No Clinical or Radiographic Differences Between Cemented Cobalt–Chromium and Titanium–Niobium Nitride Mobile-Bearing Unicompartmental Knee Arthroplasty

Riccardo D’Ambrosi 1,, Rafael Loucas 2, Marios Loucas 2, Riccardo Giorgino 3, Nicola Ursino 1, Giuseppe Maria Peretti 1,3
PMCID: PMC8586226  PMID: 34824720

Abstract

Purpose

This study aimed to compare the clinical and radiographic outcomes of patients with positive patch tests undergoing a medial mobile-bearing titanium–niobium nitride (TiNbN) unicompartmental knee arthroplasty (UKA) to patients undergoing standard UKA (cobalt–chromium [CoCr] implants).

Methods

Two successive groups of patients, amounting to a total of 246 individuals, who received Oxford (Zimmer-Biomet, Warsaw, Indiana, USA) UKA were included. The first group was composed of a series of 203 consecutive standard CoCr UKAs (Standard Group), while the second group comprised 43 consecutive hypoallergenic TiNbN UKAs (HA group). The patients of the second group had a positive epicutaneous patch test result for metals. Each patient was evaluated using the Oxford Knee Score (OKS) and Knee Society Score (KSS) a day prior to the surgery (T0) and at two consecutive follow-ups, namely T1 (minimum follow-up of 12 months) and T2 (minimum follow-up of 34 months). Radiographic measurements were performed at the final follow-up (T2).

Results

No statistical differences were noted between the two groups regarding demographic data (p > 0.05). No clinical or radiographic differences were found between the HA and standard groups at any follow-up (p > 0.05). A statistically significant improvement was found at any follow-up for both OKS and KSS (p < 0.05).

Conclusions

No clinical or radiographic differences between the hypoallergenic and standard cobalt–chromium groups at any follow-up were found, with a clinically significant improvement being experienced by both groups during the entire follow-up.

Level of evidence

Level II—comparative prospective study.

Supplementary Information

The online version contains supplementary material available at 10.1007/s43465-021-00486-3.

Keywords: Oxford mobile-bearing unicompartmental knee arthroplasty, Metal allergy, Knee osteoarthritis

Introduction

Medial unicompartmental knee arthroplasty (UKA) is a safe and reliable option for the treatment of symptomatic end-stage anteromedial osteoarthritis of the knee. Compared to total knee arthroplasty (TKA), medial UKA tends to offer superior functional outcomes, a better range of movement and more physiological knee kinematics [1, 2]. Moreover, patients have a lower rate of morbidity and mortality [1]. Despite the high success rate of knee prosthetic surgery [1, 2], a considerable number of failures, ranging from 10 to 20%, still requires revision, along with the main causes for UKAs which include biomechanical failure and infection [38]. In recent years, it has been observed that metal hypersensitivity (MHS) may be another cause of failure after UKA [9].

MHS after arthroplasty is an uncommon condition in which the body immunologically reacts to the metals used in UKA implants [911]. Skin hypersensitivity to metals, such as nickel, cobalt and chromium, in the general population has been estimated to range from 1 to 13% [1215].

Since UKA is made of several metals, patients could develop a post-operative allergic reaction to these implants, especially those with allergies. Patients with MHS might present pain and effusions, as well as skin abnormalities [1618] on the operated joint; however, MHS is still a diagnosis of exclusion since current studies report an absence of accuracy in the methods mentioned above [10, 1315].

Although MHS is a rare condition, the number of UKA patients who test positive for MHS has increased over the last 20 years. The percentage has been estimated up to 25%, which is in contrast with the percentage in the general population that ranges from 10 to 15% [11, 1921]. Furthermore, in patients with painful arthroplasties, the risk of MHS may be as high as 60% [4]. However, the relationship between MHS and UKA is currently unclear [11, 20, 22, 23].

Due to an increase in the use of UKA, different prostheses with hypoallergenic features should be developed. A substitute to the standard cobalt–chromium (CoCr), titanium–niobium nitride (TiNbN), has been developed and used in recent years [23, 24]. TiNbN has a high surface resistance to abrasion and corrosion and is composed of a fine ceramic coating. These features result in possible long-term improvements in several biological processes such as biocompatibility and osteointegration [23, 24]. However, there are no clinical studies that compare the clinical and radiographic outcomes of patients who underwent hypoallergenic TiNbN to those who underwent standard UKA.

Therefore, the main aim of this study was to compare the clinical and radiographic outcomes of patients with proven metal allergy who undergo hypoallergenic TiNbN UKA and to compare these results to those of a group of patients with standard UKAs (regular cobalt–chromium implants). It was hypothesised that TiNbN implants lead to excellent outcomes comparable to standard UKAs.

Materials and Methods

The study was performed according to the STROBE checklist for group studies [25].

All patients signed an informed consent document before being enrolled in the study. Ethical approval was obtained from the local ethical committee.

Two cohorts of patients who had undergone cemented medial mobile-bearing UKA between January 2015 and December 2017 were included in this study. The first group involved 203 consecutive UKA cobalt–chromium medial mobile-bearing with the Oxford (Zimmer-Biomet, Warsaw, IN, USA) Microplasty instrumentation (Standard Group). The second group was composed of 43 consecutive UKA TiNbN (HA group), with a single femoral peg, using the Microplasty instrumentation in patients with proven positive patch test for metals.

Patients were carefully selected, and the standardised treatment algorithm was meticulously performed in order to decide who should undergo the patch test as previously published [11]. Adverse reactions to metals were recorded according to the International Contact Dermatitis Research Group (ICDRG) criteria [26].

A standardised epicutaneous patch test method was performed to identify allergic contact sensitisation in patients with suspected metal allergy due to the high reliability [27].

A total of 43 patients resulted positive to the metal patch test and were included in this study.

All surgeries were performed by one of the authors (NU) who has considerable experience in knee arthroplasty [28].

In both groups, the indication for surgery was the isolated anteromedial OA or avascular necrosis (AVN) of the medial femoral condyle. Pre-operative clinical examination and magnetic resonance imaging (MRI) ensured ligament integrity. For radiographic inclusion, the Oxford Group criteria were followed [29]. The main exclusion criteria for participation in this study were missing data or radiographs that were unsuitable for obtaining an exact measurement, revision surgery or previous surgery of the affected knee (except arthroscopy for meniscectomy).

Clinical Evaluation

All patients were evaluated by two independent surgeons who were not involved in the index surgery. The clinical evaluation included the examination of each patient’s OKS and KSS. Each patient was evaluated on the day before surgery (T0) and at two consecutive follow-ups, namely T1 (minimum follow-up of 12 months) and T2 (minimum follow-up of 34 months) [3032].

Radiographic Evaluation

The UKA’s position was assessed only at the final follow-up (T2) according to the Oxford Partial Knee Surgical Technique operating manual.

In accordance with the manufacturer’s manual, the following parameters were evaluated [29, 33]:

  • Femoral component varus/valgus. The angle between the femoral component and the femoral axis in the coronal plane.

  • Tibial component varus/valgus. The angle between the tibial axis and a line drawn along the tibial tray in the coronal plane.

  • Anteroposterior slope. The angle between a line drawn along the tibial tray and perpendicular to the tibial axis in the lateral view.

Statistical Analysis

A sample of 240 subjects, including 40 hypoallergenic and 200 standard patients, was considered sufficient to observe a five-point group difference in both KSS and OKS score. The assumptions included a standard deviation of 10 for both groups, a 5% two-tailed alpha, an 80% power and the employment of a Wilcoxon–Mann–Whitney test. We also assumed a group disproportion of 5% due to hypersensitivity to metals. Three additional subjects were recruited in both groups to ensure statistical significance in case of unexpected events.

This sample was also adequate to detect a difference of two points at subsequent time points in both KSS and OKS score in non-allergic patients and a five-point difference in both KSS and OKS score in allergic patients, assuming a standard deviation (SD) of 10, a 5% two-tailed alpha, an 80% power and the employment of a Wilcoxon signed-rank test as above [30, 31].

The normal distribution of variables was tested with the Shapiro–Wilk test and pre- and postoperative scores compared to the paired t test (parametric data) or the Wilcoxon-rank sum test (non-parametric distribution). The two cohorts (hypoallergenic and standard groups) were compared by employing an independent-samples t test for continuous variables and chi-square analyses for categorical variables to report differences between the two cohorts. A difference of p < 0.05 was statistically significant. To identify interacting factors and correlations, multivariate analyses (Spearman and Pearson/Point biserial analysis) were utilised. All statistical analyses were performed using an IBM SPSS® statistics software (Version 25.0; Chicago, IL, USA).

Results

A total of 281 patients completed the study, out of which 246 patients satisfied all inclusion and exclusion criteria (43 in the HA group and 203 in the Standard group). Out of the 35 patients excluded, 28 reported previous knee surgery (25 anterior cruciate ligament reconstructions, 1 multi-ligaments reconstruction and 2 tibial osteotomies), while seven patients had radiographic data, which was unfavourable to achieve an exact measurement.

In the HA group, all patients completed the follow-up, while in the standard group, all patients completed T1 but only 200 patients could complete the entire follow-up due to failure of the arthroplasty. In all the three failed cases, a mobilisation of the tibial component was reported. No statistical differences were found among the groups regarding demographic data (p > 0.05). Table 1 shows the detailed results.

Table 1.

Demographic data for the study groups

Hypoallergenic
Mean ± std (min, max)
Standard
Mean ± std (min, max)
p value Test
Age at surgery 67.43 ± 6.35 (54; 78), n = 43 70.13 ± 8.35 (44; 89), n = 203 0.122 MWU
First follow-up time (T1) 12.94 ± 0.94 (12; 15), n = 43 13.30 ± 1.29 (12; 16), n = 203 0. 241 MWU
Last follow-up time (T2) 69.79 ± 17.49 (34; 95), n = 43 67.13 ± 16.20 (34; 95), n = 200 0.994 MWU
Gender

26 female

17 male

122 female

81 male

0.964 Fisher
Side

24 right

19 left

105 right

98 left

0.626 Fisher
Weight 72.11 ± 9.40 (54; 95), n = 43 75.50 ± 13.88 (50; 116), n = 203 0.001 MWU
Height 161.30 ± 10.04 (155; 170), n = 43 165.22 ± 9.87 (145; 188), n = 203 0.001 MWU
Body mass index 27.70 ± 3.35 (22; 33), n = 43 27.66 ± 4.23 (19; 39), n = 203 0.969 MWU

MWU Mann–Whitney U test, T1 first follow-up, T2 last follow-up

Comparison Between the HA Group and Standard Group

No clinical or radiographic variances were seen between the HA and standard groups at any follow-up (p > 0.05). Table 2 shows the detailed results.

Table 2.

Diagnosis and pre- and postoperative scores of the 246 patients available for personal

Hypoallergenic Standard p value
Oxford Knee Score T0

23.06 ± 2.70 (18; 28)

n = 43

22.89 ± 2.64 (16; 28)

n = 203

0.977
Oxford Knee Score T1

42.06 ± 1.62 (39; 47)a

n = 43

42.09 ± 2.87 (20; 48)a

n = 203

0.897
Oxford Knee Score T2

45.04 ± 1.76 (40; 48)a,b

n = 43

44.64 ± 1.93 (40; 48)a,b

n = 200

0.929
KSS T0

51.25 ± 4.85 (44; 62)

n = 43

50.80 ± 4.61 (44; 62)

n = 203

0.997
KSS T1

87.65 ± 2.03 (82; 91)a

n = 43

86.76 ± 5.46 (44; 91)a

n = 203

0.995
KSS T2

91.95 ± 3.44 (88; 100)a,b

n = 43

91.39 ± 6.84 (9; 100)a,b

n = 200

0.814
Tibial angle T2

3.06 ± 2.05 (0; 8)

n = 43

2.84 ± 2.46 (− 6; 9)

n = 200

0.948
Femoral angle at T2

7.18 ± 4.56 (− 2; 16)

n = 43

7.06 ± 5.13 (− 6; 20)

n = 200

0.529
Tibial slope at T2

4.97 ± 3.34 (− 3; 11)

n = 43

5.46 ± 3.13 (− 6; 16)

n = 200

0.064

Follow-up. Values are given as mean ± SD [95% confidence interval] or (minimum; maximum)

aStatistical significant difference versus t0

bStatistical significant difference versus T1

Hypoallergenic Group

A significant difference was found at T0, T1 and T2, as well as between T1 and T2 for both OKS and KSS (p < 0.05). Please refer to Table 2 for detailed results.

Standard Group

A significant difference was found at T0, T1 and T2, as well as between T1 and T2 for both OKS and KSS (p < 0.05). Detailed results are reported in Table 2.

Complications/Failure

In the standard group, three (1.47%) patients reported a revision with TKA due to the mobilisation of the tibial component.

Discussion

The main results of the present research illustrate that no clinical or radiographic variances were seen between hypoallergenic and standard CoCr Oxford UKAs at any time points. These results should be considered extremely encouraging, considering the increasing number of successful reports on UKA; in the current literature, Oxford UKA shows a 91% survival rate at 20 years [34].

The possibility of being able to guarantee similar results to patients with metal allergies further increases the interest in this surgery, and therefore, the hypoallergenic UKA should be considered as the ideal treatment in selected sensitive patients with anteromedial knee OA.

Our data significantly contribute to recent interest. The failure of implant continues to occur in a fair percentage of cases, and among them, hypersensitivity to the metal is increasingly becoming a factor. However, even today, the nature and causative mechanism of the relationship between MHS and prosthesis failure remain undefined.

Despite the evidence in the literature of significant incidences of both intolerance to metals and the need for replanting prostheses, our research supports the view that the relation between metal hypersensitivity and revision is presumably not clear and demonstrated.

In light of the significant data of Desai et al., it could be hypothesised that the materials contained within the implant were responsible for hypersensitive reactions [5]. Once the patients underwent the patch test, their bodies reacted to chromium, cobalt and nickel. These are precisely the materials used in our standard group, and the clinical and radiological results are shown to be the same in the HA group, which suggests that the reaction is probably related to other mechanisms and not specifically to a particular material.

Contact with substances that could give rise to hypersensitive reactions was also hypothesised for those who used the coating. The study of Beyer et al., which divided the patients into two cohorts and subjected them to TKA with the coated and uncoated implant variant, presented the same outcomes in both groups at four years’ follow-up: the OKS score increased in a comparable manner in both groups, which is in accordance with our study’s results [35].

The role of the coating of a prosthetic implant was also analysed by Thomas et al. who compared almost 200 patients undergoing knee arthroplasty to a 5-year follow-up, focusing not only on the clinical outcomes but also on the serum cytokine levels. Based on their analysis, proinflammatory cytokines, such as Interleukin-8 and Iinterleukin-10, had higher average values in patients with uncoated prostheses, which suggests their possible role in implant failure processes [36].

On the other hand, data differed at several times. It is well known that after excluding various causes of implant failure, with infection topping the list, metal hypersensitivity must always be taken into consideration as a reason for failure. The results of Thakur et al.’s study are confirmed by the fact that after revision with other materials, all patients experienced improvement [16]. Similar results were achieved by Zondervan et al.’s study in which patients with hypersensitivity experienced a significant benefit after a revision with a hypoallergenic component [37].

Hence, the questions arise: how should we proceed? Does preoperative testing make sense to identify the patients who might develop hypersensitivity reactions? Currently, there are no indications for preoperative tests to detect patients with metal hypersensitivity, while only the postoperative test is recommended to patients who develop certain clinical characteristics [38]. This finding seems to have already been encouraged by the study of Frigerio et al. which explained how preoperative tests may not be so useful in light of the discrepancy between the pre- and postoperative results [39].

Important clarifications were provided by Münch et al.’s study in which the patch test did not seem to be the key tool to reveal the patients who would then develop hypersensitivity, but an association between the number of revisions and metal allergy was found [40].

In this respect, the difference between allergy and hypersensitivity should imperatively be clarified given that these terms are often mistakenly used as synonyms. As suggested by Middleton et al., the diagnosis of allergy can be formulated by justifying several specific criteria, whereas the reported ‘hypersensitivity’ does not justify the use of specific components [41].

