Abstract
Purpose: The current explored the impact of heat inactivation of blood samples on the results of a particular clinical test and its potential application value during the SARS-CoV-2 pandemic. We have aimed at providing a reference for clinical testing methods during the pandemic.
Methods: Blood samples were selected from our department's routine clinical examination between January 2021 and June 2021. The levels of these samples for quantitative detection of these indicators in each group (n = 90 cases/group) covered normal reference ranges and medically determined levels. For qualitative testing of the indicators, the specimens were additionally classified as negative, weakly positive, and positive (n = 20 cases/group). The specimens were then inactivated, and the differences in relevant indicators before and after inactivation were evaluated.
Results: A statistically significant difference was evident between the levels of TSH, T3, FT4, FT3, AFP, NSE, CYFRA211, IRI, IL-1β, IL-6, IL-8, IL-10, IL-2R, and TNF-α in the non-inactivated group 1 and the inactivated group 1 (P < 0.05). Among them, there was a strong correlation between TSH, T3, FT4, FT3, CYFRA211, IRI, IL-1β, IL-6, IL-8, and IL-2R levels in the two groups (P < 0.05), however, there was no correlation between AFP (P = 0.256) and NSE (P = 0.352) levels between the two groups (P > 0.05). The detected values of low-level AFP (<4 ng/mL), IL-10, and TNF-α after inactivation were all lower than the detection limit. There was not any statistically significant difference in the levels of tumor markers, such as CEA, CA125, CA724, CA199, CA153, and the quantitative levels of T4, Vit. D, HCG, CPS, and five items of hepatitis B virus (P > 0.05). The positive rate of anti-nuclear antibodies after inactivation was not statistically different from the ones observed before inactivation (P > 0.05). Upon correction by the regression equation, the observed levels of TSH, T3, FT4, FT3, CYFRA211, IRI, IL-1β, IL-6, IL-8, and IL-2R were not significantly different from those before inactivation (P > 0.05).
Conclusion: The heat inactivation of blood samples had different various effects on different test indicators, and some indicators could be corrected by employing regression equations. This detection method could potentially be employed during the SARS-CoV-2 pandemic, thereby effectively preventing iatrogenic infections.
Keywords: SARS-CoV-2, test indicator, cytokines, pandemic, serum inactivation
Introduction
The Severe Acute Respiratory Syndrome-related Coronavirus 2 (SARS-CoV-2) is the B subgroup of Betacoronavirus and is essentially an enveloped single-stranded positive-stranded RNA virus (GenBank No.MN908947). It possesses a helical structure and is highly infectious (1–3). At the moment, it is known that SARS-CoV-2 spreads by respiratory droplets, close contact, and prolonged exposure to high aerosol concentrations (4–6). The majority of the Coronavirus Disease 2019 (COVID-19) patients suffer from co-morbidities and underlying disorders, such as tumors, cardiovascular diseases, thyroid, and hepatitis B virus. In addition to SARS-CoV-2 nucleic acid detection and antibody detection, routine hematology indicators are fundamental for clinical typing as well as for the comprehensive treatment and monitoring of patients (7, 8). SARS-CoV-2 is highly infectious, which exposes the investigators and clinicians to great risks and challenges during their efforts for the prevention and control period of the pandemic, especially during routine clinical inspections of confirmed and suspected patients (9, 10). According to GB19489-2008 “General Requirements for Laboratory Biosafety,” relevant laboratory tests for patients with COVID-19 should be carried out in a biosafety tertiary laboratory (BSL-3). However routine clinical laboratory testing is limited by the level of protection and limited space. As a potential solution to this problem, if the patient's blood sample is centrifuged and inactivated in the BSL-3 laboratory before testing, then the inactivated blood or serum sample can be tested in a routine laboratory. Such inactivation would not only eliminate the need for sample testing in BSL-3 laboratory conditions but also greatly substantially reduce the risk of infection of the laboratory personnel handling the specimen (11). There have been quite a few reports in the literature on the impact of exposing a blood sample to 56°C for 30 min, on the analysis of SARS-CoV-2 nucleic acid, routine biochemistry, blood routine, and coagulation detection. However, the influence of 30 min inactivation treatment at 56°C upon the detection of tumor markers, thyroid function, cytokines, and autoantibodies is rather ambiguous. Herein, we investigated whether the inactivation treatment at 56°C for 30 min had any influence on the evaluation of relevant blood samples.
