Skip to main content
Polymers logoLink to Polymers
. 2021 Oct 15;13(20):3564. doi: 10.3390/polym13203564

Electro-Optical Biosensor Based on Embedded Double-Monolayer of Graphene Capacitor in Polymer Technology

Ary V R Portes 1, Ana J L Martins 1, Jesus Alvarez Guerrero 2, Mauricio M Carvalho 3, Ferney O Amaya-Fernandez 4, Lúcia A M Saito 3, Jhonattan C Ramirez 1,*
Editor: Naseem Abbas
PMCID: PMC8537356  PMID: 34685322

Abstract

In this work, we present an interferometric polymer-based electro-optical device, integrated with an embedded double-monolayer graphene capacitor for biosensing applications. An external voltage across the capacitor applies an electric field to the graphene layers modifying their surface charge density and the Fermi level position in these layers. This in turn changes the electro-optic properties of the graphene layers making absorption in the waveguide tunable with external voltages. Simultaneously, it is possible to appreciate that this phenomenon contributes to the maximization of the light-graphene interaction by evanescent wave in the sensing area. As a result, it is obtained large phase changes at the output of the interferometer, as a function of small variations in the refractive index in the cladding area, which significantly increasing the sensitivity of the device. The optimum interaction length obtained was 1.24 cm considering a cladding refractive index of 1.33. An absorption change of 129 dB/mm was demonstrated. This result combined with the photonic device based on polymer technology may enable a low-cost solution for biosensing applications in Point of Care (PoC) platform.

Keywords: polymeric photonic biosensors, electro-optical components, graphene-based capacitor, Fermi level control

1. Introduction

Graphene has gained considerable attention when its unique properties have been reported in 2004 [1,2]. Since then, graphene has become a very versatile material due to its excellent electrical, optical and mechanical properties, and has been implemented in a wide range of applications [3,4,5,6]. These properties include widely tunable conductivity, dielectric constant, and real and imaginary parts of refractive index and enable electro-optic (EO) modulation [7,8,9,10].

Graphene can provide a high and very broadband optical absorption extending from visible to infrared range [11]. This absorption can be tuned over a very wide range by tuning the Fermi level using externally applied electric fields [12,13]. It also has very high carrier mobility, which can enable high speed operation. Graphene can also be incorporated into CMOS technology, enabling low cost and high functionality applications [14]. More recently, and due to the electrical characteristics previously described, graphene has been used for biosensing applications, where its high conductivity has been exploited for the generation of surface plasmon resonance [15,16,17], or for obtaining conventional electrochemical transducers, due to the high electron mobility that this material provides [18,19,20,21,22]. Additionally, because it is an organic material, it has been shown to be beneficial for this type of application, especially in its oxidized form [23,24], since conventional graphene, depending on the substrate used, can be perceived as hydrophobic [25], which can negatively impact.

Some features such as immunity to electromagnetic interference, high-speed operation, low power consumption, potential use in harsh environments, miniaturization, integration and multiplexing capability, mechanical stability, low-cost for fabrication, and real-time and label-less detection [26,27,28], have provided photonic biosensors a place of prominence before the scientific community, and not only thanks to the benefits previously listed, but also because this type of biosensors have demonstrated the best performance among all types of existing biosensors [29].

Most of the integrated photonic biosensors are based on silicon, due to the excellent optical characteristics that these materials present, demonstrating high sensitivities and a high reproducibility rate. However, despite all the benefits that silicon has, effects such as Two-Photon Absorption (TPA) and Free-Carrier Dispersion (FCD) can become evident, which would compromise the performance of said biosensor [30]. Therefore, devices based on polymeric materials are presented as a viable alternative, which despite having very low refractive indices and having shown low sensitivity, efforts are made by the scientific community to improve this scenario, compensating for the unfavorable optical characteristics of the material, with an efficient and reliable design [31,32,33]; our proposal goes in this direction. On the other hand, characteristics such as the biocompatibility of these materials enable simpler and more durable biofunctionalization processes, and its low production cost provides conditions for these devices to be massively produced for single-use [34], these are points to keep in mind when justifying its implementation.

The implementation of polymeric photonic devices, integrated with a graphene-based capacitive structure in a single device, will provide the conditions for obtaining an electro-optical biosensor with a wide evanescent field in the sensing area, and with low loss due to tunable absorption based on Fermi level control.

In this paper, we present a low-loss electro-optical polymer-based device, with an embedded double-layer graphene capacitor for biosensing applications. Our proposal demonstrates through theoretical analysis and numerical simulations the widening of the evanescent field and consequently, the increase in the sensitivity of the proposed device, due to the interaction generated between the electric field in the capacitive structure and the confined optical waveform. In this sense, a bulk sensitivity of 2828 nm/RIU and a surface sensitivity of 2021.85 nm/RIU were demonstrated for the proposed integrated photonic biosensor, both working with a gate voltage of Vg = 20 V, being RIU the (Refractive Index Unit).

2. Simulation Modeling

An efficient use of graphene in these applications requires geometries that provide the correct interaction between the applied electrical fields in the graphene layers [35], and the optical waveform, for this reason, the need to obtain an optimized device for the proposed application is essential.

