. 2021 Nov 20;20:82–87. doi: 10.1016/j.phro.2021.11.002

Comparison of Library of Plans with two daily adaptive strategies for whole bladder radiotherapy

Duncan den Boer a,, Mariska D den Hartogh a,1, Alexis NTJ Kotte a, Jochem RN van der Voort van Zyp a, Juus L Noteboom a, Gijsbert H Bol a, Thomas Willigenburg a, Anita M Werensteijn-Honingh a, Ina M Jürgenliemk-Schulz a, Astrid LHMW van Lier a, Petra S Kroon a
PMCID: PMC8609047  PMID: 34849413

Abstract

Background and purpose

Whole bladder radiotherapy is challenging due to inter- and intrafraction size and shape changes. To account for these changes, currently a Library of Plans (LoP) technique is often applied, but daily adaptive radiotherapy is also increasingly becoming available. The aim of this study was to compare LoP with two magnetic resonance imaging guided radiotherapy (MRgRT) strategies by comparing target coverage and volume of healthy tissue inside the planning target volume (PTV) for whole bladder treatments.

Methods and materials

Data from 25 MRgRT lymph node oligometastases treatments (125 fractions) were used, with three MRI scans acquired at each fraction at 0, 15 and 30 min. Bladders were delineated and used to evaluate three strategies: 1) LoP with two plans for a 15 min fraction, 2) MRgRT15min for a 15 min fraction and 3) MRgRT30min for a 30 min fraction. The volumes of healthy tissue inside and bladder outside the PTV were analyzed on the simulated post-treatment images.

Results

MRgRT30min had 120% and 121% more healthy tissue inside the PTV than LoP and MRgRT15min. For LoP slightly more target outside the PTV was found than for MRgRT30min and MRgRT15min, with median 0% (range 0–23%) compared to 0% (0–20%) and 0% (0–10%), respectively.

Conclusions

Taking into account both target coverage and volume of healthy tissue inside the PTV, MRgRT15min performed better than LoP and MRgRT30min for whole bladder treatments. A 15 min daily adaptive radiotherapy workflow is needed to potentially benefit from replanning compared to LoP.

Keywords: Bladder cancer, MR-guided radiotherapy, Daily online adaptive radiotherapy, Library of Plans

1. Introduction

Patients with muscle invasive bladder cancer typically have a five-year survival prognosis of about 50% [1]. Cystectomy has been the golden standard of treatment, but bladder sparing treatments involving radiotherapy have shown similar outcomes [1]. Bladder sparing treatments have the advantage of lower complication rates and improved quality of life by preserving normal bladder and sexual functions.

The wide variations in bladder size, shape and position have traditionally made it a challenging site for radiotherapy [2]. A commonly used bladder treatment strategy is the Library of Plans (LoP) strategy. Here, a “library” of planning target volumes (PTVs) and corresponding treatment plans is generated prior to treatment and the best fitting PTV is selected daily, typically based on an onboard conebeam CT (CBCT) [3], [4], [5], [6], [7], [8], [9], [10], [11], [12]. For this approach it was shown that irradiated normal tissue volume and treatment margins could be reduced.

Another recent development in the field is the clinical availability of daily adaptive radiotherapy systems, CBCT-guided [13], [14] as well as MR-guided systems [15], [16], [17], [18]. These have already been shown suitable to deal with interfraction changes through daily adaptation of the treatment plan for several treatment sites and are therefore promising for bladder treatments. Vestergaard et al. have illustrated that MR-guided radiotherapy (MRgRT) has the potential to reduce irradiated normal tissue volume for bladder treatment [19] and both CBCT-guided [14] and MRgRT [20] have recently been demonstrated to be a feasible treatment modality for this tumor site. However, these studies were done with a low number of bladder patients (three [14] and five [20]) and results were not compared to the current clinical practice, i.e. LoP.

Both strategies, LoP and daily adaptive radiotherapy, have their own advantages and disadvantages. A disadvantage of LoP is the chance that none of the PTVs generated prior to treatment is suitable for a specific daily situation. For daily adaptive RT the PTV is generated from the bladder of the day and the daily shape and volume can be taken into account. However, the treatment time with the presently available MRgRT systems from imaging to end of treatment delivery is typically a factor of two to three longer than the treatment time for a CBCT-based LoP strategy [5], [16], [20], [21]. Consequently, relatively large PTV margins are needed to take into account the intrafraction filling of the bladder. Current margins reported for intrafraction changes of the bladder are typically for shorter time intervals (8–18 min) [8], [11], [18], [21]. To speed up the daily MRgRT workflow technical developments are expected in the near future. Examples of these developments include faster auto-contouring [22] and Artificial Intelligence-assisted treatment planning reoptimization [23]. For daily adaptive CBCT-guided radiotherapy systems, times of 18 min from acceptance of first CBCT to start of irradiation have been reported for tumor sites in the pelvic region [14].

An advantage of MRgRT is the potential improvement of visualization of the soft tissues which can be taken into account with daily replanning to reduce the dose to the organs at risk (OAR). The possibility of intrafraction target monitoring and gating is also a reason to apply MRgRT [24].

The purpose of this study was to compare LoP with two daily adaptive MRgRT workflows with a total treatment time of 15 and 30 min per fraction, using clinically obtained MRgRT data. A retrospective analysis was performed to validate which strategy performs better in sparing healthy tissue and achieving target coverage. While our work focused on MRgRT, the conclusions are also relevant for CBCT-guided daily online adaptive radiotherapy.

2. Methods and materials

2.1. Patients

For the retrospective simulation, data from twenty-five lymph node metastases patients treated on a 1.5 T MR-linac (Unity, Elekta AB, Stockholm, Sweden) were used. This unique cohort was selected since bladder patients had not yet been treated with MRgRT in our institute. The included patients received clinical treatment with five fractions between August 2018 and August 2019. They provided written informed consent for use of their data as part of an ethics review board approved observational study. No specific drinking instructions were provided.

2.2. Imaging data and delineation

For each patient a pre-treatment planning CT (0.9 × 0.9 × 2.0 mm3; Brilliance CT big bore, Philips Medical Systems, Best, NL) was used. Additionally, three MRI scans were acquired during each fraction on the MRgRT system, MRI1, MRI2 and MRI3 respectively (total N = 15 for each patient) [16]. MRI image acquisition time was median two minutes. For each patient a T1 or T2 sequence was selected for the MRgRT workflow by the radiation oncologist, based on tumor visibility assessed during a pre-treatment MRI session on a 1.5 T Philips Ingenia MRI scanner (Philips Medical Systems, Best, NL). For 19/25 treatments a 3D T1-weighted scan (reconstructed voxel size 0.8 × 0.8 × 1.0 mm3; FOV 400 × 400 × 300 mm3) was selected and for 6/25 treatments a 3D T2-weighted scan (reconstructed voxel size 0.6 × 0.6 × 2.0 mm3; FOV 400 × 400 × 300 mm3). The geometrical inaccuracy in and closely around the bladder is expected to be <0.7 mm for a 1.5 T MR-linac system [25]. Bladders were manually and retrospectively delineated on all CT and MRI images leading to BladderCT, Bladder1, Bladder2 and Bladder3.

2.3. Clinical MRgRT workflow

For each treatment fraction of the patient group mentioned above, daily online replanning was performed following the Adapt to Shape workflow of the Elekta Unity as described previously by Werensteijn-Honingh et al. [16]. Each fraction, contours from the pre-treatment planning CT were propagated on the daily MRI using rigid and deformable registration.

Electron density was based on assignment to structures. If necessary, contours of GTV and OARs were manually adapted. The treatment plan, consisting of seven IMRT beams, was reoptimized for each fraction. A secondary dose calculation was performed with a separate treatment planning system, Oncentra version 4.5.2 (Elekta AB, Stockholm, Sweden). Irradiation was done with 7MV FFF. In the manuscript we will refer to this workflow as the “current” MRgRT workflow, illustrating the current capabilities of the MRgRT commercial system. The steps in this workflow are expected to take the same time for this patient group as for bladder patients [17], [18]. As only a small part of the time is spent on irradiation [16], this time is not expected to change drastically when a different dose per fraction is given.

2.4. Simulated MRgRT approach

Two MRgRT daily optimization strategies were simulated: 1) a workflow based on the median time between MRI1 and MRI3 of 30 min (range 21–57 min), MRgRT30min, illustrating the current capabilities of an MRgRT commercial system and 2) a workflow, based on the median time between MRI1 and MRI2 of 15 min (range 8–25 min), MRgRT15min, illustrating a hypothetical optimized workflow (Fig. 1). For MRgRT15min all online adaptive steps including radiation delivery were assumed between MRI1 and MRI2 and for MRgRT30min these steps were between MRI1 and MRI3 (with the online adaptive steps between MRI1 and MRI2 and radiation delivery between MRI2 and MRI3 [16]).

Fig. 1.

Fig. 1

Illustration of the LoP strategy and the two MRgRT strategies (MRgRT30min and MRgRT15min). 1) illustrates the LoP strategy. a) The bladder was delineated on the planning CT and two PTVs were constructed (PTV LoP Tight and PTV LoP Wide). b) Based on Bladder1 the most suitable PTV was selected (in this case PTV LoP Wide). c) MRI2 was considered the simulated post-treatment image. Evaluation of PTV LoP Selected was done with Bladder2. 2) illustrates the MRgRT30min workflow. d) A PTV was constructed from Bladder1 with population-based margins. e) This PTV was evaluated with the MRI3 as simulated post-treatment image using Bladder3. 3) illustrates the MRgRT15min workflow. f) Similar to the MRgRT30min, a PTV was constructed from Bladder1. g) Evaluation was now done with MRI2 as simulated post-treatment image using Bladder2.

PTVs were constructed from Bladder1 on MRI1 by population-based margins for MRgRT15min and MRgRT30min. Published data on margins are typically not validated for 30 min, but for shorter time intervals [8], [11], [19], [26]. Therefore the margins to account for the intra-fraction bladder changes (intra-fraction patient motion, centroid motion and bladder filling) were constructed from the MRgRT dataset. To determine appropriate margins the delineated bladders on the MRI1 were asymmetrically expanded in six directions (left, right, anterior, posterior, inferior and superior) in steps of 1 mm until the delineated bladder on the MRI3 or MRI2 scan was completely encompassed for MRgRT30min and MRgRT15min strategy, respectively. Ten of the twenty-five patients were randomly selected. Population-based PTV margins were derived from these ten patients (N = 50 fractions). Margins were constructed for each direction by placing a 90% inclusion criterion on individual treatment fractions. The most commonly used margin recipe [27] uses 90% of patients instead of fractions, but in our case a new treatment plan would be generated for each fraction. This random selection and subsequent margin calculation was done to minimize bias and repeated 1000 times. Finally, the mean values were used for the analysis in this manuscript (see Supplementary Material A for more details).

2.5. Simulated LoP approach

The bladder was contoured on the planning CT and expanded to create two PTVs to fill the library. The expansion for PTV LoP Tight was 6/6 mm left/right, 10/10 mm ant/post and 6/10 mm inf/sup and for PTV LoP Wide: 10/10 mm left/right, 15/15 mm ant/post and 10/15 mm inf/sup. These values were based on literature [3], [10] and clinical practice in our institute for CBCT-guided radiotherapy. The methodology is a further adaptation of the work as described previously by Willems et al. [28]. Daily plan selection was based on the contoured bladder on MRI1, the smallest LoP PTV was selected that encompassed the complete bladder, named PTV LoP Selected. If the bladder on MRI1 did not fit completely within either of the LoP PTVs or if the bladder was too empty (if there was more than one cm distance between the bladder and the PTV in all directions except caudal), the fraction was not included in the analysis in line with clinical practice. In these cases, the patient would be asked to void the bladder or to drink additional liquid. Plan selection and radiotherapy delivery were assumed between MRI1 and MRI2 with a time interval of about 15 min which corresponded well to reported CBCT LoP studies with time intervals between 9 and 19 min [12], [29], [30].

2.6. Treatment strategy evaluation and statistics

To evaluate the efficacy of the LoP and MRgRT approaches, the volume of healthy tissue in the PTV and the relative volume of the bladder outside the PTV were analyzed on the simulated post-treatment MRI images: MRI2 for LoP and MRgRT15min, and MRI3 for MRgRT30min.

Volumes were determined with an in-house developed software package, Volumetool [31]. Fractions were only included in the statistical comparison when an LoP PTV was selected. The analysis we present was by pooling over the different scans for the patients. The differences in volumes between the strategies were tested by the two-tailed Wilcoxon matched-pairs signed-rank test; p < 0.05 was considered significant.

3. Results

Margins (mean ± standard deviation) for the two MRgRT strategies were 11 ± 2/17 ± 2 mm left/right, 24 ± 6/13 ± 2 mm ant/post and 36 ± 5/5 ± 1 mm sup/inf for MRgRT30min and 8 ± 1/10 ± 1 mm left/right, 12 ± 2/11 ± 1 mm ant/post and 18 ± 2/4 ± 1 mm sup/inf for MRgRT15min. The margins were asymmetric with the highest values for anterior and superior directions.

For the comparison of LoP vs. MRgRT strategies 107/125 fractions were included. For 7/125 fractions the PTV LoP Wide did not completely encompass the delineated bladder on MRI1 and for 11/125 fractions the bladder was too empty. PTV LoP Tight and Wide were selected 73 and 34 times. The median and range of the bladder volumes at CT, MRI1, MRI2 and MRI3, as well as the PTVs for all three strategies are presented in Table 1. The median bladder filling speed was respectively 1.6 cm3/min (range 0.3–12.0 cm3/min) and 1.3 cm3/min (range 0.1–9.3 cm3/min) between MRI3 and MRI1, and MRI2 and MRI1.

Table 1.

Median and range are shown for bladder volumes on simulated pre-treatment scan and on three time points during simulated whole bladder radiotherapy based on 125 fractions of 25 patients treated with MRgRT, as well as median and range of PTVs of the three investigated radiotherapy approaches: Library of Plans (LoP) with fraction time of 15 min, MRgRT15min with fraction time of 15 min and MRgRT30min with fraction time of 30 min.

Structure N [–] Volume [cm3]
Bladder
Planning CT 25 97 (37–239)
MRI1 1071 83 (31–284)
125 83 (31–593)
MRI2 1071 103 (46–411)
125 104 (46–680)
MRI3 1071 136 (53–570)
125 136 (53–703)



PTV
LoP Tight 25 245 (105–438)
LoP Wide 25 348 (164–581)
LoP Selected 1071 270 (105–546)
MRgRT15min 1071 265 (134–620)
125 261 (134–1127)
MRgRT30min 1071 466 (263–970)
125 462 (263–1646)
1

Based on 107/125 fractions for which one of the two LoP plans was adequate for the observed bladder volume and patient would not be asked to void or drink.

For the MRgRT30min workflow the volumes of the PTVs were up to 75% larger than for the other two strategies, with a median value of 466 cm3 compared to 270 cm3 for the selected LoP PTVs and 265 cm3 for the PTVs of MRgRT15min.

The volumes of healthy tissue inside the PTV and target outside the PTV on the simulated post-treatment image are presented in Table 2. The median volume of healthy tissue inside the PTV was 332 cm3 for MRgRT30min and corresponded to 71% of the PTV. This volume was significantly larger (p < 0.01) than for MRgRT15min and LoP with values of 151 and 150 cm3 respectively (Fig. 2a). For MRgRT15min and LoP the volumes of healthy tissue inside the PTV did not significantly differ, p = 0.11.

Table 2.

Median and range of volumes for healthy tissue inside PTV and missed target on simulated post-treatment images for whole bladder radiotherapy simulated with 125 fractions of 25 patients treated with MRgRT. Volumes are shown for three radiotherapy approaches: Library of Plans (LoP) with fraction time of 15 min, MRgRT15min with fraction time of 15 min and MRgRT30min with fraction time of 30 min.

Method N
[–]
Healthy tissue in PTV
[cm3]
Missed target
[cm3]
Missed target
[%]
LoP
1071 150 (28–299) 0 (0–48) 0 (0–23)



MRgRT15min
1071 151 (59–260) 0 (0–28) 0 (0–10)
125 151 (59–447) 0 (0–28) 0 (0–10)



MRgRT30min
1071 332 (136–537) 0 (0–100) 0 (0–20)
125 315 (135–941) 0 (0–100) 0 (0–20)
1

Based on 107/125 fractions for which one of the two LoP plans was adequate for the observed bladder volume and patient would not be asked to void or drink.

Fig. 2.

Fig. 2

Simulated volume of healthy tissue inside PTV and missed target volume in post-treatment image (MRI2 for LoP and MRgRT15min, MRI3 for MRgRT30min) for whole bladder treatment using three treatment strategies MRgRT30min, MRgRT15min and LoP based on 25 treatments of 5 fractions. Data shown for 107 fractions (18 were excluded due to too full or too empty bladder filling for LoP). ** indicates significance of p < 0.01, by two-tailed Wilcoxon matched-pairs signed-rank test a) The volume of healthy tissue inside the PTV in the post-treatment image. b) The volume of bladder outside the PTV in the post-treatment image. c) Relative volumes of the bladder (in %) outside the PTV in the post-treatment image.

The median value of target volume, i.e. bladder volume, outside the PTV on the simulated post-treatment image was 0.0 cm3 for all three methods (Fig. 2b). The MRgRT15min strategy did not perform significantly better than the MRgRT30min strategy, p = 0.39, however both performed better than the LoP strategy, p < 0.01 (Fig. 2b). The relative volume of bladder outside the PTV on the post-treatment MRI was lower for MRgRT15min than for LoP. The 75%–91% whisker in the boxplot, corresponding to 19 fractions was between 0.5 and 6.5% for LoP and 0.0–0.3% for MRgRT15min (Fig. 2c). The LoP strategy also showed outliers up to 21–23% of missed target volume, whereas the highest value for MRgRT15min was 10%. Some practical examples of bladder outside the PTV (missed target) are given in Fig. 3, where 17–28 cm3 and 100 cm3 of the target (bladder) were missed.

Fig. 3.

Fig. 3

Two examples shown in the sagittal plane, illustrating exceptions; one or more of the strategies led to part of the bladder being outside the PTV in the simulated post-treatment image. Time of the image indicated in top right corner. a–d) Example 1. Planning CT (a) and T1 weighted images (b–d). The selected LoP PTV (yellow) was sufficient in MRI1, but missed a part (29 cm3) of the bladder in MRI2 (indicated by the yellow arrow in c). (e–h) Example 2. Planning CT (a) and T2 weighted images (f–h). The PTVs of all three strategies missed a part of the bladder in the simulated post-treatment MRI. For LoP and MRgRT15min: 17 cm3 and 28 cm3 of the Bladder2 was outside the PTV respectively (indicated by the yellow arrow in g). For the MRgRT30min workflow: 100 cm3 of the Bladder3 was outside the PTV (indicated by the yellow arrow in h).

4. Discussion

To the best of our knowledge, this is the first study comparing LoP with MRgRT for whole bladder treatments using clinical MRgRT data. Two daily adaptive MRgRT strategies with a median time of 15 min, MRgRT15min, and 30 min, MRgRT30min, between first image and end of treatment were compared to our clinically used LoP strategy. The MRgRT30min workflow had more healthy tissue inside the PTV than the LoP strategy. The MRgRT15min workflow had a similar volume of healthy tissue inside the PTV as LoP and missed less target.

Limitations of our study were the use of data of patients that did not have bladder cancer and the absence of a drinking protocol. The filling can be described as a “comfortably filled bladder”. However, the observed starting bladder volumes (median 97 cm3 at start of treatment) corresponded well with reported valued of bladder patients with drinking protocols, with a range of 100–130 cm3 [6], [7], [11], [19], [32]. The bladder filling speed of median 1.6 cm3/min between MRI1 and MRI3 and 1.3 cm3/min between MRI1 and MRI2 were also in the range of reported values ranging from 0.8 to 2.1 cm3/min [26], [32], [33], [34], [35]. Even with a drinking protocol, large interpatient and interfraction variations typically still occur [33], [34]. It has been shown that for empty and full bladder, the same values could be used in terms of margins [26], [36].

The PTV margins found in this study for MRgRT15min strategy corresponded very well to margins reported in literature for studies with similar intrafraction time intervals [8], [11], [19], [32], with values of 5–10/5–10 mm left/right, 8–14/9–14 mm ant/post and 13–33/5–6 mm sup/inf. The MRgRT15min margins fit well within the clinically used expansions for generating the LoP PTVs.

The PTV margins determined for the MRgRT30min workflow were relatively large, compared to other publications [6], [8], [11], [19], [20], [26], [36]. This was mostly due to the relatively long treatment time in comparison to the investigated time frames in other studies (8–18 min). In the study of Hunt et al. [20], the PTV margins (5/5 mm left/right, 15/10 mm ant/post and 15/5 mm sup/inf) covered 90% of the bladder at the 30 min time point, instead of the 100% for MRgRT30min in our study. The median intrafraction filling volume of the five patients studied in Hunt et al. [20], including one with a catheter, was 30 cm3 compared to 48 cm3 in our study.

The PTV margins of the MRgRT workflows were constructed from the same dataset as used for the comparison with LoP, this is a limitation of our study and could introduce bias for MRgRT30min and MRgRT15min. To mitigate this, margin construction was done with a repeated randomly selected subset of the patients and using the obtained mean values.

The advantage of the daily reoptimization using the MRgRT30min workflow compared to the LoP strategy was canceled out due to the needed relatively large population-based margins. The use of patient-specific margins could lead to a reduction in PTV volume and radiation exposure to healthy tissue. Published work showed promising results in this direction [8], [37].

For the LoP strategy, fractions were excluded following the clinical workflow, when the patients bladder was too full or too empty and no appropriate LoP PTV was available. This occurred in 14% (18/125) of the treatment fractions. This shows an advantage of the MRgRT strategies, where the treatment would have continued, as there is no dependency on a predefined library. While this approach is in line with the clinical workflow, the fact that these fractions were excluded from the analysis could potentially introduce a bias for the LoP workflow.

The differences between the LoP and MRgRT15min strategies in terms of target coverage and normal tissue in the PTV were small and bladder patients can be treated adequately with a CBCT-guided LoP strategy. However there are other advantages for using MRgRT. With the soft tissue contrast combined with daily adaptive radiotherapy, it has the potential to decrease toxicity of surrounding OARs [38]. In addition, the soft tissue contrast can visualize the tumor and allow a boost of this area without the need for implanted markers (which would be the case for CBCT-guided RT). MRgRT also allows for intrafraction monitoring and gating.

In conclusion, MRgRT15min performed best, closely followed by LoP and then MRgRT30min for whole bladder treatments, taking into account the target coverage and the amount of healthy tissue inside the PTV. To be able to use MRgRT for whole bladder treatments including its functionalities, it is necessary to further optimize and accelerate the MRgRT workflow.

Conflict of Interest Notification

The overarching University Medical Center Utrecht MR-linac scientific project, including employment of one of the authors, has been partly funded by Elekta AB (Stockholm, Sweden). Elekta did not have any part in the design, execution or analysis of this study. The authors declared that there is no other conflict of interest.

Funding

None.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Acknowledgements

The authors want to acknowledge Simon Woodings for providing language help during the preparation of this manuscript.

Footnotes

Appendix A

Supplementary data to this article can be found online at https://doi.org/10.1016/j.phro.2021.11.002.

Appendix A. Supplementary data

The following are the Supplementary data to this article:

Supplementary Data 1
mmc1.docx (530.1KB, docx)

References

  • 1.Arcangeli G., Arcangeli S., Strigari L. A systematic review and meta-analysis of clinical trials of bladder-sparing trimodality treatment for muscle-invasive bladder cancer (MIBC) Crit Rev Oncol Hematol. 2015;94:105–115. doi: 10.1016/j.critrevonc.2014.11.007. [DOI] [PubMed] [Google Scholar]
  • 2.Turner S.L., Swindell R., Bowl N., Marrs J., Brookes B., Read G., et al. Bladder movement during radiation therapy for bladder cancer: implications for treatment planning. Int J Radiat Oncol Biol Phys. 1997;39:355–360. doi: 10.1016/S0360-3016(97)00070-9. [DOI] [PubMed] [Google Scholar]
  • 3.Lalondrelle S., Huddart R., Warren-Oseni K., Hansen V.N., McNair H., Thomas K., et al. Adaptive-Predictive Organ Localization Using Cone-Beam Computed Tomography for Improved Accuracy in External Beam Radiotherapy for Bladder Cancer. Int J Radiat Oncol Biol Phys. 2011;79:705–712. doi: 10.1016/j.ijrobp.2009.12.003. [DOI] [PubMed] [Google Scholar]
  • 4.Foroudi F., Wong J., Kron T., Rolfo A., Haworth A., Roxby P., et al. Online adaptive radiotherapy for muscle-invasive bladder cancer: results of a pilot study. Int J Radiat Oncol Biol Phys. 2011;81(3):765–771. doi: 10.1016/j.ijrobp.2010.06.061. [DOI] [PubMed] [Google Scholar]
  • 5.Meijer G.J., van der Toorn P.P., Bal M., Schuring D., Weterings J., de Wildt M. High precision bladder cancer irradiation by integrating a library planning procedure of 6 prospectively generated SIB IMRT plans with image guidance using lipiodol markers. Radiother Oncol. 2012;105:174–179. doi: 10.1016/j.radonc.2012.08.011. [DOI] [PubMed] [Google Scholar]
  • 6.McDonald F., Lalondrelle S., Taylor H., Warren-Oseni K., Khoo V., McNair H.A., et al. Clinical Implementation of Adaptive Hypofractionated Bladder Radiotherapy for Improvement in Normal Tissue Irradiation. Clin Oncol (R Coll Radiol) 2013;25:549–556. doi: 10.1016/j.clon.2013.06.001. [DOI] [PubMed] [Google Scholar]
  • 7.Vestergaard A., Kallehauge J.F., Petersen J.B.B., Høyer M., Søndergaard J., Muren L.P. An adaptive radiotherapy planning strategy for bladder cancer using deformation vector fields. Radiother Oncol. 2014;112:371–375. doi: 10.1016/j.radonc.2014.07.012. [DOI] [PubMed] [Google Scholar]
  • 8.Grønborg C., Vestergaard A., Høyer M., Söhn M., Pedersen E.M., Petersen J.B., et al. Intra-fractional bladder motion and margins in adaptive radiotherapy for urinary bladder cancer. Acta Oncol. 2015;54:1461–1466. doi: 10.3109/0284186X.2015.1062138. [DOI] [PubMed] [Google Scholar]
  • 9.Lutkenhaus L.J., Visser J., de Jong R., Hulshof M.C.C.M., Bel A. Evaluation of delivered dose for a clinical daily adaptive plan selection strategy for bladder cancer radiotherapy. Radiother Oncol. 2015;116:51–56. doi: 10.1016/j.radonc.2015.06.003. [DOI] [PubMed] [Google Scholar]
  • 10.Murthy V., Masodkar R., Kalyani N., Mahantshetty U., Bakshi G., Prakash G., et al. Clinical Outcomes With Dose-Escalated Adaptive Radiation Therapy for Urinary Bladder Cancer: A Prospective Study. Int J Radiat Oncol Biol Phys. 2016;94:60–66. doi: 10.1016/j.ijrobp.2015.09.010. [DOI] [PubMed] [Google Scholar]
  • 11.Foroudi F., Pham D., Bressel M., Gill S., Kron T. Intrafraction bladder motion in radiation therapy estimated from pretreatment and posttreatment volumetric imaging. Int J Radiat Oncol Biol Phys. 2013;86:77–82. doi: 10.1016/j.ijrobp.2012.11.035. [DOI] [PubMed] [Google Scholar]
  • 12.Collins S.D., Leech M.M. A review of plan library approaches in adaptive radiotherapy of bladder cancer. Acta Oncol. 2018;57:566–573. doi: 10.1080/0284186X.2017.1420908. [DOI] [PubMed] [Google Scholar]
  • 13.Archambault Y., Boylan C., Bullock D., Morgas T., Peltola J., Ruokokoski E., et al. Making on-Line Adaptive Radiotherapy Possible Using Artificial Intelligence and Machine Learning for Efficient Daily Re-Planning. Med Phys Int. 2020;8:77–86. [Google Scholar]
  • 14.Sibolt P., Andersson L.M., Calmels L., Sjöström D., Bjelkengren U., Geertsen P., et al. Clinical implementation of artificial intelligence-driven conebeam computed tomography-guided online adaptive radiotherapy in the pelvic region. Phys Imaging Radiat Oncol. 2021;17:1–7. doi: 10.1016/j.phro.2020.12.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Henke L.E., Contreras J.A., Green O.L., Cai B., Kim H., Roach M.C., et al. Magnetic Resonance Image-Guided Radiotherapy (MRIgRT) A 4.5-Year Clinical Experience. Clin Oncol (R Coll Radiol) 2018;30:720–727. doi: 10.1016/j.clon.2018.08.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Werensteijn-Honingh A.M., Kroon P.S., Winkel D., Aalbers E.M., van Asselen B., Bol G.H., et al. Feasibility of stereotactic radiotherapy using a 1.5 T MR-linac: Multi-fraction treatment of pelvic lymph node oligometastases. Radiother Oncol. 2019;134:50–54. doi: 10.1016/j.radonc.2019.01.024. [DOI] [PubMed] [Google Scholar]
  • 17.de Muinck Keizer D.M., Kerkmeijer L.G.W., Willigenburg T., van Lier A.L.H.M.W., Hartogh M.D.D., van der Voort van Zyp J.R.N., et al. Prostate intrafraction motion during the preparation and delivery of MR-guided radiotherapy sessions on a 1.5T MR-Linac. Radiother Oncol. 2020;151:88–94. doi: 10.1016/j.radonc.2020.06.044. [DOI] [PubMed] [Google Scholar]
  • 18.Tetar S.U., Bruynzeel A.M.E., Lagerwaard F.J., Slotman B.J., Bohoudi O., Palacios M.A. Clinical implementation of magnetic resonance imaging guided adaptive radiotherapy for localized prostate cancer. Phys Imaging Radiat Oncol. 2019;9:69–76. doi: 10.1016/j.phro.2019.02.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Vestergaard A., Hafeez S., Muren L.P., Nill S., Høyer M., Hansen V.N., et al. The potential of MRI-guided online adaptive re-optimisation in radiotherapy of urinary bladder cancer. Radiother Oncol. 2016;118:154–159. doi: 10.1016/j.radonc.2015.11.003. [DOI] [PubMed] [Google Scholar]
  • 20.Hunt A., Hanson I., Dunlop A., Barnes H., Bower L., Chick J., et al. Feasibility of Magnetic Resonance Guided Radiotherapy for the Treatment of Bladder Cancer. Clin Transl Radiat Oncol. 2020;25:46–51. doi: 10.1016/j.ctro.2020.09.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Hafeez S., Warren-Oseni K., McNair H.A., Hansen V.N., Jones K., Tan M., et al. Prospective study delivering simultaneous integrated highdose tumor boost (≤ 70 Gy) with image guided adaptive radiation therapy for radical treatment of localized muscle-invasive bladder cancer. Int J Radiat Oncol Biol Phys. 2016;94:1022–1030. doi: 10.1016/j.ijrobp.2015.12.379. [DOI] [PubMed] [Google Scholar]
  • 22.Sharp G., Fritscher K.D., Pekar V., Peroni M., Shusharina N., Veeraraghavan H., et al. Vision 20/20: perspectives on automated image segmentation for radiotherapy. Med Phys. 2014;41:050902. doi: 10.1118/1.4871620. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Kontaxis C., Bol G.H., Lagendijk J.J.W., Raaymakers B.W. DeepDose: Towards a fast dose calculation engine for radiation therapy using deep learning. Phys Med Biol. 2020;65 doi: 10.1088/1361-6560/ab7630. [DOI] [PubMed] [Google Scholar]
  • 24.Mutic S., Dempsey J.F. The ViewRay system: Magnetic Resonance-Guided and controlled radiotherapy. Semin Radiat Oncol. 2014;24:196–199. doi: 10.1016/j.semradonc.2014.02.008. [DOI] [PubMed] [Google Scholar]
  • 25.Tijssen R.H.N., Philippens M.E.P., Paulson E.S., Glitzner M., Chugh B., Wetscherek A., et al. MRI commissioning of 1.5T MR-linac systems – a multi-institutional study. Radiother Oncol. 2019;132:114–120. doi: 10.1016/j.radonc.2018.12.011. [DOI] [PubMed] [Google Scholar]
  • 26.Dees-Ribbers H.M., Betgen A., Pos F.J., Witteveen T., Remeijer P., van Herk M. Inter- and intra-fractional bladder motion during radiotherapy for bladder cancer: A comparison of full and empty bladders. Radiother Oncol. 2014;113:254–259. doi: 10.1016/j.radonc.2014.08.019. [DOI] [PubMed] [Google Scholar]
  • 27.Van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol. 2004;14:52–64. doi: 10.1053/j.semradonc.2003.10.003. [DOI] [PubMed] [Google Scholar]
  • 28.Willems N.J.W., Kroon P.S., Boer J.C.J., Meijer G.J., Zyp J.R.N., Noteboom J.L. EP-1844: Clinical introduction of simple adaptive radiotherapy for transitional cell bladder carcinoma. Radiother Oncol. 2017;05:S1009. doi: 10.1016/S0167-8140(17)32279-X. [DOI] [Google Scholar]
  • 29.Adil K., Popovic M., Cury F.L., Faria S.L., Duclos M., Souhami L. Anisotropic bladder planning target volume in bladder radiation therapy. Pract Radiat Oncol. 2019;9:24–28. doi: 10.1016/j.prro.2018.07.006. [DOI] [PubMed] [Google Scholar]
  • 30.Krishnan A., Tsang Y.M., Stewart-Lord A. The impact of intra-fractional bladder filling on “Plan of the day” adaptive bladder radiotherapy. Tech Innov Patient Support Radiat Oncol. 2019;9:31–34. doi: 10.1016/j.tipsro.2019.01.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Bol G.H., Kotte A.N.T.J., van der Heide U.A., Lagendijk J.J.W. Simultaneous multi-modality ROI delineation in clinical practice. Comput Methods Programs Biomed. 2009;96:133–140. doi: 10.1016/j.cmpb.2009.04.008. [DOI] [PubMed] [Google Scholar]
  • 32.Mangar S.A., Scurr E., Huddart R.A., Sohaib S.A., Horwich A., Dearnaley D.P., et al. Assessing intra-fractional bladder motion using cine-MRI as initial methodology for Predictive Organ Localization (POLO) in radiotherapy for bladder cancer. Radiother Oncol. 2007;85:207–214. doi: 10.1016/j.radonc.2007.04.037. [DOI] [PubMed] [Google Scholar]
  • 33.McBain C.A., Khoo V.S., Buckley D.L., Sykes J.S., Green M.M., Cowan R.A., et al. Assessment of bladder motion for clinical radiotherapy practice using cine-magnetic resonance imaging. Int J Radiat Oncol Biol Phys. 2009;75:664–671. doi: 10.1016/j.ijrobp.2008.11.040. [DOI] [PubMed] [Google Scholar]
  • 34.Søndergaard J., Olsen K.O., Muren L.P., Elstrøm U.V., Grau C., Høyer M. A study of image-guided radiotherapy of bladder cancer based on lipiodol injection in the bladder wall. Acta Oncol. 2010;49:1109–1115. doi: 10.3109/02841861003789491. [DOI] [PubMed] [Google Scholar]
  • 35.McBain C.A., Green M.M., Stratford J., Davies J., McCarthy C., Taylor B., et al. Ultrasound imaging to assess inter- and intra-fraction motion during bladder radiotherapy and its potential as a verification tool. Clin Oncol (R Coll Radiol) 2009;21:385–393. doi: 10.1016/j.clon.2009.01.016. [DOI] [PubMed] [Google Scholar]
  • 36.Wilson C., Moseshvili E., Tacey M., Quin I., Lawrentschuk N., Bolton D., et al. Assessment of intrafraction motion of the urinary bladder using magnetic resonance imaging (cineMRI) Clin Oncol (R Coll Radiol) 2020;32:101–109. doi: 10.1016/j.clon.2019.09.056. [DOI] [PubMed] [Google Scholar]
  • 37.Krywonos J., Fenwick J., Elkut F., Jenkinson I., Liu Y.H., Brunt J.N.H., et al. MRI image-based FE modelling of the pelvis system and bladder filling. Comput Methods Biomech Biomed Engin. 2010;13:669–676. doi: 10.1080/10255840903446961. [DOI] [PubMed] [Google Scholar]
  • 38.Winkel D., Kroon P.S., Werensteijn-Honingh A.M., Bol G.H., Raaymakers B.W., Jürgenliemk-Schulz I.M. Simulated dosimetric impact of online replanning for stereotactic body radiation therapy of lymph node oligometastases on the 1.5T MR-linac. Acta Oncol. 2018;57:1705–1712. doi: 10.1080/0284186X.2018.1512152. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Data 1
mmc1.docx (530.1KB, docx)

Articles from Physics and Imaging in Radiation Oncology are provided here courtesy of Elsevier