Abstract
The ability of biomedical imaging data to be of quantitative nature is getting increasingly important with the ongoing developments in data science. In contrast to conventional attenuation-based X-ray imaging, grating-based phase contrast computed tomography (GBPC-CT) is a phase contrast micro-CT imaging technique that can provide high soft tissue contrast at high spatial resolution. While there is a variety of different phase contrast imaging techniques, GBPC-CT can be applied with laboratory X-ray sources and enables quantitative determination of electron density and effective atomic number. In this review article, we present quantitative GBPC-CT with the focus on biomedical applications.
Keywords: X-ray imaging, Phase contrast imaging, Grating interferometer, Grating interferometry, CT, Computed tomography, Phase contrast tomography, Quantitative imaging, Electron density, Effective atomic number
Introduction
Phase contrast imaging techniques enable visualization of subtle soft tissue differences in comparison to the conventional attenuation signal. Over the last 25 years, various phase contrast imaging (PCI) methods have been developed and established in modern X-ray imaging enabling high-resolution imaging especially of biomedical samples [1–10]. Different X-ray phase contrast methods include crystal interferometry [11, 12], propagation- based imaging (PBI) [13–17], analyzer-based imaging [18–22], grating interferometry (GI) [23–27], edge illumination (EI) [28–31], or tracking methods [32–35]. Each of those methods presents with individual requirements regarding beam coherence, detector resolution, or need of specific optical elements.
Although PCI originates from highly coherent synchrotron X-ray sources, some of the methods have been successfully translated to laboratory sources with limited coherence like PBI [36], GI [27], EI [29], or tracking methods like speckle-based imaging (SBI) [35]. The combination of phase contrast imaging with computed tomography methods extended the range of phase contrast imaging to three-dimensional sample characterization. This gave rise to numerous biomedical studies on ex vivo tissue samples like brain or lung with PBI [38, 39] next to liver, lung, and esophagus with EI [40], or brain as well as breast with GI [41, 67], to name only a few.
While many imaging applications including phase contrast methods have been focusing mainly on revealing subtle relative signal differences, quantitative imaging is gaining more and more interest. In general, relative contrast signals like the relaxation times in magnetic resonance imaging (MRI) can be sufficient for the desired information. However, absolute signals of physical nature like e.g. the electron density—this is what we describe in this work with quantitativeness—are in theory independent of the method of contrast formation. This does not only increase the comparability and reproducibility of the results but absolute quantitative signals can also be used for standardized diagnostics or for data science on a larger scale [42]. In clinical use for example, quantitative X-ray imaging is emerging shown by the development of dual-energy or spectral computed tomography [43–48]. There, decomposition of spectral CT data into the contributions from photoelectric effect and Compton scattering allows to calculate the electron density, the effective atomic number, virtual monoenergetic images, virtual non-contrast images, iodine maps, or other material decomposition of interest [47]. Clinical applications of spectral CT are for example the quantification of iodine contrast agent for increased tumor discrimination [49–51], enhanced characterization of kidney stones [52, 53], or improved stroke diagnosis [54].
Achieving quantitative phase contrast imaging is however challenging. Among the various PCI techniques available, grating interferometry is not only one of the most sensitive phase contrast methods [55, 56] but the tomographic phase contrast data can also be transferred directly to the electron density. In detail, when performing grating-based phase contrast computed tomography (GBPC-CT), one can convert the experimentally determined phase contrast signal to the electron density. In the case of polychromatic, laboratory sources’ additional effective energy calibration has to be applied [57–59]. This allows the determination of an absolute physical quantity, which enables comparing electron density results of different GI setups with different imaging methods like dual-energy CT or with reference and theoretical values [60]. Application of quantitative GBPC-CT covers myocardial infarct [61], atherosclerotic plaque [62], and tumors of kidney [63], liver [64], pancreas [65], brain [67–69], testis [70], and breast [41, 57, 72, 73]. Moreover, biochemical or biomedical fluids can be characterized quantitatively [66, 74]. In addition, Hounsfield units for the phase (HUp) [57, 75] or the effective atomic number Zeff can be calculated [76–78]. The complementarity of the attenuation and phase signal of GBPC-CT can be used to perform material decomposition [79] or a decomposition similar to spectral CT into electron density and effective atomic number [80–82].
In this review, we will focus on quantitative imaging with grating-based phase contrast tomography. In comparison to other phase contrast imaging techniques, GBPC-CT enables quantitative high spatial resolution imaging not only at synchrotron sources, but also relatively robust at polychromatic laboratory high-flux X-ray sources. After a presentation of the theoretical basics of X-ray phase contrast imaging, we will outline how we can determine electron density and effective atomic number with monochromatic and even polychromatic X-ray sources using GBPC-CT. This results in various biomedical applications of GPBC-CT and a brief presentation of related X-ray PCI methods in context of ability to provide quantitative data.
Phase contrast imaging
The quantity underlying changes in the phase Φ of electromagnetic waves is the complex index of refraction n. The different phase contrast methods can either directly access phase differences Φ of electromagnetic waves like in crystal interferometry, the gradient of the phase ∇Φ like in ABI, EI, GI, and tracking-based methods, or the second derivative of the phase ∇2Φ as measured in PBI.
Complex index of refraction
The complex index of refraction is a material specific quantity comprised by the refractive index decrement δ and the imaginary part of the refractive index β
1 |
and depends on the energy of the electromagnetic wave E [4, 84]. Figure 1 illustrates the complementarity of the attenuation coefficient μ and the refractive index decrement δ of an exemplary GBPC-CT phantom measurement. The imaginary part of the refractive index β is related to the attenuation coefficient μ as
2 |
with k = 2π/λ being the magnitude of the wave number. The refractive index decrement δ represents the phase-shift properties and is related to the electron density ρe in absence of absorption edges as
3 |
where r0 is the classical electron radius, is the reduced Planck constant, and c is the speed of light. In theory, the electron density can be calculated if the mass density ρ and composition of the material of interest are known [57]:
4 |
The weights wi account for the fraction of atom i. Zi represents the atomic number, Ai the atomic mass number, and NA Avogadro’s number. Changes in the phase induced by an object in the X-ray beam are proportional to the refractive index decrement δ as:
5 |
Grating interferometry
The focus of this review lies on quantitative grating-based phase contrast computed tomography (GBPC-CT). A laboratory GBPC-CT setup is illustrated in Fig. 2a. In detail, an X-ray Talbot interferometer uses the Talbot self-imaging effect, which has first been described for reappearing periodic structures with visible light and originates from Fresnel diffraction [83–85]. At specific distances d from those periodic structures, e.g., an optical grating, a moiré interference pattern appears. The period of the pattern p depends on the optical grating type which can in the case of X-rays be categorized as an attenuation or a phase grating, usually inducing a phase shift of either Δπ or Δπ/2. Due to angular refraction, a sample in the beam creates a lateral shift S of the interference pattern. For X-rays, the angular shift is typically in the order of micro-radians. Therefore, another grating—the so-called analyzer grating—is used to resolve the lateral shift with common detector pixel sizes and reasonable propagation distances. By moving one of the gratings in discrete steps over one period of the interference pattern, the lateral shift induced by a sample can be extracted as the lateral phase-shift of the resulting stepping curve φ. This process is called phase stepping [25].
This way, three different signals can be retrieved with GI, the conventional attenuation signal T, the differential phase contrast φ, and the dark-field signal, which represents the coherence reduction by the sample [23, 24, 86, 87]. More on the dark-field small-angle scattering signal can be found in Pfeiffer et al. [87]. There are also other extraction methods of the phase shift like Fourier transform–based single shot methods using one- or two-dimensional gratings [88, 89].
The phase shift φ of the interference pattern is related to the phase shift Φ of the wave front via the refraction angle α using small-angle approximation:
6 |
with p2 being the period of the analyzer grating and d the distance between the phase and the analyzer grating. An illustration of the effects taking place is shown in Fig. 2b. The phase shift Φ can then be directly related to the refractive index decrement δ as already shown in Eq. 5.
Signal extraction and CT reconstruction
Phase stepping
Phase stepping is recommended for stable retrieval of the differential phase contrast (DPC) signal φ for high angular sensitivity and thus high electron density resolution. The measured intensity within the phase-stepping approach is a convolution of the source shape, the intensity pattern, and the attenuation profile of the analyzer grating. This intensity in dependency of the grating position xg can be approximated to
7 |
Two phase-stepping curves are measured—one without the sample being the reference scan and one with the sample [25].
Attenuation contrast
The attenuation contrast is the average value of the intensity a0 of the stepping curve. Including sample and reference scan, the transmission signal is
8 |
with superscript “s” denoting the sample scan and superscript “r” the reference scan.
Differential phase contrast
The two stepping curves for the sample and the reference have different phase offsets. The difference between these signals is the differential phase contrast (DPC) signal
9 |
Advanced signal extraction methods to reduce jitter manifested in the stepping curve can be applied for homogeneous artifact-free DPC projections [90–94]. Additional phase ramp correction of the raw DPC signal is often needed to retrieve homogeneous DPC projections [95]. In order to reduce phase-wrapping artifacts, which occur if the phase shift is larger than 2π, putting the sample in a water container reduces phase contrast differences at the sample borders and thus also minimizes phase-wrapping artifacts [58, 59]. In addition, spectral phase unwrapping algorithms can be applied [96, 97].
Computed tomography
Grating interferometry can be combined with computed tomography [25, 26, 67]. The differential phase contrast projections can be reconstructed using filtered backprojection (FBP) with a Hilbert filter for phase integration [37]. Noise analysis in GBPC-CT reconstructions using FBP has been performed intensively in Raupach and Flohr [98], Köhler et al. [99], Chen et al. [100], and Weber et al. [101]. In direct combination with computed tomography, one can also apply reverse projection methods [67, 102–104] or phase contrast tomography with interlaced phase steps [105]. Moreover, statistical iterative reconstruction methods can be used for grating-based phase contrast computed tomography (GBPC-CT) when noise has to be reduced or the number of projections is limited [106–109] or the reconstruction should be performed without phase stepping [110, 111]. One can furthermore increase GBPC-CT image quality with bilateral filtering of the phase and the attenuation signal [112].
Laboratory GBPC-CT
GI can be used in a laboratory environment by implementing an additional grating directly behind the X-ray source as illustrated in Fig. 2a. This so-called source grating is an absorption grating which provides sufficient partial coherence for the method to work with extended X-ray sources like clinical X-ray tubes [27]. As an alternative to the source grating, one can also utilize micro focus X-ray source with small focal spot sizes [113, 114] or a structured anode [115, 116].
However, when using polychromatic X-rays, the performance of GBPC-CT suffers from spectral dependencies and from beam hardening as known from conventional attenuation CT [117–119]. The already presented water container acts also as beam filtration reducing spectral differences between the measurements with and without sample. Eventually one measures the relative differential phase contrast (DPC) signal. According to the relations shown in Eqs. 3 and 5, the relative DPC signal can be converted into the relative refractive index decrement δrel. In order to get absolute quantitative values, the corresponding refractive index decrement of the water has to be added to the experimentally determined relative signal δrel as
10 |
The value of is determined based on the electron density of water of 334e/nm3 and the effective energy Eeff [58, 59], see Eq. 3. Although the determination of quantitative values is requires a monochromatic spectrum as available at monochromatic synchrotron sources, the determination of accurate quantitative values is also possible with polychromatic sources [57–59, 76]. Therefore, an effective interaction energy has to be determined, either by calibration materials or by defining a mean energy from measured or simulated spectra. While the energy of the usually monochromatic beam at a synchrotron is known, one has to determine an effective energy for polychromatic X-ray sources. One way to determine the effective energy in the latter case is to use calibration materials like PMMA [57]. As a final GBPC-CT result, the electron density is calculated based on Eq. 3
11 |
Hounsfield units
In clinical CT, relative quantification of the attenuation is performed using the so-called Hounsfield units (HU), which express the attenuation compared to water
12 |
with the energy-dependent attenuation coefficients of water and air. Although the HU values are still energy dependent, this definition allows the comparison of HU values from measurements at different clinical CT systems in a range that is sufficient for reliable diagnostics. In a similar manner, Hounsfield units for the phase contrast signal can be defined based on the refractive index decrement δ [74, 75]:
13 |
The definition of HUp provides a normalized physical quantity comparable to the well known clinical HU. In contrast to HU, HUp do not show energy dependence and are therefore truly comparable between different setups.
Effective atomic number
With the electron density and attenuation coefficient at hand, as it is the case with GBPC-CT, one can determine also the effective atomic number Zeff as shown in refs. [57, 76, 78, 81]. The effective atomic number is like the electron density an absolute physical quantity that can be used for material characterization and decomposition, although theoretical calculation of the effective atomic number varies [120–122].
For calculation of the effective atomic number Zeff, there are two main approaches. In Qi et al. [76], the experimentally determined refractive index and attenuation coefficient are used for a calibration-based exponential fit of the effective atomic number:
14 |
with the three fit parameters p, q, and n. The fit parameters have to be determined in a prior GBPC-CT calibration experiment of a phantom with the same spectrum as the later GBPC-CT measurement of the sample and known reliable (effective) atomic number. The effective atomic number is then accessed by comparing the experimental ratio on the left side with the fit data.
In the method presented by Willner et al. [57], one can avoid this additional calibration step by using tabulated cross section data. The attenuation coefficient μ is related to the total atomic interaction cross section as
15 |
where ρ is the mass density and NA is Avogadro’s constant. The total atomic interaction cross section σtot(E,Z) depends on both the energy and the atomic number and is proportional to the electronic cross section as σtot(E,Z) = Z ⋅ σe,tot(E,Z). Rearranging 15 in combination with the formula for the electron density shown in Eq. 4 leads to
16 |
The ratio on the left side is determined in a GBPC-CT experiment and compared to tabulated data on the right side [123–127], which is interpolated to access effective atomic numbers Zeff.
High angular sensitivity in GBPC-CT
Reaching high electron density resolution with GBPC-CT is challenging, especially in the laboratory case. Depending on the desired application and spatial resolution, the performance of a GBPC-CT setup depends on numerous factors. Next to the choice of X-ray source and detector, the design of the GBPC-CT setup with position of gratings and sample have to be optimized [56, 119, 128–133]. The gratings form central elements of the interferometer featuring periods in the micrometer range. While phase gratings are comparably easy to fabricate, absorption gratings for the relevant X-ray energies are challenging to fabricate as they need high attenuating lamellae at small periods [23, 134–138]. Photon counting detectors [139–141] and spectral detectors [142] are able to reduce noise in GBPC-CT and thus increase the electron density resolution [82, 143].
Experimentally, the electron density resolution Δρe can be determined based on the standard deviation of the reconstructed refractive index decrement in a homogeneous sample volume as
17 |
similar to Eq. 11.
Applications of quantitative GBPC-CT
Due to the increased soft tissue contrast at high resolution, GBPC-CT is a promising technique for biomedical imaging. Various biomedical studies have shown applications of quantitative GBPC micro-CT of ex vivo samples at both synchrotron and laboratory ranging from the quantitative classification of different soft tissues and body fluids to the visualization of pathological changes in biopsy samples. While specific applications also aim towards clinical in vivo imaging, most studies presented in this section focus on small ex vivo tissue samples, which could be used for improving clinical histology or to address basic research questions (Fig. 3).
One of the first quantitative GBPC-CT imaging results are shown in Pfeiffer et al. [67], which features a rat brain with an induced gliosarcoma tumor. The presented electron density resolution of this synchrotron measurement was 0.18e/nm3, while the contrast in the corresponding attenuation signal was limited. Several further GBPC-CT experiments from synchrotron sources have illustrated the potential of quantitative biomedical imaging [105, 144–149]. It was even possible to visualize human cerebellar structures with high quantitative soft tissue contrast comparable to MRI but at much higher spatial resolution [68, 69]. High-resolution GBPC-CT imaging at subtle soft tissue differentiation based on the electron density allowed a big step towards X-ray 3D quantitative phase contrast histology as published in Zanette et al. [70]. More recent results by Thalmann et al. allow to differentiate quantitatively nerve fibers at high spatial resolution [71].
The transition of quantitative GBPC-CT results from synchrotron facilities to laboratory X-ray sources allowed an impact in research activity extending the range of applications [58, 75, 132]. In Herzen et al., first characterization of fluids was performed in a laboratory environment where the quantitative material values of the GBPC-CT were validated with tabulated reference values. By that the complementarity of phase and attenuation and the feasibility of quantitative measurements at a conventional X-ray source were demonstrated [59]. It could also be shown that GBPC-CT at a polychromatic source operating at 70 kVp, which is comparably high, provides good quantitative results [150]. However, increasing the performance of GI for higher energies depends strongly on grating fabrication. Absorption gratings at small periods and large attenuation material thickness of the lamellae are technically challenging [138]. Laboratory GI setups operating at 160 kVp have been successfully realized as proof of concept [151].
Further development of GBPC-CT imaging increased the performance with respect to the electron density resolution in the lab [56, 141, 152, 153]. In Birnbacher et al., highly sensitive electron density resolution could be realized with laboratory sources, as visualized by the discrimination of cerebellar tissue in Fig. 4. The electron density resolution was 0.45e/nm3, which is comparable to synchrotron experiments [56].
One focus of high-resolution quantitative GBPC-CT imaging is breast imaging. Early architectural changes of breast cancer or its precursor forms can be visualized based on the electron density with both synchrotron and laboratory studies [41, 72]. Willner et al. demonstrated the clear differentiation of adipose and fibroglandular tissue based on their HUp, where a differentiation in conventional CT would not be possible due to very similar HU [57]. A further study investigated the discrimination of fibroadenoma from other breast lesions also based on HUp values [73] (see Fig. 5). The corresponding histograms illustrate the clear separation of tissues of interest. Exemplary quantitative GBPC-CT data of a ductal carcinoma in situ (DCIS) sample is illustrated in Fig. 3. In comparison to the corresponding attenuation image, which does not reveal specific soft tissue contrast like thickening of the ductal wall, the phase contrast image reveals similar findings as the corresponding histological image [41].
Atherosclerotic plaque was analyzed quantitatively in Hetterich et al. [62]. There, calcified, lipid-rich, and fibrous tissue associated with the formation of atherosclerotic plaque was characterized. Similar results were published in Winklhofer et al. [154] and in Bonnano et al. [155]. In Notohamiprodjo et al., HUp changes in myocardial tissue types were compared with grades of myocardial lesions [61]. A multi-modal comparison of different imaging methods for the characterization of cartilage degeneration revealed additional quantitative information on morphology and ultra-structure by GBPC-CT compared to MRI [156]. Khimchenko et al. presented electron density GBPC-CT images of cartilage and bone using an adapted commercial micro-CT system [114]. Perfusion of ischemic kidneys was analyzed with X-ray phase contrast imaging in [157]. In Braunagel et al., renal cell carcinoma components were analyzed with GBPC-CT to characterize the subtypes clear cell, papillary, and chromophobe renal carcinoma based on HUp values [63]. Exemplary GBPC-CT slices compared with histology and quantitative HUp results of this study are depicted in Fig. 6.
With the development of compact inverse Compton sources, the gap between large scale synchrotron facilities and laboratory sources has been closed. These sources provide quasi-monochromatic X-rays with tunable energy [158]. Eggl et al. [159] demonstrated tomographic GBPC-CT experiments at an early prototype of such a compact light source, which allowed the visualization of brown adipose tissue in mice [160]. Similar results were achieved at a laboratory setup, which demonstrated transition from brown to white adipose tissue in mice utilizing three-dimensional anatomical electron density distribution. The biochemical changes in adipose tissue could thereby be related quantitatively to the changes in mitochondrial concentration [161].
Since GBPC-CT setups also enable the determination of the attenuation coefficient, this additional information can be used for quantitative material decomposition based on signal complementarity. In Willner et al. [79], a decomposition of GBPC-CT data based on phase contrast and attenuation data into lipid, protein, and water allowed to quantify the components of soft tissue, which is visualized by way of example in Fig. 7. Quantitative mass density measurements of differences in human teeth composition including dentin and enamel were realized in synchrotron experiments at 53 keV. The mass density was thereby approximated based on the electron density [162]. An analysis study of blood and pathological fluids complemented research in that direction. Different types of infectious fluids were differentiated based on their composition like cell, protein, and water content [66]. GBPC-CT experiments are typically performed with fixed samples to prevent degradation. The tissue fixation with formaldehyde or phosphate buffered saline (PBS) changes quantitative values slightly in comparison to reference electron density values [60, 74]. A similar GBPC-CT study of simulated liver lesion analyzed fluids based on the renal cyst composition (protein rich, hemorrhagic, and enhanced) [64].
As already mentioned, the combination of the electron density and the attenuation coefficient enables calculation of the effective atomic number [57, 76, 78]. As the measured attenuation and phase contrast signals deliver two independent data sets, they can be used for quantitative material decomposition as known from dual-energy imaging. The calculation of iodine maps, virtual non-contrast images, and virtual monoenergetic images has been demonstrated with measurements at the Munich Compact Light Source (MuCLS) [80], as illustrated in Figs. 1 and 8. Corresponding GBPC-CT quantitative values of the phantom used in this work are presented in Table 1.
Table 1.
Material | μm[cm− 1] | μl[cm− 1] | δm[10− 7] | δl[10− 7] | Zeff,m | ρe,m[1029m− 3] | ρe,l[1029m− 3] | ||
---|---|---|---|---|---|---|---|---|---|
NaI | 0.597 ± 0.006 | 0.609 | 4.08 ± 0.07 | 4.07 | 7.94 ± 0.06 | – | – | 3.34 ± 0.07 | – |
Blood | 0.568 ± 0.006 | 0.574 | 4.31 ± 0.07 | 4.27 | 7.60 ± 0.06 | – | 7.74 | 3.54 ± 0.08 | – |
Ethanol | 0.323 ± 0.006 | 0.325 | 3.23 ± 0.06 | 3.27 | 6.52 ± 0.07 | 6.35 | – | 2.69 ± 0.07 | 2.68 |
PMMA | 0.470 ± 0.006 | 0.470 | 4.70 ± 0.06 | 4.70 | 6.58 ± 0.03 | 6.47 | 6.56 | 3.86 ± 0.06 | 3.86 |
POM | 0.628 ± 0.006 | 0.628 | 5.55 ± 0.06 | 5.55 | 7.05 ± 0.05 | 6.95 | 7.03 | 4.56 ± 0.06 | 4.56 |
Nylon | 0.423 ± 0.006 | 0.419 | 4.61 ± 0.06 | 4.58 | 6.24 ± 0.06 | 6.12 | 6.21 | 3.79 ± 0.06 | 3.76 |
Water | 0.523 ± 0.006 | 0.523 | 4.07 ± 0.06 | 4.07 | 7.51 ± 0.05 | 7.42 | 7.51 | 3.34 ± 0.06 | 3.34 |
The table shows quantitative results of the attenuation coefficient μ and refractive index decrement δ as well as the determined electron density ρe and effective atomic number Zeff compared to reference data marked by “l.” Experimental GBPC-CT data is marked by “m.” Data adapted from Braig et al. (CC BY) [80]
In Braig et al., three-material decomposition was performed using electron density, effective atomic number, and the dark-field signal provided by laboratory GBPC-CT. The study presented there lead to precise calculation iodine contrast agent concentration [81]. A further step is the combination of spectral detector information with GBPC-CT, which enables noise reduction and consequent increase in quantitative electron density resolution [143]. Latest results of laboratory GBPC-CT combined with a spectral detector with two energy bins improved contrast enhancement in three-material GBPC-CT decomposition [82].
Related phase contrast imaging methods
In this section, we present shortly related PCI techniques and their ability to provide quantitative results since GBPC-CT is only one among many X-ray PCI methods.
Free-space propagation-based imaging (PBI) [13–17] can be achieved either with multiple distances as so-called holotomography [163, 164] or as a single distance experiment [165]. High resolution close to histology can be achieved with PBI [166]. However, with single distance PBI tomography, the quantitativeness of heterogeneous samples is limited [55]. PBI does not only work with highly coherent synchrotron sources but also with polychromatic laboratory sources like micro focus [36, 167, 168] or liquid jet sources [38, 39]. Generally speaking for PBI, the source size and the detector pixel size should be relatively small; hence, clinical X-ray imaging components cannot be used. With respect to quantitative imaging, the broad spectrum does not limit the application, but cone beam geometry renders quantitative multi-distance phase contrast results more challenging. One large research focus is breast imaging including tomography [169, 170].
In analyzer-based imaging (ABI) or diffraction enhanced imaging (DEI), the X-ray beam is refracted by a crystal, which is rotated for maximum intensity forming a so-called rocking curve. A sample in the beam changes this curve, which is related to the differential phase contrast signal of the sample [18–22]. Application of ABI includes especially many studies of breast imaging in projection and tomography [171–174]. To our knowledge, ABI is available only at synchrotron facilities and does not provide absolute quantitative values.
Edge illumination (EI) or coded aperture is a non- interferometric phase contrast method. Two gratings serve as sample and detector mask and allow the determination of the refractive angle and thus the differential phase signal [28–31]. The method features relaxed requirements in comparison to GI with respect to the grating periods and can be applied with X-ray sources with large focal spots, which renders the method interesting for the use of clinical X-ray sources [28, 29]. Quantitative EI can be realized either by two full individual scans in opposite direction or within a certain range of accuracy in a single scan if the sample attenuation is low [175]. Applications include breast imaging [176] or tomography of biological samples [177]. Quantitative EI tomography also works with laboratory X-ray sources [40].
Speckle-based X-ray phase contrast imaging uses a wave front modulator pattern, which is changed by an object in the beam. This modulator can be quite simple like e.g. sandpaper [35]. Compared to this, the fabrication of the gratings in grating interferometry is quite challenging [178]. This tracking-based method works with polychromatic sources; however, microfocus sources have to be used [35, 178–180]. Promising tomographic electron density results of biomedical samples were shown at high spatial resolution, however only at synchrotron sources [181–183]. More information can be found in the review articles presented in the introduction of this article.
Conclusion
X-ray grating-based phase contrast computed tomography provides the energy-independent quantities electron density and effective atomic number. The use of a source grating enables the extraction of the material quantities with high accuracy at laboratory sources. GBPC-CT data can also be converted into phase contrast Hounsfield units (HUp) in similarity to the conventional energy-dependent Hounsfield units HU in clinical CT. This possibility for improved tissue characterization based on absolute quantitative values increases the range of phase contrast micro-CT aside from mere relative contrast-based images. In this review, we summarized theory and applications of quantitative GBPC-CT imaging.
The range of GBPC-CT method goes from differentiation of pathological tissue like myocardial infarct or atherosclerotic plaque to tumors of kidney, brain, and especially breast. Thereby, subtle differences in electron density can be determined. Moreover, biochemical or biomedical fluids can be characterized quantitatively. The complementarity between the attenuation and phase signal improves the distinction of objects and enables quantitative material decomposition. Additional consideration of the dark-field signal allows the decomposition into up to three materials.
A potential application in this direction could be pre-histology GBPC-CT, where quantitative GBPC-CT data could be used for advanced tissue characterization leading to more precise and efficient sectioning in histology. Having the ability to access absolute quantitative GBPC-CT data could be used for deep learning and could also lead to improved histological diagnosis [184–186].
While many studies focus on tissue characterization in GBPC micro-CT at high spatial and electron density resolution, one promising development of GBPC-CT towards clinical application could be phase contrast breast tomography [187, 188]. The needed energy range of breast imaging is still in a range feasible for complementarity of the attenuation and phase contrast signal, as the tube voltage for breast phase contrast CT would be lower than conventional CT tube voltages. In combination with spectral detectors and iterative reconstruction, this would allow quantitative assessment of breast tissue and characterization of breast lesions [143].
In order for GBPC-CT to play a larger role in high-resolution biomedical micro-CT imaging, the spatial resolution laboratory environment has to increase aside from further optimization with respect to imaging time. Yet, the availability of the electron density with high accuracy renders GBPC-CT eminently interesting for quantitative biomedical and material science micro-CT applications. Further development of spectral detectors [143], using a combination of analyzer grating and detector [189], or the implementation of a dual phase grating approach [190–192] is expected to increase the performance of laboratory GBPC-CT in the near future.
Acknowledgements
We thank all colleagues for providing figures. We acknowledge the support through the European Research Council (AdG 695045) and the German Research Foundation (DFG) under the Gottfried Wilhelm Leibniz Program and the Research Training Group grant GRK 2274. This work was carried out with the support of the Karlsruhe Nano Micro Facility (KNMF, www.kit.edu/knmf), a Helmholtz research infrastructure at Karlsruhe Institute of Technology (KIT).
Author contributions
LB, EMP, DP, FP, and JH had the idea for the article; LB and EMP performed literature search and data analysis; LB and EMP drafted the work; and DP, FP, and JH critically revised the work.
Funding
Open Access funding enabled and organized by Projekt DEAL. European Research Council under grant AdG 695045, German Research Foundation (DFG) under the Gottfried Wilhelm Leibniz Program and under grant GRK 2274.
Declarations
Ethics approval and consent to participate
This review article does not contain any studies with human participants or animals performed by any of the authors.
Conflict of interest
The authors declare no competing interests.
Footnotes
This article is part of the Topical Collection on Technology
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Fitzgerald R. Phase-sensitive X-ray imaging. Phys Today. 2000;53(7):23–26. [Google Scholar]
- 2.Momose A. Recent advances in X-ray phase imaging. Jap J Appl Phys. 2005;44:6355–6367. [Google Scholar]
- 3.Zhou SA, Brahme A. Development of phase-contrast X-ray imaging techniques and potential medical applications. Phys Med. 2008;24(3):129–48. doi: 10.1016/j.ejmp.2008.05.006. [DOI] [PubMed] [Google Scholar]
- 4.Nugent KA. Coherent methods in the X-ray sciences. Adv Phys. 2010;59:1, 1–99. [Google Scholar]
- 5.Pfeiffer F. 2012. Milestones and basic principles of grating-based x-ray and neutron phase-contrast imaging. AIP Conf Proc 1466(2). 10.1063/1.4742261.
- 6.Bravin A, Coan P, Suortti P. X-ray phase-contrast imaging: from pre-clinical applications towards clinics. Phys Med Biol. 2013;58(1):R1–35. doi: 10.1088/0031-9155/58/1/R1. [DOI] [PubMed] [Google Scholar]
- 7.Pfeiffer F, Herzen J, Willner M, Chabior M, Auweter S, Reiser M, Bamberg F. Grating-based X-ray phase contrast for biomedical imaging applications. Z Med Phys. 2013;3:176–85. doi: 10.1016/j.zemedi.2013.02.002. [DOI] [PubMed] [Google Scholar]
- 8.Wilkins S, Nesterets YI, Gureyev TE, Mayo SC, Pogany A, Stevenson AW. On the evolution and relative merits of hard X-ray phase-contrast imaging methods. Philos Trans A Math Phys Eng Sci. 2014;372(2010):20130021. doi: 10.1098/rsta.2013.0021. [DOI] [PubMed] [Google Scholar]
- 9.Endrizzi M. X-ray phase-contrast imaging. Nucl Instrum Methods Phys Res A. 2018;878:88–98. [Google Scholar]
- 10.Zdora MC. State of the art of X-ray speckle-based phase-contrast and dark-field imaging. J Imaging. 2018;4(60):2018. [Google Scholar]
- 11.Bonse U, Hart M. An X-ray interferometer with long interfering beam paths. Appl Phys Lett. 1965;7:99–101. [Google Scholar]
- 12.Momose A, Takeda T, Itai Y, Hirano K. Phase-contrast X-ray computed tomography for observing biological soft tissues. Nat Med. 1996;2(4):473–5. doi: 10.1038/nm0496-473. [DOI] [PubMed] [Google Scholar]
- 13.Snigirev A, Snigireva I, Kohn V, Kuznetsov S, Schelokov I. On the possibilities of X-ray phase contrast microimaging by coherent high-energy synchrotron radiation. Rev Sci Instrum. 1995;66:5486. [Google Scholar]
- 14.Nugent KA, Gureyev TE, Cookson DJ, Paganin D, Barnea Z. Quantitative phase imaging using hard X rays. Phys Rev Lett. 1996;77(14):2961–2964. doi: 10.1103/PhysRevLett.77.2961. [DOI] [PubMed] [Google Scholar]
- 15.Cloetens P, Barrett R, Baruchel J, Guigay JP, Schlenker M. Phase objects in synchrotron radiation hard x-ray imaging. J Phys D Appl Phys. 1996;29:133. [Google Scholar]
- 16.Wilkins S, Gureyev T, Gao D, Pogany A, Stevenson AW. Phase-contrast imaging using polychromatic hard X-rays. Nature. 1996;384:335–338. [Google Scholar]
- 17.Paganin D, Nugent KA. Noninterferometric phase imaging with partially coherent light. Phys Rev Lett. 1998;80:2586. [Google Scholar]
- 18.Davis TJ, Gao D, Gureyev TE, Stevenson AW, Wilkins S. 1995. Phase-contrast of weakly absorbing materials using hard X-rays. Nature 373595–8. 10.1038/373595a0.
- 19.Chapman D, Thomlinson W, Johnston RE, Washburn D, Pisano E, Gmür N, Zhong Z, Menk R, Arfelli F, Sayers D. Diffraction enhanced x-ray imaging. Phys Med Biol. 1997;42(11):2015–25. doi: 10.1088/0031-9155/42/11/001. [DOI] [PubMed] [Google Scholar]
- 20.Gureyev TE, Wilkins S. Regimes of X-ray phase-contrast imaging with perfect crystals. Nouv Cim D. 1997;19:545–552. [Google Scholar]
- 21.Mayo S, Davis T, Gureyev T, Miller P, Paganin D, Pogany A, Stevenson A, Wilkins S. X-ray phase-contrast microscopy and microtomography. Opt Express. 2003;11(19):2289–302. doi: 10.1364/oe.11.002289. [DOI] [PubMed] [Google Scholar]
- 22.Nesterets YI, Gureyev TE, Paganin D, Pavlov KM, Wilkins S. Quantitative diffraction-enhanced x-ray imaging of weak objects. J Phys D, Appl Phys. 2004;37:1262–1274. [Google Scholar]
- 23.David C, Nöhammer B, Solak H, Ziegler E. Differential X-ray phase contrast imaging using a shearing interferometer. Appl Phys Lett. 2002;81:3287–3289. [Google Scholar]
- 24.Momose A. Phase-sensitive imaging and phase tomography using X-ray interferometers. Opt Express. 2003;11(19):2303–14. doi: 10.1364/oe.11.002303. [DOI] [PubMed] [Google Scholar]
- 25.Weitkamp T, Diaz A, David C, Pfeiffer F, Stampanoni M, Cloetens P, Ziegler E. X-ray phase imaging with a grating interferometer. Opt Express. 2005;13(16):6296–304. doi: 10.1364/opex.13.006296. [DOI] [PubMed] [Google Scholar]
- 26.Momose A, Yashiro W, Maikusa H, Takeda Y. High-speed X-ray phase imaging and X-ray phase tomography with Talbot interferometer and white synchrotron radiation. Opt Express. 2009;17(15):12540–5. doi: 10.1364/oe.17.012540. [DOI] [PubMed] [Google Scholar]
- 27.Pfeiffer F, Weitkamp T, Bunk O, David C. Phase retrieval and differential phase-contrast imaging with low brilliance X-ray sources. Nat Phys. 2006;2:258–261. [Google Scholar]
- 28.Olivo A, Arfelli F, Cantatore G, Longo R, Menk RH, Pani S, Prest M, Poropat P, Rigon L, Tromba G, Vallazza E, Castelli E. An innovative digital imaging set-up allowing a low-dose approach to phase contrast applications in the medical field. Med Phys. 2001;28(8):1610–9. doi: 10.1118/1.1388219. [DOI] [PubMed] [Google Scholar]
- 29.Olivo A, Speller R. Modelling of a novel x-ray phase contrast imaging technique based on coded apertures. Phys Med Biol. 2007;52(22):6555–73. doi: 10.1088/0031-9155/52/22/001. [DOI] [PubMed] [Google Scholar]
- 30.Munro PR, Ignatyev K, Speller R, Olivo A. Source size and temporal coherence requirements of coded aperture type x-ray phase contrast imaging systems. Opt Express. 2010;18(19):19681–92. doi: 10.1364/OE.18.019681. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Endrizzi M, Vittoria FA, Kallon G, Basta D, Diemoz PC, Vincenzi A, Delogu P, Bellazzini R, Olivo A. Achromatic approach to phase-based multi-modal imaging with conventional X-ray sources. Opt Express. 2015;23(12):16473–80. doi: 10.1364/OE.23.016473. [DOI] [PubMed] [Google Scholar]
- 32.Cerbino R, Peverini L, Potenza M, et al. X-ray-scattering information obtained from near-field speckle. Nature Phy. 2008;4:238–243. [Google Scholar]
- 33.Morgan KS, Paganin DM, Siu KK. Quantitative single-exposure x-ray phase contrast imaging using a single attenuation grid. Opt Express. 2011;19(20):19781–9. doi: 10.1364/OE.19.019781. [DOI] [PubMed] [Google Scholar]
- 34.Morgan KS, Paganin DM, Siu KK. Quantitative x-ray phase-contrast imaging using a single grating of comparable pitch to sample feature size. Opt Lett. 2011;36(1):55–7. doi: 10.1364/OL.36.000055. [DOI] [PubMed] [Google Scholar]
- 35.Zanette I, Zhou T, Burvall A, Lundström U, Larsson DH, Zdora M, Thibault P, Pfeiffer F, Hertz HM. Speckle-based x-ray phase-contrast and dark-field imaging with a laboratory source. Phys Rev Lett. 2014;112(25):253903. doi: 10.1103/PhysRevLett.112.253903. [DOI] [PubMed] [Google Scholar]
- 36.Pogany A, Gao D, Wilkins S. Contrast and resolution in imaging with a microfocus x-ray source. Rev Sci Instr. 1997;68(7):2774–2782. [Google Scholar]
- 37.Pfeiffer F, Kottler C, Bunk O, David C. Hard x-ray phase tomography with low-brilliance sources. Phys Rev Lett. 2007;98(10):108105. doi: 10.1103/PhysRevLett.98.108105. [DOI] [PubMed] [Google Scholar]
- 38.Töpperwien M, van der Meer F, Stadelmann C, Salditt T. Three-dimensional virtual histology of human cerebellum by X-ray phase-contrast tomography. Proc Natl Acad Sci USA. 2018;115(27):6940–6945. doi: 10.1073/pnas.1801678115. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Krenkel M, Töpperwien M, Dullin C, Alves F, Salditt T. Propagation-based phase-contrast tomographyfor high-resolution lung imaging with laboratory sources. AIP Adv. 2016;035007:6. [Google Scholar]
- 40.Hagen CK, Maghsoudlou P, Totonelli G, Diemoz PC, Endrizzi M, Rigon L, Menk RH, Arfelli F, Dreossi D, Brun E, Coan P, Bravin A, De Coppi P, Olivo A. High contrast microstructural visualization of natural acellular matrices by means of phase-based x-ray tomography. Sci Rep. 2015;5:18156. doi: 10.1038/srep18156. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Hellerhoff K, Birnbacher L, Sztrókay-Gaul A, Grandl S, Auweter S, Willner M, Marschner M, Mayr D, Reiser MF, Pfeiffer F, Herzen J. Assessment of intraductal carcinoma in situ (DCIS) using grating-based X-ray phase-contrast CT at conventional X-ray sources: an experimental ex-vivo study. PLoS One. 2019;14(1):e0210291. doi: 10.1371/journal.pone.0210291. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Sardanelli F. Trends in radiology and experimental research. Eur Radiol Exp. 2017;1(1):1. doi: 10.1186/s41747-017-0006-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Alvarez RE, Macovski A. Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol. 1976;21(5):733–44. doi: 10.1088/0031-9155/21/5/002. [DOI] [PubMed] [Google Scholar]
- 44.Kalender WA, Perman WH, Vetter JR, Klotz E. Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys. 1986;13(3):334–9. doi: 10.1118/1.595958. [DOI] [PubMed] [Google Scholar]
- 45.Torikoshi M, Tsunoo T, Sasaki M, Endo M, Noda Y, Ohno Y, Kohno T, Hyodo K, Uesugi K, Yagi N. Electron density measurement with dual-energy x-ray CT using synchrotron radiation. Phys Med Biol. 2003;48(5):673–85. doi: 10.1088/0031-9155/48/5/308. [DOI] [PubMed] [Google Scholar]
- 46.Schlomka JP, Roessl E, Dorscheid R, Dill S, Martens G, Istel T, Bäumer C, Herrmann C, Steadman R, Zeitler G, Livne A, Proksa R. Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Phys Med Biol. 2008;53(15):4031–47. doi: 10.1088/0031-9155/53/15/002. [DOI] [PubMed] [Google Scholar]
- 47.McCollough CH, Leng S, Yu L, Fletcher JG. Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology. 2015;276(3):637–53. doi: 10.1148/radiol.2015142631. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.van Abbema JK, van Goethem MJ, Greuter MJ, van der Schaaf A, Brandenburg S, van der Graaf ER. Relative electron density determination using a physics based parameterization of photon interactions in medical DECT. Phys Med Biol. 2015;60(9):3825–46. doi: 10.1088/0031-9155/60/9/3825. [DOI] [PubMed] [Google Scholar]
- 49.Lee SH, Hur J, Kim YJ, Lee HJ, Hong YJ, Choi BW. Additional value of dual-energy CT to differentiate between benign and malignant mediastinal tumors: an initial experience. Eur J Radiol. 2013;82(11):2043–9. doi: 10.1016/j.ejrad.2013.05.040. [DOI] [PubMed] [Google Scholar]
- 50.Mileto A, Marin D, Alfaro-Cordoba M, Ramirez-Giraldo JC, Eusemann CD, Scribano E, Blandino A, Mazziotti S, Ascenti G. Iodine quantification to distinguish clear cell from papillary renal cell carcinoma at dual-energy multidetector CT: a multireader diagnostic performance study. Radiology. 2014;273(3):813–20. doi: 10.1148/radiol.14140171. [DOI] [PubMed] [Google Scholar]
- 51.Kaltenbach B, Wichmann JL, Pfeifer S, Albrecht MH, Booz C, Lenga L, Hammerstingl R, D’Angelo T, Vogl TJ, Martin SS. Iodine quantification to distinguish hepatic neuroendocrine tumor metastasis from hepatocellular carcinoma at dual-source dual-energy liver CT. Eur J Radiol. 2018;105:20–24. doi: 10.1016/j.ejrad.2018.05.019. [DOI] [PubMed] [Google Scholar]
- 52.Boll DT, Patil NA, Paulson EK, Merkle EM, Simmons WN, Pierre SA, Preminger GM. Renal stone assessment with dual-energy multidetector CT and advanced postprocessing techniques: improved characterization of renal stone composition–pilot study. Radiology. 2009;250(3):813–20. doi: 10.1148/radiol.2503080545. [DOI] [PubMed] [Google Scholar]
- 53.Leng S, Huang A, Cardona JM, Duan X, Williams JC, McCollough CH. Dual-energy CT for quantification of urinary stone composition in mixed stones: a phantom study. AJR Am J Roentgenol. 2016;207(2):321–9. doi: 10.2214/AJR.15.15692. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Tijssen MP, Hofman PA, Stadler AA, van Zwam W, de Graaf R, van Oostenbrugge RJ, Klotz E, Wildberger JE, Postma AA. The role of dual energy CT in differentiating between brain haemorrhage and contrast medium after mechanical revascularisation in acute ischaemic stroke. Eur Radiol. 2014;24(4):834–40. doi: 10.1007/s00330-013-3073-x. [DOI] [PubMed] [Google Scholar]
- 55.Zhou T, Lundström U, Thüring T, Rutishauser S, Larsson DH, Stampanoni M, David C, Hertz HM, Burvall A. Comparison of two x-ray phase-contrast imaging methods with a microfocus source. Opt Express. 2013;21(25):30183–95. doi: 10.1364/OE.21.030183. [DOI] [PubMed] [Google Scholar]
- 56.Birnbacher L, Willner M, Velroyen A, Marschner M, Hipp A, Meiser J, Koch F, Schröter T, Kunka D, Mohr J, Pfeiffer F, Herzen J. Experimental realisation of high-sensitivity laboratory x-ray grating-based phase-contrast computed tomography. Sci Rep. 2016;6:24022. doi: 10.1038/srep24022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Willner M, Herzen J, Grandl S, Auweter S, Mayr D, Hipp A, Chabior M, Sarapata A, Achterhold K, Zanette I, Weitkamp T, Sztrókay A, Hellerhoff K, Reiser M, Pfeiffer F. Quantitative breast tissue characterization using grating-based x-ray phase-contrast imaging. Phys Med Biol. 2014;59(7):1557–71. doi: 10.1088/0031-9155/59/7/1557. [DOI] [PubMed] [Google Scholar]
- 58.Bech M, Jensen TH, Feidenhans’l R, Bunk O, David C, Pfeiffer F. Soft-tissue phase-contrast tomography with an x-ray tube source. Phys Med Biol. 2009;54(9):2747–53. doi: 10.1088/0031-9155/54/9/010. [DOI] [PubMed] [Google Scholar]
- 59.Herzen J, Donath T, Pfeiffer F, Bunk O, Padeste C, Beckmann F, Schreyer A, David C. Quantitative phase-contrast tomography of a liquid phantom using a conventional x-ray tube source. Opt Express. 2009;17(12):10010–8. doi: 10.1364/oe.17.010010. [DOI] [PubMed] [Google Scholar]
- 60.Woodard HQ, White DR. The composition of body tissues. Br J Radiol. 1986;59(708):1209–1218. doi: 10.1259/0007-1285-59-708-1209. [DOI] [PubMed] [Google Scholar]
- 61.Notohamiprodjo S, Webber N, Birnbacher L, Willner M, Viermetz M, Herzen J, Marschner M, Mayr D, Bartsch H, Saam T, Auweter S, Pfeiffer F, Reiser MF, Hetterich H. Qualitative and quantitative evaluation of structural myocardial alterations by grating-based phase-contrast computed tomography. Invest Radiol. 2018;53(1):26–34. doi: 10.1097/RLI.0000000000000408. [DOI] [PubMed] [Google Scholar]
- 62.Hetterich H, Willner M, Habbel C, Herzen J, Hoffmann VS, Fill S, Hipp A, Marschner M, Schüller U, Auweter S, Massberg S, Reiser MF, Pfeiffer F, Saam T, Bamberg F. X-ray phase-contrast computed tomography of human coronary arteries. Invest Radiol. 2015;50(10):686–94. doi: 10.1097/RLI.0000000000000169. [DOI] [PubMed] [Google Scholar]
- 63.Braunagel M, Birnbacher L, Willner M, Marschner M, De Marco F, Viermetz M, Notohamiprodjo S, Hellbach K, Auweter S, Link V, Woischke C, Reiser MF, Pfeiffer F, Notohamiprodjo M, Herzen J. Qualitative and quantitative imaging evaluation of renal cell carcinoma subtypes with grating-based x-ray phase-contrast CT. Sci Rep. 2017;7:45400. doi: 10.1038/srep45400. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Fingerle AA, Willner M, Herzen J, Münzel D, Hahn D, Rummeny EJ, Noël PB, Pfeiffer F. Simulated cystic renal lesions: quantitative X-ray phase-contrast CT–an in vitro phantom study. Radiology. 2014;272(3):739–48. doi: 10.1148/radiol.14130876. [DOI] [PubMed] [Google Scholar]
- 65.Tapfer A, Braren R, Bech M, Willner M, Zanette I, Weitkamp T, Trajkovic-Arsic M, Siveke JT, Settles M, Aichler M, Walch A, Pfeiffer F. X-ray phase-contrast CT of a pancreatic ductal adenocarcinoma mouse model. PLoS One. 2013;8(3):e58439. doi: 10.1371/journal.pone.0058439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Richter V, Willner MS, Henningsen J, Birnbacher L, Marschner M, Herzen J, Kimm MA, Noël PB, Rummeny EJ, Pfeiffer F, Fingerle AA. Ex vivo characterization of pathologic fluids with quantitative phase-contrast computed tomography. Eur J Radiol. 2017;86:99–104. doi: 10.1016/j.ejrad.2016.11.014. [DOI] [PubMed] [Google Scholar]
- 67.Pfeiffer F, Bunk O, David C, Bech M, Le Duc G, Bravin A, Cloetens P. High-resolution brain tumor visualization using three-dimensional x-ray phase contrast tomography. Phys Med Biol. 2007;52(23):6923–30. doi: 10.1088/0031-9155/52/23/010. [DOI] [PubMed] [Google Scholar]
- 68.Schulz G, Weitkamp T, Zanette I, Pfeiffer F, Beckmann F, David C, Rutishauser S, Reznikova E, Müller B. High-resolution tomographic imaging of a human cerebellum: comparison of absorption and grating-based phase contrast. J R Soc Interface. 2010;7(53):1665–76. doi: 10.1098/rsif.2010.0281. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Schulz G, Waschkies C, Pfeiffer F, Zanette I, Weitkamp T, David C, Müller B. Multimodal imaging of human cerebellum - merging X-ray phase microtomography, magnetic resonance microscopy and histology. Sci Rep. 2012;2:826. doi: 10.1038/srep00826. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Zanette I, Weitkamp T, Le Duc G, Pfeiffer F. X-ray grating-based phase tomography for 3D histology. RSC Adv. 2013;3(43):19816. [Google Scholar]
- 71.Thalmann P, Bikis C, Hipp A, Müller B, Hieber SE, Schulz G. Single and double grating-based X-ray microtomography using synchrotron radiation. Single Appl Phys Lett. 2017;110:061103. [Google Scholar]
- 72.Sztrókay A, Herzen J, Auweter SD, Liebhardt S, Mayr D, Willner M, Hahn D, Zanette I, Weitkamp T, Hellerhoff K, Pfeiffer F, Reiser MF, Bamberg F. Assessment of grating-based X-ray phase-contrast CT for differentiation of invasive ductal carcinoma and ductal carcinoma in situ in an experimental ex vivo set-up. Eur Radiol. 2013;23(2):381–7. doi: 10.1007/s00330-012-2592-1. [DOI] [PubMed] [Google Scholar]
- 73.Grandl S, Willner M, Herzen J, Sztrókay-Gaul A, Mayr D, Auweter SD, Hipp A, Birnbacher L, Marschner M, Chabior M, Reiser M, Pfeiffer F, Bamberg F, Hellerhoff K. Visualizing typical features of breast fibroadenomas using phase-contrast CT: an ex-vivo study. PLoS One. 2014;9(5):e97101. doi: 10.1371/journal.pone.0097101. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Willner M, Fior G, Marschner M, Birnbacher L, Schock J, Braun C, Fingerle AA, Noël PB, Rummeny EJ, Pfeiffer F, Herzen J. 2015. Phase-contrast Hounsfield units of fixated and non-fixated soft-tissue samples, Vol. 10. 10.1371/journal.pone.0137016. [DOI] [PMC free article] [PubMed]
- 75.Donath T, Pfeiffer F, Bunk O, Grünzweig C, Hempel E, Popescu S, Vock P, David C. Toward clinical X-ray phase-contrast CT: demonstration of enhanced soft-tissue contrast in human specimen. Invest Radiol. 2010;45(7):445–52. doi: 10.1097/RLI.0b013e3181e21866. [DOI] [PubMed] [Google Scholar]
- 76.Qi Z, Zambelli J, Bevins N, Chen GH. Quantitative imaging of electron density and effective atomic number using phase contrast CT. Phys Med Biol. 2010;55(9):2669–77. doi: 10.1088/0031-9155/55/9/016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Willner M, Bech M, Herzen J, Zanette I, Hahn D, Kenntner J, Mohr J, Rack A, Weitkamp T, Pfeiffer F. Quantitative X-ray phase-contrast computed tomography at 82 keV. Opt Express. 2013;21(4):4155–66. doi: 10.1364/OE.21.004155. [DOI] [PubMed] [Google Scholar]
- 78.Birnbacher L, Willner M, Marschner M, Pfeiffer D, Pfeiffer F, Herzen J. Accurate effective atomic number determination with polychromatic grating-based phase-contrast computed tomography. Opt Express. 2018;26(12):15153–15166. doi: 10.1364/OE.26.015153. [DOI] [PubMed] [Google Scholar]
- 79.Willner M, Viermetz M, Marschner M, Scherer K, Braun C, Fingerle A, Noël P, Rummeny E, Pfeiffer F, Herzen J. Quantitative three-dimensional imaging of lipid, protein, and water contents via X-ray phase-contrast tomography. PLoS One. 2016;11(3):e0151889. doi: 10.1371/journal.pone.0151889. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Braig E, Böhm J, Dierolf M, Jud C, Günther B, Mechlem K, Allner S, Sellerer T, Achterhold K, Gleich B, Noël P, Pfeiffer D, Rummeny E, Herzen J, Pfeiffer F. Direct quantitative material decomposition employing grating-based X-ray phase-contrast CT. Sci Rep. 2018;8(1):16394. doi: 10.1038/s41598-018-34809-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Braig EM, Pfeiffer D, Willner M, Sellerer T, Taphorn K, Petrich C, Scholz J, Petzold L, Birnbacher L, Dierolf M, Pfeiffer F, Herzen J. Single spectrum three-material decomposition with grating-based x-ray phase-contrast CT. Phys Med Biol. 2020;65(18):185011. doi: 10.1088/1361-6560/ab9704. [DOI] [PubMed] [Google Scholar]
- 82.Ji X, Zhang R, Li K, Chen GH. Dual energy differential phase contrast CT (DE-DPC-CT) imaging. IEEE Trans Med Imaging. 2020;39(11):3278–3289. doi: 10.1109/TMI.2020.2990347. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Cloetens P, Guigay JP, De Martino C, Baruchel J, Schlenker M. Fractional Talbot imaging of phase gratings with hard x rays. Opt Lett. 1997;22(14):1059–61. doi: 10.1364/ol.22.001059. [DOI] [PubMed] [Google Scholar]
- 84.Paganin DM. Coherent X-ray optics. Oxford: Oxford University; 2009. [Google Scholar]
- 85.Yashiro W. Hard X-ray imaging microscopy with self-imaging phenomenon. Microscopy (Oxf) 2018;67(6):303–316. doi: 10.1093/jmicro/dfy040. [DOI] [PubMed] [Google Scholar]
- 86.Pfeiffer F, Bunk O, Schulze-Briese C, Diaz A, Weitkamp T, David C, van der Veen JF, Vartanyants I, Robinson IK. Shearing interferometer for quantifying the coherence of hard x-ray beams. Phys Rev Lett. 2005;94(16):164801. doi: 10.1103/PhysRevLett.94.164801. [DOI] [PubMed] [Google Scholar]
- 87.Pfeiffer F, Bech M, Bunk O, Kraft P, Eikenberry EF, Brönnimann Ch, Grünzweig C, David C. Hard-X-ray dark-field imaging using a grating interferometer. Nat Mater. 2008;7(2):134–7. doi: 10.1038/nmat2096. [DOI] [PubMed] [Google Scholar]
- 88.Bevins N, Zambelli J, Li K, Qi Z, Chen GH. Multicontrast x-ray computed tomography imaging using Talbot-Lau interferometry without phase stepping. Med Phys. 2012;39(1):424–8. doi: 10.1118/1.3672163. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Wen HH, Bennett EE, Kopace R, Stein AF, Pai V. Single-shot x-ray differential phase-contrast and diffraction imaging using two-dimensional transmission gratings. Opt Lett. 2010;35(12):1932–4. doi: 10.1364/OL.35.001932. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Wen HH, Miao H, Bennett EE, Adamo NM, Chen L. Flexible retrospective phase stepping in x-ray scatter correction and phase contrast imaging using structured illumination. PLoS One. 2013;8(10):e78276. doi: 10.1371/journal.pone.0078276. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.Pelzer G, Rieger J, Hauke C, Horn F, Michel T, Seifert M, Anton G. Reconstruction method for grating-based x-ray phase-contrast images without knowledge of the grating positions. J Instrum. 2015;10(12):P12017. [Google Scholar]
- 92.Seifert M, Kaeppler S, Hauke C, Horn F, Pelzer G, Rieger J, Michel T, Riess C, Anton G. Optimisation of image reconstruction for phase-contrast x-ray Talbot-Lau imaging with regard to mechanical robustness. Phys Med Biol. 2016;61(17):6441–64. doi: 10.1088/0031-9155/61/17/6441. [DOI] [PubMed] [Google Scholar]
- 93.Kaeppler S, Rieger J, Pelzer G, Horn F, Michel T, Maier A, Anton G, Riess C. Improved reconstruction of phase-stepping data for Talbot-Lau x-ray imaging. J Med Imaging (Bellingham) 2017;4(3):034005. doi: 10.1117/1.JMI.4.3.034005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.De Marco F, Marschner M, Birnbacher L, Noël P, Herzen J, Pfeiffer F. Analysis and correction of bias induced by phase stepping jitter in grating-based X-ray phase-contrast imaging. Opt Express. 2018;26(10):12707–12722. doi: 10.1364/OE.26.012707. [DOI] [PubMed] [Google Scholar]
- 95.Tapfer A, Bech M, Velroyen A, Meiser J, Mohr J, Walter M, Schulz J, Pauwels B, Bruyndonckx P, Liu X, Sasov A, Pfeiffer F. Experimental results from a preclinical X-ray phase-contrast CT scanner. Proc Natl Acad Sci USA. 2012;109(39):15691–6. doi: 10.1073/pnas.1207503109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Epple FM, Potdevin G, Thibault P, Ehn S, Herzen J, Hipp A, Beckmann F, Pfeiffer F. Unwrapping differential x-ray phase-contrast images through phase estimation from multiple energy data. Opt Express. 2013;21(24):29101–8. doi: 10.1364/OE.21.029101. [DOI] [PubMed] [Google Scholar]
- 97.Epple FM, Ehn S, Thibault P, Koehler T, Potdevin G, Herzen J, Pennicard D, Graafsma H, Noël PB, Pfeiffer F. Phase unwrapping in spectral X-ray differential phase-contrast imaging with an energy-resolving photon-counting pixel detector. IEEE Trans Med Imaging. 2015;34(3):816–23. doi: 10.1109/TMI.2014.2349852. [DOI] [PubMed] [Google Scholar]
- 98.Raupach R, Flohr TG. Analytical evaluation of the signal and noise propagation in x-ray differential phase-contrast computed tomography. Phys Med Biol. 2011;56(7):2219–44. doi: 10.1088/0031-9155/56/7/020. [DOI] [PubMed] [Google Scholar]
- 99.Köhler T, Brendel B, Roessl E. Iterative reconstruction for differential phase contrast imaging using spherically symmetric basis functions. Med Phys. 2011;38(8):4542–5. doi: 10.1118/1.3608906. [DOI] [PubMed] [Google Scholar]
- 100.Chen GH, Zambelli J, Li K, Bevins N, Qi Z. Scaling law for noise variance and spatial resolution in differential phase contrast computed tomography. Med Phys. 2011;38(2):584–8. doi: 10.1118/1.3533718. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Weber T, Pelzer G, Rieger J, Ritter A, Anton G. Report of improved performance in Talbot-Lau phase-contrast computed tomography. Med Phys. 2015;42(6):2892–6. doi: 10.1118/1.4921022. [DOI] [PubMed] [Google Scholar]
- 102.Zhu P, Zhang K, Wang Z, Liu Y, Liu X, Wu Z, McDonald SA, Marone F, Stampanoni M. Low-dose, simple, and fast grating-based X-ray phase-contrast imaging. Proc Natl Acad Sci USA. 2010;107(31):13576–81. doi: 10.1073/pnas.1003198107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103.Wu Z, Gao K, Wang Z, Ge X, Chen J, Wang D, Pan Z, Zhang K, Zhu P, Wu Z. A new method to retrieve phase information for equiangular fan beam differential phase contrast computed tomography. Med Phys. 2013;40(3):031911. doi: 10.1118/1.4791672. [DOI] [PubMed] [Google Scholar]
- 104.Marschner M, Birnbacher L, Willner M, Chabior M, Herzen J, Noël PB, Pfeiffer F. Revising the lower statistical limit of x-ray grating-based phase-contrast computed tomography. PLoS One. 2017;12(9):e0184217. doi: 10.1371/journal.pone.0184217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105.Zanette I, Bech M, Rack A, Le Duc G, Tafforeau P, David C, Mohr J, Pfeiffer F, Weitkamp T. Trimodal low-dose X-ray tomography. Proc Natl Acad Sci USA. 2012;109(26):10199–204. doi: 10.1073/pnas.1117861109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Xu Q, Sidky EY, Pan X, Stampanoni M, Modregger P, Anastasio MA. Investigation of discrete imaging models and iterative image reconstruction in differential X-ray phase-contrast tomography. Opt Express. 2012;20(10):10724–49. doi: 10.1364/OE.20.010724. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Nilchian M, Vonesch C, Modregger P, Stampanoni M, Unser M. Fast iterative reconstruction of differential phase contrast X-ray tomograms. Opt Express. 2013;21(5):5511–28. doi: 10.1364/OE.21.005511. [DOI] [PubMed] [Google Scholar]
- 108.Hahn D, Thibault P, Fehringer A, Bech M, Koehler T, Pfeiffer F, Noël PB. Statistical iterative reconstruction algorithm for X-ray phase-contrast CT. Sci Rep. 2015;5:10452. doi: 10.1038/srep10452. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109.Burger K, Koehler T, Chabior M, Allner S, Marschner M, Fehringer A, Willner M, Pfeiffer F, Noël P. Regularized iterative integration combined with non-linear diffusion filtering for phase-contrast x-ray computed tomography. Opt Express. 2014;22(26):32107–18. doi: 10.1364/OE.22.032107. [DOI] [PubMed] [Google Scholar]
- 110.Brendel B, von Teuffenbach M, Noël PB, Pfeiffer F, Koehler T. Penalized maximum likelihood reconstruction for x-ray differential phase-contrast tomography. Med Phys. 2016;43(1):188. doi: 10.1118/1.4938067. [DOI] [PubMed] [Google Scholar]
- 111.Teuffenbach MV, Koehler T, Fehringer A, Viermetz M, Brendel B, Herzen J, Proksa R, Rummeny EJ, Pfeiffer F, Noël PB. Grating-based phase-contrast and dark-field computed tomography: a single-shot method. Sci Rep. 2017;7(1):7476. doi: 10.1038/s41598-017-06729-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Allner S, Koehler T, Fehringer A, Birnbacher L, Willner M, Pfeiffer F, Noël PB. Bilateral filtering using the full noise covariance matrix applied to x-ray phase-contrast computed tomography. Phys Med Biol. 2016;61(10):3867–56. doi: 10.1088/0031-9155/61/10/3867. [DOI] [PubMed] [Google Scholar]
- 113.Thuering T, Zhou T, Lundström U, Burvall A, Rutishauser S, David C, Hertz H, Stampanoni M. X-ray grating interferometry with a liquid-metal-jet source. Appl Phys Lett. 2013;091105:103. [Google Scholar]
- 114.Khimchenko A, Schulz G, Thalmann P, Müller B. Implementation of a double-grating interferometer for phase-contrast computed tomography in a conventional system nanotom(R) m. APL Bioeng. 2018;2(1):016106. doi: 10.1063/1.5022184. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Momose A, Yashiro W, Kuwabara H, Kawabata K. Grating-based X-ray phase imaging using multiline X-ray source. Jap J Appl Phys. 2009;48(7):076512. [Google Scholar]
- 116.Guo J, Xin L, Zhou B, Du Y, Lei Y, Niu H. Development of key devices of grating-based X-ray phase-contrast imaging technology at Shenzhen University. AIP Conf Proc 2012. 2012;1466:61–66. [Google Scholar]
- 117.Engelhardt M, Kottler C, Bunk O, David C, Schroer C, Baumann J, Schuster M, Pfeiffer F. The fractional Talbot effect in differential x-ray phase-contrast imaging for extended and polychromatic x-ray sources. J Microsc. 2008;232(1):145–57. doi: 10.1111/j.1365-2818.2008.02072.x. [DOI] [PubMed] [Google Scholar]
- 118.Chabior M, Donath T, David C, Bunk O, Schuster M, Schroer C, Pfeiffer F. Beam hardening effects in grating-based x-ray phase-contrast imaging. Med Phys. 2011;38(3):1189–95. doi: 10.1118/1.3553408. [DOI] [PubMed] [Google Scholar]
- 119.Hipp A, Willner M, Herzen J, Auweter S, Chabior M, Meiser J, Achterhold K, Mohr J, Pfeiffer F. Energy-resolved visibility analysis of grating interferometers operated at polychromatic X-ray sources. Opt Express. 2014;22(25):30394–409. doi: 10.1364/OE.22.030394. [DOI] [PubMed] [Google Scholar]
- 120.Taylor ML, Smith RL, Franich RD. Robust calculation of effective atomic numbers: the Auto-Zeff software. Med Phys. 2012;39(4):2473–4209. doi: 10.1118/1.3689810. [DOI] [PubMed] [Google Scholar]
- 121.Murty RC. Effective atomic numbers of heterogeneous materials. Nature. 1965;207:398–399. [Google Scholar]
- 122.Spiers W. Effective atomic number and energy absorption in tissues. Br J Radiol. 1946;19:52–63. doi: 10.1259/0007-1285-19-218-52. [DOI] [PubMed] [Google Scholar]
- 123.Hubbell JH. 1969. Photon cross sections, attenuation coefficients, and energy absorption coefficients from 10 keV to 100 GeV. Natl Stand Ref Data Ser NSRDS 29.
- 124.McCullough EC. Photon attenuation in computed tomography. Med Phys. 1975;2(6):307–20. doi: 10.1118/1.594199. [DOI] [PubMed] [Google Scholar]
- 125.Saloman EB, Hubbell JH. 1986. X-ray attenuation coefficients (total cross sections): comparison of the experimental data base with recommended values of Henke and the theoretical values of Scofield for energies between 0.1-100 keV. National Bureau of Standards Report NBSIR. 86-3431.
- 126.Saloman EB, Hubbell JH, Scofield JH. 1988. X-ray attenuation cross sections for energies 100 eV to 100 keV and elements Z = 1 to Z = 92 atomic data and Nucl. Data Tables 38(1).
- 127.Berger MJ, Hubbell JH, Seltzer SM, Chang J, Coursey JS, Sukumar R, Zucker DS, Olsen K. 2010. XCOM: photon cross section database. National Bureau of Standards Report NBSIR. 10.18434/T48G6X.
- 128.Engelhardt M, Baumann J, Schuster M, Kottler C, Pfeiffer F, Bunk O, David C. 2007. High-resolution differential phase contrast imaging using a magnifying projection geometry with a microfocus x-ray source. Appl Phys Lett 90. 10.1063/1.2743928.
- 129.Yashiro W, Takeda Y, Momose A. Efficiency of capturing a phase image using cone-beam x-ray Talbot interferometry. J Opt Soc Am A Opt Image Sci Vis. 2008;25(8):2025–39. doi: 10.1364/josaa.25.002025. [DOI] [PubMed] [Google Scholar]
- 130.Modregger P, Pinzer BR, Thüring T, Rutishauser S, David C, Stampanoni M. Sensitivity of X-ray grating interferometry. Opt Express. 2011;19(19):18324–38. doi: 10.1364/OE.19.018324. [DOI] [PubMed] [Google Scholar]
- 131.Chabior M, Schuster M, Schroer C, Pfeiffer F. Grating-based phase-contrast computed tomography of thick samples. Nucl Instrum Methods Phys Res. 2012;693:138–142. [Google Scholar]
- 132.Donath T, Pfeiffer F, Bunk O, Groot W, Bednarzik M, Grünzweig C, Hempel E, Popescu S, Hoheisel M, David C. Phase-contrast imaging and tomography at 60 keV using a conventional x-ray tube source. Rev Sci Instrum. 2009;80(5):053701. doi: 10.1063/1.3127712. [DOI] [PubMed] [Google Scholar]
- 133.Thuering T, Stampanoni M. Performance and optimization of X-ray grating interferometry. Philos Trans A Math Phys Eng Sci. 2014;372(2010):20130027. doi: 10.1098/rsta.2013.0027. [DOI] [PubMed] [Google Scholar]
- 134.David C, Ziegler E, Nöhammer B. Wet-etched diffractive lenses for hard X-rays. J Synchrotron Radiat. 2001;8(3):1054–5. doi: 10.1107/s0909049501002746. [DOI] [PubMed] [Google Scholar]
- 135.David C, Weitkamp T, Pfeiffer F, Diaz A, Bruder J, Rohbeck T, Groso A, Bunk O, Stampanoni M, Cloetens P. Hard X-ray phase imaging and tomography using a grating interferometer. Spectrochimica Acta Part B. 2007;62:626–630. [Google Scholar]
- 136.Matsumoto M, Takiguchi K, Tanaka M, Hunabiki Y, Takeda H, Momose A, Utsumi Y, Hattori T. Fabrication of diffraction grating for X-ray Talbot interferometer. Microsyst Technol. 2007;13:543–546. [Google Scholar]
- 137.Noda D, Tanaka M, Shimada K, Yashiro W, Momose A, Hattori T. Fabrication of large area diffraction grating using LIGA process. Microsyst Technol. 2008;14:1311–1315. [Google Scholar]
- 138.Mohr J, Grund T, Kunka D, Kenntner J, Leuthold J, Meiser J, Schulz J, Walter M. High aspect ratio gratings for X-ray phase contrast imaging. AIP Conf Proc. 2012;1466:41. [Google Scholar]
- 139.Bech M, Bunk O, David C, Kraft P, Brönnimann C, Eikenberry EF, Pfeiffer F. X-ray imaging with the PILATUS 100k detector. Appl Radiat Isot. 2008;66(4):474–8. doi: 10.1016/j.apradiso.2007.10.003. [DOI] [PubMed] [Google Scholar]
- 140.Gkoumas S, Thuering T, Taboada AG, Arne Jensen A, Michael Rissi M, Broennimann C, Zambon P. Dose-independent near-ideal DQE of a 75 mu m pixel GaAs photon counting spectral detector for breast imaging. Proc SPIE Med Imaging. 2019;2019:10948. [Google Scholar]
- 141.Scholz J, Birnbacher L, Petrich C, Riedel M, Heck L, Gkoumas D, Sellerer T, Achterhold K, Herzen J. Biomedical x-ray imaging with a GaAs photon-counting detector: a comparative study. APL Photonics. 2020;5(10):10.1063/5.0020262. [Google Scholar]
- 142.Pelzer G, Weber T, Anton G, Ballabriga R, Bayer F, Campbell M, Gabor T, Haas W, Horn F, Llopart X, Michel N, Mollenbauer U, Rieger J, Ritter A, Ritter I, Sievers P, Wölfel S, Wong WS, Zang A, Michel T. Grating-based x-ray phase-contrast imaging with a multi energy-channel photon-counting pixel detector. Opt Express. 2013;21(22):25677–84. doi: 10.1364/OE.21.025677. [DOI] [PubMed] [Google Scholar]
- 143.Mechlem K, Sellerer T, Viermetz M, Herzen J, Pfeiffer F. A theoretical framework for comparing noise characteristics of spectral, differential phase-contrast and spectral differential phase-contrast x-ray imaging. Phys Med Biol. 2020;65(6):065010. doi: 10.1088/1361-6560/ab7106. [DOI] [PubMed] [Google Scholar]
- 144.Zanette I, Weitkamp T, Lang S, Langer M, Mohr J, David C, Baruchel J. Quantitative phase and absorption tomography with an X-ray grating interferometer and synchrotron radiation. Phys Stat Solidi A. 2011;208(11):2526–2532. [Google Scholar]
- 145.Zanette I, Lang S, Rack A, Dominietto M, Langer M, Pfeiffer F, Weitkamp T, Müller B. Holotomography versus X-ray grating interferometry: a comparative study. Appl Phys Lett. 2013;103(24):244105. [Google Scholar]
- 146.Ruiz-Yaniz M, Zanette I, Sarapata A, Birnbacher L, Marschner M, Chabior M, Olbinado M, Pfeiffer F, Rack A. Hard X-ray phase-contrast tomography of non-homogeneous specimens: grating interferometry versus propagation-based imaging. J Synchrotron Radiat. 2016;23(Pt 5):1202–9. doi: 10.1107/S1600577516009164. [DOI] [PubMed] [Google Scholar]
- 147.Zdora MC, Vila-Comamala J, Schulz G, Khimchenko A, Hipp A, Cook AC, Dilg D, David C, Grünzweig C, Rau C, Thibault P, Zanette I. X-ray phase microtomography with a single grating for high-throughput investigations of biological tissue. Biomed Opt Express. 2017;8(2):1257–1270. doi: 10.1364/BOE.8.001257. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 148.Kaneko Y, Shinohara G, Hoshino M, Morishita H, Morita K, Oshima Y, Takahashi M, Yagi N, Okita Y, Tsukube T. Intact imaging of human heart structure using X-ray phase-contrast tomography. Pediatr Cardiol. 2017;38(2):390–393. doi: 10.1007/s00246-016-1527-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 149.McDonald SA, Marone F, Hintermüller C, Mikuljan G, David C, Pfeiffer F, Stampanoni M. Advanced phase-contrast imaging using a grating interferometer. J Synchrotron Radiat. 2009;16(Pt 4):562–72. doi: 10.1107/S0909049509017920. [DOI] [PubMed] [Google Scholar]
- 150.Sarapata A, Willner M, Walter M, Duttenhofer T, Kaiser K, Meyer P, Braun C, Fingerle A, Noël PB, Pfeiffer F, Herzen J. Quantitative imaging using high-energy X-ray phase-contrast CT with a 70 kVp polychromatic X-ray spectrum. Opt Express. 2015;23(1):523–35. doi: 10.1364/OE.23.000523. [DOI] [PubMed] [Google Scholar]
- 151.Thüring T, Abis M, Wang Z, David C, Stampanoni M. X-ray phase-contrast imaging at 100 keV on a conventional source. Sci Rep. 2014;4:5198. doi: 10.1038/srep05198. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 152.Viermetz M, Birnbacher L, Willner M, Achterhold K, Pfeiffer F, Herzen J. High resolution laboratory grating-based X-ray phase-contrast CT. Sci Rep. 2018;8(1):15884. doi: 10.1038/s41598-018-33997-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 153.Birnbacher L, Viermetz M, Noichl W, Allner S, Fehringer A, Marschner M, von Teuffenbach M, Willner M, Achterhold K, Noël PB, Koehler T, Herzen J, Pfeiffer F. Tilted grating phase-contrast computed tomography using statistical iterative reconstruction. Sci Rep. 2018;8(1):6608. doi: 10.1038/s41598-018-25075-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 154.Winklhofer S, Peter S, Tischler V, Morsbach F, von Werdt M, Berens S, Modregger P, Buser L, Moch H, Stampanoni M, Thali M, Alkadhi H, Stolzmann P. Diagnostic accuracy of quantitative and qualitative phase-contrast imaging for the ex Vivo characterization of human coronary atherosclerotic plaques. Radiology. 2015;277(1):64–72. doi: 10.1148/radiol.2015141614. [DOI] [PubMed] [Google Scholar]
- 155.Bonanno G, Coppo S, Modregger P, Pellegrin M, Stuber A, Stampanoni M, Mazzolai L, Stuber M, van Heeswijk RB. Ultra-high-resolution 3D imaging of atherosclerosis in mice with synchrotron differential phase contrast: a proof of concept study. Sci Rep. 2015;5:11980. doi: 10.1038/srep11980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 156.Herzen J, Karampinos DC, Foehr P, Birnbacher L, Viermetz M, Burgkart R, Baum T, Lohoefer F, Wildgruber M, Schilling F, Willner M, Marschner M, Noël PB, Rummeny EJ, Pfeiffer F, Jungmann PM. 2019. 3D grating-based X-ray phase-contrast computed tomography for high-resolution quantitative assessment of cartilage: an experimental feasibility study with 3T MRI, 7T MRI and biomechanical correlation, Vol. 14. 10.1371/journal.pone.0212106. [DOI] [PMC free article] [PubMed]
- 157.Velroyen A, Bech M, Zanette I, Schwarz J, Rack A, Tympner C, Herrler T, Staab-Weijnitz C, Braunagel M, Reiser M, Bamberg F, Pfeiffer F, Notohamiprodjo M. 2014. X-ray phase-contrast tomography of renal ischemia-reperfusion damage, Vol. 9. 10.1371/journal.pone.0109562. [DOI] [PMC free article] [PubMed]
- 158.Günther B, Gradl R, Jud C, Eggl E, Huang J, Kulpe S, Achterhold K, Gleich B, Dierolf M, Pfeiffer F. The versatile X-ray beamline of the Munich Compact Light Source: design, instrumentation and application. J Synchrotron Radiat. 2020;27(5):1395–1414. doi: 10.1107/S1600577520008309. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 159.Eggl E, Dierolf M, Achterhold K, Jud C, Günther B, Braig E, Gleich B, Pfeiffer F. The Munich Compact Light Source: initial performance measures. J Synchrotron Radiat. 2016;23(Pt 5):1137–42. doi: 10.1107/S160057751600967X. [DOI] [PubMed] [Google Scholar]
- 160.Eggl E, Schleede S, Bech M, Achterhold K, Loewen R, Ruth RD, Pfeiffer F. X-ray phase-contrast tomography with a compact laser-driven synchrotron source. Proc Natl Acad Sci USA. 2015;112(18):5567–72. doi: 10.1073/pnas.1500938112. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 161.Birnbacher L, Maurer S, Scheidt K, Herzen J, Pfeiffer F, Fromme T. Electron density of adipose tissues determined by phase-contrast computed tomography provides a measure for mitochondrial density and fat content. Front Physiol. 2018;9:707. doi: 10.3389/fphys.2018.00707. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 162.Gradl R, Zanette I, Ruiz-Yaniz M, Dierolf M, Rack A, Zaslansky P, Pfeiffer F. 2016. Mass density measurement of mineralized tissue with grating-based X-ray phase tomography, Vol. 11. 10.1371/journal.pone.0167797. [DOI] [PMC free article] [PubMed]
- 163.Cloetens P, Ludwig W, Baruchel J, Van Dyck D, Van Landuyt J, Guigay JP, Schlenker M. Holotomography: quantitative phase tomography with micrometer resolution using hard synchrotron radiation x rays. Appl Phys Lett. 1999;75(19):2912–2914. [Google Scholar]
- 164.Bronnikov AV. Theory of quantitative phase-contrast computed tomography. J Opt Soc Am A Opt Image Sci Vis. 2002;19(3):472–80. doi: 10.1364/josaa.19.000472. [DOI] [PubMed] [Google Scholar]
- 165.Paganin D, Mayo SC, Gureyev TE, Miller PR, Wilkins S. Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object. J Microsc. 2002;206(Pt 1):33–40. doi: 10.1046/j.1365-2818.2002.01010.x. [DOI] [PubMed] [Google Scholar]
- 166.Frohn J, Pinkert-Leetsch D, Missbach-Güntner J, Reichardt M, Osterhoff M, Alves F, Salditt T. 3D virtual histology of human pancreatic tissue by multiscale phase-contrast X-ray tomography. J Synchrotron Radiat. 2020;27(Pt 6):1707–1719. doi: 10.1107/S1600577520011327. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 167.Gao D, Pogany A, Stevenson AW, Wilkins S. Phase-contrast radiography. Radiographics. 1998;18(5):1257–67. doi: 10.1148/radiographics.18.5.9747618. [DOI] [PubMed] [Google Scholar]
- 168.Bidola PM, Zanette I, Achterhold K, Holzner C, Pfeiffer F. Optimization of propagation-based phase-contrast imaging at a laboratory setup. Opt Express. 2015;23(23):30000–13. doi: 10.1364/OE.23.030000. [DOI] [PubMed] [Google Scholar]
- 169.Longo R, Arfelli F, Bellazzini R, Bottigli U, Brez A, Brun F, Brunetti A, Delogu P, Di Lillo F, Dreossi D, Fanti V, Fedon C, Golosio B, Lanconelli N, Mettivier G, Minuti M, Oliva P, Pinchera M, Rigon L, Russo P, Sarno A, Spandre G, Tromba G, Zanconati F. Towards breast tomography with synchrotron radiation at Elettra: first images. Phys Med Biol. 2016;61(4):1634–49. doi: 10.1088/0031-9155/61/4/1634. [DOI] [PubMed] [Google Scholar]
- 170.Brombal L, Golosio B, Arfelli F, Bonazza D, Contillo A, Delogu P, Donato S, Mettivier G, Oliva P, Rigon L, Taibi A, Tromba G, Zanconati F, Longo R. Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction. J Med Imaging (Bellingham) 2019;6(3):031402. doi: 10.1117/1.JMI.6.3.031402. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 171.Zhao Y, Brun E, Coan P, Huang Z, Sztrókay A, Diemoz PC, Liebhardt S, Mittone A, Gasilov S, Miao J, Bravin A. High-resolution, low-dose phase contrast X-ray tomography for 3D diagnosis of human breast cancers. Proc Natl Acad Sci USA. 2012;109(45):18290–4. doi: 10.1073/pnas.1204460109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 172.Sztrókay A, Diemoz PC, Schlossbauer T, Brun E, Bamberg F, Mayr D, Reiser MF, Bravin A, Coan P. High-resolution breast tomography at high energy: a feasibility study of phase contrast imaging on a whole breast. Phys Med Biol. 2012;57(10):2931–42. doi: 10.1088/0031-9155/57/10/2931. [DOI] [PubMed] [Google Scholar]
- 173.Mittone A, Gasilov S, Brun E, Bravin A, Coan P. A single-image method for x-ray refractive index CT. Phys Med Biol. 2015;60(9):3433–40. doi: 10.1088/0031-9155/60/9/3433. [DOI] [PubMed] [Google Scholar]
- 174.Grandl S, Sztrókay-Gaul A, Mittone A, Gasilov S, Brun E, Bravin A, Mayr D, Auweter SD, Hellerhoff K, Reiser M, Coan P. Detection of post-therapeutic effects in breast carcinoma using hard X-ray index of refraction computed tomography - a feasibility study. PLoS One. 2016;11(6):e0158306. doi: 10.1371/journal.pone.0158306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 175.Hagen CK, Diemoz PC, Endrizzi M, Rigon L, Dreossi D, Arfelli F, Lopez FC, Longo R, Olivo A. Theory and preliminary experimental verification of quantitative edge illumination x-ray phase contrast tomography. Opt Express. 2014;22(7):7989–8000. doi: 10.1364/OE.22.007989. [DOI] [PubMed] [Google Scholar]
- 176.Olivo A, Gkoumas S, Endrizzi M, Hagen CK, Szafraniec MB, Diemoz PC, Munro PR, Ignatyev K, Johnson B, Horrocks JA, Vinnicombe SJ, Jones JL, Speller R. Low-dose phase contrast mammography with conventional x-ray sources. Med Phys. 2013;40(9):090701. doi: 10.1118/1.4817480. [DOI] [PubMed] [Google Scholar]
- 177.Hagen CK, Munro PR, Endrizzi M, Diemoz PC, Olivo A. Low-dose phase contrast tomography with conventional x-ray sources. Med Phys. 2014;41(7):070701. doi: 10.1118/1.4884297. [DOI] [PubMed] [Google Scholar]
- 178.Kashyap Y, Wang H, Sawhney K. Experimental comparison between speckle and grating-based imaging technique using synchrotron radiation X-rays. Opt Express. 2016;24(16):18664–73. doi: 10.1364/OE.24.018664. [DOI] [PubMed] [Google Scholar]
- 179.Zdora MC, Zanette I, Walker T, Phillips NW, Smith R, Deyhle H, Ahmed S, Thibault P. X-ray phase imaging with the unified modulated pattern analysis of near-field speckles at a laboratory source. Appl Opt. 2020;59(8):2270–2275. doi: 10.1364/AO.384531. [DOI] [PubMed] [Google Scholar]
- 180.Zhou T, Zanette I, Zdora MC, Lundström U, Larsson DH, Hertz HM, Pfeiffer F, Burvall A. Speckle-based x-ray phase-contrast imaging with a laboratory source and the scanning technique. Opt Lett. 2015;40(12):2822–5. doi: 10.1364/OL.40.002822. [DOI] [PubMed] [Google Scholar]
- 181.Zdora CM, Thibault P, Kuo W, Fernandez V, Deyhle H, Vila-Comamala J, Olbinado M, Rack A, Lackie P, Katsamenis O, Lawson M, Kurtcuoglu V, Rau C, Pfeiffer F, Zanette I. X-ray phase tomography with near-field speckles for three-dimensional virtual histology. Optica. 2020;7:1221–1227. [Google Scholar]
- 182.Wang H. Kashyap Y, Sawhney K. Quantitative X-ray dark-field and phase tomography using single directional speckle scanning technique. Appl Phys Lett. 2016;124102:108. [Google Scholar]
- 183.Bérujon S, Ziegler E. X-ray multimodal tomography using speckle-vector tracking. Phys Rev Appl. 2016;5(4):044014. [Google Scholar]
- 184.Madabhushi A, Lee G. Image analysis and machine learning in digital pathology: challenges and opportunities. Med Image Anal. 2016;33:170–175. doi: 10.1016/j.media.2016.06.037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 185.Jin YW, Jia S, Ashraf AB, Hu P. Integrative data augmentation with u-net segmentation masks improves detection of lymph node metastases in breast cancer patients. Cancers (Basel) 2020;12(10):2934. doi: 10.3390/cancers12102934. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 186.Muñoz-Aguirre M, Ntasis VF, Rojas S, Guigó R. 2020. PyHIST: a histological image segmentation tool, Vol. 16. 10.1371/journal.pcbi.1008349. [DOI] [PMC free article] [PubMed]
- 187.Taba ST, Gureyev TE, Alakhras M, Lewis S, Lockie D, Brennan PC. X-Ray phase-contrast technology in breast imaging: principles, options, and clinical application. AJR Am J Roentgenol. 2018;211(1):133–145. doi: 10.2214/AJR.17.19179. [DOI] [PubMed] [Google Scholar]
- 188.Heck L, Herzen J. Recent advances in X-ray imaging of breast tissue: From two- to three-dimensional imaging. Phys Med. 2020;79:69–79. doi: 10.1016/j.ejmp.2020.10.025. [DOI] [PubMed] [Google Scholar]
- 189.Kagias M, Cartier S, Wang Z, Bergamaschi A, Dinapoli R, Mozzanica A, Schmitt B, Stampanoni M. Singleshot x-ray phase contrast imaging using a direct conversion microstrip detector with single photon sensitivity. Appl Phys Lett. 2016;234102:108. [Google Scholar]
- 190.Wen HH, Gomella AA, Patel A, Lynch SK, Morgan NY, Anderson SA, Bennett EE, Xiao X, Liu C, Wolfe DE. Subnanoradian X-ray phase-contrast imaging using a far-field interferometer of nanometric phase gratings. Nat Commun. 2013;4:2659. doi: 10.1038/ncomms3659. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 191.Miao H, Gomella AA, Chedid N, Chen L, Wen HH. Fabrication of 200 nm period hard X-ray phase gratings. Nano Lett. 2014;14(6):3453–8. doi: 10.1021/nl5009713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 192.Miao H, Gomella AA, Harmon KJ, Bennett EE, Chedid N, Znati S, Panna A, Foster BA, Bhandarkar P, Wen HH. Enhancing tabletop X-ray phase contrast imaging with nano-fabrication. Sci Rep. 2015;5:13581. doi: 10.1038/srep13581. [DOI] [PMC free article] [PubMed] [Google Scholar]