Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 2000 Apr;38(4):1339–1346. doi: 10.1128/jcm.38.4.1339-1346.2000

Evaluation of the Wider System, a New Computer-Assisted Image-Processing Device for Bacterial Identification and Susceptibility Testing

Rafael Cantón 1,*, María Pérez-Vázquez 1, Antonio Oliver 1, Begoña Sánchez Del Saz 1, M Olga Gutiérrez 1, Manuel Martínez-Ferrer 1, Fernando Baquero 1
PMCID: PMC86442  PMID: 10747104

Abstract

The Wider system is a newly developed computer-assisted image-processing device for both bacterial identification and antimicrobial susceptibility testing. It has been adapted to be able to read and interpret commercial MicroScan panels. Two hundred forty-four fresh consecutive clinical isolates (138 isolates of the family Enterobacteriaceae, 25 nonfermentative gram-negative rods [NFGNRs], and 81 gram-positive cocci) were tested. In addition, 100 enterobacterial strains with known β-lactam resistance mechanisms (22 strains with chromosomal AmpC β-lactamase, 8 strains with chromosomal class A β-lactamase, 21 broad-spectrum and IRT β-lactamase-producing strains, 41 extended-spectrum β-lactamase-producing strains, and 8 permeability mutants) were tested. API galleries and National Committee for Clinical Laboratory Standards (NCCLS) microdilution methods were used as reference methods. The Wider system correctly identified 97.5% of the clinical isolates at the species level. Overall essential agreement (±1 log2 dilution for 3,719 organism-antimicrobial drug combinations) was 95.6% (isolates of the family Enterobacteriaceae, 96.6%; NFGNRs, 88.0%; gram-positive cocci, 95.6%). The lowest essential agreement was observed with Enterobacteriaceae versus imipenem (84.0%), NFGNR versus piperacillin (88.0%) and cefepime (88.0%), and gram-positive isolates versus penicillin (80.4%). The category error rate (NCCLS criteria) was 4.2% (2.0% very major errors, 0.6% major errors, and 1.5% minor errors). Essential agreement and interpretive error rates for eight β-lactam antibiotics against isolates of the family Enterobacteriaceae with known β-lactam resistance mechanisms were 94.8 and 5.4%, respectively. Interestingly, the very major error rate was only 0.8%. Minor errors (3.6%) were mainly observed with amoxicillin-clavulanate and cefepime against extended-spectrum β-lactamase-producing isolates. The Wider system is a new reliable tool which applies the image-processing technology to the reading of commercial trays for both bacterial identification and susceptibility testing.


Automatic or semiautomatic commercial systems for bacterial identification and susceptibility testing were introduced in clinical microbiology laboratories more than 20 years ago (1, 11). These systems are specifically designed to allow reliable bacterial identification by using a number of biochemical tests and MIC determinations that are interpreted according to the susceptibility and resistance criteria established by different committees (7). Most of these systems are highly automated, particularly for MIC determinations and interpretations. The final report should offer an acceptable accuracy and should reproduce the values obtained by reference methods (4, 7, 9).

Bacterial identification and susceptibility testing systems vary in the methods that they use to detect bacterial growth and/or to determine endpoints. Either turbidimetric monitoring of bacterial growth and fluorometric detection of the fluorescent indicator or the hydrolysis of fluorogenic substrates is extensively used (7). In contrast, image analysis technology has rarely been applied to these systems and is limited to use with certain devices that analyze inhibition zones obtained by disk susceptibility test methods (3, 10).

The Wider system (Francisco Soria Melguizo, S.A., Madrid, Spain) is a newly developed computer-assisted image-processing device. With the assistance of a video camera it recovers a complete image of commercial microdilution panels used for bacterial identification and susceptibility testing. After image digitization, the Wider system automatically generates the bacterial name and the susceptibility profile, which depend on the analysis of growth and color changes in the identification wells and the interpretation of growth parameters in the susceptibility testing wells, respectively. We report here the results of an evaluation of the newly developed Wider system, which has been adapted to read MicroScan panels (Dade-MicroScan, West Sacramento, Calif.). This work was specifically designed to study the accuracy of this system as a routine tool in clinical microbiology laboratories. Moreover, the susceptibility testing performance of this instrument was also determined with isolates of the family Enterobacteriaceae with known resistance mechanisms.

MATERIALS AND METHODS

Bacterial isolates.

A total of 244 fresh bacterial clinical isolates, prospectively and consecutively collected in our hospital during January and February 1999, were studied. They included 138 isolates of the family Enterobacteriaceae, 25 nonfermentative gram-negative rods (NFGNRs), 51 Staphylococcus spp., 2 Micrococcus spp., 22 Enterococcus spp., and 6 β-hemolytic streptococcal isolates. Moreover, 100 clinical Enterobacteriaceae isolates with known resistance mechanisms were also included (Table 1). Prior to identification and susceptibility testing, the organisms were subcultured twice onto 5% sheep blood agar plates.

TABLE 1.

Enterobacteriaceae with known β-lactam-resistance mechanisms

Resistance mechanism
Organism (no. of isolates tested)
Primary Secondary
Chromosomal AmpC β-lactamase, basally expressed Enterobacter cloacae (1)
Chromosomal AmpC β-lactamase, inducibly expressed Enterobacter aerogenes (1)
Enterobacter cloacae (2)
Citrobacter freundii (1)
Morganella morganii (2)
Chromosomal AmpC β-lactamase, constitutively hyperproduced Enterobacter aerogenes (1)
Enterobacter cloacae (2)
Citrobacter freundii (1)
Morganella morganii (2)
Escherichia coli (9)
Chromosomal class A β-lactamase Proteus vulgaris (2)
Kluyvera spp. (2)
Citrobacter diversus (2)
Klebsiella oxytoca (2)
TEM-1 β-lactamase Escherichia coli (4)
Proteus mirabilis (2)
Salmonella spp. (2)
TEM-1 β-lactamase, hyperproduced Escherichia coli (4)
OXA-1 β-lactamase Escherichia coli (1)
Salmonella spp. (1)
SHV-1 β-lactamase Klebsiella pneumoniae (4)
IRT β-lactamase Escherichia coli (3)
Porin-deficient mutant (OmpF) Escherichia coli (3)
Porin-deficient mutant (OmpF) TEM-1 β-lactamase Escherichia coli (4)
Porin-deficient mutant (OmpK36) SHV-1 β-lactamase Klebsiella pneumoniae (1)
Extended spectrum β-lactamase Porin-deficient mutant (OmpF) Escherichia coli (1)
Extended spectrum β-lactamase Chromosomal AmpC β-lactamase, inducibly expressed Enterobacter cloacae (3)
Enterobacter aerogenes (1)
Citrobacter freundii (1)
Extended-spectrum β-lactamase Chromosomal AmpC β-lactamase, constitutively hyperproduced Enterobacter cloacae (1)

Wider system.

