Modeling the Effects of Glaucoma Surgery on Intraocular Pressure

Nicholas Farrar 1, David B Yan 2, Mark Johnson 1,3,4,*
PMCID: PMC8595518  NIHMSID: NIHMS1712105  PMID: 34048778

Abstract

Minimally invasive glaucoma surgeries (MIGS) offer an effective way to lower intraocular pressure without inducing extensive trauma to the anterior segment. In order to predict their efficacy, an analytical model of the conventional aqueous humor outflow pathway is developed using a resistor network. The model describes outflow through the normal eye and allows for the effects of geometric changes in the outflow pathway as IOP changes. By selectively removing these resistors, the model can be used to examine and predict the outcomes of several surgical procedures currently used to treat glaucoma. Treatments examined include traditional trabeculectomy, several ab interno methods for trabeculotomy and trabeculectomy, as well as recently developed trabecular stents that bypass the trabecular meshwork and dilate Schlemm canal. The model’s predictions for the efficacy of these procedures generally matched well with the efficacy determined in experimental studies, although it tended to somewhat overestimate the efficacy of these procedures. Matching the model to experimental data indicated that a partial trabeculotomy substantially increases flow to collector channels within that region and approximately 1.5 clock hour past the ends of the trabeculotomized region. Similarly, trabecular bypass stents substantially increase flow to collector channels up to 1.5 clock hour past the open ends of the stent. The resistor model we have developed can be used to predict the efficacy of a variety of MIGS procedures. Circumferential flow in Schlemm canal extends the efficacy of MIGS, but this effect is limited to a few clock hours.

Keywords: model, MIGS, stents, trabeculotomy

1. Introduction

Lowering of intraocular pressure (IOP) is the only established treatment for glaucoma (Weinreb et al., 2014), and pharmacological treatment is usually the first choice for medical therapy. However, for many patients, surgery is required to sufficiently lower IOP so as to either slow or stop the disease process. Traditional glaucoma surgery creates a full thickness passageway between the anterior chamber to the subconjunctival or sub-Tenon’s space, either by partial thickness dissection of the scleral (in a trabeculectomy) or insertion of a silicon tube connected to an outer plate (in a tube shunt). The extensive trauma to the anterior segment and relatively uncontrolled rate of fluid flow directly to the outside of the eye is associated with a high rate of postoperative complications (Gedde et al., 2012). There is a desire to adopt procedures that minimize such trauma including ab interno trabeculotomy, small shunts placed into Schlemm canal, or localized laser ablation (Kaplowitz et al., 2014; Richter and Coleman, 2016). Predicting the potential success of such therapies would aid in their development and adoption.

The extent to which IOP is elevated over episcleral venous pressure is determined primarily by the rate at which aqueous humor is produced by the ciliary body and by the outflow resistance generated as aqueous humor passes through the conventional outflow pathway*. This outflow resistance is generated as the aqueous humor flows sequentially through 1) trabecular meshwork, 2) inner wall of Schlemm canal, 3) Schlemm canal itself, 4) collector channels, 5) aqueous veins, and finally 6) anastomosing back into the episcleral venous system. Under normal conditions, approximately half of the flow resistance is generated by the trabecular meshwork and inner wall of Schlemm canal, and half by the collector channels and aqueous veins, with little flow resistance or pressure drop occurring in Schlemm canal itself (Rosenquist et al., 1989). The elevated IOP characteristic of glaucoma is thought to be caused primarily by increased flow resistance in the trabecular meshwork and inner wall of Schlemm canal (Johnson, 2006).

Glaucoma surgery can change the path taken by aqueous humor through the conventional outflow pathway. Normally, aqueous humor flows through the trabecular meshwork and inner wall of Schlemm canal, with some regions permitting high flow and others permitting less flow (de Kater et al., 1989). Aqueous humor then briefly passes through Schlemm canal and drains through the nearest collector channel. As shown in Fig. 1, a 3-hour trabeculectomy where only the trabecular meshwork was removed would significantly alter this flow distribution. With the aqueous humor now having free access to Schlemm canal, most of the flow will enter the canal through the trabeculectomy and drain through the exposed collector channels. Intuitively, if this free access to collector channels removed the vast majority of the outflow resistance, IOP would drop close to episcleral venous pressure with a localized removal of the trabecular meshwork. However, this is not observed in surgical studies that aim to bypass the trabecular meshwork, and other components of the conventional outflow pathway are likely to play a greater role in generating outflow resistance when the system has been altered surgically. A significant fraction of the flow may travel through Schlemm canal and distribute itself among the other collector channels, with the collector channels closest to the trabeculectomy getting the bulk of this flow.

Figure 1:

Figure 1:

Schematic of aqueous humor outflow pathway after 3-hour trabeculotomy showing flow distribution (more flow is indicated by thicker arrows).

Without a model, predicting the flow distribution and the resulting IOP reduction following MIGS would be challenging even if the flow resistance of all components of this network were constant. But as IOP changes, the flow resistance of Schlemm canal, collector channels and aqueous veins also change (Johnstone and Grant, 1973; Rosenquist et al., 1989). We developed a simple resistor model of the aqueous humor outflow pathway that allows for predictions of the effect of various surgical interventions on outflow resistance (see supplement for software implementation). Our results show that while even a small opening into Schlemm canal can significantly lower IOP, the flow resistance of the collector channels and aqueous veins is sufficiently high as to generate significant circumferential flow through Schlemm canal that limits the potential IOP reduction achievable by procedures that open limited aspects of Schlemm canal. This model could aid in the future development and refinement of angle-based MIGS procedures to optimize their efficacy in reducing IOP. A program implementing our model is currently online and can be accessed at Aqueous Humor Outflow Model.

2. Methods

2.1. Baseline model of eye.

The conventional aqueous outflow pathway is modeled as a network of flow resistors. Three different types of flow resistors are defined, representing the (i) trabecular meshwork and inner wall of Schlemm canal (TM), (ii) Schlemm canal (SC), and (iii) the collector channels and aqueous veins (CCs). For a given IOP or aqueous humor flow rate (QT), the pressure at each node in the system is determined using Kirchoff’s laws for current (flow) and voltage (pressure).

