Abstract
Americans spend most of their time indoors at home, but comprehensive characterization of in-home air pollution is limited by the cost and size of reference-quality monitors. We assembled small “Home Health Boxes” (HHBs) to measure indoor PM2.5, PM10, CO2, CO, NO2, and O3 concentrations using filter samplers and low-cost sensors. Nine HHBs were collocated with reference monitors in the kitchen of an occupied home in Fort Collins, Colorado, USA for 168 h while wildfire smoke impacted local air quality. When HHB data were interpreted using gas sensor manufacturers’ calibrations, HHBs and reference monitors (a) categorized the level of each gaseous pollutant similarly (as either low, elevated, or high relative to air quality standards) and (b) both indicated that gas cooking burners were the dominant source of CO and NO2 pollution; however, HHB and reference O3 data were not correlated. When HHB gas sensor data were interpreted using linear mixed calibration models derived via collocation with reference monitors, root-mean-square error decreased for CO2 (from 408 to 58 ppm), CO (645 to 572 ppb), NO2 (22 to 14 ppb), and O3 (21 to 7 ppb); additionally, correlation between HHB and reference O3 data improved (Pearson’s r increased from 0.02 to 0.75). Mean 168-h PM2.5 and PM10 concentrations derived from nine filter samples were 19.4 μg m−3 (6.1% relative standard deviation [RSD]) and 40.1 μg m−3 (7.6% RSD). The 168-h PM2.5 concentration was overestimated by PMS5003 sensors (median sensor/filter ratio = 1.7) and underestimated slightly by SPS30 sensors (median sensor/filter ratio = 0.91).
Keywords: indoor air quality, household air pollution, electrochemical gas sensors, particulate matter, NO2
Graphical Abstract

1. INTRODUCTION
Exposure to air pollution is a leading risk factor for human morbidity and mortality [1]. To protect human health, government agencies, such as the United States Environmental Protection Agency (US EPA), have promulgated standards for outdoor concentrations of air pollutants including particulate matter (PM), carbon monoxide (CO), nitrogen dioxide (NO2), and ozone (O3) [2]; however, Americans typically spend 60% to 85% of their time at home and, as a result, more than half of personal exposure to air pollutants such as fine particulate matter (PM2.5) typically occurs in the indoor home environment [3-7]. Outdoor air pollution (e.g., from industrial sources, motor vehicles, and wildfires) can enter the home via open windows, mechanical ventilation systems, and infiltration of the building envelope [8]. Additionally, indoor sources can produce airborne pollutant concentrations that exceed outdoor levels [8-11]. For example, use of natural gas-fueled cooking burners increases indoor CO2, CO, NOx, and particle number concentrations [9-11]. Singer et al. found that use of gas-fueled burners and ovens during simulated cooking activities in homes with the windows closed, forced-air ventilation turned off, and the range exhaust hood turned off often resulted in 1-h average kitchen NO2 concentrations above the US EPA outdoor 1-h standard of 100 ppb [10]. Even when electric or induction burners are used, PM2.5 emitted from cooking food can lead to peak indoor concentrations > 100 μg m−3 [12,13].
Time-resolved data on home air quality can inform occupants, along with health and air quality professionals, about trends in and sources of air pollutants in a specific home and might help them take direct action, seek resources, and/or advocate for policies to mitigate those pollutants; however, efforts to collect such data are limited by the costs of instrumentation and sample analysis. Low-cost sensors can reduce the hardware costs associated with acquiring time-resolved pollutant concentration data, but low-cost sensor data must be interpreted with care. For example, the inaccuracies associated with sensors that convert particle light scattering signals to PM mass concentrations are well-documented [14-20]. The relationship between scattered light intensity and aerosol mass concentration depends on the shape, refractive index, size distribution, and density of the particles [21]. Without a mechanism for drying incoming particles, low-cost sensors can overestimate dry PM mass concentrations when hygroscopic particles take up water in high humidity environments [18,22]. In addition, many low-cost light scattering sensors fail to respond to increases in coarse (2.5 to 10 μm) particle mass—most likely because of particle losses in the flow path between the inlet and the light source/detector as well as decreases in the amount of light scattered over the angular detection range per unit mass [16,20,23,24].
Several recent publications have documented the challenges associated with converting raw outputs from low-cost NDIR CO2 sensors—as well as electrochemical CO, NO2, and O3 sensors—into accurate pollutant concentrations. Most studies describe outdoor monitoring applications in which temperature- and humidity-sensitive gas sensors were subjected to variable environmental conditions coupled with low pollutant concentrations [25-32]. In many of these studies, low-cost sensors were collocated with reference (i.e., research- or regulatory-grade) monitors for several months and the collocation data were then used to train linear and/or machine learning calibration models. Though effective, these calibration approaches present barriers to users who lack access to reference monitors and increase the effective “cost” of the sensors.
Few publications have described indoor monitoring with low-cost gas sensors [33,34]. We are not aware of any studies in which (a) low-cost CO2, CO, NO2, and O3 sensors were collocated with a comprehensive suite of reference instruments in an indoor environment or (b) data on the efficacy of sensor manufacturer-supplied calibrations were reported. Low-cost gas sensors might provide more reliable data indoors, where temperature and relative humidity (RH) vary less than outdoors and where indoor sources can elevate CO2, CO, and NO2 concentrations above outdoor levels.
In the present study, we developed a prototype platform—called the “Home Health Box” (HHB)— that used time-integrated pollutant samplers and low-cost sensors to measure PM2.5, PM10, CO2, CO, NO2, O3, and volatile organic compound (VOC) concentrations. We deployed nine HHBs alongside reference monitors in the kitchen of an occupied home in Fort Collins, CO, USA for one week to investigate: (1) how CO2, CO, NO2, O3, and PM2.5 concentrations in the kitchen varied with outdoor pollution levels and in-home activities; (2) whether CO2, CO, NO2, and O3 concentrations calculated using the low-cost sensor manufacturers’ calibration data agreed with concentrations reported by reference monitors; (3) whether linear calibration models developed through collocation with reference monitors improved the accuracy of CO2, CO, NO2, and O3 concentrations derived from low-cost sensor data; (4) the accuracy and precision of PM concentrations measured using filter samplers and low-cost sensors in collocated HHBs; (5) whether reference monitor and low-cost sensor data led to the same conclusions about whether pollutant concentrations were low, elevated, or high relative to various air quality standards; as well as (6) whether reference monitor and low-cost sensor data led to the same conclusions regarding which in-home activities contributed disproportionately to the time-integrated concentration of each pollutant.
2. MATERIAL AND METHODS
2.1. Home Health Box
The HHB (Figure 1) was designed to sample PM2.5, PM10, and VOCs while measuring PM, CO2, CO, NO2, and O3 concentrations continuously. Inside each HHB, two modified ultrasonic personal aerosol samplers (UPAS, Access Sensor Technologies, Fort Collins, CO, USA) collected PM2.5 and PM10 onto 37-mm PTFE filters (PT37P-PF03, MTL, Minneapolis, MN, USA) at 2 L min−1 [35,36]. A third modified UPAS sampled VOCs onto a thermal desorption tube at 0.003 L min−1. Two low-cost sensors (PMS5003, Plantower, Beijing, China and SPS30, Sensirion, Stäfa, Switzerland) employed light-scattering techniques to measure real-time variations in PM concentrations. Concentrations of CO2 were measured using a low-cost NDIR sensor (SCD30, Sensirion, Stäfa, Switzerland). Concentrations of CO, NO2, and O3 were measured using electrochemical sensors (CO-B4, NO2-B43F, and OX-B431 Alphasense, Great Notley, Essex, United Kingdom) mounted to Individual Sensor Boards (000-0ISB-00 and 000-0ISB-02, Alphasense). The CO2, CO, NO2, and O3 sensors were isolated from the rest of the internal components by a 3D-printed housing (Figure S1). This housing had one inlet and one outlet on the outside of the main HHB enclosure. The NO2 and O3 sensors were installed immediately downstream of the inlet to minimize losses of these species (but we did not characterize losses). Air flowed through the gas sensor housing continuously with the aid of a small fan installed on the outlet (UF3H3-710, Sunon, Kaohsiung City, Taiwan). Temperature and RH inside the gas sensor housing were measured by a sensor included on the SCD30 circuit board.
Figure 1.
Left: The Home Health Box enclosure had exterior dimensions of 220 × 170 × 130 mm (excluding the feet, sample inlets, exhaust vents, hinge, and latch) and weighed approximately 2 kg. Right: A top view of the Home Health Box with the cover removed to show internal components.
The HHB enclosure (YH-080604, Polycase, Avon, OH, USA) had exterior dimensions of approximately 220 × 170 × 130 mm. The HHB was designed to plug into a 120-V wall outlet but was equipped with a backup battery (V25, Voltaic Systems, Brooklyn, NY, USA) to prevent sample interruption during a short-term (< 6 h) loss of line power. The HHB weighed ~2 kg. Firmware running on an internal microcontroller (STM32L475VGT6, STMicroelectronics, Geneva, Switzerland) controlled sample flow rates, the sample duration, and data logging (Table S1).
2.2. In-Home Experiment
Approval for in-home sampling with the Home Health Boxes was obtained from the Institutional Review Board at Colorado State University (protocol number 19-9443H).
2.2.1. Experimental Setup
The experiment took place in the kitchen of a 235-m2 detached house in Fort Collins, Colorado, USA. The house—which was built in 1991—featured a ground floor (where the kitchen was located), a partially-finished basement, an attached two-car garage, and central air conditioning.
The 4.3 × 5.1-m kitchen was enclosed on three sides (north, west, and south) by two exterior walls and one interior wall (Figure S2). The ceiling height ranged from 2.5 m (at the west exterior wall) to 3.6 m (on the east side where the kitchen was connected to the main living area). The wall on the fourth (east) side of the kitchen extended from the floor to a height of 2.2 m (i.e., there was a >1 m gap between the top of this wall and the ceiling). A 1.6-m-wide by 2.1-m-tall opening (with no door) in the east wall served as a walkway between the kitchen and the living area.
Unvented natural gas cooking burners were present in the stove installed along the south wall (MGR7662WQ, Maytag, Benton Harbor, MI, USA). The left front, right front, left rear, and right rear cooktop burner firing rates were specified as 13.5, 12.5, 9.5, and 5.0 kbtu hr−1, respectively. The oven and broiler firing rates were specified as 16.5 and 10.0 kbtu hr−1, respectively. The hood installed above the cooktop (233.5338910, Kenmore, Hoffman Estates, IL, USA) exhausted back into the kitchen, instead of to the outdoors, and was not typically used by home occupants.
A portable air cleaner (AP-1512HH, Coway, Seoul, Korea) was also present in the kitchen. This air cleaner, which was rated for a 33.5 m2 room, passed air through a carbon “pre-filter”, a HEPA filter, and a bipolar ionizer. The carbon and HEPA filters were replaced just before the experiment started. A button on the unit allowed the ionizer to be switched on and off separately.
We deployed nine Home Health Boxes in the kitchen to test their ability to measure a wide range of PM, CO2, CO, NO2, and O3 concentrations resulting from indoor sources (e.g., emissions from natural gas cooking burners and cooking food) as well as infiltration of outdoor air pollution. Each HHB wrote timestamped data to a microSD card at 30-s intervals.
Reference measurements of indoor CO2, CO, NO, NOx, O3, and PM2.5 levels were logged at 60-s intervals using an LI-820 CO2 Gas Analyzer (LI-COR Biosciences, Lincoln, NE, USA), two QTrak Indoor Air Quality Monitors (7575-X with model 982 probe, TSI Incorporated, Shoreview, MN, USA), a Trace Level Chemiluminescence NO-NO2-NOx Analyzer (Model 42C, Thermo Environmental Instruments, Franklin, MA, USA), a UV Photometric O3 Analyzer (Model 49C, Thermo Environmental Instruments, Franklin, MA, USA), and a Tapered Element Oscillating Microbalance (TEOM 1405, Fisher Scientific, Waltham, MA, USA). The TEOM sample flow rate was set to 4 L min−1 and a GK2.05 (KTL) cyclone (2.5 μm cutpoint at 4 L min−1; Mesa Laboratories, Butler, NJ, USA) was installed on the sample inlet (i.e., the 16.7 L min−1 inlet typically used for regulatory monitoring was not used). See Supporting Information (SI) Section S1 for additional information on how these instruments were calibrated and operated.
We aimed to place the indoor monitors as close together as possible given the constraints imposed by the kitchen layout and the fact that the home was occupied during the experiment. The nine HHBs were stacked three high and arranged into three columns spaced 0.91 m apart on a shelf 1.3 m above the cooktop (Figure 2). The center column was aligned with the cooktop. The LI-820 and the two QTraks were set on top of a microwave 1.1 m to the right of the cooktop (from the perspective of a person standing in front of the oven). The 42C NO-NO2-NOx Analyzer and 49C O3 Analyzer were installed along the north wall. A PTFE sample line leading to these analyzers was secured 1.1 m above and 3.2 m away from the cooktop (Figure S2). The TEOM was placed on the same shelf as the HHBs, 1.9 m to the left of the cooktop.