2.1. Materials and Methods

In this study, the Finite Element Method (FEM) was used for analyzing the sensor’s behavior by using commercial software COMSOL Multiphysics. The refractive index used in the core of the waveguide was 1.57 @ 800 nm wavelength, in the substrate it was 1.45 @ 800 nm wavelength, in the cladding of the sensor area it was varied between 1.33 and 1.38, and in the PMMA cladding it was 1.49 @ 800 nm.

Our simulations were performed in two stages, first, the interaction between the projected waveguide and the graphene-based capacitor was analyzed using modal analysis. The number of elements used for this stage was 40,437. In addition, two physics were implemented to simulate the device. The first one, electrostatics, which is responsible for the generation of the surface current, through a potential applied on the ohmic contacts, as can be seen in Figure 1a,c. In this sense, the Gate terminal has a variable voltage Vg, while source terminal has a constant voltage of Vs = 1 V and both Drain and Ground terminals has a constant voltage of 0 V.

Figure 1.

Figure 1

Design of the proposed polymeric sensor. (a) 3D view of a short section in the sensing area, where the interaction between the graphene-based capacitor and the polymeric waveguide, is presented. (b) Schematic of Mach-Zehnder interferometer biosensor. (c) Cross-section view of the sensing area, where the proposed device configuration is shown. (d) Mode distribution in SU-8 waveguide, interacting with the capacitive structure.

In addition, the electromagnetic waves physics was implemented to generate the modal distribution at λ = 633 nm, λ = 800 nm, and λ = 1.2 μm for fundamental TE and TM modes, that would interact with the capacitive structure, whose composition based on graphene was modeled as a surface current boundary condition, which follows the Equation (1):

j=σ·E (1)

where j is the surface current density, σ is graphene conductivity and E is the electric field applied in the region, both by the voltage terminals and by the electromagnetic waves, demonstrating the perturbation generated to the optical mode by externally applied electric fields.

Similarly, in the second stage we use the same electromagnetic waves physics with 92,174 elements, to analyze the spectral response of the projected MZI with all the elements previously mentioned, presenting the expected phase change in the proposed photonic device, as a function of small variations in the sensing area.

2.2. Design of the Graphene-Based Capacitive Structure

The proposed simulation model is composed of a Mach Zehnder Interferometer (MZI) based on SU-8 polymer with 700 nm width and 450 nm height, on a Silicon Dioxide (SiO2) substrate, as illustrated in Figure 1a,b. The reference arm is unaltered, maintaining a PMMA cladding, on the other hand, the cladding in the sensor area is water, and a capacitive structure based on two monolayers of graphene separated by 5 nm of SiO2, was embedded below the polymer waveguide, as detailed in Figure 1c. In addition, each monolayer of graphene has a couple of ohmic contact gold terminals, Gate and Ground for the upper layer and Source and Drain for the lower layer. By applying an electrical voltage to these terminals, it is possible to control the Fermi level in this capacitor, modifying the optical mode in the waveguide, as presented in Figure 1d.

The integration of waveguides and graphene-based capacitors is a technology that has been widely exploited for electro-optical modulators, due to the tunable absorption of the graphene [8,9,36,37]. However, by using the same principle of operation for controlling graphene optical properties, it is possible to optimize the device, significantly improving the performance of biosensor components. Our hypothesis is based on the fact that the same electric field that controls the fermi energy in graphene, generates changes in the optical mode propagated inside of the optimized optical waveguide, altering the orientation and intensity of the electric field lines increasing the penetration depth of the evanescent field in the sensing area. Thus, it is possible to appreciate the dependence that exists of the permittivity of graphene with the variation of the Fermi energy (Ef), as presented in Equation (2).

ϵ(ω,Ef)=1iσ(ω,Ef)ωϵ0δ (2)

where δ is the graphene thickness, ω is the angular wave frequency, ϵ0 is the vacuum permittivity and σ is the graphene conductivity, which is a function of the angular wave frequency and the Fermi level [38]. By using the capacitive structure, graphene Fermi level and the applied voltage can be related by Equation (3).

μc=vfπη|Vg| (3)

where vf is the Fermi velocity and Vg is the gate voltage. The variable η is derived from the parallel-plate capacitor model, and it is defined by Equation (4), [39]:

η=ϵ0ϵrde (4)

where ϵr is the relative permittivity, d is the dielectric thickness between the capacitor plates and e is the electron charge. By using the Equation (4), the graphene conductivity was calculated for three different operating wavelengths, as presented in Figure 2. It is possible to observe in that figure that for the analyzed operating wavelengths, the conductivity of the graphene can be tuned from 6×105 S down to almost 0 S, setting 1 V in the Source-Drain interface, and increasing the Gate Voltage (Vg) from 0 V up to 40 V, consequently changing the permittivity of the graphene as well.

Figure 2.

Figure 2

Graphene conductivity response as a function of Vg applied voltage for 1.2 μm, 800 nm and 633 nm.

In Figure 2, it was shown that higher wavelengths have the conductivity transition at lower voltages, which is a key factor to reduce the power required. In order to reduce thermal effects in the polymeric device due to the high power required to tune the conductivity in the capacitive structure, and at the same time avoid the absorption of biological tissues in the infrared spectrum, we selected 800 nm as the operating wavelength for further studies [40]. It is important to highlight that some biological processes, especially related to imaging, are performed in IR wavelengths, however, the integrated photonic components are used mostly to carry out detection through antibody-antigen binding or, failing that, using DNA, RNA, etc. In these cases, the target protein is specifically detected on the device in the sensing area without the need for markers (label-free detection of diseases), this is our case, and when absorption by the analyzed protein is evidenced, we will have an abrupt reduction of the propagating light in the waveguide, along the sensing area, disabling the operation of the proposed interferometer, since the only contribution perceived at the output will be that of the control arm, this is observed in systems working in IR wavelengths.

According to Equations (2) and (3), it is possible to control the permittivity and the optical losses in graphene, modifying the confined optical mode collaterally, as can be seen in Figure 3. The electric field distribution in the TE and TM modes, were evaluated for 0 V and 20 V at the Vg, respectively. In this study it was observed that when applying 20 V in the Vg, the propagating TE mode shows a greater penetration depth of the evanescent wave in the cladding, reducing the interaction with the substrate previously identified in 0 V. This is evidenced by the reduction in the effective refractive index achieved. In the case of TM mode, in addition to the previously described changes, we have a slight modification in the orientation of the electric field lines, because of the applied voltage.

Figure 3.

Figure 3

Electric field distribution at λ = 800 nm, with and without the impact generated by the variation of the Fermi level when Vg=20 V. In this figure it is possible to appreciate changes in the calculated field lines for the TE and TM modes, when 20 V at the gate is applied and a constant voltage of 1 V is maintained between source and drain, however, only the TE mode presents significant variation in the effective refractive index in the presence of the surface current generated by the electric field applied to the proposed capacitor.

Next, we calculated the real and imaginary parts of the effective refractive index for the TE and TM modes, in order to choose the most suitable mode of operation for our study. In Figure 4a,b, we can see these results. As expected, the impact generated due to the voltage variation in Vg is much greater in the propagation of the TE mode, which was demonstrated through complementary simulations, where the variation of the effective refractive index, as a function of small structural changes, was of the order of 103 in favor of it, which would strongly impact, when we talk about applications in biosensors. The results obtained through our numerical simulations can be seen in Figure 4.

Figure 4.

Figure 4

Variation of the (a) real and (b) imaginary parts of the effective refractive index for TE and TM modes as a function of the gate voltage (Vg) variation.

2.3. Design of the Polymer-Based Component

Achieving the best evanescent field in the sensing area while keeping the optical mode as confined as possible in the reference arm, is essential, Equation (5). In this sense, we computationally guided our photonic component, analyzing the effective area as a function of the height of the waveguide, having as a limitation the number of modes that we want to propagate and the minimum width that we can define for reasons of manufacturing restrictions. The results obtained can be seen in Figure 5.

Aeff=(|E(x,y)|2dxdy)2|E(x,y)|4dxdy (5)

Figure 5.

Figure 5

(a) Study of the behavior of the evanescent field, evaluating the effective area as a function of the height of the waveguide. (b) Analysis of the real part of the effective refractive index as a function of the height variation, for cladding values at 1.33 and 1.34, and the Δneff as a function of the reference arm.

Considering the Equation (5), and setting the width of the device in 700 nm after being optimized, based on manufacturing restrictions and required fundamental mode propagation, in Figure 5 the height of the device was varied between 400 nm and 800 nm in order to define the height at which the proposed device enhances its efficiency.

The expected result should guarantee a wide evanescent field in the sensing area allowing the largest possible Δneff as a function of small variations in the refractive index in the cladding. In Figure 5a, it is possible to observe that the minimum effective area is achieved with a device with a height of 600 nm and the maximum one at 400 nm. This parabolic behavior is because in the region comprised between 400 nm and 600 nm height, the waveguide has a smaller size than the confined mode, for this reason as the height of the device is increased the evanescent field decreases, reducing the effective area. On the other hand, when the height is greater than 600 nm, we observe the increase of the confined optical mode accompanying the growth of the waveguide. According to the previous discussion, and the results presented in Figure 5b, where the Δneff is shown as a function of small variations of the refractive index in the cladding, i.e., variations between 1.33 and 1.34, we can affirm that the most suitable height for our component is 450 nm, since a long penetration depth of the evanescent wave and the largest possible Δneff for the small variation induced in the refractive index of the cladding, is guaranteed.

Previously, the impact of the capacitive structure in the propagated optical mode was observed, in this section it is necessary to define the thickness d of dielectric layer of the capacitor in order to tune the Fermi energy of the graphene with low power, maximizing the sensitivity of the proposed sensor. As described by Equations (3) and (4), Vg is directly proportional to η, so in order to reduce the applied voltage, smaller thicknesses are desired, as shown in Figure 6.

Figure 6.

