Ultra-Short Bone Conserving Cementless Femoral Stem

Young-Hoo Kim 1,
PMCID: PMC8654591  PMID: 34938687

Abstract

Excellent long-term results have been reported with conventional length cementless femoral stems in total hip arthroplasty; however, proximal stress shielding and thigh pain are still a concern. Metaphyseal engaging bone conserving short stems provide theoretical benefits compared with conventional length cementless stems, including avoiding proximal-distal mismatch, decreasing proximal stress shielding, and limiting perioperative fractures. The purpose of the ultra-short bone conserving cementless stem was to reproduce natural load transfer with an ultra-short stem obtaining optimal stability using the morphology of the proximal femur. Loss of stability of the stem and failure of osseous ingrowth is a potential concern with the use of ultra-short proximal loading cementless femoral stems. Ultra-short, metaphyseal-fitting anatomic or non-anatomic cementless femoral stems provided stable fixation without relying on diaphyseal fixation in young and elderly patients, suggesting that metaphyseal-fitting alone is sufficient in young and elderly patients who have good bone quality.

Keywords: Ultra-short cementless stem, Total hip arthroplasty, Young and elderly patients

INTRODUCTION

Satisfactory clinical and radiographic outcomes at long-term follow-up may be achieved with use of conventional length cementless femoral stems (stem engaging metaphysis and diaphysis of the proximal femur)1,2,3,4,5,6,7). Despite excellent long-term results with most designs, stress-shielding and thigh pain may occur4,5,8,9). Several bone conserving femoral stems have been designed during the last two decades, including mid-short stem (engaging metaphysis and junction of the metaphysis and diaphysis of the proximal femur), and ultra-short stem (engaging only metaphysis of the proximal femur) and some authors have advocated their use, particularly in young patients with high activity interests10,11). The goals of bone conserving femoral stems include the following: sparing of the trochanteric bone stock; a more physiological loading in the proximal femur reducing the risk of stress shielding; and to avoid a long stem into the diaphysis preventing impingement with the femoral cortex and thigh pain12,13). Several systems for classification of bone conserving femoral stems have been developed, considering features such as length of the stem, location of loading, osteotomy level for the femoral neck resection and implant fixation principles (Table 1)14,15,16,17,18).

Table 1. Classification Systems for Femoral Stems.

Study Publication year Classes Description Rationale
McTighe et al.14) 2013 Resurfacing Resurfacing Assessment of length and method of achieving primary stability of the stem
Neck stabilized Short curved neck-sparing stems, and standard-length stems preserving femoral neck, but engaging the neck, metaphysis and diaphysis
Metaphyseal stabilized Short metaphyseal stems including anatomical, straight and tapered designs
Conventional (metaphyseal/diaphyseal) stabilized Conventional stems engaging both metaphysis and diaphysis
Feyen and Shimmin17) 2014 Type I Resurfacing Assessment of the osteotomy level for the neck resection and implant fixation principles
Type II Mid-head resection stems
Type III Short stems with subcapital (IIIA) or standard (IIIB) osteotomy
Type IV Traditional stems
Type V Diaphyseal fixation stems
Van Oldenrijk et al.18) 2014 Collum Conical or cylindrical ultra-short stems, with complete anchorage in the femoral neck Assessment of the osteotomy level for the neck resection and implant fixation principles
Partial collum Partial femoral neck-sparing curved designs
Trochanter-sparing Trochanter-sparing but not neck-sparing, and shortened tapered stems
Khanuja et al.15) 2014 Type I Femoral neck fixation stems (from IA to IC according to the stem geometry) Assessment of location of loading and implant fixation principles
Type II Calcar loading stems (from IIA to IID according to the stem geometry)
Type III Calcar loading with lateral flare stems
Type IV Shortened tapered stems
Falez et al.16) 2015 Collum Conical or cylindrical ultra-short stems, with complete anchorage in the femoral neck Assessment of the osteotomy level for the neck resection and implant fixation principles
Partial collum Partial femoral neck-sparing curved designs
Trochanter-sparing Trochanter-sparing but not neck-sparing, and shortened tapered stems
Trochanter-harming Short stems interrupting the circumferential integrity of the femoral neck section and violating trochanteric region

McTighe et al.14) proposed the term ‘short’ for stems that do not extend below the metaphyseal region of the proximal femur. In this respect, they proposed three types of stems: head stabilized (resurfacing); neck stabilized; and metaphyseal stabilized. Khanuja et al.15) recently classified bone conserving short stems according to fixation principles and location of proximal loading. They proposed four categories: femoral neck fixation; calcar loading; lateral flare and calcar loading; and shortened taper stems. Similar prosthesis survival rates and functional outcomes in primary total hip arthroplasty (THA) were observed for the majority of bone conserving short femoral stems demonstrated, compared with conventional length cementless stems. However, superior bone remodeling and preservation of more proximal bone stock in the short and long-term may be achieved with a bone conserving short femoral stem. In addition, a bone conserving short femoral stem may be applied to any type of femoral morphology10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27,28).

The purpose of this review was to focus exclusively on type III (classification by Feyen and Shimmin17)) or type III (classification by Khanuja et al.15)) bone conserving short femoral stems, providing a description of their features and an analysis of their clinical and radiological results, and survival rates.

ANATOMIC BASIS FOR BONE CONSERVING SHORT FEMORAL STEM

Dorr et al.29) observed that the poor correlation between the proximal and distal dimensions of the femoral canal necessitates the selection of stems based on their fit in the proximal rather than the distal canal, thereby optimizing the metaphyseal load transfer. They also found that the mediolateral diameter of the femoral canal at a point 20 mm distal to the lesser trochanter has the most predictable relationship with external femoral dimensions. These findings provide an anatomical basis for the metaphyseal fixation of certain types of cementless stem.

BIOMECHANIC BASIS FOR BONE CONSERVING SHORT FEMORAL STEM

Walker et al.30) suggested that extending the femoral stem beyond the lesser trochanter is unnecessary for a cementless anatomic femoral component with a lateral flare, and that a short, metaphyseal-fitting is sufficient. Leali et al.21) found that a proximally fixed cementless femoral component with a lateral flare provided solid initial stability. When using a cementless stem, normal patterns of strain are approached when a tight proximal fit of the stem is achieved31,32), whereas a tight distal fit can significantly reduce proximal strains31). The closer the contact of the distal part of the stem, the more proximal stress shielding occurs, whereas the absence of contact between the stem and the distal cortex may reduce stress shielding, bone resorption and thigh pain31). Hence the length of the stem plays is a critical component in the transfer of forces to the femoral bone. Conceptually, reducing the length of the stem reduces proximal stress shielding, at the cost of a reduced contact area for fixation and load transfer. Bieger et al.33) and Arno et al.34) suggested that shortening a femoral stem reduces proximal stress shielding without compromising primary stability. They also concluded that a metaphyseal only design biomechanically provides the best match of the native femur.

CLINICAL STUDIES ON BONE CONSERVING SHORT FEMORAL STEMS

1. IPS Stem (Lateral Flare Calcar Loading Anatomic Stem with Distal Stem)

Considering that most cementless femoral stems are applied in young patients, preservation of bone stock and reduction of thigh pain and osteolysis when possible would be advantageous. Conservative metaphyseal-fitting anatomic cementless femoral stems with an alternative bearing surface such as an alumina-on-alumina bearing meet this requirement. Metaphyseal-engaging short stems provide theoretical benefits compared with conventional length cementless stems, including avoiding proximal-distal mismatch, decreasing proximal stress shielding, and limiting perioperative fractures.

New total hip prosthesis (Immediate Postoperative Stability [IPS]; DePuy, Leeds, UK) was developed by Kim in 199535) (Fig. 1). The intention was to reproduce natural load transfer with a short stem while obtaining optimal stability using the morphology of the proximal femur. In this design of the stem, vertical stability was provided by the wedge shape of the prosthesis with the addition of a lateral flare. This increases the load on the proximal femur, medially and laterally, and decreases load transmission to the femoral diaphysis. The transition zone between the loadbearing and nonloadbearing section of the stem is short, avoiding metal-to bone contact below the metaphysis. The polished distal stem is short and narrow and placed centrally in the femoral canal to avoid distal contact with the femur. The proximal 30% of the stem is porous-coated with sintered titanium beads with a mean pore size of 250 µm to which a hydroxyapatite coating is applied to a thickness of 30 µm.

Fig. 1. (A, B) Photos of an IPS (Immediate Postoperative Stability; DePuy) stem and radiographs of both hips taken 20 years after the operation.