Figure 2.
Placement of the Home Health Boxes, LI-820 CO2 Gas Analyzer, QTrak Indoor Air Quality Monitors (used to measure CO), and TEOM 1405 (used to measure PM2.5) relative to the natural gas cooking burners along the south interior kitchen wall. The >1-m gap between the top of the wall and the ceiling on the east side of the kitchen is visible around the TEOM. The 42C NO-NO2-NOx Analyzer and 49C O3 Analyzer were installed in the same room but are not visible from this perspective. Occupants’ personal effects have been blurred in the photo.
Outdoor particulate matter concentrations and meteorological conditions were recorded at 30-s intervals using a second-generation Aerosol Sensor Plus Environmental Node (ASPEN) Box [17] installed 1.5 m outside the north kitchen exterior wall (Figure S2). This ASPEN Box included a PM2.5 filter sampler, a PM10 filter sampler, a PMS5003 sensor, an SPS30 sensor, as well as a temperature and RH sensor (SHT30, Sensirion) installed in a radiation shield.
City-level, 10-minute average temperature, RH, and solar radiation data were obtained from the Fort Collins Weather Station located 4.0 km from the home on the Colorado State University main campus [37]. City-level, 1-h ambient CO and O3 concentrations were obtained from monitors (48i-TLE, Thermo Scientific, Franklin, MA, USA and 400E, Teledyne API, San Diego, CA, USA, respectively) operated by the Colorado Department of Public Health and Environment (CDPHE) at EPA AQS site 08-069-1004 (4.3 km from the home). City-level, 1-h ambient PM2.5 and PM10 concentrations were obtained from a monitor (EDM180, GRIMM Aerosol Technik, Ainring, Germany) operated by CDPHE at AQS site 08-069-0009 (3.7 km from the home).
2.2.2. Activities
The 168-h experiment spanned October 8 to 15, 2020 UTC. During the experiment, outdoor PM concentrations were elevated due to the Cameron Peak fire burning west of Fort Collins and the Mullen fire burning 130 km northwest of Fort Collins (Figure S3) [38,39]. The Cameron Peak fire grew from 131,000 acres on October 7th to 164,000 acres on October 14th [38,40].
The house was occupied by three unrelated adults and two dogs. All adults were employed outside the home, but one worked almost exclusively from home due to the COVID-19 pandemic. During the experiment, all monitors measured air pollution from normal occupant activities as well as scripted activities that were designed to purposefully increase indoor NO2 or O3 concentrations.
Occupants kept a log of normal (i.e., non-scripted) activities that might have affected indoor pollutant concentrations (e.g., cooking, opening windows). After the experiment, this log was reviewed alongside time-series reference monitor data to see if occupants could identify any pollution-generating or pollution-depleting events that were missing from the original log.
To modulate the NO2 concentration in the kitchen, an occupant completed six scripted activities with natural gas cooking burners. The occupant boiled 3 L of water in a pot on the left front cooktop burner on five separate occasions (Table 1): (1) while all windows and doors were closed, the air conditioner was off, and the air cleaner was off; (2, 3) while the air cleaner HEPA filter was on; (4) while two windows and a sliding glass door were open; and (5) while the air conditioner was on. On October 14th, the occupant operated the broiler for approximately 8 h with the oven door closed and with no food inside the oven. While the broiler was on, all windows and doors to the outside remained closed, the air conditioner remained off, and the air cleaner remained off. This last activity was designed to capture the high indoor NO2 concentration that could result from use of a gas burner during long-duration cooking (e.g., in preparation for a holiday meal) [41,42].
Table 1.
A summary of the scripted activities that an occupant completed to purposefully increase NO2 and O3 concentrations in the kitchen. All times are the local time in Fort Collins, CO, USA (UTC-06:00 on the dates listed). The “HEPA filter” and “Ionizer” columns represent two separate functions of the indoor air cleaner.
| Target pollutant |
Description | Start time | End time | Door & window |
AC | HEPA filter |
Ionizer |
|---|---|---|---|---|---|---|---|
| NO2 | Boiled 3 L of water | 08 Oct. 11:25 | 08 Oct. 13:05 | Closed | Off | Off | Off |
| NO2 | Boiled 3 L of water | 12 Oct. 09:40 | 12 Oct. 10:20 | Closed | Off | On | Off |
| NO2 | Boiled 3 L of water | 12 Oct. 10:45 | 12 Oct. 11:35 | Closed | Off | On | Off |
| NO2 | Boiled 3 L of water | 12 Oct. 13:30 | 12 Oct. 15:55 | Open | Off | Off | Off |
| NO2 | Boiled 3 L of water | 13 Oct. 09:50 | 13 Oct. 13:05 | Closed | On | Off | Off |
| NO2 | Turned on broiler | 14 Oct. 10:30 | 14 Oct. 16:50 | Closed | Off | Off | Off |
| O3 | Opened door and window | 08 Oct. 14:30 | 08 Oct. 16:00 | Open | Off | Off | Off |
| O3 | Opened door and window | 11 Oct. 09:00 | 11 Oct. 12:00 | Open | Off | Off | Off |
| O3 | Turned on ionizer | 10 Oct. 13:30 | 11 Oct. 10:00 | Closed | Off | On | On |
To modulate the O3 concentration in the kitchen, an occupant completed two scripted activities in which two windows and a sliding glass door were opened while all cooking burners remained off. The air conditioner and air cleaner were both off on these occasions. To investigate whether the bipolar ionizer increased the O3 concentration in the kitchen, we scripted a separate activity in which an occupant turned on the HEPA filter and ionizer for 21.5 hours while the doors and windows remained closed and the air conditioner remained off.
2.2.3. Sample Analyses
Ten PM2.5 filter samples and ten PM10 filter samples (nine HHB + one ASPEN) were collected during the experiment. The 37-mm PTFE filters (PT37P-PF03, Measurement Technology Laboratories, Minneapolis, MN, USA) used to collect these samples were pre- and post-weighed on a balance with 1-μg resolution (XS3DU, Mettler-Toledo, Columbus, OH, USA) [43].
The VOC samples collected using the HHBs were not analyzed and, thus, data on indoor VOC levels are not presented herein.
2.2.4. Data Analyses
2.2.4.1. Reference Monitor Data
The two CO concentrations measured using the two QTrak monitors during each minute were averaged. This average concentration was used as the reference CO concentration. Additionally, at each 60-s interval, the indoor NO2 concentration was calculated as the difference between the NOx and NO concentrations reported by the 42C NO-NO2-NOx Analyzer.
2.2.4.2. Gas Sensor Calibrations
First, CO, NO2, and O3 concentrations were estimated using algorithms and sensor-specific calibration data provided by Alphasense. During this process, the “corrected working electrode voltage” (WEc; mV) had to be calculated for each sensor serial number i at each 30-s log interval j. Alphasense provided four possible algorithms for calculating WEc,ij [44]:
| (1) |
| (2) |
| (3) |
| (4) |
The pollutant concentration measured by sensor serial number i at time j (cij; ppb) was then calculated from the corrected working electrode voltage using Equation 5 [44]:
| (5) |
In Equations 1-5, WEu,ij and AEu,ij were the uncorrected working and auxiliary electrode voltages recorded by sensor serial number i at time j (mV), WEe,i and AEe,i were the working and auxiliary electrode electronic offsets for individual sensor board i (mV); WE0,i and AE0,i were the working and auxiliary electrode zero voltages for sensor i (mV); nT,ij, kT,ij, k′T,ij and k″T,ij were temperature-dependent correction factors; rT,ij was the relative sensitivity as a function of temperature; and si was the sensitivity of sensor serial number i to the relevant pollutant (mV ppb−1). WEe, AEe, WE0, AE0, and s were calibration constants provided by Alphasense for each sensor serial number i. The correction factors nT, kT, k′T, k″T, and rT were provided by Alphasense, in tabular form (nT, kT, k′T, k″T) or graphical form (rT), as a function of sensor part number (CO-B4, NO2-B43F, or OX-B431) and temperature [44-47]. These correction factors were evaluated at the temperature measured inside gas sensor housing i at time j (Tij; °C) by interpolating between values in the table or graph.
Because the OX-B431 sensor responds to both O3 and NO2, it must be used in conjunction with the NO2-B43F sensor to determine O3 concentrations. To calculate O3 concentrations, all instances of (WEu,ij – WEe,i) in Equations 1-4 were replaced with (WEu,ij – WEe,i – WEc,NO2,ij), where WEc,NO2,ij (mV) was the product of the sensitivity of the OX-B431 sensor to NO2 (sNO2,i; mV ppb−1) and the NO2 concentration calculated from the NO2-B43F data at time j using Equations 4 and 5 (cNO2,ij) [44]:
| (6) |
Each gas concentration (CO, NO2, O3) was calculated using each of the four algorithms at each 30-s interval in the 168-h experiment. If the concentration (cij) calculated using Equation 5 was below 0, the calculated value was replaced with zero.
Second, we fit our own linear mixed models to predict the 1-h average gas concentration measured at time j (cj; ppb) by the reference monitor(s) for the pollutant of interest (cCO,j for CO-B4 sensors, cNO2,j for NO2-B43F sensors, and cNO2,j + cO3,j for OX-B431 sensors) from 1-h average working electrode voltages, temperatures, and RH values:
| (7) |
| (8) |
| (9) |
where ϵij was the random error. These models included the working electrode voltage as a predictor, and not the auxiliary electrode voltage, because (a) (WEu,ij – WEe,i) was strongly correlated with the reference gas concentration for all three sensors and (b) Equation 4, which did not use the auxiliary electrode voltage, was the best-performing Alphasense algorithm for the CO-B4 and NO2-B43F sensors (see Section 3.2.1). Coefficient α was a fixed intercept, β was a fixed slope, ai was a random intercept for sensor serial number i, and bi was a random slope for sensor serial number i. The random intercept and slope were included because (a) Alphasense provided a unique sensitivity (si in Equation 5; mV ppb−1) for each serial number, thereby implying that each individual sensor had a unique response to the pollutant of interest, and (b) model results indicated that there was random variation in the intercept and slopes between serial numbers. Tij and RHij were the temperature (°C) and relative humidity (%), respectively, inside gas sensor housing i at time j. The coefficients γT and γRH remained constant across all sensors with a given part number (CO-B4, NO2-B43F, or OX-B431). Temperature and RH were included as predictors in Equations 8 and 9, respectively, because documents from Alphasense [44-48], and prior studies [25,28-31], indicated that temperature and RH influenced sensor performance.
We fit the linear mixed model shown in Equation 10 to relate 1-h average CO2 concentrations reported by the LI-820 (cCO2,LI–820,j; ppm) to 1-h average CO2 concentrations reported by the low-cost NDIR sensors (CCO2,SCD30,ij; ppm):
| (10) |
where α was a fixed intercept, ai was a random intercept for sensor serial number i, γ was a fixed slope, and ϵij was the random error. Sensirion’s documentation for the SCD30 did not suggest that a random slope would be required for each sensor serial number. A model that also included a random slope for each serial number was considered, but the between-serial number variance associated with the random slope was zero—indicating that it was not needed.
The models shown in Equations 7-10 were trained and tested using seven-fold cross validation. Each fold consisted of a continuous 24-h period spanning midnight to midnight UTC. Models fit to the other 144 h of data were used to predict 1-h average concentrations in a fold. This process was repeated seven times to obtain predictions for the full 168-h duration of the experiment. The models shown in Equations 7-9 were used to predict 1-h average CO, NO2, and NO2 + O3 concentrations. The predicted O3 concentration was calculated by subtracting the predicted NO2 concentration from the predicted NO2 + O3 concentration. The model shown in Equation 10 was used to predict each 1-h average CO2 concentration by subtracting the intercepts from the sensor-reported CO2 concentration and dividing the difference by the slope: (cCO2,SCD30,ij – α – ai)/γ.
We compared 1-h average concentrations (a) obtained directly from the SCD30 CO2 sensors in the HHBs, (b) calculated from HHB data using the Alphasense-recommended algorithms shown in Equations 1-4, and (c) predicted from HHB data using the linear mixed models shown in Equation 7-10 to 1-h average concentrations measured by the reference monitors to determine the root-mean-square error (RMSE), mean absolute error (MAE), mean bias error (MBE), normalized mean absolute error (NMAE), Pearson correlation coefficient, precision, and bias for each sensor serial number i (SI Section S1.2.5) [27-29,31,49,50].
Electrochemical gas sensor data from one HHB were excluded from the model fitting process and all results because those data consisted of unusually high working and auxiliary electrode voltages. Working and auxiliary voltages read from the excluded NO2-B43F sensor were always > 2 V, whereas working and auxiliary voltages read from the other eight NO2-B43F sensors were always < 0.4 V. In addition, working electrode voltages read from the excluded CO-B4 sensor were several hundred mV higher than those read from the other eight CO-B4 sensors.
2.2.4.3. PM Sensor Calibrations
For each PMS5003 or SPS30 sensor i, the corrected PM2.5 concentration measured at 30-s interval j (cPM2.5,ij; μg m−3) was calculated as:
| (11) |
where was the 168-h average PM2.5 concentration derived from the filter sample collected using HHB i (μg m−3), was the time-averaged PM2.5 concentration reported by sensor i over the duration of the filter sample (μg m−3), and cPM2.5,sensor,ij was the PM2.5 concentration reported by sensor i at time j (μg m−3). The ratio () was the gravimetric correction factor for sensor i. For PMS5003 sensors, “PM2.5 CF=1” values were used as cPM2.5,sensor.
We compared 1-h average values of (a) cPM2.5,sensor,ij and (b) cPM2.5,ij obtained from HHB data to 1-h average PM2.5 concentrations measured using the TEOM during the first 48 h of the experiment to determine the RMSE, MAE, MBE, NMAE, and Pearson correlation coefficient for each sensor serial number i. This analysis only spanned the first 48 h of the experiment because the TEOM stopped operating reliably after that time.
We also compared 24-h average values of cPM2.5,ij calculated from HHB PMS5003 sensor data using (a) Equation 11 and (b) a United States-wide correction recently developed for outdoor PurpleAir monitors by Barkjohn et al. [51] to 24-h average PM2.5 concentrations measured by the TEOM during the first 48 h of the experiment (Equation S9). For each of these two correction approaches, the RMSE, MAE, MBE, and NMAE were calculated from the eighteen 24-h PM2.5 concentrations obtained from the nine HHBs over this 48-h period.
2.2.4.4. Classification of Pollutant Concentrations
Prior to the experiment, we developed a tiered system to classify the level of each pollutant (CO2, CO, NO2, O3, PM2.5, and PM10) as low, elevated, or high based on guidelines from the World Health Organization (WHO); US EPA; American Society of Heating, Refrigerating, and Air-Conditioning Engineers (ASHRAE); American Conference of Governmental Industrial Hygienists (ACGIH); and National Research Council (NRC) (see Table 2 and SI Section S1.2.5) [2,52-58]. For a sample that did not include scripted activities, this system could inform home occupants of whether, over the course of a full week, any pollutants were detected at levels that might motivate mitigation. Each gaseous pollutant was classified based on the highest 1-h block average or 8-h rolling average concentration measured during the experiment. The classification obtained from CO2 reference monitor data was compared to the classification obtained from uncorrected CO2 concentrations reported by the low-cost NDIR sensors in the HHBs. The classifications obtained from CO, NO2, and O3 reference monitor data were compared to the classifications obtained from CO, NO2, and O3 concentrations derived from HHB data using the best-performing Alphasense algorithms (Equations 4, 4, and 3, respectively). Each PM size fraction was classified based on the 168-h average concentrations derived from HHB filter samples.
Table 2.
Columns 3–6: Averaging periods and concentration ranges used to classify the level of each pollutant as “low,” “elevated,” or “high.” Columns 7–8: The highest 1-, 8-, or 168-h average concentration of the given pollutant. Columns 9–10: The classifications determined using the reference monitor and the Home Health Boxes. Concentrations shown in the “HHB” column were obtained using uncorrected CO2 data; the best-performing Alphasense algorithms for CO, NO2, and O3 (Equation 4, Equation 4, and Equation 3, respectively); as well as PM2.5 and PM10 filter samples. For “HHB” concentrations, the first value represents the median and the values in parentheses represent the range across all eight (for CO, NO2, and O3) or nine (for CO2, PM2.5, and PM10) boxes. O3 was classified using the median HHB.
| Pollutant | Units | Averaging period (h) |
Range for each classification |
Highest average measured |
Classification |
||||
|---|---|---|---|---|---|---|---|---|---|
| Low | Elevated | High | Reference | HHB | Reference | HHB | |||
| CO2 | ppm | 8 | < 1200 | 1200 – 5000 | > 5000 | 2198 | 1597 (1494, 1745) | Elevated | Elevated |
| CO | ppm | 8 | < 2 | 2 – 9 | > 9 | 6 | 6 (6, 6) | Elevated | Elevated |
| CO | ppm | 1 | < 6 | 6 – 35 | > 35 | 10 | 8 (8, 9) | Elevated | Elevated |
| NO2 | ppb | 1 | < 53 | 53 – 100 | > 100 | 219 | 246 (211, 271) | High | High |
| O3 | ppb | 8 | < 20 | 20 – 70 | > 70 | 31 | 59 (2, 74) | Elevated | Elevated |
| PM2.5 | μg m−3 | 168 | < 12 | 12 – 25 | > 25 | NA | 19 (17, 21) | NA | Elevated |
| PM10 | μg m−3 | 168 | < 20 | 20 – 50 | > 50 | NA | 40 (35, 46) | NA | Elevated |
For CO2, CO, NO2, and O3, we also compared each 1-h block or 8-h rolling average concentration obtained using each HHB to the corresponding reference concentration to determine the fraction of 1-h block or 8-h rolling averages for which the two concentrations resulted in the same classification. This analysis was completed twice using: (1) uncorrected CO2 concentrations reported by the low-cost NDIR sensors in the HHBs as well as CO, NO2, and O3 concentrations derived from HHB data using the best-performing Alphasense algorithms (Equations 4, 4, and 3, respectively) and (2) CO2, CO, NO2, and O3 concentrations calculated from HHB data using the best-performing empirical calibration models (Equations 10, 9, 8, and 9, respectively).
To investigate how occupant activities influenced indoor pollutant concentrations, we grouped activities into the following categories: door/windows open, air conditioner on, air cleaner HEPA filter on, air cleaner HEPA filter + ionizer on, scripted natural gas cooking burner use (with no food cooking), cooking (with one or more natural gas burners on), and none. Scripted burner use and cooking took precedence over all other activities. An open door and windows took precedence over the air cleaner being on. In other words, if scripted burner use occurred while the door and windows were open, that period was classified as “burner on.” If the door and windows were open while the air cleaner was on, that period was classified as “door/windows open.” Our experimental design dictated that scripted burner use and normal cooking never took place at the same time. Due to normal occupant behavior, the door and windows were never open while the air conditioner was on, and the air cleaner was never on while the air conditioner was on.
We calculated a concentration-time ratio (c: t)ik for each combination of pollutant i and activity k:
| (12) |
where cij was the concentration of pollutant i at time j (ppb or ppm) and (tj – tj–1) represents the data logging interval for pollutant i (s). In the numerator of Equation 12, cij(tj – tj–1) is summed over all times associated with activity category k (i.e., j ⊆ Sk) and then divided by cij(tj – tj–1) summed over the entire duration of the experiment (i.e., j = 1 … J). In the denominator, (tj – tj–1) is summed over all times associated with activity category k and then divided by (tj – tj–1) summed over the entire duration of the experiment [4]. A concentration-time ratio >1 indicated that activity k was associated with a disproportionately high fraction of the time-integrated concentration of pollutant i, relative to the amount of time associated with activity k [4].
We calculated concentration-time ratios using: (1) CO2, CO, NO2, and O3 concentrations measured using reference monitors at 60-s intervals; (2) uncorrected CO2 concentrations reported by the low-cost NDIR sensors in the HHBs as well as CO, NO2, and O3 concentrations derived from HHB data using the best-performing Alphasense algorithms (Equations 4, 4, and 3, respectively) at 30-s intervals; and (3) 1-h average CO2, CO, NO2, and O3 concentrations derived from HHB data using the best-performing empirical calibration models (Equations 10, 9, 8, and 9, respectively). One-hour averages spanned each clock hour (e.g., 00:00:01 to 01:00:00 HH:MM:SS). If an activity occurred during any portion of a given hour, the full hour was associated with that activity for the purpose of calculating time-activity ratios from 1 -h average concentrations. The eight (for CO, NO2, and O3) or nine (for CO2) concentration-time ratios calculated for each pollutant in steps 2 and 3 were averaged. Then, we compared concentration-time ratios calculated in step 1 to concentration-time ratios calculated in steps 2 and 3, respectively, to see if reference monitor and HHB data led to the same conclusions about the relative contributions of different activity categories to indoor air pollution.
3. RESULTS AND DISCUSSION
3.1. Pollutant Concentrations over Time
Time-series plots illustrating the range of in-kitchen pollutant concentrations measured during the 168-h experiment are shown in Figure 3. As expected, scripted use of and normal cooking with natural gas burners resulted in elevated CO2, CO, and NO2 concentrations. The highest 15-minute average reference CO2 concentration (3116 ppm; Table S2 and Figure S4) was measured when the broiler burner remained on for 8.3 h to simulate an all-day cooking event on October 14th. The highest 15-minute average reference CO concentration (14 ppm) was measured during normal cooking on the evening of October 11th.
Figure 3.
Concentrations of CO2, CO, NO2, O3, and PM2.5 measured over the duration of the experiment. Solid lines represent 15-minute average indoor concentrations reported by reference monitors. Dashed lines and gray shaded areas represent the median and total range, respectively, of 15-minute average concentrations measured using eight (for CO, NO2, and O3) or nine (for CO2 and PM2.5) Home Health Boxes (HHBs). HHB data include uncorrected CO2 concentrations, CO and NO2 concentrations calculated using Equation 4, O3 concentrations calculated using Equation 3, as well as filter-corrected PMS5003-reported PM2.5 concentrations. Dotted lines represent 1-h average concentrations reported at Colorado Department of Public Health and Environment (CDPHE) monitoring sites located 4.3 km (CO and O3) and 3.7 km (PM2.5) from the home. Horizontal bars along the top of each panel indicate when doors and/or windows were opened, when a central air conditioning unit was on, when an indoor air cleaner was on, when scripted use of natural gas cooking burners (which typically involved boiling water) occurred, and when normal cooking activities occurred. Bars accompanied by a shaded backdrop of the same color represent activities noted in occupants’ original log data. The reference PM2.5 monitor (TEOM) stopped operating reliably approximately 48 h after the experiment began.
Elevated NO2 concentrations were the clearest indicator of natural gas cooking burner use. The median 15-minute average reference NO2 concentration was 15 ppb, but 15-minute average concentrations above 100 ppb were measured during two normal cooking activities and 4/6 scripted burner uses. One of the scripted burner uses for which the 15-minute average reference NO2 concentration remained below 100 ppb occurred while the door and windows were open (on the afternoon of October 12th). The 1-h average reference NO2 concentration was above the US EPA standard of 100 ppb for 17/168 hours (Figure S5). Two of these 17 hours were associated with normal cooking—indicating that typical use of natural gas cooking burners resulted in indoor NO2 concentrations that exceeded outdoor air quality standards.
The highest 15-minute average indoor reference O3 concentrations (50 to 52 ppb) were measured when the door and windows were open but no natural gas cooking burners were on (i.e., no scripted use or normal cooking). When the experiment took place, Fort Collins was in a US EPA 8-h O3 (2015) nonattainment area and the CDPHE monitor 4.3 km from the home reported 1-h average O3 concentrations ranging from 3 to 75 ppb [59]. On October 8th, a peak solar radiation of 698 W m−2 was measured at the weather station at 13:10 (Figure S6), the CDPHE monitor reported a 1-h average O3 concentration of 75 ppb at 15:00, and a peak 15-minute average indoor reference concentration of 50 ppb was measured at 16:00. Transient decreases in the indoor O3 concentration were observed whenever natural gas cooking burners were operated (for scripted activities or normal cooking) with the door and windows closed (Figure 3). Under these conditions, O3 reacted with nitric oxide (NO) emitted by the natural gas cooking burners to form O2 and NO2.
The 168-h averaged PM2.5 concentrations derived from all nine HHB filter samples ranged from 17.2 to 20.8 μg m−3 (median = 19.5 μg m−3; mean = 19.4 μg m−3; relative standard deviation [RSD] = 6.1%; Table S3). The TEOM reference monitor and the real-time PM sensors inside the HHBs detected elevated indoor PM2.5 concentrations during normal cooking (e.g., shortly after 18:00 on October 9th, 12th, and 13th; Figure 3) as well as when the door and/or windows were opened. The impact of wildfire smoke on outdoor air quality throughout Fort Collins was illustrated by peak 1-h average PM2.5 concentrations of 174, 113, 77, and 194 μg m−3 reported at the CDPHE monitoring site on October 8th, 9th, 10th, and 14th, respectively. On the evening of October 8th, the reference PM2.5 concentration increased steadily before reaching a peak 15-minute average value of 66 μg m−3 at 02:15 on October 9th. This increase presumably resulted from occupants opening doors and/or windows to ventilate or cool the home. The summer and early autumn weather in Fort Collins is characterized by hot afternoons and cool nights. The temperature measured outside of the home reached a maximum 15-minute average value of 32 °C on October 8th and a minimum 15-minute average value of 8 °C in the early morning hours of October 9th (Figure S6). Residents often cool their homes by opening windows in the evening—either because they lack air conditioning or want to reduce the expense of operating it. A prior study also indicated that this practice can lead to elevated indoor PM2.5 concentrations during wildfire events [8].
3.2. Sensor Calibration
3.2.1. Gas Sensors
One-hour average SCD30-reported CO2 concentrations had strong, positive, linear correlations with the reference CO2 concentration (Pearson’s r ≥ 0.97); however, mean bias error (MBE) ranged from −495 to −334 ppm. After 1-h CO2 concentrations were corrected using the linear mixed model (Equation 10 and Table S3), MBE ranged from 0 to 2 ppm (Table 3 and Figure S7).
Table 3.
Agreement between 1-h average concentrations calculated from low-cost sensor data and reported by reference monitors. For each pollutant/equation/HHB combination, the root-mean-square error (RMSE) was calculated using all 1-h concentration pairs in the 168-h experiment (for CO2, CO, NO2, and O3) or the first 48 h of the experiment (for PM2.5). The same was done for the mean absolute error (MAE), mean bias error (MBE), normalized mean absolute error (NMAE), and Pearson’s r. All values listed below represent the median and total range of each metric across N HHBs. For CO2 and PM2.5, rows with no equation number represent uncorrected concentrations output by the sensor. For CO, NO2, and O3, bold text denotes the best-performing Alphasense algorithm (Equations 1-4) and empirical calibration model (Equations 7-10).
| Pollutant | Unit | Sensor | Eq. | N | RMSE (unit) | MAE (unit) | MBE (unit) | NMAE (%) | Pearson’s r (−) | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Med. | Min. | Max. | Med. | Min. | Max. | Med. | Min. | Max. | Med. | Min. | Max. | Med. | Min. | Max. | |||||
| CO2 | ppm | SCD30 | - | 9 | 408 | 349 | 504 | 398 | 335 | 495 | −398 | −495 | −334 | 46 | 38 | 57 | 0.99 | 0.98 | >0.99 |
| CO2 | ppm | SCD30 | 10 | 9 | 58 | 40 | 103 | 37 | 28 | 65 | 1 | 0 | 2 | 4 | 3 | 6 | 0.99 | 0.97 | >0.99 |
| CO | ppb | CO-B4 | 1 | 8 | 1159 | 512 | 1654 | 969 | 369 | 1425 | 795 | 66 | 1384 | 125 | 38 | 191 | 0.88 | 0.83 | 0.96 |
| CO | ppb | CO-B4 | 2 | 7 | 650 | 542 | 1072 | 507 | 419 | 927 | −140 | −909 | 245 | 52 | 39 | 87 | 0.94 | 0.92 | 0.97 |
| CO | ppb | CO-B4 | 3 | 8 | 710 | 505 | 784 | 543 | 365 | 600 | 268 | −234 | 387 | 64 | 29 | 76 | 0.92 | 0.90 | 0.97 |
| CO | ppb | CO-B4 | 4 | 8 | 645 | 458 | 779 | 482 | 339 | 606 | 230 | −192 | 446 | 57 | 28 | 79 | 0.94 | 0.92 | 0.97 |
| CO | ppb | CO-B4 | 7 | 8 | 657 | 405 | 725 | 496 | 314 | 563 | −33 | −42 | −23 | 52 | 30 | 57 | 0.92 | 0.90 | 0.97 |
| CO | ppb | CO-B4 | 8 | 8 | 664 | 405 | 730 | 499 | 318 | 568 | −34 | −43 | −26 | 53 | 31 | 57 | 0.92 | 0.90 | 0.97 |
| CO | ppb | CO-B4 | 9 | 8 | 572 | 381 | 624 | 445 | 263 | 494 | −34 | −38 | −27 | 47 | 24 | 52 | 0.94 | 0.93 | 0.97 |
| NO2 | ppb | NO2-B43F | 1 | 8 | 21 | 20 | 43 | 18 | 17 | 39 | −14 | −16 | 39 | 84 | 83 | 248 | 0.94 | 0.92 | 0.96 |
| NO2 | ppb | NO2-B43F | 2 | 8 | 20 | 18 | 30 | 17 | 15 | 24 | −12 | −16 | 24 | 83 | 70 | 150 | 0.95 | 0.93 | 0.97 |
| NO2 | ppb | NO2-B43F | 3 | 8 | 19 | 16 | 30 | 15 | 12 | 22 | −9 | −21 | −3 | 77 | 55 | 90 | 0.95 | 0.89 | 0.97 |
| NO 2 | ppb | NO2-B43F | 4 | 8 | 22 | 15 | 29 | 14 | 10 | 23 | −2 | −13 | 21 | 63 | 44 | 139 | 0.94 | 0.91 | 0.96 |
| NO2 | ppb | NO2-B43F | 7 | 8 | 18 | 15 | 20 | 11 | 9 | 13 | −1 | −1 | −1 | 41 | 35 | 50 | 0.92 | 0.89 | 0.95 |
| NO 2 | ppb | NO2-B43F | 8 | 8 | 14 | 11 | 16 | 8 | 7 | 19 | −1 | −1 | −1 | 26 | 23 | 33 | 0.96 | 0.93 | 0.97 |
| NO2 | ppb | NO2-B43F | 9 | 8 | 18 | 15 | 20 | 11 | 9 | 13 | −2 | −2 | −1 | 40 | 35 | 52 | 0.92 | 0.90 | 0.95 |
| O3 | ppb | OX-B431 | 1 | 8 | 21 | 18 | 22 | 19 | 14 | 20 | −15 | −20 | 0 | 318 | 98 | 3650 | −0.17 | −0.29 | −0.03 |
| O3 | ppb | OX-B431 | 2 | 8 | 21 | 14 | 23 | 16 | 9 | 20 | −7 | −20 | 11 | 1630 | 121 | 6826 | −0.15 | −0.31 | 0.07 |
| O 3 | ppb | OX-B431 | 3 | 8 | 21 | 16 | 26 | 15 | 12 | 20 | −5 | −20 | 13 | 667 | 96 | 4858 | 0.02 | −0.27 | 0.31 |
| O3 | ppb | OX-B431 | 4 | 8 | 22 | 15 | 23 | 20 | 13 | 21 | −18 | −20 | −4 | 999 | 100 | 5545 | −0.27 | −0.41 | 0.27 |
| O3 | ppb | OX-B431 | 7 | 8 | 8 | 8 | 9 | 6 | 6 | 7 | −1 | −2 | −1 | 1224 | 412 | 2031 | 0.56 | 0.48 | 0.62 |
| O3 | ppb | OX-B431 | 8 | 8 | 7 | 7 | 8 | 5 | 5 | 6 | −2 | −2 | −2 | 335 | 129 | 744 | 0.69 | 0.62 | 0.75 |
| O 3 | ppb | OX-B431 | 9 | 8 | 7 | 6 | 7 | 5 | 5 | 5 | −2 | −2 | −2 | 175 | 96 | 1284 | 0.75 | 0.73 | 0.78 |
| PM2.5 | μg m−3 | PMS5003 | − | 9 | 32 | 17 | 41 | 25 | 13 | 33 | 24 | 12 | 33 | 69 | 39 | 93 | 0.96 | 0.96 | 0.97 |
| PM2.5 | μg m−3 | PMS5003 | 11 | 9 | 6 | 5 | 8 | 5 | 4 | 7 | 1 | −2 | 4 | 20 | 19 | 23 | 0.96 | 0.96 | 0.97 |
| PM2.5 | μg m−3 | SPS30 | − | 5 | 5 | 4 | 8 | 4 | 3 | 6 | −1 | −6 | 0 | 17 | 14 | 23 | 0.97 | 0.96 | 0.97 |
| PM2.5 | μg m−3 | SPS30 | 11 | 5 | 5 | 4 | 10 | 4 | 3 | 8 | 1 | −3 | 5 | 16 | 15 | 26 | 0.97 | 0.96 | 0.97 |
We hypothesize that the SCD30 underestimated the reference CO2 concentration because the atmospheric pressure (~85 kPa), and thus the ambient air density (~1.0 kg m−3 at 25 °C), in Fort Collins (elevation 1525 m) is lower than at sea level. Lower air density translates into fewer molecules per unit volume. The SCD30 sensor includes a pressure compensation feature that should account for deviations between the ambient pressure and atmospheric pressure at sea level if the sensor is provided with ambient pressure data from an external source [60]. During this experiment, pressure data were not provided to the SCD30 and the pressure compensation feature remained deactivated. The accuracy of SCD30-reported CO2 concentrations could likely be improved by providing the sensor with ambient pressure data.
One-hour average CO concentrations calculated using all four Alphasense algorithms were strongly correlated with the reference CO concentration (Pearson’s r ≥ 0.83; Table 3; Figure 4). Algorithm 1 overestimated the reference CO concentration (MBE > 0 for all eight HHBs), likely because (AEu – AEe) was almost always negative and the correction factor nT was always positive. Equation 2 could not be applied to one CO-B4 sensor with AE0 = 0. The 1-h average CO concentrations calculated using Equation 4—which did not account for the auxiliary electrode output—had the smallest median RMSE and MAE values. Of the empirical calibration models, only Equation 9—which included the working electrode voltage and RH as predictors—produced smaller median RMSE and MAE values than Equation 4.
Figure 4.
Comparison of 1-h average concentrations calculated from low-cost sensor data in the Home Health Boxes and measured using reference monitors (CO: TSI QTrak 7575-X with model 982 probe; NO2: Thermo Environmental Instruments 42C; O3: Thermo Environmental Instruments 49C). Equations 1-4 were Alphasense algorithms. Equations 7-9 were empirical calibration models. Each HHB is represented by a different color. The dashed line is y = x.
One-hour average NO2 concentrations calculated using all four Alphasense algorithms were strongly correlated with the reference NO2 concentration (Pearson’s r ≥ 0.89). Of the Alphasense algorithms, Equation 4—which did not account for the auxiliary electrode output—produced the smallest median MAE, MBE, and NMAE values. When natural gas cooking burners were used, peak 15-minute and 1-h average NO2 concentrations calculated from HHB data using Algorithm 4 were higher than the corresponding reference NO2 concentrations (Figure 3 and Figure S5). These discrepancies might have been observed because the HHBs sampled from a location directly above the stove, whereas the reference NO2 monitor sampled from a similar height at a location 3.2 m north of the front burners; however, the 15-minute average CO concentrations calculated from HHB data using Algorithm 4 were not consistently higher than the corresponding reference CO concentrations, even though the QTrak CO monitors were installed close to the cooktop height and 1.6 m west of the left burners. One-hour NO2 concentrations predicted using the linear mixed model shown in Equation 8—which included the working electrode voltage and temperature as predictors—had the smallest median RMSE, MAE, MBE, and NMAE values.
One-hour average O3 concentrations calculated using the Alphasense algorithms were weakly—and often negatively—correlated with the reference O3 concentration. We might have had trouble obtaining reliable O3 measurements from OX-B431 data because the sensor responds to both NO2 and O3. In fact, calibration data from Alphasense indicated that each OX-B431 sensor was more sensitive to NO2 than to O3. Average NO2 and O3 concentrations in the kitchen were similar (median 1-h NO2 and O3 = 15 and 20 ppb, respectively) but the highest NO2 concentrations were much greater than the highest O3 concentrations (maximum 1-h NO2 and O3 = 219 and 46 ppb, respectively). As a result, much of the OX-B431 sensor response was likely due to NO2 rather than O3; however, the negative MBE values shown in Table 3 suggest that the Alphasense algorithms over-corrected for the NO2 response. It’s also possible that some O3 was lost to the surface of the gas sensor housing inlet and, as a result, the OX-B431 sensors were less sensitive to O3 when installed in the HHBs than when calibrated by Alphasense.
One-hour average O3 concentrations predicted using empirical calibration models were moderately- to strongly-correlated with reference O3 concentrations (Pearson’s r = 0.48 to 0.78). The linear mixed model shown in Equation 9—which included the working electrode voltage and RH as predictors—produced the strongest correlation as well as the smallest median NMAE.
Overall, we hypothesize that the following results related to gas sensor calibration might be generalizable to other studies:
Users of NDIR CO2 sensors should account for differences between local atmospheric pressure and atmospheric pressure at sea level.
Equation 2 is not an ideal calibration algorithm for Alphasense electrochemical sensors because it cannot be used if AE0 = 0 (i.e., if the sensor auxiliary electrode output is zero when measuring zero air, which Alphasense calibration data indicated was the case for one CO-B4 sensor tested in this study).
We successfully predicted gas concentrations without using electrochemical sensor auxiliary electrode output. This result might have been due to our use of the sensors in an indoor environment with relatively constant temperature (20 to 27 °C) and low RH (< 50%); however, Cross et al. reached a similar conclusion after using Alphasense CO-B4, NO2-B43F, and OX-B421 sensors in an outdoor study that spanned multiple seasons [28].
Even in an indoor setting with relatively constant temperature and low RH, electrochemical gas sensor performance might be influenced by environmental conditions, as suggested by the fact that our best-performing empirical calibration model for each sensor included either temperature or RH as a predictor.
We compared performance metrics for the low-cost gas sensors in the HHBs to results from several studies in SI Section S2.2 [27-29,31,34,61]. The RMSE for 1-h average NO2 concentrations calculated from Alphasense NO2-B43F sensor data using Equation 8 (median = 14 ppb) was higher than the RMSE reported by Chatzidiakou et al. [34] for 1-minute indoor NO2 concentrations obtained from Alphasense NO2-A43F sensors calibrated using a linear model (3 ppb). A key difference is that we measured a maximum 1-minute NO2 concentration (250 ppb) that was 6× the maximum 1-minute NO2 concentration (40 ppb) measured by Chatzidiakou et al. The NMAEs for 1-h average CO2 (median = 4%, range = 3% to 6%) and NO2 (median = 26%, range = 23% to 33%) concentrations calculated using Equations 10 and 8, respectively, were similar to relative errors reported by Zimmerman et al. [29] for 15-minute average outdoor CO2 (2%) and NO2 (29%) concentrations obtained from low-cost NDIR and NO2-B43F sensors, respectively, that had been calibrated using random forest models. We attributed the favorable performance of the linear mixed models we used to predict CO2 and NO2 concentrations, compared to the more sophisticated machine learning models developed by Zimmerman et al., to several factors. We suspect that the strength of each sensors’ response to the pollutant of interest was higher and the impact of temperature and RH on sensor outputs was mitigated during our experiment because (1) normal and scripted use of natural gas cooking burners resulted in high CO2 and NO2 concentrations and (2) we operated the sensors in an indoor environment where temperature and RH spanned relatively narrow ranges. We also averaged data over a 4× longer period (1 h) than Zimmerman et al. (15 minutes). Results from multiple prior studies indicated that hybrid calibration models for low-cost NDIR and electrochemical gas sensors—in which a machine learning model predicted concentrations below a certain threshold (e.g., 90% of the maximum in the training dataset) and a linear model predicted higher concentrations—predicted reference gas concentrations more accurately than linear calibration models [30,31].
3.2.2. PM Sensors
The 168-h average uncorrected PM2.5 concentrations reported by the nine PMS5003 sensors in the HHBs ranged from 26.0 to 38.5 μg m−3 (median = 33.2 μg m−3; mean = 33.4 μg m−3; RSD = 10.5%). The PM2.5 gravimetric correction factor for the PMS5003 ranged from 0.54 to 0.66 (median = 0.59; mean = 0.58). Correcting PMS5003-reported PM2.5 concentrations to the filter samples reduced the median MBE associated with 1-h average concentrations, relative to TEOM data recorded during the first 48-h of the experiment, from 24 to 1 μg m−3 (Table 3 and Figure S9).
Given that the RH in the kitchen remained <40% for >99% of the experimental duration, and all PM2.5 filter samples were pre- and post-weighed in an environment with RH between 30% and 40% [43], we do not suspect that the PMS5003 sensors in the HHBs overestimated in-kitchen PM2.5 concentrations due to hygroscopic particle growth.
The (a) minimum (0.54) and (b) maximum (0.66) gravimetric correction factors calculated across the nine PMS5003 sensors in the HHBs during the full 168-h experiment were approximately equal to (a) the 75th percentile correction factor reported for outdoor PMS5003 sensors that measured smoke from the 2018 Camp Fire in California, USA (0.53) [62] and (b) the gravimetric correction factor calculated for the PMS5003 sensor in the outdoor ASPEN Box during 33 h at the end of the experiment (0.67; due to a firmware error, the ASPEN PM2.5 filter sample did not run for the full 168 h). Additionally, the RMSE (3.9 μg m−3), MAE (3.1 μg m−3), MBE (1.5 μg m−3), and NMAE (10%) values associated with 24-h average PM2.5 concentrations obtained via gravimetric correction of the HHB PMS5003 data were only slightly lower than the RMSE (4.5 μg m−3), MAE (3.5 μg m−3), MBE (2.6 μg m−3), and NMAE (11%) values associated with 24-h average PM2.5 concentrations obtained by applying the U.S.-wide correction equation developed by Barkjohn et al. [51] for outdoor PurpleAir monitors to the HHB PMS5003 data. Overall, PM2.5 corrections for the PMS5003 sensors in the HHBs were similar to PM2.5 corrections for PMS5003 sensors installed outdoors during this and prior experiments; however, outdoor correction factors might not be applicable to all indoor PM2.5 measurements. We suspect that much of the indoor PM2.5 sampled during this experiment was wildfire smoke from outdoors. In the absence of a large, outdoor source of PM2.5 pollution, indoor PM2.5 pollution could be dominated by indoor sources (e.g., emissions from cooking).
The 168-h average uncorrected PM2.5 concentrations reported by the nine SPS30 sensors varied substantially (range = 3.9 to 19.2 μg m−3). Four SPS30 sensors reported 168-h average uncorrected PM2.5 concentrations between 3.9 and 11.4 μg m−3, while the other five reported 168-h averages between 16.4 and 19.2 μg m−3. During prior laboratory testing of the SPS30, we observed high precision between eight replicate sensors [18]; as a result, we suspected that the SPS30 sensors in four of the nine HHBs were not functioning properly—potentially due to the way the sensors were installed in or powered by the HHBs. PM2.5 gravimetric correction factors for the five properly functioning SPS30 sensors ranged from 0.95 to 1.27 (median = 1.09). Correcting PM2.5 concentrations reported by these five SPS30 sensors to the filter samples had little effect on the RMSE, MAE, MBE, or NMAE (Table 3 and Figure S9).
3.3. Classification of Pollutant Concentrations
Pollutant concentrations derived from HHB gas sensor data—using uncorrected CO2 concentrations, CO concentrations calculated using Equation 4, NO2 concentrations calculated using Equation 4, and O3 concentrations calculated using Equation 3—resulted in the same classifications as concentrations derived from reference instrument data (Table 2). CO2 concentrations were “elevated”, CO concentrations were “elevated”, and NO2 concentrations were “high”. The indoor O3 concentration was classified as “elevated” based on the reference data. The fact that the highest 8-hour O3 concentration from the median HHB also fell within the “elevated” range might have been coincidental, given that 1-h O3 concentrations calculated using Equation 3 were weakly correlated with 1-h reference concentrations (Table 3).
The results shown in Table 2 do not guarantee that classifications derived from HHB data and reference monitor data would agree for any given 168-h sample. In SI Section S2.3, we discuss how this agreement might be affected by the range of concentrations measured during a sample.
3.4. Concentration-Time Ratios
Both plots shown in Figure 5 feature concentration-time ratios obtained from reference data on the x-axis. The plot in the left panel of features concentration-time ratios obtained from HHB data consisting of uncorrected CO2 concentrations as well as CO, NO2, and O3 concentrations calculated using the best-performing Alphasense algorithms (Equations 4, 4, and 3, respectively) on the y-axis; the plot in the right panel features concentration-time ratios obtained from HHB data consisting of CO2, CO, NO2, and O3 concentrations calculated using the best-performing empirical calibration models (Equations 10, 9, 8, and 9, respectively) on the y-axis. Data along the diagonal dashed line indicated that equivalent ratios were obtained from reference monitors and HHBs. For points in the upper-right quadrant formed by the two dotted lines, reference and HHB data both indicated that the activity was associated with a disproportionately high fraction of the 168-h integrated concentration of the pollutant. For points in the upper-left quadrant, HHB data suggested that the activity was associated with a disproportionately high fraction of the 168-h integrated concentration of the pollutant, but reference data indicated that it was not (i.e., HHB data produced a “false-positive” result). For points in the bottom-right quadrant, reference data indicated that the activity was associated with a disproportionately high fraction of the 168-h integrated concentration of the pollutant, but HHB data did not (i.e., HHB data produced a “false-negative” result). For points in the bottom-left quadrant, reference and HHB data both indicated that the average concentration of the pollutant during the activity was lower than the 168-h average concentration of the pollutant.
Figure 5.
Left: Concentration-time ratios were obtained from HHB data consisting of uncorrected CO2 concentrations as well as CO, NO2, and O3 concentrations calculated using the best-performing Alphasense algorithms (Equations 4, 4, and 3, respectively). Right: Concentration-time ratios were obtained from HHB data consisting of CO2, CO, NO2, and O3 concentrations calculated using the best-performing empirical calibration models (Equations 10, 9, 8, and 9, respectively). In both panels, the dashed diagonal line is y = x and dotted lines represent concentration-time ratios = 1.
Whether or not empirical calibration models were applied to HHB data, reference monitors and HHBs both indicated that (a) scripted use of gas cooking burners contributed disproportionately to time-integrated CO2, CO, and NO2 concentrations and (b) normal cooking with gas burners contributed disproportionately to time-integrated CO and NO2 concentrations. When empirical calibration models were applied to HHB data, agreement between the concentration-time ratios calculated from reference and HHB data improved—especially for O3.
4. CONCLUSIONS
One week of data collected in a home kitchen in Fort Collins, CO, USA indicated that, relative to ASHRAE, ACGIH, US EPA, and WHO air quality standards, CO2 concentrations were elevated (maximum 8-h reference = 2198 ppm), CO concentrations were elevated (maximum 8-h and 1-h reference = 6 and 10 ppm, respectively), and NO2 concentrations were high (maximum 1-h reference = 219 ppb) due, in part, to scripted use of natural gas cooking burners; however, when only data affected by normal occupant activities—which included cooking with natural gas burners—were considered, CO2 concentrations remained elevated (maximum 8-h reference = 1241 ppm), CO concentrations remained elevated (maximum 8-h and 1-h reference = 2 and 9 ppm, respectively), and NO2 concentrations remained high (maximum 1-h reference = 139 ppb). Ozone, PM2.5, and PM10 concentrations were elevated (maximum 8-h reference O3 = 31 ppb; median 168-h PM2.5 = 19.5 μg m−3, RSD = 6.1%, n = 9 HHB filter samples; median 168-h PM10 = 40.0 μg m−3, RSD = 7.6%, n = 9 HHB filter samples) due primarily to infiltration of outdoor air pollution—which included wildfire smoke—when windows and doors were opened.
Sensor manufacturers’ calibrations of the low-cost CO2, CO, and NO2 sensors inside the Home Health Boxes were accurate enough to produce the same qualitative conclusions as reference monitors regarding (a) whether concentrations of these pollutants were low, somewhat elevated, or concerningly high relative to air quality guidelines and (b) which in-home activities were associated with disproportionately high fractions of the 168-h integrated concentrations of these pollutants. Our results do suggest, however, that empirical calibration (e.g., through collocation with reference monitors) would be necessary to obtain accurate quantitative data from low-cost gas sensors. Empirical linear calibration models with sensor-specific coefficients (a) reduced the error associated with CO2, NO2, and O3 concentrations obtained from the HHB data and (b) improved agreement between the HHB and reference data regarding the extent to which different activities contributed to the time-integrated concentration of each pollutant.
This study had several limitations. First, the CO “reference” monitors (TSI QTrak) relied on the same measurement technique (electrochemical) as the low-cost CO sensors; the low-cost CO sensors in the HHBs might not agree as well with a regulatory-grade NDIR CO analyzer. Second, the monitors did not all sample from the same exact location within the kitchen. Data were averaged over 15-minute intervals prior to plotting results and over 1-h intervals prior to calculating sensor performance metrics, and these averaging times likely attenuated differences in instrument response timing resulting from differences in location; however, we would have preferred to install the instruments closer together. Third, we only collected data over a single week in a single American home; therefore, we did not investigate how well the calibration models tested here transfer to other homes or retain their predictive abilities over time (e.g., months following initial calibration). Poorer sensor and calibration model performance might be observed in homes with no or less consistent climate control.
Finally, our study evaluated the accuracy and precision of pollutant concentrations measured using HHBs but did not evaluate whether knowledge gained from HHB data helped occupants or homeowners take actions to reduce occupants’ exposure to air pollution. As our data demonstrate, indoor air pollutant concentrations are influenced by occupant activities, but some pollutant sources are outside of occupant control. For example, occupants cannot control outdoor air pollution from wildfires, traffic, and industrial sources—all of which can infiltrate the home. In the case of indoor pollution sources and sinks, such as natural gas cooking burners and exhaust hoods, renters might have no control over the presence or absence of these items and be limited in their ability to move. Homeowners might not be able to afford to replace a gas stove or install a new exhaust hood. Despite these limitations, HHB data might help air quality professionals and home occupants identify the largest sources of indoor air pollution in a home. This knowledge, in turn, could inform cost-effective behavioral changes, ventilation strategies, and filtration strategies to mitigate occupants’ exposure to indoor air pollution.
