Skip to main content
Sage Choice logoLink to Sage Choice
. 2021 Nov 5;27(5):191–204. doi: 10.1177/14690667211055701

Poly(2-vinylpyridine) as a reference compound for mass calibration in positive-ion matrix-assisted laser desorption/ionization-mass spectrometry on different instrumental platforms

Jürgen H Gross 1,
PMCID: PMC8586192  PMID: 34738841

Abstract

Butyl-terminated poly(2-vinylpyridine) (P2VP), C4H9(C7H7N) n H, is evaluated for use as an external and internal mass calibrant in positive-ion matrix-assisted laser desorption/ionization-mass spectrometry (MALDI-MS). P2VP oligomers covering the m/z 450–4500 range are employed to calibrate a time-of-flight (TOF) mass spectrometer in linear and reflector mode, an ion mobility-quadrupole-time-of-flight (IM-Q-TOF) mass spectrometer, and a Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometer. The proton affinity of P2VPs introduced by the numerous pyridyl groups leads to the almost exclusive formation of [M + H]+ ions with common acidic matrices like α-cyano-4-hydroxycinnamic acid (CHCA) and 2,5-dihydroxybenzoic acid (DHB) as well as with the non-acidic and aprotic matrices 1,8-dihydroxy-10H-anthracen-9-on (dithranol) and 2-[(2E)-3-(4-tert-butylphenyl)-2-methylprop-2-enylidene]malonitrile (DCTB). This prevalence of [M + H]+ ions evenly spaced at Δ(m/z) = 105.0578 renders butyl-terminated P2VP oligomers as convenient mass calibrants. The mass accuracies achieved across various m/z ranges with different mass analyzers and modes of operation are evaluated by using established standard compounds. Results as obtained by internal or external calibration are presented. Further, the compilation of mass reference lists tailored to suit the respective analyzer modes is discussed and those reference files are provided.

Keywords: Matrix-assisted laser desorption/ionization (MALDI), time-of-flight (TOF), ion mobility-quadrupole-time-of-flight (IM-Q-TOF), Fourier transform ion cyclotron resonance (FT-ICR), poly(2-yinylpyridine) (P2VP), accurate mass, mass calibration, small molecules, polymers

Introduction

Matrix-assisted laser desorption/ionization (MALDI) 1 is still prevalently implemented in combination with axial time-of-flight (TOF) mass spectrometers, the latest generation of which is capable of providing sufficiently high mass resolving power (R = 10 000–25 000) for their use in accurate mass measurement, and thus, molecular formula determination. In contrast to Fourier transform ion cyclotron resonance (FT-ICR) or Orbitrap analyzers, axial TOF analyzers tend to exhibit a drift in mass calibration on the minute-time scale. While these drifts on the minute-time scale are comparatively small on the latest generation instruments, they are still in the order of 5–20 ppm, and thus, preclude or at least restrict the use of external mass calibration for accurate mass measurements. In case of MALDI-MS with axial TOF analyzers, drift in m/z is not solely affected by instrumental fluctuations but even more so by local changes in sample layer thickness and matrix crystallization. The goal of accurate mass measurement in MALDI-TOF-MS is thus best achieved by employing an internal mass calibrant.

The level of mass accuracy required to unambiguously derive a molecular formula from a measured accurate mass much depends on the actual m/z of the ion, the variety of elements to be considered, and the number of atoms of these elements to be taken into account.25 For example, depending on the particular restrictions, an unequivocal formula assignment by accurate mass alone may be achieved up to about m/z 500 6 but can become extremely demanding in other situations. 7 Despite the fact that there is no single generally valid level of mass accuracy to apply to any analytical quest, chemistry journals tend to define and require these limits according to their editors’ own opinion. The Journal of Organic Chemistry published by the American Chemical Society, for example, states “… a found value within 0.003 m/z unit of the calculated value of a parent-derived ion … is usually adequate”. 8 Angewandte Chemie (and other Wiley journals) requires that “high resolution … data should be provided to an accuracy within … ±0.003 of the calculated values.” (whatever the unit maybe). 9 The Royal Society of Chemistry journals are a bit more conservative in asking that “Exact masses quoted for identification purposes should be accurate to within 5 ppm (EI and CI) or 10 ppm (FAB or LSIMS).”. 10 To sum it up in brief and simplicity: a mass accuracy of 1 ppm would be desirable, 3 ppm do cover most journals’ (and thus our clients’) requested levels, and 5 ppm still solve many problems, and at times, can be sufficient for publication.

A reference compound intended to be used for mass calibration should have the following properties: (i) it has to yield intensive signals evenly spaced across an ideally wide m/z range, (ii) it should essentially provide a single series of peaks to minimize the risk of erroneous peak assignments, (iii) when used as an internal mass calibrant it should not suppress the analyte or, vice versa, be suppressed by the analyte, and (iv) in case of MALDI, it needs to be compatible with the preferred matrix (and ideally with others, too). 11

To establish mass calibration over a wide m/z range in combination with desorption/ionization techniques numerous procedures based on cluster ion series have been described. They all bear the advantage that they are typically generated from readily available compounds additionally offering a year-long shelf life even when stored in ready-to-use solutions. For example, [arginine n  + H]+ and [arginine n  − H] cluster ions can be used for mass calibration in both positive-ion and negative-ion electrospray ionization (ESI), respectively.12,13 For use in direct analysis in real time (DART), the ionic liquid (IL) 1-butyl-3-methylimidazolium tricyanomethide delivers ions covering m/z 100–4000 in positive-ion and m/z 100–2000 in negative-ion DART-MS. The IL provides a wide distribution of cluster ions at Δ(m/z) 229.1330 reflecting the mass of the pair of cation and anion.14,15 Alternatively, saccharose cluster ions may be employed for mass calibration in positive-ion DART-MS across the m/z 100–2000 range. 16 In MALDI-MS, [Cs n I n –1]+ and [Cs n –1I n ] cluster ions are also well established for mass calibration. These caesium salt cluster ions of either polarity are effectively formed from caesium triiodide, CsI3, in 2-[(2E)-3-(4-tert-butylphenyl)- 2-methylprop-2-enylidene]malonitrile (DCTB) matrix. 17 Due to the versatility of the [Cs n I n –1]+ and [Cs n –1I n ] cluster ion series some variations of this approach have been developed.18,19 While generally allowing for their use in positive-ion and negative-ion modes alike, the generation of cluster ions demands their constituents to be present at comparatively high concentration. Mutual interference with analyte ion formation then either tends to cause suppression of analyte ion or calibrant cluster ion generation, thereby often precluding the use of cluster ions as internal mass calibrants. Thus, cluster ion-based mass calibration procedures are generally limited to use for external mass calibration. Additionally, cluster ion series often have the disadvantage of showing a notable decrease of higher-mass cluster ion abundances.

Polymers or branched molecules of defined molecular mass are superior in this regard. In positive- and negative-ion ESI-MS, a mixture composed of ammonium trifluoroacetate, betaine, 2,4,6-tris(heptafluoropropyl)- 1,3,5-triazine, and various symmetrical hexakis-(fluoroalkoxy)-phosphazenes, also known as Agilent Tune Mix, is well established.2022 Agilent Tune Mix may also serve for mass calibration in negative-ion DART-MS. 23 Other mass calibration procedures for MALDI use poly(ethylene glycols) (PEGs),24,25 basic poly(propylene glycols) (Jeffamines) 11 or monodisperse dendrimers (SpheriCal)26,27 for wide range mass calibration. 28 Dendrimers with a tailored mass defect to avoid interferences with typical organic analyte ions are also available. 29 Polyalanine has been presented as a calibrant in MALDI-MS as it covers the range m/z 1000–4000 in both positive-ion and negative-ion mode. 30 Unfortunately, it is neither compatible with DCTB matrix 31 nor does it allow to yield spectra at low laser fluence crucial for optimum resolution in MALDI-TOF-MS.

The first publication dealing with MALDI-MS of poly(2-vinylpyridine) (P2VP) presented the spectrum of a sample of an average molecular weight of about 28 kDa prepared in dithranol matrix doped with sodium chloride. 32 Other work dealt with P2VP block copolymers 33 or poly(4-vinylpyridine) (P4VP) with highly polar end groups. 34 Here, butyl-terminated P2VP, C4H9(C7H7N) n H, (n = 4–48), is introduced as a convenient mass calibrant for positive-ion MALDI-MS to be used as an external or internal reference, in combination with common MALDI matrices, and with different mass analyzer configurations.

