Ultra-Low-Dose Whole-Body Computed Tomography Protocol Optimization for Patients With Plasma Cell Disorders: Diagnostic Accuracy and Effective Dose Analysis From a Reference Center

Davide Tore 1,*,, Osvaldo Rampado 2,, Carla Guarnaccia 1, Roberto Mina 3, Maria Oronzio 2, Ambra Santonocito 1, Alessandro Serafini 1, Giulio Antonino Strazzarino 1, Laura Gianusso 2, Sara Bringhen 4, Paolo Fonio 1,, Alessandro Depaoli 1,
PMCID: PMC8635628  PMID: 34869000

Abstract

Background

The whole-body low-dose CT (WBLDCT) is the first-choice imaging technique in patients with suspected plasma cell disorder to assess the presence of osteolytic lesions. We investigated the performances of an optimized protocol, evaluating diagnostic accuracy and effective patient dose reduction using a latest generation scanner.

Methods and Materials

Retrospective study on 212 patients with plasma cell disorders performed on a 256-row CT scanner. First, WBLDCT examinations were performed using a reference protocol with acquisition parameters obtained from literature. A phantom study was performed for protocol optimization for subsequent exams to minimize dose while maintaining optimal diagnostic accuracy. Images were analyzed by three readers to evaluate image quality and to detect lesions. Effective doses (E) were evaluated for each patient considering the patient dimensions and the tube current modulation.

Results

A similar, very good image quality was observed for both protocols by all readers with a good agreement at repeated measures ANOVA test (p>0.05). An excellent inter-rater agreement for lesion detection was achieved obtaining high values of Fleiss’ kappa for all the districts considered (p<0.001). The optimized protocol resulted in a 56% reduction of median DLP (151) mGycm, interquartile range (IQR) 128–188 mGycm vs. 345 mGycm, IQR 302–408 mGycm), of 60% of CTDIvol (2.2 mGy, IQR 1.9–2.7 mGy vs. 0.9 mGy, IQR 0.8–1.2 mGy). The median E value was about 2.6 mSv (IQR 1.7–3.5 mSv) for standard protocol and about 1.5 mSv (IQR 1.4–1.7 mSv) for the optimized one. Dose reduction was statistically significant with p<0.001.

Conclusions

Protocol optimization makes ultra-low-dose WBLDCT feasible on latest generation CT scanners for patients with plasma cell disorders with effective doses inferior to conventional skeletal survey while maintaining excellent image quality and diagnostic accuracy. Dose reduction is crucial in such patients, as they are likely to undergo multiple whole-body CT scans during follow-up.

Keywords: ultra-low-dose whole-body CT, multiple myeloma, plasma cell disorders, effective dose, dose reduction

Introduction

Multiple myeloma (MM) is the second most common hematological malignancy, characterized by bone marrow infiltration by monoclonal plasma cells that ultimately leads to end-organ damage (1). Monoclonal gammopathy of undetermined significance (MGUS) is a premalignant plasma cell disorder preceding MM, with a risk of progression to MM of about 1% per year. Smoldering MM (SMM) represents an intermediate clinical condition between MGUS and MM, characterized by a bone marrow plasma cell infiltration ≥10% without signs or symptoms related to MM itself, with a risk of progression to symptomatic myeloma of about 10% per year (1).

The diagnosis of MM, based on the International Myeloma Working Group (IMWG) criteria, requires a bone marrow plasma cell infiltration ≥10% and at least one CRAB syndrome feature (hypercalcemia, renal failure, anemia, and bone lesions), or ≥60% monoclonal plasma cell infiltration on bone marrow biopsy, serum involved to uninvolved free light chain ratio ≥100 or more than one >5 mm focal lesion at magnetic resonance imaging or a biopsy-proven bone or extramedullary plasmacytoma.

Bone disease is the most frequent symptom related to MM; up to 80% of patients with newly diagnosed MM have osteolytic lesions (2, 3), and about 90% of patients develop bone lesions throughout the course of their disease (4).

Imaging plays a crucial role in plasma cell disorders for the diagnosis, monitoring, and management of the disease. Conventional skeletal survey (CSS) has been replaced by more advanced and sensible imaging modalities (57) such as whole-body MRI (WBMRI), whole-body low-dose CT (WBLDCT), or FDG/PET-CT (811). WBLDCT currently represents the first-choice imaging technique in patients with a suspected plasma-cell disorder to assess the presence and the extension of osteolytic lesions (12). WBLDCT has several advantages over WBMRI, as it is widely available, cheap, simple to perform, and well tolerated by patients due to a very short scan time.

Since the introduction of CT for bone evaluation in patients with MM and other plasma cell disorders, many efforts were implemented to reduce radiation exposure. As such population is likely to undergo multiple whole-body exams during treatment (13) and follow-up, it is of the utmost importance to reduce radiation dose, as performing exams with standard-dose protocol would cause unacceptably high cumulative exposures.

The aim of this study was to investigate the performances of a whole-body ultra-low-dose CT protocol, evaluating diagnostic accuracy and effective patient dose reduction achievable through a latest generation CT scanner.

Materials and Methods

Phantom Study and Protocol Optimization

The study was performed using a 256-slice CT scanner (Revolution CT, GE, USA). First, WBLDCT examinations in our department were performed in 2018 starting from a reference protocol with acquisition parameters set considering literature indications (14) (120 kV; collimation, 80 mm; rotation time, 0.28 s; pitch, 0.9; slice thickness, 1.25 mm; noise index (NI), 25; and iterative, ASIR-V 50%).

In February 2019, a phantom study was conducted to investigate the minimum dose that allowed to maintain an optimal diagnostic accuracy in lesion detection (15).

Starting from the reference protocol, the phantom was repeatedly scanned varying several acquisition parameters. In particular, in a first measurement session, the impact of different levels of iterative reconstruction ASIR-V (50%, 70%, 80%, and 100%) combined with two noise index values (25 and 50) and different kilovoltages (100, 120, and 140 kV) was investigated and used to identify the best ASIR-V percentage.

A second measurement session was performed varying kilovoltages (100, 120, and 140 kV) and the noise index (21, 25, 28, 35, and 48) in order to obtain a CTDI vol in a range from 0.3 to 1.4 mGy.

A noise analysis was performed using an automatic multiple region of interest (ROI) evaluation made by a homemade tool (implemented as a plugin of the software ImageJ, https://imagej.nih.gov/ij/index.html). Each slice was subdivided in square ROIs of 10 per 10 voxels. Objective image noise was calculated as the standard deviation of the pixel values for each ROI that was located entirely within a soft-tissue-simulating material. The median noise value was calculated for each image and compared with the set noise index, along all the phantom length. The median noise slice by slice trend and the tube current modulation trend (obtained extracting current values from the images DICOM header) were plotted over the z-axis.

The noise power spectrum in homogeneous regions and the bone edge profiles were analyzed for some images.

Each data set was scored individually by two senior radiologists (AD and CG) and a resident (DT) using a Likert scale (from 1, non-diagnostic to 3, average still diagnostic to 5, optimal), evaluating four separate anatomical areas: skull base, thoracic spine, pelvis, and distal femora.

Effective Dose Evaluation

All CT acquisition parameters and dose indexes [CTDIvol and dose length product (DLP)] were collected by a dose registration system (Physico, Emme Esse, Italy) integrated with a software for CT effective dose (E) estimation (Virtual Phantom CT). For each patient, a virtual phantom with the correspondent gender and body size was selected among the available options, considering the patient body mass index (BMI) with the following thresholds: ≤25 kg m−2 normal weight adult, >25 and ≤30 kg m−2 overweight adult, >30 and ≤ 35 kg m−2 obese level I adult, >35 and ≤40 kg m−2 obese level II adult, and >40 kg m−2 morbidly obese adult.

Figure 1 shows an example of tube current modulation trend along the patient length. Tube current values were at the minimum settable of 10 mA over the patient head and lower limbs, whereas typical values in the range of 50–100 mA were observed for the patient chest and abdomen. To account of this dose modulation, the E for each patient was calculated by Virtual Phantom CT considering the relative contribution to average CTDIvol and total DLP of each anatomical district (head, neck, chest, abdomen, pelvis, and lower limbs). The contributions for each district were summed to obtain the total E.

Figure 1.

Figure 1

(A) Median noise trend with different acquisition parameters superimposed to the phantom scout image. (B) Noise (expressed in HU) and noise power spectrum (NPS) for the skull and pelvis districts calculated with different acquisition parameters and doses; detail of the bone edge profile in the pelvis with different acquisition parameters.

In Vivo Validation

The study, approved by the Ethics Committee of our Institution, was conducted on consecutive patients who underwent WBLDCT at our department between 2018 and 2021 for diagnostic workup of plasma cell disorders, divided in two groups depending on the scanning protocol: group 1 standard protocol (23 patients, 10 women) and group 2 low-dose protocol (189 patients, 84 women). Complete patient characteristics are reported in Table 1 .

Table 1.

Characteristics of the patients.

Characteristics of the patients
Group 1 standard protocol Group 2 low-dose protocol
Number 23 189
Sex 10 female
13 male
84 female
105 male
Age (years) 66.74 ± 2.51 68.94 ± 0.80
Number of patients by age:
 • <40 years 0 (0%) 1 (0,5%)
 • 40–50 years 3 (13%) 11 (5,8%)
 • 50–60 years 4 (17.4%) 28 (13.2%)
 • 60–70 years 5 (21.7%) 47 (24.9%)
 • 70–80 years 8 (34.7%) 68 (36%)
 • 80–90 years 3 (13%) 33 (17.4%)
 • Over 90 years 0 (0%) 1 (0.5%)
BMI (kg m−2) 25.6 ± 5.7 25.1 ± 5.4
Plasma cell disorder:
 • Multiple myeloma 12 (52.1%) 85 (44.9%)
 • Smoldering myeloma 1 (4.3%) 20 (10.6%)
 • MGUS 9 (39.1%) 82 (43.4%)
 • Plasmacytoma 0 (0%) 1 (0.5%)
 • Plasma cell leukemia 1 (4.3%) 0 (0%)
 • Other 0 (0%) 1 (0.5%)

Each data set was evaluated for image quality by three radiology residents (ASa, ASe and GS) proficient in oncologic imaging with the Likert scale used for the phantom study (from 1, non-diagnostic to 3, average still diagnostic to 5, optimal) considering four anatomic districts: skull base, thoracic spine, pelvis, and distal femora.

WBLDCT exams were independently analyzed by the three readers looking for cortical lytic bone lesions and extraosseous localizations, reported for the following anatomic districts: skull, cervical spine, thoracic spine, lumbar spine, humeral head, humeral diaphysis, scapula, clavicula, coxae-sacrum, femoral head, femoral diaphysis, sternum, ribs, and non-osseous localizations.

Cortical lytic lesions were assessed with a Likert scale from 1 to 5 depending on diagnostic confidence (1, definite non-osteolysis; 2, probably non-osteolysis; 3, probable osteolysis; 4, highly suspicious for osteolysis; 5, definite osteolysis) and reported for the same districts considered for lesion detection.

Statistics

Statistical analysis was performed with open-source software jamovi [The jamovi project (2020); jamovi (Version 1.6.14.0) (Computer Software), retrieved from https://www.jamovi.org].

Continuous variables were expressed as mean ± standard deviation (SD); CTDIvol, DLP, and E were also expressed as median and interquartile ranges (IQR) to minimize the effect of extreme values.

Discrete variables were expressed as absolute numbers and percentages.

Statistical tests Student’s t, repeated measures ANOVA, and Fleiss’ kappa were used when appropriate.

A p<0.05 was considered statistically significant.

Results

Phantom Study Results

In the first measurement session, it was stated that the optimal setting for ASIR-V iterative reconstruction was 80%, regardless the other acquisition parameters.

In the second measurement session by consensus of two radiologists, it was established that the optimal combination of acquisition parameters to obtain the best achievable image quality with minimum dose of 120 kV with automatic mAs modulation, ASIR-V of 80%, and NI of 28 resulting in an average computed tomography dose index (CTDIvol) of 0.7 mGy for the phantom WBLDCT. With such acquisition setting, images received a Likert score of 5 by both readers for all districts considered.

The median noise trend with different acquisition parameters is plotted in Figure 1A superimposed to the phantom scout image.

The noise quantified with multiple automatic ROIs with different acquisition parameters is shown in Figure 1B for the skull and pelvis.

The noise power spectrum (NPS) of the different acquisitions for the skull and pelvis districts is reported in Figure 1B . NPS at different kVs and doses presented no statistically significant differences in the skull district (p>0.05), whereas a different distribution of NPS was observed in particular in the pelvis district for CTDIvol values below 0.7 mGy. These lower-dose acquisitions also highlighted a degradation of the analyzed bone edge profiles. In summary, the quantitative analysis was consistent with the subjective image evaluation and supported the choice of a NI level of 28.

The new optimized protocol was compliant with the recommendations of the IMWG bone working group, with thin slice acquisition (1.25 mm); images were acquired from the cranial vault to proximal tibial metaphysis including humeri in the field of view and reconstructed with sharp (bone, slice thickness, 0.625 mm) and standard algorithm (12).

This low-dose protocol was implemented in clinical practice since April 2019.

Effective Dose Evaluation

Figures 2 and 3 show the dose indicators DLP and CTDI distributions for the original protocol and for the optimized protocol. DLP median value was 345 mGycm for the original protocol (mean value, 342.5 ± 102 mGycm), with an interquartile range (IQR) of 302–408 mGy cm. With the optimized protocol, a 56% reduction of the median DLP was observed, resulting in a value of 151 mGy cm (IQR 128–188 mGy cm, mean 160.5 ± 55 mGy cm). A dose reduction of about 60% was observed also for CTDIvol, from an initial median value of 2.2 mGy (IQR 1.9–2.7 mGy, mean 2.2 ± 0.6 mGy cm) down to a median CTDIvol of 0.9 mGy (IQR 0.8–1.2 mGy, mean 1 ± 0.4 mGy cm). Dose reduction was statistically significant for both DLP and CTDIvol (p<0.001 in both cases).

Figure 2.

Figure 2

DLP distributions for the original protocol (1) and for the optimized protocol (2).

Figure 3.

Figure 3

CTDI distributions for the original protocol (1) and for the optimized protocol (2).

Figure 4 shows the temporal trend of the patient effective doses, highlighting the positive effect of the introduction of the optimized protocol. The median E value was about 2.6 mSv (IQR 1.7–3.5 mSv, mean 3.2 ± 1.1 mSv) in the first 8 months of examinations and about 1.5 mSv (IQR 1.4–1.7 mSv, mean 1.6 ± 0.5 mSv) after the new protocol introduction. Almost all values were comprised between 1 and 2 mSv.

Figure 4.

Figure 4

Temporal trend of the patient effective doses. The median E-value was about 2.6 mSv in the first 8 months of examinations with the standard protocol and about 1.5 mSv after the introduction of the optimized protocol.

Image Quality

We retrospectively identified 212 patients who underwent WBLDCT for plasma cell disorder workup.

There was an overall very good agreement on image quality scores between the three readers for both groups as demonstrated by repeated measures ANOVA test with p values always superior to 0.05.

Overall image quality assessed with the same anatomic regions evaluated for the previous phantom study was very high in general and considering both groups singularly.

No statistically significant differences were demonstrated at Student’s t-test comparing mean Likert scores between standard and low-dose exams (respectively, mean score for skull, 4.94 vs. 4.93, p = 0.708; for thoracic spine, 4.88 vs. 4.89, p = 0.765; for pelvis, 4.94 vs. 4.94, p = 0.882; and for distal femora, 5 vs. 4.96, p = 0.097). Complete results are shown in Table 2 . Sample images of a low-dose and standard protocol study are shown in Figure 5 .

Table 2.

Image quality scores of patients’ datasets.

Likert score of image quality of patients’ datasets
Group 1 standard protocol Repeated measures ANOVA Group 2 low-dose protocol Repeated measures ANOVA Standard vs. low-dose protocol group repeated measure ANOVA Student’s t-test
Reader 1 Reader 2 Reader 3 p Reader 1 Reader 2 Reader 3 p p Mean value standard protocol group Mean value low-dose protocol group p
Likert score skull base 4.88 ± 0.34 4.96 ± 0.2 5 0.174 4.93 ± 0.25 4.92 ± 0.27 4.95 ± 0.22 0.405 0.146 4.94 ± 0.231 4.9 ± 3 0.250 0.708
Likert score thoracic spine 4.88 ± 0.34 4.83 ± 0.38 4.92 ± 0.28 0.651 4.84 ± 0.38 4.91 ± 0.29 4.91 ± 0.29 0.073 0.352 4.88 ± 0.333 4.89 ± 0.322 0.765
Likert score pelvis 4.96 ± 0.2 4.96 ± 0.2 4.92 ± 0.28 0.779 4.94 ± 0.24 4.93 ± 0.26 4.95 ± 0.21 0.531 0.224 4.94 ± 0.231 4.94 ± 0.238 0.882
Likert score distal femora 5 5 5 4.95 ± 0.21 4.97 ± 0.16 4.96 ± 0.19 0.494 0.416 5 4.96 ± 0.189 0.097

Figure 5.

Figure 5

(A, B) Coronal and sagittal images of a patient acquired with the optimized protocol, total DLP of 157 mGy cm. (C, D) Coronal and sagittal images of a patient acquired with the standard protocol, total DLP of 434 mGy cm.

Osteolytic Lesions Detection

A total of 78, 79, and 78 lesions were detected by the three readers in the standard protocol group, while in the low-dose group, 360, 365, and 354 lesions were detected.

An excellent inter-rater agreement for lesion detection was achieved evaluating both groups singularly (complete results in Table 3 ) and also considering the whole patient population, obtaining high values of Fleiss’ kappa for all districts, ranging from 0.851 for humeral head to 1 for thoracic spine with values of p always <0.001.

Table 3.

Osteolytic lesions detection inter-rater reliability.

Osteolytic lesions detection inter-rater reliability
Group 1 standard protocol Group 2 low-dose protocol group Whole population
Reader 1 Reader 2 Reader 3 Agreement Fleiss’ kappa p reader 1 Reader 2 Reader 3 Agreement Fleiss’ kappa p Reader 1 Reader 2 Reader 3 Agreement Fleiss’ kappa p
Skull lesions 10 10 10 100 1 <0.001 36 35 34 98.9 0.977 <0.001 46 45 44 99.1 0.981 <0.001
Cervical spine lesions 6 6 6 100 1 <0.001 31 32 32 98.4 0.962 <0.001 37 38 38 98.6 0.968 <0.001
Thoracic spine lesions 9 9 9 100 1 <0.001 38 38 38 100 1 <0.001 47 47 47 100 1 <0.001
Lumbar spine lesions 7 7 7 100 1 <0.001 34 34 33 99.5 0.988 <0.001 41 41 40 99.5 0.99 <0.001
Humeral head 2 3 2 95.8 0.842 <0.001 10 11 10 99.5 0.966 <0.001 12 14 12 99.1 0.944 <0.001
Humeral diaphysis 5 5 5 100 1 <0.001 23 22 22 99.5 0.983 <0.001 28 27 27 99.5 0.986 <0.001
Scapular lesions 5 5 5 100 1 <0.001 25 26 26 99.5 0.985 <0.001 30 31 31 99.5 0.987 <0.001
Clavicular lesions 5 5 5 100 1 <0.001 22 20 20 98.9 0.964 <0.001 27 25 25 99.1 0.97 <0.001
Hip bone lesions 6 6 6 100 1 <0.001 40 41 41 97.9 0.958 <0.001 46 47 47 98.1 0.963 <0.001
Femoral head lesions 4 4 4 100 1 <0.001 13 15 11 96.3 0.807 <0.001 17 19 15 96.7 0.851 <0.001
Femoral diaphysis 6 6 6 100 1 <0.001 19 22 20 98.4 0.945 <0.001 25 28 26 98.6 0.957 <0.001
Sternum lesions 6 6 6 100 1 <0.001 24 22 22 97.4 0.916 <0.001 30 28 28 97.6 0.933 <0.001
Rib lesions 7 7 7 100 1 <0.001 41 42 41 98.9 0.979 <0.001 48 49 48 99.1 0.982 <0.001
Extraosseous lesion 0 0 0 100 4 5 4 99.5 0.921 <0.001 4 5 4 99.5 0.921 <0.001

A high inter-rater agreement was also found for Likert scale on diagnostic confidence for osteolytic lesions with no statistically significant differences among the readers found at Fleiss’ kappa test with p values superior to 0.05 in all districts in both groups and also considering the whole patient population (complete results in Table 4). Considering the mean Likert score of all readers for every single district in both groups, no statistically significant difference could be found with Student’s t-test with p-values ranging from 0.31 for femoral diaphysis lesions to 0.99 for thoracic spine lesions (complete results in Table 5). Sample images of osteolytic lesions are shown in Figures 68.

Table 4.

Likert scores on diagnostic confidence for osteolytic lesions detection.

Likert score on diagnostic confidence for osteolytic lesions detection
Group 1 standard protocol Repeated measures ANOVA Group 2 low-dose protocol Repeated measures ANOVA
Reader 1 Reader 2 Reader 3 p Reader 1 Reader 2 Reader 3 p
Skull lesions 4.8 ± 0.422 4.90 ± 0.316 5 0.368 4.94 ± 0.232 4.91 ± 0.284 4.79 ± 0.729 0.717
Cervical spine lesions 5 5 4.83 ± 0.408 0.368 4.87 ± 0.434 4.81 ± 0.397 4.75 ± 0.568 0.307
Thoracic spine lesions 4.67 ± 0.500 4.89 ± 0.333 5 0.174 4.76 ± 0.542 4.87 ± 0.343 4.89 ± 0.315 0.497
Lumbar spine lesions 5 4.82 ± 0.387 5 0.135 4.91 ± 0.379 4.82 ± 0.387 4.82 ± 0.465 0.122
Humeral head 5 5 5 0.368 4.90 ± 0.316 4.82 ± 0.405 4.7 ± 0.483 0.472
Humeral diaphysis 4.8 ± 0.447 4.2 ± 0.837 4.8 ± 0.447 0.368 4.87 ± 0.344 4.82 ± 0.395 4.91 ± 0.294 0.607
Scapular lesions 4.6 ± 0.894 4.8 ± 0.447 4.4 ± 0.894 0.497 4.83 ± 0.482 4.81 ± 0.402 4.77 ± 0.587 0.756
Clavicular lesions 5 4.6 ± 0.548 4.8 ± 0.447 0.368 4.91 ± 0.294 4.75 ± 0.444 4.85 ± 0.366 0.18
Hip bone lesions 4.83 ± 0.408 4.83 ± 0.408 4.5 ± 0.548 0.368 4.56 ± 0.788 4.46 ± 0.897 4.44 ± 0.808 0.825
Femoral head lesions 5 4.75 ± 0.500 5 0.135 4.85 ± 0.555 4.87 ± 0.352 5 0.368
Femoral diaphysis 4.67 ± 0.516 4.67 ± 0.516 5 0.264 4.95 ± 0.229 4.86 ± 0.351 4.90 ± 0.308 0.549
Sternum lesions 5 4.83 ± 0.408 5 0.135 4.96 ± 0.204 4.77 ± 0.429 4.77 ± 0.429 0.06
Rib lesions 4.86 ± 0.378 4.71 ± 0.488 4.43 ± 0.787 0.549 4.51 ± 0.746 4.57 ± 0.630 4.34 ± 0.825 0.265
Extraosseous lesion 4.71 ± 0.488 4.60 ± 0.894 4.5 ± 0.577 0.717

Table 5.

Mean results for whole population Likert scores on diagnostic confidence for osteolytic lesions detection and comparison of mean scores for both groups.

Mean score whole population Repeated measures ANOVA Student’s t for mean results of all readers
Reader 1 Reader 2 Reader 3 p Mean standard protocol group Mean low dose protocol group p
Skull lesions 4.91 ± 0.285 4.91 ± 0.288 4.84 ± 0.645 1 4.9 ± 0.305 4.89 ± 0.466 0.924
Cervical spine lesions 4.89 ± 0.398 4.84 ± 0.370 4.76 ± 0.542 0.264 4.94 ± 0.236 4.81 ± 0.470 0.414
Thoracic spine lesions 4.74 ± 0.530 4.87 ± 0.337 4.91 ± 0.285 0.155 4.85 ± 0.362 4.84 ± 0.413 0.99
Lumbar spine lesions 4.93 ± 0.346 4.83 ± 0.381 4.85 ± 0.427 0.078 4.95 ± 0.218 4.85 ± 0.41 0.476
Humeral head 4.92 ± 0.289 4.86 ± 0.363 4.75 ± 0.452 0.472 5 4.81 ± 0.402 0.384
Humeral diaphysis 4.86 ± 0.356 4.7 ± 0.542 4.89 ± 0.32 0.264 4.6 ± 0.632 4.87 ± 0.344 0.051
Scapular lesions 4.79 ± 0.559 4.81 ± 0.402 4.71 ± 0.643 0.819 4.60 ± 0.632 4.80 ± 0.490 0.336
Clavicular lesions 4.93 ± 0.267 4.72 ± 0.458 4.84 ± 0.374 0.067 4.8 ± 0.414 4.84 ± 0.371 0.924
Hip bone lesions 4.6 ± 0.751 4.51 ± 0.856 4.45 ± 0.775 0.578 4.72 ± 0.461 4.49 ± 0.828 0.427
Femoral head lesions 4.88 ± 0.485 4.84 ± 0.375 5 0.135 4.92 ± 0.289 4.9 ± 0.384 0.984
Femoral diaphysis 4.88 ± 0.332 4.82 ± 0.390 4.92 ± 0.272 0.417 4.78 ± 0.428 4.90 ± 0.300 0.31
Sternum lesions 4.97 ± 0.183 4.79 ± 0.418 4.82 ± 0.390 0.061 4.94 ± 0.236 4.84 ± 0.371 0.441
Rib lesions 4.56 ± 0.712 4.59 ± 0.61 4.35 ± 0.812 0.607 4.67 ± 0.577 4.48 ± 0.738 0.455
Extraosseous lesion 4.71 ± 0.577 4.60 ± 0.894 4.5 ± 0.577 0.717 4.63 ± 0.619

Figure 6.

Figure 6

Patient acquired with standard protocol. (A) Cervical spine osteolytic lesions. (B) Thoracic spine and sternal and costal lesions. (C) Sacrum and hip bone lesions.

Figure 8.

Figure 8

Patient acquired with low-dose protocol. (A, B) Thoracic spine and costal osteolytic lesion. (C, D) Sacrum and hip bone lesions.

Figure 7.

Figure 7

Same patient in Figure 6. Multiple osteolytic lesions can be appreciated on maximum intensity projection (MIP) reconstruction with thick slab (125 mm).

Discussion

The phantom study allowed us to introduce and validate a new optimized protocol for WBLDCT for patients with plasma cell disorders with a significant reduction in radiation dose compared with other experiences reported in the literature. We obtained low levels of exposure in a large cohort of patients with doses that are even lower than those reported in literature for conventional skeletal survey (ranging from 1.5 to 2.5 mSv, depending on patient BMI and equipment used) (14).

Despite a reduction of about 60% in radiation dose administered to the patients, the new lower dose protocol generated images of non-inferior diagnostic quality compared to the previous protocol, with very good or excellent quality scores. The difference in image quality scores was statistically not significant in all the anatomic districts considered.

We observed an excellent inter-rater reliability in identifying cortical osteolytic lesions in both groups considered individually and also considering the whole population with very high values of Fleiss’ kappa, thus demonstrating the non-inferiority of the low-dose protocol in lesion detection compared with the previous one.

Mean Likert scores on diagnostic confidence for lesion detection in the low-dose group ranged from 4.48 for rib and 4.49 for hip bone lesions to 4.9 for femoral head and femoral diaphysis lesions with a statistically significant difference (p < 0.001 and p = 0.003, respectively). The values of diagnostic confidence for the ribs and the hip bone regions did not present a statistically significant difference when compared to the standard protocol group (p = 0.261 and 0.246, respectively). These results are probably due to the higher noise that is present in such districts because of attenuation from surrounding structures (i.e., the spine and the arms and the hip bones and sacrum), which makes it more challenging to detect osteolytic lesions. Nevertheless, the diagnostic confidence in such regions is still very good, and it is excellent in the other districts.

Effective dose in CT is often estimated by multiplying the DLP by a proper multiplication coefficient (16). However, several recent studies pointed out the importance of considering the patient size in CT effective dose estimation, since with the same DLP, the effective dose will be significantly higher for small BMI patients and lower for large BMI patients (1720). The use of a software with a library of virtual phantoms with different sizes allows a more accurate dose estimation. Another crucial point for an accurate E estimation for WBLDCT is the proper account of tube current modulation. We have verified that the E calculation without considering the modulation and using the average CTDIvol constant over all the patients resulted in a value about 30% lower. In our study, the average ratio between the E and DLP was 0.011 mSv mGy−1 cm−1. This coefficient should be considered comparing different studies providing E-values.

Since 1985, several studies aimed at reducing radiation dose in total body CT scans for patients with plasma cell disorders. The study by Schreiman et al. demonstrated that standard-dose CT could identify more osteolytic lesions than CSS in patients with MM (5). However, using a standard-dose whole-body CT protocol resulted in unacceptable high cumulative radiation doses for patients. Mahnken et al. described a conventional high-dose CT of the thoracic and lumbar spine to evaluate bone disease in a population of patients affected by MM with effective doses ranging from 25.5 and 36.6 mSv (21). In 2005, Horger et al. published the first feasibility study for a multidetector WBLDCT protocol for MM patients, as an alternative to CSS with doses ranging from 4.1 to 7.5 mSv depending on the tube current applied with a non-significant decrease in image quality (22). Kropil et al. proposed a low-dose protocol using a tube voltage of 100 kV and an effective time–current product of 100 mAs, with automatic tube current modulation resulting in an effective dose of approximately 4.8 mSv (23). Gleeson et al. applied several different WBLDCT protocols on a single cadaver achieving a dose of 1.74 mSv, comparable to CSS while still providing sufficient diagnostic quality (24). Hemke et al. presented a retrospective study on different CT scanners with effective dose ranging from 4.34 to 8.37 mSv (25). Greffier et al. performed a phantom study where they were able to obtain images with sufficient diagnostic quality with a dose level as low as 3.4 mGy, using high levels of iterative reconstruction (26).

In our study, we obtained a median E of 1.5 mSv in a real-life population of patients, which is significantly lower than the doses reported in the works cited above. The cadaver study by Gleeson is the only other article with doses in a similar order of magnitude (24); however, a study conducted on a single human body cannot represent an accurate estimation of effective dose in a wider population with different BMIs.

Only one recent study (27) showed the possibility to perform WBLDCT with doses lower than those observed in our patient sample (average CTDIvol of 0.3 mGy and DLP of 52 mGy cm), with the use of a dual source CT with spectral shaping over a population of 30 patients; however, the authors did not provide information about the patients’ BMIs, which has a considerable impact on dose estimation.

Even if most patients with MM are older adults, according to the literature, there is a non-negligible number of younger adults affected by plasma cell disorders, being about 10% of such patients younger than 50 years and 2% younger than 40 years, respectively (28). In our study population, 15 patients were 50 years old or younger (13% in group 1, 6.3% in group 2); many patients had smoldering myeloma or MGUS (respectively, 4.3% and 39.1% in group 1, 10.6% and 43.4% in group 2). Such subgroup of patients is particularly likely to undergo very long follow-ups with repeated WBLDCT scans for early identification of bone lesions (13), making it even more important to minimize radiation exposure. Dose reduction should be mandatory even in older patients, as the outcomes of plasma cell disorders are improving with current and future therapies (29); they may experience periods of disease remission or stability, and they may be treated with different lines of therapy over time. Patients with plasma cell disorders may also need to undergo exams with hybrid scanners to assess response to treatments such as PET-CT or PET-MRI, contributing to increase the radiation burden to which they are exposed.

This study has some limitations: the results cannot be directly transposed on different CT scanners, using the same acquisition parameters (i.e., kV, mA range and modulation) on a different machine, and setting a high level of iterative reconstruction may result in different performances in terms of radiation dose and image quality. For example, the minimum CTDIvol achievable with tube current modulation on a GE OPTIMA CT 660 with 64-rows detector was 1.4 mGy, with greater noise and a triple total scan time (15 s) compared to the revolution CT used in this study. However, the methodology described can be applied to every CT scanner with iterative reconstruction algorithms to obtain the best compromise between radiation dose and image quality that can be achieved on different scanners. Certainly, the best performances could be expected with state-of-the-art hardware and software, but a discrete dose reduction may be achieved even on older CT scanners.

The lowest achievable dose for WBLDCT can be assessed with phantom studies and proper image quality metrics (including noise power spectrum, task transfer function, and subjective image quality scores), thus demonstrating that routine ultra-low-dose WBLDCT is feasible on latest generation CT scanners with a proper balance between tube current modulation parameters and iterative reconstruction strength, resulting in excellent image quality and diagnostic accuracy with significant dose reduction for patients respecting the low as reasonably practicable (ALARA)/as low as reasonably achievable (ALARP) principles (as low as reasonably achievable).

Technological evolution of CT scanners, the use of iterative reconstruction and adequate protocol optimization allow to obtain high-quality total body studies of patients with MM and other plasma cell disorders with radiation doses that are inferior to CSS but with an excellent sensibility in bone lesion detection.

Data Availability Statement

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author.

Ethics Statement

The studies involving human participants were reviewed and approved by Comitato Etico Interaziendale A.O.U. Città della Salute e della Scienza di Torino—A.O. Ordine Mauriziano—A.S.L. Città di Torino. Written informed consent for participation was not required for this study in accordance with the national legislation and the institutional requirements.

Author Contributions

DT and AD: conception, design, phantom study subjective image analysis, and statistical analysis. OR: conception, design, and dose analysis. CG: conception, design, and phantom study subjective image analysis. PF, RM, and SB: conception and design. MO: dose analysis and statistical analysis. ASa, ASe, and GS: in vivo image analysis and management of clinical data. LG: dose analysis. All authors contributed to the article and approved the submitted version.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Acknowledgments

The study was approved by the Ethics Committee of our Institution. Part of the results, regarding the phantom study, were presented at RSNA 2019 Scientific Assembly and Annual Meeting by OR (15).

References

  • 1. Vincent Rajkumar S, Dimopoulos MA, Palumbo A, Blade J, Merlini G, Mateos M-V, et al. Review International Myeloma Working Group Updated Criteria for the Diagnosis of Multiple Myeloma. (2014). Available at: www.thelancet.com/oncology. [DOI] [PubMed]
  • 2. Terpos E, Ntanasis-Stathopoulos I, Dimopoulos MA. Myeloma Bone Disease: From Biology Findings to Treatment Approaches. (2019). Available at: http://ashpublications.org/blood/article-pdf/133/14/1534/1552921/blood852459.pdf. [DOI] [PubMed]
  • 3. Terpos E, Zamagni E, Lentzsch S, Drake MT, García-Sanz R, Abildgaard N, et al. Treatment of Multiple Myeloma-Related Bone Disease: Recommendations From the Bone Working Group of the International Myeloma Working Group. Lancet Oncol (2021) 22:e119–30. doi:  10.1016/S1470-2045(20)30559-3 [DOI] [PubMed] [Google Scholar]
  • 4. Ippolito D, Talei Franzesi C, Spiga S, Besostri V, Pezzati S, Rossini F, et al. Diagnostic Value of Whole-Body Ultra-Low Dose Computed Tomography in Comparison With Spinal Magnetic Resonance Imaging in the Assessment of Disease in Multiple Myeloma. Br J Haematol (2017) 177:395–403. doi:  10.1111/bjh.14545 [DOI] [PubMed] [Google Scholar]
  • 5. Schreiman JS, McLeod RA, Kyle RA, Beabout JW. Multiple Myeloma: Evaluation by CT. Radiology (1985) 154:483–6. doi:  10.1148/radiology.154.2.3966137 [DOI] [PubMed] [Google Scholar]
  • 6. Lambert L, Ourednicek P, Meckova Z, Gavelli G, Straub J, Spicka I. Whole-Body Low-Dose Computed Tomography in Multiple Myeloma Staging: Superior Diagnostic Performance in the Detection of Bone Lesions, Vertebral Compression Fractures, Rib Fractures and Extraskeletal Findings Compared to Radiography With Similar Radiation Exposure. Oncol Lett (2017) 13:2490–4. doi:  10.3892/ol.2017.5723 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Filho AGO, Carneiro BC, Pastore D, Silva IP, Yamashita SR, Consolo FD, et al. Whole-Body Imaging of Multiple Myeloma: Diagnostic Criteria. Radiographics (2019) 39:1077–97. doi:  10.1148/rg.2019180096 [DOI] [PubMed] [Google Scholar]
  • 8. di Giuliano F, Picchi E, Muto M, Calcagni A, Ferrazzoli V, da Ros V, et al. Radiological Imaging in Multiple Myeloma: Review of the State-of-the-Art. Neuroradiology (2020) 62:905–23. doi:  10.1007/s00234-020-02417-9 [DOI] [PubMed] [Google Scholar]
  • 9. Gleeson TG, Moriarty J, Shortt CP, Gleeson JP, Fitzpatrick P, Byrne B, et al. Accuracy of Whole-Body Low-Dose Multidetector CT (WBLDCT) Versus Skeletal Survey in the Detection of Myelomatous Lesions, and Correlation of Disease Distribution With Whole-Body MRI (WBMRI). Skeletal Radiol (2009) 38:225–36. doi:  10.1007/s00256-008-0607-4 [DOI] [PubMed] [Google Scholar]
  • 10. Gariani J, Westerland O, Natas S, Verma H, Cook G, Goh V. Comparison of Whole Body Magnetic Resonance Imaging (WBMRI) to Whole Body Computed Tomography (WBCT) or 18F-Fluorodeoxyglucose Positron Emission Tomography/CT (18f-FDG PET/CT) in Patients With Myeloma: Systematic Review of Diagnostic Performance. Crit Rev Oncol/Hematol (2018) 124:66–72. doi:  10.1016/j.critrevonc.2018.02.012 [DOI] [PubMed] [Google Scholar]
  • 11. Regelink JC, Minnema MC, Terpos E, Kamphuis MH, Raijmakers PG, Pieters - van den Bos IC, et al. Comparison of Modern and Conventional Imaging Techniques in Establishing Multiple Myeloma-Related Bone Disease: A Systematic Review. Br J Haematol (2013) 162:50–61. doi:  10.1111/bjh.12346 [DOI] [PubMed] [Google Scholar]
  • 12. Hillengass J, Usmani S, Rajkumar V, Durie BGM, Mateos M-V, Lonial S, et al. Review International Myeloma Working Group Consensus Recommendations on Imaging in Monoclonal Plasma Cell Disorders. (2019). Available at: www.thelancet.com/oncology. [DOI] [PubMed]
  • 13. Gavriatopoulou M, Boultadaki A, Koutoulidis V, Ntanasis-Stathopoulos I, Bourgioti C, Malandrakis P, et al. The Role of Low Dose Whole Body CT in the Detection of Progression of Patients With Smoldering Multiple Myeloma. Blood Cancer J (2020) 10. doi:  10.1038/s41408-020-00360-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Moulopoulos LA, Koutoulidis V, Hillengass J, Zamagni E, Aquerreta JD, Roche CL, et al. Recommendations for Acquisition, Interpretation and Reporting of Whole Body Low Dose CT in Patients With Multiple Myeloma and Other Plasma Cell Disorders: A Report of the IMWG Bone Working Group. Blood Cancer J (2018) 8. doi:  10.1038/s41408-018-0124-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Rampado O, Depaoli A, Lidonni L, Gianusso La, Guarnaccia C, Franchi M, et al. “Protocol Optimization of Whole-Body Low-Dose CT in Patients With Multiple Myeloma: How Low Is Too Low?” In: Oral Presentation, RSNA 2019 Scientific Assembly and Annual Meeting. Chicago, USA: Radiological Society of North America; (2019). [Google Scholar]
  • 16. Deak PD, Smal Y, Kalender WA. Multisection CT Protocols: Sex-And Age-Specific Conversion Factors Used to Determine Effective Dose From Dose-Length Product. Radiology (2010) 257(1):158–66. doi:  10.1148/radiol.10100047 [DOI] [PubMed] [Google Scholar]
  • 17. Wang J, Duan X, Christner JA, Leng S, Yu L, McCollough CH. Attenuation-Based Estimation of Patient Size for the Purpose of Size Specific Dose Estimation in CT. Part I. Development and Validation of Methods Using the CT Image. Med Phys (2012) 39:6764–71. doi:  10.1118/1.4754303 [DOI] [PubMed] [Google Scholar]
  • 18. Wang J, Christner JA, Duan X, Leng S, Yu L, McCollough CH. Attenuation-Based Estimation of Patient Size for the Purpose of Size Specific Dose Estimation in CT. Part II. Implementation on Abdomen and Thorax Phantoms Using Cross Sectional CT Images and Scanned Projection Radiograph Images. Med Phys (2012) 39:6772–8. doi:  10.1118/1.4757586 [DOI] [PubMed] [Google Scholar]
  • 19. Klosterkemper Y, Appel E, Thomas C, Bethge OT, Aissa J, Kröpil P, et al. Tailoring CT Dose to Patient Size: Implementation of the Updated 2017 ACR Size-Specific Diagnostic Reference Levels. Acad Radiol (2018) 25:1624–31. doi:  10.1016/j.acra.2018.03.005 [DOI] [PubMed] [Google Scholar]
  • 20. Franck C, Vandevoorde C, Goethals I, Smeets P, Achten E, Verstraete K, et al. The Role of Size-Specific Dose Estimate (SSDE) in Patient-Specific Organ Dose and Cancer Risk Estimation in Paediatric Chest and Abdominopelvic CT Examinations. Eur Radiol (2016) 26:2646–55. doi:  10.1007/s00330-015-4091-7 [DOI] [PubMed] [Google Scholar]
  • 21. Mahnken AH, Wildberger JE, Gehbauer G, Schmitz-Rode T, Blaum M FU, Günther RW. Multidetector CT of the Spine in Multiple Myeloma: Comparison With MR Imaging and Radiography. Available at: www.ajronline.org. [DOI] [PubMed]
  • 22. Horger M, Claussen CD, Bross-Bach U, Vonthein R, Trabold T, Heuschmid M, et al. Whole-Body Low-Dose Multidetector Row-CT in the Diagnosis of Multiple Myeloma: An Alternative to Conventional Radiography. Eur J Radiol (2005) 54:289–97. doi:  10.1016/j.ejrad.2004.04.015 [DOI] [PubMed] [Google Scholar]
  • 23. Kröpil P, Fenk R, Fritz LB, Blondin D, Kobbe G, Mödder U, et al. Comparison of Whole-Body 64-Slice Multidetector Computed Tomography and Conventional Radiography in Staging of Multiple Myeloma. Eur Radiol (2008) 18:51–8. doi:  10.1007/s00330-007-0738-3 [DOI] [PubMed] [Google Scholar]
  • 24. Gleeson TG, Byrne B, Kenny P, Last J, Fitzpatrick P, O’Gorman P, et al. Image Quality in Low-Dose Multidetector Computed Tomography: A Pilot Study to Assess Feasibility and Dose Optimization in Whole-Body Bone Imaging. Can Assoc Radiologists J (2010) 61:258–64. doi:  10.1016/j.carj.2010.01.003 [DOI] [PubMed] [Google Scholar]
  • 25. Hemke R, Yang K, Husseini J, Bredella MA, Simeone FJ. Organ Dose and Total Effective Dose of Whole-Body CT in Multiple Myeloma Patients. Skeletal Radiol (2020) 49:549–54. doi:  10.1007/s00256-019-03292-z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Greffier J, Frandon J, Pereira F, Hamard A, Beregi JP, Larbi A, et al. Optimization of Radiation Dose for CT Detection of Lytic and Sclerotic Bone Lesions: A Phantom Study. Eur Radiol (2020) 30:1075–8. doi:  10.1007/s00330-019-06425-z [DOI] [PubMed] [Google Scholar]
  • 27. Baldi D, Tramontano L, Alfano V, Punzo B, Cavaliere C, Salvatore M. Whole Body Low Dose Computed Tomography Using Third-Generation Dual-Source Multidetector With Spectral Shaping: Protocol Optimization and Literature Review. Dose-Response (2020) 18. doi:  10.1177/1559325820973131 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Kazandjian D. Multiple Myeloma Epidemiology and Survival: A Unique Malignancy. Semin Oncol (2016) 43:676–81. doi:  10.1053/j.seminoncol.2016.11.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Teoh PJ, Chng WJ. CAR T-Cell Therapy in Multiple Myeloma: More Room for Improvement. Blood Cancer J (2021) 11:84. doi:  10.1038/s41408-021-00469-5 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author.


Articles from Frontiers in Oncology are provided here courtesy of Frontiers Media SA