Fig. 1

A summary of the clinical results on IPS short anatomic cementless stems is shown in Table 220,24,35,36,37,38,39,40,41,42,43,44,45,46). Mild stress-shielding (calcar round-off) was observed and none of the patients experienced thigh pain. With an abundance of papers describing the use of IPS cementless stems with short follow-up, one paper reported long-term results (Table 2). The question of whether stable fixation can be obtained without diaphyseal fixation is a potential concern with the use of short, metaphyseal-fitting anatomic cementless femoral components. In our studies, osseointegration was reliable with an IPS stem20,37,38,39,40). Walker et al.30) and Leali et al.21) suggested that the femoral stem below the lesser trochanter would be unnecessary for a cementless anatomic femoral stem with a lateral flare and that a short stem would suffice.

Table 2. Demographic Data and IPS Stem Survivorship.

Study Level of evidence No. of hips No. of patients Mean age (yr) Mean follow-up (yr) Survivorship (%)
Kim et al.37) Level I 100 50 45.3 6.6 100
Kim et al.38) Level II 601 471 52.7 8.8 99.7
Kim et al.20) Level IV 630 500 52.7 15.8 100
Kim et al.39) Level I 140 120 45.3 6.4 100
Kim et al.35) Level III 60 50 46.6 6.3 100
Cinotti et al.40) Level IV 72 64 68 9 100
Kim et al.41) Level I 93 64 38.2 11.1 100
Kim et al.42) Level I 200 100 45.3 5.6 100
Kim et al.43) Level IV 73 71 45.5 8.5 100
Kim et al.24) Level IV 110 55 46.3 15.6 100
Kim et al.44) Level IV 127 96 24 14.6 100
Kim et al.45) Level IV 60 50 28.3 10.8 100
Kim et al.46) Level I 200 100 45.3 12.4 100
Kim et al.36) Level I 100 50 51 4.8 100

IPS: Immediate Postoperative Stability.

2. Proxima Stem (Lateral Flare Calcar Loading Anatomic without Distal Stem)

A new ultra-short anatomic cementless femoral stem (Proxima; DePuy) was developed by Kim in 200123) (Fig. 2). One of the main reasons for developing a new ultra-short metaphyseal-fitting porous-coated anatomic cementless femoral stem was to preserve bone and to provide more physiological loading. The ultra-short Proxima cementless femoral stem is designed to have a close fit within the proximal femur with the aim of maximizing primary stability, particularly in torsion, thereby limiting bone resorption due to stress shielding. It is manufactured using titanium alloy and is entirely porous-coated with sintered titanium beads having a mean pore size of 250 µm, to which a 30 µm thick hydroxyapatite coating is applied, except for the distal tip. The design features include a longer proximomedial portion of the stem, a highly pronounced lateral flare and preservation of the femoral neck. The question arises, at the time of development, as to whether it is possible to obtain rigid fixation of this stem without diaphyseal anchoring.

Fig. 2. (A, B) Photo of a Proxima (DePuy) stem and a radiograph of the right hip taken 15 years after the operation.

Fig. 2

A summary of the clinical results on Proxima ultra-short anatomic cementless stems is shown in Table 323,24,25,26,27,28,47,48,49,50,51,52,53). Mild-stress shielding (calcar round-off) was observed and none of the patients experienced thigh pain. All of the previous studies23,24,25,26,27,28,47,48,49) on Proxima stem suggested excellent clinical results in the short or long-term follow-up. Kim et al.47) obtained similar long-term results using ultra-short and conventional length cementless anatomic femoral stems in patients <65 years old, in terms of clinical and radiographic results, survival rates, and complication rates. However, significantly higher incidence of thigh pain and stress shielding-related periprosthetic bone resorption was observed in the conventional length stem group compared with the ultra-short stem group.

Table 3. Demographic Data and Proxima Stem Survivorship.

Study Level of evidence No. of hips No. of patients Mean age (yr) Mean follow-up (yr) Survivorship (%)
Kim et al.28) Level IV 84 84 78.9 4.6 100
Kim et al.27) Level I 60 50 54.3 3.35 100
Kim et al.26) Level IV 144 126 53.9 4.5 100
Kim and Oh25) Level II 70 70 74.9 4.1 100
Kim et al.24) Level IV 256 230 65 5.5 98.2
Kim et al.23) Level IV 226 200 43.9/78.9 7.5 100
Kim et al.47) Level III 858 759 56.3 16.5 97.6
Renkawitz et al.48) Level IV 200 200 18-70 15 100
Rastogi and Marya49) Level IV 50 41 45 4 95.1
Kim et al.50) Level I 524 262 53 11.8 99.6
Gombár et al.51) Level IV 86 81 50 7 97
Melisík et al.52) Level IV 130 121 <60 9.8 98.5
Kim and Jang53) Level IV 284 280 72.8 9.3 97

It has been suggested that stress shielding may be minimized by a low-modulus, intimately fit proximally device that does not bypass the proximal medial regions with distal fixation20). Using the ultra-short Proxima cementless anatomic femoral stem, a level of fixation in the proximal femur that was as adequate as that of the conventional length cementless anatomic femoral stem was achieved, but it provided significantly less stress shielding bone resorption than the conventional length cementless anatomic femoral stem47).

It is believed that short-stemmed components are associated with a higher rate of coronal malalignment15) when compared with femoral stems of conventional length. There was no significant difference in survivorship of varus components compared with neutrally implanted components. The findings of Kim et al.50) concur with those of this systemic review (98.6% survivorship at 12 years).

The Australian Orthopaedic Association National Joint Replacement Registry54) reported that the cumulative incidence of aseptic loosening for the short-stemmed THAs was more than twice that of other femoral components at 10 years (2.5% compared with 1.2%). In a long-term study, Kim et al.50) found that the survival rate of the ultra-short cementless anatomic stem (97.6%) was comparable to that of the conventional length cementless anatomic stem (96.6%). They believed that the satisfactory results using the ultra-short cementless anatomic stem can be attributed to several factors, that is, good quality of bone, optimal preparation of the proximal femur along with preservation of the femoral neck, and circumferential metaphyseal fitting.

3. SMF Stem (Ultra-short Non-anatomic Calcar Loading Stem)

Among numerous short bone conserving proximal loading cementless stems, ultra-short anatomic and ultra-short non-anatomic proximal loading cementless femoral stems were introduced to facilitate osseointegration of the stem without diaphyseal stem fixation. In the ultra-short anatomic cementless stem (Proixma; DePuy) vertical stability is provided by the wedge shape of the stem with the addition of a lateral flare and preservation of the femoral neck. In the ultra-short non-anatomic proximal loading cementless stem (Short Modular Femoral [SMF]; Smith & Nephew, Memphis, TN, USA) (Fig. 3), vertical stability is provided by the wedge shape of the stem with 3-point fixation in the femoral canal and preservation of the femoral neck. Preservation of the femoral neck and the wedge shape of the stem provide greater torsional stability and reduce distal migration of the femoral stem. Absence of distal stem fixation is allowed because of the effective stability provided by the wedge shape of the stem with preservation of the femoral neck. The absence of diaphyseal stem fixation attempts proximal load transfer to reduce stress shielding and thigh pain. In addition, it attempts preservation of the femoral canal and femoral elasticity, and ease of revision. In the current study, mild stress shielding (calcar round-off) was observed and none of the patients experienced thigh pain.

Fig. 3. (A, B) Photos of SMF (Short Modular Femoral; Smith & Nephew) stems and a radiograph of both hips taken seven years after the operation.

Fig. 3

McCalden et al.55) conducted a randomized controlled trial comparing the patterns of migration of a SMF stem with metaphyseal fixation in 22 patients with those of a conventional length femoral stem with metaphyseal fixation (Synergy; Smith & Nephew) in 21 patients. At 24 months after surgery, no statistically significant difference in mean migration was observed between the groups: total migration was 1.09±1.74 mm and 0.73±0.72 mm, respectively. A total migration <0.6 mm subsidence <0.5 mm and rotation <1.0° was observed for the majority of stems in both groups. In the group with SMF, early migration >1.0 mm which stabilized within six months was observed for three stems, and one stem had an early progressive migration requiring revision three years after surgery. In a study by Kim et al.56), in the short-term (3.5 years) follow-up, rigid fixation of the stem occurred in all of the 56 hips with a SMF stem. The findings of McCalden et al.55) and Kim et al.56) support the assumption that torsional loads can be controlled without diaphyseal stem fixation by preservation of the neck and tight fixation in the metaphysis of the femur.

CONCLUSION

Loss of stability of the stem and failure of osseous ingrowths are potential concerns with the use of an ultra-short proximal loading cementless femoral stem. However, ultra-short, metaphyseal-fitting anatomic or non-anatomic cementless femoral stems provided stable fixation without relying on diaphyseal fixation in young and elderly patients, suggesting that metaphyseal-fitting alone is sufficient in young and elderly patients who have good bone quality.

ACKNOWLEDGEMENTS

The author expresses his gratitude to Dr. Jang-Won Park, MD at the Joint Replacement Center, Ewha Womans University Seoul Hospital for his review of literature.

Footnotes

CONFLICT OF INTEREST: The authors declare that there is no potential conflict of interest relevant to this article.

References

  • 1.Bojescul JA, Xenos JS, Callaghan JJ, Savory CG. Results of porous-coated anatomic total hip arthroplasty without cement at fifteen years: a concise follow-up of a previous report. J Bone Joint Surg Am. 2003;85:1079–1083. doi: 10.2106/00004623-200306000-00015. [DOI] [PubMed] [Google Scholar]
  • 2.Capello WN, D'Antonio JA, Feinberg JR, Manley MT. Ten-year results with hydroxyapatite-coated total hip femoral components in patients less than fifty years old. A concise follow-up of a previous report. J Bone Joint Surg Am. 2003;85:885–889. doi: 10.2106/00004623-200305000-00017. [DOI] [PubMed] [Google Scholar]
  • 3.Capello WN, D'Antonio JA, Jaffe WL, Geesink RG, Manley MT, Feinberg JR. Hydroxyapatite-coated femoral components: 15-year minimum follow-up. Clin Orthop Relat Res. 2006;453:75–80. doi: 10.1097/01.blo.0000246534.44629.b2. [DOI] [PubMed] [Google Scholar]
  • 4.Engh CA, Jr, Claus AM, Hopper RH, Jr, Engh CA. Long-term results using the anatomic medullary locking hip prosthesis. Clin Orthop Relat Res. 2001;(393):137–146. doi: 10.1097/00003086-200112000-00016. [DOI] [PubMed] [Google Scholar]
  • 5.Engh CA, Jr, Mohan V, Nagowski JP, Sychterz Terefenko CJ, Engh CA., Sr Influence of stem size on clinical outcome of primary total hip arthroplasty with cementless extensively porous-coated femoral components. J Arthroplasty. 2009;24:554–559. doi: 10.1016/j.arth.2008.02.002. [DOI] [PubMed] [Google Scholar]
  • 6.Gul R, Jeer PJ, Oakeshott RD. Twenty-year survival of a cementless revision hip arthroplasty using a press-fit bulk acetabular allograft for pelvic discontinuity: a case report. J Orthop Surg (Hong Kong) 2008;16:111–113. doi: 10.1177/230949900801600126. [DOI] [PubMed] [Google Scholar]
  • 7.Kim YH. Long-term results of the cementless porous-coated anatomic total hip prosthesis. J Bone Joint Surg Br. 2005;87:623–627. doi: 10.1302/0301-620X.87B5.15554. [DOI] [PubMed] [Google Scholar]
  • 8.Lavernia C, D'Apuzzo M, Hernandez V, Lee D. Thigh pain in primary total hip arthroplasty: the effects of elastic moduli. J Arthroplasty. 2004;19(7 Suppl 2):10–16. doi: 10.1016/j.arth.2004.06.023. [DOI] [PubMed] [Google Scholar]
  • 9.Faraj AA, Yousuf M. Anterior thigh pain after cementless total hip arthroplasty. Int Orthop. 2005;29:149–151. doi: 10.1007/s00264-005-0642-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Goebel D, Schultz W. The Mayo cementless femoral component in active patients with osteoarthritis. Hip Int. 2009;19:206–210. doi: 10.1177/112070000901900303. [DOI] [PubMed] [Google Scholar]
  • 11.Morrey BF, Adams RA, Kessler M. A conservative femoral replacement for total hip arthroplasty. A prospective study. J Bone Joint Surg Br. 2000;82:952–958. doi: 10.1302/0301-620x.82b7.10420. [DOI] [PubMed] [Google Scholar]
  • 12.Rometsch E, Bos PK, Koes BW. Survival of short hip stems with a “modern”, trochanter-sparing design - a systematic literature review. Hip Int. 2012;22:344–354. doi: 10.5301/HIP.2012.9472. [DOI] [PubMed] [Google Scholar]
  • 13.Banerjee S, Pivec R, Issa K, Harwin SF, Mont MA, Khanuja HS. Outcomes of short stems in total hip arthroplasty. Orthopedics. 2013;36:700–707. doi: 10.3928/01477447-20130821-06. [DOI] [PubMed] [Google Scholar]
  • 14.McTighe T, Stulberg SD, Keppler L, et al. A classification system for short stem uncemented total hip arthroplasty. Bone Joint J. 2013;95:260 [Google Scholar]
  • 15.Khanuja HS, Banerjee S, Jain D, Pivec R, Mont MA. Short bone-conserving stems in cementless hip arthroplasty. J Bone Joint Surg Am. 2014;96:1742–1752. doi: 10.2106/JBJS.M.00780. [DOI] [PubMed] [Google Scholar]
  • 16.Falez F, Casella F, Papalia M. Current concepts, classification, and results in short stem hip arthroplasty. Orthopedics. 2015;38(3 Suppl):S6–S13. doi: 10.3928/01477447-20150215-50. [DOI] [PubMed] [Google Scholar]
  • 17.Feyen H, Shimmin AJ. Is the length of the femoral component important in primary total hip replacement? Bone Joint J. 2014;96-B:442–448. doi: 10.1302/0301-620X.96B4.33036. [DOI] [PubMed] [Google Scholar]
  • 18.van Oldenrijk J, Molleman J, Klaver M, Poolman RW, Haverkamp D. Revision rate after short-stem total hip arthroplasty: a systematic review of 49 studies. Acta Orthop. 2014;85:250–258. doi: 10.3109/17453674.2014.908343. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Patel RM, Smith MC, Woodward CC, Stulberg SD. Stable fixation of short-stem femoral implants in patients 70 years and older. Clin Orthop Relat Res. 2012;470:442–449. doi: 10.1007/s11999-011-2063-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Kim YH, Park JW, Kim JS, Kang JS. Long-term results and bone remodeling after THA with a short, metaphyseal-fitting anatomic cementless stem. Clin Orthop Relat Res. 2014;472:943–950. doi: 10.1007/s11999-013-3354-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Leali A, Fetto J, Insler H, Elfenbein D. The effect of a lateral flare feature on implant stability. Int Orthop. 2002;26:166–169. doi: 10.1007/s00264-002-0355-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Dabirrahmani D, Hogg M, Kohan L, Gillies M. Primary and long-term stability of a short-stem hip implant. Proc Inst Mech Eng H. 2010;224:1109–1119. doi: 10.1243/09544119JEIM704. [DOI] [PubMed] [Google Scholar]
  • 23.Kim YH, Park JW, Kim JS. Behaviour of the ultra-short anatomic cementless femoral stem in young and elderly patients. Int Orthop. 2013;37:2323–2330. doi: 10.1007/s00264-013-2044-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Kim YH, Park JW, Kim JS. Is diaphyseal stem fixation necessary for primary total hip arthroplasty in patients with osteoporotic bone (Class C bone)? J Arthroplasty. 2013;28:139–146.e1. doi: 10.1016/j.arth.2012.04.002. [DOI] [PubMed] [Google Scholar]
  • 25.Kim YH, Oh JH. A comparison of a conventional versus a short, anatomical metaphyseal-fitting cementless femoral stem in the treatment of patients with a fracture of the femoral neck. J Bone Joint Surg Br. 2012;94:774–781. doi: 10.1302/0301-620X.94B6.29152. [DOI] [PubMed] [Google Scholar]
  • 26.Kim YH, Kim JS, Joo JH, Park JW. A prospective short-term outcome study of a short metaphyseal fitting total hip arthroplasty. J Arthroplasty. 2012;27:88–94. doi: 10.1016/j.arth.2011.02.008. [DOI] [PubMed] [Google Scholar]
  • 27.Kim YH, Choi Y, Kim JS. Comparison of bone mineral density changes around short, metaphyseal-fitting, and conventional cementless anatomical femoral components. J Arthroplasty. 2011;26:931–940.e1. doi: 10.1016/j.arth.2010.10.001. [DOI] [PubMed] [Google Scholar]
  • 28.Kim YH, Kim JS, Park JW, Joo JH. Total hip replacement with a short metaphyseal-fitting anatomical cementless femoral component in patients aged 70 years or older. J Bone Joint Surg Br. 2011;93:587–592. doi: 10.1302/0301-620X.93B5.25994. [DOI] [PubMed] [Google Scholar]
  • 29.Dorr LD, Faugere MC, Mackel AM, Gruen TA, Bognar B, Malluche HH. Structural and cellular assessment of bone quality of proximal femur. Bone. 1993;14:231–242. doi: 10.1016/8756-3282(93)90146-2. [DOI] [PubMed] [Google Scholar]
