Short-Time Impedance Spectroscopy Using a Mode-Switching Nonsinusoidal Oscillator: Applicability to Biological Tissues and Continuous Measurement

Tomiharu Yamaguchi 1, Emiyu Ogawa 2, Akinori Ueno 1,*
Editors: Maria Giovanna Trivella, Abdelhamid Errachid, Francesca G Bellagambi
PMCID: PMC8587290  PMID: 34770258

Abstract

Herein, we propose an impedance spectroscopy method using a mode-switching nonsinusoidal oscillator and apply this method for measuring the impedance of biological tissues and continuous impedance measurement. To obtain impedance spectra over a wide frequency range, we fabricated a novel nonsinusoidal oscillator incorporating binary counters and analog switches. This oscillator could periodically switch oscillation frequency through the mode switching of the feedback resistor. From the oscillation waveform at each oscillation frequency of this circuit (oscillator), we determined the impedance spectrum of a measured object using the discrete-time Fourier transform. Subsequently, we obtained the broad impedance spectrum of the measured object by merging odd-order harmonic spectral components up to the 19th order for each oscillation frequency. From the measured spectrum, the resistive and capacitive components of the circuit simulating bioimpedance were estimated with high accuracy. Moreover, the proposed method was used to measure the impedance of porcine myocardium; changes in the impedance spectrum of the myocardial tissue due to coagulation could be measured. Furthermore, rapid variations in the resistance value of a CdS photocell could be continuously measured using the proposed method.

Keywords: capacitive coupling, impedance spectroscopy, nonsinusoidal oscillator, DFT, frequency switching, biological application, continuous impedance measurement

1. Introduction

Bioelectrical impedance analysis has become an important technique in the medical and healthcare fields [1,2,3,4,5,6,7,8,9]. Because the impedance of a living body (bioimpedance) depends on body composition, such as the water content of the body, it reflects medical conditions [10,11,12,13]. Thus, the absolute value and phase of bioimpedance are used to diagnose diseases, and this process is known as bioelectric impedance vector analysis [14,15,16,17,18,19]. Impedance measurement is also used in catheter ablation for treating arrhythmia [20,21,22,23]. By measuring the local impedance at an ablation site with electrodes integrated into an ablation catheter, the cauterization of myocardial tissue can be monitored. In the medical field, impedance may be measured using a single frequency or a few frequencies. However, bioimpedance has frequency characteristics over a wide range of frequencies, which originate from impedance components such as intracellular fluid resistance, extracellular fluid resistance, and the capacitance of cell membranes. Because cell membranes are highly insulating and cell fluids have ionic conductivity, in a low-frequency range, the current barely flows within cells but flows through the gaps among cells. Conversely, in a high-frequency range, the impedance caused by the capacitance of cell membranes decreases, meaning the current flows uniformly inside and outside cells. Therefore, if the impedance spectrum over a wide frequency range could be measured rapidly and conveniently, more detailed information about biological tissues could be obtained, which would be useful for diagnosing and treating medical conditions.

Impedance measurement generally involves coupling an electrode to a measured object through resistive coupling. Therefore, noninvasive impedance spectroscopy targeting insulated resistors, for instance, a living body covered with insulator-like clothing, has rarely been performed. To address this gap, we previously proposed capacitive coupling impedance spectroscopy (CIS) as a noninvasive technique for measuring impedance spectra through an insulation coating [24]. In this method, the measured object and capacitive coupling part are incorporated into an oscillation circuit. The oscillation frequency of the circuit depends on the coupling capacitance and the impedance of the measured object. Thus, even if the equivalent circuit model of the measured object is unknown, the change in these parameters can be monitored by measuring the oscillation frequency. This is applicable to monitoring a body movement and the breathing effort of a living body [25]. However, the equivalent circuit parameters of the measured object cannot be uniquely determined by only the oscillation frequency.

The frequency spectrum of the complex impedance of the measured object is obtained in a short time by analyzing the obtained oscillation waveform using the discrete-time Fourier transform (DFT), which allows estimating the resistance and capacitance components of the measured object. Therefore, CIS can be applied to not only measurement applications such as ablation by resistively contacting electrodes to biological tissue but also the measurement of biological impedance through an insulator such as clothing. Consequently, it is suitable for fabricating portable/wearable measurement systems. Furthermore, because the impedance spectrum is obtained from the oscillation waveform in every cycle, it enables an almost continuous impedance measurement. Continuous measurement of the impedance spectrum can be used to monitor biological processes such as cell differentiation [26].

In our previous CIS measurement, we used a nonsinusoidal oscillator incorporated into the measured object to estimate the impedance spectrum. DFT analysis of the nonsinusoidal wave obtained from this oscillator can determine the impedance at the fundamental frequency of the waveform and its odd harmonic frequencies. If the coupling capacitance is constant, and the measurement system has a high signal-to-noise ratio, the equivalent circuit parameters can be determined via less frequency points. However, in practical application, the coupling capacitance is often unknown and may fluctuate. To determine the equivalent circuit parameters of the measured object with high accuracy, we must estimate the impedance spectrum in the frequency range where the impedance has high frequency dependence. However, the oscillation frequency range of the nonsinusoidal oscillator is limited by the feedback resistance of it. Therefore, the feedback resistance cannot be set appropriately for an unknown coupling capacitance, and the impedance spectrum in the frequency range required for circuit parameter estimation may not be measured.

To solve this problem, we designed a nonsinusoidal oscillator that can change the oscillation frequency drastically by switching several feedback resistors. Using the significantly different values of the feedback resistances, even if the coupling capacitance fluctuates, one of the feedback resistors can cover the frequency range required for circuit parameter estimation and thus leads to a stable and high estimation accuracy.

Furthermore, high electrical noise may reduce the measurement accuracy of the impedance spectrum. In this measurement environment, it is difficult to estimate the circuit parameters accurately with less frequency points. Our proposed frequency-switchable (mode-switching) oscillator enables impedance measurement with many frequency points by merging the frequency spectra for each feedback resistance.

To apply our proposed nonsinusoidal oscillator to bioimpedance measurement, this study aimed to demonstrate the following four outcomes using CIS:

  • (1)

    Design a nonsinusoidal oscillation circuit capable of periodically switching oscillation frequency;

  • (2)

    Show that the proposed method can estimate circuit parameters for circuit models close to actual measured objects;

  • (3)

    Apply the proposed method to the impedance measurement of biological tissues;

  • (4)

    Show that the continuous measurement of impedance is possible.

2. Capacitive Coupling Impedance Spectroscopy

2.1. Mode-Switching Nonsinusoidal Oscillator

In CIS, a nonsinusoidal oscillator circuit, such as a Schmitt trigger inverter oscillator, is used. Although this oscillation circuit does not have a high-accuracy oscillation frequency, it allows constructing a system that is small, inexpensive, and easy to handle because it has few parts, and the oscillation is stable. In our previous CIS study, the impedance of the measured object at the fundamental frequency (oscillation frequency) and odd-order harmonics were determined using DFT analysis of the oscillation waveforms. However, in some cases, the impedance spectra of sufficient data points cannot be obtained with a single oscillation frequency, and the estimation accuracy decreases when the circuit parameters are estimated using an equivalent circuit model. To address this problem, we designed a novel mode-switching nonsinusoidal oscillator that can switch among multiple oscillating frequencies.

Figure 1 shows a schematic of the mode-switching nonsinusoidal oscillator. A nonsinusoidal oscillation waveform corresponding to the impedance of the measured object is obtained by incorporating the measured object, capacitively coupled with the electrode, into the oscillation circuit. Z˙X represents the impedance of the measured object, including capacitive coupling, whose circuit parameters differ depending on the equivalent circuit model of the measured object. For example, in the case of bioimpedance, it comprises extracellular fluid resistance RX, intracellular fluid resistance RPX, and the capacitance of cell membranes CPX [27,28]. In particular, RPX= 0 Ω for skin impedance [29]. Z˙X1 is the impedance of the measured object excluding capacitive coupling. Capacitances CSX1 and CSX2 correspond to the coupling capacitance between the insulator and electrode. Z˙A is the combined impedance of resistance RA and parasitic capacitance CA. The current i12t flowing into Z˙X is calculated from the voltage v1t generated in Z˙A (I/V conversion). The resistance RA represents the resistance for measuring the current. CA represents a parasitic component caused by the parasitic capacitance of the circuit board, input capacitance of the probe of the oscilloscope, and other factors. The feedback resistance RF of the oscillator circuit comprises resistors RF1, RF2, and RF3 connected in series. The Schmitt trigger inverter switches the high and low levels of the voltage vOSCt, VH and VL, by the input voltage v2t. The voltage v2t changes by the charge/discharge of the capacitors connected on the input side of the Schmitt trigger inverter, which causes the square wave of the voltage vOSCt. In addition, square-like and triangle-like waveforms are generated as v1t and v2t, respectively. Because the charging/discharging speed of the capacitors is determined by the current i12t flowing through Z˙X, the oscillation frequency depends on Z˙X, Z˙A, and RF. If Z˙X is a resistive element RX and capacitive coupling CSX in series, and parasitic capacitance CA can be ignored, the oscillation frequency f0 can be expressed with the following [24]:

f0=1RF+RA+RXCSXlnVHVAVHVP+lnVBVLVNVL, (1)
VA=VN+RA+RXRF+RA+RXVHVL, (2)
VB=VPRA+RXRF+RA+RXVHVL, (3)

where CSX=CSX1CSX2/CSX1+CSX2 denotes the combined capacitance of CSX1 and CSX2. VP and VN are the threshold voltages of the Schmitt trigger inverter. If Z˙X has a more complex circuit network, these equations are not strictly valid, but the oscillation frequency can be roughly estimated using these equations.

Figure 1.

Figure 1

Mode-switching nonsinusoidal oscillator. The oscillation frequency f0 and the oscillation period T0 of the oscillator are switched between four different frequencies by 4-bit binary counters and bilateral analog switches. CLK of 4-bit binary counters denotes a clock signal, and RST represents a reset signal. Q0, Q1, Q2, and Q3 of 4-bit binary counters denote zero bits, and the first, second, and third bits of the output signals, respectively. CO represents a bypass capacitor to remove the switching noise in the oscillator, and its value is 0.47 nF. CTR1 and CTR2 are binary counters to control the repeat count and the mode number of the mode switching, respectively.

Because the operation of the bilateral switches can cause an unstable oscillation such as switching noise, we used the 4-bit binary counter CTR1 to stabilize the oscillation. The counter CTR1 controls the repeat count for switching the feedback resistors. The square wave signal vOSCt of the oscillator is input to the counter CTR1 as the clock signal. The 0-bit output Q0 of the counter CTR1 has half the frequency of vOSCt. For this reason, the bilateral switches operate every two cycles of the oscillation waveform for each RF, and the oscillation of the second half cycle becomes stable.

Subsequently, the frequency-divided square wave signal is input as a clock signal to the counter CTR2. CTR2 determines the mode number obtained from the mode switching of the oscillator. Moreover, the 0-bit and 1-bit outputs Q0 and Q1  of the counter CTR2 are input as the control signals of the bilateral switches SW1 and SW2, respectively. Because RF2  or RF3  is short circuited when the switch is ON, RF is switched every two cycles of the oscillation waveform in four modes: RF1, RF1+RF2, RF1+RF3, and RF1+RF2+RF3 (Table 1). Therefore, the oscillation frequency can be switched periodically.

Table 1.

Four different modes of the feedback resistance RF.

Mode Q1 Q0 SW1 SW2 RF
1 0 0 OFF OFF RF1+RF2+RF3
2 0 1 OFF ON RF1+RF3
3 1 0 ON OFF RF1+RF2
4 1 1 ON ON RF1

0 and 1 indicate a low level and a high level of a square wave, respectively.

The procedure for estimating impedance is determined using a previously reported procedure [24]. The voltages v1t  and v12t  measured in the circuit, shown in Figure 1, form the periodic nonsinusoidal waveforms of the oscillation frequency f0. Therefore, the Fourier series expansion of the one cycle of the observed waveform comprises a sine wave of the fundamental frequency f0 and its integer multiple kf0 (k=1, 2, 3, ), and the coefficients of each frequency component are obtained as the amplitude and phase spectra through the DFT of the section concerned. From the obtained amplitude and phase spectra, Z˙X  can be calculated using the following equation:

Z˙Xkf0=V˙12kf0I˙12kf0=V˙12kf0V˙1kf0·Z˙Akf0ejθV12kf0θV1kf0+θZAkf0, (4)

where V˙12kf0 and V˙1kf0 denote the complex voltages of v12t and v1t, respectively; I˙12kf0 denotes the complex current of i12t; V˙12kf0 and θV12kf0 represent the amplitude and phase of the complex voltage at Z˙X; V˙1kf0  and θV1kf0 represent the amplitude and phase of the complex voltage at Z˙A; and Z˙Akf0 and θZAkf0 represent the magnitude and phase of the impedance Z˙A, respectively. The impedance estimation using Equation (4) is valid for any periodic waveform of v12t and v1t because Equation (4) is applied to each frequency component obtained from the oscillation waveform of v12t and v1t. The real and imaginary parts of the impedance Z˙X, ReZ˙X and ImZ˙X, are obtained with Equations (5) and (6):

ReZ˙X=V˙12nf0V˙1nf0·Z˙Anf0cosθV12nf0θV1nf0+θZAnf0, (5)
ImZ˙X=V˙12nf0V˙1nf0·Z˙Anf0sinθV12nf0θV1nf0+θZAnf0. (6)

We can obtain the oscillation waveform with multiple oscillation frequencies using the mode-switching oscillator. Therefore, one synthesized impedance spectrum is obtained by merging the spectral data for each oscillation frequency.

If it is known that Z˙X is a series connection of the resistor RX and the capacitive coupling CSX, RX and CSX can be estimated using the least squares method for a finite number of frequency spectra nf0 (n=1, 2, 3, , N).

ReZ˙X=RX, (7)
ImZ˙X=12πnf0CSX. (8)

Furthermore, when Z˙X is the common bioimpedance (Figure 1) equivalent circuit model,

ReZ˙X=2πnf0CPX2RXRPX2+RX2RPX+RX2πnf0CPX2RPX+RX2+1, (9)
ImZ˙X=12πnf0CSX2πnf0CPXRX22πnf0CPX2RPX+RX2+1. (10)

Similarly, the circuit parameters can be estimated using the least squares method. When RPX= 0 Ω, Equations (9) and (10) represent the real and imaginary parts of the circuit comprising the parallel-connected RX and CPX (RC parallel circuit) and coupling capacitance CSX, respectively.

2.2. Experimental Method

2.2.1. CIS of Parallel RC Circuits and Bioimpedance Models

A readily available 74HC14AP (Toshiba, Japan) was used as the Schmitt trigger inverter integrated circuit (IC), as illustrated in Figure 1. The typical VP and VN values of 74HC14AP in Equations (1)–(3) are 3.5 V and 2.3 V at 6 V of the supply voltage, respectively. A TC4520BP (Toshiba, Japan) was used as the 4-bit binary counter IC, and a TC74HC4066AP (Toshiba, Japan) was used as the bilateral analog switch IC. The power supply voltage of these ICs was set to 6 V, and VH and VL were set to 6 V and 0 V, respectively. Z˙A was considered to consist only of the resistor RA. RA was set to 0.20 Ω. There is typically CA of several tens of pF due to the parasitic components such as the probe capacity of the oscilloscope, but CA is negligible because RA is sufficiently small. The oscillator was designed to measure the impedance of biological tissue with a resistance of less than 1 kΩ and a capacitance of several nF. We selected four fundamental frequencies so that they contribute to an accurate estimation of the measured impedance parameters based on preliminary experiments. Two sets of the fundamental frequencies obtained by combining RF1, RF2, and RF3 are shown in Table 2.

Table 2.

Two sets of the oscillation frequencies obtained by combining RF1, RF2, and RF3.

Set RF1
(kΩ)
RF2
(kΩ)
RF3
(kΩ)
4 Modes of RF
(kΩ)
Estimated Oscillation
Frequencies (kHz)
1 3.3 1.8 3 8.1, 6.3, 5.1, 3.3 16.9, 22.3, 28.4, 47.9
2 5.1 1 2 8.1, 7.1, 6.1, 5.1 16.9, 19.5, 23.1, 28.4

The oscillation frequencies were estimated using Equations (1)–(3) with RX= 0 Ω. The actual oscillation frequencies depend on Z˙X.

First, circuit parameters comprising the RC parallel circuit (RX and CPX) and coupling capacitance CSX were estimated using CIS. Metal film resistors and polypropylene (PP) film capacitors with known resistance and capacitance values were incorporated into the circuit as resistance RX and capacitance CPX, CSX. To demonstrate that our proposed method can estimate CPX with various values, we set the capacitance of 1.016 or 9.85 nF as CPX. The coupling capacitance CSX was set to 9.76 nF. We used the combination of set 1 in Table 2 as RF1, RF2, and RF3. These feedback resistors enable determining the impedance spectrum in a wider frequency range.

Further, we estimated the circuit parameters of the bioimpedance model (RX, RPX, CPX, and CSX) with the coupling capacitance using CIS. The estimation of RPX is more difficult than that of RX because it is affected by the accuracy of the impedance estimated using higher-order harmonics. For this reason, we evaluated the estimation accuracy of RPX in more detail. Resistance with a constant value of 0.30 kΩ was used as RX. CPX was set to 3.22 nF, which simulates the capacitance with biological tissue. CSX was set to 9.76 nF. To obtain many frequency points at lower frequencies, we used the combination of set 2 in Table 2 as RF1, RF2, and RF3.

In these experiments, we prepared ten resistors as RX or RPX with different resistances from 0.10 to 1.0 kΩ and set the coupling capacitance CSX to 9.76 nF.

2.2.2. Application of CIS to Biological Tissues

To demonstrate that the impedance spectrum of biological tissues can be measured using this CIS system, we measured the impedance of porcine myocardium. The coagulation of the myocardium causes tissue degeneration and decreases the impedance of the coagulated part [30]. Therefore, porcine myocardium samples were coagulated through electrocautery to determine whether the change in impedance before and after the coagulation could be measured. Three porcine hearts were prepared, and the myocardium of each left ventricle was ablated in four locations (Figure 2a). Counter electrodes were placed across the back of the porcine myocardium, and it was coagulated through electrocautery for 30 s at a power of 30 W. The electrocautery was set to the soft coagulation mode. Moreover, measuring electrodes were set up, as illustrated in Figure 2b, with circular electrodes on either side of the ablation point. The electrodes were spaced 10 mm apart and directly contacted at the same position of the myocardium before and after each coagulation. The impedance generated when the electrode and myocardial tissue come into contact forms a resistive coupling. Therefore, the coupling capacitance is extremely small, meaning the oscillating frequency of the mode-switching nonsinusoidal oscillator increases. A PP film capacitor of CSX= 9.76 nF was connected in series between the electrode and oscillation circuit so that the oscillation frequency was reduced to approximately 20 kHz. RF1, RF2, and RF3 were 3.3, 1.8, and 3.0 kΩ (set 1), respectively.

Figure 2.

Figure 2

Impedance measurement of porcine myocardium: (a) photograph of the porcine left ventricle coagulated using electrocautery; (b) measurement electrode fixture.

2.2.3. Continuous Impedance Measurement Using CIS

Furthermore, a CdS photocoupler (LCR0203, Nanyang Senba Optical and Electronic, Shenzhen, China) was incorporated into the oscillator circuit as a part of Z˙X, and the temporal change in RX and RPX could be measured using the proposed method. CdS photocells can be regarded as simple resistors. Moreover, increasing the current IPh flowing through a light-emitting diode (LED) built in the CdS photocoupler decreases the resistance value of the CdS photocell. Therefore, the resistance value can be changed over time by changing the current IPh of the LED. This current was controlled via a constant current circuit using an operational amplifier (NJU7032D, NJR, Tokyo, Japan) and an N-channel field-effect transistor (2N7000, ON Semiconductor, Phoenix, USA), as illustrated in Figure 3. In this circuit, the current IPh can be controlled in the range 0.6–10 mA by adjusting the voltage VC between 0.12 and 2 V. The capacitive coupling CSX was set to 9.76 nF. First, it was measured when only the photocell was connected as a resistor RX via CSX (Figure 3a). In this case, the counter IC and analog switch IC of the mode-switching nonsinusoidal oscillator were not used because the parameters could be estimated sufficiently accurately with a single mode of RF. Only the RF1 of 8.2 kΩ was connected as RF. Subsequently, the temporal change in the impedance spectrum was determined from the oscillation waveform of the oscillator using the bioimpedance model and CSX (Figure 3b,c). RF3 is short circuited so that the oscillation frequency is switched in two modes to increase the temporal resolution. RF1 and RF2 were 3.3 and 1.8 kΩ, respectively.

Figure 3.

Figure 3

Constant current circuit for the LED current control of the photocoupler. The CdS photocell was used as (a) a single resistor, (b) a resistor RX of the bioimpedance model, or (c) RPX of its model.

2.2.4. Data Acquisition and DFT Analysis

Time waveform data of voltages were collected using a digital oscilloscope (PicoScope 5444D, Pico Technology, Cambridgeshire, UK). v1t and v2t in Figure 1 were measured, and time waveform data for v12t were obtained from the difference between them. Thereafter, DFT was performed for the obtained waveform data. When using the mode-switching function, the oscillation frequency is switched in four or two modes every two cycles. DFT was performed individually for the waveform of the second half cycle of each oscillation frequency, and the merged spectral data for the set of oscillation frequencies (one segment) were obtained. DFT was performed for every cycle of the oscillation waveform when mode switching was not used. A rectangular window was used for the window function of DFT. Furthermore, the impedance spectra for 10 segments were averaged, except for experiments using a CdS photocoupler. The real and imaginary parts of Z˙X were calculated from the obtained DFT data, and RX, RPX,CPX, and CSX were estimated via the least squares method using Equations (7) and (8) or (9) and (10). In the estimation using Equations (9) and (10), first, RX, RPX, and CPX were estimated using Equation (9) from the DFT data of the real part of Z˙X. The estimated RX, RPX, and CPX were then substituted into Equation (10), and CSX was estimated from the DFT data of the imaginary part of Z˙X. In the circuit parameter estimation, odd-order harmonic components up to the 19th order obtained through DFT are used. However, because the frequency characteristics of the PP film capacitor and parasitic capacitance of the circuit board affect the estimation result in high-frequency ranges, harmonic components greater than 500 kHz when CPX was less than 2.0 nF and harmonic components greater than 350 kHz when CPX was 2.0 nF or higher were not used for estimation.

3. Results and Discussion

3.1. CIS of Parallel RC Circuits and Bioimpedance Models

Figure 4 illustrates oscillation waveforms when  RX = 0.10 kΩ, RPX = 0 Ω, CPX = 1.016 nF, and CSX = 9.76 nF are connected as Z˙x in the mode-switching nonsinusoidal oscillator. RF switched every two cycles, and the oscillation frequency changed in four modes. Oscillating frequencies when RF was 8.1, 6.3, 5.1, and 3.3 kΩ were 14.0, 18.2, 23.2, and 38.4 kHz, respectively. The smaller values than the values estimated using Equation (1) may be because the actual threshold voltages of the Schmitt trigger IC are different from the typical values in the datasheet specification. The spike noise observed at 0.4 ms is due to the switching noise of the Schmitt trigger inverter. However, the spike noise has no effect on the estimation of the impedance spectrum because we used only the waveform of the second half cycle for DFT.

Figure 4.

Figure 4

Four-mode oscillation waveforms of v1t and v12t obtained through the mode-switching nonsinusoidal oscillator. RX, RPX, CPX, and CSX were set to 0.10 k, 0 Ω, 1.016 nF, and 9.76 nF, respectively. The oscillation frequency was switched every two cycles of each RF. The DFT analysis was performed on the oscillation waveforms of the second half of two cycles.

Figure 5 shows the spectra of Z˙X calculated using DFT analysis. By applying DFT to the oscillation waveform for each RF, four modes of frequency spectra were obtained. These spectra agreed well with the theoretical values calculated from Equations (9) and (10) with RPX = 0 Ω. Even when the resistance and capacitance were large, such as RX = 0.91 kΩ and CPX = 9.85 nF, impedance spectra agreed well with the theoretical values (Figure 6).

Figure 5.

Figure 5

Frequency spectra of the (a) real part ReZ˙X and (b) imaginary part ImZ˙X of the impedance Z˙X comprising the resistance RX of 0.10 kΩ, resistance RPX of 0 Ω, capacitance CPX of 1.016 nF (parallel RC circuit), and capacitance CPX of 9.76 nF. The dashed lines represent the theoretical curves calculated from Equations (9) and (10).

Figure 6.

Figure 6

Frequency spectra of the (a) real part ReZ˙X and (b) imaginary part ImZ˙X of the impedance Z˙X comprising the resistance RX of 0.91 kΩ, resistance RPX of 0 Ω, capacitance CPX of 9.85 nF (parallel RC circuit), and capacitance CSX of 9.76 nF. The dashed lines represent the theoretical curves calculated from Equations (9) and (10).

Further, the values of RX,  CPX, and CSX were estimated from the impedance spectrum. Figure 7 illustrates the estimation values of RX, CPX, and CSX. To verify the usefulness of the mode-switching function, the estimation was performed using the spectrum obtained from the four combinations of RF and the spectrum obtained from one mode and two modes of RF. The estimation error sometimes exceeded 2% when only DFT data obtained from one mode of RF were used; however, the estimation accuracy improved when two modes of RF were used. Moreover, each circuit parameter could be estimated with high accuracy, regardless of the value of RX and CPX, using the four modes of RF. This demonstrates that the superposition of the impedance spectra through the mode-switching function improves the estimation accuracy of the circuit parameters.

Figure 7.

Figure 7

Absolute relative estimation errors of the circuit parameters of parallel RC circuits: (a) RX = 0.10 kΩ, CPX = 1.016 nF, and CSX = 9.76 nF; (b) RX = 0.91 kΩ, CPX = 9.85 nF, and CSX = 9.76 nF.

Subsequently, with CPX = 1.016 nF and CSX = 9.76 nF, we measured the impedance spectrum by changing RX from 0.1 to 1 kΩ. Figure 8 shows some of the impedance spectra obtained. The impedance spectra according to the values of RX were obtained, and the fitting curves obtained from Equations (9) and (10) using the least squares method agreed well with the DFT data.

Figure 8.

Figure 8

Frequency spectra of the (a) real part ReZ˙X and (b) imaginary part ImZ˙X of the impedance Z˙X comprising the parallel RC circuits for CPX= 1.016 nF and coupling capacitance CSX= 9.76 nF. The DFT data were fitted to Equations (9) and (10) to calculate RX, CPX, and CSX (dashed lines).

When the circuit parameters were estimated, the estimation error of RX was within 12 Ω, and RX could be accurately estimated from 0.1 to 1 kΩ (Figure 9a). Furthermore, the errors of CPX and CSX were within 1.28% and 0.82%, respectively, which are sufficiently small errors (Figure 9b,c). These results indicate that the circuit parameters of the RC parallel circuit can be estimated with high accuracy through capacitive coupling.

Figure 9.

Figure 9

RX, CPX, and CSX of the parallel RC circuits estimated using oscillation waveforms and DFT: (a) the estimated RX, (b) relative error of the estimated CPX, and (c) relative error of the estimated CSX. The insets are the circuit model used for fitting, and the estimated parameters are highlighted in pink.

The impedance characteristics of a measured object are frequently discussed using Cole–Cole plots. The proposed method can obtain a wide frequency range spectrum using multiple values of RF, and the Cole–Cole plots, as illustrated in Figure 10, can be easily obtained. The Cole–Cole plot of the impedance Z˙X1 excluding capacitive coupling showed semicircles, and the radius of the semicircle increased with resistance RX.

Figure 10.

Figure 10

Cole–Cole plot of Z˙X1 for the parallel RC circuits with CPX=1.016 nF and CSX=9.76 nF. The dashed lines represent the fitting curves obtained from Equations (9) and (10).

Moreover, we investigated whether the circuit parameters of the bioimpedance model could be estimated. Figure 11 shows the impedance spectrum of the bioimpedance model when RPX was 0.20, 0.62, or 1.0 kΩ. Despite the small change in the real part compared with the RC parallel circuit due to the existence of RPX, the fitting results using Equation (9) agreed well with the DFT data. The imaginary part could also be fitted accurately using Equation (10). Figure 12 shows the estimation results of RX, RPX, CPX, and CSX. The estimation errors of RX and RPX were within 6 and 11 Ω, respectively, which are sufficiently small estimation errors. The estimation errors of CPX and CSX were also within 3.79% and 1.03%, respectively.

Figure 11.

Figure 11

Frequency spectra of the (a) real part ReZ˙X and (b) imaginary part ImZ˙X of the impedance Z˙X comprising the bioimpedance models for RX=0.30 kΩ, CPX=3.22 nF, and coupling capacitance CSX=9.76 nF. DFT data were fitted to Equations (9) and (10) to determine RX, RPX, CPX, and CSX (dashed lines).

Figure 12.

Figure 12

RX, CPX, and CSX of the bioimpedance models estimated using oscillation waveforms and DFT: (a) the estimated RX, (b) the estimated RPX, (c) relative error of the estimated CPX, and (d) relative error of the estimated CSX. The insets are the circuit models used for fitting, and the estimated parameters are highlighted in pink.

Figure 13 illustrates the Cole–Cole plots of the bioimpedance models excluding capacitive coupling. The Cole–Cole plot of Z˙X1 showed semicircles expected from Equations (9) and (10), and the radius of the semicircle decreased with resistance RPX.

Figure 13.

Figure 13

Cole–Cole plot of Z˙X1 for the bioimpedance models with CPX=3.22 nF and CSX=9.76 nF. The dashed lines represent the fitting curves obtained from Equations (9) and (10).

These results demonstrate that the circuit parameters can be estimated using the proposed method, even for complex equivalent circuit models such as bioimpedance. In this study, RF over 1 kΩ was not estimated because we designed the mode-switching oscillator to apply to biological tissue. Our proposed oscillator generates a stable oscillation waveform if the feedback resistance is sufficiently larger than the magnitudes of Z˙X and Z˙A. Therefore, the range of load that our circuit can carry depends on the feedback resistance. Resistance of less than several kΩ can be estimated when the feedback resistance is about 10 kΩ. If the feedback resistance is higher, we will be able to estimate higher resistance of about 10 kΩ.

3.2. Application of CIS to Biological Tissues

An example of the application of impedance measurement in the biomedical field is the monitoring of cauterization in catheter ablation. We, therefore, used CIS to measure the local change in impedance when porcine myocardium was coagulated. Stable oscillations with four modes of oscillating frequencies were obtained when the electrodes in Figure 2b and the surface of the porcine myocardium were in contact. Figure 14 shows the spectrum of the real part of the impedance Z˙X calculated using DFT from the obtained waveform. The equivalent circuit of biological tissues, such as porcine myocardium, is approximated using a bioimpedance model. Therefore, the real part of the impedance decreased with increasing frequency as the capacitance of cell membranes decreases. Moreover, because the measurement area included regions other than the ablation site, the impedance was larger than the local impedance (approximately 0.10 kΩ) measured through a commercial catheter ablation system. When electrocautery was performed, the impedance of the porcine myocardium decreased, regardless of the frequency range. Because cell membranes at the ablation site have high ion permeability, this decrease in impedance indicates that a part of the myocardium was degenerated through coagulation. The DFT data partially agree with the fitting curve produced by Equation (9), but the DFT data tended to be larger than the fitted data in low-frequency ranges. Furthermore, when CSX was estimated using Equation (10) from the DFT data of the imaginary part of the impedance, the value before ablation was 7.98 nF and the value after ablation was 8.29 nF, both of which are low values compared to the capacitance of the PP film capacitor used, which was 9.76 nF. This suggests that resistance and capacitance components also exist between the surface of the porcine myocardium and electrodes. Therefore, to estimate circuit parameters with high accuracy, applying an equivalent circuit model is necessary considering the impedance of the interface of the myocardium and electrodes.

Figure 14.

Figure 14

Frequency spectrum of the real part ReZ˙X of the impedance Z˙X of the porcine myocardium and coupling capacitance. DFT data were fitted to Equation (9) to determine RX, RPX, and CPX (dashed lines).

Figure 15 shows the frequency spectra of the porcine myocardium. Z˙X1 in Figure 15 represents the impedance of the porcine myocardium only, excluding the impedance of the capacitive coupling CSX from Z˙X. Here, CSX uses a value estimated through fitting, not the capacitance value of the PP film capacitor. The Cole–Cole plot of the porcine myocardium formed a semicircle, as expected from the bioimpedance model, but the semicircle was distorted (Figure 15a). This may be because the impedance between the surface of the porcine myocardium and electrodes was not considered or because the impedance of the myocardial tissue was not uniform. When the myocardium was ablated, the magnitude of the imaginary part of the impedance also decreased, similar to the real part (Figure 15b,c), as did the radius of the semicircle in the Cole–Cole plot.

Figure 15.

Figure 15

Frequency spectra of the impedance Z˙X1 of the porcine myocardium: (a) the Cole–Cole plot, (b) the real part ReZ˙X1, and (c) the imaginary part ImZ˙X1. Z˙X1 represents the impedance obtained by subtracting the impedance of the estimated CSX from Z˙X. ReZ˙X1 equals ReZ˙X in Figure 14.

Furthermore, impedance was measured at four locations using CIS for each of the three porcine myocardium samples. Figure 16 shows the result of estimating RX, RPX, and CPX from the determined impedance spectra. Ablation decreased the resistive components RX and RPX and increased the capacitive component CPX in all 12 samples. A paired t-test on the obtained samples also showed significant differences at a significance level of 0.1%, indicating that CIS can measure changes in impedance associated with myocardial tissue degeneration.

Figure 16.

Figure 16

Comparison of equivalent circuit parameters between the untreated and ablated left ventricles of porcine hearts: (a) RX, (b) RPX, and (c) CPX. The sample size is 12.

3.3. Continuous Impedance Measurement Using CIS

Figure 17 shows the oscillation waveform when a CdS photocoupler and a 9.76 nF PP film capacitor are connected in series as Z˙X (Figure 3a). Measurement was started with a current IPh of 0.6 mA flowing through the LED, and the current IPh was increased instantaneously to 2.0 mA at 10 ms after the start of the measurement. Thereafter, the current IPh was returned to 0.6 mA again at 23 ms after the start of the measurement. Figure 17a shows that the switching of the current IPh changed the peak-to-peak value of the voltage v1t, indicating that the resistance of the CdS cell changed. The peak-to-peak value VPP of the voltage v1t is expressed by the following equation:

VPP=RARFVPVN+RARF2RFRA+RX+RF1VHVL. (11)

Figure 17.

Figure 17

Single-mode oscillation waveforms of v12t and v1t for the nonsinusoidal oscillator incorporating the CdS photocell (Figure 3a): (a) full waveform and (b) enlarged waveform. The photocurrent IPh of the photocoupler is changed at 10 and 23 ms. The high and low levels of IPh are 2.0 and 0.6 mA, respectively. Circuit parameters are estimated for each cycle (blue dashed line). Representative points of time in each cycle are the ends of the cycles such as t1 and t2.

As shown in this equation, VPP increases with decreasing resistance RX. Figure 17b indicates that a stable oscillation was obtained.

Continuous frequency spectra were obtained through DFT from the waveforms of the respective cycles of Figure 17a, and the resistance RX of the CdS photocell could be estimated by applying Equation (7). The temporal waveform of RX obtained by estimation is presented in Figure 18a. When IPh was increased instantaneously, RX gradually decreased and then saturated at approximately 10 ms after the current increased. Because the response time of the CdS photocell used is relatively short, approximately 2.5 ms, this change in RX reflects the optical response of the CdS photocell. The increase from 24 ms indicates that the resistance of the photocell increased with the reduction in IPh. Generally, the recovery time of the photocell is slower than the response time. Therefore, the gradual increase in the resistance is a reasonable response. Figure 18b compares the saturated and true values of RX at various IPh values. The error between the estimated and true RX values was 29 Ω at the maximum, and thus RX could be estimated accurately.

Figure 18.

Figure 18

Resistance RX of the photocell estimated using oscillation waveforms in Figure 17a and DFT: (a) time–resistance curves and (b) saturation values at a high IPh period. Time of the time–resistance curves corresponds to the representative points of time in Figure 17b.

Moreover, a CdS photocell was incorporated as RX or RPX in the bioimpedance model (Figure 3b,c), and the resistance of the CdS photocell was changed over time. Figure 19 shows the oscillation waveform when the resistance of the CdS photocell was changed temporally as RX. Measurement began with a current IPh of 4.0 mA flowing through the LED, and the current IPh was increased instantaneously to 8.0 mA at 10 ms after the measurement began. Thereafter, IPh was returned to 4.0 mA again at 23 ms after the start of the measurement. RF was 3.3 or 5.1 kΩ, and mode switching was performed in two modes.

Figure 19.

Figure 19

Two-mode oscillation waveforms of v12t and v1t for the nonsinusoidal oscillator incorporating the CdS photocell and circuit elements of the bioimpedance model (Figure 3b): (a) full waveform and (b) enlarged waveform. The photocurrent IPh of the photocoupler is changed at 10 and 23 ms. The high and low levels of IPh are 8.0 and 4.0 mA, respectively. Circuit parameters are estimated for each cycle (blue dashed line). Representative points of time in each cycle are the ends of the cycles such as t1 and t2. The frequency f is the oscillation frequency obtained from the second half cycle of the oscillation waveform at RF = 3.3 kΩ.

A stable oscillation was obtained, as illustrated in Figure 19b, and the oscillation frequency switched in two modes. A VPP change was not observed because the change in IPh was smaller than that shown in Figure 17a. Although Equation (11) is not strictly valid when CPX is included in Z˙X, VPP depends on RX connected in parallel with RPX and CPX. However, because the resistance change due to the switching of IPh was only about 0.1 kΩ, the VPP change was negligibly small. Further, when the temporal change in RX, RPX, and CPX was estimated through the DFT analysis of the oscillation waveform, only RX changed according to the change in the CdS photocell resistance (Figure 20a). Furthermore, even when RPX and CPX are connected in parallel with RX, the oscillation frequency changes with RX of the measured object and is not significantly affected by electrical noise. Therefore, when the equivalent circuit model of the measured object is unknown or the signal-to-noise ratio in the circuit is low, the oscillation frequency can be used for monitoring the circuit parameters. However, the change in RPX was not reflected in the oscillation frequency when the CdS photocell was incorporated as RPX (Figure 20b). The oscillation frequency also depends on the impedance Z˙X including RPX. However, RPX is connected in series with CPX, and CPX has a relatively large impedance at the oscillation frequency. Therefore, the change in RPX does not significantly contribute to the changes in Z˙X, which leads to no change in the oscillation frequency.

Figure 20.

Figure 20

Time–resistance, time–capacitance, and time–frequency curves of the resistance of the photocell and circuit elements of the bioimpedance model estimated using oscillation waveforms and DFT. The photocell was used as (a) RX or (b) RPX. The other resistance is 0.20 kΩ; the capacitance CPX is 3.22 nF; and the capacitance CSX is 9.76 nF. The times of these curves correspond to the representative points of time in Figure 19b. The insets are photocell-incorporated circuits.

DFT analysis indicated that only RPX changed, but the dispersion of the data was particularly large for RPX. The estimation accuracy of RPX depends on the measurement accuracy of the impedance in high-frequency ranges, but the spectra of high-frequency ranges are susceptible to noise because the signal of the harmonic component is smaller than the fundamental frequency component. Therefore, the estimation accuracy of RPX may be poor. However, because one segment of the oscillation waveform is sufficiently short and the time resolution is high, the estimation accuracy can be improved by considering the moving average. The above results indicate that the proposed method can be used to continuously monitor the temporal variation in impedance.

4. Conclusions

We fabricated a mode-switching nonsinusoidal oscillator to accurately estimate the circuit parameters of impedance, including multiple resistances and capacitances, using CIS. The oscillation waveform of the fabricated circuit had multiple oscillation frequencies caused by the periodic switching of the feedback resistor. This enabled us to obtain a broader frequency spectrum range with a higher resolution than a single oscillation frequency through DFT analysis. Further, the circuit parameters of complicated circuit models such as RC parallel circuits and bioimpedance models could also be estimated accurately from the oscillation waveforms of this circuit.

We applied CIS using the mode-switching nonsinusoidal oscillator to impedance measurement of porcine myocardium. Consequently, the changes in the local impedance spectrum caused by the coagulation of the porcine myocardium could be measured. Moreover, the resistive components decreased and the capacitive components increased during electrocautery by fitting the impedance spectra with the bioimpedance model. By measuring and analyzing the continuous oscillation waveform when a CdS photocell was incorporated into the mode-switching nonsinusoidal oscillator, measuring the changes in impedance over time was possible.

These findings demonstrate that temporal changes in the impedance of biological tissues can be monitored using our proposed method. However, because this study excluded the impedance measurements of biological tissues through capacitive coupling using an insulator, future studies should aim to apply CIS to biological tissues covered with an insulator, such as the human body with clothing.

Author Contributions

Conceptualization, A.U.; methodology, A.U.; software, T.Y.; validation, T.Y.; formal analysis, T.Y.; investigation, T.Y., E.O., and A.U.; resources, E.O. and A.U.; data curation, T.Y.; writing—original draft preparation, T.Y. and A.U.; writing—review and editing, T.Y., E.O., and A.U.; visualization, T.Y.; supervision, A.U.; project administration, A.U.; funding acquisition, A.U. All authors have read and agreed to the published version of the manuscript.

Funding

This research was supported, in part, by the SECOM Science and Technology Foundation and, in part, by the Private University Research Branding Project supported by MEXT, Japan.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author, A.U., upon reasonable request.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Grossi M., Riccò B. Electrical impedance spectroscopy (EIS) for biological analysis and food characterization: A review. J. Sens. Sens. Syst. 2017;6:303–325. doi: 10.5194/jsss-6-303-2017. [DOI] [Google Scholar]
  • 2.Szuster B., Roj Z.S.D., Kowalski P., Sobotnicki A., Woloszyn J. Idea and measurement methods used in bioimpedance spectroscopy. Adv. Intell. Syst. Comput. 2017;623:70–78. [Google Scholar]
  • 3.Davies S.J., Davenport A. The role of bioimpedance and biomarkers in helping to aid clinical decision-making of volume assessments in dialysis patients. Kidney Int. 2014;86:489–496. doi: 10.1038/ki.2014.207. [DOI] [PubMed] [Google Scholar]
  • 4.Piccoli A. Bioelectric impedance measurement for fluid status assessment. Contrib. Nephrol. 2010;164:143–152. doi: 10.1159/000313727. [DOI] [PubMed] [Google Scholar]
  • 5.Jaffrin M.Y., Morel H. Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods. Med. Eng. Phys. 2008;30:1257–1269. doi: 10.1016/j.medengphy.2008.06.009. [DOI] [PubMed] [Google Scholar]
  • 6.Kyle U.G., Bosaeus I., De Lorenzo A.D., Deurenberg P., Elia M., Gómez J.M., Heitmann B.L., Kent-Smith L., Melchior J.-C., Pirlich M., et al. Bioelectrical impedance analysis—Part I: Review of principles and methods. Clin. Nutr. 2004;23:1226–1243. doi: 10.1016/j.clnu.2004.06.004. [DOI] [PubMed] [Google Scholar]
  • 7.Kyle U.G., Bosaeus I., De Lorenzo A.D., Deurenberg P., Elia M., Gómez J.M., Heitmann B.L., Kent-Smith L., Melchior J.-C., Pirlich M., et al. Bioelectrical impedance analysis—Part II: Utilization in clinical practice. Clin. Nutr. 2004;23:1430–1453. doi: 10.1016/j.clnu.2004.09.012. [DOI] [PubMed] [Google Scholar]
  • 8.Demura S., Sato S., Kitabayashi T. Percentage of total body fat as estimated by three automatic bioelectrical impedance analyzers. J. Physiol. Anthropol. Appl. Hum. Sci. 2004;23:93–99. doi: 10.2114/jpa.23.93. [DOI] [PubMed] [Google Scholar]
  • 9.Cha K., Chertow G.M., Gonzalez J., Lazarus J.M., Wilmore D.W. Multifrequency bioelectrical impedance estimates the distribution of body water. J. Appl. Physiol. 1995;79:1316–1319. doi: 10.1152/jappl.1995.79.4.1316. [DOI] [PubMed] [Google Scholar]
  • 10.Groenendaal W., Lee S., van Hoof C. Wearable bioimpedance monitoring: Viewpoint for application in chronic conditions. JMIR Biomed. Eng. 2021;6:e22911. doi: 10.2196/22911. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Sel K., Osman D., Jafari R. Non-invasive cardiac and respiratory activity assessment from various human body locations using bioimpedance. IEEE Open J. Eng. Med. Biol. 2021;2:210–217. doi: 10.1109/OJEMB.2021.3085482. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Zink M.D., König F., Weyer S., Willmes K., Leonhardt S., Marx N., Napp A. Segmental bioelectrical impedance spectroscopy to monitor fluid status in heart failure. Sci. Rep. 2020;10:3577. doi: 10.1038/s41598-020-60358-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Collins C.T., Reid J., Makrides M., Lingwood B.E., McPhee A.J., Morris S.A., Gibson R.A., Ward L.C. Prediction of body water compartments in preterm infants by bioelectrical impedance spectroscopy. Eur. J. Clin. Nutr. 2013;67:S47–S53. doi: 10.1038/ejcn.2012.164. [DOI] [PubMed] [Google Scholar]
  • 14.Piccoli A., Rossi B., Pillon L., Bucciante G. A new method for monitoring body fluid variation by bioimpedance analysis: The RXc graph. Kidney Int. 1994;46:534–539. doi: 10.1038/ki.1994.305. [DOI] [PubMed] [Google Scholar]
  • 15.Nwosu A.C., Mayland C.R., Mason S., Cox T.F., Varro A., Ellershaw J. The association of hydration status with physical signs, symptoms and survival in advanced cancer—The use of bioelectrical impedance vector analysis (BIVA) technology to evaluate fluid volume in palliative care: An observational study. PLoS ONE. 2016;11:e0163114. doi: 10.1371/journal.pone.0163114. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Toso S., Piccoli A., Gusella M., Menon D., Crepaldi G., Bononi A., Ferrazzi E. Bioimpedance vector pattern in cancer patients without disease versus locally advanced or disseminated disease. Nutrition. 2003;19:510–514. doi: 10.1016/S0899-9007(02)01084-5. [DOI] [PubMed] [Google Scholar]
  • 17.Nescolarde L., Piccoli A., Román A., Núñez A., Morales R., Tamayo J., Doñate T., Rosell J. Bioelectrical impedance vector analysis in haemodialysis patients: Relation between oedema and mortality. Physiol. Meas. 2004;25:1271–1280. doi: 10.1088/0967-3334/25/5/016. [DOI] [PubMed] [Google Scholar]
  • 18.Piccoli A., Italian CAPD-BIA Study Group Bioelectric impedance vector distribution in peritoneal dialysis patients with different hydration status. Kidney Int. 2004;65:1050–1063. doi: 10.1111/j.1523-1755.2004.00467.x. [DOI] [PubMed] [Google Scholar]
  • 19.Bozzetto S., Piccoli A., Montini G. Bioelectrical impedance vector analysis to evaluate relative hydration status. Pediatr. Nephrol. 2010;25:329–334. doi: 10.1007/s00467-009-1326-3. [DOI] [PubMed] [Google Scholar]
  • 20.Da Fonseca R.D., Santos P.R., Monteiro M.S., Fernandes L.A., Campos A.H., Borges D.L., Rosa S.S.R.F. Parametric evaluation of impedance curve in radiofrequency ablation: A quantitative description of the asymmetry and dynamic variation of impedance in bovine ex vivo model. PLoS ONE. 2021;16:e0245145. doi: 10.1371/journal.pone.0245145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Barkagan M., Rottmann M., Leshem E., Shen C., Buxton A.E., Anter E. Effect of baseline impedance on ablation lesion dimensions: A multimodality concept validation from physics to clinical experience. Circ. Arrhythm. Electrophysiol. 2018;11:e006690. doi: 10.1161/CIRCEP.118.006690. [DOI] [PubMed] [Google Scholar]
  • 22.Sulkin M.S., Laughner J.I., Hilbert S., Kapa S., Kosiuk J., Younan P., Romero I., Shuros A., Hamann J.J., Hindricks G., et al. Novel measure of local impedance predicts catheter—Tissue contact and lesion formation. Circ. Arrhythm. Electrophysiol. 2018;11:e005831. doi: 10.1161/CIRCEP.117.005831. [DOI] [PubMed] [Google Scholar]
  • 23.Chinitz J.S., Michaud G.F., Stephenson K. Impedance-guided radiofrequency ablation: Using impedance to improve ablation outcomes. Innov. Card. Rhythm. Manag. 2017;8:2868–2873. doi: 10.19102/icrm.2017.081003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Yamaguchi T., Ueno A. Capacitive-coupling impedance spectroscopy using a non-sinusoidal oscillator and discrete-time Fourier transform: An introductory study. Sensors. 2020;20:6392. doi: 10.3390/s20216392. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Takano M., Ueno A. Noncontact in-bed measurements of physiological and behavioral signals using an integrated fabric-sheet sensing scheme. IEEE J. Biomed. Health Inform. 2019;23:618–630. doi: 10.1109/JBHI.2018.2825020. [DOI] [PubMed] [Google Scholar]
  • 26.Elghajiji A., Wang X., Weston S.D., Zeck G., Hengerer B., Tosh D., Rocha P.R.F. Electrochemical impedance spectroscopy as a tool for monitoring cell differentiation from floor plate progenitors to midbrain neurons in real time. Adv. Biol. 2021;5:2100330. doi: 10.1002/adbi.202100330. [DOI] [PubMed] [Google Scholar]
  • 27.Sato H., Nakamura T., Kusuhara T., Kenichi K., Kuniyasu K., Kawashima T., Hanayama K. Effectiveness of impedance parameters for muscle quality evaluation in healthy men. J. Physiol. Sci. 2020;70:53. doi: 10.1186/s12576-020-00780-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Lu F., Wang C., Zhao R., Du L., Fang Z., Guo X., Zhao Z. Review of stratum corneum impedance measurement in non-invasive penetration application. Biosensors. 2018;8:31. doi: 10.3390/bios8020031. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Hernández-Balaguera E., López-Dolado E., Polo J.L. Obtaining electrical equivalent circuits of biological tissues using the current interruption method, circuit theory and fractional calculus. RSC Adv. 2016;6:22312. doi: 10.1039/C5RA24535D. [DOI] [Google Scholar]
  • 30.Münkler P., Gunawardene M.A., Jungen C., Klatt N., Schwarzl J.M., Akbulak R.Ö., Dinshaw L., Hartmann J., Jularic M., Kahle A.K., et al. Local impedance guides catheter ablation in patients with ventricular tachycardia. J. Cardiovasc. Electrophysiol. 2020;31:61–69. doi: 10.1111/jce.14269. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author, A.U., upon reasonable request.


Articles from Sensors (Basel, Switzerland) are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES