Abstract
Objective:
Detrusor overactivity (DO) is a urodynamic observation characterized by fluctuations in detrusor pressure (Pdet) of the bladder. Although detecting DO is important for the management of bladder symptoms, the invasive nature of urodynamic studies (UDS) makes it a source of discomfort and morbidity for patients. Ultrasound bladder vibrometry (UBV) could provide a direct and noninvasive means of detecting DO, due to its sensitivity to changes in elasticity and load in the bladder wall. In this study, we investigated the feasibility and applying UBV toward detecting DO.
Approach:
UBV and urodynamic study (UDS) measurements were collected in 76 neurogenic bladder patients (23 with DO). Timestamped group velocity squared (cg2) data series were collected from UBV measurements. Concurrent Pdet data series were identically analyzed for comparison and validation. A processing approach is developed to separate transient fluctuations in the data series from the larger trend of the data and a DO index is proposed for characterizing the transient peaks observed in the data.
Main Results:
Applying the DO index as a classifier for DO produced sensitivities and specificities of 0.70 and 0.75 for cg2 data series and 0.70 and 0.83 for Pdet data series respectively.
Significance:
It was found that DO can be feasibly detected from data series of timestamped UBV measurements. Collectively, these initial results are promising, and further refinement to the UBV measurement process is likely to improve and clarify its capabilities for noninvasive detection of DO.
Keywords: Bladder, Detrusor Overactivity, Lamb Waves, Urodynamics, Ultrasound Elastography
I. INTRODUCTION
Detrusor overactivity (DO) is the urodynamic observation of involuntary contractions of the detrusor muscle of the bladder during the filling phase of urinary storage. To meet the criteria of detrusor overactivity the involuntary contractions must produce fluctuations in the detrusor pressure (Pdet) of larger than 15 cm of H2O over the baseline detrusor pressure as measured by urodynamic study (UDS) (Drake et al., 2018, Rosier et al., 2019). DO is common in neurogenic bladder patients (Panicker et al., 2015, Goldmark et al., 2014) and is important to diagnose in this population due to the risk of upper urinary tract issues (Panicker et al., 2015, Liao, 2015). In both neurogenic and non-neurogenic bladder patient populations, DO is also frequently accompanied by overactive bladder syndrome (OAB), which is characterized by symptoms of urinary urgency as well as frequency and nocturia (Abrams et al., 2002). OAB is believed to occur in 10% of the population (Irwin et al., 2011) and is associated with substantial reductions in quality-of-life (Sacco et al., 2010).
Although assessment of DO in patients with OAB may be insightful for understanding the underlying etiology of said symptoms (Peyronnet et al., 2019), clinical guidelines typically recommend assessing DO only when there is a substantial clinical justification. For instance, the American Urological Association guidelines recommend that DO only be assessed in non-neurogenic OAB after conservative treatments and drug therapies have failed (Winters et al., 2012), a recommendation that is consistent with recent reviews such as (Aoki et al., 2017, Al Mousa et al., 2019). This is because of the potential for discomfort and morbidity associated with UDS, due to the catheterization of the urinary tract that is required for measuring bladder pressure. For instance, although rates of symptomatic urinary tract infections following UDS evaluations are highly variable and dependent on a variety of factors, a recent meta-analysis (Wu et al., 2021) reported rates of symptomatic UTI across 6 studies of 23% (78/340) without administration of prophylactic antibiotics and 11% (48/418) if the test was performed with antibiotic prophylaxis.
Because of the potential complications of UDS, a variety of noninvasive techniques are being investigated for assessment of lower urinary tract symptoms such as OAB (Farag and Heesakkers, 2011, Gammie et al., 2020). Ultrasound-based techniques, in particular, are promising as ultrasound is already a familiar tool in evaluation of lower urinary tract symptoms (Panicker et al., 2015) and are also portable, safe, and cost-effective. Approaches to using ultrasound for assessment of DO have primarily focused on measuring geometric characteristics of the bladder. For instance, ultrasound has been used to measure bladder wall thickness, which has been proposed to correlate with DO due to detrusor hypertrophy (Bright et al., 2010). Likewise, changes in bladder shape during contraction have been measured through ultrasound imaging (Gray et al., 2019, Glass Clark et al., 2018). However, the former has been shown to have limited diagnostic capabilities for DO (Latthe et al., 2017, Rachaneni et al., 2016), whereas the latter’s diagnostic potential has yet to be determined. Furthermore, the physical link between shape changes and DO is indirect and complex.
Recent work from our group has introduced ultrasound bladder vibrometry (UBV) as a new approach to assessing bladder elasticity through the measurement of induced Lamb waves along the wall of the bladder (Nenadic et al., 2013, Bayat et al., 2017, Nenadic et al., 2016, Fatemi and Nenadic, 2016). Because UBV is an elastographic technique, it has unique potential relative to other ultrasound-based methods, as it offers a more direct means of characterizing bladder elasticity and loading. Elastographic techniques have a well-documented sensitivity to the changes in tissue elasticity produced by tissue loading, including those produced by large deformations (Barr and Zhang, 2012) and active muscle contraction (Gennisson et al., 2005, Hug et al., 2015). This sensitivity is increasingly leveraged as a potential means of assessing and tracking the fluid pressures within a variety of biological compartments, as such pressures induce mechanical tension within the walls of the biological compartment. Biological compartments where this association between pressure and elastographic measurement include the ventricles of the heart (Vejdani-Jahromi et al., 2019), the diaphragm muscle and the lungs (Bachasson et al., 2019), and large blood vessels (Wang et al., 2019). For the bladder, previous work on UBV has shown medium-to-strong correlations of UBV measured Lamb wave group velocity squared (cg2) with UDS determined Pdet(Bayat et al., 2017).
Because UBV is sensitive to the elasticity and loading state of the bladder wall, we hypothesized that the temporal fluctuations in detrusor pressure can be usefully detected by UBV measurement. In this preliminary study, we investigated the feasibility of detecting DO by analysis of transient peaks in time-resolved series of UBV measurements on a cohort of neurogenic bladder patients undergoing routine UDS. To this end, we develop an approach to identifying fluctuations associated with DO in timestamped series of cg2 measurements. We then propose an index that characterizes the intensity of such fluctuations which we compare statistically between DO and non-DO bladders.
II. METHODS
In order to investigate the feasibility of detecting DO through analysis of UBV measurements, the following approach was used. cg2 and Pdet data series were gathered from experiments in which neurogenic bladder patients underwent concurrent UBV and UDS measurements. The collected cg2 series were first preprocessed in order to remove statistically anomalous measurements and measurements with identified sources of variation unrelated to DO. Processing was then applied to the preprocessed data series in order to identify DO associated transient peaks. Finally, to characterize the decomposed series collected from a given experiment by a single number for statistical comparisons, a DO index, I, is proposed. A non-parametric statistical test and ROC curve analysis are then applied to measurement of I to evaluate the feasibility of detecting DO from UBV measurements. These steps are described in detail in the proceeding subsections.
A. Patient Population
From April 2013 to January 2018, we evaluated 76 patient volunteers who were scheduled for a routine UDS were considered for the study. The study was composed of 76 adult patients with neurogenic bladder (16 females; 60 male). Inclusion criteria for the study were that patients be 21 years of age and have at least one of the following conditions: Neurogenic bladder, stress incontinence, benign prostatic hyperplasia, or voiding dysfunction. Patients were excluded in the case of a BMI greater than 35, a known neurologic disease other than those in the inclusion criteria, prolonged catheter drainage, previous radiation therapy, previous surgery in either the pelvis or bladder and pregnancy or breast-feeding. Patients' ages ranged from 21 to 82 years, with the mean age of 50±18 years. The body mass index (BMI) ranged from 17.3 to 34.2, with a mean BMI of 26.3±4.60. Etiologies of neurogenic bladder symptoms included traumatic spinal cord injuries in 31% (23/76), multiple sclerosis in 31% (23/76), progressive neuropathy from an unknown etiology in 14% (11/76), spina bifida 11% (8/76), neural impingement from spinal column disc disease 9% (7/76), and Parkinson’s disease in 4% (3/75). The study was Health Insurance Portability and Accountability Act Compliant and was approved by the Mayo Clinic Institutional Review Board (IRB). A signed written informed consent with permission for publication was obtained from each enrolled patient prior to the study. Review of the urodynamic studies obtained within these volunteers revealed that 23 of the patients had documented DO during the performance of the UDS. All patients that consented to the study and completed both the UDS and UBV evaluations were included in the analysis.
B. UDS and UBV measurements
UBV measurements were collected during the patient’s routine cystometry and UDS. Regarding the urodynamic techniques we used standardized techniques as outlined by the International Continence Society at a filling rate of 25 mL/minute in either a supine or sitting position (Rosier et al., 2017). It is noteworthy that prior studies in our laboratory revealed no difference in UBV measurements between the supine and sitting positions (Adusei et al., 2021). During the filling phase of the urodynamic study, filling was stopped every 50 ml up to the maximum patient capacity. The pressure sustained by the detrusor muscle of the bladder (i.e. Pdet) was recorded in a continuous fashion throughout bladder filling with a dense sampling rate (⩾2 Hz) that is in line with the International Continence Society’s standards for urodynamic equipment (Gammie et al., 2014). DO on the UDS was defined by standard urodynamic criteria (Drake et al., 2018, Rosier et al., 2019). UBV measurements were obtained episodically during the time of bladder filling with 2-3 acquisitions performed at each volume. At each stoppage volume, the probe was placed on the lower abdomen targeting the anterior bladder wall to collect measurements and was removed during filling to the next volume. At the time of acquisition of the UBV measurement, a concurrent Pdet measurement was marked on the UDS chart that was used for DO diagnosis and timestamps were recorded for both measurements. It is noteworthy that the UBV measurements and marked Pdet measurements are episodic and will be substantially undersampled compared the near continuously sampled urodynamic evaluations. the resulting series of Pdet measurements correspond to the Pdet channel from the UDS chart but undersampled identically to the UBV measurements. Figure 1 illustrates the experimental set-up during these measurements.
Figure 2 shows histograms of the means and standard deviations of time intervals between the episodic UBV measurements for each experiment. The time interval statistics are separated into time intervals between measurements collected at the same stoppage volume and between stoppage volumes. Separate means and standard deviations are collected for measurements taken at the same filling volume and measurement taken between filling volumes.
The UBV system and measurement sequence has been previously reported in (Bayat et al., 2017) and (Nenadic et al., 2016). UBV measurements were collected using a programmable ultrasound system (Verasonics, Redmond, WA), and a curvilinear array transducer (C4-2, ATL/Philips, Bothewell, WA) with a center frequency of 2.5 MHz. A single 600-900 microsecond tone burst acoustic pulse (2.5 MHz driving frequency) was applied to the wall of the bladder to induce a broad spectrum of waves in the tissue through the mechanism of acoustic radiation force. An ultrafast plane wave imaging sequence then captured the resulting Lamb wave propagation along the bladder wall. Ultrasonic plane waves were transmitted at a pulse repetition frequency of 7500 Hz and 3 steering angles (−4°,0°,4°) were used for coherent compounding of the ultrafast images (effective frame rate 2500 Hz). The transmit frequency and pulse duration of the imaging pulses were 3 MHz and 0.33 microseconds respectively. Coherent compounding was used to limit the effect of sidelobes that is known to distort the displacement tracking in heterogenous tissues. See (Montaldo et al., 2009) for a description of the benefits of coherent compounding in the tracking of transverse waves for elastography. Particle motion velocity was estimated from the resulting in-phase and quadrature (IQ) data using an autocorrelation technique (Loupas et al., 1995) and an ensemble length of 2 pulses. The group velocity (cg) of the Lamb wave propagating along the wall of the bladder was then estimated as it was in (Bayat et al., 2017): by collecting the time-to-peak of the particle velocity wave profile for each lateral location along the bladder wall and fitting a regression line to estimate the slope of the arrival times relative to position. Because elasticity is expected to correlate with the square of the wave speed, as is more rigorously the case with bulk waves, the group velocity was squared (i.e. cg2) for analysis, as was also done in the previous paper (Bayat et al., 2017).
C. Data Preparation and Peak Identification
Here the peak identification process is described, wherein the cg2 or Pdet data series collected from the concurrent UBV/UDS experiments are preprocessed and decomposed into two series: one containing transient peaks identified by the processing and hypothesized to correspond to DO, and another containing effectively the lower envelope that tracks the larger time trend of the data series during bladder filling. All processing and subsequent statistical analysis of cg2 and Pdet data series were applied using MATLAB 2019a (Mathworks Inc., Natick, MA).
All of the same preprocessing and processing applied to cg2 measurement series for peak identification and subsequent statistical analysis was also applied to concurrently collected Pdet measurement series (i.e. Pdet measurements marked on the UDS at the timepoints of UBV acquisitions). This was done based on the hypothesis that detrusor contractions due to overactivity produce similar responses in cg2 and Pdet measurements (this hypothesis is illustrated in Figure 1 (b) and (c).) As such, applying the same analysis to concurrent Pdet measurements as that applied to cg2 measurements would be expected to produce similar statistical discrimination of DO. It also provides validation of the proposed analysis, as the concurrent Pdet measurements are undersampled as compared to the full UDS chart used by the urologist to diagnose DO.
Prior to the peak identification process, several pre-processing steps were applied to the cg2 and Pdet data series. These were done in order to avoid fluctuations in the series that are expected to be unassociated with DO. Values of the given data series which were marked at the time of the experiment as having occurred during a bladder leak were removed from the series. Measurements collected at the largest fill volume of each experiment were also removed to avoid the inclusion of normal detrusor contraction associated with the onset of voiding (Wyndaele et al., 2011). Occasionally, a severe outlier measurement would be present in either the cg2 or Pdet data series, as illustrated in Figure 3. Such outliers could be due to, for instance, patient motion during measurement or unusually weak UBV signals. These outliers were handled by applying a statistical outlier filter to all of the data series (regardless of whether or not an outlier measurement was suspected). The statistical filter used was a Hampel filter (Pearson et al., 2016) with an 11-point window width and an outlier cut-off of 6 standard deviations. Details of the implementation of this outlier filter are provided in Appendix A.
Because the cg2 and Pdet data series often showed large time-varying trend due to bladder filling, accompanied by smaller fluctuations due to either DO-associated contractions or measurement noise, DO associated peak identification required some means of separating these fluctuations from the larger time-varying trend. The approach we propose for making this separation is illustrated in Figure 4 and is analogous to detrending of a time series before analysis (Casella et al., 2008). However, rather than assuming some common functional form of the large time-varying trend, we collect points that constitute the lower portion of the trend by first smoothing the data (See solid red curve in Figure 4 (a)) through a 2nd-order Savitzky-Golay(Savitzky and Golay, 1964) filter with a smoothing factor of 0.95. The points falling below the smoothed series are then collected, as shown in Figure 4 (b). After interpolation the time-points between these lower points, the resulting series is what we will hereafter refer to as the Lower Envelope (LE) series. Figure 4 (c) shows the LE series plotted over the original series. It can be seen that by subtracting the LE series from the original series, a new series that captures transient peaks observed on top of the larger trend of the measurement series is collected. This series is shown in Figure 4 (d) in blue and will be referred to as the Ensemble of Transient Peaks (ETP) series. A detailed description of the processing steps involved in this method is provided in Appendix B.
D. DO Index
In order to make statistical and diagnostic comparisons between the peak identification of DO and non-DO bladders, we proposed a DO index that characterizes the increased signal energy associated with fluctuations due to DO. To be applied to these current experiments, such a DO index should be, in principle, comparable between experiments of variable durations as different patients will have different maximum capacities. Likewise, it should also be comparable between experiments with varied ranges in UBV measurement. For this purpose, we defined the proposed DO index, denoted as I, as follows:
(1) |
Here xETP denotes the ETP data series and xLE the LE data series, for a given experiment. Likewise, xETP(t) and xLE(t) correspond to their continuous time counterparts. Essentially, I is an approximation of the ratio of root signal energies (denoted as and respectively) from the transient peaks and the lower envelope. As such, it is expected that data series from bladders with DO will have a larger signal energy in xETP(t) relative to xLE(t), and thus a larger I, than bladders without DO. Additionally, normalizing by should mitigate for differences between bladders in terms of the response of their cg2 series to filling and DO contraction. Such differences between bladders could be due to differences in bladder wall thickness and passive elasticity. Finally, because I is unitless, it is reasonable to make comparisons between experiments, as longer experimental durations and wider ranges in measured cg2 or Pdet are accounted for by normalizing RMS(xETP) by RMS(xLE) as is done in Equation 1.
E. Statistical Analysis
Statistically significant differences in the DO index between DO and non-DO bladders was assessed using a non-parametric hypothesis test, i.e. the Wilcoxon Test (Mann and Whitney, 1947). Receiver operator characteristic (ROC) analysis was also applied as a preliminary evaluation of diagnostic potential. Optimal cut-offs were estimated for the DO index using closest-to-(0,1) criterion (Perkins and Schisterman, 2006) (i.e., the point on the curve nearest to the upper left corner of the ROC plot). We also investigated the potential correlation between the DO index and the patients age and bladder compliance measured from the UDS as they could be covariant with the DO index. The Spearman’s correlation coefficient was calculated for DO for all bladders. The bladder compliance was defined as in (Wyndaele et al., 2011) and had units of ml/cmH2O.
III. RESULTS
Results are presented by first demonstrating the proposed characterization and associated observations on example UBV and UDS measurements from a bladder with DO (Figure 5 (a)) and a bladder without DO (Figure 5 (b)). The characterization is then verified in the remaining bladders and evaluated for its diagnostic potential by statistical analysis of the proposed DO index.
Figure 5 presents example analysis of cg2 and Pdet data series for the DO and non-DO bladders collected over the course of the UDS study. Note that for the cg2 and Pdet data series from the example bladder with DO, several transient peaks are observed (indicated by red arrows in Figure 5a). Furthermore, these peaks appear to be separable from the larger trend of the data series when decomposing the series into ETP and LE signals respectively (See red and blue curves in Figure 5 (d,h)). Although transient fluctuations are observable in the non-DO bladder, their associated signal energy is comparatively low. This pattern is captured in our proposed DO index, I. In particular, a higher value of I is observed in the example DO bladder for both cg2 (I=0.69; See Figure 5(c,d)) and Pdet (I=0.59; See Figure 5 (g,h)) than in the example non-DO bladder (I=0.20 for cg2 and I=0.08 for Pdet; See Figures 5(e,f) and 1(i,j) respectively).
Figure 6 graphically presents these statistical results through a scatter plot of DO indices calculated for UBV/UDS experiments from all 76 patients (Figure 6 a) as well as accompanied ROC curves for cg2 and Pdet (Figure 6 (b,c)). In agreement with the example bladder shown in Figure 5, median DO indices for cg2 and Pdet were larger in bladders with DO (0.43 and 0.70 respectively) than they were in bladders without DO (0.25 and 0.31). The Wilcoxon test suggests statistically significant differences between DO and non-DO bladders for I calculated from cg2 (p<0.01) and from Pdet (p<0.01). Figure 6 (b,c) present the associated ROC curves of I calculated for cg2 and Pdet respectively. The corresponding cut-off values were 0.327 for I calculated for cg2 and 0.567 for I calculated from Pdet. Associated diagnostic performance metrics for these cutoffs, including sensitivities and specificities, are reported in Table 1.
Table 1.
cg2 | Pdet | |
---|---|---|
AUC | 0.76 | 0.81 |
Optimal Cut-off | > 0.327 | > 0.567 |
Sensitivity | 0.70 | 0.70 |
Specificity | 0.75 | 0.83 |
Accuracy | 0.72 | 0.78 |
Although classification characteristics of peak characterization were comparable between cg2 and Pdet (i.e., comparable sensitivity, specificity, and AUC; See Table 1), from Figure 6 (a), it can be seen that a portion of their respective classifications was in disagreement for a minority of the data (i.e., upper left and lower right quadrants marked by the respective cut-off lines). This disagreement corresponded to 31.6 % of the bladders (8/23 DO bladders and 16/53 non-DO bladders).
Age and bladder compliance had nonsignificant Spearman’s correlation coefficients for the DO index calculated on cg2 (ρ = −0.134 and p-value 0.25 for Age and DO-index and ρ = −0.23 and p-value 0.051 for bladder compliance). This was similarly the case for Spearman’s correlations with the DO index calculated from Pdet (ρ = −0.023 and p-value 0.83 for Age and DO-index and ρ = −0.12 and p-value 0.31 for bladder compliance).
IV. DISCUSSION
In this study, we investigated the feasibility of detecting DO through analysis of transient peaks in UBV measurements. We characterized peaks in these measurements through a proposed DO index that characterizes the differing signal energies between peaks and LE. Our results show a statistically significant difference in I calculated for cg2 between DO and non-DO bladders. ROC analysis showed this analysis of cg2 series produced a sensitivity of 0.70 and a specificity of 0.75 for detecting DO. This is comparable to the sensitivity and specificity produced by applying the same analysis to the concurrent Pdet measurements recorded from the UDS (0.70 and 0.83, respectively). Recall that the concurrent Pdet measurement series is effectively the Pdet channel from the UDS chart but with undersampling that matches cg2 and that the analysis was applied to it for validation of the processing approach as well as comparison. Therefore, this comparability suggests promising potential for UBV in detecting DO.
Although the differences in DO classification between cg2 and Pdet observed in Figure 6(a) may be due to random variations, it may also be a consequence of the localized nature of the UBV measurement. Because UBV measures wave propagation along a segment of the bladder wall, it likely only captures the contractile activity of the tissue along the measured segment. In contrast, transient fluctuations of Pdet are a function of the net contractile activity of the detrusor muscle over the full bladder. Currently, there is some preclinical evidence for spatiotemporal variation of detrusor contraction over the bladder wall (Lentle et al., 2015), where bladder contractions appear to propagate spatially along the bladder wall. This might explain why the ROC characteristics for I were similar when calculated from cg2 versus Pdet, while showing differing classifications in 31.6% of the bladders. The development of UBV to allow for spatially resolved measurements along the bladder wall would offer the possibility of investigating this hypothesis in vivo.
The UBV measurements in our study were conducted concurrently with UDS measurements at increasing filling volumes. The advantage of this approach is that the UBV and concurrent UDS measurements were collected at the same time as the UDS charts used for urologist diagnosis of DO (ground truth), which removes the variability that would have been introduced if the UDS examination used to confirm the presence of DO were performed at different times. Another benefit of this approach is that we were able to compare our peak analysis to both cg2 and Pdet that were acquired at the same time. Our results showed that applying our analysis to Pdet from UDS data and cg2 from UBV data provide similar diagnostic performance. This suggests that UBV and UDS provided similar information relative to the proposed diagnostic index. This supports the potential of UBV as a noninvasive alternative to UDS for DO diagnosis.
The benefits of our approach came at the price of some limitations. One such limitation is the study population which included only patients with neurogenic DO. Although patients with idiopathic DO were not investigated here, we expect the UBV response to be the same as observed here for neurogenic DO. The only clinical distinction between idiopathic DO and neurogenic DO is that the latter has no known underlying pathology associated with it. However, the urodynamic observation is the same. Because UBV and UDS responses are driven by the same underlying mechanisms (i.e. changes in bladder wall elasticity and loading), idiopathic DO should produce the same UBV response as neurogenic DO.
Another limitation is that the transducer was episodically removed from the abdomen during the experiment. It can be expected that, because of changes in the probe position as well as the changes in the bladder volume, the ultrasound beam may not target exactly the same location of the bladder for each measurement. However, such errors can be acceptable assuming relatively uniform elasticity and loading along the anterior bladder wall.
A more substantial limitation for this study is that the sampling intervals between UBV measurements were relatively large, especially between filling volumes (Figure 2b). Furthermore, because UBV measurements were collected episodically at the discrete stoppage filling volumes, they were irregularly sampled in time and there was variation in the time intervals between measurements (Figure 2a). Although the results of this study indicate that the components of the processing and preprocessing appear to be robust to such a limitation on average, it is likely that many of the false positives and false negatives observed in this investigation are due to this limitation. For instance, with sparse sampling points and a bladder with infrequent detrusor contractions, the outlier filter may classify a measurement collected at the peak of a contraction as an anomalous outlier. Likewise, the Savitzky-Golay filter used to define the lower envelope of the data series may be overly influenced by a few measurements of large values when sampling is limited, though this is less likely given the effect of the outlier filter that is applied before it. A related source of error is that transient peaks in the full UDS may be missed due to unusually large delays between UBV measurements.
The limitations that accompany the large sampling intervals are acceptable for the purposes of this current study, which was to prove the concept of noninvasive detection of DO through UBV measurement. However, precise characterization of the amplitudes and durations of DO contractions through their associated UBV response will require denser time sampling than was collected here. Refinements to the UBV experimental setup that include dense and regular temporal sampling could potentially produce greater diagnostic strength. Furthermore, such refinements would permit time series and frequency-domain based analysis to be applied to the UBV measurement series, such as was adopted in a study by (Cullingsworth et al., 2018). It should be noted that the peak identification process and DO index proposed in this investigation could be readily applied to measurements with dense and regular temporal sampling with minor adaption and that such improvements would reduce the occurrence of false positives and false negatives due to the effects of sparse sampling on the filters used in the current processing. With all of this in mind, these current results should be considered as a first step in applying UBV toward the diagnosis of DO. Future studies are planned to implement such improvements to our experimental approach to UBV.
Keeping in mind the advantages and limitations of this initial investigation and the need for future refinements to UBV targeted toward assessment of DO, these initial results are promising and motivating for future investigation. While UBV was coupled with UDS in this study for verification purposes, we envision that practical application of UBV toward DO diagnosis would be completely non-invasive. In such a case, the cystometric filling used in these current experiments would be replaced by the natural filling of the bladder after oral hydration as has been done for UBV in (Nenadic et al., 2016). It should be noted that the application of non-invasive UBV still need to be monitored by a clinician in order to position the probe and evaluate factors affecting the measurement. Many factors to effectively applying such a procedure are yet to be determined, including the ideal sampling duration and rate and the preferred filling volume for the detection of DO. Additionally, it should be noted that UDS provides more information to urologists than assessment of DO, such as bladder compliance. Though further development of UBV may provide non-invasive assessment for many of these components due to its correlation with Pdet (Bayat et al., 2017), measurements such as urinary flow rate would require added techniques beyond UBV in order to replace all or most of the its diagnostic applications. Therefore, as UBV’s scope of application is explored relative to lower urinary tract symptoms, it is likely that the design of practical UBV procedures will develop to incorporate such applications.
The potential benefit of UBV over UDS toward assessment of DO is the possibility of sensing DO without the need for catheterization. This would allow for its use in DO diagnosis when UDS cannot be justified due to the risks of morbidity and discomfort (Aoki et al., 2017, Al Mousa et al., 2019). Catheterization may also be a source of irritation to the bladder wall. As noted in (Abrams, 2003), such irritation could explain the high rates of DO (up to 60%) observed in ambulatory UDS (Brown and Hilton, 1997). UBV could offer a means of testing this hypothesis as well as offer an alternative assessment of DO that is closer to everyday living (Abrams et al., 2002). Because ultrasound is portable, and potentially even wearable (Pashaei et al., 2020), UBV may offer an attractive avenue for assessing DO during ambulation which would substantially increase the urologist’s acumen. We envision future investigations of UBV-based detection of DO performed without the use of concurrent UDS and on a larger patient population to investigate its utility under a greater variety of conditions and sources of DO.
V. CONCLUSION
This study investigated the feasibility of detecting DO through UBV, which is a noninvasive ultrasound technique. We found that DO could be detected from UBV measurements using the proposed peak identification processing and DO index. Although preliminary, these initial results suggest that UBV has potential as a noninvasive tool for detecting DO.
ACKNOWLEDGMENTS
The authors would like to thank Dr. Mohammad Mehromommadi, PhD, Dr. Max Denis, PhD, Dr. Mahdi Bayat, PhD, Dr. Viksit Kumar, PhD, Dr. Bae-Hyung Kim, PhD, Dr. Rohit Nayak, Ms. Adriana Gregory and Mr. Jeremy Webb for their assistance in data acquisition at the different time periods during the course of the study. The authors also would like to thank Mr. Duane Meixner for his assistance in collecting patient ultrasound scans; Ms. Cynthia Andrist, Ms. Barbara Foreman and Ms. Pamela Draayer for their assistance in patient recruitment. Research reported in this publication was supported by the National Institute of Diabetes And Digestive And Kidney Disease of the National Institute of Health under Award Number R01DK099231. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
Appendix
Appendix A: Description of the statistical filter used for outlier removal
In order to remove outliers, a statistical outlier filter called a Hampel filter (Pearson et al., 2016) was applied to the cg2 and Pdet data series. This filter works by replacing points determined to be outliers with the median of the sample window. The filter uses the following criteria for identifying outliers: A given sample from the data series, xi, which is the center point of a sliding window with a median of mi, is considered an outlier if
Here, nσ is a threshold number of standard deviations and σi is the estimated standard deviation of the sampling window. The standard deviation of the sampling window is estimated from the median absolute deviation as
(A2) |
Where w is the number of pixels on either side of the centerpoint of the window and the parameter, k, scales the median absolute deviation to the standard deviation. For a standard normal distribution k ≈ 1.483(Hampel, 1974, Rousseeuw and Croux, 1993). For analysis of the UBV data, nσ was set to 6 and w was set to 5 for the following reasons. Although the filter does not permit a fixed time window for unevenly sampled data, it does not modify the signal except where an outlier is detected. Likewise, the window is set to be large relative to the length of the signal (11 points within the window for w = 5 versus ~16 points based on the median sampling interval and median experimental duration.) By setting the outlier threshold conservatively in terms of the number of anomalous outliers it is likely to identify, the filter alters only extreme outlying points in the series and leaves the remaining points unaltered (as illustrated in Figure 3). Because few points are modified by the filter and the sample window is large, the effect of unfixed time window is expected to be minimal. Setting nσ to 6 means that, for a Gaussian distribution, 99.9997% of the data is left unaltered. While we do not speculate that data within the filter window should follow a Gaussian distribution a prior, the filter offers a reliable and objective heuristic for identifying anomalous outliers within the data series that does not rely on human judgment.
Appendix B: Detailed description of ETP and LE series decomposition
The following is a detailed description of the processing steps used to decompose data series of UBV and UDS measurements into ETP and LE series. These steps are illustrated in Figure 4.
(1) A smoothing filter was applied to the raw data series (x) to capture any trend in the data series associated with bladder filling (Figure 4 (a)) Smoothing is applied to the series through a 2nd-order Savitzky-Golay filter(Savitzky and Golay, 1964) through MATLAB’s ‘smooth’ function. The Savitzky-Golay filter applies a (2nd-order) polynomial regression to data within a sliding window. Because the time intervals are irregularly spaced (Figure 2 a,b), a smoothing factor is used in place of a discrete window size. The smoothing factor takes on a value between 0 and 1, which corresponds to the percent of signal energy being attenuated by smoothing (e.g. a smoothing factor of 0.5 would attenuate 50% of the signal energy). The smoothing factor was set to 0.95 in order to filter all fluctuations in the data series except for the main trend of the data.
(2) After smoothing, x is differenced with the filtered data series (xs) and time-points associated with negative values of the differenced series are identified (i.e. time points were (x-xs)<0).The start and end time-points were also included in the identified time points, so that increases at the time boundary are not treated as transient peaks. (See Figure 4 (b) for illustration)
(3) Values from x at the identified time-points are then collected. Linear interpolation is then used to assign values at the time-points that were not identified in step two. The resulting series has the same number of elements as x. We refer to the resulting series as the LE series (xLE). (See Figure 4 (c) for illustration)
(4) The ETP series (xETP) is then determined by differencing the xLE from x. (See Figure 4 (d) for illustration)
The ETP and LE series can then be used for further analysis.
Footnotes
CONFLICT OF INTEREST DISCLOSURE
Mostafa Fatemi is an inventor of US patent 9,345,448 on the technology used in this study. Mostafa Fatemi and the other authors declare no relationship with any company whose products or services may be related to the subject matter of the article. The authors also affirm that they do not have any potential financial interest related to the technology used in this study.
References
- Abrams P 2003. Describing bladder storage function: Overactive bladder syndrome and detrusor overactivity. Urology, 62, 28–37. [DOI] [PubMed] [Google Scholar]
- Abrams P, Cardozo L, Fall M, Griffiths D, Rosier P, Ulmsten U, Van Kerrebroeck P, Victor A & Wein A 2002. The standardisation of terminology of lower urinary tract function: Report from the standardisation sub-committee of the International Continence Society. Neurourology and Urodynamics, 21, 167–178. [DOI] [PubMed] [Google Scholar]
- Adusei SA, Gregory AV, Polley EC, Alizad A & Fatemi M 2021. Does Body Position Affect Ultrasound Measurements of Bladder-Wall Elasticity? Ultrasound in Medicine & Biology, 47, 1115–1119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Al Mousa RT, Al Dossary N & Hashim H 2019. The role of urodynamics in females with lower urinary tract symptoms. Arab Journal of Urology, 17, 2–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Aoki Y, Brown HW, Brubaker L, Cornu JN, Daly JO & Cartwright R 2017. Urinary incontinence in women. Nature Reviews Disease Primers, 3, 19. [DOI] [PubMed] [Google Scholar]
- Bachasson D, Dres M, Niérat M-C, Gennisson J-L, Hogrel J-Y, Doorduin J & Similowski T 2019. Diaphragm shear modulus reflects transdiaphragmatic pressure during isovolumetric inspiratory efforts and ventilation against inspiratory loading. Journal of Applied Physiology, 126, 699–707. [DOI] [PubMed] [Google Scholar]
- Barr RG & Zhang Z 2012. Effects of Precompression on Elasticity Imaging of the Breast. Journal of Ultrasound in Medicine, 31, 895–902. [DOI] [PubMed] [Google Scholar]
- Bayat M, Kumar V, Denis M, Webb J, Gregory A, Mehrmohammadi M, Cheong M, Husmann D, Mynderse L, Alizad A & Fatemi M 2017. Correlation of ultrasound bladder vibrometry assessment of bladder compliance with urodynamic study results. PLOS ONE, 12, e0179598. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bright E, Oelke M, Tubaro A & Abrams P 2010. Ultrasound estimated bladder weight and measurement of bladder wall thickness–useful noninvasive methods for assessing the lower urinary tract? J. Urol, 184, 1847. [DOI] [PubMed] [Google Scholar]
- Brown K & Hilton P 1997. Ambulatory monitoring. International Urogynecology Journal, 8, 369–376. [DOI] [PubMed] [Google Scholar]
- Casella G, Fienberg S & Olkin I 2008. Time Series Analysis, New York, NY, Springer Science+Business Media, LLC. [Google Scholar]
- Cullingsworth ZE, Kelly BB, Deebel NA, Colhoun AF, Nagle AS, Klausner AP & Speich JE 2018. Automated quantification of low amplitude rhythmic contractions (LARC) during real-world urodynamics identifies a potential detrusor overactivity subgroup. PLOS ONE, 13, e0201594. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Drake MJ, Doumouchtsis SK, Hashim H & Gammie A 2018. Fundamentals of urodynamic practice, based on International Continence Society good urodynamic practices recommendations. Neurourol Urodyn, 37, S50–s60. [DOI] [PubMed] [Google Scholar]
- Farag FF & Heesakkers JP 2011. Non-invasive techniques in the diagnosis of bladder storage disorders. Neurourology and Urodynamics, 30, 1422–1428. [DOI] [PubMed] [Google Scholar]
- Fatemi M & Nenadic IZ 2016. SYSTEM AND METHOD FOR NON-INVASIVE DETERMINATION OF TISSUE WALL VISCOELASTICITY USING ULTRASOUND VIBROMETRY. US Patent 9345448 B2.
- Gammie A, Clarkson B, Constantinou C, Damaser M, Drinnan M, Geleijnse G, Griffiths D, Rosier P, Schäfer W & Van Mastrigt R 2014. International continence society guidelines on urodynamic equipment performance. Neurourology and Urodynamics, 33, 370–379. [DOI] [PubMed] [Google Scholar]
- Gammie A, Speich JE, Damaser MS, Gajewski JB, Abrams P, Rosier P, Arlandis S, Tarcan T & Finazzi Agro E 2020. What developments are needed to achieve less-invasive urodynamics? ICI-RS 2019. Neurourol Urodyn, 39 Suppl 3, S36–S42. [DOI] [PubMed] [Google Scholar]
- Gennisson JL, Cornu C, Catheline S, Fink M & Portero P 2005. Human muscle hardness assessment during incremental isometric contraction using transient elastography. Journal of Biomechanics, 38, 1543–1550. [DOI] [PubMed] [Google Scholar]
- Glass Clark S, Nagle AS, Bernardo R, Vinod N, Carucci L, Carroll A, Speich J & Klausner AP 2018. Use of Ultrasound Urodynamics to Identify Differences in Bladder Shape Between Individuals With and Without Overactive Bladder. Female Pelvic Med Reconstr Surg, Publish Ahead of Print. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Goldmark E, Niver B & Ginsberg DA 2014. Neurogenic Bladder: from Diagnosis to Management. Current Urology Reports, 15, 448. [DOI] [PubMed] [Google Scholar]
- Gray T, Phillips L, Li W, Buchanan C, Campbell P, Farkas A, Abdi S & Radley S 2019. Evaluation of bladder shape using transabdominal ultrasound: Feasibility of a novel approach for the detection of involuntary detrusor contractions. Ultrasound, 27, 167–175. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hampel FR 1974. The Influence Curve and Its Role in Robust Estimation. Journal of the American Statistical Association, 69, 383–393. [Google Scholar]
- Hug F, Tucker K, Gennisson JL, Tanter M & Nordez A 2015. Elastography for Muscle Biomechanics: Toward the Estimation of Individual Muscle Force. Exercise and Sport Sciences Reviews, 43, 125–33. [DOI] [PubMed] [Google Scholar]
- Irwin DE, Kopp ZS, Agatep B, Milsom I & Abrams P 2011. Worldwide prevalence estimates of lower urinary tract symptoms, overactive bladder, urinary incontinence and bladder outlet obstruction. BJU International, 108, 1132–1138. [DOI] [PubMed] [Google Scholar]
- Latthe P, Middleton L, Rachaneni S, Mccooty S, Daniels J, Coomarasamy A, Balogun M, Duckett J, Thakar R, Goranitis I, Roberts T, Deeks J & The BUSCG 2017. Ultrasound bladder wall thickness and detrusor overactivity: a multicentre test accuracy study. BJOG: An International Journal of Obstetrics & Gynaecology, 124, 1422–1429. [DOI] [PubMed] [Google Scholar]
- Lentle RG, Reynolds GW, Janssen PWM, Hulls CM, King QM & Chambers JP 2015. Characterisation of the contractile dynamics of the resting ex vivo urinary bladder of the pig. BJU International, 116, 973–983. [DOI] [PubMed] [Google Scholar]
- Liao L 2015. Evaluation and Management of Neurogenic Bladder: What Is New in China? International journal of molecular sciences, 16, 18580–18600. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Loupas T, Powers JT & Gill RW 1995. An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the Doppler equation by means of a two-dimensional autocorrelation approach. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 42, 672–688. [Google Scholar]
- Mann HB & Whitney DR 1947. On a Test of Whether one of Two Random Variables is Stochastically Larger than the Other. The Annals of Mathematical Statistics, 18, 50–60. [Google Scholar]
- Montaldo G, Tanter M, Bercoff J, Benech N & Fink M 2009. Coherent Plane-Wave Compounding for Very High Frame Rate Ultrasonography and Transient Elastography. IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control, 56, 489–506. [DOI] [PubMed] [Google Scholar]
- Nenadic I, Mynderse L, Husmann D, Mehrmohammadi M, Bayat M, Singh A, Denis M, Urban M, Alizad A & Fatemi M 2016. Noninvasive Evaluation of Bladder Wall Mechanical Properties as a Function of Filling Volume: Potential Application in Bladder Compliance Assessment. PLOS ONE, 11, e0157818. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nenadic IZ, Qiang B, Urban MW, De Araujo Vasconcelo LH, Nabavizadeh A, Alizad A, Greenleaf JF & Fatemi M 2013. Ultrasound bladder vibrometry method for measuring viscoelasticity of the bladder wall. Physics in Medicine and Biology, 58, 2675–2695. [DOI] [PubMed] [Google Scholar]
- Panicker JN, Fowler CJ & Kessler TM 2015. Lower urinary tract dysfunction in the neurological patient: clinical assessment and management. The Lancet Neurology, 14, 720–732. [DOI] [PubMed] [Google Scholar]
- Pashaei V, Dehghanzadeh P, Enwia G, Bayat M, Majerus SJA & Mandal S 2020. Flexible Body-Conformal Ultrasound Patches for Image-Guided Neuromodulation. IEEE Transactions on Biomedical Circuits and Systems, 14, 305–318. [DOI] [PubMed] [Google Scholar]
- Pearson RK, Neuvo Y, Astola J & Gabbouj M 2016. Generalized Hampel Filters. Eurasip Journal on Advances in Signal Processing, 18. [Google Scholar]
- Perkins NJ & Schisterman EF 2006. The inconsistency of "optimal" cutpoints obtained using two criteria based on the receiver operating characteristic curve. American journal of epidemiology, 163, 670–675. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Peyronnet B, Mironska E, Chapple C, Cardozo L, Oelke M, Dmochowski R, Amarenco G, Game X, Kirby R, Van Der Aa F & Cornu JN 2019. A Comprehensive Review of Overactive Bladder Pathophysiology: On the Way to Tailored Treatment. European Urology, 75, 988–1000. [DOI] [PubMed] [Google Scholar]
- Rachaneni S, Mccooty S, Middleton LJ, Parker VL, Daniels JP, Coomarasamy A, Verghese TS, Balogun M, Goranitis I, Barton P, Roberts TE, Deeks JJ & Latthe P 2016. Bladder ultrasonography for diagnosing detrusor overactivity: test accuracy study and economic evaluation. Health Technol Assess, 20, 1–150. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rosier PFWM, Kuo H-C, De Gennaro M, Gammie A, Finazzi Agro E, Kakizaki H, Hashim H & Toozs-Hobson P 2019. International Consultation on Incontinence 2016; Executive summary: Urodynamic testing. Neurourology and Urodynamics, 38, 545–552. [DOI] [PubMed] [Google Scholar]
- Rosier PFWM, Schaefer W, Lose G, Goldman HB, Guralnick M, Eustice S, Dickinson T & Hashim H 2017. International Continence Society Good Urodynamic Practices and Terms 2016: Urodynamics, uroflowmetry, cystometry, and pressure-flow study. Neurourology and Urodynamics, 36, 1243–1260. [DOI] [PubMed] [Google Scholar]
- Rousseeuw PJ & Croux C 1993. Alternatives to the Median Absolute Deviation. Journal of the American Statistical Association, 88, 1273–1283. [Google Scholar]
- Sacco E, Tienforti D, D'addessi A, Pinto F, Racioppi M, Totaro A, D'agostino D, Marangi F & Bassi P 2010. Social, economic, and health utility considerations in the treatment of overactive bladder. Open access journal of urology, 2, 11–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Savitzky A & Golay MJE 1964. Smoothing and Differentiation of Data by Simplified Least Squares Procedures. Analytical Chemistry, 36, 1627–1639. [Google Scholar]
- Vejdani-Jahromi M, Freedman J, Trahey GE & Wolf PD 2019. Measuring Intraventricular Pressure Using Ultrasound Elastography. Journal of Ultrasound in Medicine, 38, 1167–1177. [DOI] [PubMed] [Google Scholar]
- Wang Y, Li H, Guo Y & Lee W-N 2019. Bidirectional Ultrasound Elastographic Imaging Framework for Non-invasive Assessment of the Non-linear Behavior of a Physiologically Pressurized Artery. Ultrasound in Medicine & Biology, 45, 1184–1196. [DOI] [PubMed] [Google Scholar]
- Winters JC, Dmochowski RR, Goldman HB, Herndon CD, Kobashi KC, Kraus SR, Lemack GE, Nitti VW, Rovner ES, Wein AJ, American Urological A, Society of Urodynamics FPM & Urogenital R 2012. Urodynamic studies in adults: AUA/SUFU guideline. J Urol, 188, 2464–72. [DOI] [PubMed] [Google Scholar]
- Wu XY, Cheng Y, Xu SF, Ling Q, Yuan XY & Du GH 2021. Prophylactic Antibiotics for Urinary Tract Infections after Urodynamic Studies: A Meta-Analysis. Biomed Res Int, 2021, 6661588. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wyndaele JJ, Gammie A, Bruschini H, De Wachter S, Fry CH, Jabr RI, Kirschner-Hermanns R & Madersbacher H 2011. Bladder Compliance What Does it Represent: Can We Measure it, and is it Clinically Relevant? Neurourology and Urodynamics, 30, 714–722. [DOI] [PubMed] [Google Scholar]