Although the results of the study of Walker et al. seem promising in terms of outcomes, we do not believe that an anamnestic questionnaire could represent the most suitable tool to select patients who deserve a particular component in the prosthetic implant [42]. In accordance with the topics covered by Saccomanno et al., we consider that the employment of hypersensitivity screening should be promoted only when a patient’s clinical history suggests that it is necessary [43]. Should such a condition be ascertained, it is advisable to think of specific ‘hypersensitivity-friendly’ implants.

Based on the results obtained from our work and the comparison with the literature, our summary recommendations are as follows: it is certainly important to be able to recognise the signs and symptoms of an implant failure and subsequently trace the ones that can direct us to a hypersensitivity reaction. Moreover, we believe that only patients with a history of metal allergies should be traced and treated properly with specific components in knee arthroplasty.

The present study presented several limitations. First, it analyses results for only 24 months after the surgery; however, clinical outcomes may change over this span of time. Besides, the HA group was relatively small although a power analysis was performed to ensure that the sample size was suitable.

Conclusions

The current study showed no clinical or radiographic differences between the hypoallergenic and standard cobalt–chromium groups at any follow-up, with a clinically significant improvement being experienced by both groups over the entire follow-up. In light of these findings, we believe that a specific hypoallergenic implant should be used for the treatment of anteromedial knee OA in selected patients who are clearly allergic to metals.

Supplementary Information

Below is the link to the electronic supplementary material.

Author Contributions

RD’A: writing, data analysis, patients evaluation, surgery, and final revision. RL: data analysis, writing, and statistical analysis. ML: data analysis, writing, and statistical analysis. RG: writing, data analysis, and revision. NU: surgery and final revision. GMP: final revision.

Funding

None.

Availability of data and materials

Raw data have been submitted as supplementary material to the Journal.

Declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

Permission for the study was obtained from the local ethical committee.

Consent to participate

Informed consent was obtained from all individual participants included in the study.

Consent to publish

All authors consent to the publication of the manuscript.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Stempin R, Kaczmarek W, Stempin K, Dutka J. Midterm results of cementless and cemented unicondylar knee arthroplasty with mobile meniscal bearing: A prospective cohort study. The Open Orthopaedics Journal. 2017;11:1173–1178. doi: 10.2174/1874325001711011173. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Kendrick BJ, Kaptein BL, Valstar ER, Gill HS, Jackson WF, Dodd CA, Price AJ, Murray DW. Cemented versus cementless Oxford unicompartmental knee arthroplasty using radiostereometric analysis: A randomised controlled trial. The Bone & Joint Journal. 2015;97-B(2):185–191. doi: 10.1302/0301-620X.97B2.34331. [DOI] [PubMed] [Google Scholar]
  • 3.Pacheco KA. Allergy to surgical implants. The Journal of Allergy and Clinical Immunology: In Practice. 2015;3(5):683–695. doi: 10.1016/j.jaip.2015.07.011. [DOI] [PubMed] [Google Scholar]
  • 4.Bao W, He Y, Fan Y, Liao Y. Metal allergy in total-joint arthroplasty: Case report and literature review. Medicine (Baltimore) 2018;97(38):e12475. doi: 10.1097/MD.0000000000012475. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Desai MM, Shah KA, Mohapatra A, Patel DC. Prevalence of metal hypersensitivity in total knee replacement. Journal of Orthopaedics. 2019;16(6):468–472. doi: 10.1016/j.jor.2019.05.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Bozic KJ, Kamath AF, Ong K, Lau E, Kurtz S, Chan V, Vail TP, Rubash H, Berry DJ. Comparative epidemiology of revision arthroplasty: Failed THA poses greater clinical and economic burdens than failed TKA. Clinical Orthopaedics and Related Research. 2015;473(6):2131–2138. doi: 10.1007/s11999-014-4078-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Vasso M, Corona K, D'Apolito R, Mazzitelli G, Panni AS. Unicompartmental knee arthroplasty: Modes of failure and conversion to total knee arthroplasty. Joints. 2017;5(1):44–50. doi: 10.1055/s-0037-1601414. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Mittal A, Meshram P, Kim WH, Kim TK. Unicompartmental knee arthroplasty, an enigma, and the ten enigmas of medial UKA. Journal of Orthopaedics and Traumatology. 2020;21(1):15. doi: 10.1186/s10195-020-00551-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Mitchelson AJ, Wilson CJ, Mihalko WM, Grupp TM, Manning BT, Dennis DA, Goodman SB, Tzeng TH, Vasdev S, Saleh KJ. Biomaterial hypersensitivity: Is it real? Supportive evidence and approach considerations for metal allergic patients following total knee arthroplasty. BioMed Research International. 2015;2015:137287. doi: 10.1155/2015/137287. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Lachiewicz PF, Watters TS, Jacobs JJ. Metal hypersensitivity and total knee arthroplasty. Journal of the American Academy of Orthopaedic Surgeons. 2016;24(2):106–112. doi: 10.5435/JAAOS-D-14-00290. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.D'Ambrosi R, Nuara A, Mariani I, Di Feo F, Ursino N, Hirschmann M. Titanium niobium nitride mobile-bearing unicompartmental knee arthroplasty results in good to excellent clinical and radiographic outcomes in metal allergy patients with medial knee osteoarthritis. Journal of Arthroplasty. 2021;36(1):140–147.e2. doi: 10.1016/j.arth.2020.07.028. [DOI] [PubMed] [Google Scholar]
  • 12.Schäfer T, Böhler E, Ruhdorfer S, Weigl L, Wessner D, Filipiak B, Wichmann HE, Ring J. Epidemiology of contact allergy in adults. Allergy. 2001;56(12):1192–1196. doi: 10.1034/j.1398-9995.2001.00086.x. [DOI] [PubMed] [Google Scholar]
  • 13.Faschingbauer M, Renner L, Boettner F. Allergy in total knee replacement. Does it exist?: Review article. HSS Journal. 2017;13(1):12–19. doi: 10.1007/s11420-016-9514-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Atanaskova Mesinkovska N, Tellez A, Molina L, Honari G, Sood A, Barsoum W, Taylor JS. The effect of patch testing on surgical practices and outcomes in orthopedic patients with metal implants. Archives of Dermatology. 2012;148(6):687–693. doi: 10.1001/archdermatol.2011.2561. [DOI] [PubMed] [Google Scholar]
  • 15.Eben R, Dietrich KA, Nerz C, Schneider S, Schuh A, Banke IJ, Mazoochian F, Thomas P. Kontaktallergieraten gegen Metalle und Knochenzementbestandteile bei Patienten mit Endoprothesenunverträglichkeit [Contact allergy to metals and bone cement components in patients with intolerance of arthroplasty] Deutsche Medizinische Wochenschrift. 2010;135(28–29):1418–1422. doi: 10.1055/s-0030-1262426. [DOI] [PubMed] [Google Scholar]
  • 16.Thakur RR, Ast MP, McGraw M, Bostrom MP, Rodriguez JA, Parks ML. Severe persistent synovitis after cobalt-chromium total knee arthroplasty requiring revision. Orthopedics. 2013;36(4):e520–524. doi: 10.3928/01477447-20130327-34. [DOI] [PubMed] [Google Scholar]
  • 17.Thomsen M, Rozak M, Thomas P. Pain in a chromium-allergic patient with total knee arthroplasty: Disappearance of symptoms after revision with a special surface-coated TKA—A case report. Acta Orthopaedica. 2011;82(3):386–388. doi: 10.3109/17453674.2011.579521. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Verma SB, Mody B, Gawkrodger DJ. Dermatitis on the knee following knee replacement: A minority of cases show contact allergy to chromate, cobalt or nickel but a causal association is unproven. Contact Dermatitis. 2006;54(4):228–229. doi: 10.1111/j.0105-1873.2006.0775o.x. [DOI] [PubMed] [Google Scholar]
  • 19.Ragone V, Canciani E, Biffi CA, D'Ambrosi R, Sanvito R, Dellavia C, Galliera E. CoCrMo alloys ions release behavior by TiNbN coating: An in vitro study. Biomedical Microdevices. 2019;21(3):61. doi: 10.1007/s10544-019-0417-6. [DOI] [PubMed] [Google Scholar]
  • 20.Granchi D, Cenni E, Giunti A, Baldini N. Metal hypersensitivity testing in patients undergoing joint replacement: A systematic review. Journal of Bone and Joint Surgery. British Volume. 2012;94(8):1126–1134. doi: 10.1302/0301-620X.94B8.28135. [DOI] [PubMed] [Google Scholar]
  • 21.Kitagawa A, Chin T, Tsumura N. Iguchi T (2013) Metal sensitivity in patients before and after total knee arthroplasty (TKA): Comparison between ceramic surfaced oxidized zirconium and cobalt-chromium implants. Hypersensitivity. 2013;1:1–3. doi: 10.7243/2052-594X-1-3. [DOI] [Google Scholar]
  • 22.Thienpont E. Titanium niobium nitride knee implants are not inferior to chrome cobalt components for primary total knee arthroplasty. Archives of Orthopaedic and Trauma Surgery. 2015;135(12):1749–1754. doi: 10.1007/s00402-015-2320-9. [DOI] [PubMed] [Google Scholar]
  • 23.Thienpont E, Berger Y. No allergic reaction after TKA in a chrome-cobalt-nickel-sensitive patient: Case report and review of the literature. Knee Surgery, Sports Traumatology, Arthroscopy. 2013;21(3):636–640. doi: 10.1007/s00167-012-2000-z. [DOI] [PubMed] [Google Scholar]
  • 24.van Hove RP, Brohet RM, van Royen BJ, Nolte PA. No clinical benefit of titanium nitride coating in cementless mobile-bearing total knee arthroplasty. Knee Surgery, Sports Traumatology, Arthroscopy. 2015;23(6):1833–1840. doi: 10.1007/s00167-014-3359-9. [DOI] [PubMed] [Google Scholar]
  • 25.von Elm E, Altman DG, Egger M, Pocock SJ, Gøtzsche PC, Vandenbroucke JP, STROBE Initiative The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: Guidelines for reporting observational studies. Journal of Clinical Epidemiology. 2008;61(4):344–349. doi: 10.1016/j.jclinepi.2007.11.008. [DOI] [PubMed] [Google Scholar]
  • 26.Fransway AF, Zug KA, Belsito DV, Deleo VA, Fowler JF, Jr, Maibach HI, Marks JG, Mathias CG, Pratt MD, Rietschel RL, Sasseville D, Storrs FJ, Taylor JS, Warshaw EM, Dekoven J, Zirwas M. North American Contact Dermatitis Group patch test results for 2007–2008. Dermatitis. 2013;24(1):10–21. doi: 10.1097/DER.0b013e318277ca50. [DOI] [PubMed] [Google Scholar]
  • 27.Dickel H, Geier J, Kreft B, Pfützner W, Kuss O. Comparing reliabilities of strip and conventional patch testing. Contact Dermatitis. 2017;76(6):342–349. doi: 10.1111/cod.12758. [DOI] [PubMed] [Google Scholar]
  • 28.Zhang Q, Zhang Q, Guo W, Liu Z, Cheng L, Yue D, Zhang N. The learning curve for minimally invasive Oxford phase 3 unicompartmental knee arthroplasty: Cumulative summation test for learning curve (LC-CUSUM) Journal of Orthopaedic Surgery and Research. 2014;9:81. doi: 10.1186/s13018-014-0081-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Hamilton TW, Pandit HG, Lombardi AV, Adams JB, Oosthuizen CR, Clavé A, Dodd CA, Berend KR, Murray DW. Radiological decision aid to determine suitability for medial unicompartmental knee arthroplasty: development and preliminary validation. The Bone & Joint Journal. 2016;98-B(10 Supple B):3–10. doi: 10.1302/0301-620X.98B10.BJJ-2016-0432.R1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Quah C, Holmes D, Khan T, Cockshott S, Lewis J, Stephen A. The variability in Oxford hip and knee scores in the preoperative period: Is there an ideal time to score? Annals of the Royal College of Surgeons of England. 2018;100(1):16–20. doi: 10.1308/rcsann.2017.0090. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Insall JN, Dorr LD, Scott RD, Scott WN. Rationale of the Knee Society clinical rating system. Clinical Orthopaedics and Related Research. 1989;248:13–14. [PubMed] [Google Scholar]
  • 32.Asif S, Choon DS. Midterm results of cemented Press Fit Condylar Sigma total knee arthroplasty system. Journal of Orthopaedic Surgery (Hong Kong) 2005;13(3):280–284. doi: 10.1177/230949900501300311. [DOI] [PubMed] [Google Scholar]
  • 33.Mujika KM, Méndez JAJ, de Miguel AF. Advantages and disadvantages in image processing with free software in radiology. Journal of Medical Systems. 2018;42(3):36. doi: 10.1007/s10916-017-0888-z. [DOI] [PubMed] [Google Scholar]
  • 34.Price AJ, Svard U. A second decade lifetable survival analysis of the Oxford unicompartmental knee arthroplasty. Clinical Orthopaedics and Related Research. 2011;469(1):174–179. doi: 10.1007/s11999-010-1506-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Beyer F, Lützner C, Kirschner S, Lützner J. Midterm results after coated and uncoated TKA: A randomized controlled study. Orthopedics. 2016;39(3 Suppl):S13–17. doi: 10.3928/01477447-20160509-10. [DOI] [PubMed] [Google Scholar]
  • 36.Thomas P, Hisgen P, Kiefer H, Schmerwitz U, Ottersbach A, Albrecht D, Summer B, Schinkel C. Blood cytokine pattern and clinical outcome in knee arthroplasty patients: Comparative analysis 5 years after standard versus "hypoallergenic" surface coated prosthesis implantation. Acta Orthopaedica. 2018;89(6):646–651. doi: 10.1080/17453674.2018.1518802. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Zondervan RL, Vaux JJ, Blackmer MJ, Brazier BG, Taunt CJ., Jr Improved outcomes in patients with positive metal sensitivity following revision total knee arthroplasty. Journal of Orthopaedic Surgery and Research. 2019;14(1):182. doi: 10.1186/s13018-019-1228-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Richards LJ, Streifel A, Rodrigues JM. Utility of patch testing and lymphocyte transformation testing in the evaluation of metal allergy in patients with orthopedic implants. Cureus. 2019;11(9):e5761. doi: 10.7759/cureus.5761. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Frigerio E, Pigatto PD, Guzzi G, Altomare G. Metal sensitivity in patients with orthopaedic implants: A prospective study. Contact Dermatitis. 2011;64(5):273–279. doi: 10.1111/j.1600-0536.2011.01886.x. [DOI] [PubMed] [Google Scholar]
  • 40.Münch HJ, Jacobsen SS, Olesen JT, Menné T, Søballe K, Johansen JD, Thyssen JP. The association between metal allergy, total knee arthroplasty, and revision: Study based on the Danish Knee Arthroplasty Register. Acta Orthopaedica. 2015;86(3):378–383. doi: 10.3109/17453674.2014.999614. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Middleton S, Toms A. Allergy in total knee arthroplasty: A review of the facts. The Bone & Joint Journal. 2016;98(4):437–441. doi: 10.1302/0301-620X.98B4.36767. [DOI] [PubMed] [Google Scholar]
  • 42.Walker T, Rutkowski L, Innmann M, Panzram B, Herre J, Gotterbarm T, Aldinger PR, Merle C. Unicondylar knee arthroplasty using cobalt-chromium implants in patients with self-reported cutaneous metal hypersensitivity. The Bone & Joint Journal. 2019;101-B(2):227–232. doi: 10.1302/0301-620X.101B2.BJJ-2018-0778.R1. [DOI] [PubMed] [Google Scholar]
  • 43.Saccomanno MF, Sircana G, Masci G, Cazzato G, Florio M, Capasso L, Passiatore M, Autore G, Maccauro G, Pola E. Allergy in total knee replacement surgery: Is it a real problem? World Journal of Orthopedics. 2019;10(2):63–70. doi: 10.5312/wjo.v10.i2.63. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

Raw data have been submitted as supplementary material to the Journal.


Articles from Indian Journal of Orthopaedics are provided here courtesy of Indian Orthopaedic Association

RESOURCES