Experimental

Poly(2-vinylpyridine) standards and sample preparation

Three samples of poly(2-vinylpyridine), i.e., PVP-670, PVP-1k, and PVP-2.1k (PSS Polymer Standards Service GmbH, Mainz, Germany), were used without further treatment. These P2VPs are butyl-terminated, and thus, obey the general formula C4H9(C7H7N) n H.

For MALDI sample preparation, stock solutions of the P2VPs at 5 mg ml–1 in tetrahydrofuran were prepared and stored at 5 °C until use. These solutions could be used for weeks without change. P2VP stock solutions were admixed to the matrix 2-[(2E)-3-(4-tert-butylphenyl)-2-methylprop-2-enylidene] malononitrile (DCTB) that was employed as solution at 10 mg ml–1 in tetrahydrofuran. The ratio of analyte-to-matrix solution was 1 : 50 and 1 µl of these solutions was used per sample spot on the stainless steel Scout 384 sample plate.

For testing the compatibility of P2VPs with the matrices α-cyano-4-hydroxy­cinnamic acid (CHCA), 2,5-dihydroxybenzoic acid (DHB), and 1,8-dihydroxy-10H-anthracen-9-on (dithranol), solutions of these matrices were prepared at 10 mg ml–1 in acetone. The ratio of analyte-to-matrix solutions was 1 : 50. Structures of the matrices and of P2VP are displayed in Figure 1.

Figure 1.

Figure 1.

Structures of the matrices 2-[(2E)-3-(4-tert-butylphenyl)-2-methylprop-2-enylidene] malononitrile (DCTB), 1,8-Dihydroxy-10H-anthracen-9-on (dithranol), 2,5-dihydroxybenzoic acid (DHB), α-cyano-4-hydroxycinnamic acid (CHCA), and of the mass calibrant poly(2-vinylpyridine) (P2VP).

MALDI-TOF-MS

The MALDI-TOF spectra were acquired using a Bruker Autoflex Speed MALDI-TOF mass spectrometer (Bruker Daltonik GmbH, Bremen, Germany) in both positive-ion linear and reflector mode. The instrument was equipped with a Bruker SmartBeam laser (Nd:YAG, frequency tripled, wavelength 355 nm). It was controlled by Bruker FlexControl 3.4 software and mass spectra were processed using Bruker FlexAnalysis 3.4.

The essential instrument settings for the m/z 600–6000 range in linear mode and two typical ranges in reflector mode, i.e., m/z 200–2000 and m/z 440–5000, are summarized in Table S1. In reflector mode, intermediate ranges like m/z 350–3500 were also used as required.

Prior to initial experiments, an independent external mass calibration for positive-ion mode was established based on [Cs n I n –1]+ cluster ions generated from caesium triiodide, CsI3, in DCTB matrix. 17 Mass calibration of the instrument was performed by manual peak assignment to the reference list and application of the cubic enhanced algorithm of Bruker FlexControl 3.4

For use as an internal mass calibrant, the solution of the P2VP best suited for the intended m/z range was admixed to the analyte-matrix solution as to achieve relative peak intensities of standard to analyte in the range of 1 : 10 to 3 : 1. The laser fluence was adjusted in order to deliver good intensity signals without sacrificing mass resolution. After assignment of as many as possible reference peaks, internal mass calibration was performed using the cubic enhanced algorithm of Bruker FlexAnalysis 3.4.

IM-Q-TOF-MS

A Bruker timsTOF flex ion mobility-quadrupole-time-of-flight (IM-Q-TOF) mass spectrometer (Bruker Daltonik GmbH, Bremen, Germany) was used. The instrument was equipped with a Bruker SmartBeam laser (Nd:YAG, frequency tripled, wavelength 355 nm). In MALDI mode the instrument was controlled by the Bruker timsControl software (V 2.0) and data analysis was performed using the Bruker DataAnalysis software (V 5.3). The instrument was set to acquire the m/z 250–5000 range. Most relevant settings are summarized in Table S2.

Prior to initial experiments, an independent external mass calibration for positive-ion mode was established based on [Cs n I n –1]+ cluster ions generated from caesium triiodide, CsI3, in DCTB matrix. 17 Mass calibration of the instrument was performed by automatic peak assignment to the reference list and application of the cubic enhanced algorithm of Bruker timsControl software (V 2.0). As this study focuses on m/z calibration, the TIMS stage was switched off and IMS was not considered in these experiments.

For use as an external mass calibrant, the solutions of the P2VPs best suited to evenly cover the intended m/z range were admixed to the matrix solution, typically PVP-670 and PVP-2.1k at a ratio of 1 : 3. The laser fluence was adjusted in order to deliver good intensity signals without sacrificing mass resolution. After assignment of as many as possible reference peaks, mass calibration was performed using the cubic enhanced algorithm of Bruker timsControl software (V 2.0).

FT-ICR-MS

A Bruker Apex-Qe Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometer (Bruker Daltonik GmbH, Bremen, Germany) equipped with a 9.4 T superconducting magnet and an ESI-to-MALDI switchable Dual Source MTP was used. The instrument was equipped with a frequency tripled Nd:YAG laser (wavelength 355 nm). The instrument was controlled by the Bruker ApexControl software (V 3.0.0) and data analysis was performed using the Bruker DataAnalysis software (V 4.3).

For MALDI-MS, ions generated by sets of 15–30 laser shots were collected for 0.05 s prior to ICR mass analysis in the RF-only accumulation hexapole (h2). Ions were excited and detected using standard settings from previous work.23,35,36 As with the TOF instrument before, an independent external mass calibration for positive-ion mode was established based on [Cs n I n –1]+ cluster ions generated from caesium triiodide, CsI3, in DCTB matrix. 17

In positive-ion DART mode, the IL 1-butyl-3-methylimidazolium tricyanomethide was used for mass calibration prior to analyzing P2VPs.14,15 In positive-ion ESI mode, the instrument was externally calibrated using Agilent Tune Mix.2022

Generally, 16–24 transients were accumulated to yield a final FT-ICR mass spectrum. When the range m/z 350–3500 was selected, a 1 M data points transient resulted in a resolving power of R = 50 000 at m/z 1000. When the range m/z 740–4500 was selected, a 1 M data points transient resulted in a resolving power of R = 95 000 at m/z 1000.

Results and discussion

General characteristics of P2VP in positive-ion MALDI-MS

First, the three P2VP standards were each admixed to any of the four mentioned matrices (DCTB, CHCA, DHB, dithranol) and the positive-ion MALDI spectra were acquired in both linear and reflector mode of the TOF instrument. In any of these combinations, all P2VPs yield useful spectra that are characterized by the prevalence of just one series of ions (Fig. S1–S8). Based on the reflector mode spectra exhibiting good isotopic separation and with external calibration delivering a mass accuracy in the order of 0.1 u these signals can be assigned to [M + H]+ ions. This tentative assignment is, for example, supported by the P2VP 9mer [M + H]+ ion, [C67H74N9]+ at m/z 1004.6 (calc. m/z 1004.6061), shown in expanded views of this signal in the spectra of PVP-1k as obtained with any of these matrices (Figure 2). The formation of [M + H]+ ions can be expected due to the proton affinity of P2VPs introduced by the numerous pyridyl groups. Advantageously, the almost exclusive formation of [M + H]+ ions does not only occur with the acidic matrices CHCA and DHB but also when the non-acidic matrix dithranol and even the aprotic matrix DCTB are employed.

Figure 2.

Figure 2.

Positive-ion reflector mode MALDI-TOF spectra of PVP-1k in (a) CHCA, (b) dithranol), (c) 2,5-DHB, and (d) DCTB as acquired using the Bruker Autoflex instrument. For comparison of the relative response all spectra were acquired with the laser power set to 45%. The highest intensities in the center of the spectra approximate (a) 4.0 × 104 a.u. (a.u.: arbitrary units of signal intensity), (b) 7.0 × 103 a.u., (c) 1.4 × 103 a.u., and (d) 8.0 × 103 a.u., respectively. The inserts on the right show an expanded view of the signal at m/z 1004.6 corresponding to P2VP 9mer [M + H]+ ions. Note that the peak picking algorithm labeled the most intensive signal of each of the isotopic patterns, and thus, starting with the 12mer, labels begin referring to those peaks. Further, in case of the 12mer and the 13mer minor fluctuations of relative intensity cause the labels to either refer to the monoisotopic or the first isotopic peak.

While any of the three P2VPs, together covering the m/z 450–5000 range, showed good compatibility with all of these common matrices, there was a notable difference in signal intensity at the same laser fluence. This is demonstrated by comparing the spectra of PVP-1k in (a) CHCA, (b) dithranol), (c) 2,5-DHB, and (d) DCTB as representative of this behavior (Figure 2). With the laser power set to 45% (term used by the software to describe laser fluence) the highest intensities in the center of the spectra approximate (a) 4.0 × 104 a.u. (a.u.: arbitrary units of signal intensity), (b) 7.0 × 103 a.u., (c) 1.4 × 103 a.u., and (d) 8.0 × 103 a.u., respectively. The linear mode spectra exhibit the same tendency. Here PVP-1k in (a) CHCA, (b) dithranol), (c) 2,5-DHB, and (d) DCTB with the laser power set to 25% the highest intensities in the center of the range approximate (a) 2.5 × 105 a.u., (b) 5.0 × 103 a.u., (c) 2.3 × 102 a.u., and (d) 8.0 × 104 a.u., respectively (Fig. S9). Obviously, 2,5-DHB requires the highest laser fluence for effective analyte ion formation followed by dithranol. CHCA and DCTB perform best in that they yield good signal intensities at low laser fluence. To achieve the coverage of any m/z range of interest, different P2VPs can simply be mixed (Fig. S10). As DCTB is the most used matrix in this laboratory, unless otherwise noted, the following work is based on DCTB.

To provide an independent confirmation of the formulas of the reference ions to be used in MALDI-MS, the positive-ion DART-FT-ICR spectrum of PVP-670 was acquired using an established procedure (Fig. S11). 14 While the m/z range covered by the DART spectrum with the temperature of the helium gas set to 450 °C did not fully coincide with the m/z range observed in MALDI-MS, the spectral data nonetheless confirmed the formulas of a subset of these ions in the central section of the m/z range of interest, i.e., from the 4mer [C32H39N4]+ at m/z 479.3175 (calc. 479.3169) to the 10mer [C74H81N10]+ at m/z 1109.6665 (calc. 1109.6640). The formula assignments and mass accuracies of all seven ions are also listed in Fig. S11. As the application of DART turned out to be limited to PVP-670, the consistency of the series was additionally checked by measuring the positive-ion ESI-FT-ICR spectrum of PVP-1k in acetonitrile : water : THF = 4 : 2 : 1 with 0.1% trifluoroacetic (Fig. S12). Here, mass calibration was established using Agilent Tune Mix.2022 The ESI-FT-ICR spectrum of PVP-1k covered the 8mer [C60H67N8]+ at m/z 899.5475 (calc. 899.5483) to the 18mer [C130H137N18]+ at m/z 1950.1269 (calc. 1950.1268). The formula assignments and mass accuracies of all eleven ions are compiled in Fig. S12. Thus, the assignment of the ion series to P2VP [M + H]+ ions was independently confirmed by either DART-FT-ICR-MS or ESI-FT-ICR-MS.

Both the prevalence of [M + H]+ ion formation and the occurrence of an ion series evenly spaced at Δ(m/z) = 105.0578 as calculated for a C7H7N monomer unit indicate that P2VPs are potentially convenient mass calibrants.

Building mass reference lists

Based on these results, a mass reference list can be compiled. As the use of P2VPs is intended in combination with different mass analyzers, one has to take into account that resolving power at a given m/z dictates whether isotopic resolution allows for the unambiguous assignment of monoisotopic [M + H]+ ions or whether average mass values are required to deal with isotopically unresolved envelopes.

Among the mass analyzers employed in this study, reflector TOF, IM-Q-TOF, and FT-ICR analyzer delivered isotopic resolution beyond m/z 5000, and thus, for use as reference values in combination with these, monoisotopic [M + H]+ ion masses were calculated (Table 1). Upon growing chain length of the P2VPs, the relative abundances of the monoisotopic ions drop relative to that of the 13C1, 13C2, and 13C3 isotopolog ions, respectively. Therefore, Table 1 provides 13C1, 13C2, and 13C3 isotopolog ion masses and shows some overlap of the four reference m/z columns in order to permit proper selection of reference m/z values with instruments of resolution characteristics differing from those used in this study. The reference list was found to work best when it provided the m/z values calculated for the most abundant isotopolog ion of the respective oligomer (discussion below). Thus, for ions larger than the 11mer, [C81H88N11]+, m/z 1214.72187 (calc.), i.e., starting from the 12mer, the mass reference list shifts to refer the 13C1 ionic compositions, e.g., [13C1C87H95N12]+, m/z 1320.78308 (calc.) as reference ions. Then, starting with the 25mer, the list shifts to refer the 13C2 ionic compositions, e.g., [13C2C177H186N25]+, m/z 2687.53847 (calc.). Finally, from the 39mer, [13C3C274H284N39]+, m/z 4159.35170, to the 48mer, [13C3C336H347N48]+, m/z 5104.87234 (calc.), the list refers to the 13C3 ionic compositions.

Table 1.

Compositions and calculated m/z values of P2VP [M + H]+ ions as used in mass reference lists for calibration.

Calculated m/z of [M + H]+
Ionic composition of 2mer to 48mer Monoisotopic 13C1-Ion 13C2-Ion 13C3-Ion Average mass
[C4H9(C7H7N)2H + H]+ 269.20123
[C4H9(C7H7N)3H + H]+ 374.25907
[C4H9(C7H7N)4H + H]+ 479.31692
[C4H9(C7H7N)5H + H]+ 584.37477
[C4H9(C7H7N)6H + H]+ 689.43262
[C4H9(C7H7N)7H + H]+ 794.49047 795.092
[C4H9(C7H7N)8H + H]+ 899.54832 900.230
[C4H9(C7H7N)9H + H]+ 1004.60617 1005.367
[C4H9(C7H7N)10H + H]+ 1109.66402 1110.505
[C4H9(C7H7N)11H + H]+ 1214.72187 1215.642
[C4H9(C7H7N)12H + H]+ 1319.77972 1320.78302 1320.780
[C4H9(C7H7N)13H + H]+ 1424.83756 1425.84086 1425.917
[C4H9(C7H7N)14H + H]+ 1529.89541 1530.89871 1531.054
[C4H9(C7H7N)15H + H]+ 1634.95326 1635.95656 1636.192
[C4H9(C7H7N)16H + H]+ 1740.01111 1741.01441 1741.329
[C4H9(C7H7N)17H + H]+ 1845.06896 1846.07226 1846.467
[C4H9(C7H7N)18H + H]+ 1950.12681 1951.13011 1951.604
[C4H9(C7H7N)19H + H]+ 2055.18466 2056.18796 2056.742
[C4H9(C7H7N)20H + H]+ 2160.24251 2161.24581 2161.879
[C4H9(C7H7N)21H + H]+ 2265.30036 2266.30366 2267.017
[C4H9(C7H7N)22H + H]+ 2370.35821 2371.36151 2372.154
[C4H9(C7H7N)23H + H]+ 2475.41605 2476.41935 2477.291
[C4H9(C7H7N)24H + H]+ 2581.47720 2582.429
[C4H9(C7H7N)25H + H]+ 2686.53505 2687.53846 2687.566
[C4H9(C7H7N)26H + H]+ 2791.59290 2792.59631 2792.704
[C4H9(C7H7N)27H + H]+ 2896.65075 2897.65416 2897.841
[C4H9(C7H7N)28H + H]+ 3001.70860 3002.71201 3002.979
[C4H9(C7H7N)29H + H]+ 3106.76645 3107.76986 3108.116
[C4H9(C7H7N)30H + H]+ 3211.82430 3212.82771 3213.253
[C4H9(C7H7N)31H + H]+ 3316.88215 3317.88556 3318.391
[C4H9(C7H7N)32H + H]+ 3421.94000 3422.94341 3423.528
[C4H9(C7H7N)33H + H]+ 3526.99784 3528.00125 3528.666
[C4H9(C7H7N)34H + H]+ 3633.05910 3633.803
[C4H9(C7H7N)35H + H]+ 3738.11695 3738.941
[C4H9(C7H7N)36H + H]+ 3843.17480 3844.078
[C4H9(C7H7N)37H + H]+ 3948.23265 3949.23601 3949.216
[C4H9(C7H7N)38H + H]+ 4053.29050 4054.29385 4054.353
[C4H9(C7H7N)39H + H]+ 4158.34835 4159.35170 4159.490
[C4H9(C7H7N)40H + H]+ 4263.40620 4264.40955 4264.628
[C4H9(C7H7N)41H + H]+ 4368.46405 4369.46740 4369.765
[C4H9(C7H7N)42H + H]+ 4473.52190 4474.52525 4474.903
[C4H9(C7H7N)43H + H]+ 4579.58310 4580.040
[C4H9(C7H7N)44H + H]+ 4684.64095 4685.178
[C4H9(C7H7N)45H + H]+ 4789.69880 4790.315
[C4H9(C7H7N)46H + H]+ 4894.75665 4895.453
[C4H9(C7H7N)47H + H]+ 4999.81450 5000.590
[C4H9(C7H7N)48H + H]+ 5104.87234 5105.727