  • 30.Walker PS, Culligan SG, Hua J, Muirhead-Allwood SK, Bentley G. The effect of a lateral flare feature on uncemented hip stems. Hip Int. 1999;9:71–80. [Google Scholar]
  • 31.Kim YH, Kim JS, Cho SH. Strain distribution in the proximal human femur. An in vitro comparison in the intact femur and after insertion of reference and experimental femoral stems. J Bone Joint Surg Br. 2001;83:295–301. doi: 10.1302/0301-620x.83b2.10108. [DOI] [PubMed] [Google Scholar]
  • 32.Hua J, Walker PS. Closeness of fit of uncemented stems improves the strain distribution in the femur. J Orthop Res. 1995;13:339–346. doi: 10.1002/jor.1100130307. [DOI] [PubMed] [Google Scholar]
  • 33.Bieger R, Ignatius A, Decking R, Claes L, Reichel H, Dürselen L. Primary stability and strain distribution of cementless hip stems as a function of implant design. Clin Biomech (Bristol, Avon) 2012;27:158–164. doi: 10.1016/j.clinbiomech.2011.08.004. [DOI] [PubMed] [Google Scholar]
  • 34.Arno S, Fetto J, Nguyen NQ, et al. Evaluation of femoral strains with cementless proximal-fill femoral implants of varied stem length. Clin Biomech (Bristol, Avon) 2012;27:680–685. doi: 10.1016/j.clinbiomech.2012.03.006. [DOI] [PubMed] [Google Scholar]
  • 35.Kim YH. Cementless total hip arthroplasty with a close proximal fit and short tapered distal stem (third-generation) prosthesis. J Arthroplasty. 2002;17:841–850. doi: 10.1054/arth.2002.33555. [DOI] [PubMed] [Google Scholar]
  • 36.Kim YH, Yoon SH, Kim JS. Changes in the bone mineral density in the acetabulum and proximal femur after cementless total hip replacement: alumina-on-alumina versus aluminaon-polyethylene articulation. J Bone Joint Surg Br. 2007;89:174–179. doi: 10.1302/0301-620X.89B2.18634. [DOI] [PubMed] [Google Scholar]
  • 37.Kim YH, Kim JS, Oh SH, Kim JM. Comparison of porous-coated titanium femoral stems with and without hydroxyapatite coating. J Bone Joint Surg Am. 2003;85:1682–1688. doi: 10.2106/00004623-200309000-00005. [DOI] [PubMed] [Google Scholar]
  • 38.Kim YH. The results of a proximally-coated cementless femoral component in total hip replacement: a five- to 12-year follow-up. J Bone Joint Surg Br. 2008;90:299–305. doi: 10.1302/0301-620X.90B3.20096. [DOI] [PubMed] [Google Scholar]
  • 39.Kim YH. Titanium and cobalt-chrome cementless femoral stems of identical shape produce equal results. Clin Orthop Relat Res. 2004;(427):148–156. doi: 10.1097/01.blo.0000142350.80416.84. [DOI] [PubMed] [Google Scholar]
  • 40.Cinotti G, Della Rocca A, Sessa P, Ripani FR, Giannicola G. Thigh pain, subsidence and survival using a short cementless femoral stem with pure metaphyseal fixation at minimum 9-year follow-up. Orthop Traumatol Surg Res. 2013;99:30–36. doi: 10.1016/j.otsr.2012.09.016. [DOI] [PubMed] [Google Scholar]
  • 41.Kim YH, Choi Y, Kim JS. Cementless total hip arthroplasty with ceramic-on-ceramic bearing in patients younger than 45 years with femoral-head osteonecrosis. Int Orthop. 2010;34:1123–1127. doi: 10.1007/s00264-009-0878-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Kim YH, Kim JS, Choi YW, Kwon OR. Intermediate results of simultaneous alumina-on-alumina bearing and alumina-on-highly cross-linked polyethylene bearing total hip arthroplasties. J Arthroplasty. 2009;24:885–891. doi: 10.1016/j.arth.2008.05.009. [DOI] [PubMed] [Google Scholar]
  • 43.Kim YH, Choi Y, Kim JS. Cementless total hip arthroplasty with alumina-on-highly cross-linked polyethylene bearing in young patients with femoral head osteonecrosis. J Arthroplasty. 2011;26:218–223. doi: 10.1016/j.arth.2010.03.010. [DOI] [PubMed] [Google Scholar]
  • 44.Kim YH, Park JW, Kim JS. Cementless metaphyseal fitting anatomic total hip arthroplasty with a ceramic-on-ceramic bearing in patients thirty years of age or younger. J Bone Joint Surg Am. 2012;94:1570–1575. doi: 10.2106/JBJS.K.00697. [DOI] [PubMed] [Google Scholar]
  • 45.Kim YH, Park JW, Kulkarni SS, Kim YH. A randomised prospective evaluation of ceramic-on-ceramic and ceramic-on-highly cross-linked polyethylene bearings in the same patients with primary cementless total hip arthroplasty. Int Orthop. 2013;37:2131–2137. doi: 10.1007/s00264-013-2036-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Kim YH, Park JW, Patel C, Kim DY. Polyethylene wear and osteolysis after cementless total hip arthroplasty with alumina-on-highly cross-linked polyethylene bearings in patients younger than thirty years of age. J Bone Joint Surg Am. 2013;95:1088–1093. doi: 10.2106/JBJS.L.01211. [DOI] [PubMed] [Google Scholar]
  • 47.Kim YH, Jang YS, Kim EJ. A prospective, randomized comparison of the long-term clinical and radiographic results of an ultra-short vs a conventional length cementless anatomic femoral stem. J Arthroplasty. 2021;36:1707–1713. doi: 10.1016/j.arth.2020.12.030. [DOI] [PubMed] [Google Scholar]
  • 48.Renkawitz T, Santori FS, Grifka J, Valverde C, Morlock MM, Learmonth ID. A new short uncemented, proximally fixed anatomic femoral implant with a prominent lateral flare: design rationals and study design of an international clinical trial. BMC Musculoskelet Disord. 2008;9:147. doi: 10.1186/1471-2474-9-147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Rastogi S, Marya S. Our experience with short stem hip replacement surgery. Reconstr Rev. 2016;6:21–29. [Google Scholar]
  • 50.Kim YH, Park JW, Kim JS. Ultrashort versus conventional anatomic cementless femoral stems in the same patients younger than 55 years. Clin Orthop Relat Res. 2016;474:2008–2017. doi: 10.1007/s11999-016-4902-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Gombár C, Janositz G, Friebert G, Sisák K. The DePuy Proxima™ short stem for total hip arthroplasty-excellent outcome at a minimum of 7 years. J Orthop Surg (Hong Kong) 2019;27:2309499019838668. doi: 10.1177/2309499019838668. [DOI] [PubMed] [Google Scholar]
  • 52.Melišík M, Hrubina M, Heřt J, Cibula Z, Čabala J, Nečas L. [Mid-term results of Proxima ultra-short anatomical stem: analysis of 130 cases] Acta Chir Orthop Traumatol Cech. 2021;88:50–57. Slovak. [PubMed] [Google Scholar]
  • 53.Kim YH, Jang YS. Long-term clinical and radiographic results of an ultra-short metaphyseal-fitting non-anatomic cementless stem in patients with femoral neck fracture. J Arthroplasty. 2021;36:2105–2109. doi: 10.1016/j.arth.2021.01.029. [DOI] [PubMed] [Google Scholar]
  • 54.Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) Annual report 2018 [Internet] Australian Orthopaedic Association National Joint Replacement Registry; 2019. Feb 08, [cited 2021 Sep 11]. Available from: https://aoanjrr.sahmri.com/annual-reports-2018 . [Google Scholar]
  • 55.McCalden RW, Korczak A, Somerville L, Yuan X, Naudie DD. A randomised trial comparing a short and a standard-length metaphyseal engaging cementless femoral stem using radiostereometric analysis. Bone Joint J. 2015;97-B:595–602. doi: 10.1302/0301-620X.97B5.34994. [DOI] [PubMed] [Google Scholar]
  • 56.Kim YH, Park JW, Kim JS. Short-term results of ultra-short anatomic vs ultra-short non-anatomic proximal loading uncemented femoral stems. J Arthroplasty. 2018;33:149–155. doi: 10.1016/j.arth.2017.07.042. [DOI] [PubMed] [Google Scholar]