Abstract
Objective:
To evaluate the diagnostic performance and image quality of the low-tube voltage and low-contrast medium dose protocol for hepatic dynamic CT.
Methods:
This retrospective study was conducted between January and May 2018. All patients underwent hepatic dynamic CT using one of the two protocols: tube voltage, 80 kVp and contrast dose, 370 mgI/kg with hybrid iterative reconstruction or tube voltage, 120 kVp and contrast dose, 600 mgI/kg with filtered back projection. Two radiologists independently scored lesion conspicuity and image quality. Another radiologist measured the CT numbers of abdominal organs, muscles, and hepatocellular carcinoma (HCC) in each phase. Lesion detectability, HCC diagnostic ability, and image quality of the arterial phase were compared between the two protocols using the non-inferiority test. CT numbers and HCC-to-liver contrast were compared between the protocols using the Mann–Whitney U test.
Results:
424 patients (70.5 ± 10.1 years) were evaluated. The 80-kVp protocol showed non-inferiority in lesion detectability and diagnostic ability for HCC (sensitivity, 85.7–89.3%; specificity, 96.3–98.6%) compared with the 120-kVp protocol (sensitivity, 91.0–93.3%; specificity, 93.6–97.3%) (p < 0.001–0.038). The ratio of fair image quality in the 80-kVp protocol also showed non-inferiority compared with that in the 120-kVp protocol in assessments by both readers (p < 0.001). HCC-to-liver contrast showed no significant differences for all phases (p = 0.309–0.705) between the two protocols.
Conclusion:
The 80-kVp protocol with hybrid iterative reconstruction for hepatic dynamic CT can decrease iodine doses while maintaining diagnostic performance and image quality compared with the 120-kVp protocol.
Advances in knowledge:
The 80- and 120-kVp protocols showed equivalent hepatic lesion detectability, diagnostic ability for HCC, image quality, and HCC-to-liver contrast.
The 80-kVp protocol showed a 38.3% reduction in iodine dose compared with the 120-kVp protocol.
Introduction
Hepatic dynamic CT with four phases—pre-contrast (pre-CT), arterial phase (AP), portal venous phase (PVP), and delayed phase (DP)—is widely used for routine detection and characterization of hepatic lesions. Hepatic lesions in high-risk patients that show typical CT findings of hepatocellular carcinoma (HCC) can be diagnosed as HCC without biopsy.1 However, this technique involves radiation exposure and the use of a nephrotoxic iodine-based contrast medium (CM), which is a common cause of iatrogenic acute kidney injury.2 Patients with chronic liver disease often share some risk factors for kidney injury, such as advanced age and diabetes mellitus.3 Moreover, they require repeated hepatic dynamic CT scans for assessment of HCC development. Therefore, it is desirable to reduce the amount of CM used without causing a loss of image quality. Low-tube voltage CT is a well-established and effective method of reducing the contrast dose. Enhancement of iodine-based CM substantially increases at lower tube voltages, as the effective energy of the X-ray beam approaches the absorption k-edge of iodine (33.2 keV).4 Although lower tube voltages can increase the image noise because of the reduced number of X-ray photons, a compensatory increase in the tube current and the use of iterative reconstruction (IR) instead of filtered back projection (FBP) can solve this problem.5,6 Currently, hybrid IR is the most popular IR method that implements a two-step noise reduction process, wherein streak artifacts are first removed from the projection data by combining the scanner model and the statistical noise model, and image reconstruction is then performed by FBP with repeated noise removal.7 This technique can improve image quality by decreasing image noise and artifacts in low-tube voltage hepatic dynamic CT.8,9 Nevertheless, although the utility of low tube voltage has been reported in multiple studies to date, the non-inferiority in the diagnostic performance of low-tube voltage dynamic liver CT has not been well clarified in comparison with standard-tube voltage CT.
We hypothesized that a protocol employing a low-tube voltage and low CM dose might be useful in routine practice because it could reduce the CM dose without causing a loss of image quality. The purpose of this study was to compare the diagnostic performance and image quality of a low-tube voltage (80 kVp) and low CM dose (370 mgI/kg) protocol with those of a conventional (120 kVp and 600 mgI/kg) protocol for hepatic dynamic CT.
Methods and materials
Patients
This single-center study was approved by the relevant institutional review board. The requirement for written informed consent was waived because this study retrospectively used the data from a clinical cohort. Between January and May 2018, 522 consecutive adult patients who underwent hepatic dynamic CT were identified as eligible participants. From this population, the further selection was performed according to the following inclusion criteria: (i) body weight ≤75 kg, (ii) tube voltage and contrast dose, 120 kVp and 600 mgI/kg, respectively (120-kVp protocol) or 80 kVp and 370 mgI/kg, respectively (80-kVp protocol), and (iii) presence of available CT data.
CT and contrast medium injection protocols
The 80-kVp protocol CT was performed using a 320-detector-row unit, and the 120-kVp protocol CT was performed using a 64-detector-row unit because of a technical limitation. Patients with reduced renal function tended to be assigned to the 80-kVp protocol to reduce the dose of CM, while other patients were assigned to whichever protocol was available at the time of examination. The detailed scanning parameters of each protocol are shown in Table 1. Automatic exposure control (Real EC, Canon Medical Systems, Otawara, Japan) was used to optimize the dose for each patient. The maximum tube current was 580 mA with the 320-detector-row unit and 450 mA with the 64-detector-row unit. Therefore, we assumed that tube current would not be sufficient if the maximum tube currents of the CT units were used for the scan. A 370 mgI/kg or 600 mgI/kg dose of CM was intravenously administered using a power injector during a fixed injection time of 30 s.
Table 1.
80-kVp protocol | 120-kVp protocol | |
---|---|---|
CT unit | 320-detector-row unit (Aquilion ONE; Canon Medical Systems, Otawara, Japan) |
64-detector-row unit (Aquilion 64, Canon Medical Systems, Otawara, Japan) |
Rotation time [s] | 0.75 | 0.5 |
Beam collimation [mm] | 80 × 0.5 | 64 × 0.5 |
Reconstructed section thickness [mm] | 5 | 5 |
Reconstructed section interval [mm] | 5 | 5 |
Helical pitch (beam pitch) | 0.813 | 0.828 |
Table movement [mm/s] | 43.6 | 52.9 |
Scanning field of view [cm] | 40 | 40 |
Tube current [mA] | 150–580 | 150–450 |
Effective mAs/slice | 139–536 | 91–272 |
Total amount of contrast medium [mgI/kg] | 370 | 600 |
Injection duration [s] | 30 | 30 |
Scan delay after injection | Pre-contrast, 40 (AP), 70 (PVP), 180 (DP) s |
Pre-contrast, 40 (AP), 70 (PVP), 180 (DP) s |
Contrast medium | Iopamiron 370 (Iopamidol), Bayer Healthcare, Osaka, Japan; Iomeron 350 (Iomeprol), Eisai Co., Tokyo, Japan; Optiray 320 (Ioversol), Guerbet Japan, Tokyo, Japan or Omnipaque 300 (Iohexol), GE Healthcare Japan, Tokyo, Japan |
AP, arterial phase; PVP, portal venous phase; DP, delayed phase.
Image reconstruction
Image reconstruction was performed in a 32–45 cm display field of view, depending on the patient’s physique. The 120-kVp images were reconstructed using a standard FBP algorithm with a standard soft tissue kernel (FC03). The 80-kVp images were reconstructed using the hybrid IR algorithm (adaptive iterative dose reduction [AIDR] 3D) and kernel FC03. The hybrid IR algorithm is a raw-data-based, statistical IR technique. During the reconstruction, a scanner model and a statistical noise model are considered to improve image quality and dose reduction.10 AIDR 3D is available at four strength levels (weak, mild, standard, and strong). The mild level was selected based on a previous study which reported half-dose CT scans compared to FBP can be achievable in small-sized patients without hampering diagnostic performance by applying the mild level of the hybrid IR.11
Qualitative image analysis
One radiologist (S.I. with 11 years of experience in abdominal radiology) reviewed all images and chose slices of AP images with or without focal liver lesions from patients with focal liver lesions for evaluation. In patients with multiple lesions, the largest three lesions were selected for evaluation. For the remaining cases, slices of AP images for which the slice positions of liver CT were randomly chosen were selected for evaluation.
The AP images were independently and randomly assessed by two radiologists (S.I. and T.S. with 11 and 5 years of experience in abdominal radiology, respectively). They were unaware whether the patients had liver lesions or not. They evaluated the scan timing for AP images using a 4-point scale and the image quality of AP scans using a 3-point scale (Figure 1). The criteria of the scan timing were as follows: 1, too early = without enhancement of the portal vein; 2, early = weak enhancement of the portal vein; 3, appropriate = adequate enhancement of the hepatic artery and portal vein without enhancement of the hepatic vein; 4, late = enhancement of the hepatic vein. The criteria of the image quality were as follows: 1, fair = image noise or beam hardening artifacts are present and slightly interfere with the depiction of intrahepatic structures; 2, good = image noise or beam hardening artifacts are present but do not interfere with the depiction of intrahepatic structures; 3, excellent = minimal or no image noise or beam hardening artifacts.
The radiologists were asked to detect any liver lesions for the assessment of lesion detectability. They were then required to identify HCCs among the lesions for the assessment of HCC diagnostic ability. These assessments were performed using a 5-point scale indicating the reviewers’ confidence in diagnosing any hepatic lesion (including HCC and other lesions) as well as HCC in the slices by reviewing all the phases (1, hepatic lesion/HCC definitely absent; 2, hepatic lesion/HCC probably absent; 3, equivocal; 4, hepatic lesion/HCC probably present; and 5, hepatic lesion/HCC definitely present). Each radiologist used the LI-RADS v. 2018 criteria12 for the assessment and assigned the scores based on their subjective judgments. Hypervascular pseudolesions and cysts were considered negative cases in the assessment of lesion detectability because they are not clinically significant. The radiologists could freely scroll all the AP images to evaluate pre-chosen slices and image quality.
Quantitative image analysis
Another radiologist (M.L.K. with 3 years of experience in abdominal radiology) measured the CT numbers of abdominal organs (liver, portal vein, abdominal aorta, pancreas, spleen, renal cortex, and renal medulla) and HCC in each phase for all patients by placing circular regions of interest (ROIs). Image noise was defined as the standard deviation of the ROI of the liver. To evaluate the contrast-to-noise ratio (CNR), we measured the attenuation of the erector spinae muscle. The contrast enhancement (CE), CNR, and HCC-to-liver contrast were calculated using the following formulae:
CE of the organs or HCC = CT number on each phase – CT number on pre-contrast scan
CNR of the organs or HCC = (CT number of organs or HCC – CT number of the erector spinae muscle)/SD of the liver
HCC-to-liver contrast = CT number of HCC – CT number of the liver
Statistical analysis
CM dose and injection rate and AP tube current were compared between the two protocols using the Mann–Whitney U test. Lesion detectability and diagnostic ability for HCC were evaluated in each protocol by calculating sensitivity, specificity, accuracy, positive and negative predictive values, and the area under the receiver operating characteristic curve (AUC). A score of ≥3 was considered to indicate a positive result. Non-inferiority of AP image quality, lesion detectability, and diagnostic ability for HCC of the 80-kVp protocol were compared with those of the 120-kVp protocol using a one-sided χ2 test with a 7.5% non-inferiority margin. The non-inferiority of the AUC of the 80-kVp protocol was compared with that of the 120-kVp protocol using a one-sided Student’s t-test with a 7.5% non-inferiority margin.13 Prior non-inferiority tests on diagnostic imaging have reported margins of 6–10%14–16 ; thus, in this study, the margin was set at 7.5%, which was within this range. Weighted κ values were calculated to assess interobserver agreement. Agreement was considered excellent for κ values (κ)>0.8, good for 0.6 < κ ≤ 0.8, moderate for 0.4 < κ ≤ 0.6, fair for 0.2 < κ ≤ 0.4, and poor for κ ≤ 0.2. All statistical analyses were performed using JMP software (v. 14.1.0; SAS Institute Inc., Cary, NC) and R (v. 3.6.2; The R Foundation for Statistical Computing, Vienna, Austria). p-values <0.05 were considered statistically significant.
Results
Patient demographics for each protocol
The final study cohort consisted of 401 patients (mean age, 70.5 ± 10.2 [range, 33–91] years) and included 270 men (71.1 ± 9.6 [40–91] years) and 131 women (69.4 ± 11.3 [33–91] years). Patients who underwent both protocols during the study period were included in both protocols; therefore, 424 examinations were included. Of these, 180 examinations were scanned using the 80-kVp protocol, and 244 examinations underwent the 120-kVp protocol. Based on the exclusion criteria, 121 patients were excluded (Figure 2). The underlying liver diseases in patients in the final study cohort are shown in Figure 2. A significant difference between the protocols was observed in age and estimated glomerular filtration rate (eGFR) (p = 0.010 and 0.027, respectively). Other factors, including sex, body weight, body mass index, presence or absence of chronic liver disease, history of liver surgery or liver lesion, and lesion size, did not show significant differences between the protocols (p = 0.251–0.604, Table 2).
Table 2.
Variables | 80-kVp protocol | 120-kVp protocol | p-value |
---|---|---|---|
Number of examinations | 180 | 244 | - |
Age [years] | 71.7 ± 10.4 | 69.6 ± 9.7 | 0.010* |
Sex [male: female] | 126:54 | 163:81 | 0.527 |
Body weight [kg] | 57.0 ± 9.0 | 58.1 ± 9.8 | 0.251 |
Body mass index [kg/m2] | 22.5 ± 3.1 | 22.8 ± 3.3 | 0.317 |
Chronic liver disease [present: absent] | 143:37 | 186:58 | 0.480 |
History of liver surgery [present: absent] | 42:138 | 51:193 | 0.555 |
eGFR [mL/min/1.73 m2] | 68.6 ± 17.0 | 72.0 ± 16.8 | 0.027* |
Liver lesionb [present: absent] | 64:116 | 81:163 | 0.604 |
Lesion size [mm] | 15.4 ± 9.5 | 14.4 ± 7.9 | 0.477 |
eGFR, estimated glomerular filtration rate.
Continuous variables were analyzed by Wilcoxon test and are expressed as mean ± standard deviation. Categorical variables were analyzed by the χ2 test and are expressed as ratios.
*p < 0.05
Except for cyst and hypervascular pseudolesions
Evaluated hepatic lesions
278 slices (235 slices with focal liver lesions and 43 slices without focal liver lesions) were chosen from 146 examinations with focal liver lesions. From the rest of the cases, 345 slices were selected. In total, 623 slices of AP images (265 slices from the 80-kVp protocol and 358 slices from the 120-kVp protocol) were selected from 424 examinations for evaluation (Figure 3). Final diagnoses of 623 slices were as follows; HCC, n = 116 (mean size, 14.4 ± 7.5 [range, 5–40] mm); other liver tumors, n = 119 (17.0 ± 9.9 [5–50] mm); hypervascular pseudolesions, n = 78; cysts, n = 97; and no lesion, n = 213 (Figure 3). 29 lesions were pathologically confirmed, while the remaining lesions were diagnosed using imaging and clinical observation. The detailed criteria for the imaging and clinical observation-based diagnoses are described in Supplementary Material 1.
Dose and injection rate of CM and tube current for the AP in each protocol
The CM volume, iodine dose, and injection rate in the 80-kVp protocol were significantly lower than those in the 120-kVp protocol (all p < 0.001). Tube current in the 80-kVp protocol was significantly higher than that in the 120-kVp protocol (p < 0.001). The proportion of cases showing insufficient tube current was not significantly different between the two protocols (p = 0.410) (Table 3).
Table 3.
Variables | 80-kVp protocol | 120-kVp protocol | P value |
---|---|---|---|
Injection condition of contrast medium | |||
Volume of contrast medium [mL] | 63.5 ± 10.3 | 101.5 ± 17.1 | <0.001* |
Iodine dose [g] | 20.9 ± 3.3 | 34.9 ± 5.9 | <0.001* |
Injection rate [mL/s] | 2.1 ± 0.3 | 3.4 ± 0.5 | <0.001* |
Tube current | |||
Tube current [mA] | 541.9 ± 45.4 | 390.6 ± 63.1 | <0.001* |
Insufficient tube current | 36.7% (66/180) | 32.8% (80/244) | 0.410 |
Image quality of arterial phase | |||
Appropriate scan timing (Reader 1) | 83.9% (151/180) | 82.0% (200/244) | <0.001* |
Appropriate scan timing (Reader 2) | 80.6% (145/180) | 81.6% (199/244) | <0.001* |
Fair image quality (Reader 1) | 1.7% (3/180) | 0.4% (1/244) | <0.001* |
Fair image quality (Reader 2) | 2.2% (4/180) | 0.8% (2/244) | <0.001* |
Contrast medium injection parameters and tube current were analyzed by Wilcoxon test and are expressed as mean ± standard deviation. Non-inferiority of image quality in the arterial phase in the 80-kVp protocol was compared to those in the 120-kVp protocol by one-sided χ2 test with 7.5% non-inferiority margin.
Categorical variables are expressed as ratios or percentage with numerators and denominators.
*p < 0.05
Qualitative image analysis
Both readers noted non-inferiority in the proportions of appropriate scan timing of AP and the proportion of fair image quality in the 80-kVp protocol in comparison with the 120-kVp protocol (p < 0.001) (Table 3).
Lesion detectability and diagnostic ability for HCC of both protocols and readers were high (sensitivity, 85.7–93.3%; specificity, 93.6–98.6%; AUC, 0.93–0.96) (Table 4). All parameters of lesion detectability and diagnostic ability for HCC in the 80-kVp protocol showed non-inferiority in comparison with those in the 120-kVp protocol (p < 0.001–0.039) (Table 4). Interobserver agreements were good for the scan timing and image quality of AP (80-kVp and 120-kVp, κ = 0.73 [95% CI, 0.65–0.81] and 0.72 [0.65–0.79], respectively) and excellent for the scores of the confidence of the presence of hepatic lesions and HCCs (κ = 0.86 [0.82–0.90] and 0.86 [0.81–0.90], respectively).
Table 4.
Reader 1 | Reader 2 | |||||
---|---|---|---|---|---|---|
80-kVp protocol | 120-kVp protocol | P value | 80-kVp protocol | 120-kVp protocol | P value | |
Lesion detectability | ||||||
Sensitivity | 89.1% (90/101) | 91.0% (122/134) | <0.001a | 86.1% (87/101) | 92.5% (124/134) | <0.001a |
Specificity | 96.3% (158/164) | 97.3% (218/224) | <0.001a | 97.0% (159/164) | 94.6% (212/224) | <0.001a |
Accuracy | 93.6% (248/265) | 95.0% (340/358) | <0.001a | 92.8% (246/265) | 93.9% (336/358) | <0.001a |
Positive predictive value | 93.8% (90/96) | 95.3% (122/128) | <0.001a | 94.6% (87/92) | 91.2% (124/136) | <0.001a |
Negative predictive value | 93.5% (158/169) | 94.8% (218/230) | <0.001a | 91.9% (159/173) | 95.5% (212/222) | <0.001a |
Area under the curve | 0.93 (0.89–0.96) | 0.95 (0.92–0.97) | 0.018a | 0.93 (0.89–0.96) | 0.95 (0.92–0.97) | 0.012a |
Diagnostic ability for hepatocellular carcinoma | ||||||
Sensitivity | 89.3% (50/56) | 93.3% (56/60) | <0.001a | 85.7% (48/56) | 91.7% (55/60) | 0.038a |
Specificity | 98.6% (206/209) | 97.0% (289/298) | <0.001a | 98.6% (206/209) | 93.6% (279/298) | <0.001a |
Accuracy | 96.6% (256/265) | 96.4% (345/358) | <0.001a | 95.8% (254/265) | 93.3% (334/358) | <0.001a |
Positive predictive value | 94.3% (50/53) | 86.2% (56/65) | <0.001a | 94.1% (48/51) | 74.3% (55/74) | <0.001a |
Negative predictive value | 97.2% (206/212) | 98.6% (289/293) | <0.001a | 96.3% (206/214) | 98.2% (279/284) | <0.001a |
Area under the curve | 0.94 (0.88–0.97) | 0.96 (0.91–0.98) | 0.039a | 0.94 (0.88–0.97) | 0.96 (0.91–0.98) | 0.036a |
Non-inferiority of sensitivity, specificity, accuracy, and positive/negative predictive values of the 80-kVp protocol were compared to those in the 120-kVp protocol by one-sided χ2 test with a 7.5% non-inferiority margin. Area under the curve of the 80-kVp protocol was compared to those of 120-kVp protocol by one-sided Student’s t-test with a 7.5% non-inferiority margin.
Categorical variables are expressed as percentage with numerators and denominators.
P < 0.05
Case examples are shown in Figures 4 and 5. Figure 4 shows the findings for a patient with HCC who underwent both 80-kVp and 120-kVp protocols with a 10 day interval, while Figure 5 shows the findings for a patient with intrahepatic cholangiocarcinoma who underwent both 80-kVp and 120-kVp protocols with a 2 month interval. Both protocols showed similar image quality and typical imaging findings of HCC and intrahepatic cholangiocarcinoma.
Quantitative image analysis
The image noise in each phase was significantly lower for the 80-kVp protocol than for the 120-kVp protocol (80-kVp protocol vs 120-kVp protocol on pre-contrast scan, 9.1 ± 1.7 vs. 9.7 ± 1.5 [p < 0.001]; AP, 9.7 ± 1.8 vs. 10.3 ± 1.7 [p = 0.001]; PVP, 10.0 ± 2.0 vs, 10.5 ± 1.8 [p = 0.002]; DP, 9.8 ± 1.8 vs, 10.2 ± 1.7 [p = 0.013]).
The mean CT numbers of the liver, portal vein, abdominal aorta, pancreas, spleen, renal cortex, renal medulla, and HCC on pre-CT were significantly higher for the 80-kVp protocol than for the 120-kVp protocol (p < 0.001–0.034) (Supplementary Material 1 and Figure 6). The CE of the abdominal aorta on AP, PVP, and DP; portal vein on PVP and DP; spleen on PVP and DP; renal cortex on AP, PVP, and DP; and renal medulla on DP were significantly higher for the 80-kVp protocol than for the 120-kVp protocol (p < 0.001–0.002) (Supplementary Material 1). The CNR of the liver on PVP; portal vein on pre-CT, PVP, and DP; abdominal aorta for all phases; pancreas on AP and PVP; spleen on pre-CT, PVP, and DP; renal cortex on AP, PVP, and DP; and renal medulla on PVP and DP were significantly higher for the 80-kVp protocol than for the 120-kVp protocol (p < 0.001–0.042) (Supplementary Material 1).
The CT number of HCC in the 80-kVp protocol was significantly higher than that in the 120-kVp protocol on pre-CT (p = 0.034). In contrast, the CE of HCC, CNR of HCC, and HCC-to-liver contrast were not significantly different for all phases between the two protocols (p = 0.112–0.795) (Figure 6).
Discussion
Our results demonstrated that in comparison with the 120-kVp protocol, the 80-kVp protocol with a reduced iodine dose showed non-inferiority in image quality, lesion detectability, and diagnostic ability for HCC. CNR of HCC and HCC-to-liver contrast did not significantly differ for all phases between the two protocols. These results will facilitate the routine use of the low-tube voltage protocol in patients with liver diseases in order to reduce the iodine dose.
Low-tube voltage hepatic dynamic CT shows high diagnostic ability for HCC,17–19 with the sensitivity, specificity, accuracy, and AUC for HCC diagnosis being 82–100%, 97–100%, 77–80%, and 0.94–0.99, respectively. Our results corresponded to those findings. In general, specificity is important for HCC diagnosis, and low-tube voltage hepatic dynamic CT showed high specificity. Thus, low-tube voltage hepatic dynamic CT is useful because of the considerable increases in hepatic enhancement and HCC conspicuity with lower radiation and CM doses. The quantitative image quality in previous reports focused on the diagnostic ability for HCC. We also investigated lesion detectability in the two protocols, and low-tube voltage hepatic dynamic CT showed high lesion detectability not only for HCC but also other hepatic lesions, while its ability was not inferior to that of the standard protocol. This is a new finding shown in our study.
Enhancement of iodine-based CM is known to substantially increase at lower tube voltages, but its effects on the CT numbers of abdominal organs on pre-contrast images have not been clarified. Our study revealed that the mean CT numbers of the liver, portal vein, abdominal aorta, pancreas, spleen, renal cortex, renal medulla, and HCC on pre-contrast CT were significantly higher for the 80-kVp protocol than for the 120-kVp protocol. In contrast, Sun et al reported no significant differences in the mean CT numbers of abdominal organs between the 100-kVp and 120-kVp protocols by using dual-source CT.20 However, they used the same CT machine for both 100- and 120-kVp protocols, while we used two different CT machines for the 80- and 120-kVp protocols. This may have influenced the differences in the results. After CM administration, the CE and CNR were significantly higher for the 80-kVp protocol than for the 120-kVp protocol for most organs in each phase. These results suggest that the 80-kVp protocol can yield better contrast despite the smaller amount of CM than the 120-kVp protocol. In contrast, the CE of HCC, CNR of HCC, and HCC-to-liver contrast of the two protocols were not significantly different for all phases. Although our results corresponded to some previous reports,18,21–23 they also differed from others.8,9,17,24–26 These differences may be caused by slight variations in the amount of CM administered and the scanning parameters. Further studies are needed to determine the optimal settings of low-tube voltage CT for hepatic dynamic CT.
Low-tube voltage CT is an effective method to reduce the CM dose. Patients with chronic liver disease often share some risk factors for kidney injury, such as advanced age and diabetes mellitus3 ; therefore, protocols with low CM doses are desirable to reduce the iodine burden in kidneys.
Our study had some limitations. First, the lack of histopathological confirmation for a large number of evaluated hepatic lesions is a potential limitation of this study. Second, the range and mean values of body weights of our patients were relatively smaller than those of North American and European individuals. In patients with large bodies, low-tube voltage scans tend to show poorer image quality due to the increased image noise attributable to radiation scattering and absorption. Thus, further studies are needed to determine whether our results are applicable for cohorts with higher body weights. Third, radiation doses between the two protocols were not comparable because the two CT units displayed the volume CT dose index differently (maximum in 120-kVp protocol [CT dose index, 32.1 ± 9.3 mGy; dose–length product, 880.2 ± 312.0 mGy·cm] vs average in 80-kVp protocol [CT dose index, 15.1 ± 3.6 mGy; dose–length product, 397.3 ± 121.8 mGy·cm]). However, phantom scans (Appendix 3) can be an alternative method for showing differences in the radiation doses for the two protocols. Fourth, about one-third of cases (36.7% [66/180] in 80-kVp protocol and 32.8% [80/244] in 120-kVp protocol) showed insufficient tube current. Finally, significant differences in age and estimated glomerular filtration rate were observed between the 80- and 120-kVp protocols due to the retrospective study design. Renal insufficiency tends to occur in patients with advanced age. However, the effects of age and renal function on our study’s findings were minimal because the major factors affecting CE (iodine dose and injection methods) were well controlled in the protocols.
In summary, the 80-kVp protocol with hybrid IR for hepatic dynamic CT can decrease iodine doses while maintaining diagnostic performance and image quality in comparison with the 120-kVp protocol.
Footnotes
Conflict of interest: The authors declare that they have no conflict of interest.
Funding: This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Contributor Information
Shintaro Ichikawa, Email: shintaro@hama-med.ac.jp.
Utaroh Motosugi, Email: umotosugi@nifty.com.
Tatsuya Shimizu, Email: december_twenty_four@yahoo.co.jp.
Marie Luise Kromrey, Email: ml.kromrey@yahoo.de.
Yoshihito Aikawa, Email: yaikawa@yamanashi.ac.jp.
Daiki Tamada, Email: dtamada@yamanashi.ac.jp.
Hiroshi Onishi, Email: honishi@yamanashi.ac.jp.
REFERENCES
- 1.Marrero JA, Kulik LM, Sirlin CB, Zhu AX, Finn RS, Abecassis MM, et al. Diagnosis, staging, and management of hepatocellular carcinoma: 2018 practice guidance by the American association for the study of liver diseases. Hepatology 2018; 68: 723–50. doi: 10.1002/hep.29913 [DOI] [PubMed] [Google Scholar]
- 2.Mohammed NMA, Mahfouz A, Achkar K, Rafie IM, Hajar R. Contrast-Induced nephropathy. Heart Views 2013; 14: 106–16. doi: 10.4103/1995-705X.125926 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Desbois A-C, Cacoub P, mellitus D. Diabetes mellitus, insulin resistance and hepatitis C virus infection: a contemporary review. World J Gastroenterol 2017; 23: 1697–711. doi: 10.3748/wjg.v23.i9.1697 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Szucs-Farkas Z, Verdun FR, von Allmen G, Mini RL, Vock P. Effect of X-ray tube parameters, iodine concentration, and patient size on image quality in pulmonary computed tomography angiography: a chest-phantom-study. Invest Radiol 2008; 43: 374–81. doi: 10.1097/RLI.0b013e3181690042 [DOI] [PubMed] [Google Scholar]
- 5.Beister M, Kolditz D, Kalender WA. Iterative reconstruction methods in X-ray CT. Phys Med 2012; 28: 94–108. doi: 10.1016/j.ejmp.2012.01.003 [DOI] [PubMed] [Google Scholar]
- 6.Solomon J, Marin D, Roy Choudhury K, Patel B, Samei E. Effect of radiation dose reduction and reconstruction algorithm on image noise, contrast, resolution, and detectability of subtle hypoattenuating liver lesions at multidetector CT: filtered back projection versus a commercial model-based iterative reconstruction algorithm. Radiology 2017; 284: 777–87. doi: 10.1148/radiol.2017161736 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Stiller W. Basics of iterative reconstruction methods in computed tomography: a vendor-independent overview. Eur J Radiol 2018; 109: 147–54. doi: 10.1016/j.ejrad.2018.10.025 [DOI] [PubMed] [Google Scholar]
- 8.Nakaura T, Nakamura S, Maruyama N, Funama Y, Awai K, Harada K, et al. Low contrast agent and radiation dose protocol for hepatic dynamic CT of thin adults at 256-detector row CT: effect of low tube voltage and hybrid iterative reconstruction algorithm on image quality. Radiology 2012; 264: 445–54. doi: 10.1148/radiol.12111082 [DOI] [PubMed] [Google Scholar]
- 9.Taguchi N, Oda S, Utsunomiya D, Funama Y, Nakaura T, Imuta M, et al. Using 80 kVp on a 320-row scanner for hepatic multiphasic CT reduces the contrast dose by 50 % in patients at risk for contrast-induced nephropathy. Eur Radiol 2017; 27: 812–20. doi: 10.1007/s00330-016-4435-y [DOI] [PubMed] [Google Scholar]
- 10.Willemink MJ, de Jong PA, Leiner T, de Heer LM, Nievelstein RAJ, Budde RPJ, et al. Iterative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol 2013; 23: 1623–31. doi: 10.1007/s00330-012-2765-y [DOI] [PubMed] [Google Scholar]
- 11.Yoon JH, Lee JM, Hur BY, Baek J, Shim H, Han JK, et al. Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study. Abdom Imaging 2015; 40: 1843–52. doi: 10.1007/s00261-014-0333-4 [DOI] [PubMed] [Google Scholar]
- 12.American College of Radiology Committee on LI-RADS®. CT/MRI LI-RADS® v2018 Core.. Available from: https://www.acr.org/-/media/ACR/Files/RADS/LI-RADS/LI-RADS-2018-Core.pdf [Accessed on July 12, 2021].
- 13.Zhou XH, Obuchowski NA, McClish DK. Statistical Methods in Diagnostic Medicine. Second Edition. New York, NY: Wiley; 2002. pp. 189–92. [Google Scholar]
- 14.Lanotte SJ, Larbi A, Michoux N, Baron M-P, Hamard A, Mourad C, et al. Value of CT to detect radiographically occult injuries of the proximal femur in elderly patients after low-energy trauma: determination of non-inferiority margins of CT in comparison with MRI. Eur Radiol 2020; 30: 1113–26. doi: 10.1007/s00330-019-06387-2 [DOI] [PubMed] [Google Scholar]
- 15.Xin Y, Zhang X, Yang Y, Chen Y, Wang Y, Zhou X, et al. A multicenter, hospital-based and non-inferiority study for diagnostic efficacy of automated whole breast ultrasound for breast cancer in China. Sci Rep 2021; 11: 13902. doi: 10.1038/s41598-021-93350-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Tiegs-Heiden CA, Adkins MC, Carter RE, Geske JR, McKenzie GA, Ringler MD. Does gadolinium improve magnetic resonance arthrography of the hip beyond fluid distension alone? Clin Radiol 2020; 75: 713.e1–713.e9. doi: 10.1016/j.crad.2020.01.019 [DOI] [PubMed] [Google Scholar]
- 17.Pregler B, Beyer LP, Teufel A, Niessen C, Stroszczynski C, Brodoefel H, et al. Low tube voltage liver MDCT with sinogram-affirmed iterative reconstructions for the detection of hepatocellular carcinoma. Sci Rep 2017; 7: 9460. doi: 10.1038/s41598-017-10095-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Noda Y, Kanematsu M, Goshima S, Kondo H, Watanabe H, Kawada H, et al. Reducing iodine load in hepatic CT for patients with chronic liver disease with a combination of low-tube-voltage and adaptive statistical iterative reconstruction. Eur J Radiol 2015; 84: 11–18. doi: 10.1016/j.ejrad.2014.10.008 [DOI] [PubMed] [Google Scholar]
- 19.Lee CH, Kim KA, Lee J, Park YS, Choi JW, Park CM. Using low tube voltage (80kVp) quadruple phase liver CT for the detection of hepatocellular carcinoma: two-year experience and comparison with Gd-EOB-DTPA enhanced liver MRI. Eur J Radiol 2012; 81: e605–11. doi: 10.1016/j.ejrad.2011.12.033 [DOI] [PubMed] [Google Scholar]
- 20.Sun H, Xue H-dan, Jin Z-yu, Wang X, Chen Y, He Y-lan, et al. Non-enhanced low-tube-voltage high-pitch dual-source computed tomography with sinogram affirmed iterative reconstruction algorithm of the abdomen and pelvis. Chin Med Sci J 2014; 29: 214–20. doi: 10.1016/S1001-9294(14)60073-0 [DOI] [PubMed] [Google Scholar]
- 21.Nakaura T, Awai K, Oda S, Funama Y, Harada K, Uemura S, et al. Low-kilovoltage, high-tube-current MDCT of liver in thin adults: pilot study evaluating radiation dose, image quality, and display settings. AJR Am J Roentgenol 2011; 196: 1332–8. doi: 10.2214/AJR.10.5698 [DOI] [PubMed] [Google Scholar]
- 22.Liu S, Sheng H, Shi H, Li W, Fan J, He J, et al. Computed tomography portography of patients with cirrhosis with normal body mass index: comparison between low-tube-voltage CT with low contrast agent dose and conventional CT. Medicine 2018; 97: e13141. doi: 10.1097/MD.0000000000013141 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Iyama Y, Nakaura T, Yokoyama K, Kidoh M, Harada K, Tokuyasu S, et al. Impact of knowledge-based iterative model reconstruction in abdominal dynamic CT with low tube voltage and low contrast dose. AJR Am J Roentgenol 2016; 206: 687–93. doi: 10.2214/AJR.15.14518 [DOI] [PubMed] [Google Scholar]
- 24.Araki K, Yoshizako T, Yoshida R, Tada K, Kitagaki H. Low-voltage (80-kVp) abdominopelvic computed tomography allows 60% contrast dose reduction in patients at risk of contrast-induced nephropathy. Clin Imaging 2018; 51: 352–5. doi: 10.1016/j.clinimag.2018.05.027 [DOI] [PubMed] [Google Scholar]
- 25.Svensson A, Thor D, Fischer MA, Brismar T. Dual source abdominal computed tomography: the effect of reduced X-ray tube voltage and intravenous contrast media dosage in patients with reduced renal function. Acta Radiol 2019; 60: 293–300. doi: 10.1177/0284185118783213 [DOI] [PubMed] [Google Scholar]
- 26.Nakaura T, Nagayama Y, Kidoh M, Nakamura S, Namimoto T, Awai K, et al. Low contrast dose protocol involving a 100 kVp tube voltage for hypervascular hepatocellular carcinoma in patients with renal dysfunction. Jpn J Radiol 2015; 33: 566–76. doi: 10.1007/s11604-015-0457-7 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.