To deal with the detection of envelopes covering the isotopic pattern in case of higher-mass ions in linear mode TOF-MS, Table 1 also provides a column with [M + H]+ ion average masses.

Reflector mode TOF-MS

Axial TOF instruments can still be considered the (gold) standard in MALDI-MS, and thus, the data obtained by reflector mode MALDI-TOF instrumentation shall be discussed first The present instrument, typically providing a mass resolving power of 15,000 at m/z 1000, easily delivered isotopic resolution over the entire range covered by the P2VPs. Based on external calibration, the achievable mass accuracy depended on the spatial distance between sample spot and reference spot and on the temporal gap between measuring these two. Provided the reference was placed on an adjacent spot and both the sample and the mass calibrant spectrum were acquired within about one minute, the mass accuracy was in the 3–10 ppm range. Polyethylene glycols (PEGs) provide well-defined series of [M + Na]+ and/or [M + K]+ ions in MALDI-MS.24,25 Here, PEG 600, PEG 1000, and PEG 1500 were prepared in DCTB matrix. The data obtained by alternating acquisition of spectra of P2VP, performing mass calibration and applying this to the next spectrum to be acquired of PEG 1000 is presented in Table 2. While both [M + Na]+ and [M + K]+ ions of PEG 1000 were formed, for clarity, only the more abundant [M + K]+ ions were considered here. Analogous results were also obtained with PEG 600 and PEG 1500 as long as the m/z range of the PEGs was covered by reference peaks of a suitable P2VP oligomer. Overall, the level of mass accuracy required for formula determination was normally not achieved by external mass calibration. Nonetheless, these calibrations were stable at ±0.2 u for weeks, and could thus be used as long as accurate mass was not required.

Table 2.

Mass accuracy by reflector mode MALDI-TOF-MS based on external calibration. Data of PEG 1000 [M + K]+ ions, five repetitions, standard deviation of m/z in ppm, average error Δ(m/z) in mu. Conservatively, the accuracy level is at 10 ppm.

Formula Calc. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Avg. Exp. m/z Std. Dev.
[C36H74O19K]+ 849.4456 849.4424 849.4408 849.4365 849.4375 849.4386 849.4392 2.8
Δ(m/z) 0.0032 0.0048 0.0091 0.0081 0.0070 0.0064 7.6
[C38H78O20K]+ 893.4718 893.4699 893.4682 893.4645 893.4655 893.4658 893.4668 2.5
Δ(m/z) 0.0019 0.0036 0.0073 0.0063 0.0060 0.0050 5.6
[C40H82O21K]+ 937.4980 937.4958 937.4921 937.4897 937.4906 937.4912 937.4919 2.5
Δ(m/z) 0.0022 0.0059 0.0083 0.0074 0.0068 0.0061 6.5
[C42H86O22K]+ 981.5242 981.5241 981.5225 981.5191 981.5196 981.5222 981.5215 2.1
Δ(m/z) 0.0001 0.0017 0.0051 0.0046 0.0020 0.0027 2.8
[C44H90O23K]+ 1025.5504 1025.5503 1025.5478 1025.5441 1025.5451 1025.5469 1025.5468 2.4
Δ(m/z) 0.0001 0.0026 0.0063 0.0053 0.0035 0.0036 3.5
[C46H94O24K]+ 1069.5767 1069.5788 1069.5765 1069.5719 1069.5739 1069.5764 1069.5755 2.5
Δ(m/z) −0.0021 0.0002 0.0048 0.0028 0.0003 0.0012 1.1
[C48H98O25K]+ 1113.6029 1113.6077 1113.6063 1113.6003 1113.6035 1113.6064 1113.6048 2.7
Δ(m/z) −0.0048 −0.0034 0.0026 −0.0006 −0.0035 −0.0020 −1.8
[C50H102O26K]+ 1157.6291 1157.6362 1157.6351 1157.6280 1157.6323 1157.6355 1157.6334 2.9
Δ(m/z) −0.0071 −0.0060 0.0011 −0.0032 −0.0064 −0.0043 −3.7
[C52H106O27K]+ 1201.6553 1201.6641 1201.6612 1201.6550 1201.6598 1201.6630 1201.6606 3.0
Δ(m/z) −0.0088 −0.0059 0.0003 −0.0045 −0.0077 −0.0053 −4.4
[C54H110O28K]+ 1245.6815 1245.6892 1245.6875 1245.6781 1245.6885 1245.6934 1245.6873 4.5
Δ(m/z) −0.0077 −0.0060 0.0034 −0.0070 −0.0119 −0.0058 −4.7
[C56H114O29K]+ 1289.7077 1289.7195 1289.7126 1289.7068 1289.7149 1289.7167 1289.7141 3.7
Δ(m/z) −0.0118 −0.0049 0.0009 −0.0072 −0.0090 −0.0064 −4.9
[C58H118O30K]+ 1333.7340 1333.7506 1333.7447 1333.7348 1333.7470 1333.7509 1333.7456 4.9
Δ(m/z) −0.0166 −0.0107 −0.0008 −0.0130 −0.0169 −0.0116 −8.7
[C60H122O31K]+ 1377.7602 1377.7706 1377.7672 1377.7533 1377.7625 1377.7730 1377.7653 5.7
Δ(m/z) −0.0104 −0.0070 0.0069 −0.0023 −0.0128 −0.0052 −3.7

When used as internal mass calibrant, the P2VPs were quite versatile. To avoid suppression of the analyte of interest one had to carefully adjust the amount of P2VP relative to the analyte to achieve an intensity ratio in the range of about 1 : 3 to 3 : 1. The positive-ion reflector mode MALDI-TOF spectrum of PVP-670 admixed to PEG 1000 exemplifies this approach. The spectrum shows [PVP + H]+ ions, [PEG + Na]+ ions, and [PEG + K]+ ions (Figure 3). Normally, the spacing of Δ(m/z) = 105.0578 between reference peaks is wide enough to encompass several analyte ion signals. However, there is some interference of calibrant and analyte could occur. In this particular spectrum there is an overlap with an isotopic peak of the [PEG + Na]+ ion at m/z 1317.7685. Proper operation provided, this technique generally delivered mass accuracies of 2–5 ppm (Table 3). Again, analogous results were also obtained with PEG 600 and PEG 1500 as long as the m/z range of the PEGs was covered by reference peaks of a suitable P2VP. Nonetheless, admixing the calibrant at a ratio that yielded well-defined reference peaks while not suppressing the analyte ions often required the preparation of several spots on the MALDI target. Further, this procedure is restricted to analytes tolerant to the basic P2VP calibrant.

Figure 3.

Figure 3.

Positive-ion reflector mode MALDI-TOF spectrum of PVP-670 admixed to PEG 1000 in DCTB matrix. [PVP + H]+ ions are marked with diamonds, [PEG + Na]+ ions with empty squares, and [PEG + K]+ ions with filled squares. The insert shows that the gap between reference peaks is wide enough to encompass several analyte ion signals. However, there is some interference of calibrant and analyte with an isotopic peak of the [PEG + Na]+ ion at m/z 1317.7685.

Table 3.

Mass accuracy by reflector mode MALDI-TOF-MS based on internal calibration. Data of PEG 1000 [M + Na]+ ions, five repetitions, standard deviation of m/z in ppm, average error Δ(m/z) in mu.

Formula Calc. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Avg. Exp. m/z Std. Dev.
[C36H74O19Na]+ 833.4717 833.4715 833.4700 833.4698 833.4700 833.4682 833.4699 1.4
Δ(m/z) 0.0002 0.0017 0.0019 0.0017 0.0035 0.0018 2.1
[C38H78O20Na]+ 877.4979 877.4987 877.5000 877.4997 877.4979 877.4985 877.4990 1.0
Δ(m/z) −0.0008 −0.0021 −0.0018 0.0000 −0.0006 −0.0011 −1.2
[C40H82O21Na]+ 921.5241 921.5251 921.5261 921.5222 921.5232 921.5216 921.5236 2.1
Δ(m/z) −0.0010 −0.0020 0.0019 0.0009 0.0025 0.0004 0.5
[C42H86O22Na]+ 965.5503 965.5483 965.5483 965.5492 965.5464 965.5476 965.5480 1.1
Δ(m/z) 0.0020 0.0020 0.0011 0.0039 0.0027 0.0023 2.4
[C44H90O23Na]+ 1009.5765 1009.5766 1009.5766 1009.5781 1009.5790 1009.5780 1009.5777 1.0
Δ(m/z) −0.0001 −0.0001 −0.0016 −0.0025 −0.0015 −0.0011 −1.1
[C46H94O24Na]+ 1053.6027 1053.6007 1053.6029 1053.6043 1053.6035 1053.6052 1053.6033 1.6
Δ(m/z) 0.0020 −0.0002 −0.0016 −0.0008 −0.0025 −0.0006 −0.6
[C48H98O25Na]+ 1097.6289 1097.6276 1097.6289 1097.6285 1097.6307 1097.6308 1097.6293 1.3
Δ(m/z) 0.0013 0.0000 0.0004 −0.0018 −0.0019 −0.0004 −0.3
[C50H102O26Na]+ 1141.6552 1141.6534 1141.6534 1141.6566 1141.6567 1141.6601 1141.6560 2.4
Δ(m/z) 0.0018 0.0018 −0.0014 −0.0015 −0.0049 −0.0009 −0.8
[C52H106O27Na]+ 1185.6814 1185.6789 1185.6789 1185.6796 1185.6828 1185.6839 1185.6808 2.0
Δ(m/z) 0.0025 0.0025 0.0018 −0.0014 −0.0025 0.0005 0.5
[C54H110O28Na]+ 1229.7076 1229.7074 1229.7080 1229.7072 1229.7123 1229.7116 1229.7093 2.0
Δ(m/z) 0.0002 −0.0004 0.0004 −0.0047 −0.0040 −0.0017 −1.4
[C56H114O29Na]+ 1273.7338 1273.7307 1273.7277 1273.7337 1273.7353 1273.7390 1273.7333 3.4
Δ(m/z) 0.0031 0.0061 0.0001 −0.0015 −0.0052 0.0005 0.4
[C58H118O30Na]+ 1317.7600 1317.7609 1317.7539 1317.7612 1317.7642 1317.7686 1317.7618 4.1
Δ(m/z) −0.0009 0.0061 −0.0012 −0.0042 −0.0086 −0.0017 −1.3
[C60H122O31Na]+ 1361.7862 1361.7837 1361.7911 1361.7796 1361.7826 1361.7866 1361.7847 3.2
Δ(m/z) 0.0025 −0.0049 0.0066 0.0036 −0.0004 0.0015 1.1

Linear mode TOF-MS

Using the TOF analyzer in linear mode introduces a limitation in mass resolving power. Thus, the positive-ion linear mode MALDI spectrum of a mixture of PVP-670 plus PVP-2.1k (1 : 3) covering the m/z 600–5000 range shows a transition from isotopic resolution over an intermediate zone to the detection of envelopes over the entire isotopic distribution (Figure 4). While the instrument used still delivered good isotopic separation up to the 10mer, the signals above about m/z 1200 showed a coalescence that finally resulted in isotopically unresolved envelopes for the 20mer and larger ions. This gradual loss of isotopic resolution is exemplified by the inserts of Figure 4 showing expanded views of signals corresponding to [M + H]+ ions of the resolved peaks related to the 6mer and 9mer, to skewed peaks of the 12mer, and finally to evenly shaped envelopes as in case of the 26mer or 32mer.

Figure 4.

Figure 4.

Positive-ion linear mode MALDI spectrum of a mixture of PVP-670 plus PVP-2.1k (1 : 3) in DCTB. The inserts show expanded views of signals corresponding to [M + H]+ ions across the m/z range exemplifying the transition from isotopic resolution (n = 6 and 9) over an intermediate zone (n = 12) to the detection of envelopes over the isotopic distribution (n = 26 and 32).

For mass calibration, this dictates that depending on the actual resolving power, monoisotopic m/z values need to be employed in the lower portion of the m/z range while average masses are to be used at higher m/z. With the particular instrument, sufficient resolution typically allowed for the assignment of the monoisotopic m/z values up to the 10mer (m/z 1109.66) whereas the transition from isotopic resolution to a sufficiently uniform envelope over the isotopic peaks caused problems in this regard as the peak position neither reflects the neat monoisotopic nor the correct average mass of the ion. Signals above the 22mer (m/z 2372.12) reflected quite clean envelopes of the isotopic distribution and could confidently be assigned to average mass values. In practice, calibration of the wide m/z range in linear mode was realized by selecting and including all monoisotopic ions up the 11mer, then omitting poorly defined peaks of the intermediate region, and finally assigning the evenly shaped envelope signals to average m/z values. The calibration was then performed using the “cubic enhanced” algorithm provided by the instrument software.

The need to interpolate the mid-range combined with limited resolution in the low-mass range resulted in a comparatively low mass accuracy as demonstrated on oligosaccharide [M + Na]+ ions (Table 4). The oligosaccharides were extracted from gummy bears and admixed to DHB matrix.37,38 While ions starting from [C84H142O71Na]+, m/z 2310.976 (calc.) and up to [C114H192O96Na]+, m/z 3121.680 (calc.) appeared at a Δ(m/z) of less than 0.15 u, the ions below tended to deviate from the calculated value by about 0.25–0.40 u. In case of linear mode TOF-MS, the established [Cs n I n –1]+ cluster ion calibration1719 admittedly provided much better accuracy, because there, isotopic resolution is not an issue.

Table 4.

Mass accuracy by linear mode MALDI-TOF-MS based on external calibration. Data of oligosaccharide [M + Na]+ ions, five repetitions, standard deviation of m/z in ppm, average error Δ(m/z) in mu.

Formula Calc. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Avg. Exp. m/z Std. Dev.
[C24H42O21Na]+ 689.568 689.251 689.255 689.253 689.252 689.237 689.250 10.4
Δ(m/z) 0.317 0.313 0.315 0.316 0.331 0.318 461.8
[C30H52O26Na]+ 851.709 851.335 851.338 851.335 851.336 851.331 851.335 3.0
Δ(m/z) 0.374 0.371 0.374 0.373 0.378 0.374 439.0
[C36H62O31Na]+ 1013.849 1013.467 1013.473 1013.470 1013.472 1013.470 1013.470 2.3
Δ(m/z) 0.383 0.377 0.380 0.378 0.380 0.379 374.1
[C42H72O36Na]+ 1175.990 1175.600 1175.610 1175.606 1175.612 1175.597 1175.605 5.4
Δ(m/z) 0.390 0.380 0.384 0.378 0.393 0.385 327.7
[C48H82O41Na]+ 1338.131 1337.759 1337.761 1337.758 1337.765 1337.748 1337.758 4.7
Δ(m/z) 0.372 0.370 0.373 0.366 0.383 0.373 278.8
[C54H92O46Na]+ 1500.272 1499.981 1499.882 1499.885 1499.896 1499.888 1499.906 28.0
Δ(m/z) 0.291 0.390 0.387 0.376 0.384 0.366 243.7
[C60H102O51Na]+ 1662.413 1661.995 1662.018 1662.036 1662.007 1661.997 1662.011 10.2
Δ(m/z) 0.418 0.395 0.377 0.406 0.416 0.402 241.9
[C66H112O56Na]+ 1824.554 1824.873 1824.883 1824.873 1824.884 1824.872 1824.877 3.3
Δ(m/z) −0.319 −0.329 −0.319 −0.330 −0.318 −0.323 −177.3
[C72H122O61Na]+ 1986.694 1986.973 1987.000 1986.990 1986.992 1986.959 1986.983 8.3
Δ(m/z) −0.279 −0.306 −0.296 −0.298 −0.265 −0.288 −145.2
[C78H132O66Na]+ 2148.835 2149.038 2149.067 2149.048 2149.111 2149.038 2149.060 14.3
Δ(m/z) −0.203 −0.232 −0.213 −0.276 −0.203 −0.225 −104.8
[C84H142O71Na]+ 2310.976 2311.091 2311.097 2311.121 2311.145 2311.104 2311.112 9.4
Δ(m/z) −0.115 −0.121 −0.145 −0.169 −0.128 −0.136 −58.7
[C90H152O76Na]+ 2473.117 2473.152 2473.208 2473.152 2473.136 2473.149 2473.159 11.3
Δ(m/z) −0.035 −0.091 −0.035 −0.019 −0.032 −0.043 −17.3
[C96H162O81Na]+ 2635.258 2635.218 2635.234 2635.255 2635.257 2635.294 2635.252 10.9
Δ(m/z) 0.040 0.024 0.003 0.001 −0.036 0.006 2.2
[C102H172O86Na]+ 2797.398 2797.244 2797.214 2797.257 2797.185 2797.284 2797.237 13.7
Δ(m/z) 0.154 0.184 0.141 0.213 0.114 0.162 57.7
[C108H182O91Na]+ 2959.539 2959.498 2959.228 2959.255 2959.273 2959.444 2959.340 41.4
Δ(m/z) 0.041 0.311 0.284 0.266 0.095 0.200 67.4
[C114H192O96Na]+ 3121.680 3121.935 3121.293 3121.932 3121.334 3121.538 3121.606 100.2
Δ(m/z) −0.255 0.387 −0.252 0.346 0.142 0.074 23.5

IM-Q-TOF-MS and FT-ICR-MS

The Bruker timsTOF flex ion mobility-quadrupole-time-of-flight (IM-Q-TOF) mass spectrometer and the Bruker Apex-Qe Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometer not only shared the capability to deliver more than sufficient resolving power (R ≥ 40.000) and to deliver long-term stability of their mass calibration, they even shared the same mass reference file. As mentioned above (Table 1), isotopolog ions yield the most intensive signal within isotopic patterns of larger calibrant ions, and thus, the common mass reference list for these instruments was based on the respective ions. In particular with the IM-Q-TOF instrument the peak distribution of positive-ion spectra of a mixture composed of PVP-067 and PVP-2.1k as well as of each of the P2VPs alone exhibited essentially the same appearance as observed in reflector TOF-MS (Fig. S13). Either instrument was able to automatically pick the correct reference peak across the entire range and to perform mass calibration (Fig. S14).

As in linear TOF-MS, gummy bears were used as a source of oligosaccharides and [M + Na]+ ions appeared as prevalent species in the spectra. The signals observed ranged from the disaccharide sucrose, [C12H22O11Na]+, m/z 365.1054 (calc.) to the 17mer oligosaccharide [C102H172O86Na]+, m/z 2795.8978 (calc.). The IM-Q-TOF instrument delivered data with clearly better than 2 ppm mass accuracy (Table 5). Additionally, spectra of [peptide + H]+ ions as delivered by the Bruker Standard Peptide Mix II were acquired from CHCA matrix. Again, all ions from the [(CHCA)2 + H]+ matrix cluster ion, m/z 379.0924 (calc.) to the largest peptide of the mixture, somatostatin 28, m/z 3147.4710 (calc.) were well within 2 ppm error (Table 6). As examples, one spectrum of each series as obtained using the Bruker timsTOF flex instrument is depicted in Figure 5.

Table 5.

Mass accuracy by IM-Q-TOF-MS based on external calibration. Data of oligosaccharide [M + Na]+ ions. Signals were observed from the disaccharide sucrose, [C12H22O11Na]+, m/z 365.1054 (calc.) to the 17mer oligosaccharide [C102H172O86Na]+, m/z 2795.8978 (calc.). Data of five repetitions, standard deviation of m/z in ppm, average error Δ(m/z) in mu.

Formula Calc. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Avg. Exp. m/z Std. Dev.
[C12H22O11Na]+ 365.1054 365.1048 365.1047 365.1048 365.1048 365.1050 365.1048 0.3
Δ(m/z) 0.0006 0.0007 0.0006 0.0006 0.0004 0.0006 1.7
[C18H32O16Na]+ 527.1583 527.1581 527.1580 527.1581 527.1581 527.1583 527.1581 0.2
Δ(m/z) 0.0002 0.0003 0.0002 0.0002 0.0000 0.0001 0.3
[C24H42O21Na]+ 689.2111 689.2113 689.2111 689.2113 689.2112 689.2115 689.2113 0.2
Δ(m/z) −0.0002 0.0000 −0.0002 −0.0001 −0.0004 −0.0002 −0.3
[C30H52O26Na]+ 851.2639 851.2640 851.2639 851.2640 851.2640 851.2643 851.2640 0.2
Δ(m/z) −0.0001 0.0000 −0.0001 −0.0001 −0.0004 −0.0001 −0.2
[C36H62O31Na]+ 1013.3167 1013.3168 1013.3168 1013.3168 1013.3169 1013.3174 1013.3169 0.3
Δ(m/z) −0.0001 −0.0001 −0.0001 −0.0002 −0.0007 −0.0002 −0.2
[C42H72O36Na]+ 1175.3695 1175.3697 1175.3695 1175.3697 1175.3697 1175.3702 1175.3698 0.2
Δ(m/z) −0.0002 0.0000 −0.0002 −0.0002 −0.0007 −0.0002 −0.2
[C48H82O41Na]+ 1337.4224 1337.4224 1337.4222 1337.4224 1337.4225 1337.4231 1337.4225 0.3
Δ(m/z) 0.0000 0.0002 0.0000 −0.0001 −0.0007 −0.0002 −0.1
[C54H92O46Na]+ 1499.4752 1499.4751 1499.4750 1499.4752 1499.4754 1499.4758 1499.4753 0.2
Δ(m/z) 0.0001 0.0002 0.0000 −0.0002 −0.0006 −0.0001 −0.1
[C60H102O51Na]+ 1661.5280 1661.5276 1661.5276 1661.5278 1661.5282 1661.5283 1661.5279 0.2
Δ(m/z) 0.0004 0.0004 0.0002 −0.0002 −0.0003 0.0001 0.1
[C66H112O56Na]+ 1823.5808 1823.5803 1823.5802 1823.5807 1823.5805 1823.5815 1823.5806 0.3
Δ(m/z) 0.0005 0.0006 0.0001 0.0003 −0.0007 0.0002 0.1
[C72H122O61Na]+ 1985.6337 1985.6329 1985.6331 1985.6319 1985.6329 1985.6343 1985.6330 0.4
Δ(m/z) 0.0008 0.0006 0.0018 0.0008 −0.0006 0.0006 0.3
[C78H132O66Na]+ 2147.6865 2147.6861 2147.6853 2147.6848 2147.6858 2147.6864 2147.6857 0.3
Δ(m/z) 0.0004 0.0012 0.0017 0.0007 0.0001 0.0008 0.4
[C84H142O71Na]+ 2309.7393 2309.7391 2309.7380 2309.7377 2309.7374 2309.7426 2309.7390 0.9
Δ(m/z) 0.0002 0.0013 0.0016 0.0019 −0.0033 0.0003 0.2
[C90H152O76Na]+ 2471.7921 2471.7888 2471.7901 2471.7883 2471.7901 2471.7919 2471.7898 0.6
Δ(m/z) 0.0033 0.0020 0.0038 0.0020 0.0002 0.0023 0.9
[C96H162O81Na]+ 2633.8450 2633.8435 2633.8428 2633.8436 2633.8455 2633.8425 2633.8436 0.4
Δ(m/z) 0.0015 0.0022 0.0014 −0.0005 0.0025 0.0014 0.5
[C102H172O86Na]+ 2795.8978 2795.8964 2795.8942 2795.8935 2795.8982 2795.9014 2795.8967 1.1
Δ(m/z) 0.0014 0.0036 0.0043 −0.0004 −0.0036 0.0010 0.4

Table 6.

Mass accuracy by IM-Q-TOF-MS based on external calibration. Data of [peptide + H]+ ions of the Bruker Standard Peptide Mix II in CHCA. Data of five repetitions, standard deviation of m/z in ppm, average error Δ(m/z) in mu.

Formula Calc. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Exp. m/z Avg. Exp. m/z Std. Dev.
[(C10H7NO3)2 + H]+ 379.0924 379.0915 379.0917 379.0918 379.0914 379.0917 379.0916 0.4
Δ(m/z) 0.0009 0.0007 0.0006 0.0010 0.0007 0.0008 2.1
Bradykinin 1–7 757.3992 757.3986 757.3989 757.3992 757.3978 757.3988 757.3987 0.7
Δ(m/z) 0.0006 0.0003 0.0000 0.0014 0.0004 0.0005 0.7
Angiotensin II 1046.5418 1046.5411 1046.5417 1046.5420 1046.5396 1046.5415 1046.5412 0.9
Δ(m/z) 0.0007 0.0001 −0.0002 0.0022 0.0003 0.0006 0.6
Angiotensin I 1296.6848 1296.6838 1296.6843 1296.6852 1296.6832 1296.6849 1296.6843 0.6
Δ(m/z) 0.0010 0.0005 −0.0004 0.0016 −0.0001 0.0005 0.4
Substance P 1347.7354 1347.7343 1347.7347 1347.7353 1347.7353 1347.7348 1347.7349 0.3
Δ(m/z) 0.0011 0.0007 0.0001 0.0001 0.0006 0.0005 0.4
Bombesin 1619.8223 1619.8209 1619.8214 1619.8219 1619.8203 1619.8213 1619.8212 0.4
Δ(m/z) 0.0014 0.0009 0.0004 0.0020 0.0010 0.0011 0.7
ACTH clip 1–17 2093.0862 2093.0832 2093.0844 2093.0857 2093.0832 2093.0852 2093.0843 0.5
Δ(m/z) 0.0030 0.0018 0.0005 0.0030 0.0010 0.0019 0.9
ACTH clip 18–39 2465.1983 2465.1954 2465.1964 2465.1977 2465.1948 2465.1969 2465.1962 0.5
Δ(m/z) 0.0029 0.0019 0.0006 0.0035 0.0014 0.0021 0.8
Somatostatin 28 3147.4710 3147.4662 3147.4672 3147.4695 3147.4656 3147.4683 3147.4674 0.5
Δ(m/z) 0.0048 0.0038 0.0015 0.0054 0.0027 0.0036 1.2

Figure 5.

Figure 5.

Positive-ion MALDI spectra of (a) Bruker Peptide Mix II in CHCA matrix and (b) oligosaccharides from gummy bears in DHB matrix as obtained using the Bruker timsTOF flex instrument. The spectra are part of the dataset used to compile Tables 5 and 6. In (a), the inserts show expanded views of the [M + H]+ ions of bombesin, m/z 1619.8223 (calc.) and of ACTH clip 1–17, m/z 2093.0862 (calc.). In (b) the inserts show expanded views of the m/z range above 2000 and of the [C78H132O66Na]+ ion, m/z 2147.6865 (calc.). Labels of ions beyond m/z 2000 may refer to the first isotopic peak.

Analogous results were obtained using the FT-ICR instrument (Table S3 and Fig. S15) the main difference being that the 14-year old instrument started to show its age. Thus, with the given instrument, MALDI-FT-ICR spectra were somewhat inferior to the IM-Q-TOF spectra, mostly in terms of signal-to-noise ratio, and as a result, also in mass accuracy. Nonetheless, either instrument worked flawlessly with the P2VP reference file.

Conclusions

P2PVs are demonstrated as versatile reference compounds for external and internal mass calibration due to the formation of clean [M + H]+ ion series. P2VPs are compatible with CHCA, DHB, dithranol, and DCTB matrix and a wide range of analyte polarities ranging from ionic to nonpolar. In our laboratory DCTB proved most useful due its wide range of analyte acceptance and low laser fluence requirements.

P2VP-based mass calibration can be applied in MALDI-MS with a linear TOF analyzer, however, the mid-range drop in mass accuracy renders it inferior to calibration compounds delivering monoisotopic ions.

The mass accuracy based on external calibration of the reflector MALDI-TOF instrument was in the range of 3–10 ppm, and thus, generally not sufficient for reliable formula determination based on accurate mass. However, using P2VPs for internal mass calibration allowed to routinely achieve mass accuracies in the 2–5 ppm range for analytes of molecular mass between 300 u and 1500 u when using the reflector MALDI-TOF instrument with pulsed ion extraction. Notably higher levels of mass accuracy, i.e., 1–2 ppm, were achieved when higher-resolving analyzers like a modern IM-Q-TOF analyzer or an FT-ICR instrument were being used. Remarkably, IM-Q-TOF analyzer and FT-ICR instrument were able to provide this 1–2 ppm mass accuracy with external calibration all day long.

The reference data (as MS Excel sheet, as *.mcl files for linear and reflector mode TOF-MS in Bruker FlexControl 3.4 format, and as *.ref files for IM-Q-TOF Bruker timsControl and FT-ICR-MS for Bruker ApexControl, respectively) can be obtained from the author upon request

Supplemental Material

sj-pdf-1-ems-10.1177_14690667211055701 - Supplemental material for Poly(2-vinylpyridine) as a reference compound for mass calibration in positive-ion matrix-assisted laser desorption/ionization-mass spectrometry on different instrumental platforms

Supplemental material, sj-pdf-1-ems-10.1177_14690667211055701 for Poly(2-vinylpyridine) as a reference compound for mass calibration in positive-ion matrix-assisted laser desorption/ionization-mass spectrometry on different instrumental platforms by Jürgen H Gross in European Journal of Mass Spectrometry

Acknowledgements

The author thanks Doris Lang (Heidelberg University, Heidelberg, Germany) for testing the calibration in routine MALDI operation. He is grateful to the Deutsche Forschungsgemeinschaft (DFG) for granting the Bruker timsTOF fleX instrument (INST 35/1640-1 FUGG) to his institution.

Footnotes

Declaration of conflicting interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: Open access funding was enabled and organized by Projekt DEAL.

ORCID iD: Jürgen H Gross https://orcid.org/0000-0003-0748-2535

Supplemental material: Supplemental material for this article is available online.

References

  • 1.Hillenkamp F, Peter-Katalinic J. MALDI MS: a practical guide to instrumentation, methods and applications. 2nd ed. Weinheim: Wiley-VCH, 2013. [Google Scholar]
  • 2.Balogh MP. Debating resolution and mass accuracy in mass spectrometry. Spectroscopy 2004; 19: 34–38, 40. [Google Scholar]
  • 3.Busch KL. Masses in mass spectrometry: perfluors and more. Part II. Spectroscopy 2005; 20: 76–81. [Google Scholar]
  • 4.Bristow AWT. Accurate mass measurement for the determination of elemental formula – a tutorial. Mass Spectrom Rev 2006; 25: 99–111. [DOI] [PubMed] [Google Scholar]
  • 5.Leslie AD, Volmer DA. Dealing with the masses: a tutorial on accurate masses, mass uncertainties, and mass defects. Spectroscopy 2007; 22: 34–32, 39. [Google Scholar]
  • 6.Kim S, Rodgers RP, Marshall AG. Truly “exact” mass: elemental composition can be determined uniquely from molecular mass measurement at approximately 0.1 mDa accuracy for molecules up to approximately 500 Da. Int J Mass Spectrom 2006; 251: 260–265. [Google Scholar]
  • 7.Spengler B. De novo sequencing, peptide composition analysis, and composition-based sequencing: a new strategy employing accurate mass determination by Fourier transform ion cyclotron resonance mass spectrometry. J Am Soc Mass Spectrom 2004; 15: 703–714. [DOI] [PubMed] [Google Scholar]
  • 8.Data Requirements, https://publish.acs.org/publish/author_guidelines?coden=joceah#data_requirements (accessed Apr. 20, 2021).
  • 9.Angewandte Chemie, Notice to Authors, https://onlinelibrary.wiley.com/page/journal/15213773/homepage/notice-to-authors (accessed April 20, 2021).
  • 10.Guide to the presentation of experimental data, https://www.rsc.org/journals-books-databases/journal-authors-reviewers/prepare-your-article/experimental-data/#data-presentation (accessed April 20, 2021).
  • 11.Gross JH. Basic poly(propylene glycols) as reference compounds for internal mass calibration in positive-ion matrix-assisted laser desorption/ionization mass spectrometry. Eur J Mass Spectrom 2017; 23: 411–416. [DOI] [PubMed] [Google Scholar]
  • 12.Soga T, Kakazu Y, Robert M, et al. Qualitative and quantitative analysis of amino acids by capillary electrophoresis-electrospray ionization-tandem mass spectrometry. Electrophoresis 2004; 25: 1964–1972. [DOI] [PubMed] [Google Scholar]
  • 13.Schug K, Lindner W. Development of a screening technique for noncovalent complex formation between guanidinium- and phosphonate-functionalized amino acids by electrospray ionization ion trap mass spectrometry: assessing ionization and functional group interaction. Int J Mass Spectrom 2004; 235: 213–222. [Google Scholar]
  • 14.Gross JH. High-mass cluster ions of ionic liquids in positive-ion and negative-ion DART-MS and their application for wide range mass calibrations. Anal Bioanal Chem 2014; 406: 2853–2862. [DOI] [PubMed] [Google Scholar]
  • 15.Gross JH. High-mass capabilities of positive-ion and negative-ion direct analysis in real time mass spectrometry. Eur J Mass Spectrom 2016; 22: 43–48. [DOI] [PubMed] [Google Scholar]
  • 16.Gross JH. Saccharose cluster ions as mass calibrants in positive-ion direct analysis in real time-mass spectrometry. Eur J Mass Spectrom 2020; 26: 324–331. [DOI] [PubMed] [Google Scholar]
  • 17.Lou X, van Dongen JLJ, Meijer EW. Generation of CsI cluster ions for mass calibration in matrix-assisted laser desorption/ionization mass spectrometry. J Am Soc Mass Spectrom 2010; 21: 1223–1226. [DOI] [PubMed] [Google Scholar]
  • 18.Mochizuki S. Effective methods for the measurement of CsI cluster ions using MALDI-MS with suitable solvent combinations and additives. J Mass Spectrom 2014; 49: 1199–1202. [DOI] [PubMed] [Google Scholar]
  • 19.Mochizuki S. Enhanced measurement of CsI cluster ions for mass calibration in MALDI-MS using sugar alcohols. Anal Methods 2015; 7: 2215–2218. [Google Scholar]
  • 20.Moini M. Ultramark 1621 as a calibration/reference compound for mass spectrometry. II. Positive- and negative-ion electrospray ionization. Rapid Commun Mass Spectrom 1621; 1994; 8: 711–714. [Google Scholar]
  • 21.Flanagan JM. Mass spectrometry calibration using homogeneously substituted fluorinated triazatriphosphorines. Patent 5872357, US, 1997.
  • 22.Wu J, McAllister H. Exact mass measurement on an electrospray ionization time-of-flight mass spectrometer: error distribution and selective averaging. J Mass Spectrom 2003; 38: 1043–1053. [DOI] [PubMed] [Google Scholar]
  • 23.Gross JH. Wide range mass calibration for negative-ion direct analysis in real time-mass spectrometry. Eur J Mass Spectrom 2014; 20: 155–161. [DOI] [PubMed] [Google Scholar]
  • 24.de Koster CG, Duursma MC, van Rooij GJ, et al. Endgroup analysis of polyethylene glycol polymers by matrix-assisted laser desorption/ionization Fourier-transform ion cyclotron resonance mass spectrometry. Rapid Commun Mass Spectrom 1995; 9: 957–962. [DOI] [PubMed] [Google Scholar]
  • 25.Montaudo G, Montaudo MS, Puglisi C, et al. Characterization of polymers by matrix-assisted Laser desorption ionization-time of flight mass spectrometry. End group determination and molecular weight estimates in poly(ethylene glycols). Macromolecules 1995; 28: 4562–4569. [Google Scholar]
  • 26.Grayson SM, Myers BK, Bengtsson J, et al. Advantages of monodisperse and chemically robust “SpheriCal” polyester dendrimers as a “universal” MS calibrant. J Am Soc Mass Spectrom 2014; 25: 303–309. [DOI] [PubMed] [Google Scholar]
  • 27.Casey BK, Grayson SM. Letter: the potential of amine-containing dendrimer mass standards for internal calibration of peptides. Eur J Mass Spectrom 2015; 21: 747–752. [DOI] [PubMed] [Google Scholar]
  • 28.Gross JH. Improved procedure for dendrimer-based mass calibration in matrix-assisted laser desorption/ionization-time-of-flight-mass spectrometry. Anal Bioanal Chem 2016; 408: 5945–5951. [DOI] [PubMed] [Google Scholar]
  • 29.Giesen JA, Diament BJ, Grayson SM. Iodine-containing mass-defect-tuned dendrimers for use as internal mass spectrometry calibrants. J Am Soc Mass Spectrom 2018; 29: 490–500. [DOI] [PubMed] [Google Scholar]
  • 30.Gruendling T, Sauerland V, Barahona C, et al. Polyalanine – a practical polypeptide mass calibration standard for matrix-assisted laser desorption/ionization mass spectrometry and tandem mass spectrometry in positive and negative mode. Rapid Commun Mass Spectrom 2016; 30: 681–683. [DOI] [PubMed] [Google Scholar]
  • 31.Ulmer L, Mattay J, Torres-Garcia HG, et al. The use of 2-[(2E)-3-(4-tert-butylphenyl)-2-methylprop-2-enylidene]malononitrile as a matrix for matrix-assisted laser desorption/ionization mass spectrometry. Eur J Mass Spectrom 2000; 6: 49–52. [Google Scholar]
  • 32.Lim H, Lee Y, Han S, et al. Investigation of polystyrene, polyisoprene, and poly(2-vinylpyridine) using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Bull Korean Chem Soc 1999; 20: 853–856. [Google Scholar]
  • 33.Corten C, Kretschmer K, Kuckling D. Novel multi-responsive P2VP-block-PNIPAAm block copolymers via nitroxide-mediated radical polymerization. Beilstein J Org Chem 2010; 6: 756–765. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Chendo C, Phan TNT, Marque SRA, et al. Mass spectrometry of nitroxide-terminated poly(4-vinylpyridine): a case of unwanted reactive MALDI. Int J Mass Spectrom 2016; 405: 50–58. [Google Scholar]
  • 35.Gross JH. Polydimethylsiloxane-Based wide range mass calibration for direct analysis in real time mass spectrometry. Anal Bioanal Chem 2013; 405: 8663–8668. [DOI] [PubMed] [Google Scholar]
  • 36.Gross JH. Analysis of silicones released from household items and baby articles by direct analysis in real time-mass spectrometry. J Am Soc Mass Spectrom 2015; 26: 511–521. [DOI] [PubMed] [Google Scholar]
  • 37.Mohr MD, Börnsen KO, Widmer HM. Matrix-assisted laser desorption/ionization mass spectrometry: improved matrix for oligosaccharides. Rapid Commun Mass Spectrom 1995; 9: 809–814. [DOI] [PubMed] [Google Scholar]
  • 38.Gross JH. Mass spectrometry – A textbook. 3rd ed. Cham: Springer International Publishing, 2017, p. 686–688. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

sj-pdf-1-ems-10.1177_14690667211055701 - Supplemental material for Poly(2-vinylpyridine) as a reference compound for mass calibration in positive-ion matrix-assisted laser desorption/ionization-mass spectrometry on different instrumental platforms

Supplemental material, sj-pdf-1-ems-10.1177_14690667211055701 for Poly(2-vinylpyridine) as a reference compound for mass calibration in positive-ion matrix-assisted laser desorption/ionization-mass spectrometry on different instrumental platforms by Jürgen H Gross in European Journal of Mass Spectrometry


Articles from European Journal of Mass Spectrometry (Chichester, England) are provided here courtesy of SAGE Publications

RESOURCES