The Wider system is basically composed of a reader module assisted by a data analysis module. The reader module is an illuminated chamber with a digitizing video camera that completely reflects the image of a commercial tray used for bacterial identification and susceptibility testing. The only action required by the operator is manual insertion of the trays into the reader module with the assistance of a special support. The rest of the process is computer controlled. The digitized image is analyzed by the Wider system's software in the data analysis module. A clear image appears on the computer screen within 5 s. The software automatically detects the type of panel, assigns identification probability scores as a result of the analysis of growth and color changes in the identification wells, and identifies each isolate by comparing the biochemical profile with the profiles in the software database. Moreover, growth parameters in susceptibility testing wells are analyzed in comparison with those in positive and negative control wells. The MIC of each antibiotic is defined as the lowest concentration with the absence of bacterial growth. For categorization purposes, MICs are interpreted by using either the guidelines of the National Committee for Clinical Laboratory Standards (NCCLS) (20) or those from the Spanish Antibiogram Committee (Mesa Española para la Normalización de la Susceptibilidad y Resistencia a los Antimicrobianos [MENSURA]) guidelines from the Spanish Antibiogram Committee (2).

Bacterial identification and susceptibility testing with the Wider system.

Wider system 6W and 3W panels containing lyophilized antibiotics and substrates were used for bacterial identification and susceptibility testing for gram-negative and gram-positive bacteria, respectively. Dade-MicroScan manufactured these panels, which are similar to those used in the overnight WalkAway system (1). Biochemical tests used for both gram-negative and gram-positive organism identification are identical to those included in the WalkAway MicroScan panels. Moreover, the biochemical identification database is the same as that used in the WalkAway system. In addition, Wider 5W panels, which only have antimicrobial agents for susceptibility testing, were also used to assay all isolates. The antibiotics used in the panels (concentration ranges) are as follows: for the Wider panel for gram-negative organisms (reference 6W), amikacin (4 to 16 μg/ml), amoxicillin (4 to 16 μg/ml), amoxicillin-clavulanate (4/2 to 32/16 μg/ml), cefazolin (2 to 16 μg/ml), cefotaxime (0.12 to 8 μg/ml), cefoxitin (4 to 16 μg/ml), ceftazidime (0.5 to 16 μg/ml), ceftazidime-clavulanate (1/4 and 8/4 μg/ml), cefuroxime (1 to 16 μg/ml), ciprofloxacin (0.12 and 1 to 4 μg/ml), fosfomycin (8 to 32 μg/ml), gentamicin (2 to 8 μg/ml), nalidixic acid (4 and 16 μg/ml), nitrofurantoin (64 μg/ml), norfloxacin (1 and 4 μg/ml), ticarcillin (16 to 64 μg/ml), trimethoprim-sulfamethoxazole (2/38 to 4/76 μg/ml), and tobramycin (2 to 8 μg/ml); for the Wider panel for gram-positive organisms (reference 3W), amikacin (4 to 16 μg/ml), amoxicillin-clavulanate (4/2 to 32/16 μg/ml), ampicillin (0.5 to 16 μg/ml), cefazolin (2 to 4 μg/ml), cefotaxime (0.06 to 4 μg/ml), cefuroxime (0.5 to 2 μg/ml), chloramphenicol (8 μg/ml), ciprofloxacin (0.5 to 4 μg/ml), clindamycin (0.5 and 2 μg/ml), erythromycin (0.12 and 0.5 to 2 μg/ml), fosfomycin (32 to 64 μg/ml), gentamicin (2 to 8 and 500 μg/ml), oxacillin (1 to 4 μg/ml), penicillin (0.06 to 8 μg/ml), rifampin (0.5 and 2 μg/ml), streptomycin (1,000 μg/ml), teicoplanin (1 to 16 μg/ml), trimethoprim-sulfamethoxazole (1/19 to 2/38 μg/ml), and vancomycin (1 to 16 μg/ml); for the supplementary panel (reference 5W), aztreonam (0.12 to 16 μg/ml), cefepime (0.12 to 16 μg/ml), ceftriaxone (0.12 to 16 μg/ml), chloramphenicol (1 to 8 μg/ml), colistin (2 to 4 μg/ml), imipenem (0.125 to 16 μg/ml), meropenem (0.12 to 16 μg/ml), ofloxacin (0.06 to 8 μg/ml), piperacillin (8 to 64 μg/ml), piperacillin-tazobactam (8/4 to 64/4 μg/ml), sulbactam (1 to 8 μg/ml), tetracycline (1 to 16 μg/ml), and trovafloxacin (0.06 to 4 μg/ml).

The panels were inoculated with a standardized inoculum by using a rehydrator-inoculator (RENOK) by following the guidelines provided by the manufacturer. The inoculum was prepared with the Prompt inoculation system (29). After overnight incubation in a conventional chamber, the panels were introduced into the Wider system. The results of the external reactions, the oxidase test for gram-negative organisms, and catalase and hemolysis tests for gram-positive organisms should be marked in a special square in the commercial tray. Consequently, the video camera reflects these results at the same time as the other biochemical reactions.

Reference methods.

API galleries (API 20E, API 20NE, API Staph, and API Strep; BioMerieux, SA, Marcy-l'Étoile, France) were used as the reference tests for bacterial identification. Conventional biochemical tests (17) were also performed when discrepancies were observed. The MIC results obtained with the Wider system were compared with those obtained by the reference broth microdilution method described by NCCLS (19). The final inoculum concentration was 5 × 105 CFU/ml. Antimicrobial powders were obtained from their respective manufacturers.

Quality control.

Quality control was assured by running every day the organisms recommended for this purpose by NCCLS (20): Escherichia coli ATCC 25922 and ATCC 35218, Staphylococcus aureus ATCC 29213, Enterococcus faecalis ATCC 29212, and Pseudomonas aeruginosa ATCC 27853. The reproducibility of the MIC readings was also analyzed with the American Type Culture Collection (ATCC) strains. These strains and Proteus vulgaris ATCC 13315, Klebsiella pneumoniae ATCC 29665, Klebsiella oxytoca ATCC 49131, Salmonella enterica subsp. arizonae ATCC 12323, Enterobacter cloacae ATCC 13047, Stenotrophomonas maltophilia ATCC 13637, Staphylococcus epidermidis ATCC 12228, and Listeria monocytogenes ATCC 19112 were run during the evaluation as quality controls for bacterial identification.

Analysis and accuracy of results.

The identification scores provided by the Wider system software were not considered in the analysis; instead, the final identification results were used. A “correctly identified” category was established when identical identifications at the species level were provided by the Wider system and by the reference method. The “partially identified” category meant that the strain identifications were coincident only at the genus level, and the “incorrectly identified” category meant that the genus assigned by the Wider system was different from that obtained assigned by the reference method.

Susceptibility testing agreements or discrepancies were analyzed by considering all the organism-antimicrobial agent combinations with 18, 12, 13, 9, and 8 antimicrobial agents for the Enterobacteriaceae, NFGNRs, Staphylococcus spp. and Micrococcus spp., Enterococcus spp., and β-hemolytic streptococcal isolates, respectively. “Essential agreement” was defined when the MICs obtained with the Wider system and by the reference method were identical or ±1 log2 dilution. Moreover, by using the interpretive NCCLS criteria (20), qualitative analysis was also performed. Those antibiotics for which the concentrations present in the evaluated panel did not allow the use of NCCLS criteria were excluded from the analysis. When the MIC was categorized as susceptible with the Wider system and as resistant by the reference method, the classification of a “very major error” was made. The endpoint “major error” was used when the MIC obtained with the Wider system was categorized as resistant and that obtained by the reference method was categorized as susceptible. The term “minor error” was used when the MIC obtained with the Wider system was categorized as intermediate and that obtained by the reference method was categorized as susceptible or resistant and when the MIC obtained with the Wider system was categorized as susceptible or resistant and that obtained by the reference method was categorized as intermediate. A similar susceptibility testing analysis was performed with isolates of the family Enterobacteriaceae with known resistance mechanisms, but only the results obtained with amoxicillin-clavulanate, ticarcillin, cefuroxime, cefoxitin, cefotaxime, ceftazidime, cefepime, and imipenem were compared.

RESULTS

Organism identification.

Table 2 shows the performance of the Wider system for bacterial identification. A total of 244 isolates routinely obtained from different clinical sources were tested. The Wider system correctly identified 97.5% of these bacterial clinical isolates: 99.3% of the isolates of the family Enterobacteriaceae, 92.0% of the NFGNRs, and 96.3% of the gram-positive organisms. Misidentifications at the genus level (incorrect identification) were limited to only two NFGNRs. In addition, one enterobacterial isolate and three staphylococcal isolates were partially identified (misidentification to the species level). It must be stressed that a Salmonella sp. isolate and an S. aureus isolate were incorrectly identified as S. enterica subsp. arizonae and a coagulase-negative staphylococcus, respectively. Bacterial strains for identification quality control were always correctly identified, including S. enterica subsp. arizonae ATCC 12323 and S. aureus ATCC 29213.

TABLE 2.

Identification of routine isolates by the Wider system

Organisms (no. of isolates tested) No. (%) of isolates:
Correctly identified Partially identified Incorrectly identified
Citrobacter freundii (1) 1
Enterobacter aerogenes (1) 1
Enterobacter asburriae (1) 1
Enterobacter cloacae (6) 6
Escherichia coli (89) 89
Klebsiella oxytoca (2) 2
Klebsiella pneumoniae (8) 8
Kluyvera spp. (1) 1
Morganella morganii (3) 3
Pantoea aglomerans (4) 4
Proteus mirabilis (14) 14
Proteus vulgaris (1) 1
Providencia stuartii (1) 1
Salmonella spp. (4) 3 1a
Serratia marcescens (1) 1
Yersinia enterocolitica (1) 1
Acinetobacter baumannii (2) 2
Acinetobacter junii (1) 1b
Acinetobacter lwofii (1) 1
Flavobacterium odoratum (1) 1
Pseudomonas aeruginosa (13) 13
Pseudomonas putida (1) 1c
Stenotrophomonas maltophilia (6) 6
Enterococcus casseliflavus (1) 1
Enterococcus durans (2) 2
Enterococcus faecalis (15) 15
Enterococcus gallinarum (1) 1
Enterococcus faecium (3) 3
Micrococcus spp. (2) 2
Staphylococcus aureus (19) 18 1d
Staphylococcus epidermidis (24) 23 1e
Staphylococcus lugdunensis (1) 1
Staphylococcus simulans (2) 1 1f
Staphylococcus warnerii (1) 1
Streptococcus agalactiae (5) 5
Streptococcus pyogenes (1) 1
Total (244) 238 (97.5) 4 (1.6) 2 (0.8)
a

Identified with the Wider system as Salmonella enterica subsp. arizonae

b

Identified with the Wider system as Empedobacter brevis

c

Identified with the Wider system as Acinetobacter anitratus-A. haemolyticus

d

Identified with the Wider system as Staphylococcus simulans

e

Identified with the Wider system as Staphylococcus auricularis

f

Identified with the Wider system as Staphylococcus hominis

It is noteworthy that five isolates (two coagulase-negative staphylococci and one isolate each of Acinetobacter baumannii, Stenotrophomonas maltophilia, and Enterococcus durans) were misidentified with the reference API system. In these cases, conventional biochemical test results for arbitration confirmed the Wider system identification result.

Susceptibility testing of routine isolates.

A total of 3,719 organism-antimicrobial agent combinations were analyzed: 2,484 for isolates of the family Enterobacteriaceae, 300 for NFGNRs, and 893 for gram-positive isolates. The overall essential agreement in MICs (±1 log2 dilution) for all these organism-antimicrobial agent combinations was 95.6%; 1.6% of the results with the Wider system were 2 or more dilutions lower than the reference MICs, and 2.8% of the results were 2 or more dilutions higher than the reference MICs.

Results for the Enterobacteriaceae, NFGNRs, and gram-positive isolates and the antimicrobial agents evaluated for each bacterial group are indicated in Tables 3, 4, and 5, respectively. The lowest essential agreement was observed with the NFGNRs, 88.0%, a value significantly lower than those obtained for isolates of the family Enterobacteriaceae (96.6%) and gram-positive cocci (95.6%). Imipenem was the antimicrobial agent tested with the enterobacterial isolates (84.0%) with the lowest essential agreement in MICs. For this group of bacterial isolates, imipenem MICs were clearly displaced to higher values when they were determined with the Wider system. In contrast, 100% essential agreement in MICs was observed with meropenem. Moreover, ciprofloxacin MICs were also displaced to higher values. Among NFGNRs, piperacillin and cefepime yielded the lowest essential agreement (80.0%). None of the antimicrobial agents analyzed yielded 100% essential agreement for NFGNRs (Table 4). For gram-positive cocci, the essential agreement obtained with penicillin was only 80.4%, but it was greater than 90.0% with the other antimicrobial agents tested (Table 5). Most of the discrepancies in penicillin MICs for gram-positive cocci were due to staphylococcal isolates, but no interpretive errors were detected. On the contrary, 100% agreement between the results obtained with the Wider system and those obtained by the standard microdilution method was observed with the enterococcal isolates when results for ampicillin resistance, high-level gentamicin resistance, and high-level streptomycin resistance were compared.

TABLE 3.

Performance of Wider system susceptibility testing with routine Enterobacteriaceae isolatesa

Antimicrobial agentb % MICs obtained with Wider system within indicated log2 reference MIC:
% Essential Agreement % Errors
>−2 −2 −1 0 +1 +2 >+2 mi M VM
Ampicillin (138) 0.0 0.0 2.1 97.8 0.0 0.0 0.0 100.0 0.7 0.0 0.0
Amoxicillin-clavulanate (138) 0.0 0.7 7.9 85.5 3.6 1.4 0.7 97.1 5.1 1.0 0.0
Ticarcillin (138) 1.4 0.7 3.6 93.5 0.7 0.0 0.0 97.8 3.6 0.0 0.0
Cefazolin (138) 1.4 0.7 0.7 93.5 2.2 1.4 0.0 96.4 2.9 0.0 6.6
Cefuroxime (138) 0.0 1.4 2.8 92.0 2.1 0.7 0.7 97.1 3.6 0.0 0.0
Cefoxitin (138) 0.0 0.0 0.7 94.2 4.3 0.7 0.0 99.3 2.2 0.0 0.0
Cefotaxime (138) 0.0 1.4 0.0 92.0 5.7 0.7 0.0 97.8 NA NA NA
Ceftazidime (138) 0.0 0.0 0.0 93.5 5.1 1.4 0.0 98.5 0.7 0.0 0.0
Aztreonam (138) 1.4 0.7 2.2 85.5 6.5 2.9 0.7 94.2 0.7 0.7 33.3
Cefepime (138) 0.0 2.1 2.9 90.6 2.9 0.7 0.7 96.4 0.7 0.0 0.0
Meropenem (138) 0.0 0.0 0.0 98.5 1.4 0.0 0.0 100.0 0.0 0.0 0.0
Imipenem (138) 0.0 0.7 2.1 71.0 11.0 5.8 9.4 84.0 1.4 0.7 0.0
Gentamicin (138) 0.0 0.0 1.4 95.0 0.7 2.9 0.0 97.1 0.0 0.0 0.0
Tobramycin (138) 0.0 0.7 0.7 94.2 2.8 1.4 0.0 97.8 4.3 0.0 0.0
Amikacin (138) 0.7 0.0 2.8 92.0 2.8 1.4 0.0 97.8 NA NA NA
Nalidixic acid (138) 0.0 0.0 0.0 97.1 0.7 2.1 0.0 97.8 0.0 0.0 0.0
Ciprofloxacin (138) 0.0 0.0 2.2 85.5 4.3 0.7 7.2 92.0 0.7 0.9 0.0
Co-trimoxazole (138) 0.0 0.0 0.0 94.9 2.1 2.9 0.0 97.1 0.0 4.8 0.0
 Total (2,484) 0.3 0.5 1.8 91.8 3.2 1.5 1.1 96.6 1.7 0.5 0.9
a

Abbreviations: NA, not applicable; mi, minor error; M, major error; VM, very major error. 

b

Values in parentheses are the number of organism-antimicrobial agent combinations tested. 

TABLE 4.

Performance on Wider system susceptibility testing with routine NFGNRsa

Antimicrobial agentb % MICs obtained with Wider system within indicated log2 reference MICs:
% Essential agreement % Errors
>−2 −2 −1 0 +1 +2 >+2 mi M VM
Piperacillin (25) 8.0 4.0 8.0 72.0 0.0 8.0 0.0 80.0 4.0 0.0 50.0
Piperacillin-tazobactam (25) 4.0 0.0 4.0 72.0 12.0 4.0 4.0 88.0 4.0 7.7 4.0
Ceftazidime (25) 0.0 8.0 4.0 72.0 12.0 4.0 0.0 88.0 4.0 5.2 0.0
Aztreonam (25) 8.0 0.0 20.0 68.0 4.0 0.0 0.0 92.0 4.0 0.0 12.5
Cefepime (25) 0.0 8.0 12.0 56.0 12.0 8.0 4.0 80.0 4.0 0.0 0.0
Imipenem (25) 0.0 0.0 0.0 76.0 12.0 4.0 8.0 88.0 4.0 0.0 0.0
Meropenem (25) 0.0 4.0 12.0 76.0 4.0 4.0 0.0 92.0 0.0 0.0 0.0
Gentamicin (25) 0.0 0.0 0.0 80.0 8.0 8.0 4.0 88.0 8.0 0.0 0.0
Tobramycin (25) 0.0 0.0 0.0 88.0 4.0 8.0 0.0 92.0 0.0 5.0 0.0
Amikacin (25) 0.0 0.0 0.0 80.0 12.0 8.0 0.0 92.0 NA NA NA
Ciprofloxacin (25) 0.0 4.0 4.0 84.0 0.0 8.0 0.0 88.0 0.0 0.0 16.6
Co-trimoxazole (25) 0.0 8.0 12.0 64.0 12.0 4.0 0.0 88.0 0.0 4.0 8.0
 Total (300) 1.7 3.0 6.3 73.6 8.0 5.7 1.7 88.0 2.9 2.2 12.1
a

Abbreviations: NA, not applicable; mi, minor error; M, major error; VM, very major error. 

b

Values in parentheses are the number of organism-antimicrobial agent combinations tested. 

TABLE 5.

Performance of Wider system susceptibility testing with gram-positive isolatesa

Antimicrobial agentb % MICs obtained with Wider system within indicated log2 reference MICs:
% Essential agreement % Errors
>−2 −2 −1 0 +1 +2 >+2 mi M VM
Penicillin (81)bcde 1.2 7.4 8.0 62.0 10.3 3.4 1.2 80.4 0.0 0.0 0.0
Ampicillin (22)d 0.0 0.0 9.1 81.8 9.1 0.0 0.0 100.0 0.0 0.0 0.0
Oxacillin (53)c 0.0 1.9 3.7 88.6 3.7 1.9 0.0 96.2 0.0 3.4 0.0
Cefazolin (53)c 0.0 5.6 13.2 79.2 0.0 1.8 0.0 92.4 NA NA NA
Cefotaxime (59)ce 0.0 3.4 55.9 33.9 5.1 1.7 0.0 94.9 NA NA NA
Vancomycin (81)cde 0.0 0.0 2.5 70.4 27.2 0.0 0.0 100.0 0.0 0.0 0.0
Teicoplanin (81)cde 0.0 0.0 6.2 81.5 7.4 4.9 0.0 95.1 1.3 0.0 0.0
Gentamicin (53)c 7.5 0.0 0.0 90.5 1.8 0.0 0.0 92.4 0.0 0.0 9.1
Amikacin (53)c 0.0 9.4 1.8 88.3 5.6 0.0 0.0 90.5 NA NA NA
Gentamicin HLAR (22)d 0.0 0.0 0.0 100.0 0.0 0.0 0.0 100.0 0.0 0.0 0.0
Streptomycin HLAR (22)d 0.0 0.0 0.0 100.0 0.0 0.0 0.0 100.0 0.0 0.0 0.0
Erythromycin (81)cde 2.5 0.0 2.5 95.1 0.0 0.0 0.0 97.5 NA NA NA
Clindamycin (59)ce 0.0 1.7 1.7 96.6 0.0 0.0 0.0 98.3 1.7 0.0 0.0
Ciprofloxacin (81)cde 0.0 0.0 2.5 87.6 7.4 2.5 0.0 97.5 1.2 2.2 0.0
SXT (59)ce 0.0 0.0 1.7 96.6 1.7 0.0 0.0 100.0 0.0 0.0 0.0
Tetracycline (75)cd 0.0 0.0 0.0 73.6 1.3 5.3 0.0 94.6 0.0 2.4 0.0
 Total (935) 0.7 1.9 6.9 82.5 6.0 1.7 0.1 95.5 0.4 1.6 0.6
a

Abbreviations: NA, not applicable; SXT, trimethoprim-sulfamethoxazole; mi, minor error; M, major error; VM, very major error; HLRA, high-level aminoglycoside resistance. 

b

Values in parentheses are the number of organism-antimicrobial agent combinations tested. 

c

Staphylococcus and Micrococcus

d

Enterococcus. 

e

β-Hemolytic streptococci. 

The overall agreement of the interpretive categories obtained with the Wider system compared with those obtained by the standard microdilution method when the NCCLS criteria were used was 95.8%, ranging from 82.8% for NFGNRs to 97.4% for gram-positive cocci. Results for cefotaxime and amikacin with the Enterobacteriaceae and gram-positive cocci and for cefazolin and erythromycin with gram-positive cocci were suppressed in the interpretive category analysis, as the antimicrobial concentrations presented in the Wider system panels do not allow the use of NCCLS interpretive criteria (Tables 3, 4, and 5). Equal to or less than 2.0% of the errors were major (16 of 2,478) or very major (12 of 582) errors, and these were mainly due to discrepancies in interpretive categories with piperacillin and co-trimoxazole for NFGNRs. Only 48 of 3,172 (1.5%) organism-antimicrobial agent combinations tested were classified as having minor discrepancies. Again, the most important percentage of minor errors were for NFGNRs. Despite the smaller number of NFGNRs tested, minor discrepancies appeared to be randomly distributed among these isolates.

The ATCC strains recommended by NCCLS were used in the quality control procedure, and these were also used to study the reproducibilities of the MIC readings. At least 25 runs were performed on different days with each ATCC strain. MICs were highly reproducible, although the MICs for certain ATCC organism-antimicrobial agent combinations were equal to or below the lowest concentration of the antimicrobial agent in the wells. With the exception of the combination penicillin-S. aureus ATCC 29213, MICs were within the expected ranges. The penicillin MICs for S. aureus ATCC 29213 exceeded the expected range in 44% of the runs. It is noteworthy that for penicillin and E. faecalis ATCC 25922, for imipenem and E. coli ATCC 25922, and for imipenem and P. aeruginosa ATCC 27853, MICs were persistently near the upper limit of the MIC range.

Susceptibility testing of Enterobacteriaceae with well-characterized β-lactam resistance mechanisms.

A total of 800 organism-antimicrobial agent combinations were analyzed. The essential agreement of susceptibility testing of isolates of the family Enterobacteriaceae with known β-lactam resistance mechanisms with the Wider system and by the standard microdilution method was 94.8% (Table 6), which is slightly lower than that obtained with routine isolates of the family Enterobacteriaceae (97.4%). The highest essential agreement was observed with chromosomal AmpC β-lactamase-producing isolates (97.7%) and permeability mutants (98.4%). On the contrary, the lowest essential agreement was obtained with chromosomal class-A β-lactamase-producing isolates (89.1%). The analysis of the different antimicrobial agents tested revealed that more than 97.0% essential agreement was observed for ticarcillin, cefuroxime, cefoxitin, cefotaxime, and ceftazidime. The corresponding values for amoxicillin-clavulanate, imipenem, and cefepime were 94.0, 90.0, and 83.0%, respectively.

TABLE 6.

Results of comparison between Wider system susceptibility testing results and reference broth microdilution MIC for isolates of the family Enterobacteriaceae with known β-lactam resistance mechanismsa

Primary resistance mechanismb % Essential agreement % Errors
mi M VM
Chromosomal AmpC β lactamase (176) 97.7 1.0 0.0 0.0
Chromosomal class A β-lactamase (64) 89.1 7.1 2.7 6.2
Broad-spectrum and IRT β-lactamases (168) 97.6 4.1 0.0 0.0
Extended-spectrum β-lactamase (328) 92.3 3.5 2.0 0.8
Permeability mutant (64) 98.4 3.7 0.0 0.0
 Total (800) 94.8 3.6 1.0 0.8
a

Abbreviations: mi, minor error; M, major error; VM, very major error. 

b

Values in parentheses are the number of organism-antimicrobial agent combinations tested. 

Interpretive category analysis showed that only two very major errors were observed (0.8%) among isolates of the family Enterobacteriaceae with known β-lactam resistance mechanisms. These were detected with amoxicillin-clavulanate and an E. coli strain that produced an extended-spectrum-β-lactamase and with ticarcillin and a Proteus vulgaris strain that produced a chromosomal class A β-lactamase. The minor errors represented 3.6% of the susceptibility test determinations. Forty percent of them were for extended-spectrum β-lactamase-producing isolates and were particularly observed for amoxicillin-clavulanate (5 of 41) and cefepime (15 of 41) determinations.

DISCUSSION

Simultaneous identification and susceptibility testing is clearly an advantage of automatic and semiautomatic commercial systems (4, 7). The Wider system is a newly developed semiautomatic device for bacterial identification and susceptibility testing which needs the classical overnight period for retrieval of results. The major attraction of this system is the image-processing technology that has been applied for the first time to the reading of commercial trays for bacterial identification and susceptibility testing. Biochemical patterns and MICs are processed by the computer and are offered to the operator both visually and, if required, on paper. Indeed, a particular advantage of image-assisted analysis is that the system greatly facilitates reading of the results, but the results for the panels can still be interpreted directly by the microbiologist by applying classical criteria. This eliminates dependence on a fully automatic device; moreover, the technologist can adjust the video-assisted readings prior to the release of a patient's results. This approach has previously been applied for disk diffusion antibiograms (3, 10) and also for an unsuccessful API Aladin instrument, marketed in the late 1980s, which used microdilution tray wells for bacterial identification and susceptibility testing (12).

In the present study, the Wider system correctly identified at the species level 95.7% of routine clinical isolates, represented by 20 genera commonly isolated in a clinical microbiology laboratory. This percentage is similar to or slightly higher than those observed in other studies with other systems (25, 31) and demonstrates that with currently accepted criteria the Wider system is an acceptable method for bacterial identification (15). For the isolates of the family Enterobacteriaceae, the only identification problem was an atypical lactose-positive S. enterica serovar Enteritidis isolate which was misidentified by the Wider system as S. enterica subsp. arizonae. This particular problem has also been delineated with other systems (18); however, the Wider system correctly identified the S. enterica subsp. arizonae ATCC 12323 strain. As with other systems, less accuracy (92.0%) than that observed with isolates of the family Enterobacteriaceae was shown when only NFGNRs were taken into account (8, 22, 24, 25, 27). In all cases, the API system was used as the “gold standard” for bacterial identification, as it is a well-recognized system for this purpose (25, 31). Nevertheless, conventional biochemical tests were also performed when discrepancies between the Wider system and the API galleries were observed before assuming that the API identification was correct.

The Wider system database contains information on 112 and 42 different taxa for gram-negative and gram-positive bacteria, respectively, which includes both organisms usually encountered in clinical laboratories and those rarely isolated. It is remarkable that, with the exception of the oxidase test for gram-negative organisms and catalase and hemolysis tests for gram-positive organisms, the Wider system does not require additional external reactions to provide the final identification or to improve the identification level. Other identification systems need external biochemical tests other than those provided in the microdilution panels to resolve identifications with low probabilities of accuracy (8, 21). As with other devices, if a biochemical profile does not match one in the database, there may be no identification, but this did not occur with the routine isolates tested with the Wider system. As the purpose of our study was to determine the accuracy of the Wider system for the identification of bacteria that would routinely be encountered in a hospital microbiology laboratory, additional studies with other relevant and uncommon organisms should be performed.

The susceptibility testing results obtained with the Wider system in tests with routine clinical isolates showed an overall essential agreement of 95.6% and an overall category interpretation error of 4.2%. Combined major and very major errors were less than 3%. These results meet the performance criteria for susceptibility testing (7, 9, 18). It is of note that these results were compiled by considering the results for both gram-negative and -positive isolates and a large number of antimicrobial agents for each isolate. When analyzed with respect to the organisms tested, the highest essential agreements were observed with isolates of the family Enterobacteriaceae (96.6%) and gram-positive organisms (95.6%). As other investigators have noted with other systems (22, 23), lower essential agreement was observed with NFGNRs (88.0%). Nevertheless, in contrast to automated instrument systems with short incubation times (14, 15), susceptibility testing results could be obtained for all organisms considered to be slowly growing organisms.

When analyzed with respect to the antimicrobial agents tested, the lowest essential agreement for isolates of the family Enterobacteriaceae was observed with imipenem (84.0%). It is remarkable that 26.2% of the imipenem MICs obtained with the Wider system were higher than those obtained by the reference method, but the percentage of interpretive errors was limited to only 1.4 and 0.7% minor and major errors, respectively, with no very major errors. This difference is clearly related to the intrinsic activity of imipenem (modal MIC, 0.25 μg/ml for isolates of the family Enterobacteriaceae) and the relatively high MIC breakpoint for susceptibility (4 μg/ml). For NFGNRs, the essential agreement for imipenem was 88.0%. Again, nearly 25% of the imipenem MICs obtained with the Wider system were higher than those obtained by the reference method. These higher imipenem MICs could be due to insufficient adjustment of inoculum size or to a decline in antimicrobial activity during storage. The former possibility was well corroborated by Doern et al. (6), who demonstrated that an inappropriately large inoculum size in automatic susceptibility testing devices significantly modified the results, particularly for cell wall-active antibiotics. False positive results for resistance may occur when the inoculum size is too large, and the numbers of major and minor errors are increased. In our study, the inoculum for the Wider system was prepared with the Prompt inoculation system (28), as recommended by the manufacturer. In contrast, the inoculum for the reference microdilution method was controlled nephelometrically. On the other hand, it has been shown that the activity of imipenem is particularly affected during storage (5, 26) and is a well-recognized cause of false-positive resistance with commercial microdilution panels (23, 28). It is not a surprise that the imipenem MICs for E. coli ATCC 25922 and P. aeruginosa ATCC 27853 were persistently near the upper limit of the MIC range recommended by NCCLS (20). The same problem of stability during storage could be responsible for the higher penicillin MICs for S. aureus ATCC 29213 and routine gram-positive isolates.

In previous evaluations of susceptibility testing devices, specific attention has been given to resistant isolates (31). In our evaluation, although a limited number of resistant isolates were included among the clinical isolates (data not shown), no problems with specific issues of resistance specifically studied with other automatic susceptibility testing devices were detected, such as oxacillin resistance in staphylococci (13) and high-level aminoglycoside resistance in enterococci (30). Moreover, when assessing susceptibility testing instrument performance, in addition to fresh or stock clinical isolates and quality control strains, a set of organisms with known resistance mechanisms should be included (7, 9). One hundred isolates of the family Enterobacteriaceae with known β-lactam resistance mechanisms were studied, and 800 organism-antimicrobial agent combinations were evaluated. Essential agreement and interpretive errors for eight β-lactam antibiotics were 94.8 and 5.4%, respectively. Interestingly, the proportion of very major errors with this set of strains was limited to 0.8%. The lowest essential agreement was observed with extended-spectrum and chromosomal class A β-lactamase-producing isolates. Slight variations in the inoculum size could affect the amount of these β-lactamases and would be the reason for the lower essential agreement. The case of cefepime is illustrative, as this antibiotic has been shown to be very stable during storage (26) but is particularly affected by an increase in inoculum size for those isolates with extended-spectrum β-lactamases (R. Cantón and A. Oliver, unpublished data). Interestingly, no decreases in the MICs with the Wider system were observed with ceftazidime and cefotaxime for extended-spectrum β-lactamase producing isolates of the family Enterobacteriaceae, thus avoiding false-positive detection of isolates with these enzymes. The design of panels is essential for retrieval of results that can serve as indicators of the presence of extended-spectrum β-lactamase-producing isolates (16). The wide range of concentrations for ceftazidime (0.5 to 16 μg/ml) and cefotaxime (0.12 to 8 μg/ml) in the Wider system panels facilitates the detection of these isolates. In addition, the Wider system panels for gram-negative isolates possess the combination of ceftazidime plus clavulanate to facilitate the detection of these β-lactamase-producing isolates.

In conclusion, our evaluation showed accurate and acceptable results with both routine clinical isolates and a set of isolates of the family Enterobacteriaceae with known β-lactam resistance mechanisms. The Wider system is a new reliable tool which applies the image-processing technology for the reading of commercial trays for bacterial identification and susceptibility testing.

ACKNOWLEDGMENTS

We are grateful to Isabel Soler for continuous technical assistance.

REFERENCES

  • 1.Baker C N, Stocker S A, Rhoden D L, Thornsberry C. Evaluation of the MicroScan antimicrobial susceptibility system with the AutoScan-4 automated reader. J Clin Microbiol. 1986;19:744–747. doi: 10.1128/jcm.23.1.143-148.1986. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Baquero F, Martínez-Beltrán J, Cantón R y los Restantes Miembros de la Mesa Española de Normalización de la Sensibilidad y Resistencia a los Antimicrobianos. Criterios del Grupo MENSURA para la definición de los puntos críticos de sensibilidad a los antibióticos. Rev Esp Quimioter. 1997;10:303–313. [PubMed] [Google Scholar]
  • 3.Berke I, Tierno P M., Jr Comparison of efficacy and cost-effectiveness of BIOMIC VIDEO and Vitek antimicrobial susceptibility test systems for use in the clinical microbiology laboratory. J Clin Microbiol. 1996;34:1980–1984. doi: 10.1128/jcm.34.8.1980-1984.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Cherubin C, Eng R, Appleman M A. A critique of semiautomated susceptibility systems. Rev Infect Dis. 1987;9:655–659. doi: 10.1093/clinids/9.3.655. [DOI] [PubMed] [Google Scholar]
  • 5.Daly J S, DeLuca D B, Hebert S R, Dodge R A, Soja D T. Imipenem stability in a predried susceptibility panel. J Clin Microbiol. 1994;32:2584–2587. doi: 10.1128/jcm.32.10.2584-2587.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Doern G V, Brueggemann A B, Perla R, Daly J, Halkias D, Jones R N, Saubolle M A. Multicenter laboratory evaluation of the bioMerieux Vitek antimicrobial susceptibility testing system with 11 antimicrobial agents versus members of the family Enterobacteriaceae and Pseudomonas aeruginosa. J Clin Microbiol. 1997;35:2115–2119. doi: 10.1128/jcm.35.8.2115-2119.1997. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ferraro M J, Jorgensen J H. Susceptibility testing instrumentation and computerized expert systems for data analysis and interpretation. In: Murray P R, Baron E J, Pfaller M A, Tenover F C, Yolken R H, editors. Manual of clinical microbiology. 7th ed. Washington, D.C.: American Society for Microbiology; 1999. pp. 1593–1600. [Google Scholar]
  • 8.Flunke G, Monnet D, DeBernardis C, von Graevenitz A, Freney J. Evaluation of the Vitek 2 system for rapid identification of medically relevant gram-negative rods. J Clin Microbiol. 1998;36:1948–1952. doi: 10.1128/jcm.36.7.1948-1952.1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Food and Drug Administration. Federal guidelines. Review criteria for assessment of antimicrobial susceptibility testing device. Rockville, Md: Food and Drug Administration; 1991. [Google Scholar]
  • 10.Hejblum G, Jarlier V, Grosset J, Aurengo A. Automated interpretation of disk diffusion antibiotic susceptibility test with the radial profile analysis algorithm. J Clin Microbiol. 1993;31:2396–2401. doi: 10.1128/jcm.31.9.2396-2401.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Isenberg H D, Reichler A, Wiseman D. Prototype of a fully automated device for determination of bacterial antibiotic susceptibility testing in the clinical laboratory. Appl Microbiol. 1971;22:980–986. doi: 10.1128/am.22.6.980-986.1971. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Jorgensen J H. Antibacterial susceptibility tests: automated or instrument-based methods. In: Balows A, Hausler W J Jr, Herrmann K L, Isenberg H D, Shadomy H J, editors. Manual of clinical microbiology. 5th ed. Washington, D.C.: American Society for Microbiology; 1991. pp. 1166–1172. [Google Scholar]
  • 13.LaTemple D, Cruz C. Evaluation of MicroScan rapid gram-positive panels for detection of oxacillin-resistant staphylococci. J Clin Microbiol. 1994;32:1058–1059. doi: 10.1128/jcm.32.4.1058-1059.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.McGregor A, Schio F, Beaton S, Boulton V, Perman M, Gilbert G. The MicroScan WalkAway diagnostic microbiology system. An evaluation. Pathology. 1995;27:172–176. doi: 10.1080/00313029500169822. [DOI] [PubMed] [Google Scholar]
  • 15.Miller J M, O'Hara C M. Manual and automated systems for microbial identification. In: Murray P R, Baron E J, Pfaller M A, Tenover F C, Yolken R H, editors. Manual of clinical microbiology. 7th ed. Washington, D.C.: American Society for Microbiology; 1999. pp. 193–201. [Google Scholar]
  • 16.Moland E S, Sanders C C, Thomson K S. Can results obtained with commercially available MicroScan microdilution panels serve as an indicator of β-lactamase production among Escherichia coli and Klebsiella isolates with hidden resistance to expanded-spectrum cephalosporins and aztreonam? J Clin Microbiol. 1998;36:2575–2579. doi: 10.1128/jcm.36.9.2575-2579.1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Murray P R, Baron E J, Pfaller M A, Tenover F C, Yolken R H, editors. Manual of clinical microbiology. 7th ed. Washington, D.C.: American Society for Microbiology; 1999. [Google Scholar]
  • 18.National Committee for Clinical Laboratory Standards. Development of in vitro susceptibility testing criteria and quality control parameters: approved guideline. Document M23-A. Wayne, Pa: National Committee for Clinical Laboratory Standards; 1994. [Google Scholar]
  • 19.National Committee for Clinical Laboratory Standards. Method for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. 4th ed. 1997. Approved standard. Document M7-A4. National Committee for Clinical Laboratory Standards, Wayne, Pa. [Google Scholar]
  • 20.National Committee for Clinical Laboratory Standards. Performance standards for antimicrobial susceptibility testing. Ninth informational supplement. Document M100-S9. Wayne, Pa: National Committee for Clinical Laboratory Standards; 1999. [Google Scholar]
  • 21.O'Hara C M, Westbrook F C, Miller J M. Parallel comparison of accuracy of API 20E, Vitek GNI, MicroScan Walk/Away Rapid ID, and Becton Dickinson Cobas Micro ID-E/NF for identification of members of the family Enterobacteriaceae and common gram-negative, non-glucose-fermenting bacilli. J Clin Microbiol. 1993;31:3165–3169. doi: 10.1128/jcm.31.12.3165-3169.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.O'Hara C M, Westbrook G L, Miller J M. Evaluation of Vitek GNI+ and Becton Dickinson Microbiology Systems Crystal E/NF identification systems for identification of members of the family Enterobacteriaceae and other gram-negative, glucose-fermenting and non-glucose-fermenting bacilli. J Clin Microbiol. 1997;35:3269–3273. doi: 10.1128/jcm.35.12.3269-3273.1997. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.O'Rourke E J, Lambert K G, Parsonnet K C, Macone A B, Goldmann D A. False resistance to imipenem with a microdilution susceptibility testing system. J Clin Microbiol. 1991;29:827–829. doi: 10.1128/jcm.29.4.827-829.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Simoons-Smit A M, MacLaren D M. Comparison of Vitek and Cobas Micro systems with semiautomated conventional microsystem for identification and susceptibility testing of gram-negative bacilli. J Clin Pathol. 1994;47:71–75. doi: 10.1136/jcp.47.1.71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Stager C E, Davis J R. Automated systems for identification of microorganisms. Clin Microbiol Rev. 1992;5:302–327. doi: 10.1128/cmr.5.3.302. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Valdezate S, Martínez-Beltrán J M, de Rafael L, Baquero F, Cantón R. β-Lactam stability in frozen microdilution PASCO MIC panels using strains with known resistance mechanisms as biosensors. Diagn Microbiol Infect Dis. 1996;26:53–61. doi: 10.1016/s0732-8893(96)00178-2. [DOI] [PubMed] [Google Scholar]
  • 27.Visser M R, Bogaards L, Rozenberg-Arska M, Verhoef J. Comparison of the autoSCAN-W/A and Vitek Automicrobic systems for identification and susceptibility testing of bacteria. Eur J Clin Microbiol Infect Dis. 1992;11:979–984. doi: 10.1007/BF01967786. [DOI] [PubMed] [Google Scholar]
  • 28.White R L, Kays M B, Friedrich L V, Brown E W, Koonce J R. Pseudoresistance of Pseudomonas aeruginosa resulting from degradation of imipenem in an automated susceptibility testing system with predried panels. J Clin Microbiol. 1991;29:398–400. doi: 10.1128/jcm.29.2.398-400.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Wicks J H, Nelson R L, Krejcarek G E. Rapid inoculum standardization system: a novel device for standardization of inocula in antimicrobial susceptibility testing. J Clin Microbiol. 1983;17:1114–1119. doi: 10.1128/jcm.17.6.1114-1119.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Woods G L, DiGiovanni B, Levison M, Pitsakis P, LaTemple D. Evaluation of MicroScan rapid panels for detection of high-level aminoglycoside resistance in enterococci. J Clin Microbiol. 1993;31:2786–2787. doi: 10.1128/jcm.31.10.2786-2787.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.York M K, Brooks G F, Fiss E H. Evaluation of the autoSCAN-W/A Rapid System for identification and susceptibility testing for gram-negative fermentative bacilli. J Clin Microbiol. 1992;30:2903–2910. doi: 10.1128/jcm.30.11.2903-2910.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES