Skip to main content
Diagnostics logoLink to Diagnostics
. 2021 Oct 26;11(11):1981. doi: 10.3390/diagnostics11111981

Tracking SARS-CoV-2: Novel Trends and Diagnostic Strategies

Linda P Guaman-Bautista 1, Erick Moreta-Urbano 1, Claudia G Oña-Arias 1, Marbel Torres-Arias 2, Nikolaos C Kyriakidis 3,4, Koray Malcı 5,6, Nestor Jonguitud-Borrego 5,6, Leonardo Rios-Solis 5,6, Espiridion Ramos-Martinez 7, Andrés López-Cortés 8, Carlos Barba-Ostria 9,*
Editor: Anna Baraniak
PMCID: PMC8621220  PMID: 34829328

Abstract

The COVID-19 pandemic has had an enormous impact on economies and health systems globally, therefore a top priority is the development of increasingly better diagnostic and surveillance alternatives to slow down the spread of the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). In order to establish massive testing and contact tracing policies, it is crucial to have a clear view of the diagnostic options available and their principal advantages and drawbacks. Although classical molecular methods such as RT-qPCR are broadly used, diagnostic alternatives based on technologies such as LAMP, antigen, serological testing, or the application of novel technologies such as CRISPR-Cas for diagnostics, are also discussed. The present review also discusses the most important automation strategies employed to increase testing capability. Several serological-based diagnostic kits are presented, as well as novel nanotechnology-based diagnostic methods. In summary, this review provides a clear diagnostic landscape of the most relevant tools to track COVID-19.

Keywords: COVID-19, nucleic acid amplification test, antigen testing, CRISPR-based diagnostics, nanotechnology-based diagnostics, automation

1. Introduction

In late 2019, multiple cases of atypical pneumonia caused by a then unknown pathogen were reported in Wuhan, province of Hubei in China. The disease named COVID-19 quickly spread globally and, after considerable research efforts, the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was identified as the pathogen causing the disease [1]. Coronaviruses are a large group of viruses that can infect a wide range of vertebrate hosts [2]. Diseases caused by each of those viruses show different clinical manifestations, usually affecting the respiratory tract in humans and digestive disorders in other animals [3]. In the last 20 years, three highly pathogenic beta coronaviruses have emerged from zoonotic events and have caused human disease [4]. In February 2003, the severe acute respiratory syndrome-related coronavirus 1 (SARS-CoV-1) was first reported in Asia. According to the World Health Organization (WHO), a total of 8,098 people worldwide became infected and 774 died (global case–fatality ratio (CFR) of 11%) [5] across two dozen countries in North America, South America, Europe, and Asia [6]. Since September 2012, WHO has been notified of 2562 laboratory-confirmed cases of infection with Middle East respiratory syndrome-related coronavirus (MERS-CoV), with a global CFR of 34.4% since 2012 [7]. Finally, the current disease caused by SARS-CoV-2, to date, has resulted in 217,914,903 confirmed cases and 4,523,984 deaths in 223 affected countries, areas, and territories [8]. There have been many attempts to estimate the global CFR of COVID-19 [9,10,11,12,13,14], but it has proven challenging, mainly because different regions of the world are experiencing different stages of the pandemic [15] affecting those estimates.

SARS-CoV-2 belongs to the Coronaviridae family, subfamily Orthocoronavirinae, genus Betacoronavirus, subgenus Sarbecovirus, and the species severe acute respiratory syndrome-related coronavirus [16]. Coronaviruses are enveloped, positive-sense, single-stranded RNA viruses that are distributed broadly among humans, other mammalian, and avian hosts [17]. The genome of SARS CoV-2 (NCBI Reference Sequence: NC_045512.2) [18] is very similar to the genomes of other known SARS-CoV and SARS-related coronaviruses, including the one that caused the SARS epidemic in 2003 (SARS CoV, NCBI Reference sequence: NC_004718.3) [19] and other related coronaviruses (e.g., MERS-CoV) [20]. Therefore, studies of protein structure and function from those coronaviruses have been important for understanding the function and mechanisms of SARS-CoV-2 proteins [21]. In addition, recent structural and functional studies on SARS-CoV-2 proteins have provided critical knowledge of SARS-CoV-2 biology [17].

The SARS-CoV-2 genome is about 30 kb (Zhou et al., 2020), and it encodes at least 29 proteins, including 16 non-structural proteins (NSP), 4 structural proteins:(S), envelope (E), membrane (M), and nucleocapsid (N), and 9 accessory proteins (Figure 1). Some structural and non-structural proteins (S, E, M, N, 3CL protease, papain-like protease, RNA polymerase [22], and helicase) have been suggested as viable antiviral drug and diagnostic targets [23,24,25]. Although coronaviruses accessory proteins are frequently considered as non-critical for viral replication in vitro, some have an important role in virus-host interactions in vivo [26] and in vitro [27]. Different studies have shown that SARS-CoV-2 is a pathogen with a high efficiency in invading the cells of its host [28,29]. Currently, there is enough evidence supporting the preponderant role of angiotensin-converting enzyme 2 (ACE2) as a key factor allowing the binding and entry of SARS-CoV-2 to host cells [30,31,32].

Figure 1.

Figure 1

Structure of SARS-CoV-2 and diagnostic targets: SARS-CoV-2 structure has two components: viral genome (RNA) and viral proteins. Tests based on the detection of nucleic acids (RT-qPCR, RT-LAMP, CRISPR-Cas-based, NGS) target different regions of the viral RNA. Antigen tests aim at the detection of viral proteins such as spike protein (S). Serological tests detect host antibodies produced in response to SARS-COV-2 infection.

2. Diagnosis

Due to the pandemic nature of COVID-19, testing has become an indispensable tool for monitoring the diagnosing and spread of the disease [33,34]. Multiple approaches and strategies for detecting SARS-CoV-2 have been developed. In this context, a clear understanding of the principles and differences between these strategies, their impact on the results, their correct interpretation, and the best conditions for their adequate use, is crucial [34]. In general, the major approaches for COVID-19 testing are: (1) detection of viral structural components (RNA and proteins) and (2) detection of molecules produced by the host immune system in response to SARS-CoV-2 infection.

RNA is a key structural component of SARS-CoV-2 and, given the previous development of several technologies allowing for the efficient isolation, amplification, and detection of specific nucleic acid sequences, the viral genome (RNA) is the most targeted molecule used for the detection of SARS-CoV-2. (Figure 1) [35,36]. This is the principle of widely used detection methods such as RT-qPCR, RT-LAMP, and next generation sequencing based assays, among others. A second group of methods have been developed targeting SARS-CoV-2 proteins in samples. These methods include viral antigen detection using proteomics-based tests and the nanosensor-based detection of viral proteins.

On the other hand, the concentration of the different types of antibodies (IgM, IgG, IgA) produced by the host immune system in response to SARS-CoV-2 can be used as a tool for surveillance to better understand how much of the population has been infected with SARS-CoV-2, as well as how the virus is spreading through the population over time [37] (Figure 2). The detection of antibodies is the common principle of all serological tests.

Figure 2.

Figure 2

Timeline of SARS-CoV-2 detection using nucleic acid-based, antigen and serological tests: Detection of viral nucleic acids is useful from 5 to 15 days after the onset of symptoms. Because many nucleic acid detection tests include nucleic acid amplification, these tests can give positive results for prolonged periods. Due to the nature of antibody production and the variability of immune responses between individuals, serological tests can be used for seroprevalence studies and surveillance from the second week after the onset of symptoms. Antigen tests are useful for the detection of infected individuals from 5 to 15 days after the onset of symptoms.

It is important to highlight that although serological tests are very useful from an epidemiological perspective, given that antibodies can take several days or even weeks to be produced and may stay in your blood for several weeks or more after recovery, serological tests should not be used to diagnose COVID-19 and are not approved for the diagnosis of acute, active SARS-CoV-2 infection [38,39].

3. Nucleic Acid Amplification-Based Tests (NAAT) for Detection of SARS-CoV-2

Since the SARS-CoV-2 genomic sequence was made public in early January 2020, several primers, probes, and detection strategies have been developed and optimized worldwide (Table 1). Despite the increasing availability of assays, most strategies are based on the same basic principle: the accurate and sensitive identification of SARS-CoV-2 RNA sequences. As is further discussed in this section, the major differences of all those assays are in the method used to measure/detect the presence of viral genetic material.

Table 1.

Diagnostic tests based on detection of nucleic acids.

Test Name Manufacture Country Technology LoD Sample Approval Ref
1copy COVID-19 qPCR Multi Kit Korea
1 drop
RT-qPCR 0.2 copies/µL Nasopharyngeal and nasal swab and wash EUA 1/5/2020 [40]
TRUPCR SARS-CoV-2 Kit India
3B Blackbio Biotech India Kilpest India subsidiary
RT-qPCR 10 copies/µL Nasopharyngeal and oropharyngeal swabs, anterior nasal swab, and mid-turbinate nasal swabs, nasopharyngeal aspirates/washes or nasal aspirates, and bronchoalveolar lavage EUA 18/6/2020 [41]
SARS-CoV-2 and Influenza A and B RT-qPCR Detection Kit China
3D Medicines
RT-qPCR 5 copies/reaction - CE mark 3/2020 [42]
Abbott RealTime SARS-CoV-2 EUA test USA
Abbott
RT-qPCR 0.1 copies/µL Nasopharyngeal swabs, Oropharyngeal swabs EUA 18/3/2020 [43]
CareStart COVID-19 MDx RT-PCR USA
Access Bio
RT-qPCR 5.4 NDU/µL Nasopharyngeal, oropharyngeal and nasal swabs, and nasopharyngeal wash/aspirate or nasal aspirate and bronchoalveolar lavage EUA 7/7/2020 [44]
Acupath COVID-19 Real-Time (RT-PCR) Assay USA
Acupath Laboratories
RT-qPCR 25 copies/µL Nasopharyngeal swab, nasopharyngeal aspirate, and bronchoalveolar lavage EUA 29/6/2020 [45]
UltraGene Combo2Screen SARS-CoV-2 assay Luxemburg
Advanced Biological Laboratories
RT-qPCR 10E-6 TCID50/mL Nasopharyngeal swab EUA submission pending, CE mark 5/2020 [46]
RealStar SARS-CoV-2 RT-PCR Kits (1.0 and U.S versions) Germany
Altona Diagnostics
RT-qPCR 1.0E-01 PFU/mL Human respiratory swabs EUA 22/4/2020, CE mark 4/2020 [47]
Altru Dx SARS-CoV-2 RT-PCR assay USA
Altru Diagnostics
RT-qPCR 0.625 copies/µL Nasal, midturbinate, nasopharyngeal, and oropharyngeal swab EUA 30/4/2020 [48]
Bosphore Novel Coronavirus (2019-nCoV) Detection Kit Turkey
Anatolia Geneworks
RT-qPCR - Human respiratory sample CE mark 2/2020 [49]
BioCode SARS-CoV-2 Assay USA
Applied BioCode
RT-qPCR 1.7E-2 TCID50/mL Nasopharyngeal swabs (NPS), oropharyngeal swabs (OPS), nasal swabs or bronchoalveolar lavage EUA 6/15/2020 [50]
Linea COVID-19 RT-PCR test USA
Applied DNA Sciences
RT-qPCR 1.25 copies/µL Nasopharyngeal and oropharyngeal swabs, mid-turbinate nasal swabs, nasopharyngeal washes/aspirates or nasal aspirates, and bronchoalveolar lavage EUA 5/13/2020 [51]
Aspirus SARS-CoV rRT-PCR Assay USA
Aspirus Reference Laboratory
RT-qPCR 0.5 copies/µL Nasal, mid-turbinate, nasopharyngeal, oropharyngeal swab specimens and bronchoalveolar lavage specimens EUA 6/1/2020 [52]
Assurance SARS-CoV-2 Panel USA
Assurance Scientific Laboratories
RT-qPCR 9 copies/µL Nasal, nasopharyngeal, or oropharyngeal swabs EUA 5/15/2020 [53]
A*STAR Fortitude 2.0 Singapore A*STAR, Tan Tock Seng Hospital of Singapore RT-qPCR 25 copies/reaction Nasal pharyngeal swab Singapore Health Sciences Authority provisional authorization [54]
Avera Institute for Human Genetics SARS-CoV-2 Assay USA
Avera Institute for Human Genetics
RT-qPCR 1.6 copies/µL Nasopharyngeal, nasal, and oropharyngeal swab specimens EUA 5/22/2020 [55]
Avellino SARS-CoV-2/COVID-19 (AvellinoCoV2) USA
Avellino Lab
RT-qPCR 18 NDU/µL Nasopharyngeal swab and oropharyngeal swab EUA 3/25/2020 [56]
Viro-Q SARS-CoV-2 kit Germany
BAG Diagnostics
RT-qPCR - - CE mark 4/2020 [57]
ThermoFisher—Applied Biosystems TaqPath COVID-19 Combo Kit USA
Rutgers University Clinical Genomics Laboratory
RT-qPCR 0.250 GCE/µL Nasopharyngeal swabs, nasopharyngeal aspirate (nasal aspirate), and nasopharyngeal aspirate (nasal aspirate), and bronchoalveolar lavage (BAL) EUA 4/10/2020 [58]
Cobas SARS-CoV-2 Test USA
Roche
RT-qPCR 0.003 TCID50/mL Nasopharyngeal and oropharyngeal swab samples EUA 3/12/2020, CE mark 2020 [59]
GSD NovaPrime SARS-CoV-2 (COVID-19) Real-Time PCR test Hungary
Gold Standard Diagnostics/Eurofins Technologies
RT-qPCR 3.75 copies/reaction Nasal wash/swab, nasopharyngeal wash/swab, oropharyngeal swab and bronchoalveolar lavage CE mark 5/2020 [60]
QiaStat-Dx Respiratory SARS-CoV-2 Panel USA
Qiagen
RT-qPCR 180 NDU/µL Nasopharyngeal swab under CDC’s EUA [61]
Rida Gene SARS-CoV-2 Germany
R-Biopharma
RT-qPCR 50 copies/reaction Human throat and nasopharyngeal swabs CE mark 5/2020 [62]
Standard M nCoV Real-Time Detection Kit Korea
SD Biosensor
RT-qPCR 0.25 copies/μL Nasoropharyngeal, nasal, and mid-turbinate nasal swab, and sputum EUA 4/23/2020 [63]
ExProbe SARS-CoV-2 Testing Kit Taiwan
TBG Biotechnology
RT-qPCR - Nasopharyngeal and oropharyngeal swabs, anterior nasal and mid-turbinate nasal swabs EUA 6/10/2020 [64]
Quick SARS-CoV-2rRT-PCR Kit USA
Zymo Research
RT-qPCR 2.5 E3 GEC/μL Nasal, nasopharyngeal, mid-turbinate or oropharyngeal swabs), and lower respiratory specimens (such as sputum, tracheal aspirates, and bronchoalveolar lavage) EUA 5/7/2020 [65]
Fulgent COVID-19 by RT-PCR Test USA
Fulgent Genetics/Fulgent Therapeutics
RT-qPCR 20 copies/µL Nasal, nasopharyngeal, and oropharyngeal swabs EUA 5/15/2020 [66]
SARS-CoV-2 RT-PCR test Germany
Centogene
RT-qPCR 5 copies/µl Dry oropharyngeal swabs EUA 7/1/2020 [67]
Xpert Xpress SARS-CoV-2 test USA
Cepheid
RT-qPCR 0.0001 PFU/µL Nasopharyngeal, oropharyngeal, nasal, or mid-turbinate swab and/or nasal wash/aspirate EUA 3/20/2020 [68]
HDPCR SARS-CoV-2 real-time PCR assay USA
ChromaCode
RT-qPCR 0.250 copies/µL Nasopharyngeal swabs oropharyngeal swabs, anterior nasal swabs, midturbinate nasal swabs, nasal aspirate, nasal wash, and bronchoalveolar lavage (BAL) specimens EUA 6/9/2020 [69]
Hymon SARS-CoV-2 Test Kit China
Dba SpectronRx
RT-qPCR 1.2 copies/µl Nasal, mid-turbinate, nasopharyngeal, and oropharyngeal swab specimens) EUA 5/22/2020 [70]
DSL COVID-19 Assay USA
Diagnostic Solutions Laboratory
RT-qPCR 18 NDU/µL Nasopharyngeal swab EUA 6/25/2020 [71]
Simplexa COVID-19 Direct USA
DiaSorin Molecular
RT-qPCR 6 NDU/µL Nasal swab, nasopharyngeal swab, nasal wash/aspirate, and BAL EUA 3/19/2020, CE mark 4/2020 [72]
Lilly SARS-CoV-2 Assay USA
Eli Lilly
RT-qPCR 1 copy/µL Nasopharyngeal swabs, oropharyngeal (throat) swabs, anterior nasal swabs, mid-turbinate nasal swabs, nasal aspirates, nasal washes and bronchoalveolar lavage EUA 7/27/2020 [73]
Ampiprobe SARS-CoV-2 Test System USA
Enzo Biochem/Enzo Life Sciences
RT-qPCR 0.280 copies/µL Nasopharyngeal swabs EUA 7/7/2020 [74]
Euroimmun/PerkinElmer USA
EuroRealTime SARS-CoV-2
RT-qPCR 0.150 copies/µL Nasal, mid-turbinate, nasopharyngeal, oropharyngeal swabs and bronchioalveolar lavage EUA 6/8/2020, CE mark 3/2020 [75]
Fosun COVID-19 RT-PCR Detection Kit China
Fosun Pharma USA
RT-qPCR 0.3 copies/µL Anterior nasal swabs, mid-turbinate nasal swabs, nasopharyngeal swabs, oropharyngeal swabs, sputum, lower respiratory tract aspirates, bronchoalveolar lavage, and nasopharyngeal wash/aspirate or nasal aspirate EUA 4/17/2020 [76]
NeoPlex COVID-19 Detection Kit Korea
GeneMatrix
RT-qPCR 5.4 NDU/µL Nasopharyngeal swabs EUA 5/14/2020 [77]
GB SARS-CoV-2 Real-Time RT-PCR Taiwan
General Biologicals
RT-qPCR 0.1 copies/µL Nasopharyngeal swabs CE mark 4/2020 [78]
Genetron SARS-CoV-2 RNA Test China
Genetron
RT-qPCR 10 copies/µL Oropharyngeal, nasopharyngeal, anterior nasal and mid-turbinate nasal swab EUA 6/5/2020 [79]
Helix COVID-19 test USA
Helix
RT-qPCR 1 GCE/µL Nasopharyngeal and oropharyngeal swabs EUA 7/23/2020 [80]
Panther Fusion SARS-CoV-2 assay USA
Hologic
RT-qPCR 1 × 10−5 TCID50/µL Nasopharyngeal (NP), nasal, oropharyngeal (OP) swab specimens and lower respiratory tract (LRT) specimens EUA 3/16/2020 [81]
Smart Detect SARS-CoV-2 rRT-PCR Kit USA
InBios International
RT-qPCR 12 GCE/reaction Nasopharyngeal swab, anterior nasal swab and mid-turbinate nasal swab EUA 4/7/2020 [82]
IDT 2019-novel coronavirus kit USA
Integrated DNA Technologies/Danaher
RT-qPCR - Oropharyngeal, nasopharyngeal, anterior nasal and mid-turbinate nasal swab under CDC’s EUA [83]
SARS-CoV-2 Assay USA
Integrity Laboratories
RT-qPCR 2.5 copies/µL Nasal, nasopharyngeal and oropharyngeal swab EUA 4/13/2020 [84]
ProTect Covid-19 kit Singapour
JN Medsys
RT-qPCR <2% Upper respiratory nasopharyngeal swabs Under CDC’s EUA, CE mark 4/2020, Singapore HAS provisional authorization; Philippines FDA [85]
COVID-19 Coronavirus Real Time PCR Kit China
Jiangsu Bioperfectus Technologies
RT-qPCR 0.350 copies/µL Nasopharyngeal swabs, oropharyngeal (throat) swabs, anterior nasal swabs, mid-turbinate nasal swabs, nasal aspirates, nasal washes, bronchoalveolar lavage (BAL) fluid and sputum EUA 18/6/2020 [86]
Curative-Korva SARS-Cov-2 Assay USA
KorvaLabs
RT-qPCR 0.200 copies/µL Oropharyngeal (throat) swab, nasopharyngeal swab, nasal swab, and oral fluid specimens EUA 16/4/2020 [87]
Idylla SARS-CoV-2 Test Belgium
Biocartis
RT-qPCR 0.5 copies/µL Nasopharyngeal swab CE mark 11/2020 [88]
BioFire Respiratory Panel 2.1-EZ (RP2.1-EZ) USA
BioMérieux/BioFire Diagnostics
RT-qPCR 6 NDU/µL Nasopharyngeal swab EUA 2/10/2020 [89]
VitaPCR Influenza A and B/SARS-CoV-2 assay Singapore
Credo Diagnostics Biomedical
RT-qPCR 2.73 copies/μl Nasopharyngeal swab CE mark 10/2020 [90]
Genetworx Covid-19 Nasal Swab Test USA
Genetworx
RT-qPCR 0.274 copies/µL Nasal swab EUA 15/12/2020 [91]
EuroRealTime SARS-CoV-2/Influenza A/B Germany
Euroimmun/PerkinElmer
RT-qPCR 1.8 NDU/µL Throat swab CE mark 12/2020 [92]
MassArray SARS-CoV-2 Panel USA
Agena Bioscience
RT-PCR/MALDI-TOF 0.3 copies/µL Nasopharyngeal swab, oropharyngeal swab, and BAL EUA 10/26/2020 CE mark 9/2020 [93]
Bio-Rad SARS-CoV-2 ddPCR Test USA
Bio-Rad Laboratories
ddPCR 0.625 copies/µL Nasopharyngeal, anterior nasal and mid-turbinate swab specimens as well as nasopharyngeal wash/aspirate and nasal aspirate specimens EUA 1/5/2020 [94]
iAMP COVID-19 Detection Kit USA
Atila BioSystems
Isothermal amplification (OMEGA amplification) 10 copies/µL Nasal, nasopharyngeal (NP), and oropharyngeal (OP) swabs EUA 10/4/2020 [95]
ID Now COVID-19 USA
Abbott
Isothermal amplification (proprietary enzymes) 0.125 GCE/µL Direct nasal, nasopharyngeal or throat swabs EUA 27/3/2020 [96]
Color SARS-CoV-2 LAMP Diagnostic Assay USA
Color
RT-LAMP 0.75 copies/µL Nasopharyngeal (NP) swabs, oropharyngeal (OP) swabs, anterior nares (AN) swabs, mid-turbinate nasal (MTN) swabs, NP wash/aspirate or nasal aspirates, and bronchoalveolar lavage specimens EUA 5/18/2020, amended 24/7/2020 [97]
2019-nCoV detection kit China
Rendu Biotechnology
RT-LAMP - Nasal, nasopharyngeal, and oropharyngeal swab China NMPA 3/2020 [98]
AQ-TOP COVID-19 Rapid Detection Kit Korea
Seasun Biomaterials
RT-LAMP 7 copies/µL Oropharyngeal and nasopharyngeal swab specimens, anterior nasal and mid-turbinate nasal swabs, nasopharyngeal wash/aspirate or nasal aspirate specimens, bronchoalveolar lavage (BAL) and sputum from individuals EUA 21/5/2020 [99]
Lucira Health USA
Lucira COVID-19 All-In-One Test Kit
RT-LAMP 0.9 copies/µL Nasal swab EUA 17/11/2020 [100]
Poplar SARS-CoV-2 TMA Pooling assay USA
Poplar Healthcare
TMA - Nasal, nasopharyngeal, and oropharyngeal swab EUA 3/8/2020 [101]
COVIDSeq Test USA
Illumina
Next generation gene sequencing 5.4 NDU/µL Nasopharyngeal (NP), oropharyngeal (OP), and mid-turbinate (MT) nasal swabs EUA 9/6/2020 [102]
Fulgent COVID-19 by NGS USA
Fulgent Genetics/MedScan Laboratory
Next generation gene sequencing 3.6 NDU/µL Nasal, nasopharyngeal, and oropharyngeal swabs EUA submission pending [103]
Sherlock CRISPR SARS-CoV-2 kit USA
Sherlock Biosciences
RT-LAMP and CRISPR-Cas 13 6.75 copies/µL Nasopharyngeal and oropharyngeal swab samples EUA 6/5/2020 [104]
SARS-CoV-2 RNA DETECTR Assay USA
UCSF Clinical Labs at China Basin
CRISPR-Cas12 20 copies/µL Nasopharyngeal swabs, oropharyngeal (throat) swabs, mid-turbinate nasal swabs, anterior nasal swabs, nasopharyngeal wash/aspirate or nasal aspirate EUA 9/7/2020 [105]

LoD: Limit of detection. TCID50: Median Tissue Culture Infectious Dose. GCE: Genome copy equivalents. NDU: RNA NAAT detectable units. PFU: plaque-forming unit.

3.1. Reverse Transcription-Polymerase Chain Reaction (RT-PCR)

RT-PCR is a widely used method in molecular biology and is considered the gold standard for the detection of SARS-CoV-2 [106] and other respiratory viruses [107,108,109]. Early detection of infection caused by SARS-CoV-2 by RT-PCR relies on the efficient detection of viral genomic sequences in clinical samples. Based on the established practices for other respiratory infections such as influenza, the nasopharyngeal swab has been widely adopted as the preferred sampling technique for SARS-CoV-2 along with other techniques such as saliva, oropharyngeal, nasal mid-turbinate, and nasal swabs [110]. However, some studies have also reported using blood, sputum, feces, and urine samples for SARS-CoV-2 detection [111,112].

Most of the described protocols for SARS-CoV-2 detection require the isolation of viral RNA from biological samples as a first step. RNA isolation must be performed in a careful and quick manner to avoid degradation caused by nucleases released from cells and present in the environment [113]. First, the sample is mixed with a lysis buffer to release the genetic material and preserve RNA integrity. The following steps are designed to eliminate DNA, protein, and other cell components in order to isolate the RNA by serial washes [113]. After RNA isolation, the next step for RT-PCR is the conversion of viral genome (RNA) into DNA. This process requires the specific binding of DNA primers to viral RNA sequences and the activity of an RNA-dependent DNA polymerase (reverse transcriptase) to generate a short complementary DNA copy (cDNA) of the viral RNA. In the real-time RT-PCR, the polymerase amplifies the viral cDNA, and the synthesis is monitored in real-time by using a fluorescent dye or a sequence-specific DNA probe labeled with a fluorescent molecule and a quencher molecule [114]. To date, most molecular diagnostic tests developed have targeted different SARS-CoV-2 genomic regions, including the ORF1b or ORF8 regions and the nucleocapsid (N), spike (S) protein, RNA-dependent RNA polymerase (RdRP), or envelope (E) genes, while the most-used genes employed as targets in diagnostic tests are N gene, ORF1b, ORF1a and the E gene [115].

Since the development of PCR in the 1980s, many assays used for detecting nucleic acid have been created, and a growing number of research and diagnostic laboratories around the world perform PCR routinely. The wide adoption of this method has been, and still is, critical because it allowed for the rapid implementation of RT-qPCR-based diagnostics facilities and increased the SARS-CoV-2 diagnostic capabilities in many regions. In addition, given that specific assays can be designed based on any DNA or RNA sequence, this technology is highly flexible and is relatively simple to apply to the detection of SARS-CoV-2 variants, and of course, for the detection of future infectious microbes. In a fashion similar to other nucleic acid detection methods, RT-qPCR-based detection heavily relies on the quantity, quality, and integrity of DNA or RNA in samples, therefore the development of simpler, cheaper, and more efficient sampling alternatives is a challenge for improving SARS-CoV-2 RT-qPCR assays.

3.2. Assays Based on Nucleic Acid Isothermal Amplification

Introduced in the early ‘90s, isothermal amplification offers an alternative to Polymerase Chain Reaction (PCR)-based detection methods [116,117]. In contrast to PCR, isothermal amplification can be performed without the need of thermocycling equipment. Instead, it can be carried at one reaction temperature and under relatively simple conditions (e.g., in a water bath) [118]. More than ten types of isothermal amplification methods have been developed [119]. Some are based on DNA replication, enzyme-free nucleic acid assembly, and the use of enzymes for digestion [118]. In this review, we focus on six different techniques, and, where specified, examples are given for its application in COVID-19 diagnostics.

Isothermal amplification detection tests have become popular because they can be performed with simpler laboratory equipment and produce results in shorter times than conventional RT-qPCR assays. Despite this, significant challenges should be addressed to fully exploit this testing technology as a massive and frequent SARS-CoV-2 (and other infectious agents) detection tool, including: (i) reducing the number of micropipetting steps needed and (ii) avoiding the need of cold chain conditions to transport or store reagents.

3.3. Nucleic Acid Sequence-Based Amplification (NASBA)

Self-sustained sequence replication, better known as Nucleic Acid Sequence-Based Amplification (NASBA) was originally designed for RNA amplification [120]. In this technique, RNA hybridizes to the forward primer for the synthesis of its complementary chain to obtain cDNA using RNAase H. The newly synthesized cDNA contains a promoter region for the formation of an intermediate double-stranded (ds) cDNA, which is formed by using a second primer [121]. Several copies of antisense RNA strands to the target RNA are produced when T7 DNA-dependent RNA polymerase carries out transcription from the dsDNA. Then, the antisense RNA and cDNA become templates for the continuous cycle carried out by reverse transcriptase. The reaction can happen at 41 °C and it usually takes 1.5 to 2 h to complete. Colorimetric assays, gel electrophoresis, and real-time fluorescence can be used for amplicon detection [118]. This method has been used in combination with next generation sequencing (NGS) for the detection of SARS-CoV-2 [121].

3.4. Loop-Mediated Isothermal Amplification (LAMP)

Loop-Mediated Isothermal Amplification or LAMP is a low-cost diagnostic technique which relies on the activity of a strand-displacement DNA polymerase and four to six primers to exponentially amplify the target sequence [122]. This method amplifies nucleic acid under isothermal conditions at about 65 °C [123]. The forward inner primer (FIP) and the backward inner primer (BIP) are crucial to the amplification. They both contain two functional sequences: the first one for priming at an early stage and the second for self-priming at a later stage. The functional sequences described for FIP and BIP correspond to the target sequence of the target DNA. The result of the amplification is stem-loop DNA that can be detected by real-time assays [118].

As a diagnostic method, LAMP has several advantages, such as avoiding the need of a thermal cycler, making LAMP simpler and cheaper [124]. Moreover, due to the use of several primers, this technique has a high specificity, allowing it to discriminate a single nucleotide difference. Besides, the amplification is highly efficient (DNA can be amplified 109–1010 times in 15–60 min) [125]. In addition, LAMP is fast due to the activity of the strand-displacement polymerase eliminating the need for a denaturation step, allowing the amplification and detection to occur in a single step. Additionally, in the case of low-resource settings, the amplification can be monitored by using a cheap turbid meter and, after, the addition of an intercalating agent [126]. LAMP has been used previously as a tool to diagnose diseases such as malaria [122] and tuberculosis [127].

In the context of COVID-19, RT-LAMP is the technology used by several assays of different companies (Table 1) worldwide. In addition, tests based on RT-LAMP have been approved by the FDA, CDC, and other regulatory agencies in many countries [128]. RT-LAMP has proven to be effective in detecting viral RNA, showing high specificity and sensitivity, thus being a potentially robust technique during the global COVID-19 pandemic [129].

In addition to its technical advantages for diagnostics, LAMP is a versatile method that has been coupled to other strategies to improve diagnostics. A clear example of this is STOP (SHERLOCK Testing in One Pot), a CRISPR -based method for the diagnosis of SARS-CoV-2 [130]. This method relies on the amplification performed by LAMP with further detection using CRISPR-Cas12 [130]. This is a very simple test that can be carried out at room temperature and, as the name implies, in a single pot.

3.5. Rolling Circle Amplification (RCA)

RCA is a simple and efficient amplification method that relies on the activity of highly processive DNA or RNA strand-displacing polymerases to generate a long single-stranded DNA (ssDNA) or RNA molecule [131]. RCA has attracted attention for diagnostics because of its simplicity and capability of being performed at a constant low temperature (room temperature to 37 °C) in a relatively short time (approximately 90 min). Importantly, due to its molecular principle, RCA avoids the generation of false-positive results, a common problem of other detection assays. This method is based the use of a circular probe (C-probe) formed by the interaction and subsequent ligation of the analyzed sequence with the 5′ and 3′ ends of a single-stranded DNA probe [132]. This C-probe hybridizes with the primer, which is elongated by the strand-displacement polymerase, resulting in the exponential amplification of the target DNA or RNA. It must be mentioned that DNA ligase catalyzes the ligation of the C-probe with the target only in the case of perfect complementarity of the 3′-and 5′-end sequences [133] avoiding false-positives [134].

RCA has been used before for detecting single nucleotide polymorphisms [135,136], M. tuberculosis genomic DNA [137], SARS-CoV [138], Ebola, and other pathogens [139]. Recently, a study described a modified RCA assay capable of detecting up to picomolar concentrations of synthetic viral DNA strands of SARS-CoV-2, Influenza A (H1N1pdm09), and Influenza B [140]. As in other nucleic acid amplification methods, RCA output can be detected by measuring fluorescence, by using colorimetric chemistry, or by coupling the results with other readout techniques [139].

3.6. Transcription-Mediated Amplification (TMA)

TMA is a technology based on isothermal amplification, that can be performed in a single tube [141] to detect target RNA sequences. This technology is more efficient than RT-PCR [141], as it can produce 10 billion amplicons in 15–30 min [142] and it does not require a thermal cycler [142,143], making TMA a more affordable alternative.

TMA employs two primers: one of them contains the promoter sequence for RNA-polymerase. The first step of the process consists of the hybridization of the primer–promoter to the target RNA. Then, a reverse transcriptase creates a copy of DNA based on the RNA target by extension. Later, an RNAse H degrades the RNA in the resulting RNA–DNA duplex. The second primer binds to the DNA copy, and the reverse transcriptase creates a new strand of DNA, resulting in a double-stranded DNA molecule. Then, an RNA polymerase initiates the transcription, since it recognizes the promoter sequence in DNA [142]. The detection is conducted by the HPA separation/detection procedure. First, the amplicon is hybridized with acridinium ester-labeled DNA probes. Then, a chemical reaction separates the hybridized probes from the unhybridized probes. The final step consists of the addition of the reagents into the reaction tubes to produce the chemiluminescent signal, which is measured by a luminometer [142].

TMA is an interesting alternative that is being used to diagnose COVID-19 [144] with a sensitivity of 98.15%, compared to RT-qPCR (96.25%) with a detection limit of 5.5 × 102 copies in 1 of 5 samples [145]. These results included 116 nasopharyngeal swabs [144,145]. The Aptima SARS-CoV-2 assay consists of target capture, TMA, and Dual Kinetic Assay (DKA). In the first step, the target RNA is isolated by capturing oligomers which have a complementary sequence to the target and a string of deoxyadenosine residues. Then, the oligomers:target complex is then captured out of the solution by decreasing the temperature of the reaction to room temperature. The deoxyadenosine region covalently attaches to magnetic microparticles which contain poly-deoxythymidine molecules. The microparticles are pulled to the side of the reaction tube by magnets. Then, the supernatant is aspirated, and the particles are washed. After the target is captured, the sample passes to TMA. This assay amplifies and detects two conserved regions of the ORF1ab gene in a single reaction [146]. The detection step is mediated by DKA detection technology, which is based on HPA. However, DKA uses two different acridinium ester molecules in two different nucleic acid probes [142].

3.7. Recombinase Polymerase Amplification (RPA)

Another polymerase-based amplification method is the recombinase polymerase amplification (RPA). This technology is an isothermal amplification method developed by Piepenburg et al., in 2006 by using proteins involved in recombination, synthesis, and DNA repair [147]. The mechanism in which this method works starts with a recombinase that binds to oligonucleotides in the presence of ATP and high molecular polyethylene glycol (also referred as crowding agent). This mixture forms a recombinase–primer complex which screens the template DNA until a homologous sequence is found. Then, a primer stop occurs so that the primer interacts with the template while single-stranded binding proteins stabilize the displaced DNA strand in order to prevent it from ejecting the complex from the target DNA. Finally, the recombinase breaks away and a polymerase binds to the 3′ end of the primer and starts to synthesize a new strand of DNA in the presence of dNTPs [148]. If this process is repeated in several cycles, linear (with one primer) and exponential (with a forward and a reverse primer) amplification can be achieved [147,149].

Certain aspects must be taken into account in the design/usage of RPA components. Primers used in RPA were thought to be 30–35 bases in length, however some studies have reported that PCR primers (≈19–21 mers) can be used, and that successful amplification can be achieved [143,150,151]. Additionally, 5-end guanines should be avoided while 3-end cytidines and guanines improve amplification performance, while GC content should be between 30% and 70%. Although RPA can work at a range of temperatures from 22–45 °C, optimal temperatures tend to be in the range of 37–42 °C [152,153], and multiple equipment can be employed to regulate the temperature such as incubators, heating blocks, or even body heat, facilitating its use as a point-of-care test in the field [154]. RPA can successfully amplify dsDNA, ssDNA, cDNA, and methylated DNA from a large variety of organisms such as viruses, bacteria, fungi, animals, and plants. Moreover, RPA has proven to be effective in different sample types such as body fluids and animal products [148].

Two main advantages of RPA over PCR have been pointed out by various studies: reaction temperature and reaction time. The latter is notorious if we compare the 15–40 min run time of RPA with the usual run times that PCR takes to show results [149]. Additionally, several detection methods have been developed to visualize results in RPA, such as solid phase RPA combined with electrochemical, ring resonators, and fluorescent and colorimetric detection [148,155,156].

Although RPA has been studied and used for the detection of various viral and bacterial pathogens, it appears to be confined to research-only activities for now [132]. Ebola detection by RPA studies has suggested that the diagnosis with recombinant polymerase amplification can be deployed in coming Ebola outbreaks due to the low amount of equipment required by this technique: a heating block and a centrifuge [157]. Some studies have used RPA as a step in CRISPR diagnostic systems, while RPA as the main detection technique has also been tested for COVID-19 [158,159].

4. Next Generation Sequencing (NGS) for Detection of SARS-CoV-2

First introduced as Massively Parallel Signature Sequencing (MPSS), next generation sequencing has been in the market since around 2004 [160]. The COVIDSeq Test is the first NGS test approved by the US Food and Drug Administration for COVID-19 diagnostics [161]. This test is designed for RNA detection of SARS-CoV-2 in nasopharyngeal, oropharyngeal, and mid-turbinate nasal swabs with a capacity of up to 3072 samples to be processed in 12 h [102].

Although NGS has proven to be a useful detection method, as in other diagnostic methods, as in the case of other diagnostic strategies, the test should not be used as a diagnostic strategy by itself, as results from the NGS test need to be confirmed with features such as clinical observation, patient history, and epidemiological information [102]. A critical disadvantage of NGS limiting its broad application for COVID diagnostics is the high cost of reagents and materials needed and the requirement of complex laboratory equipment [162,163].

5. CRISPR-Based Tests

Since the early days of the pandemic, the high transmission rate of SARS-CoV-2, as well as the high frequency of asymptomatic, but still infectious, cases, took the world by surprise. Even more than a year after the first cases, the lack of a high-throughput, high sensitivity, simple and low-cost diagnostic test still limits the massive detection of the positive individuals, inhibiting not only early medical intervention but also allowing for further disease transmission [164].

Rapid and reliable diagnostic challenges can be tackled by new-generation Clustered Regularly Interspaced Short Palindromic Repeats (CRISPR)-based platforms that offer fast pathogen-tailored detection assays, as well as the possibility to simultaneously detect hundreds of pathogens per individual in a robust multiplex fashion [165] as a point-of-care molecular diagnostic test [166].

Clustered Regularly Interspaced Short Palindromic Repeats are a group of DNA sequences found in bacteria and archaea which serve as an immune system or an immune “hard drive” [167]. Although this immune hard drive was first discovered more than three decades ago [168], it was not until 2012 that their potential as genetic editing scalpels for prokaryotic, eukaryotic, and even mammalian cells was appreciated [169,170]. Therefore, it comes as no surprise that, after 2012, a group of academic laboratories working in collaboration with pharma counterparts leverage CRISPR-Cas systems to develop the DNA Endonuclease-Targeted CRISPR Trans Reporter (DETECTR) [171] and Specific High Sensitivity Enzymatic Reporter UnLOCKing (SHERLOCK) [172] developed the first pathogen diagnostic platforms based on this novel technology.

The DETECTR system uses “off-the-shelf” reagents, portable heat blocks, and a lateral flow readout, similar to pregnancy tests, which can achieve a point-of-care setup that can detect HPV16 and HPV18 in patient samples within 1 h [171]. The reported sensitivity of DETECTR assay is 95%, with a specificity close to 100% [171,173]. Due to the simple principle of targeting Cas proteins, CRISPR-Cas detection assays can be easily programmed to detect practically any pathogen by modifying the guide DNA or RNA that recognizes and directs the Cas enzyme to a specific sequence of the pathogen’s genetic material, including coronaviruses [171,174].

In a recent study, a research group from UCSF reported the design and validation of a SARS-CoV-2 DETECTR diagnostic assay [175]. As in other nucleic acid-based diagnostics, the assay’s initial steps are comprised of nasopharyngeal or oropharyngeal swab sampling and the ensuring of RNA extraction and a one-step reverse transcription and loop-amplification of viral genetic material (RT–LAMP) with primers that target the E and N genes of SARS-CoV-2 [141]. Then, the CRISPR step involves incubation, with Cas12-guide RNAs targeting sequences found in three SARS-like coronaviruses (SARS-CoV-2, bat-SL-CoVZC45, and SARS-CoV), and the detection of target sequences inducing a conformational change and cleavage by lbCas12a enzyme [175]. Activated lbCas12a starts cleaving nearby single-stranded DNA nonspecifically, including the fluorescent-quencher probe, releasing the fluorophore and increasing the fluorescence in the final readout [176]. The limit of detection of this method is reported to be around 10 copies per μL of reaction [175]. According to the authors, the assay showed no cross-reaction with other coronaviruses or influenza strains and showed a lower sensitivity than RT-qPCR; nine out of eleven RT-qPCR positive confirmed samples were positive using this assay [177,178]. Furthermore, the authors carried out an additional step, running 30 RT-qPCR positive and 30 negative samples, and demonstrated a positive predictive value of 95% and a negative predictive value of 100% for the SARS-CoV-2 DETECTR assay [175].

An alternative CRISPR-based detection method named All-In-One-Dual CRISPR-Cas12a (AIOD-CRISPR) was recently developed. The method uses a dual guide RNA approach to augment specificality and in an isothermic, one-pot amplification; the CRISPR readout reaction enables its use in a point-of-care setup, delivering results in 40, 20, or 1 min depending on the target and conditions [179]. As a proof of concept, HIV and SARS-CoV-2 RNA derived from plasmids were detected at a load of 1.2 and 4.6 copies/µL, respectively [180]. The method was further evaluated by detecting HIV RNA extracted from human plasma samples, exhibiting a performance comparable to RT-qPCR tests [180]. A study has used AIOD-CRISPR to test clinical swabs and the results showed to be consistent with the RT-qPCR methods approved by the CDC [179,181].

In 2017, a group of scientists from MIT and Broad Institute reported the development of Specific High Sensitivity Enzymatic Reporter UnLOCKing (SHERLOCK) [182]. This CRISPR-based technique for the detection of pathogens uses a similar approach as the DETECTR, while harnessing Cas13a and Cas13b (instead of Cas12a) enzymatic activity to produce a detectable fluorescence as a readout [182]. Cas13 orthologs recognize single-strand RNA sequences and exhibit a promiscuous RNase activity detected by including a fluorochrome-quencher probe. This method has been previously used to detect specific strains of Zika and Dengue virus to distinguish pathogenic bacteria, genotype human DNA, and to identify cell-free tumor DNA mutations [172].

Recently, Zhang and colleagues have presented a detailed SHERLOCK-based protocol for detecting SARS-CoV-2, starting with RNA purified from patient samples [158]. According to the presented data, the test can detect SARS-CoV-2 target sequences at a concentration of 10–100 copies per microliter of input within 60 min. In addition, once RNA has been extracted, the test can be performed using a dipstick without requiring elaborate instrumentation [158]. Currently, there is a SARS-CoV-2 SHERLOCK detection kit, as indicated in Table 1. Additionally, a high-throughput platform called CARMEN-Cas13 was developed by Pardis Sabeti’s lab. CARMEN-Cas13 simultaneously differentiates all 169 human-associated viruses, with at least 10 published genome sequences including SARS-CoV-2 [183]. Using a similar approach, the same lab developed a CRISPR-based surveillance tool for the detection of 67 viral species and subspecies, including SARS-CoV-2, phylogenetically related viruses, and viruses with similar clinical presentation [184]. This protocol has been tested with clinical samples [185].

Another CRISPR-based SARS-CoV-2 detection method worth mentioning is Cas13-based, Rugged, Equitable, Scalable Testing (CREST). CREST was designed to address three of the main hurdles—reagent accessibility, equipment availability, and cost—that limit the scalability of Cas13-based testing particularly in low-resource areas [186]. The rationale of CREST is using affordable reagents and equipment for RNA extraction, amplification, and detection. Final output is displayed in low-cost and easy-to-use fluorescent visualizers. According to the authors, the sensitivity of CREST is equivalent to RT-qPCR. This method has been successfully used as a rapid SARS-CoV-2 surveillance tool in asymptomatic college students [186].

Cas-based detection depends heavily on the quality of the sample and the integrity and quality of the extracted RNA. In addition, due to the involvement of various enzymatic steps, the robustness of the method still has room for improvement. To massively deploy sensitive, reliable, and easy-to-use Cas-based detection assays, it is important to optimize clinical performance and enhance the robustness of the assays. Given the simple principle of nucleic acid detection by Cas-based methods detection, it is expected that the next generations of SARS-CoV-2 tests will show enhanced clinical performance and will be optimized for the detection, not only of the original SARS-CoV-2 sequences, but of novel variants in the population as well.

6. Serology-Based Tests

Since the early days of epidemiology, serological tests have been a very useful tool to assess previous infections and track the community spreading of diseases [187], including SARS-CoV-2 [188]. In this context, serological tests can be useful for (1) estimating epidemiological variables such as the degree of transmission in the community and its burden [189], (2) performing mass testing to reduce the exposure of susceptible individuals to the virus [190], and (3) identifying individuals with a strong immune response to the virus and whose isolated antibodies can, potentially, be used for a convalescent plasma trial [4,191].

Several antibody detection platforms have been developed over the last 30 years, including enzyme-linked immunosorbent (ELISA) assays and lateral-flow assays [192]. Serological tests are designed to detect antibodies produced by the host in response to specific pathogens, in this case SARS-CoV-2, and, ideally, to minimize cross-reactivity to antibodies generated in response to other viruses or bacterial pathogens [193]. However, given the nature of the antigen–antibody interactions and the structural similarities between antigens from related viruses, cross-reactivity may exist.

Although the most common serologic tests available measure antibodies that bind to viral S and N proteins, not all of these antibodies are considered neutralizing/protective, therefore results from these kind of tests should not be used for identifying individuals who are protected against SARS-CoV-2 infection. In general, serological tests have lower sensitivity and specificity than average RT-qPCR-based tests. Negative results should be interpreted with caution, as negative serology does not definitively rule out acute infection [194].

Serologic assays for SARS-CoV-2 are currently being cleared by the U.S. Food and Drug Administration (FDA) for emergency use and are available and include indirect or lateral flow ELISA assays [195].

During the current pandemic, serological tests have been important for detecting the previous infection of SARS-CoV-2 and became popular because some are cheaper and simple to use. Despite this, they should not be used to substitute molecular testing for detecting acute infections. Some of the advantages of this kind of test are its low-price and simplicity to use. To improve SARS-CoV-2 serological testing, the main challenges are improving the sensitivity and specificity of the assays to enhance its clinical performance. Table 2 shows some serological tests developed to detect SARS-CoV-2.

Table 2.

Serological tests that have some certification to date, FDA. Include: % sensitivity, %, specificity, cross reactivity, interferences, mark, country.

Diagnostic Country
Manufacturer
Tech Specificity Sensitivity Cross Reactivity Interferences Date EUA Issued Ref
Babson Diagnostics aC19G1 USA
Babson Diagnostics, Inc.
IgG CLIA 100% 100% Anti-HIV 1 + 2, Anti-HCV, CMV IgG, Anti-HBs, Anti-HAV EIA N/A 23/6/2020 [196]
Access SARS-CoV-2 IgG USA
Beckman Coulter, Inc.
IgG CLIA 99.6% 96.8% Human chorionic gonadotropin (hCG), HIV antibody, Influenza antibody, Influenza A antibody, Influenza B antibody, Measles antibody, Mycoplasma pneumoniae IgG, Parvovirus B19 antibody, Respiratory pathogen antibodies, Respiratory syncytial virus (RSV) antibody Hemoglobin, Bilirubin (conjugated), Bilirubin (unconjugated), Triglycerides (Intralipid) 26/6/2020 [197]
Diazyme DZ-Lite SARS-CoV-2 IgG CLIA Kit USA, Diazyme Laboratories, Inc. IgG CLIA 97.4% 100% Influenza A H1N1 IgM/IgG, Influenza A H7N9 IgM/IgG, Rhinovirus Type A IgM/IgG, Rotavirus IgM/IgG, Human coronavirus HKU1 IgM/IgG, Human coronavirus NL63 IgM/IgG, ANA Triglycerides, Hemoglobin, Rheumatoid Factor, Anti-Mitochondrial, HAMA, Total IgG, Total IgM, Interferon α, Ribavirin 8/7/2020 [198]
SCoV-2 Detect IgG ELISA USA, InBios International, Inc. IgG ELISA 98.9% 97.8% VIH 1 y 2, hepatitis C and B Hemoglobin Bilirubin Triglycerides Cholesterol 10/6/2020 [199]
SARS-CoV-2 RBD IgG test USA, Emory Medical Laboratories IgG ELISA 96.4% 100% Anti-Influenza B, Anti-HCV, Anti-HBV, Anti-Haemophilus, Anti-Rhinovirus influenzae, ANA, Anti-HIV N/A 15/6/2020 [200]
xMAP SARS-CoV-2 Multi-Antigen IgG Assay USA, Luminex Corporation IgG FMIA 99.3% 96.3% N/A (dipotassium EDTA) 16/7/2020 [201]
ADVIA Centaur SARS-CoV-2 IgG (COV2G) Germany, Siemens Healthcare Diagnostics Inc. IgG Semi-quantitative 99.9% 100% Chlamydia trachomatis IgM, Cytomegalovirus (CMV) IgM, Epstein Barr virus (EBV) IgG, Epstein Barr virus (EBV) IgM, Hepatitis A virus (HAV) IgM, Hepatitis B core (Anti-HBc) IgM, Hepatitis B core (Anti-HBc) total antibody, Hepatitis C virus (HCV) antibody, Herpes simplex virus (HSV) IgM Hemoglobin, Bilirubin (conjugated), Bilirubin (unconjugated), Triglycerides (Intralipid), Biotin 31/7/2020 [202]
Atellica IM SARS-CoV-2 IgG (COV2G) Germany, Siemens Healthcare Diagnostics Inc. IgG Semi-quantitative 99.9% 100% Human chorionic gonadotropin (hCG), Human immunodeficiency virus (HIV) antibody, Influenza antibody, Influenza A antibody, Influenza B antibody, Measles antibody Hemoglobin, Bilirubin (conjugated), Bilirubin (unconjugated), Triglycerides (Intralipid), Biotin, Cholesterol, Protein, total 7/31/2020 [203]
Anti-SARS-CoV-2 ELISA (IgG) Germany, EUROIMMUN US Inc. Serology IgG 99.6% 86.7% Influenza, Acute EBV infection, Rheumatoid factor, other Human coronavirus Hemoglobin, triglycerides, and bilirubin 4/5/2020 [204]
SARS-CoV-2 IgG assay USA, Abbott Laboratories Inc. Serology IgG 99% 100% CMV, Hepaitis A,B, RSV, Rubella, Herpes virus Adenovirus, pregnant female, lupus 26/4/2020 [205]
LIAISON SARS-CoV-2 S1/S2 IgG Italy, DiaSorin Inc. Serology IgG 99.3% 97.6% Anti-Human CoV OC43; Anti-Human CoV HKU1, 4 Anti-Human CoV unknown strain. Triglycerides, Hemoglobin, Unconjugated bilirubin 24/4/2020 [206]
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack USA, Ortho-Clinical Diagnostics, Inc. Serology IgG 100% 83.3% Adenovirus Antibody, Influenza A IgG, Influenza A IgM, Influenza B IgG, Influenza B IgM, Coxsackie Virus Antibody, Echovirus Antibody, Polio Virus, Anti-respiratory syncytial virus (RSV), HCV Antibody, Anti Nuclear Antibody Bilirubin (conjugated), Bilirubin (unconjugated), Biotin, Hemoglobin, Intralipid 24/4/2020 [207]
COVID-19 ELISA IgG Antibody Test USA, Mount Sinai Laboratory Serology IgG 100% 92% Varicella, Infleunza, Hepatitis, HIV, CMV Ascorbic Acid, Hemoglobin, Bilirubin, Albumin, Triglyceride 15/4/2020 [208]
SCoV-2 Detect IgM ELISA USA, InBios International, Inc. IgM ELISA 100% 100% Anti-Influenza A/B, Anti-Hepatitis B, Anti-Hepatitis C, Anti-Nuclear Antibody, Rheumatoid Factor, Human Anti-Mouse Antibody, Anti-HIV, Anti-Respiratory Syncytial Virus, Normal Human Sera Hemoglobin, Bilirubin (conjugated and unconjugated), Triglycerides, Buffer (SDB), Cholesterol 30/6/2020 [199]
VIDAS SARS-CoV-2 IgM France, bioMérieux SA IgM ELFA 99.4% 100% SARS-CoV(-1) Infection (2005), SARS-CoV(-1) Infection (2020), HCoV-NL63 Infection, HCoV-229E Infection, HCoV-OC43 Infection HCoV-HKU1 Infection MERS-CoV Infection, Acute EBV infections with heterophile antibodies, ANA Hemoglobin, Lipids, Albumin, Bilirubin (conjugated), Bilirubin (unconjugated) 6/8/2020 [209]
RightSign COVID-19 IgG/IgM Rapid Test Cassette China, Hangzhou Biotest Biotech Co., Ltd. IgM and IgG Lateral Flow 100% 93.3% Anti-FLU A, Anti-FLU B, Anti-Respiratory Syncytial, Virus Anti-Adenovirus, Anti-HBsAg, Anti-Syphilis, Anti-H. Pylori, Anti-HIV, Anti-HCV, Anti-SARS-COV Acetaminophen, Caffeine, Albumin, Acetylsalicylic Acid, Gentisic Acid, Ethanol, Ascorbic Acid 4/6/2020 [210]
LYHER Novel Coronavirus (2019-nCoV) IgM/IgG Antibody Combo Test Kit (Colloidal Gold) China, Hangzhou Laihe Biotech Co., Ltd. IgM and IgG Lateral Flow 98.8% 96.7% H1N1-1~H1N1-3, H7N9-1~H7N9-2, ANA-1, Staphyl. -1~Staphyl.-2, EBV-1~EBV-5, RSV-1~RSV-2, Chlamydia-1~Chlamydia-3 N/A 19/6/2020 [211]
Assure COVID-19 IgG/IgM Rapid Test Device China, Assure Tech. (Hangzhou Co., Ltd.) IgM and IgG Lateral Flow 100% 98.8% Human coronavirus HKU1 IgM/IgG, Human coronavirus NL63 IgM/IgG Icteric (Bilirubin), Lipemicicines Acetylsalicylic acid Ascorbic acid (Vitamin C), Amoxicillin 6/7/2020 [212]
Sienna-Clarity COVIBLOCK COVID-19 IgG/IgM Rapid Test Cassette Finland, Salofa Oy IgM and IgG Lateral Flow 99.2% 94.9% HCV, HBV, ANA Metapneumovirus Ascorbic Acid, Hemoglobin, Bilirubin, Albumin, Triglyceride 13/7/2020 [213]
Rapid COVID-19 IgM/IgG Combo Test Kit USA, Megna Health, Inc. IgM and IgG Lateral Flow 97.5% 100% Influenza A Virus, Influenza B Virus, Adenovirus, Rotavirus and Mycoplasma Pneumoniae. Rheumatoid Factor, Bilirubin, Triglyceride, Hemoglobin 17/7/2020 [211]
CareStart COVID-19 IgM/IgG USA, Access Bio, Inc. IgM and IgG Lateral Flow 100% 98.8% Anti-Influenza A, Anti-HCV, Anti-229E (alpha coronavirus), Anti-OC43 (beta coronavirus), Anti-HKU1 (beta coronavirus), Antinuclear antibodies (ANA), Anti-respiratory syncytial virus, Anti-Rhinovirus Acetaminophen, HAMA, Acetylsalicylic acid, Hemoglobin, Albendazole, Ibuprofen, Chloroquine diphosphate, Rifampicin 24/7/2020 [214]
BIOTIME SARS-CoV-2 IgG/IgM Rapid Qualitative Test China, Xiamen Biotime Biotechnology Co., Ltd. IgM and IgG Lateral Flow 98.8% 100% Anti-OC43 (beta coronavirus), Anti-HKU1 (beta coronavirus), Antinuclear antibodies (ANA) N/A 24/7/2020 [215]
Vibrant COVID-19 Ab Assay USA, Vibrant America Clinical Labs IgM and IgG CLIA 98.3% 97.1% Anti-Influenza A, Anti- Influenza B, Anti-HCV, Anti-HBV, ANA, Anti-respiratory syncytial virus, Anti-Haemophilus influenzae Bilirubin, Triglycerides Hemoglobin Rheumatoid Factor (RF) Cholesterol, HAMA, Ribavirin, Levofloxacin, Azithromycin, Ceftriaxone sodium 4/6/2020 [216]
Anti-SARS-CoV-2 Rapid Test China, Autobio Diagnostics Co. Ltd. Serology IgM and IgG 97% 93% Other human coronavirus, Influenza HAMA positive sample, Rheumatoid factor, Antinuclear antibody (ANA), Anti-mitochondrial antibody (AMA), Bilirubin 24/4/2020 [217]
DPP COVID-19 IgM/IgG System Usa, Chembio Diagnostic System, Inc. Serology IgM and IgG 100% 95% Human coronavirus, Zika, Influenza A, B, Mononucleosis, Chikungunya, Yellow fever virus, Dengue virus Biotin Hemoglobin 14/4/2020 [218]
qSARS-CoV-2 IgG/IgM Rapid Test USA, Cellex Inc. Serology IgM and IgG 96.0% 93.8% Human coronavirus, Adenovirus, Influenza A, B, Rhinovirus, Chlamydia pneumoniae, Streptococcus pneumoniae, Mycobacterium tuberculosis, Mycoplasma pneumoniae Hemoglobin, Bilirubin (conjugated), Triglycerides, Cholesterol 1/4/2020 [219]
COVID-19 IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) USA, Healgen Scientific LLC Serology IgM and IgG 96.7% 97% Influenza A virus IgG, Influenza B virus IgG, Respiratory syncytial virus IgG, Adenovirus IgG, Rhinovirus IgG, Human metapneumovirus IgG, Mycoplasma pneumoniae IgG, Chlamydia pneumoniae IgG, HCV IgG, Haemophilus influenza IgG, HBV core antibody IgG Ascorbic Acid, Hemoglobin, Bilirubin, Albumin, Triglyceride 29/5/2020 [210]
Dimension EXL SARS-CoV-2 Total antibody assay (CV2T) USA, Siemens Healthcare Diagnostics Inc. Total Antibody CLIA 99.9% 100% Anti-influenza A, Anti-influenza B, AntiHBV Antinuclear, antibody (ANA), Hepatitis B core antigen (anti-HBc) IgM, Hepatitis B surface antigen (HBs Ag), Hepatitis C virus (HCV) antibody, HIV antibody, influenza antibody Hemoglobin, Bilirubin, conjugated, Bilirubin, Lipemia (Intralipid) 8/6/2020 [220]
WANTAI SARS-CoV-2 Ab ELISA China, Beijing Wantai Biological Pharmacy Enterprise Co., Ltd. Total Antibody ELISA 100% 94.5% HBsAg and antibodies to HIV 1/2, HCV, TP N/A 5/8/2020 [221]
WANTAI SARS-CoV-2 Ab Rapid Test China, Beijing Wantai Biological Pharmacy Enterprise Co., Ltd. Total Antibody Lateral Flow 98.8% 100% alpha COV 229E, alpha COV NL63, beta COV OC43, beta COV HKU1 Flu A, Flu B, HCV, HBV, ANA N/A 10/7/2020 [222]
New York SARS-CoV Microsphere Immunoassay for Antibody Detection USA, Wadsworth Center, New York State Department of Health Serology Total Antibody 99.1% 99% HCV, HIV, Measles, Mumps, Rubella, Varicella Zoster virus, WNV, Herpes Simplex virus, ZIKV, Enterovirus, Rheumatoid factor Biotin Hemoglobin 30/4/2020 [223]
Platelia SARS-CoV-2 Total Ab assay USA, Bio-Rad Laboratories, Inc. Serology Total Antibody 98% 99% Influenza, Mycoplasma pneumoniae, Rheumatoid factor, Other human coronavirus 29/4/2020 [224]
Elecsys Anti-SARS-CoV-2 Germany, Roche Diagnostics Serology Antibody 99.8% 100% Other human coronavirus, Influenza Biotin Hemoglobin 2/5/2020 [225]

Antigen Testing

Antigen testing is a type of immunoassay designed to detect the presence of one or more specific proteins of the pathogen, allowing the crucial identification of infected individuals (symptomatic or asymptomatic), so that they can receive the appropriate care, follow public health measures, such as self-isolation, and so that contact tracing can be implemented without delay [226].

In the case of COVID-19, The FDA has issued an EUA for the first antigen assay, the Sofia 2 SARS Antigen FIA test, which employs immunofluorescence-based lateral flow technology in a sandwich design used to detect viral nucleocapsid protein in nasopharyngeal and nasal swab samples with test results within 15 min [227].

Antigen tests may not be as accurate as molecular tests, but they are more accessible in terms of availability and ease of use, and can be used to amplify tests without laboratories, including frequent retesting if necessary. Modeling studies have shown that an availability of results within 15–20 min and the frequency of testing are more important than sensitivity [226].

SARS-CoV-2 antigen tests developed during 2020 have comparable clinical performance to RT-qPCR [228], being useful for the detection of symptomatic and asymptomatic individuals [227]. Moreover, since antigen testing is significantly cheaper (5 to 10 times cheaper than the average RT-qPCR test) [229], simpler, and faster to use (approximately 15 min) than other diagnostic tests, this technology has become a powerful tool for reducing the spread of COVID-19, particularly in community settings such as universities, schools, and offices [230]. Moreover, a frequent method of antigen testing has been implemented as a safe way to reopen economies while controlling the spread [226]. In a recent study, Chaimayo et al., demonstrated that SARS-CoV-2 antigen detection (Standard™ Q COVID-19 Ag kit) shows a sensitivity (98.33%; 95% CI, 91.06–99.96%) and specificity (98.73%; 95% CI, 97.06–99.59%) comparable with the real-time RT-PCR assay [228]. Another advantage of antigen testing is that this type of test is the most likely to detect a positive person when at the peak of their infection, where people are most able to transmit the virus to others.

Regarding its use, WHO [8] has suggested the direct detection of SARS-CoV-2 viral proteins by nasal swabs and other respiratory secretions by lateral flow immunoassay, giving results in less than 30 min, as they offer the possibility of rapid, inexpensive, and early detection of the most infectious cases of COVID-19 in suitable settings. At the time of revision of this article, 13 tests have been granted Emergency Use Authorization from the FDA [231] (Table 3).

Table 3.

FDA-approved antigen-based diagnostic tests.

Diagnostic Country,
Manufacturer
Technology Specificity Sensitivity Ref
Lateral Flow, Visual Read USA, Ortho clinical Diagnostic, Inc. Chemiluminescence immunoassay 98.50% 97.10% [232]
LumiraDx SARS·CoV-2 Ag Test Spain, LumiraDx UK Ltd. Chromatographic Digital immunoassay 96.50% 98% [233]
BD Veritor System for Rapid Detection England, Becton Dickinson Chromatographic immunoassay 100% 84% [234]
Clip COVID Rapid Antigen Test USA, Luminostics, Inc. Lateral Flow, Immunoluminescent Antigen 100% 100% [235]
CareStart COVID-19 Antigen test USA, Access Bio, Inc. Lateral Flow, Antigen 98% 99% [44]
Ellume COVID-19 Home Test Australia, Ellume Limited Test Lateral Flow, Fluorescence 100% 100% [236]
Sofia 2 SARS Antigen FIA Germany, Quidel Corporation Lateral Flow, Fluorescence Antigen 100% 80% [237]
QuickVue SARS Antlgen Test USA, Quidel Corporatlon Lateral Flow, Visual Read 98% 98% [238]
BinaxNOW COVID-19 Ag Card Home USA, Abbott Diagnostics Lateral Flow, Visual Read 98% 98% [239]
BinaxNOW COVID-19 Ag card USA, Abbott Diagnostics Scarborough, Inc. Lateral Flow, Visual Read 98.50% 97.10% [240]
Sofía 2 Flu+ SARS Antlgen FIA Germany, Quidel Corporatlon Lateral Flow, Fluorescence, Instrument Read, Multl-Analyte 100% 80% [241]
Sampinute COVID-19 Antlgen MIA USA, Celltrion, Inc. Magnetic force assisted electrochemical 100% 100% [242]
Simoa SARS-CoV-2 N Protein Antigen test USA, Quanterix Paramagnetic Microbead-based 98.50% 97.10% [243]

7. Nanotechnology for COVID-19 Diagnostics

The synthesis of nanoparticles (NPs), the development of nanosensors and microchips, and the use of devices have become essential tools for the diagnosis of various diseases, including COVID-19 [244]. In recent months, an increasing number of studies have shown that nucleic acid sequences can be detected and play an essential role in diagnosis. However, the detection of single nucleotide variations, such as polymorphisms or small nucleotide changes in viral variants, has proven to be challenging [245].

A particularly relevant technology useful for pathogen and antibody detection purposes are gold NP-based immunoassays. In these assays, antibodies are coupled to gold NPs that, in the presence of a viral antigen or antibody, form a tertiary complex, leading to the immobilization and aggregation of the complex [246]. Other examples of the use of nanotechnology are: Biosensors based on graphene field effect transistors (FETs) coupled to a specific antibody against SARS-CoV-2 spike protein [247]; Dual-function plasmonic biosensors combining plasmonic photothermal effect (PPT) and localized surface plasmon resonance to detect Au nanomaterials coupled to complementary DNA sequences to detect SARS-CoV-2 hybridized cDNA [248]; Dendrimer-like DNA nanostructures (DL-DNA) for DNA detection (Li Y) system-on-a-chip (SoC) technology, whose detection mechanism is based on RNA aptamer conjugated to QDs [249] and materials such as silica that have been used to identify bioimaging and biodetection, and pathogen detection [250] (Figure 3).

Figure 3.

Figure 3

Nanomaterial tools for fighting future pandemics: (1) diagnostics, (2) vaccine delivery, (3) cell tramps, and (4) microfluidics are the future to fight against the pandemic. Created with BioRender.com (accessed date 15 January 2021).

In addition, gold nanoparticles coated with sulfated ligands, silver nanoparticles, hybrid silver, and copper nanoparticles can bind to the HIV envelope glycoprotein gp120 and inhibit HIV-1 infection in vitro in cell models [251]. Single wall carbon nanotubes (SWCNT) were also proposed as antiviral carriers, testing the increase in antiviral activity of hybrid materials compared to that of antiviral drugs [252]. Even antiviral and antibacterial textiles were based on nanomaterials; however, the commercialization of new functional nanomaterials is still not clear yet, and seems to be limited by nanotoxicology concerns or various other practical aspects (off-target and/or any other unpredictable effects, costs, performance production, durability, environmental impact, etc.). The emergence of new diseases has generated increased development and innovative tools in various areas of biotechnology and nanomedicine. The development of Au nanoparticles has mainly been useful for detecting different coronaviruses and respiratory infections, as presented in Table 4.

Table 4.

Nanoparticle-based tests to detect several coronaviruses.

Virus NPs Readout/Output LoD T (min) Ref
SARS-CoV AuNPs Immunosensing 10 fmol 30 [253]
PEDV AU NPs Nano nest PCR 2.21 × 10−7 ngμl−1 45 [254]
Mers CoV AgNPs Colorimetric 1.53 nM 20 [255]
Mers CoV AgNPs Electrochemiluminescence 1.0 pg/mL−1 20 [256]
Mers CoV S.aureus nanobioparticles Agglutination test 0.93055556 20 [257]
IBV MoS2 nanosheets Immunosensing 4.6 × 102 EID50 per mL 20 [258]
IBV Qd-MP NPs and Zr NPs Photoluminesce 79.15 EID/50 mL 20 [259]
IBV CAu NPs Chiroimmunosensing 47.91 EID/50 mL 20 [260]
IBV Colloidal Au NPs ICS 10 4.4 EID/50 mL 45 [257]
HCoV AgNPs Electrochemiluminescence 0.4 pgml−1 20 [257,261]

8. High-Throughput and Automated Screening for COVID-19

Given that test, track, and trace are crucial to tackle the COVID-19 pandemic, these processes are the core recommendation by CDC, European Centre for Disease Prevention and Control (ECDC), WHO, and other health agencies [262]. The rapid spread of COVID-19 has led to a huge demand for diagnostic screening [263], and to meet this goal while increasing the testing capacity of cities and countries, automated workflows have become key to enable high-throughput screening at an increased speed while improving diagnostic precision and reducing labor costs [264].

Several alliances between academic laboratories and companies such as Analytik Jena, Beckman Coulter, Hamilton, Tecan, Roche, Opentrons, among others, have focused their time and efforts in an impressive way to provide automated testing solutions for COVID-19 diagnosis [265]. These automated solutions were designed to cater to the various alternatives of molecular diagnosis protocols, such as the SHERLOCK [130] and LAMP [266] assays, although the RT-qPCR assay is still considered the “gold standard” for automated for SARS-CoV-2 testing [267].

8.1. Automation Platforms for SARS-CoV-2 Testing

An impressive example of automation support for COVID-19 testing and screening was the Huo-Yan Air Laboratory or “Fire Eye Lab’’ built in Wuhan, where the COVID-19 outbreak started in China. The Huo-Yan Lab uses automated nucleic acid extraction as a part of the RT-qPCR testing workflow, reaching 14.000 testing capacity per day on February 9 in a 2000 m2 site. This testing capacity was further increased to 20.000 per day by March 1 [268]. To achieve this, MGISP-960 platforms produced by the Chinese biotech company, MGI, were used to automate RNA extraction using the MGIEasy Magnetic Beads Virus DNA/RNA Extraction Kit to achieve high-throughput and standardized clinical testing, as shown in Figure 4 [81,269].

Figure 4.

Figure 4

General RT-qPCR test workflow and comparison of manual and automated nucleic acid extraction in Huo-Yan Lab. The use of MGISP-960 platform to automate nucleic acid extraction increased the testing capacity, simplified the steps for extraction, and reduced the time per test. Adapted from [146,269].

Manual RNA extractions using the QIAamp Viral RNA Mini Kit were compared to the platform automation, showing that the manual RNA extraction of 24 samples took on average 1 h and 50 min, whereas the automated extraction could finish processing 192 samples in 1 h and 8 min with a single run. The Ct values between the manual automated extraction resulted in no significant statistical difference, which was necessary to validate the automated platform [269].

In parallel to Huo-Yan Laboratory and MGI automated alternatives, several other biotech companies focusing on automation platforms rapidly adapted their technologies for COVID-19 screening. For instance, Hamilton offered an automated RNA extraction solution with its MagEx STARlet platform, enabling it to process 96 samples simultaneously [270]. Moreover, this platform is compatible with five commercially available magnetic bead-based RNA extraction kits, including MagMAX™ Viral/Pathogen NA Extraction Kit (Thermo Fisher, Waltham, MA, USA), Quick-DNA/RNA™ Viral MagBead Kit (Zymo Research, Irvine, CA, USA), and the Maxwell® HT Viral TNA Kit (Promega, Madison, WI, USA). In addition to RNA extraction, PCR Prep STARlet for automated qPCR mix prep solution is also offered by the company as a compatible automation platform with FDA-approved kits such as TaqPath COVID-19 Combo Kit (Thermo Fisher), 2019-nCoV Kit, (Integrated DNA Technologies, Coralville, IA, USA), and the 2019-nCoV CDC Probe and Primer Kit (BioSearch Technologies, Hoddesdon, UK) [270].

The popular Fluent and Freedom EVO platforms of Tecan were adapted for use in automated RNA extraction and PCR preparation processes [271]. A Freedom EVO platform allowing for a fully automated qPCR reaction set-up works in compatibility with the real-time thermocyclers of Thermo Fisher, Bio-Rad, and Roche [271].

These technologies are currently being improved at a very rapid scale through innovative partnerships such as the collaboration of Zymo Research, focusing on molecular biology tools with Tecan and Opentrons to optimize the magnetic bead-based automated RNA extraction process [272,273,274,275].

8.2. Biofoundries Role against COVID-19

Biofoundries, which are non-profit organizations with state-of-art automation platforms and high-throughput equipment focusing on accelerating and prototyping biological designs for bioengineering applications [276]. Such facilities were uniquely positioned to contribute to the development of novel automated platforms for SARS-CoV-2.

The London Biofoundry configured its existing automated liquid handling infrastructure to COVID-19 screening platforms using three different workflows, encompassing RT-qPCR, CRISPR-Cas, and LAMP detection methods [276]. These three methods were automated and validated using various liquid handling platforms [276]. Figure 5 illustrates the automated workflows for the three different SARS-CoV-2 assays [276].

Figure 5.

Figure 5

Schematic illustrations of three different diagnostic workflows from patient samples: A thermal cycler device is necessary in the case of RT-qPCR tests, while plate reader or other alternatives are available for measuring the readout in the case of CRISPR-Cas and LAMP tests. Adapted from [277].

To test the platform performance, the researchers also tested two different RNA extraction kits from Analytik Jena and Promega and three different qPCR master mixes from ThermoFisher (TaqPath and Fast Virus) and NEB (Luna). Ct values of 103 copies/mL of virus-like particles cwere observed around 30 cycles, while Ct values of 104 copies/mL of VLP were approximately 27 cycles in both RNA extraction kits. For the LAMP workflow, however, it was observed that at least 42.5 copies of VLP were necessary for appropriate detection. The RT-qPCR workflow was validated with 173 patient samples obtained from the North West London Pathology using two different RNA extraction kits, obtaining an adequate R2 correlation of 0.8310 and R2 = 0.9357 for the innuPREP Virus DNA/RNA Kit (Analytik Jena) and Maxwell HT Viral RNA extraction kit (Promega), respectively. This automated platform employing the RT-qPCR workflow became operational in two London hospitals from May 2020 with a 2000 testing capacity per day [276].

There are also commercially available alternatives from several suppliers such as Hologic and Roche. For example, the Cobas® 8800 (Roche, Basel, Switzerland) platform can process 1056 samples in eight hours or 384 samples in 4 h [59]. On the other hand, Panther Fusion from Hologic can run 336 samples simultaneously when 28 cartridges are used at the same time [278]. However, these platforms are also relatively expensive, and they mainly use proprietary reagents and cartridges which are tailored to their own platforms.

8.3. Free and Open Source Scientific and Medical Hardware (FOSH)

Even though the combination of the FeliX and Echo 525 were proven to be effective automation platforms for SARS-CoV-2, their accessibility by many laboratories is limited due to the higher costs of the equipment. To counteract this, the free open-source hardware and software community has played a prominent role for COVID-19 diagnostics, contributing to ease the burden on the global health systems. This has been achieved by the development and sharing of tools freely available for the community for further study, customization, and commercialization at a lower cost [279]. The best example of this for automation is Opentrons’ platform OT-2, which is an affordable and open-source lab automation system allowing to scale-up and automate hundreds of assays and workflows.

The OT-2 has gained popularity in the scientific community day by day, thanks to its flexibility and affordability [280]. While the Felix and Echo 525 cost are in the order of hundreds of thousands of pounds, the customizable OT-2 costs between 5 and 15 thousand, depending on the selected configuration.

Researchers at the Biomedical Diagnostic Center (CBD) of the Hospital Clinic of Barcelona have developed an automated COVID-19 screening circuit named ROBOCOV, using four OT-2 robots from Opentrons, a KingFisher Flex extraction instrument from ThermoFisher, and an ABI 7500 qPCR device [280]. In this circuit, the initial setup, sample preparation, plate filling for RNA extraction, and qPCR mix preparation steps were carried out by four OT-2 robots while RNA extraction was performed by the KingFisher Flex and real-time qPCR process was done by the ABI 7500 qPCR device, as shown in Figure 6.

Figure 6.

Figure 6

ROBOCOV circuit: Run preparation is done by using open-source Python codes. Initial sample setup, sample preparation, and plate filling are performed by OT-2 Station A, B1, and B2, respectively. RNA extraction is processed by KingFisher Flex. Then, qPCR mic preparation is done by OT-2 Station C, and, finally, the qPCR process is run by ABI 7500. The analysis results are exported as a user-friendly R file. Adapted from [280].

This study run preparation was done by using open-source Python codes. Initial sample setup, sample preparation, and plate filling are performed by OT-2 Station A, B1, and B2, respectively. RNA extraction is processed by KingFisher Flex [280]. Then, qPCR mic preparation is done by OT-2 Station C, and, finally, the qPCR process is run by ABI 7500 [280]. The analysis results are exported as a user-friendly R file. Due to this parallel and automated molecular testing circuit consisting of six platforms, the researchers tram obtained a high screening capacity completing the testing process of 96 samples in 4 h while starting each new run every 70 min. This set-up was further assessed and validated by the European Molecular Genetics Quality Network, which provides external quality assessment, and the findings revealed a comparable performance and consistent Ct levels compared with other validated automated platforms from Roche and Hamilton. A similar ROBOCOV is currently being used for high-throughput COVID-19 screening as an FDA-approved method at the Biomedical Diagnostic Center of the Hospital Clinic of Barcelona, with a 2400 test capacity per day [281]. The CLIAHUB facility, which intends to aid on COVID-19 testing in the San Francisco Bay Area, have also demonstrated that research institutions can contribute to COVID-19 testing and that the workflow provided by their report can process a vast number of tests by using eight automated liquid handles into the RNA extraction step [282].

The studies mentioned in this section highlight the key role that the automation of molecular diagnosis has had in the scaling and speeding up of COVID-19 testing and screening. It also showcases the need to develop novel and affordable automated diagnostic workflows to be deployed in lower income settings where testing is currently done predominantly manually. The automated platforms developed during the COVID-19 outbreak are very relevant, as they can easily be adapted for diagnostics for many other types of biological threats and infectious diseases, allowing the global community to be better prepared for future pandemics.

9. Conclusions

The COVID-19 pandemic has been and continues to be a threat to economies and health systems worldwide. Currently, no standard treatment for COVID-19 is globally available. Despite the enormous efforts that have allowed the unprecedented development of multiple highly effective vaccines against COVID-19 in record time, the broad access and distribution of vaccines in many countries is still a significant challenge. Therefore, the implementation of efficient and versatile COVID-19 testing policies remains a key strategy to tackle the effects of COVID-19, to its epidemiological features, and to prevent further damage from this virus and from additional SARS-CoV-2 variants.

Given the structure of the SARS-CoV-2 virus, and due to the previous development of technologies for the detection of nucleic acids and proteins, the major diagnostic strategies for COVID-19 are based on the detection of these molecules. The main objective of this review is to present a clear and useful diagnostic landscape from the most common strategies and technologies to novel diagnostic alternatives, describing their principles, as well as highlighting their advantages and disadvantages. It is also intended to review some of the most relevant automation platforms useful for boosting diagnostics capabilities, as well as diagnostic alternatives that are based on the application of nanotechnology principles.

Author Contributions

C.B.-O. and L.P.G.-B. were responsibles for the conceptualization of this work. C.B.-O., E.M.-U. and C.G.O.-A. were in charge of drafting the document in all of the stages. E.M.-U. and C.G.O.-A. compiled the information and organized the Table 1, Table 2, Table 3. M.T.-A. completed the Section 6 (Serology-based tests) and Section 7 (Nanotechnology for COVID-19 Diagnostics). N.C.K. and L.P.G.-B. completed the Section 5 (CRISPR-Based Tests) and the antigen testing section. K.M. and L.R.-S. completed the Section 8 (Automation Platforms for SARS-CoV-2 Testing). C.B.-O. and N.J.-B. completed the Section 3 and Section 4. A.L.-C. and E.R.-M. completed the Section 1 (Introduction) and Section 2 (Diagnosis). A.L.-C. and C.B.-O. completed all the figures for this work. All authors have read and agreed to the published version of the manuscript.

Funding

This work was funded by The Royal Academy of Engineering, Pandemic Preparedness Phase 1 grant EXPP2021\1\207, The University of Edinburgh SFC-GCRF Covid-19 Fund, COV_27 and the PAPIIT project, UNAM number: IN216720.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Conflicts of Interest

The authors declare having no competing or financial interests or personal relationships related to this manuscript.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References


Articles from Diagnostics are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES