Accuracy of high-frequency oscillations recorded intraoperatively for classification of epileptogenic regions

Shennan A Weiss 1,2,3, Richard J Staba 6, Ashwini Sharan 11, Chengyuan Wu 11, Daniel Rubinstein 10, Sandhitsu Das 12, Zachary Waldman 10, Iren Orosz 6, Gregory Worrell 4,5, Jerome Engel Jr 6,7,8,9, Michael R Sperling 10,
PMCID: PMC8560764  PMID: 34725412

Abstract

To see whether acute intraoperative recordings using stereo EEG (SEEG) electrodes can replace prolonged interictal intracranial EEG (iEEG) recording, making the process more efficient and safer, 10 min of iEEG were recorded following electrode implantation in 16 anesthetized patients, and 1–2 days later during non-rapid eye movement (REM) sleep. Ripples on oscillations (RonO, 80–250 Hz), ripples on spikes (RonS), sharp-spikes, fast RonO (fRonO, 250–600 Hz), and fast RonS (fRonS) were semi-automatically detected. HFO power and frequency were compared between the conditions using a generalized linear mixed-effects model. HFO rates were compared using a two-way repeated measures ANOVA with anesthesia type and SOZ as factors. A receiver-operating characteristic (ROC) curve analysis quantified seizure onset zone (SOZ) classification accuracy, and the scalar product was used to assess spatial reliability. Resection of contacts with the highest rate of events was compared with outcome. During sleep, all HFOs, except fRonO, were larger in amplitude compared to intraoperatively (p < 0.01). HFO frequency was also affected (p < 0.01). Anesthesia selection affected HFO and sharp-spike rates. In both conditions combined, sharp-spikes and all HFO subtypes were increased in the SOZ (p < 0.01). However, the increases were larger during the sleep recordings (p < 0.05). The area under the ROC curves for SOZ classification were significantly smaller for intraoperative sharp-spikes, fRonO, and fRonS rates (p < 0.05). HFOs and spikes were only significantly spatially reliable for a subset of the patients (p < 0.05). A failure to resect fRonO areas in the sleep recordings trended the most sensitive and accurate for predicting failure. In summary, HFO morphology is altered by anesthesia. Intraoperative SEEG recordings exhibit increased rates of HFOs in the SOZ, but their spatial distribution can differ from sleep recordings. Recording these biomarkers during non-REM sleep offers a more accurate delineation of the SOZ and possibly the epileptogenic zone.

Subject terms: Epilepsy, Diagnostic markers

Introduction

For patients with medically refractory epilepsy who are candidates for epilepsy surgery, invasive intracranial EEG (iEEG) monitoring is often required. Intraoperative electrocorticography (ECoG) using subdural electrodes1, and occasionally intrahippocampal depth electrodes2,3, are often used in place of prolonged iEEG recordings in childrenn4,5, and in adults with a clear pathological substrate1,6,7. It is unclear if acute recordings from stereo EEG (SEEG) electrodes could similarly be used to reduce or obviate the need for prolonged epilepsy monitoring unit (EMU) evaluations to capture seizures.

ECoG can identify brain regions with high rates of inter-ictal epileptiform spikes1 and help localize the epileptogenic zone (EZ), which is the hypothetical region necessary and sufficient for seizure generation8. Results from several studies suggest high-frequency oscillations (HFOs: ripples 80–250 Hz, fast ripples 250–600 Hz)3,4,6,7,911 is a more specific biomarker of EZ than interictal spikes. Ripples occur more frequently than fast ripples but are thought to be less specific, because they are more often generated by physiological mechanisms9,10. Several groups have reported that unresected fast ripples in intraoperative ECoG recordings predict poor post-operative seizure outcome3,4,6,7,1115. When spikes and HFOs coincide, or ripples and fast ripples coincide15, the biomarker is thought to be even more specific for EZ3,1618.

Few studies have compared the spatial accuracy of spikes and HFOs in intraoperative ECoG recordings to extraoperative prolonged recordings for identifying the EZ1923. Studies have found anesthesia and analgesia can affect the rate of these biomarkers24, and also the spatial distribution with respect to the EZ23. A recent meta-analysis22 suggests that these studies report conflicting results. Whether anesthesia affects the morphology of HFOs is also unknown. Since electrocorticography is commonly performed, and HFOs may replace or supplement spikes in tailoring resection during ECoG25, the accuracy of spikes and HFOs for the EZ in intraoperative recordings must be further clarified with respects to the influence of anesthesia and analgesia.

Prolonged recordings using subdural or SEEG electrodes are advantageous compared to intraoperative recordings, because they are much more likely to capture spontaneous seizures that specify the seizure onset zone (SOZ). The EZ may not always encompass all of the SOZ due to both volume conduction and a possible ictal core region26,27, and conversely localizing the SOZ may not identify the complete EZ because of limited temporal sampling during the EMU evaluation. Evidence suggests that HFOs recorded from prolonged implants during non-rapid eye movement (non-REM) sleep may identify the EZ better than the SOZ28,29. One contentious issue is whether the spatial distribution of HFO rates during non-REM sleep is stable15,30,31. One recent large study found that the classification accuracy of fast ripples for the EZ was higher when a full night of non-REM sleep was analyzed as opposed to 5-min segments32. This issue is of particular importance to comparing intraoperative to sleep recordings, since the duration of the intraoperative recordings are intrinsically limited.

In this study, we used intraoperative and extra-operative non-REM sleep recordings from SEEG electrodes, to determine whether the morphology of HFOs is altered by anesthesia. We also compared the rates of HFOs and sharp-spikes (i.e. spikes that when digitally filtered have HFO frequency content 80–600 Hz, but do not have a distinct HFO on time frequency analysis)3,33,34, recorded in the two conditions, with each other and the location of both the clinically defined SOZ and the EZ. The EZ was defined as those electrode contacts with the highest rates of events, and then assessing whether removal of these electrodes or EZ was predictive of the post-operative seizure outcome. Our goal was to determine whether intraoperatively recorded HFOs using SEEG electrodes could obviate the need for prolonged iEEG.

Results

Patient characteristics, anesthesia, and analgesia

A total of 16 patients with medically refractory epilepsy were included in this study, 11 of them were male (Table S1). The patients had diverse etiologies of their epilepsy (Table S1) with 25% who had non-lesional epilepsy on MRI. Four patients had mesial-temporal lobe epilepsy (MTLE); 2 patients had lateral temporal lobe epilepsy; 4 patients had temporal lobe epilepsy plus other region(s); and 6 patients had extra-temporal lobe epilepsy (Table S1). In the operating room, 2 of the patients were anesthetized with ~ 1% exhaled sevoflurane, 12 patients with ~ 2% exhaled sevoflurane, and 2 patients with propofol 90 mcg/kg/min. As analgesia 7 patients received a continuous infusion of ~ 0.8 ug/kg/h remifentanil, 3 patients received ~ 1.6 ug/kg/h remifentanil, 1 patient received 0.05 ug/kg/h sufentanil. Five patients did not receive analgesia after induction (Table S1).

Differences in HFO power and frequency during intraoperative recordings compared to non-REM sleep

Since we quantified the power and spectral frequency of each HFO event, we first asked if anesthesia effected these measures. We used a generalized linear mixed-effects model to account for inter-contact differences in the matched electrodes (n = 530). We found that HFO power was decreased in the intraoperative recordings relative to the non-REM recordings for all HFO types, except for fRonO (p < 0.01, Fig. 1, Table 1). We next asked if this difference was influenced by anesthesia type. For RonO, we found the largest relative differences in the mean power for sevoflurane 2% and propofol compared to sevoflurane 1% (Figure S1, Table S2, p < 0.05). However, this comparison was indirect such that anesthesia related differences in ripple power were not compared across the intraoperative condition. Furthermore, the two patients that received sevoflurane 1% had smaller RonO power during sleep than the others, which could be related to their poor outcome (Table S1).

Figure 1.

Figure 1

Differences in HFO average power and frequency by recording condition (intraoperative vs. non-REM sleep). Violin plots of ripples on oscillations (RonO, A), ripples on spikes (RonS, B), fast ripples on oscillations (fRonO, C), and fast ripples on spikes (fRonS, D) average power. HFO power was greater for all the event types, except fRonO, during non-REM sleep (red) than intraoperative recording (blue, p < 0.01). (E) Violin plots of HFO spectral frequency. There was a small, but significant, increase in RonO and RonS frequency (p < 0.01) and decrease in fRonO and fRonS frequency (p < 0.01) during non-REM sleep than during intraoperative recording. Asterisk indicates mean.

Table 1.

Results of generalized linear mixed-effects models fitting HFO frequency and power.

Response variable Intercept estimate Intercept p value Condition estimate Condition p value SOZ estimate SOZ p value d.f
RonO Power 12.287 [12.2–12.4] < 1e−4 1.291 [1.14–1.3] < 1e−4 0.263 [0.08–0.4] < 0.005 196,470
RonO Freq 4.734 [4.71–4.75] < 1e−4 0.03 [0.027–0.032] < 1e−4 − 0.025 [− 0.05 to 0] 0.08 196,470
RonS Power 14.912 [14.6–15.2] < 1e−4 0.97 [0.77–1.17] < 1e−4 − 0.101 [− 0.46 to .26] 0.58 31,479
RonS Freq 4.678 [4.66–4.68] < 1e−4 0.022 [0.016–0.029] < 1e−4 0.003 [− 0.013 to .02] 0.7 31,479
fRonO Power 12.251 [11.7–13.3] < 1e−4 0.3 [− 0.5 to 1.11] 0.47 0.303 [0.03–0.57] 0.03 8416
fRonO Freq 5.755 [5.74–5.77] < 1e−4 − 0.1877 [− 0.2 to − 0.17] < 1e−4 − 0.001 [− 0.02 to − 0.004] 0.003 8416
fRonS Power 13.602 [12.6–14.5] < 1e−4 1.208 [0.4–2.02] 0.004 − 0.209 [− 1.0 to .63] 0.62 2803
fRonS Freq 5.706 [5.68–5.72] < 1e−4 − 0.115 [− 0.13 to − 0.09] < 1e−4 0.029 [0.013–0.045] < 1e−4 2803

The random-effect term was electrode, the fixed effects were recording condition (intraoperative vs. non-REM sleep), and SOZ. Brackets indicate 95% confidence interval.

RonO and RonS spectral frequency was slightly decreased in the intraoperative recordings relative to the sleep recordings, and fRonO and fRonS frequency was increased in the intraoperative recordings (p < 0.01, Fig. 1, Table 1). We also examined whether the SOZ influenced HFO power and frequency, and found a relatively weak but sometimes significant correlation (Table 1).

Effects of non-REM sleep and anesthesia dose and type on HFO rate in SOZ and non-SOZ

We next asked for each of the HFO subtypes and sharp-spikes: (1) whether rates differed in the intraoperative and non-REM sleep condition; (2) whether rates were increased in the SOZ; (3) if the intraoperative and non-REM sleep recordings differed in the magnitude or significance of this effect; and (4) whether anesthesia and analgesia selection were significant factors interacting with this effect. We found that rates of some of the HFO subtypes were comparatively increased during the non-REM sleep recordings compared to the intraoperative recordings in the matched contacts (repeated measures two-way ANOVA RonO: F = 113.9, p < 0.001; fRonO: F = 36.15, p < 0.001; fRonS: F = 26.34, p < 0.001, Fig. 2A, Figure S2 d.f. = 524,1,2,1,2). However, the rate of RonS (F = 6.79, p < 0.01) and sharp-spikes (F = 12.68, p < 0.001) was higher in the intraoperative recording than during the sleep recording. The rate of HFOs and sharp-spikes was higher in the SOZ than in the non-SOZ during the combined sleep and intraoperative recordings (RonO: F = 41.16, p < 0.001; RonS: F = 9.97, p < 0.005; sharp-spike: F = 13.81, p < 0.001; fRonO: F = 7.0124, p < 0.01; fRonS: F = 12.10, p < 0.001, Fig. 2A). The difference in rates was larger for all event during non-REM sleep than intraoperative recording (RonO:F = 35.65, p < 0.001; RonS: F = 4.52, p < 0.05; sharp-spike: F = 9.66, p < 0.005; fRonO: F = 5.44, p < 0.05; fRonS: F = 14.05, p < 0.001, Fig. 2A).

Figure 2.

Figure 2

Mean HFO and spike rates in relation to the SOZ, recording condition, and anesthesia. (A1–5) A comparison of mean event rates in the non-SOZ (blue) and SOZ (orange) in the intraoperative (left) and sleep (right) condition measured from matched electrode contacts. The rate of RonO (A1), fRonO (A3), and fRonS (A4) was higher during sleep than intraoperative recording (rmANOVA, p < 0.05, N = 16 patients, d.f. = 523,1,1,2 contacts). For both conditions combined, the rate was higher in the SOZ than the non-SOZ for all the individual biomarkers (p < 0.01). However, the difference in the rate of HFOs between SOZ and NSOZ was larger during sleep than intraoperative recordings (A1–5, p < 0.05). (B) Differences in HFO and spike rates (intraoperative minus sleep) for each matched electrode contact with respect to anesthesia type and dose. In most cases, rates of HFO were lower during anesthesia than sleep, indicated by negative values in bar chart. Note RonS and spikes were higher during 2% sevoflurane than non-REM sleep. Error bars indicate standard error of the mean (S.E.M).

HFO generation has been shown to vary by neuroanatomic region3539. We investigated if HFOs and sharp-spikes rates varied by neuroanatomical lobe, and if this location interacted with the effect of the recording condition and the SOZ in the matched contacts. We found that the neuroanatomic location, by lobe, significantly influenced HFO and sharp-spike rates (Figure S3, Table S3). In contrast to other regions, higher ripple and sharp-spike rates, were seen in the frontal lobe, and to a less extent the parietal lobe, in the intraoperative recording relative to the sleep recording (Figure S3). Consequently, for all biomarker rates, except fRonO, the neuroanatomic location of the contact significantly interacted with the recording condition (Table S3). A weak but significant (p < 0.05) interaction was found between location and the SOZ for ripples and sharp-spikes only (Table S3). For all the biomarkers, the three-way interaction between SOZ, location, and condition was not significant. General anesthesia dose and type interacted with recording condition (intraoperative vs. non-REM sleep) to influence HFO rates for all HFO subtypes (RonO:F = 15.35, p < 0.001; RonS: F = 8.43, p < 0.001 sharp-spike: F = 8.24, p < 0.001; fRonO: F = 4.52, p < 0.05; fRonS: F = 5.45, p < 0.01, Fig. 2B). For sharp-spikes, fRonO, and fRonS, but not RonO or RonS, there was also a three-way interaction (p < 0.05) between SOZ, anesthesia type, and the recording condition. However, we did not investigate this interaction further because of the small number of patients in this study.

The greatest decrease in HFO and spike rates was seen with propofol anesthesia, whereas higher concentrations of sevoflurane increased the rates of RonS and sharp-spikes (Fig. 2B). Visual analysis of the annotated iEEG revealed that ~ 2% sevoflurane anesthesia often resulted in high amplitude sharply contoured rhythmic oscillations that were sometimes recognized as spikes by the detector (Figure S4), whereas propofol at a dose of 90 mcg/kg/min did not produce a burst suppression pattern.

Analgesia infusion selection also significantly interacted with the condition of the recording for all HFO types and spikes, except RonO (p < 0.05). Paradoxically RonS and sharp-spike relative rates were higher in the patients who received no analgesia after induction, Figure S5. We could not investigate the interaction between choice of analgesia and anesthesia because the two were unmatched. There was no significant interaction between analgesia selection, condition, and the SOZ.

Classification of the SOZ using intraoperative and sleep HFO rates

Since HFO rates were increased in the SOZ in the intraoperative and non-REM sleep recordings, we next used receiver-operating characteristic (ROC) curve analysis to assess how well the rate of HFOs on each matched electrode contact (n = 530) during intraoperative and non-REM sleep recordings classified the SOZ. Using boot-strap analysis, we found that the area under the ROC (AUROC) curve was larger for RonO and RonS during non-REM sleep recordings than during the intraoperative recordings, and was significantly larger for sharp-spikes, fRonO, and fRonS (p < 0.05, n = 1000 surrogates, Fig. 3A–E). RonO had the largest AUROC and was 0.72 [95% confidence interval (CI) 0.66–0.77] for intraoperative recordings, and 0.80 [95% CI 0.75–0.84] for sleep recordings. During the intraoperative and sleep recordings, the AUROC curve was significantly smaller for fRonO and fRonS than RonO (p < 0.05), but this may be due to a plateau effect at a limited specificity seen in the curves for fRonO and fRonS.

Figure 3.

Figure 3

Comparison of receiver-operating characteristic (ROC) curves for different HFO subtype and sharp spike rates (AE) in classifying the SOZ during intraoperative recordings (n = 530 contacts, black) and the same matched electrode contacts during non-REM sleep recordings (n = 530, red). Dashed lines indicate 95% confidence intervals (CI) for the ROC curves. Only sharp-spikes and fRonS had a significantly larger area under the ROC curve (AUROC) in the sleep recordings (p < 0.05). (F) ROC curves for the HFO subtype and sharp spike rates using all contacts (n = 2630), including those unmatched, for classifying the SOZ sites during the non-REM sleep recordings (F1). At high specificities (F2), the partial AUROC of RonO and fRonO were inferior to sharp-spikes and HFOs on spikes (p < 0.5). Black lines indicate 95% CI.

Since only a subset of the total number of electrodes were recorded in the intraoperative setting and in some cases, were outside the SOZ, we repeated the ROC curve analysis using rates of HFOs recorded from all electrodes during non-REM sleep (n = 2630), including those that were unmatched. We found that at specificities > 80%, RonO and fRonO were inferior to sharp-spikes and HFOs on spikes for classifying the SOZ (Fig. 3F, p < 0.05).

Differences in the spatial distribution of HFOs and sharp-spikes in the intraoperative and non-REM sleep condition

To better understand the within-patient differences in HFO generation in the intraoperative and non-REM sleep recordings we calculated the scalar product for all pairs of HFO vectors during the two conditions using the paired contacts. We found that in some patients, HFOs and spikes were very reliable between the two conditions (Fig. 4A, p < 0.05, n = 2000 surrogates). However, in other patients, the spatial distribution was less reliable or completely different (Fig. 4B, Figure S6). Overall, fRonO were the least reliable of all the event types across the 16 patients (Fig. 4A).

Figure 4.

Figure 4

Evaluation of reliability in the spatial distribution of HFOs and sharp-spikes during the intraoperative and non-REM sleep recordings from matched electrode contacts. (A) The reliability measure computed from the scalar product for all pairs of HFO vectors between the two conditions for the 16 patients. Asterisk indicates significant reliability (p < 0.05) with respect to a random distribution (n = 2000 surrogates). (B) Intraoperative recording from patient IO015 shows RonS (pink and blue) in lateral temporal contacts (LB7,8,LC7), but during non-REM sleep recordings RonS, RonOs (pink), and fRonOs (green) appeared on in a mesial temporal contact (LC1), which was part of the SOZ.

Comparison of resection of HFO and sharp-spike positive regions in individual patients with post-operative seizure outcome

To determine if HFOs and sharp-spikes recorded intraoperatively and during sleep define the EZ, we used an established method6,7,15 to define the contacts that exhibited the highest rate of events in each patient as the putative EZ. In eight patients, post-resection imaging was available, and we asked if the putative EZ was completely within the resection cavity. Four of the patients were not seizure free after surgery (Table S1). In two of the eight patients the resected region was not sampled by the intraoperative recording. In the intraoperative recordings, we found that the EZ defined by fRonO and fRonS trended the most sensitive for predicting failure. However, specificity and PPV were low (Table 2). Interestingly, these same events recorded from the same matched contacts during sleep performed slightly worse, but this difference was not statistically significant (Table 2). In the sleep recordings from all contacts, including those unmatched, fRonO trended the most sensitive and accurate among all the biomarkers, and also trended better than almost all of the biomarkers in the intraoperative recordings which had a limited montage (Table 2). However, even in the sleep recordings from all the electrodes, two seizure free patients had electrodes with fRonO that were outside the margins of resection.

Table 2.

Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV, and accuracy of remaining HFO and sharp-spike (SSpike) positive regions (see “Methods” section) for predicting surgical failure (Engel Class ≥ 2) in the intraoperative and electrode matched non-REM sleep recordings (n = 6), and the sleep recordings from all contacts (all, n = 8).

Sensitivity Specificity PPV NPV Accuracy
RonO intraop 0.67 [0.11–0.88] 0.67 [0.22–0.96] 0.67 [0.22–0.96] 0.67 [0.36–1.0] 0.67 [0.22–0.96]
RonO sleep 0.67 [0.11–0.88] 0.67 [0.22–0.96] 0.67 [0.22–0.96] 0.67 [0.36–1.0] 0.67 [0.22–0.96]
RonO all 1.0 [0.63–1.0] 0.0 [0.0–0.36] 0.50 [0.03–0.65] 0.0 [0.0–0.0] 0.50 [0.09–0.75]
RonS intraop 0.67 [0.12–0.89] 1.0 [0.54–1.0] 1.0 [0.54–1.0] 0.75 [0.23–0.96] 0.83 [0.36–0.99]
RonS sleep 0.33 [0.0–0.64] 0.67 [0.23–0.96] 0.50 [0.12–0.88] 0.50 [0.04–0.78] 0.50 [0.12–0.88]
RonS all 0.75 [0.35–0.97] 0.25 [0.0–0.52] 0.50 [0.16–0.84] 0.50 [0.09–0.76] 0.50 [0.16–0.84]
SSpike intraop 0.67 [0.36–1.0] 0.67 [0.36–1.0] 0.67 [0.11–0.88] 0.67 [0.36–1.0] 0.67 [0.0–0.64]
SSpike sleep 0.33 [0.22–0.96] 0.67 [0.23–0.96] 0.50 [0.12–0.88] 0.50 [0.04–0.77] 0.50 [0.12–0.88]
SSpike all 0.75 [0.35–0.97] 0.25 [0.0–0.52] 0.50 [0.16–0.84] 0.50 [0.09–0.76] 0.50 [0.16–0.84]
fRonO intraop 1.0 [0.54–1.0] 0.33 [0.04–0.77] 0.60 [0.54–1.0] 1.0 [0.54–1.0] 0.67 [0.22–0.96]
fRonO sleep 0.67 [0.54–1.0] 0.33 [0.04–0.78] 0.50 [0.0–0.64] 0.50 [0.0–0.64] 0.5 [0.12–0.88]
fRonO all 1.0 [0.63–1.0] 0.50 [0.16–0.84] 0.67 [0.35–0.97] 1.0 [0.63–1.0] 0.75 [0.63–1.0]
fRonS intraop 1.0 [0.54–1.0] 0.33 [0.0–0.64] 0.60 [0.04–0.77] 1.0 [0.54–1.0] 0.67 [0.12–0.88]
fRonS sleep 0.67 [0.22–0.96] 0.33 [0.12–0.88] 0.50 [0.22–0.96] 0.50 [0.04–0.77] 0.50 [0.0–0.64]
fRonS all 0.75 [0.25–0.92] 0.25 [0.09–0.76] 0.50 [0.35–0.97] 0.50 [0.09–0.76] 0.50 [0.09–0.76]

Brackets indicate 95% confidence intervals.

Discussion

In summary, intraoperative recordings under anesthesia altered HFO morphology, rate, and in some patients, the spatial distribution compared to HFO from the same electrode contacts recorded during non-REM sleep. Overall, HFOs and sharp-spikes delineated the SOZ more accurately in the non-REM sleep than intraoperative recordings (Table 3). The neuroanatomic location of the SOZ did not significantly influence this effect. When the EZ was defined using all contacts during non-REM sleep, fRonO trended the best predicting post-operative seizure outcome, but the specificity and PPV were relatively low (Table 3). Thus, intraoperative SEEG recording is not a substitute for prolonged non-REM sleep recordings for identifying epileptogenic regions using inter-ictal biomarkers.

Table 3.

Summary of findings. Minus sign indicates negative change, plus sign indicates positive change, minus and equals sign indicates negative trend, but no significant change.

Bio-marker Effect of anesthesia on HFO power Effect of anesthesia on HFO spectral content Effect of anesthesia (sevoflurane 2%) on event rate Overall accuracy for SOZ (both conditions) Sensitivity for SOZ at high specificities (sleep) Effect of anesthesia on SOZ accuracy EZ accuracy, (non-REM sleep)
RonO *** ** −/= *
RonS + ** *** −/= *
Sharp−spike n/a n/a + * *** *
fRonO −/= ++ * * **
fRonS ++ * *** *

Single asterisk indicates relatively low value, double asterisk medium value, and triple asterisk relatively high value.

HFOs exhibited decreased power under anesthesia, but the difference did not meet significance for fRonO. Also, ripples exhibited a slight but significant decrease in frequency, while fast ripples a large increase in frequency under anesthesia. Fast ripples are thought to be generated by clusters of pathologically interconnected neurons (PIN clusters)40 or out of phase firing of pyramidal neurons41. Ripples are generated by synchronized inhibitory post-synaptic potentials42. Propofol and isoflurane are both GABAA receptor agonists, but isoflurane also inhibits NMDA receptors43. Our results suggest that, during HFO generation, propofol and isoflurane decrease the number of recruited synchronized neurons. In the case of fast ripples, this theory is supported by the notion that event spectral frequency is inversely proportional to the size of the network44. More work is required to determine whether these changes alter the specificity of the biomarkers for epileptogenic regions. One possibility is, that in the intraoperative recordings, fast ripples may behave more like ripples generated during non-REM sleep because anesthesia increases the spectral frequency.

HFOs except for RonS exhibited lower rates under anesthesia. Notably, RonS and sharp-spike rates were elevated in patients administered 2% isoflurane. Some studies have shown that higher concentrations of isoflurane increase the rate of spikes45, but in our study the increase may have been artifactual due to sharply contoured oscillations that would not have necessarily been marked as spikes on visual review. We did not redact these detections because of the ambiguity of the events, and the coinciding ripple events. Neuroanatomical differences may also have contributed, since RonS and sharp-spike rates were elevated during the intraoperative recording, relative to the sleep recording, in the frontal lobe and to a lesser degree the parietal lobe.

In comparing the anesthetics given in this study, propofol resulted in the largest decrease in HFO and spike rates in the absence of a burst-suppression pattern. A decrease in HFO rates with the use of propofol has been reported previously, and it is recommended to wean the propofol off before recording HFOs24. Opioid analgesia also decreased the rate of some HFO types and sharp-spikes. Typically, in electrocorticography recordings, these agents are given as boluses to increase spike rates46. In this study these agents were given as a constant infusion which has been shown in other studies to not increase spike rates47.

With respect to delineating the SOZ, HFO and sharp-spikes rates were relatively higher in the SOZ during non-REM sleep than during intraoperative recordings. Very few studies have compared the accuracy of intraoperative-recorded spikes and HFOs to localize the SOZ19,2224. In one study23, however, HFOs recorded under propofol could define the SOZ, but accuracy between intra- and extra-operative recordings was not performed. Anesthesia could disrupt the normal expression of HFOs and spikes by epileptogenic tissue, while in contrast non-REM sleep facilitates the generation of spikes and HFOs in the SOZ and the EZ29,48. Frontal and parietal lobe seizures occur more frequently during non-REM sleep49,50, and pathological ripples there preferentially occur during the transition between the up-down state5153. In other regions, such as the hippocampus10, the facilitation of pathological HFOs and spikes by the up-down state is less pronounced. To examine potential causes of the observed decrease in accuracy of HFOs for epileptogenic regions in the intraoperative condition, future experiments could examine changes to HFO phase amplitude coupling to slower oscillations under anesthesia.

Among the 8 patients with post-operative MRI, fRonO had the highest accuracy for localizing the EZ, which is consistent with many past studies relating resection of HFOs with outcome4,6,7,15,32,54,55. By contrast, fRonO rates had the lowest accuracy of all the HFO types for identifying the SOZ in both intraoperative and sleep recordings. This inconsistency highlights the differences between the SOZ and the EZ, and also the dangers of using the SOZ as a metric by which to evaluate the utility of HFOs for surgical planning8. Also, HFOs on spikes and sharp-spike rates were more sensitive for the SOZ at high specificities, but notably these events trended less accurate for delineating that EZ. The study was underpowered to detect significant differences between HFOs on oscillations and HFOs on spikes for delineating the EZ. Several studies have suggested that HFOs on spikes are more accurate3,54,56, and are tightly correlated with increases in single unit firing rate57. Results are consistent with the hypothesis pathological HFOs, and specifically fRonO, are a electrophysiological biomarker for localizing the EZ in epilepsy surgery.

This study was also underpowered to detect significant differences between the intraoperative and sleep recording of HFO to localize the EZ. The intraoperative recordings trended more accurate than the matched contact sleep recordings, this result may be spurious since the SOZ was often only partially sampled by the limited montage of the intraoperative electrode contacts. The accuracy of fRonO in the sleep recordings using all the contacts trended better than the intraoperative recordings. Thus, fRonOs in sleep recordings, with adequate spatial sampling, may be more accurate than intraoperative recordings for defining the EZ. Another indication that supports this conclusion is that fRonO were the least spatially reliable biomarker between the two recording conditions.

In several patients during intraoperative recordings, high rates of, HFOs on spikes and sharp-spikes were found outside the SOZ. One possible explanation is that these spikes were injury potentials from electrode implantation that have been previously reported in studies of intraoperative depth electrode recordings58. However, in the hippocampus, these injury potentials were found to last only 1 min after implantation59. In the majority of patients, high rates of ectopic spikes were not present. This is consistent with a prior report that found that spike rates in intraoperative recordings from hippocampal depth electrodes correlated with hippocampal sclerosis2. Besides the possibility of injury potentials, another potential cause of the poor spatial reliability of HFOs and sharp-spikes was that the sleep recordings were often longer in duration than the intra-operative recordings. Poor temporal sampling should influence low-rate events such as fRonO and fRonS most, which is consistent with our results.

Results from this study were drawn from relatively few patients. For the GLMM, ANOVA, and ROC analysis we increased statistical power by making contact-matched comparisons. We could not use this approach to analyze the accuracy of HFOs and sharp-spikes for the EZ. Also, less than half of the patients in this study had resections followed by post-operative MRI. A larger study is required to prove that the accuracy of HFOs and sharp-spikes is inferior in the intraoperative condition, and would also help to resolve the effects of specific types and doses of anesthesia on delineating the SOZ and the EZ.

As noted previously only a subset of the total number of electrodes were recorded in the operating room and in some cases, were outside the SOZ and EZ. This could be a reason why fRonOs recorded from all electrodes during sleep more accurately defined the EZ than the fRonOs recorded from the subset of electrodes in the operating room. Future experiments may require recording from all SEEG electrodes, but this should be carefully planned since this would prolong the patient’s exposure to anesthesia while connecting all the electrodes. Alternatively, anesthesia could be weaned while the electrodes are being connected which may also improve delineation of the SOZ and the EZ.

In summary, our results suggest that localization of the SOZ and likely the EZ is more accurate using HFOs recorded during sleep than during HFOs recorded in the operating room (Table 3). An appropriately powered, likely multi-centered, study is needed to verify these results. Anesthesia affects the morphology and rate of HFOs. The utility of intraoperative SEEG recordings could be improved by increasing recording duration, tapering patients off anesthesia, and perhaps utilizing opioid boluses. Prolonged SEEG recordings remain essential because they permit electroclinical correlation of the SOZ. However, for patients who cannot tolerate long-term monitoring or have circumscribed lesions on MRI, the utility of intraoperative SEEG deserves further exploration. A lack of spatial reliability, and the possibility of injury potentials remain a concern.

Methods

All patients underwent intracranial monitoring with depth electrodes between 2016 and 2018 at Thomas Jefferson University (TJU) for the purpose of localization of the SOZ (Table 1). Intraoperative recordings were obtained from the anesthetized patients 10–15 min following implantation of all SEEG electrodes. In the operating room, 10-min recordings were obtained from 3 to 7 depth electrodes per patient. We connected only a subset of the total number of electrodes in the operating room so as not to prolong the patient’s exposure to anesthesia. Depth electrodes selection was based on the results from the patient’s previous scalp EEG study. One-two days after implantation, for each patient a 10–60 min iEEG recording from all the depth electrodes containing large amplitude, delta-frequency slow waves (i.e., non-REM sleep) was selected for analysis. Only iEEG that was at least 4 h of seizure-free, and was free of low levels of muscle contamination, and other artifacts, was selected.

All recordings were referenced to an iEEG electrode contact positioned in the white matter. Clinical iEEG recordings (0.016–600 Hz) were acquired, at a 2 kHz sampling rate, from the depth electrode contacts using a Nihon–Kohden 256-channel JE-120 long-term monitoring system (Nihon-Kohden America, Foothill Ranch, CA, USA). The study was approved by the TJU institutional review boards (IRB Control #16F.592), and patients gave informed consent prior to participating in this research. All research was in accord with the office for human research protections.

The seizure onset zone was defined by the attending epileptologist for each patient and did not include areas of early propagation. The non-SOZ included all remaining contacts and was often separated from the SOZ by sub-centimeter distances (Table 1). In all the patients the SOZ was not completely sampled by the intraoperative recordings, in 4 patients, the SOZ was not sampled at all by the intraoperative recording contacts.

HFO detection

HFOs and sharp-spikes were detected in the intraoperative and sleep iEEG using previously published methods3,34,60 implemented in Matlab (Mathworks, Natick, MA, USA). In brief, the HFO detector reduced muscle and electrode artifacts in the iEEG recordings using a independent component analysis (ICA)-based algorithm60. After applying this ICA-based method, ripples and fast ripples were detected in the referential and bipolar montage iEEG recordings per contact by utilizing a Hilbert detector, in which a 1000th order symmetric finite impulse response (FIR) band-pass filter in the (80–600 Hz) for ripples and (250–600 Hz) for fast ripples was applied, and (ii) a Hilbert transform was applied to calculate the instantaneous amplitude of this time series according to the analytic signal z(t)

zt=ateiϕt 1

where a(t) is the instantaneous amplitude and ø(t) is the instantaneous phase of z(t). Following the Hilbert transform, the instantaneous HFO amplitude function [a(t)] was smoothed using moving window averaging, the smoothed instantaneous HFO amplitude function was normalized using the mean and standard deviation of the time series, and a custom statistical threshold defined by the skewness of the normalized time series was used to detect the onset and offset of discrete/potential events.

HFO-like events can arise due to Gibb's phenomenon, i.e., high-pass filtering sharp transients, including epileptiform spikes33. To distinguish authentic HFOs from authentic HFOs on EEG spikes or spurious HFO due to filter ringing, we used a custom algorithm that performed topographic analysis of time–frequency plots for each HFO34. The algorithm also measured the power, spectral content, and duration of each HFO and categorized the HFO as an HFO on oscillation or HFO on spike. Following automatic detection of HFO and sharp-spikes, false detections of clear muscle and mechanical artifact were deleted by visual review in Micromed Brainquick (Venice, Italy).In almost all the patients, less than 3–5% of the detections were deleted3.

Neuroimaging

T1- pre-implant and post-resection MRIs were obtained for each patient. Post-implantation CT scans were then co-registered with the MRIs using Advanced Neuroimaging Tools (ANTs)61 with neuroradiologist supervision, using an in-house pipeline (https://github.com/pennmem/neurorad_pipeline). The position of each electrode contact was localized to the Desikan-Killiany atlas62. Identification of the electrode contacts in the resection cavity was performed manually in itk-SNAP (http://www.itksnap.org/pmwiki/pmwiki.php).

Statistics

HFO frequency and power values were fit with generalized linear mixed-effects models in Matlab with contact as the random-effects term, and condition and SOZ as fixed-effects predictors. A comparison of HFO and sharp spike rates, by electrode contact, was carried out in Matlab using the RANOVA function with factors including SOZ, anesthesia or analgesia or neuroanatomic lobe of the electrode contact, and two levels for the recording condition. Receiver operating curves were generated using the perfcurve function in Matlab, and 95% confidence intervals were estimated using 1000 boot-strap replicas. We quantified the reliability of the distribution of HFO rates between the intraoperative recording and extraoperative recording. For each interval pair we computed the normalized scalar product of the spatial distribution of the HFO rates. To test the magnitude of the true scalar product against chance, we permuted (N = 2000) the order of channels for the intra-operative and sleep conditions15. For the outcome analysis, HFO and sharp spike positive regions were defined as the channels whose rates exceeded the 95th percentile of the distribution of all contacts during the recording (intra-op, matched sleep contacts, all sleep contacts). We defined as true positive a patient whose event area was not fully located in the margins of the resection and the post-operative seizure outcome was Engel IIa or worse. The positive predictive value was calculated as PPV = TP/(TP + FP), negative predictive value as NPV = TN/(TN + FN), sensitivity = TP/(TP + FN), specificity = TN/(TN + FP), and accuracy = (TP + TN)/N. Estimates of the 95% confidence intervals (CI) used the binomial method.

Supplementary Information

Acknowledgements

The authors would like to thank Dale and Ted Wyeth for their technical contribution as well as Dr. Lin Li. This work was fully supported by NIH K23 NS094633 (SAW), a Junior Investigator Award from the American Epilepsy Society (SAW), R01 NS106958 (RS), and R01 NS033310 (JE).

Author contributions

S.A.W.: conceptualization, methodology, software, investigation, writing original draft, review and editing, visualization, supervision, funding acquisition. R.J.S.: conceptualization, investigation, writing, funding acquisition. A.S.: investigation. C.W.: investigation. D.R.: investigation, visualization, writing S.D.: software. Z.W.: software, investigation, visualization. I.O.: investigation, writing. G.W.: conceptualization, writing, funding acquisition. J.E.: conceptualization, investigation, writing, funding acquisition. M.S.: conceptualization, methodology, software, investigation, writing, visualization, supervision, funding acquisition.

Data availability

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.

Competing interests

MS has served as Consultant/advisor: Medtronic, Neurelis; Speaker: International Medical Press, Medscape, Projects in Knowledge, Eisai, UCB Pharma Research support: Cavion, Cerevel, Eisai, Engage, Medtronic, Neurelis, SK Life Science, Takeda, UCB Pharma, Xenon. The other authors do not declare competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-021-00894-3.

References

  • 1.Keene DL, Whiting S, Ventureyra EC. Electrocorticography. Epileptic Disord. Int. Epilepsy J. Videotape. 2000;2:57–63. [PubMed] [Google Scholar]
  • 2.Cascino GD, Sharbrough FW, Jack CR, Parisi JE, O’Brien PC. Acute depth electrode investigations in temporal lobe epilepsy: Correlation with magnetic-resonance-imaging-based volume studies and pathology. J. Epilepsy. 1992;5:49–54. doi: 10.1016/S0896-6974(05)80020-9. [DOI] [Google Scholar]
  • 3.Weiss SA, et al. Visually validated semi-automatic high-frequency oscillation detection aides the delineation of epileptogenic regions during intra-operative electrocorticography. Clin. Neurophysiol. 2018;129:2089–2098. doi: 10.1016/j.clinph.2018.06.030. [DOI] [PubMed] [Google Scholar]
  • 4.Wu JY, et al. Removing interictal fast ripples on electrocorticography linked with seizure freedom in children. Neurology. 2010;75:1686–1694. doi: 10.1212/WNL.0b013e3181fc27d0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Gelinas JN, Battison AW, Smith S, Connolly MB, Steinbok P. Electrocorticography and seizure outcomes in children with lesional epilepsy. Child’s Nerv. Syst. 2011;27:381–390. doi: 10.1007/s00381-010-1279-7. [DOI] [PubMed] [Google Scholar]
  • 6.van’t Klooster MA, et al. Residual fast ripples in the intraoperative corticogram predict epilepsy surgery outcome. Neurology. 2015;85:1. doi: 10.1212/WNL.0000000000001727. [DOI] [PubMed] [Google Scholar]
  • 7.van’t Klooster MA, et al. Tailoring epilepsy surgery with fast ripples in the intraoperative electrocorticogram. Ann. Neurol. 2017;81:664–676. doi: 10.1002/ana.24928. [DOI] [PubMed] [Google Scholar]
  • 8.Jehi L. The epileptogenic zone: Concept and definition. Epilepsy Curr. 2018;18:12–16. doi: 10.5698/1535-7597.18.1.12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Engel J, Bragin A, Staba R. Nonictal EEG biomarkers for diagnosis and treatment. Epilepsia Open. 2018;3:120–126. doi: 10.1002/epi4.12233. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Weiss SA, et al. Ripples have distinct spectral properties and phase-amplitude coupling with slow waves, but indistinct unit firing, in human epileptogenic hippocampus. Front. Neurol. 2020;11:174. doi: 10.3389/fneur.2020.00174. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Fedele T, et al. Automatic detection of high frequency oscillations during epilepsy surgery predicts seizure outcome. Clin. Neurophysiol. 2016;127:3066–3074. doi: 10.1016/j.clinph.2016.06.009. [DOI] [PubMed] [Google Scholar]
  • 12.Burnos S, et al. The morphology of high frequency oscillations (HFO) does not improve delineating the epileptogenic zone. Clin. Neurophysiol. 2016;127:2140–2148. doi: 10.1016/j.clinph.2016.01.002. [DOI] [PubMed] [Google Scholar]
  • 13.Hussain SA, et al. Intraoperative fast ripples independently predict postsurgical epilepsy outcome: Comparison with other electrocorticographic phenomena. Epilepsy Res. 2017;135:79–86. doi: 10.1016/j.eplepsyres.2017.06.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Jacobs J, et al. Removing high-frequency oscillations. Neurology. 2018;91:e1040–e1052. doi: 10.1212/WNL.0000000000006158. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Fedele T, et al. Resection of high frequency oscillations predicts seizure outcome in the individual patient. Sci. Rep. 2017;7:13836. doi: 10.1038/s41598-017-13064-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Wang S, et al. Ripple classification helps to localize the seizure-onset zone in neocortical epilepsy. Epilepsia. 2013;54:370–376. doi: 10.1111/j.1528-1167.2012.03721.x. [DOI] [PubMed] [Google Scholar]
  • 17.Weiss SA, et al. Ripples on spikes show increased phase-amplitude coupling in mesial temporal lobe epilepsy seizure-onset zones. Epilepsia. 2016;57:1916–1930. doi: 10.1111/epi.13572. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Kumar U, Li L, Bragin A, Engel J. Spike and wave discharges and fast ripples during posttraumatic epileptogenesis. Epilepsia. 2021 doi: 10.1111/epi.16958. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Asano E, et al. Is intraoperative electrocorticography reliable in children with intractable neocortical epilepsy? Epilepsia. 2004;45:1091–1099. doi: 10.1111/j.0013-9580.2004.65803.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Kurita N, et al. The effects of sevoflurane and hyperventilation on electrocorticogram spike activity in patients with refractory epilepsy. Anesth. Analg. 2005;101:517–523. doi: 10.1213/01.ANE.0000158606.31021.1F. [DOI] [PubMed] [Google Scholar]
  • 21.Orihara A, et al. Effects of sevoflurane anesthesia on intraoperative high-frequency oscillations in patients with temporal lobe epilepsy. Seizure. 2020;82:44–49. doi: 10.1016/j.seizure.2020.08.029. [DOI] [PubMed] [Google Scholar]
  • 22.Bayram AK, et al. Effect of anesthesia on electrocorticography for localization of epileptic focus: Literature review and future directions. Epilepsy Behav. 2021;118:107902. doi: 10.1016/j.yebeh.2021.107902. [DOI] [PubMed] [Google Scholar]
  • 23.Inada T, et al. Effects of a stable concentration of propofol on interictal high-frequency oscillations in drug-resistant epilepsy. Epileptic Disord. 2021;23:299–312. doi: 10.1684/epd.2021.1264. [DOI] [PubMed] [Google Scholar]
  • 24.Zijlmans M, et al. Epileptic high-frequency oscillations in intraoperative electrocorticography: The effect of propofol. Epilepsia. 2012;53:1799–1809. doi: 10.1111/j.1528-1167.2012.03650.x. [DOI] [PubMed] [Google Scholar]
  • 25.van’t Klooster MA, et al. High frequency oscillations in the intra-operative ECoG to guide epilepsy surgery (“The HFO Trial”): Study protocol for a randomized controlled trial. Trials. 2015;16:422. doi: 10.1186/s13063-015-0932-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Weiss SA, et al. Ictal high frequency oscillations distinguish two types of seizure territories in humans. Brain. 2013;136:3796–3808. doi: 10.1093/brain/awt276. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Weiss SA, et al. Seizure localization using ictal phase-locked high gamma. Neurology. 2015;84:2320–2328. doi: 10.1212/WNL.0000000000001656. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Frauscher B, et al. High-frequency oscillations: The state of clinical research. Epilepsia. 2017;58:1316–1329. doi: 10.1111/epi.13829. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Klimes P, et al. NREM sleep is the state of vigilance that best identifies the epileptogenic zone in the interictal electroencephalogram. Epilepsia. 2019;60:2404–2415. doi: 10.1111/epi.16377. [DOI] [PubMed] [Google Scholar]
  • 30.Gliske SV, et al. Variability in the location of high frequency oscillations during prolonged intracranial EEG recordings. Nat. Commun. 2018;9:2155. doi: 10.1038/s41467-018-04549-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Fedele T, Ramantani G, Sarnthein J. High frequency oscillations as markers of epileptogenic tissue—End of the party? Clin. Neurophysiol. 2019;130:624–626. doi: 10.1016/j.clinph.2019.01.016. [DOI] [PubMed] [Google Scholar]
  • 32.Nevalainen P, et al. Association of fast ripples on intracranial EEG and outcomes after epilepsy surgery. Neurology. 2020;95:e2235–e2245. doi: 10.1212/WNL.0000000000010468. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Bénar CG, Chauvière L, Bartolomei F, Wendling F. Pitfalls of high-pass filtering for detecting epileptic oscillations: A technical note on “false” ripples. Clin. Neurophysiol. 2010;121:301–310. doi: 10.1016/j.clinph.2009.10.019. [DOI] [PubMed] [Google Scholar]
  • 34.Waldman ZJ, et al. A method for the topographical identification and quantification of high frequency oscillations in intracranial electroencephalography recordings. Clin. Neurophysiol. 2018;129:308–318. doi: 10.1016/j.clinph.2017.10.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Alkawadri R, et al. The spatial and signal characteristics of physiologic high frequency oscillations. Epilepsia. 2014;55:1986–1995. doi: 10.1111/epi.12851. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Frauscher B, et al. High-frequency oscillations in the normal human brain. Ann. Neurol. 2018;84:374–385. doi: 10.1002/ana.25304. [DOI] [PubMed] [Google Scholar]
  • 37.Guragain H, et al. Spatial variation in high-frequency oscillation rates and amplitudes in intracranial EEG. Neurology. 2018;90:e639–e646. doi: 10.1212/WNL.0000000000004998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Motoi H, et al. Quantitative analysis of intracranial electrocorticography signals using the concept of statistical parametric mapping. Sci. Rep. 2019;9:17385. doi: 10.1038/s41598-019-53749-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Salami P, et al. Seizure onset location shapes dynamics of initiation. Clin. Neurophysiol. 2020;131:1782–1797. doi: 10.1016/j.clinph.2020.04.168. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Bragin A, Mody I, Wilson CL, Engel J. Local generation of fast ripples in epileptic brain. J. Neurosci. 2002;22:2012–2021. doi: 10.1523/JNEUROSCI.22-05-02012.2002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Foffani G, Uzcategui YG, Gal B, de la Prida LM. Reduced spike-timing reliability correlates with the emergence of fast ripples in the rat epileptic hippocampus. Neuron. 2007;55:930–941. doi: 10.1016/j.neuron.2007.07.040. [DOI] [PubMed] [Google Scholar]
  • 42.Stark E, et al. Pyramidal cell-interneuron interactions underlie hippocampal ripple oscillations. Neuron. 2014;83:467–480. doi: 10.1016/j.neuron.2014.06.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Brohan J, Goudra BG. The role of GABA receptor agonists in anesthesia and sedation. CNS Drugs. 2017;31:845–856. doi: 10.1007/s40263-017-0463-7. [DOI] [PubMed] [Google Scholar]
  • 44.Brunel N, Wang X-J. What determines the frequency of fast network oscillations with irregular neural discharges: I. Synaptic dynamics and excitation–inhibition balance. J. Neurophysiol. 2003;90:415–430. doi: 10.1152/jn.01095.2002. [DOI] [PubMed] [Google Scholar]
  • 45.Ito BM, Sato S, Kufta CV, Tran D. Effect of isoflurane and enflurane on the electrocorticogram of epileptic patients. Neurology. 1988;38:924–924. doi: 10.1212/WNL.38.6.924. [DOI] [PubMed] [Google Scholar]
  • 46.Wass CT, et al. The effects of remifentanil on epileptiform discharges during intraoperative electrocorticography in patients undergoing epilepsy surgery. Epilepsia. 2001;42:1340–1344. doi: 10.1046/j.1528-1157.2001.05901.x. [DOI] [PubMed] [Google Scholar]
  • 47.Herrick IA, Craen RA, Blume WT, Novick T, Gelb AW. Sedative doses of remifentanil have minimal effect on ECoG spike activity during awake epilepsy surgery. J. Neurosurg. Anesth. 2002;14:55–58. doi: 10.1097/00008506-200201000-00011. [DOI] [PubMed] [Google Scholar]
  • 48.von Ellenrieder N, Dubeau F, Gotman J, Frauscher B. Physiological and pathological high-frequency oscillations have distinct sleep-homeostatic properties. Neuroimage Clin. 2017;14:566–573. doi: 10.1016/j.nicl.2017.02.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Herman ST, Walczak TS, Bazil CW. Distribution of partial seizures during the sleep–wake cycle. Neurology. 2001;56:1453–1459. doi: 10.1212/WNL.56.11.1453. [DOI] [PubMed] [Google Scholar]
  • 50.Loddenkemper T, et al. Circadian patterns of pediatric seizures. Neurology. 2011;76:145–153. doi: 10.1212/WNL.0b013e318206ca46. [DOI] [PubMed] [Google Scholar]
  • 51.Frauscher B, et al. Facilitation of epileptic activity during sleep is mediated by high amplitude slow waves. Brain. 2015;138:1629–1641. doi: 10.1093/brain/awv073. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Ellenrieder N, Frauscher B, Dubeau F, Gotman J. Interaction with slow waves during sleep improves discrimination of physiologic and pathologic high-frequency oscillations (80–500 Hz) Epilepsia. 2016;57:869–878. doi: 10.1111/epi.13380. [DOI] [PubMed] [Google Scholar]
  • 53.Song I, et al. Bimodal coupling of ripples and slower oscillations during sleep in patients with focal epilepsy. Epilepsia. 2017;58:1972–1984. doi: 10.1111/epi.13912. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Jacobs J, et al. High-frequency electroencephalographic oscillations correlate with outcome of epilepsy surgery. Ann. Neurol. 2010;67:209–220. doi: 10.1002/ana.21847. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Haegelen C, et al. High-frequency oscillations, extent of surgical resection, and surgical outcome in drug-resistant focal epilepsy. Epilepsia. 2013;54:848–857. doi: 10.1111/epi.12075. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Jacobs J, Kobayashi K, Gotman J. High-frequency changes during interictal spikes detected by time-frequency analysis. Clin. Neurophysiol. 2011;122:32–42. doi: 10.1016/j.clinph.2010.05.033. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Guth TA, et al. Interictal spikes with and without high-frequency oscillation have different single-neuron correlates. Brain. 2021 doi: 10.1093/brain/awab288. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Engel J, Driver MV, Falconer MA. Electrophysiological correlates of pathology and surgical results in temporal lobe epilepsy. Brain. 1975;98:129–156. doi: 10.1093/brain/98.1.129. [DOI] [PubMed] [Google Scholar]
  • 59.Ulbert I, et al. In vivo laminar electrophysiology co-registered with histology in the hippocampus of patients with temporal lobe epilepsy. Exp. Neurol. 2004;187:310–318. doi: 10.1016/j.expneurol.2003.12.003. [DOI] [PubMed] [Google Scholar]
  • 60.Shimamoto S, et al. Utilization of independent component analysis for accurate pathological ripple detection in intracranial EEG recordings recorded extra- and intra-operatively. Clin. Neurophysiol. 2018;129:296–307. doi: 10.1016/j.clinph.2017.08.036. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Avants BB, Epstein CL, Grossman M, Gee JC. Symmetric diffeomorphic image registration with cross-correlation: Evaluating automated labeling of elderly and neurodegenerative brain. Med. Image Anal. 2008;12:26–41. doi: 10.1016/j.media.2007.06.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Desikan RS, et al. An automated labeling system for subdividing the human cerebral cortex on MRI scans into gyral based regions of interest. Neuroimage. 2006;31:968–980. doi: 10.1016/j.neuroimage.2006.01.021. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES