Inter-rater reliability of a wright respirometer to measure vital capacity in neuromuscular disorders

Ryuji Yoshinaga 1,2,3,*, Haruki Futsuhara 1
PMCID: PMC8575478  PMID: 34776615

Abstract

[Purpose] We examined the inter-rater reliability of the Wright respirometer between therapists who assessed the vital capacity of patients with neuromuscular disorders. [Participants and Methods] We examined 18 patients with neuromuscular disorders. We performed a test-retest method after measuring their vital capacities. Two physical therapists experienced in the use of the Wright respirometer, specifically, therapist A with six years of clinical experience and therapist B with one, each took one measurement of the vital capacity of the same patients. The measurements between the therapists were taken at intervals of 3–7 days. We made a manual to standardize the measurements between the therapists. [Results] The vital capacities were 905 ± 490 mL for therapist A and 897 ± 483 mL for therapist B. The inter-rater reliability of ICC2,1 was 0.96 (95% CI: 0.89–0.98). Bland-Altman analyses revealed neither a fixed nor proportional bias. [Conclusion] These results demonstrated good to excellent inter-rater reliability of the Wright respirometer for patients with neuromuscular disorders using a manual and instructions.

Keywords: Wright respirometer, Reliability, Neuromuscular disorder

INTRODUCTION

Neuromuscular disorders (NMD) are characterized by the progressive weakening of skeletal, respiratory, and/or bulbar-innervated muscles1,2,3). Patients with NMDs have decrements in vital capacity (VC), lung and chest wall compliance, and coughing capacity4). The pulmonary rehabilitation of patients with NMDs is designed to maintain lung and thorax mobility and ameliorate the elasticity reduction that are associated with reduced VC5). Pulmonary rehabilitation for NMD patients often includes the introduction of maximum insufflation capacity (MIC) instead of deep inspiration to maintain the patients’ lung and thorax function from an early stage6). The MIC is attained when the patient takes a deep breath and holds it in; the air is then stacked using a bag-valve mask device7). Measurements of the VC and MIC in the pulmonary rehabilitation of patients with NMD are important to accurately grasp the progress of their disease.

NMD guidelines for respiratory management recommend measuring the MIC when the patient’s %VC is <50%8). For measurements of the MIC and VC in clinical settings, a Wright respirometer is frequently used9). The principle of the Wright respirometer is that the air flow hits the impeller and the volume is measured by the number of rotations10). The Wright respirometer can be used for NMD patients who cannot hold the mouthpiece of a spirometer in their mouth. When a Wright respirometer is used clinically for a patient, a face mask is attached to the respirometer for measurement. However, this evaluation may involve an air leak from the face mask. The reliability of the Wright respirometer had not been well studied. In a previous study, we investigated the intra-rater reliability of the Wright respirometer to determine the variability of the results obtained between measurements taken by a single physical therapist11), and we observed good-to-excellent results. The inter-rater reliability of the Wright respirometer for patients with NMDs has not been known.

Our hypothesis was that the Wright respirometer measurements would show high inter-rater reliability among examiners. We conducted the present study to examine the inter-rater reliability of the Wright respirometer between therapists assessing the VC of patients with NMDs.

PARTICIPANTS AND METHODS

The participants were NMD patients who were admitted to our hospital or outpatient clinic. The eligibility criteria were (1) diagnosed with NMD, (2) VC <2,000 mL, and (3) %VC <50%. The exclusion criteria were (1) tracheotomy and (2) patients judged to be unable to voluntarily control their respiratory movements. The study protocol was approved by the institutional review board of National Hospital Organization Miyazaki Higashi Hospital (approval no. 25-13). All patients signed an informed consent form.

We performed a test-retest method after measuring each patient’s VC. Two physical therapists experienced in the use of the Wright respirometer, i.e., therapist A (sixth year of clinical experience) and therapist B (first year of clinical experience) each took one measurement of the VC of the same patients. The measurement sequence of therapists A and B was randomized using a random number table. The measurements between the therapists were taken at intervals of 3–7 days. We used the Wright respirometer (Halo/scale Wright respirometer standard type, nSpire Health, Hertford, UK), with a silicone face mask (No.4/5 + Multifunctional mask cover, Laerdal Medical Japan, Tokyo) (Fig. 1). First, the measurement procedure and precautions regarding the measurement method were explained to the patient. The patient then practiced a few times with the respirometer, and after confirming that there was no problem in the calling method, the VC measurement was started. We created a manual to standardize the measurements among therapists11) (Fig. 2). A set of standard instructions was given to all patients. The patient’s posture during the measurements was in a wheelchair sitting position in all cases. The measurement location was a rehabilitation room or at the patient’s bedside. The patients were not informed of the measured values.

Fig. 1.

Fig. 1.

Wright respirometer.

Fig. 2.

Fig. 2.

Voice credit at the time of the vital capacity (VC) measurements used in this study.

All measurements took the following four points into account. First, the patient should have no fever or abnormal vital signs. Second, measurements should be in the same environment and at the same time of day as much as possible. Third, we did not implement inspiratory muscle training, which causes fatigue in the respiratory muscles. Fourth, the patient had had enough rest between the measurements.

We analyzed the relative and absolute inter-rater reliability using the intraclass correlation coefficient (ICC2,1) with 95% confidence intervals (95% CIs). We conducted Bland-Altman analyses12) to identify systematic bias between the two therapists. We investigated the measurement variability by calculating the standard error of measurement (SEM) and the 95% CI of the minimal detectable change (MDC). The MDC exhibits the smallest change that indicates a real improvement for a single individual. The ICC criteria were as follows: ≥0.9 as ‘excellent’, ≥0.8 and <0.9 as ‘good’, ≥0.7 and <0.8 as ‘normal’, ≥0.6 and <0.7 as ‘possible’, and <0.6 as ‘reconsidered’13, 14). The significance level was <5% on both sides in all analyses. All statistical analyses were done with R 2.8.1 (CRAN, free software).

RESULTS

We examined 18 patients (age median, range: 51, 14–83 years; 12 females, 6 males; height 151.8 ± 7.2 cm; weight 45.1 ± 12.0 kg; body mass index [BMI] 19.5 ± 4.5 kg/m2) with NMDs including myotonic dystrophy (MD) (n=6), Duchenne muscular dystrophy (DMD) (n=3), limb girdle muscular dystrophy (LGMD) (n=2), progressive muscular dystrophy (n=2), myopathy (n=2), Fukuyama congenital muscular dystrophy (FCMD) (n=1), facioscapulohumeral muscular dystrophy (FSHD) (n=1), and Gene carriers of DMD (n=1) (Table 1). Eleven patients were fitted with noninvasive positive pressure ventilation (NPPV), and the median wearing time (interquartile range) was 24 (9–24) hr. The measured VC values were 905 ± 490 mL by therapist A and 897 ± 483 mL by therapist B. The inter-rater reliability of ICC2,1 was 0.96 (95% CI: 0.89–0.98) (Table 2). The Bland-Altman analyses revealed neither fixed bias nor proportional bias (Fig. 3). The values for SEM and MDC between therapists A and B were 102 mL and 282 mL, respectively.

Table 1. Characteristics of the 18 patients with neuromuscular disorders.

No. Category Disease Age (years) Gender Height (cm) Weight (kg) BMI (kg/m2) Stage NPPV time (hr) Vital capacity (mL) Sequence (random)

Examiner A Examiner B
1 Hospitalization MD 62 female 143.0 43.1 21.1 9 350 440 A→B
2 Hospitalization PMD 65 female 151.0 37.4 16.4 0 1,110 1,120 A→B
3 Outpatient DMD 14 male 135.0 36.3 19.9 9 1,220 1,500 B→A
4 Hospitalization Gene carriers of DMD 41 female 150.0 39.7 17.6 24 320 330 B→A
5 Hospitalization MD 52 female 150.0 57.6 25.6 24 1,000 1,030 B→A
6 Hospitalization MD 46 female 153.8 58.5 24.7 20 730 880 B→A
7 Outpatient FCMD 15 female 140.0 40.0 20.4 2 1,260 1,210 B→A
8 Hospitalization DMD 32 male 158.0 33.2 13.3 24 160 220 B→A
9 Outpatient FSHD 58 female 154.2 43.7 18.4 0 1,900 1,750 A→B
10 Outpatient LGMD 54 female 160.0 65.6 25.6 0 1,130 1,210 B→A
11 Hospitalization MD 60 male 152.7 30.8 13.2 24 670 550 A→B
12 Outpatient DMD 31 male 160.0 70.2 27.4 24 320 310 A→B
13 Hospitalization MD 59 male 151.5 38.9 17.0 0 1,200 850 B→A
14 Outpatient MD 50 female 158.0 58.0 23.2 9 790 790 A→B
15 Outpatient LGMD 28 female 146.3 40.8 19.1 0 1,870 1,900 A→B
16 Hospitalization Myopathy 34 female 154.0 28.9 12.2 12 840 720 A→B
17 Hospitalization PMD 83 female 152.0 47.8 20.7 0 600 720 B→A
18 Hospitalization Myopathy 61 male 163.0 40.8 15.4 0 820 620 A→B

BMI: Body mass index; stage: Ministry of Health and Welfare research group functional classification; NPPV: noninvasive positive pressure ventilation; MD: myotonic dystrophy; PMD: progressive muscular dystrophy; DMD: Duchenne muscular dystrophy; FCMD: Fukuyama congenital muscular dystrophy; LGMD: Limb girdle muscular dystrophy; FSHD: facioscapulohumeral muscular dystrophy.

Table 2. The inter-rater reliability of the vital capacity measurement between therapists A and B.

Bland-Altman analysis

ICC2,1 (95% CI) Fixed bias Proportional bias SEM (mL) MDC (mL)


95% CI of the d p value Slope of regression line p value
0.96 (0.89–0.98) −79 to 64 p=0.82 −0.02 p=0.84 102 282

ICC: intraclass correlation coefficients; CI: confidence interval; d: the mean of difference between the two therapists; SEM: standard error of measurement; MDC: 95% confidence interval of the minimal detectable change.

Fig. 3.

Fig. 3.

The Bland-Altman plots showing the differences between measures of VC (mL) from the two therapists’ sessions. The upper and lower horizontal lines represent the mean of the differences and 95% limits of agreement, respectively.

DISCUSSION

The results of the analyses support our hypothesis, i.e., that measurements of NMD patients’ VC by the Wright respirometer would show high inter-rater reliability. The ICC of this study was 0.89–0.98. Our present results were thus good-to-excellent. In our previous investigation, we observed that the intra-rater reliability of the ICC by the same method as that used herein was nearly the same at 0.86–0.9811). On the other hand, another research group reported that the ICC of VC measurements taken with an electronic spirometer was 0.9815). Compared to that study, the reliability of the Wright respirometer observed herein is slightly inferior to that of an electronic spirometer.

The results of the present Bland-Altman analysis did not reveal systematic bias, i.e., fixed bias and proportional bias. According to the Japanese Society of Respiratory Science spirometry handbook, pulmonary function measurements are to be repeated at least twice, ideally three times, until the difference is within 100 mL or within 5%16). In the present investigation, the SEM was 102 mL. Since our Wright respirometer measurement method’s SEM is nearly within the standard value, the method can be used for evaluations among multiple therapists. However, it is desirable for the measurer to take multiple measurements in order to reduce the SEM. The MDC in this study was 282 mL. We also reported that the intra-examiner reliability MDC was 233 mL11). It is necessary to judge the inter-rater reliability by estimating the clinical effect more strictly than the intra-rater reliability.

The strength of this study is that the creation of a unified manual gives a high degree of reliability among therapists (even first-year physical therapists). The small and lightweight Wright respirometer does not require electricity, and it is thus not easily affected by the environment. The evaluation method described herein is very useful for NMD patients living at home because it can be conducted in a short time and easily.

This study has two limitations. First, in the present measurement method, the instructions/script (Fig. 2) is performed only under a standardized condition, and it is not clear whether the reliability actually decreases without the instructions/script. Second, the MDC in this study is a statistical judgment (by a distribution-based method), and in the future it will be necessary to consider including the anchor-based method.

In conclusion, our findings demonstrated the good-to-excellent inter-rater reliability in the use of the Wright respirometer for NMD patients along with the use of our new manual and the instructions.

Co nference presentation

We presented this study for the World Confederation for Physical Therapy Congress 2015 in Singapore.

Co nflict of interest

The authors declare that there is no conflict of interest.

REFERENCES

  • 1.Katz SL, Barrowman N, Monsour A, et al. : Long-term effects of lung volume recruitment on maximal inspiratory capacity and vital capacity in Duchenne muscular dystrophy. Ann Am Thorac Soc, 2016, 13: 217–222. [DOI] [PubMed] [Google Scholar]
  • 2.Buu MC: Respiratory complications, management and treatments for neuromuscular disease in children. Curr Opin Pediatr, 2017, 29: 326–333. [DOI] [PubMed] [Google Scholar]
  • 3.Dohna-Schwake C, Ragette R, Teschler H, et al. : IPPB-assisted coughing in neuromuscular disorders. Pediatr Pulmonol, 2006, 41: 551–557. [DOI] [PubMed] [Google Scholar]
  • 4.Tzeng AC, Bach JR: Prevention of pulmonary morbidity for patients with neuromuscular disease. Chest, 2000, 118: 1390–1396. [DOI] [PubMed] [Google Scholar]
  • 5.Kang SW, Bach JR: Maximum insufflation capacity: vital capacity and cough flows in neuromuscular disease. Am J Phys Med Rehabil, 2000, 79: 222–227. [DOI] [PubMed] [Google Scholar]
  • 6.Bach JR, Bianchi C, Vidigal-Lopes M, et al. : Lung inflation by glossopharyngeal breathing and “air stacking” in Duchenne muscular dystrophy. Am J Phys Med Rehabil, 2007, 86: 295–300. [DOI] [PubMed] [Google Scholar]
  • 7.Bach JR, Mahajan K, Lipa B, et al. : Lung insufflation capacity in neuromuscular disease. Am J Phys Med Rehabil, 2008, 87: 720–725. [DOI] [PubMed] [Google Scholar]
  • 8.Hull J, Aniapravan R, Chan E, et al. : British Thoracic Society guideline for respiratory management of children with neuromuscular weakness. Thorax, 2012, 67: i1–i40. [DOI] [PubMed] [Google Scholar]
  • 9.Park ES, Park JH, Rha DW, et al. : Comparison of the ratio of upper to lower chest wall in children with spastic quadriplegic cerebral palsy and normally developed children. Yonsei Med J, 2006, 47: 237–242. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Hatch DJ, Williams GM: The Haloscale “Infanta” Wright respirometer. An in vitro and in vivo assessment. Br J Anaesth, 1988, 60: 232–238. [DOI] [PubMed] [Google Scholar]
  • 11.Yoshinaga R, Futsuhara H: Relative and absolute intra-rater reliability of a Wright respirometer to measure vital capacity for neuromuscular disorders. Jpn J Natl Med Serv, 2015, 69: 62–68 (in Japanese). [Google Scholar]
  • 12.Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res, 1999, 8: 135–160. [DOI] [PubMed] [Google Scholar]
  • 13.Kuwahara Y, Saito T, Inagaki Y: Intra- and inter-rater reliability. Kokyu To Junkan, 1993, 41: 945–952 (in Japanese). [PubMed] [Google Scholar]
  • 14.Overend T, Anderson C, Sawant A, et al. : Relative and absolute reliability of physical function measures in people with end-stage renal disease. Physiother Can, 2010, 62: 122–128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Tschopp JM, Roulin JP, Juilland A, et al. : [Evaluation of the reliability of 2 portable electronic spirometers]. Schweiz Med Wochenschr, 1988, 118: 1382–1385 (in French). [PubMed] [Google Scholar]
  • 16.Ogawa H, Aizawa H, Ichinose M: Spirometry handbook—easy respiratory function tests that can be performed in daily medical care. Tokyo: Medical Review Company, 2007 (in Japanese). [Google Scholar]