003;20(2):870‐888. [DOI] [PubMed] [Google Scholar]
- 37. Avants BB, Tustison NJ, Song G, Cook PA, Klein A, Gee JC. A reproducible evaluation of ANTs similarity metric performance in brain image registration. Neuroimage. 2011;54(3):2033‐2044. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Klein A, Andersson J, Ardekani BA, et al. Evaluation of 14 nonlinear deformation algorithms applied to human brain MRI registration. Neuroimage. 2009;46(3):786‐802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Diedrichsen J, Balsters JH, Flavell J, Cussans E, Ramnani N. A probabilistic MR atlas of the human cerebellum. Neuroimage. 2009;46(1):39‐46. [DOI] [PubMed] [Google Scholar]
- 40. Power JD, Barnes KA, Snyder AZ, Schlaggar BL, Petersen SE. Spurious but systematic correlations in functional connectivity MRI networks arise from subject motion. Neuroimage. 2012;59(3):2142‐2154. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Keltner JR, Carlson JW, Roos MS, Wong ST, Wong TL, Budinger TF. Electromagnetic fields of surface coil in vivo NMR at high frequencies. Magn Reson Med. 1991;22(2):467‐480. [DOI] [PubMed] [Google Scholar]
- 42. Kumar A, Edelstein WA, Bottomley PA. Noise figure limits for circular loop MR coils. Magn Reson Med. 2009;61(5):1201‐1209. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Fracasso A, Luijten PR, Dumoulin SO, Petridou N. Laminar imaging of positive and negative BOLD in human visual cortex at 7T. Neuroimage. 2018;164:100‐111. [DOI] [PubMed] [Google Scholar]
- 44. Hendriks AD, D'Agata F, Raimondo L, et al. Pushing functional MRI spatial and temporal resolution further: high‐density receive arrays combined with shot‐selective 2D CAIPIRINHA for 3D echo‐planar imaging at 7 T. NMR Biomed. 2020;33(5):3420–3433. 10.1002/nbm.4281 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Weaver JB. Simultaneous multislice acquisition of MR images. Magn Reson Med. 1988;8(3):275‐284. [DOI] [PubMed] [Google Scholar]
- 46. Poser BA, Koopmans PJ, Witzel T, Wald LL, Barth M. Three dimensional echo‐planar imaging at 7 Tesla. Neuroimage. 2010;51(1):261‐266. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. de Zwart JA, Ledden PJ, Kellman P, van Gelderen P, Duyn JH. Design of a SENSE‐optimized high‐sensitivity MRI receive coil for brain imaging. Magn Reson Med. 2002;47(6):1218‐1227. [DOI] [PubMed] [Google Scholar]
- 48. Breuer FA, Blaimer M, Heidemann RM, Mueller MF, Griswold MA, Jakob PM. Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi‐slice imaging. Magn Reson Med. 2005;53(3):684‐691. [DOI] [PubMed] [Google Scholar]
- 49. Stirnberg R, Stöcker T. Segmented K‐space blipped‐controlled aliasing in parallel imaging for high spatiotemporal resolution EPI. Magn Reson Med. 2021;85(3):1540–1551. 10.1002/mrm.28486 [DOI] [PubMed] [Google Scholar]
- 50. Wiggins GC, Polimeni JR, Potthast A, Schmitt M, Alagappan V, Wald LL. 96‐channel receive‐only head coil for 3 tesla: design optimization and evaluation. Magn Reson Med. 2009;62(3):754‐762. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Van de Moortele PF, Akgun C, Adriany G, et al. B 1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil. Magn Reson Med. 2005;54(6):1503‐1518. [DOI] [PubMed] [Google Scholar]
- 52. Zhu Y. Parallel excitation with an array of transmit coils. Magn Reson Med. 2004;51(4):775‐784. [DOI] [PubMed] [Google Scholar]
- 53. Cloos MA, Boulant N, Luong M, et al. kT‐points: short three‐dimensional tailored RF pulses for flip‐angle homogenization over an extended volume. Magn Reson Med. 2012;67(1):72‐80. [DOI] [PubMed] [Google Scholar]
- 54. Gras V, Poser BA, Wu X, Tomi‐Tricot R, Boulant N. Optimizing BOLD sensitivity in the 7T Human Connectome Project resting‐state fMRI protocol using plug‐and‐play parallel transmission. Neuroimage. 2019;195:1‐10. [DOI] [PubMed] [Google Scholar]
- 55. Roos K, van der Zwaag W. Improving 7T brain scans for anatomical and functional modalities by reducing B1 inhomogeneity; pads or universal pulses? Paper presented at: ESMRMB 2019; October 3‐5, 2019; Rotterdam, The Netherlands.
- 56. Mountcastle VB. The columnar organization of the neocortex. Brain. 1997;120(4):701‐722. [DOI] [PubMed] [Google Scholar]
- 57. King M, Hernandez‐Castillo CR, Poldrack RA, Ivry RB, Diedrichsen J. Functional boundaries in the human cerebellum revealed by a multi‐domain task battery. Nat Neurosci. 2019;22(8):1371‐1378. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. van Es DM, van der Zwaag W, Knapen T. Topographic maps of visual space in the human cerebellum. Curr Biol. 2019;29(10):1689, e1683‐1694. [DOI] [PubMed] [Google Scholar]
- 59. van der Zwaag W, Kusters R, Magill A, et al. Digit somatotopy in the human cerebellum: a 7T fMRI study. Neuroimage. 2013;67:354‐362. [DOI] [PubMed] [Google Scholar]
- 60. Boillat Y, Bazin PL, O'Brien K, et al. Surface‐based characteristics of the cerebellar cortex visualized with ultra‐high field MRI. Neuroimage. 2018;172:1‐8. [DOI] [PubMed] [Google Scholar]
- 61. Boillat Y, Bazin PL, van der Zwaag W. Whole‐body somatotopic maps in the cerebellum revealed with 7T fMRI. Neuroimage. 2020;211. [DOI] [PubMed] [Google Scholar]
- 62. Marques JP, van der Zwaag W, Granziera C, Krueger G, Gruetter R. Cerebellar cortical layers: in vivo visualization with structural high‐field‐strength MR imaging. Radiology. 2010;254(3):942‐948. [DOI] [PubMed] [Google Scholar]
- 63. van der Zwaag W, Jorge J, Butticaz D, Gruetter R. Physiological noise in human cerebellar fMRI. Magn Reson Mater Phys Biol Med. 2015;28(5):485‐492. [DOI] [PubMed] [Google Scholar]
- 64. Triantafyllou C, Hoge RD, Krueger G, et al. Comparison of physiological noise at 1.5 T, 3 T and 7 T and optimization of fMRI acquisition parameters. Neuroimage. 2005;26(1):243‐250. [DOI] [PubMed] [Google Scholar]
- 65. Bazin PL, Alkemade A, van der Zwaag W, Caan M, Mulder M, Forstmann BU. Denoising high‐field multi‐dimensional MRI with local complex PCA. Front Neurosci. 2019;13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66. Caan MWA, Bazin PL, Marques JP, de Hollander G, Dumoulin SO, van der Zwaag W. MP2RAGEME: T1, T2*, and QSM mapping in one sequence at 7 tesla. Hum Brain Mapp. 2019;40(6):1786‐1798. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67. Wald LL. The future of acquisition speed, coverage, sensitivity, and resolution. Neuroimage. 2012;62(2):1221‐1229. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Figure S1. Maximum intensity projections of B1 normalized for 2 W of input power. A‐B, Snake antenna array (no shim/shimmed). C‐D, Loop coil array (no shim/shimmed). Note that the snake antenna provides a more homogeneous transmit profile over the cerebellum (white boxes).
Figure S2. Maximum intensity projections of SAR10g for the Duke model (Sims4Life). A, Worst‐case phase shim cerebellar transmit (maximum SAR10g = 1.68 W/kg/W of input power), B, Phase shimmed for maximum B1 cerebellar transmit (maximum SAR10g = 1.58 W/kg/W of input power).
Figure S3. A, Successive slices of the receive array detuning ratio (flip angle map with the surface receive array present divided by the flip angle map with the surface receive array replaced with a pad). The ratio was calculated over the area with sufficient signal in the DREAM images and is overlaid as a colormap over the second magnitude image of the DREAM acquisition. B, example slices of the magnitude 1 and 2 of the DREAM acquisition. C, Examples slices of the resulting flip angle maps with and without the receive array present. D, Histogram of the detuning ratio values within the ROI (colormap).
Figure S4. Example sagittal, coronal and axial slices (columns) of the EPI for each coil. A, Cerebellar coil. B, Head coil unshimmed. C, Head coil shimmed.
Figure S5. B1 + normalized SNR maps for one participant. A, coils and their difference (columns) along axial slices at the height of the cerebellum (rows). B, SNR distribution within the cerebellum for the different coil setups. C, SNR plotted against the distance from the skull.
Figure S6. (Finger tapping > Rest) BOLD fMRI significant clusters (z > 3.1) for all participants (rows) and coils (columns). Axial slices are shown at the level of the anterior (A) and posterior cerebellum (B).
Data Availability Statement
The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.