High efficiency laser resonance ionization of plutonium

Alfredo Galindo-Uribarri 1,2,, Yuan Liu 1,3,, Elisa Romero Romero 1,2,4,, Daniel W Stracener 1,
PMCID: PMC8648762  PMID: 34873192

Abstract

Three-step resonance photoionization spectra of plutonium have been studied with Ti:Sapphire lasers for the development of efficient laser ionization schemes for ultra-trace analysis of Pu isotopes by resonance ionization mass spectrometry. We observed eighteen intermediate excited states of even parity in the energy range 35568–36701 cm-1, thirteen of them have not been previously documented, and a larger number of high-lying excited states and autoionizing states of odd-parity between 48238 and 49510 cm-1. Three-color, three-photon ionization schemes via six intermediate states were evaluated under similar ion source operating conditions. This led to a highly efficient three-step scheme with an overall ionization efficiency of 51.1±1.3%, which is an order of magnitude improvement over the previously reported ionization efficiency for Pu.

Subject terms: Planetary science, Optics and photonics, Physics

Introduction

Plutonium (Pu) is one of the most important actinide members for its use in nuclear weapons and nuclear energy and its ubiquitous presence in the Earth’s surface environment as a result of anthropogenic activities, such as nuclear weapons tests, accidents at nuclear power plants, release from nuclear fuel reprocessing facilities, poor disposal of nuclear wastes, and loss of nuclear weapons13. Although the amount of Pu in the environment is small, it is considered a highly hazardous pollutant due to its high radiotoxicity and potential risk to human health. Pu could also be naturally produced, like all the heavy elements making up the Earth, by rapid neutron-capture process (r-process) in stellar nucleosynthesis4. The observation of ultra-trace 244Pu, the longest-lived Pu isotope with a half-life of about 80 million years, in deep-sea crust and sediment is indicative of possible stellar nucleosynthesis sites such as neutron star mergers5. Hence, detection of Pu concentration and the abundance ratios of its various isotopes, in particular 238-242,244Pu, in environmental samples and nuclear materials has vitally important applications in nuclear forensics and safeguards6, environmental science3, radiation protection at nuclear facilities7, and cosmochemistry and astrophysics5,8. However, the Pu amount in the samples is usually at very low trace levels. The most sensitive analytical techniques are required for these applications.

Resonance ionization mass spectrometry (RIMS) is one of the most sensitive techniques for ultra-trace analysis of long-lived radionuclides such as the isotopes of actinides914. In RIMS, ions are formed by stepwise resonant absorption of two or three photons through allowed atomic levels to photoionization. The multistep resonance process is extremely selective for the specific element of interest. Thus, in comparison with the commonly used mass spectrometry techniques15, such as accelerator mass spectrometry (AMS), inductively coupled plasma mass spectrometry (ICP-MS), and thermal ionization mass spectrometry (TIMS), RIMS is superior in suppressing the isobaric interferences in the subsequent mass separation and promises high detection efficiencies10,12,16 and thus high sensitivities10,17,18. In this paper, we present the studies of efficient resonance ionization schemes for ultra-trace analysis of Pu isotopes by RIMS.

Trace detection of Pu isotopes by RIMS was first demonstrated by Donohue at al.19,20 and Krönert, et al.21. In these early studies, Pu atoms were ionized by single-color, one or two-photon excitations. Three-color, three-step resonant ionization was first reported by Peuser et al.22 using three pulsed dye lasers, which showed a detection efficiency of 10-7. Since then, there have been extensive studies on RIMS as an analytical instrument for trace analysis of Pu and other radioactive elements17,2331, for fundamental studies of actinides32, and in combination with other techniques such as SIMS30,32,33, fast chemical separation29, and laser ablation34. Worden et al.35 studied two- and three-step resonance ionization pathways using dye lasers. Grüning et al.26 and Kunz et al.27 investigated two-color and three-color, three-step ionization schemes using tunable solid-state Ti:Sapphire (Ti:Sa) lasers. Due to these efforts, the RIMS detection limit has been substantially improved to 104-105 Pu atoms17 and work is underway towards the detection limit of 1000 atoms for 244Pu in interstellar dust grains18.

In spite of the prior work, an efficient ionization scheme for Pu is still lacking. To our knowledge, the best overall efficiency for Pu is 5% by in-source RIMS17 and 5% ion yield with a RIMS instrument18. Pu has the atomic ground-state electronic configuration of [Rn]5f67s2 with a partially occupied 5f-shell and open 6d-shell. Like other actinides, it has complicated spectra with high density of energy levels due to many possible electronic configurations and strong configuration interactions36. On the other hand, the large number of bound and autoionizing (AI) electronic states could offer pathways for more efficient photoionization. Recent studies have shown highly efficient resonance ionization of actinides16,37 and lanthanides38,39. Therefore, using all of this prior work as a foundation, we conducted spectroscopy surveys of intermediate, high-lying, and AI states in Pu, in order to identify suitable multi-step ionization pathways for Ti:Sa lasers, and evaluated different three-step schemes under similar experimental conditions10. As a result, a three-step scheme is demonstrated with an overall ionization efficiency of over 50%. The experimental procedure and results are described.

Experimental

The experiment was conducted with the hot-cavity resonance ionization laser ion source (RILIS) and the Injector for Radioactive Ion Species 2 (IRIS2) mass separator40 at the Oak Ridge National Laboratory (ORNL). The experimental setup is illustrated in Fig. 1 and the detailed components have been previously described10,41,42. Hence a brief description is given here.

Figure 1.

Figure 1

Schematic view of the experimental setup.

The RILIS consisted of a hot-cavity ion source and a system of three Ti:Sapphire (Ti:Sa) lasers. The ion source had a Ta tubular cavity of 3 mm inner diameter (ID), 1 mm wall, and 30 mm in length, which was connected to a closed-end Ta sample tube of 8.5 mm ID and about 100 mm in length that served as a crucible for the sample. The ion source assembly was resistively heated to high temperatures. With a heating current up to 450 A, the cavity could be heated to temperatures exceeding 2300 K while the sample tube could reach 1800-2000 K. Pu atoms were produced by evaporating a Pu sample inside the sample tube. The evaporated atoms effused into the cavity where they were irradiated by laser beams which were tuned to selectively ionize Pu atoms. The ions were extracted from the cavity, accelerated to 40 keV, and transported to a magnetic mass separator with a nominal mΔm= 1000 mass resolving power for mass number A = 100. All Pu isotopes could be ionized. The mass separator was optimized for transmission of the A = 242 ion beam. The 242Pu ion beam intensity passing through the mass separator was measured with a Faraday cup (FC). All the FCs in the experimental setup were secondary-electron suppressed with a repelling voltage of -300 V. The laser beams were delivered into the hot Ta cavity through a quartz window on the “straight-through” port in the vacuum chamber of the mass separator magnet, traveling in the opposite direction of the ion beam. The three laser beams were collimated and merged into a single laser beam inside the laser room, which was then transported to IRIS2 and focused over a distance of more than 10 meters into the 3-mm cavity of the ion source. The laser beam size entering the hot-cavity was chosen to be slightly larger than the cavity ID (3 mm) in order to fill the cavity for high ionization efficiency. The typical efficiency of laser transportation into the cavity was measured to be 70–80% after the beams were merged. Accordingly, the full width at half-maximum (FWHM) of the laser beams in the laser-atom interaction region was estimated to be 2.4–3 mm.

Ionization scheme

Three-color, three-step resonant excitation of Pu was investigated (Fig. 2). The first step was a known transition26 from the 5f67s2 7F0 ground state to the first excited state 5f67s7p 7D1 at 23765.98(2) cm-1 for 242Pu induced by the first laser (λ1 = 420.77 nm). The excited Pu atoms were subsequently excited by the second laser (λ2) to the second excited state (SES) from where they were ionized by absorbing another photon from the third laser (λ3). The three laser beams were provided by a system10,42 of three pulsed Ti:Sa lasers tunable between 720 nm and 960 nm. All the wavelengths stated in this paper are given in vacuum. The first photon at 402.77 nm was produced by second harmonic generation (SHG) of the fundamental radiation of the Ti:Sa laser. The second and third steps employed the Ti:Sa laser fundamental output. The typical laser power in this study was 200 mW for the first step and 1.5 W for the second and third steps. The three laser beams were individually collimated and then merged into a single laser beam. The merged laser beam was focused into the 3-mm cavity of the ion source with an average injection efficiency of 60-70%. The Ti:Sa lasers were individually pumped by three Q-switched Nd:YAG lasers at 532 nm at 10-kHz pulse repetition rate. The pulse lengths of the Ti:Sa lasers were on the order of 30 ns. The three laser outputs were synchronized in time to better than 5 ns via synchronizing the individual pump lasers.

Figure 2.

Figure 2

Three-step ionization scheme for 242Pu investigated. The first excited state energy 23765.98(2) cm-1 is from Ref.26 and the IP value is from the NIST database43. The laser wavelength is in vacuum.

The experiment was conducted by first scanning the wavelength of the second laser λ2 to search for available intermediate excited states E2. Then, with λ2 tuned to a specific wavelength to populate a selected E2, the third laser wavelength λ3 was scanned to obtain three-step photoionization spectra. In all cases, the first excitation step was the same. The fundamental wavelengths of the three Ti:Sa lasers were measured simultaneously with a calibrated wavelength meter (HighFinesse WS6-600) equipped with a 4-channel opto-mechanical switcher. The absolute accuracy of the wavelength meter was 0.02 cm-1 according to a 3-sigma criterion, which was taken as the systematic uncertainty in measuring the laser wavelengths.

Pu sample preparation

The Pu samples were prepared from Certified Reference Material for 242Pu (CRM 130, plutonium-242 spike assay and isotopic standard) purchased from New Brunswick Laboratory (NBL). The CRM 130 contained 4.1403 ± 0.003 μmoles of 242Pu in the form of evaporated plutonium nitrate with 99.94623 ± 0.00065 % purity and traces of other Pu isotopes, each below 0.025%. The isotopic standard is designed to prepare a master solution having a known concentration of plutonium on a weight basis. The sample and the 30 ml Teflon container were handled under strict radiologically controlled conditions in a dedicated hot laboratory. The standard was first dissolved in 1 ml of 8 M nitric acid solution at 100 C, which was then diluted to 10 ml with de-ionized water to a final concentration of 100 ppm 242Pu. Different sizes of Pu samples were prepared with different amounts of the 100 ppm 242Pu master solution. For spectroscopy studies, the sample was made with 400 μl of the 100 ppm 242Pu solution containing a total of about 1017 242Pu atoms. The sample solution was dried on a thin zirconium (Zr) metal foil of approximately 8x8 mm size and then wrapped in the foil. At elevated temperatures, the plutonium nitrate was expected to reduce to atomic Pu using Zr as the metallic reductant. In this study, laser ionized 242Pu ions were observed at sample temperatures starting around 1300 K. The Pu sample for the efficiency measurement contained 40 ± 0.4 μl of the 100 ppm 242Pu solution, dried on a 0.001 inch thick Zr foil (approximately 6x8 mm in size) and then wrapped in the foil. To verify the concentration of the 100 ppm master solution, three samples of 1 μl of the 100 ppm solution were measured by gross alpha counting using a Canberra Series 5 XLB low background counting system with an efficiency calibrated with a NIST traceable standard. The average activity measured with the three samples was 1020.15 ± 13.37 dpm, very close to the expected activity of 1046.5 ± 10.8 dpm per μl. It is worth noting that the shorter lived Pu isotopes with high specific activity account for 16% of the alpha activity even when present at trace levels.

Results

Intermediate excited states

We scanned λ2 over the range of 767–850 nm with λ1and λ3 fixed. The scan was conducted two times with λ3 = 761.95 and 763.07 nm, respectively, which were chosen arbitrarily such that the combined energy of the three photons exceeded the ionization energy of the atom. The intermediate states observed in the first scan with λ3 = 761.95 nm were confirmed in the second scan (λ3 = 763.07 nm). Eighteen resonance lines were observed, corresponding to resonant transitions from the first excited state to intermediate excited states of even-parity in the energy range 35568–36701 cm-1. The line centroid and the full width at half maximum (FWHM) of the spectral lines were obtained by fitting with Gaussian profiles. The results are presented in Table 1. Column 2 is the average line centroid of several scans, using both increasing and decreasing photon energy, for each resonance. The uncertainty of the centroid position includes the standard deviation of the measurements and the 0.02 cm-1 systematic error, added in quadrature. Column 3 is the centroid laser wavelength. Column 4 is the intermediate level energy E2 obtained by adding the energy of the first excited state, 23765.98 ± 0.02 cm-1 for 242Pu26, to the line centroid. The total angular momentum J of these intermediate states of even-parity could be 0, 1, and 2. Five of the observed intermediate states could match with the previously known states with J = 2 or 0. Columns 5 and 6 give the level energy of these known states in 242Pu from the corresponding energy for 240Pu and the isotope shifts of 240-242Pu36 and their J value, respectively. As shown, the measured energy positions agree well with the literature values. The other intermediate levels have not been previously documented. The average FWHM of these resonances was 0.25 ± 0.06 cm-1, which was larger than the Doppler broadening (approximately 1.5 GHz) and limited by the bandwidths of the lasers.

Table 1.

Measured resonance centroids and energy positions of the intermediate excited states of even-parity observed in λ2 scans.

Peak # Centroid (cm-1) λ2 (nm) E2 (cm-1) Refs.25,35
E (cm-1) J
1* 11802.75(2) 847.26 35568.73(3) 35568.70(3) 2
2* 12024.67(4) 831.62 35790.65(4)
3 12160.72(4) 822.32 35926.70(4)
4* 12211.30(4) 818.91 35977.28(4)
5 12264.24(4) 815.38 36030.22(4)
6 12371.68(4) 808.30 36137.66(4) 36137.60 2
7 12455.62(4) 802.85 36221.60(4)
8* 12547.64(4) 796.96 36313.62(4)
9 12570.12(4) 795.54 36336.10(4)
10 12577.22(2) 795.09 36343.20(3) 36343.05 0
11 12602.38(5) 793.50 36368.36(5)
12 12685.63(3) 788.29 36451.61(3)
13* 12686.36(3) 788.25 36452.34(3)
14 12736.96(3) 785.12 36502.94(3) 36502.88 2
15 12790.12(3) 781.85 36556.10(4)
16 12843.84(3) 778.58 36609.82(4) 36609.74 2
17* 12923.51(3) 773.78 36689.49(3)
18* 12934.51(3) 773.13 36700.50(3)

The last two columns are the energy and J values of the known levels for 242Pu from direct measurement26 and the 240-242Pu isotope shifts36.

Seven λ2 resonances that were clearly visible in the first scan with λ3 = 761.95 nm were chosen as SES for further investigation of three-step schemes, as marked with (*).

Two of the intermediate levels, E2 = 35568.73 ± 0.03 cm-1 (λ2 = 847.26 nm) and 36137.66 ± 0.04 cm-1 (λ2 = 808.3 nm), have been previously used as the SES in three-step resonant ionization of Pu17,27,29,30,44. We selected seven of the relatively strong λ2 resonances as SES for further investigation of three-step schemes, as marked with (*) in column 1 of Table 1.

Three-step resonant ionization

Three-step photoionization spectra were obtained by scanning the third laser wavelength λ3, while the first and second lasers were tuned to the respective transitions. The results are presented in Figs. 3 and 4 as a function of the total excitation energy obtained by adding the third photon energy to the level energy of the SES. Hundreds of resonance lines were observed from different intermediate states. The spectral line centroids are determined by fitting the resonance peaks to Gaussian or Fano profiles45. Each spectrum was measured 2–3 times by scanning λ3 with increasing and decreasing photon energies. The final level energies are given by the average values of the scan-up and scan-down measurements. Hundreds of high-lying excited states and autoionizing states of odd-parity in the energy range of 48238–49510 cm-1 were observed.

Figure 3.

Figure 3

Three-step photoionization spectra excited from four selected SES (#1–#4). Inverted red triangle are selected lines for further evaluation. In the top plot, red star is the line at λ3 = 767.53 nm.

Figure 4.

Figure 4

Three-step photoionization spectra excited from selected SES #5–#7. Inverted red triangle are selected lines for further evaluation.

As noted, the ion intensities from different SESs are significantly different. This results from the fact that the measurements were conducted at different times and under different ion source operation conditions. It may also reflect the overall ionization efficiencies of the different excitation schemes. Therefore, in order to identify the most efficient scheme, we needed to compare the schemes under similar experimental conditions and with the lasers optimized for the respective schemes. For this purpose, we selected the strongest lines in the spectra from different SESs for further evaluation. A total of eleven resonances were chosen, as marked in Figs. 3 and 4. Table 2 lists the corresponding second and third excitations as well as the total excitation energy of the three-step schemes. Nine of the selected resonances corresponded to ionization via AI states above the IP. No resonance was selected for the SES at λ2 = 773.13 nm for the following reasons. First, it yielded small ion currents with significant non-resonant contribution. As a results, the relative resonance enhancement was smaller than other SESs. Second, we also observed large fluctuations in the ion current, mostly due to the non-resonant contribution. Such fluctuations were not observed with other SESs and not desired for the intended applications. It is noted that the relative ion yields depend not only on the ionization schemes but also on other factors such as the laser setup, laser-neutral overlapping and interaction, ion transportation and detection, etc.

Table 2.

Comparison of the strong resonances from selected SESs with relative ion intensity normalized to the first scheme.

SES (cm-1) λ2 (nm) Step 3 (cm-1) λ3 (nm) Total E (cm-1) Rel. int.
35568.73(3) 847.26 13329.10(3) 750.24 48897.83(4) 1
35790.65(4) 831.62 12806.93(2) 780.83 48597.58(5) 1.2
831.62 12875.04(2) 776.70 48665.68(5) 1.1
35977.28(4) 818.91 12613.62(2) 792.79 48590.90(5) 4.5
818.91 12659.64(2) 789.91 48636.92(5) 4.0
36313.62(4) 796.96 12594.47(2) 794.00 48908.07(5) 4.2
796.96 12707.15(2) 786.96 49020.77(5) 4.1
36452.34(3) 788.25 12306.02(2) 812.61 48758.29(6) 4.1
36689.49(3) 773.78 12751.88(2) 784.20 49441.36(4) 2.6
773.78 12520.34(2) 798.70 49209.82(4) 2.5
773.78 12545.75(2) 797.08 49235.23(4) 2.1

The first excitation was the same for all the schemes (Fig. 2).

Two of the previously known three-step ionization schemes for Pu trace analysis are marked in the top spectrum from SES #1 (λ2 = 847.26 nm) in Fig. 3 as they are only different in the 3rd-step excitation: (inverted red triangle) λ3 = 750.24 nm and (red star) λ3 = 767.53 nm. The former was used by Raeder et al.17 to demonstrate an overall detection limit of 104 - 105 Pu atoms by in-source RIMS. It was the strongest resonance from SES #1 observed in this work. However, the latter was a relatively small resonance in the spectrum.

The selected three-step schemes were evaluated using the procedure very similar to our previous work10,16. Briefly, the ion source was maintained at a heating current of 300 A, where the Pu sample temperature was estimated to be on the order of 1600 K. The lasers were first tuned to a reference ionization scheme, optimized in wavelength and power, and the mass-selected 242Pu ion intensity was recorded. Then, the second and third lasers were quickly tuned to another scheme, optimized, and the corresponding ion current was measured, and so on. Periodically, the lasers were tuned back to the reference scheme to account for possible variations in the evaporation rate of the Pu sample and drifts in the ion source condition and beamline tuning. The reference scheme was the known scheme with λ2 = 847.26 nm and λ3 = 750.24 nm17. The measured ion intensities obtained with other schemes were normalized to this reference scheme and the resulting relative intensities are given in Column 6 of Table 2.

All the selected schemes were found to produce more ion currents than the reference scheme. Five schemes from three different SESs showed similar relative intensities of 4 - 4.5. The scheme that gave the largest relative ion intensity of 4.5 was λ2 = 818.91 nm and λ3 = 792.79 nm, which excited the Pu atoms to a high-lying excited state at 48590.90 ± 0.05 cm-1 below the continuum followed by ionization by infrared radiation, electric field, collisions, etc., with the exact mechanism(s) not identified.

Ionization efficiency

The overall ionization efficiency for Pu was measured using the three-step scheme that gave the largest relative ion intensity in this study, that is,

7s27F0λ17s7p7D1oλ235977.28±0.04cm-1λ348590.90±0.05cm-1,

where λ1 = 402.77 nm, λ2 = 818.91 nm, and λ3 = 792.79 nm. The measurement procedure was as described below and the three Ti:Sapphire lasers were tuned and optimized to these three wavelengths. A quantified sample with a known number of 242Pu atoms was gradually heated in the ion source while the 242Pu ion beam intensity extracted from the ion source and mass-selected was continuously measured until the sample was completely evaporated out of the ion source. The ionization efficiency was then given by the ratio of the integrated total number of detected ions to the total number of neutral atoms initially in the sample.