Figure 6

(a) Real part of the effective refractive index as a function of the Vg for 5 nm and 20 nm thickness in the dielectric between the two monolayers of graphene. (b) Analysis at Vg=20 V of the real part of the effective refractive index as a function of the d variation, for cladding values at 1.33 and 1.34, and the Δneff as a function of the reference arm. (c) Analysis at d=5 nm of the real part of the effective refractive index as a function of the Vg variation, for cladding values at 1.33 and 1.34, and the Δneff as a function of the reference arm. (d) Attenuation as a function of the Fermi level variation, for TE and TM propagation modes.

By reducing d from 20 nm to 5 nm thickness, it is possible to observe a shift in the maximum value reached by the effective refractive index from 44 V to 11 V, observing a reduction in the power required to tune the optical absorption of the graphene in approximately 40 V. This can be seen in Figure 6a. Additionally and according to Figure 6b, lower thicknesses can provide also an increase in the refractive index variation, demonstrating the highest Δneff at 1.405×103 for d=5 nm. For this value of thickness it was obtained a surface charge density of 1.1fF/m.

Similarly, we analyzed the biosensor response as a function of the Vg variation in Figure 6c. We demonstrated that when the real part of the effective refractive index is maximum, the Δneff calculated for small variations of the refractive index in the cladding, i.e., variations between 1.33 and 1.34 is minimum. In addition, we observed that increasing the Vg further than 11 V the Δneff also increase. A Vg at 20 V guarantees a good variation of the Δneff for biosensing applications, and banish the possibility of polymeric deformation due to the high temperatures that can arise from the implementation of high voltage levels.

We reported, by simulations, that tuning the Fermi energy level of the graphene enables the electro-optical modification in the sensor, according to Equation (3). The absorption of TE mode change from 143.14 dB/mm at 0.1 eV to 13.83 dB/mm for 1.25 eV such values demonstrate the high interaction of the evanescent wave in the sensing area contributing to optimize the sensibility of the sensor, as presented in Figure 6d.

3. Results and Discussion

To evaluate the performance of the proposed sensor, we conducted simulations in frequency domain, evaluating the complete layout using Finite Element Method (FEM) in COMSOL Multiphysics.

The effective refractive indices calculated in the previous section for the corresponding sensing area and for the complementary section of the MZI, were incorporated in this simulation system, considering the variation that the refractive index presents between 400 nm and 800 nm wavelengths.

In order to optimize the total length of the sensing area (Lsens), we simulated the transmission response of the sensor as a function of the Lsens, evaluating the spectral response for Lsens ranging from 100 μm up to 1.8 cm length. In Figure 7a, it is possible to appreciate the maximum phase shift when Lsens is 1.24 cm, considering 1.33 refractive index in the cladding. In accordance with the above, in Figure 7b the phase shift was verified as a function of the variation of the wavelength between 400 nm and 900 nm, observing an excellent result at 800 nm wavelength, in accordance with the results discussed in the previous section.

Figure 7.

Figure 7

Propagation analysis in the proposed MZI biosensor, based on double monolayer of graphene capacitor. (a) Transmission response where the phase shift is evaluated as a function of the variation of the length of the sensing area (Lsens) at 800 nm wavelength and 1.33 refractive index in the cladding (b) Transmission response where the phase shift is evaluated as a function of the wavelength variation between 400 nm and 900 nm, for 1.24 cm length in the sensing area. (c) Close-view of the redshifted normalized transmission in the region where the maximum phase shift occurs, for refractive indices between 1.33 and 1.38 with a step of 0.01 RIU. (d) Sensitivity obtained through a linear regression calculated on the simulations carried out in (c).

Setting the total length of the sensing area at 1.24 cm, and propagation wavelength at 800 nm, we analyzed the perceived transmittance at the output of the component for different values of refractive index in the cladding, varying from 1.33 up to 1.38 with a step of 0.01 RIU. According to Figure 7c, it is possible to observe that despite maintaining the phase shift at close wavelengths, the resonance wavelength shows a redshift as a function of the increase in the refractive index in the sensing area. This can mean that small variations in bulk can be detected with high precision due to the high sensitivity demonstrated, i.e., 2828 nm/RIU. The calculation of the sensitivity achieved can be seen in Figure 7d, where by applying a linear regression to the calculated points, it was possible to obtain from the curve slope a sensitivity of 2828 nm/RIU for the simulated device. As Table 1 shows, our sensor obtained improvements in relation to the sensitivity of similar biosensors, even in relation to silicon-based components.

Table 1.

Comparison between the proposed MZI biosensor based on double-layer graphene capacitor, with other optical biosensors presented to the scientific community, based on polymer or silicon technology.

Sensor Type Waveguide Material Mode Sensitivity
Ridge waveguide Polymer—PSQ-Ls TE 49.75 nm/RIU [41]
Grating-based waveguide Polymer—FSU-8 TE 1606.2 nm/RIU [42]
Microring resonator Polymer—ZPU13-430 TE 200 nm/RIU [43]
MZI Silicon Nitride TE 1864π/RIU [44]
Slot Waveguide Silicon Nitride TE 1730(2π)/RIU [45]
MZI Silicon TE 740 nm/RIU [46]
MZI Silicon TM 460(2π)/RIU [47]
MZI Silicon TE 300(2π)/RIU [48]
Photonic crystal Silicon TM 425 nm/RIU [49]
Microring resonator Silicon TM 579.5 nm/RIU [50]
Photonic crystal Silicon TE 404.11 nm/RIU [51]
Photonic crystal Silicon TE 300 nm/RIU [52]
MZI biosensor based on double-layer graphene capacitor Polymer—SU-8 TE 2828 nm/RIU [This work]

Finally, we carry out sensitivity studies on the surface, where a biolayer with a refractive index of 1.45 and 10 nm thick [32] was added to our model, in order to observe the performance of the polymeric transducer. The result observed for the sensitivity on the surface, i.e., 2021.85 nm/RIU at Vg = 20 V, demonstrate consistency with those presented in the bulk device, suffering a reduction of only 28% in comparison with the latter, which is the main differential of this technology. This decrease in sensitivity is due to the fact that the presence of the biological layer prevents a greater penetration depth of the evanescent field in the sensing area, reducing the intensity of the external field and consequently reducing the efficiency of the proposed sensor. However, and against the forecasts, the performance of our transducer shows a decrease much less than that presented in other types of devices, where it is observed that the reduction between the sensitivity obtained in bulk and surface, is around 40% for devices in silicon technology [32,46] and greater than 60% in polymer technology [32], limiting them in terms of Limit of Detection (LoD). This good performance is due to the interaction induced between the propagated optical mode and the electric field generated by the embedded capacitor, which shows significant improvements compared to previously proposed sensors.

4. Conclusions

In summary, a MZI sensor based on embedded double-monolayer graphene capacitor in polymer technology is presented and demonstrated. The embedded capacitive structure was modeled, guaranteeing the tuning of the Fermi energy level, aiming to reduce the optical absorption of the graphene, and simultaneously guaranteeing an increase in the evanescent field in the sensing area, due to the interaction induced between the optical mode propagated in the polymer waveguide and the electric field in the proposed capacitive structure. The optimization of the sensing area in the MZI allows increasing the sensitivity that can be achieved, because it allows strong confinement of the light in the general waveguide and a long penetration depth of the evanescent field in the sensing area. Our device was evaluated as a biosensor, presenting a bulk sensitivity of 2828 nm/RIU and a surface sensitivity of 2021.85 nm/RIU, both with a power Vg = 20 V, demonstrating a sensitivity greater than that achieved by similar devices available from experimental results in the scientific literature [41,42,43,44,45,46,47,48,49,50,51,52].

Due to the high sensitivity demonstrated and the materials used, these devices represent an important advance in obtaining Label-Free disposable optical biosensors for Lab on a Chip (LoC) platform, for the early detection of genetic and infectious diseases.

Author Contributions

A.V.R.P. Theoretical and numerical studies, data analysis, and writing of the manuscript. A.J.L.M. Numerical studies, data analysis, and writing of the manuscript. J.A.G. Conception of the idea and writing of the manuscript. M.M.C. Numerical studies, data analysis, and writing of the manuscript. F.O.A.-F. Conception of the idea, supervision, and writing of the manuscript. L.A.M.S. Conception of the idea, supervision and writing of the manuscript. J.C.R. Conception of the idea, supervision, theoretical and numerical studies, data analysis, and writing of the manuscript. All authors have read and agreed to the published version of the manuscript.

Funding

The authors would like to thank the financial support of CAPES (Print process 88887.310281/2018-00), MackPesquisa (201037), Pró-Reitoria de Pesquisa—Universidade Federal de Minas Gerais (UFMG) (03/2019), Processo nº 2018/08988-9 Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP), and the Ministry of Science, Technology and Innovation from Colombia for funding the doctoral project entitled “Diseño de transductores fotónicos usando materiales poliméricos y grafeno para aplicaciones de biosensado”.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available on request from the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Novoselov K.S. Electric field effect in atomically thin carbon films. Science. 2004;306:666–669. doi: 10.1126/science.1102896. [DOI] [PubMed] [Google Scholar]
  • 2.Geim A.K., Novoselov K.S. The rise of graphene. Nat. Mater. 2007;6:183–191. doi: 10.1038/nmat1849. [DOI] [PubMed] [Google Scholar]
  • 3.Schwierz F. Graphene transistors. Nat. Nanotechnol. 2010;5:487–496. doi: 10.1038/nnano.2010.89. [DOI] [PubMed] [Google Scholar]
  • 4.Bonaccorso F., Sun Z., Hasan T., Ferrari A. Graphene photonics and optoelectronics. Nat. Photonics. 2010;4:611–622. doi: 10.1038/nphoton.2010.186. [DOI] [Google Scholar]
  • 5.Liao L., Lin Y.C., Bao M., Cheng R., Bai J., Liu Y., Qu Y., Wang K.L., Huang Y., Duan X. High-speed graphene transistors with a self-aligned nanowire gate. Nature. 2010;467:305–308. doi: 10.1038/nature09405. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Avouris P., Chen Z.H., Perebeinos V. Carbon-based electronics. Nat. Nanotechnol. 2007;2:605–615. doi: 10.1038/nnano.2007.300. [DOI] [PubMed] [Google Scholar]
  • 7.Luo S., Wang Y., Tong X., Wang Z. Graphene -based optical modulators. Nanoscale Res. Lett. 2015;10:199. doi: 10.1186/s11671-015-0866-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Liu M., Yin X., Ulin-Avila E., Geng B., Zentgraf T., Ju L., Wang F., Zhang X. A graphene-based broadband optical modulator. Nat. Lett. 2011;474:64–67. doi: 10.1038/nature10067. [DOI] [PubMed] [Google Scholar]
  • 9.Sorianello V., Midrio M., Contestabile G., Asselberghs I., Campenhout J.V., Huyghebaert C., Goykhman I., Ott A.K., Ferrari A.C., Romagnoli M. Graphene-silicon phase modulators with gigahertz bandwidth. Nat. Photonics. 2018;12:40–44. doi: 10.1038/s41566-017-0071-6. [DOI] [Google Scholar]
  • 10.Carvalho M.M., Thoroh de Souza E.A., Saito L.A.M. Graphene-based PAM-4 modulator compatible with CMOS platform operating over DWDM C-Band. Results Opt. 2021;5:100110. doi: 10.1016/j.rio.2021.100110. [DOI] [Google Scholar]
  • 11.Thongrattanasiri S., Koppens F.H.L., de Abajo F.J.G. Complete Optical Absorption in Periodically Patterned Graphene. Phys. Rev. Lett. 2012;108:047401. doi: 10.1103/PhysRevLett.108.047401. [DOI] [PubMed] [Google Scholar]
  • 12.Wang F., Zhangchuanshan Y., Giritalex T., Crommie Z., Shen Y.R. Gate-variable optical transitions in graphene. Science. 2008;320:206–209. doi: 10.1126/science.1152793. [DOI] [PubMed] [Google Scholar]
  • 13.Li Z.Q., Henriksen E.A., Jiang Z., Hao Z., Martin M.C., Kim P., Stormer H.L., Basov D.N. Dirac charge dynamics in graphene by infrared spectroscopy. Nat. Phys. 2008;4:532–535. doi: 10.1038/nphys989. [DOI] [Google Scholar]
  • 14.Goossens S., Navickaite G., Monasterio C., Gupta S., Piqueras J.J., Pérez R., Burwell G., Nikitskiy I., Lasanta T., Galán T., et al. Broadband image sensor array based on graphene—CMOS integration. Nat. Photonics. 2017;11:366–371. doi: 10.1038/nphoton.2017.75. [DOI] [Google Scholar]
  • 15.Wu L., Chu H.S., Koh W.S., Li E.P. Highly sensitive graphene biosensors based on surface plasmon resonance. Opt. Express. 2010;18:14395. doi: 10.1364/OE.18.014395. [DOI] [PubMed] [Google Scholar]
  • 16.Xu S., Zhan J., Man B., Jiang S., Yue W., Gao S., Guo C., Liu H., Li Z., Wang J., et al. Real-time reliable determination of binding kinetics of DNA hybridization using a multi-channel graphene biosensor. Nat. Commun. 2017;8:1–10. doi: 10.1038/ncomms14902. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Alfonso F.S., Zhou Y., Liu E., McGuire A.F., Yang Y., Kantarci H., Li D., Copenhaver E., Zuchero J.B., Müller H., et al. Label-free optical detection of bioelectric potentials using electrochromic thin films. Proc. Natl. Acad. Sci. USA. 2020;117:17260–17268. doi: 10.1073/pnas.2002352117. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Afsahi S., Lerner M.B., Goldstein J.M., Lee J., Tang X., Bagarozzi D.A., Pan D., Locascio L., Walker A., Barron F., et al. Novel graphene-based biosensor for early detection of Zika virus infection. Biosens. Bioelectron. 2018;100:85–88. doi: 10.1016/j.bios.2017.08.051. [DOI] [PubMed] [Google Scholar]
  • 19.Seo G., Lee G., Kim M.J., Baek S.H., Choi M., Ku K.B., Lee C.S., Jun S., Park D., Kim H.G., et al. Rapid Detection of COVID-19 Causative Virus (SARS-CoV-2) in Human Nasopharyngeal Swab Specimens Using Field-Effect Transistor-Based Biosensor. ACS Nano. 2020;14:5135–5142. doi: 10.1021/acsnano.0c02823. [DOI] [PubMed] [Google Scholar]
  • 20.Li Z., Zhang W., Xing F. Graphene Optical Biosensors. Int. J. Mol. Sci. 2019;20:2461. doi: 10.3390/ijms20102461. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Horng J., Balch H.B., McGuire A.F., Tsai H.Z., Forrester P.R., Crommie M.F., Cui B., Wang F. Imaging electric field dynamics with graphene optoelectronics. Nat. Commun. 2016;7:13706. doi: 10.1038/ncomms13704. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Balch H.B., McGuire A.F., Horng J., Tsai H.Z., Qi K.K., Duh Y.S., Forrester P.R., Crommie M.F., Cui B., Wang F. Graphene Electric Field Sensor Enables Single Shot Label-Free Imaging of Bioelectric Potentials. Nano Lett. 2021;21:4944. doi: 10.1021/acs.nanolett.1c00543. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Morales-Narváez E., Merkoçi A. Graphene Oxide as an Optical Biosensing Platform: A Progress Report. Adv. Mater. 2019;31:1–12. doi: 10.1002/adma.201805043. [DOI] [PubMed] [Google Scholar]
  • 24.Salamon J., Sathishkumar Y., Ramach K., Lee Y.S., Yoo D.J., Kim A.R., Gnana-Kumar G. One-pot synthesis of magnetite nanorods/graphene composites and its catalytic activity toward electrochemical detection of dopamine. Biosens. Bioelectron. 2015;64:269–276. doi: 10.1016/j.bios.2014.08.085. [DOI] [PubMed] [Google Scholar]
  • 25.Feng J., Guo Z. Wettability of graphene: From influencing factors and reversible conversions to potential applications. Nanoscale Horizons. 2019;4:526–530. doi: 10.1039/C8NH00348C. [DOI] [PubMed] [Google Scholar]
  • 26.Gavela A.F., García D.G., Ramirez J.C., Lechuga L.M. Last advances in silicon-based optical biosensors. Sensors. 2016;16:285. doi: 10.3390/s16030285. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Estevez M.C., Alvarez M., Lechuga L.M. Integrated optical devices for lab-on-a-chip biosensing applications. Laser Photonics Rev. 2012;6:463–487. doi: 10.1002/lpor.201100025. [DOI] [Google Scholar]
  • 28.Liu Q., Shin Y., Kee J.S., Kim K.W., Rafei S.R.M., Perera A.P., Tu X., Lo G.Q., Ricci E., Colombel M., et al. Mach–Zehnder interferometer (MZI) point-of-care system for rapid multiplexed detection of microRNAs in human urine specimens. Biosens. Bioelectron. 2015;71:365–372. doi: 10.1016/j.bios.2015.04.052. [DOI] [PubMed] [Google Scholar]
  • 29.Ruiz-Vega G., Soler M., Lechuga L.M. Nanophotonic biosensors for point-of-care COVID-19 diagnostics and coronavirus surveillance. J. Phys. Photonics. 2021;3:011002. doi: 10.1088/2515-7647/abd4ee. [DOI] [Google Scholar]
  • 30.Aldaya I., Gil-Molina A., Pita J.L., Gabrielli L.H., Fragnito H.L., Dainese P. Nonlinear carrier dynamics in silicon nano-waveguides. Optica. 2017;4:1219. doi: 10.1364/OPTICA.4.001219. [DOI] [Google Scholar]
  • 31.Hiltunen M., Hiltunen J., Stenberg P., Aikio S., Kurki L., Vahimaa P., Karioja P. Polymeric slot waveguide interferometer for sensor applications. Opt. Express. 2014;22:7229. doi: 10.1364/OE.22.007229. [DOI] [PubMed] [Google Scholar]
  • 32.Ramirez J.C., Lechuga L.M., Gabrielli L.H., Hernez-figueroa H.E. Study of a low-cost trimodal polymer waveguide for interferometric optical biosensors. Opt. Express. 2015;23:11985. doi: 10.1364/OE.23.011985. [DOI] [PubMed] [Google Scholar]
  • 33.Ramirez J.C., Gabrielli L.H., Lechuga L.M., Hernandez-figueroa H.E. Trimodal Waveguide Demonstration and Its Implementation as a High Order Mode Interferometer for Sensing Application. Sensors. 2019;19:2821. doi: 10.3390/s19122821. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Ramirez J.C., Schianti J.N., Souto D.E.P., Kubota L.T., Hernez-Figueroa H.E., Gabrielli L.H. Dielectric barrier discharge plasma treatment of modified SU-8 for biosensing applications. Biomed. Opt. Express. 2018;9:2168. doi: 10.1364/BOE.9.002168. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Rezaeifar F., Ahsan R., Lin Q., Chae H.U., Kapadia R. Hot-electron emission processes in waveguide-integrated graphene. Nat. Photonics. 2019;13:843–848. doi: 10.1038/s41566-019-0524-1. [DOI] [Google Scholar]
  • 36.Ramirez J.C., Dagli N. Electro Optical Waveguide Based on Embedded Double-Monolayer Graphene Capacitor. In: Kang J., Tomasulo S., Ilev I., Müller D., Litchinitser N., Polyakov S., Podolskiy V., Nunn J., Dorrer C., Fortier T., et al., editors. Proceedings of the Conference on Lasers and Electro-Optics; San Jose, CA, USA. 9–14 May; Washington, DC, USA: Optical Society of America; 2021. OSA Technical Digest. Paper STh4B.4. [Google Scholar]
  • 37.Portes A.V.R., Shimabuko H.H., Saito L.A.M., Ramirez J.C. Highly Efficient Fermi Level Tunning in EO Waveguide Based on Double Layer Graphene Capacitor; Proceedings of the 2021 SBFoton International Optics and Photonics Conference (SBFoton IOPC); Online. 31 May–2 June 2021; pp. 1–4. [DOI] [Google Scholar]
  • 38.Gungor A., Bhasker P., Kim N.K., Kim J.H., Dagli N. Advanced Photonics 2016 (IPR, NOMA, Sensors, Networks, SPPCom, SOF) Optical Society of America; Washington, DC, USA: 2016. A Technique for Accurate Analysis of Graphene Containing Optical Waveguides. OSA Technical Digest. Paper IW2B.3. [Google Scholar]
  • 39.Hu X., Zhang Y., Chen D., Xiao X., Yu S. Design and Modeling of High Efficiency Graphene Intensity/Phase Modulator Based on Ultra-Thin Silicon Strip Waveguide. J. Light. Technol. 2019;37:2284–2292. doi: 10.1109/JLT.2019.2901916. [DOI] [Google Scholar]
  • 40.Jacques S.L. Optical properties of biological tissues: A review. Phys. Med. Biol. 2013;58:37–61. doi: 10.1088/0031-9155/58/11/R37. [DOI] [PubMed] [Google Scholar]
  • 41.Wang L., Ren J., Han X., Claes T., Jian X., Bienstman P., Baets R., Zhao M., Morthier G. A Label-Free Optical Biosensor Built on a Low-Cost Polymer Platform. IEEE Photonics J. 2012;4:920–930. doi: 10.1109/JPHOT.2012.2200671. [DOI] [Google Scholar]
  • 42.Wang W., Yi P., Li J., Dong H., Chen C., Zhang D., Shen H., Fu B. Polymer Optical Waveguide Grating-Based Biosensor to Detect Effective Drug Concentrations of Ginkgolide A for Inhibition of PMVEC Apoptosis. Biosensors. 2021;11:264. doi: 10.3390/bios11080264. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Kim G.D., Son G.S., Lee H.S., Kim K.D., Lee S.S. Integrated photonic glucose biosensor using a vertically coupled microring resonator in polymers. Opt. Commun. 2008;281:4644–4647. doi: 10.1016/j.optcom.2008.06.006. [DOI] [Google Scholar]
  • 44.Liu Q., Tu X., Kim K.W., Kee J.S., Shin Y., Han K., Yoon Y.J., Lo G.Q., Park M.K. Highly sensitive Mach–Zehnder interferometer biosensor based on silicon nitride slot waveguide. Sens. Actuators B Chem. 2013;188:681–688. doi: 10.1016/j.snb.2013.07.053. [DOI] [Google Scholar]
  • 45.Tu K., Song J., Liow T.Y., Park M.K., Yiying J.Q., Kee J.S., Yu M., Lo G.Q. Thermal independent Silicon-Nitride slot waveguide biosensor with high sensitivity. Opt. Express. 2012;20:2640–2648. doi: 10.1364/OE.20.002640. [DOI] [PubMed] [Google Scholar]
  • 46.Taha A.M., Paredes B., Khilo A., Dahlem M.S. SOI-based centimeter-scale Mach–Zehnder interferometers for fluid sensing; Proceedings of the SPIE 10106, Integrated Optics: Devices, Materials, and Technologies XXI; San Francisco, CA, USA. 28 January–2 February 2017; [DOI] [Google Scholar]
  • 47.Densmore A., Xu D.-X., Janz S., Waldron P., Mischki T., Lopinski G., Delâge A., Lapointe J., Cheben P., Lamontagne B., et al. Spiral-path high-sensitivity silicon photonic wire molecular sensor with temperature-independent response. Opt. Lett. 2008;33:596–598. doi: 10.1364/OL.33.000596. [DOI] [PubMed] [Google Scholar]
  • 48.Densmore A., Xu D.-X., Waldron P., Janz S., Cheben P., Lapointe J., Delage A., Lamontagne B., Schmid J.H., Post E. A Silicon-on-Insulator Photonic Wire Based Evanescent Fiel. IEEE Photonics Technol. Lett. 2006;18:2520–2522. doi: 10.1109/LPT.2006.887374. [DOI] [Google Scholar]
  • 49.Ahlam H., Abdesselam H. Design of high-sensitive biosensor based on cavity-waveguides coupling in 2D photonic crystal. J. Electromagn. Waves Appl. 2015;29 doi: 10.1080/09205071.2015.1012597. [DOI] [Google Scholar]
  • 50.Luan E., Yun H., Laplatine L., Dattner Y., Ratner D.M., Cheung K.C., Chrostowski L. Enhanced Sensitivity of Subwavelength Multibox Waveguide Microring Resonator Label-Free Biosensors. IEEE J. Sel. Top. Quantum Electron. 2019;25:7300211. doi: 10.1109/JSTQE.2018.2821842. [DOI] [Google Scholar]
  • 51.Jindal S., Sobti S., Kumar M., Sharma S., Pal M.K. Nanocavity-Coupled Photonic Crystal Waveguide as Highly Sensitive Platform for Cancer Detection. IEEE Sens. J. 2016;16:3705–3710. doi: 10.1109/JSEN.2016.2536105. [DOI] [Google Scholar]
  • 52.Zadeh F.R., Kaatuzian H., Danaie M. Hybrid Photonic Crystal Cavity as a Sensitive Label-Free Biosensor; Proceedings of the 2019 27th Iranian Conference on Electrical Engineering (ICEE); Yazd, Iran. 30 April–2 May 2019; pp. 18–22. [DOI] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data presented in this study are available on request from the corresponding author.


Articles from Polymers are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES