Invertibility of multi‐energy X‐ray transform

Yijun Ding 1,, Eric W Clarkson 2, Amit Ashok 3
PMCID: PMC8568641  NIHMSID: NIHMS1733072  PMID: 34390587

Abstract

Purpose

The goal is to provide a sufficient condition for the invertibility of a multi‐energy (ME) X‐ray transform. The energy‐dependent X‐ray attenuation profiles can be represented by a set of coefficients using the Alvarez–Macovski (AM) method. An ME X‐ray transform is a mapping from N AM coefficients to N noise‐free energy‐weighted measurements, where N2.

Methods

We apply a general invertibility theorem to prove the equivalence of global and local invertibility for an ME X‐ray transform. We explore the global invertibility through testing whether the Jacobian of the mapping J(A) has zero values over the support of the mapping. The Jacobian of an arbitrary ME X‐ray transform is an integration over all spectral measurements. A sufficient condition for J(A)0 for all A is that the integrand of J(A) is 0 (or 0) everywhere. Note that the trivial case of the integrand equals 0 everywhere is ignored. Using symmetry, we simplified the integrand of the Jacobian to three factors that are determined by the total attenuation, the basis functions, and the energy‐weighting functions, respectively. The factor related to the total attenuation is always positive; hence, the invertibility of the X‐ray transform can be determined by testing the signs of the other two factors. Furthermore, we use the Cramér–Rao lower bound (CRLB) to characterize the noise‐induced estimation uncertainty and provide a maximum‐likelihood (ML) estimator.

Results

The factor related to the basis functions is always negative when the photoelectric/Compton/Rayleigh basis functions are used and K‐edge materials are not considered. The sign of the energy‐weighting factor depends on the system source spectra and the detector response functions. For four special types of X‐ray detectors, the sign of this factor stays the same over the integration range. Therefore, when these four types of detectors are used for imaging non‐K‐edge materials, the ME X‐ray transform is globally invertible. The same framework can be used to study an arbitrary ME X‐ray imaging system, for example, when K‐edge materials are present. Furthermore, the ML estimator we presented is an unbiased, efficient estimator and can be used for a wide range of scenes.

Conclusions

We have provided a framework to study the invertibility of an arbitrary ME X‐ray transform and proved the global invertibility for four types of systems.

Keywords: invertibility, multi‐energy X‐ray imaging, spectral X‐ray imaging, X‐ray

1. INTRODUCTION

Multi‐energy (ME) X‐ray imaging, also referred to as spectral or energy‐selective X‐ray imaging, has long been used to image the chemical composition of the object being scanned. 1 , 2 , 3 , 4 , 5 , 6 In X‐ray imaging, the chemical composition of a material is characterized by the energy dependence of the X‐ray attenuation profile. As an X‐ray attenuation profile can be represented as a linear combination of basis functions with known energy dependences, it can be summarized by a few energy‐independent coefficients as in the Alvarez–Macovski method. 1 We refer to these coefficients as AM coefficients. Imaging AM coefficients requires multiple energy‐weighted measurements, for example, energy integration with varying source tube voltages or photocounting with multiple energy bins. We refer to the mapping from the AM coefficients to the energy‐weighted measurements an X‐ray transform. The question whether an X‐ray transform is invertible has only been explored recently for dual‐energy (DE) 7 , 8 and ME measurements. 9 The purpose of this work is to provide a sufficient condition for the invertibility of a general ME X‐ray transform from a different perspective.

With the recent developments in detectors, ME X‐ray imaging is becoming more tangible. DE X‐ray imaging recovers two AM coefficients 1 that represent contributions from photoelectric absorption and Compton scattering to the linear attenuation profile, respectively. The contribution from Rayleigh scattering has been considered negligible or assumed to be captured by the other two AM coefficients in DE X‐ray imaging. 10 , 11 , 12 However, it is difficult to predict the effect caused by ignoring the Rayleigh scattering term due to the nonlinear nature of the X‐ray transform, especially for security‐ and industrial‐screening applications where the materials of interest are not necessarily low‐Z materials. With ME detectors, the AM coefficient corresponding to the Rayleigh scattering can be recovered. Furthermore, ME X‐ray imaging systems can image materials containing K‐edges in the spectral range used for imaging. 5

With broad‐spectrum X‐ray sources, measurements of many X‐ray systems are naturally energy‐weighted. 13 ME measurements can be acquired with varying source settings 14 , 15 or with detectors with varying energy responses, such as sandwich detectors, 16 counting and integrating X‐ray (CIX) detectors, 17 and multibin photon‐counting (PC) detectors. 18 More specifically, the recent advancement in PC detectors with pulse‐height analysis, which output signals in multiple energy levels, provides a paradigm shift in X‐ray detector technology and is enabling many new applications. 19 , 20

The invertibility of a transform is a fundamental question in inverse problems. The invertibility problem considers noise‐free measurements and determines whether a unique solution exists. A system of N linear equations of N unknowns has a unique solution (as long as the forward matrix is invertible); this is not necessarily true for nonlinear transforms. Levine et al 7 demonstrated a case of DE X‐ray imaging with nonunique solutions. Alvarez et al 8 have applied a two‐dimensional global inverse theorem 21 to DE X‐ray transforms. Bal et al 9 provided an invertibility criteria for an ME X‐ray transform by placing strong orientation constraints on the Jacobian matrices and demonstrated the equivalence of global and local invertibility for some examples through numerical experiments. We apply a global inverse function theorem for an N‐dimensional map and prove that, for an ME X‐ray transform, local invertibility is equivalent to global invertibility. Our global invertibility criteria is local invertibility, which is a weaker sufficient condition than the criteria provided by Bal et al. 9 This is proved through topological properties of the definition region and inequalities of the X‐ray transform. Furthermore, we provide a sufficient condition for the global invertibility by taking advantage of the symmetries in the expression of the Jacobian. With its simple expression, this condition can be applied to the design of ME X‐ray imaging systems and detectors.

In this paper, we provide a framework to study the invertibility of an arbitrary ME X‐ray transform and prove the invertibility for four special cases of energy‐weighted detectors. Furthermore, we consider Poisson noise in the measurement data and present the Cramér–Rao lower bound (CRLB) on the estimation of AM coefficients. Lastly, we provide a fast maximum‐likelihood (ML) algorithm for coefficients estimation and demonstrate its application in an X‐ray reconstruction problem.

2. FORWARD PROBLEM: ME X‐RAY TRANSFORM

In the energy range 20‐200 keV, which is commonly used for X‐ray transmission imaging, the interaction between X‐ray photons and the medium can be categorized into the following three processes: photoelectric absorption, Compton (incoherent) scattering, and Rayleigh (coherent ) scattering. Correspondingly, the X‐ray linear attenuation coefficient profiles can be represented accurately by a summation of N terms as:

μ(E)=iNaifi(E)=a·f(E), (1)

where each component of f is a function of energy E, the coefficients a are determined by the material composition, and the N terms include photo electric, Compton scattering, Rayleigh scattering, and K‐edges. Here “photo electric” refers to the smooth energy dependence of the photo electric effect and “K‐edges” refers to the discontinuities in the energy dependence of the photo electric effect just above the binding energy of the K‐shell electrons. We use this set of fi(E) functions as basis functions and the coefficients a as the AM coefficients.

For materials that do not contain K‐edges in the energy range of interest, the number of basis functions needed is N=3. Approximated expressions of photo electric and Rayleigh scattering term have been provided in Williamson et al 22 by fitting to DLC‐146 cross‐section data 23 and the Klein–Nishina function 1 describes the Compton scattering term:

f1(E)=c1E3.088,f2(E)=c21+αα22(1+α)1+2α1αln(1+2α)+12αln(1+2α)1+3α(1+2α)2,f3(E)=c3E1.737, (2)

where α=E/(510.975keV), the subscripts 1‐3 refer to the photoelectric effect, the Compton scattering and the Rayleigh scattering, respectively, and ci are normalization factors so that fi(E)2=1. The normalized basis functions are presented in Figure 1 (left). The usefulness of these functions in representing attenuation coefficient profiles is well known. We generated attenuation profiles for 128 materials based on the NIST XCOM data. 24 As an example, the fitted attenuation profile and the XCOM data for water are presented in Figure 1 (right).

FIGURE 1.

FIGURE 1

Shape of fi(E) for i=1,2,3 (left) and fitted attenuation profile of water (right)

In a tomographic imaging or measurement system, the total attenuation τ(E) is the line integral of the X‐ray attenuation coefficient μ(E) along the ray path

τ(E)=dlμ(E)=i=1NAifi(E)=A·f(E), (3)

where

Ai=dlai(R) (4)

is a sinogram of the ith AM coefficient. For a parallel‐beam system, Ai(θ,ρ) is the Radon transform of ai(R), where θ is the rotation angle of the ray path and ρ is the position along the detector plane.

The object ai(R) can be reconstructed from the sinograms Ai(θ,ρ), and the line integrals Ai(θ,ρ) can be estimated from ME measurements of the corresponding ray path. Consider an ME X‐ray imaging system producing M energy‐weighted measurements with a source photon budget I0 (total number of photons emitted by the source across the energy range of interest). To describe the mth energy‐weighted measurement, where m=1,,M, denote Dm(E) as the detector response and Sm(E) as the normalized source spectrum of the mth measurement. For a given ray path, the mean signal of the mth measurement can be described by

Im=I00dEDm(E)Sm(E)expA·f(E),=I00dEpm(E)expA·f(E), (5)

where pm(E)=Dm(E)Sm(E) is the combined energy‐weighting function. This equation can be used to describe many energy‐weighted measurements, such as a PC binning detector and an energy‐integrating detector. In the most general case, the source spectra may vary across measurements and the combined weighting functions are arbitrary and can take on any real nonnegative values at each energy. The basis functions f(E) can contain components describing K‐edges as well. Therefore, Equation (5) describes a general ME X‐ray transform. In the following sections, we study the invertibility of the mapping defined by this equation in the domain Ai0 for i=1,2,,N.

A special case of an ME detector is a CIX detector that counts the number of photons and integrates both the energy and the momentum of the photons (PC/EI/MI), providing measurements with detector response D1(E)=1, D2(E)E, and D3(E)E, as shown in Figure 2(a). As a CIX PC/EI detector has been developed, 17 it is reasonable to assume that it is feasible to build a CIX PC/EI/MI detector. A second special case is a binning detector where the weighting functions are arbitrary and nonoverlapping as shown in Figure 2(b). Another special case is an ideal PC detector as illustrated in Figure 2(c), where the detector response of each bin can be considered as rect functions and there may be overlaps between different bins. Binning detectors in real life tend to have nonoverlapping bins. Here, for the comprehensiveness, we include detectors with overlapping bins. Another special case considers a slightly overlapping three‐bin detector, where the overlap is introduced by the finite energy resolution of the detector. The detector response functions of such a detector are plotted in Figure 2(d).

FIGURE 2.

FIGURE 2

Energy weighting functions for four special cases: (a) CIX‐PC/EI/MI, (b) nonoverlapping bins with arbitrary response, (c) three rect bins with ideal energy resolution, and (d) a slightly overlapping three‐bin detector, where the overlap is introduced by the finite energy resolution of the detector

3. INVERTIBILITY

We explore the invertibility of the mapping from the AM coefficients A to the noise‐free ME measurement data I. Suppose that we have M=N measurements. The coefficients A and the mean photon count I are both subsets of N‐dimensional Euclidean space RN. We define the ME X‐ray transform from A to I as X:M1M2, where the domain of the mapping is A in M1 and the range of the mapping is I in M2.

The Hadamard's global inverse function theorem: 25 Let M1 and M2 be smooth, connected N‐dimensional manifolds and let X:M1M2 be a C1 function. If (1) X is proper, (2) the Jacobian of X vanishes nowhere, and (3) M2 is simply connected, then X is a homeomorphism. A homeomorphism is one‐to‐one and onto, which implies global invertibility, whereas nonvanishing Jacobian implies local invertibility.

In the following sections, we will use the Hadamard's global inverse function theorem to prove the equivalence of global and local invertibility for an ME X‐ray transform. We will first construct a simply‐connected range M2, then prove that the mapping X:M1M2 is proper through inequality relations. Lastly, we will provide a simplified expression for the Jacobian determinant and a sufficient condition for the Jacobian to vanish nowhere.

3.1. Simply connected

We briefly summarize the property of the mapping X. The first‐order derivative of the X‐ray transform can be expressed as follows:

ImAi=I00dEpm(E)fi(E)expA·f(E). (6)

The first‐order derivative exists and is continuous; therefore, the mapping X is a C1 mapping. The values X(0), which represents the mean signals of an air scan, are finite and equal to the maximum (mean) count values. We further define a normalization factor dm=0dEpm(E). With this definition, the maximum mean count measured by the mth detector is I0dm. As the magnitude of A approaches infinity, the counts approach zeros. We will use these properties and the assumption that the Jacobian is nonvanishing in RN to construct a simply connected range M2 with the corresponding domain M1 connected. Furthermore, we will justify that the interior of M1 and M2 are both smooth N‐dimensional manifolds.

The ME X‐ray transform, as defined in Equation (5), has physical meaning when A is in the positive subspace of RN, denoted as PN, where Ai0 for all i. However, the transform is mathematically valid over the domain RN. To construct a simply connected M2, we will expand the domain of the mapping from PN.

Let U0=PN and V0 be the image of U0 under the mapping X. In V0, the mean photon count I is bounded by 0<ImI0dm. Furthermore, as U0 is path connected and X is a continuous mapping, V0 is path connected 26 (p. 150). From every point Ii in V0, draw a straight line to the maximum‐count point I0d and define this line with end points as Vi. Every point in Vi is bounded by IimImI0dm. Define M2 as the union of all Vi. As V0 and Vi are all path connected, M2 is path connected. The space M2 is simply connected if every closed curve in M2 can be contracted to a point. 25 Define a closed curve ϕ(s):[0,1]M2. We can contract ϕ(s) to the maximum‐count point I0d through the following continuous function H: [0,1]×[0,1]M2,

H(s,t)=tϕ(s)+(1t)I0d. (7)

As M2 is path connected, the closed curve ϕ(s) can be contracted to any points in M2 26 (p. 332). Therefore, M2 is simply connected.

We assume that the Jacobian vanishes nowhere in RN. In other words, the mapping X is a local homeomorphism, which means that every point of ARN has a neighborhood that is homeomorphic to an open subset in the range. For every straight line Vi(I), the corresponding preimage ui(A) in the domain RN can be constructed by successive local inverses X1(I). The maximum‐count point I0d corresponds to only one point in U0, and this point is the origin of the coefficient space. Therefore, the origin is one end point of all preimages. The point Ii may have multiple local inverses, which we can index with subscript j. The jth local inverse introduces an inverse curve uij, where the jth local inverse is the second end point of the corresponding preimage uij. Each uij is connected and connected to U0. We define the union of all uij(A) as Ui. Furthermore, we define M1 as the union of all Ui. M1 is connected and a superset of PN. The expansion of the range and domain for N=2 is illustrated in Figure 3.

FIGURE 3.

FIGURE 3

Constructing a simply connected range M2 by expanding the domain of the map X: (a) The initial domain U0 and range V0. (b) Vi, defined by points Ii and I0d, and its corresponding preimages uij. Ui is defined as the union of all uij. (c) The expanded domain M1 and range M2. (d) Every closed curve ϕ(s) (blue loop) can be shrunk down to the point I0d through the function H(s,t). Color varying from blue to green represents t from 1 to 0. The black‐dashed line is the trajectory of a point s in the loop ϕ as t decreases from 1 to 0

The interior points of M1 and M2 (excluding the boundaries) are both smooth N‐dimensional manifolds, because they are open subsets of RN 27 (p. 19). We will limit our proof to the interior of M1 and M2 and discuss the boundary points in Section 6.

3.2. Proper

We derive the bounds on the coefficients A for given measurement data I. Using Jensen's inequality, we have

lnImdmI0>0dEpmEdmA·fE, (8)

where dm has been defined previously and dmI0 is the maximum count in the mth measurement. These inequalities can be written as

A·nm>ln[(dmI0)/Im], (9)

where

nm=0dEpmEfE/dm. (10)

The vector nm has all nonnegative components. Furthermore, the mean photon count Im in M2, is always less than or equal to the maximum count, dmI0. Therefore, the right‐hand side of Equation (9) is always larger than or equal to 0. Each of the inequalities in Equation (9) forces the vector A to be on the side that is opposite to the origin of the hyperplane defined by to A·nm=ln[Im/(dmI0)], as shown in Figure 4(a) for the case of N=2.

FIGURE 4.

FIGURE 4

(a) For given noise‐free measurement data I, the vector A is bounded in the area indicated in gray. Each energy‐weighted measurement generates a pair of red and blue hyperplanes, which bound the vector A. (b) Illustration of the sign‐switched mirror symmetry in function det[F(E1,E2)] along line E1=E2

We define the support of the weighting functions pm(E) as Ωm. Using the Schwarz inequality, we have

Im2I02ΩmdEpm2EΩmdEexp2A·fE, (11)

with equality if and only if pm(E)exp[A·f(E)]. In many occasions, the equality condition is not attainable. For example, when the three basis functions given in Equation (2) are used, exp[A·f(E)] is not proportional to the pm(E) of the detectors illustrated in Figure 2(b)–(d). If we define

γm=Im2I02ΩmdEpm2E1, (12)

then we have

γmΩmdEexp2A·fE. (13)

Assume that the length |Ωm| of each support set is finite. Replacing the integrand with its maximum possible value gives

γmexp2minEΩmA·fEΩm. (14)

Therefore, we have another set of inequalities

minEΩmA·fE12lnΩmγm. (15)

Now we may choose an energy Em such that

A·fEmlnΩmγm. (16)

The right‐hand side satisfies ln(|Ωm|/γm)ln(dmI0/Im)0 through the Schwarz inequality as follows:

dm2=ΩmdEpmE2|Ωm|ΩmdEpm2E. (17)

Each of the inequalities in Equation (16) forces the vector A to be on the same side of the corresponding hyperplane as the origin.

Therefore, for given mean photon count I, where dmI0Im>0, the inequalities defined by Equations (9) and (16) force A to be in a bounded set defined by the first set of hyperplanes and the second set of hyperplanes. A typical picture of this scenario for N=2 is shown in Figure 4(a). Note that for a physical measurement, the corresponding coefficients A are further bounded by the coordinate planes. Here we focus on demonstrating that A is bounded for a given I even without the positivity constraints on A.

Now we show that the mapping X:M1M2 is a proper mapping. If we have a compact set C in the data space M2, where all of the data vectors are located, then there are maximum and minimum values for each Im over all I in C. The maximum value for Im determines the hyperplane A·nm=ln[(dmI0)/Im] that is close to the origin. The minimum value for Im determines the hyperplane A·f(Em)=ln(|Ωm|/γm) that is furthest away from the origin. Therefore, the set of A that are mapped into C is contained in a region bounded by these two sets of hyperplanes. This bounded region together with its boundary form a closed and bounded set in RN, hence a compact set. As the map X(A) is continuous, the set of A that are mapped into the closed set C is closed. The set of A that are mapped into C is a closed subset of a compact set. This set is therefore also compact. As a result, the mapping X(A) is proper.

3.3. Jacobian

The Jacobian of the mapping is J(A)=|det(AI)|, where |·| represents the absolute value and det(·) is the determinant of a matrix. The matrix inside the determinant is

AI=I1A1I1A2I1ANI2A1I2A2I2ANIMA1IMA2IMAN. (18)

As M=N, AI is a square matrix. Defining the set Ω=Ω1×Ω2×Ωm, the determinant of a square matrix can be expressed in Leibniz formula as

det(AI)=σsgn(σ)m=1MImAσm,=(I0)Mσsgn(σ)m=1MΩmdEmpm(Em)×fσm(Em)expA·f(Em)=(I0)MΩdMEm=1Mpm(Em)eA·f(Em)×σsgn(σ)m=1Mfσm(Em), (19)

where the sum is computed over all permutations σ of the set {1,2,,M}, and the sign of the permutation σ, sgn(σ), is +1 or −1 for even or odd permutations, respectively. Invoking the Leibniz formula again, we can simplify the Jacobian to:

det(AI)=(I0)MΩdMEm=1Mpm(Em)eA·f(Em)det[F(E)],=(I0)MΩdMEm=1Mfm(Em)eA·f(Em)det[P(E)], (20)

where the second line follows a similar derivation as Equation (19) by swapping the subscripts of p(E) and f(E). The matrix F as a function of E=(E1,E2,,EM) is

F(E)=f1(E1)f2(E1)fN(E1)f1(E2)f2(E2)fN(E2)f1(EM)f2(EM)fN(EM), (21)

and the matrix P as a function of E is

P(E)=p1(E1)p2(E1)pN(E1)p1(E2)p2(E2)pN(E2)p1(EM)p2(EM)pN(EM). (22)

The integrand in Equation (20) has several interesting symmetry properties. The factor m=1MeA·f(Em)=eA·m=1Mf(Em) has mirror symmetry about all hyperplanes Ei=Ej for i,j{1,2,..,M}. The other factor, det[F(E)], has sign‐switching mirror symmetry about the same hyperplanes, which can be described mathematically as:

det[F(E1,E2,,EM)]=sgn(σ)det[F(Eσ1,Eσ2,,EσM)]. (23)

A sign‐switching mirror symmetry means that, when we switch the positions of two coordinates (odd permutation), the sign of the function changes but the absolute value of the function is preserved. For example, with two coordinates E1 and E2, det[F(E1,E2)]=f1(E1)f2(E2)f2(E1)f1(E2)=det[F(E2,E1)]. To illustrate the sign‐switching mirror symmetry, we plotted det[F(E1,E2)] for the case when f1(E) and f2(E) are both Gaussian functions in Figure 4(b).

Now we can divide the space occupied by Ω into M! subspaces with hyperplanes Ei=Ej for i,j{1,2,..,M}. One of the subspace has property E1<E2<<EM and we define this subspace as Ω1,2M. For every point (E1,E2,,EM) in the subspace Ω1,2M, there is a corresponding point (Eσ1,Eσ2,,EσM) in each of the remaining subspaces. Applying the sign‐switching mirror symmetry of the determinant, we can further simplify the Jacobian to:

det(AI)=(I0)MΩ12..MdMEm=1MeA·f(Em)×det[F(E)]σsgn(σ)m=1Mpm(Eσm),=(I0)MΩ12..MdMEm=1MeA·f(Em)×det[F(E)]det[P(E)]. (24)

The symbol M emphasizes that the integration is over all spectral measurements. The three factors of the integrand, m=1MeA·f(Em), det[F(E)] and det[P(E)], depend on the total attenuation, the basis functions, and the energy‐weighting functions, respectively.

When the Jacobian is nonvanishing everywhere in RN, we can construct the domain M1 M1 and the range M2 M2 and prove that the mapping X:M1M2 X:M1M2 is globally invertible. As a result, the ME X‐ray transform defined on the domain PN PN, which is a subset of M1 M1, is globally invertible. Therefore, we have proved the equivalence of global invertibility with local invertibility for an ME X‐ray transform. This equivalence holds when K‐edge basis functions are considered.

3.4. A sufficient condition for invertibility

A sufficient but not necessary condition for J(A)0 is that the integrand of J(A), which is given in Equation (24), has the same sign over the subspace S12N and has nonzero values. The first factor in the integrand, which depends solely on the total attenuation, is always positive. If we ignore the trivial case that det[F(E)]det[P(E)]=0 everywhere, a sufficient condition for the invertibility of the ME X‐ray transform is det[F(E)]det[P(E)]0 (or 0) for all E in S12N. As the sign of the integrand does not depend on A, if the sufficient condition is satisfied, the Jacobian is nonvanishing for all A in RN.

When the photoelectric/Compton/Rayleigh basis functions are used, the basis‐function determinant det[F(E)] is always negative in the subspace Ω123. This set of three basis functions is sufficient to describe an object when the materials of interest have no K‐edges in the energy range used for imaging, for example, soft tissue and bone. The proof of det[F(E)]<0 for all E that satisfy E1<E2<E3 is provided in Online Appendix A. When K‐edge materials are considered, the values of det[F(E)] can be calculated numerically and the positive regions of det[F(E)] can be avoided by adjusting the detector sensitivity or source spectrum in pm(E).

Now we apply the sufficient condition for invertibility to the DE scenario that has nonunique solutions discussed by Levine. 7 For DE X‐ray imaging, a set of two basis functions, photoelectric and Compton, can be used. The basis‐function determinant is always positive in the subspace S12. Levine assumed the same detector response for the two measurements; hence, the sign of det[P(E)] is the same with the sign of det[S(E)], which is the source‐spectra determinant. Similar to the definition in Equation (22), the matrix S(E) can be written as

S(E1,E2)=S1(E1)S2(E1)S1(E2)S2(E2), (25)

where (E1,E2) can be any combinations of two energies. With these, det[S(E)]=S1(E1)S2(E2)S1(E2)S2(E1). Levine assumed that both source spectra S1(E) and S2(E) are not zero only at three energy points (30, 60, 100) keV. Hence, det[S(E)] is not zero only when (E1,E2) are combinations of the set {30,60,100} keV. Within the subspace S12, where E1 is always less than E2, det[S(E)] is nonvanishing only at three points (E1,E2)=(30,60), (30,100), and (60,100) keV. Given the two source spectra as S1(E)=(1,1,1) and S2(E)=(0.93,1.71,0.30) at E=(30,60,100) keV, respectively. The values of det[S(E)] at (E1,E2)=(30,60), (30,100), and (60,100) keV are 0.78, −0.63, and −1.41, respectively. Therefore, det[S(E)] does not have constant sign over the subspace S12N; hence, the invertibility of the DE X‐ray transform for the proposed scenario is not guaranteed. Note that this analysis only shows that the existence of nonunique solutions is possible, it does not prove their existence.

We then consider the sign of det[P(E)] for the four types of detectors illustrated in Figure 2. Assuming that the source spectrum S(E) is same for different m, the weighting‐function determinant det[P(E)] has the same sign with the detector‐response determinant det[D(E)], where the (i,j) element of matrix D(E) is the detector response of the ith measurement at Ej denoted as Di(Ej). The sign of det[D(E)] for the four types of detectors is studied in Online Appendix A and the main results are presented as follows.

  • (a)

    An CIX‐PC/EI/MI detector: det[D(E)]<0 for all EΩ123.

  • (b)

    A three bin detector, where the three bins are not overlapping and the energy‐response functions are arbitrary: det[D(E)]0 for all EΩ123.

  • (c)

    A three bin PC detector with rect‐response functions and possible overlaps between bins: if Bin 1 and Bin 3 has no overlap, the lower edges of the three bins satisfy l1<l2<l3 and the upper edges of the three bins satisfy u1<u2,<u3, the determinant det[D(E)]0 for all EΩ123.

  • (d)

    A nonoverlapping three bin PC detector with finite energy resolution: if the energy resolution of the detector can be modeled by a narrow truncated function (for mathematical description, see Online Appendix A) and there is no overlap between Bin 1 and Bin 3, det[D(E)]0 in Ω123.

In conclusion, the mapping X:AI is globally invertible for these four types of detectors when measuring attenuation profiles without K‐edges. For arbitrary detectors or systems with varying source spectra, the values of det[P(E)] can be calculated numerically. One can alway maintain det[F(E)]det[P(E)]0 for any E in S12N by adjusting the energy‐response function or the bin boundaries of the detectors.

4. ESTIMATION UNCERTAINTIES FOR POISSON DATA

From a practical point of view, it is also crucial to consider the uncertainty in the estimation under the presence of noise. In this section, we consider PC detectors with nonoverlapping bins. If only inherent quantum noise is considered, the data of the mth measurement at a given ray path, gm, is a Poisson random variable with mean equals to Im,

gm(A)=Poiss(Im(A)), (26)

where Im(A) is the mean photon count of the mth measurement given in Equation (5). Combining all M measurements, we get the measurement data g. The probability density function of the data g given the AM coefficient A along the ray path is

pr(g|A)=m=1MIm(A)gmeIm(A)gm!. (27)

The log‐likelihood function of the AM coefficients A is

L(A|g)=lnpr(g|A)=m=1MgmlnIm(A)Im(A)lngm!. (28)

The first derivative of the log‐likelihood function is

LAi(A)=m=1MgmIm(A)Im(A)ImAi(A)=m=1MgmIm(A)Im(A)[AI]mi. (29)

The Hessian, or second derivative, of the log‐likelihood function is given by

[A2L]ij=2LAiAj=m=1MgmImIm2ImAiAjgmIm2ImAiImAj. (30)

The components of the Fisher information matrix are

FIMij(A)=2LAiAjg|A=dMgpr(g|A)×m=1MgmImIm2ImAiAjImAiImAjgmIm2=m=1M1ImImAiImAj. (31)

Therefore, the Fisher information matrix is

FIM(A)=(AI)TΛ1(AI), (32)

where Λ is a diagonal matrix with the mth diagonal element equals to Im.

The Crámer–Rao bounds 28 , 29 characterize the limit on the estimation uncertainties induced by noise. It states that for an unbiased estimate of the ith parameter, its variance must be at least as large as the ith diagonal element of the inverse of the Fisher information matrix. Mathematically, the CRLBs are

Var(A^iAi)[FIM1]ii=[(AI)1Λ(AI)1,T]ii, (33)

where the symbol A^ indicates an estimate of A. Note that the uncertainty in the estimation is inversely related with the source photon budget I0, which agrees with our intuition. Also, note that the uncertainty of an unbiased estimation is inversely related with the Jacobian J(A)=|det(AI)|. If the Jacobian J(A) is close to zero, the estimation uncertainty is close to infinity and the coefficients cannot be estimated accurately in practice.

5. ESTIMATION ALGORITHM AND ILLUSTRATIVE RESULTS

In this section, we develop an ML algorithm for Poisson data. The goal of the algorithm is to estimate AM‐coefficients A from noisy data g. The assumption for the algorithm is that both Equations (5) and  (27) are valid. First, consider L as a function of the mean signal I. The first derivation of this function is

LIi=giIi1. (34)

Hence, the point I=g is a critical point. The Hessian of the function L(I|g) is

[I2L]ij=2LIiIj=giIi2σij, (35)

where σij=1 when i=j, and σij=0 otherwise. This Hessian is a diagonal matrix with all negative elements when gm>0. Therefore, the function L(I|g) is a concave function and the critical point at I=g is the global maximum for L. When the mapping X:AI is invertible and g is within the range of the mapping, the maximum value of L corresponds to a point Ag that satisfies I(Ag)=g.

We then consider L as a function of the AM coefficients A. When the matrix AI is invertible, which is true when the X‐ray transform is invertible, Ag is a critical point of the likelihood function L(A|g), as L(Ag|g) is the ML value. Furthermore, when A is located within the region defined by Im(A)gm, the likelihood function L(A|g) is a concave function of A. An ML algorithm can be developed based on Newton's method 30 , 31 with iterations described as

Ak+1=Ak+tkΔAk,andΔAk=[A2L(Ak)]1AL(Ak), (36)

where Ak is the attenuation coefficients at iteration k and tk is the step size chosen with an Armijo‐type (or back‐tracking) line search to enforce sufficient increase in L and the negativeness of the Hessian A2L. Note that the enforcement of negative definiteness of the Hessian is important, as the algorithm may not converge otherwise. The likelihood function is convex in the region 0AAg; hence, the algorithm works the best when the initialization point A0 is less than Ag. Furthermore, during the iterations, the Hessian may become rank‐deficient due to numerical accuracy. In this case, a gradient‐descent step can be used instead of the Newton step. For example, in our implementation, we used an initialization of A0=(0,0,0) and back‐tracking parameters α=0.1 and β=0.1 (α and β are defined as in Ref. 30), and the convergence of the algorithm required less than 15 iteration steps for every case we have tested in the following sections.

5.1. AM coefficients estimation uncertainties

To demonstrate the applications of this ML algorithm, we considered an ideal three‐bin PC detector with non‐overlapping rect response functions, as shown in Figure 2(c) but with equal heights and no overlapping. The source is operated at 160 kVp and generates a broad X‐ray spectrum. 32 The material attenuation profiles are extracted from the NIST XCOM data. X‐ray attenuation was simulated according to Beer's law. X‐ray scattering and detector imperfections are not considered. The data were simulated at source photon budget I0=107 with Poisson noise.

We simulated a single X‐ray path with different lengths of water as the attenuating media. The energy‐bin boundaries of the detector were set at [30, 75, 100, 160] keV. The length of water ranged from 1 to 28 cm. For each length, 1000 sets of noisy data were generated, and the AM coefficients were estimated for each set of data. We calculated the mean and the variance of the estimated coefficients and compared the estimation uncertainty to the CRLB. The results are presented in Figure 5.

FIGURE 5.

FIGURE 5

Uncertainties in the three estimated photoelectric/Compton‐scattering/Rayleigh‐scattering coefficients for different lengths of water

To check if the algorithm works for materials that are very different from water, we changed the material in the X‐ray path to iron and the detector bin boundaries to [30, 110, 140, 160] keV. The bin boundaries were changed so that there are sufficient number of photons collected in all three bins. The length of iron ranged from 0.1 to 3 cm. The mean and variance of the 1000 repeated estimations at each length are presented in Figure 6.

FIGURE 6.

FIGURE 6

Uncertainties in the three estimated photoelectric/Compton‐scattering/Rayleigh‐scattering coefficients for different lengths of iron

In both scenarios, the mean of the estimates (black line) matches well with the true coefficient (red circle). The slight deviation between the true coefficient and the mean estimation at high attenuation region can be attributed to sampling error in Monte‐Carlo simulation. The standard deviation of the estimates (purple area) is almost perfectly aligned with the CRLB. These results demonstrate that our estimation algorithm is unbiased and efficient. Furthermore, the AM coefficient corresponding to the Rayleigh scattering is estimable and the uncertainty in A^rs is comparable to the uncertainty in A^pe, which corresponds to the photoelectric effect. However, the A^pe and A^rs are anticorrelated, which is demonstrated by a negative value of the element [FIM1]13. This correlation is probably due to the fact that the basis functions fpe(E) and frs(E) have similar shapes. This correlation may partially explain why A^pe and A^rs have more variance than A^cs, as shown in Figure 5. As a result of the correlation, the estimated total attenuation τ^(E)=A^pefpe(E)+A^csfcs(E)+A^rsfrs(E), which is the ultimate physical quantity we are interest in, is not very noisy, as will be shown in the reconstruction results of the phantom.

5.2. Phantom reconstruction

We further applied the ML estimation algorithm for an image reconstruction. The reconstruction problem in X‐ray computed tomography (CT) is to estimate the distribution a(R) from the estimated line integrals A^ for each ray path. The AM coefficients a at each location R correspond to an attenuation profile μ(E). For a two‐dimensional scene, the object μ(E,R) and the reconstruction μ^(E,R) are both three‐dimensional data cubes. To present the reconstruction result, we plot μ(E,R) and μ^(E,R) at an arbitrary energy.

We simulate a two‐dimensional fan‐beam CT system (62 fan‐beam angle) with 360 views and 245 detectors. The field of view is 256×256 pixels with a pixel pitch of 1–1.5 mm. The same source, detector, and material database described in the previous section are used. X‐ray attenuation is simulated according to Beer's law, whereas scattering and detector imperfections are not considered. The detector energy bin boundaries are [30, 75, 100, 160] keV. The source photon budget for each beam path is 107. The AM coefficients A are estimated for each beam path and the object represented by a(R) is reconstructed from A^ using a filtered‐back projection (FBP) algorithm.

The first phantom reconstructed is a circular water phantom of diameter about 30 cm with pixel‐pitch 0.15 cm. This phantom and its reconstruction are plotted at E=75.5 keV in Figure 7. The reconstruction matches well with the object. As shown in the centerline plot in the right panel of Figure 7, the reconstruction has no cupping artifacts, which is typically associated with beam hardening.

FIGURE 7.

FIGURE 7

Reconstruction of the circular phantom. The object μ(E,R) (left), the reconstruction μ^(E,R) (middle), and the centerline plots (right) are all presented at E=75.5 keV

A second phantom is a multi‐material resolution phantom design inspired by Gong et al. 33 The length of the phantom is 20 cm with pixel pitch around 0.1 cm. The phantom, as shown in Figure 8 (left), is a Delrin block with 25 circular inserts in five rows and five columns. In each row, the inserts are made from the same material; in each column, the inserts have the same diameter. From top to bottom, the five materials for the inserts are water, polyvinyl chloride, cast magnesium, acrylic, and methanol. The diameters of the inserts are from 0.6 to 1.8 cm with 0.3 cm step. The reconstruction of the resolution phantom is plotted at E=75.5 keV in Figure 8. In the plots, different shades of grey represent different materials. The reconstruction results show that the ML estimation algorithm works for a broad range of AM coefficients

FIGURE 8.

FIGURE 8

Reconstruction of the resolution phantom. The object μ(E,R) (left), the reconstruction μ^(E,R) (middle), and the centerline plots (right) are all presented at E=75.5 keV

Each reconstruction, which calls the ML estimation algorithm 88,200 times, takes approximately 130 s on a desktop with a quad‐core central processing unit (CPU). The reconstruction can be further sped up using a graphic processing unit (GPU).

6. DISCUSSION

In our proof of invertibility, we focused on the interior points of M1 and M2 and proved that, for an ME X‐ray transform, the global invertibility is equivalent to local invertibility for A in the interior of M1 and I in the interior of M2. This equivalence can be extended to the boundaries of M1 and M2 by invoking Theorem 6 in Sandberg et al. 34 As mentioned in the introduction, Bal et al 9 have also provided a sufficient condition for the invertibility of ME X‐ray transform. Their sufficient condition is that the Jacobian is a P‐matrix in a rectangle in RN. P‐matrix, which is a concept related to the preservation of orientation, requires the matrix and a few submatrices to be all positive. For the definition of P‐matrix, please refer to Bal et al 9 or Gale and Nikaido. 35 Bal et al studied the invertibility of different ME X‐ray systems by numerically calculating the Jacobian matrix on a grid of A values for each system. Based on numerical simulations, they have suggested that an ME X‐ray system may become invertible as soon as the mapping is locally invertible in the rectangle. Our sufficient condition for global invertibility is that the Jacobian is nonvanishing for all points in M1. In comparison to Bal et al, our sufficient condition is weaker (better), but the domain where the Jacobian matrix needs to be checked is larger. For ease of computation, we also provide a sufficient condition for nonvanishing Jacobian, which requires the integrand of the Jacobian to have constant sign over all energy combinations in S12N. From a practical point of view, the latter condition is significantly easier to use, as (1) the sign of the Jacobian integrand does not depend on A, and (2) the properties of the basis functions and the detector response functions can be studied separately.

Invertibility only requires that the Jacobian J(A)0 for all coefficients A. Nonetheless, a smaller J(A) leads to a worse‐conditioned inverse problem and hence more uncertainty in the estimation as discussed in Section 4. Take the binning detector depicted in Figure 2(c) as an example, when Bin 1 and Bin 3 do not overlap, the system is invertible for N=3 (when imaging materials with no K‐edges). However, the overlap between bins would result in a reduction in the Jacobian and hence more uncertainty in the coefficient estimation, which has also been observed in other works. 36 , 37 One can employ the CRLB to optimize bin boundaries for a given set of AM coefficients A. As the CRLB varies with A, the optimum bin boundaries depend on the prior distribution of the objects. An optimum energy‐weighting strategy that is not object‐dependent has been proposed by Wang et al. 38 Their strategy is to set the weights pm(E) same as the attenuation basis functions fi(E). They have proved that this measurement strategy provided a sufficient statistic to the X‐ray spectral flux. From Equation (24), we can prove that this strategy is globally invertible, because det[F(E)]det[P(E)]={det[F(E)]}20, for all E.

Conventional DECT systems reconstruct the effective atomic number (Ze) and the electron density (ρ) 39 from two energy‐weighted measurements. However, Ze and ρ may not capture all of the information about material composition measurable from attenuation‐based X‐ray systems. Based on principal component analysis (PCA), Bornefalk et al 40 have suggested that the intrinsic dimensionality of the attenuation profiles of low‐Z materials in the XCOM data base is four. Midgley et al 41 also showed similar degrees of freedom in the parameterization of the X‐ray linear attenuation profiles. However, whether these intrinsic dimensions are accessible or not is still up to debate. 42 , 43 There is potential value in collecting ME X‐ray data, but the benefits may depend on the task of the imaging system and the experimental setup.

We used a set of basis functions that describe photoelectric, Compton scattering and Rayleigh scattering, because we wanted to investigate whether the Rayleigh coefficient is estimable or not. Rayleigh scattering has often been ignored in DE imaging due to its small contribution in the X‐ray attenuation profile. Our results show that the Rayleigh component, Ars, is solvable and the uncertainty in its estimation is comparable to that of the photoelectric coefficient. However, we did not specifically study how important Ars is for the task of material discrimination. Other basis functions that are based on materials of interest (such as water and bone) or on PCA 40 , 44 , 45 , 46 can be used as well. As pointed out by Alvarez et al, 8 the choice of a particular basis set does not affect the invertibility. The uncertainty in the estimated attenuation profile should not be affected by the choice of the basis set either.

We demonstrated a two‐step reconstruction algorithm that consists of an ML estimation of the AM coefficients and the FBP reconstruction. Many work have been done in DECT and MECT reconstruction. Reconstruction algorithms are currently available in three main flavors: object‐domain based, 47 , 48 projection‐domain based, 49 , 50 and one‐step statistical algorithms 51 , 52 , 53 , 54 , 55 that estimate a^(R) from the raw data directly. Our ML estimation algorithm was designed for Poisson likelihood and ideal ME X‐ray transform where Equation (5) is valid. When the Poisson likelihood model or the ideal forward model are not accurate, the estimation algorithm needs modification.

Our ML estimation algorithm is almost unbiased and achieves the CRLB. The ML estimation is an established paradigm for nonlinear estimation tasks. At high signal‐to‐noise ratio (SNR), ML estimates are asymptotically unbiased and efficient (achieves CRLB). At low SNR (e.g., short exposure time), however, the ML estimates tend to be skewed and the variance is often larger than the Cramer–Rao bound. 56 The reason why our estimator is efficient is probably that our simulation was carried out in the high SNR regime. In our experiment, the smallest photon count collected in an energy bin is about 300 photons, which still has a relatively high SNR. To analyze a realistic system, one need to first identify if the system is operating at low SNR regime. If that is the case, instead of using the Crámer–Rao bound to characterize the variance of the ML estimates, one can apply other measures such as χpdfML2‐isocontours 57 to describe the distribution of the ML estimates.

There are several limitations in our work. First, the physical process considered by the mapping X:AI includes only the attenuation of the X‐ray photons, which follows Beer's law, but not signals due to scattered radiation or background radiation. Although scattered radiation and background radiation can be significantly mitigated through antiscatter grids, 58 those signals should be characterized and accounted for, as they may affect the invertibility of a realistic system. Second, effects that limit detector performance, such as charge sharing, 59 , 60 charge trapping, 60 , 61 and pulse pileup, are ignored. Such details should be considered in the system model when the framework is applied to a specific imaging system. For a given realistic detector response function, the invertibility of the system can be studied by calculating det[P(E)] numerically over the subspace Ω12N. In this case, det[P(E)]0 (or 0) over Ω12N may not be guaranteed; hence, the invertibility is not guaranteed. However, one can apply the invertibility framework in the optimization of detector designs. Lastly, we assumed the energy‐weighting functions, including the source spectrum and the detector energy response functions, are known exactly. In reality, one can measure the energy‐weighting functions experimentally 62 within some uncertainty.

7. CONCLUSION

We have provided a sufficient condition for the global invertibility of an ME X‐ray transform for attenuation‐based X‐ray imaging. The ME X‐ray transform is defined as the mapping from N (N2) AM coefficients to N energy‐weighted noise‐free measurements. The invertibility of this transform depends greatly on the weighting schemes used in the measurements. Considering scenes with no K‐edge materials, we represented the X‐ray attenuation profiles with N=3 AM coefficients and proved the global invertibility of the transform for four commonly used weighting schemes. The same framework can be used to examine the invertibility of an arbitrary ME X‐ray system, such as a system with nonideal detectors, a system with multiple source emission spectra, and scenes with K‐edge materials. This mathematical framework can be applied broadly in the design of X‐ray detectors and systems.

We also considered Poisson noise in the measurement data and presented the CRLB on the estimation uncertainty. Furthermore, we presented an ML estimation algorithm and applied the algorithm to estimate AM coefficients for varying lengths of water and varying lengths of iron. The results have shown that the coefficient corresponding to Rayleigh scattering is estimable. Last but not least, we demonstrated the application of the ML estimator in reconstruction. Two phantoms imaged through a simulated fan‐beam CT with ideal three‐energy discriminating PC detectors were reconstructed. The reconstructed images match well with the objects and are free of the “cupping artifacts” induced by beam hardening.

CONFLICT OF INTEREST

The authors declare that there are no conflicts of interest related to this article.

Supporting information

Invertibility of Multi‐Energy X‐ray Transform–Appendix A

ACKNOWLEDGMENTS

Dr. Clarkson acknowledges the support of NIH R01‐ EB000803 and P41‐ EB002035.

Ding Y, Clarkson EW, Ashok A. Invertibility of multi‐energy X‐ray transform. Med Phys. 2021;48:5959–5973. 10.1002/mp.15168

DATA AVAILABILITY STATEMENT

The data and code that support the findings of this study are available from the corresponding author upon reasonable request.

REFERENCES

  • 1. Alvarez RE, Macovski A. Energy‐selective reconstructions in X‐ray computerised tomography. Phys Med Biol. 1976;21: 733‐744. [DOI] [PubMed] [Google Scholar]
  • 2. Lehmann LA, Alvarez RE, Macovski A, et al. Generalized image combinations in dual kVp digital radiography. Med Phys. 1981;8: 659‐667. [DOI] [PubMed] [Google Scholar]
  • 3. Barnes GT, Sones RA, Tesic MM, Morgan DR, Sanders JN. Detector for dual‐energy digital radiography. Radiology. 1985;156: 537‐540. [DOI] [PubMed] [Google Scholar]
  • 4. Cardinal HN, Fenster A. Theoretical optimization of a split septaless Xenon ionization detector for dual‐energy chest radiography. Med Phys. 1988;15: 167‐180. [DOI] [PubMed] [Google Scholar]
  • 5. Roessl E, Proksa R. K‐edge imaging in X‐ray computed tomography using multi‐bin photon counting detectors. Phys Med Biol. 2007;52: 4679‐4696. [DOI] [PubMed] [Google Scholar]
  • 6. Fredenberg E. Spectral and dual‐energy X‐ray imaging for medical applications. Nucl Instrum Methods Phys Res, Sect A: Accel Spectrometers Detectors Associated Equip. 2018;878: 74‐87. [Google Scholar]
  • 7. Levine ZH. Nonuniqueness in dual‐energy CT. Med Phys. 2017;44: e202‐e206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Alvarez RE. Invertibility of the dual energy X‐ray data transform. Med Phys. 2019;46: 93‐103. [DOI] [PubMed] [Google Scholar]
  • 9. Bal G, Terzioglu F. Uniqueness criteria in multi‐energy CT. Inverse Prob. 2020;36: 065006. [Google Scholar]
  • 10. Sukovle P, Clinthorne NH. Basis material decomposition using triple‐energy X‐ray computed tomography. In: IMTC/99: Proceedings of the 16th IEEE Instrumentation and Measurement Technology Conference (Cat. No. 99CH36309) , IEEE; 1999;3:1615‐1618. [Google Scholar]
  • 11. Lehmann L, Alvarez R, Macovski A, et al. Energy‐selective reconstructions in X‐ray computerized tomography. Med Phys. 1981;8: 659‐667. [DOI] [PubMed] [Google Scholar]
  • 12. Macovski A, Alvarez RE, Chan JL‐H, Stonestrom JP, Zatz LM. Energy dependent reconstruction in X‐ray computerized tomography. Comput Biol Med. 1976;6: 325‐336. [DOI] [PubMed] [Google Scholar]
  • 13. Tapiovaara MJ, Wagner R. SNR and DQE analysis of broad spectrum X‐ray imaging. Phys Med Biol. 1985;30: 519‐529. [Google Scholar]
  • 14. Flohr TG, McCollough CH, Bruder H, et al. First performance evaluation of a dual‐source CT (DSCT) system. Eur Radiol. 2006;16: 256‐268. [DOI] [PubMed] [Google Scholar]
  • 15. Zhang Da, Li X, Liu B. Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode. Med Phys. 2011;38: 1178‐1188. [DOI] [PubMed] [Google Scholar]
  • 16. Altman A, Carmi R. TU‐E‐210A‐03: a double‐layer detector, dual‐energy CT–principles, advantages and applications. Med Phys. 2009;36: 2750‐2750. [Google Scholar]
  • 17. Kraft E, Fischer P, Karagounis M, et al. Counting and integrating readout for direct conversion X‐ray imaging: concept, realization and first prototype measurements. IEEE Trans Nucl Sci. 2007;54: 383‐390. [Google Scholar]
  • 18. Karg J, Niederlöhner D, Giersch J, Anton G. Using the Medipix2 detector for energy weighting. Nucl Instrum Methods Phys Res, Sect A: Accel Spectrometers Detectors Associated Equip. 2005;546: 306‐311. [Google Scholar]
  • 19. Taguchi K, Iwanczyk JS. Vision 20/20: Single photon counting X‐ray detectors in medical imaging. Med Phys. 2013;40: 100901. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Fredette NR, Kavuri A, Das M. Multi‐step material decomposition for spectral computed tomography. Phys Med Biol. 2019;64: 145001. [DOI] [PubMed] [Google Scholar]
  • 21. Fulks W. Advanced Calculus: An Introduction to Analysis. John Wiley & Sons; 1978. [Google Scholar]
  • 22. Williamson JF, Li S, Devic S, Whiting BR, Lerma FA. On two‐parameter models of photon cross sections: application to dual‐energy CT imaging. Med Phys. 2006;33: 4115‐4129. [DOI] [PubMed] [Google Scholar]
  • 23. Trubey DK. HUGI VI Notes Documentation for the DLC‐146 Code Package. Radiaton Safety Information Computational Center; 1989. [Google Scholar]
  • 24. Berger MJ, Hubbell JH, Seltzer SM, et al. XCOM: photon cross sections database. NIST Standard Reference Database 8 (XGAM). http://physics.nist.gov/PhysRefData/Xcom/Text/XCOM.html. 1998.
  • 25. Krantz SG, Parks HR. The Implicit Function Theorem: History, Theory, and Applications. Springer Science & Business Media; 2012. [Google Scholar]
  • 26. Munkres JR. Topology;2. Prentice Hall; 2000. [Google Scholar]
  • 27. Lee J. Introduction to Smooth Manifolds. Springer; 2012. [Google Scholar]
  • 28. Cramer H. Mathematical Methods of Statistics. Princeton University Press; 1946. [Google Scholar]
  • 29. Rao CR. Information and the accuracy attainable in the estimation of statistical parameters. Bull Calcutta Math Soc. 1945;37: 81‐89. [Google Scholar]
  • 30. Boyd S, Boyd SP, Vandenberghe L. Convex Optimization. Cambridge University Press; 2004. [Google Scholar]
  • 31. Nocedal J, Wright S. Numerical Optimization. Springer Science & Business Media; 2006. [Google Scholar]
  • 32. Ding Y, Ashok A. X‐ray measurement model and information‐theoretic metric incorporating material variability with energy correlations in Anomaly Detection and Imaging with X‐rays (ADIX) IV . J Int Soc Opt Photonics. 2019;10999: 109990. [Google Scholar]
  • 33. Gong Q, Stoian R‐I, Coccarelli DS, Greenberg JA, Vera E, Gehm ME. Rapid simulation of X‐ray transmission imaging for baggage inspection via GPU‐based ray‐tracing. Nucl Instrum Methods Phys Res, Sect B: Beam Interactions Materials Atoms. 2018;415:100–109. [Google Scholar]
  • 34. Sandberg I. Global inverse function theorems. IEEE Trans Circuits Syst. 1980;27: 998‐1004. [Google Scholar]
  • 35. Gale D, Nikaido H. The Jacobian matrix and global univalence of mappings. Mathematische Annalen. 1965;159: 81‐93. [Google Scholar]
  • 36. Ren Y, Xie H, Long W, Yang X, Tang X. On the conditioning of spectral channelization (energy binning) and its impact on multi‐material decomposition based spectral imaging in photon‐counting CT. IEEE Trans Biomed Eng. 2021:1. [DOI] [PubMed] [Google Scholar]
  • 37. Tang X, Ren Y. On the conditioning of basis materials and its impact on multi‐material decomposition based spectral imaging in photon‐counting CT. Med Phys. 2021. [DOI] [PubMed] [Google Scholar]
  • 38. Wang AS, Pelc NJ. Sufficient statistics as a generalization of binning in spectral X‐ray imaging. IEEE Trans Med Imaging. 2010;30: 84‐93. [DOI] [PubMed] [Google Scholar]
  • 39. Martz HE, Seetho IM, Champley KE, Smith JA, Azevedo SG. CT dual‐energy decomposition into X‐ray signatures ρe and Ze in Anomaly Detection and Imaging with X‐Rays (ADIX) . Int Soc Opt Photonics. 2016;9847:98470. [Google Scholar]
  • 40. Bornefalk H. XCOM intrinsic dimensionality for low‐Z elements at diagnostic energies. Med Phys. 2012;39: 654‐657. [DOI] [PubMed] [Google Scholar]
  • 41. Midgley SM. A parameterization scheme for the x‐ray linear attenuation coefficient and energy absorption coefficient. Phys Med Biol. 2004;49: 307‐325. [DOI] [PubMed] [Google Scholar]
  • 42. Midgley SM. Materials analysis using x‐ray linear attenuation coefficient measurements at four photon energies. Phys Med Biol. 2005;50: 4139‐4157. [DOI] [PubMed] [Google Scholar]
  • 43. Levine ZH, Peskin AP, Holmgren AD, Garboczi EJ. Preliminary X‐ray CT investigation to link Hounsfield unit measurements with the International System of Units (SI). PLoS One. 2018;13: e0208820. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Alvarez RE. Dimensionality and noise in energy selective X‐ray imaging. Med Phys. 2013;40: 111909. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Weaver JB, Huddleston AL. Attenuation coefficients of body tissues using principal‐components analysis. Med Phys. 1985;12: 40‐45. [DOI] [PubMed] [Google Scholar]
  • 46. Xie H, Ren Y, Long W, Yang X, Tang X. Principal component analysis in projection and image domains Another form of spectral imaging in photon‐counting CT. IEEE Trans Biomed Eng. 2021;68: 1074‐1083. [DOI] [PubMed] [Google Scholar]
  • 47. Brooks RA. A quantitative theory of the Hounsfield unit and its application to dual energy scanning. J Comput Assisted Tomogr. 1977;1: 487‐493. [DOI] [PubMed] [Google Scholar]
  • 48. Maaß C, Baer M, Kachelrieß M. Image‐based dual energy CT using optimized precorrection functions: a practical new approach of material decomposition in image domain. Med Phys. 2009;36: 3818‐3829. [DOI] [PubMed] [Google Scholar]
  • 49. Abascal Juan FPJ, Ducros N, Peyrin F. Nonlinear material decomposition using a regularized iterative scheme based on the Bregman distance. Inverse Prob. 2018;34: 124003. [Google Scholar]
  • 50. Wu D, Zhang L, Zhu X, Xu X, Wang S. A weighted polynomial based material decomposition method for spectral x‐ray CT imaging. Phys Med Biol. 2016;61: 3749‐3783. [DOI] [PubMed] [Google Scholar]
  • 51. Mory C, Sixou B, Si‐Mohamed S, Boussel L, Rit S. Comparison of five one‐step reconstruction algorithms for spectral CT. Phys Med Biol. 2018;63: 235001. [DOI] [PubMed] [Google Scholar]
  • 52. Mechlem K, Ehn S, Sellerer T, et al. Joint statistical iterative material image reconstruction for spectral computed tomography using a semi‐empirical forward model. IEEE Trans Med Imaging. 2017;37: 68‐80. [DOI] [PubMed] [Google Scholar]
  • 53. Long Y, Fessler JA. Multi‐material decomposition using statistical image reconstruction for spectral CT. IEEE Trans Med Imaging. 2014;33: 1614‐1626. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Barber RF, Sidky EY, Schmidt TG, Pan X. An algorithm for constrained one‐step inversion of spectral CT data. Phys Med Biol. 2016;61: 3784‐3818. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55. Kazantsev D, Jørgensen JS, Andersen MS, Lionheart WRB, Lee PD, Withers PJ. Joint image reconstruction method with correlative multi‐channel prior for x‐ray spectral computed tomography. Inverse Prob. 2018;34: 064001. [Google Scholar]
  • 56. Mueller SP, Kijewski MF, Kappeler C, Moore SC. Estimation performance at low SNR: predictions of the Barankin bound in Medical Imaging 1995: Physics of Medical Imaging. Int Soc Opt Photonics. 1995;2432: 157‐166. [Google Scholar]
  • 57. Müller SP, Abbey CK, Rybicki FJ, Moore SC, Kijewski MF. Measures of performance in nonlinear estimation tasks: prediction of estimation performance at low signal‐to‐noise ratio. Phys Med Biol. 2005;50: 3697‐3715. [DOI] [PubMed] [Google Scholar]
  • 58. Tang C‐M, Stier E, Fischer K, Guckel H. Anti‐scattering X‐ray grid. Microsyst Technol. 1998;4: 187‐192. [Google Scholar]
  • 59. Shikhaliev PM, Fritz SG, Chapman JW. Photon counting multienergy X‐ray imaging: effect of the characteristic X rays on detector performance. Med Phys. 2009;36: 5107‐5119. [DOI] [PubMed] [Google Scholar]
  • 60. Xu C, Danielsson M, Bornefalk H. Evaluation of energy loss and charge sharing in cadmium telluride detectors for photon‐counting computed tomography. IEEE Trans Nucl Sci. 2011;58: 614‐625. [Google Scholar]
  • 61. Knoll GF. Radiation Detection and Measurement. John Wiley & Sons; 2010. [Google Scholar]
  • 62. Ha W, Sidky EY, Barber RF, Schmidt TG, Pan X. Estimating the spectrum in computed tomography via Kullback–Leibler divergence constrained optimization. Med Phys. 2019;46: 81‐92. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Invertibility of Multi‐Energy X‐ray Transform–Appendix A

Data Availability Statement

The data and code that support the findings of this study are available from the corresponding author upon reasonable request.


Articles from Medical Physics are provided here courtesy of Wiley