Abstract
Microstructure imaging seeks to noninvasively measure and map microscopic tissue features by pairing mathematical modeling with tailored MRI protocols. This article reviews an emerging paradigm that has the potential to provide a more detailed assessment of tissue microstructure—combined diffusion‐relaxometry imaging. Combined diffusion‐relaxometry acquisitions vary multiple MR contrast encodings—such as b‐value, gradient direction, inversion time, and echo time—in a multidimensional acquisition space. When paired with suitable analysis techniques, this enables quantification of correlations and coupling between multiple MR parameters—such as diffusivity, , , and . This opens the possibility of disentangling multiple tissue compartments (within voxels) that are indistinguishable with single‐contrast scans, enabling a new generation of microstructural maps with improved biological sensitivity and specificity.
Keywords: diffusion, multidimensional MRI, quantitative MRI, relaxometry
1. INTRODUCTION
This article reviews the current capabilities and future potential of an emerging paradigm in microstructure imaging: combined diffusion‐relaxometry. Diffusion MRI (dMRI) indirectly assesses tissue microstructure by measuring water diffusion. 1 MR relaxometry, while also sensitive to small‐scale tissue structures, 2 additionally offers information on the chemical composition of tissue 3 through the estimation of transverse and longitudinal relaxation times. Conventionally, diffusion and relaxation properties are measured and analyzed independently. Combined diffusion‐relaxometry techniques measure and analyze diffusion and relaxation properties jointly, based on scans that vary both diffusion (eg, b‐value and gradient direction) and relaxation (eg, inversion time [TI], flip angle, repetition time [TR] and echo time [TE]) sensitizing sequence parameters in multiple combinations. This yields images where contrast reflects both diffusion properties (eg, diffusivity and anisotropy) and relaxation times (eg, , , ). The opening up of a multidimensional acquisition space enables exploration of the correlations between these complementary MR contrasts. Analyzing such multidimensional data with appropriate techniques can potentially reveal unique information on tissue microstructure. In particular, by identifying and disentangling the unique MR signatures of different components, we can precisely characterize multiple tissue environments within a single voxel.
In this article, we focus on imaging, but the concepts and approaches we discuss have a strong foundation in multidimensional NMR techniques developed in the context of porous media analysis. Such experiments typically sample diffusion and relaxation sequence parameters in a multidimensional acquisition space, and hence estimate multidimensional correlation spectra (ie, multivariate distributions of NMR properties). First used to estimate ‐ distributions (eg, Ref. [4]) they were later extended to calculate ‐diffusivity distributions. 5 , 6 , 7 The approach was further enabled by development of efficient techniques 8 , 9 for calculating multidimensional correlation spectra, allowing combined diffusion‐relaxometry NMR to be deployed in a wide variety of applications, including geology 10 , 11 , 12 and food science. 13 , 14 , 15 The uptake of these techniques shows that combining diffusion and relaxation information in multidimensional scans can provide heightened sensitivity to chemical composition and microstructural features (see references 16 , 17 for in depth reviews of multidimensional correlation NMR in porous media). The new insights into porous media demonstrated by these techniques motivated their translation to quantify microstructure in biological systems. Early work applied combined diffusion‐relaxometry to ex‐vivo biological systems, such as frog sciatic nerve, 18 rat brain, 19 yeast cells, 20 and muscle 21 ; alongside in vivo studies exploring the interrelationship between diffusion and relaxation properties in human brain. 22
There are two main strengths of combined diffusion‐relaxometry driving its expansion in the microstructure imaging field: (i) it accesses complementary measurements with the potential to separate and quantify multiple microstructural environments and (ii) it accounts for inherent biases in such measurements, which are present as relaxation properties of complex tissue environments in an MRI voxel influence the estimation of the corresponding diffusion properties. These biases stem from the intrinsic dependence of the diffusion MRI signal on sequence parameters, such as TE and TR, which hinders the separation and quantification of tissue compartments with distinct chemical and microstructural properties. For example, if , , and/or vary across tissue compartments, then volume fractions inferred from diffusion‐only scans will be additionally weighted by the corresponding relaxation times. This means that while it may be possible to satisfactorily quantify some microstructural environments by measuring multiple MR contrasts in a series of independent 1D acquisitions (ie, dots in Figure 1) other microstructural environments can only be comprehensively characterized by measuring multiple contrasts in combination, that is, by varying diffusion and relaxation encoding parameters in a 2D (ie, planes in Figure 1) or higher acquisition space. We highlight this with a ‐diffusion example in Figure 2. Two tissue structures are shown in (A) and (B), comprising two and three distinct microstructural compartments respectively. These tissue structures cannot be distinguished with single contrast experiments that estimate relaxation time and diffusivity separately. While there are numerous real‐world examples, in , 23 24 , 25 and diffusion 1 domains, where tissue microenvironments can be distinguished with single‐contrast experiments, combined multicontrast experiments may be required to fully assess the tissue of interest. An example is the proposed separation of intra‐ and extra‐axonal compartments in the brain at clinically accessible diffusion weightings (), which is facilitated by combining both and diffusivity measures 26 ; using only relaxation or diffusion data separately leads to possible failure of the model to represent reality and/or to very high uncertainty on the intra‐axonal vs extra‐axonal estimates.
FIGURE 1.

The parameter space relevant for combined diffusion‐relaxometry experiments. Diffusion parameters, for example, b‐value and gradient direction, are represented on the x‐axis. The echo times (TE) are on the y‐axis, and the inversion times (TI) on the z‐axis. The green dots represent a conventional diffusion acquisition at fixed TE with multiple diffusion preparations. The red dots illustrate a scan sampling several TEs without diffusion weighting to achieve / maps. The blue dots illustrate a scan sampling multiple TIs without diffusion weighting to achieve maps. The transparent cyan, yellow and magenta planes depict the acquisition parameter space sampled in hypothethical ‐diffusion, ‐diffusion (equivalently ‐diffusion), and T1‐T2 experiments, respectively
FIGURE 2.

Separate diffusion and relaxation experiments can confound distinct tissue microenvironments. A,B, display simplified tissue structures comprising two and three distinct microenvironments, respectively. Each point denotes a distinct microenvironment with fixed , diffusivity values, and a percentage volume fraction. The projected and diffusivity distributions are shown on the top and right‐hand sides. The 1D distributions are equivalent for both (A) and (B) despite the different tissue structures, showing that 1D measurements can confound distinct tissue microenvironments. Combined diffusion‐relaxometry can disentangle the contributions from the distinct microenvironments due to its ability to quantify correlations between multiple contrasts
This review provides an introduction to combined diffusion‐relaxometry imaging with focus on three overarching themes: acquisition, analysis, and applications. Suitable acquisition techniques are crucial for diffusion‐relaxometry, particularly to address the severe increase in scan time that moving from 1D to 2D (and higher dimensions) necessitates. The acquisition section reviews the additional contrast encoding parameters that can be combined with diffusion to make a diffusion‐relaxometry experiment, including TE in spin echo and gradient echo sequences, and TI in inversion recovery sequences. We also review additional diffusion‐weighting parameters beyond b‐values and gradient directions that can be combined with relaxometry, such as diffusion time, b‐tensor shape, mixing time, and B0 field strength. Given these higher dimensional datasets, new analysis techniques that account for correlations between relaxation and diffusion properties can offer exciting new perspectives on tissue microstructure. We discuss the range of modeling approaches in the analysis section, including continuum modeling methods that make minimal assumptions about tissue structure to calculate multidimensional correlation spectra, other signal representation approaches based on the cumulant expansion, and microstructure modeling approaches where a fixed number of water pools are assumed a priori. In line with the ultimate goal of combined diffusion‐relaxometry techniques that make biologically and clinically meaningful observations, in the final section we review application areas where diffusion‐relaxometry techniques have contributed novel insights.
2. ACQUISITION
We first review techniques suitable for acquiring combined diffusion‐relaxometry data. The essential property of such techniques is that they sample a multidimensional parameter space including diffusion (eg, b‐value) and relaxation (eg, TE, TI) encoding parameters. This section proceeds as follows. In “Background and Motivation,” we first explain a significant limitation of dMRI—the confounds that diffusion preparations introduce into MRI acquisition sequences and the resulting intertwined relationship between dMRI and relaxation, then introduce some basic combined diffusion‐relaxometry acquisition strategies. We next introduce and discuss advanced acquisition techniques that seek to acquire more informative diffusion‐relaxometry data with higher efficiency in “Advanced combined diffusion‐relaxometry acquisition.” We finish by introducing some advanced diffusion MRI and MR relaxometry techniques that are particularly relevant for combined diffusion‐relaxometry in “Combined diffusion‐relaxometry and complimentary advanced acquisition techniques.”
2.1. Background and motivation
2.1.1. Intrinsic relaxation in diffusion MRI experiments
The standard technique for dMRI is single‐shot echo planar imaging (ssEPI). In ssEPI the diffusion preparation is followed by an EPI read‐out train as shown in Figure 3. In this illustrated example, the diffusion preparation, described by the gradient waveform, is a standard pulsed gradient experiment (ie, pulsed gradient spin echo [PGSE]) with two gradient lobes of equal length , equal polarity, and equal gradient strength . One lobe lies between the excitation and refocusing pulse, with the other between the refocusing pulse and the start of the EPI read‐out. The spacing between the gradient lobes () determines the diffusion time. In this classic setup, the b‐value is determined by , a special case of the more generic formulation
| (1) |
where
| (2) |
describes the gradient waveform, and denotes the field‐strength‐dependent gyromagnetic ratio over 2. Typically the echo time (TE) is minimized, which achieves the highest possible SNR. Therefore, the key limitations are the length of EPI read‐out train before the TE ( in Figure 3), and the available gradient strength . Figure 3 also illustrates that the diffusion acquisition does not happen in isolation, but in parallel to the / decay. It follows that the measured ssEPI signal is influenced both by decay and by the effect of diffusion gradients and can be written, assuming monoexponential relaxation and diffusion decays
| (3) |
where D is the apparent diffusion coefficient and is the relaxation time. It follows that the choice of the highest b‐value in a dMRI acquisition, and hence the minimum TE possible due to the restrictions outlined above (assuming fixed TE across all volumes), has a direct effect on the signal attenuation for all b‐values in the acquisition. Therefore, these choices influence quantitative diffusion‐related metrics derived from dMRI experiments. Two artifacts arising from the influence of the transverse relaxation are shine‐through and blackout, where variations in the time (eg, in hematomas or myelin water) influence the dMRI signal intensity. 27
FIGURE 3.

Standard pulsed gradient spin echo (PGSE) ssEPI dMRI acquisition. From top down the pulse diagram shows: the excitation and refocusing RF pulses, a standard diffusion preparation in all three directions alongside the EPI read‐out train, and the and decay occurring during the spin echo acquisition. A whole dMRI scan comprises multiple repeats of these PGSE ssEPI blocks, one for each slice and each diffusion preparation used. The repetition time TR is defined as the time required to sample each slice in one stack. denotes the readout time, the available gradient strength, the length of the gradient lobe, and the separation of the gradient pulses
2.1.2. Basic combined diffusion‐relaxometry acquisition
The inherent relationship between diffusion and relaxation acquisition sequences, alongside the opportunity to probe tissue microstruture in more detail, motivates MRI acquisition sequences that probe diffusion and relaxation properties and their correlations. We will now introduce and review such acquisition sequences. For clarity, here and throughout we avoid using the terms “multicontrast” and “multimodal”, which can refer to any scan (or series of individual scans) that measures multiple MR contrasts. Instead we use “combined diffusion‐relaxometry,” to mean an experiment where diffusion and relaxation encoding parameters are varied in a 2D (or higher) acquisition space (eg, a diffusion‐prepared sequence repeated at multiple TEs), and “simultaneous combined diffusion‐relaxometry” for the special case where multiple diffusion and relaxation encoding parameters are varied within a single repetition time (TR) (eg, diffusion‐prepared sequence where multiple TEs are measured in a single TR).
Consecutive acquisitions
The most common way to acquire combined diffusion‐relaxometry data is by repeating a diffusion‐weighted scan with varying relaxation‐encoding parameters. For example, since relaxation time can be estimated by acquiring multiple spin echoes with different TEs, 28 typical ‐diffusion experiments comprise acquiring multiple diffusion prepared spin echoes with varying TEs (eg, Figure 4A). With this approach, each sample in the high‐dimensional acquisition space (eg, Figures 1 and 4) is obtained with its own acquisition block of length TR. This is inefficient, as the time per sample is limited by the time required to achieve the specific diffusion preparation. This can lead to large idle times compared to typical dMRI—where TE is minimized subject to the constraints of the required b‐value and imaging gradients (eg, Figure 5A)—as data are acquired at a range of TEs with some longer than the minimal TE (eg, Figure 5B). Consequently, typical consecutive combined diffusion‐relaxometry ‐diffusion acquisitions, such as in Figure 4A, have taken 1 hour to scan the whole brain with around 500 volumes at 2.5 mm isotropic resolution. 26 , 29 As well as being slow and inefficient, this approach adds to the risk of inconsistencies and bias through motion and modifying the diffusion time.
FIGURE 4.

Example ‐diffusion acquisitions with three existing techniques. Each color represents a distinct excitation of length TR. A, The consecutive approach: repeat diffusion‐prepared scans with different TEs (illustrative acquisition time: 50 TRs). B, Simultaneous approach: applying techniques allowing multiple samples in the parameter space within the same excitation of length TR (illustrative acquisition time: 10 TRs). C, MADCO approach 58 : Fully sampled 1D data are augmented with sparsely sampled multidimensional data (illustrative acquisition time: 23 TRs). While we only show 2D ‐diffusion (or ‐diffusion) experiments for convenience, the same principles apply when extending these techniques to 3D and higher
FIGURE 5.

Illustration of separate vs integrated acquisition strategies using the example of multiple diffusion‐encoded scans with varying TE, that is, ‐diffusion. A,B, show separate acquisitions at different TEs, together with the included dead time and change in diffusion preparation. Acquiring (A) and (B) to sample the 2D space is a basic ‘consecutive’ acquisition. Integrated multi‐echo acquisitions are shown in (C) for spin echoes, hence acquiring combined diffusion‐relaxometry data ‘simultaneously’
The effect for ‐diffusion experiments is even more drastic. A basic ‐diffusion acquisition comprises spin echo acquisitions preceded with a global inversion pulse, as in a typical inversion recovery sequence, 30 with the time between the global inversion pulse and the first spin echo pulse (the inversion time, TI) varied to yield sensitivity. This introduces significant delays compared to typical dMRI sequences, as the time between inversion and excitation, which sets the achieved TI (see Figure 6B), is typically in the range of 0‐3000 ms. Moreover, the inversion recovery sequence is typically performed slice‐by‐slice, with the global inversion pulse followed by reading out a single slice. An early in vivo ‐diffusion combined diffusion‐relaxometry experiment gave whole brain ‐diffusion coverage with 2 mm isotropic voxels in 1 hour. 31
FIGURE 6.

Illustration of several alternatives to the conventional PGSE acquisition. A, Replacing the bipolar gradient pair by free waveforms used for q‐space sampling. B, Addition of a global inversion pulse to achieve a set inversion time TI. C, STEAM acquisition including and depicting the mixing time, introducing sensitivity and changing the diffusion time
This intrinsic inefficiency in sequences, together with the huge number of possible data points—two previous 2D diffusion‐relaxometry studies using uniform sampling utilized 1024 32 and 225 000 33 data points—hampers multidimensional scanning, and in the past prevented translation of such techniques from NMR to imaging. It is therefore clear that methods for faster and more efficient combined diffusion‐relaxometry acquisition protocols are of paramount importance. Two commonly employed strategies for working toward this goal are:
modifying the pulse sequence to acquire data points more efficiently (ie, efficient sampling schemes)
selecting acquisition parameters that maximize the information content for subsequent analysis and interpretation (ie, optimized sampling schemes)
2.2. Advanced combined diffusion‐relaxometry acquisition
2.2.1. Efficient sampling schemes
There are multiple techniques that apply the aforementioned first strategy, often inspired by efficient simultaneous diffusion and relaxation acquisitions pioneered in the NMR domain. 34 , 35 , 36 These bespoke MR sequences typically improve acquisition efficiency by combining multiple contrast preparations and multiple read‐outs within a single TR. In particular, for transverse relaxometry, time can be saved by acquiring multiple echoes (either spin or gradient) after the initial diffusion preparation and spin echo readout (eg, Figure 4B). This is further illustrated for the multiple spin echo case, which captures information, in Figure 5C. Similarly, additional gradient echoes would capture information, akin to a typical multi‐echo gradient echo scan for measuring relaxation. 37 In both spin and gradient echo cases, this approach means that a single diffusion preparation is shared across multiple echoes, rather than being repeated for each echo, with the benefits of a more efficient scan and a consistent diffusion preparation across echoes. This results in a more complete sampling of the TE‐diffusion preparation space within a fixed scan time. The exact speed up depends on the number of TEs chosen, and the required extension of the shot length and the duty cycle requirements set by the heating of the gradients. For large b‐values, the TR might be extended due to this requirement of cooling periods, potentially offsetting some time gains when acquiring multiple echoes.
For relaxometry experiments where a global inversion pulse is inserted before the first excitation (ie, Figure 6B), again there are NMR‐inspired techniques that more efficiently sample multiple TIs over the imaging volume. Typical inversion recovery experiments acquire multiple spatial slices sequentially, meaning that each slice has a different effective TI value, allowing the measurement of many points on the inversion recovery curve, albeit at different spatial locations, within a single TR. The addition of slice‐shuffling as originally proposed in 1990 and recently done in a number of studies, 38 , 39 , 40 where the acquisition order of spatial slices is changed in subsequent TRs, allows the efficient sampling of each TI at each spatial slice location. Efficient slice shuffling has recently been utilized to yield whole brain T1‐diffusion coverage in 13 minutes for 2.6 mm isotropic voxels. 40 An alternative approach avoids inversion recovery altogether by only varying TR to yield T1 sensitivity. 41
Further acquisition flexibility comes from applying the slice shuffling principle to the diffusion preparation, as demonstrated recently by Hutter et al 42 who, instead of acquiring all spatial slices with the same diffusion preparation in a single TR, changed the diffusion preparation on a slice level. The ZEBRA (Z‐location shuffling, multiple echoes and B‐interleaving for relaxometry‐diffusion acquisitions) technique combines three strategies 42 , 43 —multiple subsequent gradient echoes are combined with slice‐shuffling and interleaved slice‐level diffusion encoding. This allows the acquisition parameter spaces required for , , and diffusion contrast to be efficiently sampled in a significantly reduced acquisition time—a ZEBRA whole‐brain ‐‐diffusion scan with 1344 volumes at 2.5 mm isotropic resolution was demonstrated in a total scan time of 52 minutes. 44 However, there are remaining challenges and constraints for such pulse sequences: acquisition times are still too long for clinical translation, the number of subsequent echoes that can be acquired is constrained by the length of the EPI read‐out train and crusher gradients, and the acquired data cannot be easily split into shells—as each gradient direction will have different TIs due to slice shuffling—preventing the use of standard analysis techniques.
Stimulated echo (STEAM) techniques, albeit commonly employed to alter the diffusion time, also have an inherent sensitivity, 45 although 50% of the signal is lost compared to a spin echo in the conventional setup (see Figure 6C for a representative STEAM sequence). However, a recent example exploiting stimulated echo pathways, MESMERISED (Multiplexed Echo Shifted Multiband Excited and Recalled Imaging of STEAM Encoded Diffusion), uses echo‐shifting to remove the dead time in STEAM acquisitions and hence allows more efficient ‐diffusion acquisition. 46
Similarly, there are multiple techniques proposed to acquire transverse relaxometry data more efficiently. These include echo planar time‐resolved imaging (EPTI), 47 which uses novel sampling strategies to exploit correlations in k‐space and time to enable faster and relaxometry. This approach was recently demonstrated for diffusion‐relaxometry with PEPTIDE (PROPELLER EPTI with dynamic encoding). 48 Also related are high strength gradient systems, as they can enable lower TEs for the same diffusion weighting 49 although it does remain the case that lower minimum TEs are possible in 1D relaxometry experiments than combined diffusion‐relaxometry acquisitions. This may inhibit the ability of diffusion‐relaxometry to measure structures with low / relaxation times, such as myelin water. 2 magnetic resonance fingerprinting (MRF) is another relevant technique. This allows simultaneous quantification of multiple tissue properties by combining a highly accelerated acquisition that varies all relevant parameters, with dictionary matching to prior computed signal curves. The simulation of the MRF dictionary 50 with additional parameters constitutes a large computational effort, limiting diffusion MRF so far largely to variation of the b‐value to allow calculation of the ADC value, for example, Ref. [51]. However, recent efforts have started to overcome this limitation by supporting this step with deep learning techniques, resulting in combined diffusion‐relaxometry MRF 52 including variation of b‐value and gradient direction. Another proposed alternative is MR multitasking, a continuous acquisition that uses using real‐time low‐rank modeling to account for motion, relaxation and other dynamics and hence efficiently quantify MR properties, 53 and was recently demonstrated for simultaneous , , and ADC mapping. 54
2.2.2. Optimized sampling schemes
In addition to sequence efficiency, the choice of acquisition parameters is of key importance. Sampling the acquisition parameter space (ie, Figures 1 and 4) with a uniform grid is the obvious place to start. However, similar to nonuniform k‐space sampling strategies, this is not required for analysis and not necessarily the most efficient approach. For example, the Cramer‐Rao lower bound (CRLB) is an estimation theoretic tool that can be used to quantify the efficiency of different encoding schemes, 55 and CRLB‐based analysis suggests that there can be many sampling locations which provide little additional information within a uniform diffusion‐relaxometry sampling grid. 56 Samples that do not provide much information can be skipped to enable high‐quality data from a very small number of samples. This approach has enabled estimation of multidimensional spectra from as few as 12 samples for 2D diffusion‐relaxation, 56 and fitting of a two compartment ‐diffusion model from a 15 minutes whole brain scan at 2.5 mm isotropic resolution. 57
The MADCO (marginal distributions constrained optimization) framework is another example of the reduced‐sampling approach. It exploits the fact that lower dimensional distributions, or spectra, of MR parameters estimated from the data are essentially projections of the corresponding multidimensional spectra. MADCO utilizes a hierarchical encoding scheme: first, fully sampled 1D data is acquired for all sampling dimensions; then 2D (or higher) datapoints are very sparsely sampled (see Figure 4C). This, when combined with analysis techniques detailed in later sections, supports estimation of multidimensional spectra using many less data points that would typically be required—MADCO was shown to achieve an acceleration factor of up to 50. 58 , 59 MADCO also has the significant advantage that acquisition times do not dramatically increase for higher dimensions, as demonstrated for the 3D case. 60
Further acceleration techniques that reduce the number of required samples, all unified by the assumption that the data is sparse, have been proposed. While strategies like MADCO leverage the marginalized axes to constrain where peaks can appear, compressed sensing approaches—traditionally applied in k‐space—impose a presumed basis set in which the data is sparse. These methods have been used to recover 2D spectra using significantly less data in NMR 61 and MRI 62 contexts. Similarly, a framework based on PCA‐like optimization has been proposed 63 and aims to retrieve the sparse basis from the data. Finally, recent frameworks using machine learning techniques such as SARDU‐net 64 have shown excellent ability to retrieve information based on fewer samples. 44
2.3. Combined diffusion‐relaxometry and complementary advanced acquisition techniques
The standard PGSE sequence discussed above can be modified in various ways to alter the sensitivity to specific microstructural features. In the following sections, we introduce some common modifications, and discuss their influence on relaxation properties. Since the temporal profile of diffusion gradients is directly linked to the information encoded into the signal, different gradient waveforms can expose (or suppress) different aspects of diffusive motion, such as restriction, 65 flow, 66 , 67 anisotropy 68 and exchange. 69
2.3.1. Tensor‐valued diffusion encoding schemes
Arbitrary diffusion‐weighting gradient waveforms (eg, Figure 6A) can be described in terms of a symmetric second‐order b‐tensor defined as 70
| (4) |
where q is defined in Equation (2).
The tensor‐valued diffusion encoding framework describes protocols where diffusion encoding is executed in more than one direction per signal preparation to enable the measurement of b‐tensors with arbitrary traces (ie, b‐values), shapes and orientations. 70 In contrast, diffusion encoding schemes based on the Stejskal–Tanner design can only modulate the trace and orientation of the b‐tensor. Consequently, the benefit of tensor‐valued encoding is that it unlocks the “shape of the b‐tensor,” a new encoding dimension that can be modulated to control the sensitivity of the detected signal to microscopic diffusion anisotropy, as reviewed in detail in Ref. [71].
The parallels between novel diffusion gradient waveforms and combined diffusion‐relaxometry are clear—both are acquisition techniques that can precisely quantify and disentangle distinct tissue microenvironments and, excitingly, both approaches are complementary. It follows that modified gradient waveforms are necessary in order to maximize the utility of combined diffusion‐relaxometry acquisitions (and vice versa). For example, a sequence combining tensor‐valued diffusion encoding with multiple TR weighting was introduced as a means to establish correlations between and diffusion properties and thus characterize complex fiber orientation layouts. 41 However, the combined relaxation and b‐tensor protocols developed thus far have utilized a rather straightforward design, particularly from the relaxation encoding perspective, with suboptimal echo and recovery time samplings. In this regard, there is potential for improvement through combining arbitrary b‐tensor shape with acquisition strategies that encode relaxation information more efficiently. We also note that, particularly when combined with relaxometry, the additional degrees of freedom describing an advanced diffusion gradient waveform put further emphasis on efficiently selecting the acquisition parameters.
2.3.2. Varying diffusion time
As well as modifying diffusion gradient shape, we can also modify the time scale at which diffusion is observed. In the conventional PGSE approach, , the spacing between diffusion encoding pulsed gradients, can be varied for this purpose. Importantly, is closely linked to a change in TE as described above, due to the intertwining between the structure used for the spin echo and the placement of the diffusion gradients after the excitation and after the refocusing pulse.
There are a number of advanced acquisition techniques which probe the diffusion time. Higher gradients with faster slew rates can be used to allow MRI systems to access short and may be critical for measurement of white matter features, such as the axonal diameter. 72 Alternative sequences, such as the oscillating gradient spin echo (OGSE) are associated with significantly shorter effective diffusion times. 73 For structures with larger radii (prostate lumen, 74 muscle tissue 75 ), PGSE cannot offer long enough diffusion times, due to signal loss from ‐weighting.
The aforementioned STEAM diffusion preparations (Figure 6C) can maintain high SNR for diffusion times above 50 ms. Although STEAM conflates and diffusion time dependencies, this may be overcome with a twice‐refocused STEAM preparation. 76 In combination with the appropriate models (as discussed in later sections) this approach can investigate the correlations between the time‐dependence of diffusivity and relaxation times.
Modifying the diffusion time yields particular sensitivity to porous structures, such as cells. 77 Such structures also affect relaxation properties, for example, differences in magnetic susceptibility between the pores and surrounding material cause spatial variations in the field. 78 Song et al demonstrated that this can be exploited using sequences that vary the TE and diffusion time to measure the decay due to diffusion in the internal field (DDIF), and hence determine the characteristic length‐scales of a porous structure. 79 Alvarez et al developed a related sequence sensitive to diffusion and relaxation properties to measure internal gradient distribution tensors, which probe local magnetic susceptibility properties. 80
These experiments show that incorporating diffusion time dependence into combined diffusion‐relaxometry models is an exciting prospect for quantifying microscopic structures. Alongside magnetic susceptibility, future acquisition methods could also probe and relaxation. These methods can potentially detect compartments with small (1 m) and impermeable features, which are not experimentally accesible with diffusion‐only sequences due to the limited diffusion times available on clinical MR scanners (20‐50 ms).
2.3.3. field strength
Higher B0 field strength offers significant advantages for combined diffusion‐relaxometry, including increased SNR, contrast, and spatial and temporal resolution. 81 However, relaxation times vary strongly with B0, and the specific relationship often depends on the tissue type, such as with in the brain. 82 Since they are affected by relaxation, diffusion metrics may also depend on , 83 particularly at ultra‐high field (ie, 7T and above). 84 , 85 In some cases, it may be important to account for these dependencies in the clinic, such as when setting cut‐off values between healthy and diseased tissue. There are also likely opportunities to exploit this dependency—in some applications distinguishing between tissue types may be easier at a certain . Ultra‐high‐field MRI seems most promising for revealing and exploiting dependencies of diffusion metrics. In particular, the reported increase of micro‐FA at ultra‐high‐field deserves further investigation. 85 , 86 Another promising avenue is the B0 dependency of the intravoxel incoherent motion (IVIM) signal curve at ultra‐high‐field, 87 , 88 where the weight of the arterial pool should further increase, combining this with additional relaxometry information can potentially make “arterial‐pool‐weighted” IVIM feasible. 89
2.3.4. Exchange
During acquisitions, some spins will inevitably move between microenvironments that have different MR properties. Traditional diffusion‐relaxometry acquisition techniques do not explicitly account for exchange or provide a means to modulate its effects, but exchange will affect the signal and hence derived diffusion and relaxation parameters. A class of related and relevant techniques can detect exchange between compartments using sequences incorporating variable mixing times. These include diffusion exchange spectroscopy (DEXSY) 69 and filter exchange imaging (FEXI) 90 , 91 for diffusion, and relaxation exchange spectroscopy (REXSY) for 92 and 93 relaxation. Such techniques can only measure the apparent exchange rate, as differences in relaxation times (for diffusion exchange) or diffusivities (for relaxation exchange) between compartments will affect the measured exchange rate. 94 Techniques that acquire combined diffusion‐relaxometry data with variable mixing times have the potential to account for these differences, and hence measure exchange rates more closely related to underlying values. However, naively merging these techniques would lead to prohibitively long acquisition times, so novel pulse sequences, such as methods that rapidly measure exchange, 95 are desirable.
2.3.5. Final thoughts
Further improvements will come from combining specific multidimensional data acquisition techniques with important parallel developments in MRI acceleration. These include multiband imaging 96 , 97 and compressed sensing approaches 62 , 98 among others. In summary, the efficient acquisition of combined diffusion‐relaxometry data is an exciting and ongoing field which will continue to be driven and influenced by a multitude of developments in both MRI and NMR. Such accelerated acquisition techniques are crucial to provide data for the appropriate analysis method while avoiding prohibitively long acquisition times.
3. ANALYSIS
3.1. Background and motivation
The basic premise of a combined diffusion‐relaxometry MRI experiment is that we are interested in understanding and quantifying the different microenvironments that coexist within an imaging voxel, under the assumption that these different microenvironments have distinctive diffusion‐relaxation characteristics that will allow them to be clearly discriminated from one another. The main goal of data analysis is therefore to identify and quantify different microenvironments. However, practical limits on the spatial resolution of MRI often mean that the signal from a single voxel will represent a partial‐volume mixture of multiple distinct microenvironments.
In the absence of exchange, the measured data can be modeled as the linear superposition of the signals that would be observed from each individual component, and data analysis generally necessitates some form of multicomponent modeling so that the contributions of different microenvironments can be disentangled. One of the main advantages of combined diffusion‐relaxometry is that the resulting multicomponent data analysis can be shown to be easier, both empirically 4 , 99 , 100 and theoretically, 55 than multicomponent diffusion analysis (without relaxation) or multicomponent relaxometry (without diffusion). Multiexponential fitting for a single parameter such as diffusivity or relaxation time is highly ill‐posed. However, joint exponential fitting for multiple parameters in 2D (or higher) is better conditioned than the corresponding 1D fits.
Separating a mixture of superposed signal components is a classical inverse problem that is not only unique to diffusion and relaxation MRI, but also appears commonly in other applications like remote sensing, spectroscopy of all kinds (including NMR), functional imaging (eg, fMRI, MEG, EEG, and PET), and dynamic contrast‐enhanced imaging (eg, MRI and PET). Due to its ubiquity, this kind of problem has been widely investigated over many decades, and a wide variety of analysis tools have been developed. Due to space limitations, the description below focuses on some of the most common unmixing approaches that are applicable to combined diffusion‐relaxometry, and is not intended to be comprehensive.
For the sake of concreteness and without loss of generality, our description will assume a simple MRI experiment that combines 1D diffusion encoding with relaxometry encoding, where each data sample is associated with one diffusion‐encoding parameter b, one relaxation‐encoding parameter TE, and one spatial position (ie, voxel) x. If we assume that M components are present in a voxel, the measured data d(x, b, TE) can be modeled (in the absence of noise) as
| (5) |
where is a function describing the contrast variations of the mth component, and is a spatial map describing how much of the mth component is present within each image voxel.
Different data analysis methods can be distinguished from each other based on the different modeling assumptions that they make. Some of the most general unmixing approaches (including nonnegative matrix factorization, 101 , 102 independent component analysis, 103 and low‐rank tensor decomposition 104 ) are sometimes called blind source separation methods because they make minimal prior assumptions about and . Instead, these approaches attempt to learn all the model parameters from the data, based only on the assumptions that M is relatively small and that the components are simple in an appropriate way. Although such methods are straightforward to implement, can be effective at automatically decomposing the data into seemingly‐meaningful components, and have been successfully used to analyze combined diffusion‐relaxometry data, 105 , 106 , 107 these approaches have some potential disadvantages. These include that they may decompose data into ambiguous constituent parts that are not biologically meaningful, and the decomposition may not be stable and/or reproducible. Such methods may also overlook short compartments due to the longer TEs required to explore the ‐diffusion space.
A commonly used stronger model assumes that the are not some arbitrary/unknown functions, but can instead be described concretely. For example, one common assumption, which we make from here onwards, is that each component is associated with independent monoexponential diffusion and relaxation decays such that
| (6) |
where is an apparent diffusion coefficient and is a ‐relaxation parameter. (Note that analogous assumptions can be applied for any experiment, eg, for T1 inversion recovery we have
| (7) |
We include this equation here for reference, and note that the approaches discussed in the following sections equally apply to any signal equation.) Given this assumption, we now explore three main approaches for analyzing combined diffusion‐relaxometry data:
Continuum modeling
Cumulant expansion
Parsimonious modeling using strong biophysical assumptions
In the next section, we discuss these three approaches in turn.
3.2. Current state‐of‐the‐art: Continuum modeling
Thus far, continuum modeling has been the preeminent technique for analyzing combined diffusion‐relaxometry data. Under the assumption of monoexponential decays, it is common to rewrite the data model from Equation (5) as an integral equation of the first kind 8 according to
| (8) |
where is the diffusion‐relaxation correlation spectrum for a voxel in spatial location x. The choice to use an infinite dimensional integral equation rather than the finite discrete sum from Equation (3) is consistent with earlier methods for 1D multicomponent relaxometry, 108 , 109 and reflects the practical reality that we often do not have prior information about the number of components M. An alternative approach, discussed later, is to incorporate prior assumptions about tissue microstructure, enabling techniques more specific to distinct tissue types, at the cost of generality. In addition, in complicated heterogeneous tissues, the number of distinct decay parameters that are present within a voxel might be very large and effectively infinite. In the literature, this integral equation has been viewed as a special case of the Laplace transform, causing the associated inverse problem to sometimes be called an inverse Laplace transform (ILT).
Since the correlation spectrum is infinite dimensional and practical MRI experiments only acquire a finite number of measurements, we should not expect the ILT to have a unique solution. In MRI and NMR spectroscopy, these problems are usually resolved by choosing the “simplest” solution, ie, the unique solution that fulfills minimum‐norm least‐squares (MNLS) criteria. 110 However, while it is straightforward to formulate the ILT within the framework of infinite dimensional Hilbert spaces and derive simple analytic expressions for the unique MNLS solution, 111 the ILT solutions obtained in this manner are usually not very satisfying or useful. This occurs because, unlike the case for MRI and NMR spectroscopy where MNLS solutions are stable and interpretable, the presence of exponential decays in Equation (3) can make the inverse problem severely ill‐posed and highly unstable. As a result, additional assumptions must usually be imposed to get reasonable ILT solutions.
There are several constraints that have been proposed to help achieve robust and interpretable correlation spectra. A widespread approach, originating in some of the early papers on diffusion‐relaxometry, 8 is to assume that the correlation spectrum should be everywhere nonnegative. This assumption can be motivated by physics and is inherited from earlier work on 1D relaxometry. 108 , 109 To numerically perform the inversion and estimate the spectrum, Equation (8) is discretized on grids of predefined ranges with and nonnegative components in the D and dimensions, respectively. The maximum and minimum grid values are chosen such that the solution is physically realistic by excluding negative fractions. The resulting matrix equation is
| (9) |
where vectors f and d are discretized versions of and d(x, b, TE), respectively, and the matrix A is a corresponding discretization of the integral equation from Equation (8). A is by , f has length , and d has length , the total number of MR‐encodings in the experiment. This leads to a nonnegative least‐squares inverse problem formulation that can be written as
| (10) |
using the standard 2‐norm. This constrained optimization problem does not have an analytic solution, but can still be solved iteratively using efficient optimization algorithms. 112 Importantly, nonnegativity constraints have good theoretical characteristics, and existence of a unique nonnegative solution can sometimes be guaranteed in inverse problems that would otherwise have infinitely many solutions. 113
In addition to nonnegativity, regularization constraints can also be used to stabilize the inverse problem, leading to the formulation
| (11) |
where R(.) is a regularization penalty function that is designed to prefer solutions that have certain desirable characteristics and is a user‐selected regularization parameter. Common choices include:
Regularization of the spectrum from each voxel with , where H is an appropriately chosen voxelwise constraint matrix, p is a user‐selected parameter, and denotes the standard p‐norm. Common choices of p include (leading to standard Tikhonov regularization and guarantees under specific conditions the existence of a unique solution to Equation (11) 8 and (leading to standard L1 regularization, which often results in sparser solutions). Depending on the choice of H, this penalty can be used to enforce the constraints that the reconstructed correlation spectrum for each voxel should have small signal energy or should have relatively smooth spectral variations. This kind of regularization is popular in combined diffusion‐relaxometry, 8 , 58 , 114 and also 1D relaxometry. 108 , 109
Spatial regularization that encourages the reconstructed correlation spectrum to be spatially smooth, 100 suited to mapping large structures over a whole image. Unlike the previous regularization penalties (which operate voxelwise and can be used to reconstruct the spectra for each voxel separately), the use of spatial regularization couples the estimation of correlation spectra from neighboring voxels, and necessitates an imaging acquisition. This approach has been used in combined diffusion‐relaxometry, 100 but was also used in earlier 1D multicomponent exponential modeling applications. 115 , 116 Estimation theoretic analysis can be used to show that spatial smoothness constraints can theoretically reduce the ill‐posedness of the ILT by orders of magnitude in both 1D 116 and higher‐dimensional 55 settings.
Data‐driven regularization, where a fixed number of correlation spectra are assumed within the image. 117 , 118 This approach is related to the previously mentioned blind source‐separation methods 105 , 106 , 107 and seeks a lower‐dimensional spectral representation of the image that is supported by the data, effectively regularizing the inversion by sharing information across voxels. This approach is appropriate when seeking to discover prominent microstructural features, at the expense of estimating spectra in every voxel.
We note that selection of the optimal regularization tuning parameter is an open research question, with many approaches and strategies suggested and tested over the course more than three decades. Notable methods include the generalized cross‐validation (GCV) method, 119 the Butler‐Reeds‐Dawson (BRD) algorithm and the L‐curve method 112 , 120 and theoretical regularization parameter selection based on the desired spatial‐spectral resolution characteristics. 55 , 121 , 122 A thorough review of some of these methods, along with alternative inversion methods, can be found here. 123
In addition to nonnegativity and regularization, other potentially useful constraints include:
Enforcing consistency between the multidimensional correlation spectrum derived from combined diffusion‐relaxometry data and the 1D diffusion and relaxation spectra derived from 1D ILTs of subsets of the data. 58 For example, the MADCO technique—as discussed in the Acquisition section—exploits the fact that lower dimensional spectra are projections of higher dimensional spectra, for example,
| (12) |
to enforce constraints on the multidimensional spectrum estimation.
3.2.1. Mapping voxelwise spectra
Unlike modeling methods that calculate a single, or small number of, values per voxel (as discussed later), the continuum modeling approach we describe calculates the spectrum, f, in each voxel. The typical way to derive meaningful maps from such voxelwise spectra is known as spectral integration. 60 , 100 , 124 In summary, spectral intergration comprises first manually identifying prominent regions of the spectrum—typically by examining a spectrum derived from the signal averaged over a large representative ROI. The proportion of each voxelwise spectra that lies within each of these prominent regions is then calculated (hence the name spectral integration), yielding scalar indices often termed apparent spectral volume fractions. The data‐driven regularization methods described above 117 , 118 provide an alternative approach for deriving maps.
In contrast to continuum modeling, a useful and widely used constraint in single‐contrast diffusion and relaxation experiments is to assume that the number of components M in Equation (1) is known in advance and is very small, while also assuming simple/parsimonious parametric models for the signal observed from each component. This approach has been used with combined diffusion‐relaxometry data, 26 , 74 , 125 and uses modeling assumptions that are substantially more restrictive than those previously discussed. However, when the modeling assumptions are accurate, they lead to a simpler inverse problem that requires substantially lower SNR data to solve than the more general formulations described previously (calculating multidimensional spectra, particularly for single voxels, requires very high SNR). We discuss two examples of this approach in the next sections.
3.3. Current state‐of‐the‐art: Signal representations
As previously discussed, a major weakness of continuum modeling is that the inversion of the Fredholm integral or Laplace transform is ill‐posed. 126 Naturally, there are representations that do not require an ILT. The most commonly used representations in diffusion MRI are diffusion tensor imaging (DTI) 127 and diffusion kurtosis imaging (DKI) 128 (although further techniques have been demonstrated, such as a decomposition in the basis of harmonic oscillator eigenmodes, 129 spherical harmonics, or extended phase graph (EPG) formulation. 130 DTI and DKI correspond to truncating the cumulant expansion, 131 which is the Taylor expansion of for small diffusion wave vector q, at the 2nd order and 4th order, respectively. Ning et al 132 adapted the cumulant expansion approach to diffusion‐relaxation experiments, hence calculating the joint moments of relaxation rate and diffusivity. This allowed the calculation of several novel microstructural metrics, including estimating diffusion properties independently from TE without solving an ILT.
Signal representations are a potential starting point to determine the number of degrees of freedom that a future microstructural model could have. For instance, the observation of bi‐exponential decay in S(b) or a nonzero kurtosis necessarily indicates that the measurement is sensitive to non‐Gaussian diffusion. 133 The observation of diffusion time dependence suggests that this non‐Gaussian diffusion occurs in one or more compartments. 134 Further study of mathematical functional forms would reveal even more information about the underlying structure.
3.4. Current state‐of‐the‐art: Parsimonious modeling using strong biophysical assumptions
Biophysical modeling moves to a biophysical description of the signal, where we assume a fixed number of tissue compartments, essentially trying to identify the relevant degrees of freedom while discarding many others. 135 In our notation we can write the signal as a sum over multiple tissue compartments
| (13) |
where we change the index from Equation (5) to reflect the fact that we assume a known number of tissue compartments, N. The choice of N and the form of each is biophysically motivated based on our knowledge of the tissue of interest (see Ref. [136] for a summary of typical compartments).
Multicompartment biophysical models are prevalent in diffusion MRI, particularly in the brain. 137 , 138 However, a significant limitation is that due to model identifiability issues, it is common to fix certain parameters to reasonable values. 139 , 140 Although this approach can produce high precision biomarkers, it also introduces biases. 141 , 142 Accounting for both the diffusion and relaxometry properties of tissue compartments during acquisition and analysis can avoid fixing parameters, as has been demonstrated in the brain 57 and prostate. 143 When combined with rich enough data and robust parameter estimation techniques, biophysical models have the potential to estimate highly specific physical features—such as permeabilities, compartment sizes, orientation dispersion, packing correlation length scales.
Metrics obtained from multicompartment diffusion MRI models can be directly compared to histology, since it can be reasonably assumed that the measured tissue microstructure remains reasonably similar in the ex‐vivo histology (with the caveats of membrane integrity and shrinkage due to tissue preparation, 144 and the geometric mismatch between 3D MRI and 2D histology 145 ). However, the extent to which this generalizes to diffusion‐relaxometry is a question attracting significant research interest.
3.5. Future promises
As can be seen, there are many different options available for the analysis of combined diffusion‐relaxometry data that have complementary strengths, with different methods being more or less powerful in different application contexts. For illustration, the effects of using different kinds of estimation constraints are shown using synthetic diffusion‐relaxometry data in Figure 7. This represents a specific toy ‐diffusion system where the ground truth for each microenvironment is represented by a simple Gaussian distribution. In reality, microenvironments will have more complex, and potentially overlapping, shapes. In cases where the ‐diffusion approach cannot disentangle the relevant microenvironments, the system could be explored with higher‐dimensional acquisition approaches detailed in this review, such as ‐‐diffusion, incorporating diffusion time dependence, etc.
FIGURE 7.

Illustration of estimating 2D diffusion‐relaxation correlation spectra using different kinds of constraints. A, Simulated ground truth spectrum from a single voxel, comprised of three spectral peaks with Gaussian lineshapes. (The simulation is identical to that described in previous work, 100 and we omit the details). B, Simulated noisy combined diffusion‐relaxometry data from one voxel (following, 100 the SNR for the highest‐SNR image was 200). C‐J, Reconstruction results using different kinds of constraints: C, MNLS; D, nonnegative least‐squares using Equation (4); E‐H, regularized solutions using Equation (5), including (E) voxelwise Tikhonov regularization (H is an identity matrix), (F) voxelwise Tikhonov regularization (H computes a finite‐difference approximation of second‐derivatives along the vertical and horizontal dimensions of the spectrum), (G) voxelwise L1‐norm regularization (H is an identity matrix), and (H) spatially regularized reconstruction (assuming the spatial distribution described in previous work 100 ). We also show fits obtained using Equation (1) with known M and the parametric model from Equation (2) for each component: (I) and (J) . Note that these spectra can look very different from one another even though they are all similarly consistent with the measured data (less than 0.4% data error in all cases). This reflects the severe ill‐posedness of the problem and the need to choose constraints carefully. As can be seen, the MNLS reconstruction completely fails to capture the true structure of the data, while the remaining constrained reconstructions have varying levels of correspondence with the ground truth spectrum. Note that it is very hard for any of these methods to estimate the lineshapes of the spectral peaks accurately (some have much sharper peaks while others have much broader peaks), and that aside from (H), the rest of the reconstructions have spectral peaks in the wrong locations and potentially incorrect spectral peak integrals. Despite these mismatches with the ground truth, many of these estimation methods have been shown to produce consistent spectral decompositions of real data that can serve as useful biomarkers for different microstructural characteristics
While the approaches described above can be quite powerful, it should be emphasized that this is still a developing field, and there are many opportunities to make combined diffusion‐relaxometry data analysis even better. For example, recent literature has demonstrated that it is possible to quantify the estimation uncertainty associated with combined diffusion‐relaxometry experiments using Cramer‐Rao theory, 55 , 57 , 146 similar to approaches used in other quantitative MRI applications. 147 , 148 , 149 While such uncertainty quantification is clearly useful for interpreting experimental results, this kind of information is also important because it can be used to design more efficient data acquisition schemes. 57 , 146 As further examples, it has recently been shown that a nonnegativity‐constrained ILT formulation can be combined together with statistical resampling methods to enable very high‐dimensional correlation spectrum estimation without additional regularization constraints. 150 We also note that we assumed time‐independent diffusion with no intra‐compartmental kurtosis in Equation (6). Mean kurtosis can be estimated by adding the relevant terms to Equation (6), 128 or different sources of kurtosis could be decoupled by combining with new approaches such as correlation tensor imaging. 151 There are also potentially many other creative approaches that are yet to be developed. Machine learning is a promising avenue; this will likely focus on methods that learn from simulations (eg, Refs. [152, 153]) or unsupervised learning (eg, Ref. [118]), as ground truth information on tissue microstructure is not typically obtainable.
4. APPLICATIONS
4.1. Brain
4.1.1. Background and motivation: Brain applications
There are a wide range of diffusion MRI brain microstructure imaging techniques, many of which show promise in the clinic. 154 Compared to relaxometry and more conventional DTI, the main advantage offered by these techniques is a more direct link between quantitative parametric maps and the underlying tissue microstructure. For example, simple diffusion models have shown remarkable value for imaging acute ischemic stroke, 155 but the exact microstructural changes driving the contrast in ADC and DTI maps remain unclear. Moreover, unaccounted for changes in relaxation properties of brain tissue due to pathological conditions (eg, increase of relaxation time in ischemic lesion 156 ) may bias parameter estimates and lead to misleading or wrong interpretation of the ongoing pathological mechanisms. Incorporating relaxation into diffusion MRI techniques can lead to more specific measures of brain microstructure, leading to improved understanding of disease and new clinical applications.
4.1.2. Current state‐of‐the‐art: Brain applications
There are a wide range of brain microstructure models, spanning signal representations and biophysical models. 138 The most common brain microstructure model encompasses a large set of white matter models consisting of Gaussian compartments, is the so‐called “Standard model”. 135 This describes the measured dMRI signal as the convolution of a fiber orientation distribution function (ODF), , and a response kernel, from a straight fiber oriented along , where is the unit gradient direction:
| (14) |
while the Standard model does not fix individual parameters, 26 , 57 , 157 there are numerous particular cases of this model that involve prior assumptions or bounds on the scalar parameters of the kernel (diffusivities, volume fractions), inclusion of free water compartment, and assumptions relating to the functional form of the ODF, most generally represented by an expansion of spherical harmonic coefficients. Fixing compartment diffusion coefficients and the ODF shape introduces an unknown a priori bias 141 on the remaining parameters, which is particularly problematic in regions with diffusivity changes and no change in the volume fractions of tissue compartments. 158 On the other hand, without any priors or constraints the standard model suffers from degeneracy, that is, multiple combinations of the model parameters can provide the same indistinguishable signals. 141 Simultaneously measuring relaxation and diffusion properties can help address this issue and relax a number of the priors and constraints. This was recently demonstrated in the brain, where an optimized 5D protocol comprising multiple TE and b‐tensor shapes was shown to enable the fit of a minimally constrained white matter model. 57
Both and have been utilized to resolve inherent ambiguities in diffusion measurements. De Santis et al used combined ‐diffusion acquisition to resolve crossing fiber populations 31 , 125 (see Figure 8A). More recently, Leppert et al investigated tract specific mapping. 40 Veraart et al used combined ‐diffusion to improve separation of presumed intra‐ and extra‐axonal compartments 26 (Figure 8B), and a similar strategy was later adopted within the NODDI model framework 29 (Figure 8C). An interesting different multidimensional approach has been recently proposed to map intra‐axonal values using the extra‐axonal water suppression provided by high diffusion‐weighting. Using a diffusion‐filtered asymmetric spin echo, Kleban et al jointly estimated diffusion and susceptibility effects in major brain white matter tracts, showing that intra‐axonal is lower in the corticospinal tract than in corpus callosum and cingulum. 159
FIGURE 8.

Examples of combined diffusion‐relaxometry in brain and spinal cord applications. A, ‐diffusion—direction encoded color map maps of tissue in a fiber crossing phantom from Ref. [125]. B, ‐diffusion—TE‐dependent diffusion imaging (TEdDI) maps for in vivo brain from Ref. [26]. Top row: intra‐ and extra‐axonal , and mean ; bottom row: diffusivity parameters. C, ‐diffusion—Multi‐TE NODDI parameter maps for in vivo brain from Ref. [29]. Parameters from left to right are: intraneurite fraction, free water volume fraction, and intra‐ and extra‐neurite values. D, ‐diffusion with b‐tensor encoding—5D diffusion‐relaxation distribution for a selected voxel of in vivo brain scan from Ref. [161]. E, ‐diffusion—spatially averaged ‐D distributions and spectral volume fraction maps in control and injured spinal cord samples from Ref. [100]. F, ‐ ‐diffusion—spectral signature of microscopic lesions in the human corpus callosum (top: injured, bottom: control). Right panel shows corresponding injury biomarker images, side‐by‐side with histology. 164 Subfigures A, C, and D are reproduced under Creative Commons licenses; subfigures B, E, and F are used with permission
Moving away from biophysical model‐based approaches, more data‐driven methods to multidimensional diffusion‐relaxation analysis of the brain tissue have been recently proposed, carrying very exciting perspectives (for a more comprehensive review of the most recent methods, we refer the reader to Refs. [55, 160]). For example, de Almeida Martins et al proposed an MRI framework to quantify the microscopic heterogeneity of the living human brain by spatially resolving five‐dimensional relaxation‐diffusion distributions from combined 5D relaxation acquisition, 161 , 162 with experiments on a healthy volunteer demonstrating that the retrieved 5D ‐D distributions can resolve distinct microscopic tissue environments within the living human brain (see Figure 8D). It is worthwhile to mention here that these data‐driven multidimensional diffusion‐relaxation approaches can be extremely useful to characterize parts of the central nervous system other than the brain, such as the spinal cord 100 , 163 (eg, Figure 8E). Recently, in a comprehensive postmortem multidimensional MRI and histopathology combined study that included multiple human subjects ranging in axonal injury severities, Benjamini et al showed that axonal injury has a multidimensional spectral signature, 164 which can be used to generate injury biomarker images that closely follow amyloid precursor protein (APP, axonal injury marker) histopathology, with strong correlation (see Figure 8F).
Of course, more general data‐driven multidimensional approaches come together with new challenges, 160 for example, how to improve the robustness (in terms of accuracy and precision) of the statistical descriptors that are estimated directly from the data, 165 or how to summarize and visualize in an effective and simple way the large amount of information that can be retrieved. 161 , 166 A particularly interesting model‐free application of multidimensional relaxation‐diffusion approach to the same problem previously tackled by De Santis et al using a model‐based approach has been recently proposed for the nonparametric estimation of fiber‐specific diffusion‐ features. 41 However, for spinal cord, the application of such methods to in vivo clinical settings is hampered by challenges involving low SNR and high motion corruption. Nevertheless, recently proposed strategies to reduce the number of samples 167 may provide useful tools to address some of these limitations, opening exciting possibilities for in vivo clinical applications.
4.1.3. Future promise: Brain applications
Combined diffusion‐relaxometry MRI is quickly evolving and moving into more and more clinically feasible applications. 160 As discussed in the previous sections, this new approach comes with new challenges, but several recent works have already highlighted interesting avenues toward the resolution of some of them. 161 , 165 , 166 Notwithstanding the current challenges, the possibilities offered by combined diffusion‐relaxometry for brain microstructure characterization are exciting, especially for pathological conditions. In particular, the rich data available from combined diffusion‐relaxometry lend itself to data‐driven analysis (eg, continuum modeling), which in principle can better generalize across diseases than typical brain microstructure models that make a priori assumptions about tissue structure.
In addition, new acquisition techniques have the ability to probe very short TEs, 49 paving the way toward very interesting applications, for example, the quantification of combined diffusion‐relaxation properties of myelin, whose is very short (typically <10 ms), raising in turns new challenges for the acquisition. Furthermore, advanced MRI hardware, utilizing stronger diffusion gradients such as those available in the CONNECTOM scanner, 49 may allow the exploration of an additional dimension, the diffusion time, that may carry unique information about microstructural features such as membrane permeability, spatial disorder and more (see further the earlier Diffusion Time Section). Promising future applications may involve the multidimensional exploration of sub‐cellular membrane structures 168 or the combined diffusion‐relaxation of brain metabolites. 169 , 170
4.2. Prostate
4.2.1. Background and motivation: Prostate applications
The multiparametric MRI (mp‐MRI) examination for diagnosis, staging, and risk stratification of prostate cancer relies on ‐weighted and diffusion MRI images. 171 The synergy between modalities is abundantly clear to clinicians; moreover, decoupling signals from multiple tissue compartments promises biomarkers with greater diagnostic value. 172 This is a monumental modeling challenge, as the prostate is heterogeneous at multiple length scales, where each voxel presents unknown mixtures of stroma (20 m), epithelium (10 m), vasculature (10 m), and lumen (100 m). The observation 173 , 174 of a nonmonotonic diffusion time dependence, necessarily means that at least 1 compartment falls within the short or long‐time diffusion limits. For this reason, it is incredibly difficult to estimate compartment fractions via diffusion, as this would require definition of higher order cumulants of the Taylor expansion of for each compartment. A potential avenue to resolve the model complexity of diffusion is to rely on relaxometry for compartment estimation, where separation of signal into cellular (stromal + epithelium) and luminal compartments using alone has been performed reliably in prostate tissue as early as 1987 175 and has been reproduced on numerous occasions. 145 , 176 , 177 , 178 The practical challenge remains on how to describe the prostate signal through diffusion‐relaxometry, without biasing parameter estimates or over complicating the model.
4.2.2. Current state‐of‐the‐art: Prostate applications
The model of Chatterjee et al 179 incorporates relaxometry and hypothesizes that prostate diffusion can be described as a sum of 3 Gaussian compartments: epithelium, stroma, and lumen (see Figure 9A). The model of Lemberskiy et al 74 circumvents the issue of higher order cumulants, by modeling DWI at low‐b (where diffusion remains sufficiently Gaussian) and performing the assumed compartmental decomposition of the diffusion tensor entirely through , thereby preserving the full diffusion tensor for each compartment (see Figure 9B). This approach allowed for the functional validation of diffusion time dependence within putative cellular (epithelium and stroma), which was found to agree with random permeable barriers, and luminal compartments were found to agree with the short‐time limit. Though high b‐values were not considered, exchange/permeability could be measured for the cellular compartment via a random permeable barrier model. 75 , 180 This approach did not account for vascularity, which while small, may be a critical microstructural biomarker. VERDICT does account for vascularity, 137 it originally did not include relaxometry but was recently expanded to account for and dependence. 143 Zhang et al recently applied continuum modeling with spatial regularization to quantify epithelium, stroma, and lumen compartments, and showed that these correlated well with results from histopathlogy 181 (see Figure 9C).
FIGURE 9.

Examples of combined diffusion‐relaxometry in the prostate. A, ‐diffusion—MR‐derived parameter maps and predicted cancer map, and corresponding histologic maps for prostate with Gleason 3+3 cancers. 179 B, ‐diffusion with varying diffusion time—parameter maps for healthy prostate from [74]. C, ‐diffusion—and corresponding histology for prostate with Gleason 4+3 cancer from [181]. Subfigures A and C are used with permission, and subfigure B under a Creative Commons license
4.2.3. Future promise: Prostate applications
The complexity of individual compartments and the potential for diffusion time‐dependence present a challenge to modeling prostate microstructure. Unlike the mapping of axonal diameter in the brain, 72 MR gradient systems are sufficiently strong to measure various microstructural features. With the hardware challenge resolved, the primary challenge is to determine the optimal combined diffusion‐relaxometry acquisition and its subsequent interpretation through appropriate biophysical modeling.
4.3. Placenta
4.3.1. Background and motivation: Placenta applications
Placental MRI is emerging as a sensitive tool which can supplement antenatal ultrasound in the clinic. There are a number of MR contrasts which show promise as markers of placenta‐related pregnancy complications. relaxometry—which relates to oxygenation levels—reflects placental dysfunction in fetal growth restriction (FGR) 182 , 183 and pre‐eclampsia. 184 Diffusion MRI is also sensitive to these changes, with ADC 185 and IVIM‐derived perfusion fraction 186 , 187 decreased in FGR cases.
The placenta is unique in that it contains two distinct and nonmixing circulations: fetal and maternal blood. Fetal blood circulates in vasculature within convoluted villous tree structures. On the other hand, maternal blood, uniquely, does not flow within vasculature, but resides in intervillous space surrounding fetal villous trees. This one‐of‐a‐kind structure is a leading motivation behind applying combined diffusion‐relaxometry in the placenta. Melbourne et al make the following initial speculations about placental tissue environments 188 : fetal and maternal blood have similar relaxation times, and hence their relative fractions cannot be determined with relaxometry; However, they can likely be differentiated by diffusivity properties, since fetal blood perfusing in vasculature travels much faster than maternal blood. Moreover, maternal blood cannot be separated from water in tissue by diffusivity, but can be separated by relaxation time. It follows that diffusion‐relaxometry is necessary to adequately disentangle these respective compartments.
4.3.2. Current state‐of‐the‐art: Placenta applications
The first example of combined diffusion‐relaxometry in the placenta was presented by Melbourne et al 188 (see Figure 10A). The DECIDE model contains three putative tissue compartments: fetal blood (characterized by high diffusivity and long ), maternal blood (low diffusivity, long ), and tissue (low diffusivity, short ). This enables estimation of fetal and maternal blood values, and the maternal to fetal blood volume ratio. These are potential biomarkers of placental dysfunction that are unavailable through single contrast measurements.
FIGURE 10.

Examples of combined diffusion‐relaxometry in the placenta. A, ‐diffusion—parameter maps from DECIDE model fit on healthy placenta. 188 B, ‐diffusion—spatially averaged ‐D spectra and spectral volume fraction maps in healthy and pre‐eclamptic placentas. 189 Both subfigures are used under a Creative Commons license
Slator et al 189 simultaneously probed and diffusivity in the placenta (see Figure 10B). The ‐ADC spectrum was calculated with an ILT, and showed multiple separated peaks potentially reflecting multiple tissue microenvironments. Encouragingly, although the sample size is small, ‐ADC spectra clearly separated normal and dysfunction placentas, demonstrating the potential of this approach to quantify and predict pregnancy complications.
4.3.3. Future promise: Placenta applications
A promising avenue for future work is to account for anisotropy in analysis techniques. It has been shown that models incorporating anisotropy—in both perfusion and diffusion regimes—explain the placental dMRI signal well 190 . Incorporating this into diffusion‐relaxometry analysis techniques has the potential to increase sensitivity to pregnancy complications since, for example, fetal vascular tree morphology is significantly altered in fetal growth restriction 191 . The ultimate aim is to develop targeted imaging tools that can supplement antenatal ultrasound in the clinic. Such a tool should predict and precisely diagnose pregnancy complications, and hence be beneficial for potential future therapies. The extent to which such a tool is sensitive to the fetal to maternal blood ratio, oxygenation levels, and placental microstructure is an open question.
5. CONCLUSIONS
All diffusion measurements are influenced to some extent by relaxometry properties. Combined diffusion‐relaxometry measurements can leverage this intrinsic link to reveal information on tissue composition and microstructure that is inaccessible through typical separate diffusion and relaxometry experiments. Despite the challenges, such as long scanning times and multidimensional datasets, continually improving acquisition and analysis techniques can unlock the full potential of combined diffusion‐relaxometry MRI, enabling the next generation of microstructure imaging techniques.
CONFLICT OF INTEREST
No conflicts of interest.
ACKNOWLEDGEMENTS
The authors thank the ISMRM for supporting their combined diffusion‐relaxometry microstructure imaging member‐initiated symposium at the ISMRM 2019 Annual Meeting & Exhibition.
Slator J, Palombo M, Miller KL, Westin C‐F, Laun F, Kim D, Haldar JP, Benjamini D, Lemberskiy G, de Almeida Martins JP, Hutter J Combined diffusion‐relaxometry microstructure imaging: Current status and future prospects. Magn Reson Med. 2021;86:2987–3011. 10.1002/mrm.28963
Funding information
This work was supported by the NIH Human Placenta Project grant 1U01HD087202‐01 (Placenta Imaging Project [PiP]); the Wellcome EPSRC Centre for Medical Engineering at Kings College London (WT 203148/Z/16/Z); the National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ NHS Foundation Trust and Kings’ College London; and the NIHR Biomedical Research Centre at University College London Hospitals NHS Foundation Trust and University College London. The views expressed are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health. PJS was funded by the European Union’s Horizon 2020 research and innovation programme under grant agreement No. 666992. MP was supported by EPSRC grant EP/N018702/1 and by UKRI Future Leaders Fellowship MR/T020296/1. CFW was funded by NIH grants P41EB015902 and R01MH074794. FBL was funded by the Deutsche Forschungsgemeinschaft (DFG LA 2804/12‐1). DK and JPH were funded by NIH grants R01NS074980 and R01MH116173. DB was supported by a grant from the U.S. Department of Defense, Program Project 308430 ‐ Uniformed Services University of the Health Sciences. JH was supported by Wellcome Trust Sir Henry Wellcome Fellowship 201374/Z/16/Z and UKRI Future Leaders Fellowship MR/T018119/1.
REFERENCES
- 1. Le Bihan D, Breton E, Lallemand D, Aubin MLL, Vignaud J, LavalJeantet M. Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging. Radiology. 1988;168:497‐505. [DOI] [PubMed] [Google Scholar]
- 2. Does MD. Inferring brain tissue composition and microstructure via MR relaxometry. NeuroImage. 2018;182:136‐148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Halle B. Molecular theory of field‐dependent proton spin‐lattice relaxation in tissue. Magn Reson Med. 2006;56:60‐72. [DOI] [PubMed] [Google Scholar]
- 4. English AE, Whittall KP, Joy ML, Henkelman RM. Quantitative Two?Dimensional time Correlation Relaxometry. Magn Reson Med. 1991;22:425‐434. [DOI] [PubMed] [Google Scholar]
- 5. Van Dusschoten D, Moonen CT, Adrie De Jager P, Van As H. Unraveling diffusion constants in biological tissue by combining Carr‐ Purcell‐Meiboom‐Gill imaging and pulsed field gradient NMR. Magn Reson Med. 1996;36:907‐913. [DOI] [PubMed] [Google Scholar]
- 6. Hürlimann MD, Venkataramanan L, Flaum C. The diffusion‐spin relaxation time distribution function as an experimental probe to characterize fluid mixtures in porous media. J Chem Phys. 2002;117:10223‐10232. [Google Scholar]
- 7. Callaghan PT, Godefroy S, Ryland BN. Diffusion‐relaxation correlation in simple pore structures. J Magn Reson. 2003;162:320‐327. [DOI] [PubMed] [Google Scholar]
- 8. Venkataramanan L, Song YQ, Hürlimann MD. Solving Fredholm integrals of the first kind with tensor product structure in 2 and 2.5 dimensions. IEEE Trans Signal Process. 2002;50:1017‐1026. [Google Scholar]
- 9. Song YQ, Venkataramanan L, Hürlimann MD, Flaum M, Frulla P, Straley C. ‐T2 correlation spectra obtained using a fast two‐dimensional Laplace inversion. J Magn Reson. 2002;154:261‐268. [DOI] [PubMed] [Google Scholar]
- 10. Hürlimann MD, Venkataramanan L. Quantitative measurement of two‐dimensional distribution functions of diffusion and relaxation in grossly inhomogeneous fields. J Magn Reson. 2002;157:31‐42. [DOI] [PubMed] [Google Scholar]
- 11. Hürlimann MD, Venkataramanan L, Flaum C. The diffusion‐spin relaxation time distribution function as an experimental probe to characterize fluid mixtures in porous media. J Chem Phys. 2002;117:10223‐10232. [Google Scholar]
- 12. Hürlimann MD, Flaum M, Venkataramanan L, Flaum C, Freedman R, Hirasaki GJ. Diffusion‐relaxation distribution functions of sedimentary rocks in different saturation states. Magn Reson Imaging. 2003;21:305‐310. [DOI] [PubMed] [Google Scholar]
- 13. Hürlimann MD, Burcaw L, Song YQ. Quantitative characterization of food products by two‐dimensional D‐T 2 and T1–T2 distribution functions in a static gradient. J Colloid Inter Sci. 2006;297:303‐311. [DOI] [PubMed] [Google Scholar]
- 14. Song YQ. A 2D NMR method to characterize granular structure of dairy products. Progress Nuclear Magn Reson Spectroscopy. 2009;55:324‐334. [Google Scholar]
- 15. van Duynhoven J, Voda A, Witek M, Van As H. Time‐domain NMR applied to food products. Annual Reports on NMR Spectroscopy., Vol. 69Academic Press Inc.; 2010:145‐197. [Google Scholar]
- 16. Galvosas P, Callaghan PT. Multi‐dimensional inverse Laplace spectroscopy in the NMR of porous media. Comptes Rendus Physique. 2010;11:172‐180. [Google Scholar]
- 17. Song YQ. Magnetic resonance of porous media (MRPM): A perspective. J Magn Reson. 2013;229:12‐24. [DOI] [PubMed] [Google Scholar]
- 18. Peled S, Cory DG, Raymond SA, Kirschner DA, Jolesz FA. Water diffusion, T2, and compartmentation in frog sciatic nerve. Magn Reson Med. 1999;42:911‐918. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Does MD, Gore JC. Compartmental study of diffusion and relaxation measured in vivo in normal and ischemic rat brain and trigeminal nerve. Magn Reson Med. 2000;43:837‐844. [DOI] [PubMed] [Google Scholar]
- 20. Silva MD, Helmer KG, Han SS, et al. Deconvolution of compartmental water diffusion coefficients in yeast‐cell suspensions using combined T1 and diffusion measurements. J Magn Reson. 2002;156:52‐63. [DOI] [PubMed] [Google Scholar]
- 21. Fan RH, Does MD. Compartmental relaxation and diffusion tensor imaging measurements in vivo in ‐carrageenan‐induced edema in rat skeletal muscle. NMR Biomed. 2008;21:566‐573. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Mulkern RV, Zengingonul HP, Robertson RL, et al. Multi‐component apparent diffusion coefficients in human brain: Relationship to spin‐lattice relaxation. Magn Reson Med. 2000;44:292‐300. [DOI] [PubMed] [Google Scholar]
- 23. Lancaster JL, Andrews T, Hardies LJ, Dodd S, Fox PT. Three‐pool model of white matter. J Magn Reson Imaging. 2003;17:1‐10. [DOI] [PubMed] [Google Scholar]
- 24. Raj A, Pandya S, Shen X, LoCastro E, Nguyen TD, Gauthier SA. Multi‐compartment relaxometry using a spatially constrained multi‐Gaussian model. PLoS ONE. 2014;9:e98391 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Sati P, van Gelderen P, Silva AC, et al. Micro‐compartment specific relaxation in the brain. NeuroImage. 2013;77:268‐278. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Veraart J, Novikov DS, Fieremans E. TE dependent Diffusion Imaging (TEdDI) distinguishes between compartmental T2 relaxation times. Neuroimage. 2018;182:360‐369. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Silvera S, Oppenheim C, Touzé E, et al. Spontaneous intracerebral hematoma on diffusion‐weighted images: Influence of ‐shine‐through and ‐blackout effects. Am J Neuroradiol. 2005;26:236‐241. [PMC free article] [PubMed] [Google Scholar]
- 28. Duncan JS, Bartlett P, Barker GJ. Technique for measuring hippocampal relaxation time. Am J Neuroradiol. 1996;17:1805‐1810. [PMC free article] [PubMed] [Google Scholar]
- 29. Gong T, Tong Q, He H, Sun Y, Zhong J, Zhang H. MTE‐NODDI: Multi‐TE NODDI for disentangling non‐‐weighted signal fractions from compartment‐specific relaxation times. NeuroImage. 2020;217:116906 [DOI] [PubMed] [Google Scholar]
- 30. Bydder GM, Young IR. MR imaging: clinical use of the inversion recovery sequence. J Comput Assisted Tomography. 1985;9:659‐675. [PubMed] [Google Scholar]
- 31. De Santis S, Assaf Y, Jeurissen B, Jones DK, Roebroeck A. relaxometry of crossing fibres in the human brain. NeuroImage. 2016;141:133‐142. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Does MD, Beaulieu C, Allen PS, Snyder RE. Multi‐component relaxation and magnetisation transfer in peripheral nerve. Magn Reson Imaging. 1998;16:1033‐1041. [DOI] [PubMed] [Google Scholar]
- 33. Seland JG, Bruvold M, Brurok H, Jynge P, Krane J. Analyzing equilibrium water exchange between myocardial tissue compartments using dynamical two‐dimensional correlation experiments combined with manganese‐enhanced relaxography. Magn Reson Med. 2007;58:674‐686. [DOI] [PubMed] [Google Scholar]
- 34. Frydman L, Scherf T, Lupulescu A. The acquisition of multidimensional NMR spectra within a single scan. Proc Nat Acad Sci USA. 2002;99:15858‐15862. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Ahola S, Zhivonitko VV, Mankinen O, et al. Ultrafast multidimensional Laplace NMR for a rapid and sensitive chemical analysis. Nat Commun. 2015;6:1‐7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Hutter J, Tournier JD, Hajnal J. Chapter 9 pulse sequences combining advanced diffusion encoding and image read‐out. in “Advanced Diffusion Encoding Methods in MRI”. The Royal Society of Chemistry; 2020:123‐153. [Google Scholar]
- 37. Chavhan GB, Babyn PS, Thomas B, Shroff MM, Mark Haacke E. Principles, techniques, and applications of ‐based MR imaging and its special applications. Radiographics. 2009;29:1433‐1449. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Ordidge RJ, Gibbs P, Chapman B, Stehling MK, Mansfield P. High‐speed multislice mapping using inversion‐recovery echo‐planar imaging. Magn Reson Med. 1990;16:238‐245. [DOI] [PubMed] [Google Scholar]
- 39. Wu H, Tian Q, Poetter C, et al. Whole brain inversion recovery diffusion weighted imaging using slice‐shuffled acquisition. In Proceedings: ISMRM 25th Annual Meeting & Exhibition, Honolulu Hawai'i, USA, 2017. Abstract #0387.
- 40. Leppert IR, Andrews DA, Campbell JSW, et al. Efficient whole‐brain tract‐specific t1 mapping at 3t with slice‐shuffled inversion‐recovery diffusion‐weighted imaging. Magn Reson Med. 2021;86:738‐753. [DOI] [PubMed] [Google Scholar]
- 41. Reymbaut A, Critchley J, Durighel G, et al. Toward nonparametric diffusion‐ characterization of crossing fibers in the human brain. Magn Reson Med. 2021;85:2815‐2827. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Hutter J, Christiaens DJ, Schneider T, et al. Slice‐level diffusion encoding for motion and distortion correction. Med Image Anal. 2018;48:214‐229. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Hutter J, Slator PJ, Christiaens D, et al. Integrated and efficient diffusion‐relaxometry using ZEBRA. Sci Rep. 2018;8:15138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Pizzolato M, Palombo M, BonetCarne E, et al. Acquiring and predicting multidimensional diffusion (MUDI) data: an open challenge, in “CDMRI’20”. Springer Science and Business Media Deutschland GmbH; 2020:195‐208. [Google Scholar]
- 45. Lundell H, Alexander DC, Dyrby TB. High angular resolution diffusion imaging with stimulated echoes: Compensation and correction in experiment design and analysis. NMR Biomed. 2014;27:918‐925. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Fritz FJ, Poser BA, Roebroeck A. MESMERISED: super‐accelerating T1 relaxometry and diffusion MRI with STEAM at 7 T for quantitative multi‐contrast and diffusion imaging. Neuroimage. 2021;239:118285. 10.1016/j.neuroimage.2021.118285. [DOI] [PubMed] [Google Scholar]
- 47. Wang F, Dong Z, Reese TG, et al. Echo planar time‐resolved imaging (EPTI). Magn Reson Med. 2019;81:3599‐3615. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48. Fair MJ, Liao C, Kim D, Varadarajan D, Haldar JP, Setsompop K. Diffusion‐PEPTIDE: rapid distortion‐free diffusion‐relaxometry imaging. ISMRM & SMRT Virtual Conference & Exhibition. 2020. Abstract #0953. [Google Scholar]
- 49. Jones DK, Alexander DC, Bowtell R, et al. Microstructural imaging of the human brain with a “super‐scanner”: 10 key advantages of ultra‐strong gradients for diffusion MRI. NeuroImage. 2018;182:8‐38. [DOI] [PubMed] [Google Scholar]
- 50. Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature. 2013;495:187‐92. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Yu AC, Badve C, Ponsky LE, et al. Development of a combined MR fingerprinting and diffusion examination for prostate cancer. Radiology. 2017;283:729‐738. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52. Pirkl CM, Gómez PA, Lipp I, et al. Deep learning‐based parameter mapping for joint relaxation and diffusion tensor MR fingerprinting. Proceedings of the Third Conference on Medical Imaging with Deep Learning, PMLR, 2020;121:638‐654. http://proceedings.mlr.press/v121/pirk20a.html. Accessed August 8, 2021. [Google Scholar]
- 53. Christodoulou AG, Shaw JL, Nguyen C, et al. Magnetic resonance multitasking for motion‐resolved quantitative cardiovascular imaging. Nat Biomed Eng. 2018;2:215‐226. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Ma S, Nguyen CT, Han F, et al. Three‐dimensional simultaneous brain , , and adc mapping with mr multitasking. Magn Reson Med. 2020;84:72‐88. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Kim D, Wisnowski JL, Nguyen CT, Haldar JP. Multidimensional correlation spectroscopic imaging of exponential decays: from theoretical principles to in vivo human applications. NMR Biomed. 2020;1‐19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56. Kim D, Wisnowski JL, Haldar JP. Improved efficiency for microstructure imaging using high‐dimensional MR correlation spectroscopic imaging. 2017 51st Asilomar Conference on Signals, Systems, and Computers.; 2017:1264‐1268. [Google Scholar]
- 57. Lampinen B, Szczepankiewicz F, Mårtensson J, et al. Towards unconstrained compartment modeling in white matter using diffusion‐relaxation MRI with tensor‐valued diffusion encoding. Magn Reson Med. 2020;84:1605‐1623. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. Benjamini D, Basser PJ. Use of marginal distributions constrained optimization (MADCO) for accelerated 2D MRI relaxometry and diffusometry. J Magn Reson. 2016;271:40‐45. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. Benjamini D, Basser PJ. Towards clinically feasible relaxation‐diffusion correlation MRI using MADCO. Microporous Mesoporous Mater. 2018;269:93‐96. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60. Benjamini D, Basser PJ. Magnetic resonance microdynamic imaging reveals distinct tissue microenvironments. NeuroImage. 2017;163:183‐196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61. Shrot Y, Frydman L. Compressed sensing and the reconstruction of ultrafast 2D NMR data: Principles and biomolecular applications. J Magn Reson. 2011;209:352‐358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. Bai R, Cloninger A, Czaja W, Basser PJ. Efficient 2D MRI relaxometry using compressed sensing. J Magn Reson. 2015;255:88‐99. [DOI] [PubMed] [Google Scholar]
- 63. Song Y, Xiao L. Optimization of multidimensional MR data acquisition for relaxation and diffusion. NMR Biomed. 2020;33. 10.1002/nbm.4238. [DOI] [PubMed] [Google Scholar]
- 64. Grussu F, Blumberg S, Battiston M, et al. “Select and retrieve via direct upsampling” network (SARDU‐Net): a data‐driven, model‐free, deep learning approach for quantitative MRI protocol design. bioRxiv. 2020;2020.05.26.116491. [Google Scholar]
- 65. Tanner JE. Self diffusion of water in frog muscle. Biophys J. 1979;28:107‐116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66. Fujita N, Harada K, Sakurai K, Akai Y, Kozuka T. Separation of diffusion and slow flow effects by use of flow rephasing and dephasing. Magn Reson Med. 1992;24:109‐122. [DOI] [PubMed] [Google Scholar]
- 67. Laun FB, Wetscherek A. Chapter 5 separating flow from diffusion using velocity‐compensated diffusion encoding. in “Advanced Diffusion Encoding Methods in MRI”. The Royal Society of Chemistry; 2020:123‐153. [Google Scholar]
- 68. Mori S, Van Zijl PC. Diffusion weighting by the trace of the diffusion tensor within a single scan. Magn Reson Med. 1995;33:41‐52. [DOI] [PubMed] [Google Scholar]
- 69. Callaghan PT, Furó I. Diffusion‐diffusion correlation and exchange as a signature for local order and dynamics. J Chem Phys. 2004;120:4032‐4038. [DOI] [PubMed] [Google Scholar]
- 70. Westin CF, Knutsson H, Pasternak O, et al. Q‐space trajectory imaging for multidimensional diffusion MRI of the human brain. NeuroImage. 2016;135:345‐362. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71. Szczepankiewicz F, Westin CF, Nilsson M. Gradient waveform design for tensor‐valued encoding in diffusion MRI. J Neurosci Methods. 2021;348:109007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72. Nilsson M, Lasič S, Drobnjak I, Topgaard D, Westin CF. Resolution limit of cylinder diameter estimation by diffusion MRI: the impact of gradient waveform and orientation dispersion. NMR Biomed. 2017;30:e3711. 10.1002/nbm.3711. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Stepinik J. Analysis of nmr self‐diffusion measurements by a density matrix calculation. Physica B+C. 1981;104:350‐364. [Google Scholar]
- 74. Lemberskiy G, Fieremans E, Veraart J, Deng FM, Rosenkrantz AB, Novikov DS. Characterization of prostate microstructure using water diffusion and NMR relaxation. Front Phys. 2018;6. 10.3389/fphy.2018.00091. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75. Fieremans E, Lemberskiy G, Veraart J, Sigmund EE, Gyftopoulos S, Novikov DS. In vivo measurement of membrane permeability and myofiber size in human muscle using time‐dependent diffusion tensor imaging and the random permeable barrier model. NMR Biomed. 2017;30:e3612. 10.1002/nbm.3612. [DOI] [PubMed] [Google Scholar]
- 76. Martin J, Endt S, Wetscherek A, et al. Twice‐refocused stimulated echo diffusion imaging: measuring diffusion time dependence at constant weighting. Magn Reson Med. 2020;83:1741‐1749. [DOI] [PubMed] [Google Scholar]
- 77. Reynaud O. Time‐dependent diffusion MRI in cancer: tissue modeling and applications. Front Phys. 2017;5. 10.3389/fphy.2017.00058. [DOI] [Google Scholar]
- 78. Zheng G, Price WS. Suppression of background gradients in ( gradient‐based) NMR diffusion experiments. Conc Magn Reson Part A. 2007;30A:261‐277. [Google Scholar]
- 79. Song YQ, Ryu S, Sen PN. Determining multiple length scales in rocks. Nature. 2000;406:178‐181. [DOI] [PubMed] [Google Scholar]
- 80. Álvarez GA, Shemesh N, Frydman L. Internal gradient distributions: a susceptibility‐derived tensor delivering morphologies by magnetic resonance. Sci Rep. 2017;7:1‐10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81. Ladd ME, Bachert P, Meyerspeer M, et al. Pros and cons of ultra‐high‐field MRI/MRS for human application. Progress Nuclear Magn Reson Spectroscopy. 2018;109:1‐50. [DOI] [PubMed] [Google Scholar]
- 82. Wright PJ, Mougin OE, Totman JJ, et al. Water proton measurements in brain tissue at 7, 3, and 1.5T using IR‐EPI, IR‐TSE, and MPRAGE: results and optimization. Magn Reson Mater Phys Biol Med. 2008;21:121‐130. [DOI] [PubMed] [Google Scholar]
- 83. Qin W, Yu CS, Zhang F, et al. Effects of echo time on diffusion quantification of brain white matter at 1.5T and 3.0T. Magn Reson Med. 2009;61:755‐760. [DOI] [PubMed] [Google Scholar]
- 84. Endt S, Martin J, Ning L, et al. An observational study on the diffusion time and field strength dependency of the diffusion coefficient in human white matter. In: Proceedings ISMRM 27th Annual Meeting & Exhibition, Montréal, QC, Canada. 2019. Abstract #3538.
- 85. Martin J. On the influence of field strength and diffusion time dependence in magnetic resonance microscopic diffusion imaging., (PhD thesis); 2019. [Google Scholar]
- 86. Szczepankiewicz F, Sjölund J, Ståhlberg F, Lätt J, Nilsson M. Tensor‐valued diffusion encoding for diffusional variance decomposition (DIVIDE): technical feasibility in clinical MRI systems. PLoS ONE. 2019;14:e0214238. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87. Rydhög AS, van Osch MJ, Lindgren E, et al. Intravoxel incoherent motion (IVIM) imaging at different magnetic field strengths: What is feasible? Magn Reson Imaging. 2014;32:1247‐1258. [DOI] [PubMed] [Google Scholar]
- 88. Lévy S, Rapacchi S, Massire A, et al. Intravoxel incoherent motion at 7 tesla to quantify human spinal cord perfusion: limitations and promises. Magn Reson Med. 2020;84:1198‐1217. [DOI] [PubMed] [Google Scholar]
- 89. Riexinger AJ, Martin J, Rauh S, et al. On the field strength dependence of bi‐ and triexponential intravoxel incoherent motion (IVIM) parameters in the liver. J Magn Reson Imaging. 2019;50:1883‐1892. [DOI] [PubMed] [Google Scholar]
- 90. Lasič S, Nilsson M, Lätt J, Ståhlberg F, Topgaard D. Apparent exchange rate mapping with diffusion MRI. Magn Reson Med. 2011;66:356‐365. [DOI] [PubMed] [Google Scholar]
- 91. Nilsson M, Lätt J, Van Westen D, et al. Noninvasive mapping of water diffusional exchange in the human brain using filter‐exchange imaging. Magn Reson Med. 2013;69:1572‐1580. [DOI] [PubMed] [Google Scholar]
- 92. Washburn KE, Callaghan PT. Tracking pore to pore exchange using relaxation exchange spectroscopy. Phys Rev Lett. 2006;97:25‐28. [DOI] [PubMed] [Google Scholar]
- 93. Chandrasekera TC, Mitchell J, Fordham EJ, Gladden LF, Johns ML. Rapid encoding of with spectral resolution in n‐dimensional relaxation correlations. J Magn Reson. 2008;194:156‐161. [DOI] [PubMed] [Google Scholar]
- 94. Cai TX, Benjamini D, Komlosh ME, Basser PJ, Williamson NH. Rapid detection of the presence of diffusion exchange. J Magn Reson. 2018;297:17‐22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95. Williamson NH, Ravin R, Cai TX, et al. Real‐time measurement of diffusion exchange rate in biological tissue. J Magn Reson. 2020;317:106782. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96. Larkman DJ, Hajnal JV, Herlihy AH, Coutts GA, Young IR, Ehnholm G. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J Magn Reson Imaging. 2001;13:313‐317. [DOI] [PubMed] [Google Scholar]
- 97. Barth M, Breuer F, Koopmans PJ, Norris DG, Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn Reson Med. 2016;75:63‐81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98. Cloninger A, Czaja W, Bar R, Basser PJ. Solving 2D fredholm integral from incomplete measurements using compressive sensing. SIAM J Imaging Sci. 2014;7:1775‐1798. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99. Celik H, Bouhrara M, Reiter DA, Fishbein KW, Spencer RG. Stabilization of the inverse Laplace transform of multiexponential decay through introduction of a second dimension. J Magn Reson. 2013;236:134‐139. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100. Kim D, Doyle EK, Wisnowski JL, Kim JH, Haldar JP. Diffusion‐relaxation correlation spectroscopic imaging: a multidimensional approach for probing microstructure. Magn Reson Med. 2017;78:2236‐2249. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101. Lee DD, Seung HS. Learning the parts of objects by non‐negative matrix factorization. Nature. 1999;401:788‐791. [DOI] [PubMed] [Google Scholar]
- 102. Donoho D, Stodden V. When does non‐negative matrix factorization give a correct decomposition into parts? In Proceedings of the 16th International Conference on Neural Information Processing Systems (NIPS'03). 2003, pp. 1141‐1148.
- 103. Choi S, Cichocki A, Park HM, Lee SY. Blind source separation and independent component analysis: a review. Neural Inform Process Lett Rev. 2005;6:1‐57. [Google Scholar]
- 104. Kolda TG, Bader BW. Tensor decompositions and applications. SIAM Rev. 2009;51:455‐500. [Google Scholar]
- 105. Tønning E, Polders D, Callaghan PT, Engelsen SB. A novel improved method for analysis of 2D diffusion‐relaxation data‐2D PARAFAC‐Laplace decomposition. J Magn Reson. 2007;188:10‐23. [DOI] [PubMed] [Google Scholar]
- 106. Kim D, Kim JH, Haldar JP. Automatic tissue decomposition using nonnegative matrix factorization for noisy MR magnitude images. In: Proceedings ISMRM 23rd Annual Meeting & Exhibition, Toronto, Ontario, Canada. 2015. p. 3701.
- 107. MolinaRomero M, Gómez PA, Sperl JI, et al. A diffusion model‐free framework with echo time dependence for free‐water elimination and brain tissue microstructure characterization. Magn Reson Med. 2018;80:2155‐2172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108. Provencher SW. A constrained regularization method for inverting data represented by linear algebraic or integral equations. Comput Phys Commun. 1982;27:213‐227. [Google Scholar]
- 109. Whittall KP, MacKay AL. Quantitative interpretation of NMR relaxation data. J Magn Reson. 1989;84:134‐152. [Google Scholar]
- 110. Liang ZP, Lauterbur PC. Principles of Magnetic Resonance Imaging: A Signal Processing Perspective. SPIE Optical Engineering Press; 2000. [Google Scholar]
- 111. Bertero M, Mol CD, Pike ER. Linear inverse problems with discrete data. I. General formulation and singular system analysis. Inverse Prob. 1985;1:301‐330. [Google Scholar]
- 112. Lawson CL, Hanson RJ. Solving least squares problems. Soc Indus Appl Math. 1995. [Google Scholar]
- 113. Slawski M, Hein M. Non‐negative least squares for high‐dimensional linear models: Consistency and sparse recovery without regularization. Elect J Statist. 2013;7:3004‐3056. [Google Scholar]
- 114. Slator PJ, Hutter J, Palombo M, et al. Combined diffusion‐relaxometry MRI to identify dysfunction in the human placenta. Magn Reson Med. 2019;82:95‐106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115. Kumar D, Nguyen TD, Gauthier SA, Raj A. Bayesian algorithm using spatial priors for multiexponential relaxometry from multiecho spin echo MRI. Magn Reson Med. 2012;68:1536‐1543. [DOI] [PubMed] [Google Scholar]
- 116. Lin Y, Haldar JP, Li Q, Conti PS, Leahy RM. Sparsity constrained mixture modeling for the estimation of kinetic parameters in dynamic PET. IEEE Trans Med Imaging. 2014;33:173‐185. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 117. Slator PJ, Hutter J, Marinescu RV, et al. InSpect: INtegrated SPECTral component estimation and mapping for multi‐contrast microstructural MRI. IPMI 2019. Cham: Springer; 2019:755‐766. [Google Scholar]
- 118. Slator PJ, Hutter J, Marinescu RV, et al. Data‐driven multi‐contrast spectral microstructure imaging with inspect. In: Martel AL, Abolmaesumi P, Stoyanov D, et al. (eds) Medical Image Computing and Computer Assisted Intervention – MICCAI 2020. 2020;12266:375‐385. [Google Scholar]
- 119. Golub GH, Heath M, Wahba G. Generalized Cross‐Validation as a Method for Choosing a Good Ridge Parameter. Technometrics. 1979;21:215. [Google Scholar]
- 120. Miller K. Least Squares Methods for Ill‐Posed Problems with a Prescribed Bound. SIAM J Math Anal. 1970;1:52‐74. [Google Scholar]
- 121. Fessler JA, Rogers WL. Spatial resolution properties of penalized‐likelihood image reconstruction: space‐invariant tomographs. IEEE Trans Image Process. 1996;5:1346‐1358. [DOI] [PubMed] [Google Scholar]
- 122. Ahn S, Leahy RM. Analysis of resolution and noise properties of nonquadratically regularized image reconstruction methods for pet. IEEE Trans Med Imaging. 2008;27:413‐424. [DOI] [PubMed] [Google Scholar]
- 123. Mitchell J, Chandrasekera TC, Gladden LF, Mitchell J, Chandrasekera TC, Gladden LF. Numerical estimation of relaxation and diffusion distributions in two dimensions. Prog Nucl Magn Reson Spectrosc. 2012;62:34‐50. [DOI] [PubMed] [Google Scholar]
- 124. Mackay A, Whittall K, Adler J, Li D, Paty D, Graeb D. In vivo visualization of myelin water in brain by magnetic resonance. Magn Reson Med. 1994;31:673‐677. [DOI] [PubMed] [Google Scholar]
- 125. De Santis S, Barazany D, Jones DK, Assaf Y. Resolving relaxometry and diffusion properties within the same voxel in the presence of crossing fibres by combining inversion recovery and diffusion‐weighted acquisitions. Magn Reson Med. 2016;75:372‐380. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 126. Istratov AA, Vyvenko OF. Exponential analysis in physical phenomena. Rev Sci Instrum. 1999;70:1233‐1257. [Google Scholar]
- 127. Basser P, Mattiello J, Lebihan D. Estimation of the effective self‐diffusion tensor from the NMR spin echo. J Magn Reson Ser B. 1994;103:247‐254. [DOI] [PubMed] [Google Scholar]
- 128. Jensen JH, Helpern JA, Ramani A, Lu H, Kaczynski K. Diffusional kurtosis imaging: the quantification of non‐gaussian water diffusion by means of magnetic resonance imaging. Magn Reson Med. 2005;53:1432‐40. [DOI] [PubMed] [Google Scholar]
- 129. Özarslan E, Koay CG, Shepherd TM, et al. Mean apparent propagator (MAP) MRI: a novel diffusion imaging method for mapping tissue microstructure. NeuroImage. 2013;78:16‐32. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130. Weigel M. Extended phase graphs: dephasing, RF pulses, and echoes–pure and simple. J Magn Reson Imaging. 2015;41:266‐295. [DOI] [PubMed] [Google Scholar]
- 131. Kiselev VG. In: Jones PhDDK, ed. The cumulant expansion: an overarching mathematical framework for understanding diffusion NMR. in “Diffusion MRI”, vol. 1. Oxford University Press; 2010:152‐168. [Google Scholar]
- 132. Ning L, Gagoski B, Szczepankiewicz F, Westin CF, Rathi Y. Joint RElaxation‐diffusion imaging moments to probe neurite microstructure. IEEE Trans Med Imaging. 2020;39:668‐677. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 133. Kiselev VG, Il’yasov KA. Is the “biexponential diffusion” biexponential? Magn Reson Med. 2007;57:464‐469. [DOI] [PubMed] [Google Scholar]
- 134. Novikov DS, Kiselev VG. Effective medium theory of a diffusion‐weighted signal. NMR Biomed. 2010;23:682‐697. [DOI] [PubMed] [Google Scholar]
- 135. Novikov DS, Fieremans E, Jespersen SN, Kiselev VG. Quantifying brain microstructure with diffusion MRI: theory and parameter estimation. NMR Biomed. 2019;32:e3998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 136. Fick RH, Wassermann D, Deriche R. The Dmipy Toolbox: Diffusion MRI Multi‐Compartment Modeling and Microstructure Recovery Made Easy. Front Neuroinformat. 2019;13:1‐26. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 137. Panagiotaki E, Schneider T, Siow B, Hall MG, Lythgoe MF, Alexander DC. Compartment models of the diffusion MR signal in brain white matter: a taxonomy and comparison. NeuroImage. 2012;59:2241‐2254. [DOI] [PubMed] [Google Scholar]
- 138. Jelescu IO, Budde MD. Design and validation of diffusion MRI models of white matter. Front Phys. 2017;5:61. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 139. Tariq M, Schneider T, Alexander DC, Gandini Wheeler‐Kingshott CA, Zhang H. Bingham‐NODDI: mapping anisotropic orientation dispersion of neurites using diffusion MRI. NeuroImage. 2016;133:207‐223. [DOI] [PubMed] [Google Scholar]
- 140. Zhang H, Schneider T, WheelerKingshott CA, Alexander DC. NODDI: practical in vivo neurite orientation dispersion and density imaging of the human brain. NeuroImage. 2012;61:1000‐1016. [DOI] [PubMed] [Google Scholar]
- 141. Jelescu IO, Veraart J, Fieremans E, Novikov DS. Degeneracy in model parameter estimation for multi‐compartmental diffusion in neuronal tissue. NMR Biomed. 2016;29:33‐47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 142. Lampinen B, Szczepankiewicz F, van Westen D, et al. Optimal experimental design for filter exchange imaging: apparent exchange rate measurements in the healthy brain and in intracranial tumors. Magn Reson Med. 2017;77:C1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 143. Palombo M, Singh S, Whitaker H, Alexander DC. Relaxed‐VERDICT: decoupling relaxation and diffusion for comprehensive microstructure characterization of prostate cancer. In Proceedings: ISMRM & SMRT Virtual Conference & Exhibition. 2020. Abstract #0710.
- 144. Aboitiz F, Scheibel AB, Fisher RS, Zaidel E. Fiber composition of the human corpus callosum. Brain Res. 1992;598:143‐153. [DOI] [PubMed] [Google Scholar]
- 145. Sabouri S, Fazli L, Chang SD, et al. MR measurement of luminal water in prostate gland: Quantitative correlation between MRI and histology. J Magn Reson Imaging. 2017;46:861‐869. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 146. Kim D, Haldar JP. Faster diffusion‐relaxation correlation spectroscopic imaging (DR‐CSI) using optimized experiment design. In Proceedings: ISMRM 25th Annual Meeting & Exhibition, Honolulu, Hawai'i, USA. 2017. p. 0176.
- 147. Cavassila S, Deval S, Huegen C, Van Ormondt D, GraveronDemilly D. Cramér‐Rao bounds: an evaluation tool for quantitation. NMR Biomed. 2001;14:278‐283. [DOI] [PubMed] [Google Scholar]
- 148. Alexander DC. A general framework for experiment design in diffusion MRI and its application in measuring direct tissue‐microstructure features. Magn Reson Med. 2008;60:439‐448. [DOI] [PubMed] [Google Scholar]
- 149. Zhao B, Haldar JP, Liao C, et al. Optimal experiment design for magnetic resonance fingerprinting: cramér‐rao bound meets spin dynamics. IEEE Trans Med Imaging. 2019;38:844‐861. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 150. De Almeida Martins JP, Topgaard D. Multidimensional correlation of nuclear relaxation rates and diffusion tensors for model‐free investigations of heterogeneous anisotropic porous materials. Sci Rep. 2018;8:2488. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151. Henriques RN, Jespersen SN, Shemesh N. Correlation tensor magnetic resonance imaging. Neuroimage. 2020;211:116605. 10.1016/j.neuroimage.2020.116605. [DOI] [PubMed] [Google Scholar]
- 152. NedjatiGilani GL, Schneider T, Hall MG, et al. Machine learning based compartment models with permeability for white matter microstructure imaging. NeuroImage. 2017;150:119‐135. [DOI] [PubMed] [Google Scholar]
- 153. Rensonnet G, Scherrer B, Girard G, et al. Towards microstructure fingerprinting: estimation of tissue properties from a dictionary of Monte Carlo diffusion MRI simulations. NeuroImage. 2019;184:964‐980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 154. Alexander DC, Dyrby TB, Nilsson M, Zhang H. Imaging brain microstructure with diffusion MRI: practicality and applications. NMR Biomed. 2019;32:1‐26. [DOI] [PubMed] [Google Scholar]
- 155. Van Everdingen KJ, Van Der Grond J, Kappelle LJ, Ramos LM, Mali WP. Diffusion‐weighted magnetic resonance imaging in acute stroke. Stroke. 1998;29:1783‐1790. [DOI] [PubMed] [Google Scholar]
- 156. Knight MJ, Damion RA, McGarry BL, et al. Determining relaxation time and stroke onset relationship in ischaemic stroke within apparent diffusion coefficient‐defined lesions. A user‐independent method for quantifying the impact of stroke in the human brain. Biomed Spectroscopy Imaging. 2019;8:11‐28. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 157. Novikov DS, Veraart J, Jelescu IO, Fieremans E. Rotationally‐invariant mapping of scalar and orientational metrics of neuronal microstructure with diffusion MRI. NeuroImage. 2018;174:518‐538. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 158. Khan AR, Chuhutin A, Wiborg O, et al. Biophysical modeling of high field diffusion MRI demonstrates micro‐structural aberration in chronic mild stress rat brain. NeuroImage. 2016;142:421‐430. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 159. Kleban E, Tax CM, Rudrapatna US, Jones DK, Bowtell R. Strong diffusion gradients allow the separation of intra‐ and extra‐axonal gradient‐echo signals in the human brain. NeuroImage. 2020;217:116793. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 160. Benjamini D, Basser PJ. Multidimensional correlation MRI. NMR Biomed. 2020;33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 161. de Almeida Martins JP, Tax CMW, Szczepankiewicz F, Jones DK, Westin CF, Topgaard D. Transferring principles of solid‐state and Laplace NMR to the field of in vivo brain MRI. Magn Reson. 2020;1:27‐43. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 162. de Almeida Martins JP, Tax CM, Reymbaut A, et al. Computing and visualising intra‐voxel orientation‐specific relaxation‐diffusion features in the human brain. Human Brain Mapping. 2021;42:310‐328. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 163. Benjamini D, Basser PJ. Water mobility spectral imaging of the spinal cord: parametrization of model‐free Laplace MRI. Magn Reson Imaging. 2019;56:187‐193. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 164. Benjamini D, Iacono D, Komlosh ME, Perl DP, Brody DL, Basser PJ. Diffuse axonal injury has a characteristic multidimensional MRI signature in the human brain. Brain. 2021;144:800‐816. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 165. Reymbaut A, Mezzani P, Almeida Martins JP, Topgaard D. Accuracy and precision of statistical descriptors obtained from multidimensional diffusion signal inversion algorithms. NMR Biomed. 2020;33. 10.1002/nbm.4267. [DOI] [PubMed] [Google Scholar]
- 166. Pas K, Komlosh ME, Perl DP, Basser PJ, Benjamini D. Retaining information from multidimensional correlation MRI using a spectral regions of interest generator. Sci Rep. 2020;10. 10.1038/s41598-020-60092-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 167. Grussu F, Battiston M, Veraart J, et al. Multi‐parametric quantitative in vivo spinal cord MRI with unified signal readout and image denoising. NeuroImage. 2020;217:116884. 10.1016/j.neuroimage.2020.116884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 168. Williamson NH, Ravin R, Benjamini D. Magnetic resonance measurements of cellular and sub‐cellular membrane structures in live and fixed neural tissue. eLife. 2019;8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 169. Palombo M, Shemesh N, Ronen I, Valette J. Insights into brain microstructure from in vivo DW‐MRS. 2018. [DOI] [PubMed]
- 170. Ligneul C, Palombo M, Valette J. Metabolite diffusion up to very high b in the mouse brain in vivo: Revisiting the potential correlation between relaxation and diffusion properties. Magn Reson Med. 2017;77:1390‐1398. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 171. Stabile A, Giganti F, Rosenkrantz AB, et al. Multiparametric MRI for prostate cancer diagnosis: current status and future directions. Nature Reviews Urology. 2020;17:41‐61. [DOI] [PubMed] [Google Scholar]
- 172. Chatterjee A, Watson G, Myint E, Sved P, McEntee M, Bourne R. Changes in epithelium, stroma, and lumen space correlate more strongly with gleason pattern and are stronger predictors of prostate ADC changes than cellularity metrics1. Radiology. 2015;277:751‐762. [DOI] [PubMed] [Google Scholar]
- 173. Lemberskiy G, Rosenkrantz AB, Veraart J, Taneja SS, Novikov DS, Fieremans E. Time‐dependent diffusion in prostate cancer. Investigat Radiol. 2017;52:405‐411. [DOI] [PubMed] [Google Scholar]
- 174. Bourne R, Liang S, Panagiotaki E, Bongers A, Sved P, Watson G. Measurement and modeling of diffusion time dependence of apparent diffusion coefficient and fractional anisotropy in prostate tissueex vivo. NMR Biomed. 2017;30:e3751. 10.1002/nbm.3751. [DOI] [PubMed] [Google Scholar]
- 175. Kjaer L, Thomsen C, Iversen P, Henriksen O. In vivo estimation of relaxation processes in benign hyperplasia and carcinoma of the prostate gland by magnetic resonance imaging. Magn Reson Imaging. 1987;5:23‐30. [DOI] [PubMed] [Google Scholar]
- 176. Gilani N, Rosenkrantz AB, Malcolm P, Johnson G. Minimization of errors in biexponential measurements of the prostate. J Magn Reson Imaging. 2015;42:1072‐1077. [DOI] [PubMed] [Google Scholar]
- 177. Storås TH, Gjesdal KI, Gadmar ØB, Geitung JT, Kløw NE. Prostate magnetic resonance imaging: multiexponential decay in prostate tissue. J Magn Reson Imaging. 2008;28:1166‐1172. [DOI] [PubMed] [Google Scholar]
- 178. Sabouri S, Chang SD, Savdie R, et al. Luminal water imaging: a new MR imaging mapping technique for prostate cancer diagnosis. Radiology. 2017;284:451‐459. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 179. Chatterjee A, Bourne RM, Wang S, et al. Diagnosis of prostate cancer with noninvasive estimation of prostate tissue composition by using hybrid multidimensional MR imaging: a feasibility study. Radiology. 2018;287:864‐873. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 180. Novikov DS, Kiselev VG. Surface‐to‐volume ratio with oscillating gradients. J Magn Reson. 2011;210:141‐145. [DOI] [PubMed] [Google Scholar]
- 181. Zhang Z, Wu HH, Priester A, et al. Prostate microstructure in prostate cancer using 3‐T MRI with diffusion‐relaxation correlation spectrum imaging: validation with whole‐mount digital histopathology. Radiology. 2020;296:348‐355. [DOI] [PubMed] [Google Scholar]
- 182. Sinding M, Peters DA, Frøkjaer JB, et al. Placental magnetic resonance imaging measurements in normal pregnancies and in those complicated by fetal growth restriction. Ultrasound Obstetri Gynecol. 2016;47:748‐754. [DOI] [PubMed] [Google Scholar]
- 183. Ingram E, Morris D, Naish J, Myers J, Johnstone E. MR imaging measurements of altered placental oxygenation in pregnancies complicated by fetal growth restriction. Radiology. 2017;285:953‐960. [DOI] [PubMed] [Google Scholar]
- 184. Hutter J, Slator PJ, Jackson L, et al. Multi‐modal functional MRI to explore placental function over gestation. Magn Reson Med. 2019;81:1191‐1204. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 185. Bonel HM, Stolz B, Diedrichsen L, et al. Diffusion‐weighted MR imaging of the placenta in fetuses with placental insufficiency. Radiology. 2010;257:810‐9. [DOI] [PubMed] [Google Scholar]
- 186. Moore RJ, Issa B, Tokarczuk P. In vivo intravoxel incoherent motion measurements in the human placenta using echo‐planar imaging at 0.5 T. Magn Reson Med. 2000;43:295‐302. [DOI] [PubMed] [Google Scholar]
- 187. Sohlberg S, MulicLutvica A, Olovsson M, et al. Magnetic resonance imaging‐estimated placental perfusion in fetal growth assessment. Ultrasound Obstetri Gynecol. 2015;46:700‐705. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 188. Melbourne A, Aughwane R, Sokolska M, et al. Separating fetal and maternal placenta circulations using multiparametric MRI. Magn Reson Med. 2019;81:350‐361. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 189. Slator PJ, Hutter J, Palombo M, et al. Combined diffusion‐relaxometry MRI to identify dysfunction in the human placenta. Magn Reson Med. 2019;82:95‐106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 190. Slator PJ, Hutter J, McCabe L, et al. Placenta microstructure and microcirculation imaging with diffusion MRI. Magn Reson Med. 2018;80:756‐766. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 191. Mifsud W, Sebire NJ. Placental pathology in early‐onset and late‐onset fetal growth restriction. Fetal Diagnosis Therapy. 2014;36:117‐28. [DOI] [PubMed] [Google Scholar]