Evaluation of Durability of Transparent Graphene Electrodes Fabricated on Different Flexible Substrates for Chronic in vivo Experiments

David Ding 1, Yichen Lu 2, Ruoyu Zhao 3, Xin Liu 4, Chawina De-Eknamkul 5, Chi Ren 6, Armaghan Mehrsa 7, Takaki Komiyama 8, Duygu Kuzum 9
PMCID: PMC8560430  NIHMSID: NIHMS1639106  PMID: 32191878

Abstract

Objective:

To investigate chronic durability of transparent graphene electrodes fabricated on polyethylene terephthalate (PET) and SU-8 substrates for chronic in vivo studies.

Methods:

We perform systematic accelerated aging tests to understand the chronic reliability and failure modes of transparent graphene microelectrode arrays built on PET and SU-8 substrates. We employ graphene microelectrodes fabricated on PET substrate in chronic in vivo experiments with transgenic mice.

Results:

Our results show that graphene microelectrodes fabricated on PET substrate work reliably after 30 days accelerated aging test performed at 87 °C, equivalent to 960 days in vivo lifetime. We demonstrate stable chronic recordings of cortical potentials in multimodal imaging/recording experiments using transparent graphene microelectrodes fabricated on PET substrate. On the other hand, graphene microelectrode arrays built on SU-8 substrate exhibit extensive crack formation across microelectrode sites and wires after one to two weeks, resulting in total failure of recording capability for chronic studies.

Conclusion:

PET shows superior reliability as a substrate for graphene microelectrode arrays for chronic in vivo experiments.

Significance:

Graphene is a unique neural interface material enabling cross-talk free integration of electrical and optical recording and stimulation techniques in the same experiment. To date, graphene-based microelectrode arrays have been demonstrated in various multi-modal acute experiments involving electrophysiological sensing or stimulation, optical imaging and optogenetics stimulation. Understanding chronic reliability of graphene-based transparent interfaces is very important to expand the use of this technology for long-term behavioral studies with animal models.

Keywords: graphene, accelerated aging, microelectrode arrays, chronic reliability, polyethylene terephthalate (PET), SU-8, neural interface

I. Introduction

Electrophysiology has been one of the major tools for basic neuroscience research and clinical practice for decades. There have been vast technological advances in electrode technologies, allowing for high-temporal resolution recording of neuronal spikes and local field potentials generated by neuronal ensembles. Optical techniques such as multi-photon imaging has opened new avenues for high-spatial resolution mapping of neural activity from hundreds of neurons [1, 2]. Combining electrical recordings with optical methods such as multi-photon microscopy or optogenetics is crucial to study neural dynamics at multiple spatial and temporal scales. To that end, graphene-based optically transparent microelectrode arrays built on flexible polymer substrates have been developed and employed in various acute in vivo experiments [38]. Graphene is an optically transparent conductor, offering several outstanding properties, such as flexibility [9], high conductivity [10] and biocompatibility [11, 12], making it attractive for neural interface applications. Optically transparent graphene microelectrode arrays have been shown to enable simultaneous multi-modal imaging and electrical recording with high spatiotemporal resolution in acute studies. However, further studies investigating the chronic reliability of graphene-based transparent neural interfaces is critical for adaptation of this technology to long-term behavioral studies with animal models [13].

Transparent graphene arrays can be fabricated on different types of flexible polymer substrates, which exhibit optical transparency in wavelength ranges from 450 nm to 850 nm, commonly used in optical imaging and stimulation experiments. Different types of polymers including polyimide [4, 14], SU-8 [15], PET [5, 6], Parylene-C [16] and polydimethylsiloxane (PDMS) [17] have been investigated for fabrication of metal-based flexible microelectrode arrays in the past. Among those, polyimide is not preferred due to its yellow color and absorption peak around 450 nm, overlapping with wavelength of interest for optical imaging and stimulation [4]. Parylene-C has been shown to work in chronic settings. However, the dry etch processes needed to pattern Parylene-C also etches graphene, leading to complex fabrication processes involving sacrificial layers to protect graphene. Stretchable PDMS substrates are widely used for wearable sensors but softness and stretchability of PDMS increases the risk of cracks forming on monolayer graphene. Therefore, in this study we focus on the remaining two polymer substrates PET and SU-8, which meet the requirements for transparency, fabrication compatibility and mechanical support to the monolayer graphene electrodes. We fabricate transparent graphene microelectrode arrays on PET and SU-8 substrates using similar process steps to perform systematic studies to understand the chronic reliability and failure modes of transparent graphene microelectrode arrays built on PET and SU-8 substrates. In order to investigate the chronic reliability of these arrays comparatively, we employ an accelerated aging test, with an aging factor of 32 times. We examine various modes of failure including delamination, cracking, leakage using systematic optical microscope and scanning electron microscopy (SEM) inspection and electrochemical impedance spectroscopy tests performed over 30 days, 2.63 years of equivalent aging. Finally, we perform in vivo chronic animal experiments on mice with simultaneous 2-photon imaging and electrical recordings using the PET graphene electrode array and evaluate the stability of the electrode impedance and the recording noise over time.

II. Methods

A. Fabrication

We fabricated 16-channel graphene microelectrode arrays with a 100 μm electrode size and 400 μm spacing on SU-8 and PET substrates (Figures 1ad). Besides the substrate fabrication, the arrays go through the same fabrication flow. For the graphene microelectrode array with an SU-8 substrate, a sacrificial layer is created by spin-coating polymethylglutarimide (PMGI) resists on a silicon wafer. A 10 μm thick SU-8 layer is spin-coated and defined with photolithography on the sacrificial layer to create the substrate. The substrate is baked incrementally from 125 °C to 135 °C in order to bond the SU-8 [18]. For the graphene microelectrode array with a PET substrate, PDMS is spin-coated on a silicon wafer as the adhesive layer. A 50 μm thick PET substrate layer is then mechanically applied onto the wafer. For both microelectrode arrays, 10 nm chromium and 100 nm gold are deposited onto the substrate and patterned into metal wires and contact pads using photolithography and wet-etching processes. Chemical vapor deposited (CVD) monolayer graphene is transferred onto the wafer using the ‘‘bubbling’’ transfer method [5, 19], and the graphene pads are patterned with AZ1512/PMGI bilayer lithography and oxygen plasma etching. A ‘‘four-step’’ cleaning process’’ is then deployed with the wafers going through 3 solvent bathes of AZ 1-Methyl-2-pyrrolidone (AZ NMP), Remover PG, acetone, and finally a 10-cycle isopropanol/deionized water rinse. A second 10 μm thick SU-8 encapsulation layer is then spin-coated, and the electrode openings/contact pads are defined using photolithography as well. The encapsulation layer is also hard-baked at 150 °C to bond the second layer of SU-8. The graphene microelectrode array fabricated on a PET substrate is mechanically lifted-off the wafer, while the graphene microelectrode array fabricated on an SU-8 substrate is immersed in AZ NMP remover for its lift-off process. A detailed schematic of both graphene array’s fabrication is shown in Figure 1e.

Figure 1.

Figure 1.

Macro and micro images of graphene microelectrodes fabricated on SU-8 and PET substrates, along with the fabrication flow. (a,b) Macro and micro image of graphene microelectrode array fabricated on PET substrate. Each array contains 16 electrode openings, each with a size of 100 µm x 100 µm. (c,d) Macro and micro image of graphene microelectrode array fabricated on SU-8 substrate. (e) i) PMGI sacrificial layer spin-coated onto a silicon wafer for graphene microelectrode array fabricated on SU-8 substrate. ii) SU-8 substrate spin-coated onto the wafer. iii) PDMS sacrificial layer spin-coated onto a silicon wafer for graphene microelectrode array fabricated on PET substrate. iv) PET substrate mechanically applied onto the wafer. v) 10 nm chromium and 100 nm gold deposited for both arrays. vi) Metal wires patterned using photolithography and excess metal etched. vii) Monolayer CVD graphene transferred onto the wafer using the ‘‘bubbling’’ method. viii) Graphene pattemed using AZ1512/PMGI bilayer lithography and etched with oxygen plasma etching. ix) SU-8 encapsulation spin-coated and openings patterned using photolithography. x) Sacrificial layer lift-off for final array.

B. Characterization and comparisons

Both graphene microelectrodes fabricated on a PET substrate and SU-8 substrate were characterized using electrochemical impedance spectroscopy (EIS) in 0.01 M phosphate buffered saline (PBS) to study the graphene/electrolyte interface (Figures 2ad). Impedance results for microelectrodes built on PET and SU-8 substrates from 1 Hz to 100 kHz are shown in Fig. 2a and Fig. 2c. Impedance histograms at 1 kHz are shown in Fig. 2b and Fig. 2d, respectively. To maintain a good signal-to-noise ratio (SNR) for electrical recordings, an electrode’s impedance at 1 kHz should be below 2 MΩ when using Intan acquisition systems [5]. The results suggest that graphene microelectrodes show low impedance and high yield (~90% working channels) regardless of the substrate used. We investigated whether use of different substrates changes the effective surface area for the graphene microelectrodes. Atomic force microscopy (AFM) scanning results for PET and SU-8 substrates are shown in Figure 2e and Figure 2f, respectively. Although PET and SU-8 show different surface morphology, the effective surface areas calculated based on AFM results have shown similar values, with SU-8 being slightly larger (7% increase compared to geometrical surface area) than PET (1.5% increase compared to geometrical surface area). We incorporated these effective surface area results in the equivalent circuit model (Figure 2g). In the equivalent circuit model fitting, Rs mimics the series resistance of the solution, the constant phase element (CPE) represents the Helmholtz double-layer capacitance (Cdl), the bounded Warburg element (Wb) simulates any diffusion processes occurring, and Rct is the charge-transfer resistance, which simulates any faradaic reactions [20]. Representative parameter values for both graphene microelectrode arrays fabricated on a PET substrate and SU-8 substrate are listed in Table I.

Figure 2.

Figure 2.

Electrochemical and surface characterization of both graphene microelectrode arrays and equivalent circuit model. (a) EIS results from 1 Hz to 100 kHz of graphene microelectrode array fabricated on PET substrate along with equivalent model fitting. (b) 1 kHz impedance bar chart for sample graphene microelectrode array on PET substrate. (c) EIS results from 1 Hz to 100 kHz of graphene microelectrode array fabricated on SU-8 substrate along with equivalent model fitting. (d) 1 kHz impedance bar chart for sample graphene microelectrode array on SU-8 substrate. (e) Tapping-mode atomic force microscopy (AFM) 3-dimensional photograph of PET substrate. (f) AFM 3D image of SU-8 substrate. (g) Equivalent circuit model for graphene microelectrode arrays, regardless of substrate. Rs is the series resistance. CPE is the constant phase element representing the Helmholtz double-layer capacitance. Cq represents quantum capacitance. Rct represents the charge-transfer resistance, and Wb is a bounded Warburg element representing any diffusion processes. (h) Accelerated aging test schematic showing the graphene microelectrode arrays suspended in 87 °C PBS solution. A hotplate is used to keep the PBS constantly heated, and a thermometer is held next to the arrays in order to confirm the temperature. The beaker is enclosed by a silicone stopper in order to prevent temperature changes and evaporation.

TABLE I. EQUIVALENT CIRCUIT MODEL FITTING VALUES FOR GRAPHENE MICROELECTRODE ARRAYS.

Substrate Rs
[KΩ]
Cdl
[µF cm−2]
A W
[MΩ s−½]
B
[s−½]
Rct
[MΩ]
Cq
[µF cm−2]
PET 4.879 ± .54 7.07 ± .14 .924 ±.085 178 ± 7.9 .565 ± 3.7e–4 74.91 ± 7.1 1.45 ± .06
SU-8 4.613 ± .39 7.41 ± 1.23 .868 ± .011 640 ± 13.6 .703 ± 3.4e–4 1.36 ± .13 3.72 ± .04

Graphene gives rise to an additional quantum capacitance component represented by Cq [6, 21] in series with Cdl. Computational values of Cdl (around 10 μF cm−2) are larger than computational values of Cq (around 1–4 μF cm−2) [21, 22]. Cdl is experimentally measured using Au electrodes [23], and Cq is computed by fitting EIS results to the equivalent circuit model shown in Fig. 2g. The quantum capacitance dominates the impedance in both graphene microelectrode arrays, regardless of the substrate. Our experimental values are thus consistent with theoretical predictions. The Rct and Wb values are both large, so the electrode behaves capacitively, which is consistent with well-known characteristics of graphene.

C. Accelerated Aging Test

Accelerated aging tests have been used to simulate in vivo lifetime for implantable devices [2427]. The general theory states that holding a device in elevated temperatures exponentially increases the rate of chemical reactions, thus speeding up the degradation of the materials to simulate long-term defects within a short experimental time. A commonly accepted approach is to use Eq. (1) as the factor of increased aging [24]. ΔT is equal to the difference between the temperature the device is held at (T) and the reference temperature (Tref). For our study, Tref is set at 37 °C, for that is the average body temperature of mice.

f=2ΔT/10 (1)

To simulate in vivo environments, we suspended the microelectrode arrays in a glass beaker and immersed them in 0.01 M PBS at pH 7.4, a buffer solution commonly used in accelerated aging tests due to the constant pH level [26]. The set up was enclosed by a silicone stopper due to its heat resistance [28], and the beaker was placed on a hot plate (Figure 2h). The solution was kept at 87 °C for an accelerated aging factor of 32 according to Eq. (1), so that 1 month in the aging test is equivalent to 32 months in vivo. Four graphene microelectrode arrays fabricated on a PET substrate and four graphene microelectrode arrays fabricated on an SU-8 substrate were used for a 30-day accelerated aging test. The impedance of working channels was tracked every three days for PET and every day for the SU-8 arrays which showed early failure.

D. Graphene Electrode Implantation for Chronic in vivo Experiments

All procedures were in accordance with protocols approved by the UCSD Institutional Animal Care and Use Committee and the guidelines of the National Institutes of Health. Adult mice (cross between CaMKIIa-tTA (JAX 003010) and tetO-GCaMP6s (JAX 024742), 2 months old) were anesthetized with isoflurance (3% for induction and 1% for maintenance) and a circular piece of scalp was removed. A custom-built head-post was implanted to the exposed skull (1 mm posterior to lambda) with cyanoacrylate glue and cemented with dental acrylic (Lang Dental). A stainless-steel screw (F000CE156, J.I. Morris) was implanted over right olfactory bulb as reference. A square craniotomy (4 × 4 mm2) was made over the left hemisphere and the graphene microelectrode array was implanted to cover visual cortices. A glass imaging window was placed on top of the electrode array and further secured with vetbond (3 M) and dental acrylic. To protect the array while the mice were in the home cage, a custom 3D-printed case was employed. Electrophysiological data were recorded by the RHD2000 amplifier board and RHD2000 evaluation system (Intan Technologies), which has a large input impedance which allows us to achieve high SNR with our relatively high impedance electrodes. It is important to note that relatively high impedance of graphene electrodes necessitates recording systems with large input impedance to prevent signal attenuation. The data was processed using custom Matlab code.

III. Results

A. Microelectrode Arrays on PET

All the microelectrode arrays fabricated on a PET substrate were working after the 30-day accelerated aging test. The average impedance of the microelectrodes over the 30-days is plotted along with the standard deviation in Figure 3a. The average impedance across all the channels from the four different arrays after the full duration of the test is 585 kΩ. The average impedance of the microelectrode arrays fabricated on PET substrate showed an initial drop but stabilized after a few days. This initial impedance drop is a common observation for electrodes soaked in solution for longer durations [29, 30]. The graphene arrays fabricated on PET substrate successfully completed the aging study with stable impedance characteristics, reaching an equivalent lifetime over 960 days (2.63 years). Figure 4a shows microscope images of a representative PET arrays captured on different days during one-month long study. Graphene microelectrodes fabricated on PET substrate did not exhibit any delamination, cracking or other forms of degradation until the end of the aging test.

Figure 3.

Figure 3.

Impedance graphs over accelerated aging test. (a) Average impedances along with the standard deviation of four graphene microelectrode arrays fabricated on PET substrate over 30 days of 2.63 equivalent years in vivo. (b) Average impedances along with the standard deviation of four graphene microelectrode arrays fabricated on SU-8 substrate over 30 days of 2.63 equivalent years in vivo. Failed arrows indicate when each array failed during the aging test. None of the four arrays made it the full 30 days.

Figure 4.

Figure 4.

Graphene microelectrode arrays fabricated on PET substrate micro images over time. (a) One graphene microelectrode array fabricated on PET substrate micro images over 30 days in the accelerated aging test. No visual differences between start and end of the experiment. (b) One graphene microelectrode array fabricated on PET substrate that displayed signs of PBS permeation. One electrode on the array is tracked over the course of 30 days. The PBS permeation gets slightly larger overtime but is never able to reach other electrode sites or the wires.

However, one of the PET arrays (array number 3) exhibited visual PBS permeation between the PET substrate and SU-8 encapsulation layer (Figure 4b). Liquid intrusion and substrate delamination are common sources of delamination for thin-film interfaces [3133]. The delamination can be attributed to the weaker adhesion between the SU-8 encapsulation layer and PET substrate, leading to the PBS leakage beneath the encapsulation surface. In order to precisely measure permeation distance, we processed microscope images in grey scale and adjusted the threshold to clearly identify the saline permeation as seen in Supplementary Figure S1. The size of the permeated region increased over the 30 days for the representative channel, with the maximum permeation distance reaching around 18 μm and percentage area increased to a maximum of ~14%. Since that is much smaller than interelectrode spacing (400 μm) between electrode sites, this minor permeation does not pose the risk of electrical shorting between the channels for our design. The impedance of all the PET arrays is stable at the end of the test, with an average impedance of 585 kΩ. The array which showed permeations has a lower average impedance (432 kΩ) consistent with the slight increase in the effective electrode area [34]. In summary, the results of the accelerated aging tests show that monolayer graphene microelectrodes fabricated on PET substrates demonstrate long-term durability required for chronic in vivo experiments.

B. Microelectrode Arrays on SU-8

All four graphene microelectrode arrays fabricated on SU-8 substrate succumbed to crack failures prior to the conclusion of the 30-day accelerated aging test. The average impedance of SU-8 arrays over the 30-days is plotted along with the standard deviation in Figure 3b. Graphene electrodes fabricated on SU-8 substrate also showed an initial drop like the PET arrays but stabilized after a few days. The final average impedance of all working channels across the arrays is 684 kΩ. None of the arrays showed any signs of PBS delamination (Figures 5ac). However, two forms of crack formations occurred: cracks over the electrode site (Figures 5df) and cracks over the gold/chromium wires (Figures 5gi). The two arrays with cracks over the electrode site had a longer equivalent aging time (1.93 and 1.49 years), while the arrays with cracks over the wires failed earlier on (0.70 and 0.61 years). On the one hand, Figure 5f shows SEM images of an example crack over the electrode site. The crack completely deforms and penetrates the electrode site. On the other hand, Figure 5i shows that the cracks over the wires are deep and wide (around 20 μm). The crack fully punctures through the SU-8 encapsulation layer as well as the gold/chromium wires. This causes open failure and prevents any recordings of electrical signals or impedance measurements, which renders the arrays nonworking.

Figure 5.

Figure 5.

Graphene microelectrode arrays fabricated on SU-8 substrate post accelerated aging results. (a,b,c) A single electrode site on a graphene microelectrode array fabricated on SU-8 substrate over 22 days. (d,e,f) A separate electrode site on the same array in (a,b,c) tracked from day 15 to day 23. A large crack spontaneously formed on day 23, causing leakage and device failure. An SEM image of the crack is shown in (f), where a noticeable vertical distortion is created on the electrode site. (g,h,i) A different graphene microelectrode array fabricated on SU-8 substrate that suffered from crack formation along its metal wires. (g) shows the cracks cleanly penetrating numerous wires. SEM images in (h,i) clearly defines the crack’s thickness and depth.

A summary of all the arrays detailing equivalent lifetime age and failure mechanisms is listed in Table II. Our study concludes that graphene arrays fabricated on PET substrate meets the lifetime and durability requirements for chronic studies in animal models. In contrast, graphene arrays fabricated on SU-8 substrate consistently failed due to crack formation, making them unsuitable for chronic animal experiments.

TABLE II. SUMMARY OF ACCELERATED AGING TEST FOR GRAPHENE MICROELECTRODE ARRAYS.

Array Final Status Time Elapsed (days in 87 °C PBS) Equivalent Lifetime (years in 37 °C) Failure Mechanism
PET #1 Working 30 2.63 No failure
PET #2 Working 30 2.63 No failure
PET #3 Working 30 2.63 Minor PBS Permeation
PET #4 Working 30 2.63 No Failure
SU-8 #1 Failed 22 1.93 SU-8 cracks over electrode
SU-8 #2 Failed 17 1.49 SU-8 cracks over electrode
SU-8 #3 Failed 8 0.70 SU-8 cracks over wire
SU-8 #4 Failed 7 0.61 SU-8 cracks over wire

Some studies in literature have suggested SU-8 as a reliable material for neural applications [15, 35, 36] but crack formation in SU-8 substrates have also been reported in many studies [3739]. In our study, both graphene microelectrode arrays have an SU-8 encapsulation layer, but only the graphene microelectrode arrays fabricated on SU-8 substrate are prone to crack formation. Mechanical properties of several soft substrates are listed in Supplementary Table I. The two factors which requires attention are the thermal expansion coefficients and the fracture strain (elongation at break) percentages [40]. PET and SU-8 have similar thermal expansion coefficients, but there is a significant distinction between PET and SU-8’s fracture strain (elongation at break percentages). While PET and SU-8 behave thermo-mechanically similarly, PET is mechanically much more compliant. Our experimental observation that the cracks only form on the top encapsulation layer suggests that the SU-8/PET combination exhibits better thermomechanical stability compared to the SU8 encapsulation/substrate configuration. It has also been reported in a separate study that a PET substrate is able to form an adaptive layer when SU-8 is fabricated on top due to its similar thermal expansion coefficient as well as less adhesion between the two layers [41], compared to an SU-8 substrate with SU-8 encapsulation layer. The 7% difference in surface roughness between PET and SU-8 also has effects on the adhesion strength and crack formation. The SU-8’s larger effective surface area leads to a stronger adhesion strength [4244], which correlates to our observation that only PET showed a minor delamination issue. On the other hand, it has been shown that surface roughness does not contribute to crack formation failure modes [4547]. The risk of cracking for a graphene microelectrode array fabricated on SU-8 is notably high (100% in our study) such that the device does not demonstrate chronic reliability.

C. Chronic In Vivo Experiments

Based on the results of the accelerated aging tests, transparent graphene microelectrode arrays fabricated on a PET substrate were determined to be more reliable for chronic in vivo studies. Therefore, for chronic in vivo experiments, we only used transparent graphene microelectrodes fabricated on PET substrates. To further evaluate the reliability of the transparent graphene microelectrode arrays in chronic imaging and recording experiments, we implanted the graphene microelectrodes over the visual cortex of transgenic mice. The transgenic mice express GCaMP6s [48], which is a green fluorescent calcium biosensor that is widely used to image neural activity, in the cortical neurons. We have previously shown that graphene microelectrode arrays do not generate light-induced artifacts under 2-photon imaging or optical stimulation [6]. The photo-induced currents are intrinsically very weak and fast in graphene enabling artifact-free integration of 2-photon imaging or optogenetics with electrical recordings [49]. Similar to the results obtained during accelerated aging tests, the impedance of the electrodes drops in the first few days after the initial implantation and stays stable throughout the course of 45 days (Figure 6a). On the other hand, the noise of the array rises slightly after the implantation and remains stable towards the later days (Figure 6b). With the low noise, we successfully recorded the spontaneous electrocorticography (ECoG) signals with high fidelity for 45 days after implantation (Figure 6c, Supplementary Figure S2).

Figure 6:

Figure 6:

Chronic in vivo implantation of graphene microelectrode array fabricated on PET substrate. (a) Average impedance of working channels of transparent graphene microelectrode arrays built on PET over the course of 45 days after implantation. (b) Root-mean-square (rms) values of the noise of the electrical recordings throughout the course of the in vivo experiment. (c) Representative electrocorticography (ECoG) recordings on days 1, 7, and 45. All recordings were taken while the animal was awake. (d) ECoG visual responses under awake state. Stimulus-induced trial-averaged ECoG response of the awake mice. There is a typical on-response ~70 ms after the given stimulus at 0 seconds. (e) Stimulus-induced trial-averaged frequency response spectrogram. A persistent power increase between [140–200 Hz] is observed. (f) Pictures of the imaging window for chronic 2-photon imaging taken 13 days after implantation. Image on the right shows 2-photon microscope image taken at 400 µm depth under the yellow outlined area.

In addition, we observed stimulus-induced responses in multiple ECoG channels, showing delayed on-responses and elevated signal power across wide frequency bands (Figures 6d, 6e). During the experiment, we gave visual stimulus to the mice. Square-wave drifting grating stimuli (100% contrast, 0.04 cycles/degree, 3 cycles/sec, covering entire contralateral receptive field) were presented on a LCD monitor (30 × 38 cm2) positioned 15 cm away from the right eye using Psychtoolbox [50]. One of 12 orientations (30° apart) was presented for 4 seconds on each trial in pseudorandom order. Inter-stimulus-interval was varied between 7–9 seconds. We presented each orientation at least 10 times in a session. We also performed simultaneous high quality 2-photon calcium imaging in Layer IV of the visual cortex at 400 μm depth without causing light-induced artifacts in recordings. The imaging frame rate is set to ~29 Hz, and the imaging area covers 786 × 846 μm2. The neuron cell bodies are clearly identified due to the high transparency of the graphene electrodes and PET substrate (Figure 6f). Our results demonstrate that transparent graphene microelectrodes on PET achieve stable impedance, low noise and artifact-free recordings in chronic in vivo experiments combining electrophysiological recordings with 2-photon microscopy.

IV. Conclusion

We performed systematic accelerated aging tests to investigate reliability of transparent graphene microelectrode arrays fabricated on PET and SU-8 substrates. Graphene microelectrodes fabricated on PET substrate show high reliability with all functioning channels after 30 days of accelerated aging or 2.63 years equivalent in vivo lifetime. On the other hand, graphene microelectrodes fabricated on SU-8 substrate exhibited failures due to different forms of crack formation across the electrodes or wires. We further demonstrated the reliability of graphene microelectrode arrays fabricated on PET in chronic 45-day in vivo studies in mice. After 45 days, graphene electrodes still exhibit low noise and impedance and can reliably record ECoG during calcium imaging. Our studies conclude that compared to other substrate materials, PET shows superior reliability and is therefore an ideal substrate of the graphene microelectrodes arrays for chronic in vivo studies, combining simultaneous electrophysiology and optical imaging.

Supplementary Material

supp1-2979475

Acknowledgments

The authors acknowledge the Office of Naval Research (N000141612531), National Science Foundation (ECCS-1752241, ECCS-1734940), and NIH (EY029466, EB026180) for funding. This work was performed in part at the San Diego Nanotechnology Infrastructure (SDNI) of UCSD, a member of the National Nanotechnology Coordinated Infrastructure, which is supported by the National Science Fountain (Grant ECCS-15421-148)

Contributor Information

David Ding, Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA 92093 USA.

Yichen Lu, Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA 92093 USA.

Ruoyu Zhao, Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA 92093 USA.

Xin Liu, Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA 92093 USA.

Chawina De-Eknamkul, Nanoengineering Department, University of California San Diego, La Jolla, CA 92093 USA..

Chi Ren, Neurobiology Section, University of California San Diego, La Jolla, CA 92093 USA.

Armaghan Mehrsa, Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA 92093 USA.

Takaki Komiyama, Neurobiology Section, University of California San Diego, La Jolla, CA 92093 USA.

Duygu Kuzum, Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, CA 92093 USA.

REFERENCES

  • [1].Denk W, Strickler JH, and Webb WW, “Two-photon laser scanning fluorescence microscopy,” Science, vol. 248, no. 4951, pp. 73–76, 1990. [Online]. Available: https://science.sciencemag.org/content/248/4951/73.long. [DOI] [PubMed] [Google Scholar]
  • [2].Kurtz R, Fricke M, Kalb J, Tinnefeld P, and Sauer M, “Application of multiline two-photon microscopy to functional in vivo imaging,” Journal of neuroscience methods, vol. 151, no. 2, pp. 276–286, 2006. [DOI] [PubMed] [Google Scholar]
  • [3].Lu Y, Liu X, and Kuzum D, “Graphene-based neurotechnologies for advanced neural interfaces,” Current Opinion in Biomedical Engineering, vol. 6, pp. 138–147, 2018. [Google Scholar]
  • [4].Kuzum D et al. , “Transparent and flexible low noise graphene electrodes for simultaneous electrophysiology and neuroimaging,” Nature communications, vol. 5, p. 5259, 2014. [Online]. Available: https://www.nature.com/articles/ncomms6259.pdf. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Thunemann M et al. , “Deep 2-photon imaging and artifact-free optogenetics through transparent graphene microelectrode arrays,” Nat Commun, vol. 9, no. 1, p. 2035, May 23 2018, 10.1038/s41467-018-04457-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Lu Y et al. , “Ultralow Impedance Graphene Microelectrodes with High Optical Transparency for Simultaneous Deep Two-Photon Imaging in Transgenic Mice,” Advanced Functional Materials, vol. 28, no. 31, p. 1800002, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Song EB et al. , “Robust bi-stable memory operation in single-layer graphene ferroelectric memory,” Applied Physics Letters, vol. 99, no. 4, p. 042109, 2011. [Google Scholar]
  • [8].Park D-W et al. , “Graphene-based carbon-layered electrode array technology for neural imaging and optogenetic applications,” Nature communications, vol. 5, p. 5258, 2014. [Online]. Available: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4218963/pdf/ncomms6258.pdf. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Lee C, Wei X, Kysar JW, and Hone J, “Measurement of the elastic properties and intrinsic strength of monolayer graphene,” science, vol. 321, no. 5887, pp. 385–388, 2008. [Online]. Available: https://science.sciencemag.org/content/321/5887/385.long. [DOI] [PubMed] [Google Scholar]
  • [10].Geim AK and Novoselov KS, “The rise of graphene,” Nature materials, vol. 6, no. 3, pp. 183–191, 2007. [DOI] [PubMed] [Google Scholar]
  • [11].Sahni D et al. , “Biocompatibility of pristine graphene for neuronal interface: Laboratory investigation,” Journal of Neurosurgery: Pediatrics, vol. 11, no. 5, pp. 575–583, 2013. [DOI] [PubMed] [Google Scholar]
  • [12].Li N et al. , “Three-dimensional graphene foam as a biocompatible and conductive scaffold for neural stem cells,” Scientific reports, vol. 3, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Kook G, Lee S, Lee H, Cho I-J, and Lee H, “Neural probes for chronic applications,” Micromachines, vol. 7, no. 10, p. 179, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Cheung KC, Renaud P, Tanila H, and Djupsund K, “Flexible polyimide microelectrode array for in vivo recordings and current source density analysis,” Biosensors and Bioelectronics, vol. 22, no. 8, pp. 1783–1790, 2007. [DOI] [PubMed] [Google Scholar]
  • [15].Luan L et al. , “Ultraflexible nanoelectronic probes form reliable, glial scar–free neural integration,” Science Advances, vol. 3, no. 2, p. e1601966, 2017. [Online]. Available: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5310823/pdf/1601966.pdf. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Rodger DC et al. , “Flexible parylene-based multielectrode array technology for high-density neural stimulation and recording,” Sensors and Actuators B: chemical, vol. 132, no. 2, pp. 449–460, 2008. [Google Scholar]
  • [17].Guo L and DeWeerth SP, “PDMS-based conformable microelectrode arrays with selectable novel 3-D microelectrode geometries for surface stimulation and recording,” in 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2009: IEEE, pp. 1623–1626. [DOI] [PubMed] [Google Scholar]
  • [18].Agirregabiria M et al. , “Fabrication of SU-8 multilayer microstructures based on successive CMOS compatible adhesive bonding and releasing steps,” Lab on a Chip, vol. 5, no. 5, pp. 545–552, 2005. [Online]. Available: https://pubs.rsc.org/en/content/articlelanding/2005/LC/b500519a. [DOI] [PubMed] [Google Scholar]
  • [19].Wang Y et al. , “Electrochemical delamination of CVD-grown graphene film: toward the recyclable use of copper catalyst,” ACS nano, vol. 5, no. 12, pp. 9927–9933, 2011. [Online]. Available: https://pubs.acs.org/doi/10.1021/nn203700w. [DOI] [PubMed] [Google Scholar]
  • [20].Cogan SF, “Neural stimulation and recording electrodes,” Annu. Rev. Biomed. Eng, vol. 10, pp. 275–309, 2008. [Online]. Available: https://www.annualreviews.org/doi/pdf/10.1146/annurev.bioeng.10.061807.160518. [DOI] [PubMed] [Google Scholar]
  • [21].Xia J, Chen F, Li J, and Tao N, “Measurement of the quantum capacitance of graphene,” Nature nanotechnology, vol. 4, no. 8, p. 505, 2009. [DOI] [PubMed] [Google Scholar]
  • [22].Ji H et al. , “Capacitance of carbon-based electrical double-layer capacitors,” Nature communications, vol. 5, p. 3317, 2014. [DOI] [PubMed] [Google Scholar]
  • [23].Franks W, Schenker I, Schmutz P, and Hierlemann A, “Impedance characterization and modeling of electrodes for biomedical applications,” IEEE Transactions on Biomedical Engineering, vol. 52, no. 7, pp. 1295–1302, 2005. [DOI] [PubMed] [Google Scholar]
  • [24].Hemmerich KJ, “General aging theory and simplified protocol for accelerated aging of medical devices,” Medical Plastic and Biomaterials, vol. 5, pp. 16–23, 1998. [Google Scholar]
  • [25].Hukins D, Mahomed A, and Kukureka S, “Accelerated aging for testing polymeric biomaterials and medical devices,” Medical engineering & physics, vol. 30, no. 10, pp. 1270–1274, 2008. [DOI] [PubMed] [Google Scholar]
  • [26].Palopoli-Trojani K, Woods V, Chiang C-H, Trumpis M, and Viventi J, “In vitro assessment of long-term reliability of low-cost μECoG arrays,” 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2016: IEEE, pp. 4503–4506. [DOI] [PubMed] [Google Scholar]
  • [27].Woods V et al. , “Long-term recording reliability of liquid crystal polymer μECoG arrays,” Journal of neural engineering, vol. 15, no. 6, p. 066024, 2018. [Online]. Available: https://iopscience.iop.org/article/10.1088/1741-2552/aae39d/pdfhttps://iopscience.iop.org/article/10.1088/1741-2552/aae39d/meta. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Mathivanan L, Selvaraj M, Azim SS, and Balakrishnan K, “Evaluation of heat resistant properties of silicone based coatings by SEM and ac impedance techniques,” Progress in organic coatings, vol. 28, no. 2, pp. 113–116, 1996. [Google Scholar]
  • [29].Prasad A and Sanchez JC, “Quantifying long-term microelectrode array functionality using chronic in vivo impedance testing,” Journal of neural engineering, vol. 9, no. 2, p. 026028, 2012. [DOI] [PubMed] [Google Scholar]
  • [30].Sankar V, Patrick E, Dieme R, Sanchez JC, Prasad A, and Nishida T, “Electrode impedance analysis of chronic tungsten microwire neural implants: understanding abiotic vs. biotic contributions,” Frontiers in neuroengineering, vol. 7, p. 13, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Ortigoza-Diaz J, Scholten K, and Meng E, “Characterization and modification of adhesion in dry and wet environments in thin-film Parylene systems,” Journal of Microelectromechanical Systems, no. 99, pp. 1–12, 2018. [Google Scholar]
  • [32].Alpern P and Lee KC, “Moisture-induced delamination in plastic encapsulated microelectronic devices: A physics of failure approach,” IEEE Transactions on Device and materials reliability, vol. 8, no. 3, pp. 478–483, 2008. [Google Scholar]
  • [33].Kim BJ, Washabaugh EP, and Meng E, “Annealing effects on flexible multi-layered parylene-based sensors,” in 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS), 2014: IEEE, pp. 825–828. [Google Scholar]
  • [34].Ahmed R and Reifsnider K, “Study of influence of electrode geometry on impedance spectroscopy,” Int. J. Electrochem. Sci, vol. 6, pp. 1159–1174, 2011. [Google Scholar]
  • [35].Cho S-H, Lu HM, Cauller L, Romero-Ortega MI, Lee J-B, and Hughes GA, “Biocompatible SU-8-based microprobes for recording neural spike signals from regenerated peripheral nerve fibers,” IEEE Sensors Journal, vol. 8, no. 11, pp. 1830–1836, 2008. [Google Scholar]
  • [36].Altuna A et al. , “SU-8 based microprobes with integrated planar electrodes for enhanced neural depth recording,” Biosensors and Bioelectronics, vol. 37, no. 1, pp. 1–5, 2012. [DOI] [PubMed] [Google Scholar]
  • [37].Bystrova S, Luttge R, and van den Berg A, “Study of crack formation in high-aspect ratio SU-8 structures on silicon,” Microelectronic engineering, vol. 84, no. 5–8, pp. 1113–1116, 2007. [Google Scholar]
  • [38].Wouters K and Puers R, “Diffusing and swelling in SU-8: insight in material properties and processing,” Journal of Micromechanics and Microengineering, vol. 20, no. 9, p. 095013, 2010. [Google Scholar]
  • [39].Das A, Sinha A, Rao V, and Jonnalagadda K, “Fracture in microscale SU-8 polymer thin films,” Experimental Mechanics, vol. 57, no. 5, pp. 687–701, 2017. [Google Scholar]
  • [40].Alkhadra MA, Root SE, Hilby KM, Rodriquez D, Sugiyama F, and Lipomi DJ, “Quantifying the Fracture Behavior of Brittle and Ductile Thin Films of Semiconducting Polymers,” Chemistry of Materials, vol. 29, no. 23, pp. 10139–10149, 2017. [Google Scholar]
  • [41].Abgrall P, Lattes C, Conédéra V, Dollat X, Colin S, and Gué AM, “A novel fabrication method of flexible and monolithic 3D microfluidic structures using lamination of SU-8 films,” Journal of Micromechanics and Microengineering, vol. 16, no. 1, p. 113, 2005. [Google Scholar]
  • [42].Meine K, Kloss K, Schneider T, and Spaltmann D, “The influence of surface roughness on the adhesion force,” Surface and Interface Analysis: An International Journal devoted to the development and application of techniques for the analysis of surfaces, interfaces and thin films, vol. 36, no. 8, pp. 694–697, 2004. [Google Scholar]
  • [43].Kim SH, Na SW, Lee N-E, Nam YW, and Kim Y-H, “Effect of surface roughness on the adhesion properties of Cu/Cr films on polyimide substrate treated by inductively coupled oxygen plasma,” Surface and Coatings Technology, vol. 200, no. 7, pp. 2072–2079, 2005. [Google Scholar]
  • [44].Awaja F, Gilbert M, Kelly G, Fox B, and Pigram PJ, “Adhesion of polymers,” Progress in polymer science, vol. 34, no. 9, pp. 948–968, 2009. [Google Scholar]
  • [45].Chakrabarti B and Chakrabarti BK, “Crack propagation in solids and crack-surface roughness,” Physica A: Statistical Mechanics and its Applications, vol. 270, no. 1–2, pp. 21–26, 1999. [Google Scholar]
  • [46].Lee HT and Tai TY, “Relationship between EDM parameters and surface crack formation,” Journal of Materials Processing Technology, vol. 142, no. 3, pp. 676–683, 2003. [Google Scholar]
  • [47].Fukahori Y and Andrews E, “Fracture surface roughness in highly deformable polymers,” Journal of Materials Science, vol. 13, no. 4, pp. 777–785, 1978. [Google Scholar]
  • [48].Chen T-W et al. , “Ultrasensitive fluorescent proteins for imaging neuronal activity,” Nature, vol. 499, no. 7458, p. 295, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [49].Liu X, Lu Y, Iseri E, Shi Y, and Kuzum D, “A compact closed-loop optogenetics system based on artifact-free transparent graphene electrodes,” Frontiers in neuroscience, vol. 12, p. 132, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]