Supplementary Material
HIGHLIGHTS.
We assembled a platform to sample/sense PM2.5, PM10, CO2, CO, NO2, and O3 indoors
Nine units were collocated with reference monitors in a home kitchen for one week
We tested manufacturer and empirical linear calibrations for low-cost gas sensors
Gas sensors correctly identified whether pollutants exceeded air quality guidelines
Cooking with natural gas burners led to 1-h indoor NO2 concentrations above 100 ppb
ACKNOWLEDGEMENTS
This work was funded by the National Institute on Minority Health and Health Disparities (NIMHD) of the National Institutes of Health (NIH) under award number R43MD014915. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH. The NIH was not involved in study design; in the collection, analysis, or interpretation of data; in writing of this article; or in the decision to submit the article for publication. The authors thank Delphine Farmer and Tyson Berg of the Colorado State University Department of Chemistry for help with calibrating the NOx and O3 reference monitors.
Footnotes
CONFLICT OF INTEREST
John Volckens is a scientific founder of Access Sensor Technologies, LLC and has an equity interest in the company. The terms of this arrangement have been reviewed and approved by Colorado State University in accordance with its conflict-of-interest policies.
DATA STATEMENT
Data related to this article are available through the open access repository service Mountain Scholar: https://hdl.handle.net/….
Declaration of interests
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
REFERENCES
- [1].Murray CJL, Aravkin AY, Zheng P, Abbafati C, Abbas KM, Abbasi-Kangevari M, Abd-Allah F, Abdelalim A, Abdollahi M, Abdollahpour I, Abegaz KH, Abolhassani H, Aboyans V, Abreu LG, Abrigo MRM, Abualhasan A, Abu-Raddad LJ, Abushouk AI, Adabi M, Adekanmbi V, Adeoye AM, Adetokunboh OO, Adham D, Advani SM, Agarwal G, Aghamir SMK, Agrawal A, Ahmad T, Ahmadi K, Ahmadi M, Ahmadieh H, Ahmed MB, Akalu TY, Akinyemi RO, Akinyemiju T, Akombi B, Akunna CJ, Alahdab F, Al-Aly Z, Alam K, Alam S, Alam T, Alanezi FM, Alanzi TM, wassihun Alemu B, Alhabib KF, Ali M, Ali S, Alicandro G, Alinia C, Alipour V, Alizade H, Aljunid SM, Alla F, Allebeck P, Almasi-Hashiani A, Al-Mekhlafi HM, Alonso J, Altirkawi KA, Amini-Rarani M, Amiri F, Amugsi DA, Ancuceanu R, Anderlini D, Anderson JA, Andrei CL, Andrei T, Angus C, Anjomshoa M, Ansari F, Ansari-Moghaddam A, Antonazzo IC, Antonio CAT, Antony CM, Antriyandarti E, Anvari D, Anwer R, Appiah SCY, Arabloo J, Arab-Zozani M, Ariani F, Armoon B, Ärnlöv J, Arzani A, Asadi-Aliabadi M, Asadi-Pooya AA, Ashbaugh C, Assmus M, Atafar Z, Atnafu DD, Atout MMW, Ausloos F, Ausloos M, Ayala Quintanilla BP, Ayano G, Ayanore MA, Azari S, Azarian G, Azene ZN, Badawi A, Badiye AD, Bahrami MA, Bakhshaei MH, Bakhtiari A, Bakkannavar SM, Baldasseroni A, Ball K, Ballew SH, Balzi D, Banach M, Banerjee SK, Bante AB, Baraki AG, Barker-Collo SL, Bärnighausen TW, Barrero LH, Barthelemy CM, Barua L, Basu S, Baune BT, Bayati M, Becker JS, Bedi N, Beghi E, Béjot Y, Bell ML, Bennitt FB, Bensenor IM, Berhe K, Berman AE, Bhagavathula AS, Bhageerathy R, Bhala N, Bhandari D, Bhattacharyya K, Bhutta ZA, Bijani A, Bikbov B, Bin Sayeed MS, Biondi A, Birihane BM, Bisignano C, Biswas RK, Bitew H, Bohlouli S, Bohluli M, Boon-Dooley AS, Borges G, Borzì AM, Borzouei S, Bosetti C, Boufous S, Braithwaite D, Breitborde NJK, Breitner S, Brenner H, Briant PS, Briko AN, Briko NI, Britton GB, Bryazka D, Bumgarner BR, Burkart K, Burnett RT, Burugina Nagaraja S, Butt ZA, Caetano dos Santos FL, Cahill LE, Cámera LLA, Campos-Nonato IR, Cárdenas R, Carreras G, Carrero JJ, Carvalho F, Castaldelli-Maia JM, Castañeda-Orjuela CA, Castelpietra G, Castro F, Causey K, Cederroth CR, Cercy KM, Cerin E, Chandan JS, Chang K-L, Charlson FJ, Chattu VK, Chaturvedi S, Cherbuin N, Chimed-Ochir O, Cho DY, Choi J-YJ, Christensen H, Chu D-T, Chung MT, Chung S-C, Cicuttini FM, Ciobanu LG, Cirillo M, Classen TKD, Cohen AJ, Compton K, Cooper OR, Costa VM, Cousin E, Cowden RG, Cross DH, Cruz JA, Dahlawi SMA, Damasceno AAM, Damiani G, Dandona L, Dandona R, Dangel WJ, Danielsson A-K, Dargan PI, Darwesh AM, Daryani A, Das JK, Das Gupta R, das Neves J, Dávila-Cervantes CA, Davitoiu DV, De Leo D, Degenhardt L, DeLang M, Dellavalle RP, Demeke FM, Demoz GT, Demsie DG, Denova-Gutiérrez E, Dervenis N, Dhungana GP, Dianatinasab M, Dias da Silva D, Diaz D, Dibaji Forooshani ZS, Djalalinia S, Do HT, Dokova K, Dorostkar F, Doshmangir L, Driscoll TR, Duncan BB, Duraes AR, Eagan AW, Edvardsson D, El Nahas N, El Sayed I, El Tantawi M, Elbarazi I, Elgendy IY, El-Jaafary SI, Elyazar IR, Emmons-Bell S, Erskine HE, Eskandarieh S, Esmaeilnejad S, Esteghamati A, Estep K, Etemadi A, Etisso AE, Fanzo J, Farahmand M, Fareed M, Faridnia R, Farioli A, Faro A, Faruque M, Farzadfar F, Fattahi N, Fazlzadeh M, Feigin VL, Feldman R, Fereshtehnejad S-M, Fernandes E, Ferrara G, Ferrari AJ, Ferreira ML, Filip I, Fischer F, Fisher JL, Flor LS, Foigt NA, Folayan MO, Fomenkov AA, Force LM, Foroutan M, Franklin RC, Freitas M, Fu W, Fukumoto T, Furtado JM, Gad MM, Gakidou E, Gallus S, Garcia-Basteiro AL, Gardner WM, Geberemariyam BS, Gebreslassie AAAA, Geremew A, Gershberg Hayoon A, Gething PW, Ghadimi M, Ghadiri K, Ghaffarifar F, Ghafourifard M, Ghamari F, Ghashghaee A, Ghiasvand H, Ghith N, Gholamian A, Ghosh R, Gill PS, Ginindza TGG, Giussani G, Gnedovskaya EV, Goharinezhad S, Gopalani SV, Gorini G, Goudarzi H, Goulart AC, Greaves F, Grivna M, Grosso G, Gubari MIM, Gugnani HC, Guimarães RA, Guled RA, Guo G, Guo Y, Gupta R, Gupta T, Haddock B, Hafezi-Nejad N, Hafiz A, Haj-Mirzaian A, Haj-Mirzaian A, Hall BJ, Halvaei I, Hamadeh RR, Hamidi S, Hammer MS, Hankey GJ, Haririan H, Haro JM, Hasaballah AI, Hasan MM, Hasanpoor E, Hashi A, Hassanipour S, Hassankhani H, Havmoeller RJ, Hay SI, Hayat K, Heidari G, Heidari-Soureshjani R, Henrikson HJ, Herbert ME, Herteliu C, Heydarpour F, Hird TR, Hoek HW, Holla R, Hoogar P, Hosgood HD, Hossain N, Hosseini M, Hosseinzadeh M, Hostiuc M, Hostiuc S, Househ M, Hsairi M, Hsieh VC, Hu G, Hu K, Huda TM, Humayun A, Huynh CK, Hwang B-F, Iannucci VC, Ibitoye SE, Ikeda N, Ikuta KS, Ilesanmi OS, Ilic IM, Ilic MD, Inbaraj LR, Ippolito H, Iqbal U, Irvani SSN, Irvine CMS, Islam MM, Islam SMS, Iso H, Ivers RQ, Iwu CCD, Iwu CJ, Iyamu IO, Jaafari J, Jacobsen KH, Jafari H, Jafarinia M, Jahani MA, Jakovljevic M, Jalilian F, James SL, Janjani H, Javaheri T, Javidnia J, Jeemon P, Jenabi E, Jha RP, Jha V, Ji JS, Johansson L, John O, John-Akinola YO, Johnson CO, Jonas JB, Joukar F, Jozwiak JJ, Jürisson M, Kabir A, Kabir Z, Kalani H, Kalani R, Kalankesh LR, Kalhor R, Kanchan T, Kapoor N, Karami Matin B, Karch A, Karim MA, Kassa GM, Katikireddi SV, Kayode GA, Kazemi Karyani A, Keiyoro PN, Keller C, Kemmer L, Kendrick PJ, Khalid N, Khammarnia M, Khan EA, Khan M, Khatab K, Khater MM, Khatib MN, Khayamzadeh M, Khazaei S, Kieling C, Kim YJ, Kimokoti RW, Kisa A, Kisa S, Kivimäki M, Knibbs LD, Knudsen AKS, Kocarnik JM, Kochhar S, Kopec JA, Korshunov VA, Koul PA, Koyanagi A, Kraemer MUG, Krishan K, Krohn KJ, Kromhout H, Kuate Defo B, Kumar GA, Kumar V, Kurmi OP, Kusuma D, La Vecchia C, Lacey B, Lal DK, Lalloo R, Lallukka T, Lami FH, Landires I, Lang JJ, Langan SM, Larsson AO, Lasrado S, Lauriola P, Lazarus JV, Lee PH, Lee SWH, LeGrand KE, Leigh J, Leonardi M, Lescinsky H, Leung J, Levi M, Li S, Lim L-L, Linn S, Liu S, Liu S, Liu Y, Lo J, Lopez AD, Lopez JCF, Lopukhov PD, Lorkowski S, Lotufo PA, Lu A, Lugo A, Maddison ER, Mahasha PW, Mahdavi MM, Mahmoudi M, Majeed A, Maleki A, Maleki S, Malekzadeh R, Malta DC, Mamun AA, Manda AL, Manguerra H, Mansour-Ghanaei F, Mansouri B, Mansournia MA, Mantilla Herrera AM, Maravilla JC, Marks A, Martin RV, Martini S, Martins-Melo FR, Masaka A, Masoumi SZ, Mathur MR, Matsushita K, Maulik PK, McAlinden C, McGrath JJ, McKee M, Mehndiratta MM, Mehri F, Mehta KM, Memish ZA, Mendoza W, Menezes RG, Mengesha EW, Mereke A, Mereta ST, Meretoja A, Meretoja TJ, Mestrovic T, Miazgowski B, Miazgowski T, Michalek IM, Miller TR, Mills EJ, Mini G, Miri M, Mirica A, Mirrakhimov EM, Mirzaei H, Mirzaei M, Mirzaei R, Mirzaei-Alavijeh M, Misganaw AT, Mithra P, Moazen B, Mohammad DK, Mohammad Y, Mohammad Gholi Mezerji N, Mohammadian-Hafshejani A, Mohammadifard N, Mohammadpourhodki R, Mohammed AS, Mohammed H, Mohammed JA, Mohammed S, Mokdad AH, Molokhia M, Monasta L, Mooney MD, Moradi G, Moradi M, Moradi-Lakeh M, Moradzadeh R, Moraga P, Morawska L, Morgadoda-Costa J, Morrison SD, Mosapour A, Mosser JF, Mouodi S, Mousavi SM, Mousavi Khaneghah A, Mueller UO, Mukhopadhyay S, Mullany EC, Musa KI, Muthupandian S, Nabhan AF, Naderi M, Nagarajan AJ, Nagel G, Naghavi M, Naghshtabrizi B, Naimzada MD, Najafi F, Nangia V, Nansseu JR, Naserbakht M, Nayak VC, Negoi I, Ngunjiri JW, Nguyen CT, Nguyen HLT, Nguyen M, Nigatu YT, Nikbakhsh R, Nixon MR, Nnaji CA, Nomura S, Norrving B, Noubiap JJ, Nowak C, Nunez-Samudio V, Oţoiu A, Oancea B, Odell CM, Ogbo FA, Oh I-H, Okunga EW, Oladnabi M, Olagunju AT, Olusanya BO, Olusanya JO, Omer MO, Ong KL, Onwujekwe OE, Orpana HM, Ortiz A, Osarenotor O, Osei FB, Ostroff SM, Otstavnov N, Otstavnov SS, Øverland S, Owolabi MO, P A M, Padubidri JR, Palladino R, Panda-Jonas S, Pandey A, Parry CDH, Pasovic M, Pasupula DK, Patel SK, Pathak M, Patten SB, Patton GC, Toroudi H. Pazoki, Peden AE, Pennini A, Pepito VCF, Peprah EK, Pereira DM, Pesudovs K, Pham HQ, Phillips MR, Piccinelli C, Pilz TM, Piradov MA, Pirsaheb M, Plass D, Polinder S, Polkinghorne KR, Pond CD, Postma MJ, Pourjafar H, Pourmalek F, Poznańska A, Prada SI, Prakash V, Pribadi DRA, Pupillo E, Quazi Syed Z, Rabiee M, Rabiee N, Radfar A, Rafiee A, Raggi A, Rahman MA, Rajabpour-Sanati A, Rajati F, Rakovac I, Ram P, Ramezanzadeh K, Ranabhat CL, Rao PC, Rao SJ, Rashedi V, Rathi P, Rawaf DL, Rawaf S, Rawal L, Rawassizadeh R, Rawat R, Razo C, Redford SB, Reiner RC, Reitsma MB, Remuzzi G, Renjith V, Renzaho AMN, Resnikoff S, Rezaei N, Rezaei N, Rezapour A, Rhinehart P-A, Riahi SM, Ribeiro DC, Ribeiro D, Rickard J, Rivera JA, Roberts NLS, Rodríguez-Ramírez S, Roever L, Ronfani L, Room R, Roshandel G, Roth GA, Rothenbacher D, Rubagotti E, Rwegerera GM, Sabour S, Sachdev PS, Saddik B, Sadeghi E, Sadeghi M, Saeedi R, Moghaddam S. Saeedi, Safari Y, Safi S, Safiri S, Sagar R, Sahebkar A, Sajadi SM, Salam N, Salamati P, Salem H, Salem MRR, Salimzadeh H, Salman OM, Salomon JA, Samad Z, Samadi Kafil H, Sambala EZ, Samy AM, Sanabria J, Sánchez-Pimienta TG, Santomauro DF, Santos IS, Santos JV, Santric-Milicevic MM, Saraswathy SYI, Sarmiento-Suárez R, Sarrafzadegan N, Sartorius B, Sarveazad A, Sathian B, Sathish T, Sattin D, Saxena S, Schaeffer LE, Schiavolin S, Schlaich MP, Schmidt MI, Schutte AE, Schwebel DC, Schwendicke F, Senbeta AM, Senthilkumaran S, Sepanlou SG, Serdar B, Serre ML, Shadid J, Shafaat O, Shahabi S, Shaheen AA, Shaikh MA, Shalash AS, Shams-Beyranvand M, Shamsizadeh M, Sharafi K, Sheikh A, Sheikhtaheri A, Shibuya K, Shield KD, Shigematsu M, Shin JI, Shin M-J, Shiri R, Shirkoohi R, Shuval K, Siabani S, Sierpinski R, Sigfusdottir ID, Sigurvinsdottir R, Silva JP, Simpson KE, Singh JA, Singh P, Skiadaresi E, Skou STS, Skryabin VY, Smith EUR, Soheili A, Soltani S, Soofi M, Sorensen RJD, Soriano JB, Sorrie MB, Soshnikov S, Soyiri IN, Spencer CN, Spotin A, Sreeramareddy CT, Srinivasan V, Stanaway JD, Stein C, Stein DJ, Steiner C, Stockfelt L, Stokes MA, Straif K, Stubbs JL, Sufiyan MB, Suleria HAR, Suliankatchi Abdulkader R, Sulo G, Sultan I, Szumowski Ł, Tabarés-Seisdedos R, Tabb KM, Tabuchi T, Taherkhani A, Tajdini M, Takahashi K, Takala JS, Tamiru AT, Taveira N, Tehrani-Banihashemi A, Temsah M-H, Tesema GA, Tessema ZT, Thurston GD, Titova MV, Tohidinik HR, Tonelli M, Topor-Madry R, Topouzis F, Torre AE, Touvier M, Tovani-Palone MRR, Tran BX, Travillian R, Tsatsakis A, Tudor Car L, Tyrovolas S, Uddin R, Umeokonkwo CD, Unnikrishnan B, Upadhyay E, Vacante M, Valdez PR, van Donkelaar A, Vasankari TJ, Vasseghian Y, Veisani Y, Venketasubramanian N, Violante FS, Vlassov V, Vollset SE, Vos T, Vukovic R, Waheed Y, Wallin MT, Wang Y, Wang Y-P, Watson A, Wei J, Wei MYW, Weintraub RG, Weiss J, Werdecker A, West JJ, Westerman R, Whisnant JL, Whiteford HA, Wiens KE, Wolfe CDA, Wozniak SS, Wu A-M, Wu J, Wulf Hanson S, Xu G, Xu R, Yadgir S, Yahyazadeh Jabbari SH, Yamagishi K, Yaminfirooz M, Yano Y, Yaya S, Yazdi-Feyzabadi V, Yeheyis TY, Yilgwan CS, Yilma MT, Yip P, Yonemoto N, Younis MZ, Younker TP, Yousefi B, Yousefi Z, Yousefinezhadi T, Yousuf AY, Yu C, Yusefzadeh H, Zahirian Moghadam T, Zamani M, Zamanian M, Zandian H, Zastrozhin MS, Zhang Y, Zhang Z-J, Zhao JT, Zhao XJG, Zhao Y, Zhou M, Ziapour A, Zimsen SRM, Brauer M, Afshin A, Lim SS, Global burden of 87 risk factors in 204 countries and territories, 1990–2019: a systematic analysis for the Global Burden of Disease Study 2019, The Lancet. 396 (2020) 1223–1249. 10.1016/S0140-6736(20)30752-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].US EPA, NAAQS Table, (2016). https://www.epa.gov/criteria-air-pollutants/naaqs-table (accessed June 15, 2020).
- [3].Quinn C, Miller-Lionberg DD, Klunder KJ, Kwon J, Noth EM, Mehaffy J, Leith D, Magzamen S, Hammond SK, Henry CS, Volckens J, Personal Exposure to PM 2.5 Black Carbon and Aerosol Oxidative Potential using an Automated Microenvironmental Aerosol Sampler (AMAS), Environ. Sci. Technol 52 (2018) 11267–11275. 10.1021/acs.est.8b02992. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [4].Koehler K, Good N, Wilson A, Mölter A, Moore BF, Carpenter T, Peel JL, Volckens J, The Fort Collins Commuter Study: Variability in Personal Exposure to Air Pollutants by Microenvironment, Indoor Air. 29 (2018) 231–241. 10.1111/ina.12533. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Levy Zamora M, Pulczinski JC, Johnson N, Garcia-Hernandez R, Rule A, Carrillo G, Zietsman J, Sandragorsian B, Vallamsundar S, Askariyeh MH, Koehler K, Maternal exposure to PM2.5 in south Texas, a pilot study, Science of The Total Environment. 628–629 (2018) 1497–1507. 10.1016/j.scitotenv.2018.02.138. [DOI] [PubMed] [Google Scholar]
- [6].Tryner J, Good N, Wilson A, Clark ML, Peel JL, Volckens J, Variation in gravimetric correction factors for nephelometer-derived estimates of personal exposure to PM2.5, Environmental Pollution. 250 (2019) 251–261. 10.1016/j.envpol.2019.03.121. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Do K, Yu H, Velasquez J, Grell-Brisk M, Smith H, Ivey CE, A data-driven approach for characterizing community scale air pollution exposure disparities in inland Southern California, Journal of Aerosol Science. (2020) 105704. 10.1016/j.jaerosci.2020.105704. [DOI] [Google Scholar]
- [8].Shrestha PM, Humphrey JL, Carlton EJ, Adgate JL, Barton KE, Root ED, Miller SL, Impact of Outdoor Air Pollution on Indoor Air Quality in Low-Income Homes during Wildfire Seasons, IJERPH. 16 (2019) 3535. 10.3390/ijerph16193535. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Mullen NA, Li J, Russell ML, Spears M, Less BD, Singer BC, Results of the California Healthy Homes Indoor Air Quality Study of 2011-2013: impact of natural gas appliances on air pollutant concentrations, Indoor Air. 26 (2016) 231–245. 10.1111/ina.12190. [DOI] [PubMed] [Google Scholar]
- [10].Singer BC, Pass RZ, Delp WW, Lorenzetti DM, Maddalena RL, Pollutant concentrations and emission rates from natural gas cooking burners without and with range hood exhaust in nine California homes, Building and Environment. 122 (2017) 215–229. 10.1016/j.buildenv.2017.06.021. [DOI] [Google Scholar]
- [11].Logue JM, Klepeis NE, Lobscheid AB, Singer BC, Pollutant Exposures from Natural Gas Cooking Burners: A Simulation-Based Assessment for Southern California, Environmental Health Perspectives. 122 (2014) 43–50. 10.1289/ehp.1306673. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Militello-Hourigan RE, Miller SL, The impacts of cooking and an assessment of indoor air quality in Colorado passive and tightly constructed homes, Building and Environment. 144 (2018) 573–582. 10.1016/j.buildenv.2018.08.044. [DOI] [Google Scholar]
- [13].Tryner J, Quinn C, Windom BC, Volckens J, Design and evaluation of a portable PM 2.5 monitor featuring a low-cost sensor in line with an active filter sampler, Environ. Sci.: Processes Impacts. 21 (2019) 1403–1415. 10.1039/C9EM00234K. [DOI] [PubMed] [Google Scholar]
- [14].Singer BC, Delp WW, Response of consumer and research grade indoor air quality monitors to residential sources of fine particles, Indoor Air. 28 (2018) 624–639. 10.1111/ina.12463. [DOI] [PubMed] [Google Scholar]
- [15].Levy Zamora M, Xiong F, Gentner D, Kerkez B, Kohrman-Glaser J, Koehler K, Field and Laboratory Evaluations of the Low-Cost Plantower Particulate Matter Sensor, Environmental Science & Technology. 53 (2019) 838–849. 10.1021/acs.est.8b05174. [DOI] [PubMed] [Google Scholar]
- [16].Sayahi T, Butterfield A, Kelly KE, Long-term field evaluation of the Plantower PMS low-cost particulate matter sensors, Environmental Pollution. 245 (2019) 932–940. 10.1016/j.envpol.2018.11.065. [DOI] [PubMed] [Google Scholar]
- [17].Tryner J, L’Orange C, Mehaffy J, Miller-Lionberg D, Hofstetter JC, Wilson A, Volckens J, Laboratory evaluation of low-cost PurpleAir PM monitors and in-field correction using co-located portable filter samplers, Atmospheric Environment. 220 (2020) 117067. 10.1016/j.atmosenv.2019.117067. [DOI] [Google Scholar]
- [18].Tryner J, Mehaffy J, Miller-Lionberg D, Volckens J, Effects of aerosol type and simulated aging on performance of low-cost PM sensors, Journal of Aerosol Science. 150 (2020) 105654. 10.1016/j.jaerosci.2020.105654. [DOI] [Google Scholar]
- [19].Zou Y, Clark JD, May AA, Laboratory evaluation of the effects of particle size and composition on the performance of integrated devices containing Plantower particle sensors, Aerosol Science and Technology. (2021) 1–13. 10.1080/02786826.2021.1905148.34732970 [DOI] [Google Scholar]
- [20].Ouimette JR, Malm WC, Schichtel BA, Sheridan PJ, Andrews E, Ogren JA, Arnott WP, Evaluating the PurpleAir monitor as an aerosol light scattering instrument, Aerosols/In Situ Measurement/Instruments and Platforms, 2021. 10.5194/amt-2021-170. [DOI] [Google Scholar]
- [21].Hinds WC, Aerosol Technology: Properties, Behavior, and Measurement of Airborne Particles, 2nd ed, Wiley, New York, 1999. [Google Scholar]
- [22].Jayaratne R, Liu X, Thai P, Dunbabin M, Morawska L, The influence of humidity on the performance of a low-cost air particle mass sensor and the effect of atmospheric fog, Atmos. Meas. Tech 11 (2018) 4883–4890. 10.5194/amt-11-4883-2018. [DOI] [Google Scholar]
- [23].Kuula J, Mäkelä T, Aurela M, Teinilä K, Varjonen S, González Ó, Timonen H, Laboratory evaluation of particle-size selectivity of optical low-cost particulate matter sensors, Atmos. Meas. Tech 13 (2020) 2413–2423. 10.5194/amt-13-2413-2020. [DOI] [Google Scholar]
- [24].Hagan DH, Kroll JH, Assessing the accuracy of low-cost optical particle sensors using a physics-based approach, Atmos. Meas. Tech 13 (2020) 6343–6355. 10.5194/amt-13-6343-2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [25].Mead MI, Popoola OAM, Stewart GB, Landshoff P, Calleja M, Hayes M, Baldovi JJ, McLeod MW, Hodgson TF, Dicks J, Lewis A, Cohen J, Baron R, Saffell JR, Jones RL, The use of electrochemical sensors for monitoring urban air quality in low-cost, high-density networks, Atmospheric Environment. 70 (2013) 186–203. 10.1016/j.atmosenv.2012.11.060. [DOI] [Google Scholar]
- [26].Masson N, Piedrahita R, Hannigan M, Quantification Method for Electrolytic Sensors in Long-Term Monitoring of Ambient Air Quality, Sensors. 15 (2015) 27283–27302. 10.3390/s151027283. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Castell N, Dauge FR, Schneider P, Vogt M, Lerner U, Fishbain B, Broday D, Bartonova A, Can commercial low-cost sensor platforms contribute to air quality monitoring and exposure estimates?, Environment International. 99 (2017) 293–302. 10.1016/j.envint.2016.12.007. [DOI] [PubMed] [Google Scholar]
- [28].Cross ES, Williams LR, Lewis DK, Magoon GR, Onasch TB, Kaminsky ML, Worsnop DR, Jayne JT, Use of electrochemical sensors for measurement of air pollution: correcting interference response and validating measurements, Atmos. Meas. Tech 10 (2017) 3575–3588. 10.5194/amt-10-3575-2017. [DOI] [Google Scholar]
- [29].Zimmerman N, Presto AA, Kumar SPN, Gu J, Hauryliuk A, Robinson ES, Robinson AL, Subramanian R, A machine learning calibration model using random forests to improve sensor performance for lower-cost air quality monitoring, Atmos. Meas. Tech 11 (2018) 291–313. 10.5194/amt-11-291-2018. [DOI] [Google Scholar]
- [30].Hagan DH, Isaacman-VanWertz G, Franklin JP, Wallace LMM, Kocar BD, Heald CL, Kroll JH, Calibration and assessment of electrochemical air quality sensors by co-location with regulatory-grade instruments, Atmos. Meas. Tech 11 (2018) 315–328. 10.5194/amt-11-315-2018. [DOI] [Google Scholar]
- [31].Malings C, Tanzer R, Hauryliuk A, Kumar SPN, Zimmerman N, Kara LB, Presto AA, Subramanian R, Development of a general calibration model and long-term performance evaluation of low-cost sensors for air pollutant gas monitoring, Atmos. Meas. Tech 12 (2019) 903–920. 10.5194/amt-12-903-2019. [DOI] [Google Scholar]
- [32].Li J, Hauryliuk A, Malings C, Eilenberg SR, Subramanian R, Presto AA, Characterizing the Aging of Alphasense NO 2 Sensors in Long-Term Field Deployments, ACS Sens. 6 (2021) 2952–2959. 10.1021/acssensors.1c00729. [DOI] [PubMed] [Google Scholar]
- [33].Gillooly SE, Zhou Y, Vallarino J, Chu MT, Michanowicz DR, Levy JI, Adamkiewicz G, Development of an in-home, real-time air pollutant sensor platform and implications for community use, Environmental Pollution. 244 (2019) 440–450. 10.1016/j.envpol.2018.10.064. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [34].Chatzidiakou L, Krause A, Popoola OAM, Di Antonio A, Kellaway M, Han Y, Squires FA, Wang T, Zhang H, Wang Q, Fan Y, Chen S, Hu M, Quint JK, Barratt B, Kelly FJ, Zhu T, Jones RL, Characterising low-cost sensors in highly portable platforms to quantify personal exposure in diverse environments, Atmos. Meas. Tech 12 (2019) 4643–4657. 10.5194/amt-12-4643-2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [35].Volckens J, Quinn C, Leith D, Mehaffy J, Henry CS, Miller-Lionberg D, Development and evaluation of an ultrasonic personal aerosol sampler, Indoor Air. 27 (2017) 409–416. 10.1111/ina.12318. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [36].Leith D, L’Orange C, Mehaffy J, Volckens J, Design and performance of UPAS inlets for respirable and thoracic mass sampling, Journal of Occupational and Environmental Hygiene. 17 (2020) 274–282. 10.1080/15459624.2020.1741595. [DOI] [PubMed] [Google Scholar]
- [37].Fort Collins Weather Station, Colorado State University. (2021). https://ccc.atmos.colostate.edu/~autowx/ (accessed January 9, 2021).
- [38].U.S. Forest Service, Cameron Peak Fire Information, InciWeb. (2020). https://inciweb.nwcg.gov/incident/6964/ (accessed September 28, 2020). [Google Scholar]
- [39].U.S. Forest Service, Mullen Fire Information, InciWeb. (2020). https://inciweb.nwcg.gov/incident/7208 (accessed September 28, 2020). [Google Scholar]
- [40].Lytle K, Cameron Peak Fire becomes largest wildfire in Colorado history, Fort Collins Coloradoan. (2020). https://www.coloradoan.com/story/news/2020/10/14/largest-wildfire-colorado-history-cameron-peak-fire/3659545001/ (accessed October 28, 2020). [Google Scholar]
- [41].Farmer DK, Vance ME, Abbatt JPD, Abeleira A, Alves MR, Arata C, Boedicker E, Bourne S, Cardoso-Saldaña F, Corsi R, DeCarlo PF, Goldstein AH, Grassian VH, Hildebrandt Ruiz L, Jimenez JL, Kahan TF, Katz EF, Mattila JM, Nazaroff WW, Novoselac A, O’Brien RE, Or VW, Patel S, Sankhyan S, Stevens PS, Tian Y, Wade M, Wang C, Zhou S, Zhou Y, Overview of HOMEChem: House Observations of Microbial and Environmental Chemistry, Environ. Sci.: Processes Impacts 21 (2019) 1280–1300. 10.1039/C9EM00228F. [DOI] [PubMed] [Google Scholar]
- [42].Brown WL, Day DA, Stark H, Pagonis D, Krechmer JE, Liu X, Price DJ, Katz EF, DeCarlo PF, Masoud CG, Wang DS, Hildebrandt Ruiz L, Arata C, Lunderberg DM, Goldstein AH, Farmer DK, Vance ME, Jimenez JL, Real-time organic aerosol chemical speciation in the indoor environment using extractive electrospray ionization mass spectrometry, Indoor Air. (2020) ina.12721. 10.1111/ina.12721. [DOI] [PubMed] [Google Scholar]
- [43].L’Orange C, Neymark G, Carter E, Volckens J, A High-throughput, Robotic System for Analysis of Aerosol Sampling Filters, Aerosol Air Qual. Res 21 (2021) 210037. 10.4209/aaqr.210037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [44].Alphasense Limited, Correcting for Background Currents in Four Electrode Toxic Gas Sensors, Great Notley, Essex, UK, 2019. www.alphasense.com. [Google Scholar]
- [45].Alphasense Limited, CO-B4 Carbon Monoxide Sensor, (2019). http://www.alphasense.com/WEB1213/wp-content/uploads/2019/09/CO-B4.pdf (accessed October 8, 2019).
- [46].Alphasense Limited, NO2-B43F Nitrogen Dioxide Sensor, (2019). http://www.alphasense.com/WEB1213/wp-content/uploads/2019/09/NO2-B43F.pdf (accessed October 8, 2019).
- [47].Alphasense Limited, OX-B431 Oxidising Gas Sensor Ozone + Nitrogen Dioxide, (2019). http://www.alphasense.com/WEB1213/wp-content/uploads/2019/09/OX-B431.pdf (accessed October 8, 2019).
- [48].Alphasense Limited, Humidity Extremes: Drying out and Water Absorption, Alphasense Limited, Great Notley, Essex, UK, 2013. http://www.alphasense.com/WEB1213/wp-content/uploads/2013/07/AAN_106.pdf. [Google Scholar]
- [49].Camalier L, Eberly S, Miller J, Papp M, Guideline on the Meaning and the Use of Precision and Bias Data Required by 40 CFR Part 58 Appendix A, U.S. Environmental Protection Agency, Research Triangle Park, NC, 2007. https://www3.epa.gov/ttn/amtic/files/ambient/qaqc/PBGuidance101007.pdf (accessed July 31, 2020). [Google Scholar]
- [50].Williams R, Kilaru V, Snyder E, Kaufman A, Dye T, Rutter A, Russell A, Hafner H, Air Sensor Guidebook, United States Environmental Protection Agency, Washington, D.C., 2014. https://cfpub.epa.gov/si/si_public_record_report.cfm?Lab=NERL&direntryid=277996. [Google Scholar]
- [51].Barkjohn KK, Gantt B, Clements AL, Development and application of a United States-wide correction for PM<sub>2.5</sub> data collected with the PurpleAir sensor, Atmos. Meas. Tech 14 (2021) 4617–4637. 10.5194/amt-14-4617-2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [52].ASHRAE, Appendix D to ANSI/ASHRAE Standard 62.1-2016: Ventilation for Acceptable Indoor Air Quality, ASHRAE, Atlanta, GA, 2018. https://www.ashrae.org/technical-resources/standards-and-guidelines/read-only-versions-of-ashrae-standards (accessed August 4, 2021). [Google Scholar]
- [53].OSHA, OSHA Annotated Table Z-1, Occupational Safety and Health Administration. (2020). https://www.osha.gov/dsg/annotated-pels/tablez-1.html (accessed November 12, 2020). [Google Scholar]
- [54].ASTM, Standard Guide for Using Indoor Carbon Dioxide Concentrations to Evaluate Indoor Air Quality and Ventilation, ASTM International, West Conshohocken, PA, 2018. [Google Scholar]
- [55].World Health Organization, ed., WHO guidelines for indoor air quality: selected pollutants, World Health Organization, Copenhagen, 2010. https://apps.who.int/iris/handle/10665/260127 (accessed November 12, 2020). [PubMed] [Google Scholar]
- [56].US EPA, Ambient Concentrations of Carbon Monoxide, U.S. Environmental Protection Agency, Washington, D.C., 2019. https://cfpub.epa.gov/roe/indicator.cfm?i=4 (accessed January 20, 2021). [Google Scholar]
- [57].National Research Council’s Committee on Toxicology, Emergency and Continuous Exposure Limits for Selected Airborne Contaminants, National Academies Press, Washington, DC, 1984. https://www.nap.edu/catalog/689/emergency-and-continuous-exposure-limits-for-selected-airborne-contaminants-volume (accessed January 20, 2021). [Google Scholar]
- [58].World Health Organization, WHO Air quality guidelines for particulate matter, ozone, nitrogen dioxide and sulfur dioxide – Global update 2005 – Summary of risk assessment, WHO, Geneva, Switzerland, 2006. https://apps.who.int/iris/handle/10665/69477 (accessed November 12, 2020). [Google Scholar]
- [59].US EPA, 8-Hour Ozone (2015) Nonattainment Area Partial County Descriptions, (2020). https://www3.epa.gov/airquality/greenbook/jnp.html#Ozone_8-hr.2015.Denver (accessed October 28, 2020).
- [60].Sensirion, Interface Description Sensirion SCD30 Sensor Module, Version 1.0–D1, (2020). https://www.sensirion.com/fileadmin/user_upload/customers/sensirion/Dokumente/9.5_CO2/Sensirion_CO2_Sensors_SCD30_Interface_Description.pdf (accessed March 12, 2021).
- [61].Borrego C, Costa AM, Ginja J, Amorim M, Coutinho M, Karatzas K, Sioumis Th., Katsifarakis N, Konstantinidis K, De Vito S, Esposito E, Smith P, André N, Gérard P, Francis LA, Castell N, Schneider P, Viana M, Minguillón MC, Reimringer W, Otjes RP, von Sicard O, Pohle R, Elen B, Suriano D, Pfister V, Prato M, Dipinto S, Penza M, Assessment of air quality microsensors versus reference methods: The EuNetAir joint exercise, Atmospheric Environment. 147 (2016) 246–263. 10.1016/j.atmosenv.2016.09.050. [DOI] [Google Scholar]
- [62].Delp WW, Singer BC, Wildfire Smoke Adjustment Factors for Low-Cost and Professional PM2.5 Monitors with Optical Sensors, Sensors. 20 (2020) 3683. 10.3390/s20133683. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.