Ions of 242Pu ions were observed starting at a heating current of 220 A, at which the estimated cavity temperature was T 1490 K and the Pu sample was at T 1300 K. At the maximum heating current 400 A, the cavity temperature was on the order of 2200 K. Figure 5 shows the 242Pu ion beam intensity recorded as a function of time during the efficiency measurement. The step increases in the ion intensity correspond to stepwise increases of the electrical heating current in 10-A steps. The peak 242Pu+ intensity was about 25 nA at heating current of 370–380 A. Increasing the heating current to 390 A and 400 A, between t = 16 to 17 h, only resulted in smaller increases followed by faster decreases in the ion intensity. This indicated that the sample was close to depletion. The measurement continued until the ion intensity dropped to about 0.9 nA while the ion source was heated at 400 A. During the measurement, the ion intensity from surface ionization was frequently checked by blocking all the laser beams. This corresponds to the vertical drops of the ion intensity down to the baseline in Fig. 4. The maximum surface ion intensity observed was on the order of 80 pA; an average of 0.35 ± 0.22% of the ions were from surface ionization. Based on weight we estimate that the 40 μl sample contained a total 9.97 ± 0.1 x 1015 242Pu atoms which is consistent with the estimate based on the measured activities, of 9.72 ± 0.16 x1015 242Pu. As a conservative estimation, the total number of 242Pu atoms is 9.97 ± 0.25 x 1015 with a larger uncertainty taking into account of the difference between the estimations by weight and by measured activities. The measured ion current, after subtracting the surface ions, gave an overall ionization efficiency of 51.1 ± 1.3 %.

Figure 5.

Figure 5

Time evolution of the efficiency measurement with a sample of 9.97×1015 atoms of 242Pu. The blue line shows the 242Pu ion current measured. The sample was gradually heated and the ion signal was recorded as a function of time. The surface ionization (non-resonant background) was measured by blocking the lasers (red points). The green line shows the heating current from 210 to 400 A. The corresponding cavity temperature was estimated in the range of 1435–2200 K.

Saturation curves

The laser output power for the efficiency measurement was about 100 mW, 1.5 W, and 1.6 W for the first, second, and third steps, respectively. The saturation behavior of the three excitation steps used for the efficiency measurement have been studied. Figure 6 shows the measured 242Pu ion intensity as a function of the laser output power for each excitation, where the solid lines represent the saturation curves obtained by fitting the experimental data to the following saturation equation

I(P)=I0+AP/Psat1+P/Psat 1

where I is the 242Pu ion intensity, P is the laser power, Psat is the saturation power, I0 is the background ion intensity due to non-laser ionization, and A is a constant. The extracted saturation power is Psat = 12 ± 2 mW, 45 ± 5 mW, and 325 ± 67 mW for the first, second, and third excitations, respectively. Thus, all three transitions were well saturated in the efficiency measurement. The selectivity of the three-step scheme was also checked by blocking the individual lasers and we found that approximately 97% of the laser ionized ions were produced by three-step, three-photon ionization and 3% by λ1 + λ2. It should be pointed out that the laser saturation power was measured in the laser room after the lasers beams were merged. From that point, about 70–80% of the laser power was injected into the hot-cavity. That is, the actual saturation power for the excitation steps in the interaction region was about 70–80% of the above fitted Psat values.

Figure 6.

Figure 6

Measured saturation curves for step 1 (λ1 = 402.77 nm), step 2 (λ2 = 818.92 nm), and step 3 (λ3 = 792.79 nm) excitations. The error bars represent one standard deviation of the measurement uncertainty.

Discussion

We have demonstrated an ionization efficiency of 51.1 ± 1.3 % for Pu with the best three-step resonance ionization scheme identified. It should be noted that several other schemes with relative intensity of 4 in Table 2 could be similarly efficient. Our result is a factor of 10 improvement over the previously reported efficiency17 using the reference scheme given in Table 2 (λ1 = 402.77 nm, λ2 = 847.26 nm, and λ3 = 750.24 nm). The overall ionization efficiency depends not only on the ionization scheme but also on other factors including sample evaporation, laser-atom overlapping, and transmission of the ion optics, etc. The efficiency of the reference scheme obtained by Raeder et al. could be limited by the performance of their apparatus. In comparison, the hot-cavity RILIS in the present study has shown to give higher overall efficiencies than similar laser ion sources due to its longer laser-atom overlapping region and near 100% ion transmission to the FC detector. Raeder et al.17 reported a detection limit of 104 - 105 Pu atoms by in-source RIMS using the reference scheme.

The ground state of Pu (Fig. 2) has several fine structure levels with substantial thermal populations at high temperatures. The selected three-step scheme can only access the 242Pu atoms in the ground state 5f67s2 7F0 at 0 cm-1. During the efficiency measurement, the ion source was gradually heated to a heating current of 400 A. The average temperature of the Pu sample was estimated to be on the order of 1800 K. At this temperature, about 55% of the Pu atoms are in the ground state 7F0, with 28% and 9% in the 7F1 level at 2203.606 cm-1 and the 7F2 level at 4299.659 cm-1, respectively. It is possible to further increase the overall ionization efficiency by adding another laser that can provide λ1 = 463.77 nm photons to excite the thermal population in the 7F1 level to the same excited state 5f67s7p 7D1o. Such enhancement has been demonstrated to increase the laser ionization yield for 26Al46 and U37.

Our study is limited to three-step ionization schemes. Two-step resonance ionization schemes have been used for Pu and other actinides4750. They are desired since only two lasers are needed. However, they use frequency-doubled photons for both steps. Studying two-step photoionization spectra requires an autotracking system for continuous scan of frequency-doubled light, which is not available for the present work.

Conclusion

Three-step resonance ionization spectroscopy on 242Pu has been studied using Ti:Sapphire lasers and a hot-cavity laser ion source. We observed eighteen intermediate excited states of even parity in the energy range 35568–36701 cm-1, thirteen of them were newly observed. Three-step photoionization spectra from seven intermediate states were measured. The strongest resonance ionization transitions from different SESs were selected and evaluated under similar ion source operating conditions. This led to the identification of several highly efficient three-step resonant ionization schemes for Pu. The scheme that yielded the largest relative ion intensity was selected for an efficiency measurement and an overall ionization efficiency of 51.1 ± 1.3% was obtained. To our knowledge, this is the highest laser ionization efficiency for plutonium and is an order of magnitude improvement over the previously reported laser ionization efficiency for Pu. This high efficiency could significantly improve RIMS analysis of ultra-trace Pu isotopes.

Acknowledgements

This material is based upon work supported by the U.S. Department of Energy, Office of Science, Office of Nuclear Physics and this research used resources of the Holifield Radioactive Ion Beam Facility of Oak Ridge National Laboratory, which was a DOE Office of Science User Facility. E.R.R. acknowledges funding from the European Research Council (ERC) under the European Union’s Horizon 2020 Research and Innovation Programme (Grant Agreement No. 819957).

Author contributions

A.G.-U. and E.R.R. planned the project. All authors made initial preparations for the experiment, Y.L. and E.R.R. made the measurements, and analysed the data. Y.L. wrote the main manuscript text and E.R.R. assisted in its preparation. D.W.S. assisted with the operation of the spectrometer. All authors discussed the data and reviewed the manuscript.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Contributor Information

Alfredo Galindo-Uribarri, Email: uribarri@ornl.gov.

Yuan Liu, Email: liuy@frib.msu.edu.

Elisa Romero Romero, Email: eromero@vols.utk.edu.

Daniel W. Stracener, Email: stracenerdw@ornl.gov

References

  • 1.Cooper N, editor. Challenges in Plutonium Science. Los Alamos Science; 2000. [Google Scholar]
  • 2.Harley JH. Plutonium in the environment—A review. J. Radiat. Res. 1980;21:83–104. doi: 10.1269/jrr.21.83. [DOI] [PubMed] [Google Scholar]
  • 3.Romanchuk AY, Kalmykov SN, Kersting AB, Zavarin M. Behaviour of plutonium in the environment. Russ. Chem. Rev. 2016;85:995–1010. doi: 10.1070/RCR4602. [DOI] [Google Scholar]
  • 4.Hohenberg C. Radioisotopes and the history of nucleosynthesis in the galaxy. Science. 1969;166:212–215. doi: 10.1126/science.166.3902.212. [DOI] [PubMed] [Google Scholar]
  • 5.Wallner A, et al. Abundance of live 244Pu in deep-sea reservoirs on earth points to rarity of actinide nucleosynthesis. Nat. Commun. 2015;6:1–9. doi: 10.1038/ncomms6956. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Kristo MJ, et al. Nuclear forensic science: Analysis of nuclear material out of regulatory control. Annu. Rev. Earth Planet. Sci. 2016;44:555–579. doi: 10.1146/annurev-earth-060115-012309. [DOI] [Google Scholar]
  • 7.Kurihara E, et al. Particulate plutonium released from the Fukushima Daiichi meltdowns. Sci. Total Environ. 2020;743:140539. doi: 10.1016/j.scitotenv.2020.140539. [DOI] [PubMed] [Google Scholar]
  • 8.Fields BD, et al. Supernova triggers for end-Devonian extinctions. Proc. Natl. Acad. Sci. 2020;117:21008–21010. doi: 10.1073/pnas.2013774117. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Trautmann N, Passler G, Wendt K. Ultratrace analysis and isotope ratio measurements of long-lived radioisotopes by resonance ionization mass spectrometry (RIMS) Anal. Bioanal. Chem. 2004;378:348–355. doi: 10.1007/s00216-003-2183-8. [DOI] [PubMed] [Google Scholar]
  • 10.Romero Romero, E. Development of Ultrasensitive Analytical Techniques to Detect Trace Elements. Ph.D. thesis (University of Tennessee Knoxville, 2019).
  • 11.Kratz J. Ultratrace analysis of long-lived radionuclides by resonance ionization mass spectrometry (RIMS) J. Radioanal. Nucl. Chem. 2015;303:1361–1366. doi: 10.1007/s10967-014-3475-5. [DOI] [Google Scholar]
  • 12.Savina MR, et al. High useful yield and isotopic analysis of uranium by resonance ionization mass spectrometry. Anal. Chem. 2017;89:6224–6231. doi: 10.1021/acs.analchem.7b01204. [DOI] [PubMed] [Google Scholar]
  • 13.Raeder S, et al. Recent developments in resonance ionization mass spectrometry for ultra-trace analysis of actinide elements. Radiochim. Acta. 2019;107:645–652. doi: 10.1515/ract-2019-0001. [DOI] [Google Scholar]
  • 14.Strashnov I, Fernando R, Izosimov I. Trace analysis of radioisotopes by laser spectroscopy and mass spectrometry. J. Radioanal. Nucl. Chem. 2019;322:1437–1445. doi: 10.1007/s10967-019-06899-7. [DOI] [Google Scholar]
  • 15.Ketterer ME, Szechenyi SC. Determination of plutonium and other transuranic elements by inductively coupled plasma mass spectrometry: A historical perspective and new frontiers in the environmental sciences. Spectrochim. Acta Part B Atomic Spectrosc. 2008;63:719–737. doi: 10.1016/j.sab.2008.04.018. [DOI] [Google Scholar]
  • 16.Liu Y, Stracener DW. High efficiency resonance ionization of thorium. Nucl. Instrum. Methods Phys. Res. Sect. B Beam Interact. Mater. Atoms. 2020;462:95–101. doi: 10.1016/j.nimb.2019.11.006. [DOI] [Google Scholar]
  • 17.Raeder S, Hakimi A, Stöbener N, Trautmann N, Wendt K. Detection of plutonium isotopes at lowest quantities using in-source resonance ionization mass spectrometry. Anal. Bioanal. Chem. 2012;404:2163–2172. doi: 10.1007/s00216-012-6238-6. [DOI] [PubMed] [Google Scholar]
  • 18.Trappitsch, R., Savina, M., Rolison, J., Harrison, L. & Dauphas, N. Towards the 1000 atom detection limit for 244Pu. In Lunar and Planetary Science Conference, vol. 2132, 2978 (2019).
  • 19.Donohue D, Young J. Detection of plutonium by resonance ionization mass spectrometry. Anal. Chem. 1983;55:378–379. doi: 10.1021/ac00253a046. [DOI] [Google Scholar]
  • 20.Donohue D, Smith D, Young J, McKown H, Pritchard C. Isotopic analysis of uranium and plutonium mixtures by resonance ionization mass spectrometry. Anal. Chem. 1984;56:379–381. doi: 10.1021/ac00267a018. [DOI] [Google Scholar]
  • 21.Krönert U, et al. Laser resonant ionization of plutonium. Appl. Phys. B. 1985;38:65–70. doi: 10.1007/BF00691772. [DOI] [Google Scholar]
  • 22.Peuser P, et al. Trace detection of plutonium by three-step photoionization with a laser system pumped by a copper vapor laser. Appl. Phys. B. 1985;38:249–253. doi: 10.1007/BF00818051. [DOI] [Google Scholar]
  • 23.Trautmann N, et al. Low-level detection of actinides by laser resonance photoionization. Radiochim. Acta. 1988;44:107–110. doi: 10.1524/ract.1988.4445.1.107. [DOI] [Google Scholar]
  • 24.Ruster W, et al. A resonance ionization mass spectrometer as an analytical instrument for trace analysis. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 1989;281:547–558. doi: 10.1016/0168-9002(89)91489-7. [DOI] [Google Scholar]
  • 25.Grüning C, et al. A high repetition rate solid state laser system for resonance ionisation mass spectrometry of actinides. AIP Conf. Proc. 1998;454:285. doi: 10.1063/1.57161. [DOI] [Google Scholar]
  • 26.Grüning C, et al. Resonance ionization mass spectrometry for ultratrace analysis of plutonium with a new solid state laser system. Int. J. Mass Spectrom. 2004;235:171–178. doi: 10.1016/j.ijms.2004.04.013. [DOI] [Google Scholar]
  • 27.Kunz P, Huber G, Passler G, Trautmann N. Efficient three-step, two-color ionization of plutonium using a resonance enhanced 2-photon transition into an autoionizing state. Eur. Phys. J. D Atomic Mol. Opt. Plasma Phys. 2004;29:183–188. [Google Scholar]
  • 28.Trautmann N, Wendt K. Fast chemical separations and laser mass spectrometry-tools for nuclear research. Radiochim. Acta. 2012;100:675–685. doi: 10.1524/ract.2012.1951. [DOI] [Google Scholar]
  • 29.Isselhardt B, et al. Improved precision and accuracy in quantifying plutonium isotope ratios by RIMS. J. Radioanal. Nucl. Chem. 2016;307:2487–2494. doi: 10.1007/s10967-015-4393-x. [DOI] [Google Scholar]
  • 30.Franzmann M, Bosco H, Hamann L, Walther C, Wendt K. Resonant laser-SNMS for spatially resolved and element selective ultra-trace analysis of radionuclides. J. Anal. Atomic Spectrom. 2018;33:730–737. doi: 10.1039/C7JA00423K. [DOI] [Google Scholar]
  • 31.Köhler S, et al. Determination of the first ionization potential of actinide elements by resonance ionization mass spectroscopy. Spectrochim. Acta Part B Atomic Spectrosc. 1997;52:717–726. doi: 10.1016/S0584-8547(96)01670-9. [DOI] [Google Scholar]
  • 32.Erdmann N, Kratz J-V, Trautmann N, Passler G. Resonance ionization mass spectrometry of ion beam sputtered neutrals for element-and isotope-selective analysis of plutonium in micro-particles. Anal. Bioanal. Chem. 2009;395:1911–1918. doi: 10.1007/s00216-009-2906-6. [DOI] [PubMed] [Google Scholar]
  • 33.Stephan T, et al. CHILI—The Chicago instrument for laser ionization—A new tool for isotope measurements in cosmochemistry. Int. J. Mass Spectrom. 2016;407:1–15. doi: 10.1016/j.ijms.2016.06.001. [DOI] [Google Scholar]
  • 34.Higuchi Y, Watanabe K, Kawarabayashi J, Iguchi T. Development of trace isotope analysis using resonance ionization mass spectrometry based on isotope selection with doppler shift of laser ablated atoms. J. Nucl. Sci. Technol. 2006;43:334–338. doi: 10.1080/18811248.2006.9711100. [DOI] [Google Scholar]
  • 35.Worden EF, Carlson LR, Johnson SA, Paisner JA, Solarz RW. Ionization potential of neutral atomic plutonium determined by laser spectroscopy. J. Opt. Soc. Am. B. 1993;10:1998–2005. doi: 10.1364/JOSAB.10.001998. [DOI] [Google Scholar]
  • 36.Blaise J, Fred M, Gutmacher RG. Term analysis of the spectrum of neutral plutonium, Pu I. J. Opt. Soc. Am. B. 1986;3:403–418. doi: 10.1364/JOSAB.3.000403. [DOI] [Google Scholar]
  • 37.Savina MR, Trappitsch R, Kucher A, Isselhardt BH. New resonance ionization mass spectrometry scheme for improved uranium analysis. Anal. Chem. 2018;90:10551–10558. doi: 10.1021/acs.analchem.8b02656. [DOI] [PubMed] [Google Scholar]
  • 38.Studer D, Dyrauf P, Naubereit P, Heinke R, Wendt K. Resonance ionization spectroscopy in dysprosium. Hyperfine Interact. 2017;238:1–11. doi: 10.1007/s10751-016-1384-4. [DOI] [Google Scholar]
  • 39.Gadelshin VM, et al. Measurement of the laser resonance ionization efficiency for lutetium. Radiochim. Acta. 2019;107:653–661. doi: 10.1515/ract-2019-3118. [DOI] [Google Scholar]
  • 40.Beene JR, et al. ISOL science at the Holifield radioactive ion beam facility. J. Phys. G Nucl. Part. Phys. 2011;38:024002. doi: 10.1088/0954-3899/38/2/024002. [DOI] [Google Scholar]
  • 41.Liu Y, et al. On-line commissioning of the HRIBF resonant ionization laser ion source. Nucl. Instrum. Methods Phys. Res. Sect. B Beam Interact. Mater. Atoms. 2013;298:5–12. doi: 10.1016/j.nimb.2012.12.041. [DOI] [Google Scholar]
  • 42.Liu Y, Stracener DW. A resonant ionization laser ion source at ORNL. Nucl. Instrum. Methods Phys. Res. Sect. B Beam Interact. Mater. Atoms. 2016;376:68–72. doi: 10.1016/j.nimb.2015.12.015. [DOI] [Google Scholar]
  • 43.Kramida, A., Ralchenko, Y. & Reader, J. NIST Atomic Spectra Database (ver. 5.7. 1). Available: https://physics.nist.gov/asd [2020, July 21] (National Institute of Standards and Technology, 2019).
  • 44.Bosco H, et al. Influence of the hyperfine structure on plutonium in resonant laser-SNMS. Hyperfine Interact. 2020;241:1–8. doi: 10.1007/s10751-020-1696-2. [DOI] [Google Scholar]
  • 45.Kieck T, et al. Resonance laser ionization spectroscopy of tellurium. Spectrochim. Acta Part B Atomic Spectrosc. 2019;159:105645. doi: 10.1016/j.sab.2019.105645. [DOI] [Google Scholar]
  • 46.Prime EJ, Lassen J, Achtzehn T, et al. TRIUMF resonant ionization laser ion source. Hyperfine Interact. 2006;171:127–134. doi: 10.1007/s10751-006-9493-0. [DOI] [Google Scholar]
  • 47.Voss A, et al. High-resolution laser spectroscopy of long-lived plutonium isotopes. Phys. Rev. A. 2017;95:032506. doi: 10.1103/PhysRevA.95.032506. [DOI] [Google Scholar]
  • 48.Kneip N, et al. Highly selective two-step laser ionization schemes for the analysis of actinide mixtures. Hyperfine Interact. 2020;241:1–7. doi: 10.1007/s10751-020-01712-4. [DOI] [Google Scholar]
  • 49.Zhang K, et al. Detection of the lowest-lying odd-parity atomic levels in actinium. Phys. Rev. Lett. 2020;125:073001. doi: 10.1103/PhysRevLett.125.073001. [DOI] [PubMed] [Google Scholar]
  • 50.Raeder S, et al. Probing sizes and shapes of nobelium isotopes by laser spectroscopy. Phys. Rev. Lett. 2018;120:232503. doi: 10.1103/PhysRevLett.120.232503. [DOI] [PubMed] [Google Scholar]

Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES