Abstract
Background: High-power radiofrequency (RF) catheter ablation was considered as a promising alternative strategy to conventional-power ablation in the treatment of patients with atrial fibrillation (AF). This study sought to compare the efficacy and safety of high-power energy delivery to that of conventional-power setting in AF catheter ablation.
Methods: We performed a systematic review of relevant literature in Pubmed, Embase, Cochrane library, and Google Scholar database. Sixteen eligible studies totaling 3,307 patients (1,929 for high-power ablation; 1,378 for conventional-power ablation) met inclusion criteria.
Results: During a median 12 month follow-up, high-power ablation showed a significantly higher AF/atrial tachycardia-free survival rate in comparison with conventional-power ablation (risk ratio [RR] 1.09, 95% CI 1.02 to 1.15, p = 0.008). Notably, a high-power strategy convincingly decreased the procedure time (weighted mean difference [WMD] −46.11 min, 95% CI −59.15 to −33.07, p < 0.001) and RF ablation time (WMD −19.19 min, 95% CI −24.47 to −13.90, p < 0.001), along with reduced fluoroscopy time (WMD −7.82 min, 95% CI −15.13 to −0.68, p = 0.036). In addition, there was no perceptible difference in the potential risk of procedure-related complications between these two approaches (RR 0.81, 95% CI 0.48 to 1.37, p = 0.428).
Conclusions: High-power RF catheter ablation was associated with an improvement in long-term sinus rhythm maintenance for treatment of AF, without exacerbating the risk of adverse events during the procedure. Impressively, high-power pulmonary vein isolation had the potential to shorten the application duration and minimize fluoroscopic exposure.
Keywords: high-power ablation, ablation lesion, atrial fibrillation, catheter ablation, meta-analysis
Introduction
Radiofrequency (RF) catheter ablation is an effective option for patients with symptomatic, drug-refractory atrial fibrillation (AF) (1). Nevertheless, durable pulmonary vein isolation (PVI) is still a clinical challenge in achieving long-term atrial arrhythmia-free survival. Emerging evidence indicates that contact force, catheter stability, and particularly RF energy settings are the main determinants of irreversible, continuous lesion creation (2, 3). Currently, PVI is commonly applied with a low power output (20–35 W) for a relatively long duration (4, 5). However, the rate of pulmonary vein reconduction remains substantial and subsequently increases the risk for recurrence.
In view of this, a high-power ablation strategy is increasingly employed to treatment of AF in order to generate more effective lesions while shortening the duration of energy delivery. Mechanistically, high-power ablation is linked with a wider, continuous tissue injury and composed of lower depth (6, 7). Remarkably, recent studies show an impressive efficacy of high-power ablation in maintenance of sinus rhythm without an increased risk for extracardiac damage in clinical setting (8–10). Compared to variation of radiofrequency output (ranged from 40 to 90 W), relatively few patients and potential selection bias introduced by observational data may collectively limit the strength and reliance of evidence.
Therefore, we conduct a comprehensive meta-analysis of available clinical studies to compare the efficacy and safety of high-power ablation with conventional-power ablation in patients undergoing treatment for AF.
Methods
The study was registered with PROSPERO (CRD42020160991) and reported in accordance with the PRISMA statement.
Literature Search
We systematically searched the electronic databases of Pubmed, Embase, Cochrane library, and Google Scholar from inception through June 2021 to identify the trials comparing radiofrequency catheter ablation with high power vs. conventional power for patients with atrial fibrillation. No language restrictions were set. The following keywords were used for search: “atrial fibrillation” or “AF,” AND “high power” or “higher power” or “high output' or “higher output.” Additional articles were searched through review of reference lists of related original and review articles which cited eligible manuscripts.
Eligibility Criteria and Data Extraction
Studies that fulfilled the following criteria were eligible for further analysis: (i) observational studies or randomized control trials (RCTs); (ii) high-power RF ablation vs. conventional-power RF ablation; and (iii) provided the AF/atrial tachycardia relapse during a minimum of 6 month follow-up. A high-power RF ablation strategy was set as energy ≥40 W during the procedure. AF/atrial tachycardia relapse was defined by any documented episode of AF, atrial flutter, or atrial tachycardia lasting >30 s after the blanking period. Studies that enrolled patients with an initially failed atrial ablation or previous cardiac surgery were excluded. Subsequently, two investigators (Zhi-Jie Mao and Yi-He Chen) screened the included studies and extracted the following data: author name, year of publication, study design, AF type, follow-up duration, clinical monitoring during follow-up, ablation protocol, and technical details. Discrepancies that were unable to be resolved were discussed by an additional investigator (Hui Lin). Baseline characteristics of the study population (i.e., mean age, gender, sample size, left atrial diameter, CHA2DS2-VASc score, left ventricular ejection fraction [LVEF], body mass index [BMI], and comorbidities) were also extracted. The primary endpoint of efficacy was AF/atrial tachycardia-free survival, while the primary endpoints of safety were procedure time, RF ablation time, fluoroscopy time, or procedure-related complications (e.g., esophageal fistula, tamponade or pericarditis, stroke or transient ischemic attack [TIA], vascular-related adverse events, phrenic nerve palsy, pulmonary vein [PV] stenosis, pulmonary edema, cardiovascular ischemic attack, and even death).
Statistical Analysis
Statistical analysis was conducted using STATA version 12.0 (StataCorp, College Station, TX, USA) and SPSS version 20.0 (SPSS Inc., IL). Categorical variables (AF/atrial tachycardia-free survival and procedure-related complications) and continuous variables (procedure time, ablation time, and fluoroscopy time) were presented as risk ratios (RRs) or weighted mean difference (WMDs) with corresponding 95% confidence intervals (CIs), respectively. Random-effect models were used according to the DerSimonian & Laird method. Heterogeneity among studies was estimated by Q test and quantified with I2 statistic. Publication bias for the pooled estimates was visually assessed by funnel plot and supplemented with Egger's test. Sensitivity analysis was performed by excluding each study for efficacy outcomes of AF/atrial tachycardia-free survival. To explore the impact of potential effect modifiers on outcomes and the possible sources of heterogeneity, subgroup analysis was conducted according to the following subsets: use of contact force (CF)-sensing catheter or ablation index (AI) guidance. A two-tailed p < 0.05 was considered as the threshold for statistical significance.
Results
The initial database search yielded a total of 1,195 articles. Of these, 36 potentially relevant articles were retained for full-text review after screening titles and abstracts. Sixteen studies comprising 3,307 patients (1,929 assigned to high-power group and 1,378 assigned to conventional-power group) met eligibility criteria for further meta-analysis (Figure 1). Details about the included studies and patient demographics are summarized in Tables 1–3. Of these studies, only two was RCT and the remaining followed either prospective (n = 8) and retrospective (n = 6) design. The majority of studies were executed in North America and Europe from 2006 to 2021. The mean age was 61.9 years. The proportion of male patients ranged from 60.4 to 84.3%. The median follow-up duration was 12 months. In the high-power group, RF application was mainly set as ≥40 W for the encirclement. Over half of included studies utilized CF-sensing catheter during the ablation. Meanwhile, additional AI guidance was conducted in the studies of Okamatsu et al. (22), Shin et al. (24), Kyriakopoulou et al. (19), Ejima et al. (20), and Hansom et al. (25). There was no significant difference in baseline patient characteristics between these two approaches (Table 3).
Figure 1.
Flowchart of the search strategy.
Table 1.
Main characteristics of included studies.
| Author | State | Design | Patients | AF type | Age | Male (%) | Ablation protocol | Catheter parameters |
Follow-up (months) |
||
|---|---|---|---|---|---|---|---|---|---|---|---|
| HP | CP | HP | CP | ||||||||
| Baher et al. (11) | USA | OS, R | 574 | 113 | ParAF, PerAF | 68.9 | 78.7 | 50 W, 5 s | 25–35 W, 10–30 s | 3.5-mm irrigated tip, CF | 30 |
| Pambrun et al. (12) | France | OS, P | 50 | 50 | ParAF | 63.8 | 65.0 | 40–50 W | 25–30 W | 3.5-mm irrigated tip, CF | 12 |
| Nilsson et al. (13) | Denmark | OS, R | 45 | 45 | ParAF, PerAF | 53.0 | 73.3 | 45 W, 20 s | 30 W, 120 s | 5-mm irrigated tip | 15 |
| Kanj et al. (14) | USA | RCT | 61 | 60 | ParAF, PerAF | 60.5 | 82.6 | 50 W | 35 W | 3.5-mm irrigated tip | 6 |
| Matiello et al. (15) | Spain | OS, P | 89 | 42 | ParAF, PerAF | 53.3 | 73.3 | 40 W | 30 W | 5-mm irrigated tip | 12 |
| Yamada et al. (16) | Japan | OS, R | 61 | 47 | ParAF | 57.0 | 84.3 | 40 W, 60 s | 30 W, 60 s | 4-mm or 8-mm tip | 6 |
| Vassallo et al. (9) | Brazil | OS, R | 41 | 35 | ParAF, PerAF | 62.6 | 73.6 | 45–50 W, 6 s | 30 W, 30 s | 3.5-mm irrigated tip, CF | 12 |
| Bunch et al. (17) | USA | OS, P | 402 | 402 | ParAF, PerAF | 66.8 | 64.1 | 50 W, 2–15 s | 30 W, 5–20 s | 3.5-mm irrigated tip, CF | 12 |
| Kottmaier et al. (18) | Germany | OS, P | 97 | 100 | ParAF | 60.8 | 59.4 | 70 W, 5–7 s | 30–40 W, 20–40 s | 4-mm irrigated tip | 12 |
| Kyriakopoulou et al. (19) | Belgium | OS, R | 80 | 105 | ParAF | 65.3 | 60.4 | 40 W | 35 W | 3.5-mm irrigated tip, CF, AI | 12 |
| Ejima et al. (20) | Japan | OS, P | 60 | 60 | ParAF | 64.9 | 71.7 | 50 W | 25–40 W | 3.5-mm irrigated tip, CF, AI | 12 |
| Yazaki et al. (21) | Japan | OS, R | 32 | 32 | ParAF, PerAF | 63.5 | 73.4 | 50 W, 8–12 s | 25–40 W, 15–40 s | 3.5-mm irrigated tip, CF | 10 |
| Okamatsu et al. (22) | Japan | OS, P | 20 | 20 | ParAF, PerAF | 66.5 | 70.0 | 40–50 W | 20–30 W | 3.5-mm irrigated tip, CF, AI | 6 |
| Yavin et al. (23) | USA | OS, P | 112 | 112 | ParAF, PerAF | 63.6 | 67.0 | 45–50 W, 8–15 s | 20–35 W, 20–30 s | 3.5-mm irrigated tip, CF | 19 |
| Shin et al. (24) | Korea | RCT | 100 | 50 | ParAF, PerAF | 58.2 | 76.0 | 40–50 W, 10–20 s | 30 W, 40 s | 3.5-mm irrigated tip, CF, AI | 12 |
| Hansom et al. (25) | Canada | OS, P | 107 | 107 | ParAF, PerAF | 62.0 | 70.1 | 50 W, 6–10 s | 20–35 W, 20–40 s | 3.5-mm irrigated tip, CF, AI | 12 |
HP, high-power ablation; CP, conventional-power ablation; AF, atrial fibrillation; OS, observational study; RCT, randomized control trail; P, prospective; R, retrospective; ParAF, paroxysmal atrial fibrillation; PerAF, persistent atrial fibrillation; W, watt; CF, contact force-sensing catheter; AI, ablation index.
Table 3.
Baseline population demographics of included patients.
| No. of studies | HP | CP | p-value | |
|---|---|---|---|---|
| Age | 16 | 61.9 ± 4.4 | 62.0 ± 5.0 | 0.917 |
| Male (%) | 16 | 71.7 ± 9.8 | 69.3 ± 8.2 | 0.471 |
| Hypertension | 12 | 53.1 ± 19.7 | 50.4 ± 18.8 | 0.733 |
| Diabetes mellitus | 9 | 20.1 ± 12.1 | 15.2 ± 8.6 | 0.333 |
| Stroke/TIA | 8 | 9.0 ± 3.7 | 8.9 ± 2.5 | 0.932 |
| CHA2DS2-VASC | 9 | 2.0 ± 0.4 | 2.1 ± 0.3 | 0.818 |
| LA diameter | 9 | 41.3 ± 2.7 | 41.0 ± 3.3 | 0.837 |
| AF duration (year) | 5 | 3.9 ± 2.0 | 4.3 ± 2.2 | 0.793 |
| Heart failure | 5 | 15.1 ± 19.1 | 14.8 ± 18.3 | 0.983 |
| LVEF (%) | 10 | 58.6 ± 4.2 | 58.6 ± 4.3 | 0.996 |
| CAD | 4 | 19.3 ± 6.7 | 16.9 ± 6.6 | 0.637 |
| BMI | 8 | 26.7 ± 2.4 | 26.8 ± 2.5 | 0.944 |
HP, high-power ablation; CP, conventional-power ablation; TIA, transient ischemic attack; LA, left atrium; AF, atrial fibrillation; LVEF, left ventricular ejection fraction; CAD, coronary artery disease; BMI, body mass index.
Table 2.
Detailed clinical follow-up information of included studies.
| Author | Clinical monitoring during follow-up | Blanking period | Definition of recurrence |
|---|---|---|---|
| Baher et al. (11) | 30 or 60 d event monitors after ablation and at the 3 month follow-up. Patients were followed at 3-, 6-, 12 month intervals. Additional home monitoring was ordered in case of AF symptoms | 3 months | ≥30-s AF or flutter |
| Pambrun et al. (12) | 10 d Holter monitoring at 1, 3, 12 months, 24 h Holter monitoring at 6 and 9 months. Additional 24 h Holter was obtained in case of AF symptoms | NA | >30-s AF or atrial tachycardia |
| Nilsson et al. (13) | NA | 1 month | Atrial arrhythmias |
| Kanj et al. (14) | Arrhythmia transmitter was used to monitor events during the first 6 months | 2 months | Atrial arrhythmias |
| Matiello et al. (15) | 24 h Holter monitoring at 1, 4, and 7 months, and every 6 months thereafter. ECG was performed in case of arrhythmia symptoms | 3 months | AF or left atrial flutter |
| Yamada et al. (16) | 24 h Holter monitoring at 2 weeks, 1 month, and every month thereafter | NA | Atrial arrhythmias |
| Vassallo et al. (9) | ECG at 3, 6, 9, and 12 months and 24 h Holter monitoring at 6 and 12 months | 3 months | Atrial arrhythmias |
| Bunch et al. (17) | Ambulatory monitoring at 3 month intervals post ablation up to the first year | 3 months | AF or atrial flutter |
| Kottmaier et al. (18) | 7 d Holter monitoring at 3, 6, and 12 months | 6 weeks | AF or atrial tachycardia |
| Kyriakopoulou et al. (19) | Holter monitoring at 12 months | 3 months | >30-s atrial tachyarrhythmias |
| Ejima et al. (20) | ECG and 24 h Holter monitoring at 3, 6, 9, and 12 months and every 6 months thereafter | 2 months | >30-s atrial tachyarrhythmias |
| Yazaki et al. (21) | ECG and 24 h Holter monitoring at 3, 6, 9, and 12 months and every 6 months thereafter | 2 months | >30-s atrial tachyarrhythmias |
| Okamatsu et al. (22) | ECG and 24 h Holter monitoring at 3, 6, 9, and 12 months | 3 months | >30-s atrial tachyarrhythmias |
| Yavin et al. (23) | 14 d continuous or patient-triggered Holter monitoring | 4 weeks | >30-s AF or atrial tachycardia |
| Shin et al. (24) | ECG and 24 h Holter monitoring at 3, 6, and 12 months. Additional ECG was performed in case of symptoms | 3 months | >30-s AF, atrial tachycardia or atrial flutter |
| Hansom et al. (25) | ECG and 14 day Holter monitoring at 3, 6, and 12 months | 3 months | >30-s atrial arrhythmias |
AF, atrial fibrillation; ECG, electrocardiogram; d, day; h, hour; NA, none available.
Primary Efficacy Endpoints
All 16 studies contributed to the primary outcome of AF/atrial tachycardia-free survival. During a median follow-up of 12 months, high-power ablation was associated with a 9% relative increase in AF/atrial tachycardia-free survival as compared with conventional-power ablation (RR 1.09, 95% CI 1.02 to 1.15, p = 0.008) (Figure 2), with moderate heterogeneity across the studies (I2 = 49.8%, p = 0.012). Notwithstanding the publication bias detected by visual inspection of funnel plot and Egger's test (p = 0.001), the trim-and-fill test showed a similar result (adjust RR: 1.10, 95% CI 1.03 to 1.17). One study removed sensitivity analysis which revealed that individual data did not alter the results of this meta-analysis. Subgroup analyses according to use of CF-sensing catheter or AI-guided RF ablation are presented in Table 4. Of note, the pronounced benefit of high-power ablation over conventional-power ablation in maintenance of sinus rhythm was compromised by utilization of the CF-sensing catheter (RR 1.04, 95% CI 0.98–1.10, p = 0.164) or aided with AI guidance (RR 1.05, 95% CI 0.98–1.11, p = 0.142).
Figure 2.
Forest plot of AF/atrial tachycardia-free survival for high-power ablation vs. conventional-power ablation.
Table 4.
Subgroup analysis of AF/atrial tachycardia-free survival.
| Subgroup | No. of studies | HP | CP | RR (95% CI) | p-value | ||
|---|---|---|---|---|---|---|---|
| Events | Patients | Events | Patients | ||||
| CF-sensing catheter | |||||||
| With | 11 | 1,006 | 1,578 | 806 | 1,084 | 1.04 (0.98, 1.10) | 0.164 |
| Without | 5 | 256 | 353 | 179 | 294 | 1.23 (1.10, 1.36) | <0.001 |
| AI guided RF ablation | |||||||
| With | 6 | 334 | 397 | 293 | 372 | 1.05 (0.98, 1.11) | 0.142 |
| Without | 10 | 928 | 1,532 | 692 | 1,006 | 1.13 (1.02, 1.25) | 0.019 |
| Total | 16 | 1,262 | 1,929 | 958 | 1,378 | 1.09 (1.02, 1.15) | 0.008 |
HP, high-power ablation; CP, conventional-power ablation; CF, contact force; AI, ablation index; RF, radiofrequency; RR, risk ratios.
Primary Safety Endpoints
As shown in Figures 3, 4, high-power ablation was associated with a remarkable reduction in procedure time (WMD −46.11 min, 95% CI −59.15 to −33.07, p < 0.001) and RF ablation time (WMD −19.19 min, 95% CI −24.47 to −13.90, p < 0.001) compared to conventional-power ablation, with high heterogeneity for both (I2 = 83.2%, p = 0.003 and I2 = 96.5%, p < 0.001, respectively). Furthermore, high-power RF energy delivery also decreased fluoroscopy time in the high-power group (WMD −7.82 min, 95% CI −15.13 to −0.52, p = 0.036) in spite of the high heterogeneity (I2 = 99.8%, p < 0.001) (Figure 5). Systemic exclusion of each study did not change the pooled estimates and p-value. Accordingly, stratified by use of the CF-sensing catheter or AI-guided RF ablation also has no significant impact on these results (Table 5).
Figure 3.
Forest plot of procedure time for high-power ablation vs. conventional-power ablation.
Figure 4.
Forest plot of RF ablation time for high-power ablation vs. conventional-power ablation.
Figure 5.
Forest plot of fluoroscopy time for high-power ablation vs. conventional-power ablation.
Table 5.
Subgroup analysis of procedure time, RF ablation time, and fluoroscopy time.
| Subgroup | Procedure time | RF ablation time | Fluoroscopy time | ||||||
|---|---|---|---|---|---|---|---|---|---|
| No. of studies | WMD (95% CI) | p-value | No. of studies | WMD (95% CI) | p-value | No. of studies | WMD (95% CI) | p-value | |
| CF-sensing catheter | |||||||||
| With | 9 | −48.29 (−63.72, −32.87) | <0.001 | 8 | −18.92 (−24.90, −12.93) | <0.001 | 8 | −3.08 (−5.31, −0.84) | 0.007 |
| Without | 3 | −38.70 (−63.13, −14.27) | 0.002 | 2 | −20.84 (−26.74, −14.94) | <0.001 | 4 | −18.51 (−36.21, −0.81) | 0.040 |
| AI guided RF ablation | |||||||||
| With | 5 | −38.40 (−57.48, −19.33) | <0.001 | 5 | −20.82 (−30.16, −11.47) | <0.001 | 5 | −3.83 (−7.07, −0.60) | 0.020 |
| Without | 7 | −51.72 (−70.15, −33.28) | <0.001 | 5 | −18.15 (−21.62, −14.69) | <0.001 | 7 | −11.00 (−22.08, 0.07) | 0.051 |
| Total | 12 | −46.11 (−59.15, −33.07) | <0.001 | 10 | −19.19 (−24.47, −19.30) | <0.001 | 12 | −7.82 (−15.13, −0.52) | 0.036 |
CF, contact force; AI, ablation index; RF, radiofrequency; WMD, weighted mean difference; CI, confidence interval.
The overall incidence of procedure-related complications was 1.59% (28 of 1,758 patients) in the high-power group and 2.22% (28 of 1,263 patients) in the conventional-power group. No significant difference was observed for patients with high-power ablation vs. conventional-power ablation (RR 0.81, 95% CI 0.48 to 1.37, p = 0.428) (Figure 6), with low heterogeneity (I2 = 0.0%, p = 0.710). Consistently, the results did not change after sensitivity and subgroup analysis (Table 6). Detailed information of complications during the procedure is presented in Table 7. No death, esophageal fistula, and PV stenosis occurred in either group. Notably, the predominant adverse events were vascular complications (e.g., groin hematoma, pseudoaneurysm, and arteriovenous fistula) with 0.74% patients in the high-power group and 1.74% patients in the conventional-power group. In addition, there was a similar prevalence of pericardial complications (0.29% vs. 0.24%), stroke/TIA (0.23% vs. 0.16%), and cardiovascular ischemic attack (0.11% vs. 0.08%) from both groups.
Figure 6.
Forest plot of procedure-related complications for high-power ablation vs. conventional-power ablation.
Table 6.
Subgroup analysis of procedure-related complications.
| Subgroup | No. of studies | HP | CP | RR (95% CI) | p-value | ||
|---|---|---|---|---|---|---|---|
| Events | Patients | Events | Patients | ||||
| CF-sensing catheter | |||||||
| With | 8 | 5 | 1,405 | 9 | 969 | 0.63 (0.23, 1.75) | 0.374 |
| Without | 5 | 23 | 353 | 19 | 294 | 0.96 (0.45, 2.02) | 0.907 |
| AI guided RF ablation | |||||||
| With | 4 | 3 | 267 | 5 | 292 | 0.68 (0.17, 2.67) | 0.586 |
| Without | 9 | 25 | 1,491 | 23 | 971 | 0.83 (0.47, 1.47) | 0.528 |
| Total | 13 | 28 | 1,758 | 28 | 1,263 | 0.81 (0.48, 1.37) | 0.428 |
HP, high-power ablation; CP, conventional-power ablation; CF, contact force; AI, ablation index; RF, radiofrequency; RR, risk ratios.
Table 7.
Details of procedure-related complications.
| Complications | HP | CP | ||||
|---|---|---|---|---|---|---|
| Events | Incidence (%) | Ratio (%) | Events | Incidence (%) | Ratio (%) | |
| Tamponade | 2 | 0.11 | 7.1 | 2 | 0.16 | 7.1 |
| Pericarditis | 3 | 0.18 | 10.8 | 1 | 0.08 | 3.6 |
| Stroke/TIA | 4 | 0.23 | 14.3 | 2 | 0.16 | 7.1 |
| Esophageal fistula | 0 | 0.00 | 0.0 | 0 | 0.00 | 0.0 |
| Vascular | 13 | 0.74 | 46.5 | 22 | 1.74 | 78.6 |
| PV stenosis (>50%) | 0 | 0.00 | 0.0 | 0 | 0.00 | 0.0 |
| Pulmonary edema | 2 | 0.11 | 7.1 | 0 | 0.00 | 0.0 |
| Cardiovascular ischemic attack | 2 | 0.11 | 7.1 | 1 | 0.08 | 3.6 |
| Death | 0 | 0.00 | 0.0 | 0 | 0.00 | 0.0 |
| PNP | 2 | 0.11 | 7.1 | 0 | 0.00 | 0.0 |
| Total | 28 | 1.59 | 100 | 28 | 2.22 | 100 |
HP, high-power ablation; CP, conventional-power ablation; TIA, transient ischemic attack; PV, pulmonary vein; PNP, phrenic nerve palsy.
Discussion
In the present analysis, high-power ablation was associated with an improved AF/atrial tachycardia-free survival rate during a 12 month median follow-up in comparison with conventional-power ablation. Importantly, high-power settings significantly reduced the procedure time and RF ablation time and alleviated fluoroscopy exposure. Of note, the incidence of procedure-related complications was comparable between these two approaches.
Biophysical Mechanism of High-Power Ablation
RF catheter ablation transmits energy of alternating current with a frequency of 500–1,000 kHz, causing permanent tissue damage (26). Electrical currents through the tissue primarily generated resistive heating in the vicinity of the electrode tip, while deeper and extracardiac damage occurs as a consequence of passive heat conduction. Commonly, the ablation power setting was 20 to 30 W for a duration of 30 to 60 s in clinical practice (4, 5), whereas a longer duration of the current ablation strategy significantly extended the conductive heating phase and thus increased the risk of thermal injury to adjacent structures (i.e., lung hematoma/hemorrhage, phrenic nerve palsy, atrioesophageal fistula) (27, 28). Lesion formation was mainly based on the power and duration of RF application, which provided the rationale for high-power RF delivery with short duration by modifying the relationship between resistive and conductive heating (3). It was intended for rapidly generating lethal heating during the resistive phase and avoiding the distant tissue damage. Furthermore, higher-power and resultant shorter RF application could overcome the challenge of catheter dislodgement or inadequate catheter–tissue contact resulting in tissue edema, and improved the lesion efficacy. The mechanistic insight into the lesion biophysics metrics of higher-power ablation was initially evaluated in both in vivo and ex vivo models by Borne et al. (29). They found that 50 W/5 s was associated with a trend toward larger lesion diameter but a less lesion depth than 20 W/30 s. In addition, experimental data from Bourier et al. also reported a similar lesion geometry of high power with reduced lesion depth and extended lesion width in porcine thigh muscle preparations when compared with the standard approach (30). Subsequently, Leshem et al. further confirmed the efficacy and safety of high-power ablation in a beating heart swine model which resembled RF catheter ablation in clinical practice (31). Histopathologic examinations demonstrated an ~50% greater width (6.02 ± 0.2 mm vs. 4.43 ± 1.0 mm) and similar depth (3.58 ± 0.3 mm vs. 3.53 ± 0.6 mm) of atrial lesions vs. lower power. In terms of the thickness of the left atrium (0.5–3.5 mm), it was theoretically plausible that high-power ablation favored the creation of contiguous, transmural lesions which may facilitate the circumferential isolation of the pulmonary vein (32).
Intriguingly, a previous systematical review by Yuyun et al. reported a significant relationship between power output and primary effectiveness outcomes of AF catheter ablation and recommended the higher-power (>45-W), shorter-duration (15–20-s) strategy for RF ablation with optimized efficacy and safety profiles (33). Early in 2006, Yamada et al. retrospectively compared high-power output with conventional parameters in patients with paroxysmal AF (16). High-power ablation was associated with a lower AF recurrence (32% vs. 47%) during a short-term follow-up. In another randomized pilot trial by Kanj et al., higher-energy delivery presented a markedly greater efficacy for maintenance of sinus rhythm at the 6 month follow-up compared with the lower-power setting (14). Subsequently, Matiello et al. ensured the favorable clinical endpoints of long-term AF/atrial tachycardia-free survival in patients who underwent RF catheter ablation with high output (15). In view of this, the high-power RF ablation strategy was increasingly utilized for treatment of AF in recent studies. Chen et al. displayed a higher first-round PVI, lower acute reconnection of 50-W ablation, which translated into almost excellent clinical endpoints of sinus rhythm maintenance (96% patients) at 6 months (34). Accordingly, Kottmaier and colleagues, likewise, demonstrated that high-power ablation was associated with a significantly fewer atrial arrhythmia recurrence after 1 year (18). The present study reconfirmed the superiority of high-power ablation with a 10% increase in freedom from the AF rate during a median follow-up of 12 months. Therefore, the aforementioned results provided robust evidence for the feasibility and efficacy of high-power ablation in the treatment of AF.
Recently, Winkle et al. retrospectively examined the adverse events in 10,284 patients receiving AF ablation with a high-power setting (45–50 W/2–15 s) (35). This study exhibited an extremely low incidence of procedure-related complications. In the high-power group, only one atrio-esophageal fistula was found in the 11,436 ablations compared with 3 of the 2,538 ablations of the conventional-power group. In line with this, Baher et al. also detected identical esophageal thermal injury patterns between these two approaches assessed by late gadolinium enhancement MRI (11). Interestingly, data from the present work unmasked a relatively higher proportion of pericardial complication and stroke for high-power ablation despite comparable prevalence vs. conventional-power ablation. To our knowledge, steam pop and catheter char due to tissue overheating were regarded as the main culprit of tamponade and thrombus, which may be partially resolved by utilization of novel multielectrode catheters with a more sophisticated temperature feedback control system and advanced cooling techniques. Moreover, an optimal CF spectrum and/or AI-guided ablation may further minimize the collateral thermal injury during the high-power RF delivery.
Notably, high-power ablation constantly reduced the procedure time and RF ablation time and shortened the radiation exposure across the studies irrespective of supplementary CF or AI application. Elongation access to the systemic circulation may trigger thrombus formation and even uninterrupted anticoagulation. A prior study reported an ~13 to 20% frequency of subtle neurocognitive impairment after ablation of AF and disclosed a causal relationship between cognitive decline and left atrial access time (36). On this account, a high power accompanied with shorter duration may imperceptibly provide a guarantee of neuroprotection. Furthermore, this shortening also limited the excessive fluid load from catheter irrigation, which in turn minimized the risk of acute heart failure in patients with impaired cardiac function. Particularly, it was necessary to highlight that shorter ablation time could remarkably weaken the impact of heart beating and even deep breathing on catheter stability while ensuring irreversible atrial lesions and favorable long-term outcomes.
Limitations
First, lack of individual patient-level data impeded the exploration of correlation between other effect modifiers (e.g., catheter, type of AF, left atrial dimension) and clinical outcomes. Second, the present results were largely driven by observational studies, which seemed more susceptible to potential biases. Third, there was a variation in the radiofrequency energy dosing of the high-power ablation strategy in different clinical centers (varied from 40 to 70 W), while in the conventional-power group, RF application was even set at 40 W from Kottmaier et al. and Ejima et al., which led to partial overlaps between these two approaches and may possibly dwarf the beneficial effect of the high-power setting (18, 20). Therefore, it was of great importance to standardize ablation parameters of high output strategy. Taking into account consistent safety outcomes in the included studies and the results of animal studies in vivo, >50 W may be an appropriate definition of “Real HPSD.” The benefits of strictly realistic HPSD may have been underestimated in this article. Finally, despite substantial heterogeneity among the studies, sensitivity and subgroup analyses demonstrated no difference in pooled results and provided robust evidence of its superiority in improving efficacy and safety outcomes.
Conclusions
High-power ablation presents an incremental long-term efficacy in maintaining the sinus rhythm vs. conventional-power ablation in patients with AF. In addition, a high-power strategy significantly reduces the procedure burden and fluoroscopic exposure without increasing the risk of procedure-related complications.
Data Availability Statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author/s.
Author Contributions
F-YX provided the idea. Y-HC designed and subsequently guided this article and is responsible for the overall content as guarantor. HL and Z-QH assisted with the guiding and revising of the article. Other authors participated in writing and data processing statistics. All authors contributed to the article and approved the submitted version.
Funding
This work was supported by the Natural Science Foundation of China (NSFC) (Grant No. 81900229) and Wenzhou Municipal Science and Technology Bureau (Grant No. Y20180079) to Y-HC. and the Natural Science Foundation of China (NSFC) (Grant No. 81900293) to Qian Wang.
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher's Note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Acknowledgments
We thank Professor Qian Wang for assisting and guiding in the revision of the article.
References
- 1.Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B, et al. 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Eur Heart J. (2016) 37:2893–62. 10.15829/1560-4071-2017-7-7-86 [DOI] [PubMed] [Google Scholar]
- 2.Haines DE. The biophysics of radiofrequency catheter ablation in the heart: the importance of temperature monitoring. Pacing Clin Electrophysiol. (1993) 16:586–91. 10.1111/j.1540-8159.1993.tb01630.x [DOI] [PubMed] [Google Scholar]
- 3.Eick OJ. Factors influencing lesion formation during radiofrequency catheter ablation. Indian Pacing Electrophysiol J. (2003) 3:117–28. [PMC free article] [PubMed] [Google Scholar]
- 4.Bahnson TD. Strategies to minimize the risk of esophageal injury during catheter ablation for atrial fibrillation. Pacing Clin Electrophysiol. (2009) 32:248–60. 10.1111/j.1540-8159.2008.02210.x [DOI] [PubMed] [Google Scholar]
- 5.Reddy VY, Dukkipati SR, Neuzil P, Natale A, Albenque JP, Kautzner J, et al. Controlled trial of the safety and effectiveness of a contact force-sensing irrigated catheter for ablation of paroxysmal atrial fibrillation: results of the TactiCath Contact Force Ablation Catheter Study for Atrial Fibrillation (TOCCASTAR) study. Circulation. (2015) 132:907–15. 10.1161/CIRCULATIONAHA.114.014092 [DOI] [PubMed] [Google Scholar]
- 6.Bhaskaran A, Chik W, Pouliopoulos J, Nalliah C, Qian P, Barry T, et al. Five seconds of 50-60 W radio frequency atrial ablations were transmural and safe: an in vitro mechanistic assessment and force-controlled in vivo validation. Europace. (2017) 19:874–80. 10.1093/europace/euw077 [DOI] [PubMed] [Google Scholar]
- 7.Barkagan M, Contreras-Valdes FM, Leshem E, Buxton AE, Nakagawa H, Anter E. High-power and short-duration ablation for pulmonary vein isolation: Safety, efficacy, long-term durability. J Cardiovasc Electrophysiol. (2018) 29:1287–96. 10.1111/jce.13651 [DOI] [PubMed] [Google Scholar]
- 8.Lee A, Gerstenfeld EP. High-power radiofrequency ablation for atrial fibrillation: establishing a standardized protocol. J Cardiovasc Electrophysiol. (2019) 30:2732–3. 10.1111/jce.14217 [DOI] [PubMed] [Google Scholar]
- 9.Vassallo F, Cunha C, Serpa E, Meigre LL, Carloni H, Simoes A, et al. Comparison of high-power short-duration (HPSD) ablation of atrial fibrillation using a contact force-sensing catheter and conventional technique: initial results. J Cardiovasc Electrophysiol. (2019) 30:1877–83. 10.1111/jce.14110 [DOI] [PubMed] [Google Scholar]
- 10.Reddy VY, Grimaldi M, De Potter T, Vijgen JM, Bulava A, Duytschaever MF, et al. Pulmonary vein isolation with very high power, short duration, temperature-controlled lesions: the QDOT-FAST Trial. JACC Clin Electrophysiol. (2019) 5:778–86. 10.1016/j.jacep.2019.04.009 [DOI] [PubMed] [Google Scholar]
- 11.Baher A, Kheirkhahan M, Rechenmacher SJ, Marashly Q, Kholmovski EG, Siebermair J, et al. High-power radiofrequency catheter ablation of atrial fibrillation: using late gadolinium enhancement magnetic resonance imaging as a novel index of esophageal injury. JACC Clin Electrophysiol. (2018) 4:1583–94. 10.1016/j.jacep.2018.07.017 [DOI] [PubMed] [Google Scholar]
- 12.Pambrun T, Durand C, Constantin M, Masse A, Marra C, Meillet V, et al. High-power (40-50 W) radiofrequency ablation guided by unipolar signal modification for pulmonary vein isolation: experimental findings and clinical results. Circulation Arrhyth Electrophysiol. (2019) 12:e007304. 10.1161/CIRCEP.119.007304 [DOI] [PubMed] [Google Scholar]
- 13.Nilsson B, Chen X, Pehrson S, Svendsen JH. The effectiveness of a high output/short duration radiofrequency current application technique in segmental pulmonary vein isolation for atrial fibrillation. Europace. (2006) 8:962–5. 10.1093/europace/eul100 [DOI] [PubMed] [Google Scholar]
- 14.Kanj MH, Wazni O, Fahmy T, Thal S, Patel D, Elayi C, et al. Pulmonary vein antral isolation using an open irrigation ablation catheter for the treatment of atrial fibrillation: a randomized pilot study. J Am Coll Cardiol. (2007) 49:1634–41. 10.1016/j.jacc.2006.12.041 [DOI] [PubMed] [Google Scholar]
- 15.Matiello M, Mont L, Tamborero D, Berruezo A, Benito B, Gonzalez E, et al. Cooled-tip vs. 8 mm-tip catheter for circumferential pulmonary vein ablation: comparison of efficacy, safety, lesion extension. Europace. (2008) 10:955–60. 10.1093/europace/eun144 [DOI] [PubMed] [Google Scholar]
- 16.Yamada T, Murakami Y, Okada T, Okamoto M, Shimizu T, Toyama J, et al. Can segmental pulmonary vein ablation reduce the recurrence of atrial fibrillation when using a higher RF power, larger tip electrode catheter, and additional RF deliveries?: the limitations of point-by-point RF ablation. Int Heart J. (2006) 47:219–28. 10.1536/ihj.47.219 [DOI] [PubMed] [Google Scholar]
- 17.Bunch TJ, May HT, Bair TL, Crandall BG, Cutler MJ, Mallender C, et al. Long-term outcomes after low power, slower movement versus high power, faster movement irrigated-tip catheter ablation for atrial fibrillation. Heart Rhythm. (2020) 17:184–89. 10.1016/j.hrthm.2019.08.001 [DOI] [PubMed] [Google Scholar]
- 18.Kottmaier M, Popa M, Bourier F, Reents T, Cifuentes J, Semmler V, et al. Safety and outcome of very high-power short-duration ablation using 70 W for pulmonary vein isolation in patients with paroxysmal atrial fibrillation. Europace. (2020) 22:388–93. 10.1093/europace/euz342 [DOI] [PubMed] [Google Scholar]
- 19.Kyriakopoulou M, Wielandts JY, Strisciuglio T, El Haddad M, Pooter J, Almorad A, et al. Evaluation of higher power delivery during RF pulmonary vein isolation using optimized and contiguous lesions. J Cardiovasc Electrophysiol. (2020) 31:1091–8. 10.1111/jce.14438 [DOI] [PubMed] [Google Scholar]
- 20.Ejima K, Higuchi S, Yazaki K, Kataoka S, Yagishita D, Kanai M, et al. Comparison of high-power and conventional-power radiofrequency energy deliveries in pulmonary vein isolation using unipolar signal modification as a local endpoint. J Cardiovasc Electrophysiol. (2020) 31:1702–8. 10.1111/jce.14532 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Yazaki K, Ejima K, Kanai M, Kataoka S, Higuchi S, Yagishita D, et al. Impedance drop predicts acute electrical reconnection of the pulmonary vein-left atrium after pulmonary vein isolation using short-duration high-power exposure. J Interv Cardiac Electrophysiol. (2020) 59:575–84. 10.1007/s10840-019-00691-z [DOI] [PubMed] [Google Scholar]
- 22.Okamatsu H, Koyama J, Sakai Y, Negishi K, Hayashi K, Tsurugi T, et al. High-power application is associated with shorter procedure time and higher rate of first-pass pulmonary vein isolation in ablation index-guided atrial fibrillation ablation. J Cardiovasc Electrophysiol. (2019) 30:2751–8. 10.1111/jce.14223 [DOI] [PubMed] [Google Scholar]
- 23.Yavin HD, Leshem E, Shapira-Daniels A, Sroubek J, Barkagan M, Haffajee CI, et al. Impact of high-power short-duration radiofrequency ablation on long-term lesion durability for atrial fibrillation ablation. JACC Clin Electrophysiol. (2020) 6:973–85. 10.1016/j.jacep.2020.04.023 [DOI] [PubMed] [Google Scholar]
- 24.Shin DG, Ahn J, Han SJ, Lim HE. Efficacy of high-power and short-duration ablation in patients with atrial fibrillation: a prospective randomized controlled trial. Europace. (2020) 22:1495–501. 10.1093/europace/euaa144 [DOI] [PubMed] [Google Scholar]
- 25.Hansom SP, Alqarawi W, Birnie DH, Golian M, Nery PB, Redpath CJ, et al. High-power, short-duration atrial fibrillation ablation compared with a conventional approach: outcomes and reconnection patterns. J Cardiovasc Electrophysiol. (2021) 32:1219–28. 10.1111/jce.14989 [DOI] [PubMed] [Google Scholar]
- 26.Nath S, DiMarco JP, Haines DE. Basic aspects of radiofrequency catheter ablation. J Cardiovasc Electrophysiol. (1994) 5:863–76. 10.1111/j.1540-8167.1994.tb01125.x [DOI] [PubMed] [Google Scholar]
- 27.Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, Aguinaga L, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Europace. (2018) 20:e1–160. 10.1093/europace/eux274 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Sauer WH, Tzou WS. With great power comes great responsibility: defining the safety of high-power short-duration atrial ablation. Circul Arrhyth Electrophysiol. (2019) 12:e007456. 10.1161/CIRCEP.119.007456 [DOI] [PubMed] [Google Scholar]
- 29.Borne RT, Sauer WH, Zipse MM, Zheng L, Tzou W, Nguyen DT. Longer duration versus increasing power during radiofrequency ablation yields different ablation lesion characteristics. JACC Clin Electrophysiol. (2018) 4:902–8. 10.1016/j.jacep.2018.03.020 [DOI] [PubMed] [Google Scholar]
- 30.Bourier F, Duchateau J, Vlachos K, Lam A, Martin CA, Takigawa M, et al. High-power short-duration versus standard radiofrequency ablation: insights on lesion metrics. J Cardiovasc Electrophysiol. (2018) 29:1570–5. 10.1111/jce.13724 [DOI] [PubMed] [Google Scholar]
- 31.Leshem E, Zilberman I, Tschabrunn CM, Barkagan M, Contreras-Valdes FM, Govari A, et al. High-power and short-duration ablation for pulmonary vein isolation: biophysical characterization. JACC Clin Electrophysiol. (2018) 4:467–79. 10.1016/j.jacep.2017.11.018 [DOI] [PubMed] [Google Scholar]
- 32.Ho SY, Sanchez-Quintana D, Cabrera JA, Anderson RH. Anatomy of the left atrium: implications for radiofrequency ablation of atrial fibrillation. J Cardiovasc Electrophysiol. (1999) 10:1525–33. 10.1111/j.1540-8167.1999.tb00211.x [DOI] [PubMed] [Google Scholar]
- 33.Yuyun MF, Stafford PJ, Sandilands AJ, Samani NJ, André Ng G. The impact of power output during percutaneous catheter radiofrequency ablation for atrial fibrillation on efficacy and safety outcomes: a systematic review. J Cardiovasc Electrophysiol. (2013) 24:1216–23. 10.1111/jce.12206 [DOI] [PubMed] [Google Scholar]
- 34.Chen S, Schmidt B, Bordignon S, Urbanek L, Tohoku S, Bologna F, et al. Ablation index-guided 50 W ablation for pulmonary vein isolation in patients with atrial fibrillation: procedural data, lesion analysis, and initial results from the FAFA AI High Power Study. J Cardiovasc Electrophysiol. (2019) 30:2724–31. 10.1111/jce.14219 [DOI] [PubMed] [Google Scholar]
- 35.Winkle RA, Mohanty S, Patrawala RA, Mead RH, Kong MH, Engel G, et al. Low complication rates using high power (45-50 W) for short duration for atrial fibrillation ablations. Heart Rhythm. (2019) 16:165–9. 10.1016/j.hrthm.2018.11.031 [DOI] [PubMed] [Google Scholar]
- 36.Medi C, Evered L, Silbert B, Teh A, Halloran K, Morton J, et al. Subtle post-procedural cognitive dysfunction after atrial fibrillation ablation. J Am Coll Cardiol. (2013) 62:531–9. 10.1016/j.jacc.2013.03.073 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author/s.