The model of the outflow pathway allows for a variable number of resistors in the network. One resistor is assigned for each of N CCs. Within Schlemm canal, M resistors were placed evenly between every two CCs. The TM is represented as having N × M resistors. The resulting network has N × M nodes in SC with the pressure and flow of each of those nodes designated as Pi and Qi, with i varying from 1 to N × M. An example of the resulting resistor network model with N = 4 and M = 2 is shown in Fig. 2.

Figure 2:

Figure 2:

Example of resistor network with N = 4 CCs and M = 2 SC resistors between each pair of collector channels. Numbers represent Schlemm canal nodes. RTMi and RSCi represent the resistances of the TM and SC resistors at a given node. RCCk represents the resistance of the kth CC resistor. IOP and PV respresent intraocular pressure and episcleral venous pressure, respectively.

The actual resistor network used to perform our baseline simulations is much larger. We used N=30 as a typical number of CCs in the human eye (Moses, 1977). The number of SC resistors between CCs (M=40) was chosen in order to optimize the accuracy of the model and minimize computational cost. Increasing M past 40 yielded minimal increases in accuracy, but significantly increased computational expense.

The resistance of each resistor in the network is designated as Ri,j where i and j are the nodes in the network connected by the resistor. Node 0 is IOP and node N × M + 1 is episcleral venous pressure (Pv). Nodes 1 to N × M are in Schlemm canal. Ri,j is infinite for any two nodes not directly connected to one another. The N × M resistors modeling the trabecular meshwork and inner wall endothelium each have a resistance of

R0,i=NMRTM (1)

where RTM is the total resistance of the trabecular meshwork and inner wall endothelium with i varying from 1 to N × M.

Schlemm canal is modeled as a compliant, porous duct that is elliptical in cross-section (Ethier et al., 2004). For such a duct with width W and height h, the pressure gradient is found as:

dPdx=64μQ(x)πWh3(x) (2)

where x is distance from the midpoint of the canal between two CCs, Q(x) is the flow rate and μ is the viscosity of the aqueous humor (White and Corfield, 2006). By assuming that the height varies linearly between adjacent SC nodes and integrating equation (2) from the ith node to the jth node (see Supplemental Information), the resistance of a given SC resistor is then found as:

Ri,j=PjPiQ(x)=32μ(hi+hj)ΔxπWhi2hj2 (3)

where Δx is the distance between adjacent SC nodes and i and j are the number of the adjacent nodes that the resistor connects.

The height of the canal is assumed to vary as a function of the pressure drop across the TM. Two regimes are considered, depending on whether the canal is being held open primarily by tension in the trabecular meshwork, or by compression of the septae in Schlemm canal (Hoffmann and Dumitrescu, 1971; Johnson and Kamm, 1983). When the canal is held open by the tension in the trabecular meshwork, the height is assumed to vary linearly with the pressure drop (IOPPi) across the trabecular meshwork and inner wall of Schlemm canal:

hi=h0(1IOPPiETM) (4)

where h0 is the undeformed height of the canal, Pi is the pressure in SC, ETM is the modulus of elasticity of the trabecular meshwork and i = 1 to N*M.

Complete canal collapse at higher pressure drops across the inner wall is prevented by support structures in the canal known as septae (Hoffmann and Dumitrescu, 1971). The septae can be thought of as beams between the inner and outer walls of Schlemm canal whose compression opposes canal collapse. The relation governing canal height in this regime is determined by an order of magnitude analysis (see Appendix) to be

hi=hs[KS(IOPPi)]13 (5)

where hs is the height of Schlemm canal at which the septae begin to oppose canal compression and KS is a stiffness parameter characterizing the septae. Recognizing that canal height given by equations (4) and (5) must be the same when hi = hs, we

Ks=ETM(1hsh0) (6)

find that Ks is not an independent parameter and is calculated as

The values of the parameters hs, ETM, and KS were determined by a best fit of the model to a pressure-resistance curve for enucleated human eyes (Brubaker, 1975) using the parameters in Table 1 with Pv = 0 mmHg.

Table 1:

Baseline model parameters

Parameter Description Value Source
N Number of CCs 30 (Moses, 1977)
M Number of SC resistors/nodes between adjacent CCs 40 Optimal resolution
QT Conventional flowrate of aqueous humor 2 μL/min (Brubaker, 1982)
RTM Overall resistance of TM 2 mmHg/μL/min (Rosenquist et al., 1989)
μ Viscosity of aqueous humor 0.007 g/cm/sec (Moses, 1977)
W Width of SC 230 μm (Zhou and Smedley, 2006)
h0 Undeformed height of SC 20 μm (Zhou and Smedley, 2006)
hs Undeformed height of septae 3.0 μm Least-squares fit
ETM Young’s modulus of TM 13 mmHg Least-squares fit
KS Stiffness of septae 11.05 mmHg Equation (6)

The N resistors modeling collector channels were each assumed to have the same flow resistance:

RiM,(NM+1)=NRCC (7)

where RCC is the total resistance of all the CCs and i = 1 to N. Rosenquist et al. (1989) measured outflow resistance after complete trabeculotomy (i.e. only CC resistance remaining) at IOP = 7 mmHg and IOP = 25 mmHg, and we used their values to estimate Rcc with the following relationship:

RCC=2.03mmHgμL/min(0.033minμL)(IOPPv) (8)

The resistances Rij were used to find the conductances Gij (the conductance is the inverse of the resistance such that Gij = 1/Rij). Using this conductance matrix, the following linear system of equations can be constructed using Kirchoff’s current and voltage laws and solved to find the pressure distribution in SC