Materials and Methods
The Source of the Specimen
The blood specimens were obtained from routine clinical examination specimens in our department. Quantitative detection indicators were collected based on high-, medium-, and low-value (n = 60 cases), and the remaining 30 cases were used for post-calibration detection. Quantitative detection indicators were collected based on negative, weakly positive, and positive (n = 60 cases). The samples chosen were all fresh specimens taken on the same day, ensuring a detection gap of no more than 2 h between specimens before and after inactivation. The testing system utilized was run strictly in accordance with the relevant ISO15189 quality management system requirements, and its precision is higher than the relevant standards indicated in the manufacturer's statement or the relevant CLIA '88 recommendations. Identical fresh reagents were used to evaluate the specimens daily before and following inactivation as well as for interior quality control (IQC). If the IQC fails or there is a significant deviation from the mean value of the IQC, the quality control products were tested continuously at two concentration levels for 20 times in accordance with the requirements of NCCLS EP15-A, and the intra-batch imprecision and total precision of the two concentrations were calculated, respectively. Only when the findings were satisfactory, we would begin testing specimens on the same day before and after inactivation (Table 1).
Table 1.
Indicators | Control Lot. (level 1, level 2) | IQC mean levels | IQC CVs(%) | Value ranges | ||
---|---|---|---|---|---|---|
level 1 | level 2 | level 1 | level 2 | |||
TSH (mIU/L) | 40,371, 40,373 | 0.364 | 32.917 | 3.77 | 3.25 | 0.2–51.01 |
40,391, 40,393 | 0.368 | 32.877 | 4.08 | 4.22 | ||
T4 (nmol/L) | 40,371, 40,373 | 58.39 | 214.45 | 6.20 | 4.59 | 4.7–368.3 |
40,391, 40,393 | 67.93 | 176.17 | 7.07 | 5.59 | ||
T3 (nmol/L) | 40,371, 40,373 | 1.30 | 5.40 | 7.69 | 5.74 | 0.63–7.23 |
40,391, 40,393 | 1.27 | 5.32 | 7.87 | 5.26 | ||
FT4 (pmol/L) | 40,371, 40,373 | 9.59 | 42.73 | 6.99 | 6.08 | 2.13–51.93 |
40,391, 40,393 | 9.84 | 35.55 | 7.11 | 7.82 | ||
FT3 (pmol/L) | 40,371, 40,373 | 4.02 | 17.12 | 5.22 | 4.38 | 0.3–26.7 |
40,391, 40,393 | 4.08 | 17.22 | 4.41 | 5.28 | ||
AFP (ng/ml) | 54,671, 54,673 | 9.98 | 235.06 | 4.91 | 4.24 | 0.61–446.6 |
CEA (ng/ml) | 54,671, 54,673 | 2.74 | 66.14 | 5.47 | 3.19 | 0.22–1904 |
CA125 (U/ml) | 54,671, 54,673 | 25.20 | 194.72 | 5.24 | 5.31 | 3.72–1775 |
CA724 (U/ml) | 54,671, 54,673 | 4.38 | 39.59 | 3.65 | 3.64 | 0.2–245.9 |
CA199 (U/ml) | 54,671, 54,673 | 21.79 | 224.00 | 4.77 | 3.89 | 3.12–3097 |
CA153 (U/ml) | 54,671, 54,673 | 22.39 | 119.78 | 5.60 | 5.25 | 3.17–45.38 |
NSE (ng/ml) | 54,671, 54,673 | 11.41 | 75.33 | 4.33 | 4.77 | 7.86–96.39 |
CYFRA21-1 (ng/ml) | 54,671, 54,673 | 6.86 | 119.78 | 3.21 | 2.87 | 2.19–17.83 |
VitD (nmol/L) | 64,911, 64,913 | 15.6 | 91.3 | 7.53 | 6.89 | 7.7–35.1 |
HCG (mIU/ml) | 40,371, 40,373 | 6.32 | 153.87 | 7.99 | 6.97 | 45.87–9104 |
40,391, 40,393 | 11.16 | 235.82 | 7.32 | 5.53 | ||
IRI (mU/L) | 40,371, 40,373 | 15.55 | 191.32 | 5.98 | 6.07 | 6.47–300 |
40,391, 40,393 | 16.57 | 170.04 | 6.70 | 6.73 | ||
Cps (ng/ml) | 40,371, 40,373 | 0.52 | 6.79 | 5.77 | 5.15 | 0.66–12.93 |
40,391, 40,393 | 0.84 | 7.00 | 4.76 | 4.71 | ||
HBs-Ag (IU/ml) | 407,779,407,774 | 91.68 | 3.76 | 5.45 | 6.32 | 1.87–1732 |
HBs-Ab (IU/ml) | 457,381, 457,380 | 89.78 | <2.00 | 3.81 | 0 | 2.12–396.8 |
HBe-Ag (COI) | 483,246, 477,333 | 12.68 | 0.87 | 6.36 | 10.89 | 0.08–14.78 |
HBe-Ab (COI) | 491,396, 491,394 | 0.58 | 1.52 | 6.52 | 4.31 | 0.02–1.53 |
HBc-Ab (COI) | 488,568, 488,565 | 0.53 | 2.28 | 6.71 | 4.55 | 0.01–2.32 |
IL-1β (pg/ml) | ILCO10036, ILCO20036 | 93.67 | 427 | 6.65 | 5.78 | 6.52–59.3 |
IL-6 (pg/ml) | ILCO10036, ILCO20036 | 94.56 | 578 | 7.32 | 6.45 | 4.35–135.8 |
IL-8 (pg/ml) | ILCO10036, ILCO20036 | 113.45 | 867.36 | 7.12 | 6.35 | 6.98–254 |
IL-10 (pg/ml) | LXPC10029, LXPC20029 | 25.78 | 246.76 | 9.98 | 6.54 | 7.21–58.6 |
IL-2R (U/ml) | ILCO10036, ILCO20036 | 423.67 | 1897.35 | 6.73 | 6.25 | 130–1990 |
TNF-α (pg/ml) | ILCO10036, ILCO20036 | 70.75 | 485.67 | 8.02 | 7.34 | 4.67–104 |
Instruments and Reagents
The quantitative detections of tumor markers and five markers of hepatitis B virus were completed using the Roche e 602 automatic chemiluminescence immunoassay analyzer. Thyroid function, C-peptide, and IRI were estimated using Siemens ADVIA Centaur XP automatic chemiluminescence immunoassay analyzer. HCG was tested using Siemens Immulite 2,000 XPi. Detection of the Cytokines was made using Siemens Immulite 1,000 whereas Vit D was detected by Liaison XL automatic chemiluminescence immunoassay analyzer. The anti-nuclear antibody detection kit was purchased from Oumeng, Germany. The water bath equipment was provided by Beijing Fuyilian Medical Equipment Co., Ltd. Interior quality control (IQC) products: Bio-Rad tumor marker plus control for tumor markers, and Bio-Rad immunoassay plus control for thyroid function, C-peptide, IRI, HCG, and Bio-Rad specialty immunoassay control for Vit D, and Original quality control products of Roche for five markers of hepatitis B virus, and Siemens Immulite cytokine control for Cytokines.
Specimen Processing Method
First, each serum sample was divided based on the value into two portions. One part was regarded as non-inactivated group 1, and the other part was inactivated group 1. Following detection, the serum samples of non-inactivated group 1 were collected in 2 ml EP tubes in the biosafety cabinet, numbered and labeled as inactivated group 1, and finally sealed with sealing film. The specimen from inactivated group 1 was heat-treated in a water bath at 56°C for 30 min, then removed and placed in a biosafety cabinet. After removing the samples to be restored to room temperature they were centrifuged and tested on the same machine and reagent as non-inactivated group 1. The remaining 30 specimens were used for calibration testing and were divided into two groups: one as the non-inactivated group 2, and the other as the inactivated group 2 which was exposed to heat inactivation at 56°C for 30 min. The operation steps were the same as above.
Statistical Analysis
In this study, SPSS 26.0 software was used for statistical analysis. The measurement data conformed to the normal distribution, which was expressed by ± s, and the comparison of the level difference of each index before and after inactivation was performed by a paired t-test. The χ2-test was used to compare the positive rate of anti-nuclear antibodies. The detection value after inactivation was used to reverse the detection prior to inactivation (true value) to calibrate the detection value of certain markers following inactivation. The detection level before and after heat inactivation was compared using Pearson correlation and regression analysis. The study's significance level was set to α = 0.05.
Results
Comparison of Detection Levels Before and After Inactivation
The levels of TSH, T3, AFP, NSE, CYFRA211, IRI, IL-6, IL-10, IL-2R, and TNF-α in inactivated group 1 were substantially low in comparison to those in non-inactivated group 1, while the levels of FT4, FT3, IL-1β, and IL-8 were significantly higher in the inactivated group, and the difference was statistically significant (P < 0.05). Among them, the deviation of TSH and T3 was small, which might be considered a clinically acceptable error. The detection of low-level AFP (<4 ng/mL), IL-10, and TNF-α following inactivation were all well below the lower limit of detection. For all other indicators, there was no statistically significant difference in the detection levels of other indicators (P > 0.05; Table 2).
Table 2.
Indicators | Non-inactivated group 1 | Inactivation group 1 | Bias (%) | t-value | P-value |
---|---|---|---|---|---|
(n = 60) | (n = 60) | ||||
TSH (mIU/L) | 7.94 ± 1.03 | 7.39 ± 0.95 | −6.47 ± 5.46 | 4.28 | 0.001 |
T4 (nmol/L) | 72.55 ± 9.69 | 69.22 ± 9.25 | −2.37 ± 12.63 | 1.752 | 0.085 |
T3 (nmol/L) | 1.11 ± 0.16 | 1.06 ± 0.15 | 14.47 ± 16.20 | 3.307 | 0.001 |
FT4 (pmol/L) | 18.08 ± 8.64 | 56.53 ± 22.63 | 220.17 ± 59.84 | 16.541 | 0.001 |
FT3 (pmol/L) | 5.27 ± 3.49 | 8.45 ± 5.74 | 59.89 ± 32.10 | 9.555 | 0.001 |
AFP (ng/ml) | 13.26 ± 37.17 | 6.30 ± 13.40 | −44.96 ± 26.52 | 7.598 | 0.001 |
CEA (ng/ml) | 85.35 ± 265.39 | 87.13 ± 294.91 | −4.48 ± 5.34 | 0.37 | 0.712 |
CA125 (U/ml) | 124.63 ± 261.58 | 125.01 ± 266.80 | −1.41 ± 4.75 | 0.319 | 0.751 |
CA724 (U/ml) | 15.78 ± 37.94 | 16.23 ± 40.06 | 2.12 ± 4.60 | 1.319 | 0.194 |
CA199 (U/ml) | 69.57 ± 117.72 | 67.81 ± 117.93 | −4.86 ± 5.57 | 1.832 | 0.073 |
CA153 (U/ml) | 10.63 ± 2.47 | 11.42 ± 2.63 | 2.05 ± 1.38 | 1.469 | 0.062 |
NSE (ng/ml) | 21.9 ± 16.64 | 0.87 ± 0.61 | −94.93 ± 4.36 | 7.724 | 0.001 |
CYFRA21-1 (ng/ml) | 5.75 ± 3.23 | 0.71 ± 0.16 | −85.02 ± 5.65 | 10.032 | 0.001 |
VitD (nmol/L) | 12.41 ± 4.28 | 14.05 ± 4.47 | 2.38 ± 2.51 | 0.537 | 0.514 |
HCG (mIU/ml) | 285.72 ± 15.64 | 287.35 ± 16.27 | 2.51 ± 1.48 | 0.629 | 0.431 |
IRI (mU/L) | 79.52 ± 76.99 | 1.39 ± 1.39 | −197.89 ± 1.19 | 5.923 | 0.001 |
Cps (ng/ml) | 7.38 ± 1.52 | 7.61 ± 1.58 | 1.52 ± 1.07 | 0.543 | 0.521 |
HBs-Ag (IU/ml) | 649.28 ± 47.32 | 653.71 ± 49.06 | 7.38 ± 4.25 | 0.574 | 0.516 |
HBs-Ab (IU/ml) | 41.63 ± 5.28 | 43.25 ± 5.41 | 2.54 ± 2.07 | 0.641 | 0.432 |
HBe-Ag (COI) | 0.17 ± 0.05 | 0.14 ± 0.03 | −0.06 ± 0.02 | 0.538 | 0.572 |
HBe-Ab (COI) | 0.75 ± 0.14 | 0.77 ± 0.16 | 0.16 ± 0.04 | 0.642 | 0.437 |
HBc-Ab (COI) | 0.24 ± 0.12 | 0.21 ± 0.09 | −0.07 ± 0.02 | 0.549 | 0.527 |
IL-1β (pg/ml) | 18.63 ± 12.35 | 33.01 ± 20.15 | 86.76 ± 58.20 | 11.709 | 0.001 |
IL-6 (pg/ml) | 32.10 ± 36.61 | 14.87 ± 18.79 | −56.28 ± 12.51 | 6.131 | 0.001 |
IL-8 (pg/ml) | 53.93 ± 59.41 | 95.37 ± 113.52 | 59.88 ± 34.89 | 5.872 | 0.001 |
IL-10 (pg/ml) | 23.85 ± 16.16 | <5.00 | - | - | - |
IL-2R (U/ml) | 796.48 ± 453.03 | 365.33 ± 13.79 | −58.25 ± 13.79 | 15.686 | 0.001 |
TNF-α (pg/ml) | 21.21 ± 21.59 | <4.00 | - | - | - |
Correction of Results of Some Test Items
The detection value after inactivation was used to calibrate the detection value of certain markers after inactivation by reversing the detection value of some items after inactivation. We used linear correlation and regression analysis to get the regression equation and correlation coefficient for the detection level before and after heat inactivation. We found that there was a correlation between TSH, T3, FT4, FT3, CYFRA211, IRI, IL-1β, IL-6, IL-8, and IL-2R between the two groups (P < 0.05). However, there was no correlation between the AFP (P = 0.256) and NSE (P = 0.352) levels of the two groups (P > 0.05, Figure 1 and Table 3).
Table 3.
Indicators | Correlation coefficient r | P-value | Regression equation | Decisive factor |
---|---|---|---|---|
r2 | ||||
TSH | 1.00 | 0.001 | Y = 0.002 + 1.075X | 1.000 |
T3 | 0.99 | 0.001 | Y = −0.22 + 1.04X | 0.790 |
FT4 | 0.907 | 0.001 | Y = −1.474 + 0.346X | 0.822 |
FT3 | 0.963 | 0.001 | Y = 0.319 + 0.586X | 0.928 |
AFP | 0.329 | 0.256 | - | - |
NSE | 0.151 | 0.35 | - | - |
CYFRA211 | 0.751 | 0.003 | Y = −5.141 + 15.245X | 0.564 |
IRI | 0.884 | 0.001 | Y = 11.403 + 48.923X | 0.781 |
IL-1β | 0.941 | 0.001 | Y = −0.389 + 0.576X | 0.885 |
IL-6 | 0.886 | 0.001 | Y = 6.419 + 1.727X | 0.786 |
IL-8 | 0.995 | 0.001 | Y = 4.242 + 0.521X | 0.991 |
IL-2R | 0.951 | 0.001 | Y = 213.51 + 1.596X | 0.905 |
Comparison of Corrected Value and the True Value of Some test Items
We used the regression equation obtained from result 2.2 to correct the inactivated group 2. Compared with the level of non-inactivated group 2, the difference between the two groups was not statistically significant (P > 0.05; Table 4).
Table 4.
Indicators | Non-inactivated group 2 | After correction for inactivation group 2 | Bias (%) | t-value | P-value |
---|---|---|---|---|---|
(n = 30) | (n = 30) | ||||
TSH (mIU/L) | 7.94 ± 1.03 | 7.95 ± 1.05 | 0.26 ± 0.14 | 0.473 | 0.625 |
T3 (nmol/L) | 1.11 ± 0.16 | 0.95 ± 0.18 | −0.72 ± 0.43 | 0.639 | 0.421 |
FT4 (pmol/L) | 18.08 ± 8.64 | 18.09 ± 8.67 | 0.17 ± 0.09 | 0.512 | 0.583 |
FT3 (pmol/L) | 5.27 ± 3.49 | 5.32 ± 3.54 | 0.54 ± 0.12 | 0.612 | 0.437 |
CYFRA211 (ng/ml) | 5.28 ± 3.27 | 5.38 ± 1.84 | 0.19 ± 0.40 | 0.197 | 0.846 |
IRI (mU/L) | 93.47 ± 88.13 | 86.28 ± 80.75 | 0.29 ± 0.80 | 1.006 | 0.327 |
IL-1β (pg/ml) | 17.26 ± 12.59 | 16.91 ± 10.06 | 0.08 ± 0.45 | 0.376 | 0.709 |
IL-6 (pg/ml) | 38.45 ± 40.72 | 35.98 ± 35.15 | 0.27 ± 0.53 | 0.572 | 0.522 |
IL-8 (pg/ml) | 57.52 ± 74.53 | 57.51 ± 73.91 | 0.09 ± 0.22 | 0.003 | 0.997 |
IL-2R (U/ml) | 820.50 ± 505.187 | 820.67 ± 464.52 | 0.13 ± 0.40 | 0.005 | 0.996 |
Comparison of the Positive Rates Before and After Anti-nuclear Antibody Inactivation
The variation in the positive rate of anti-nuclear antibodies before and after inactivation was not statistically significant in this study (P > 0.05; Table 5).
Table 5.
Inactivation | Result | n | Positive rate | x2 | P-value |
---|---|---|---|---|---|
method | (%) | ||||
Non-inactivated group 1 (n = 60) | + | 39 | 65.00 | 0.036 | 0.849 |
- | 21 | ||||
Inactivation group 1 (n = 0) | + | 38 | 63.33 | ||
- | 22 |
Conclusion
SARS-CoV-2 is sensitive to ultraviolet light and heat. Studies have shown that heating at 56°C for 30 min or exposure to ether, 75% ethanol, chlorine-containing disinfectant, peracetic acid, and chloroform can effectively inactivate the virus (12, 13). Laboratory staff can become infected after accidental extended exposure to aerosols at high concentrations when they come into contact with patients' blood, urine, feces, or other body fluids (14). Along with adequate biosafety protection, if the sample can be inactivated prior to testing without compromising the test result, the inspector's risk of infection can be significantly decreased. Additionally, this enables routine clinical testing of blood samples from COVID-19 patients in biosafety secondary laboratories, which will aid in the comprehensive treatment and prognosis monitoring of clinical COVID-19 patients. In this investigation, blood samples examined at our hospital were inactivated, and the changes in key indicators and the positive rate of detection of anti-nuclear antibodies were detected before and after inactivation. The purpose of this study was to increase the safety of serum sample testing during the pandemic and to minimize the risk of iatrogenic SARS-CoV-2 infection.
Recent studies have reported the influence of heat inactivation treatment upon the test results of routine clinical test indicators. Studies have shown (15–18) that following heat inactivation treatment, there is no statistical difference in the values of TP, TBIL, Cr, TC, TG, K+, Cl-, CRP, BUN, Glu, D-dimer, hemoglobin, white blood cells, platelet count, serological markers of infectious diseases, such as five indicators of hepatitis B virus, TP-Ab, and HIV. However, the values pertaining to some heat-labile enzymes, such as ALB, GLO, PA, AST, ALT, LDH, ALP, γ-GT, CK were found to be significantly reduced. Liu Changde et al. (19) showed that TCO2, GLO, TBIL, IBIL, HDL, PA, β2-MG, TP, ALB, and enzyme indicators were more affected by inactivation, while other conventional biochemical experience less difference. The disparity in study results was most likely driven by the disparity in detecting platforms and thermal treatment techniques. Several studies have also shown that (16, 20) inactivating blood samples at 56°C for 30 min had almost no effect on the results of SARS-CoV-2 antibody detection.
In this study, after the serum samples were inactivated at 56°C for 30 min, the levels of CEA, CA125, CA724, CA199, CA153, and other tumor markers, T4, VitD, HCG, CPS, five typical markers of hepatitis B virus, and anti-nuclear antibodies did not vary change significantly. However, the levels of TSH, T3, AFP, NSE, CYFRA211, IRI, IL-6, IL-10, IL-2R, and TNF-α were observed to be significantly reduced. This was most likely the effect of heat inactivation destroying the protein structure, resulting in reduced antigenicity. The levels of FT4 and FT3 increased significantly since presumably heat inactivation cause a part of T4 and T3 to be dissociated from thyroid-binding globulin. The levels of IL-1β and IL-8 were significantly increased, which may be related to the exposure of some antigen epitopes after heat treatment, and further studies are required to find out the specific reasons. Although the levels of TSH and T3 before and after inactivation were statistically different, the bias was relatively small and was within the clinical acceptable range. In this study, the levels of TSH, T3, FT4, FT3, CYFRA211, IRI, IL-1β, IL-6, IL-8, and IL-2R changed significantly. However, after correction by the regression equation, its level was not significantly different from that before inactivation. Therefore, clinical regression equation analysis could be performed on the indicators that had changed significantly after partial inactivation to reduce the difference caused by inactivation and improve the clinical value of the detected indicators.
The inactivation of blood samples had only minimal influence on the results of the clinical tests. Additionally, regression equations can be used to adjust for specific indications. This detection approach might be utilized successfully to restrict the spread of sample aerosols during the SARS-CoV-2 pandemic. This study successfully reinforced previous reports on the impact of serum inactivation upon tumor markers, thyroid function, IRI, Cps, cytokines, and anti-nuclear antibodies, completing the examination of the influence of serum inactivation on clinical tests findings. However, the study was not without limitations. For instance, the research was constrained by certain circumstances; the number of selected samples was limited, and large-scale data exploration was difficult. Future research should aim to increase the sample size as much as possible to derive more accurate findings. Simultaneously, the extinguishing detection technique needs enhancement. This technique has the potential to reduce the need for protective equipment and personnel and increase the efficiency of inspections, allowing it to be deployed on a broad scale during the SARS-CoV-2 pandemic when medical resources are scarce.
Data Availability Statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.
Author Contributions
EX and TL contributed to conception and design of the study. EX and QC performed the statistical analysis and wrote sections of the manuscript. EX wrote the first draft of the manuscript. All authors contributed to manuscript revision, read, and approved the submitted version.
Funding
Natural Science Research Project of Universities in Anhui Province (KJ2020ZD24).
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher's Note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
- 1.Zhang JC, Jiang H, Deng L, Wang K, Sun WM, Zhou P, et al. Recommendations on metabolic support and nutritional therapy for patients with severe new coronavirus pneumonia (COVID-19). Chin J Emerg Med. (2020) 29:456–60. 10.3760/cma.j.issn.1671-0282.2020022.003 [DOI] [Google Scholar]
- 2.Behzad S, Velez E, Najafi MH, Gholamrezanezhad A. Coronavirus disease (2019). (COVID-19) pneumonia incidentally detected on coronary CT angiogram: a do-not-miss diagnosis. Emerg Radiol. (2020). 27:721–6. 10.1007/s10140-020-01802-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Di Maio P, Iocca O, Cavallero A, Giudice M. Performing the nasopharyngeal and oropharyngeal swab for 2019-novel coronavirus (SARS-CoV-2) safely: how to dress, undress, and technical notes. Head Neck. (2020) 42:1548–51. 10.1002/hed.26230 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Zhang Y, Tang DZ, Shu B, Li WX, Zhang JL, Li Y, et al. Evaluation and analysis on the SARS-CoV-2-induced injuries in multiple organs and the intervention of traditional Chinese medicine based on renin-angiotensin system. Mod Traditional Chinese Med Mater Med. World Sci Technol. (2020) 22:264–69. 10.11842/wst.20200214004 [DOI] [Google Scholar]
- 5.Huang D, Wang T, Chen Z, Yang H, Yao R, Liang Z. A novel risk score to predict diagnosis with coronavirus disease 2019 (COVID-19) in suspected patients: a retrospective, multicenter, and observational study. J Med Virol. (2020). 92:2709–17. 10.1002/jmv.26143 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Qu J, Yang W, Yang Y, Qin L, Yan F. Infection control for CT equipment and radiographers' personal protection during the coronavirus disease (COVID-19) outbreak in China. AJR Am J Roentgenol. (2020) 215:940–4. 10.2214/AJR.20.23112 [DOI] [PubMed] [Google Scholar]
- 7.Goh Y, Chua W, Lee JKT, Ang BWL, Liang CR, Tan CA, et al. Operational Strategies to Prevent Coronavirus Disease 2019 (COVID-19) spread in radiology: experience from a singapore radiology department after severe acute respiratory syndrome. J Am Coll Radiol. (2020) 17:717–23. 10.1016/j.jacr.2020.03.027 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Xu XH, He N, Shen N, Xu J, Zhai SD. Opinions on safety of chloroquine phosphate dosing regimen for novel coronavirus pneumonia. Adverse Drug React J. (2020) 22:142–6. 10.3760/cma.j.cn114015-20200310-00179 [DOI] [Google Scholar]
- 9.Yin DL, Liu LX. Management strategy of prevention and control for surgeons to asymptomatic 2019-nCoV RNA positive corona virus diseases 2019 (COVID-19). Chinese J Pract Surg. (2020) 40:248–9. 10.19538/j.cjps.issn1005-2208.2020.03.03 [DOI] [Google Scholar]
- 10.Medicine A S F R. Patient management and clinical recommendations during the coronavirus (COVID-19) pandemic (as of March 17, 2020). Obstet Gynecol Surv. (2020) 75:38. 10.1097/01.ogx.0000666100.94243.bc [DOI] [Google Scholar]
- 11.Kampf G, Todt D, Pfaender S, Steinmann E. Corrigendum to “Persistence of coronaviruses on inanimate surfaces and their inactivation with biocidal agents”. J Hosp Infect. (2020) 105:587. 10.1016/j.jhin.2020.06.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Pan F, Yang L, Li Y, Liang B, Li L, Ye T, et al. Factors associated with death outcome in patients with severe coronavirus disease-19 (COVID-19): a case-control study. Int J Med Sci. (2020) 17:1281–92. 10.7150/ijms.46614 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Zou XB. The Importance of infection control in clinical practice of dermatology from the outbreak of the novel coronavirus pneumonia. Chinese J Dermatovenereol. (2020) 34:495–9. 10.13735/j.cjdv.1001-7089.202001118 [DOI] [Google Scholar]
- 14.Ye L, Yang S, Liu C. Infection prevention and control in nursing severe coronavirus disease (COVID-19) patients during the pandemic. Crit Care. (2020) 24:338. 10.1186/s13054-020-03076-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Li XD, Liao H, Liu Y, Li L, Chen RJ, Li Q, et al. Study on the effect of virus inactivation method heated at 60? for 1 hour on routine clinical detection indicators. Infect Dis Inf. (2016) 29:160–6. 10.1016/j.hydromet.2016.10.007 [DOI] [Google Scholar]
- 16.Guo WY, Tang M, Shi WY, Tan T, Lu S, Hu Z, et al. The effect of inactivation at 56řC for 30 minutes on the results of new coronavirus antibody and biochemical immunoassay. Lab Med Clin. (2021) 18:97–100. 10.3969/j.issn.1672-9455.2021.01.029 [DOI] [Google Scholar]
- 17.She YQ, Liu Q, Zhou JH, Dai LZ, Deng Y, Li N. Effect of heat inactivation on blood transfusion-related infection markers. Chinese J Infect Control. (2020) 19:498–503. 10.12138/j.issn.1671-9638.20206436 [DOI] [Google Scholar]
- 18.Liu ZY, Li XW, Zhou Y, Xiao Z. The influence of heat-inactivated virus method on the detection results of biochemical and inflammatory indicators[J]. CChin J Sociol. (2020) 30:1485–1489. [Google Scholar]
- 19.Liu CD, Jiao MY, Gao C, Zhang DH, Cui LY. Effects of inactivation of serum at 56° for 30 min on biochemical test results. Int J Lab Med. (2020) 41:1366–70. 10.3969/j.issn.1673-4130.2020.11.020 [DOI] [Google Scholar]
- 20.Xue XY, Zhu CL, Huang SZ, Pan LH, Li WX. Effect of heat inactivation of blood samples on the efficacy of three detection methods of SARS-CoV-2 antibodies. J Southern Med Univ. (2020) 40:316–20. 10.12122/j.issn.1673-4254.2020.03.03 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.