Highlights
-
•
Identification and quantitation of fentanyls is needed in the clinical practice setting.
-
•
A UPLC-MS/MS method has been validated in plasma and whole blood.
-
•
Application of the method in medical examiner casework shows a range of fentanyls involved in the fatalities including 4-ANPP and beta-hydroxy fentanyl.
-
•
Application in pharmaceutical administration of fentanyl shows no evidence of the fentanyl precursor (4-ANPP) and the presence of beta-hydroxy fentanyl.
-
•
Evidence points to beta-hydroxy fentanyl as a metabolite of fentanyl and 4-ANPP associated with illicit fentanyl and fentanyl analog use.
Abbreviations: ACN, Acetonitrile; 4-ANPP, 4-Anilino-N-phenethylpiperidine (despropionyl-fentanyl); β-OH fentanyl, beta-hydroxy fentanyl; LC-MS, liquid chromatography-mass spectrometry; LLOQ, lower limit of quantitation; MRM, multiple reaction monitoring; PFBF, para-fluorobutyryl fentanyl; UPLC-MS/MS, ultra-performance liquid chromatography-tandem mass spectrometry
Keywords: Fentanyl, Fentanyl analogs, Fentanyl metabolites, Fentanyl precursors, Quantitative UPLC-MS/MS analysis, Clinical and postmortem casework applications
Abstract
The opioid crisis is linked to an increased misuse of fentanyl as well as fentanyl analogs that originate from the illicit drug market. Much of our current understanding of fentanyl and fentanyl analog use in our communities comes from postmortem toxicology findings. In the clinical settings of addiction medicine and pain management, where the opioid abuse potential is high, the use of fentanyl, as well as specific fentanyl analogs, may be underestimated due to limited plasma testing and limited availability of assays with suitable analytical sensitivity and selectivity to detect misuse of fentanyls. We report plasma and blood assays for 17 fentanyls (these include fentanyl, fentanyl analogs, fentanyl metabolites and synthetic precursors) in clinical, and medical examiner, casework. A mixed-mode solid phase extraction of diluted plasma or precipitated blood was optimized for maximum recovery of the fentanyls with minimized matrix effects. Analysis was performed using a Waters ACQUITY UPLC I-Class interfaced with a Waters Xevo TQ-S micro tandem quadrupole mass spectrometer. Method parameters were optimized and validated for precision, accuracy, carryover, linearity and matrix effects. Application studies were performed in postmortem blood obtained in 44 fentanyl-related fatalities and in serial plasma samples from 18 surgical patients receiving intravenous fentanyl therapy while undergoing parathyroidectomy. Fentanyls found in postmortem cases included fentanyl, norfentanyl, despropionyl-fentanyl (4-ANPP), beta-hydroxy fentanyl (β-OH fentanyl), acetyl fentanyl, acetyl norfentanyl, methoxyacetyl fentanyl, furanyl fentanyl, cyclopropyl fentanyl, and para-fluorobutyryl fentanyl, with fentanyl, norfentanyl, 4-ANPP and β-OH fentanyl predominating in frequency. Fentanyl concentrations ranged from 0.2 to 56 ng/mL and fentanyl was nearly always found with 4-ANPP, norfentanyl and β-OH fentanyl. Concentrations of other fentalogs ranged from <1 to 84 ng/mL (extrapolated). In the surgical cases, fentanyl was detected and quantified along with norfentanyl and β-OH fentanyl, but without detection of 4-ANPP in any of the samples. The association and relative concentrations of β-OH fentanyl, fentanyl and norfentanyl in the postmortem and clinical studies indicated a metabolic, rather than an illicit, source of β-OH fentanyl.
1. Introduction
On a global level, overdose deaths due to opioids have dramatically increased in the past several years [1], [2], [3]. In North America, the situation remains particularly challenging; while deaths due to heroin and other natural and semi-synthetic opioids have increased substantially, deaths due to synthetic opioids such as fentanyl and fentanyl analogs (fentanyls) increased three-fold from 2015 to 2017 and 10-fold from 2013 to 2017 [3]. The increased confiscations of fentanyl and fentanyl analogs by law enforcement agencies during this period points to illicit production as a major factor in fueling the opioid crisis [4], [5]. These data also point to the growing need for identification and quantitation of the fentanyls in forensic as well as clinical toxicology.
Sensitive and selective identification and quantitation of a growing number of structurally similar fentanyls in biological matrices is analytically challenging. In medical examiner casework and clinical toxicology, where urine and postmortem blood are often the primary drug test specimens, significant progress has been made in the development and application of definitive methods for the analysis of the fentanyls [6], [7], [8], [9]. As a result, many fentanyl-related, as well as fentanyl analog-related fatalities, have been reported along with methods of analysis of these agents in forensic casework [10], [11], [12], [13], [14], [15], [16], [17]. Methods for alternative matrices, such as oral fluid, have also been reported recently [18], [19]. In addition to the analytical challenge is a lack of certainty in interpretation regarding the metabolic versus synthetic-byproduct origin of a number of fentanyls found in postmortem cases. It is known for example, that 4-ANPP is a byproduct of fentanyl synthesis and may be present as an impurity in preparations. However, it has also been proposed as a minor inactive metabolite of fentanyl and some other fentanyl analogs [20], [21], [22], [23], [24], consequently the contribution of production or metabolism to the 4-ANPP levels measured in postmortem blood remains a question. In addition, while it is well established that norfentanyl is a major metabolite of fentanyl, hepatic microsomal studies suggest that hydroxy fentanyl may also originate from metabolism of fentanyl [21], [25]. Clinical measurement of 4-ANPP and β-OH fentanyl in fentanyl treated patients is limited but could add to our knowledge of a metabolic versus illicit source of these agents. Therefore, in addition to postmortem blood testing is the need for plasma determination of the fentanyls by sensitive and selective methods that will allow clinical monitoring of the high opioid risk population and may further improve our understanding of the fentanyls.
We describe methods for the analysis of fentanyls in plasma and blood, with subsequent application to 44 fatalities involving the fentanyls and to sequential plasma samples from 18 surgical patients receiving intravenous fentanyl therapy while undergoing parathyroidectomy.
2. Materials and methods
2.1. Reference material and reagents
The fentanyls (acetyl norfentanyl, norfentanyl, norcarfentanil, remifentanil acid, remifentanil, acetyl fentanyl, 4-ANPP, fentanyl, furanyl fentanyl, 3-methylfentanyl, carfentanil, methoxyacetyl fentanyl, butyryl fentanyl, para-fluorobutyryl fentanyl, sufentanil) and the internal standards (acetyl norfentanyl-13C6, norfentanyl-d5, acetyl fentanyl-13C6, fentanyl-d5, sufentanil-d5) were obtained from Cerilliant (Round Rock, TX, USA). β-OH fentanyl and cyclopropyl fentanyl were obtained from Cayman Chemical (Ann Arbor, Michigan, USA). Ammonium acetate (trace metal grade), strong ammonia solution (28–30%), ammonium formate (trace metal grade), zinc sulfate heptahydrate, formic acid (Optima grade, 99%), and 85% phosphoric acid solution were obtained from Fisher Scientific (Waltham, MA). LC-MS grade acetonitrile (ACN) and methanol were also from Fisher Scientific. A stock standard of the fentanyls was prepared at analyte concentrations of 10 µg/mL in methanol. Internal standard stock solutions were also prepared in methanol at a concentration of 2 µg/mL. A combined working internal standard solution (1 ng/mL) was prepared daily in an aqueous solution containing 0.1 M zinc sulfate and 0.1 M ammonium acetate for the whole blood analysis. For the plasma analysis, a working internal standard solution (0.2 µg/mL) was prepared in 4% aqueous phosphoric acid. Human whole blood and plasma for quality control pool preparations were obtained from Lampire Biological Supply (Pipersville, PA, USA).
2.2. Postmortem and clinical samples
Discarded postmortem whole blood samples from fentanyl and fentanyl analog-related fatalities were initially collected under medical examiner orders and tested by the Forensic Toxicology Laboratory at the Albany Medical Center (Albany, NY) between December 2018 and August 2019. After all medical examiner testing was completed, in accordance with the laboratory’s forensic protocol, the samples were de-identified and tested in the study.
Discarded, de-identified plasma samples used in the study were initially collected intra-operatively under physician orders for the measurement of parathyroid hormone levels from patents undergoing minimally invasive parathyroidectomy at the Albany Medical Center Hospital and Albany Medical Center South Clinical Campus. After the measurement of parathyroid hormone was complete, the samples were used for this current study. The study was performed with review and approval by the Albany Medical College Institutional Review Board.
Subjects received 50–150 µg of intravenous fentanyl upon induction of anesthesia, with further administration as needed for the care of the patients. A baseline specimen was collected prior to the initial administration of fentanyl in all but case 17 where administration occurred before the baseline sample was drawn. Following the requested clinical testing and after sample discard, the samples were de-identified and stored frozen until analysis. Each patient had between 2 and 6 specimens drawn after the initial fentanyl administration; samples were collected intra-operatively (approximately 1–2 h duration) and therefore represent acute administration conditions.
2.3. Blood assay
Analyte negative blood bank blood (300 µL) was added to 300 µL of internal standard reagent containing 0.1 M zinc sulfate and ammonium acetate. Samples were vortex-mixed and allowed to sit for 5 min. Six-hundred microliters of ice-cold ACN:methanol (50:50 v/v) were then added and the samples were briefly vortex-mixed and allowed to sit for a further 5 min before centrifugation for 10 min at 7000×g. Four-hundred microliters of the supernatant were diluted with 1 mL of 4% phosphoric acid and then added, in two aliquots, to a Waters Oasis PRiME MCX µElution plate (Waters Corp, Milford, MA, USA). After loading the plate, the wells were washed with 200 µL of 0.1 M ammonium formate containing 2% formic acid, followed by 200 µL of methanol. Samples were eluted with two 25 µL volumes of ACN:methanol (50:50 v/v) containing 5% strong ammonia solution. All samples were then diluted with 50 µL of water:ACN:formic acid (97:2:1 v/v/v). Five microliters of the diluted eluates were injected onto the UPLC-MS/MS system. Calibration curves for whole blood samples ranged from 50 to 20000 pg/mL. Low, medium, and high quality control pools were prepared at combined analyte concentrations of 75, 2500, and 15000 pg/mL, respectively.
Blood extracts were analyzed on a Waters TQ-S micro tandem mass spectrometer (Waters Corp) equipped with a Waters ACQUITY I-Class UPLC system. Analytical separation was performed using a Waters BEH C18 column (1.7 µm, 2.1 × 100 mm) together with a solvent gradient of mobile phase A (MPA) comprising 0.1% formic acid in water, and mobile phase B (MPB), comprising 0.1% formic acid in ACN. The chromatographic flow rate was 0.6 mL/minute. The gradient started at 15% MPB, increased over 2 min to 55% MPB and then to 90% MPB over the next 0.5 min prior to returning to initial conditions. Electrospray analysis was performed in positive ionization mode with a capillary voltage at 1.5 kV. Nitrogen was used as the desolvation gas at 500 °C with a flow of 1000 L/hr and as a cone gas at a flow of 150 L/hr. Mass spectrometer parameters and retention times of the fentanyls and internal standards are listed in Table 1. Phospholipid content was profiled by performing a precursor ion scan of m/z 184.2, over the range m/z 400–900 at a cone voltage of 30 V and a collision energy of 30 eV. Compound identification was based upon retention time tolerances of ± 2% and product ion ratios were based on the draft OSAC guidelines for mass spectral data acceptance [26], as defined below:
| Relative abundance | Acceptance criteria |
|---|---|
| Greater than 50% | 20% |
| 20–50% | 25% |
| Less than 20% | 30% |
Table 1.
Individual MS parameters of the fentanyls and internal standards analyzed in postmortem whole blood samples. Retention times (R.T.) correspond to the BEH C18 column. The first product ion is the quantitative transition.
| R.T. | Name | M + H+ | Product Ions | Cone voltage (V) | Collision energy (eV) | Internal standard |
|---|---|---|---|---|---|---|
| 0.68 | Acetyl Norfentanyl | 219.1 | 84.1 136.1 |
4 4 |
18 20 |
Acetyl Norfentanyl-13C6 |
| 0.67 | Acetyl Norfentanyl-13C6 | 225.1 | 84.1 | 4 | 18 | |
| 0.91 | Norfentanyl | 233.1 | 84.1 150.1 |
5 5 |
15 15 |
Norfentanyl-d5 |
| 0.90 | Norfentanyl-d5 | 238.2 | 84.1 | 5 | 15 | |
| 1.03 | Norcarfentanil | 291.1 | 231.2 259.2 |
5 5 |
14 10 |
Norfentanyl-d5 |
| 1.06 | Remifentanil acid | 363.2 | 214.1 113.0 |
5 5 |
30 18 |
Acetyl Fentanyl-13C6 |
| 1.18 | Remifentanil | 377.2 | 113.0 228.1 |
5 5 |
30 20 |
Acetyl Fentanyl-13C6 |
| 1.24 | Methoxyacetyl Fentanyl | 353.2 | 188.1 105.1 |
10 10 |
22 37 |
Acetyl Fentanyl-13C6 |
| 1.29 | Acetyl Fentanyl | 323.1 | 188.1 105.0 |
5 5 |
34 22 |
Acetyl Fentanyl-13C6 |
| 1.28 | Acetyl Fentanyl-13C6 | 329.2 | 188.1 | 5 | 22 | |
| 1.33 | β-OH Fentanyl | 353.2 | 186.1 204.2 |
10 10 |
26 25 |
Acetyl Fentanyl-13C6 |
| 1.48 | 4-ANPP | 281.2 | 188.1 105.0 |
8 8 |
16 30 |
Fentanyl-d5 |
| 1.50 | Fentanyl | 337.2 | 188.1 105.0 |
5 5 |
22 35 |
Fentanyl-d5 |
| 1.49 | Fentanyl-d5 | 342.2 | 188.1 | 5 | 22 | |
| 1.55 | Furanyl Fentanyl | 375.2 | 188.1 105.0 |
5 5 |
22 38 |
Fentanyl-d5 |
| 1.59 | Cyclopropyl Fentanyl | 249.2 | 188.1 105.1 |
10 10 |
25 35 |
Fentanyl-d5 |
| 1.65 | (±)-cis-3-Methylfentanyl | 351.2 | 202.2 105.0 |
5 5 |
22 35 |
Sufentanil-d5 |
| 1.66 | Carfentanil | 395.2 | 355.2 113.1 |
10 10 |
16 32 |
Sufentanil-d5 |
| 1.69 | Butyryl Fentanyl | 351.2 | 188.1 105.0 |
5 5 |
22 38 |
Sufentanil-d5 |
| 1.76 | para-Fluorobutyryl Fentanyl | 369.2 | 188.1 105.0 |
5 5 |
24 40 |
Sufentanil-d5 |
| 1.80 | Sufentanil | 387.2 | 238.1 355.2 |
10 10 |
18 18 |
Sufentanil-d5 |
| 1.79 | Sufentanil-d5 | 392.2 | 238.1 | 10 | 18 |
2.4. Plasma assay
Plasma samples (250 µL) were diluted with 250 µL plasma internal standard solution (4% formic acid containing internal standards at 200 pg/mL) and loaded onto a Waters Oasis PRiME MCX µElution plate. As with the postmortem blood assay, wells were then washed with 200 µL of 100 mM ammonium formate containing 2% formic acid, followed by 200 µL of methanol. Samples were eluted with two 25 µL volumes of ACN:methanol (50:50 v/v) containing 5% strong ammonia solution. All samples were then diluted with 50 µL of water:ACN:formic acid (97:2:1 v/v/v). Five microliters of diluted extract were injected onto the UPLC-MS/MS system. Plasma analysis was also performed on a Waters TQ-S micro mass spectrometer. Separation was achieved on a Waters CSH C18 analytical column (1.7 µm, 2.1 × 100 mm) and the mobile phase gradient was the same as that used for whole blood analysis.
Calibration ranges for fentanyl and norfentanyl were 4–1000 pg/mL with low, medium and high control pools prepared at concentrations of 30, 150 and 600 pg/mL. Calibration ranges for the remaining fentanyls were 2–500 pg/mL with the three quality control pools prepared at 15, 75 and 300 pg/mL.
2.5. Method validation
Both methods were validated according to SWGTOX guidelines for recovery, matrix effects, accuracy, precision, linearity, specificity, selectivity, sensitivity [27]. Recovery and matrix effects were determined quantitatively by analyte addition experiments [28], [29]. Whole blood matrix effects and recoveries were evaluated at three different concentrations, 250, 1000, and 10000 pg/mL. Plasma recoveries and matrix effects were also determined by analyte addition experiments. Due to the more limited range of the plasma assay, recoveries and matrix effects were determined at a single concentration of 1000 pg/mL.
Quantitative validation was achieved by analyzing 6 replicates for inter-day accuracy and precision and 5 replicates between days at 4 QC levels. Lower limits of quantitation were defined as those concentrations at which accuracy was within 20% of target concentrations and precision was less than 20%. Accuracy and precision requirements for all other concentrations were within 15% of target and less than 15% RSD.
3. Results and discussion
3.1. Whole blood assay: Method development and validation studies
Assay optimization and validation was conducted for all pre-analytical and analytical procedures. Pre-analytical preparation of postmortem blood employed zinc sulfate and ammonium acetate treatment with protein precipitation prior to solid phase extraction. Precipitation solvent conditions were optimized for analyte recovery and a methanol:ACN (50:50 v/v) precipitation was selected because it resulted in visibly clearer supernatants, tightly compact precipitate pellets and consistent recoveries. The addition of phosphoric acid to the sample, prior to the solid phase extraction loading step, was also found to improve recoveries, especially among the more polar fentanyls that included the nor-metabolites and remifentanil acid. Fig. 1 shows the mean recoveries and standard deviations resulting from analysis with four pools of analyte-negative blood supplemented with the fentanyls at a concentration of 1000 pg/mL. Recoveries ranged from 53.6 to 84.5% with relative standard deviations under 10% for all analytes demonstrating extraction efficiency. Overall recovery was a balance between developing a simplified workflow that permitted direct loading of the precipitated sample onto the µElution plate. The relatively high percentage of organic solvent (20%) that resulted from this pretreatment step led to lower recoveries for some of the early eluting compounds. Stable isotopically labelled internal standards were available for the two most affected analytes and recoveries were found to be consistent and reproducible. No conditioning or equilibration was necessary for either matrix, further streamlining the workflow.
Fig. 1.
Mean recovery and matrix effects in the postmortem blood assay. Four lots of blank whole blood were evaluated for recovery and matrix effects as described in the Materials and Methods section. Solid bars represent the mean recovery (blue) or matrix effects (orange). Error bars indicate standard deviations. Recoveries ranged from 53.6 to 84.5%. Matrix effects were all less than 13%. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
An additional advantage of the solid phase extraction method was the reduction of matrix effects to less than 13%, also shown in Fig. 1. Matrix effects in blood can be the result of the co-elution of phospholipids that cause ion suppression [29], [30], [31], [32] and build up on the analytical column and within the source region of the mass spectrometer. Mixed-mode solid phase extraction significantly reduced the levels and effects of these phospholipids. Fig. 2 shows the phospholipid profiles of whole blood samples prepared by protein precipitation only in comparison with the results after combined use of precipitation and solid phase extraction. The combination eliminated greater than 99% of the residual phospholipids in the final extract. Recovery and matrix effect studies at lower (250 pg/mL) and higher (10000 pg/mL) concentrations of analytes were also carried out; no concentration-dependent effects were seen, indicating consistent analyte recovery and low matrix effects across the analytical range of the assay.
Fig. 2.
Comparison of phospholipid content with, and without, solid phase extraction (SPE). On the left are total ion chromatograms of residual phospholipids in blood samples subjected to protein precipitation only (upper) and protein precipitation followed by solid phase extraction (lower). The mean area from each set of samples is shown on the right. SPE removed greater than 99% of residual phospholipids from whole blood samples compared to protein precipitation only.
A rapid chromatographic separation method was developed and optimized for analyte selectivity. Representative chromatograms showing separation of the fentanyls in a postmortem blood extract is displayed in Fig. 3. All fentanyls eluted from the column within two minutes; the total cycle time of four minutes included elution and re-equilibration. The functional isomers butyryl fentanyl and cis-3-methyl fentanyl were not fully baseline-separated, but each isomer produced a unique product ion fragment during collision induced dissociation allowing unambiguous identification and quantification for these analytes (Table 1). Chromatographic interferents were evaluated but not found in either analyte-negative blood or the internal standard preparation.
Fig. 3.
Chromatography of fentanyls extracted from blank whole blood samples. Separation was achieved on a Waters BEH C18 column, 1.7 µm; 2.1 × 100 mm as described in Materials and Methods. All analytes eluted in under 2 min.
Only 5 stable isotopically labelled internal standards were available at the time of assay development. Analytes were assigned to internal standards based upon similarities in extraction efficiency (recovery) and matrix effects, followed by retention time.
The assay was further validated for linearity, accuracy, precision, sensitivity, specificity, and carryover according to SWGTOX guidelines [27]. Calibration curves for the whole blood samples (50–20000 pg/mL) were linear with r2 values ≥ 0.997. All analytes were validated for lower limits of quantitation (LLOQ) of 50 pg/mL, apart from remifentanil acid and 4-ANPP, which both validated for LLOQs of 100 pg/mL. Mean (range) inter-assay biases (N = 5 batches) for the fentanyls at 750, 7500 and 15000 pg/mL concentrations in blood were −1.0% (−4.5 to 3.0%), −1.8% (−8.3 to 0%) and −0.2% (−4.4 to 1.5%) respectively. Mean inter-assay relative standard deviations (range) for the fentanyls in replicate low, medium and high control pool concentrations were 3.3% (1.1–16.1%), 2.0% (1.0–3.0%) and 4.5% (3.1–8.9%). Mean carryover (N = 6) was 0.05% (0.00–0.65%) for negative control sample analysis following the analysis of the high calibrator concentration at a concentration of 20000 pg/mL Proficiency of the blood assay was further validated by correlation with an external reference laboratory using 30 postmortem blood samples assayed in the reference lab for 4-ANPP, para-fluorobutyryl fentanyl, and furanyl fentanyl. Fig. 4 shows composite comparison data for the three fentanyl analogs with a regression equation of y = 0.87x – 0.004 and a correlation coefficient of 0.96.
Fig. 4.
Blood concentrations of 4-ANPP, para-fluorobutyryl fentanyl, and furanyl fentanyl in postmortem cases compared with reference laboratory results. The equation represents the linear regression of the results of all positive samples combined.
3.2. Plasma assay: Method development and validation studies
The plasma assay was initially validated for acetyl norfentanyl, norfentanyl, methoxyacetyl fentanyl, acetyl fentanyl, β-OH fentanyl, 4-ANPP, and fentanyl only. A low, but consistent level carryover of 4-ANPP in the whole blood assay (mean 0.65%, range 0.55–0.73%) was observed but met the validation criteria for the concentration range used in the postmortem casework. In the subsequent development of a plasma assay, however, a lower limit of quantitation was required and was achieved by use of an alternate analytical column. Studies with a CSH C18 analytical column revealed a mean (N = 6) 4-ANPP carryover of only 0.15% (range 0.11–0.18%) in analyte negative plasma analyzed after analysis of plasma containing a high 4-ANPP concentration of 2500 pg/mL. The CSH C18 column was therefore adopted for use in the plasma assay and performed with comparable selectivity and slightly shortened retention time for the individual fentanyls compared to the BEH column.
Analytical performance studies with the plasma assay resulted in extraction recoveries ranging between 88 and 95% and standard deviations under 9.5% across all the fentanyls. Matrix effects ranged from −8% to −22% with standard deviations of less than 14%. All calibrations were linear with r2 values ≥ 0.997. The lower limits of quantitation (LLOQ) were 2 pg/mL for acetyl norfentanyl, 4 pg/mL for norfentanyl and fentanyl, and 5 pg/mL for the remaining compounds. Average inter-assay biases at low, medium and high quality control pool concentrations were 0.3% (range −4.3–6.7%), 1.6% (−12.0–13.40%) and −2.3% (-5.5–0.7%), respectively. Precision studies showed inter-assay relative standard deviations for corresponding control pool concentrations of 8.0% (3.0–17.1%), 5.3% (2.8–9.3%) and 3.3% (2.1–5.2%).
3.3. Application of assays in postmortem blood and clinical plasma
Availability of validated blood and plasma assays allowed a comparative study of postmortem blood and clinical samples. Analysis of fentanyls was conducted in a cohort of postmortem blood samples from 44 fatalities involving fentanyl and fentanyl analogs. Table 2 shows the diversity of fentanyls detected in postmortem blood along with their concentration. The frequency of positive findings were as follows: 4-ANPP (38), fentanyl (36), norfentanyl (35), β-OH fentanyl (28), acetyl fentanyl (13), para-fluorobutyryl fentanyl (9), methoxyacetyl fentanyl (6), furanyl fentanyl (5), cyclopropyl fentanyl (3) and acetyl norfentanyl (2), and resulted from the multiple fentanyl findings in each case. Fentanyl was found in over 80% of the cases at a mean concentration of 14710 pg/mL (median 11000 pg/mL) and in close association with norfentanyl, its major metabolite. A steep progression in mean concentration (pg/mL) for the other fentanyls including acetyl norfentanyl (284), β-OH fentanyl (444), acetyl fentanyl (1096), norfentanyl (2609), 4-ANPP (3583), para-fluorobutyryl fentanyl (4284), cyclopropyl fentanyl (4753). furanyl fentanyl (5522), fentanyl (14705) and methoxyacetyl fentanyl (>20000 pg/mL) was found along with a significant variation in blood concentration between cases. The diversity of fentanyls seen in opioid misuse, together with the most adverse of drug use outcomes, reinforces the requirement for definitive methods, of appropriate selectivity and specificity, to be used in testing for synthetic opioid use, not only in fatalities, but in the clinical setting as well.
Table 2.
Fentanyls and their concentration (pg/mL) in de-identified postmortem blood collected from fatalities involving fentanyl and fentanyl analogs.
| Case | 4-ANPP | Fentanyl | Norfentanyl | β–OH fentanyl | Acetyl fentanyl | Acetyl norfentanyl | Furanyl fentanyl | Methoxyacetyl fentanyl | Cyclopropyl fentanyl | para-Fluorobutyryl fentanyl |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 21,500 | 16,700 | ||||||||
| 2 | 1970 | 885 | 751 | |||||||
| 3 | 1100 | 282 | 71 | 867 | ||||||
| 4 | 7190 | 1860 | 697 | |||||||
| 5 | 15,700 | 84,500 | ||||||||
| 6 | 118 | 5330 | 1120 | 134 | 15,100 | |||||
| 7 | 5700 | 1010 | 426 | |||||||
| 8 | 147 | 2860 | 4350 | 313 | 490 | |||||
| 9 | 6750 | 579 | 319 | 30,000 | ||||||
| 10 | 1350 | 2340 | 287 | 118 | ||||||
| 11 | 394 | 9130 | 1860 | 169 | 225 | |||||
| 12 | 3500 | 36,500 | ||||||||
| 13 | 8620 | 24,700 | ||||||||
| 14 | 390 | 10,400 | 5030 | 483 | 4710 | |||||
| 15 | 116 | 6130 | ||||||||
| 16 | 124 | 6620 | 1810 | 188 | 141 | |||||
| 17 | 117 | 18,800 | 3640 | 297 | 74 | |||||
| 18 | 1470 | 36,700 | 12,500 | 954 | ||||||
| 19 | 403 | 24,600 | 1410 | 265 | 69 | |||||
| 20 | 294 | 13,000 | 20,100 | |||||||
| 21 | 182 | 14,400 | 1550 | 272 | ||||||
| 22 | 742 | 51,000 | 10,200 | 2560 | ||||||
| 23 | 212 | 4100 | 1050 | 116 | 121 | 3420 | ||||
| 24 | 1390 | 24,800 | 4030 | 247 | ||||||
| 25 | 45,300 | 142 | 69,900 | |||||||
| 26 | 677 | 20,900 | 3550 | 352 | 143 | |||||
| 27 | 263 | 6480 | 1280 | 277 | ||||||
| 28 | 416 | 28,500 | 5260 | 1200 | 364 | |||||
| 29 | 1860 | 56,400 | 3270 | 738 | 445 | |||||
| 30 | 185 | 12,300 | 3240 | 436 | ||||||
| 31 | 828 | 31,600 | 1440 | 338 | ||||||
| 32 | 226 | 17,800 | 13,600 | 905 | 299 | 87 | ||||
| 33 | 10,500 | 212 | 53 | 7300 | ||||||
| 34 | 447 | 15,100 | 2420 | 519 | ||||||
| 35 | 451 | 41,300 | 311 | |||||||
| 36 | 2050 | 146 | 52 | 909 | ||||||
| 37 | 347 | 12,100 | 1840 | 208 | 8130 | 481 | ||||
| 38 | 528 | 11,600 | 357 | 109 | ||||||
| 39 | 194 | 9360 | 564 | 139 | 192 | |||||
| 40 | 7790 | 910 | 4060 | |||||||
| 41 | 12,500 | 308 | 125 | |||||||
| 42 | 2420 | 413 | 156 | 80 | ||||||
| 43 | 139 | 7020 | 1240 | 447 | 231 | |||||
| 44 | 3290 | 892 | 53 |
In total, 175 identifications of fentanyls were made in the 44 postmortem cases. While fentanyl was present in many of the cases, the fentanyl analogs were significant among these findings. The analogs para-fluorobutyryl fentanyl, methoxyacetyl fentanyl, furanyl fentanyl, and cyclopropyl fentanyl were present in the blood from 20 of the fatalities and appeared consistent with fentanyl analog administration and overdose. These four analogs were found along with fentanyl use in 12 cases but were also found independent of fentanyl use in eight additional cases where analogs were the only opioids detected in blood. The four analogs also predominated in concentration relative to other detected fentanyls in over 70% of the fatalities in which they were involved. Methoxyacetyl fentanyl was found in cases collected early in the study at concentrations that exceeded the calibration range of the assay. Concentration ranges for cyclopropyl fentanyl (3420–6130 pg/mL), furanyl fentanyl (867–16700 pg/mL) and para-fluorobutyryl fentanyl (141–20000 pg/mL) also varied considerably but remained within the limits of quantitation of the blood assay. It should also be noted that the acetyl fentanyl analog was also found in 13 postmortem cases, two with concentrations exceeding 4000 pg/mL and the rest with levels less than 500 pg/mL.
The detection patterns of β-OH fentanyl and 4-ANPP raise further questions regarding potential metabolic or precursor sources in these cases. β-OH fentanyl has been referenced as an illicitly produced fentanyl analog associated with toxicity [33], but fentanyl metabolism also has been suggested by in vitro microsomal and hepatocyte studies [20], [21]. In vivo studies in support of fentanyl metabolism to β-OH fentanyl have not been previously reported to our knowledge. A metabolic source of 4-ANPP has been suggested by an early report of three patients with qualitative identification of 4-ANPP following intravenous administration of fentanyl [20] and by a hepatic microsomal study [21]. It is known, however, that 4-ANPP is a precursor intermediary in the synthesis of fentanyl. A metabolic versus illicit production source of β-OH fentanyl and 4-ANPP in our cases is not clear from the postmortem data alone but several observations can be made. First, β-OH fentanyl was present in 28 of the fatality cases but did not predominate in opioid concentration in any of the cases. Importantly, β-OH fentanyl showed a striking 100% association with the presence of fentanyl and also was not detected in the eight cases where fentanyl was not detected. On average, β-OH fentanyl concentrations were 3.4% of the fentanyl concentrations and ranged up to 11%. The close association of fentanyl with β-OH fentanyl and the lower levels of β-OH fentanyl relative to fentanyl in these overdoses are consistent with a metabolic source of the β-OH fentanyl but the possibility of illicit exogenous use cannot be ruled out. Second, 4-ANPP was present in blood from 38 cases and was detected in cases with, and without, evidence of fentanyl use. The concentration of 4-ANPP ranged widely and did not appear to show any correlation with fentanyl or any of the analogs. The furanyl fentanyl and methoxyacetyl fentanyl positive samples had substantial concentrations of 4-ANPP compared with the cyclopropyl fentanyl cases suggesting a possible variation in synthesis or clean up efficiency in illicit laboratory production. These findings suggest a synthesis byproduct origin of 4-ANPP rather than metabolism, but again, the data is only suggestive.
The plasma assay was also applied in a study of clinical samples from patients receiving pharmaceutical grade fentanyl and these results may shed additional light on the questions raised in the postmortem casework, particularly with respect to the source of acetyl fentanyl, β-OH fentanyl and 4-ANPP. We analyzed 70 baseline and post-dose plasma samples from 18 individuals receiving intravenous fentanyl during a surgical procedure. Methoxyacetyl fentanyl, acetyl fentanyl, noracetyl fentanyl and 4-ANPP were not detectable. Fentanyl, norfentanyl and β-OH fentanyl, however, were present as shown in Table 3. As anticipated, the level of fentanyls determined with clinical use is much lower than observed in the fatal overdose cases and the lower limit of quantitation used in the plasma assay was clearly needed. Consistent with the dose and recency of fentanyl administration in these clinical cases, fentanyl was found in highest concentration, with a mean of 397 pg/mL and ranging from 7 to 1500 pg/mL. Even with acute administration of fentanyl, metabolism to norfentanyl was quantifiable in 52 of the 53 samples containing fentanyl. The mean concentration of norfentanyl was 14 pg/mL with a range of 4–45 pg/mL. Norfentanyl represented 6% of the fentanyl concentration on average, with a range up to 17%. β-OH fentanyl was also measurable in 24 of the fentanyl positive samples but at lower concentration than norfentanyl with an average β-OH fentanyl concentration of 9 pg/mL and a range of 5 to 20 pg/mL. β-OH fentanyl quantitation averaged 44% (19 to 86%) of norfentanyl and 2% (range 0.5–4%) of fentanyl. For comparison, postmortem blood β-OH fentanyl levels averaged 19% (5 to 42%) of the corresponding norfentanyl concentration and 3% (1–11%) of the fentanyls concentration. These data would support the conclusion that β-OH fentanyl is a significant metabolite of fentanyl that is present in the circulation following acute fentanyl administration in either therapeutic or fatal doses.
Table 3.
Fentanyls and their concentration in plasma from de-identified serial blood samples collected for parathyroid hormone measurement in 18 parathyroidectomy cases at the Albany Medical Center (AMC).
| Sample | Norfentanyl (pg/mL) | β-OH Fentanyl (pg/mL) | Fentanyl (pg/mL) | |
|---|---|---|---|---|
| Case 1 Baseline | AMC Sample 001B | nd | nd | nd |
| AMC Sample 002 | 20.7 | 5 | 203.2 | |
| AMC Sample 003 | 28.4 | 7.5 | 172 | |
| Case 2 Baseline | AMC Sample 004B | nd | nd | nd |
| AMC Sample 005 | 17.6 | nd | 526.3 | |
| AMC Sample 006 | 27.2 | 8.2 | 369 | |
| AMC Sample 007 | 32.6 | 12.9 | 366.2 | |
| Case 3 Baseline | AMC Sample 008B | nd | nd | nd |
| AMC Sample 009 | 6.8 | nd | 341.9 | |
| AMC Sample 010 | 10.3 | nd | 327.1 | |
| AMC Sample 011 | 11.4 | nd | 289.1 | |
| Case 4 Baseline | AMC Sample 012B | nd | nd | nd |
| AMC Sample 013 | 20.7 | 6.7 | 480 | |
| AMC Sample 014 | 23.5 | 5 | 322 | |
| AMC Sample 015 | 24.4 | nd | 283.7 | |
| Case 5 Baseline | AMC Sample 016B | nd | nd | nd |
| AMC Sample 017 | 11.1 | nd | 467 | |
| AMC Sample 018 | 15.3 | nd | 330.5 | |
| AMC Sample 019 | 16.9 | nd | 448.4 | |
| Case 6 Baseline | AMC Sample 020B | nd | nd | nd |
| AMC Sample 021 | 5.8 | 134.6 | ||
| AMC Sample 022 | 19.7 | 5.8 | 307.7 | |
| Case 7 Baseline | AMC Sample 023B | nd | nd | nd |
| AMC Sample 024 | 17.5 | nd | 221.7 | |
| AMC Sample 025 | 26.1 | 6.5 | 322.8 | |
| Case 8 Baseline | AMC Sample 026B | nd | nd | nd |
| AMC Sample 027 | 25.3 | 5.5 | 561.9 | |
| AMC Sample 028 | 36.2 | 19.6 | 1250.3 | |
| AMC Sample 029 | 35.2 | 13.2 | 963.2 | |
| Case 9 Baseline | AMC Sample 030B | nd | nd | nd |
| AMC Sample 031 | 5.3 | nd | 465.1 | |
| AMC Sample 032 | 14.8 | nd | 347.6 | |
| AMC Sample 033 | 12.6 | nd | 406.9 | |
| Case 10 baseline | AMC Sample 034B | nd | nd | nd |
| AMC Sample 035 | 19.6 | nd | 616.5 | |
| AMC Sample 036 | 22.3 | 9.8 | 592.6 | |
| AMC Sample 037 | 26.5 | 5.1 | 512.3 | |
| Case 11 Baseline | AMC Sample 038B | nd | nd | nd |
| AMC Sample 039 | 44.9 | nd | 310.9 | |
| AMC Sample 040 | 36.7 | nd | 349.3 | |
| Case 12 Baseline | AMC Sample 041B | nd | nd | nd |
| AMC Sample 042 | 3.4 | nd | 194.1 | |
| Case 13 Baseline | AMC Sample 043B | nd | nd | nd |
| AMC Sample 044 | 12.6 | nd | 99.8 | |
| AMC Sample 045 | 13 | nd | 92.4 | |
| AMC Sample 046 | 22.4 | nd | 175.7 | |
| Case 14 Baseline | AMC Sample 047B | nd | nd | nd |
| AMC Sample 048 | 10.5 | nd | 404.2 | |
| AMC Sample 049 | 13.3 | nd | 311.6 | |
| AMC Sample 050 | 15.1 | nd | 282 | |
| Case 15 Baseline | AMC Sample 051B | nd | nd | nd |
| AMC Sample 052 | 11.5 | nd | 644.9 | |
| AMC Sample 053 | 10.4 | nd | 816.5 | |
| Case 16 Baseline | AMC Sample 054B | nd | nd | 7.3 |
| AMC Sample 055 | 40.9 | 14.8 | 352.1 | |
| AMC Sample 056 | 37.9 | 13 | 294.4 | |
| Case 17 * | AMC Sample 057 | 7.6 | 5.6 | 1494.1 |
| AMC Sample 058 | 13.1 | 10.8 | 621.4 | |
| AMC Sample 059 | 12.1 | 7.4 | 500.3 | |
| AMC Sample 060 | 13 | 9.1 | 551.2 | |
| AMC Sample 061 | 13.6 | 11.7 | 411.7 | |
| AMC Sample 062 | 15.5 | 12 | 392.2 | |
| AMC Sample 063 | 19.8 | 11.3 | 295 | |
| Case 18 Baseline | AMC Sample 064B | nd | nd | nd |
| AMC Sample 065 | 12.8 | 5.1 | 196 | |
| AMC Sample 066 | 13.4 | nd | 163.2 | |
| AMC Sample 067 | 14.9 | nd | 159.7 | |
| AMC Sample 068 | 16.6 | 5.2 | 151.9 | |
| AMC Sample 069 | 20.1 | nd | 136.9 |
nd = none detected. LLOQs were 5 pg/mL for 4-ANPP and β-OH Fentanyl and 4 pg/mL for norfentanyl and fentanyl.
*Baseline specimen not available for testing.
The complete absence of detectable 4-ANPP in the clinical study may be a result of a lower fentanyl concentration achieved in therapeutic use independent of a metabolic or synthesis byproduct origin. Therapeutic concentration of fentanyl averages 3% of the mean as well as median concentration in fentanyl-related fatalities. A proportionate reduction in 4-ANPP levels would still result in concentrations measurable with the greater sensitivity of the plasma assay and should be detectable if 4-ANPP were metabolic in origin. The data is not conclusive and may be explained by a lower 4-ANPP concentration in pharmaceutical fentanyl due to manufacturing practices and governmental oversight. If further studies substantiate this explanation, then 4-ANPP may be considered a concentration-based marker of illicitly sourced fentanyl.
4. Conclusions
Sensitive and selective methods for the analysis of a panel of fentanyls have been developed and validated for use in blood or plasma testing. The assays have been validated based on current standards of practice and are applicable to both forensic and clinical testing. Application in postmortem blood analysis shows concomitant use of fentanyl and fentanyl analogs along with a high frequency of 4-ANPP associated with the cases. Clinical studies in plasma from patients receiving pharmaceutical grade fentanyl revealed an association of fentanyl with norfentanyl and β-OH fentanyl. The postmortem and clinical findings indicate a metabolic source for the β-OH fentanyl determined in these cases.
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
References
- 1.Jannetto P.J., Helander A., Garg U., Janis G.C., Goldberger B., et al. The fentanyl epidemic and evolution of fentanyl analogs in the United States and the European Union. Clin. Chem. 2019;65:242–253. doi: 10.1373/clinchem.2017.281626. [DOI] [PubMed] [Google Scholar]
- 2.Mounteney J., Giraudon I., Denissov G., Griffiths P. Fentanyls: are we missing the signs? Highly potent and on the rise in Europe. Int. J. Drug Policy. 2015;26:626–631. doi: 10.1016/j.drugpo.2015.04.003. [DOI] [PubMed] [Google Scholar]
- 3.Centers for Disease Control and Prevention. 2019 annual surveillance Report of Drug Related Risks and Outcomes - United States Surveillance Special Report. Centers for DiseaseControl and Prevention, U.S. Department of Health and Human Services. November 1, 2019. https://www.cdc.gov/drugoverdose/pdf/pubs/2019-cdc-drug-surveillance-report.pdf (accessed 10th February 2020).
- 4.USDEA National Drug Threat Assessment, DEA-DCT-DIR-040–17. https://www.hsdl.org/?abstract&did=805532 (accessed 10th February 2020).
- 5.Centers for Disease Control and Prevention. CDC Health Advisory: Increases in fentanyl drug confiscations and fentanyl-related overdose fatalities. HAN Health Advisory. October 26, 2015. http://emergency.cdc.gov/han/han00384.asp (accessed February 2020).
- 6.Huynh N.H., Tyrefors N., Ekman L., Johansson M. Determination of fentanyl in human plasma and fentanyl and norfentanyl in human urine using LC-MS/MS. J. Pharm. Biomed. Anal. 2005;37(5):1095–1100. doi: 10.1016/j.jpba.2004.09.024. [DOI] [PubMed] [Google Scholar]
- 7.Thevis M., Geyer H., Bahr D., Schanzer W. Identification of fentanyl, alfentanil, sufentanil, remifentanil and their major metabolites in human urine by liquid chromatography/tandem mass spectrometry for doping control purposes. Eur J. Mass Spectrom. 2005;11(4):419–427. doi: 10.1255/ejms.761. [DOI] [PubMed] [Google Scholar]
- 8.Gergov M., Nokua P., Vuori E., Ojanpera I. Simultaneous screening and quantification of 25 opioid drugs in post-mortem blood and urine by liquid chromatography-tandem mass spectrometry. Forensic Sci. Int. 2009;186(1–3):36–43. doi: 10.1016/j.forsciint.2009.01.013. [DOI] [PubMed] [Google Scholar]
- 9.Cooreman S., Deprez C., Martens F., Van Bocxlaer J., Croes K. A comprehensive LC-MS-based quantitative analysis of fentanyl-like drugs in plasma and urine. J. Sep. Sci. 2010;33(17–18):1654–1662. doi: 10.1002/jssc.201000330. [DOI] [PubMed] [Google Scholar]
- 10.Sofalvi S., Schueler H.E., Lavins E.S., Kaspar C.K., Brooker I.T., et al. An LC-MS-MS method for the analysis of carfentanil, 3-methylfentanyl, 2-furanyl fentanyl, acetyl fentanyl, fentanyl and norfentanyl in postmortem and impaired-driving cases. J. Anal. Toxicol. 2017;41:473–483. doi: 10.1093/jat/bkx052. [DOI] [PubMed] [Google Scholar]
- 11.McIntyre I.M., Trochta A., Gary R.D., Wright J., Mena O. An acute butyr-fentanyl fatality: a case report with postmortem concentrations. J. Anal. Toxicol. 2016;40:162–166. doi: 10.1093/jat/bkv138. [DOI] [PubMed] [Google Scholar]
- 12.Algren D.A., Monteilh C.P., Punja M., Schier J.G., Belson M., et al. Fentanyl-associated fatalities among illicit drug users in Wayne County, Michigan (July 2005–May 2006) J. Med. Toxicol. 2013;9:106–115. doi: 10.1007/s13181-012-0285-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.T.G. Rosano, M. Wood, Tof-MRM for the confirmation of fentanyl analogues, Waters Application Note 720006515EN (2019) 1-6. https://www.waters.com/webassets/cms/library/docs/720006515en.pdf.
- 14.Fogarty M.F., Papsun D.M., Logan B.K. Analysis of fentanyl and 18 novel fentanyl analogs and metabolites by LC-MS-MS, and report of fatalities associated with methoxyacetylfentanyl and cyclopropylfentanyl. J. Anal. Toxicol. 2018;42:592–604. doi: 10.1093/jat/bky035. [DOI] [PubMed] [Google Scholar]
- 15.Mohr A.L.A., Friscia M., Papsun D., Kacinko S.L., Buzby D., et al. Analysis of novel synthetic opioids U-47700, U-50488 and furanyl fentanyl by LC-MS/MS in postmortem casework. J. Anal. Toxicol. 2016;40:709–717. doi: 10.1093/jat/bkw086. [DOI] [PubMed] [Google Scholar]
- 16.G. Roda, F. Faggiani, C. Bolchi, M. Pallavicini, M. Dei Cas, Ten years of fentanyl-like drugs: a technical-analytical review. Anal. Sci.35 (2019) 479-491. https://doi.org/10.2116/analsci.18R004 [DOI] [PubMed]
- 17.Sofalvi S., Lavins E.S., Brooker I.T., Kaspar C.K., Kucmanic J., et al. Unique structural/stereo-isomer and isobar analysis of novel fentanyl analogues in postmortem and DUID whole blood by UHPLC–MS-MS. J. Anal. Toxicol. 2019;43:673–687. doi: 10.1093/jat/bkz056. [DOI] [PubMed] [Google Scholar]
- 18.Coulter C.A., Moore C.M. Analysis of drugs in oral fluid using LC-MS/MS. Methods Mol. Biol. 1872;2019:237–259. doi: 10.1007/978-1-4939-8823-5_22. [DOI] [PubMed] [Google Scholar]
- 19.Griswold M.K., Chai P.R., Krotulski A.J., Friscia M., Chapman B.P., et al. A novel oral fluid assay (LC-QTOF-MS) for the detection of fentanyl and clandestine opioids in oral fluid after reported heroin overdose. J. Med Toxicol. 2017;13(4):287–292. doi: 10.1007/s13181-017-0632-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Goromaru T., Matsuura H., Yoshimura N., Miyawaki T., Sameshima T., et al. Identification and quantitative determination of fentanyl metabolites in patients by gas chromatography–mass spectrometry. Anesthesiology. 1984;61:73–77. [PubMed] [Google Scholar]
- 21.Labroo R.B., Paine M.F., Thummel K.E., Kharasch E.D. Fentanyl metabolism by human hepatic and intestinal cytochrome P450 3A4: implications for interindividual variability in disposition, efficacy and drug interactions. Drug Metab. Dispos. 1997;25:1072–1080. [PubMed] [Google Scholar]
- 22.Kuip E.J.M., Zandvliet M.L., Koolen S.L.W., Mathijssen R.H.J., Rijt C.C.D. A review of factors explaining variability in fentanyl pharmacokinetics; focus on implications for cancer patients. Br. J. Clin. Pharmacol. 2017;83:294–313. doi: 10.1111/bcp.13129. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Armenian P., Vo K.T., Barr-Walker J., Lynch K.L. Review: fentanyl, fentanyl analogs and novel synthetic opioids: a comprehensive review. Neuropharmacology. 2018;134:121–132. doi: 10.1016/j.neuropharm.2017.10.016. [DOI] [PubMed] [Google Scholar]
- 24.Martucci H.F.H., Ingel E.A., Hunter M.D., Rodda L.N. Distribution of furanyl fentanyl and 4-ANPP in an accidental acute death: a case report. Forensic Sci. Int. 2018;283:13–17. doi: 10.1016/j.forsciint.2017.12.005. [DOI] [PubMed] [Google Scholar]
- 25.Kanamori T., Iwata Y.T., Segawa H., Yamamuro T., Kuwayama K., et al. Metabolism of fentanyl and acetylfentanyl in human-induced pluripotent stem cell-derived hepatocytes. Biol. Pharm. Bull. 2018;41:106–114. doi: 10.1248/bpb.b17-00709. [DOI] [PubMed] [Google Scholar]
- 26.OSAC for Forensic Science – ASB/ANSI 098 Standard for Mass Spectral Data Acceptance in Forensic Toxicology. Draft https://www.nist.gov/system/files/documents/2019/04/22/chsac_-_tox_-_identification_in_forensic_toxicology_-_for_asb_and_website_1.pdf
- 27.Scientific Working Group for Forensic Toxicology (SWGTOX) – Recommendations of the Research, Development, Testing, and Evaluation Committee. J. Anal. Toxicol.37 (2013) 187-191. [DOI] [PubMed]
- 28.Chambers E., Diehl-Wagrowski D.M., Lu Z., Mazzeo J.R. Systematic and comprehensive strategy for reducing matrix effects in LC/MS/MS analyses. J. Chromatogr. B. 2007;852:22–34. doi: 10.1016/j.jchromb.2006.12.030. [DOI] [PubMed] [Google Scholar]
- 29.Matuszewski B.K., Constanzer M.L., Chavez-Eng C.M. Matrix effect in quantitative LC/MS/MS analyses of biological fluids: a method for determination of finasteride in human plasma at picogram per milliliter concentrations. Anal. Chem. 1998;70:882–889. doi: 10.1021/ac971078+. [DOI] [PubMed] [Google Scholar]
- 30.Danaceau J.P., Wood M., Calton L.J. Extraction of synthetic fentanyl compounds and removal of phospholipids from whole blood using Oasis PRiME MCX for forensic toxicology. Waters Technol. Brief 720006330EN. 2018:1–2. [Google Scholar]
- 31.Ahmad S., Kaira H., Gupta A., Raut B., Hussain A., et al. HybridSPE: a novel technique to reduce phospholipid-based matrix effect in LC–ESI-MS Bioanalysis. J. Pharm. Bioallied Sci. 2012;4:267–275. doi: 10.4103/0975-7406.103234. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Little J.L., Wempe M.F., Buchanan C.M. Liquid chromatography–mass spectrometry/mass spectrometry method development for drug metabolism studies: Examining lipid matrix ionization effects in plasma. J. Chromatogr. B. 2006;833:219–230. doi: 10.1016/j.jchromb.2006.02.011. [DOI] [PubMed] [Google Scholar]
- 33.Hendrickson R.G., Akpunonu P., Hughes A.R., Gerona R. Highly potent fentanyl analogs: apnea from exposure to small quantities of ß-hydroxyfentanyl and furanylfentanyl. Clin. Toxicol. 2019;57:813–815. doi: 10.1080/15563650.2018.1558233. [DOI] [PubMed] [Google Scholar]




