CSF gusher complicating cochlear implant surgery

Soorya Pradeep 1, Arun Alexander 1,
PMCID: PMC8634216  PMID: 34848412

Description

A cerebrospinal fluid (CSF) gusher may be defined as a profuse pulsatile egress of clear fluid which occurs on making an opening in the cochlea.1 2 A gusher may be encountered in 1 in 100 cochlear implantations,3 but in patients with inner ear malformations, its incidence increases to 40%–50%.4 5 CSF leaks are believed to occur due to an abnormal communication between the CSF in the internal acoustic meatus and the perilymph in the inner ear.1 2 A careful analysis of preoperative CT and MRI is crucial to look for anatomic abnormalities which may predispose to gushers. However, gushers may be encountered even in cases where the preoperative scans appear normal (1.5%–2.5%),3 6 7 as current imaging modalities are sometimes unable to pick up minor communications between the internal acoustic canal and the cochlea.

A 14-month-old prelingually deaf boy underwent cochlear implantation with an Advanced Bionics-slim J electrode. Preoperative CT showed normal cochlear anatomy with normal dimensions of the cochlear and vestibular aqueduct. No defects could be detected in the lamina cribrosa or the modiolus. MRI showed the presence of bilateral vestibulocochlear nerves of appropriate diameter (figure 1). Implantation was done through the facial recess via the extended round window approach.

Figure 1.

Figure 1

Preoperative scans showing normal anatomy of the cochlea and the internal acoustic meatus. (A) CT temporal bone axial section, (B) CT temporal bone coronal section, (C) T2-weighted MRI axial section and (D) T2-weighted MRI coronal section.

On opening the round window, a CSF leak was detected (video 1). Controlled hyperventilation was initiated and intravenous propofol and mannitol were administered. The electrode array was inserted into the cochleostomy to stop the leak, but the pressure of the gusher was such that the electrode kept getting extruded. Pieces of periosteum were initially inserted around the electrode, but the leak continued to persist. Thin strips of temporalis fascia were then harvested and packed into the cochleostomy and around the electrode. Periosteum was further packed around the electrode to achieve a water-tight seal. Fibrin glue was applied.

Video 1.

DOI: 10.1136/bcr-2021-245766.video01

Intraoperative neural response imaging showed normal evoked compound action potential in all 16 electrodes. A postoperative X-ray confirmed the accurate positioning of the electrode (figure 2). A lumbar drain was not used. The postoperative period was uneventful, and the child was discharged on the 10th postoperative day. The device was switched on after 3 weeks and the patient continues to receive auditory-verbal therapy to date.

Figure 2.

Figure 2

Postoperative X-ray in modified Stenver view showing normal placement of the electrode.

The successful management of a CSF gusher relies on the surgeon’s ability to create a water-tight seal around the cochleostomy. This may be achieved by the folowing.

  1. Placing the patient in a reverse trendelenburg position.1 3 5

  2. Initiating controlled hyperventilation to decrease the venous return to the heart and hence decrease cerebral perfusion, hypercapnia results in vasoconstriction of cerebral vessels.3 5

  3. Administration of hypotensive agents such as propofol and diuretics like mannitol.3

  4. Insertion of the electrode after waiting for the flow of CSF to slow down.1–3

  5. Packing periosteum around the electrode array.3 5

  6. Packing strips of fascia into the basal turn and around the electrode.1 3 5

  7. Post-operative acetazolamide.6

  8. Continuous postoperative lumbar drainage.1 6

In rare cases where the above measures fail to control the leak, the surgeon may be required to completely obliterate the middle ear and close the Eustachian tube with connective tissue,5 failing which a subtotal petrosectomy may be necessary.1 6

Learning points.

  • Cerebrospinal fluid gushers during cochlear implantation are mostly encountered in patients with inner ear malformations.

  • However, gushers may be encountered even in patients with normal preoperative scans.

  • In most cases, a gusher may be managed by creating a water-tight seal around the cochleostomy site.

Footnotes

Contributors: SP: drafting the manuscript, acquisition of data and editing the video. AA: patient care and editing the manuscript.

Funding: The authors have not declared a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

Case reports provide a valuable learning resource for the scientific community and can indicate areas of interest for future research. They should not be used in isolation to guide treatment choices or public health policy.

Competing interests: None declared.

Provenance and peer review: Not commissioned; externally peer reviewed.

Ethics statements

Patient consent for publication

Consent obtained from parent(s)/guardian(s).

References

  • 1.Papsin BC. Cochlear implantation in children with anomalous cochleovestibular anatomy. Laryngoscope 2005;115:1–26. 10.1097/00005537-200501001-00001 [DOI] [PubMed] [Google Scholar]
  • 2.Sennaroglu L. Cochlear implantation in inner ear malformations--a review article. Cochlear Implants Int 2010;11:4–41. 10.1002/cii.416 [DOI] [PubMed] [Google Scholar]
  • 3.Wootten CT, Backous DD, Haynes DS. Management of cerebrospinal fluid leakage from cochleostomy during cochlear implant surgery. Laryngoscope 2006;116:2055–9. 10.1097/01.mlg.0000240286.43289.87 [DOI] [PubMed] [Google Scholar]
  • 4.Loundon N, Rouillon I, Munier N, et al. Cochlear implantation in children with internal ear malformations 2005;26:6. [DOI] [PubMed] [Google Scholar]
  • 5.Mehanna AMA, Fathala MF, Elwany MS. A stepwise algorithm for the management of cerebrospinal fluid gusher during cochlear implantation. Egypt J Otolaryngol 2015;31:19–29. 10.4103/1012-5574.152704 [DOI] [Google Scholar]
  • 6.Vashist S, Singh S. CSF Gusher in cochlear implant Surgery-does it affect surgical outcomes? Eur Ann Otorhinolaryngol Head Neck Dis 2016;133(Suppl 1):S21–4. 10.1016/j.anorl.2016.04.010 [DOI] [PubMed] [Google Scholar]
  • 7.Eftekharian A, Amizadeh M. Cerebrospinal fluid gusher in cochlear implantation. Cochlear Implants Int 2014;15:179–84. 10.1179/1754762814Y.0000000069 [DOI] [PubMed] [Google Scholar]

Articles from BMJ Case Reports are provided here courtesy of BMJ Publishing Group

RESOURCES