G(p(k))p(k+1)=q (9)

where G(p) is the matrix of conductances, p is the pressure vector [p0Pv, p1Pv,…, pNM–1Pv, IOPPv]T, and q is the flow rate vector [0,0,… ,0, QT]T. When IOP is specified instead of QT, IOP is moved to the righthand side of the equation, so that the last element of p is removed, and q becomes [−(IOPPv)/R0,0, – (IOPPv)/R0,1,…, – (IOPPv)/R0,NM–1]. Note that the conductance matrix is a function of the pressure vector. This has two consequences. Before the system can be solved the first time, an initial guess for the pressure vector must be made. The value used for this guess has essentially no effect on the final result, but does affect computation time, so it is most efficient to select a reasonable guess. The initial guess used for simulations where IOP is specified (constant pressure mode) is what the pressure in the canal would be assuming that SC offers no resistance to fluid flow, yielding

Pguess=IOP(RCCRTM+RCC) (10)

Since this value is dependent on IOP, an initial guess is also needed for IOP when only QT is specified (constant flow mode). This guess is

IOPguess=QT(RTM+RCC) (11)

The other consequence of G being a function of p is that the linear system in equation (9) must be solved iteratively, recalculating the conductance matrix with each new pressure vector. The indices k and (k + 1) refer to the kth and (k + 1)th iterations of the solution, respectively. Equation (9) is repeatedly solved until p converges on a value. The metric used for determining convergence is

Residual=1IOPi=0NM1(pi(k+1)pi(k))2NM (12)

and convergence is said to have occurred when Residual < 10−4. For most cases, Gaussian elimination is used to solve the system; however, occasionally when the canal is in transition between the two height regimes, Gaussian elimination is unable reach a stable solution, and in such cases, the iterative Gauss-Seidel method is used.

The model parameters used for the baseline model describing the normal human eye are shown in Table 1. The flow rate chosen, Q = 2 μl/min, is the flow through the conventional outflow pathway, and does not include flow through the unconventional pathway or any posteriorly directed flow.

2.2. Model for surgically treated eye.

Several surgical methods for lowering outflow resistance in glaucoma were modeled. Trabeculotomy or trabeculectomy, in which trabecular meshwork and the inner wall of Schlemm canal are either cut or removed in a portion of the eye, respectively, were modeled by removing the TM and SC resistors in the region where the surgery was performed. The effects of varying degrees of trabeculotomy in enucleated eyes were examined using the model at pressures of either 7 or 25 mm Hg and comparing to results of corresponding perfusion experiments by Rosenquist et al. (1989). In addition to traditional trabeculotomy, the model was also compared with several minimally invasive approaches to trabeculotomy and trabeculectomy. The Trabectome (NeoMedix, Tustin, CA) is a device for performing ab interno partial trabeculectomy, which allows the surgeon to ablate between 90° and 120° of the trabecular meshwork through a small corneal incision (Vold, 2011). The effect of the Trabectome was modeled by removing the trabecular meshwork and Schlemm canal resistors in a 90° to 120° region of Schlemm canal. Predicted IOP was determined for a number of trabeculotomies, randomly varying the length from 90 to 120 degrees as well as the location around the canal. These predicted IOPs were then used to calculate a mean and standard deviation. Gonioscopy-assisted transluminal trabeculotomy (GATT) is a minimally invasive method for performing 360° ab interno trabeculotomy (Grover et al., 2014). The TRAB360 (Sight Sciences, Menlo Park, CA) is a device which can perform 360° ab interno trabeculotomy using a method similar to GATT (Sarkisian et al., 2019). Both GATT and the TRAB360 are modeled by removing all TM and SC resistors, leaving only the CCs.

Additionally, three trabecular bypass stents were modeled. The iStent (Glaukos Corp., Laguna Hills, CA), a 1 mm long titanium microstent (Figure 3A) (Hunter et al., 2014); the 8 mm Hydrus Microstent (Ivantis Inc., Irvine, CA), an 8 mm scaffold made from nitinol (Figure 3B) (Grierson et al., 2015; Yuan et al., 2016); and the iStent inject, the smaller, second generation of the iStent (Figure 3C) (Bahler et al., 2012). When the iStent inject is used clinically, almost always two of the stents rather than one are implanted 1-3 clock-hours apart. For our study comparing model predictions to clinical results (Figure 8), two iStent injects were modeled as being 2 clock hours apart from one other. In a study modeling how stent separation affects IOP reduction (Figure 11), the distance between iStent injects was varied from 1-6 clock hours.

Figure 3:

Figure 3:

Stents modeled in study annotated with dimensions given Table 2. (A) First-generation iStent (Magazine, 2018) (B) 8 mm Hydrus Microstent (Gulati et al., 2013), (C) iStent inject (Glaukos).

Figure 8:

Figure 8:

Comparison of model predictions for IOP reduction with results of clinical studies (Fea et al., 2017; Fea et al., 2014; Katz et al., 2015) for trabecular bypass stents. In Katz et al. (2015) (iStent), patients underwent medication washout prior to both the baseline and 12-month IOP measurements. In Fea et al. (2017) (Hydrus), IOP reducing medication was uncontrolled at both the baseline and 12-month measurements. In Fea et al. (2014) (iStent inject), patients underwent medication washout prior to the baseline measurement, however, many patients resumed medication prior to the 12-month measurement. Legend is same as Figure 7.

Figure 11:

Figure 11:

Predicted IOP following implantation of two iStent injects separated by 1-6 clock hours compared to the preoperative baseline of 25.2 mm Hg. Error bars are standard deviations.

Trabecular bypass stents have an opening, referred to as the inlet, that connects the anterior chamber to the lumen of Schlemm canal. This opening is modeled exactly the same as trabeculotomy, i.e. by removing the trabecular meshwork resistors where the stent is placed. Second, as these stents dilate Schlemm canal to well above its natural undeformed height, this part of the stent is modeled by holding the canal (at the locations filled by the stent) at a height determined by the size of the stent. This dilation is reflected in the model by replacing h and W in equation (3) with the corresponding heights and widths listed in Table 2 for each device. In the iStent and iStent inject, the portion of the stent dilating the canal also blocks flow through the trabecular meshwork, so no flow is allowed to enter this part of the canal through the TM resistors. In the Hydrus, however, there are “window” regions, which allow flow to enter through the trabecular meshwork, but still keep Schlemm canal dilated, and “spine” regions which block flow through the trabecular meshwork.

Table 2:

Dimensions of stents

DEVICE Description Value Source
iStent
L Length of the device in SC 1 mm (Hunter et al., 2014)
w Width of the device 120 μm (Katz et al., 2015)
h Height of device in SC 60 μm (Katz et al., 2015)
8mm Hydrus
L Length of device in SC 7.2 mm (Yuan et al., 2016)
Lw Length of window region 1.1 mm (Yuan et al., 2016)
Ls Length of spine region 0.9 mm (Yuan et al., 2016)
Lin Length of inlet spine region 1.1 mm (Yuan et al., 2016)
Nw Number of windows 3 (Yuan et al., 2016)
Aw Area of window region 17553 μm2 (Yuan et al., 2016)
As Area of spine region 22955 μm2 (Yuan et al., 2016)
Ain Area of inlet spine region 29841 μm2 (Yuan et al., 2016)
w Width of device 292 μm FDA product insert
hw Height of window portion 76.5 μm 4Awπw
hs Height of spine portion 100.0 μm 4Asπw
hin Height of inlet spine region 130.1 μm 4Ainπw
iStent inject
L Length of device in SC 230 μm (Glaukos, 2018)
w Diameter of device outlets 50 μm (Glaukos, 2018)
h Same as w for the iStent 50 μm (Glaukos, 2018)
din Diameter of inlet 80 μm (Glaukos, 2018)
hd Height of device in SC 150 μm Analysis of device photo (Glaukos, 2018)

Similar to the modeling studies of the Trabectome, a number of simulations were run randomly varying the location where each stent was placed around the canal, and the resulting IOPs were then used to calculate a mean and standard deviation for the predicted postoperative IOP. The resulting variability represents the variability due to the location of surgical intervention relative to the collector channel ostia.

Both the iStent and iStent inject are bidirectional bypasses, meaning that once aqueous humor enters through the inlet, it can travel in either direction through Schlemm canal. This is reflected in the model by allowing flow to travel in either direction from the inlet. The Hydrus, on the other hand, is unidirectional, only allowing flow to travel in the direction of the scaffold from the inlet.

We excluded from our modeling any clinical studies that included phacoemulsification due to the IOP-lowering effect of cataract surgery (Mansberger et al., 2012).

2.3. Model for Glaucomatous Eye.

In order to accurately compare the model predictions for treatments to the results of clinical studies, baseline IOPs from the studies must be matched by the model. Since the source of increased outflow resistance in primary open-angle glaucoma is generally thought to be in the trabecular meshwork (and/or inner wall of Schlemm canal), the TM resistance was the parameter used to match clinical baseline IOP. Episcleral venous pressure was assumed to be 8 mmHg (Podos et al., 1968) for all studies compared. The values of RTM used for each study and device are shown in Table 3. The aqueous flowrate passing through the outflow pathway was assumed to be 2 μL/min (Brubaker, 1982) for all simulations. The RTM values shown in Table 3 were determined by adjusting RTM so that the model returned the same baseline IOP as was reported in the corresponding clinical study assuming the conditions outlined above.

Table 3:

Baseline resistance for clinical studies.

Treatment Study Mean Baseline
IOP (mmHg)
RTM
(mmHg/μL/min)
Trabectome (Vold, 2011) 23.6 5.58
GATT (Grover et al., 2014) 25.6 6.42
TRAB360 (Sarkisian et al., 2019) 23.7 5.62
iStent (Katz et al., 2015) 25 6.16
8mm Hydrus (Fea et al., 2017) 23.1 5.42
iStent inject (Fea et al., 2014) 25.2 6.25

3. Results

3.1. Model Validation.

Figure 4 shows the best fit of the model to data for flow resistance as a function of pressure in enucleated human eyes (Brubaker, 1975). The small decrease in outflow resistance between 5 and 15 mmHg in the model is due to decreasing CC resistance with increasing IOP; the increasing resistance at higher pressures is due to collapse of Schlemm canal. This small decrease in resistance is not seen in the experiment data likely due to natural variability in the human outflow system, as reflected in the measured standard deviations.

Figure 4:

Figure 4:

Flow resistance of enucleated human eyes (normalized to the mean value at 10 mm Hg) as a function of IOP (Brubaker, 1975). The solid line is the best fit of the baseline model to the data (see text). Error bars are standard errors of the experimental measurements.

The model for canal collapse was further validated by comparing the height distributions in Schlemm canal, shown in Figure 5, with observations of canal collapse at different pressures. Johnstone and Grant (1973) examined Schlemm canal in enucleated human eyes at several levels of IOP. At an IOP of 5 mmHg, they observed the canal to be open, consistent with the model's prediction that the canal is at ~80% of its undeformed height. At 15 mmHg, Johnstone and Grant observed the canal to be in a transition between collapsed and uncollapsed states, this agrees with the model’s prediction that the canal is collapsed down to ~35% of its undeformed height, but is not quite collapsed enough for the septae to begin supporting the canal. Finally, at 30 mmHg, they observed the canal to be fully collapsed, consistent with the model’s prediction that ~70% of the canal is collapsed and supported by the septae at this pressure.

Figure 5:

Figure 5:

Model predictions for the normalized height of Schlemm canal, h(x)/h0 as a function of the distance from a collector channel, by the normalized distance between adjacent collector channels, XCC. The two collector channels are located at x/XCC = 0 and x/XCC = 1. The canal height is shown for an enucleated eye at IOP = 5 mmHg (dot-dashed line), 15 mm Hg (solid line) and 30 mm Hg (dashed line). h0 is the undeformed canal height (at IOP = 0 mmHg).

3.2. Simulation of trabeculotomy and trabeculectomy.

Model predictions of percent reduction of outflow resistance after 1-hour, 4×1-hour, and complete 12-hour trabeculotomy at constant IOPs of 7 mmHg and 25 mmHg were compared with the results of corresponding constant pressure perfusion experiments in enucleated human eyes (Rosenquist et al., 1989). The resulting resistance data is shown in Figure 6. The model predictions are generally good, although they somewhat overpredict the effect of partial trabeculotomy at 25 mmHg. Interestingly, the model predicts almost no difference in resistance between the 4× 1-hour and complete 12-hour trabeculotomies, which is not observed experimentally. This is likely due to the fact that the model does not account for weakening of the trabecular meshwork caused by the procedure that would be expected to increase the resistance in the 4×1-hour trabeculotomy due to increased Schlemm canal collapse.

Figure 6:

Figure 6:

Model predictions (red) of resistance reduction for varying degrees of trabeculotomy in enucleated eyes as compared with experimental results (blue) (Rosenquist et al., 1989). Error bars are standard deviations.

Model predictions for the effects of Trabectome, GATT and TRAB360 procedures on IOP as compared to clinical results are shown in Figure 7. Agreement between the model and clinical results are also generally good, although the model overpredicts the success of the procedures. In these cases, the overpredicted pressure reduction is due, at least in part, to fewer IOP-reducing medications taken by the patients following successful surgery (Grover et al., 2014; Sarkisian et al., 2019; Vold, 2011). As well, the model does not account for other surgical factors such as inflammation, bleeding/hyphema and the release of debris that will affect flow into the collector channels that have been unroofed by the trabeculectomy procedures.

Figure 7:

Figure 7:

Comparison of model predictions for IOP reduction with results of clinical studies (Grover et al., 2014; Sarkisian et al., 2019; Vold, 2011) for ab interno trabeculotomy procedures. In all three clinical studies, IOP reducing medication was uncontrolled at both the baseline and 12-month IOP measurements. Baseline pressures for the clinical studies are shown blue and 12-month post-operative results are shown in green; baselines for the model (red) are the same as for the model as for the clinical studies, and model predictions for effect of the procedures are shown in purple. Error bars are standard deviations.

3.3. Simulation of Stents.

Model predictions for the effects of inserting stents into Schlemm canal are shown in Figure 8. Similar to the predictions of trabecular surgery described above, the model showed good agreement with clinical results 12 months after implantation of the stents but somewhat overpredicted procedure success for the Hydrus. This overprediction is, again, likely due to the fact that patients were taking fewer IOP reducing medications at 12 months than at the baseline measurement (Fea et al., 2017). The agreement between the model predictions and clinical results was good for the iStent. For the iStent injects, the model slightly underpredicts the success of the procedure. This underprediction is likely due to the fact that patients in this study underwent medication washout for the baseline IOP measurement, but not for the IOP measurement at 12 months (Fea et al., 2014).

4. Discussion

The goal of minimally invasive glaucoma surgery is to adequately lower IOP while minimizing surgical trauma to the anterior segment of the eye. Accurate estimation of the effect of surgical procedures on outflow resistance can greatly facilitate achievement of this goal. The model we have developed treats the outflow pathway as a network of resistors and uses circuit theory to predict the result of removing various resistors from the network.

Predictions from the model gave generally good agreement with results from a variety of surgical procedures although the predicted reductions in patients tended to be greater than those seen clinically (Figs 7, 8). A likely explanation for at least part of this discrepancy is that medication was not a controlled variable in the clinical trials of the ab interno trabeculotomy. As patients' pressure control improved following surgery, their pressure reducing medication was reduced, likely offsetting some of the IOP reduction from the surgery (Grover et al., 2014; Sarkisian et al., 2019; Vold, 2011); simulations of reductions in topical aqueous suppressants using the resistor model confirm this speculation (data not shown). It is also important to note that the clinical results that we compared our model predictions to are for a time point 12 months after the procedure during which time healing would have occurred that could have lessened the effect of the procedure.

All of the IOP reducing procedures modeled here involve some form of opening into Schlemm canal, thus allowing direct communication between the anterior chamber and the canal. If Schlemm canal had no resistance to circumferential flow, then only a small opening in the inner wall would be required to reduce the outflow resistance to that of the collector channels. However, due to the flow resistance that the canal presents, procedures that opened only part of the canal (limited trabeculotomy or use of stents) fall short of this result. Figure 9 shows the predicted flow distribution through the collector channels after a 1-hour trabeculotomy at constant flow for a glaucomatous eye (RTM increased from normal eye by 3-fold), with substantially increased flow to collector channels within approximately 1.5 clock hours of the trabeculotomy site, while those further away carry equal or less flow than they would without surgery.

Figure 9:

Figure 9:

Distribution of flow through collector channels with 1-hour trabeculotomy (marked by capped line) in a glaucomatous eye with RTM = 6 mmHg/μL/min and a constant flow of 2 μL/min.

Trabecular bypass stents lower outflow resistance by providing direct access from the anterior chamber to Schlemm canal as well as dilating parts of Schlemm canal to well above its undeformed height. Dilation of Schlemm canal effectively eliminates circumferential resistance in the region occupied by the stent. Similar to trabeculotomy, the collector channels in the region occupied by a stent as well as approximately 1.5 clock hour past the open ends of a stent carry the bulk of the flow, while those further away carry less flow.

Our model predicted a somewhat larger pressure reduction than was seen clinically (at 12 months) for the Hydrus (Figure 8). As with the ab interno trabeculotomies, this discrepancy can be partly explained by the fact that medication was not a controlled variable in this clinical study of the Hydrus (Fea et al., 2017). The model predictions matched well for the iStent. For the iStent injects, the model slightly underpredicted the IOP reduction seen clinically (Figure 8). Similar to the Hydrus and ab interno trabeculotomies, medication is a confounding variable in this clinical study of the iStent inject. Patients underwent medication washout prior to their baseline IOP measurement, however many patients resumed some medication prior to their IOP measurement at 12 months, leading to further reduction of IOP (Fea et al., 2014). This underprediction may also be due to another mechanism by which the stents could reduce outflow resistance. By expanding Schlemm canal to well above its undeformed height, resistance to circumferential flow is effectively eliminated not only within the stent but likely beyond the end of the stent, such that collector channels in this dilated region can carry additional flow. This is consistent with observations from scanning electron microscopy studies that Schlemm canal is dilated past the region dilated by the stent (Bahler et al., 2012; Johnstone et al., 2014), although this effect was not included in the model results presented. This dilation would have the largest effect for the iStent inject, because two stents are placed into the canal.

The model predicts a 4×1-hour trabeculotomy to be equally effective as complete trabeculotomy, in contrast to experimental results indicating a meaningful difference in resistance between the two procedures. This is likely due to the model not accounting for any meshwork weakening effects that the procedure might have, which would increase the resistance in the 4×1-hour trabeculotomy (Figure 6). The 4x1-hour trabeculotomy would create eight points of direct communication between the anterior chamber and Schlemm canal, but these openings could be partly obstructed by debris and collapse of the canal structure, thus leading to the over-prediction of the resistance reduction by the model. Similarly, the model prediction (Figure 7) for pressure reduction following use of the Trabectome, which removes a segment of the trabecular meshwork, was greater than seen clinically, although this was likely due in large part to reduction in pressure-lowering medications take by the patients following successful pressure reduction (Vold, 2011). Future work could model the reduction in IOP when combining a MIGS procedure with reduction in medications by altering the rate of aqueous production in the model predictions. Topical aqueous suppressants usually reduce IOP (and presumably aqueous production) by 20-25% when used as monotherapy, but when used in combination the effect is not additive. As such, the reduction in aqueous production when reducing medical therapy from multiple drugs is harder to model accurately.

When the trabecular meshwork is fully removed in a complete 360° trabeculotomy (GATT and TRAB360), the model predicts that the resistance to outflow should be reduced to that of the collector channels. However, the model overestimated the IOP reduction for these ab interno trabeculotomy procedures (Figure 7) suggesting that these procedures do not eliminate all trabecular meshwork and Schlemm canal flow resistance. This can again be partly explained by the reduction in pressure-lowering medications, but another consideration is the mechanism of action for GATT trabeculotomy, which is to tear open the trabecular meshwork rather than remove it. The flap of trabecular meshwork left behind after surgery is commonly referred to as a trabecular shelf (Grover et al., 2014). Clinically, the presence of a trabecular shelf could be associated with a positive surgical outcome, as the flap of trabecular meshwork may attach to the iris and hold the canal open; however, the presence of the torn trabecular meshwork debris after GATT may add flow resistance by blocking the collector channels.

In a previous study modeling trabecular bypass stents, Yuan et al. (2016) argued that a further mechanism by which Schlemm canal stents lower outflow resistance is by dilation of the collector channel ostia. While this assumption worked well for the 8mm Hydrus, with their model predicting a 35% IOP reduction from a 25 mmHg baseline similar to the 28% reduction from a 23.1 mmHg baseline found in a clinical study by Fea et al. (2017), they greatly underestimated the IOP reduction brought on by the iStent, predicting only a 16% reduction from a 25 mmHg baseline, as compared to a 40% reduction seen in a clinical study (Katz et al., 2015). Collector channel ostia have large openings (Bentley et al., 2016), and so it is hard to understand how dilation at just the ostia would significantly alter their resistance. Furthermore, scanning electron microscopy studies have found that Schlemm canal scaffolds like the 8mm Hydrus may even have the potential to occlude some collector channel ostia, rather than dilate them (Johnstone et al., 2014).

We view the utility of the model we have developed not primarily as one of verifying the results of previous experimental studies, but instead of prospectively predicting the usefulness of procedures for lowering outflow resistance. To demonstrate its utility, we performed three hypothetical analyses: first we consider the progressive lowering of IOP that would be expected if one to four iStent injects were inserted into Schlemm canal 90° apart; as a second example, we investigate how the IOP reduction from two iStent injects is affected by the separation between the two devices implanted; finally, we ask how much the Hydrus efficiency would be increased if it were made bi-directional. The IOP reductions reported in these hypothetical scenarios use the baseline IOPs from the iStent inject and Hydrus clinical studies, respectively. Note that these are only model predictions, as we are unaware of studies that have investigated these questions experimentally.

Predictions of the IOP reduction when one, two, three, and four iStent injects are implanted 90° apart is shown in Figure 10. IOP is reduced the most by the first device implanted, and the addition of each new device lowers IOP further, but to a lesser degree.

Figure 10:

Figure 10:

Comparison of predicted IOP using one, two, three, and four iStent injects inserted into Schlemm canal 90° apart, from a baseline pressure of 25 mm Hg. Error bars are standard deviations.

Figure 11 shows the IOP reduction when two iStent injects are implanted 1-6 clock hours apart from each other. IOP reduction is improved by each addition clock hour of separation; however, the impact on IOP is minimal beyond 3 clock hours of separation. This is a reflection of the findings described above that stents increase flow through collector channels up to 1.5 clock hours beyond the region occupied by the stent. When the stents are placed within 3 clock hours of each other, the 1.5 clock hour regions overlap, decreasing the number of collector channels which carry excess flow. Separating the stents by three or more clock hours ensures that there is no overlap of the regions of Schlemm canal affected by the two stents.

Figure 12 shows predictions of the potential increased pressure reduction of a bi-directional Hydrus. Making the Hydrus bi-directional decreases IOP by an additional 5% as compared with a unidirectional Hydrus.

Figure 12:

Figure 12:

Comparison of predicted IOP between the unidirectional Hydrus stent with a hypothetical bi-directional version of the Hydrus stent from a baseline of 23 mm Hg. Error bars are standard deviations.

The effect on outcome of the relative location of collector channels is another parameter that could be examined using our model. An optimally located trabeculotomy might be performed on the region of Schlemm canal containing the highest concentration of patent collector channel entrances. Similarly, an optimally placed stent maximizes the number of patent collector channel entrances in the stent region and surrounding 2-3 clock hours. The model we have developed could be used to predict where these optimal locations might be. In order to achieve this, recent advances in imaging technology for visualizing Schlemm canal and collector channels might be utilized (Ren et al., 2011). In all of the simulations discussed here, a uniform collector channel distribution was used, however in the human eye, collector channels have neither uniform size nor location (Hann et al., 2011). IOP reductions from both ab interno trabeculotomies and trabecular bypass stents in an eye with nonuniform collector channels were determined and are included in the Supplementary Information. Altering the collector channel distribution had minimal effect on the average IOP reductions, but greatly increased the variability due to the location of the surgical interventions.

The aqueous outflow human model developed and validated in this study for surgical alterations to the conventional outflow pathway at the level of the trabecular meshwork can guide the development of new surgical devices and procedures for the treatment of glaucoma. Implants can be customized according to the desired degree of IOP reduction either through the number of devices implanted or their circumferential spacing in the angle. Procedures that remove trabecular meshwork can be customized in terms of the extent of tissue removal to balance the targeted IOP reduction with the potential for complications resulting from excessive tissue disruption. This modeling will optimize procedures and improve patient outcomes in glaucoma surgery that alters the aqueous outflow system.

Supplementary Material

1

Highlights.

  • Novel resistor network model developed for aqueous outflow system

  • Partial trabeculotomy increases flow to collector channels near the surgery

  • Trabecular bypass stents increase flow to collector channels near the surgery

Acknowledgements:

We gratefully acknowledge useful discussions with Prof. Roger Kamm of MIT.

Funding:

This work was supported by NIH [1R01EY019696].

Abbreviations and Acronyms

IOP

intraocular pressure

TM

trabecular meshwork and inner wall of Schlemm canal

SC

Schlemm canal

CC

collector channel or aqueous vein

GATT

Gonioscopy-assisted transluminal trabeculotomy

MIGS

minimally invasive glaucoma surgery

Appendix A

Relationship between canal height and IOP when the canal is collapsed.

In enucleated human eyes, the total outflow resistance has been observed to vary linearly with IOP(Brubaker, 1975) for IOP greater than approximately 15 mmHg, at which point Schlemm canal is observed to be partially collapsed in enucleated eyes (with zero episcleral venous pressure)(Johnstone and Grant, 1973). We assume that in this pressure range, the increasing flow resistance is largely due to progressively increasing Schlemm canal collapse. As such, the resistance in Schlemm canal (RSC) must vary linearly with IOP.

RSCIOPconstant (A4)

Equation (3) shows that RSC varies with respect to canal height as

RSC1h3, (A5)

and thus we conclude that:

1h4hIOPconstant (A6)

or,

IOP1h3 (A7)

This relationship applies to enucleated eyes in which episcleral venous pressure is zero. In live, eyes, this relationship become

IOPPSC1h3 (A8)

Solving for h and by introducing the constants hs and KS, equation (5) results.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

*

Approximately 10% (range: 0-14%) of aqueous humor outflow passes through an alternate pathway known as the unconventional outflow pathway (Johnson et al., 2017). We do not further consider that pathway here.

References

  1. Bahler CK, Hann CR, Fjield T, Haffner D, Heitzmann H, Fautsch MP, 2012. Second-generation Trabecular Meshwork Bypass Stent (iStent inject) Increases Outflow Facility in Cultured Human Anterior Segments. American Journal of Ophthalmology 153, 1206–1213. [DOI] [PubMed] [Google Scholar]
  2. Bentley MD, Hann CR, Fautsch MP, 2016. Anatomical Variation of Human Collector Channel Orifices. Investigative ophthalmology & visual science 57, 1153–1159. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Brubaker RF, 1975. The effect of intraocular pressure on conventional outflow resistance in the enucleated human eye. undefined. [PubMed] [Google Scholar]
  4. Brubaker RF, 1982. The flow of aqueous humor in the human eye. Trans Am Ophthalmol Soc 80, 391–474. [PMC free article] [PubMed] [Google Scholar]
  5. de Kater AW, Melamed S, Epstein DL, 1989. Patterns of Aqueous Humor Outflow in Glaucomatous and Nonglaucomatous Human Eyes: A Tracer Study Using Cationized Ferritin. Archives of Ophthalmology 107, 572–576. [DOI] [PubMed] [Google Scholar]
  6. Ethier CR, Read AT, Chan D, 2004. Biomechanics of Schlemm's canal endothelial cells: influence on F-actin architecture. Biophys J 87, 2828–2837. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Fea AM, Ahmed IIK, Lavia C, Mittica P, Consolandi G, Motolese I, Pignata G, Motolese E, Rolle T, Frezzotti P, 2017. Hydrus microstent compared to selective laser trabeculoplasty in primary open angle glaucoma: one year results. Clinical & Experimental Ophthalmology 45, 120–127. [DOI] [PubMed] [Google Scholar]
  8. Fea AM, Belda JI, Rękas M, Jünemann A, Chang L, Pablo L, Voskanyan L, Katz LJ, 2014. Prospective unmasked randomized evaluation of the iStent inject (®) versus two ocular hypotensive agents in patients with primary open-angle glaucoma. Clin Ophthalmol 8, 875–882. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Gedde SJ, Schiffman JC, Feuer WJ, Herndon LW, Brandt JD, Budenz DL, 2012. Treatment Outcomes in the Tube Versus Trabeculectomy (TVT) Study After Five Years of Follow-up. American Journal of Ophthalmology 153, 789–803.e782. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Glaukos, iStent inject, in: GLKOS-15559-Single-iStent.jpg (Ed.) Glaukos, glaukos.com. [Google Scholar]
  11. Glaukos, 2018. GLAUKOS ® CORPORATION ISTENT inject ® TRABECULAR MICRO-BYPASS SYSTEM PMA P170043. [Google Scholar]
  12. Grierson I, Saheb H, Kahook MY, Johnstone MA, Ahmed IIK, Schieber AT, Toris CB, 2015. A Novel Schlemm's Canal Scaffold: Histologic Observations, Journal of Glaucoma, pp. 460–468. [DOI] [PubMed] [Google Scholar]
  13. Grover DS, Godfrey DG, Smith O, Feuer WJ, Montes de Oca I, Fellman RL, 2014. Gonioscopy-Assisted Transluminal Trabeculotomy, Ab Interno Trabeculotomy: Technique Report and Preliminary Results. Ophthalmology 121, 855–861. [DOI] [PubMed] [Google Scholar]
  14. Gulati V, Fan S, Hays CL, Samuelson TW, Ahmed IIK, Toris CB, 2013. A Novel 8-mm Schlemm's Canal Scaffold Reduces Outflow Resistance in a Human Anterior Segment Perfusion Model. Investigative Opthalmology & Visual Science 54, 1698. [DOI] [PubMed] [Google Scholar]
  15. Hann CR, Bentley MD, Vercnocke A, Ritman EL, Fautsch MP, 2011. Imaging the aqueous humor outflow pathway in human eyes by three-dimensional micro-computed tomography (3D micro-CT). Experimental Eye Research 92, 104–111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Hoffmann F, Dumitrescu L, 1971. Schlemm’s Canal under the Scanning Electron Microscope. Ophthalmic Research 2, 37–45. [Google Scholar]
  17. Hunter KS, Fjield T, Heitzmann H, Shandas R, Kahook MY, 2014. Characterization of micro-invasive trabecular bypass stents by ex vivo perfusion and computational flow modeling. Clinical ophthalmology (Auckland, N.Z.) 8, 499–506. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Johnson M, 2006. ‘What controls aqueous humour outflow resistance?’. Experimental Eye Research 82, 545–557. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Johnson M, Kamm RD, 1983. The role of Schlemm's canal in aqueous outflow from the human eye. Investigative Ophthalmology & Visual Science 24, 320–325. [PubMed] [Google Scholar]
  20. Johnson M, McLaren JW, Overby DR, 2017. Unconventional aqueous humor outflow: A review. Experimental Eye Research 158, 94–111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Johnstone MA, Grant WG, 1973. Pressure-dependent changes in structures of the aqueous outflow system of human and monkey eyes. American journal of ophthalmology 75, 365–383. [DOI] [PubMed] [Google Scholar]
  22. Johnstone MA, Saheb H, Ahmed IIK, Samuelson TW, Schieber AT, Toris CB, 2014. Effects of a Schlemm canal scaffold on collector channel ostia in human anterior segments. Experimental Eye Research 119, 70–76. [DOI] [PubMed] [Google Scholar]
  23. Kaplowitz K, Schuman JS, Loewen NA, 2014. Techniques and outcomes of minimally invasive trabecular ablation and bypass surgery. The British journal of ophthalmology 98, 579–585. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Katz LJ, Erb C, Carceller GA, Fea AM, Voskanyan L, Wells JM, Giamporcaro JE, 2015. Prospective, randomized study of one, two, or three trabecular bypass stents in open-angle glaucoma subjects on topical hypotensive medication. Clinical ophthalmology (Auckland, N.Z.) 9, 2313–2320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. Magazine MPO, 2018. Study: Glaukos iStent Trabecular Micro-Bypass Stent Reduces IOP, Lowers Medication; Burden. [Google Scholar]
  26. Mansberger SL, Gordon MO, Jampel H, Bhorade A, Brandt JD, Wilson B, Kass MA, 2012. Reduction in intraocular pressure after cataract extraction: the Ocular Hypertension Treatment Study. Ophthalmology 119, 1826–1831. [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Moses RA, 1977. The effect of intraocular pressure on resistance to outflow. Survey of Ophthalmology 22, 88–100. [DOI] [PubMed] [Google Scholar]
  28. Podos SM, Minas TF, Macri FJ, 1968. A New Instrument to Measure Episcleral Venous Pressure: Comparison of Normal Eyes and Eyes With Primary Open-Angle Glaucoma. JAMA Ophthalmology 80, 209–213. [DOI] [PubMed] [Google Scholar]
  29. Ren J, Gille HK, Wu J, Yang C, 2011. Ex Vivo Optical Coherence Tomography Imaging of Collector Channels with a Scanning Endoscopic Probe. Investigative Ophthalmology & Visual Science 52, 3921–3925. [DOI] [PubMed] [Google Scholar]
  30. Richter GM, Coleman AL, 2016. Minimally invasive glaucoma surgery: current status and future prospects. Clinical ophthalmology (Auckland, N.Z.) 10, 189–206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Rosenquist R, Epstein D, Melamed S, Johnson M, Grant WM, 1989. Outflow resistance of enucleated human eyes at two different perfusion pressures and different extents of trabeculotomy. Current eye research 8, 1233–1240. [DOI] [PubMed] [Google Scholar]
  32. Sarkisian SR, Mathews B, Ding K, Patel A, Nicek Z, 2019. 360° ab-interno trabeculotomy in refractory primary open-angle glaucoma. Clinical ophthalmology (Auckland, N.Z.) 13, 161–168. [DOI] [PMC free article] [PubMed] [Google Scholar]
  33. Vold SD, 2011. Ab interno trabeculotomy with the trabectome system: what does the data tell us? International ophthalmology clinics 51, 65–81. [DOI] [PubMed] [Google Scholar]
  34. Weinreb RN, Aung T, Medeiros FA, 2014. The Pathophysiology and Treatment of Glaucoma: A Review. JAMA 311, 1901–1911. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. White FM, Corfield I, 2006. Viscous fluid flow. McGraw-Hill; New York. [Google Scholar]
  36. Yuan F, Schieber AT, Camras LJ, Harasymowycz PJ, Herndon LW, Rand Allingham R, 2016. Mathematical modeling of outflow facility increase with trabecular meshwork bypass and schlemm canal dilation. Journal of Glaucoma 25, 355–364. [DOI] [PubMed] [Google Scholar]
  37. Zhou J, Smedley GT, 2006. Trabecular bypass: Effect of Schlemm canal and collector channel dilation. Journal of Glaucoma 15, 446–455. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1

RESOURCES